The invention concerns inter alia an iris cover implant.
The young human eye can naturally focus from far to near objects. The mechanism underlying this focus is called the accommodation process. It is composed of three main aspects: first, a convergence of the eye ball as objects get closer, then—for a given luminosity—an increase in the pupil constriction, and finally an adjustment of the crystalline lens shape modulated by the ciliary muscle contraction.
As the eye naturally ages, it becomes harder and harder for it to adjust to near objects. While aging, the subject loses progressively its near vision and is able to create a sharp picture only for far objects. This natural vision affect is called presbyopia. It is assessed that by 2020, around two billion people will have to deal with this optical affect. Presbyopia is experienced as a blurring of the person's vision and frequently arises when the person is reading a document or working at a computer. An untreated patient may compensate by moving the viewed material farther away than would be their previous practice. Roots of presbyopia lays in the aging of the crystalline lens that becomes larger and stiffer, and therefore harder to deform whereas convergence and pupil constriction seem not to be affected.
Many attempts have been done to correct presbyopia. The most common and legacy solution is the use of a pair of reading glasses: as far vision is preserved, the subject can wear glasses to be able to read a book for instance. While this solution is not invasive at all, wearing glasses can be seen as neither aesthetic nor practical for the patient, as he/she needs to put them on and off depending on the distance of the object he is looking at. To solve these problems, multifocal and progressive glasses have been developed. They offer, at least, a first refractive correction for distance viewing and a second refractive correction for viewing near objects; sometimes multiple distance vision by progressive curvature changes.
Surgical attempts have been performed to correct permanently in a more invasive manner the symptoms of presbyopia and restore an amplitude of vision without the need of glasses, often considered as an aging symbol. A common treatment is to induce a static adjustment of the refractive power of one eye of the person. In this way the other eye is used for far vision, and the brain processes the two signals of both eyes to create a new depth of vision that is suitable for near and far images (monovision).
There are different ways to correct an eye for near vision. Changes can be performed on the cornea, by reshaping the latter with laser for instance, or placing a refractive implant inside the corneal tissues (Corneal Inlays). The other solution is to replace the crystalline lens with an implant, adapted to a set of specific distances of vision, named Multifocal Intra Ocular Lens (or MIOLs). This type of device comprises different corrections for a least two specific distances. However this type of treatment suffers drawbacks: it induces a loss of luminosity, halos, glares and flares in night vision for instance.
Original attempts that have been done to try to restore the amplitude of vision may also be noticed such as scleral implants which, for instance, focus on a release of the zonular tension without getting the expected performances.
However, none of the above solutions has proved to be satisfactory.
Taking account of the above, the inventors have searched for a new effective implant to compensate presbyopia.
When searching for such a new implant the inventors have conceived a new implant configuration and have discovered other applications for this new implant configuration.
According to an aspect of the invention, a new iris implant configuration is an iris cover implant that is intended to cover at least partially the iris of an eye, the cover implant comprising a body having a first face that is an opaque face and, on an opposite second face, at least one attachment member that extends outwardly from the body and is fixed thereto, the at least one attachment member comprising at least two clamping portions that are able to attach the body to an iris of an eye by clamping.
The iris cover implant is adapted to be located on an iris of a human eye and attached thereto.
The attachment member or each attachment member is fixed or connected to the cover implant body which means that the attachment member(s) is(are) not a tool which serves only for the placement and attachment of the implant on an iris and which is removed afterwards. In contrast, the attachment member(s) remains in situ in a clamped engagement with an iris portion during the whole duration of the use of the implant in the eye. In other words, the attachment member(s) remains attached in a permanent manner to the anterior face of the iris during the use of the implant contrary to a tool. Thus, the use of such attachment member(s) does not mean that no external tool can be used for helping the positioning of the cover implant on the iris.
The clamping portions of the attachment member or each attachment member may carry out a clamping function on the iris, i.e. on a portion of the anterior surface of the iris, in particular a ridge thereof (natural iris fold) to clamp an iris portion or ridge between the two facing portions. Securing the cover implant body to the iris is therefore possible without piercing or traversing the iris tissue. This makes it possible to remove the cover implant body without strongly damaging the tissue with holes.
The two clamping portions are close to each other in their initial position (clamping position) so as to be able to clamp an iris ridge between the two facing portions. The two clamping portions may be located at a distance from each other that is less than 3 mm, preferably less than or equal to 2 mm, e.g. less than or equal to 2 mm. If the distance between the two facing portions of the attachment member or each attachment is greater clamping an iris ridge will not be possible or will prove to be very difficult.
