Not applicable.
The use of isometric as compared to rhythmic exercise in the general field of athletic strength development, as well as a therapy for strength recovery has been the subject of somewhat controversial discourse over the past decades. In general, such exercise has been considered to promote, for example, coronary risk factors. See generally:
However, as such attitudes persisted, some investigators commenced to observe contradictions to these generally accepted beliefs. See for, example, the following publications:
With the approach of protocol developed by Wiley, the isometric regimen is closely controlled both in terms of exerted force and in the timing of trials or exertions. The system and method described by Wiley are known to be useful for treating hypertension. Hypertension is associated with an increased risk of a wide range of disease and disorder, including stroke, organ failure, and particularly cardiopathy. The exact causes of hypertension are rarely known with certainty, but risk factors for hypertension include obesity, genetic factors, smoking, diet and inactivity. As Wiley has shown, not all forms of exercise provide equivalent therapeutic benefit to the cardiovascular system and for the treatment of hypercholesterolemia, with a protocol for brief maximal isometric exercise providing a clear benefit.
Hypertension, hypercholesterolemia, atherosclerosis and cardiovascular disease hare interrelated in their causes, treatment and effect on the body. The class of drugs known as HMG-CoA reductase inhibitors or statins is widely prescribed for treatment of hypercholesterolemia and associated cardiovascular disease, including the debilitating effects of progressive atherosclerosis. Various statins that have been clinically utilized include atovastatin, cerivastatin, fluvastatin, ovastatin simvastatin among others. The statin drugs were initially prescribed to relieve hypercholesterolemia, and to reduce the blood concentrations of low-density lipoprotein (LDL) and triglycerides. It has become apparent that the statin drugs apparently have additional therapeutic benefits that are independent and or interrelated with the effects of reduction in blood cholesterol concentration. The effect of statin drugs thus includes a reduction in vascular inflammation, and a protection of the heart against ischemic disorders. For more information on the pleiotropic effects of the statin drugs see generally:
Although the exact mechanism of statin drug action for cardioprotection is not fully known, it is widely believed that statin drugs stimulate nitric oxide synthase activity in vascular endothelium, and presumably in other tissues. Disorders of the vascular endothelium related to in nitric oxide metabolism are believed to play a crucial role in the pathogenesis of atherosclerosis in hypercholesterolemia. Thus, certain cardioprotective effects of statin drugs can be mimicked in part by other physiological stimuli that induce nitric oxide synthase and increase nitric oxide availability. Moreover, nitric oxide metabolism is interconnected with the metabolism and regulation of LDL, cholesterol and triglycerides, and with the progress of atherosclerosis. As one example of this interrelationship, changes in nitric oxide levels have been inversely correlated with changes in LDL-cholesterol concentrations.
Nitric oxide (NO) has been identified as a signaling molecule in mammalian and other systems. NO, is a labile, endogenously produced gas that is enzymatically synthesized, can rapidly diffuse, and quickly disappear. NO is known to be a potent regulator of blood pressure due to its activity as a vasodilator, but has a diverse action on a wide variety of organ systems. Endothelial nitric oxide synthase (eNOS) is induced to synthesize NO by blood vessel wall shear stress. Upon the activation of eNOS and induction of NO synthesis, NO is released by endothelial cells. Based on the position of endothelial cells lining the inner surface of blood vessels, NO can be released into the blood stream, where it can act both locally and systemically. NO induces vasodilation by a reduction in the contraction of smooth muscle cells lining blood vessels. NO acts as a negative feedback for mean arterial pressure, since as arterial pressure increases, wall shear stress increases, inducing eNOS and increasing the NO concentration. As NO concentration increases, smooth muscle contraction is decreased, blood vessel lumen diameter increases, arteriole resistance decreases and arterial pressure decreases. The modulating action of wall shear stress on eNOS activity and NO production serves to maintain wall shear stress at a constant level. A diagram highlighting some of the interactions between NO, local metabolites, wall shear stress and smooth muscle contraction is shown in
Prolonged elevation of wall shear stress, in addition to activation of eNOS, leads to the transcriptional activation of the eNOS gene in endothelial cells. After several hours, eNOS enzyme levels increase due to the induced transcription of the eNOS gene. Increased levels of eNOS enzyme in endothelial cells increases those cells ability to release NO following induction of eNOS activity. It is thus expected that those cells which have experienced prolonged elevation of wall shear stress will have an increased ability to synthesize NO, and the same levels of wall shear stress will result in a greater synthesis of NO. One effect of increased eNOS levels is a reduction in the amount of wall shear stress that is required to induce biologically significant NO levels. Blood vessels that have been entrained by prolonged elevation of wall shear stress will release more NO relative to shear stress, and the vasodilation effect of NO will be increased, Higher relative NO concentration leads to reduced smooth muscle contraction, increased blood vessel lumen diameter and decreased arteriole resistance. Assuming that the cardiac output of the heart does not change, the net effect of a lower “set point” for responding to wall shear stress is a reduction in total peripheral resistance in blood vessels and a reduction in mean arterial pressure.
