The present invention relates generally to 3D imaging, and more particularly to 3D imaging of the human breast for medical screening and diagnostic purpose.
Breast cancer affects many women worldwide. When caught early is often treatable, but current screening and diagnostic tools have considerable drawbacks. X-ray mammography Requires painful compression of the breast while imaging, uses ionizing radiation, and has poor sensitivity in women with dense breasts, who are most at risk for breast cancer. In the alternative case of traditional sonography, early-stage cancers are often missed, and image quality notably depends on the skill of the clinician. Magnetic resonance imaging (MRI) is expensive, time-consuming to operate, and is characterized by poor specificity, leading to unnecessary further tests or treatment.
Wavefield imaging offers a promising solution to overcome such shortcomings of the conventional screening and diagnostic tools. Wavefield imaging can be used to create a 3D image of physical properties of an object of interest, given a limited amount of data collected by sensors placed outside the object. In the case of microwave imaging (MWI), antennas illuminate the target and collect electromagnetic field data, and the aim is to reconstruct complex-valued permittivity of the object of interest. In the case of ultrasound imaging (USI), the target object is surrounded by piezoelectric transducers, and the aim is to reconstruct acoustic speed and/or attenuation.
Breast imaging using microwaves has been studied for several decades. It is well accepted that MWI lacks the resolution required for early detection of breast tumours, although quantitative MWI does have better specificity than some other modalities. Achieving accurate quantitative images of the complex-valued permittivity of the breast and improving the resolution remains an active area of research. One approach to enhancing microwave image quality (accuracy and resolution) is to incorporate information obtained from a second, complementary imaging modality into the microwave (MW) inversion algorithm. The integration of multi-modal imaging data may be accomplished via joint inversion, where data from the two imaging modalities are inverted simultaneously, or sequential inversions, where information about the object of interest (Ol) is extracted from the first modality and used as prior information in the second inversion. Several studies have looked at methods for extracting prior information, for use in MWI, from other imaging modalities as well as integrating this information into inversion algorithms.
Quantitative USI for the breast is also an active area of research in its own right. USI uses smaller wavelengths, allowing for higher resolution reconstructions, and Quantitative USI can also provide improved specificity compared to traditional modalities. Some clinical quantitative USI systems for breast imaging exist, but these are expensive and highly specialized.
Accordingly, there remains a need for a wavefield breast imaging solution that at least partially addressing the foregoing shortcomings of the prior art.
According to a first aspect of the invention, there is provided a multi-modal imaging system comprising:
According to a second aspect of the invention, there is provided a method of imaging of an imaging target, said method comprising:
Preferred embodiments of the invention will now be described in conjunction with the accompanying drawings in which:
Preferred embodiments of the present invention adopt the aforementioned sequential inversion approach to multi-modal imaging, where information about the object of interest (Ol) is extracted from the first modality and used as prior information in the second inversion. More specifically, such embodiments utilize prior information, in the form of relatively low-resolution reconstructions derived from a relatively simple USI system that is combined with a known MWI system. In contrast, the other aforementioned approach, in which the MW and ultrasound (US) data are simultaneously inverted, generally requires that more extensive US data be obtained in order to match the resolution on the single reconstruction mesh. In some embodiments, as is the case of the prototyped embodiment, the inversion algorithm is the finite-element method contrast-source inversion (FEM-CSI) technique that allows prior information to be included as either an initial guess or as an inhomogeneous numerical background, as described in below cited references [16]-[19]. The US sound-speed reconstructions use a ray-based technique with a polynomial basis approximation, per below cited reference [20].
