The instant application contains a Sequence Listing which has been submitted electronically in ASCII format and is hereby incorporated by reference in its entirety. Said ASCII copy, created on Oct. 28, 2021, is named 00138-012WO1_SL.txt and is 114,277 bytes in size.
The disclosure provides for hybrid protocols and barcoding schemes that allow for sequencing of targeted polynucleotides in multiple types of sequencing platforms, and applications thereof, including for metagenomic analysis.
Early detection of causative microorganisms in patients with severe infections is important to informing clinical interventions and administering appropriately targeted antibiotics. Timely and accurate diagnosis, however, remains highly challenging for many hospitalized patients. As most infectious syndromes present with indistinguishable clinical manifestations, broad-based, multiplexed diagnostic tests are urgently needed but not yet available for the vast majority of potential pathogens. Some microorganisms are difficult to grow in culture (e.g., Treponema pallidum, Bartonella sp.), or unculturable (e.g., some viruses), while others (e.g., mycobacteria and molds) can take weeks to grow and speciate. Accurate molecular detection by PCR provides an alternative diagnostic approach to culture, but is hypothesis-driven and thus requires a priori suspicion of the causative pathogen(s).
Metagenomic analysis by next-generation sequencing of random, “shotgun” reads has a number of applications, including (1) clinical diagnosis, (2) pathogen discovery, (3) de novo genome assembly, (4) whole-exome sequencing, (5) targeted gene panel sequencing, (5) transcriptome profiling, and (6) whole-genome resequencing. Disclosed herein is a metagenomic next-generation sequencing (mNGS) method using cell-free DNA from body fluids to identify pathogens. The performance of mNGS testing of 182 body fluids from 160 acutely ill patients was evaluated using two sequencing platforms in comparison to microbiological testing using culture, 16S bacterial PCR, and/or 28S-ITS fungal PCR. Test sensitivity and specificity of detection were 79% and 91% for bacteria and 91% and 89% for fungi, respectively, by Illumina sequencing; 75% and 81% for bacteria and 91% and 100% for fungi, respectively, by nanopore sequencing. In a case series of 12 patients with culture/PCR-35 negative body fluids but for whom an infectious diagnosis was ultimately established, 7 (58%) were mNGS-positive. Real-time computational analysis enabled pathogen identification by nanopore sequencing in a median 50-minute sequencing and 6-hour sample-to-answer time. The Rapid mNGS methods of the disclosure are promising tools for diagnosis of unknown infections from body fluids.
The disclosure provides an oligonucleotide comprising barcodes for use in multiple types of next generation sequencing technologies, the barcodes comprising at least about 18 to about 160 nucleotides in length having a first nucleotide domain and at least one second nucleotide domain; wherein the first nucleotide domain comprises 4-12 nucleotides (4-12mer) of the barcode located at either end of the barcode and wherein the 4-12mer are compatible with a next generation sequencing technology that utilizes bridge amplification, wherein the second nucleotide domain comprises 14-35 nucleotides (14-35mer) of the barcode and wherein the 14-35mers are compatible with a next generation sequencing that utilizes nanopores, wherein at least a minimum Levenshtein distance between a pair of 4-12mers is utilized, and wherein the Levenshtein distance has been maximized between a pair of barcodes in order to minimize barcode “crosstalk”. In one embodiment, the oligonucleotide further comprises a flow cell attachment domain. In a further embodiment, the flow cell attachment domain comprises a sequence selected from SEQ ID NO:1, 2, 3 or 4. In another embodiment, the oligonucleotide further comprises a sequencing primer binding domain. In another embodiment, the barcode is comprised of the 4-12mer and the second domain comprises 3 sets of 10mers that when concatenated together form a 34-42mer, wherein the last nucleotide is removed to form the 33-41mer barcode. In another embodiment, the oligonucleotide comprises a sequence selected from any one of SEQ ID Nos: 226-416 and 417. In another embodiment of any of the foregoing embodiments, oligonucleotide consists of 47-80 nucleotides. In another embodiment, the oligonucleotide is 62-83 nucleotides in length.
The disclosure also provides an oligonucleotide comprising barcodes for use in multiple types of next generation sequencing technologies, the barcodes comprising at least about 18 to about 39 nucleotides in length having a first nucleotide domain and at least one second nucleotide domain; wherein the first nucleotide domain comprises 4-9 nucleotides (4-9mer) of the barcode located at either end of the barcode and wherein the 4-9mers are compatible with a next generation sequencing technology that utilizes bridge amplification, wherein the second nucleotide domain comprises 14-35 nucleotides (14-35mer) of the barcode and wherein the 14-35mers are compatible with a next generation sequencing that utilizes nanopores, wherein at least a minimum Levenshtein distance between a pair of 4-9mers is utilized, and wherein the Levenshtein distance has been maximized between a pair of barcodes in order to minimize barcode “crosstalk”.
The disclosure also provides an oligonucleotide barcode sequence for use in multiple types of next generation sequencing, wherein the oligonucleotide barcode is about 24 to 39 nucleotides in length and comprises a first oligonucleotide barcode domain of about 4-12 nucleotides (4-12mer) at the 5′ or 3′ end of the oligonucleotide barcode and a second oligonucleotide barcode domain of about 10-29 nucleotides in length operably linked to the first oligonucleotide barcode domain, wherein the Levenshtein distance has been maximized between a pair of oligonucleotide barcodes in order to minimize barcode “crosstalk”; wherein the first oligonucleotide barcode domain is compatible with next generation sequencing using bridge amplification; wherein the second oligonucleotide barcode domain is compatible with next generation sequencing using nanopores; and wherein the oligonucleotide has a minimum Levenshtein distance between a pair of 4-9mers. In one embodiment, the barcode is about 36-39 nucleotides in length. In still another or further embodiment, the oligonucleotide comprises a sequence selected from the group consisting of SEQ ID Nos: 226-416 and 417.
The disclosure also provides a set of oligonucleotides comprising a barcode as set forth herein. In another embodiment, each barcode is located between a pair of sequencing adaptors. In still a further embodiment, the pair of sequencing adaptors have sequences selected from (i) or (ii): (i) CAAGCAGAAGACGGCATACGAGAT (SEQ ID NO:1), and GTGACTGGAGTTCAGACGTGTGCTCTTCCGATC*T (SEQ ID NO:2); or (ii) AATGATACGGCGACCACCGAGATCTACAC (SEQ ID NO:3), and ACACTCTTTCCCTACACGACGCTCTTCCGATC*T (SEQ ID NO:4), wherein * indicates a phosphorothioate bond between the nucleotides. In still another embodiment, the set of oligonucleotides are PCR primers used for sequencing library barcoding.
The disclosure also provides a sequencing library comprising the set of barcodes as described herein. In another embodiment, the sequencing library is used for an application selected from: pathogen discovery, environmental metagenomics, de novo genome assembly, whole-exome sequencing, transcriptomics sequencing, targeted gene panel sequencing or whole-genome resequencing.
The disclosure also provides a method for rapid pathogen detection in a sample using metagenomic next-generation sequencing (mNGS), comprising: obtaining one or more samples comprising cell-free DNA (cfDNA); generating a plurality of sequencing reads comprising a barcode from the set of barcodes of the disclosure using next-generation sequencing; performing metagenomic analysis on the plurality of sequencing read data using a clinical bioinformatics software pipeline that can rapidly analyze sequencing read data for pathogenic DNA; determining and identifying pathogen(s) in the one or more samples based upon the metagenomic analysis of the sequencing read data. In another embodiment, the one or more samples comprises a body fluid sample from a subject. In a further embodiment, the body fluid sample is an infected body fluid sample. In still another or further embodiment, the body fluid sample is selected from cerebrospinal fluid, urine, semen, pericardial fluid, pleural fluid, peritoneal fluid, synovial fluid, amniotic fluid, fetal fibronectin, saliva, sweat, eye vitreous humor, eye aqueous humor, bronchoalveolar lavage fluid, breast milk, bile, and ascites fluid. In still a further embodiment, the one or more samples further comprise a blood serum sample. In another embodiment, the next-generation sequencing comprises sequencing technology that utilizes bridge amplification. In another or further embodiment, the next-generation sequencing comprises or further comprise sequencing technology that utilizes nanopores. In still another embodiment, the clinical bioinformatics software pipeline that can rapidly analyze sequencing read data for pathogenic DNA is SURPI+ or SURPIrt. In still another embodiment, the pathogen(s) comprise one or more pathogenic bacteria. In another embodiment, the pathogen(s) comprise one or more pathogenic fungi.
The disclosure provides a set of paired 37mer barcodes comprising dual indexes that are configured for dual use in multiple types of next generation sequencing technologies, wherein the Levenshtein distance has been maximized between each pair of 37mer barcodes in order to minimize barcode “crosstalk”; wherein the first 8 nucleotides (8mer) of each pair of 37mer barcodes is compatible with a next generation sequencing technology that utilizes bridge amplification, and wherein at least a minimum Levenshtein distance between each pair of 8mers is utilized; wherein at least a minimum Levenshtein distance between each pair of 37mers barcodes is used so that the 37mer barcode is compatible with a next generation sequencing technology that utilizes nanopores.
In a certain embodiment, the disclosure provides for a composition or method as substantially described and/or illustrated herein.
As used herein and in the appended claims, the singular forms “a,” “an,” and “the” include plural referents unless the context clearly dictates otherwise. Thus, for example, reference to “a body fluid” includes a plurality of such body fluids and reference to “the organism” includes reference to one or more organisms and equivalents thereof known to those skilled in the art, and so forth.
Also, the use of “or” means “and/or” unless stated otherwise. Similarly, “comprise,” “comprises,” “comprising” “include,” “includes,” and “including” are interchangeable and not intended to be limiting.
It is to be further understood that where descriptions of various embodiments use the term “comprising,” those skilled in the art would understand that in some specific instances, an embodiment can be alternatively described using language “consisting essentially of” or “consisting of.”
Unless defined otherwise, all technical and scientific terms used herein have the same meaning as commonly understood to one of ordinary skill in the art to which this disclosure belongs. Although many methods and reagents are similar or equivalent to those described herein, the exemplary methods and materials are disclosed herein.
All publications mentioned herein are incorporated herein by reference in full for the purpose of describing and disclosing the methodologies, which might be used in connection with the description herein. Moreover, with respect to any term that is presented in one or more publications that is similar to, or identical with, a term that has been expressly defined in this disclosure, the definition of the term as expressly provided in this disclosure will control in all respects.
It should be understood that this disclosure is not limited to the particular methodology, protocols, and reagents, etc., described herein and as such may vary. The terminology used herein is for the purpose of describing particular embodiments only and is not intended to limit the scope of the disclosure, which is defined solely by the claims.
Other than in the operating examples, or where otherwise indicated, all numbers expressing quantities of ingredients or reaction conditions used herein should be understood as modified in all instances by the term “about.” The term “about” when used to described the present invention, in connection with percentages means±1%.
As used herein, the term “amount” or “level” in reference to a targeted biomolecule, refers to a quantity of the targeted molecule that is detectable or measurable in a sample and/or control.
As used herein, the term “biological sample” includes any sample(s) that is taken from a subject which contains one or more targeted biomolecules described herein. Suitable samples in the context of the present disclosure include, for example, blood, plasma, serum, amniotic fluid, vaginal excretions, saliva, and urine. In a particular embodiment, biological samples used in a method disclosed herein comprise a blood plasma sample and a body fluid sample. In a further embodiment, biological samples used in a method disclosed herein comprise cell-free DNA (cfDNA) from body fluids.
