The present invention relates to medical devices for carrying out internal examination and relates particularly to laryngoscopes to assist intubation of a tracheal tube that have disposable sections.
Insertion of a tracheal tube is an important procedure in providing an airway to an anaesthetist prior to a surgical operation. Tracheal tubes also often need to be inserted in an emergency situation into the airway of an unconscious patient by paramedics or doctors. Insertion of a tracheal tube requires significant skill, and laryngoscopes are generally used to assist the insertion of the tube by restraining the patient's tongue and allowing a clear view of the larynx and the entrance to the trachea. Considerable skill and care is required in carrying out this procedure in order to avoid damage to the patient's teeth and soft tissue of the throat.
Often problems occur when a practitioner is attempting to intubate a patient using a laryngoscope as it is often difficult for the practitioner to see what is going on.
Figures show that in approximately 12% of cases trauma occurs during intubation (which affects a large number of people when you consider there are over 12 million intubations carried out annually).
A UK wide study found 50% of apparently clean laryngoscope handles waiting for re-use to be contaminated with blood from previous procedures. This is due to design flaws in the traditional laryngoscope and ineffective cleaning practice within hospitals. Where disposable single use surgical instruments have already been introduced, 12% of hospitals are actually re-using them due to the high cost of replacement. Making the situation worse, these elements are often never cleaned due to the assumption of single use sterility.
Obviously in order to use a laryngoscope on a patient, it is important to know that the laryngoscope is cleaned sufficiently and there is no risk of cross contamination between patients. There is evidence to show that standard cleaning procedures are not always fully effective at removing contaminants such as bacteria from the laryngoscope (J R Hall. ‘Blood contamination of equipment . . . ’ Anaesthesia and Analgesia. 1994; 78:1136-9 M D Ester, L C Baines, D J Wilkinson & R M Langford. ‘Decontamination of Laryngoscopes: a survey of national practice.’ Anaesthesia, 1999, 54).
Typically, in order to clean a laryngoscope, the blade is soaked and autoclaved. The handle can undergo a similar procedure or can simply be wiped down as it does not make contact with the patient as the blade does. The cleaning takes a significant amount of time, which means that it is necessary to have a number of handles and blades in rotation to ensure that there are always clean laryngoscopes available if required. This results in a time consuming and costly procedure needing to be put in place.
In order to try and overcome the problems associated with laryngoscope disposable blades have been suggested for use. Unfortunately, it is common in practice that the blades are still used multiple times before being discarded. Flexible protective sheaths can also be used which slip over a standard laryngoscope blade to act as a guard. While useful, it is optional to a user whether the sheath is used or not. For the user, existing blades perform better without the sheath, which distorts light output and, as a result, existing sheaths are rarely used.
The present invention attempts to improve upon the prior art.
Throughout this Application the term blade should be read in a broad sense to cover not only laryngoscope blades but also to cover speculums or elements that are inserted into body cavities.
According to a first aspect of the present invention, there is provided a medical device comprising a body portion and a blade portion, wherein the blade portion is separable from the body portion characterised by the medical device being provided with spoiling means for indicating if a particular blade portion has already been attached to the body portion previously.
Optionally, the spoiling means will prevent reattachment of a blade that has already been attached to a body portion previously.
Preferably, the spoiling means comprises transmitting and receiving means in the body, which send and receive signals respectively with receiving and transmitting means in the blade.
Preferably the receiving means in the body can determine from the transmitting means in the blade, if the blade has previously been brought into close proximity of a body.
Preferably the transmitting means is a radio frequency transmitter.
Preferably the receiving means is a radio frequency receiver.
Optionally, the receiving means in the body will count the number of blades it contacts in its life.
Optionally, the spoiling means comprises a device for tripping electrical contacts to prevent their continued use.
Optionally the spoiling means comprises elements that cause a physical change if a particular blade portion has already been attached to the body portion previously.
Alternatively, the spoiling means comprises locking elements that break off when the blade and body are separated.
Preferably the locking elements comprise a male protrusion and a female ingression, one of which is provided on the blade and one of which is provided on the body.
Preferably the male protrusion is provided with a weakened section.
Preferably the female ingression is provided with a weakened section.
