The application generally relates to digital x-ray imaging methods/system, and more specifically, to methods and/or systems for acquiring multiple image information of an object (e.g., medical radiographic imaging) using a grating-based differential phase contrast imaging technique with a slot-scanning configuration.
Conventional medical x-ray imaging devices are based on the attenuation through photoelectric absorption of the x-rays penetrating the object to be imaged. However, for soft tissues including vessels, cartilages, lungs, and breast tissues with little absorption, this provides poor contrast compared with bone images. This problem of low contrast in soft tissues can be addressed with phase contrast imaging (PCI) techniques.
The principle of PCI is based on the wave nature of x-rays, where refraction and diffraction properties need to be considered. As an electromagnetic wave, the x-ray is usually characterized by its frequency, amplitude, and phase. When an electromagnetic wave penetrates a medium, its amplitude is attenuated and its phase is shifted. In x-ray technology, the refractive index n of a material can be expressed by a complex number
n=1−δ+iβ (1)
The imaginary part β contributes to the attenuation of the amplitude and the real part δ is responsible for the phase shift. It has been shown that δ is about 103 to 104 times larger than β. But in conventional medical imaging, only the information of β is recorded while the information of δ is completely lost. In recent years, several PCI techniques have been explored to make use of the phase shift to form the image, which is expected to provide more information about the object. These include (i) the interferometer technique, (ii) the diffraction-enhanced imaging (DEI) technique, and (iii) the free-space propagation technique.
However, there are various practical problems associated with all three techniques such as efficiency and limited field of view. In the case of perfect crystal interferometers and crystal diffractometers, high temporal coherence (i.e., a high degree of monochromaticity) is required; as a result, only a synchrotron or a well-defined wavelength of the whole spectrum from a radiation source is used. A synchrotron radiation source is costly and incompatible with a typical clinical environment. Both techniques are also limited by the accepted beam divergence of only a very small angle (a few mrad) due to the use of crystal optics. The free-space propagation technique is limited in efficiency since it requires high spatial coherence, which can only be obtained from an x-ray source with a very small focal spot. The three PCI techniques differ greatly in the way the image is recorded, the instrumental setup, and the requirements on the radiation source (especially its spatial and temporal coherence). Although some of the techniques yield excellent results for specific applications, none is very widely used and none has so far found application in medical diagnostics.
The grating-based PCI method with a standard x-ray tube is limited by the loss of visibility of the interference fringes at the detector due to the broad spectrum of the x-ray tube. A standard polychromatic x-ray tube generates soft x-rays (<15 keV) that barely penetrate the skin at the low-energy portion of the spectrum, as well as hard x-rays (>50 keV) that penetrate through both bones and tissues at the high-energy portion of the spectrum. The use of an energy filter is thus preferred to obtain a narrow-bandwidth x-ray beam to reduce the radiation dose significantly by eliminating the unnecessary soft and hard x-rays and increase the clearness of the image.
For applications requiring a large FOV, a large-size phase grating G1 and analyzer grating G2 are needed. For example, a typical mammogram has a size of 24 cm×30 cm. This means that a phase grating and an analyzer grating having the same size are required. Given the limitations of the current grating fabrication techniques (e.g., silicon wafer size, structure height, and grating uniformity), the manufacturing cost of such large gratings will be extremely high.
For a grating-based PCI system having a divergent cone beam (or fan beam) geometry and a large FOV, the phase contrast image quality is generally inferior in the edge regions of the detector. Toward the edges of plane gratings, the angle subtended by the grating bars with the incoming x-ray beam becomes larger. Since the bar height of the phase and analyzer gratings increase approximately linearly with the x-ray energy (E), the aspect ratio of bar height to gap width would be very large (>10:1 for E>20 keV). As a result, these gratings would cause a shadowing effect of the phase grating and the scan effect of the analyzer grating at larger angles, degrading the image quality.