Similar to the effects of the statin drugs, an improvement in endothelial function is interconnected with LDL and cholesterol blood levels and NO bioavailability. LDL and cholesterol have been shown to prevent the down-regulation of eNOS. In turn, down-regulation of eNOS is apparently mediated by the stimulation of levels of caveolin-1 by LDL. Caveolin-1 is an important inhibitor of eNOS catalytic activity. Modulation of NO is expected to affect the interrelated blood lipid concentrations of VHDL, HDL, LDL, and cholesterol. To the extent that the pleiotropic activity of the cholesterol lowering statin drugs is modulated by NO levels, stimulation of NO bioavailability is expected to affect blood lipid composition. For additional background on the interrelationship between LDL and NO, see generally:
As described above, it has been known for some time that exercise can provide relief from hypertension in certain individuals. As the modulation of nitric oxide levels is part of a feedback system that responds in part to the stretching and extensibility of blood vessels of the body, it is hypothesized that exercise in general plays a role in stimulating cycles of NO release, and effectively providing some of the benefits of statin drugs, including improvement in endothelial function, increased nitric oxide bioavailability, anti-oxidant effects, anti-inflammatory protection, and stabilization of atherosclerotic plaques. Notwithstanding the effects of the modulation of NO bioavailability, exercise is known to modulate blood cholesterol and blood lipid composition.
There is widespread discourse on the relative benefits of particular forms of exercise. There is an ongoing need for patients suffering from hypertension, hypercholesteremia, atherosclerosis, and other cardiovascular and cardiopulmonary diseases to obtain the maximum benefit from the exercise utilized. Patients who are suffering from severe cardiovascular disease may be unable to engage in intense exercise, and many patients may be unable to engage in other forms of exercise due to limitations in time or facility availability. The invention disclosed herein provides for a device, system and method of exercise that can be optimized to provide an improved benefit to the patient in stimulating endothelial function, overall blood vessel health, and cardiovascular benefit while at the same time limiting the dangerous side effects of intensive exercise.
In contrast to the approach of Wiley, earlier subjects or trainees undergoing isometric exercise stressed the involved musculature to their full or maximum capability (publication (11)) or at some submaximal force as long as it could be sustained, in either case only terminating with the onset of unendurable fatigue. Such approaches often have incurred somewhat deleterious results as evidenced by the injuries sustained in consequence of improper weightlifting procedures. Weightlifting procedures or endeavors exhibit a significant isometric factor. See generally:
The diagnosis of patient hand-arm strength using isometric-based testing has been employed by physiologists, physical therapists and medical personnel for over three decades. These procedures function to evaluate hand-arm trauma or dysfunction and involve the patient use of a handgrip-based dynamometer. The dynamometer is grasped by the patient and squeezed to a maximum capability under the verbal instruction of an attending therapist or diagnostician. The hand dynamometer most widely used for these evaluations incorporates a grip serving to apply force through closed circuit hydraulics to a force readout provided by an analog meter facing outwardly so as to be practitioner readable. Adjustment of the size of the grip of the dynamometer is provided by inward or outward positioning of a forwardly disposed grip component. The dynamometers currently are marketed under the trade designation: “Jamar Hydraulic Hand Dynamometer” by Sammons Preston of Bolingbrook, Ill. An extended history of use of these dynamometers has resulted in what may be deemed a “standardization” of testing protocols. For instance, three of the above-noted grip length adjustments are employed in a standardized approach and verbal instructions on the part of the testing attendant, as well as the treatment of force data read from the analog meter are now matters of accepted protocol. In the latter regard, multiple maximum strength values are recorded whereupon average strengths, standard deviations and coefficients of variation are computed by the practitioner. In one test, the instrument is alternately passed between the patient's right and left hands to derive a maximum strength output reading each 1.5 seconds or 2.5 seconds. Reading and hand recording strength values for such protocols has remained problematic. The protocols, for example, have been the subject of recommendations by the American Society of Hand Therapist (ASHT) and have been discussed in a variety of publications including the following:
In about 1998, the above-noted Wiley protocols as described in connection with publication (12) above were incorporated in a compact, lightweight isometric device. Described in detail in U.S. Pat. No. 5,904,639 entitled “Apparatus, System, and Method for Carrying Out Protocol-Based Isometric Exercise Regimens” by Smyser, et al., the hand-held dynamometer has a hand grip which incorporates a load cell assembly. Extending from the hand grip is a liquid crystal display and two user actuated control switches or switch buttons. The display is mounted in sloping fashion with respect to the grip such that the user can observe important visual cues or prompts while carrying out a controlled exercise regimen specifically structured in terms of force values and timing in accordance with the Wiley protocols. This device is therapeutic as opposed to diagnostic in nature and is microprocessor driven with archival memory. External communication with the battery powered instrument is made available through a communications port such that the device may be configured by programming and, additional data, such as blood pressure values and the like may be inserted into its memory from an external device. Visual and audible cueing not only guides the user through a multi-step protocol but also aids the user in maintaining pre-computed target level grip compression levels.