To be noted that the clamping portions may have each at their free end (opposite the end that is connected to the second face) an enlarged part or head with a shape that is appropriate to firmly and safely clamp an iris ridge, e.g. with a flat surface facing the flat surface of the other clamping portion.
The iris has a shape of a disc or annulus and the iris cover implant is intended to cover at least a portion of the iris whatever the shape of the implant. Generally speaking, the implant may be located on a radius of the iris and oriented towards the center of the iris.
The iris cover implant may have different shapes, such as an overall shape of a segment, an angular sector or area in some instances etc.
Generally speaking, the iris cover implant body has two opposite faces: a first face is opaque to light and a second one is provided with the attachment member that is intended to cooperate with the iris of the eye (more particularly, the anterior surface of the iris) so as, for the iris cover implant, to be attached to the iris (two functional zones of the body). The opaque face is then intended to be visible through the cornea of the eye. The first and second faces are defined as opposite in that they are faces of the body that are both oriented in opposite directions, one being intended to face the cornea when the implant is attached on the iris, while the other is intended to face the surface of the iris. Both faces may not strictly extend one above the other. One face may even extend further than the other so that the extended portion of this face does not overlap the other one.
Such an iris cover implant is efficient and easy to install on the iris (pupillary margin) due to the at least one attachment member.
In addition, it is not intended to be in contact with the iridocorneal angle, which avoids ocular hypertension and following sever complications. It is neither intended to be in contact with the endothelium of the cornea to avoid endothelial failure and following severe complications.
With such a cover implant, more or less severe complications of iris can be avoided: infections, anterior uveitis, ocular hypertension, glaucoma, endothelial failure, corneal oedema and decompensation.
Example of complication secondary to cosmetic artificial iris anterior chamber implants can be found in a case report from Yusrah Shweikh, Sally Ameen and Ali Mearza, BMC Ophthalmology, 201515:97.
The cover implant is intended to be placed into the anterior chamber of the eye, that is, within aqueous humor.
No biocompatibility issue will arise on a long term basis with this new implant configuration. No inflammatory reaction will take place after the installation of such a cover implant.
More generally, such a cover implant may be used for the optical enhancement of an eye, including the compensation of presbyopia, regular ametropia or regular astigmatism.
More particularly, such a cover implant may be used for presbyopia correction or compensation by forming, generally together with other cover implants simultaneously, a diaphragm that is intended to reduce the diameter of the aperture of a constricted iris.
Such a cover implant may alternatively be used for visual or cosmetic enhancement, alteration or repair of the iris anterior surface (visible through the cornea as the “color of the eye”).
In such applications a cover implant may be used alone or in combination with other cover implants, e.g. to hide an iris deformity or to locally or wholly modify the color.
In these applications, the iris cover implant or implants are attached on the iris in a position such that the opaque face is above the iris. Here the cover function does not require for the opaque face to extend beyond the iris inner edge.
When searching for a way to compensate presbyopia, the inventors have noted that it could be interesting to reduce the aperture of the eye (pupil) so as to increase the depth of field. In particular, they have noted that the ability of the lens to change its shape decreases with time, but the pupil constriction keeps working. The inventors have also noted that elderly persons who can no longer accommodate with their crystalline lens exacerbate the phenomenon of pupil constriction. Similarly, to the way a photographer increases the depth of field of its camera, pupil constriction leads to a reduction in the diameter of the pupil, thereby suppressing marginal rays. The paraxial rays that remain and pass through the fully constricted pupil are thus able to create a sharp picture on the retina for a wider range of distances of vision. The suppression of these marginal rays using a small aperture is called stenopeic effect or “pinhole effect”. However, the inventors have realized that, even though the phenomenon of pupil constriction is exacerbated with elderly persons, this is not sufficient to compensate for the other effects of the eye aging, including presbyopia.
From these statements the inventors have thought of further reducing the diameter of a constricted iris (enhancement of natural pupil constriction) through the use of several iris cover implants, e.g. of the above-mentioned type, to correct presbyopia.
When installed on an iris the plurality of cover implants are arranged in a radial distribution around the central aperture of the iris (pupil) and attached to the iris. The plurality of cover implants may radially converge towards the center of the central aperture.
Each cover implant may be attached on the anterior surface of the iris in a position such that at least one zone of its opaque face that has an optical function forms a cantilevered portion with respect to the inner edge of the iris that bounds outwardly the pupil. Therefore, the opaque optical face extends in the central aperture (the pupil) beyond the iris inner edge when viewed through the cornea. The cantilevered portion with the opaque optical zone may be viewed as a diaphragm portion.
