None.
The present invention relates generally to pump-down diagnostic testing, a technique for assessing near wellbore conditions and treatment isolation characteristics. It is more limited in scope than some of the previously discussed methods but requires no additional equipment or personnel to implement, is very economical and can be implemented on a large scale.
In recent years, plug-and-perf has become a widely adopted hydraulic fracturing technique in horizontal wells completed in unconventional reservoirs (Weijers et al. 2019). It consists of using a wireline cable to run in the well with a frac plug for temporarily sealing previously treated intervals and multiple select-fire perforating guns for creating multiple new fracture-initiation intervals known as perforation clusters. The frac plug and perforating guns are moved down the lateral part of the well by pumping water at a sufficient rate to create drag force as the wellbore fluid flows past the guns, frac plug setting tool, and frac plug as it is being displaced into previously treated clusters. Once pumped to the desired location in the lateral and uphole from all preexisting perforations, pumping is stopped, the frac plug is set, new perforation clusters are created, and the wireline with spent perforating guns and frac plug setting tool is retrieved (see
While fracturing design has evolved significantly, featuring larger fluid and proppant volumes, reduced frac-stage interval length and tighter perforation cluster spacing, there are still concerns about each perforation cluster receiving a reasonably proportionate amount of fluid and proppant (Ugueto et al. 2016; Cramer et al. 2020). If slurry volume is unevenly distributed among the perforation clusters, the treatment is considered to have low perforation cluster efficiency.
Treatments have been designed with varying perforation cluster spacing for dealing with rock heterogeneity, so that rocks that are similar in mechanical behavior are treated simultaneously (Walker et al. 2012). However, this tactic does not address local stress variations resulting from rock displacements generated by previously created or actively propagating multiple fractures, known as stress shadowing (Sneddon 1946). To deal with the inevitable but hard-to-predict variations in rock stress, the limited entry method has been widely applied to achieve high backpressure in casing through choked flow to ensure fluid and proppant entry into most or all perforation clusters, which preferably connect to separate large-scale hydraulic fractures (Cramer et al. 2020).
Effective limited entry treatments require isolation within and outside of the casing (i.e., behind pipe), with intact solid cement filling the drilled hole/casing annular void between perforation clusters. For well-cement-bonded cases, tortuous pathways initially exist from perforation clusters to primary transverse fractures due to narrow, longitudinal starter fractures that are perpendicular to intermediate principal stress and result from drilled-hole hoop stress (Wright et al. 1997). This mechanism has been observed in laboratory block tests, as demonstrated in
To improve perforation cluster efficiency, various diagnostic methods have been applied to characterize the degree of treatment variation among the perforation clusters. These methods include fiber optic measurements, tracers, perforation imaging and treatment pressure analysis.
Distributed acoustic and temperature sensing fiber optic technology (DAS, DTS) has been applied in both in-well (active) and cross-well (passive) applications (Ugueto et al. 2016; Zhang et al. 2017). In-well application refers to the data acquisition and interpretation collected on an instrumented well during the stimulation of the same well. With cross-well measurements, data is measured at a passive instrumented well during the treatment of an adjacent well. Successful implementation of fiber optic sensing projects requires intensive project planning, field operation and data interpretation efforts. The cost of project execution is very high. Thus, its application is generally limited to appraisal wells in new geologic areas or for comparing multiple treatment or completion techniques.
Tracers are used to evaluate perforation cluster efficiency by enabling identification of radioactively tagged particles in treated intervals. These particles are pumped with proppant during a fracturing treatment. After a post-treatment wellbore clean out, a spectral gamma-ray logging tool is run to identify tracer distribution and by association proppant placement in the near-wellbore region (Leonard et al. 2015). Up to three separate radioactive isotopes can be alternately used to estimate the lateral treatment coverage originating from the various frac stages. However, the depth of radioactive logging investigation is only about 2 ft, limiting the ability to determine the presence and concentration of tracers in discreet fractures. This limitation also makes tracer logging prone to detecting near-wellbore deposits of proppant in channels or low spots not associated with fractures or tracer material from a different frac stage that migrated to the logged interval during post-treatment clean out operations.
Downhole video and acoustic based imaging enable investigating individual perforations after stimulation. Significant perforation entry hole erosion has been sometimes indicated from the downhole images, providing evidence of variable slurry distribution across perforations (Cramer et al. 2020; Robinson et al. 2020). However, downhole imaging is typically limited to a small set of wells since additional wellbore preparation efforts are required for a successful operation and the total project cost is significant.