Of course, it will be beneficial to incorporate improved diagnostic features for hand-arm evaluation techniques with therapist or practitioner designed therapeutic protocols specifically tailored to the condition of a given patient and which provide a control over such therapies clearly establishing such therapies as beneficial to strength development and recovery. One particular diagnostic and therapeutic feature that would be beneficial to incorporate is protocol that modulates wall shear stress of blood vessels so as to increase the bioavailability of nitric oxide and foster a reduction in total peripheral resistance in blood vessels and a reduction in mean arterial pressure, along with the other physiologic benefits associated with stimulation of NO signaling pathways, including reduction in LDL and cholesterol concentrations and an increase. in arterial flexibility.
The present invention is addressed to a system method and apparatus for carrying out a controlled isometric regimen by a user. Being microprocessor driven, the instrument is programmed to carry out established diagnostic as well as newly developed grip-based isometric regimens. When employed for carrying out a diagnostic maximum grip test, the diagnostician selects configuration parameters and the instrument provides both visual and audible prompts and cues throughout the procedure. Maximum grip forces for each of the sequence of trials of this procedure are selected typically by the diagnostician and when so selected are recorded in instrument memory along with calendar data, and processor computed values for average grip force, standard deviation of the force values throughout a sequence of tests and corresponding coefficients of variation. At the termination of the diagnostic procedure, memory recorded test data are displayable to the diagnostician and may be downloaded through a communications port to a computer facility.
For each of the diagnostic procedures, the widthwise extent of the instrument grip may be both varied in standard ½ inch increments from a minimum width. The grip is further configured such that the visually perceptible readout of the instrument may be viewed only by the diagnostician where deemed appropriate.
An important aspect of the therapeutic method associated with the instrument of the invention resides in the limiting of user performance to carry out the regimen of trials. In this regard, the instrument is programmed to perform only within predetermined and mandated test limits. Each therapeutic regimen is based upon an initial evaluation of the maximum gripping force capability of the user. Under that limitation, target load factors, hold on target load intervals, intervening rest intervals and trial repetition numbers may be elected only from pre-established and mandated memory retained ranges. The program also nominates rest intervals and hold on target intervals in correspondence with user elected target force factors. Thus, valuable strength recovery and development may be achieved but only within safe limits.
During each of the above therapeutic regimens, an audible warning is elicited whenever the user grip force value exceeds a computed upper limit. During each timed interval wherein the user is prompted to grip at a target force value computed with respect to the pre-tested maximum grip force, a dynamic bar graph and center point display is provided as a visual cue related to desired grip performance. Additionally, a rapid succession of score values are computed and the average thereof recorded at the end of each trial of a given regimen. These scores permit a therapist to access the quality of the performance of the user. In general, trial data is recorded in conjunction with calendar data and, as before, may be downloaded to a computer facility from an instrument contained communications port.
An additional object of the invention is to influence biological parameters of the user by selecting target loads that stimulate particular biological pathways. In one target force mode, the invention allows stimulation of nitric oxide bioavailability, which directly influences resting blood pressure and overall cardiovascular health. In other instances, the target force mode can be directed to maximize the exercise benefit for modulating blood lipid composition, including reducing low density lipoprotein (LDL) and cholesterol in the blood.
Other objects of the invention will, in part, be obvious and will, in part, appear hereinafter.
The invention, accordingly, comprises the method, system and apparatus possessing the construction, combination of elements, arrangement of parts and steps which are exemplified in the following detailed description.
For a fuller understanding of the nature and objects of the invention, reference should be made to the following detailed description taken in connection with the accompanying drawings.
Isometric exercise apparatus under which the methodology of the invention may be carried out is lightweight, portable, battery powered and sufficiently rugged to withstand the compressive pressures which it necessarily endures during use. The instrument is programmable such that it may be utilized by a therapeutic practitioner for diagnostic purposes employing established grip test modalities. Strength measurements carried out during these modes are compiled in memory and the practitioner is afforded calculated values for average grip force, standard deviation and coefficient of variation with respect to grip force trials. Furthermore, individual strength measurements compiled in these averages, whether taken rapidly or slowly, are stored in memory and may be reviewed by the therapist.
Additionally, the instrument is employable as a therapeutic device. First a protocol is nominated by prescribing nominal parameters of the effort. Each isometric regimen is controlled initially by requiring that a maximum grip strength be established for each individual patient or user. Then, the practitioner may elect parameters of grip force and timing under mandated memory contained parameter limits. Accordingly, the user will be unable to carry out strength enhancement therapies which would otherwise constitute an excessive grip force regimen. For carrying out the noted diagnostic procedures as well as therapy activities, the grip widthwise extent is variable from 1⅞ inches to 2⅞ inches, such variation being adjustable in ½ inch increments. This is in keeping with standardized diagnostic practices. Further with respect to diagnostic procedures, the display or readout of the instrument can be adjusted with respect to the grip structuring such that only the practitioner or therapist may observe the data which is being developed during a diagnostic protocol.