The cover implants may be arranged on the iris so that when pupil is dilated (for far vision and/or in dark conditions) the cover implants are away from each other.
When pupil is constricted (this phenomenon of constriction takes place for near vision and/or under daylight conditions) the cover implants are located one close to another so as to form all together a diaphragm and further reduce the diameter of the natural reduced central aperture of the iris.
By further reducing the aperture of the naturally constricted iris, suppression of the marginal rays can be obtained and the depth of field can be increased (pinhole effect) without significantly and permanently reducing the light adaptation ability of the implanted eye (e.g. luminosity in far vision must be preserved).
Such a diaphragm configuration makes it possible to create a stenopeic effect which efficiently improves near vision.
This pinhole effect is obtained only when needed, i.e. the effect is non permanent whatever the position of the pupil while the cover implants are always present on the iris and move with the latter. When pupil is dilated, the cover implants are away from each other and, therefore, do not lead to a pinhole effect. This effect may be more or less progressive when pupil is constricting. The implant may therefore be considered as dynamic since its position in the eye changes with the iris movement and its technical effect is produced dynamically, i.e. during the constriction movement of the iris.
Thanks to this new iris implant configuration the amount of light is not significantly reduced for distance vision or low light conditions (mesopic) vision. This implant makes it possible to adapt the conditions of luminosity and focus. This implant is adapted to the patient's eye specificities and adapts dynamically so that the vision in night or dark conditions is not limited while a pinhole diameter could be perfectly fitted (customized to the eye).
To be noted that the part of the body that carries out an optical function is not necessarily a detachable or removable part of the body but may be integrated in the cover implant.
According to other possible features:
the opaque face extends at least in a plane that is substantially perpendicular to the direction of extension of the at least one attachment member relative to the second face of the body;
the at least two portions are able to move away from each other under the action of an external force in a plane containing the direction of extension of the at least one attachment member relative to the second face of the body (this direction of extension is generally perpendicular to the second face of the body, and possibly to the first face if both faces are parallel), the body being able to elastically bend within the plane when submitted to an external force so as to cause said at least two portions to move away from each other from an initial position, said at least two portions being able to return to their initial position (a clamping position) in the absence of any external force (this elastic bending movement can me made a repeated number of times without causing any damage to the implant); the two facing portions are spaced apart from each other in their initial position by a distance that allows to clamp an iris ridge therebetween and can be moved away from each other in a direction that tends to increase the initial distance therebetween; this direction can be a combination of an axial direction (perpendicular to the two facing portions) and the direction of outward extension of the attachment member; in the bent position, the two portions are spaced apart from each other by a distance that enables placement of the portions on either side of an iris edge; to be noted that the bending movement is made possible due to the inherent properties of the whole body material (the whole body deforms during this bending, not only one zone of the body even though some part of the body may be conceived for facilitating bending of the body; bending of the body is generally not possible beyond a given angular range, e.g. 90°; generally speaking, the body cannot be bent so that the two extreme bent parts thereof come into contact with each other (as in folding movement of a paper sheet)), i.e. this is not a folding movement; the bending movement is made about a longitudinal direction (perpendicular to the above plane) and the bending force may be applied on two opposite peripheral edges of the body that are spaced apart from each other along a transverse direction (these two edges run parallel to the longitudinal direction); attachment/securing of the body to one or more portions of the iris anterior surface is therefore quite simple since this is not necessary to cause complex deformation to the body in different directions to achieve attachment
the body has at least two receiving portions that are each adapted to receive an instrument or a portion of an instrument for elastically bending the body; each receiving portion may take the shape of a groove that opens out towards the exterior of the body, i.e. in a direction away from the body;
the at least one attachment member is integral with the body; this simplifies the conception of the cover implant compared to an attachment member which is fixed to the cover implant body afterwards, e.g. in a removable manner;
the at least two attachment members are spaced apart from each other along a first direction that is substantially perpendicular to the direction of extension of the at least one attachment member, the body having a thickness along the direction of extension of the at least one attachment member relative to the second face of the body, the thickness of the body being reduced in a zone of the body that is located between the two zones where the at least two attachment members are disposed;
the body comprises a frame provided with the at least one attachment member and a cover member assembled with the frame, the cover member including the opaque face; the two-part body may be easily manufactured; such a configuration makes it possible to customize the cover member, e.g. by selecting appropriate materials, pigmentation or the like, whereas the frame may be common to all the cover implants (ex: one frame per cover member); when the cover member is entirely opaque it masks the at least one attachment member from the above (according to a view taken through the cornea, facing the iris); when assembled the cover member extends above the frame and covers at least a portion thereof, in particular at least the portion where the attachment member(s) is(are) located; the frame may extend perpendicularly to the direction of extension of the attachment member(s) in two planar directions; the cover member (e.g. it may be curved as a tile) may also extend in the same two planar directions and has a longitudinal extension that covers (overlaps or is over) the whole longitudinal extension of the frame or at least the portion where the attachment member(s) is(are) located;
