The field of art to which this invention relates is surgical sutures, in particular, surgical sutures having both bioabsorbable and nonabsorbable components.
Surgical sutures are well known medical devices in the art. The sutures may have braided or monofilament constructions, and may be provided in single-armed or double-armed configurations with a surgical needle mounted to one or both ends of the suture, or may be provided without surgical needles mounted. The sutures are used in a variety of conventional medical and surgical procedures to approximate tissue, affix or attach implants to tissue, etc. Surgical sutures may be made from a variety of known bioabsorbable and nonabsorbable materials. For example, sutures are known to be made from aromatic polyesters such as polyethylene terephthalate, nylons such as nylon 6 and nylon 66, polyolefins such as polypropylene, silk, and other nonabsorbable polymers. In addition, sutures may be made from polymers and copolymers of p-dioxanone (also known as 1,4-dioxane-2-one), ε-caprolactone, glycolide, L(−)-lactide, D(+)-lactide, meso-lactide, trimethylene carbonate, and combinations thereof. Of particular utility is polydioxanone homopolymer.
Surgical sutures are typically available in a range of conventional sizes for a variety of conventional surgical procedures. The size of the suture used by the surgeon for any particular procedure is dictated in part by the type of tissue to be sutured, the relative size of the tissue structure, as well as the forces that will be applied to the sutures by the approximated tissue after the surgical procedure has been completed. Similarly, the type of suture selected will be dictated by the procedure. Nonabsorbable sutures are typically used for applications such as cardiovascular, vascular, orthopedic, gastrointestinal and the like wherein a nonabsorbable suture is desired or required because a permanent or an extended period of fixation is required during the healing period, e.g., implantation of a heart valve prostheses. Bioabsorbable sutures are typically used for applications such as plastic surgery, skin fixation and certain soft tissue approximation, and the like. A bioabsorbable suture may be used when extended tissue approximation or fixation is not required as long as the suture maintains adequate strength during the healing period, and it is desirable to replace the suture with autologous tissue such as skin or soft tissue during the healing process.
In certain applications where significant tensile forces will be applied on a suture, it is desirable to use materials that will provide high tensile strength with minimal suture size or cross-section. The tensile strength of a suture is known to be a function of several parameters including material, suture size, filament diameter, type of construction (i.e., braided versus monofilament), ratio of sheath to core in a braided construction, type of material in the core, etc. In particular it is known to use ultra high molecular weight polyethylene (hereinafter referred to as UHMWPE) to construct sutures. Sutures made from ultra high molecular polyethylene have the advantage of having high tensile strength while having a smaller suture size. A disadvantage of such sutures is that ultra high molecular weight polyethylene is a relatively slippery material. This inherent slipperiness may affect the knot integrity of a suture. In a typical surgical procedure, after the surgeon makes multiple passes of the suture through tissue so that it is properly approximated and stitched, one or more conventional knots are placed in the suture by the surgeon to maintain the suture in place and thereby prevent it from loosening. Loosening of a suture may allow the approximated tissue to split apart, thereby compromising the surgical procedure and preventing healing with the potential for causing catastrophic effects (e.g., anasotomosis graft blow out in cardiovascular repairs, implant dislodgement, organ failure, soft tissue dislocation from a bone, etc.).
In order to improve the knot holding characteristics (e.g., knot strength) of ultra high molecular polyethylene sutures, it is known to combine the ultra high molecular weight polyethylene yarn component with another yarn component that has surface friction higher that the ultra high molecular weight yarn component. For example it is known to provide a ultra high molecular weight polyethylene suture having an nonabsorbable core, an ultra high molecular weight polyethylene core or an aromatic polyester core with an outer sheath surrounding the core braided from a combination of ultra high molecular weight polyethylene fiber yarns and yarns made from fibers of a nonabsorbable material such as an aromatic polyester. It is also known to have a suture that has a nonabsorbable ultra high molecular weight polyethylene core or an aromatic polyester core surrounded by an outer braided sheath braided from a yarn made from ultra high molecular weight polyethylene fibers and a yarn made from fibers of a bioabsorbable polymer.