Looking to
Referring to
Looking to
Disposed centrally within the cavity defined by gripping portion sides 52 and 54 is a steel thrust plate 82 having a thickness and rigidity elected to withstand compressive gripping forces which may range, for example, up to about 205 pounds. Plate 82 is configured with two holes 81 and 83 which are used to restrain the plate from disengaging from the assembly when fitted over respective posts 53 and 55. Elongate side 84 of thrust plate 82 is configured for insertion within an elongate groove 86 of a base grip component 88. Grip component 88 is formed of a rigid plastic and includes an outwardly disposed base grasping surface 90 upwardly located in adjacency with the grasping surface 90 is one component of a base connector assembly represented generally at 92 and which is seen to be integrally molded with the grip component 88 and incorporates a slot or opening 94 in conjunction with a tab receiving trough 96. A tab component (not shown) of the base connector assembly feature of the base grip component 88 will be seen to extend from the end thereof opposite connector assembly component 92.
Two oppositely disposed edge extensions 98 and 100 of the trust plate 82 are configured for operative association with a load cell assembly represented generally at 102. Load cell assembly 102 includes an elongate steel base 104 incorporating two slots for receiving extensions 98 and 100, one such slot being revealed at 106. Connection between the base 104 and thrust plate 82 is provided by pins (not shown) which extend through mated bores 108 and 110 and 112 and 114. The load cell assembly 102 further includes an elongate outer force component 116. Two field plate-form load cells 118 and 120 are mounted from load cell mount structures shown, respectively at 122 and 124 formed within base 104. Such mounting is in cantilever fashion, the load cell 118 being attached to mount 122 by a screw and mounting plate assembly 126. Similarly, load cell 120 is attached in cantilever fashion to mount structure 124 by a screw and mounting plate assembly 128. Outer force component 116 is seen to have a centrally disposed rectangular post portion 120 which is attached by a connector plate assembly to the mutually inwardly extending ends of the load cells 118 and 120. the attachment plate assembly for this union is seen in general at 132. Assembly 132 is seen to be formed of two plate components 132a and 132b coupled, in turn, to load cells 120 and 118. Screws are used to effect the attachment.
The base grip component positioned oppositely of base grip component 88 is shown at 134. In similar fashion as component 88, the base grip component 134 is configured with a base connector assembly having one component at 136 which incorporates a slot and trough (not shown) in similar fashion as described at 92 in connection with component 88. a tab protrusion of generally cylindrical configuration shown at 138 is disposed oppositely from connector assembly component 136. The rigid plastic base component 134 is attached to elongate outer force component 116 of the load cell assembly 102. This attachment is provided by the insertion and crimping of two posts 134a and 134b (
Auxiliary grip component 34 is shown in the figure in spaced adjacency with respect to the base grip component 134. Auxiliary component 34 is configured with an outwardly disposed auxiliary grasping surface of generally half cylindrical cross section with a grasping surface profile curved concavely outwardly, for example, at region 140. This curvature is provided for enhancing grip contact with the palm of the user hand and for applying force centrally to the load cell assembly. Component 34 is formed with an auxiliary connector assembly which includes a flexible engaging tab 150 configured for insertion within the connector component 136 of base grip component 134. Connection at the opposite end is provided by a curved slot (not shown) which receives the tab protrusion 138 of base grip component 134. The connector assemblies are universal such that each of the auxiliary grip components may be mounted upon either of the base grip components 88 or 134. In this regard, not that a similar flexible engaging tab 152 is positioned upwardly upon auxiliary grip component 36. Similarly, the component 36 is configured having a curved slot 154 at its opposite end which receives tabs, for example, as at 138. The mounting of either auxiliary grip component 36 or 34 will increase the widthwise extent of the grip by one half inch. Accordingly, with both auxiliary grip components installed, the widthwise extent of the grip is increased to 2⅞ inches.
Interacting region 16 also includes a top cover 156. Formed, as the other components, of ABS plastic, the cover 156 includes a rectangular bezel opening 158 within which the LCD 18 is positioned. Integrally formed with top cover 156 is a downwardly depending switch cover 160 through which two rectangular openings 162 and 164 are provided. The switching function 20 is mounted upon a separate circuit board 166 which is seen to carry two push actuated switches as earlier described at 22 and 24 and identified by the same numeration in the instant figure. Located over the switches 22 and 24 is a flexible polymeric cover 168 formed of a flexible polymeric material such as Santoprene, a thermoplastic elastomer marketed by General Polymers of Charlotte, N.C. Circuit board 166 is supported between two slots formed in the interior of side components 56 and 58, one of these slots is seen at 170. The LCD 18 is mounted upon a circuit board 172 supported in turn, from interactive components 56 and 58. A bus-type wiring harness electrically associates the switching function 20, LCS 18, load cell assembly 102, the battery within compartment 74 and the circuitry carried by circuit board 76.
A sectional view of the instrument 10 is provided at
In
Turning to
Load cells 118 and 120 are represented with that numeration in
Each of the instruments 10 is calibrated using nineteen combinations of six standard weights. A best fit is determined and the instrument is called upon to have a root mean square error (RMS) of 0.1 pounds or less to pass calibration requirements. Once the calibration constants has been determined, the system is loaded with two redundant copies of the calibration constants. The zero point of the load cell is monitored at all times during the use of the instrument 10. If a drift is found, then a warning is shown at the LCD display 18. If any lead wire to the load cell becomes disconnected, then the built-in monitoring detects this occurrence, shows an error message, and disables further use of instrument 10 until the power is reset. These features insure that the force reading shown is accurate and true. Absolute values of the outputs of load cells 118 and 120 are summed to provide a force output signal. In general, the load measurement accuracy of instrument 10 is better than 0.1 pound of 0.1% of applied force whichever is greater.