the frame and the cover member are mechanically engaged with each other; the cover member may be engaged inside the frame or vice versa; the cover may be inserted in a slot defined in the frame;
the frame has two receiving portions that are each adapted to receive the cover member; such receiving portions may be inwardly oriented relative to the frame and, e.g. may be facing each other; such receiving portions may take the shape of rails or grooves; the two receiving portions may be located on, or close to, two opposite edges of the frame that are spaced apart from each other along a transverse direction (these edges run parallel to the longitudinal direction of the body when it has a longitudinal extension);
the frame comprises two parallel spaced apart beams with two attachment members located on the two beams respectively; the two parallel beams may be perpendicular to the longitudinal direction of the body and perpendicular to the above two opposite edges of the frame (when the current feature is combined with the previous feature);
the cover member also includes a mechanical portion that is assembled with the frame;
the mechanical portion is opaque, transparent or pigmented in a way to obtain a specific visual appearance;
the cover member and the frame are each able to be elastically deformed; this may make it possible to easily mount the cover member to the frame; the cover member may take the shape of a flexible foil or tile;
the at least one attachment member is a clamping member; a clamping member makes it possible to clamp a portion of the anterior surface of the iris; several clamping members may be used;
the body comprises a frame provided with the at least one attachment member and a cover member including the opaque face, the frame and the cover member being made of a single piece; when the cover member is entirely opaque it masks the at least one attachment member from the above (according to a view taken through the cornea, facing the iris);
the opaque face has at least one zone that is an opaque optical face; this feature corresponds to an iris cover implant that is, inter alia, a diaphragm implant and has an optical function when implanted on the iris of an eye;
the opaque optical face belongs to an optical portion that extends beyond the frame (in a longitudinal direction); here the optical portion is intended to be located above the pupil margin (beyond the iris inner edge) so as to reduce the diameter of the latter when the cover implant is attached to the iris; to be noted that the entire opaque face of the cover member extends both above the frame and beyond the latter, only the zone extending beyond the frame having an optical function (diaphragm); in another embodiment, the opaque optical face constitutes the only opaque zone of the cover member; the optical portion may longitudinally or axially extend the mechanical portion;
the opaque optical portion and the mechanical portion may form together a single piece, which proves to be easy to manufacture.
Another aspect of the invention concerns an iris cover implant intended to cover at least partially the iris of an eye, the cover implant comprising a body having a first face that is an opaque face and, on an opposite second face, at least one attachment member that extends outwardly from the body, the at least one attachment member being able to attach the body to an iris of an eye so as to move together with the natural movement of the iris. Such an iris cover implant (diaphragm implant) is thus a dynamic implant (or part of a whole implant) that is controlled by the movement of the iris. Such an iris cover implant may have an optical correction function and protrudes inside the central aperture of the iris to carry out optical correction. The protrusion may intervene only when optical correction is necessary, i.e. when the pupil is in a constricted position, for example to reduce the pupil size. The iris cover implant may therefore act as a dynamic diaphragm together with the natural movement of the pupil. Any of the explanation and features given in relation to the above first aspect of the invention may also be applied here.
The invention also concerns a surgical kit comprising a plurality of cover implants (e.g. two or more than two or three) as defined above in any of the two aspects (with the main features of the aspect or also with at least some of the possible other features). These cover implants are then placed inside the eye, e.g. through a corneal incision, and next attached to the anterior surface of the iris by means of a specifically designed surgical instrument such as an appropriate surgical forceps thanks to the above-described attachment members. The plurality of cover implants forms a multi-component iris (intraocular) implant acting as a diaphragm when attached to an iris of an eye (presbyopia treatment). The plurality of cover implants provide all together a stenopeic effect as described above when all attached to the iris and the iris is constricted enough. The multi-component iris implant is a dynamic implant that is controlled by the natural movement of the iris.
The invention also concerns a surgical kit comprising a plurality of cover implants (e.g. two or more than two or three) as defined above in any of the two aspects (with the main features of the aspect or also with at least some of the possible other features), the plurality of cover implants forming a multi-component iris (intraocular) implant acting as a partial or total artificial iris when attached to an iris of an eye thanks to the above-described attachment members. Here the cover implants are intended to replace at least in part the visual aspect of the anterior surface on an iris. In practice, the cover implants are intended to at least partially cover the natural iris. This makes it possible to replicate, alter or enhance the cosmetic/visual appearance of the natural iris surface. The multi-component iris implant is a dynamic implant that is controlled by the natural movement of the iris.
