The invention relates to an on-chip laboratory comprising a coplanar micro-fluidic network and electronebulization nose. In particular, it relates to the coupling of an on-chip laboratory with a mass spectrometer.
For now ten years, many ways have been explored for coupling different micro-fluid devices to mass spectrometers. Indeed, optical detection methods like spectrophotometry or fluorescence are not suitable for detecting biomolecules such as proteins or peptides, a detection which particularly concerns the field of proteomics. The limits are either the sensitivity or the necessity of preparing the sample (fluorescent marking), which, in the case of identifying proteins after enzymatic digestion, has a problem as the obtained peptides are a priori not known. Mass spectrometry is therefore often retained as it gives information on the nature of the analyzed samples (intensity spectrum according to the mass/charge ratio) with very good sensitivity (femtomol/μl), and it enables complex mixtures of molecules to be analyzed. For this, it is often necessary that pre-treatment of the sample be performed upstream from the analysis. For example, this pre-treatment consists in separating chemical and/or biological compounds, preceded and/or followed by concentration of the species.
In order to perform this continuous pre-treatment together with the analysis in a minimum time and minimize the volumes of reagents used, it is possible to take advantage of the recently accomplished advances in the field of microfluidics. As examples, microfluidic devices for enzymatic digestion (Lian Ji Jin, “A microchip-based proteolytic digestion system driven by electroosmotic pumping”, Lab Chip, 2003, 3, 11-18), for capillary electrophoresis (B. Zhang et al., “Microfabricated Devices for Capillary Electrophoresis-Electrospray Mass Spectrometry”, Anal. Chem., Vol. 71, No. 15, 1999, 3259-3264) or for 2D separation (J. D. Ramsey, “High-efficiency Two-dimensional Separations of Protein Digests on Microfluidic Devices”, Anal. Chem., 2003, 75, 3758-3764 ou N. Gottschlich et al., “Two-Dimensional Electrochromatography/Capillary Electrophoresis on a Microchip”, Anal. Chem. 2001, 73, 2669-2674) have already been presented.
The microfluidics/mass spectrometry coupling may be based on a technique for ionizing the sample by electronebulization or electrospray (ElectroSpray Ionization (ESI)). At atmospheric pressure and immersed in a strong electric field, the pre-treated liquid sample leaving the microfluidic chip is nebulized into a gas of ions or into a multitude of charged droplets which may enter the mass spectrometer (SM) for analysis. This nebulization requires deformation of the interface formed between the leaving liquid and the surrounding gas (liquid/gas meniscus) and the liquid <<drop>> assumes a conical shape called a <<a Taylor cone>>. The volume of this cone forms a dead volume for the leaving liquid (a geometrical space in which the chemical compounds may mix), which is not desirable, especially when the last pre-treatment step in fact consists in separating the chemical compounds from the sample. This is why one always seeks to minimize the size of this cone, and this requires i.a. reduction of the inner and outer dimensions of the outlet channel of the microfluidic chip.
Conventionally, during analysis by mass spectrometry, the sample is pre-treated <<outside the ESI device>> and manually placed (with the pipette) in a hollow needle, the end of which is electrically conducting (the <<PicoTip emitter>> from New Objective for example). An electrical field is imposed between the conducting portion of the PicoTip and the entrance of the SM, with which a Taylor cone may be formed at the outlet of the PicoTip and the sample nebulized. The <<pointed>> cylindrical geometry of the PicoTips is ideal for forming a small Taylor cone, but the limits on minimization of their size (conventionally with an outer diameter of 360 um and an inner diameter of 10 μm), those limits on obtaining good reproducibility with the manufacturing techniques used (drawing process) and their brittleness upon use are the main reasons for trying to make other types of nebulization devices.
In the literature, when these devices are developed with micro-technologies such as silicon planar technologies for example (etching, machining, thin film depositions and photolithography of different materials on substrates having very large side dimensions relatively to their thickness), these are often designated as <<electrospray nozzles>> (Tai et al., “MEMS electrospray nozzle for mass spectroscopy”, WO-A-98/35376). The stake of such realizations is double.
On the one hand, with micro-technologies, ESI interfaces may be made by defining tip type structures (like the PicoTips) but smaller (to limit the volume of the Taylor cone), more reproducible and less brittle structures, which are of interest per se (see document WO-A-00/30167).