In summary, the diagnostic methods discussed above can provide valuable information for evaluating and modifying plug-and-perf designs for potential improvement in treatment efficiency. However, due to complicated operational procedures and significant cost requirements, their application is usually limited to a few selected candidates for appraisal testing.
The present invention relates generally to a multi-component diagnostic process of evaluating pressure responses associated with customized pump-down and perforating operations in plug-and-perf fracturing treatments. Results from field applications can be instrumental in assessing frac plug and cement sheath integrity, the degree of isolation from previous frac stages and perforation-cluster spacing efficiency, and in improving injectivity in all perforation clusters within a frac stage.
The method is based on analyzing wellbore pressure responses occurring at key segments of the pump-down and perforating operation and correlating the results among multiple frac stages and wells in a field or play. A special requirement is that the frac plug ball-check is run in with the tool string and pumped to seat prior to performing perforating operations. Optionally, a solid bridge plug (of any material composition, including composite type) or poppet-type frac plug (featuring a pre-installed check device) can be used in lieu of a ball-check type frac plug. The key segments are: 1.) pumping-down the wireline, frac plug, perforating guns and accessories to the desired location in the wellbore, then briefly monitoring shut in pressure, 2.) pressure testing the wellbore after setting the frac plug with the ball-check on seat (or simply setting the alternative wellbore-plugging devices noted previously), 3.) selectively shooting perforations in the cluster closest to the toe of the well, and 4.) selectively injecting wellbore fluid into that perforation cluster. For the segments 3 and 4, the pressure response is compared to the pressure decline trend at the end of pump-down (segment 1), looking for characteristic responses associated with isolation from or communication to previous frac stages. A complementary benefit of this process is that selectively establishing injectivity in most distant perforation cluster facilitates spearhead or wireline acid coverage across all perforation intervals for uniform reduction in near-wellbore tortuosity.
In one embodiment, a hydrocarbon well is tested by running a wireline and bottomhole assembly (BHA) consisting of frac plug, setting tool, multiple perforating guns and casing collar locator, with the frac ball (ball check) preinstalled in the frac plug until said BHA reaches the build section of the well; pumping water into the well at a rate of 5-15 bbl/min (0.79-2.4 m3/min) to drag the BHA to the desired location in the lateral; shutting down the pump to obtain an instantaneous shut in pressure (ISIP) and 3 to 5 minutes of shut-in pressure for establishing a pump-down pressure-falloff trend line; activating the setting tool to set the frac plug; moving wireline up the well to place the gun string at the first perforating location; pumping at 1-2 bbl/min (0.16-0.32 m3/min) to seat the frac ball in the frac plug, wherein seating said frac ball isolates previously treated intervals and forms a closed wellbore chamber from the frac plug to surface treating lines; pressurizing the wellbore to at least 1000 psi (6.9 MPa) above the pump-down shut-in pressure; closing a plug valve in the surface treating line to isolate the pumping equipment and optional safety relief valve during this pressure test; monitoring pressure for 3 to 5 minutes to check for pressure-tightness of the closed wellbore chamber; maintaining pressure, selectively perforate the first (toc-ward) cluster interval only and observe pressure falloff response for 3 to 5 minutes; evaluating for communication with or isolation from the previously treated intervals; moving the wireline up the well to locate the BHA away from the perforations; reopen the plug valve; injecting into the first cluster at an injection rate of 2 bbl/min (0.32 m3/min) until treating pressure stabilizes or breaks back; increasing the injection rate to 5-6 bbl/min (0.79-0.95 m3/min), continuing to pump until pressure re-stabilizes; pumping for at least an additional minute; shutting down to obtain ISIP and evaluate pressure falloff response for 3-5 minutes; optionally evaluating for indication of communication with or isolation from the previously treated intervals; perforating the remaining clusters, re-establishing injection at 2 bbl/min (0.32 m3/min) while perforating; discontinue injection after perforating is complete; and retrieving wireline, setting tool and spent guns and prepare the well for the next frac stage.