The US fixture 12 of the illustrated embodiment is a multi-piece assembly composed of two hollow semi-cylindrical pieces 22A, 22B plus a semi-spherical cap 24. The US fixture 12 is shown exploded in an unassembled state in
In the assembled state of the US fixture 12, the semi-spherical fixture cap 24 forms an internally concave, closed bottom end of the assembled US fixture 12, thus capping off the bottom end of the fixture’s upright cylindrical wall 28. The assembled US fixture 12 houses an array of single-element piezoelectric transducers, schematically illustrated at 36, of which a respective subset of these transducers 36 is possessed by each of the thee fixture components 22A, 22B, 24. In the prototyped embodiment, there are sixty-four piezoelectric transducers 36, arranged in a helical fashion around a central longitudinal axis of the assembled US fixture 12, though the transducers 36 are shown in simplified schematic form without accurate reproduction of such helical layout. In use of the inventive multi-modal imaging apparatus 10, the multi-piece US fixture 12 is easily taken apart to remove the fixture from the imaging target (e.g. human breast) after US data collection. An accompanying US data measurement system (schematically shown in
The US fixture 12 is designed to fit snugly around a flexible breast-support cup 38, which is shown in cross-section and in exploded from its normally installed position in
The breast cup 38 is installed in a patient table 40, or other comparable type of patient support, atop which a patient may lie in a prone position. The patient table 40 has a hole 42 therein that penetrates through the patient table 40 from a topside thereof to an opposing underside thereof. The breast cup 38 is installed at the hole 42 of the patient table 40, from which the installed breast cup 38 hangs down from the underside of the patient table. In use, the breast of a patient lying prone atop the table 40 is received within the breast cup 38 and through the hole 42 in the table.
The MWI chamber 14 resides beneath the patient table 40 in aligned relation to the breast cup 38 and the patient table hole 42 at which the breast cup 38 is installed. The MWI chamber 14 is movable relative to the patient table 40, and more particularly is movable relative thereto into and out of a position of surrounding working relation to the breast cup 38 for the purpose of imaging the patient breast received therein. In the illustrated embodiment, the MWI chamber 14 is linearly displaceable upwardly and downwardly on an upright longitudinal axis AL shared by the MWI chamber 14, breast cup 38 and patient table hole 42 by one or more linear actuators 44. In the illustrated example, the push rod 44A of each linear actuator 44 is attached to an out-turned flange 46 of the MWI chamber 14 of radially projecting relationship to the chamber wall 20, for example at a top end thereof. In this example, the actuator(s) 44 thus perform(s) direct lifting and lowering of the MWI chamber 14, but it will be appreciated that the linear actuator(s) 44 or other type of actuation means may alternatively perform lifting and lowering of the MWI chamber 14 via indirect connection thereto, for example by lifting and lowering a chamber support platform atop which the MWI chamber 14 is mounted.
With the MWI chamber 14 retracted in
Throughout the US and MW measurement procedures, and throughout the transition of the MWI chamber 14 and US fixture 12 into and out of their respective working positions surrounding the breast cup 38, the table-supported breast cup 38, and the patient’s breast received therein, remain statically positioned. With reference to
While the forgoing paragraphs describe working placement of the MWI chamber 14 and taking of the MW measurements as occurring prior to the working placement of the US fixture 12 and taking of the US measurements, it will be appreciated that the order may be reversed. While the illustrated embodiment features machine powered lifting and lowering of the MWI chamber, for example using linear actuators or other powered lifting means, and lacks any powered means of placing and removing the US fixture components 22A, 22B, 24 into and out of their working relationship to the breast cup 38, it will be appreciated that other degrees or combinations of powered actuation and/or human performed placement of the MWI chamber 14 and US fixture 12 may alternatively be employed. For example, fixture-manipulating actuators may be attached the underside of the patient table 40, or otherwise supported therebeneath, for the purpose of laterally sliding the two semi-cylindrical pieces 22A, 22B together from initially withdrawn positions that are situated radially far enough from the breast cup 38 to reside outside the MWI chamber 14 when raised. The bottom cap 24 of the US fixture 12 may likewise be placeable by one or more actuatros into mating relationship with the bottoms 32 of the mated semi-cylindrical pieces 22A, 22B from an initially withdrawn position of likewise non-obstructive relationship to the raising and lowering of the MWI chamber 14, for example being laterally slid into aligned relation under the bottom ends of the semi-cylindrical pieces 22A, 22B by a respective fixture actuator.
