The present invention relates to improvements relating to the profiling of particles, in particular, proteins, in microfluidic devices.
Non-covalent interactions are predominately responsible for the folding, binding and assembly of many proteins. Protein interactions with other partners are often associated with its specific amino acid sequence and its post-translational modification. These unique properties can lead to either hydrophobic or electrostatic noncovalent interactions. The overall charge of proteins and complexes in solution can be dependent on the total compositions of accessible charged groups and can usually be determined by the isoelectric point (pI), the pH value where the net charge is zero. Protein interactions, oligomerisation and assembly are highly regulated processes in organisms and influence the individual function of each protein. Uncontrolled and unregulated misfolding and interaction of proteins is an important class of malfunction, and often leads to protein aggregation. Therefore, understanding protein-protein interactions in complex mixtures is of key relevance in modern protein science.
In proteomics, the understanding of protein-protein interactions is key and therefore there is a strong driver within the industry to understand these interactions. In order to probe and predict these reactions, each protein needs to be profiled. This means, within the context of this patent, that the individual and fundamental biophysical properties of the protein need to be determined. These include, but are not limited to, the isoelectric point (pI), hydrodynamic radius, hydrophobicity, molecular weight, stoichiometry of binding partners and binding affinity.
Techniques capable of characterising complex mixtures are limited in the field and these techniques are only commonly used to provide unidimensional information. As protein mixtures are highly dynamic and composition depends on various exogenous factors such as temperature, pH, local protein/salt concentration, viscosity and matter of state, it can be challenging to accurately determine the physiological and biophysical properties of the proteins within a mixture.
One of the most powerful methods capable of profiling these particles is mass spectrometry. This technique requires the transfer of the analyte from liquid to gaseous phase. This diminishes or completely destroys biologically relevant interactions such as non-covalent protein-protein interactions and also protein-ligand interactions. This technique is therefore destructive of the sample that it seeks to profile and is capable of providing only limited information about the native state of the proteins present.
The destructive and incomplete analysis available using mass spectrometry has led to a focus on protein characterisation methods that are able to determine protein properties in solution. There are many such techniques, including nuclear magnetic resonance spectroscopy, circular dichroism, isothermal titration calorimetry, fluorescence spectroscopy, dynamic light scattering, multi-angle light scattering and analytical centrifugation. Most of these techniques are capable of measuring a single attribute at any one time. These attributes include the structure, binding properties, Stokes radius or charge. This is very limiting in terms of profiling throughput as a number of separate experiments are needed in order to characterise a protein mixture fully.
Therefore, there is a requirement to provide experimental apparatuses and methods for measuring multiple parameters of one or more particles such as proteins in a sample mixture. However, techniques providing an experimental set-up for measurements of multiple parameters are rare and often non-applicable for a wide variety of proteins. Conventional approaches, including but not limited to, size-exclusion chromatography coupled multi-angle light scattering (SEC-MALS), 2D Gel electrophoresis, liquid chromatography coupled mass spectrometry (LC-MS), Liquid chromatography coupled nuclear magnetic resonance spectroscopy (LC-NMR), high performance anion exchange coupled pulsed amperometric detection (HPAECPAD) and electrochromatography have been used to obtain multiple parameters of a protein in the sample mixture. While these techniques have their advantages, there are known limitations for using these techniques. For instance, some approaches require special probes for example isotopes, oxidisable functional groups, protein tags or, elevated temperature conditions, sample vaporization, or use of high sample concentration for analysis. Other methods often show non-specific binding and/or low signal to noise ratio for example, NMR spectroscopy and capillary electrophoresis.
Microfluidic devices, sometimes referred to as lab on a chip devices, enable the manipulation and control of small volumes of fluid, typically in the range of picolitres to microliters, in microfabricated structures. Microfluidic systems have a compact footprint and can therefore parallelise experiments thus reducing overall time.
It is against this background that the present invention has arisen.
According to the present invention there is provided a device for profiling particles such as proteins, the device comprising: a liquid chromatography column; a plurality of microfluidic analysis modules; wherein the microfluidic analysis modules are configured to provide multi-dimensional analysis of the particles; and wherein the flow of fluid through the device is smoothed to provide a consistent and continuous fluid flow.