The invention also concerns an assembly of several cover implants (e.g. two or more than two or three) as defined above in any of the two aspects (with the main features of the aspect or also with at least some of the possible other features), the plurality of cover implants forming a single iris (intraocular) implant acting as a diaphragm when attached to an iris of an eye thanks to the above-described attachment members. This assembly is intended to be put in place as a whole on the iris. The multi-component iris implant is a dynamic implant that is controlled by the natural movement of the iris.
The invention also concerns an assembly of several cover implants (e.g. two or more than two or three) as defined above in any of the two aspects (with the main features of the aspect or also with at least some of the possible other features), the plurality of cover implants forming a single iris (intraocular) implant acting as a partial or total artificial iris when attached to an iris of an eye thanks to the above-described attachment members. This assembly is intended to be put in place as a whole on the iris. The multi-component iris implant is a dynamic implant that is controlled by the natural movement of the iris.
The detailed aspects which will be described with reference to the drawings may complete the above-mentioned more general aspects and therefore may be combined with them. To be noted that the above-mentioned more general aspects may be applied to other embodiments than those described thereafter.
Other features and advantages will emerge in the course of the remainder description, given by way of non-limiting example only, with reference to the following drawings, in which:
In a first embodiment an iris (intraocular) cover implant carries out a cover or overlap function relative to the iris on which it is intended to be fixed.
The terms “cover implant” will be used in the remainder of the description whatever the embodiments and the function/application of the implant since the iris will always be covered by at least a portion or component of the implant.
Here the cover implant is made in two parts as illustrated in
As illustrated in
The frame is made of a material or a combination of materials that allow elastic deformation of the frame. PMMA may be an example of appropriate material (transparent or died in the mass).
In the present embodiment there are two attachment members 14, 16 that are disposed on one and the same side or face of the frame that is here oriented downwards. The attachment members are here made integral with the cover implant body, in particular the frame. In a variant embodiment a single attachment member or more than two can be envisaged.
The frame 12 has an overall axial or longitudinal extension along a longitudinal axis X and transverse extension along a transverse axis Y. The frame 12 also extends perpendicularly to the plane X, Y along axis Z that defines its thickness (
The attachment members 14, 16 extend outwardly from the frame 12 along an overall direction (axis Z) that is substantially perpendicular to the plane XY.
The frame 12 comprises here two transverse beams 18, 20 that are axially spaced apart from each other along axis X so as to leave there between a central opening 22. Cross beams 18, 20 extend along transverse axis Y (fig.1).
The two transverse beams 18, 20 are connected at each of their two opposite ends to two parallel longitudinal or axial supporting members 24, 26. In another embodiment, more than two transverse beams may be envisaged or a single beam connecting the two supporting members. The single beam may have a longitudinal axis (X) that has not necessarily the same extension as the supporting members.
For instance, the frame is made of a single piece and may be manufactured by an injection molded process or standard machining.
The following description will be made for cross beam 18 for the sake of simplicity and is identical for beam 20.
When viewed in a front or top view (
When viewed in a transverse cross section (AA cross section of
The thinned configuration of the cross beams in their central and main portions as described above makes it possible to elastically deform the frame in the plane of
In the present embodiment, the two attachment members 14, 16 are located on the underside of the beams 18, 20 respectively in the central portion 18a, 20a thereof.
Each attachment member may be a clamping member and may comprise at least two clamping portions or jaws 14a, 14b and 16a, 16b.
In the present embodiment, the two jaws are spaced apart from each other in their stable, non-deformed initial position as represented in
When an external bending force is exerted on the two supporting members, here along axis Y, the frame 12 bends in the plane of
When the bending force does no longer exert, the frame returns to its initial position of
The axial supporting members 24, 26 (
In the present embodiment the axial supporting members 24, 26 may perform a first function for accommodating a surgical instrument or part thereof (ex: forceps) that makes it possible to exert an external bending force on the frame to cause its bending and jaws opening as explained above.
In this respect, the axial supporting members 24, 26 may each comprise a first receiving portion 24a, 26a outwardly oriented relative to the frame (
Here the first receiving portion takes the shape of a longitudinal or axial groove opening out to the outside of the frame and running along the length of the supporting members. The two grooves are opening out on two opposite directions respectively. Each groove has a concave shape which may be adapted to that of the instrument or part thereof in a complementary manner. The opening of the groove may be centered on a plane XY in which the two beams 18, 20 extend. In alternate configurations, the groove may rather be oriented upwardly, i.e. with an opening that makes an angle with the above-mentioned plane.