A notable feature of the dual-modal imaging apparatus 10 detailed above is that the US fixture 12 is easily installed and removed, by coupling and decoupling of component pieces of the US fixture into and out of snugly surrounding relation to a statically supported breast cap, thereby denoting an ability to obtain scattered-field data from the MW and US measurement systems that, as schematically shown in
In a non-limiting example reflective of the prototyped embodiment of the present invention, the arbitrary waveform generator may be comprised of four PC-based waveform generator cards (e.g. Signatec PXDAC4800), each with four channels, and a pulsed sinusoid truncated after five periods may be used to drive the US transducers at their observed resonant frequency of 1.4 MHz. The oscilloscope may be, for example, a 32-channel digital oscilloscope, composed of four PC-based oscilloscope cards (e.g. GaGe Octopus 8387 CompuScope digitizer boards), each with 8 channels. The US transducers 36 (e.g. from Sonometrics Corporation, London, Canada) may be single-element piezoelectric transducers, whose elements consists of a cylindrical lead zirconate titanate (PZT-5H) crystal, connected to a twisted pair of wires and dipped in epoxy to provide better coupling to the background material, assumed to be water or similar. The twisted pair may be surrounded by a ground shield and encased inside a waterproof coating. The usable bandwidth of the transducers may be approximately 1.0-1.8 MHz.
In place of the waveform generator 102 and oscilloscope 104 found in the US measurement system, the MW measurement system 200 instead features a vector network analyzer 202, whose ports are connected to the MW field probe antennae 16A of the MWI chamber 14 via another switch 206. Via execution of an MW measurement algorithm of the control software on the computer 300, e.g. via selection of an MW measurement start command in the GUI once the MWI chamber 14 is in the working position (with the US fixture 12 absent from the breast cup 38), the field probe antennae 16A illuminate the target object (the patient breast inside the breast cup 38) with electromagnetic energy in the microwave frequency range, typically in the range of 0.3 - 9 GHz for biomedical applications, and field data are collected from all antenna positions in the MWI chamber, and stored by the computer 300.
The MWI chamber may feature forty-four facets, approximating a hemi-ellipsoidal shape, with S-parameter measurements being made using the magnetic field probes (half-loop antennas) installed on 24 of the facets. The antennas may be connected via coaxial cables to a 24-port vector network analyzer (VNA), for example comprised of four 6-port Keysight PXle VNA modules (Keysight Technologies, Santa Rosa, CA, USA). The air-based MWI chamber 14 simplifies registration of the target object (patient breast) between the two imaging systems, as the variable buoyancy of breasts can lead to unpredictability of the exact position and shape of the breast in a fluid-filled chamber.
Before an MW inversion algorithm of the control software 300 is executed, an intermediate mapping step is performed, in which the reconstructed values of acoustic speed obtained from the US inversion algorithm are mapped to complex permittivity values, different mapping techniques may be used, for example segmented mapping, linear mapping or tissue-range mapping. After such mapping has been performed, the MW inversion algorithm is then executed to reconstruct the complex-valued permittivity, related to the dielectric constant and loss, of tissues within the imaging domain. This MW inversion algorithm takes as input both the stored MW measurement data set as well as the mapped permittivity values derived from the US inversion algorithm, the latter of which is the “prior information″” that links the USI and MWI modes of the inventive dual-mode system. The US-derived prior information may be used by the MW inversion algorithm in different ways, for example as an “initial guess” or as an “inhomogeneous numerical background”. The resultant output from the MW inversion algorithm is a final image, which shows the complex-valued permittivity inside the target object (breast), and can be displayed to the operator in the GUI. While the illustrated embodiment of
In the prototype, the patient table 40 was substituted for a smaller support rested atop the out-turned flange 46 at the top of the MWI chamber 14, by which the breast cup 38, and the assembled US fixture 12, when present during the US measurement stage, could be suspended within the MWI chamber 14. In this prototype apparatus, the MWI chamber 14, instead of being linearly displaced relative to the resting support, remained static, with reliance being made on manual lifting of the resting support (and attached breast cup 28 and US fixture 12) from off the flanged top of the MWI chamber 14 to enable installation and removal of the US fixture 12. Testing was conducted using a breast phantom consisting of four distinct tissue regions: skin, fat, fibroglandular and tumour, with the tumour embedded inside the fibroglandular region. For each tissue type, both the complex-valued permittivity and acoustic sound-speed of the phantom material were designed to approximate the values of real breast tissue. The skin