Although PCR tests targeting the conserved 16S ribosomal RNA (rRNA) gene (“16S PCR”) and 28S-internal transcribed ribosomal gene spacer (“28S-ITS PCR”) regions of bacteria and fungi, respectively, have been developed, concerns have been raised regarding detection sensitivity. Failure or delay in diagnosing infections results in extended hospitalizations, readmissions, and increased mortality and morbidity. In addition, undiagnosed patients nearly always require empiric broad-spectrum therapy, with increased risk of adverse side effects and antimicrobial drug resistance.
Metagenomic next-generation sequencing (mNGS) enables detection of nearly all known pathogens simultaneously from clinical samples. Previous work in this area has focused on a single, generally non-purulent body fluid type, and few studies to date have demonstrated clinical validation and/or utility. Methodology and sample types are also highly variable, making it difficult to evaluate comparative performance across different studies. In particular, purulent fluids, which often suggest an infectious etiology, can be challenging to analyze by mNGS due to high human host DNA background, which can decrease assay sensitivity.
Methods exist to enrich for pathogen-specific reads from metagenomic data, such as differential lysis of human cells, but the scope of detection using these approaches is largely restricted to bacteria and/or fungi. Rapid identification of pathogens from infected body fluid compartments is important because empiric antimicrobial treatment is often suboptimal, contributing to increased morbidity and mortality. Most metagenomic studies have employed Illumina™ sequencing platforms, with sequencing run times exceeding 16 hours and overall sample-to-answer turnaround times of 48-72 hours. In contrast, nanopore sequencing (MinION™ sequencer by Oxford Nanopore Technologies) can detect microbes within minutes of starting sequencing and with a <6-hour turnaround time. Nanopore sequencing has been extensively used for genomic surveillance of emerging viruses, but clinical metagenomic applications of the technology for pathogen detection have been limited. One published study describes the use of a saponin-based differential lysis enrichment method for metagenomic nanopore sequencing-based detection of bacteria in respiratory infections with 96.6% sensitivity yet only 41.7% specificity.
Provided herein are simple, rapid, and universal methods for pathogen detection by mNGS analysis of cell-free DNA (cfDNA) from a variety of different body fluids, ranging from low-cellularity cerebrospinal fluid (CSF) to purulent fluids with high human host DNA content (e.g., abscesses). An innovative dual-use protocol, suitable for either Oxford Nanopore Technologies' nanopore or Illumina™ sequencing platforms, is used to evaluate the diagnostic accuracy of mNGS testing against traditional culture and PCR-based testing. A case series evaluating the performance of mNGS testing in 12 patients with culture- and PCR-negative body 95 fluids is described herein. For all cases, there was either high clinical suspicion for an infectious etiology or a confirmed microbiological diagnosis by orthogonal laboratory testing.
Described herein are rapid diagnostic assays for unbiased metagenomic detection of DNA-based pathogens from body fluids. Some advances underlying the approaches presented herein, include: (i) detection across a broad range of sample types, (ii) compatibility with input cfDNA concentrations varying across 6 orders of magnitude (100 pg/mL-100 ug/mL), (iii) a dual-use barcoding system enabling deployment on Illumina and nanopore sequencing platforms, and (iv) clinically validated bioinformatics pipelines for automated analysis and interpretation of mNGS data. Importantly, it was found that sensitivities and specificities for bacterial and fungal detection across Illumina and nanopore sequencing platforms were comparable. The potential utility of the methods of the disclosure are highlighted by detection of pathogens in 7 of 12 (58.3%) selected cases for which culture and PCR testing of the body fluid were negative, with subthreshold detection of pathogen reads in an additional two cases (9 of 12, 75%) (Table 11).
In the studies presented herein, mNGS testing failed to detect S. aureus at higher rates than other bacteria, a finding that was statistically significant for nanopore but not for Illumina sequencing. The lower sensitivity of S. aureus detection by nanopore sequencing was attributed to higher levels of human host background DNA. Notably, the median body fluid white blood cell (WBC) count for S. aureus was 70,250×109/L (IQR 42,800-137,500), an approximately 100-fold increase over median WBC counts for other microorganisms (p<0.00001 by Mann-Whitney U-test). Other factors contributing to lower sensitivity for nanopore sequencing may be the lower read depths achieved in the current study and higher error rates relative to Illumina sequencing. These limitations are addressed by increasing average sequencing throughput per sample or making improvements in nanopore read accuracy over time.
The methods disclosed herein utilize pathogen-specific cfDNA sequences in body fluid supernatant. Intact pathogen DNA from high human DNA background samples, such as respiratory or joint fluids, can be obtained using differential lysis protocols. However, as the supernatant containing pathogen cfDNA is removed during the differential lysis protocol, such enrichment methods may not work as well for low cellularity samples such as plasma and CSF. Differential lysis can also hinder detection of other microorganisms such as viruses and parasites. In addition, these methods involve multiple steps of lysis and centrifugation, which can increase method complexity and prolong assay turnaround times. The methods disclosed herein also forego the use of mechanical processing steps such as bead-beating. Bead-beating may improve the detection of intact fungi and some bacteria containing rigid cell walls, but is laborious for routine use in the clinical laboratory and can reduce detection sensitivity by increasing host background from the release of human DNA.
While other studies have used metagenomic sequencing for pathogen detection in sepsis and pneumonia, the reported test specificities of 63% and 42.7% respectively, limiting broad clinical application, as it can be challenging to evaluate the clinical significance of false-positive results. In direct contrast, an overall specificity ranging from 83% to 100% was achieved using the methods and compositions of the disclosure.
Pathogen cfDNA analysis from blood has been used to diagnose deep-seated infections. However, bacterial DNA is often present at low levels in blood, with a lower quartile of 5 bacterial genome copies per mL in patients with sepsis. In matched pairs of samples, it was shown herein that there was an observed 160-fold higher pathogen cfDNA burden in body fluids. Similarly, tumor cfDNA is higher in adjacent body fluids than in blood. Higher levels of pathogen cfDNA in body fluids can increase analytical sensitivity and decrease sequencing depths required for accurate detection, thereby lowering the cost of testing. In addition, direct identification of a pathogen from a body fluid can localize the source of an infection, which is important to guiding definitive management and treatment.
In comparing mNGS with bacterial 16S or fungal 28S-ITS PCR, occult pathogens were detected solely by mNGS in 5 of 14 cases. False-negative 16S PCR results have been previously reported, and are generally attributed to suboptimal primer design or decreased assay sensitivity from background contamination. However, discordant results between 16S PCR and mNGS may also be due to short pathogen read lengths in cell-free body fluids. Notably, size ranges for bacterial 16S PCR amplicons span 300-460 nt, whereas those for fungal 28S-ITS PCR amplicons span 250-650 nt. Decreases in sensitivity due to fragmented cfDNA that are not amenable to long-read amplicon PCR have also been observed for detection of EBV virus in clinical samples.
The mNGS methods of the disclosure expand the scope of conventional diagnostic testing to multiple body fluid types. The achievable <6-hour turnaround time using nanopore sequencing may also be important for infections such as sepsis and pneumonia that demand a rapid response and timely diagnosis. The results presented herein indicate that mNGS testing methods disclosed herein are useful for a plurality of scenarios, including: (i) for identification of culture-negative or slow-growing pathogens, (ii) for diagnosis of rare or unusual infections that were not considered by the health care provider a priori, (iii) as a first-line test in critically ill patients, and (iv) as an early alternative to the large number of send out tests that would otherwise be ordered as part of the diagnostic workup.
The studies presented herein have focused on clinical development and validation of metagenomic sequencing technologies, including pathogen detection and gene expression profiling, to diagnose infections in clinical samples from patients. There are key advantages and disadvantages regarding the choice of sequencing technologies for the metagenomic sequencing approach. For instance, nanopore sequencing (currently available on the MinION™, GridION™, or PromethION™ instruments by Oxford Nanopore Technologies™, or ONT) enables longer reads and “real-time” sequencing analysis; the latter aspect enables more rapid sequencing protocols and shorter turnaround times, albeit with lower throughput and higher error rates. Illumina™ sequencing, in contrast, has much higher throughput (number of reads per given unit time) and lower costs, albeit at greater turnaround times.
Presented herein is the development and validation of a hybrid approach and barcoding schemes in which sequencing libraries can be constructed from samples that would be compatible (e.g., can be sequenced) on a variety of different sequencing platforms. Most sequencing technologies utilize “adapter-ligation” protocols for barcoding and sequencing, whereby an indexed adapter is attached to the end(s) of free DNA or cDNA molecules in order to barcode multiplexed samples and facilitate a subsequent sequencing reaction. The hybrid approach for use with ONT and Illumina platforms can use an adapter-ligation approach coupled to the same or different-sized barcodes (e.g., 37mers for the ONT and 8mers—the first or last 8 bases of the 37mer for Illumina) to generate barcoded, dual- or singly-indexed libraries that are compatible with both platforms.
In a particular embodiment, the disclosure provides at least one, typically a set including 2, 3, 4 or more pairs of a barcode Xmer (wherein X is an integer selected 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42 or more) barcodes comprising an index (e.g., dual indexes comprising a first domain or bridge domain index and a second domain or nanopore domain index) that is configured for use in multiple types of next generation sequencing technologies, wherein the Levenshtein distance has been maximized between each pair of Xmer barcodes in order to minimize barcode “crosstalk”; wherein the first or last, e.g., 4 to 9 nucleotides (4-9mer) of each pair of Xmer barcodes is compatible with a next generation sequencing technology that utilizes bridge amplification (e.g., iSeq100, MiniSeq, MiSeq, HiSeq, NovaSeq, and NextSeq from Illumina™), and wherein at least a minimum Levenshtein distance between each pair of, e.g., 4-9mers is utilized; wherein at least a minimum Levenshtein distance between each pair of Xmer barcodes is used so that the Xmer barcode is compatible with a next generation sequencing technology that utilizes nanopores (e.g., Flongle, MinION, MinION Mk1C, GridION, and promethION from Oxford Nanopore Technologies™). In a further embodiment, the Xmer barcodes are comprised of the, e.g., 4-9mer and, e.g., 3 sets of 10mer barcodes that concatenated together to form, for example, a Xmer of 33-39 nucleotides, wherein the last nucleotide is removed to form the Xmer barcodes of 32-38 nucleotides. In regards to Levenshtein distance, the Levenshtein distance can be computed using the methods presented herein, or the Levenshtein distance calculations described in detailed in Bushmann et al., (“Levenshtein error-correcting barcodes for multiplexed DNA sequencing.” BMC Bioinformatics 14: 272 (2013)), the disclosure of which is incorporated herein in full.
It should be recognized that the second ‘nanopore’ domain index can completely overlap and encompass the first ‘bridge’ domain index. The overall length can have a higher upper limit, such as 160 nucleotides. The exemplary oligonucleotides described in the Examples below used two 37mers, for a total of 74 nucleotides. Moreover, the first ‘bridge’ domain can go up to two 12mers, so the minimum can be high at 24 or 25 nucleotides total. Although the Examples, use a 37mer, an exact 37mer is not necessary, e.g., 36mer or 38mer will also work. The second ‘nanopore’ barcode index can be at least a total of 24 nucleotides (all locations combined). Alternatively, the second ‘nanopore’ barcodes are at least double in length the size of the bridge amplification barcodes. In addition, paired barcodes are not required. The barcodes can be arbitrarily shifted between the two sides, all the way on one side or the other, to effectively have single-end barcodes. Index barcodes can also be easily shifted into other locations—currently, in the Illumina and nanopore configuration, there are 4 convention locations, so the total can be quadruple barcodes rather than paired. In addition, although the Examples below used an 8mer first ‘bridge’ domain index it does not have to be a precise 8mer. For example, bridge amplification systems such as that on Illumina systems also use 6mers, 7mers, 8mers or 9mers.