In the discourse to follow, the sequences of the program protocol carried out by instrument 10 are represented in flow chart fashion. In general, these flow charts commence with a configuration sequence if desired and then look to two diagnostic protocols followed by two therapeutic protocols.
Turning to
With the publication of the screen as represented at block 246, then as represented at line 252 and block 254 the practitioner or user is called upon to determine whether to enter a configuration sequence or to progress to a diagnostic grip test. To enter the latter diagnostic grip test sequence, as represented at line 256 and block 258 by pressing switch 24 display 18 will prompt the user to press the select switch 24 to commence a diagnostic grip test sequence. Where the select switch 24 is actuated, then the program enters the diagnostic grip test sequence as represented at line 260 and node A.
Where a determination on the part of the practitioner or user is made to enter a configuration sequence, then as represented at line 262 and block 264 the configuration sequence is entered by actuating switch 22. As represented at line 266 and block 268 the initial configuration looks to units. Recall from block 246 that the instrument 10 defaults to a units evaluated in pounds. As represented at line 270 and block 272 by actuating select switch 24 the units parameter can be converted to kilograms instead of pounds. The program then continues upon depressing or actuating menu switch 22 as represented at either lines 274 or 276 leading to block 278. As represented at block 278, the user then is given the opportunity to delete the audible tone. In this regard, by actuating select switch 24, as represented at line 280 and block 282, the tone is deleted, display 18 showing the term “tone” in connection with the letter N.
The configuration sequence then continues as represented at either lines 283 or 284 with the actuation of menu switch 22. This actuation of switch 22 provides for the establishing of a rapid exchange diagnostic test cycle time change. As set forth at block 286 the default cycle time is 1.5 seconds. However, by actuation of select switch 24, as represented at line 288 and block 290 the operator may change the cycle time to 2.5 seconds. The program then continues by actuating the menu switch 22 as represented at either of lines 292 or 294. These lines lead to the configuration alteration represented at block 296. Recall from block 246 that the default number of exchanges for the rapid exchange diagnostic procedure is 10.
However, as represented at line 298 and block 300 the operator may change the number of exchanges from 10 to 20 by actuation of select switch 24. The program then returns to line 244 by actuation of the menu switch 22 as represented at lines 302 and 304. As described in connection with block 258, line 260 and node A, the operator may elect to proceed with a diagnostic grip test.
Referring to
Returning to block 316, the maximum strength grip test can be carried out with 10 maximum squeezing force trials. At the conclusion of a given number of such trials, the practitioner actuates select switch 24, whereupon computations are carried out. Accordingly, as represented at line 320 and block 322 the user is prompted with the message “squeeze hard!!!” at the readout 18. The program will elect the highest force applied during such squeezing activity, whereupon the user releases the grip force as represented at line 324 and block 326. Then instrument 10 will publish the maximum force applied by the user as represented at line 328 and block 330, a first maximum grip evaluation being shown as an example as 64.4 pounds. Block 330 also indicates that the user is prompted to either actuate the select switch 24 to accept the published maximum squeeze evaluation as set forth at block 330 or to squeeze the grip 14 again. Such squeezing again will provide a substitute maximum grip force evaluation. Then, as represented at line 332 and block 334 the query is posed as to whether the select switch 24 has been actuated. In the event that it has not, then the program loops as represented at line 336 extending to line 320, whereupon a maximum grip effort again is undertaken. Where the operator elects the maximum first trial grip force evaluation, then as represented at line 338 and block 340, the program will compute an average of force values, standard deviation and coefficient variation, albeit it for one trial at this junction in the procedure.