In order to provide a better understanding of the present invention, embodiments will now be described by way of example only, and with reference to the following Figures, in which:
In the preferred embodiment of the present invention, the medical device is a laryngoscope that can be used for intubation of a tracheal tube.
According to the present invention, there is a provided a laryngoscope 1 which has a disposable blade 3. The blade 3 is the section that comes into contact with the patient during examination. Therefore, after use the blade 3 can be disposed of and the remaining parts of the laryngoscope 1 re-used.
One of the benefits of the disposability of the blade 3 is that there will be no cross-contamination to patients, and no lengthy cleaning procedures are required. However, to ensure that a blade 3 is not reused, a spoiling mechanism is incorporated into the laryngoscope laryngoscope 1.
In one embodiment, a mother chip 4 is provided in the body 2 of the laryngoscope 1. A corresponding radio frequency tag 5 that can be recognised by the mother chip 4 is provided in the disposable blade 3. The radio frequency tag 5 is provided with a serial number and each radio frequency tag 5 has a unique serial number. When the radio frequency tag 5 and the mother chip 4 are brought into close contact i.e. by the blade 3 being attached to the body section 2, the mother chip 4 can read the serial number. If the mother chip recognises the serial number from a previous occasion this will be indicated to the user.
Alternatively, the radio frequency tag does not have a serial number, but the mother chip electronically marks the radio frequency tag, so that it can be recognised as having previously being in contact with another body. This is the preferred embodiment.
In the preferred embodiment, the body 2 is provided with a monitor 6. If the mother chip 4 recognises a serial number, or any other electronic signal on a radio frequency tag 5 of a blade that is being attached, a computer program will be in place to display a message on a monitor which is attached to the body 2 informing the user. The mother chip 4 will typically be able to both read the blade chip and write on it. This would allow a blade chip to be electronically written on in a manner that means it would be recognised by any body section that it is attached to. This is the preferred embodiment as it ensures that in emergencies for example when a paramedic is called to a scene with a significant number of casualties, a blade 3 can be re-used if absolutely necessary. It would also allow the storage of the date, time, patient details and other information, if required. An alternative to the radio frequency messaging described above would be to use a mother chip 4a that is an optical reader. There would be a mechanism for physically marking a blade 3 that has previously been used, and this physical mark would be recognised by the mother chip 4a. The physical Marks may be surface scoring, discolouring, exposure to light or faint fracture points/lines designed to appear after pressure has been applied during use.
The benefit of the abovementioned options is that the mother chip 4 or 4a could be programmed to allow a certain number of re-uses which may be within a defined period, or allow a manual override in emergency situations.
Alternatively, the spoiling mechanism can take the form of a breaking of electrical connections when the blade 3 and body 2 are parted, such that if the same blade 3 and body 2 are reconnected, no power is provided to anything inserted into the core 6 of the blade 3. A further alternative is that the blade 3 may comprise protrusions which are able to fix into ingressions in the body 2 of the laryngoscope 1, such that the protrusions break off when the blade 3 is removed from the body 2, such that the blade 3 cannot then be reused. These alternatives may be more useful in non-emergency areas such as operating theatres for routine surgery where the need to re-use a blade 3 in extreme circumstances is less likely to occur.
It can be seen that the current invention has a number of benefits over the prior art and a number of possible uses. Although the examples above relate to a laryngoscope, it can be seen that the concept can be extended to other medical and veterinary devices and still stay within the scope of the present invention. The fact that the blade is fully disposable is of great importance, as it means that practitioners are required to change blades and the product is both simple to use and cheap to manufacture.
It will be appreciated by persons skilled in the art that the above embodiment has been described by way of example only, and not in any limiting sense, and that various alterations and modifications are possible without departure from the scope of the invention as defined by the appended Claims.