In a particular embodiment, the disclosure provides for a set of oligonucleotides comprising a set of Xmer barcodes (e.g., a 37mer) disclosed herein. In a further embodiment each Xmer barcode is located between a pair of sequencing adaptors. In yet a further embodiment, the pair of sequencing adaptors have sequences selected from (i) or (ii):
In a certain embodiment, the disclosure also provides a sequencing library comprising a set of paired Xmer barcodes (wherein X is between 15 and 42 nt) disclosed herein. In a further embodiment, the sequencing library is used for an application selected from: pathogen discovery, environmental metagenomics, de novo genome assembly, whole-exome sequencing, transcriptomics sequencing, targeted gene panel sequencing or whole-genome resequencing. In a further embodiment, the sequencing library is generated using a library preparation kit. In yet a further embodiment, the library preparation kit is from Illumina, Inc (e.g., AmpliSeq™ kits, COVIDSeq™ kit, Illumina DNA prep kits, Illumina RNA prep kits, Nextera™ Kits, SureCell WTA™ Kits, TruSeq™ kits, and TruSight™ kits).
In a particular embodiment, the disclosure also provides a method for rapid pathogen detection in a sample using metagenomic next-generation sequencing (mNGS), comprising: obtaining one or more samples comprising cell-free DNA (cfDNA); generating a plurality of sequencing read data comprising a Xmer barcode (wherein X is between 15 and 42 nt) from a set of paired Xmer barcodes wherein the Levenshtein distance has been maximized between each pair of Xmer barcodes in order to minimize barcode “crosstalk”; wherein the first or last 4 to 9 nucleotides (4-9mer) of each pair of Xmer barcodes is compatible with a next generation sequencing technology that utilizes bridge amplification, and wherein at least a minimum Levenshtein distance between each pair of 4-9mers is utilized and wherein at least a minimum Levenshtein distance between each pair of Xmer barcodes is used so that the Xmer barcode is compatible with a next generation sequencing technology that utilizes nanopores; performing metagenomic analysis on the plurality of sequencing read data using a clinical bioinformatics software pipeline that can rapidly analyze sequencing read data for pathogenic DNA; identifying pathogen(s) in the one or more samples based upon the metagenomic analysis of the sequencing read data. In another embodiment, the one or more samples comprises a body fluid sample from a subject. In yet another embodiment, the body fluid sample is a purulent body fluid sample. In a certain embodiment, the body fluid sample is selected from cerebrospinal fluid, urine, semen, pericardial fluid, pleural fluid, peritoneal fluid, synovial fluid, amniotic fluid, fetal fibronectin, saliva, sweat, eye vitreous humor, eye aqueous humor, bronchoalveolar lavage fluid, breast milk, bile, and ascites fluid. In another embodiment, the one or more samples further comprise a blood serum sample. In yet another embodiment, the next-generation sequencing comprises sequencing technology that utilizes bridge amplification. In a further embodiment, the next-generation sequencing comprises or further comprise sequencing technology that utilizes nanopores. In yet a further embodiment, the next-generation sequencing comprises sequencing technology that utilizes bridge amplification and sequencing technology that utilizes nanopores. In another embodiment, the clinical bioinformatics software pipeline that can rapidly analyze sequencing read data for pathogenic DNA is SURPI+ or SURPIrt. In a further embodiment, the pathogen(s) comprise one or more pathogenic bacteria. In an alternate embodiment, the pathogen(s) comprise one or more pathogenic fungi.
Methods using the hybrid approach described herein allows for short read, high-throughput, slower sample-to-sequence technologies, such as Illumina, to be performed simultaneously with long read, lower-throughput, rapid sequencing technologies, such as ONT. The methods disclosed herein by using such a hybrid approach, can leverage key advantages of each sequencing technology (e.g., ONT nanopore sequencing—speed; Illumina sequencing—throughput). In the studies presented herein, the hybrid approach described herein was successfully run with 37mer barcoding for ONT nanopore sequencing and 8mer barcoding for Illumina sequencing. Accordingly, the disclosure has provided methodologies where two or more sequencing platforms can be used simultaneously and successfully for metagenomic analysis in a number of applications, including, but not limited to, clinical diagnosis, pathogen discovery, de novo genome assembly, whole-exome sequencing, targeted gene panel sequencing, transcriptome profiling, and whole-genome resequencing.
Accordingly, the disclosure further provides for integrated assays to simultaneously use multiple sequencing platforms for metagenomic analysis, such as assay kits. Such assay kits can be used for applications, including but not limited to, clinical diagnosis with initial sequencing for rapid diagnosis (e.g., ONT platform) followed by more complete reflex sequencing for high sensitivity (e.g., Illumina platform); generating hybrid libraries for all sequencing applications, including, but not limited to, pathogen discovery, environmental metagenomics, de novo genome assembly whole-exome sequencing, transcriptomics sequencing (e.g., RNA-Seq); targeted gene panel sequencing; and whole-genome resequencing (e.g., cancer genome sequencing). Such assay kits provide a “one stop” kit to perform metagenomic analysis on samples, include primers, sequencing reagents, analysis software, etc. In a particular embodiment, the kit comprises, consists essentially of, or consists of dual use barcode primers that have been designed using the methods disclosed herein that can be used in both Illumina and Oxford Nanopore Technologies instruments. In another embodiment, a kit described herein is used to determine pathogenic microorganism(s) in patient sample(s) using the methods disclosed herein.
The assay kit will comprise a plurality of detection/quantification tools specific to each targeted biomolecule detected by the kit (e.g., pathogenic nucleic acid). Many of the targeted biomolecules disclosed herein comprise DNA, which may be detected by next generation sequencing and like technologies. The detection/quantification tools may comprise a set of dual use barcode primers, each barcode primer directed to the selective amplification by NGS of a targeted biomolecule(s) in a sample.
In yet another embodiment, the assay kits of the disclosure further comprise reagents or enzymes which can be used for next generation sequencing and like technologies. Assay kits may further comprise elements such as reference DNAs (e.g., positive and negative controls), washing solutions, buffering solutions, reagents, printed instructions for use, and containers.
The following examples are intended to illustrate but not limit the disclosure. While they are typical of those that might be used, other procedures known to those skilled in the art may alternatively be used.
Sample selection and processing. All body fluid samples were obtained from patients at the University of California San Francisco (UCSF) hospitals and clinics for three years. The study only used residual body fluid samples after standard-of-care clinical laboratory testing was performed. Body fluid samples were collected in sterile tubes or using swabs as part of routine clinical care and included abscess, joint, peritoneal, pleural, cerebrospinal, urine, bronchoalveolar lavage and other fluids (see Table 1). Swabs were stored in charcoal gel columns (Swab Transport Media Charcoal 220122, BD) and reconstituted in 0.5 mL of Universal Transport Media (350C, Copan Diagnostics, Murrieta, CA); the media liquid was subsequently used for culture, PCR, and mNGS analyses. Cultures for bacteria, fungi, and AFB from body fluid samples were done in-house at UCSF. Clinical 16S rDNA and 28S-ITS PCR for bacterial and fungal detection were performed by a reference laboratory at the University of Washington. Residual samples were stored at 4° C. and tested within 14 days of collection or centrifuged at 16,000 relative centrifugal force for 10 minutes and the supernatant stored at −80° C. until time of extraction.
Plasma samples were obtained by collecting blood from hospitalized patients as part of routine clinical testing into EDTA Plasma Preparation Tubes (BD) or standard EDTA Tubes (BD). The tubes were centrifuged (4000-6000 rcf for 10 minutes) within 6 hours, and plasma was isolated from the buffy coat and red cells. The plasma component was further aliquoted and centrifuged at 16,000 rcf for 10 minutes in microcentrifuge tubes. Plasma samples were stored at −80° C. until the time of extraction.
In the study of test performance, body fluids samples were included if they were culture positive or PCR positive for bacterial or fungal pathogen(s) with pathogen(s) identified to genus/species level. Body fluids from patients with ambiguous laboratory findings (e.g., a positive culture that was judged clinically to be a contaminant) or from patients with an established infectious diagnosis and already receiving targeted treatment at the time of body fluid collection were excluded. Negative control body fluid samples were selected from patients who had clear alternative non-infectious diagnoses (e.g., cancer, trauma) and negative for infection by culture and clinical adjudication (CYC and WG).
In the series of 12 cases, body fluid samples were included if (i) they were culture and PCR negative and (ii) from a patient with a microbiologically established infection (by orthogonal testing such as serology or testing of a different body fluid/tissue) or clinically probable infection based on review of the clinical charts by an infectious disease specialist (CYC) and clinical pathologist (WG) (Table 11).
Staphylococcus
aureus
Enterobacter
aerogenes
Streptococcus
Candida
parapsilosis
Staphylococcuss
Staphylococcus
aureus
aureus x1
Pseudomonas
aeruginosa
Sraphylococcus
aureus
Staphylococcus
Staphylococcus
aureus
aureus x10+
Haemophilus
Haemophilus
infuenzae
influenzae x1
Staphylococcus
aureus
Staphylococcus
aureus
Staphylococcus
aureus
Pseudomonas
aeruginosa
Staphylococcus
aureus
Serratia
marcescens
Candida
tropicalis.
bacteremia this
Staphylococcus
epidermidis
Serratia
marcescens
tropicalis
Enterococcus
faecium
Enterococcus
faecium
Candida
albicans,
Candida
glabrata
Staphylococcus
aureus,
Escherichia
coli
Staphylococcus
aureus
Enterococcus
faecium,
Candida
albicans
Candida
glabrata
Streptococcus
pyogenes
Enterobacter
cloacae
Candida
albicans
Staphylococcus
aureus
Streptococcus
mitis group
Staphylococcus
aureus
Streptococcus
anginosus
Corynebacterium
Staphylococcus
aureus
Enterococcus
Enterococcus
faecalis,
faecalis x2
Entercoccus
faecium,
Candida
knaei
Staphylococcus
aureus
Staphylococcus
aureus
Streptococcus
pyogenes
Escherichia
coli
Staphylococcus
aureus
Staphylococcus
aureus
Staphylococcus
aureus
Staphylococcus
aureus
Excherichia
coli,
Klebsiella
pneumaniae
Staphylococcus
aureus
Enterococcus
Candida
albicans
Escherichia
coli
Aspergillus
fumigatus
Staphylococcus
aureus
Klebsiella
pneumaniae,
Citrobacter
Citrobacter
freundi
freundii
Escherichia
coli
Staphylococcus
aureus
Klebsiella
pneumoniae
Enterococcus
faecium
Aspergillus
fumigatus
Pseudomonas
aeruginosa,
Candida
glabrata,
Candida
krusei
Pseudomonas
aeruginosa
Staphylococcus
lugdimensis
Escherichia
coli
Staphylococcus
lugdimensis
Mycoplasma
hominis
Streptococcus
pyogenes
Mycobacterium
tuberculosis
Staphylococcus
aureus
Staphylococcus
aureus
Candida
albicans
Staphylococcus
aureus
Staphylococcus
aureus
Streptococcus
mitis group
Escherichia
coli
Candida
albicans
Escherichia
coli
Coccidioides
immitis
Coccidioides
immitis
Streptococcus
pyogenes
Propionibacterium
acnes)
Aspergillus
terreus,
Aspergillus
fumigatus
Streptococcus
pyogenes
Staphylococcus
Staphylococcus
aureus
aureus
Finegoldia
magna
infection
Streptococcus
pneumoniae
Streptococcus
Nocardia
farcinica
Caccidioides
immitis
Staphylococcus
aureus
Staphylococcus
aureus
Candida
glabrata
Streptococcus
Streptococcus
pneumoniae
Staphylococcus
epidermidis
Candida
albicans
Staphylococcus
aureus
Escherichia
coli
Candida
glabrata
Cryptococcus
neoformans
Candida
parasilopsis
Candida
parasilopsis
Candida
parasilopsis
Cryptococcus
neoformans
Aspergillus
niger, rare A.
flavus, rare A.