On the other hand, with micro-technologies, devices may be made which integrate a fluidic network with which pre-treatment of the sample and an interface of the ESI type may be provided. In addition to the aforementioned advantages (reduction in the output dead volumes, reproducibility, robustness of the ESI interface), one benefits from those related to an integrated pre-treatment device (continuous pre-treatment protocol with analysis, reduction in the global analysis times, minimization of the volumes of reagents).
Nevertheless, such integration poses three major technological design problems:
As early as 1997, R. S. Ramsey and J. M. Ramsey (“Generating Electrospray from Microchip Devices Using Electroosmotic Pumping”, Anal. Chem., 1997, 69, 1174-1178) proposed a glass microfluidic chip, the liquid flows of which are generated by electro-osmotic pumping and the output channel of which opens into the section of the component with a planar geometry. Under the assistance of overpressure imposed upstream, a sample drop of 12 nl forms at the chip outlet, which drop under the action of a strong electric field forms a Taylor cone and is nebulized. This simple approach poses the problem of a significant liquid dead volume (12 nl), whence a sensitivity limit of the device.
More recently, K. Huikko et al. (“Poly(dimethylsiloxane) electrospray devices fabricated with diamond-like-carbon poly (dimethylsiloxane) coated SU-8 masters”, Lab Chip, 2003, 3, 67-72) proposed a poly(dimethylsiloxane) (PDMS) chip, it also having opening channels intended to be put opposite a SM for nebulizing the sample. The authors take advantage of the hydrophobicity of PDMS for obtaining a small Taylor cone, whence the limitation of the dead volume at the outlet. Nevertheless, the proposed device does not integrate any nebulization electrode. The tests were carried out by using a platinum wire dipping in the inlet reservoir of the ESI channel; this cannot be a good solution, i.e., without adding any dead volume, for possible integration to a pre-treatment fluid network. Moreover, PDMS technology remains a limited technology which does not yet allow the design of complex microfluidic networks with a characteristic size of the order of one micrometer. This imposes a strong limitation as to the pattern of the microfluidic entities required for pre-treatments of samples (concentration, separation . . . ).
M. Svederberg et al. (“Sheathless Electrospray from Polymer Microchips”, Anal. Chem., 2003, 75, 3934-3940) propose polymer devices which have very interesting geometries for making electrospray nozzles (2D or 3D tips) but the dimensions of the outlet channel (width 100 μm×height 70 μm) obtained by their machining technology are redhibitory for making a device with small dead volumes. Indeed, it is recalled that the outlet diameter of a PicoTip is only 10 μm. In addition, the use of polymer materials imposes strong limits as to possible chemical for biological functionalizations of the internal walls of the outlet channel or of a possible sample pre-treatment fluidic network. Indeed, up to now, most of these functionalizations were developed on silicon or on glass. Moreover, the proposed manufacturing technology is not collective and the nebulization electrode is made on the outer portion of the ESI tip.
V. Gobry et al. (“Microfabricated polymer injector for direct mass spectrometry coupling”, Proteomics 2002, 2, 405-412), J. Kameoka et al. (“An electrospray ionization source for integration with microfluidic”, Anal. Chem. 2002, 74, 5897-5901) and J. Wen et al. (Electrophoresis 2000, 21, 191-197) also propose making electrospray nozzles in polymer materials with a two-dimensional geometry suitable for forming a stable Taylor cone and limiting the dead volumes. But the technology used does not propose the integration of a nebulization electrode. The tests are made by means of a gold wire dipping in an inlet reservoir of the device.
Another approach consists in adapting the outlet of the separation channel so as to be able to receive a commercial PicoTip (Y. Tachibana et al., “Robust and simple interface for microchip electrophoresis-mass spectrometry”, J. of Chromatography, 1011 (2003), 181-192). This requires the use of a metal and/or plastic part playing a linking role in assembling both entities. This kind of assembly has significant dead volumes and does not solve the problem of using commercial PicoTips having a certain irreproducibility in dimensions and great brittleness upon use.