Although such composite ultra high molecular weight polyethylene sutures provide increased tensile strength, there are several disadvantages associated with their use. One disadvantage is the possibly diminished knot characteristics that they provide. Typically such sutures produce a large knot having increased mass, size and/or volume. A knot in a nonabsorbable suture remains in the patient's body after the tissue has been approximated and the healing has been completed, and may irritate soft tissue and cause patient discomfort and pain. It is known that from a medical perspective it is desirable to minimize the mass of an implant in tissue. In addition, it is advantageous for an implant to allow for autologous tissue migration or in-growth to provide for better healing. A bioabsorbable implant such as an absorbable suture permits tissue ingrowth as the mass of the implant is resorbed or degraded by the patient's body to provide for better healing. This healing process allows the patient's tissue to assume the tissue loads as the absorbable portion of the implant or suture resorbs/degrades and loses structural integrity and strength. The ultra high molecular weight polyethylene sutures of the prior art have a dense nonabsorbable volume and knot mass, even when combined with a bioabsorbable component, which for the most part remains behind after the tissue healing process has occurred, contributing to problems discussed above.
Accordingly, what is needed in this art are novel composite sutures constructed from materials providing high tensile strength, and bioabsorbable materials, wherein the composite sutures provide optimal tensile strength with minimal mass of high tensile strength material, and have good knot strength and knot security, and reduced in vivo knot profile over time, while allowing tissue ingrowth.
Therefore novel surgical sutures are disclosed. The sutures have a core of yarn made from filaments of a bioabsorbable polymer. The core is surrounded by a sheath braided from a yarn made from filaments of ultra high molecular weight polyethylene and a yarn made from bioabsorbable polymer fibers.
Yet another aspect of the present invention is a surgical suture. The surgical suture has a core and a braided sheath surrounding the core. The core is made from a yarn made from filaments of a bioabsorbable polymer. The sheath is braided from a yarn made from ultra high molecular weight polyethylene fibers and a yarn made from bioabsorbable polymer fibers. The bioabsorbable core and the bioabsorbable fibers include one or more active agents including pharmaceuticals, bioceramics, and anti-microbials.
Still yet another aspect of the present invention is a method of approximating tissue. A surgical suture is provided. The surgical suture consists of a core and a braided outer sheath surrounding the core. The core consists of at least one yarn made from filaments of a bioabsorbable polymer. The sheath is braided from at least one yarn made from ultra high molecular weight polyethylene filaments and at least one yarn made from bioabsorbable polymer filaments. The suture is stitched into a tissue to approximate the tissue and at least one surgical knot is optionally applied to the suture.
Still yet another aspect of the present invention is a method of improving absorption profile of the absorbable component of the suture.
Yet another aspect of the present invention is a method of reducing the knot profile of a knot in a surgical suture. A surgical suture is provided. The surgical suture consists of a core and a braided outer sheath surrounding the core. The core consists of at least one yarn made from filaments of a bioabsorbable polymer. The sheath is braided from at least one yarn made from filaments of ultra high molecular weight polyethylene and at least one yarn made from bioabsorbable polymer filaments. The suture is stitched into a tissue to approximate the tissue. A surgical knot is made in the suture to secure the suture. The knot has a knot profile that is reduced as the bioabsorbable component is absorbed or degraded in vivo over time.
A further aspect of the present invention is a method of providing the release of active components including pharmaceuticals, and antimicrobials through coatings applied to said sutures.
Yet another aspect of the present invention is a combination of a suture anchor and the above-described surgical suture. The suture anchor has an anchor body and a suture mounting opening. The suture is mounted to the suture mounting opening.
Still yet a further aspect of the present invention is a method of affixing soft tissue to the surface of a bone using the suture anchor and suture combination of the present invention. The anchor is mounted in bone, and the suture is used to affix the soft tissue to the surface of the bone.