fumigatus), Few
Oronosal flora
Staphylococcus
aureus
Enterococcus
faecium
Staphylococcus
aureus
Coccidioides
immitis
Streptococcus,
Corynebacterium
diphthering
Cryptococcus
neoformens
Salmonella
typhi group D
Mycobacterium
tuberculosis
Escherichia
coli
Enterobacter
cloacae,
Candida
albicans
Coccidioides
immitis
Achromobacter
xylosaxidans
Cryptococcus
gattii
Coccidioides
immitis
Histoplasma
capsulatum
Pneumnocytis
jireoveci
Coccidioides
immitis
Staphylococcus
aureus
Streptococcus
Klebsiella
pneumoniae
Klebsiella
pneumoniae
Enterobacter
aerogenes
Staphylococcus
aureus
Staphylococcus
aureus
Staphylococcus
aureus
Aspergillus
fumnigatus
Klebsiella
pneumoniae
Staphylococcus
aureus
Enterococcus
faecium
Staphylococcus
lugdunensis
Candida
parasilopsis
Candida
tropicalis
Mycobacterium
avium
Mycobacterium
avium
Nocardia
blacklockiae
Listeria
monocytogenes
Listeria
monocytogenes
Staphylococcus
aureus
Staphylococcus
aureus
Enterococcus
faecium,
Enterococcus
faecalis
Staphylococcus
aureus
Staphylococcus
aureus; also
Mycobacterium
tuberculosis
Mycobacterium
tuberculosis
Escherichia coli
Staphylococcus
epidermidis
Enterobacter
aerogenes
Cryptococcus
pneumonia
Staphylococcus
Staphylococcus
aureus
aureus
Enterobacter
Klebsiella
aerogenes
aerogenes
Streptococcus
Streptococcus
agalactiae
agalactiae
Candida
Candida
parapsilosis
parapsilosis
Saccharomyces
cerevisiae, off
Saccharomyces
cerevisiae
Pseudomonas
Pseudomonas
aeruginosa
aeruginosa
Pseudomonas
aeruginosa
Staphylococcuss
Staphylococcuss
aureus
aureus
Haemophilus
Haemophilus
influenzae
influenzae
Rothia
dentocariosa
Staphylococcus
aureus
Staphylococcus
aureus
Pseudomonas
Pseudomon
aeruginosa
deruginosa
Staphylococcus
Staphylococcus
aureus
aureus
Serratia sp.
Serratia
marcescens,
Enterococcus
Enterococcus
faecium
faecium
Staphylococcus
Staphylococcus
epidermidis
epidermidis
Serratia sp.
Serratia
marcescens
Enterococcus
Enterococcus
faecium
faecium
Enterococcus
Enterococcus
faecium
faecium,
Stenotrophomonas
maltophilia
Enterococcus
Enterococcus
faecium
faecium
Candida
Candida
albicans,
glabrata,
Candida
Candida
glabrata
albicans
Staphylococcus
aureus
Enterococcus
Enterococcus
faecium
faecium
Candida
Candida
glabrata,
glabrata
Candida
albicans
Escherichia
Escherichia
coli
coli
Enterobacter
Enterobacter
cloacae
kobei
Staphylococcus
Staphylococcus
aureus
aureus
Klebsiella
Klebsiella
pneumoniae
pneumoniae
Staphylococcus
aureus
Streptococcus
Streptococcus
constellatus,
anginosus
Streptococcus
anginosus
Streptococcus
intermedius,
Parvimonas
micra
Staphylococcus
aureus
Enterococcus
Enterococcus
faecalis,
faecium,
Enterococcus
Enterococcus
faecalis,
faecalis
Prevotella
malaningenica,
Lactobacillus
gasseri,
Campylobacter
curvus,
Peptoclostridiam
difficile,
Campylobacter
conrisus
Staphylococcus
aureus
Achromobacter
xylosoidans
Streptococcus
Streptococcus
pyogenes
pyogenes
Escherichia
Escherichia
coli
coli
Staphylococcus
Staphylococcus
aureus
aureus
Staphylococcus
Staphylococcus
aureus
aureus
Staphylococcus
Staphylococcus
aureus
aureus
Staphylococcus
Staphylococcus
aureus
aureus
Klebsiella
Klebsiella
pneumaniae,
pneumaniae,
Excherichia
Excherichia
coli
coli
Enterococcus
Enterococcus
faecium,
faecium,
Candida
Candida
albicans
albicans
Klebsiella
Klebsiella
pneumoniae
pneumoniae
Streptococcus
pneumoniae,
Aspergillus
fumigatus,
Prevotella
melaminogenica
Staphylococcus
Staphylococcus
aureus
aureus
Citrobacter
Klebsiella
freundii,
pneumaniae,
Klebsiella
Citrobacter
pneumoniae,
freundii
Salmonella
enterica,
Bacteroides
xylanisolvens,
Bacteroides
thetaiotaomicron
Escherichia
Escherichia
coli
coli
Staphylococcus
aureus
Klebsiella
Klebsiella
pneumoniae,
pneumoniae
Enterococcus
faecium,
Klebsiella
Klebsiella
pneumoniae
pneumoniae
Aspergillus
Aspergillus
fumigatus
fumigatus
Pseudomonas
Pseudomonas
aeruginosa,
aeruginosa,
Staphylococcus
Candida
epidermidis,
glabrata,
Candida
Staphylococcus
glabrata,
epidermidis,
Pichia kluyveri,
Neisseria
Lactococcus
sicca,
lactis,
Leuconostoc
Campylobacter
citreum
concisus,
Lactobacillus
acidophilus,
Veillonella
parvuila
Pseudomonas
Pseudomonas
aeruginosa
aeruginosa
Staphylococcus
Staphylococcus
lugdimensis
lugdimensis
Escherichia
Escherichia
coli
coli
Mycoplasma
Mycoplasma
hominis
hominis
Streptococcus
Streptococcus
pyogenes
pyogenes
Mycobacterium
Mycobacterium
tuberculosis
tuberculosis
Staphylococcus
Staphylococcus
aureus
aureus
Staphylococcus
Staphylococcus
aureus
aureus
Candida
albicans
Staphylococcus
Staphylococcus
aureus
aureus
Enterococcus
Enterococcus
faecalis,
faecalis,
Escherichia
Escherichia
coli,
coli,
Staphylococcus
Staphylococcus
aureus,
aureus
Streptococcus
mitis,
Bifidobacterium
breve,
Peptoclostridsum
difficile
Streptococcus
Streptococcus
mitis group
gordonii
Escherichia
Escherichia
coli
coli
Escherichia
Escherichia
coli
coli
Corynebacterium
striatum
Coccidioides
Coccidioides
immitis
immitis
Coccidioides
Coccidioides
immitis
immitis
Streptococcus
Streptococcus
pyogenes
pyogenes
Aspergillus
Aspergillus
terreus,
terreus,
Aspergillus
Aspergillus
fumigatus
fumigatus
Streptococcus
Streptococcus
pyogenes,
pyogenes
Talaromyces
marneffei
Finegoldia
magna,
Bartonella
henselae
Klebsiella
Klebsiella
aerogenes
aerogenes
Polymicrobial
anaerobes
Streptococcus
pneumoniae
Aspergillus
fumigatus
Coccidioides
immitis
Staphylococcus
aureus
Staphylococcus
aureus,
Enterococcus
faecalis
Candida
glatrata
Streptococcus
pyogenes
Streptococcus
pneumoniae
Staphylococcus
epidermis
Candida
parapsilosis,
Candida
albicans
Staphylococcus
aureus
Escherichia
coli
Candida
glabrata
Cryptococcus
neoformans
Candida
parasilopsis
Candida
parasilopsis
Candida
parasilopsis
Cryptococcus
neoformans
Pseudomonas
aeruginosa
Staphylococcus
aureus
Enterococcus
faecium
Staphylococcus
aureus
Coccidioides
immitis,
Methylobacterium
radiotolerans,
Methylobacterium
extarquens,
Burkholderia
gladioli,
Methylobacterium
populi,
Sphingomonas
taxi
Streptococcus
pyogenes,
Corynebacterium
diphtheriae
Cryptococcus
neoformans,
Streptococcus
parasanguinis
Salmonella
enterica
Mycobacterium
tuberculosis
Escherichia
coli
Enterobacter
cloacae,
Candida
albicans
Coccidioides
immitis
Achromobacter
xylosoxidans,
Staphylococcus
epidermidis,
Pseudomonas
pseudoalcoliganes
Cyptococcus
gattii
Coccidioides
immitis
Pneumocystis
jirovecii
Coccidioides
immitis
Aggregatibacter
aphrophilus
Staphylococcus
aureus
Streptococcus
agalactiae
Klebsiella
pneumoniae
Enterobacter
aerogenes
Staphylococcus
aureus,
Bacillus
thuringiensis
Staphylococcus
aureus
Staphylococcus
aureus
Aspergillus
fumigatus
Klebsiella
pneumoniae,
Sodatis sp.
Veilionella
parvula
Staphylococcus
aureus
Enterococcus
faecium,
Bordetella
petrii
Staphylococcus
lugdunensis
Candida
tropicalis
Mycobacterium
avium
Listeria
monocytogenes
Staphylococcus
aureus
Staphylococcus
aureus
Enterococcus
faecium,
Enterococcus
faecalis
Peptoclostridium
difficile
Staphylococcus
aureus
Propionibacterium
acnes
Streptococcus
Streptococcus
oralis,
Veillonella
parvula, Rothia
muculaginosa
DNA extraction. Samples were processed in a blinded fashion In a CLIA (Clinical Laboratory Improvement Amendments)-certified clinical microbiology laboratory with physically separate pre- and post-PCR rooms. Cells were first removed through centrifugation to minimize host background. 400 μL of body fluid supernatant or plasma then underwent total nucleic acid extraction to 60 μL extract using the EZ1 Advanced XL BioRobot and EZ1 Virus Mini Kit v2.0 (QIAGEN) according to the manufacturer's instructions.
Library preparation and PCR amplification. Library preparation was performed using the NEBNext Ultra II DNA Library Prep Kit (New England Biolabs), with the use of 25 μL of extracted DNA input and half of the reagent volumes suggested by the manufacturer's protocol. Briefly, extracted DNA from most samples was quantified on a NanoDrop spectrophotometer (ThermoFisher) and diluted to 10-100 ng of input as recommended by the manufacturer. Plasma or CSF DNA was not quantified or diluted as typical input concentrations of <10 ng/μL could not be reliably detected using a spectrophotometer. The DNA was then end-repaired, ligated with the NEBNext Adapter (0.6 μM final concentration) to enrich for short-fragment pathogen DNA (100-800 nt) relative to residual human genomic DNA (>1 kb), and cleaned using AMPure beads. In addition to the initial manual preparation of 17 samples, an automated protocol using the epMotion 5075 liquid handler (Eppendorf) was used to process the remaining 165 samples, with 16-48 samples batch-processed per run.
PCR amplification was performed using a 40 μL mix consisting of adapter-ligated DNA, premixed custom index primers at 3 μM final concentration (see Table 2), and a quantitative PCR master mix (KAPA RT-kit, KK2702, Roche). DNA amplification was performed to saturation of the fluorescent signal on a qPCR thermocycler (Lightcycler 480, Roche) using the following PCR conditions: initiation at 98° C.×45 s, then 24 cycles of 98° C.×15 s/63° C.×30 s/72° C.×90 s, and a final extension step of 72° C.×60 s. Ct values were continually monitored until the libraries were fully amplified to saturation. Final DNA libraries were cleaned up using Ampure beads (Beckman) at a 0.9×volumetric ratio and eluted in 30 μL EB buffer (Qiagen).