Two documents from a team of the California Institute Of Technology may also be mentioned: Tai et al., “MEMS electrospray nozzle for mass spectroscopy”, WO-A-98/35376 and Tai et al., “Polymer-based electrospray nozzle for mass spectrometry”, WO-A-00/30167. The claimed technologies for making an electrospray nozzle provided with an upstream filter are surface technologies with which hollow structures may be made in silicon nitride in the first case and in parylene in the second. These surface technologies are based on the use of a sacrificial layer (in phosphosilicate glass (PSG)), which as indicated by its name, is not retained up to the end of the technological continuity. Removal of this layer, performed by chemical etching, determines the hollow structures. From a geometrical point of view, these technologies are of interest (tip shapes of the nozzle), but they do not propose integration of nebulization electrodes and the authors use the standard way with a platinum wire dipping in an inlet reservoir in order to test their system, which is redhibitory for obtaining a complete fluidic system (pre-treatment and electrospray nozzle) with small dead volumes.
Finally, J. E. Moon et al. in U.S. Pat. No. 6,464,866 claim a chemical analysis system made with micro-technology from two substrates, preferably in silicon, and comprising a liquid chromatography system and an electrospray device. The device disclosed in this document includes a tip of the electrospray nozzle, perpendicular to the plane of the substrates used. So this arrangement does not prevent dead volumes due to changes in direction.
The present invention proposes a microfluid device allowing different treatments of samples and having a good interface with an ESI type mass spectrometer, which requires:
The object of the invention is therefore an on-chip laboratory comprising a support, at least one fluidic network, at least one inlet fluid orifice connected to the fluidic network and at least one fluid outlet orifice connected to the fluidic network, a thin layer integral with the support and in which the fluidic network and an electronebulization nozzle are made, the electronebulization nozzle overhanging relatively to the support and comprising a channel, an end of which is connected to the fluidic network and the other end of which forms said fluid outlet orifice, the channel being fitted with electrical conduction means forming at least one electrode, characterized in that the thin layer is a layer fixed by direct sealing on the support.
The rear face of the support, i.e., the one which does not support the thin layer, may advantageously be of an inert nature. It is then not involved in the operation of the device. In particular, it does not then have any electrical connection.
If the support is a semiconducting material, the electric conduction means may be a doped portion of said support. The support may be in a conducting material.
This laboratory may comprise a cover sealably covering the fluidic network, this cover being provided with a fluid access means at the fluid inlet orifice.
According to another arrangement, the on-chip laboratory may comprise a cover sealably covering the fluidic network, this cover being provided with a fluid access means at the fluid inlet orifice and provided with said electric conduction means.
The cover may be in a conducting material. It may be in a semiconducting material, the electric conduction means may then comprise a doped portion of the cap.
By using a cover, it is possible to seal off the fluidic network.
The electric conduction means may therefore be located both in the support and in the cover and may also be made either with the support or the cover in a conducting material, or with metal tracks deposited on the support or the insulating cover, or may be doped portions of the support or the cover in semiconducting material.
The invention will be better understood and other advantages and features will be apparent upon reading the description which follows, given as a non-limiting example, accompanied by the appended drawings wherein:
The Fluidic Network
Firstly the fluidic network intended for preparing a complex biological sample in order to identify the proteic contents is described. This fluidic network consists of a set of reservoirs and channels, an enzyme digestion reactor, a pre-concentration reactor and a reactor for separation by liquid. electro-chromatography. The basic structure of all these reactors is a deep cavity provided with a large number of square or hexagonal section pads . . . . This kind of structure is known as COMOSS (“COllocated MOnolith Support Structures”). Reference may be made on this subject to the article of Bing He et al. entitled “Fabrication of nanocolumns for liquid chromatography”, Anal. Chem. 1998, 70, 3790-3797. For all these reactors, advantage is taken of the large surface/volume ratios developed by these COMOSS structures, ratios which increase the probabilities of encounter between the molecules of the mobile phases (for example proteins for the enzyme digestion reactor) and those of the stationary phases (trypsin for the enzyme digestion reactor).