Microfluidic device can be an optimal solution to utilise without the requirement for large sample volumes. Microfluidic devices may allow manipulation and control of small quantities of one or more fluid samples, usually in the range of pico- to microliters. A wide range of liquid chromatography methods for example, size-exclusion, reversed phase, ion-exchange and affinity chromatography can be used in conjunction with microfluidic devices. Microfluidic devices may be used for mapping out physiological protein complexes from endogenous samples. For chromatography methods, the bedding material, known as the stationary phase, can influence the purification of proteins within a mixture and can comprise biomolecules such as dextran, agarose or cellulose and synthetic polymers such as polyacrylamide, polystyrene or silica-based polymers. The selection of the mobile phase may control the interplay between the separating molecules and the matrix and is usually organic or buffered. By combining liquid chromatography (LC) protein separation and microfluidics, the present invention is provided to enable multidimensional characterisation of complex mixtures and sample fractionation. By measuring the sample composition in the condensed phase, single proteins and/or protein complex formation under native conditions may be analysed. Moreover, the microfluidic systems according to the disclosure of the present invention may allow for the simultaneous determination of multiple parameters such as hydrodynamic radius and electrophoretic mobility of molecules in a quantitative manner in complex mixtures.
In another advantage, the use of LC with a plurality of microfluidic analysis modules may only require a small fraction of sample for analysis whereas the main volume of the sample can be collected and be used for further evaluation. Due to the compact size of microfluidic devices, the addition of microfluidic systems/microfluidic analytical modules into existing LC systems can be relatively simple to implement. Thus, this can be highly advantageous as it provides a low cost, simple and effective apparatus and method for profiling particles such as proteins.
By “consistent and continuous flow”—we mean that the flow rate will vary by no more than 10%, no more than 5% or even less than 2%. Consistent flow rate is important because it is required to enable the sizing of the particles by diffusion and measure the electrophoretic mobility.
The consistent flow rate can be at least partially achieved through selection of liquid chromatography column. For example, some chromatography columns are provided with two pumps in anti-phase to improve the flow stability.
Alternatively, or additionally, the consistent flow rate can be achieved by introducing some compliance into the system. This is counter intuitive within the context of microfluidic devices because there is usually a strong driver to reduce the compliance within the system as this leads to slow fluidic response times and results in variable hydrodynamic resistance which, in turn, makes it difficult to predict flow at a junction.
Alternatively, or additionally, the consistent flow rate can be achieved by providing a smooth buffer flow on dilution of the sample flow. The buffer may be provided in a volume around ten times the volume of the sample flow and therefore the provision of a continuous flow of buffer will ensure that the combined, diluted flow has a consistent flow rate.
Multi-dimensional analysis, within the context of this patent application, means analysis of a particle, such as a protein using two or more different attributes of the particle, substantially simultaneously.
Multidimensional analysis may be achieved by a number of different permutations and combinations of key techniques. In some embodiments and examples, simultaneous acquisition of multidimensional characteristics can be advantageous as sequential measurements can be taken to show different states and composition of unequilibrated molecular mixtures.
In some embodiments a single image is taken and from that image multiple data points can be ascertained. An example of this may be the calculation of the charge of the particle calculable from an image showing diffusional broadening and electrophoretic motion giving the electrophoretic mobility of the particle. In some embodiments, multiple separate measurements can be synchronized in order to facilitate the multidimensional analysis. In some embodiments, data can be combined and then extrapolated to provide the multidimensional analysis.
In a case when the sample is not well separated, i.e. has multiple molecular peaks overlapping with each other, it is very difficult to measure the biophysical properties of molecules within these peaks. By using the low noise signal at the first absorbance detector and taking the second derivative of absorption, we can estimate the molecular elution positions for the different molecular species. The corresponding molecular elution volumes are then used to pick data points from the continuous size and electrophoretic mobility measurement dataset and give an estimate of the biophysical properties of molecules which is difficult to obtain otherwise.
Multidimensional analysis may include calculation of the isoelectric point. Calculation of the isoelectric point enables predictions of the behaviour of the particle, especially protein, under different conditions. Typically, a pH gradient has been generated and then it has been determined the position of the particle across that gradient and that pH has been identified as the isoelectric point. However, this workflow does not permit the charge to be determined. However, this process is considerably improved by taking a multidimensional approach and measuring the deflection at two discrete pH values and combining these measurements with diffusional sizing. By making this multidimensional combination, a calculation of the isoelectric point can be achieved by the measurement of deflection at only two different pH values. The multidimensional analysis therefore simplifies the overall analysis and reduces the number of data points required in order to calculate the isoelectric point.
A fractionation device may be provided. The fractionation device may be downstream of the liquid chromatography column and upstream of the microfluidic analysis modules. In many embodiments, the fractionation of the fluid flow precedes the multi-dimensional microfluidic analysis. However, in some embodiments, it may be desirable to collect doubly separated fractions and this could be achieved by the provision of the fractionation device downstream of at least one of the microfluidic analysis modules.