These and other aspects and advantages of the present invention will become more apparent by the following description and accompanying drawings.
The ultra high molecular weight polyethylene used to manufacture the sutures of the present invention will typically have a weight average molecular weight of about 500,000 g/mole to about 5,000,000 g/mole, more typically about 1,000,000 g/mole to about 3,000,000 g/mole, and preferably about 2,000,000 g/mole to about 2,500,000 g/mole. The ultra high molecular weight polyethylene typically has a tenacity of about 20 cN/dtex to about 40 cN/dtex, more typically about 27.5 cN/dtex to about 38.7 cN/dtex.
The ultra high molecular weight polyethylene will preferably be in the form of a conventional yarn having multiple filaments. Such yarns are commercially available for use in the manufacture of medical devices such as surgical sutures. For example, an ultra high molecular weight polyethylene yarn useful in the sutures of the present invention is 110 dtex Dyneema Purity yarn available from DSM, Netherlands.
The bioabsorbable polymers useful to manufacture the sutures of the present invention include conventional bioabsorbable and biodegradable polymers and copolymers. Of particular utility are the synthetic bioabsorbable polymers prepared from lactone monomers selected from the group consisting of p-dioxanone (also known as 1,4-dioxane-2-one), ε-caprolactone, glycolide, L(−)-lactide, D(+)-lactide, meso-lactide, trimethylene carbonate, and combinations thereof. These polymers include polyglycolide, poly(caprolactone-co-glycolide), poly(glycolide-co-lactide), polylactide, polycapro-lactone and the like and equivalents thereof; advantageously the bioabsorbable polymer is polydioxanone homopolymer. The term “bioabsorbable” as used herein is defined to mean both bioabsorbable and biodegradable. It is particularly preferred to use a polydioxanone polymer for the bioabsorbable component of the sutures of the present invention. The bioabsorbable component of the sutures will preferably consist of a yarn containing multiple filaments made from the bioabsorbable polymer. Methods of manufacturing such yarns are known in this art. If desired, although not preferred, the bioabsorbable yarn component may be made of filaments of more than one bioabsorbable polymer and/or copolymer.
The ultra high molecular weight polyethylene composite sutures of the present invention will typically contain about 40 wt. % to about 70 wt. % of bioabsorbable polymer, more typically about 50 wt. % to about 65 wt. % of bioabsorbable polymer, and preferably about 52 wt. % to about 62 wt. % of bioabsorbable polymer, and more preferably about 62 wt. % of bioabsorbable polymer. For example, the braided sheath may consist of 45 wt. % ultra high molecular weight polyethylene and 55 wt. % of polydioxanone, while the core is 100 wt. % polydioxanone, yielding a suture of the present invention consisting of about 38 wt. % ultra high molecular weight polyethylene and 62 wt. % of polydioxanone.
The composite ultra high molecular weight polyethylene sutures of the present invention may have a standard range of conventional suture sizes ranging from but not limited to USP standard sizes 9/0 to 5. The selection of a particular suture size by the surgeon will depend upon several factors including type of tissue, size of the wound being approximated, tissue load, post operative stresses on the suture, location of the suture, the type of surgical procedure, etc.
The bioabsorbable component of the sutures of the present invention may include various active agents. The active agents that may be added to the bioabsorbable component of the sutures of the present invention include a variety of pharmaceutical compositions and bone-inducing compositions, as well as biologics and bio-active materials. The active agents include, but are not limited to, known anti-inflammatories, anti-infectives, and analgesics or pain killers. Examples of specific pharmaceutical compounds include inderal, diclofenac, fenoprofen, ketoprofen, ketorolac, naproxen, bupivicaine, lidocaine, mepivicaine. Examples of antimicrobials include triclosan and chlorhexidine digluconate. The bone inducing compositions may include known materials such as tricalcium phosphate, bioceramics, bioglasses, calcium hydroxy apatite, combinations thereof and the like. If desired, the coatings used for the sutures of the present invention may contain such active agents
The amount of such active agents included in the bioabsorbable component of the sutures of the present invention will be sufficient to provide a therapeutically effective dose. Similarly, the amount of active agent used in a coating will be sufficient to provide a therapeutically effective dose.