Dual-use protocol for Illumina and nanopore sequencing platforms. Multiplexing barcodes on the Illumina platform typically have the lengths of 8 nt flanking the sequence read on both ends, but they are not ideal for multiplexing samples being sequenced on a nanopore instrument (Oxford Nanopore Technologies) due to the higher error rate of this platform. A dual-use barcode system was designed that contains, in an exemplary embodiment, a distinct 37 nucleotide (nt) barcode on each side of sequencing adaptor (the first 8 nt of which were used for Illumina multiplexing), which enables the multiplexed DNA library to be sequenced on both Illumina and nanopore platforms. The barcodes were designed using an in-house developed R script to generate 8 nt and 29 nt barcodes that maximized the Levenshtein distance between any given pair of barcodes. Specifically, the DNA Barcodes package in BioConductor was first used to generate a set of 1,014 unique 10mers with minimum Levenshtein distance of 4 and a set of 283 unique 8mers with minimum Levenshtein distance of 3, as computational limitations prevented design of 37mers directly. An 8mer and three 10mers were then concatenated together with stripping of the last nt to generate a 37mer index primer. The final set of 192 37mer (
Pipeline for Designing 37mer Barcodes for Dual Use of Illumina and Oxford Nanopore Technologies (Nanopore) Sequencing Instruments. 37mer barcodes that could effectively be used for both Illumina and Nanopore were designed with the following goals in mind: generate 192 37mer barcodes such that unique dual indexes (UDIs) can be used for all 96 samples on a 96-well plate; maximizing Levenshtein distance between each pair of barcodes in order to minimize barcode “crosstalk”; Ensuring a minimum Levenshtein distance between each pair of 8mers (the first 8 nt (nucleotides) of each 37mer) for Illumina sequencing; and ensuring a minimum Levenshtein distance between each pair of 37mers for nanopore sequencing on Oxford Nanopore Technologies (ONT) instruments. It was determined that it was too computationally expensive to determine DNA barcodes that maximize the Levenshtein distance between them for barcodes of >12 nt in length. Instead, the strategy was to design a set of 8mer barcodes and 3 sets of unique 10mer barcodes that are concatenated together to form 37mer barcodes.
A custom algorithm was developed in python to carry out the above strategy. For the python code, the following scripts were utilized: Run the R script “generate_8mer barcodes.R”. This script uses the DNABarcodes tool in the Bioconductor software package (v.3.1.1). These scripts were run in R v4.0. Required additional R packages include Matrix and parallel. Note that setting the Levenshtein distance threshold to 3 will not yield a sufficient number of candidate barcodes (n=187). Thus, barcodes generated from setting the Levenshtein distance threshold to 3 and then rerunning the R script with the Hamming distance threshold set to 3 were combined.
This R script generates a total of 914 candidate barcodes at a Levenshtein or Hamming distance threshold of 3. Note that the DNABarcodes algorithm is non-coordinated, meaning that it will not generate identical results when the program is rerun. The pairwise distances between any two barcodes can be calculated using Linux commands that will also auto-generate an R script.
The in-house Python script “parse_lev_distance.py” can then be used to pull out all barcodes at a predefined minimum Levenshtein distance threshold of 3.
It was found that there were 283 8mer barcodes at a Levenshtein threshold cutoff of 3 generated in total (“8mer_barcodes.txt”). The minimum Levenshtein distance for any given set of barcodes can be separately checked with a Bash shell script “check_distance_lev.sh”.
Next, to generate the remaining 29 nt stretch of the 37mer barcodes, the DNABarcodes tool was used to generate 12mer (“generate_12mer_barcodes.R”) and 10mer (“generate_10mer_barcodes.R”) barcodes and then concatenate 1 12mer and 2 10mer barcodes to generate a 32mer barcode, from which 3 nucleotides can be stripped to generate 29mer barcodes.
This R script generated 232 12mer barcodes and 1,014 10 mer barcodes. The pairwise distances between any two barcodes can be calculated using Linux commands, and which also auto-generate an R script.
An in-house Python script “parse_lev_distance.py” can then be used to pull out all barcodes at a predefined minimum Levenshtein distance threshold of 4.
Use of the script results in 232 12mer barcodes (“12mer_barcodes.txt”) and 1,014 10mer barcodes (“10mer_barcodes.txt”). Note the minimum Levenshtein distances cannot be increased by 1 to 6 and 5, because this will decrease the number of usable barcodes to less than 192 (45 and 72, respectively).
Next, randomly from the pool of 232 12mer barcodes (“12mer_barcodes.txt”) and 1,014 10mer barcodes (“10mer_barcodes.txt”), barcodes were selected and concatenated these chosen barcodes to generate 232 32mer barcodes (12mer-10 mer-10 mer) (“32mer_barcodes.txt”). 3 nucleotides were stripped off of the 3′ end and then concatenated to a random 8mer barcode from the pool of 238 8mer barcodes (“8mer_barcodes.txt”) to the 5′ end to generate 37mer barcodes. Finally, 192 of the “best” 37mer barcodes (“37mer_barcodes.txt”) were selected on the basis of maximum pairwise Levenshtein distance selected using similar Linux commands as described above. The minimum pairwise Levenshtein distance for this set of 192 37mer barcodes is 14 using the “check_distance_lev.sh” shell script.
Illumina sequencing. DNA libraries were pooled in equal volumes and the sequencing library pool was quantified using the Qubit fluorometer (ThermoFisher). Illumina sequencing was performed on MiSeq (2×150 nt paired-end)(with capacity for up to 5 samples per run) or HiSeq 1500/2500 instruments (140 nt single or 2×140 nt paired-end, with capacity for up to 40 samples per lane), according to the manufacturer's protocol.
Nanopore sequencing. Stringent procedures were adopted to prevent cross-contamination between samples during the library preparation steps, including unidirectional workflow, separating pre-PCR and post-PCR workspaces, and regular cleaning of the workbenches and biosafety cabinets with 5% sodium hypochlorite. Amplified DNA libraries were prepared for nanopore sequencing using the 1D library preparation kit (Oxford Nanopore Technologies) either manually or on an epMotion 5075 liquid handler biorobot (Eppendorf), with the processing of 8-16 samples per batch. The input DNA ranged from 200-1000 ng. The DNA was then sequenced using either R9.4 or R9.5 flow cells on a MinION or GridION X5 instrument (Oxford Nanopore Technologies). The MinION has a single flow cell position for processing of a single sample at a time, while the GridION has 5 flow cell positions for processing of up of 5 samples simultaneously. Up to five individually barcoded samples per flow cell were sequentially loaded on the nanopore instrument for sequencing. Between each sample, flow cells were washed according to the manufacturer's instructions to minimize carryover contamination. The estimated cost for reagents per sample (excluding labor) was $27.20-$61.40 and $269.70 for Illumina and nanopore sequencing, 589 respectively.
Positive and negative external controls. Negative controls were from the same batch of pooled plasma from healthy donors (Golden West Biologicals, CA). Positive controls consist of the negative control plasma spiked with sheared (to 150-200 base pair range) DNA extracted from cultured non-pathogenic microorganisms (American Type Culture Collection, VA): Koi herpesvirus (virus, VT-1592D), Streptococcus uberis (gram-positive bacterium, ATCC strain 0140J BAA-854D-5), Rhodobacter sphaeroides (gram-negative bacterium, ATCC BAA-808D-5), Millerozyma farinosa (yeast, ATCC MYA-4447D-5), Aspergillus oryzae (mold, ATCC 42149D-2), and Neospora caninum (parasite, ATCC 50843D) (see Table 3). All controls underwent the same wet lab procedure and bioinformatics analysis as the clinical samples.
Thermus
scotoductus
Propionibacterium
acnes
Methylobacterium
radiotolerans
Variovorax
paradoxus
Propionibacterium
acnes
Achromobacter
xylosoxidans
Pseudomonas
pseudoalcaligenes
Thermus
scotoductus
Verminephrobacter
eiseniae
Comamonas
testosteroni
Achromobacter
xylosoxidans
Cutibacterium
acnes
Pseudomonas
fluorescens
Streptococcus
uberis
Rhodobacter
sphaeroides
Millerozyma
farinosa
Aspergillus oryzae
Neospora caninum
Thermus
scotoductus
Achromobacter
xylosoxidans
Propionibacterium
acnes
Pseudomonas
pseudoalcaligenes
Streptococcus
uberis
Rhodobacter
sphaeroides
Millerozyma
farinosa
Aspergillus oryzae
Neospora caninum
Cutibacterium
acnes
Thermus
scotoductus
Limits of Detection and Linearity. To evaluate the limit of detections for bacteria and fungi for this assay, DNA was spiked from non-pathogenic microorganisms acquired from ATCC into healthy donor negative plasma in a series of 4-fold dilutions from 1:1 (no dilution) to 1:4096 (see Table 4). Each concentration of microorganism was tested on mNGS with 4 replicates for reproducibility. The bacteria and fungi tested include Streptococcus uberis, Rhodobacter sphaeroides, Millerozyma farinosa, and Aspergillus oryzae. Thresholds were chosen based on the nPRM corresponding to Youden's index on the training data ROC curve and using the composite standard. The bacterial nRPM thresholds were 2.6 and 0.54 for Illumina and nanopore sequencing, respectively; the fungal nRPM was 0.10 for Illumina and nanopore sequencing. The LoD was defined as the dilution at which mNGS testing detected the pathogen at levels above the nRPM threshold in 4 of 4 replicates. To evaluate assay linearity, a linear regression was performed on the same four sets of 615 serially diluted positive controls used in the LoD. The nRPM values were plotted against the input concentration (copies, or genome equivalents per mL). The best fit regression line along with the linear equation and R2 value was added to the plotted values (see
Aspergillus
fumigatus
Candida
albicans
Candida
glabrata
Candida
albicans
Candida
albicans
Candida
krusei
Candida
albicans
Aspergillus
fumigatus
Candida
albicans
Candida
albicans
Candida
glabrata
Pichia
kluyveri
Candida
krusei
Coccidioides
immitis
Coccidioides
immitis
Aspergillus
terreus
Aspergillus
fumigatus
Talaromyces
marneffei
Candida
parapsilosis
Saccharomyces
cerevisiae, which
Candida
albicans
Candida
glabrata
Coccidioides
immitis
Blastomyces
dermatitidis
Lodderomyces
elongisporus
Lodderomyces
elongisporus
Aspergillus
fumigatus
Coccidioides
immitis
Cryptococcus
neoformans
Coccidioides
immitis
Candida
parapsilosis
Candida
tropicalis
Candida
glabrata
Candida
albicans
Candida sp.
Candida
glabrata
Cryptococcus
neoformans
Candida
parapsilosis
Candida
parapsilosis
Candida
parapsilosis
Cryptococcus
neoformans
Aspergillus spp
Aspergillus
oryzae
Cryptococcus
gattii
Malassezia
globosa
Coccidioides
immitis
Penicillium
rubens
Candida sp.