Further, the axial supporting members 24, 26 may each comprise a second receiving portion 24b, 26b that is generally opposite the first receiving portion. Both second receiving portions are facing each other and define there between a transverse space. They more particularly define there between together with the upper side or face of transverse beams 18, 20 (the upper side lies in an XY plane) a housing or slot that is adapted to accommodate a cover member.
Second receiving portions 24b, 26b may take the shape of rails or grooves that are inwardly oriented relative to the frame, i.e. substantially towards the beams. Second receiving portions 24b, 26b are more particularly located at a distance from the upper side of the beams (
The cover implant body 10 also comprises a cover member 30 (
The cover member 30 comprises two portions (see
a mechanical portion 32, and
an opaque optical portion 34.
The two portions form together a single piece in the present embodiment. They may have each the same thickness along axis Z (
The mechanical portion 32 is to be assembled with the frame 12 thanks to the second receiving portions 24, 26 described above.
The mechanical portion 32 and the opaque optical portion 34 are aligned to each other along the longitudinal axis X. Both mechanical portion 32 and opaque optical portion 34 extend in a plane XY (parallel to the plane of extension of the two beams 18, 20).
Here the cover member visible on
When viewed from above (as in
The opaque optical portion 34 axially extends mechanical portion 32 at its front zone by an appropriate shape that may vary according to the embodiments. The portion 34 has a function of being opaque, i.e. non transparent to light so as to produce a stenopeic or pinhole effect. Here this is the second zone 34a of the upper face of the cover member that is opaque. Generally, the portion 34 may be opaque across its whole thickness (
Here the opaque optical portion 34 has two converging edges 34b 34c that extend from longitudinal edges 32a, 32b respectively and terminate at an end edge 32d that extends transversally and connects both inclined edges 34b, 34c.
Edges 34b and 34c have a geometry that adapts to the geometry of other opaque optical portion edges of adjacent cover implants when all are installed on the iris of an eye around the pupil and the latter is in its most constricted position as will be seen subsequently. Here the inclination angle of the edges is the same for all the opaque optical portion edges of the cover implants so that two adjacent converging edges of two adjacent cover implants are in contact. Thus, no light will be able to pass between two adjacent edges in this embodiment.
Put it another way, when all the cover implants are in a close relationship (most constricted position of the pupil) adjacent to each other, the opaque optical portions form, and act as, a whole diaphragm that is as homogeneous as possible (continuity in the opacity).
Different other possible shapes of opaque optical portions will be described later on.
As illustrated in
As represented in
The cover member is able to be elastically deformed, i.e. here by bending around its longitudinal axis (parallel to axis X). The bending movement takes place in the plane Y, Z.
In a general manner, the cover member 30 overall extends in a plane (XY plane). It is however elastically deformed as represented in
As represented in
Here cover member 30 is flexible and may be made, e.g. of silicon or hydrophilic acrylics or hydrophobic acrylic or elastomer.
In
Prior to any assembling operation the cover member 30 is bent as explained above (this is done through using conventional forceps as represented in
The longitudinal edges 32a and 32b are engaged or inserted into second receiving portions 24b, 26b above cross beams 18, 20 and then are caused to slide along these portions (ex: in the rails or grooves) as illustrated in
The cover member 30 is axially pushed until the opaque optical portion 34 substantially reaches the free ends of second receiving portions 24b 26b (
When the cover member has been assembled with the frame, these two parts are maintained together through friction and contact pressure.
Thus, opaque optical portion 34 axially extends (along axis X) beyond the frame as a cantilevered beam. This holds true whatever the shape of the cover member, in particular the opaque optical portion. This arrangement makes it possible to attach the cover implant on the iris in such a position that only the opaque optical portion extends beyond the iris edge or pupil margin (see e.g. the relative position between cover implant 10 and iris edge E on
Other embodiments not depicted here may cover an inverted configuration in which the frame is engaged inside the cover member. The cover member may be designed so as to incorporate receiving portions for accommodating the frame. For example, two receiving portions may extend respectively from two substantially parallel edges of the cover member (as edges 32a and 32b), e.g. in a perpendicular direction to the cover member, and may cover the frame placed below on three or four sides as a cap: the upper side and two downwardly extending lateral sides with a possible angled terminating portion extending under the frame so as to form a lower side. Assembling the cover member and the frame member may be achieved by axially engaging the frame into the lower receiving portions of the cover member.