After complete pre-filling of the fluidic network with buffer, the biological sample (protein) is deposited in the reservoir R1, and then pumped by electro-osmosis from the reservoir R1 to the reservoir R2 through the enzyme digestion reactor 2. Reservoirs with large volumes are positioned between the different reactors of the fluidic network in order to provide a change of buffer between two consecutive steps of the protocol. Thus, R1 contains ammonium bicarbonate ([NH4HCO3]=25 mM; pH=7.8), R2, R3 and R4 contain a water/acetonitrile ACN/formic acid TFA (95% ; 5% ; 0.1%) mixtures, whereas R5 contains a water/acetonitrile/formic acid (20% ; 80% ; 0.1%) mixture. The recovered digest in the reservoir R2 has to be concentrated before separation. For this, it is pumped by electro-osmosis towards the reservoir (R3) (rubbish bin). The whole of the peptides resulting from the enzymatic digestion is then <<captured>> by the small volume pre-concentration reactor 3, whence the concentration. An acetonitrile gradient made by mixing the buffer of R4 with that of R5 in the <coil>> type (length 2 cm) structure 4, will then selectively “unhook” the peptides according to their affinity with the stationary phase (for example C18) of the pre-concentration reactor 3. The latter are again <<captured>> by the chromatography column 5, denser than the pre-concentration reactor 3. By enriching the mixture with CAN, it is again possible to selectively “unhook” these peptides from the chromatography column 5, and to carry them off separated, towards the outlet 6 of the chip 1 where the liquid is nebulized towards the inlet of a mass spectrometer not shown.
A reactor with affinity to a given protein (not shown) may be used for sensing the latter in a multi-proteic mixture conveyed through this reactor. For this, a reservoirs/affinity reactor/concentration reactor assembly operating according to the same fluidic principles as described above may be integrated upstream from the fluidic network as described above. The affinity reactor may be functionalized with antibodies and the elution buffer may consist of proteins which are competitive (with regard to the antibody) with the one which one desires to <<capture>> in the multi-proteic complex.
✓ The Upstream Affinity Reactor
With a COMOSS structure, it is intended to specifically sense a protein, a family of proteins, or a multi-proteic complex in the complex biological sample. The tools used for this step may be antibodies, but also small molecules for example which have an interaction specificity with the sought-after protein(s).
✓ The Enzyme Digestion Reactor
The COMOSS structure of the enzyme digestion reactor, illustrated in
With this structure, it is optionally possible to organize functionalized (with trypsin for example) <<beads>> of silica of a few micrometers (beads from Bangs Laboratories distributed by Serotec France par example), in order to provide the reactor with its enzymatic properties or to enhance them.
As an example, the enzyme grafted on the pads may be trypsin. The protocol used is the one described in document FR-A-2 818 662.
✓ The Pre-Concentration Reactor
The COMOSS structure of the pre-concentration reactor, illustrated in
With this structure, it may be possible to organize functionalized silica beads in order to provide the reactor with affinity properties or enhance them (C18 grafting for example).
✓ The Reactor for Separation by Liquid Electrochromatography
The COMOSS structure of separation reactor, illustrated in
For gaining space, the reactor may be made in three portions each with a length of 12 mm as shown in
With this structure, functionalized silica beads may possibly be organized in order to provide the reactor with its affinity properties or to enhance them (C18 grafting for example).
ESI Interface
The section of the outlet channel 40 may be adapted by preferentially working on the transverse sides (in the plane of the substrate) of the latter, which provides the possibility of achieving <<mild restrictions>> preventing dead volumes. In
At the end of the device, the outlet channel 40 opens into a structure of the tip type 42, a structure with a variable external section with which the surface of the liquid/gas and liquid/solid interfaces may be limited, as exhibited by the out-flowing liquid with its environment, by means of its end with small inner and outer sections, while retaining robustness during its use by its end with a wide section.
Finally, the inside of the outlet channel 40 is provided with an electrode 43 with which an electrical potential may be imposed to the liquid which appears at the outlet of the device, which is necessary for nebulizing the sample (stability of the Taylor cone) and/or participating in its electroosmotic pumping.
The whole of these components provides a complete planar ESI interface, since it is robust, without dead volumes for connecting to fluidic networks and with which a Taylor cone with good stability may be formed.
Different embodiments of the microfluidic device provided with an electro-nebulization structure according to the invention will now be described. Only the preferred embodiment, the fifth one, will be described in detail.
For more clarity, these descriptions are made at the scale of a chip (a device), but the various technological systems are made on substrates which may include several devices (circular 100 mm substrate for example).
