The present invention relates generally to medical devices. More particularly, the present invention relates to intraoperative non-lifting peripheral nerve action potential recording.
Intraoperative nerve action potential (NAP) or compound nerve action potential (CNAP) recording is a proven useful tool to guide surgeon's decisions about surgical approaches during various nerve repairs. Using this tool in combination with other assessments, the surgical team can determine the exact location of a nerve lesion. Clinical data from retrospective studies have shown that the surgical outcome is improved by utilizing this tool to select the repair method.
Intraoperative NAP recording can be technically challenging although the current methods are simplified by the use of handheld electrodes applied to surgically exposed nerves. The typical recommendation is to lift the nerve out of the surgical field during testing. After dissection and isolation, a surgically exposed nerve is typically lifted by two pairs of electrodes (one for recording and another for stimulation) and the nerve segment between the two electrodes is often suspended with no direct contact with tissue underneath. The main limitations noted are the length of the exposed nerve segment and the presence of a large stimulus artifact. Despite these major challenges, no fundamental understanding and/or improvement of the methods have been reported. Despite considerable improvements in hardware and software employed for neurophysiological recordings, the new generation of IONM (intraoperative neurophysiological monitoring) machines do not resolve the existing technical challenges of NAP recordings.
Accordingly, there is a need in the art for a device and method for intraoperative non-lifting peripheral nerve action potential recording.
The foregoing needs are met, to a great extent, by the present invention which provides a system and method including an electrode for stimulation or recording of a compound nerve action potential from surgically exposure peripheral nerve. The electrode includes a number of insulated metal prongs, such that the nerve can be sandwiched between the prongs, without lifting the nerve. The prong has an area for contact for nerve stimulation or recording. The electrode also includes an elastic base or body, configured to provide the nerve with a predetermined tension when the nerve is sandwiched by the prongs.
In accordance with an aspect of the present invention, the prongs include insulation that covers all of the prong- except for the area for contact for nerve stimulation or recording. The base is formed from a plastic or silicone and has elastic properties. In another configuration when a clamp body is used, the body is formed from a silicone and has elastic properties, In addition, the body includes two elastic clips to be used to open and close the electrode. The prongs include a metal core.
In accordance with an aspect of the present invention, an elastic base has at least three projections. At least one of the at least three projections is on a top side of a nerve and at least two of the at least three projections are on a bottom side of the nerve. The nerve is sandwiched by the at least three projections. The system includes an electrode for stimulation or recording of a nerve action potential comprising at least two metal contacts. The elastic base and electrode are self-insulating and self-holding, such that a non-lifting nerve action potential can be stimulated and/or recorded.
In accordance with another aspect of the present invention, the middle prong and the two outer prongs include insulation that covers all of the middle prong and the two outer prongs except for an area of the metal contact for nerve stimulation or recording. The base is formed from a plastic or silicone. A pressure is applied to the nerve, because it is sandwiched by the at least three projections. The pressure is less than what would be damaging to the nerve for a duration of a test of the compound nerve action potential. The at least three projections can include grooves at a location of the electrode contact with the nerve, such that the nerve is self-positioned on the grooves. The projections can be opened by applying pressure to the elastic base using fingers or a surgical tool. The elastic base returns to its original shape after a deformation. The system can also include a device for applying deformation force to the elastic base. An application of deformation force to the elastic base releases pressure on the nerve. An application of deformation force also allows the elastic base to be repositioned.
In accordance with another aspect of the present invention, the system further includes an impedance monitor. The impedance monitor measures impedance in real time. The impedance monitor couples the electrode to a neural stimulator device. The impedance monitor can include a light emitting diode (LED). The LED is configured to indicate whether the electrode has a good connection to the nerve. The impedance monitor further includes a switch. The switch can be engaged to convert the electrode from stimulation and recording mode to impedance monitoring mode. The impedance monitor is configured to deliver short voltage pulses through the electrode. The impedance monitor further includes an op-amp. The op-amp measures output from a voltage divider. A low impedance measurement indicates the electrode is shorted through saline in the body and a high impedance measurement indicates that the electrode is not connected.