The sutures of the present invention will typically be braided from conventional fiber-containing yarns using conventional braiding equipment. Processes and methods for manufacturing surgical sutures are known in this art and are disclosed, for example, in U.S. Pat. Nos. 5,314,446 and 6,045,071, both of which are incorporated by reference. In addition to braiding, sutures typically require one or more additional conventional process steps after braiding including but not limited to hot stretching, scouring, annealing, coating, tipping, cutting, needle, attachment, packaging and sterilization.
Hot stretch is needed for reducing braid diameter and heat setting the yarns in the braid to prevent so-called “brooming”, when braid edge would open up or unravel after cutting. The scouring process is required for removal of spin finishes that are applied to the yarns during extrusion. Annealing is required to remove residual monomer and for increasing crystallinity. One purpose of a coatings is to provide good suture slidability without unduly compromising knot security. Tipping is optionally needed for a good needle attachment, depending upon the suture. In addition to spin finish removal, the scouring process also helps in removing dust particles and oil droplets that may be deposited during braiding. At the same time, the scouring agent should not have significant adverse effects on suture properties.
It is also known that, in general, coating provides good suture slidability, e.g., through tissue. Coatings may provide the lubricity necessary to achieve smooth tiedown and knot slide performance. While this lubricity is required for smooth tiedown, the coating level(s) must be precisely controlled in order to maintain knot security.
After a suture is optionally tipped in a conventional manner and cut to length, needle attachment takes place. Then, the needle/suture combinations are packaged, and finally sterilized, for example with an Ethylene Oxide process, or other conventional sterilization processes depending on the type of material and packaging, e.g., gamma radiation, autoclaving, plasma, etc.
Referring to
When used in a surgical procedure, the sutures of the present invention are typically knotted, ultimately, by the surgeon after stitches have been made in tissue to emplace the suture, in order to secure the suture. The surgical knots that are utilized for this purpose are well known to surgeons and medical practitioners and include the following types of knots: surgeon's throw, Revo knot, SNC knot, Tennessee Slider, Modified Roeder, square knot, half-hitch and reverse half-hitch, Duncan loop, etc. As seen in
The sutures of the present invention are preferably used in conjunction with suture anchors to repair soft tissue injuries. Suture anchors and procedures using suture anchors are disclosed in the following United States patents which are incorporated by reference: U.S. Pat. Nos. 4,632,100, 4,999,074, 5,814,051, 5,709,708, 5,782,864, 6,270,518, 5,540,718, 6,264,674, 6,270,518, 6,306,158, 5,961,538, 5,782,863, 5,683,418, 5,554,171, 5,078,730, 4,632,100, 5,217,486, 5,011,473, 4,898,156, 4,899,743, 4,946,468, 4,968,315, 5,002,550, 5,046,513, 5,192,303, 5,207,679, 5,358,511. As seen in
It is also known to those skilled in the art that polydioxanone absorbable yarns with a relatively small filament diameter (about 3 denier per filament or smaller) will lose strength faster than yarns of the same total denier that have filaments with substantially larger denier per filament (e.g. 12 denier per filament). In many surgical indications, substantial strength needs to be maintained during the critical wound healing process. There is thus a need to use larger diameter polydioxanone filaments—higher denier per filament yarns. On the other hand, it is a well-known principle that column strength and bending strength of cylinders are both very dependent on diameter. It is thus surprising and unexpected that the bi-component suture that is made with larger diameter 12 denier per filament (dpf) yarn does not become stiffer and still retains desirable handling characteristics similar to the suture made with the much smaller filament diameter yarns (3 dpf).