Candida
albicans
Aspergillus
fumigatus
Tetrapisispora
blattae
Coccidioides
immitis
Histoplasma
capsulatum
Pneumocystis
jirovecii
Candida
parapsilosis
Candida
glabrata
Candida
albicans
Aspergillus
fumigatus
Aspergillus
terreus
Aspergillus
fumigatus
Candida
albicans
Candida
albicans
Candida
krusei
Aspergillus
fumigatus
Candida
albicans
Candida
albicans
Coccidioides
immitis
Candida
glabrata
Candida
albicans
Candida
glabrata
Candida
krusei
Coccidioides
immitis
Bioinformatics analysis. Illumina sequencing data were analyzed for pathogens using the clinically validated SURPI+(sequence based ultra-rapid pathogen identification) computational pipeline v1.0.63-dev 7,20,58. SURPI+ uses the entirety of the NCBI GenBank nt database (March 2015 distribution) as the reference database and incorporates taxonomic classification algorithms for accurate identification of pathogens as described in Miller et al. (Genome Res 29:831-842 (2019)). Nanopore sequencing data were analyzed using SURPIrt (SURPI “real-time”) software (SURPIrt research 1.0.14-build.86). Raw fast5 files were base called using MinKNOW software v3.1.20 installed on the GridION in real-time mode without polishing. The base called reads were run through in-house developed scripts for sample demultiplexing using the BLASTn (v2.7.1+) aligner at a significance E-value threshold of 10-2. After trimming adapters and removing low-quality and low-complexity sequences, the first 450 nt of the preprocessed read was partitioned into three 150 nt segments, followed by rapid low-stringency identification of candidate pathogen reads using SNAP (version 1.0dev100) alignment to microbial reference databases (viral portion of 2019 NCBI nt; bacterial RefSeq; fungal and parasitic pathogens in the fungal RefSeq and parasitic RefSeq databases), using an edit distance of 5059. Candidate reads were then filtered and taxonomically classified as described in Miller et al. Real-time analysis was performed by running the SURPIrt pipeline in continuously looping mode, with ˜100k-200k nanopore reads analyzed per batch.
Computational algorithm for pathogen identification. A pathogen identification algorithm that was applicable for both Illumina and 642 nanopore datasets outputted by SURPI or SURPIrt (see above) was developed to assess and optimize performance accuracy. An initial reference database was manually tabulated based on pathogens detected in body fluids by culture and/or PCR testing. The algorithm calculated a nRPM pathogen count, filtered out taxonomically related microorganisms, and defined criteria for pathogen detection, as explained in detail below.
(1) Calculating a normalized RPM. A nRPM metric was developed to standardize microorganisms across samples with uneven sequencing depths and input DNA concentrations. For Illumina sequencing, the RPM was defined as the number of pathogen reads divided by the number of preprocessed reads (reads remaining after adapter trimming, low-quality filtering, and low-complexity filtering), while for nanopore sequencing, the RPM was defined as the number of pathogen reads divided by the number of base called reads. A nRPM was calculated that adjusted the RPM with respect to background based on the Ct value (to the nearest 0.5 increment) during the PCR amplification step of library preparation. As the average Ct value across all samples was 7, the nRPM was defined as nRPM=RPM/2 (Ct-7). Receiver-operating characteristic (ROC) and precision-recall curves were plotted using the Python software package and pandas data analysis library. The optimal nRPM threshold was obtained by plotting the ROC curve at varying nRPM values and determining the nRPM at Youden's Index. The incorporation of a nRPM metric is based on a previous observation of a log-linear relationship between the qPCR Ct value and the RPM of representative, presumed background contaminant microorganisms such as Achromobacter xylosoxidans (see
(2) Filtering out closely-related microorganisms. Taxonomic classification using metagenomics data commonly yields a minority fraction of reads that map to related taxa with the same family or genus as the microorganism truly present in the sample. In order to minimize cross-species misalignments for closely related microorganisms, the nRPM of microorganisms that share a genus or family designation was penalized (reduced). A penalty of 10% and 5% was used for genus and family respectively, based 675 on the empirical maximization of specificity from the ROC curve of the training set. For example, 676 if Escherichia coli had an nRPM of 100 and Shigella sonnei (from same Enterobacteriaceae 677 family) had an nRPM of 5, the nRPM of Shigella sonnei would be reduced to zero. In the current study, better performance was achieved in the training dataset using this filter.
(3) Criteria for pathogen detection. Two criteria were developed for pathogen detection. The candidate pathogen was required to 683 (i) have a minimum number of pathogen-specific reads identified (≥3 for bacteria and >1 for fungi) (see
Statistical methods. To evaluate accuracy, two criteria were applied: (i) a clinical gold standard based on culture and 16S PCR results obtained through routine clinical care, and (ii) a composite standard based on a combination of clinical testing (culture and 16S/28S-ITS PCR), orthogonal testing (e.g., digital PCR, serology), and clinical adjudication. The specific scoring algorithm is outlined as follows (see Table 5): Based on the clinical or composite standard, true positives (TP) or false negatives (FN) were scored for each microorganism that was detected or not detected by mNGS, respectively. For each sample, a true-negative (TN) was scored if no other microorganism(s) other than the expected ones based on the clinical or composite standard were detected by mNGS; otherwise, a false-positive (FP) was scored. Multiple FP results in a sample were counted as one FP overall. p-values were calculated using a two-sided Welch's t-test at a significance p-value threshold of 0.05. All data points in the study were performed once, except the LoD studies which were performed in four replicates at each dilution.
Confidence intervals for the ROC curves. To evaluate the reliability of the validation set data, a custom python script was coded that bootstrapped the dataset by randomly resampled the dataset with replacement to generate a replicate dataset of the same size for 2000 iterations. The resultant distribution was used to produce a 95% confidence interval (CI) for the ROC curve (see
Orthogonal confirmation of mNGS results. Digital PCR (dPCR) for orthogonal confirmation of mNGS results was performed using the Biorad QX200 Droplet Digital PCR System. The advantages of dPCR include the ability for absolute quantification, improved detection of very low-abundance nucleic acids with high precision, and higher tolerance to the presence of inhibitors and/or contaminants in the body fluid samples. Thus, the use of dPCR was deemed to be a more robust indicator for the presence of pathogen-specific DNA in the body fluids than conventional PCR. All primer and probe pairs were synthesized by Integrated DNA Technologies, Inc. and first validated using positive control microorganisms (see Table 6). Genomic DNA from positive control microorganisms was purchased from ATCC and mechanically sheared (MiniTUBE, Covaris) to an average of 200-300 base pairs. For Sanger sequencing, DNA was first cloned into colonies using a TOPO TA Cloning Kit (ThermoFisher). Sanger sequencing of the clones was then performed at Elim Biopharmaceuticals, Inc. Sequencing traces were analyzed on Geneious software (version 10.2.3) and aligned to the National Center for Technology Information nt database using BLAST. Serology confirmation of the Bartonella case was performed by Quest Diagnostics.
Saccharomyces
Cerevisiae
cerevisiae
Streptococcus
Escherichia
coli
indicates data missing or illegible when filed
Analysis of pathogen and human DNA lengths. Pathogen-specific length distributions in mNGS data were obtained by aligning paired-end Illumina reads or single-end nanopore reads to individual pathogen genomes (see
For characterization of human DNA length distributions from Illumina data, FASTQ files were first trimmed for Illumina adapters with cutadapt (v1.16), followed by alignment with BWA 742 (v0.7.12) to the hg38 human reference genome. This revealed a previously described peak of ˜160 nt that corresponds to nuclear DNA wrapped around a single histone (see
Length distributions were assessed from 58 bacterial and 10 fungal pathogens by histogram analysis, with the inclusion criteria of at least 10 paired-end reads aligned to each pathogen genome (see
Data availability. Metagenomic sequencing data (FASTQ files) after removal of human genomic reads have been deposited into the NCBI Sequence Read Archive (SRA) (PRJNA558701, under umbrella project PRJNA234047).
Software and code accessibility. SURPI+v1.0 (github.com/chiulab/SURPI-plus-dist) and SURPIrt v1.0 software 761 (github.com/chiulab/SURPIrt-dist) have been deposited on GitHub and are available for download for research use only. Linux (Ubuntu 16.04.6) and Python (python 2.7.12) scripts used for construction of dual-use Illumina and nanopore barcodes are provided below. Other custom scripts for ROC curve and read length analysis have been deposited on Github (github.com/wei2gu/2020-NGSInfectedBodyFluids/).
Sample Collection. A total of 182 body fluid samples from 160 patients, including 25 abscess, 21 joint, 32 pleural, 27 peritoneal, 35 cerebrospinal, 13 bronchoalveolar lavage (BAL), and 29 other body fluids (see Table 7 and Table 1), were collected as residual samples after routine clinical testing in the microbiology laboratory. Among the 182 samples, 170 were used to evaluate the accuracy of mNGS testing by Illumina sequencing (see
Study Patients. Among 158 patients out of 160 with available clinical data, 144 (91%) were hospitalized, of whom 61 (39%) required intensive care unit (ICU) management and 45 (28%) met clinical criteria for sepsis (32%) were immunocompromised due to organ transplantation, recent chemotherapy, human immunodeficiency virus (HIV) infection, or drug-induced immunosuppression, and 71 (45%) were on antibiotics at the time of body fluid collection (Table 7). According to usual standard-of-care practices, bacterial cultures were obtained for all body fluids, with 63 (35%) and 81 (45%) having additional cultures done for acid-fast bacilli (AFB) and fungi, respectively.
Staphylococcus aureus
Streptococcus sp.
Enterococcus sp.
aSIRS, systemic inflammatory response syndrome
bvitreous fluid, perihepatic fluid, surgical swab, subgaleal fluid, heel fluid swab, peri-graft fluid swab, anterior mediastinal fluid, chest fluid, chest wall mass, wound swab, synovial fluid, breast fluid, back fluid, fine needle aspirate (FNA), left thigh bursal fluid, peri-gastric fluid, thoracic spine seroma, peri-tonsillar drainage, knee swab, ililpsoas collection fluid, iliac wing fluid, retrogastric fluid, and urine
Metagenomic Sequencing Analysis. A dual-use barcoding protocol was developed for mNGS testing that was cross-compatible on both nanopore and Illumina sequencing platforms, suitable for all body fluids, and automated in the clinical microbiology laboratory on liquid handling workstations. The amount of input DNA varied over 6 logs from approximately 100 pg/mL in low cellularity fluids such as CSF to 100 μg/mL in purulent fluids. The median read depths for Illumina and nanopore sequencing were 7.2M (interquartile range or IQR 4.0-8.3M, range 0.15-35M) and 1.1M (IQR 1.0-1.5M, 137 range 0.29-6.7M), respectively (see Table 1). Metagenomic analysis for pathogen detection from Illumina data was performed using clinically validated SURPI+ software. Nanopore sequencing yielded 1 million reads per hour on average, with real-time data analysis performed using SURPIrt software, a new in-house developed bioinformatics pipeline for pathogen detection from metagenomic nanopore sequence data. After a 5-hour library preparation, nanopore sequencing detected pathogens in a median time of 50 minutes (IQR 23 143-80 minutes; range 21-320 minutes) (see
Test Accuracy. The accuracy evaluation focused on the performance of mNGS relative to gold standard culture and/or PCR testing for pathogen detection (see
Staphylococcus aureus
Staphylococcus aureus,
Pseudomonas aeruginosa
Serratia sp. SCBI
Serratia sp. SCBI
Enterococcus faecium
Serratia sp. SCBI
Serratia sp. SCBI
Enterococcus faecium
Streptococcus mitis
Klebsiella pneumoniae
Escherichia coli
Escherichia coli,
Klebsiella pneumoniae
Enterococcus faecium
Enterococcus faecium,
Klebsiella pneumoniae
Staphylococcus aureus
Staphylococcus aureus,
Enterococcus faecalis,
Escherichia coli
Pseudomonas aeruginosa,
Pseudomonas aeruginosa,
Candida glabrata,
Candida glabrata,
Candida krusei
Candida krusei,
Staphylococcus epidermidis
Aspergillus spp
Pseudomonas aeruginosa,
Aspergillus spp
At the optimal Youden's index (nRPM threshold of 2.6 and 0.54 for Illumina and nanopore sequencing, respectively) derived from the training set ROC curve, the sensitivity and specificity of mNGS testing for bacterial detection based on the validation set using the clinical gold standard were 79.2% (95% confidence interval CI 73.5-85.2%), and 90.6% (95% CI 87.3-93.8%), respectively, for Illumina sequencing, compared to 75.0% (95% CI 65.0-85.7%) and 81.4% (95% CI 74.1-89.3%), respectively, for nanopore sequencing (see
Among the 34 negative control samples that were negative by culture and 16S-PCR (see
Among false-negative cases from both the training and validation sets using the composite threshold, the most common missed organism was Staphylococcus aureus (see Table 1). Illumina sequencing missed 10 of 40 (25%) cases of Staphylococcus aureus, but this was not statistically significant compared to missed cases of infection from other bacteria (12 of 81, 18%) (p=0.21, Fisher's Exact Test). Nanopore sequencing missed 10 of 26 (38.5%) cases of Staphylococcus aureus, statistically significant compared to missed cases from other bacteria (4 of 50, 8%) (p=0.0034, Fisher's Exact Test) (see Table 9).