The device may further comprise a controller configured to use the multi-dimensional analysis obtained from the microfluidic analysis modules in order to assess the quality of the liquid chromatography column.
This feedback loop provides quality assurance for the liquid chromatography column and also aids in the identification of unknown eluting species.
The controller may be further configured to use the multi-dimensional analysis obtained from the microfluidic analysis modules in order to control the fractionation device.
Furthermore, according to the present invention there is provided a device for conducting multi-dimensional profiling of particles such as proteins, the device comprising: a liquid chromatography column; more than one microfluidic analysis modules; and wherein the device is configured to provide continuous real-time data acquisition.
The device may further comprise a detector configured to, detect and record data from each microfluidic analysis module.
The detector may include a microscope and a detector such as a camera for recording the data. The detector may be optical or non-optical. If the detector is non-optical it may be selected from the following non-exhaustive list of sensors: a biosensor; an electrochemical sensor; a point detector that is scanned across the region to be sensed; a mass sensor such as a quartz crystal microbalance or cantilever system. Alternatively, the detector may work using chemiluminescence which does not rely on an illumination source but rather the chemical stimulation leading to light emission with subsequent capture via an optical sensor.
The detector may be further configured to illuminate at least part of the microfluidic analysis module. The illumination source may be an LED or a laser.
The data may be recorded using a CCD camera, CMOS camera or other optical data recording device.
The device may further comprise a flow adapter. The flow adapter is provided to smooth the fluid flow so that it is constant. Constant within this context means that there is less than 10% variation in flow over time. In some embodiments, this threshold may be reduced to less than 5% or even less than 2%. Alternatively or additionally, the flow adapter connects the chromatographic column with a plurality of microfluidic analysis devices.
The device may further comprise a device for measuring optical absorption. The device for measuring optical absorption, such as UV absorption, may provide an absorption measurement cell for monitoring the separated molecules. The continuous data received from the microfluidic modules can be matched to the absorption measurements and thereby the measured peaks can be assigned to data points. In some embodiments, the second derivative of the spectrum can be used to identify the most significant sub-peaks. In circumstances where there is prior information about the peaks, for example, which peak is which molecule, it will then be possible to assign the observed data to the known molecules.
The microscope may be an intrinsic fluorescence microscope or an epifluorescence microscope. The epi-fluorescence microscope may be matched to the wavelength of the fluorescent label deployed.
All of the data to be observed and recorded may be configured to fall within the field of view of the microscope.
Furthermore according to the present invention there is provided a method of multi-dimensional profiling of particles such as proteins present in a fluid sample; the method comprising the steps of: introducing the fluid sample containing the particles to be profiled into a liquid chromatography column; consistently and continuously flowing the fluid output from the column into each of a plurality of microfluidic analysis modules in parallel; detecting data pertaining to multiple characteristics of the particles by observing the fluid within the microfluidic analysis modules; and combining the data to calculate one of more attributes of the particle profile.
The characteristics detected may include the hydrodynamic radius and the electrophoretic mobility and the attribute calculated may therefore be the effective charge.
Alternatively or additionally, the characteristics detected may be the mobility and diffusional size and the attribute calculated may therefore be the isoelectric point.
The method may further comprise fractionating the fluid containing the particles. The fractionation of the fluid may include 90% of the sample eluting from the liquid chromatography column.
In some embodiments, the initial sample may be a 40 ul sample size and may be a high concentration protein solution. After separation, each of the proteins is diluted 5-10 times depending on the chromatography conditions. Typical operation time of a column is 1-3 hours depending on the flow rate. Laminar flow is established both on chip and at the flow adapter, prior to the introduction of the proteins which are subsequently introduced onto the chip sequentially. In this way, the device of the present invention minimises or even eliminates losses of sample due to transient processes. The flow adapter samples continuously all of the sample coming from the chromatography column so that the whole sample is analysed.
The invention will now be further and more particularly described, by way of example only, and with reference to the accompanying drawings, in which:
The present invention relates to a serial combination of a preparative analyte separation technique with parallel (or parallel and serial) analytical and preparative fluidic device network capable of sample fractionation and simultaneous multidimensional analyte characterisation.
Referring to
Referring to
The fluidic network can be performing simultaneous measurements on the physical properties of the separated analytes e.g. hydrodynamic radius, electrophoretic mobility, effective charge, isoelectric point. A detection device 18 yields a signal readout representing the analyte concentration variation in time after separation in the separator 16 and can be used to reference and match the sequence of the measured analyte physical properties in the parallelised fluidic network. A detection signal from the detection device 18 can also be used to introduce weight for the confidence/importance in each of the measured analyte physical properties. The network may include an arbitrary number of further separation modules e.g. Capillary, Free-flow electrophoresis enabling more complex analyte mixture separation and characterisation.