The accompanying drawings provide visual representations, which will be used to more fully describe the representative embodiments disclosed herein and can be used by those skilled in the art to better understand them and their inherent advantages. In these drawings, like reference numerals identify corresponding elements and:
The presently disclosed subject matter now will be described more fully hereinafter with reference to the accompanying Drawings, in which some, but not all embodiments of the inventions are shown. Like numbers refer to like elements throughout. The presently disclosed subject matter may be embodied in many different forms and should not be construed as limited to the embodiments set forth herein; rather, these embodiments are provided so that this disclosure will satisfy applicable legal requirements. Indeed, many modifications and other embodiments of the presently disclosed subject matter set forth herein will come to mind to one skilled in the art to which the presently disclosed subject matter pertains having the benefit of the teachings presented in the foregoing descriptions and the associated Drawings. Therefore, it is to be understood that the presently disclosed subject matter is not to be limited to the specific embodiments disclosed and that modifications and other embodiments are intended to be included within the scope of the appended claims. It should be noted that any dimensions are included simply by way of example and are not meant to be considered limiting. Any suitable dimensions known to or conceivable to one of skill in the art could also be used.
The present invention is directed to a GMG-electrode for recording nerve action potential (NAP) and methods for using such an electrode. The GMG-electrode is so-named for the inventors. Electrophysiological methods are used during repair surgery of peripheral nerve trauma (PNT). PNT is a major medical problem with an annual incidence similar to that of epilepsy. Surgical intervention is provided based on the severity of nerve injury which is determined preoperatively and intraoperatively mainly by electrophysiological assessments. Among those, intraoperative nerve action potential (NAP) or compound action potential (CNAP) recording is preferred for direct assessment of functional continuity of the nerve.
Unsatisfactory, poor intraoperative NAPs associated with large stimulus artifact are recorded with the commonly employed methods using standard hook electrodes. This large stimulus artifact is the consequence of a ‘loop effect’ which is caused by lifting the surgically exposed nerve from the tissue underneath. Based on the loop hypothesis, ‘non-lifting’ nerve recording has been introduced as a new method and been confirmed to result in NAP recordings of satisfactory and significantly improved quality.
‘Non-lifting’ nerve recordings are essential to obtain interpretable NAP recordings. The currently available nerve electrodes, however, do not easily allow clinicians to record proper NAPs. The newly designed electrodes described in the following material solve the problem with stimulus artifact seen in the clinic.
The innovations for ‘non-lifting’ nerve recording, described below, include 1) new electrodes with self-insulated and self-holding features as well as how to easily apply and remove them from the nerve, and 2) procedures for electrode placement and signal verification (such as the stimulus polarity switch test and the intensity-function test), and instrument parameters.
An electrode according to an embodiment of the present invention includes three metal prongs and one elastic base to which the prongs are attached. It is named a ‘sandwich’ electrode because a nerve is sandwiched between the middle prong and the two outer prongs for stimulation or recording. The elastic or plastic base is orientated in parallel with the nerve to provide the nerve with certain tension when it is sandwiched by the prongs. With appropriate elasticity of the base and appropriate size relationship between the nerve and the prongs, the electrode can be held by the nerve itself without any additional support or hand holding which may excessively stretch the nerve. Another feature of sandwich electrodes is that the metal prongs are insulated except for a small area of contact with the nerve for stimulation or recording. Thus, the nerve does not need to be lifted and can remain in situ to keep good contact with underneath tissue during stimulation and recording to achieve a so called ‘non-lifting’ nerve recording.
Electrodes are disposable and used for acute recordings only. Medical grade stainless steel is a primary option for metal prongs used in the electrodes. Other materials include tungsten, platinum, platinum-iridium alloys, iridium oxide, and titanium nitride. For elastic base, medical grade plastic can be used, and other materials include PDMS silicone.