Optionally, inhibited HCl acid (wireline acid) can be injected during the pump down process and spotted at a sufficient distance uphole from the location of the most distant perforation cluster, to prevent placing the acid in a dead space downstream of that perforation cluster. Injection time is extended during the diagnostic injection test until the wireline acid enters the just-perforated distant perforation cluster. As a result, the wireline acid is placed across all perforation clusters to be subsequently perforated for the upcoming fracturing stage. After perforating all clusters, water is again injected to displace the wireline acid from the wellbore into the formation. This operation can be performed immediately, or at the start of the main fracturing treatment.
In a preferred embodiment, an optimized automated hydraulic integrity system may be used to adjust integrity analysis parameters in real-time.
As used herein, hydraulic fracturing or fracturing (abbreviated as “frac”) is the propagation of fractures through layers of rock using pressurized fracturing fluid. This technique is primarily used in the extraction of resources from low permeability reservoirs, but may be used in a variety of reservoir types where stimulation is required.
As used herein, BHA or bottomhole assembly refers to a fracturing BHA which includes a frac plug, setting tool, perforating guns, CCL, and other downhole tools that may be used for a completion. A bottomhole assembly may also include gauges, sensors, pumps, switches, valves, and other tools that facilitate completion of the well bore.
As used herein, distributed acoustic sensing or DAS is the measure of Rayleigh scatter distributed along the fiber optic cable. A coherent laser pulse is sent along the optic fiber, and scattering sites within the fiber cause the fiber to act as a distributed interferometer with a gauge length approximately equal to the pulse length. The intensity of the reflected light is measured as a function of time after transmission of the laser pulse. When the pulse has had time to travel the full length of the fiber and back, the next laser pulse can be sent along the fiber. Changes in the reflected intensity of successive pulses from the same region of fiber are caused by changes in the optical path length of that section of fiber. This type of system is very sensitive to both strain and temperature variations of the fiber and measurements can be made almost simultaneously at all sections of the fiber.
As used herein, low frequency DAS or LF-DAS refers to a frequency component of the DAS signal that has a period of about 1 second or greater for an interferometer length of a few meters. By using the phase of the low frequency components of the DAS signal, changes along the fiber can be estimated and monitored in real time and with much higher precision than is possible with a conventional measurements. The processor may be configured to process DAS signal data to separate out the low frequency oscillations present in DAS signals.
As used herein, distributed temperature sensing or DTS is a way of measuring temperature in a continuous manner. DTS systems are optoelectronic devices that measure temperature by means of optical fibers functioning as linear sensors. Temperatures are recorded along the optical sensor cable, thus not at discrete widely separated points, but as a continuous profile. Temperature determination is achieved over great distances. Typically the DTS systems can locate the temperature to a spatial resolution of 1 m with accuracy to within ±1° C. at a resolution of 0.01° C. Measurement distances of greater than 30 km can be monitored and some specialized systems can provide even tighter spatial resolutions.
As used herein, diagnostic fracture injection test or DFIT, comprises injecting a relatively small volume of fluid into the subsurface and creating a hydraulic fracture. After the end of injection, the pressure in the wellbore is monitored. The pressure measurements are used to infer properties of the formation, including the leak-off coefficient, permeability, fracture closure pressure (which is related to the magnitude of the minimum principal stress and the net pressure), formation pressure, and the like. These parameters are utilized for hydraulic fracture design and reservoir engineering.
As used herein, instantaneous shut-in pressure ISIP is in adjacent wells, in the same well, or at a surface pressure gauge.
The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee. A more complete understanding of the present invention and benefits thereof may be acquired by referring to the follow description taken in conjunction with the accompanying drawings.
Turning now to the detailed description of the preferred arrangement or arrangements of the present invention, it should be understood that the inventive features and concepts may be manifested in other arrangements and that the scope of the invention is not limited to the embodiments described or illustrated. The scope of the invention is intended only to be limited by the scope of the claims that follow.
The general procedure for a pump-down diagnostic test is outlined below and depicted in the rate-pressure treatment chart shown in
1. Run in the well with wireline and bottomhole assembly (BHA) consisting of frac plug, setting tool, multiple perforating guns and casing collar locator, with the frac ball (ball check) preinstalled in the frac plug (
2. When wireline reaches the build section of the well, initiate pumping water into the well at a rate of 5-15 bbl/min (0.79-2.4 m3/min) to drag the BHA to the desired location in the lateral (i.e., pump-down process). During the pump-down, wellbore fluid is entering previously treated perforation clusters.
3. Shut down to obtain an instantaneous shut-in pressure (ISIP) and 3 to 5 minutes of shut-in pressure for establishing a pump-down pressure-falloff trend line.