region was a 2 mm thick, 10 cm diameter cylindrical shell with a hemispherical cap. The shell was made from a graphite-urethane mixture, and its total height was 9.64 cm. Construction details are available in below cited reference [21]. The shell had a measured relative permittivity of 11-j1.2 at 1 GHz. The fat region consisted of canola oil, having a measured relative permittivity of 2.9-j0.23 at 1 GHz and sound-speed of 1463 m/s at 1.4 MHz, the center frequency of the US transducers 36. The fibroglandular and tumour regions were composed of gelatin-based mixtures. The recipes were derived from the acoustic breast phantom described in below cited reference [22], modified so that the phantoms also exhibit dielectric properties representative of the respective breast tissue. The fibroglandular region was composed of water, gelatin, agar, glycerin, and n-propanol, whereas the tumour consisted of water, gelatin, agar, glycerin and table salt. The liquid mixtures were heated just until boiling, and left to set inside custom molds. The fibroglandular region was asymmetric, as depicted in Applicant’s related provisional application, incorporated herein by reference. The tumour is ellipsoidal in shape, with diameters of 2 and 2.5 cm. The relative permittivities of the fibroglandular and tumour regions at 1 GHz were 41-j8.4 and 71-j19, respectively, and measured sound-speeds are 1595 and 1587 m/s, respectively. Note that the tumour and fibroglandular sound-speeds are approximately the same, considering the accuracy of the present USI system. This is contrary to the fact that the sound-speed of tumour tissue is known to be higher than that of fibroglandular tissue. Photographs of the phantom components are shown in Applicant’s related provisional application, in which the relative complex permittivities of the fat, fibroglandular and tumour-mimicking materials are also plotted as a function of frequency.
The US data was collected first, with the US fixture 12 holding the US transducers 36 in positions making direct contact with the skin of the phantom, which doubled as the breast support cup in the experimental setup. Good contact was ensured by using ultrasonic coupling gel, which is preferably also used on the exterior of the breast support cup in preferred working embodiments of the present invention. The US data consists of time-domain transmission pulse waveforms from which time-of-flight (TOF) between each Tx/Rx pair is extracted. For the results presented herein, a pulsed sinusoid having a frequency of 1.4 MHz truncated after five periods was used. The TOF data was used within a ray-based reconstruction algorithm that utilizes whole-domain polynomial basis functions of variable order to approximate the sound-speed within the breast. The whole breast was used as the imaging domain. Details of the reconstruction algorithm are provided in below cited reference [20]. Once the US time-domain data was acquired, the US fixture 12 was carefully removed, and frequency-domain MW data was acquired at several frequencies between 1-2 GHz.
To invert the MW data and reconstruct an image of the complex-valued permittivity within the breast phantom, the FEM-CSI algorithm was used. Although there are several methods that can be used to integrate the prior (in this case ultrasound) information into FEM-CSI reconstructions, two have been prevalent in past work: a) using it as an initial guess, or b) introducing it as a numerical background that changes the contrast being sought. The latter technique was implemented in a previous MWI breast phantom study, published in below cited reference [16]. Both require that the reconstructed sound-speed values be converted to complex-valued permittivity. There are various ways of performing this mapping, which is the subject of ongoing study. Below are provided initial results of using the prior information as both an initial guess, and as an inhomogeneous background.
US reconstruction was obtained using 5th order basis functions, resulting in creation of permittivity maps, which are shown in Applicant’s related provisional application. The conversion from sound-speed to permittivity was accomplished using a tissue-range mapping technique similar to that published in below cited reference [23]. For an MW-only case, single-frequency results at 1.0 and 1.3 GHz were obtained. For these cases, perfect prior information (both geometric and permittivity) of the skin and fat regions were incorporated into the inversion algorithm as a numerical inhomogeneous background. No prior information related to the fibroglandular region was assumed. These MW reconstructions (without US-derived prior, but with perfect knowledge of the skin and fat) were created, and are also shown in Applicant’s related provisional application. The imaging domain was limited to the fat region. These results were quite poor given that no fibroglandular prior was included, as was expected from the study performed in below cited reference [16]. In that prior work, tumour detection was shown to be quite good with the inclusion of perfect fibroglandular prior. Of course, the purpose in the testing context of the present invention was to determine just how “perfect” the fibroglandular prior must be, that is, using imperfect prior obtained from the coarse US reconstructions.