Dimensions of the electrodes can be adjusted based on the size of nerve to be recorded. These dimensions are included simply by way of example and are not meant to be considered limiting. Any dimensions known to or conceivable by one of skill in the art could also be used. Peripheral nerves from normal children and adults range from 1 to 13 mm in diameter. The horizontal inter-prong distance can be varied from 3 to 7 mm. The vertical distance can be varied from 0 to 9 mm for example. Table 1 lists combinations of different dimensions of the electrodes for different sizes of nerves, for example. A set of these prong dimensions, together with the flexibility provided by the elastic base, provide an electrode with an appropriate fit for a nerve of a given size.
A relatively large contact area is used for all nerve sizes to ensure modest charge-injection densities to address a safety concern with nerve stimulation. Such a concern does not apply to nerve recording and therefore contact sizes can be smaller on recording electrodes.
The stretch and/or pressure on the nerve is less than 20 grams, which is controlled by the elastic base or body of the electrodes. The pressure on the nerve is controlled as for the GMG electrode by adjusting the bridge height during fabrication process.
The electrodes were designed using Solidworks. An electrode clip was cast using suing Slyguard 184 (10:1 curing agent) in a 3-part mold and was cured at 150 C. for 15 minutes Electrodes were insulated with the Same PDMS as the clip design and insulation was removed from appropriate areas. The GMG body of the electrode was molded using GMG design and was tested to ensure it applies less than 10 grams/cm2when applied to the nerve.
The electrodes are used for whole nerve stimulation and compound nerve action potential recording from a surgically exposed peripheral nerve, nerve root, or other neural structure. In peripheral nerve recordings, for example, a nerve segment (about 10 cm or longer) is exposed and isolated from the surrounding tissue. The grounding is placed via a surface or needle electrode positioned near the surgical area. For a good contact with underneath tissue between the two pairs of electrodes it is important that the nerve is kept in situ and not lifted as shown in
The electrodes are positioned to align all the contacts along the nerve and to fully cover each contact by the nerve. Both stimulation and recording electrodes are free from extra holding by hand or stand. Electrodes with appropriate dimensions are chosen to ensure appropriate stretch/tension on the nerve.
The recording method using a combination of a sandwich electrode for recording and a pen electrode for stimulation keeps the feature of ‘non-lifting’ nerve recording and improves the flexibility for moving the stimulation site along the nerve. Moving stimulation site along the nerve is performed in nerve inching to collect NAP data around the nerve injury site to localize the site of conduction failure.
The recording and stimulation configurations when using GMG electrodes are the same as those for sandwich electrodes. However, GMG electrodes are designed to be easily applied to and removed from the nerve and for reducing application variability when they are applied to peripheral nerves simply by pressing and releasing the clip. Thus, GMG electrodes can be easily released and moved to a different location along the nerve. In the same way, GMG electrodes can also be used to standardize the peripheral nerve recording procedures for improved consistency in single or repetitive electrode applications. ‘Non-lifting’ nerve recording of clean NAPs using GMG electrodes has been demonstrated in in vitro and in vivo experiments conducted on isolated and exposed animal peripheral nerves.
A differential amplifier and an isolated constant current stimulator (biomedical safety) are two major components of equipment. The bandwidth is 120 Hz to 2.5 kHz (10 Hz to 3 kHz, optional). An analog or digital stimulus polarity switch is provided. Impedance measurement and/or control of electrodes can be implemented. The electrodes are connected to the headbox of a standard IONM (intraoperative neurophysiological monitoring) machine.
In exemplary implementations of the present invention, which are not meant to be considered limiting, but are included as further illustration of the invention, five (3 females and 2 males) adult pigtail monkeys (Macaca nemestrina) weighing 7 to 20 kg were used for recordings under anesthesia. From two additional male monkeys, weighing 4 kg each, peripheral nerves were acquired postmortem and used for recordings in a model system.
Following initial sedation with intramuscular ketamine (12 mg/kg, Phoenix Pharmaceutical, Inc., St., Joseph, Miss.), anesthesia was induced by an intravenous bolus of pentobarbital (6 mg/kg, Ovation Pharmaceuticals, Inc., Deerfield, Ill.), and animals were intubated. Anesthesia was maintained by pentobarbital (continuous infusion at 4-6 mg/kg/h) or isoflurane (0.5-2%). Neuromuscular blockade (NMB) was induced and maintained with pancuronium bromide (0.1 mg/kg every 2 h, SICOR Pharmaceuticals, Irvine, Calif.). Animals were ventilated to maintain a pCO2 of 35-40 mmHg. Heart rate was monitored with an ECG. Core temperature was maintained near 38° C. using feedback-controlled warm-water heating pads. An intravenous drip of 5% dextrose was continuously administered.