4. Following this brief shut-in period, activate the setting tool to set the frac plug. Then move wireline up the well to place the gun string at the first perforating location, which is the perforation cluster closest to the toe of the well.
5. Pump at 1-2 bbl/min (0.16-0.32 m3/min) to seat the frac ball (ball check) in the frac plug, isolating the previously treated intervals and forming a closed wellbore chamber from the frac plug to surface treating lines (or simply set one of the alternative wellbore-plugging devices noted previously).
6. Pressure the wellbore to at least 1000 psi (6.9 MPa) above the pump-down shut-in pressure. Then close a plug valve in the surface treating line to isolate the pumping equipment and safety relief valve (if applicable) during this pressure test. This will prevent potential bleed-back of fluid at the surface to enable selective evaluation of frac plug pressure integrity. Also monitor the wireline lubricator seal throughout the process to check for indications of fluid bleed-back and make adjustments if necessary to achieve a leak-tight seal.
7. Monitor pressure for 3 to 5 minutes to check for pressure-tightness of the closed wellbore chamber.
8. Maintaining pressure, selectively perforate the first (toe-ward) cluster interval only and observe pressure falloff response for 3 to 5 minutes, evaluating for communication with or isolation from the previously treated intervals. Then move wireline up the well to locate the BHA away from the perforations.
9. Reopen the plug valve. Inject into the first cluster at an injection rate of 2 bbl/min (0.32 m3/min) until treating pressure stabilizes or breaks back. Then increase the injection rate to 5-6 bbl/min (0.79-0.95 m3/min), continuing to pump until pressure re-stabilizes. Pump for at least an additional minute. Note: when using inhibited HCl acid (wireline acid) in conjunction with the pump-down process, extend the fluid injection until the wireline acid arrives at the open perforation cluster, then discontinue injection. Doing this will result in the acid being spotted across all subsequent perforation cluster locations, for more uniform reduction of near-wellbore tortuosity among the newly perforated clusters during acid displacement.
10. Shut down to obtain ISIP and evaluate pressure falloff response for 3-5 minutes, once again evaluating for indication of communication with or isolation from the previously treated intervals.
11. Perforate the remaining clusters, re-establishing injection at 2 bbl/min (0.32 m3/min) while perforating. Note: if wireline acid was spotted across perforations, it can optionally be displaced from the wellbore after perforating all clusters-if casing corrosion is a concern. Injection rate can be increased at this time.
12. Discontinue injection after perforating (and acid displacement if implemented) is complete; retrieve wireline, setting tool and spent guns and prepare the well for the next fracturing stage.
During the pump-down diagnostic process, surface pressure, injection rate and wireline data should be recorded to file at a fixed increment of 1 second. Pressure gauge resolution of 0.1 psi (689 Pa) or better is required.
The primary objectives of performing pump-down diagnostics are to evaluate the sealing characteristics of the frac plug, the capacity of the cement sheath to provide isolation from the previously treated intervals in the wellbore, and the impact of cluster spacing on treatment isolation. Secondary objectives include: ability to spot inhibited HCl acid (wireline acid) across the entire perforated interval, evaluation the components of pressure drop in the wellbore system including friction across the bottomhole assembly (BHA) during the pump-down operation for identification of restrictions in the wellbore, comparing pump-down ISIP, leak-off characteristics and water hammer responses among frac stages for assessing in-situ stress and near-wellbore fracture conductivity and locating areas of reservoir pressure depletion and enhanced permeability. For a variation of the last secondary objective, see Roark et al 2017.
The following examples of certain embodiments of the invention are given. Each example is provided by way of explanation of the invention, one of many embodiments of the invention, and the following examples should not be read to limit, or define, the scope of the invention.
Diagnostic fracture injection tests (DFIT's) conducted from a single initiation site near the toe of cased/cemented horizontal wells are characterized by an elevated instantaneous shut-in pressure (ISIP) followed by steep pressure falloff after shut-in. An example of a horizontal-well DFIT is exhibited in
The rate-pressure record of a pump-down diagnostic sequence conducted after the second fracturing stage on a well in the Baltic Basin is shown in
The increase in surface pressure during the post-perforating shut-in period is the result of fluid expansion due to thermal recovery from wellbore cooling. The cooling resulted from the large volume of water injected during previous treatment stages. The pressure buildup indicates that closed-chamber conditions prevailed due to excellent wellbore tubular integrity and effective scaling from previously treated intervals by the frac plug. It also indicates that minimal if any fluid is leaving the wellbore due to lack of behind-pipe communication to previous intervals and the extremely low permeability of the contacted reservoir rock.