The program then continues as represented at line 342 and block 344 to display computed values which, as noted above, for the first trial are irrelevant. However, as the number of trials increases, those computed values gain significance. Next, as represented at line 346 and block 348 the program commences to carry out a next maximum grip test by providing a prompt at readout 18 which advises the user to “squeeze hard!!!” and indicates that this is a second trial as represented by the terms: “MAX 2”. Following a squeezing of the grip region 14, as represented at line 350 and block 352 the user releases the grip force and, as represented at line 354 and block 356 the maximum force asserted by the user is published, for example, showing 60 pounds for a “MAX 2” trial. This prompt further advises the user to actuate select switch 24 to elect the published grip force value or to squeeze again to carry out a next trial. The program then continues as represented at line 360 and block 362 to determine whether or not select switch 24 had been actuated. In the event that it had not been actuated then the program loops as represented at lines 364 and 346 whereupon the user again may carry out the second maximum grip trial. Where switch 24 has been actuated, then as represented at line 366 and block 368, the program carries out a computation of the average of the maximum forces asserted and computes standard deviation and coefficient of variation which are submitted to memory. The program then continues as represented at line 370 and block 372 whereupon the values computed in connection with block 368 are published at display 18. The above maximum grip test trials may be reiterated for 10 trials. Accordingly, as represented at line 374 and block 376 the maximum test trials are reiterated for a total of N tests (10 maximum) and the computed values of average force, standard deviation and coefficient of variation are both submitted to memory and published at display 18. As represented at line 378 and block 380 the user may restart this max test sequence following the Nth trial by actuating select switch 24, whereupon the program returns as represented at line 382 to 310 (
Returning to
Looking back to the query posed at block 334, where the menu switch 22 is actuated as opposed to electing a maximum grip value, then as represented at line 410 and block 412 the program will reconfigure for restarting the grip test mode. Once at this point in the program as represented at block 412, by again actuating select switch 24, the program reverts as represented at line 414 to line 320 to carry out another maximum grip trial. On the other hand, where menu switch 22 is actuated, as represented at line 416 and block 418 an indication will be given to the operator that to elect previous menu, select switch 24 is to be actuated. As represented at line 419, the program then reverts to node C. Node C again appears in
Looking again to
The diagnostic performance mode of the instrument 10 also provides for the carrying out of a rapid exchange (RE) test. With the rapid exchange test, the user may grip instrument 10 in the manner shown in
As represented at line 458 and block 460 the user will have squeezed the grip region 14 and the maximum hand force value evolved will be submitted to memory. Then as represented at line 462 and block 464 a determination is made as to whether the menu switch 22 has been actuated. In the event that it has not, as represented at line 466 and block 468 the program determines whether the Nth, i.e., 10th or 20th trial has been completed. In the event that it has not, then as represented at line 470 and block 472 the rapid exchange test has not been completed and an audible tone cue (time hack) is provided indicating that the instrument should be switched to the opposite hand. A short dwell occurs as represented at line 474 and block 476 wherein the instrument determines whether or not a squeeze force has been asserted. In the event that it has not, then the program loops as represented at line 478. Where the user has imparted a squeezing force to the instrument, the program continues or loops as represented at line 480 extending to line 458 leading to a next trial in an alternate hand.
Returning to block 464 where menu switch 22 is actuated in the course of carrying out rapid exchange trials, an affirmative determination will be made with respect to the query posed at that block. Accordingly, as represented at line 482 and block 484 the user is prompted to restart the rapid exchange test by actuating select switch 24. Where select switch 24 is actuated, then as represented at line 486 the program reverts to line 444 and block 446. On the other hand, where menu switch 22 is actuated, then as represented at line 488 and block 490 the user is prompted to revert to the previous menu by actuating select switch 24. Where select switch 24 is so actuated, then the program reverts to node C as represented at line 492. Note, additionally, that if menu switch 22 is actuated in conjunction with the prompt provided at block 442, then as represented at line 494 the program reverts to line 488. Returning to block 490, where menu switch 22 is actuated then as represented at line 496 and block 498 the program computes and displays the overall average of the maximum trial values, standard deviation and coefficient of variation for the N trials. That data is submitted to memory. Should menu switch 22 be actuated at this juncture, then as represented at lines 500 and 482, the program returns to block 484. Where the select switch 24 is actuated, however, as represented at line 502 and block 504 the maximum force value for trial N and the average SE and CD for all trials is displayed. On the other hand, where the menu switch 22 is actuated, then as represented at lines 506 and 482, the program reverts to block 484.
Where the select switch 24 is actuated repetitively, then as represented at line 508 and block 510 the succession of trials 1 through N is displayed. Additionally, the unchanging average for all those trials is displayed for convenience. Further, a query is posed as to whether the Nth trial has been displayed. Where it has not, then the display program loops as represented at line 512 extending to line 502. On the other hand, where the Nth trial has been displayed, then as represented at line 514, the program loops to line 502 to repeat the succession of displays.
It may be recalled that in conjunction with block 398 in
Following the target load computation, as represented at line 572 and block 573 the program displays the newly computed target force value at readout 18 along with the default values for number of repetitions (which defaults at 4), and the nominated hold on target interval and the rest interval (Table 1). As a prompt, the readout “4 REP” blinks to indicate that adjustment is available to the user. The program then continues as represented at line 574 which reappears in
Turning momentarily to
Returning to
Looking momentarily to
Returning to
Node G reappears in conjunction with line 708 (
Referring to
The number of steps elected adjusts the percentage of maximum grip force factor in accordance with a preordained schedule. That schedule is provided in Table 2 below. For example, if only one step is elected, that target grip factor will be 20%. On the other hand if five steps are elected, the first trial will be at 100% of maximum grip force. The second step will be at 80% of maximum grip force and so forth. On the other hand, if four steps are elected, the initial trial will be in conjunction with an 80% maximum grip force factor; the second step will be at 60% and so forth as set forth in Table 2. For each of these percentages as set forth in Table 2, the corresponding hold on target or effort interval and rest intervals will follow the values given above in Table 1.