To be noted that the above-described assembling process equally applies to any other cover implant embodiment with a frame, and possibly a cover member, having different configurations. For example, the frame may have only one transverse beam or more than two transverse beams as already envisaged above.
Clamping member 42 has two axially elongated seizing portions or jaws that behave as jaws 14a, 14b in
In
In
Other possible configurations may be envisaged, e.g. with different head shapes, different number of seizing portions or jaws etc.
As for the previous embodiments the attachment members may be located under the two spaced apart transverse beams in the central portion thereof or may be arranged at another location. As in
To be noted that in
However,
The attachment members of
In
The opaque optical portion 34′ has a substantially curved shape, e.g. a bean shape, with a central curved zone 34′a and two side zones 34′b and 34′c that laterally extend beyond portion 32 and frame 12 in a flared manner viewed from above. Central curved zone 34′a is axially aligned with mechanical portion 32 when viewed from the above.
As for portion 34 in
Side zones 34′b and 34′c are not lying in a plane as that of
This arrangement makes it possible to appropriately dispose adjacent opaque optical portions during the diaphragm closing (when the pupil is being constricted) so that they do not mechanically interfere with other. The adjacent opaque optical portions may thus partially overlap when seen from the above.
Such a configuration allows to provide an efficient homogeneous opaque diaphragm or cover with all the adjacent opaque optical portions of all the cover implants.
In
Opaque optical portion 34″ has a curved end edge 34″c that is substantially concave along its length so as to reproduce a substantially circular outline when all the cover implants are adjacent to each other in the most closed configuration of the whole diaphragm implant (pupil fully constricted).
More particularly, opaque optical portion 34″ has a double concavity instead of a single one for opaque optical portions 34 and 34′.
This arrangement makes it possible to appropriately dispose adjacent opaque optical portions during the diaphragm closing (when the pupil is being constricted) so that they do not mechanically interfere with other. The adjacent opaque optical portions may thus partially overlap when seen from the above.
Such a configuration allows to provide an efficient continuous opaque diaphragm or cover with all the adjacent opaque optical portions of all the cover implants. All that has been described above in relation with
In a variant the surgical kit may comprise separate cover members and separate frames (when the cover implants are not in a single piece) that need to be assembled together prior to any iris implantation by the surgeon.
In the initial dilated position (fully open) of
The central aperture O1 bounded by the iris sphincter edge E is at its maximum diameter (ex: 7 mm).
The cover implants 10 have a rather small opaque optical portion 34 such that in this dilated position where they are far from each other they may be considered as not affecting the vision.
In
The cover implants 10 are caused to be moved by the iris movement and get closer to the center of the aperture and from each other (dynamic implant the movement of which is controlled by the natural movement of the iris). In
The cover implants 10 come into contact with each other two by two or in very close proximity so that the opaque optical portions 34 form all together a substantially continuous opaque diaphragm or zone (ex: here a band-shaped zone) around reduced aperture O3. This produces a pinhole or stenopeic effect which increases the depth of field.
Here the close contact or proximity between the adjacent cover implants 10 is made possible thanks to the shapes of the opaque optical portions 34 that are adjusted to be positioned side by side in a same plane.
In contrast to cover implants 10, cover implants 30″ are not lying in a planar configuration due to their upwardly inclined or raised side zones 34″b (see
In the dilated position of
In
In
In these Figures the opaque optical portions may overlap (since they have not a planar configuration) two by two not necessarily in a regular manner. These opaque optical portions have all end or terminating edges that are at least partially facing the central aperture O3 of the pupil when the corresponding cover implants are close to each other.
What matters is that the outline of the central aperture O3 bounded by these end edges be located in an annular zone AZ centered on a diameter or circle D1 that corresponds to the targeted diameter for the diaphragm in the most constricted pupil position so as to obtain the desired stenopeic effect. The zone Z is outwardly bounded by two circular lines C1 and C2 that represent tolerances margins with respect to the target D1.
In
In
In
In
In the present embodiment the cover implants may form part of a surgical kit.
In a first step (
In a further step (
The forceps include a pair of arms 112, 114 that flank the cover implant 10 along its longitudinal edges (supporting members 24, 26). In particular, arms 112, 114 are engaged into the respective first receiving portions 24a, 26a (the latter have shapes that are adapted to that of the arms, i.e. a semi-cylindrical shape to house a cylindrical or substantially cylindrical shape) to carry the cover implant 10.
Alternatively, the cover implant 10 may be loaded into a dedicated cartridge injector and then inserted through the corneal or scleral tunnel incision into the anterior chamber of the eye.