An oscillatory pressure signature (known as a water hammer) was observed upon perforating. It was generated by the gun detonation shockwave in a wellbore system lacking the ability to discharge fluid through the new perforations. This led to a strong change in momentum of the detonation pulse which favored the formation of the water hammer (Nguyen et al. 2021). A distinguishing characteristic is that the frequency of water hammer oscillations in this closed-chamber environment (see
A rate of 6 bbl/min (0.95 m3/min) was achieved during the injectivity test, with a tortuous near-wellbore flow restriction indicated by the very high surface treating pressure. When this characteristic is combined with the high, unstable ISIP and significant separation of shut-in pressure from the pump-down pressure-falloff trend line [˜1700 psi (˜11.7 MPa)], the injectivity test pressure response resembles a toe DFIT, providing strong indication that the new treatment interval is completely isolated from the previously treated intervals and that cement sheath quality is adequate in this part of the lateral.
The rate-pressure record of a pump-down diagnostic testing sequence conducted after the tenth fracturing stage on the same Baltic Basin well is shown in
The rate-pressure record of a pump-down diagnostic testing sequence done in a well in south Texas is shown in
An abnormally high injection rate [over 8 bbl/min (1.3 m3/min)] was required to seat the ball for performing the pre-perforating closed chamber pressure test, indicating a defect in the ball or its seat. Surface treating pressure was very low throughout the injectivity test with total friction pressure of 210 psi (1.4 MPa), well less than the predicted friction pressure for injecting through three 0.40 in (10.2 mm) entry holes [495 psi (3.41 MPa)] but slightly greater than the predicted friction for injecting through an 1 in. (25.4 mm) opening in the frac plug [100 psi (0.69 MPa)].
Water hammer events were exhibited when pumping was ended following the injectivity test and the injection during perforating the remaining clusters. In this case, the water hammer oscillations appeared to be of the same frequency as the water hammer oscillations following the pump-down, another indication of strong connection to the hydraulic fracture system associated with previously treated intervals. Water hammer events associated with large conductive hydraulic fracture systems are indicative of through-pipe communication, since behind-pipe channels usually have some degree of flow-path restriction or tortuosity. Tortuosity results in the buildup of back pressure in the casing during injection, leading to restricted, slowly declining flow through the perforations after surface shut in, which suppresses water hammer development.
The rate-pressure record of a pump-down diagnostic testing sequence conducted in an offsetting well in south Texas is shown in
The rapid drop in pressure to the pump-down pressure-falloff trend line following the perforating event coupled with a water hammer oscillatory pulse indicated the frac ball instantly fell off seat or broke apart upon perforating. This enabled the decompressing wellbore fluid to surge through the unplugged opening in the frac plug, inducing a brief but strong rate pulse into a previously treated interval and causing a water hammer after the rate pulse terminated.
Subsequent activities exhibited an identical pattern to the previous test with no evidence that the frac ball reseated. Surface treating pressure was very low throughout the injectivity test with total friction pressure of 210 psi (1.4 MPa), well less than the predicted friction pressure for injecting through three 0.40 in (10.2 mm) perforations [605 psi (4.2 MPa)] but slightly greater than the predicted friction for injecting through an 1 in. (25.4 mm) opening in the frac plug [123 psi (0.85 MPa)].
The rate-pressure record of a pump-down diagnostic testing sequence conducted on a different treatment stage on the same well as above is shown in
Cases of direct communication through the wellbore due to an unseated frac ball or failed frac plug have been infrequently observed in pump-down diagnostics testing. These examples are included to show potential situations that could be encountered which could give misleading results in terms of measuring cement containment between stages.
The pressure test and perforating portion of pump-down diagnostics testing on another horizontal well in south Texas is shown in
Due to pre-installation of the frac ball in the frac plug, a complete perforation gun misfire following setting of the frac plug will result in a job delay, since the ability to do another pump-down is lost once frac plug is set. In those cases, an additional perforating gun run must be made to complete the perforating process by using a wireline tractor or coiled tubing. Using a perforating system featuring addressable-switch gun firing significantly reduces the chance of total misfire, since a gun that fails to detonate can be bypassed and the next gun in sequence can be fired. When using perforating systems with a diode-switch firing mechanism, any misfire in the gun sequence prevents firing of additional guns and forces premature retrieval of the BHA.