Referring to
In
Referring to
The label-free mixture 62 comprises thyroglobulin, conalbumin and lysozyme. The second heterogeneous seven component mixture 64, which is prepared by mixing streptavidin, biotinylated BSA and Atto-488 molecules, comprises five Atto-488 labelled complexes (molecules I-V).
To demonstrate the functionality of the method according to the present invention, a mixture of three proteins varying in size and isoelectric point (pI): bovine thyroglobulin (Mw=670 kDa, pI=4.5, GE Healthcare, 28-4038-42), chicken conalbumin (Mw=76 kDa, pI=6.7, GE Healthcare, 28-4038-42,) and chicken lysozyme (Mw=14.3 kDa, pI=9.3, Sigma-Aldrich, L6876) may be selected. The proteins can be diluted in a 100 mM sodium HEPES buffer (pH=7.3) at a ratio of 4.6:33:110 μM respectively; total sample volume 40 μL. The second system can be used to generate a heterogeneous sample based on Streptavidin-Biotin complex formation, which may be one of the strongest known non-covalent interactions between a protein and a ligand. A mixture can be prepared by incubating Streptavidin (Prospec, Israel, PRO-791), biotinylated bovine serum albumin (Generon, UK, 7097-5) and a biotinylated Atto-488 (ATTO-TEC GmbH, Germany) dye at a ratio of 1:1:3 (20:20:60 μM, total volume 40 μL) for typically about an 1 hour at room temperature in 10% Phosphate buffered saline solution (0.1×PBS, pH=7.3). The mixture is expected to form seven distinct complexes with sizes ranging from 1 kDa to 300 kDa, as shown in
In
The flow adapter 66 interface may enable standard liquid chromatography (LC) fractionation and simultaneous multi-dimensional eluting molecule characterisation. The LC separation on AKTA Pure can be driven by two high pressure pumps maintaining a highly stable flow of 10 μL/min=600 μL/h with a 1-5% fluctuation level depending on the buffer and the age of separation column. The microfluidic flow adapter 66 with carefully adjusted resistances can be used for distributing the incoming fluid from the LC absorption cell between one or more e.g. two microfluidic sample inlets and a fractionation outlet. The flow rates at the chip ports 3 and 6 89 may be measured to be at 40.0±0.7 μL/h and 37.4±0.7 μL/h respectively, which may represent the electrophoresis device and diffusional sizing sample inlets on the chip. The rest of the post LC separation fluid about 90% can be collected via the fractionation outlet 65.
Referring to
As shown in
The microfluidic device may be custom designed for fitting two distinct analytical blocks in one fluorescence microscope field of view. The first block—the diffusional sizing device—may have a long diffusion channel of length LD=43 mm, width of WD=300 μm and height of HD=55 μm, as illustrated in
Referring to
where v is the transverse velocity, E—the applied electric field, x—the measured profile deflection, V—the applied voltage across the channel, d—the distance of the profile deflection measurement along the channel and h—the channel height.
Referring to
As illustrated in
Referring to
As shown in
In one example, the charge of a biotinylated Atto-488 dye may be measured to be (−0.99±0.11) e which appears to be in agreement with the expected charge of μe close to neutral pH conditions. Streptavidin with 4 bound dyes resulted in the size of (3.51±0.13) nm and the effective charge of around (−2.83±0.28) e. BSA at normal pH conditions has been known to have an effective charge of around −7e. Therefore, the expected effective charge of the Streptavidin-BSA complexes III-V, as shown in
Referring to
The system, apparatus, device and methods of the present invention have established a direct coupling between size exclusion chromatography with a parallelised microfluidic analysis while being able to fractionate about 90% of the total sample volume. The multidimensional characterisation of distinct complexes may be used to yield simultaneous size, electrophoretic mobility and effective charge measurements. As described herein, the operation principle of the present invention can be used for determining the biophysical properties of unlabelled standard analytes such as proteins within a mixture, as well as analysing multiple partially separated peaks after chromatographic separation and predicting the effective charge and molecular size of complexes of a heterogeneous labelled molecule within a mixture.