Using aseptic techniques, an incision was made on the upper arm. A segment of 10-15 cm length of either median or ulnar nerve was exposed and, under a microscope, carefully dissected from the surrounding tissue while keeping the epineurium intact. During and after dissection, the nerve was irrigated frequently with normal saline solution.
Compound action potentials were recorded from the whole nerve. The animal was grounded through a needle electrode positioned subcutaneously near the incision. Signals were filtered (low frequency 120 Hz; high frequency 3 kHz; Krohn-Hite 3700 Filter, Krohn-Hite Corp., Avon, Mass.), amplified differentially (EG&E Princeton Applied Research 5113 Amplifier, Ametek, Berwyn, Pa.), and digitized at a sampling rate of 25 kHz (Digital Acquisition Processor board, Microstar Laboratories Inc.). Gain varied from 1000 to 25,000. Signals were recorded on a PC using DAPSYS software (v.8; Brian Turnquist, see www.dapsys.net). The nerve was stimulated with constant current, monophasic square pulses (0.05 or 0.1 ms, up to 20 mA, Digitimer DS7A Stimulator, Digitimer Ltd., Welwyn Garden City, England) delivered at 0.25 Hz (controlled by DAPSYS). No signal averaging was carried out for recordings.
Standard IONM hook electrodes (Cadwell, Kennewick, Wash.) and in-house made ‘self-holding’ electrodes were used to perform bipolar recording and tripolar stimulation. The stimulation electrode was either placed underneath the nerve or used to ‘sandwich’ the nerve. For sandwiching, the middle prong of the IONM electrode was bent upwards. The inter-prong distance of IONM electrodes was 6 and 5 mm for recording and stimulation, respectively. The in-house made electrodes for stimulation and recording had the same inter-prong distance (6.5 mm). The two outer prongs were connected to a common outlet and served as an anode while the middle prong served as a cathode for tripolar stimulation with normal stimulus polarity. For the stimulation with reversed polarity, the two outer prongs served as a cathode and the middle prong anode. Standard recording and stimulating electrodes were held by magnetic stands during most recordings.
In a simple model system, saline-soaked gauze (4″×4″) was placed on a platform mimicking the surrounding tissue and body. A saline-soaked gauze strip (three strand braid, 10-12 cm long and 3 mm in diameter) or segment of actual nerve was held by recording and stimulation electrodes such that it was suspended in the air between electrodes, but that the ends touched the ‘tissue’ underneath. This arrangement therefore mimics the recording situation in the operating room. Electrodes, equipment, parameters and configurations were identical to those used for the animal experiments. The system was grounded through a clipper electrode attached to the ‘body’. For nerve recordings in the model system, saline was replaced with synthetic interstitial fluid composed as described previously. Recordings were made at room temperature (about 24° C.).
The findings were applied to an adult patient in whom routine IONM procedures were used to evaluate a left brachial plexus injury during surgical nerve repair. The patient was under general anesthesia with no NMB agents given after induction. Intraoperative NAPs were recorded with an IONM machine (Neuromaster MEE-2000 Intraoperative Monitoring System; Nihon Kohden, Tokyo, Japan). The settings were 100 Hz and 3 kHz for low- and high frequency filter, respectively. The sampling rate was 10 kHz. The same type of Cadwell IONM electrodes were used for nerve stimulation and recording but were held by hand. Constant current stimulation was triggered manually. Stimulation duration was 0.1 ms (monophasic square pulse) and intensity was 0, 2, 5 or 10 mA. The patient was grounded via a sterile subdermal needle electrode (Rhythmlink, Columbia, S.C.) positioned in the shoulder.