Another potential risk is failure to achieve injectivity into any perforation cluster without spotting HCl acid. This would require a coiled tubing run to spot and inject acid into the perforations. This is a very rare occurrence when tubulars and wellhead with high pressure ratings are used. To access this risk, prior treatments in the region should be researched to assess the potential for injectivity problems.
An infrequent problem with pump-down diagnostic injectivity testing was experienced in an early application and exhibited in
Time is money. The primary cost of performing pump-down diagnostics is the incremental time required to perform the work. Incremental time is calculated by determining the elapsed time between the start of the frac plug pressure test and the end of the injectivity pressure-falloff period. The incremental time required for a pump-down diagnostics project performed in south Texas is shown in
This case study is based on a horizontal well completed in the Ordovician Sasino formation in the Baltic Basin of northcentral Poland. The well was drilled to a true vertical depth of 9269 ft (2825.2 m) and had 4910 ft (1497 m) of lateral coverage within the Sasino interval. The casing long-string cement job design specified mixing and pumping 50 bbls (7.9 m3) of 15.0 lb/gal (1.80 kg/L) weighted spacer and 425 bbls (67.6 m3) of 16.0 lb/gal (1.92 kg/L) Class G cement, to be displaced with 3 bbls (0.5 m3) of weighted spacer and 343 bbls (54.5 m3) of 2% KCl water at a rate of 6 bbl/min (0.95 m3/min). Fluids and cement slurry were mixed and pumped as per plan. However, the top cementing plug failed to launch during job execution, leading to severe channeling within the lateral part of the wellbore as the lighter, lower viscosity displacement fluid fingered through and migrated above the denser, more viscous spacer and cement. This channeling phenomenon was evidenced by a leaking shoe joint and layer of set cement in the bottom part of the lateral which necessitated extensive cleanout work prior to doing the fracturing treatments. The plug-and perf treatment process was combined with limited entry treatment methods in performing 25 frac stages with 6 perforation clusters per frac stage. Clusters were spaced at 32.8 ft (10 m) intervals, which was well beyond the expected breath of the longitudinal starter fracture and associated transverse fractures. Pump-down diagnostics were performed during all 24 pump-downs. Rate/pressure plots for two of these pump-down diagnostic tests were previously shown in
Treating pressure tended to be much higher during the injection tests in fracturing stages demonstrating behind-pipe isolation from the previous treated intervals. This relationship is exhibited in
Incremental time for doing the diagnostic testing was calculated by determining the elapsed time between the beginning of the frac plug pressure test and the end of the injectivity pressure-falloff period. Here are the statistical calculations on incremental time for the 24 stages: mean=19 minutes; median=16 minutes; mode=16 minutes; minimum=13 minutes; maximum=34 minutes.
Cement quality in the annular gap between casing and drilled hole was evaluated prior to the fracturing treatments with an acoustic radial bond tool. Log results for a lateral section near the toe of the well, from 4200-4260 m (13,780-13,976 ft) measured depth are shown in
Using the method listed in previous sections of the paper, pump-down diagnostics were performed on 27 horizontal wells drilled in a south Texas reservoir. Interpretable data was collected on a total of 304 stages for determining if there was pressure isolation from the previous stage.