To demonstrate the functionality of the method according to the present invention, a mixture of three proteins varying in size and isoelectric point (pI): bovine thyroglobulin (Mw=670 kDa, pI=4.5, GE Healthcare, 28-4038-42), chicken conalbumin (Mw=76 kDa, pI=6.7, GE Healthcare, 28-4038-42,) and chicken lysozyme (Mw=14.3 kDa, pI=9.3, Sigma-Aldrich, L6876) may be selected. The proteins can be diluted in a 100 mM sodium HEPES buffer (pH=7.3) at a ratio of 4.6:33:110 μM respectively; total sample volume 40 μL. The second system can be used to generate a heterogeneous sample based on Streptavidin-Biotin complex formation, which is one of the strongest known non-covalent interactions between a protein and a ligand. A mixture can be prepared by incubating Streptavidin (Prospec, Israel, PRO-791), biotinylated bovine serum albumin (Generon, UK, 7097-5) and a biotinylated Atto-488 (ATTO-TEC GmbH, Germany) dye at a ratio of 1:1:3 (20:20:60 μM, total volume 40 μL) for 1 hour at room temperature in 10% Phosphate buffered saline solution (0.1×PBS, pH=7.3). The mixture is expected to form seven distinct complexes with sizes ranging from 1 kDa to 300 kDa, as shown in
Two different buffers may be used for the sample elution through the HPLC column. In a first instance, a 100 mM sodium HEPES buffer (pH=7.3) can be used for the label-free sample characterisation and streptavidin-biotin mixture may be eluted in a 0.1×PBS (pH=7.3) buffer. Both buffers may also contain 0.01% Sodium azide and 0.1% Tween to reduce sample adhering to microfluidic channels. A Superdex 200 Increase 3.2/300 column (GE Healthcare, UK) at a flow of 10 μL/min may be operated on an A KTA Pure System (GE Healthcare, UK). The eluting sample absorption at 280 nm and 500 nm wavelengths may be monitored simultaneously with a 10 mm path length absorption monitor U9-M (GE Healthcare, UK). The absorption intensity may be used for matching the molecular elution volume with the image sequence on a fluorescence microscope. The flow from the liquid chromatography (LC) separation can be connected to a microfluidic flow adapter.
A microfluidic junction (P-722, IDEX Health & Science, USA) with carefully pre-cut polyether ether ketone (PEEK) capillaries (IDEX Health & Science, USA) and flow sensors can be built, directing only a fraction of the flow coming from chromatographic separation into multiple microfluidic devices, as shown in
The microfluidic device may be custom designed for fitting two distinct analytical blocks in one fluorescence microscope field of view. The first block—the diffusional sizing device—may have a long diffusion channel of length LD=43 mm, width of WD=300 μm and height of HD=55 μm, as illustrated in
Referring to
where v is the transverse velocity, E—the applied electric field, x—the measured profile deflection, V—the applied voltage across the channel, d—the distance of the profile deflection measurement along the channel and h—the channel height.
Hollow metal with approximately 1.5 mm ID electrodes may be inserted into device ports 8 and 9 where a power supply (EA Elektro—Automatik 6230207, Germany) is connected to the chip via a digital multimeter (Agilent 34410A, USA) recording a current flowing through the circuit. The two microfluidic devices were operating continuously and a measurement of the hydrodynamic radius, electrophoretic mobility and charge were obtained for every 3.3 μL of the eluting sample (every 20 seconds) from the column while still fractionating 90% of the total volume.
The devices may be fabricated using a standard polydimetylsiloxane (PMDS) soft-lithography approach. The master for the replica molding of PDMS may be fabricated with an SU-8 photolithography process. After mixing PDMS (Sylgard184, Dow Corning, two components 10:1 ratio and degassed) and casting it onto the photo-lithographically defined structure, it is cured at 70° C. for 1 h. A carbon black nanopowder (Sigma-Aldrich) may be added to the PMDS before curing to create black devices, thus minimizing background noise and the unwanted autofluorescence from PDMS under 280 nm-LED illumination during the measurements. The PDMS replica of each master may then cut, and the connection holes may be made with a biopsy punch. The PDMS device may be sonicated for 3 min in isopropanol, blow dried with N2, and placed in an oven at 70° C. for 10 min. Finally, the replica may be activated using O2 plasma at a 40% power for 10 seconds (Diener etcher Femto, Germany) and bonded to a clean quartz slide (Alfa Aesar, 76.2 25.4 1.0 mm).