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Streptococcus
Enterococcus
Staphylococcus aureus
Pseudomonas aeruginosa
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Salmonella enterica
Staphylococcus aureus
Enterococcus faecium
Staphylococcus
Staphylococcus aureus
Enterococcus faecalis
Staphylococcus
Methylobacterium
Nocardia farcinica
Bartonella
Mycobacterium
Mycobacterium avium
Mycobacterium
Staphylococcus
Nocardia nova
Mycobacterium avium
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Staphylococcus aureus
Enterococcus faecium
Enterococcus faecium
Neisseria
Staphylococcus
Staphylococcus aureus
Klebsiella pneumoniae
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Klebsiella pneumoniae
Streptococcus pneumoniae
Enterococcus faecium
Klebsiella pneumoniae
Staphylococcus aureus
Staphylococcus aureus
Staphylococcus aureus
Escherichia coli
Salmonella enterica
Staphylococcus aureus
Enterococcus faecium
Staphylococcus
Staphylococcus aureus
Enterococcus faecalis
Staphylococcus
Methylobacterium
Nocardia
Mycobacterium tuberculosis
Mycobacterium avium
Mycobacterium tuberculosis
Staphylococcus
Nocardia nova
Mycobacterium avium
Staphylococcus aureus
Enterococcus faecium
Enterococcus faecium
Klebsiella pneumoniae
indicates data missing or illegible when filed
For fungal pathogen detection, a clinical gold standard consisting of available culture and 28S-ITS PCR results was used. On average, fungal DNA was at a significantly lower concentration based on nRPM counts than bacterial DNA (p=0.0049) (see
Limits of Detection (LoD) and Linearity. DNA was spiked from a mixture of 4 organisms that were non-pathogenic to humans (Streptococcus uberis, Rhodobacter sphaeroides, Millerozyma farinosa, and Aspergillus oryzae) into healthy donor plasma matrix for LoD evaluation. Samples were spiked in 4-fold dilutions, ranging from 1:1 (no dilution) to 1:4096 dilution, with 4 replicates at each dilution. The LoD for bacterial detection using this assay was estimated to be between 400-700 genome equivalents (GE) per mL for bacteria and 4 GE per mL for fungi (see Table 10). A strong linear correlation between the organism titer (GE/mL) and nRPM values by mNGS was observed (R2=0.89-0.98; see
Streptococcus
uberis
Rhodobacter
sphaeroides
Aspergillus
oryzae
indicates data missing or illegible when filed
Case Series. To assess the potential clinical utility of body fluid mNGS for diagnosis of infection, 12 patients were selectively enrolled with clinically probable or established infection despite negative culture and/or PCR testing of the body fluid (see Table 11). An infectious diagnosis had been made by direct detection from a different body fluid/tissue or by serology/chemistry in 8 and 3 cases respectively. A peritoneal fluid from a patient with bowel perforation and suspected abdominal infection was also included. Presumptive causative pathogens (Klebsiella aerogenes, Aspergillus fumigatus, Streptococcus pneumoniae, Streptococcus pyogenes, Cladophialophora psammophila, Candida parapsilosis, and anaerobic gastrointestinal microbiota) were identified in 7 of 12 cases using mNGS (see Table 11; and Clinical Vignettes in the Examples). Two additional cases of Treponema pallidum (neurosyphilis) and Coccidioides immitis (coccidioidomycosis), diagnosed by serology, had reads detected but present at levels below pre-established nRPM reporting thresholds. Among the remaining 3 cases, mNGS testing was unable to detect Cryptococcus neoformans in pleural fluid (diagnosis made from a culture-positive BAL fluid), Mycobacterium tuberculosis in pleural fluid (diagnosis made from a positive lymph node culture), and Sporothrix sp. in CSF (diagnosis made from serum and CSF IgM antibody positivity), presumably due to a lack of DNA representation from absent or very low pathogen titers and/or high human host background in the body fluid.
Klebsiella aerogenes:
Streptococcus pneumoniae:
Streptococcus pyogenes:
Aspergillus fumigatus:
Coccidioides immitis:
Cladophialophora bantiana:
Cladophialophora bantiana
Cryptococcus neoformans:
Cryptococcus neoformans.
Treponema pallidum:
Treponema pallidum
Mycobacterium tuberculosis:
Candida parasilopsis
Candida parasilopsis:
Candida parasilopsis.
Sporothrix schenkii:
Sporothrix antibody positive.
aBacterial 16S PCR or fungal 28S-ITS PCR
bThe top 5 anaerobes were Faecalibacterium prausnitzii, Eubacterium rectale, Akkermansia muciniphila, Acidaminococcus intestini, and Bifidobacterium adolescentis.
cInfectious diagnosis missed by mNGS testing
Comparison of mNGS with bacterial 16S and fungal 28S-ITS PCR. Out of the 160 patients, the performance of mNGS relative to bacterial 16S PCR or fungal 28S-ITS PCR was compared in 14 cases that had 16S or 28S-ITS PCR testing performed out of 160 (see
Haemophilus
Haemophilus
influenzae
influenzae
Klebsiella
Streptococcus mitis
pneumoniae
Staphylococcus
Staphylococcus aureus
aureus
Streptococcus
Streptococcus
pyogenes
pyogenes (16S PCR)
Streptococcus
Streptococcus
pyogenes
pyogenes (16S PCR)
Mycobacterium
Mycobacterium
tuberculosis
tuberculosis (16S PCR)
Streptococcus
Streptococcus
pyogenes
pyogenes (16S PCR)
Klebsiella
aerogenes
Aspergillus
Aspergillus
fumigatus
fumigatus
Coccidioides
Candida
aall samples in this table had negative testing results by culture, but the pathogen was detected by 16S PCR, mNGS, and/or orthogonal testing, and all were clinically adjudicated if there was a discrepancy between tests.
bdoes not take into account any dilution of DNA extract prior to library preparation.
c
Klebsiella pneumoniae pathogen was detected by mNGS, but not by 16S PCR. To resolve the discrepancy, this sample was also confirmed positive for the pathogen by digital PCR (dPCR) and Sanger sequencing from the original pleural fluid and the contralateral pleural fluid. See FIG. 3B and Clinical Vignettes in the Examples for further details.
indicates data missing or illegible when filed
The first of 3 discordant bacterial cases was a case of an immunocompromised child with necrotizing pneumonia (see case S31 in Table 12, see
In all 3 discordant fungal cases, body fluid mNGS was able to find the causative organism, whereas fungal 28S-ITS PCR testing was negative (see
Comparison of diagnostic yield of mNGS testing from body fluids versus plasma. Seven patients in the study harboring a total of 9 pathogens had paired body fluid and plasma samples available for comparative mNGS testing (See Table 13). Pathogen cfDNA burden based on nRPM was a median 160-fold higher (IQR 34-298) in the local body fluid than in plasma from the same patient (see
Haemophilus
influenzae
Asperigillus
fumigatus
Fusobacterium
Escherichia coli
Escherichia coli
Escherichia coli
Staphylococcus
aureus
adoes not take into account any dilution of DNA extract prior to library preparation.
bdifference in days between plasma and body fluid collection; for example, 4.51 refers to plasma collection done 4.51 days before the body fluid collection
indicates data missing or illegible when filed
Detection of Anaerobic Bacteria and Viruses. Anaerobic bacteria were not included in the accuracy assessment, as anaerobic culture was not always performed and cultured anaerobes were typically not speciated. However, the one sample in the accuracy study that was culture-positive for an anaerobic bacterium (Finegoldia magna from a soft tissue abscess, Case S87 of Table 13) was successfully detected by mNGS testing (See Table 14).
Rothia dentocariosa
Peptoclostriudium difficile
Peptoclostriudium difficile
Prevotella denticola
Campylobacter curvus
Veillonella parvula
Lactococcus lactis
Geobacillus sp. WCH70
Parvimonas micra
Bacillus coagulans
Campylobacter concisus
Bifidobacterium breve
Lactobacillus gasseri
Bacteroides xylanisolvens
Lactococcus lactis
Prevotella melaninogenica
Prevotella melaninogenica
Veillonella parvula
Fusobacterium nucleatum
Fingoidia magna
Streptococcus
Peptoclostriudium difficile
Veillonella parvula
Rothia
Veillonella parvula
indicates data missing or illegible when filed
DNA viruses were also excluded in the accuracy assessment due to lack of routine clinical testing for viruses. Applying previously validated clinical mNGS thresholds of 3 non-overlapping reads for viral detection viruses were detected from the Anelloviridae (n=5), Herpesviridae (n=9), and Adenoviridae (n=2) families (See Table 15). Four of the 5 (80%) anellovirus detections were from immunocompromised patients, consistent with the reported association of anelloviruses as non-pathogenic markers of active inflammation in this population. Among the 11 remaining viruses detected by body fluid mNGS, 6 of 6 (100%) were orthogonally confirmed as true-positive cases by virus-specific PCR.
Detected: low
and IgA deficiency who
copies/mL
indicates data missing or illegible when filed
Clinical Vignettes. The first set of clinical vignettes comprised 5 cases where both culture and 16S PCR was negative, but a clinical diagnosis was made through other means (also shown in Table 11). The cases were sourced from a combination of physician referral and positive microbiological results.
The second set of clinical vignettes is comprised of 7 cases from the accuracy study where mNGS was able to find incidental new bacteria and fungi that were not known at the time of testing. In each case, follow-up orthogonal testing using 16S/ITS PCR or digital PCR was performed and clinical adjudication after mNGS was able to subsequently confirm the new organism.
Set 1: Prospective Case Series of Body Fluid mNGS Testing in Patients with Probable Infection but Negative Clinical Microbiological Testing
Case S88
CSF (2 days prior to surgical removal of the implant):
Deep brain stimulator implant material removed from the brain:
Clinical adjudication: Klebsiella aerogenes
Case S88 is a man in his 70s with a background of Parkinson's disease, deep brain stimulator (DBS) placement, and mechanical aortic valve replacement on warfarin. The DBS was placed 3 years prior to admission and the electrode was repositioned 9 months prior to admission. The patient was admitted for fever and reduced consciousness with a history of recent traumatic head injury and a scalp wound. He was treated for meningitis with empirical vancomycin, ceftriaxone, and ampicillin, with clinical improvement after six days of treatment. A prompt lumbar puncture was not possible due to the anticoagulation, but this was performed four days into antibiotic treatment. CSF bacterial culture and 16S rDNA PCR were both negative at the time. Fourteen days after stopping antibiotic treatment, the patient was readmitted to the hospital for reduced consciousness.
As fever was noted, meningeal doses of vancomycin, cefepime, and ampicillin were commenced. Once again, a lumbar puncture could not be immediately performed due to anticoagulation. The scalp wound he previously sustained was noted to be close to the DBS lead. A brain CT with contrast showed streak artifact associated with DBS leads, but no acute intracranial pathology.
The CSF was hazy macroscopically, with a high WBC of 760×106/L (63% lymphocytes, 11% lymphocytes, 25% monocytes/histiocytes, 1% basophils), RBC 28×106/L, protein 58 mg/dL, glucose 48 mg/dL (corresponding serum glucose 75 mg/dL). CSF culture, HSV/VZV PCR, and 16S rDNA PCR were all negative. Three days after admission, the DBS was removed surgically, and bacterial culture of the prosthetic material was positive for Klebsiella aerogenes. The patient had complete resolution of the infection and a good clinical outcome.