Analysis of pressure responses. A summary of the testing outcomes for all wells is shown in
The variable that had the strongest correlation with indicated pressure isolation between stages was cluster spacing. Stages that implemented wider cluster spacing had a higher likelihood of exhibiting good pressure isolation between stages. This relationship is illustrated in
It is notable that abundant data was available for cases with 15 ft (4.6 m) cluster spacing (256 stages), less data was available for cases with 25 ft (7.6 m) cluster spacing (43 stages), and even less data was available for cases with 35 ft (10.7 m) cluster spacing (5 stages). However, stages with wider cluster spacing regularly had much larger surface pressure separations from the pump-down pressure-falloff trend line following the injectivity test, which supports the notion of cluster spacing being a dominant factor in pressure isolation between stages. The examples shown in
To further evaluate the impact of cement quality on isolation characteristics, treatment isolation was correlated with the distance between the tested perforation cluster and the closest casing collar or centralizer. In a horizontal wellbore, the casing string tends to lay on the low side of the wellbore. A casing centralizer or casing collar with its larger OD forms an external upset that helps support the string and increases clearance on the low side, improving cement quality along the corresponding lateral interval (Haut and Crook 1979). Treatment isolation as a function of the distance between tested perforation clusters and the closest casing collar or centralizer is shown in
Well production analysis. Of the wells completed with 15 ft (4.6 m) cluster spacing, 16 wells have been on production for a minimum of 9 months and were compared against expected productivity, i.e., mass production rate divided by calculated reservoir pressure drawdown. The findings are shown in
Despite these findings, when treatments are confined to targeted perforations and even better, targeted perforation clusters, the created far-field transverse fractures along the lateral will be more uniform in extent and distribution. Reservoir simulations indicate that hydraulic fracture uniformity leads to more uniform drainage of reservoir rock and superior long-term field economics. Short-term production results do not always capture the negative effects of irregular fracture coverage, especially when observations are limited to a small set of wells. For the South Texas case study wells, 15 ft cluster spacing may be too close to prevent inter-stage communication during the subsequent high-volume fracturing treatments, even when a high-quality cement sheath is present. Determining volume-dependent inter-stage communication tendencies is not within the scope of pump-down diagnostics testing.
By developing a model of previous integrity analyses along with real-time data, an automated hydraulic integrity system can be deployed to 1.) generate a pump-down trend line and compute the amount of separation between that trend line and post-perforation and injectivity-test pressure responses and 2.) calculate the periodicity and decay rate of water hammer oscillations. This information is then processed using an algorithm routine to determine the existence or absence of inside-pipe and behind-pipe isolation of the new treatment stage from previously treated intervals. This same automated system can process databased operational data to dramatically reduce the time required to analyze historical pump-down data.
Pump-down diagnostics provide a means of checking if communication is occurring between a just-perforated fracturing stage and previously treated intervals, which can serve as a key performance indicator for treatment control and cement sheath integrity. For moderate perforation cluster spacing (e.g., 33 ft (10 m) between clusters), pump-down diagnostics have been shown to provide a more reliable diagnosis of cement sheath quality along the lateral than cement bond log evaluation. For close perforation cluster spacing (e.g., 15 ft (4.6 m) between clusters), pump-down diagnostic results for stage isolation may be more affected by the breadth of the longitudinal starter fracture and associated hydraulic fracturing activity than by the cement sheath quality. The timing and oscillatory frequency of water hammer events observed during pump-down diagnostic operations offer additional clues to the nature of inter-stage communication. Pump-down diagnostics are time efficient and economical, typically requiring about 15 minutes per frac stage. Pump-down diagnostics risk factors can be effectively mitigated by using an addressable-switch select-fire perforating system, applying area experience to assess fracture initiation behavior in the absence of spotting HCl, and modifying or foregoing the injectivity test if the frac plug fails the pressure test.
In closing, it should be noted that the discussion of any reference is not an admission that it is prior art to the present invention, especially any reference that may have a publication date after the priority date of this application. At the same time, each and every claim below is hereby incorporated into this detailed description or specification as a additional embodiments of the present invention.
Although the systems and processes described herein have been described in detail, it should be understood that various changes, substitutions, and alterations can be made without departing from the spirit and scope of the invention as defined by the following claims. Those skilled in the art may be able to study the preferred embodiments and identify other ways to practice the invention that are not exactly as described herein. It is the intent of the inventors that variations and equivalents of the invention are within the scope of the claims while the description, abstract and drawings are not to be used to limit the scope of the invention. The invention is specifically intended to be as broad as the claims below and their equivalents.
All of the references cited herein are expressly incorporated by reference. The discussion of any reference is not an admission that it is prior art to the present invention, especially any reference that may have a publication data after the priority date of this application. Incorporated references are listed again here for convenience:
This application is a continuation application which claims benefit under 35 USC § 120 to U.S. application Ser. No. 17/577,848 filed Jan. 18, 2022, entitled “Hydraulic Integrity Analysis,” which is a non-provisional application which claims benefit under 35 USC § 119(c) to U.S. Provisional Application Ser. No. 63/138,138 filed Jan. 15, 2021, entitled “Hydraulic Integrity Analysis,” which is incorporated herein in its entirety.
Number | Date | Country | |
---|---|---|---|
63138138 | Jan 2021 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17577848 | Jan 2022 | US |
Child | 18657136 | US |