1. Field of the Invention
The field of the invention relates to an inhalation anaesthesia delivery systems generally, and more particularly to an inhalation anaesthesia delivery system comprising a fresh gas feeding arrangement connected to a breathing circuit, a monitor device, a control device and an interface unit, whereby the fresh gas feeding arrangement and the ventilator are configured to deliver a desired concentration of gas to the breathing circuit, and to a method of operating the same.
2. Description of Prior Art
The inhalation anaesthesia delivery systems are used to maintain oxygen (O2) and carbon dioxide (CO2) exchange of the patient during anaesthesia. During inspiration oxygen enriched gas flows into the lungs where it is diffused into circulation. At the same time CO2 is diffusing from the circulation into the lungs. During expiration the oxygen depleted and CO2 enriched gas flows out from the lungs. Inspiration and expiration together form a breath. The amount of delivered gas in single breath is called tidal volume. Inspiration may be either spontaneous action of the patient or artificial when a ventilator pressurizes the lungs with fresh gas. Expiration is spontaneous in both cases and effected by lung elastic forces.
Breathing gas is often mixed with anaesthesia gases to provide The inhalation anaesthesia. These gases are nitrous oxide (N2O) or air, and volatile anaesthetics. Patient concentration determines gas exchange status (O2, CO2) and the depth of anaesthesia of the patient. Normal range for N2O is 30%-75% and for volatile agents depending on the agent from 0.7% (halothane) to 3% (sevoflurane, enflurane, isoflurane) and up to 6%-12% with desflurane. CO2 concentration is typically about 5% and O2 concentration 25%-75%. Patient concentration is the best measured as end-expiration (=end-tidal) breathing gas concentration. However, oxygen delivery is often measured with inspiration gas O2 concentration.
Anaesthesia is delivered using rebreathing circuit where expired gas is circulated after removal of carbon dioxide and adding fresh gas back to inspiration. The proportion of circulated gas increases with decreasing fresh gas flow. To save anaesthetic gases, the fresh gas flow is minimized. In low flow, minimal flow, and closed circuit anaesthesia the circulated gas conforms the majority of the new breath. During anaesthesia, oxygen delivery is fitted with patient oxygen demand. In case delivery is low compared to demand patient oxygen reservoir is emptying and vice versa. Difficulty to control the oxygen reservoirs arises in low fresh gas flows where the system time constant for the control may be measured in tens of minutes. This can be shortened with temporary major increase of the fresh gas flow. Similar increase boosts also anaesthetic agent delivery change when required, but if this is not expected, vaporizer setting needs to be compensated for the fresh gas flow change. These back-and-forth controls made manually grab the attention of the anaesthesiologist from the patient. Recent development asks for increase in safety and efficacy, which promotes automatic control loops controlling the breathing actuators in response to the measured values. Increased automation free up the human resources in operating room to concentrate on patient instead of the machine, or even take care of more patients at the same time.
The filling grade of body oxygen reservoirs can be measured with end-expiration gas oxygen concentration (EtO2), arterial blood hemoglobin oxygen saturation (SpO2), or arterial blood oxygen partial pressure (PaO2). EtO2 and SpO2 are continuous non-invasive measurements whereas the PaO2 is discrete and minimally invasive. SpO2 is insensitive on changes when the blood hemoglobin oxygen reservoir is filled and gives a delayed response only when that reservoir is already emptying. EtO2 reflects the status of all body oxygenation reservoirs including the blood hemoglobin saturation and dissolved oxygen content in blood. However, EtO2 suffers of the problem that some lung ventilation/blood perfusion mismatch conditions may disturb the connection between the measured EtO2 and body oxygen storage status. The same may occur in case the circulatory status of the patient becomes disturbed e.g. due to reduced heart pumping action. As a result of such disturbances, EtO2 measurement may overestimate the filling grade of the body oxygen reservoirs.