In all x-ray imaging systems, scattered radiation from the object has been shown to degrade the image quality in terms of subject contrast and contrast-to-noise ratio significantly. Currently, anti-scatter grid is the most widely used device for scatter rejection with most radiography and mammography systems. In mammography, with anti-scatter grid the amount of scattered radiation measured by the scatter-to-primary ratio can be reduced to between 0.1 and 0.3 from about 0.25 to 1.2. However, intrinsic to the anti-scatter grid method is the attenuation of a significant fraction of the primary x-rays.
An aspect of this application is to advance the art of medical radiographic imaging.
Another aspect of this application to address in whole or in part, at least the foregoing and other deficiencies in the related art.
It is another aspect of this application to provide in whole or in part, at least the advantages described herein.
Another aspect of the application is to provide methods and/or apparatus embodiments for digital radiographic medical imaging. Another aspect of the application is to provide methods and/or apparatus embodiments for mammographic medical imaging. Another aspect of the application is to provide methods and/or apparatus embodiments for slot-scanning phase contrast imaging for large field of view (FOV) (e.g., greater than 100 mm square) radiographic medical imaging.
In accordance with one embodiment, the invention can provide a slot-scanning phase-contrast digital mammography system that can include a polychromatic x-ray source for mammography imaging; a beam shaping assembly including a collimator, a source grating, an x-ray grating interferometer including a phase grating and an analyzer grating; and an area x-ray detector; wherein the three gratings are positioned so that the plane and the grating bars of these gratings are aligned to each other.
In accordance with one embodiment, the invention can provide a phase-contrast digital radiographic imaging system that can include a radiation source for imaging, a beam shaping assembly including a collimator and a source grating G0, an x-ray grating interferometer including a phase grating G1 and an analyzer grating G2, and an area x-ray detector, where a pitch and a position of the analyzer grating G2 relative to a pitch of an interference pattern produced by the phase grating G1 produce at least one fringe pattern over a width of the analyzer grating G2.
In accordance with one embodiment, the invention can provide a method that can include providing a beam shaping assembly comprising a beam limiting apparatus and a source grating G0, providing an x-ray grating interferometer comprising a phase grating G1, and an analyzer grating G2, and offsetting a pitch of the analyzer grating G2 relative to a pitch of an interference pattern produced by the phase grating G1 at a prescribed distance from the phase grating G1.
These objects are given only by way of illustrative example, and such objects may be exemplary of one or more embodiments of the invention. Other desirable objectives and advantages inherently achieved by the disclosed invention may occur or become apparent to those skilled in the art. The invention is defined by the appended claims.
The foregoing and other objects, features, and advantages of the invention will be apparent from the following more particular description of the embodiments of the invention, as illustrated in the accompanying drawings. The elements of the drawings are not necessarily to scale relative to each other.
The following is a detailed description of exemplary embodiments according to the application, reference being made to the drawings in which the same reference numerals identify the same elements of structure in each of the several figures.
To be useful for clinical imaging, the phase contrast imaging systems must meet various requirements including: (i) use of a standard broadband x-ray source; (ii) a large field of view (FOV) of many centimeters (e.g., 24 cm×30 cm for a typical mammography system); (iii) a reasonably compact design comparable to current radiographic imaging systems (e.g., the source-to-detector distance is about 65 cm for a typical mammography system); and/or (iv) a reasonable exposure time and dose (e.g., the mean exposure for a typical mammography system is about 5 mR).
1. System Configuration
As shown in
2. System Components
(a) X-Ray Source
As shown in
(b) Filter and Monochromator
Beside inherent filtration associated with the x-ray tube 110, additional filtration (e.g., by the filter B) can be optionally used to spectrally shape the x-ray beam into a narrow-bandwidth beam to reduce or eliminate the unnecessary soft x-rays that are mostly absorbed by the patient and contribute to the radiation dose received during the examination, and/or the hard x-rays that can reduce the quality of the image. Exemplary typical filter materials are aluminum (Al), molybdenum (Mo), rhodium (Rh), silver (Ag), and other metals.
Alternatively, the filter B can be a tunable monochromatic x-ray filter that can be used with a divergent polychromatic x-ray source to produce monochromatic x-rays with a narrow spectrum centered at a selectable energy with a bandwidth of 1-3 keV.