The step value is elected by actuation of select switch 24 and the program continues as represented at line 758 and block 760. Block 760 replicates a display at readout 18 which prompts the user by indicating that the maximum elected gripping force selected was 90 pounds and that A steps were selected and a further prompt is provided showing blinking or intermittent display of “4 REPS”. Then, as represented at line 762 and block 764 the operator may adjust the number of repetitions of the program to a value within a preordained number of 1 through 10 by actuating menu switch 22. The elected number of repetitions then is selected by actuation of switch 24 and, as represented at line 766 and block 768 the system displays the now selected parameters of a maximum grip force, for example, 90 pounds, an election of A steps in the regimen and an election of “B” repetitions. Next, as represented at line 770 and block 772 the stepped exercise therapy is entered. Upon entry into this stepped exercise trial mode, target values are computed based upon the number of steps elected and the hold on target and rest intervals will be acquired, such data with respect to target factors being set forth in Table 2 and the latter hold on target and rest intervals being set forth in Table 1. This function is represented in block 776. Line 778 reappears in
Returning to block 780, where switch 24 has been actuated, then as represented at line 798 and block 800 the user is prompted to hold the grip force at the computed target level for 100%. Additionally, the prompt term “SQUEEZE” is provided within the readout 18. Next, as represented at line 802 and block 804 a determination is made as to whether the grip force exerted by the user is within 10% of the computed target value. Where it is not, then the system dwells as represented at loop line 806 and the display represented at block 800 continues. Where the asserted grip force is within 10% of the target load, then as represented at line 808 and block 810 the mandated hold on target interval timeout set forth in Table 1 commences and, as represented at line 812 and block 814 a dynamic comparison value is derived for dynamic bar graph cueing. Next, as represented at line 814 and block 816 a computation then is made as to whether the instantaneous grip force is at or above 125% of the target value. Where that is the case, then as represented at line 820 and block 822 an audible warning cue is sounded. The program then continues as represented at lines 824 and 826 when the excessive force has been lessened. Line 826 is directed to block 828 which provides for carrying out a computation of a score value as a percentage of target for a sequence of time increments. Computation of this score has been discussed in connection with
Line 830 reappears in
Following the generation of this audible cue, as represented at line 870 and block 872 the program reiterates the trial sequence following the mandates of Tables 1 and 2 and the elected parameters. As represented at line 874 and block 876, a query then is made as to whether the repetitions and associated efforts are complete. This value is the product of the elected number of steps A multiplied by the elected number of repetitions, B. Where that number of reiterations has not occurred, then the program continues as represented by look line 878 extending to line 870. Where the number of repetitions is completed, then as represented at line 880 and block 882 a final score is computed and submitted to memory with calendar and force data. Next, as represented at line 884 and block 886 the program selects a message to the user which will be based upon the final score. For example, the user may be advised to consult a therapist or the program directions in the event of the low score and is congratulated in the event of a good score. As represented at line 888 and block 890 those messages are selected. Where the user actuates select switch 24, the program continues as represented at line 892 and node H.
Turning again to
The user has the option of powering down instrument 10 by pressing select switch 24 for an interval of at least 2 seconds. This power off sequence is represented in the flow chart of
The protocol based isometric exercise approach of the invention has applicability to a broad range of muscle groups of the user. By employing the protocol which, inter alia, involves the evaluation of maximum muscle group strength as a precondition to then applying a factor related protocol, one of those factors may apply to the measured maximum strength value. The remaining factors which involve, for example, variations of target loads, hold times, rest intervals and exercise regimen planning in terms of calendar days achieves a safe and effective utilization of isometric activities. The exercisable anatomical features to be strengthened are generally identifiable as muscle groups of the human anatomy which may include but are not limited: jaw muscles, neck muscles, shoulder muscles, upper arm muscles, lower arm muscles, hand muscles, finger muscles, diaphragm muscles, abdominal muscles, lower back muscles, upper leg muscles, lower leg muscles, ankle muscles, foot muscles, and tow muscles.
Looking to
Through use of the invention, cardiac function and a variety of physiologic effects are produced, including effects on the endothelium and the release of biological active signaling molecules, including nitric oxide. The following two studies demonstrate the measureable biochemical and biophysical effect of utilization of the system, method, and apparatus of the invention.
Among many other factors, both hypertension and arterial distensibility are independent risk factors for cardiovascular disease. The research of Wiley et al. (1992) and Taylor et al. (2003) demonstrated that isometric training is effective for reducing resting blood pressure (RBP). NO is a potent vasodilator, and crucial component of the regulation of vascular tension (See
As NO is a rapidly diffusible gas, release of NO in the blood vessels of the arms or other muscle groups, in response to isometric training is expected to have both a local and systemic effect on vasodilation, arterial distensibility and resting blood pressure. Thus an increase of arterial distensibility in response to isometric exercise may contribute to reduction in RBP and increased NO bioavailability. A wide variety of muscle groups such as from exercisable regions of the musculature of the user including jaw muscles, neck muscles, shoulder muscles, upper arm muscles, lower arm muscles, hand muscles, finger muscles, diaphragm muscles, abdominal muscles, lower back muscles, upper leg muscles, lower leg muscles, ankle muscles, foot muscles, and toe muscles may provide a therapeutic benefit by utilization of the isometric exercise protocols of the invention.