In these descriptions, the fluidic network is simplified and reduced to an inlet reservoir, an inlet channel, a microreactor, and an outlet channel with a constant section opening into the tip type structure. One skilled in the art will design the fluidic network as desired, for example the one described earlier.
This embodiment is illustrated by
The doping of the thin silicon layer 53 is achieved through a photoresist (or silicon oxide) mask over the integrality of the thickness of this layer.
The etching of the thin silicon layer 53 is partial (20 μm) in the portion intended to form the fluidic network in order to retain a portion of doped silicon track (5 μm) at the bottom of certain areas of the fluidic network (in particular close to the outlet for making the nebulization electrode). The achieved fluidic network comprises an inlet reservoir 61, an inlet channel 62, a microreactor 63 and an outlet channel 64. At the tip type structure, the outlet channel defined herein then exhibits two side walls and a horizontal wall (the ground). It is noted that an end 58 of the doped area 55 is located at the bottom of the inlet reservoir 61 and that an end 59 of the doped area 57 is located at the bottom of a portion of the outlet channel 64.
This step is a key step as it allows in-depth continuity of the fluidic network and of the outlet channel. Thus, <<zero dead volume>> connector technology is made possible. This will be the case in all the other embodiments.
After this etching, the tip type structure 65 is freed and forms an overhang above the support 51. It should be noted that the outlet channel 64 always includes the bottom 66.
A step for electrically insulating the fluidic network is then performed. This is obtained by thermal oxidization over 3 μm of the thickness of the silicon of the thin layer 53. The silicon support 51 should not be oxidized otherwise the tip type structure 65 would no longer overhang.
This thermal oxidization step is required in order to electrically insulate the liquid present in the fluidic network from the outside. This electrical insulation is necessary, for example when electro-osmotic pumping is used or when a separation by electrophoresis or an electrochemical reaction takes place in the fluidic network.
The next step consists of clearing the electrical contacts. In order to clear the electrodes (ends 58 and 59) and the contact connections (areas 54 and 56), it is necessary to locally etch the thermal SiO2 layer (3 μm) made earlier. This step may be performed by a laser etching technique as proposed by NovaLase from Pessac (Gironde, France).
In order to obtain the on-chip laboratory according to the invention, support 51 is cleaved as shown in
According to this embodiment, the device 70 (see
Once the seal is obtained, one proceeds with three cleavages. With a first cleavage of the plate 71 and a cleavage of the support 51 of the device 70, the electronebulization nozzle may be freed. With a second cleavage of the plate 71 the contact connections 54 and 56 may be freed.
This embodiment is illustrated by
The etching of the upper silicon layer 83 is:
Moreover, in the latter case, the etching may optionally be continued through the oxide layer 82 and then into the silicon support 81 in order to make a fluidic network with a great depth.
The achieved fluidic network comprises an inlet reservoir 91, an inlet channel 92, a microreactor 93 and an outlet channel 94. The etching of the thin layer 83 also defines the tip type structure 95.
The tip type structure 95 is then cleared by total chemical etching of the portion of the oxide layer 82 which has been exposed by etching of the layer 83 and also of the one which is found under the tip type structure 95 (see
A step for electrically insulating the fluidic network is then performed. This is achieved by thermal oxidization over 3 μm of thickness of the silicon of the thin layer 83.
The contact connections 84 and 86, the electrodes 88 (at the bottom of the inlet reservoir) and 89 (in the channel of the electro-nebulization nozzle) are then made by lifting off metal, as well as the conducting tracks 85 and 87 connecting each electrode to its corresponding contact connection (see
This is an alternative to the second embodiment wherein the use of a SOI substrate is replaced by the use of two silicon substrates.
The substrate 100 is then subjected to reactive ion etching (RIE) or chemical etching by means of KOH from the face 102, in order to obtain a recess 101 for providing the tip type structure and the cleaving of the substrate (see
Another silicon substrate 110 is then fixed by directly sealing it onto the face 102 of the substrate (see
The substrate 110 is then thinned until a thin layer 111 is obtained (see
The fluidic network is then made as shown in
As for the preceding embodiments, a step for electrically insulating the fluidic network is then performed. This is obtained by thermal oxidization.