(c) Gratings
As shown in
The source grating G0 can allow the use of a large incoherent x-ray source as the x-ray source 110 because the source grating G0 can create an array of individual line sources that each can provide sufficient spatial coherence for the interferometric contrast. The images created by the source grating G0 generated line sources can be superimposed congruently in the detector plane at the detector 140 leading to a gain in intensity (e.g., controllable interference).
The phase grating G1 can operate as a beam splitter and divide the incoming beam essentially into the ±1 diffraction orders. These two ±1 diffracted beams can interfere and form a periodic interference pattern in the plane of the second grating G2 through the Talbot self-imaging effect. When an object is inserted in the x-ray beam path, the position of the fringe pattern would change. As the change of the fringe position in the micron range is not determined with a common detector, an analyzer second grating G2 can be placed at a specific Talbot distance from the phase first gating G1 to enable the transform of fringe positions into intensity modulations on the detector 140 located directly behind the second grating G2 with the phase stepping technique.
As the source grating G0 is disposed close to the x-ray source 110 and the collimator C, the size the source grating G0 can be small (e.g., about 1 cm×0.5 cm) because of the small angle subtended by the x-ray fan. For an exemplary (e.g., mammography) application, the FOV can be 24 cm×30 cm. Since the object is located close to the interferometer formed by gratings G1 and G2, the size of these gratings should match the FOV. Given the state of art for standard photolithography techniques, repeatable fabrications of such large-area gratings G1 and G2 (e.g., 24 cm×30 cm) with high or sufficient yield and an acceptable uniformity are not trivial. To address this fabrication problem, a standard 6 or 8 inch-silicon wafer can be used to fabricate multiple small gratings (e.g., each with an area of 8 cm×1 cm) within a square of 8 cm×8 cm. By abutting three pieces of small gratings together, a long and narrow grating (e.g., 24 cm×1 cm) can be repeatedly obtained with acceptable uniformity.
The source grating G0 allows the use of large incoherent x-ray sources since it creates an array of individual line sources each providing enough spatial coherence for the interferometric contrast. The images created by each line source are superimposed congruently in the detector plane leading to a gain in intensity. The phase grating G1 acts as a beam splitter and divides the incoming beam essentially into two first diffraction orders that interfere and form periodic fringe patterns in planes perpendicular to the optical axis (z). Based on the Talbot effect, the periodic fringe pattern, which is called the self image of the phase grating G1, will have its highest contrast at the first Talbot distance d1 behind G1. Assuming that the phase shift undergone by x-rays passing through the grating bars of G1 is π, the first Talbot distance is given by
where p1 is the period of G1 and λ is the wavelength of x-ray for plane waves. The period of the fringe pattern (p2) at the plane of the analyzer grating G2 placed at a distance of d1 from G1 is approximately half the period of G1. The analyzer grating G2 has approximately the same period of the fringe pattern (p2).
When an object is placed in the beam path, the incoming x-ray wavefront can be locally distorted by the object. Where the wavefront is distorted, the fringes formed by the phase grating G1 are displaced from their unperturbed positions. The fringe displacements are transformed into intensity variations by the analyzer grating G2 placed at a distance d1 from the phase grating G1. This allows the use of an x-ray detector placed just behind the analyzer grating G2 with much larger pixels than the spacing of the fringes. Using the phase stepping technique, scanning the lateral position xg of one of the gratings over one period of the grating (here the analyzer grating G2) causes the recorded signal in each pixel to oscillate as a function of xg as shown in
(d) Detector
For the detector 140, either an indirect or a direct flat-panel x-ray detector can be used. An indirect flat panel detector can include a layer of scintillator made of CsI, Gd2O2S, or other scintillating phosphors coupled with an array of photodiodes (e.g., a-Si photodiodes) and switches (e.g., thin-film transistor (TFT) switches). The thickness of the scintillator layer can be between 80 um and 600 um. The pixel pitch of the detector is ranged from 20 to 200 um. On the other hand, a direct detector can include a photoconductor such as amorphous selenium (a-Se) or PbI2 to produce electrical charges on the detection of an x-ray. The electromagnetic radiation detection process is considered direct because the image information is transferred from x-rays directly to electrical charges with no intermediate stage.