The dual-mode inversion results using the US-derived initial guess are shown in Applicant’s related provisional application, along with those using the US-derived background. It is noted that the scales of the colour bars in the permittivity reconstructions of the provisional figures are not all the same: the MW-only reconstructions achieved a peak-value that was considerably lower than the true value. From the results, it was clear that the MWI produces some structure within the US fibroglandular prior that was supplied. When the prior was incorporated as an inhomogeneous numerical background, there seemed to be more irregular structures and, in addition, it was found that the permittivity values saturated to the maximum value that was imposed as a constraint in the FEM-CSI inversion. On the other hand, when using the prior as an initial guess, the permittivity values within the fibroglandular region still did not reach the measured values of the tumour. So, it was concluded that there was “undershooting” in the initial guess case, and “overshooting” in the inhomogeneous background case. From the obtained images, it was unclear which of the two techniques provides better tumour detection, but it was reasonable to say that either provides better detection compared to when no prior is provided.
At the time of testing, a systematic tumour detection algorithm, such as one of the techniques applied in below cited reference [16], had not yet been implemented for the dual-mode results. Applicant’s related provisional application includes image examples of tumour detection using a simple thresholding technique for one set of parameters that produced better reconstructions; that is, choosing over the reconstructions obtained with different inversion parameters: inversion frequency and the order of the US basis functions used. Before the images were thresholded, artefacts at the top of the image were manually removed. Such artefacts were observed to be typical in the sound-speed reconstructions, and thus could be safely removed when present. In the thresholded images found in Applicant’s related provisional application, the threshold value was set to 96% of the peak value in the reconstruction.
The test results were clear: that the use of the US prior information, as either an initial guess or as a background, has resulted in an improved detectability of the tumour. The results presented demonstrated the ability of the novel, dual-mode microwave-ultrasound system to reconstruct quantitative images using a preliminary attempt at a realistic dual-mode breast phantom.
Since various modifications can be made in my invention as herein above described, and many apparently widely different embodiments of same made, it is intended that all matter contained in the accompanying specification shall be interpreted as illustrative only and not in a limiting sense.
The following references, and any and all other references cited or mentioned elsewhere herein, are incorporated herein by reference in their entirety.
J.-C. Bolomey, “Crossed viewpoints on microwave-based imaging for medical diagnosis: From genesis to earliest clinical outcomes,” in The World of Applied Electromagnetics. Springer, 2018, pp. 369-414.
B. M. Moloney, D. O’Loughlin, S. Abd Elwahab, and M. J. Kerin, “Breast cancer detection-a synopsis of conventional modalities and the potential role of microwave imaging,” Diagnostics, vol. 10, no. 2, p. 103, 2020.
P. Mojabi and J. LoVetri, “Experimental evaluation of composite tissue-type ultrasound and microwave imaging,” IEEE Journal on Multiscale and Multiphysics Computational Techniques, vol. 4, pp. 119-132, 2019.
D. O’Loughlin, M. O’Halloran, B. M. Moloney, M. Glavin, E. Jones, and M. A. Elahi, “Microwave breast imaging: Clinical advances and remaining challenges,” IEEE Transactions on Biomedical Engineering, vol. 65, no. 11, pp. 2580-2590, 2018.
X. Song, M. Li, F. Yang, S. Xu, and A. Abubakar, “Study on joint inversion algorithm of acoustic and electromagnetic data in biomedical imaging,” IEEE Journal on Multiscale and Multiphysics Computational Techniques, vol. 4, pp. 2-11, 2019.
Y. Qin, T. Rodet, M. Lambert, and D. Lesselier, “Joint inversion of electromagnetic and acoustic data with edge-preserving regularization for breast imaging,” IEEE Transactions on Computational Imaging, vol. 7, pp. 349-360, 2021.
G. Boverman, C. E. Davis, S. D. Geimer, and P. M. Meaney, “Image registration for microwave tomography of the breast using priors from nonsimultaneous previous magnetic resonance images,” IEEE journal of electromagnetics, RF and microwaves in medicine and biology, vol. 2, no. 1, pp. 2-9, 2017.
D. Kurrant and E. Fear, “Regional estimation of the dielectric properties of inhomogeneous objects using near-field reflection data,” Inverse Problems, vol. 28, no. 7, p. 075001, 2012.