The next step has the purpose of clearing the contact connections 104 and 106 (see
Step 9G illustrates the direct sealing of a cover plate 131 on the thin layer 111. The cover plate 131 includes an overhanging end portion 132 so that the plate 131 does not cover the tip type structure 125. It also includes a through-hole 133 intended for providing fluid communication with the inlet reservoir 121. The cover plate 131 may be a pyrex substrate.
Once the sealing is obtained, one proceeds with clearing. the tip type structure 125 and the contact connections 104 and 106. With a first cleavage of the plate 131 and a cleavage of the substrate 100 the electronebulization nozzle may be freed. With a second cleavage of the plate 131, the contact connections 104 and 106 may be freed.
This embodiment uses an SOI substrate and a pyrex substrate (<<Corning>> 7740) as a cover. The electrodes, the conducting tracks and the electrical contact connections are made by depositing metal (aluminium, platinum, gold . . . ) and by photolithography on the lower face of the pyrex cover, wherein they are <<inlaid>>.
This etching step is standard in microtechnology. It uses a silicon oxide mask with a thickness of 5,000 Å produced in an oven at 1,050° C. in a humid atmosphere. A 1.3 μm layer of photoresist <<Shipley S 1813 SP15>> is then spread out over a <<SVG>> track (adherence promoter: HMDS vapor). The 1× patterns are insolated, and then developed with <<Shipley MIF 310>> on an <<SVG>> track. The oxide mask may then be etched with reactive ion etching (RIE) under a CHF3/O2 mixture, by <<Nextral 330>> for example. The resin is then removed (by the so-called stripping method) with Posistrip or fuming HNO3. Silicon is then DRIE-etched under an SF6/O2 mixture at 110° C. with <<Alacatel ICP 601E>> for example. Finally, the oxide mask is stripped with 10% HF until dewetting occurs.
The tip type structure is then cleared by chemically etching the oxide layer 142. This chemical etching may be performed in a bath called BOE (“Buffer Oxide Etchant”): HF/NH4F). The device illustrated in
As for the previous embodiments, electrical insulation of the fluidic network is obtained by thermal oxidization. This oxidization takes place in an oven at 1,150° C. in a humid atmosphere.
This direct sealing step is performed at 400° C. It requires proper preparation of the surfaces, i.e:
Structuration of the pyrex cover (<<etching>> and <<inlaying>> the metal track) is performed according to the following technological steps:
Performing Etchings for a Cut-Out Recess and <<Box for Metal Track>>:
Producing the <<Inlaid>> Metal Strip:
Opening the Contacts:
Once the sealing is obtained, one proceeds with clearing the tip type structure 155 and the contact connection 166 by cleaving the support 141 (see
This is an alternative to the fifth embodiment in which the use of an SOI substrate is replaced with the use of two silicon substrates.
The fluidic network is then achieved as shown in
As for the previous embodiments, a step for electrically insulating the fluidic network is then performed. This is obtained by thermal oxidization.
Once the sealing is achieved (see
Circuit for Electro-Osmotic Pumping
In order to externally impose electro-osmotic pumping in the different reactors of the micro-fluid device, it is possible to use a card with tips produced by MESATRONIC S.A. of Voiron (Isere, France). Such a card is an electrical circuit which may withstand high voltages (10 kV) and is provided with a set of platinum tips which will simultaneously dip in different reservoirs of the device. Different electrical potentials may therefore be imposed at different points of the device in order to manage the different flows thereof.
Use of the Invention
In this configuration, the fluidic networks are patterned radially, along the radii of the circular substrate 210. N electro-nebulization nozzles are then distributed along the circumference for the substrate, and it is sufficient to manually or automatically rotate the latter in order to provide a continuous series of analyses with a mass spectrometer 212. For this, the support of the substrate may be mounted on a rotary axis. Preparation of the samples may itself be achieved beforehand in parallel on the N devices.
The possible applications of the invention are all those which use as a detection method, mass spectrometry with an “electrospray ionization” (ESI) technique as an interface.
As an example the analysis of samples may be mentioned in the biomedical sector and in the pharmaceutical industry:
Number | Date | Country | Kind |
---|---|---|---|
0450173 | Jan 2004 | FR | national |
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/FR05/50053 | 1/28/2005 | WO | 7/28/2006 |