As an alternative to the flat-panel detectors, a charge-coupled device (CCD) based x-ray detector can be used as the detector 140. For example, the CCD based x-ray detector can include a scintillating screen.
For a slot-scanning system, a tiled CCD detector array operated in time delay integration (TDI) mode is preferred to enable continuous scanning motion and x-ray illumination during each scan. The detector array can be formed by tiling two or more CCD devices together and can be coupled to a scintillator layer and a fiber optic plate (FOP). The FOP is used to protect the CCD array from radiation damage.
A slot-scanning system with a beam width comparable to the pixel width would require an extremely high tube output. The TDI operating mode of the CCD can allow a significantly wider beam to be used. The detected x-rays are first transformed into light photons via the scintillator layer. The light photons are then transmitted to the CCD through the FOP producing electrons in the CCD in response to the light emission from the scintillator upon x-ray absorption. By moving the electronic charges from pixel-to-pixel across the CCD width (e.g., columns), in synchrony with (e.g., at the same velocity) but in the opposite direction of the scanning motion, the TDI mode can enable x-ray integration across the CCD width while maintaining the pixel resolution. When the charges reach the last row of the CCD, the accumulated charge is read out and digitized. For example, the detector array can include four CCDs, each having a size of 6 cm×1 cm, abutted along their narrow dimension to form a long and narrow detector (e.g., 24 cm×1 cm). Again, the typical pixel size is between 20 um and 200 um.
As another alternative to the flat-panel detectors, a linear photon counting gaseous detector using avalanche amplification process can be also used as the detector 140. Besides the use of gaseous detectors in photon counting technique, crystalline Si, CdTe, and CdZnTe can also be used in direct-conversion photon-counting detectors.
This exemplary single photon counting detection technique can discriminate noise in the detector 140 from a true x-ray photon interaction. By counting signals above a predefined threshold, an electronic noise free and efficient counting of single x-ray photons is achieved. When this detector type is used in a slot-scanning system according to embodiments of the application, significant reduction of patient dose and scattered radiation and/or a considerable increase in image quality in terms of contrast and spatial resolution can be obtained, as compared to the integrating detectors (such as direct and indirect flat-panel detectors and CCD devices).
3. Selection of System and Grating Parameters
Selections of grating parameters and the geometric system parameters in exemplary embodiments can be restricted by the choice of x-ray source, the limitation of the grating fabrication process, the practicality of the system size, the system performance requirements, and the conformation of physical laws. In summary, for a spherical x-ray wave, the system parameters and grating parameters should satisfy the following equations.
1. Spatial Coherence Requirement
2. Period of Gratings
3. Phase Grating Requirement
The structure height of the silicon phase grating G1 has to be such that the x-rays passing through the grating bars undergo a prescribed phase shift or a phase shift of π (as an example), which results in the splitting of the beam into the ±1 diffraction orders.
Also, the structure height of gratings G0 and G2 should be large enough to provide sufficient absorption of x-ray (e.g., >75%) for selected or optimum system performance.
4. Talbot Self-Imaging Condition
The parameters shown in Eqs. (3)-(7) are as follows.
lc=coherence length=
λ=mean wavelength of x-ray radiation
L=distance between G0 and G1
s=slit width of G0
n=integer (Talbot order)
dn=Talbot distance between G1 and G2
p0=period of G0
p1=period of G1
p2=period of G2
h0=structure height of G0
h1=structure height of G1
h2=structure height of G2
δSi=refractive index decrement of silicon
By first selecting n, p2, λ, and L based on system requirements and limitations on grating fabrication, other parameters, namely, s, p0, p1, h1, h2, h3, and dn can then be determined. As an example, Table 1 lists exemplary system design parameters and grating parameters for an embodiment of a slot-scanning phase-contrast digital mammography system.