To demonstrate the systemic effect of practicing the system and method off the invention, the impact of isometric arm and leg exercise on RBP and central and peripheral arterial distensibility was tested in patients being medicated for hypertension. Resting blood pressure was measured by brachial oscillometry, and arterial distensibility, as measured by Doppler ultrasound and applanation tonometry in the carotid, brachial and femoral arteries. Study participants were directed to perform isometric handgrip (IHG) exercise (n=10), or isometric leg press (ILP) exercise (n=6) according to the method of the invention three times per week for eight weeks. Exercise intensity was maintained at 30% of maximal voluntary contraction.
Following eight weeks of IHG exercise, systolic blood pressure decreased significantly (from 140.2 mmHg+/−3.82 to 132.3 mmHg+/−3.97), while no decrease was observed after isometric leg press exercise. Diastolic blood pressure did not change after either IHG or ILP exercise. Measurement of carotid arterial distensibility showed a significant improvement following IHG exercise (from 0.1105 mmHg−/−1×10−2 0.0093 to 0.1669 mmHg+/−1×10−2 0.0221), while no such changes occurred in the ILP exercise group. Peripheral arterial distensibility did not change following either IHG or ILP exercise. These studies demonstrate that the isometric handgrip exercise according to the invention improves resting systolic blood pressure and carotid arterial distensibility. As arterial tension is under the direct control of the NO/LDL-cholesterol signaling system, the system and method of the invention allows modulation of NO and indirectly of the LDL-cholesterol components.
As described previously, hypertension is associated with endothelial dysfunction, reduced NO bioavailability, and the development of coronary artery disease among other effects on the body of the patient. The isometric hand grip exercise protocol of the invention further reduces blood pressure even in patients already medicated for hypertension. In order to demonstrate the mechanisms of IHG affect on hypertension, endothelial function was studied in patients practicing the system and method of the invention. Study participants (n=8, 62+/−3.5 years) performed 4 sets of 2-minute isometric contractions at 30% of their maximal voluntary contraction. Ulnar reactivity was assessed in alternate hands, 3×/week for 8 weeks. Resting blood pressure was measured using automated brachial oscillometry. Vascular reactivity was measured in both arms using ultrasound imaging to determine brachial artery flow-mediated dilation (FMD). Following utilization of the IHG protocol of the invention, systolic blood pressure decreased (137 mm Hg+/−5.3 to 121.7 mm Hg=/−4.8 mmHg, p=0.03). Relative FMD increased (1.6%+/−0.3 to 4.5%+/−0.5 and normalized to average shear rate, 0.007%+/−0.001 to 0.02+/−0.004%/s−1). Reactive hyperemic flow decreased (peak, 344.3+/−36.5 to 258.2+/−27.2 ml/min and average, 301.6+/−33.1 to 239.0+/−28.4 ml/min). Average resting blood vessel diameter and resting flow rates remained unchanged. As systemic shear stress is known to induce the activity of NO as a vasodilator, the IHG training apparently causes the release of NO, as shown by an increase in flow mediated arterial dilation. The IHG exercise protocol produced a reduced reactive hyperemic flow, accompanied by improvements in normalized FMD, and a heightened vasoreactive sensitivity to the reactive hyperemic stimulus. The IHG protocol by providing an improved cardiovascular function, demonstrates a modification of the wall shear stress setpoint for the activation of eNOS to produce biologically active levels of NO.
The statin class of drugs used in the treatment of hypercholesterolemia surprisingly has a pleiotropic effect on a variety of other systems of the body, including on the bioavailability of NO. Thus, by down modulating cholesterol biosynthesis, statin drugs affect systems that are controlled by NO dependent signaling systems. The method and apparatus of the invention, surprisingly, by stimulating changes in the structure of the vasculature, and by creating increased wall shear stress in the blood vessels experiencing the effects of the inventive protocol also induces broad effects on signaling systems including those that regulate the bioavailability of NO and serum cholesterol and LDL-cholesterol levels.
Since certain changes may be made in the above-described apparatus, method and system without departing from the scope of the invention herein involved, it is intended that all matter contained in the description thereof or shown in the accompanying drawings shall be interpreted as illustrative and not in a limiting sense.
This application is a continuation-in-part of copending application Ser. No. 10/268,363 filed Oct. 10, 2002 and claims the benefit of Provisional Application No. 60/330,265, filed Oct. 18, 2001.
Number | Name | Date | Kind |
---|---|---|---|
4884445 | Sadoff et al. | Dec 1989 | A |
4949729 | Haski | Aug 1990 | A |
5398696 | Wiley | Mar 1995 | A |
5904639 | Smyser et al. | May 1999 | A |
6026684 | Calder | Feb 2000 | A |
6217504 | Phillips | Apr 2001 | B1 |
6358187 | Smith | Mar 2002 | B1 |
6962569 | Smyser et al. | Nov 2005 | B2 |
20050101461 | Johnson | May 2005 | A1 |
20080132388 | Clem et al. | Jun 2008 | A1 |
Number | Date | Country |
---|---|---|
WO87 02567 | May 1987 | WO |
WO2004 032701 | Apr 2004 | WO |
Number | Date | Country | |
---|---|---|---|
20060035762 A1 | Feb 2006 | US |
Number | Date | Country | |
---|---|---|---|
60330265 | Oct 2001 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 10268363 | Oct 2002 | US |
Child | 11204001 | US |