Recordings were made using standard IONM hook electrodes from either median or ulnar nerve in the upper arm of the anaesthetized monkey with full NMB. Following the recommended method, stimulation and recording electrodes were used to lift the nerve from the tissue underneath. A large electrical artifact contaminated or completely obscured the neural signal in all recordings performed in this manner. The stimulus artifact was suppressed and a NAP became recognizable after placing a saline-soaked gauze underneath the nerve between the stimulating and recording electrodes to bridge the nerve and the surrounding tissue (i.e., ‘bridge grounding’). Signal specimen and the recording setup are shown in
In
Further,
The model system recordings using a nerve demonstrate that the loop effect influences stimulus artifacts but not neural signals. This might be explained by the fact that these two types of signals travel differently in the loop with neural signals reaching the recording electrode only by one-way conduction along the nerve.
Based on the loop hypothesis, intraoperative NAP recordings were thought to significantly improve if the nerve is not lifted. ‘Non-lifting’ nerve recordings in the model system using modified IONM electrodes were executed with prongs fully insulated except for an area to be in contact with the nerve. This abolished the need to lift the nerve or use the ‘bridge grounding’ technique.
In
‘Non-lifting’ nerve recordings were performed in an anesthetized monkey with newly designed, in-house made ‘self-holding’ sandwich electrodes. As shown in
Further,
Previous publications recommend lifting the nerve out of the surgical field to achieve better intraoperative NAP recordings and to avoid current spread from the stimulating electrode to the tissue underneath. A recording distance of at least 4 cm was also recommended. In practice, however, recordings are often difficult to interpret due to large stimulus artifacts. Results from this study suggest that exaggerated stimulus artifacts may be caused by a loop effect created by lifting the nerve from the tissue underneath. In the model system, the artifact substantially reduced when the loop was cut open, indicating that it is as a major factor contributing to exaggerated stimulus artifact. The loop effect also explains why ‘bridge grounding’ in experiments and in clinical practice reduced the stimulus artifact in recordings from a lifted nerve. Minimizing the loop effect with the ‘non-lifting’ technique of the present invention also allowed for shorter recording distances of less than 4 cm.
This is believed to be the first study showing that the loop effect is the primary cause of a large stimulus artifact that often ruins NAP recordings from a surgically exposed nerve. The findings clearly imply that, contrary to what is typically recommended for NAP recordings, the nerve should not be lifted out of the surgical field and should stay in place (in situ) to have good contact with tissue underneath. However, good contact of the nerve segment tween the pairs of electrodes is difficult to achieve because the recording and stimulation electrodes have to be lifted away from tissue underneath. Furthermore, when there is not enough length of the nerve segment between the two pairs of electrodes, it may be impossible to allow the segment to touch the tissue underneath. Therefore, in this type of recording, the nerve segment is often entirely suspended in air and not touching the tissue underneath. In this case ‘bridge grounding’ needs to be applied.
Interestingly, an animal study previously reported that NAPs could only be recorded when the nerve between the recording and stimulating site was not dissected and isolated from the surrounding tissue. This configuration may be similar to the lifting nerve recordings with ‘bridge grounding’ and to clinical situations when the nerve between stimulation and recording electrodes are left in situ. In the same study, recordings were unsuccessful even when a nerve that did not touch the surrounding tissue was grounded between recording and stimulating site. This observation can be explained by the loop model because the electrical current can travel to the recording electrode via the loop formed by tissue underneath even when the direct pathway along the nerve is shunted by grounding.
The artifact issue has been discussed specifically with regard to the validity of intraoperative NAP recordings in brachial plexus lesion surgeries in infants. The infant plexus is small and relatively short. In a clinical report, Pondaag and colleagues concluded that intraoperative NAP recordings are not useful. They suggested a possibly different pathophysiology of the injured infant brachial plexus rather than suspecting technical challenges. However, as Kline and Happel pointed out, a representative NAP recording presented by Pondaag and colleagues seems to contain a large stimulus artifact outlasting the time within which, based on conduction time, a true NAP would have been expected. It is worth noting that Pondaag and colleagues reported that the nerve under study was fully dissected from the surrounding tissue and isolated from it with a dry surgical patty in their recordings. Kline and Happel successfully recorded NAPs from infants and argued that, although difficult, the recordings were useful for this type of surgery in infants. They did not provide details about how large artifacts were suppressed and neural signals were recognized. However, they suggested that improved instrumentation and technique should be developed for use in the pediatric population to allow application of the same decision-making principles used in the adult population. Avoiding the loop effect may help in the clinic with NAP monitoring in infants in whom the recording distance is short.