4. Exemplary System Operations
As shown in
When the determination in block 850 is affirmative because a predetermined number of cycles N (e.g., typically 5 to 8) of scanning and stepping are completed, the image data set can be extracted, processed, and displayed on a monitor (operation blocks 870, 880, 890). For example, the same image data set can be processed by an image processing unit of the computer to construct multiple images of the object including absorption contrast, differential phase contrast, phase shift contrast, and dark-field images, as described herein.
These absorption contrast, differential phase contrast, phase shift contrast, and dark-field images are complementary to each other can provide the necessary specificity to visualize subtle details in the object.
There are alternate ways to implement the phase stepping described in the method embodiment of
As shown in
Then, the swing arm 160 is stepped to a current step position (operation block 933), the x-ray beam is fired to expose and capture an image of a portion of the object (operation block 940). Then, it is determined whether the image series is complete for that step (e.g., N images have been captured) in operation block 945. When the determination in block 945 is negative, using the phase stepping technique, as an example, the analyzer grating G2 (e.g., mounted on a piezo translation stage) is then moved laterally by a predetermined distance (e.g., p2/N such as 2 mm/8=250 nm) and the process jumps back to block 940 where the x-ray beam is fired to expose and capture an image of a portion of the object.
When the determination in block 945 is affirmative because a predetermined number of cycles N (e.g., typically 5 to 8) of stepping and scanning are completed, the image data set can be stored and it can be determined in operation block 955 whether scanning is complete for the whole object. When the determination in block 955 is negative, the swing arm 160 is stepped to the next position (operation block 933) and operation blocks 940, 945 and 950 can be repeated. When the determination in block 955 is affirmative because the whole object has been scanned, the image data set can be extracted, processed, and displayed on a monitor (operation blocks 960, 965, 970). For example, the same image data set can be processed by an image processing unit of the computer to construct multiple images of the object including absorption contrast, differential phase contrast, phase shift contrast, and dark-field images, as described herein.
5. Image Formation and Image Retrieval
Without the object in place, the x-ray beam passes through the phase grating G1 and form interference fringes. Having the object in the beam path, the incoming x-ray wavefront is locally distorted by the object causing an angular deviation of the x-ray beam:
Where the wavefront is distorted, these fringes are displaced from their unperturbed position by
D(x,y)=dn·α(x,y) (9)
The fringe displacements are transformed into intensity values by an analyzer grating G2 placed at a distance dn from the phase grating G1. A two-dimensional detector with much larger pixels than the spacing of the fringes can be used to record the signal. Scanning the lateral position xg of one of the gratings (e.g., the analyzer grating G2) causes the recorded signal in each pixel to oscillate as a function of xg. For each pixel (i, j), the signal oscillation curve can be expressed by a Fourier series,
From Eqs. (10) and (11), the following images of the object can be retrieved. The transmission image is given by
The differential phase contrast image is given by
Also, the phase shift image of the object can be obtained by simple one-dimensional integration along the pixel direction perpendicular to the grating bars, e.g.,
Furthermore, a dark-field image is formed from higher-angle diffraction intensities scattered by the object. The information about the scattering power of the object is contained in the first Fourier amplitude coefficient, bs(i, j) of Is(i, j, xg). Thus, the dark-field image can be obtained by
These four different images of the object can be derived from the same data set and can be complementary to each other to provide multiple information of the object enabling the visualization of subtle details in the object.
As described herein, embodiments of phase-contrast digital imaging systems and/or methods of using the same can provide various advantages according to the application. Embodiments of a hybrid slot-scanning grating-based differential phase contrast mammography system have various advantages (e.g., compared to a full-field digital mammography system).
Embodiments of a grating-based differential phase contrast imaging technique can use conventional x-ray tubes instead of expensive and huge synchrotron radiation sources to provide multiple image information (e.g., absorption contrast image, differential phase contrast image, phase shift image, and dark-field image) of the object from a single image capture process.
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can significantly enhance the contrast of low absorbing tissues (e.g., the contrast between healthy and diseased tissues), which can be especially useful for mammography and orthopedic joints.
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can allow the use of small gratings and detectors to produce a large-area image. Embodiments can provide reduction in motion blur, scattered radiation, and patient dose without using a grid.