Regarding fluorescence microscope set up according to the present invention, one or more different fluorescence microscopes may be used for the experiments as described in the present invention. For example, an intrinsic fluorescence microscope for a label free protein detection and a green label epifluorescence measurement setup may be used. In one example, an auto-fluorescence measurement of proteins containing the aromatic amino acid tryptophan may be carried out on a quartz-based intrinsic fluorescence visualisation platform. In short, the proteins may be illuminated with a 25 mW 280 nm LED (M280L3, Thorlabs, UK) through an excitation filter (FF01-280/20-25, Semrock, USA) centered at a λex=280 nm and a dichroic mirror (FF310-Di01-25x36, Semrock, USA). The fluorescence from the sample may be collected through an emission filter (FF01-357/44-25, Semrock, USA) centered at λem=357 nm, and can be focused onto an EMCCD camera (Rolera EMC2, QImaging, Canada). The green epifluorescence microscope, as illustrated in
In some embodiments, there may be a slight delay between the molecule absorption measurement after the LC separation and the detection on chip. The delay volume from the absorption measurement cell to the flow adapter can be 70 μL and the volume from the flow adapter to the chip detection channel is around 8 μL causing 20-30 min delay time depending on the system flow. The elution volume with the microscope image sequence may be matched by comparing absorption intensity on the absorbance detector (280 nm and 500 nm) and the fluorescence intensity of the eluting sample on chip.
The diffusion coefficient D can be used to quantify the fluctuations of a particle under Brownian motion and is described for a spherical particle by the Stokes-Einstein equation:
where η is the viscosity of the solution, Rh is the hydrodynamic radius and kB and T are the Boltzmann constant and absolute temperature, respectively. The measured diffusion constant D and the electrophoretic mobility μe can be used to estimate the complex charge:
where κ—the inverse Debye length and f1 is a function of Rh that describes the effect of the electric field distribution around the particle. For most of the proteins in high salt buffers f1 (κRh)≈1 since κRh<<1. Hence, the charge can be estimated by the Nernst-Einstein relation
Referring to
The flow systems that are often used for liquid chromatography techniques for purification or separation of complex mixtures may not readily be compatible with microfluidic devices. In order to match macrofluidic flows of high pressure flow systems with microfluidic flows found in chips with micron sized features, the flow adapter 54 can be scalable to various rates. In some embodiments, the flow adapter 54 can be a macrofluidic or a microfluidic flow adapter.
As illustrated in
The microfluidic flow adapter may be provided with resistances. The resistances of the flow adapter may be adjusted such that the flow adapter is able to distribute the incoming eluted fluid flow following the LC absorption cell between two or more microfluidic sample inlets and a fractionation outlet. In some embodiments, the flow can be tailored to the requirements of each downstream microfluidic device, for example each of the microfluidic devices may receive a different fluid flow i.e. not the same fluid flows. Alternatively, each of the microfluidic devices may receive the same or similar fluid flows. In some examples, the flow rates at the microfluidic inlets for the free-flow electrophoresis and the diffusional sizing devices are measured to be around 6.7±0.1% and around 6.2±0.1%, respectively, of the initial flow rate. The remaining portion of the post LC separation fluid (eluted flow) may not be used for further characterisation. Additionally or alternatively, the remaining portion of the post LC separation fluid can be collected via the fractionation outlet 55. By way of example only, the remaining portion of the post LC separation fluid can be 90% or it may be more than 90%. In some examples, the portion may be less than 90%, 80%, 70%, 60%, 50%, 40%, 30%, 20% or 10%. These ratios may be adapted to the microfluidic application used or separation procedure applied. The flow rates of the LC system can be dependent on the column used.
Referring to
Referring to
In
In order to detect and analyse unlabelled proteins in a mixture following size exclusion chromatography (SEC), a microscope 56 e.g. a UV-microscope set up capable to excite amino acids and in particular, aromatic amino acids is used to measure intrinsic protein fluorescence as illustrated in
The eluting samples can be continuously loaded into a plurality of microfluidic analysis modules such as a free flow electrophoresis device and/or a diffusional sizing device as shown in
Referring to
In
For the experimental data shown in
A further second derivative analysis on the absorption signal can be applied. In this case, only 4 distinct Atto488 labelled molecules and the free dye can be assigned. The first major peak with elution volume between 1 ml and 1.5 ml has three sub peaks. However, using the second derivative analysis of absorption at 500 nm, the approximate elution volumes for streptavidin with one, two and three BSA molecules are founded to be 1.11 ml, 1.20 ml and 1.33 ml respectively, as shown in
Referring to
The electrophoretic mobility of the molecules over a range of different pH conditions can be used to identify the pI value of individual molecules. Referring to
A Superdex 200 Increase 3.2/300 column (GE Healthcare, UK) at a flow of 10 μL/min is operated on an AKTA Pure System (GE Healthcare, UK). The samples are monitored at absorption at 280 nm and 500 nm wavelengths simultaneously with a 10 mm path length absorption monitor U9-M (GE Healthcare, UK). The absorption intensity is used for matching the molecular elution volume with the image sequence on a fluorescence microscope. The flow from the LC separation is connected to the microfluidic flow adapter.