Change of stimulus polarity and an intensity-response function survey were used to identify neural signals and to differentiate neural signals from stimulus artifacts. This is the first report of NAP recognition with the aid of the stimulus polarity switch. It is easily performed (some IONM machines have a polarity switch button), time-efficient and produces a consistent result. It is recommended to use this routinely in intraoperative NAP recordings. Because the stimulation was performed with a tripolar electrode, the site of NAP generation changed slightly with polarity switching, and the signal latency always became shorter. The amount of latency shortening depended on the conduction velocity along the nerve and the inter-prong distance of the electrode. In addition, the intensity for maximal NAP increased slightly upon reversal of stimulus polarity reflecting a change in current density.
The loop hypothesis suggests that the ‘non-lifting’ configuration may be the best for intraoperative NAP recording. A new electrode for proof-of-concept of ‘non-lifting’ nerve recording in non-human primate was designed and tested.
The electrodes of the present invention are self-holding and self-insulated. Self-holding is made possible by the three-prong design. Three-pronged hook electrodes are commonly used for nerve stimulation in intraoperative NAP recording. Other types are also used, including straight and ball tip electrodes. NAP recordings are often made with bipolar electrodes. The use of a tripolar recording electrode was also reported with a ground-active-referential arrangement. However, the recordings for the present invention were differential.
When using non-insulated hook electrodes, the nerve needs to be lifted from the wet field potentially stretching the nerve and thereby affecting the recording. ‘Non-lifting’ nerve recording was reported using APS electrodes (the vagus nerve stimulation electrodes). In that report, only small NAP signals were shown following large stimulus artifacts in recordings made in monopolar stimulation and recording mode as bipolar mode was unsuccessful because of limited nerve length.
The results from tests with the new electrodes of the present invention in animals suggest that intraoperative NAP recordings may be substantially advanced by development of electrodes with self-insulation and self-holding features, equipped with multiple contact arrays and an electrical circuitry that allows automation of the stimulus polarity switch and intensity-response function tests.
Currently recommended clinical methods for intraoperative NAP recording originated from experiments in nonhuman primates. Exaggerated stimulus artifacts are identified as a major problem and found ‘bridge grounding’ to be a simple and effective solution. Ultimately, the new methodology of the present invention was brought forward into clinical practice where clinical rather than research equipment was used. The outcome was the same, validating the principle concept shared by recordings in these different settings.
The results of this study provide for the first time compelling, qualitative evidence that suggests that the loop effect may be the primary cause of an exaggerated stimulus artifact that often ruins NAP recordings from a short segment of surgically exposed peripheral nerves. This study was carried out using a limited number of animals and focused on qualitative descriptions of experimental findings. The study included a single clinical case where the ‘bridge grounding’ technique was applied for the first time in the OR to demonstrate and strengthen the significance of the findings related to the present invention. Since then, the technique was applied in two more cases and observed the same effect consistently. Nevertheless, for systematic and quantitative evaluations of these methods additional investigations in healthy and, more importantly, chronically injured nerves would be helpful and are needed.
Intraoperative nerve action potential (NAP) recordings are often affected by exaggerated stimulus artifacts. The results from experiments in nonhuman primate, a model system and clinical recordings, suggest that this technical challenge is embedded in the methodology where the exposed nerve is lifted from the surgical field, leading to a loop effect. This difficulty was resolved by application of ‘bridge grounding’ and configuration of ‘non-lifting’ nerve recording.
The proposed loop hypothesis and the reported solutions may significantly improve intraoperative NAP recording and clinical outcomes in various nerve repair surgeries.