The following examples are illustrative of the principles and practice of the present invention, although not limited thereto.
A suture of the present invention was manufactured in the following manner. The suture was made using a conventional maple (hollow) braider. The composite suture had a sheath of ultra high molecular weight polyethylene yarn and polydioxanone yarn. The sheath contained 45 wt. % ultra high molecular weight polyethylene and 55 wt. % polydioxanone. The core of the suture consisted of 100 wt. % polydioxanone yarn. Overall, the suture consisted of 38 wt. % ultra high molecular weight polyethylene and 62 wt. % polydioxanone. The nonabsorbable component of the suture construction consisted of eight 100-denier Dyneema Purity ultra high molecular weight polyethylene multifilament yarns (manufactured by DSM) that were loaded on the braider in the counterclockwise direction, and the polydioxanone component was a conventional multifilament polydioxanone yarn, such as that used to manufacture polydioxanone sutures by Ethicon, Inc., Somerville, N.J. The polydioxanone component consisted of eight polydioxanone 60×2 denier dyed multifilament yarns that were loaded on the braider in the clockwise direction. The braid had a core construction that consisted of dyed polydioxanone 60×⅔ multifilament yarns. Pick count is selected at around 47-48 ppi. The absorbable core was selected to minimize suture mass in the body after absorption and provide better suture cutting ability. The braid construction is described in the Table 1.
The off-braider material required additional conventional processing in order to become a suitable suture material. The process flow can be described as follows: hot stretching, annealing, scouring, coating, tipping, and cutting.
The suture was hot stretched at 90° C., which is 5 degrees above the annealing temperature. Such arrangement is needed in order to maintain the heat set memory acquired by braid during the hot stretch. Due to a very low elongation of Dyneema yarn (about 4%), the percentage stretch was set at 8%. Such relatively low percentage of stretch does not damage the braid, and at the same time provides a sufficient force for consolidating the braid structure.
The material was wound onto the racks, and scoured for about 80 min in acetone. Spin finish residuals were reduced to acceptable levels.
The material was coated with a conventional absorbable coating with a concentration level of 4.5% (wt./wt.) of a 90/10 copolymer of caprolactone/glycolide dissolved in ethyl acetate using a drip process. The suture's unique absorbable/non-absorbable construction has led to the evaluation of both absorbable and non-absorbable coatings.
In certain surgical circumstances, it may be preferred that the coating absorbs prior to the absorption of the polydioxanone component of the suture.
Since polydioxanone (hereinafter referred to as PDS) material is susceptible to UV light and exposure to ambient air in general, the exposure time limits have to be maintained during suture processing.
The tensile strength of the suture of Example 1 was tested in accordance with the United States Pharmocopoeia using the method for a constant rate of extension. In addition, the tensile strengths of the following commercially available sutures were also determined: an aromatic polyester (hereinafter referred to as PET) suture such as one manufactured by Ethicon, Inc., Somerville, N.J., and an ultra high molecular weight polyethylene/PET suture such as FIBERWIRE manufactured by Arthrex, Inc., Naples, Fla. The results of the testing are presented in Table 2.
This example compared straight and knot tensile strength of suture of the present invention to the FiberWire suture and the PET suture. The comparison shows that the use of polydioxanone in the core of the suture of the present invention had no adverse effect upon the physical properties of the suture. In order to compare strength of the sutures that have slightly different diameters, straight and knot strength was normalized to the cross-sectional area of the suture and expressed in Kpsi.
The test results showed the suture of the present invention had equivalent or superior tensile strength, and knot break tensile strength than the ultra high molecular weight polyethylene/an aromatic polyester nonabsorbable suture and the an aromatic polyester suture.