A microfluidic junction (P-722, IDEX Health & Science, USA) with polyether ether ketone (PEEK) capillaries (IDEX Health & Science, USA) and flow sensors (MF2 7 μL/min, Elveflow, France) are designed and manufactured to direct only a fraction of the flow coming from chromatographic separation into multiple microfluidic devices. The lengths of the capillaries are as follows: the fractionator output is made of a capillary with L1=10.2 cm and 125 μm ID and the outputs A and B are made of two capillaries (L1=10 cm with 125 μm ID and L2=8.1 cm with 67.8 μm ID. Outputs A and B are connected to microfluidic devices operating at flow rates of between 90 to 100 μL/h. In general, the flow from the liquid chromatography (LC) protein separation can be in the range of 10 μL/min 1 mL/min (600 μL/h−60 ml/h) depending on the pressure and column used and, therefore, the capillary resistances can be tuned for the desired flow splitting ratio. The flow sensors may be integrated into AKTA Pure system with an I/O-box E9 for real time flow monitoring. Stable flow splitting may be achieved by directing approximately 10% of the total flow to different parts of the microfluidic chip. The flow rates at the diffusional sizing and the electrophoresis device sample inlets are measured to be 40.0±0.7 μL/h and 37.4±0.7 μL/h, respectively.
The microfluidic device is designed to fit two distinct analytical parts in one fluorescence microscope field of view. By way of example only, one part contains the diffusional sizing device and comprises a long diffusion channel of a length of LD=43 mm, a width of WD=300 μm and a height of HD=55 μm. The positions for the diffusion profile acquisition are provided such that it may allow for a high sizing dynamic range and fixed to distances of 1.4 mm, 2.0 mm, 10.7 mm, 11.3 mm, 19.9 mm, 20.5 mm and 39.2 mm from the sample injection point. A degassed co-flow buffer (same as the LC mobile phase) is injected into the device at a flow rate of 150 μL/h, typically with the use of a neMESYS syringe pump (CETONI GmbH, Germany). The outlet of the microfluidic flow adapter is connected to the sample inlet on the diffusional sizing device. The diffusion profile of the injected sample is recorded and an analysis is performed via a fit to the numerical diffusion simulations.
The second part of the microfluidic chip is a free-flow electrophoresis device with one or more liquid electrodes. The free-flow electrophoresis device can be designed to create up to 60 V/cm transverse electric fields on the microfluidic chip whilst avoiding bubble formation and a build-up of electrolysis product(s) on the microfluidic chip. A conductive electrolyte solution e.g. 3 M KCl is injected into the free-flow electrophoresis device into an inlet from the side of the device at flow rate of 150 μL/h. The sample buffer solution is injected as a co-flow of the sample at a flow rate of 150 μL/h using a neMESYS syringe pump. The second output of the fluidic adapter is connected to the sample inlet of the free-flow electrophoresis device. The mobility of a charged particle is given by equation below:
where v is the transverse velocity, E—the applied electric field, x—the measured profile deflection, V—the applied voltage across the channel, d—the distance of the profile deflection measurement along the channel and h—the channel height. Hollow metal 1.5 mm ID electrodes are inserted into the liquid electrode channels on the sides of the device and connected to a power supply (EA Elektro-Automatik 6230207, Germany). The power supply is connected to the chip via a multimeter (Agilent 34410A, USA) recording a current flowing through the circuit.
The two microfluidic devices are operating continuously, and a measurement of the hydrodynamic radius, electrophoretic mobility and charge can be obtained for every 3.3 μL of the eluting sample (every 20 seconds) from the column while still fractionating 90% of the total volume.
The microfluidic devices can be fabricated using a standard polydimethylsiloxane (PDMS) soft-lithography approach. The master for the replica molding of PDMS is fabricated with an SU-8 photolithography process. After mixing PDMS (Sylgard184, Dow Corning, two components 10:1 ratio and degassed, mixed with carbon black nanopowder (Sigma-Aldrich) to create black devices) and casting it onto the photo-lithographically defined structure, it is cured at 70° C. for 1 h. Black devices can be advantageous as they can be used for minimizing background noise and the unwanted autofluorescence from PDMS under 280 nm-LED illumination during the measurements. Each PDMS replica of every master is then cut, and connection holes are made with a biopsy punch. To clean the PDMS devices they are sonicated for 3 min in isopropanol, blow dried with N2, and placed in an oven at approximately 70° C. for 10 min. The replica is activated using O2 plasma at a 40% power for 10 s (Diener etcher Femto, Germany) and bonded to a clean quartz slide (Alfa Aesar, 76.2×25.4×1.0 mm) for UV measurements or a simple glass slide for fluorescence measurements.