M. Omer, P. Mojabi, D. Kurrant, J. LoVetri, and E. Fear, “Proof-of-concept of the incorporation of ultrasound-derived structural information into microwave radar imaging,” IEEE Journal on Multiscale and Multiphysics Computational Techniques, vol. 3, pp. 129-139, 2018.
N. Abdollahi, D. Kurrant, P. Mojabi, M. Omer, E. Fear, and J. LoVetri, “Incorporation of ultrasonic prior information for improving quantita-tive microwave imaging of breast,” IEEE Journal on Multiscale and Multiphysics Computational Techniques, vol. 4, pp. 98-110, 2019.
P. Mojabi, N. Abdollahi, M. Omer, D. Kurrant, I. Jeffrey, E. Fear, and J. LoVetri, “Tissue-type imaging for ultrasound-prior microwave inver-sion,” in 2018 18th International Symposium on Antenna Technology and Applied Electromagnetics (ANTEM), 2018, pp. 1-3.
N. Duric, P. Littrup, L. Poulo, A. Babkin, R. Pevzner, E. Holsapple, O. Rama, and C. Glide, “Detection of breast cancer with ultrasound tomography: First results with the computed ultrasound risk evaluation (cure) prototype,” Medical physics, vol. 34, no. 2, pp. 773-785, 2007.
N. Duric, P. Littrup, and C. Kuzmiak, “Breast ultrasound tomography,” Breast Imaging, vol. 6, 2018.
N. Duric, M. Sak, S. Fan, R. M. Pfeiffer, P. J. Littrup, M. S. Simon, D. H. Gorski, H. Ali, K. S. Purrington, R. F. Brem et al., “Using whole breast ultrasound tomography to improve breast cancer risk assessment: a novel risk factor based on the quantitative tissue property of sound speed,” Journal of clinical medicine, vol. 9, no. 2, p. 367, 2020.
K. Nemez, A. Baran, M. Asefi, and J. LoVetri, “Modeing error and calibration techniques for a faceted metallic chamber for magnetic field microwave imaging,” IEEE Transactions on Microwave Theory and Techniques, vol. 65, no. 11, pp. 4347-4356, 2017.
M. Asefi, A. Baran, and J. LoVetri, “An experimental phantom study for air-based quasi-resonant microwave breast imaging,” IEEE Transactions on Microwave Theory and Techniques, vol. 67, no. 9, pp. 3946-3954, 2019.
I. J. Amer Zakaria, J. LoVetri, and A. Zakaria, “Full-vectorial parallel finite-element contrast source inversion method,” Progress In Electromagnetics Research, vol. 142, pp. 463-483, 2013.
D. Kurrant, A. Baran, J. LoVetri, and E. Fear, “Integrating prior information into microwave tomography part 1: Impact of detail on image quality,” Medical physics, vol. 44, no. 12, pp. 6461-6481, 2017.
D. Kurrant, E. Fear, A. Baran, and J. LoVetri, “Integrating prior information into microwave tomography part 2: Impact of errors in prior information on microwave tomography image quality,” Medical physics, vol. 44, no. 12, pp. 6482-6503, 2017.
M. Hughson, “Quantitative transmission tomography for non-destructive imaging of stored grain and biological tissue,” Master’s thesis, University of Manitoba, 2021.
J. D. Garrett, “Average dielectric property analysis of non-uniform structures: Tissue phantom development, ultra-wideband transmission measurements, and signal processing techniques,” Master’s thesis, University of Calgary, 2014.
L. Medina-Vald′es, M. P′erez-Liva, J. Camacho, J. Ud’ias, J. Herraiz, and N. Gonz′alez-Salido, “Multi-modal ultrasound imaging for breast cancer detection,” Physics Procedia, vol. 63, pp. 134-140, 2015.
C. Kaye, I. Jeffrey, and J. LoVetri, “Improvement of multi-frequency microwave breast imaging through frequency cycling and tissue-dependent mapping,” IEEE Transactions on Antennas and Propagation, vol. 67, no. 11, pp. 7087-7096, 2019.
This application claims benefit under 35 U.S.C. 119(e) of U.S. Provisional Application No. 63/325,586, filed Mar. 30, 2022, the entirety of which is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
63325586 | Mar 2022 | US |