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can use a phase grating (G1) and an analyzer grating (G2) with a long and narrow geometry that can be formed by abutting two or more short and narrow (e.g., 8 cm×1 cm) gratings together and will cost significantly lower than the ones with a large full-field size (24 cm×30 cm for typical mammography). Thus, embodiments of a tiled detector can be made and will cost much less than a large full-field two-dimensional detector (e.g., 24 cm×30 cm for typical mammography).
Embodiments of an imaging system can require a long and narrow detector, which can be formed by abutting two or more short and narrow (e.g., 8 cm×1 cm) detectors together. Smaller detectors with high sensitivity and low noise are commercially available at low cost relative to a large full-field two-dimensional detector (24 cm×30 cm for typical mammography).
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can use curved gratings and detectors circularly around the source focus to enable the design of a more compact system and reduce or eliminate the shadowing effect of the phase grating and/or the scan effect of the analyzer grating occurred in the edge regions of the image.
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can use an x-ray tube with rotating-anode (higher output), a short distance between the x-ray source and the object (higher x-ray fluence), and a detector with a CsI scintillator coupled with a tiled TDI-mode CCD array (higher detection sensitivity). As a result, the exposure time can be significantly reduced.
Certain exemplary embodiments for slot-scanning phase-contrast digital imaging systems and/or methods for using the same, e.g., see
Both exemplary scanning embodiments described in
To implement continuous motion of the swing arm with fixed G1 and G2 gratings, exemplary embodiments of phase contrast imaging systems have to be detuned. In one exemplary embodiment, a detuned phase contrast imaging system can be understood to be an imaging system in which the pitch p2 of the analyzer grating G2 is purposely controlled or fabricated to be unequal to a period of interference pattern pint at a Talbot distance behind the phase grating G1. In another exemplary embodiment, a detuned phase contrast imaging system can be understood to be an imaging system in which the pitch p2 of the analyzer grating G2 is controlled or fabricated to be equal to a period of interference pattern pint at a Talbot distance behind the phase grating G1, but the analyzer grating G2 is positioned away from the corresponding Talbot distance. In certain exemplary embodiment, a detuned phase contrast imaging system can generate a periodic fringe pattern, where the fringe pattern occurs over a width or a portion of the width of the analyzer gating G2. Although a number of exposures for detuned grating based PCI system embodiments in a complete or partial scan of an object is about the same, positional errors and/or scanning time can be reduced relative to a tuned grating based PCI systems.
I
s
=MTF(f)·[cos(2πfintx)·cos(2πf2x)]=MTF(f)·[cos(2π(fint+f2)x)+cos(2π(fint−f2)x]/2 (16)
For example, MTF is a detector's modulation transfer function that can be approximated by: MTF(f)=0.5·erfc[α ln(f/f0)], where α is a slope of the MTF curve and f0 is the spatial frequency at which MTF drops by 50%. The spatial frequency at p2=2 um pitch of the analyzer grating is 500 cyc/mm. When summed with comparable frequency of interference pattern, it doubles, e.g., fint+f2=1000 cyc/mm. Exemplary values of f0 in indirect charge integrating detectors can be typically between 1 and 2 cyc/mm, while value of f0 can reach 5 cyc/mm in the case of direct photon counting detectors. That said, the detector will measure no signal at 1000 cyc/mm. Therefore, the only detectable signal is:
MTF(f)·cos(2π(fint−f2)x)/2 (17)
In the case of a tuned phase contrast imaging system (fint=f2), the signal is increased or maximum. When measuring the open field in such configuration, the detector yields the uniform image. In the case of detuned phase contrast imaging system, the detected image will have a cosine pattern with a lower contrast caused by detector's MTF. The loss of the contrast depends on how strongly the system is detuned, i.e. Δf=fint−f2.
The response of the detector as a function of the spatial frequency is important.