Expired gas control principle is not a novelty regarding anaesthesia gas delivery. Preliminary research concepts to control volatile agent vaporizers based on measured end-tidal concentrations have been presented already on 1980'ies and is today commercially available. However, known O2 control systems match either inspired gas O2 concentration US 2005/0103338A1; US2006/0090757) or patient blood measured hemoglobin (Hb) oxygen saturation (SpO2) (US 2005/0109340; U.S. Pat. No. 6,761,165) with respective setting. Disadvantage of the inspired gas control is vague correlation with patient oxygenation state, specially during instable situations where the patient O2 level is changed. Actually, O2 inspired-to-expired difference remains even at steady state whereas the difference disappears with anaesthesia gases. Controlling SpO2 involves a problem of poor sensitivity when patient oxygenation is normal, i.e. SpO2 is above 95%. On the other side, below 90% body O2 reserves are already depleted and SpO2 changes rapidly in any changes in body oxygenation state. Time constant to affect the SpO2 by changing the O2 delivery is counted at minimum tens of seconds and responding this severe situation becomes thus delayed. The problems in SpO2 rely back on S-shaped Hb saturation dissociation curve describing the saturation on the ordinate and blood O2 partial pressure on abscissa. EtO2 is closely correlated with blood O2 partial pressure and is easily measurable at the point of care with any fast responding O2 sensor capable to separate inspiration and expiration concentrations.
An embodiment of the invention provides a system by which the disadvantages of the prior art can be eliminated. This is obtained with the invention. The system of the invention is characterized in that the monitor device is configured to monitor expired inorganic gas oxygen concentration by comparing the measured expired breathing gas concentration with the set target value and changing the fresh gas delivery accordingly to meet the target value. The method of the invention is characterized in that expired inorganic gas concentration is monitored by comparing the measured expired breathing gas concentration with the set target value and the fresh gas delivery is changed accordingly to meet the target value.
An advantage of the invention is in its simplicity and ability to produce reliable results, i.e. end tidal oxygen (EtO2) is closely correlated with blood O2 partial pressure and is easily measurable at the point of care with any fast responding O2 sensor capable to separate inspiration and expiration concentrations.
As described above in this invention expired gas mixture concentration is controlled automatically with a control system comparing the measured end-tidal concentration with the target value and changing the fresh gas composition accordingly to meet the target. Expired gas mixture is measured for O2, N2O, CO2, and AA concentrations. User selects the gas delivered with O2 from the two of N2O and air and the end-tidal target concentrations for O2 and anaesthetic agent. Control device determines from the differences between the measured and target values set the demand to change fresh gas composition. When the EtO2 concentration change is required, fresh gas flow rate is increased and the concentration is adjusted respectively to meet the demanded change. At this time, if anaesthetic concentration change is not expected, vaporized setting is dialled down. When the measured O2 concentration approaches the target, fresh gas flow is decreased to close the circuit for savings in anaesthetics, and vaporizer setting is increased to compensate for this. Respectively, change in anaesthetic agent (AA) concentration is accompanied with increased fresh gas flow to boost the AA change while controlling the delivered fresh gas mixture to preserve the end-tidal O2 concentration. In some cases, user may want to have also control over the speed of concentration change. For this purpose, a user control for maximum fresh gas flow may be provided.
To safeguard the control delivery system against the potential disturbances on connection between EtO2 and body oxygen reservoir filling grade, the control system is backed up with SpO2. For this purpose, user gives the minimum SpO2 the controller allows for. Decreasing below this minimum level oxygen delivery is automatically increased despite of the matching of target and measured end-tidal oxygen concentrations. This oxygen delivery increase may be effected by automatically increasing the EtO2 target value.