Samples of the suture of Example 1 for In Vitro Knot Security were degraded in a buffer of pH 7.27 and a water bath with temperature 57° C. Knots were secured with 5 throws and placed in the buffer container. Five knot-secured sutures from each lot were tested for each time period. Sutures were tested before and after degradation using the Instron Tensile Tester employing a Hermann's loop fixture, 2″/min XH speed to obtain breaking strength. Note: Because this test was a loop, the knot strength was approximately twice that for the typical USP knot tensile test. The mass remaining was obtained by drying and weighing the sutures after different times in the buffer solution.
In addition to the suture of Example 1 consisting of about 40 wt. % high molecular weight polyethylene and about 60% polydioxanone, another suture was tested consisting of 100 wt % polydioxanone. Table 3 provides the in vitro knot strength results versus days in the buffer. The 60/40 polydioxanone/ultra high molecular weight polyethylene suture reached approximately 30 lb strength after 18 days and was relatively constant thereafter. By design the ultra high molecular weight polyethylene strength after loss of the polydioxanone was intended to be approximately 30 lb to match the strength of nonabsorbable an aromatic polyester sutures. Thus, the strength retention of the polydioxanone component of the polydioxanone/ultra high molecular weight polyethylene suture was estimated by subtracting 30 lb from the strength for the suture. This data was also used to present the data as a percent remaining in Table 3 and graphically in
It was not anticipated that the polydioxanone in the polydioxanone/polyethylene suture would maintain higher strength over time than for a 100% polydioxanone braided suture.
In Table 4 and
Note that the weight loss for the polydioxanone component of the bicomponent suture was not atypical of that obtained for the 100% polydioxanone suture. Thus it becomes even more surprising that the knot strength retention in the polydioxanone component of the bicomponent suture remains higher than expected while the mass loss is as expected. This feature is highly desired to maintain good strength yet decrease mass.
In vivo breaking strength retention was conducted for the suture of Example 1 in the following manner.
The method used for in vivo knot tensile was the same as for the in vitro testing except four knot-secured sutures were implanted from each lot, for each time period, using two rats. One of each knot secured sutures were implanted right and left posterior dorsal subcutis. Explanted sutures were examined visually and microscopically: it was noted that there was slight growth of tissue around the suture as a result of time, not as a result of suture composition. Sutures were tested on the Instron Tensile Tester using Herrmann's loop fixture, 2″/min XH speed to obtain breaking strength.
Analysis of this data is displayed in Table 5 and
An accelerated absorption test (boil-out) in 0.05N Sodium Hydroxide solution was conducted using coated suture of Example 1 to verify whether suture knot profile was being reduced after the polydioxanone component was absorbed.
Five samples were prepared by tying loops in the suture, each having a knot consisting of five surgeon's throws. Each sample was labeled accordingly. The thickness of each knot was measured and recorded using a Mitutoyo digital gauge. The measurement was performed with the suture loop lying on the anvil and with the knot being placed between the center of the anvil and the micrometer foot. A 500 ml Pyrex bottle with 0.05N Sodium Hydroxide solution was prepared. The knots were placed into the bottle, with the bottle sealed using a tight lid. The bottle was then placed into a water bath at 90° C. for about twenty-four hours.
After twenty-four hours “boil-out”, the samples were removed from the bottle and dried for one hour. The thickness of each knot was measured again. The results are exhibited in the Table 6 below.
This experiment demonstrates that the knot profile was reduced by about 50% after the bioabsorbable polydioxanone component is absorbed. The suture flattens, making the knot less palpable and thus, reducing tissue irritation.
In another experiment, an in vitro absorption test was conducted using the coated suture of Example 1.
Samples were prepared by tying loops in the suture, each having a knot consisting of six half hitch knots, alternating throw, alternating posts. Each sample was labeled accordingly. The samples were placed in phosphate buffered solution at 57° C. for a period of 26 weeks. The samples were weighed and the volume of the samples was calculated. These results were compared to Day 0 samples. The results are shown in Table 7 below.