Number | Date | Country | Kind |
---|---|---|---|
0309754.0 | Apr 2003 | GB | national |
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/GB2004/001844 | 4/29/2004 | WO | 00 | 12/28/2006 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2004/096031 | 11/11/2004 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
3643654 | Felbarg | Feb 1972 | A |
4086919 | Bullard | May 1978 | A |
4113137 | Wind | Sep 1978 | A |
4126127 | May | Nov 1978 | A |
4306547 | Lowell | Dec 1981 | A |
4384570 | Roberts | May 1983 | A |
4406280 | Upsher | Sep 1983 | A |
4556052 | Muller | Dec 1985 | A |
4573451 | Bauman | Mar 1986 | A |
4742819 | George | May 1988 | A |
4832003 | Yabe | May 1989 | A |
4832020 | Augustine | May 1989 | A |
4930495 | Upsher | Jun 1990 | A |
4934773 | Becker | Jun 1990 | A |
4982729 | Wu | Jan 1991 | A |
5003963 | Bullard et al. | Apr 1991 | A |
5203320 | Augustine | Apr 1993 | A |
5233426 | Suzuki et al. | Aug 1993 | A |
5239983 | Katsurada | Aug 1993 | A |
5261392 | Wu | Nov 1993 | A |
5339805 | Parker | Aug 1994 | A |
5349943 | Ruiz | Sep 1994 | A |
5355870 | Lacy | Oct 1994 | A |
5363838 | George | Nov 1994 | A |
5373317 | Salvati et al. | Dec 1994 | A |
5381787 | Bullard | Jan 1995 | A |
D358471 | Cope et al. | May 1995 | S |
5413092 | Williams, III et al. | May 1995 | A |
5443058 | Ough | Aug 1995 | A |
5513627 | Flam | May 1996 | A |
5551946 | Bullard | Sep 1996 | A |
5594497 | Ahern et al. | Jan 1997 | A |
5645519 | Lee et al. | Jul 1997 | A |
5676635 | Levin | Oct 1997 | A |
5701904 | Simmons et al. | Dec 1997 | A |
5702351 | Bar-Or et al. | Dec 1997 | A |
5734418 | Danna | Mar 1998 | A |
5734718 | Prafullchandra | Mar 1998 | A |
5754313 | Pelchy et al. | May 1998 | A |
5762605 | Cane et al. | Jun 1998 | A |
5800342 | Lee et al. | Sep 1998 | A |
5800344 | Wood, Sr. et al. | Sep 1998 | A |
5827178 | Berall | Oct 1998 | A |
5827428 | Chang | Oct 1998 | A |
5846186 | Upsher | Dec 1998 | A |
5873818 | Rothfels | Feb 1999 | A |
5879289 | Yarush et al. | Mar 1999 | A |
5879304 | Shuchman et al. | Mar 1999 | A |
5895350 | Hori | Apr 1999 | A |
6056716 | D'Antonio et al. | May 2000 | A |
6080101 | Tatsuno et al. | Jun 2000 | A |
6095972 | Sakamoto | Aug 2000 | A |
6354993 | Kaplan et al. | Mar 2002 | B1 |
6543447 | Pacey | Apr 2003 | B2 |
6676598 | Rudischhauser et al. | Jan 2004 | B2 |
6826422 | Modell et al. | Nov 2004 | B1 |
6847490 | Nordstrom et al. | Jan 2005 | B1 |
6964637 | Dalle et al. | Nov 2005 | B2 |
7001330 | Kobayashi | Feb 2006 | B2 |
7001366 | Ballard | Feb 2006 | B2 |
7128710 | Cranton et al. | Oct 2006 | B1 |
7156091 | Koyama et al. | Jan 2007 | B2 |
7182728 | Cubb et al. | Feb 2007 | B2 |
7214184 | McMorrow | May 2007 | B2 |
7448377 | Koyama et al. | Nov 2008 | B2 |
20010051766 | Gazdzinski | Dec 2001 | A1 |
20020022769 | Smith et al. | Feb 2002 | A1 |
20020038076 | Sheehan et al. | Mar 2002 | A1 |
20020087050 | Rudischhauser et al. | Jul 2002 | A1 |
20020107436 | Barton et al. | Aug 2002 | A1 |
20020198554 | Whitman et al. | Dec 2002 | A1 |
20030195390 | Graumann | Oct 2003 | A1 |
20040015132 | Brown | Jan 2004 | A1 |
20040127770 | McGrath | Jul 2004 | A1 |
20050090712 | Cubb | Apr 2005 | A1 |