When the cover implant 10 is above the iris (
Thus the cover implant 10 moves towards the radial ridge (ex: downward movement if the patient is in an horizontal position) until the latter lies between the two clamping portions or jaws. Once correctly positioned on either part of the ridge P, the clamping effort can stop so that the two clamping portions or jaws can clamp the ridge as illustrated by
The same process is repeated (from
To be noted that radial and longitudinal positioning of the cover implant may be guided by means of a dedicated corneal marker impregnated with sterile biocompatible ink to indicate the central optical clear zone (minimum diaphragm diameter) and any adequate number of equally spaced radii or by any other optical means including the projection of a pattern of visible laser light through the cornea onto the iris surface.
More generally, the surgeon may use a positioning assist system, including inter alia an optical guide target (e.g. a laser with a visible wavelength) that will represents the optimal position of the implants on the iris, through the cornea.
The method may be performed with the pupil in its more constricted position so as to optimize implants positioning in order to have in this position an artificial diaphragm with a minimum diameter and with relative positioning of the implants that ensures no mechanical blocking.
Generally speaking, the opaque optical portion (or at least the opaque optical face of the cover member) of the iris cover implant may be designed according to various geometrical shapes (see for example
Cover implant 120 comprises a cover member 122 corresponding to cover member 32 of
The axial extension (along axis X) of cover member 122, and possibly frame 124, may be lengthened so as to cover a greater area of the iris anterior surface.
The two cover implants 132, 134 are assembled with each other through a sliding arrangement. However, other alternative assembling arrangements may be envisaged.
Each cover implant comprises a body having a frame 136 with attachment members 138 (
Each cover member 140, 142 has two portions:
a main identical portion 144 as mechanical portion 32 in
a secondary portion that has two laterally extending arms 148, 150 and 152, 154 both connected to main portion 144.
The two arms of the two cover members are in register with each other.
One of the two arms lies substantially in a plane as that of
This arrangement makes it possible for each raised arm of a cover member to overlap, to a small or greater extent, the corresponding planar arm of the other cover member (see
The raised arms are each provided with a protruding member, e.g. a lug, a rib etc., 148a, 154a (
The two arms of each cover member comprise each a first zone Z1 corresponding to an opaque optical zone that is intended to play the role of a diaphragm (this is an optical portion with an optical face).
The two arms of each cover member also comprise each a second zone Z2 and Z2′ corresponding to an opaque non optical zone. Zone Z2, Z2′ has two spaced apart portions that flank the first central zone Z1.
The above described sliding arrangement is carried by second zone Z2, Z2′, in particular, by the two spaced apart portions that flank the first central zone Z1.
In
In
In the present embodiment the main portions 144 and the zones Z2 and Z2′ have each a transparent face or an opaque one (pigmented, textured and/or colored or not) so as to cover the iris either for locally masking a deformity, etc. and/or for aesthetic reasons (color change etc.).
The zones Z1 have both an opaque optical face (the whole zones in their entirety may be opaque), e.g. a black face to play the role of a diaphragm when moved close to each other as in
Other configurations may be envisaged depending on the anatomy of the patient's eye and his/her optical disorders. In particular, different shapes of iris cover implants, arms etc. may be envisaged to more or less cover the natural iris.
The features and advantages of assembly also apply here to assembly 160.
In
This volume represents a typical volume V in which the cover implant(s) according to the invention may be installed, being understood that a security margin has to be taken so as to avoid placing an implant which would be in contact or too close to the iridocorneal angle (angle θ in
The volume V has been represented in an offset position (this does not correspond to any actual design) to highlight the width C that is not authorized for implant installation.
Generally speaking, the angle a of the allowed volume is less than the angle θ, for instance less than 60% of this angle so as to avoid any contact between the implant(s) and the cornea Cr.
The length or width L is also preferable less than D-C for the same reasons.
The height h is less than the maximum axial distance A of the anterior chamber and for instance less than 60% of this distance for the same reasons as above.
Put it another way, the constraints that have to be taken into account before placing a cover implant on an iris depend on the actual diameter of the iris, the actual travel of the iris during its two extreme positions and the height of the cornea.
By way of example, for an average implant (i.e. an implant that is based on an average of biological data measurements obtained on a plurality of patients) its length or width L and height should not go beyond 7.5 mm and 1.5 mm for space physiological reasons. Anyway, the size of the implant is to be made as small as possible.
Number | Date | Country | Kind |
---|---|---|---|
17162905.8 | Mar 2017 | EP | regional |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2018/057661 | 3/26/2018 | WO | 00 |