Various types of fluorescence microscopes can be used for the experiments as described in the present invention. For example, an intrinsic fluorescence microscope for a label free protein detection and a green label epifluorescence measurement setup can be provided. The autofluorescence measurements of proteins containing aromatic amino acids can be measured on an intrinsic fluorescence visualisation platform. Component in this platform may be made from quartz. Alternatively or additionally, components in this platform may be made from other glass-like and/or plastic materials with high transmittance and low autofluorescence for UV light. The proteins can be illuminated with a 25 mW 280 nm LED (M280L3, Thorlabs, UK) through an excitation filter (FF01-280/20-25, Semrock, USA) centred at a λex=280 nm and a dichroic mirror (FF310-Di01-25x36, Semrock, USA). The fluorescence from the sample may be collected through an emission filter (FF01-357/44-25, Semrock, USA) centred at λexm=357 nm, and focused onto an EMCCD camera (Rolera EMC2, QImaging, Canada).
The green epifluorescence microscope, optimised for Green Fluorescent protein (GFP)/Alexa488 detection, comprises a 490 nm LED (M490L4, Thorlabs, UK), an excitation filter at 482±9 nm, a dichroic mirror (350-488 nm/502-950 nm) and the emission filter at 520±14 nm (filter set MDF-GFP2, Thorlabs, UK). The microscope may have a xyz stage for accurate chip positioning in the field of view of a 2.5× objective, and the pictures can be taken with a CCD camera (Retiga R1, QImaging, USA).
A small delay between the molecule absorption measurement after the LC separation and the detection on chip may be possible. The delay volume from the absorption measurement cell to the flow adapter is approximately 70 μL and the volume from the flow adapter to the chip detection channel is around 8 μL causing 20 to 30 minutes delay time depending on the system flow. The time delay can be matched by comparing the absorption intensity on the absorbance detector (280 nm and 500 nm) of the LC and the fluorescence intensity of the eluting sample on the microfluidic chip.
As illustrated in the equation below, a voltage V0 can be applied to the electrophoresis device electrodes. Mobility measurements can be performed while the current flowing through the circuit I is recorded. The electrophoresis chamber of the device can be filled with a conductive electrolyte solution and the current I0 is measured whilst applying the same voltage V0.
V0=I(Relec+Rch),
V0=I0Relec,
⇒V=IRch=Relec(I0−I)=V0(1−I/I0).
The equation below is used to express the mobility of the sample
The total flow to the device Q=337 μL/h, V0=60 V, d=2880 μm, h=55 μm, I=0.267±0.002 mA, I0=0.283±0.001 mA.
The diffusion coefficient D quantifies the fluctuations of a particle under Brownian motion and is described for a spherical particle by the Stokes-Einstein equation as shown below:
where η is the viscosity of the solution, Rh is the hydrodynamic radius and kB and T are the Boltzmann constant and absolute temperature, respectively. The measured diffusion constant D and the electrophoretic mobility μe can be used to estimate the complex charge:
where κ—the inverse Debye length and f1 is a function of Rh that describes the effect of the electric field distribution around the particle. For most of the proteins in high salt buffers f1(κRh)≈1 since κRh<<1. Hence, the charge can be estimated by the Nernst-Einstein relation:
As shown in
Various further aspects and embodiments of the present invention will be apparent to those skilled in the art in view of the present disclosure.
“and/or” where used herein is to be taken as specific disclosure of each of the two specified features or components with or without the other. For example “A and/or B” is to be taken as specific disclosure of each of (i) A, (ii) B and (iii) A and B, just as if each is set out individually herein.
Unless context dictates otherwise, the descriptions and definitions of the features set out above are not limited to any particular aspect or embodiment of the invention and apply equally to all aspects and embodiments which are described.
It will further be appreciated by those skilled in the art that although the invention has been described by way of example with reference to several embodiments. It is not limited to the disclosed embodiments and that alternative embodiments could be constructed without departing from the scope of the invention as defined in the appended claims.
Number | Date | Country | Kind |
---|---|---|---|
1815360.1 | Sep 2018 | GB | national |
1910277.1 | Jul 2019 | GB | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/GB2019/052635 | 9/19/2019 | WO | 00 |