1. Field of the Invention
This invention relates to the need to simplify and accurately dispense the application of immersion fluid for immersion optical systems.
2. Description of the Prior Art
No mechanism exists to accomplish the task of dispensing the immersion fluid. Presently, the fluid is dispensed manually utilizing an eye dropper or its equivalent. For upright microscopes, It is not uncommon for excessive fluid to be dispensed on the specimen using this technique. When using an inverted microscope, the user must apply the drop of fluid on the objective lens which can be partially obscured as it is nested inside, or below, the microscope stage.
It is the object of this invention to provide the microscope user with a device that will accurately dispense the precise amount of immersion fluid at the proper location.
It is another object of this invention to accomplish these tasks in either a manual mode with minimal mechanical demands on the user, or in an automated configuration.
The satisfaction of these objectives will result in a cleaner working environment, less waste of the immersion fluids, and the elimination of the awkward and inaccurate application procedures currently in use.
In the drawings:
The mechanical actuator is a two-stage device with an upper section 4 and a lower section 5. Both sections share a common pivotal axis. The upper section contains a constrained spring 6 that initially forces the upper section to rotate in concert with the lower section.
As the flexible driving plunger 7 is initially displaced, it rotates the complete assembly about the pivot and positions the output port of the drop dispenser into position above the front objective lens 8. At this point, the upper section encounters the fixed stop 9 and ceases rotating. Further displacement of the plunger causes the lower section to overcome the spring's static force. The lower section continues to rotate and a linear actuator 10 drives the ratcheting roller bearing assembly 11. By peristaltic action, the immersion fluid is squeezed from the peristaltic chamber 12 out and through the dispenser outlet port.