In contrast to embodiments of tuned phase contrast imaging systems, embodiments of detuned system can only be implemented according to schematics shown in
When the width of the analyzer grating G2 is chosen, for example 1 cm, it might be challenging to precisely fabricate the grating with the pitch that would form expected frequency of the fringe pattern at the detector plane, for example 0.1 cyc/mm. In one embodiment, when the pitch G2 is slightly off of the desired or selected dimensions, the phase contrast imaging system can be tweaked by shifting the analyzer grating G2 along the beam axis (e.g., axis z) relative to the phase grating G1. By shifting the analyzer grating G2 along the beam axis, the analyzer grating G2 can peak at different z position of the interference pattern formed by phase grating G1. In other words, in certain exemplary embodiments, the different frequency of interference pattern, fint, is used to form the desired fringe pattern at the detector plane.
As described herein, in embodiments of tuned phase contrast imaging systems, the phase retrieval algorithm can require multiple x-ray exposures at different lateral positions of analyzer grating, which allows forming a cosine shaped intensity curve shown in
The functional diagram in
In accordance with certain exemplary embodiments, there can be provided methods that can include providing an x-ray generator for radiographic imaging, providing a beam shaping assembly comprising a beam limiting apparatus and a source grating G0, providing an x-ray grating interferometer comprising a phase grating G1, and an analyzer grating G2, and offsetting a pitch of the analyzer grating G2 relative to a pitch of an interference pattern produced by the phase grating G1 at a prescribed distance from the phase grating G1. In one method embodiment, the relative position of the phase gratings G1 and the analyzer grating G2 does not change for a scan of an object, and where the prescribed distance is a Talbot distance. One method embodiment can include producing a fringe pattern that is greater than 0.1 cm or over a significant portion of the analyzer grating G2. In one method embodiment, the grating G1, the grating G2 and the detector D can be fixed at one relative position, attached to the swing arm and moved to image the object, where the relative position of the grating G1 and the grating G2 provide a non-zero Δf. In one method embodiment, a fringe pattern is produced by the pitch of the analyzer grating G2 being unequal to the pitch of an interference pattern produced by the phase grating G1 at a position of the analyzer grating G2, or the fringe pattern is produced by the position of the analyzer grating G2 being offset from a Talbot distance when the pitch of the analyzer grating G2 is equal to a pitch of the interference pattern.
Embodiments of slot-scanning grating-based differential phase contrast systems and/or methods can provide a wide range of potential applications including medical imaging, small-animal imaging, security screening, industrial non-destructive testing, and food inspection. Embodiments according to the application can also be used for phase-contrast applications using other forms of radiation such as neutron and atom beams. Embodiments according to the application can provide a robust and low-cost phase-contrast mammography system with high efficiency and large field of view for clinical applications.
Further, when embodiments according to the application (e.g., grating-based PCI) are combined with a tomographic scan, the three-dimensional distribution of x-ray refraction index in the object as well as the distribution of absorption coefficient commonly obtained in absorption tomography can be reconstructed.
While the invention has been illustrated with respect to one or more implementations, alterations and/or modifications can be made to the illustrated examples without departing from the spirit and scope of the appended claims. In addition, while a particular feature of the invention can have been disclosed with respect to only one of several implementations, such feature can be combined with one or more other features of the other implementations as can be desired and advantageous for any given or particular function. The term “at least one of” is used to mean one or more of the listed items can be selected. The term “about” indicates that the value listed can be somewhat altered, as long as the alteration does not result in nonconformance of the process or structure to the illustrated embodiment. Finally, “exemplary” indicates the description is used as an example, rather than implying that it is an ideal. Other embodiments of the invention will be apparent to those skilled in the art from consideration of the specification and practice of the invention disclosed herein. It is intended that the specification and examples be considered as exemplary only, with a true scope and spirit of the invention being indicated by the following claims.
This is application claims priority to U.S. provisional patent application 61/617,948, filed Mar. 30, 2012, entitled HYBRID SLOT-SCANNING GRATING-BASED DIFFERENTIAL PHASE CONTRAST IMAGING SYSTEM FOR MAMMOGRAPHY, which is hereby incorporated by reference in its entirety.
Number | Date | Country | |
---|---|---|---|
61617948 | Mar 2012 | US |