At this point, a CSF sampled 2 days before the surgery removal of the infected hardware was retrospectively enrolled. CSF testing by mNGS was positive for Klebsiella aerogenes, which was further confirmed by digital PCR of the sequencing library (see
Case S89
Retrouterine fluid:
Clinical adjudication: predominately anaerobic GI flora
A woman in her 20s with inflammatory bowel disease and past GI surgery presents with free air seen on an abdominal CT and was confirmed to have small bowel perforation during corrective surgery. One week after her operation, the patient continued to have leukocytosis and a CT scan showed a rectouterine fluid collection that was drained the next day. The rectouterine fluid was visually purulent (cloudy) and thick, but was negative on culture, including anaerobic culture. Testing by mNGS of this fluid drainage shows multiple, anaerobic, gastrointestinal bacteria: Faecalibacterium prausnitzii, Eubacterium rectale, Akkermansia muciniphila, Acidaminococcus intestini, and Bifidobacterium adolescentis as the top 5 most common organisms, whereas anaerobic culture results were negative. The primary surgical team started empiric antibiotics (piperacillin/tazobactam only) after drainage. CT imaging showed persisting rectouterine fluid collection a few days later, although slightly decreased and the patient's elevated white count normalized. Antibiotics were discontinued after nearly a week and the patient was discharged. In this case, mNGS could have suggested the addition of metronidazole to cover the undocumented anaerobic organisms.
Case S92
Bronchoalveolar lavage (BAL) fluid:
Clinical adjudication: Aspergillus fumigatus (Probable by EORTC/MGS* international standards)
A man in his 60s with anaplastic large cell lymphoma and bladder cancer was admitted electively for chemotherapy. His clinical course was complicated by MRSA bacteremia and endocarditis from a PICC line source (treated with vancomycin) and ischemic bowel requiring primary resection and anastomosis.
A CT chest (without contrast) performed for persistent fevers and streaky opacities on CXR revealed multiple bilateral pulmonary nodules, nodular areas of consolidation, and a left pleural effusion, with unchanged supraclavicular, mediastinal, and hilar lymphadenopathy. Serum beta-D-glucan was raised at 316 picograms/mL (reference range <60) and serum aspergillus galactomannan index was raised at 4.501 (reference range <0.5).
The patient met the EORTC/MGS* criteria for probable invasive aspergillosis, and voriconazole was commenced.
*European Organization for Research and Treatment of Cancer
BAL and FNA of a pulmonary nodule were collected 3 days into voriconazole treatment. BAL Gram stain and cultures were negative. FNA revealed malignant lymphoma cells on cytology, consistent with the patient's known lymphoma, but also negative cultures.
At this point, the BAL sample was included in this series given that the patient was culture-negative but had a clinically probable invasive Aspergillus infection. mNGS of the BAL demonstrated the presence of Aspergillus fumigatus.
Voriconazole was changed to posaconazole after 8 days due to liver toxicity concerns. Follow-up of serum galactomannan index demonstrated a treatment response (falling to 0.24 mg/mL at 15 weeks of posaconazole treatment) and follow-up CT scans showed a continued decrease in size of the multiple pulmonary nodules representing resolving infection. The patient was subsequently discharged from the hospital with hematology and infectious diseases clinic follow-up.
Case S90
Blood culture: Streptococcus pneumoniae
Pleural fluid:
Clinical adjudication: Streptococcus pneumoniae
A woman in her 50s, with a history of a hematopoietic stem cell transplant 1 year ago, presented with fever, productive cough, and tachycardia, and was subsequently found to be blood culture positive for Streptococcus pneumoniae. CT imaging showed a left loculated pleural effusion. The effusion was drained, but the culture results of the pleural fluid were negative. The sample was enrolled into this study and mNGS results of the same pleural fluid showed Streptococcus pneumoniae as the top species at a normalized RPM of 57,856 and were more than 99.99% of all the microbial reads not classified into the same family or genus.
Case S91
Blood culture: Group A Streptococcus
Pleural fluid:
Clinical adjudication: Streptococcus pyogenes
A woman in her 50s who presented with fever, malaise, and syncope in the setting of sepsis, was admitted with Group A Streptococcus bacteremia and pneumonia. She placed on ceftriaxone and quickly improved. She developed a complicated parapneumonic effusion (LDH (lactate dehydrogenase)>2700 U/L). The effusion was drained, but the culture of the pleural fluid was negative. The sample was enrolled into this study and mNGS of the same pleural fluid showed that Streptococcus pyogenes was the top species identified at a normalized RPM of 57,856 and were more than 99.99% of all the microbial reads not classified into the same family or genus.
Set 2: Additional Pathogens Incidentally Detected by Body Fluid mNGS Testing in Patients with Microbiologically Proven Infection
Case S31
Pleural fluid:
Clinical adjudication: Klebsiella pneumoniae
A child with congenital CMV and myelodysplastic syndrome was admitted for chemotherapy. He developed febrile neutropenia with septic shock and coagulopathy. Despite empirical cefepime, the sepsis worsened with the development of ARDS and worsening abdominal distension, leading to an intensive care admission. His antibiotics were changed empirically to meropenem, ciprofloxacin, and vancomycin. Caspofungin was also initiated for antifungal cover.
CT imaging revealed necrotizing pneumonia involving all lobes of both lungs and moderate bilateral pleural effusions. Asymmetric enhancement of the small intestine may have indicated bowel inflammation/infection or septic shock physiology.
Blood, BAL, and pleural fluid were all negative on bacterial culture. The pleural fluid was exudative by Light's criteria. 16S rDNA PCR of the pleural fluid was positive for Streptococcus mitis group, with no other organisms detected.
Despite the specific 16S rDNA PCR result, a decision was made to continue the broad-spectrum antibiotic combination, including meropenem, to cover the range of possible organisms contributing to the necrotizing pneumonia. The patient improved clinically with no signs of sepsis after treatment. A chest CT verified resolution of the infection.
Pleural fluid mNGS by both Illumina and Nanopore sequencing showed Klebsiella pneumoniae. This was subsequently confirmed by digital PCR of both the sequencing library and the original DNA extract and Sanger sequencing of the DNA extract (see
Case S65
Ascitic fluid:
A previously healthy woman in her 30s presented to the hospital with diffuse abdominal pain, nausea, vomiting, watery diarrhea, fever, leukocytosis, and acute kidney injury four days after IUD placement. CT abdomen and pelvis demonstrated inflammation of the caecum, sigmoid colon, and rectum, with peritoneal enhancement and intra-abdominal ascites. Chlamydia and gonorrhea NAAT testing were negative.
A percutaneous drain was inserted five days into antibiotic treatment with piperacillin-tazobactam. The ascitic fluid showed WBC 14.375×109/L (74% neutrophils, 5% lymphocytes, 21% others), a high total protein of 3.8 g/dL, and a serum albumin albumin gradient (SAAG) of 0.4 g/dL, consistent with infected ascites. Direct microbiological cultures of the ascitic fluid, however, yielded no growth.
This case was referred by a hospitalist physician and mNGS on the same ascitic fluid was positive for Streptococcus pyogenes. 16S PCR of a later ascitic fluid was previously sent and it was negative. The same ascitic fluid was then sent for 16S PCR that underwent mNGS and the 16S test was also positive for Streptococcus pyogenes. The patient was treated with 14 days of piperacillin-tazobactam, along with percutaneous drainage of subsequent loculated collections. She clinically improved was discharged from the hospital 20 days after admission.
Case S10
A woman in her 30s who was a smoker and previous intravenous drug user presented with one week of productive cough and dyspnea. She developed type 1 respiratory failure and cardiogenic shock, requiring intubation, ventilation, and ECMO support. She was then transferred from the outside hospital to a tertiary care hospital. CT chest revealed bilateral upper lobe consolidation with patchy regions of nodular consolidation throughout the remaining lung fields, with diffuse mediastinal and hilar lymphadenopathy. The blood cultures and BAL cultures collected three days after the initiation of antibiotics at the outside hospital were all negative.
This case was referred by an infectious disease specialist and mNGS of the BAL fluid was positive for Haemophilus influenzae. One of 2 blood cultures prior to antibiotics at an outside hospital was also positive for Haemophilus influenzae. Subsequent 16S rDNA PCR was positive for Haemophilus influenzae.
The patient was commenced empirically on ceftriaxone, vancomycin, and azithromycin. These were subsequently changed to ceftriaxone monotherapy based on Haemophilus influenzae sensitivity results. Her workup revealed a new diagnosis of B-cell acute lymphoblastic leukemia. The patient improved clinically, completed induction chemotherapy, and has been disease-free for over a year.
Case S42
A woman in her 30s with a history of chronic endometriosis and laparotomy for ruptured appendicitis and tubo-ovarian abscess two months prior to admission, was readmitted for severe lower abdominal pain, vaginal bleeding, nausea, low-grade fevers, and chills. CT abdomen and pelvis showed multiple loculated abdominal and pelvic abscesses (the largest measuring 9×7×17 cm) interspersed between bowel loops and mesentery—these could not be surgically drained due to the dense adhesions and bowel loops surrounding the fluid collections. Piperacillin-tazobactam was commenced empirically and an abdominal drain was placed into the large abscess. Both blood cultures from admission and abscess fluid cultures grew pan-sensitive Escherichia coli only.
Plasma and abscess fluid mNGS both showed DNA reads to Fusobacterium nucleatum and Escherichia coli. Follow-up CT two weeks later showed resolution of multiple abscesses, with minimal residual collections remaining.
Fusobacterium nucleatum is an anaerobe commonly found in polymicrobial intra-abdominal abscesses. This was detected by mNGS and was not detected by conventional bacterial culture.
Case S64
AV graft tissue culture: negative
Peri-graft swab:
A woman in her 50s presented with fever and tenderness in the area over a Polytetrafluoroethylene (PTFE) arteriovenous graft. She had a background of end-stage renal failure with a renal transplantation two months prior to admission. Intra-operative findings during graft excision revealed that the graft was completely thrombosed, with surrounding purulent fluid and extension of the infection along the graft to disrupt the arterial anastomosis.
AV graft tissue cultures were negative, but a peri-graft swab grew pinpoint colonies of gram-negative rods after 6 days. Identification of the colonies was difficult as MALDI-ToF (matrix-associated laser desorption/ionization—time of flight) and biochemical testing were inconclusive. Send-out 16S sequencing eventually identified the colonies as Mycoplasma hominis 16 additional days later. mNGS from the original pen-graft swab (available on day 0) was also positive for Mycoplasma hominis. Nanopore real-time sequencing took less than 10 minutes for organism identification after the initiation of sequencing.
The patient was discharged back to her referring hospital before final culture results were available, as the Mycoplasma hominis required 16S PCR for identification. This is a case where an earlier result (such as by mNGS) would have had an impact on clinical management as vancomycin is an ineffective treatment for Mycoplasma hominis.
It will be understood that various modifications may be made without departing from the spirit and scope of this disclosure. Accordingly, other embodiments are within the scope of the following claims.
This application is a U.S. National Phase Application filed under 35 U.S.C. § 317 and claims priority to International Application No. PCT/US2021/051924, filed Sep. 24, 2021, which application claims priority under 35 U.S.C. § 119 from Provisional Application Ser. No. 63/083,868 filed Sep. 26, 2020, the disclosures of which are incorporated herein by reference.
This invention was made with government support under grants HL105704, R21 AI120977 and R33 AI129077, awarded by The National Institutes of Health. The Government has certain rights in the invention.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2021/051924 | 9/24/2021 | WO |
Number | Date | Country | |
---|---|---|---|
63083868 | Sep 2020 | US |