In some embodiments, the present invention includes an electrode system for stimulation or recording of a compound nerve action potential having an elastic base or body with a minimum of two projections on one side and one on the opposing side of the nerve. The electrode placement would allow for “sandwiching” the electrode around the trunk of the nerve. The electrode is self-insulated and self-holding by allowing easy applications and a consistent and reliable ‘non-lifting’ nerve recording from surgically exposed peripheral nerve. The electrode includes a minimum of two metal contacts configured for an appropriate contact size, with none, one, or several on each projection.
The pressure on the nerve as the result of sandwiching it between the projections would apply enough pressure on the nerve to ensure sufficient and consistent electrical contact between the electrode contacts and the nerve. The pressure of the electrode on the nerve would apply less pressure than would be damaging to the nerve for the duration of the test. The projections comprise divots/grooves at the location of the electrode contact with an appropriate configuration and alignment to allow self-orienting positioning on the nerve once the electrode is positioned around the nerve.
The electrode is elastic and is formed in a way that would allow it to be open by squeezing either by fingers or with the use of common surgical tools (e.g. surgical tweezers) or special applicators to change the body of the electrode in a way that would fully open the electrode for positioning on the nerve. Once released from holding, the electrode would return to its original shape, fitting over the nerve with the self-orienting divots/grooves. The electrode would maintain its contact with the nerve in situ once it is released completely and positioned around the nerve. By squeezing the elastic body, the electrode can be opened in a way that would release pressure on the nerve completely allowing the electrode to be removed or repositioned.
Alternatively, the switch can connect the electrodes from GMG to RTIM. RTIM is a microcontroller-controlled device that delivers short voltage pulses via Dout, allowing instant impedance measurement of electrodes as a standalone device. The resulting current is delivered through the resistor to the Electrode. When the electrodes (Electrode and Ref Electrode) are connected across the tissue with the GMG electrode, the impedance will change. Low impedance corresponds to the electrodes shorted through body saline. High impedance corresponds to the electrodes not being connected. The impedance is measured as an output of the voltage divider via the op-amp. The output of the op-amp is sampled by the ADC of the microcontroller. Depending on the voltage output of the voltage divider that measures the impedance, the microcontroller will light the LEDs as red or green depending on the measured impedance. The LEDs are placed visible to the surgeon to provide visual aids to appropriate electrode placement.
The present invention carried out using a computer, non-transitory computer readable medium, or alternately a computing device or non-transitory computer readable medium incorporated into the NAP recording console. Indeed, any suitable method of calculation known to or conceivable by one of skill in the art could be used. It should also be noted that to the extent specific equations are detailed herein, variations on these equations can also be derived, and this application includes any such equation known to or conceivable by one of skill in the art.
A non-transitory computer readable medium is understood to mean any article of manufacture that can be read by a computer. Such non-transitory computer readable media includes, but is not limited to, magnetic media, such as a floppy disk, flexible disk, hard disk, reel-to-reel tape, cartridge tape, cassette tape or cards, optical media such as CD-ROM, writable compact disc, magneto-optical media in disc, tape or card form, and paper media, such as punched cards and paper tape. The computing device can be a special computer designed specifically for this purpose. The computing device can be unique to the present invention and designed specifically to carry out the method of the present invention. It is not a standard business or personal computer that can be purchased at a local store. Additionally this computer carries out communications with the NAP recording device through the execution of proprietary custom built software that is designed and written by the scanner manufacturer for the computer hardware to specifically operate the scanner hardware.
The many features and advantages of the invention are apparent from the detailed specification, and thus, it is intended by the appended claims to cover all such features and advantages of the invention which fall within the true spirit and scope of the invention. Further, since numerous modifications and variations will readily occur to those skilled in the art, it is not desired to limit the invention to the exact construction and operation illustrated and described, and accordingly, all suitable modifications and equivalents may be resorted to, falling within the scope of the invention.
This application claims the benefit of U.S. Provisional Patent Application No. 62/858,418 filed on Jun. 7, 2019, which is incorporated by reference, herein, in its entirety.
This invention was made with government support under AR070875 awarded by the National Institutes of Health. The government has certain rights in the invention.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2020/036599 | 6/8/2020 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62858418 | Jun 2019 | US |