To use this invention, the operator selects the gas to be mixed with O2, and gas concentrations. Because the fresh gas is mixture of O2, secondary gas (N2O, N2 or xenon), and volatile agent, target Et concentrations may be given to any two of these gases. Advantageously the given target concentrations are end-tidal values for oxygen and volatile agent the secondary gas conforming the rest. Given also the secondary gas, the controller adjusts O2 flow, secondary gas flow, and vaporizer setting to meet the target values. The flows are adjusted within the limits of total fresh gas flow that may also be given to limit the rate of change in gas compositions, that may be used to titrate the correct end-tidal settings. When changing the secondary gas from one to another, the controller needs to control the concentrations of all three gases to replace the previous secondary gas with the new one from the breathing circuit. Finally, when the measured end-tidal values match with the target values, fresh gas flow is gradually reduced to minimize the gas, especially volatile agent, consumption. For full performance of the oxygen control, the user gives also the minimum allowed SpO2 value.
In the following description, embodiments of the invention will be described in greater detail by means of examples shown in the attached drawings, in which:
Alternatively patient may be breathing spontaneously. In spontaneous breathing the ventilator comprises reservoir collecting the exhalation gas and therefrom patient breathing action receives inspiration gas.
Monitor device 15, i.e. gas monitor may be of a side-stream type drawing a sample gas stream from the sampling port 10 through sampling line 16 for analysis with the sensors within the monitor. Alternatively the monitor device may be of mainstream type where the gas analysis sensors are located directly at the patient limb instead of the sampling port.
Monitor device 15 is further electrically connected to control device 17, which is further connected to the actuators (gas regulating unit 13, and/or vaporizer 14) closing the control loop. This controller compares the measured values with the user set target and tunes the actuators to match the measured values with the setting. The user can set the target by using an interface unit 18.
To control the end-tidal gas mixture, the control device needs to increase the fresh gas flow rate as a response to differences in end-tidal to target gas concentration. This comparison takes into consideration the differences between:
Measured Et O2−target Et O2;
Measured Et AA−target Et AA;
3a) N2O selected as a secondary gas: N2 concentration target is zero. Direct N2 concentration measurement is often not available, thus that needs to be calculated using the other gas mixture components as EtN2=100−EtO2−EtCO2−EtN2O−EtAA.
3b) N2 (air) selected as secondary gas: N2O concentration target is zero.
Fresh gas flow demand is determined for each of these differences. This determination advantageously increases fresh gas flow demand as a response to increased difference to boost the concentration change speed. Respectively decreased difference would end up to decreased flow demand to save the gases. Such response can be achieved e.g. with linear relationship equation between the flow demand and difference. The final target fresh gas flow value is determined from the comparison results 1-3. As an example, the target fresh gas flow can be the largest of the flows 1-3. Providing possibility to limit fresh gas flow maximum, user may control the concentration change speed, which is useful in titrating the optimum fresh gas mixture. Respectively, minimum fresh gas flow can be given to assure minimum breathing circuit clearance from patient- or system derived gas compounds that may accumulate in closed circuit.
Now, using the fresh gas flow value and difference 1), fresh gas oxygen concentration is calculated. In this context the fresh gas flow value may be either the measured or target value. Respectively, for anaesthesia gas a new fresh gas AA concentration is calculated from difference 2) and the flow.
Finally, using the fresh gas flow and fresh gas oxygen concentration, the target settings for the fresh gas oxygen flow and secondary gas flow is calculated. This final calculation is straightforward when having N2O as secondary gas, but with air the mixture content of 21% O2 has to be taken into consideration in this final flow calculation. The calculated flow values are then sent to the gas mixer and vaporizer for new settings. Alternatively, if the gas mixer operates with the principle of total flow & O2 % control, then these are used for respective target settings instead of the flows.
The embodiment of the invention described above is by no means intended to restrict the invention but the invention can be varied completely freely within the scope of the claims. Therefore it is obvious that the system or its details do not necessary have to be exactly the same as shown in the figures but other solutions are possible, too.
Number | Date | Country | Kind |
---|---|---|---|
07102935.9 | Feb 2007 | EP | regional |