In this example, an antimicrobial, Triclosan, is blended with a powdered form of a biodegradable polymer such as polydioxanone. The blending is carried out in a Brabender, Baker Perkins Blender, or a V-blender, and is generally although not necessarily done at an elevated temperature. Solvents may be added if desired to help improve the blending; the dispersing solvent is then generally removed prior to extrusion. The resulting composition is extruded using a single or twin extruder, or pelletized. Once blended, the material was re-extruded using a conventional single or twin screw extruder with a die that is capable of forming a monofilament or multifilament fiber. This fiber is either collected directly from the spinneret with no other modification, or it can be optionally post processed by annealing and drawing. Resin throughput, take-up speed and the extent of drawing dictate the diameter and degree of orientation of the fiber. The fiber is spun into a yarn.
The absorbable multifilament yarn of Example 6 is used to make a composite suture of the present invention by following the process of Example 1 and substituting the yarn of Example 6 for the bioabsorbable yarn of Example 1.
Five suture strands 1.5 inch in length were prepared from each lot for testing. A special fixture (described below) was used with the Instron Tensile Tester (Model 4201) to measure the resistance of the force required to pull the suture through the opening in the fixture. A load cell of 500 g was used for this test. The crosshead speed was one inch per minute. The information from the test was gathered electronically.
The upper portion of this fixture is illustrated in
Five samples of 1.5 inch each of the suture of Example 1 were prepared. Each sample was inserted through passage 508 in the metal rod until length of suture on each side of opening was about the same. Then, the crosshead was getting started and the suture was pulled through opening in the bracket. When the suture sample was pulled through the opening in the bracket, the resistance in grams is being recorded. Five samples were measured and the results were averaged to provide the reported values.
The results as presented in Table 8 demonstrate that the polyethylene/polydioxanone suture of the present invention had better bending stiffness than FiberWire polyethylene/an aromatic polyester suture, and had similar properties to an aromatic polyester suture.
A patient is prepared for rotator cuff surgery in a conventional manner. The patient is anesthetized using conventional anesthesia. The pathology is identified and diagnosed arthroscopically using a conventional trocar and a conventional arthroscope. The surgeon views the site of the injury using a conventional arthroscope inserted through a cannula that is emplaced in the shoulder adjacent to the injury site. The surgeon releases any adhesions, debrides the lateral margin of the rotator cuff, immobilizes the tendon and prepares the bone bed. The surgeon locates an optimal suture anchor placement site on the ‘downside’ of the greater tuberosity of the patient's humeral head and drills a bone bore hole for anchor emplacement in the following manner. A drill guide is inserted through the trocar. A drill bit is placed through the drill guide and the bore hole is drilled for the suture anchor. A conventional suture anchor such as the suture anchor illustrated in
The sutures of the present invention have many advantages. The advantages include maximized suture mass reduction over time, reduced knot profile over time, improved BSR profile, and high tensile strength, and improved handling characteristics.
Although this invention has been shown and described with respect to detailed embodiments thereof, it will be understood by those skilled in the art that various changes in form and detail thereof may be made without departing from the spirit and scope of the claimed invention.
This application is a continuation of Ser. No. 14/062,090, filed Oct. 24, 2013, which is a divisional of Ser. No. 13/330,741, filed Dec. 20, 2011, now U.S. Pat. No. 8,568,449 which is a divisional of Ser. No. 12/102,477, filed on Apr. 14, 2008, now U.S. Pat. No. 8,109,967 which is a continuation of application Ser. No. 10/740,023, filed on Dec. 18, 2003, now U.S. Pat. No. 7,357,810. The complete disclosures of the aforementioned related U.S. patent applications are hereby incorporated herein by reference for all purposes.