20060129031 | Roberts et al. | Jun 2006 | A1 |
20060276694 | Acha Gandarias | Dec 2006 | A1 |
Number | Date | Country |
---|---|---|
0653180 | Oct 1998 | EP |
0901772 | Mar 1999 | EP |
1188420 | Mar 2002 | EP |
1 285 623 | Feb 2003 | EP |
1285623 | Feb 2003 | EP |
2830428 | Apr 2003 | FR |
2 086 732 | May 1982 | GB |
61141342 | Jun 1986 | JP |
2000175867 | Jun 2000 | JP |
8301373 | Apr 1983 | WO |
WO-8911305 | Nov 1989 | WO |
WO-9104703 | Apr 1991 | WO |
WO-9513023 | May 1995 | WO |
WO-9819589 | May 1998 | WO |
WO-9841137 | Sep 1998 | WO |
WO-9927840 | Jun 1999 | WO |
WO-9944490 | Sep 1999 | WO |
WO-0074556 | Dec 2000 | WO |
WO-0110293 | Feb 2001 | WO |
WO-0178582 | Oct 2001 | WO |
WO-0178582 | Oct 2001 | WO |
WO-0211608 | Feb 2002 | WO |
WO-02056756 | Jul 2002 | WO |
WO-02095675 | Nov 2002 | WO |
WO-03015619 | Feb 2003 | WO |
WO-03077738 | Sep 2003 | WO |
WO-03077738 | Sep 2003 | WO |
Entry |
---|
Woodson, “Retropalatal Airway Characteristics in Uvulopalatopharyngoplasty Compared With Transpalatal Advacement Pharyngoplasty”; Laryngoscope 107, Jun. 1997, pp. 735-740. |
Satava, “A technologic framework for the future”; Surgical Endoscopy (1993) 7: 111-113. |
Meinke, et al, “What is the learning curve for laparoscopic appendectomy?”; Surgical Endoscopy (1994) 8: 371-375. |
Cunningham, “Laparpscopic surgery—anesthetic implications”; Surgical Endoscopy (1994) 8: 1272-1284. |
Siker ES, “A Mirror Laryngoscope”, Anaesthesiology 17:38-42, 1956. |
Australian Search Report. |
International Search Report. |
Murphy, et al. “Rigid and Semirigid Fiberoptic Intubation”, Manual of Emergency Airway Management, The Airways, Lippincott Williams & Wilkins, 2004. |
Crosby, et al., “The Unanticipated Difficult Airway With Recomendations for Management”, Can J Anaesth 1998 pp. 757-776. |
Biro, et al., “Comparison of Two Video-Assisted Techniques for the Difficult Intubation”, 2001, pp. 761-765. |
Bellhouse, et al., “An Angulated Laryngoscope for Routine and Difficult Trachael Intubation”, 1997, pp. 126-129. |
Pearce, et al., “Evaluation of the Upsherscope”, 1996, pp. 561-564. |
Smith, et al., The Complexity of Trachael Intubation Using Rigid Fiberoptic Laryngoscopy (WuScope), Anesth Analg, 1999, pp. 236-239. |
Cooper, “Use of a New Videolaryngoscope (GlideScope®) in the Management of a Difficult Airway”, Can J Anesth, 2003, pp. 611-613. |
Dullenkopf, et al., “Video-enhanced Visualization of he Larynx and intubation with teh Bullard Laryngoscope—equipment report” Can J Anesth 2003, pp. 507-510. |
Thompson, “A New Video Laryngoscope”, Anaesthesia, 2004, pp. 410. |
Esler, et al., “Decontaminationoflaryngoscopes: asurveyofnational Practice”, Anaesthesia, 1999, pp. 582-598. |
Morrell, et al., “A Survey of Laryngoscope Contamination at a University and a Community Hospital”, Anesthesiology, 1994, pp. 960. |
Weiss, M. “Video-intuboscopy: a new aid to routine and difficult tracheal intubation”, British Journal of Anaesthesia 1998; 80: 525-527. |
U.S. Appl. No. 60/067,205. |
U.S. Appl. No. 60/074,355. |
U.S. Appl. No. 09/060,891. |
U.S. Appl. No. 60/223,330. |
U.S. Appl. No. 09/732,129. |
Number | Date | Country | |
---|---|---|---|
20070299313 A1 | Dec 2007 | US |