Number | Name | Date | Kind |
---|---|---|---|
3463158 | Schmitt | Aug 1969 | A |
3942532 | Hunter | Mar 1976 | A |
4105034 | Shalaby | Aug 1978 | A |
4455273 | Harpell | Jun 1984 | A |
4624256 | Messier | Nov 1986 | A |
4632100 | Somers | Dec 1986 | A |
4898156 | Gatturna | Feb 1990 | A |
4899743 | Nicholson | Feb 1990 | A |
4946468 | Li | Aug 1990 | A |
4968315 | Gatturna | Nov 1990 | A |
4999074 | Afeyan | Mar 1991 | A |
5002550 | Li | Mar 1991 | A |
5011473 | Gatturna | Apr 1991 | A |
5046513 | Gatturna | Sep 1991 | A |
5078730 | Li | Jan 1992 | A |
5183701 | Jacobs | Feb 1993 | A |
5192303 | Gatturna | Mar 1993 | A |
5207679 | Li | May 1993 | A |
5217486 | Rice | Jun 1993 | A |
5261886 | Chesterfield | Nov 1993 | A |
5306289 | Kaplan | Apr 1994 | A |
5314446 | Hunter | May 1994 | A |
5358511 | Gatturna | Oct 1994 | A |
5456697 | Chesterfield | Oct 1995 | A |
5540718 | Bartlett | Jul 1996 | A |
5554171 | Gatturna | Sep 1996 | A |
5584836 | Ballintyn | Dec 1996 | A |
5683418 | Luscombe | Nov 1997 | A |
5709708 | Thal | Jan 1998 | A |
5782863 | Bartlett | Jul 1998 | A |
5782864 | Lizardi | Jul 1998 | A |
5814051 | Wenstrom, Jr. | Sep 1998 | A |
5961538 | Pedlick | Oct 1999 | A |
5980559 | Bonutti | Nov 1999 | A |
6045071 | Emsens | Apr 2000 | A |
6045571 | Hill | Apr 2000 | A |
6264674 | Washington | Jul 2001 | B1 |
6270518 | Pedlick | Aug 2001 | B1 |
6306158 | Bartlett | Oct 2001 | B1 |
6406479 | Justin | Jun 2002 | B1 |
6689153 | Skiba | Feb 2004 | B1 |
6712838 | D'Aversa | Mar 2004 | B2 |
6716234 | Grafton | Apr 2004 | B2 |
7329271 | Koyfman | Feb 2008 | B2 |
7357810 | Koyfman | Apr 2008 | B2 |
8109967 | Koyfman | Feb 2012 | B2 |
8568449 | Koyfman | Oct 2013 | B2 |
20030050666 | Grafton | Mar 2003 | A1 |
20040015187 | Lendlein | Jan 2004 | A1 |
20090299408 | Schuldt-Hempe | Dec 2009 | A1 |
Number | Date | Country |
---|---|---|
10219860 | Nov 2003 | DE |
557894 | Sep 1993 | EP |
561108 | Sep 1993 | EP |
1171172 | Jan 2001 | EP |
1293218 | Mar 2003 | EP |
1506789 | Feb 2005 | EP |
49-14800 | Feb 1974 | JP |
54004493 | Jan 1979 | JP |
59130316 | Jul 1984 | JP |
63-95060 | Apr 1988 | JP |
3-80868 | Apr 1991 | JP |
2002540901 | Dec 2002 | JP |
2003339849 | Dec 2003 | JP |
2005-177500 | Jul 2005 | JP |
WO 03092758 | Nov 2003 | WO |
WO 2004082466 | Sep 2004 | WO |
Entry |
---|
Mead et al., “Analysis of Several Ultrahigh Molecular Weight Polyethylene Fibers.” ANTEC, Plastics: Plastics Saving Planet Earth, 1997, p. 543, vol. 2; Materials, Society of Plastics Engineers. |
Number | Date | Country | |
---|---|---|---|
20150133868 A1 | May 2015 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13330741 | Dec 2011 | US |
Child | 14062090 | US | |
Parent | 12102477 | Apr 2008 | US |
Child | 13330741 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14062090 | Oct 2013 | US |
Child | 14602601 | US | |
Parent | 10740023 | Dec 2003 | US |
Child | 12102477 | US |