The present invention relates to detecting leaks in a containment system of the type that has a primary container, such as a tank or a pipe, contained within a secondary container.
Systems for storing or transporting fluids that may damage the environment, such as caustic or volatile fluids, are often double-wall systems. These systems include a primary container that contains the fluid and a secondary container that contains the primary container. As a result, if the primary container leaks, the fluid is contained within the secondary container.
After such systems are installed, they are typically tested, which may be required by governmental regulations, to make sure that the primary container is functioning properly, i.e., not leaking. Sometimes these systems are buried underground. For example, petroleum dispensers are typically connected to underground storage tanks that contain the petroleum. The underground tanks as well as the pipe connecting the tanks to the dispensers are almost always primary containers that are surrounded and sealed within secondary containers. Even though the primary container is intended for liquids, the integrity of the system is typically determined by testing for vapor leaks.
Testing these, and other underground systems, after installation poses challenges because most of the containment system is buried and therefore not accessible except for portions that communicate with the surface, such as manways, risers and the like. Another challenge, regardless of the system location, is that the primary container may be made of a polymeric material that is somewhat permeable to any vapors it contains. In other words, there is a continuous, albeit small, transmission of vapor within the primary container into the interstitial space between the primary and secondary containers. This may need to be taken into account to accurately determine whether there is a vapor leak over and above that expected as a result of permeation.
Furthermore, when a primary system is first pressurized for testing, typically with a tracer gas, very little fluid permeates into the interstice between the primary and secondary containers. But permeation increases over time. When a primary vessel under test has been pressurized for a period before the test begins, concentrations of interstitial tracer gas are higher at the start of the test than if the vessel was pressurized immediately prior to the test start. These higher concentrations that result from permeation should be considered when determining the integrity of the primary vessel. Aside from the permeation issues, when a primary vessel is a pipe, especially a buried one, it would be desirable to approximate the location of any leak detected so that only a relatively small portion of the pipe system needed to be exposed by digging and then repaired.
Turning now to
End 14 of pipe 12 is connected to a helium tank 26 via a hose 28 and one or more valves, regulators, and gauges, as shown, for delivering helium from tank 26 to the interior of pipe 12. A valve 29 seals the interior of pipe 12 from the surrounding atmosphere when closed and provides open communication when opened. A T-joint 36 extends from vacuum chamber 18 and provides communication annular space 24 and a helium leak detector 38. The present implementation uses a Model 979 Series Helium Mass Spectrometer Leak Detector manufactured by Varian, although other types of helium sensors may be equally well used. Helium sensor 38 generates electrical signals that are proportional to helium concentration. These signals are provided to a computer 40, a laptop computer in the present embodiment, via a cable 42. In addition, helium sensor 38 includes a vacuum pump that, when operating, can create a powerful vacuum within annular space 24.
Before considering containment systems in the field, and how they are tested, a description of how setup 10 is used to determine permeation of polymeric material that is used in such containment systems is provided. After the polymeric material of interest, in the form of pipe 12, is configured as shown in setup 10, helium leak detector 38 is turned on to warm up. When the helium detector is ready, valve 29 is opened and helium tank 26 is placed in communication with hose 28 thus flushing out the interior of pipe 12. Valve 29 is then shut thereby pressurizing the interior of pipe 12 with helium at a positive pressure. Next the helium leak detector is activated to begin sensing the rate of change of helium in volume per unit time.
A program in computer 40 receives this data from helium sensor 38, which is in the form of periodic measurements indicating current concentrations of helium. These samples are shown on the permeation curve of
Typically a variety of pipes, like pipe 12, from different manufacturers and having different thicknesses are tested as described above. The test results produce different curves, like the curve of
Turning now to
In setup 46, a hose 56 connects to fittings 58, 60 on either end of hose 48. Fitting 58 is conventional except for a quick-release connection 62, on fitting 58, which permits hose 56 to be connected to and disconnected from the fittings.
Before considering testing of the system of
Considering first testing of the system of
This is true even though any helium in space 72 is flushed out as just described. It will be recalled that permeation increases over time, If the tank has been charged with helium for long enough, permeation of helium into space 72 will migrate into space 72 at the beginning of the test and will contribute to any helium measured, which may include helium contributed by one or more leaks or other defects in primary vessel 68 through which helium moves into space 72.
In the present embodiment, if there is a leak rate greater than 0.005 gallon per hour, the system fails and cannot be returned to or put into service. Although 0.005 gallon per hour is used as a test standard in this description, it should be appreciated that different jurisdictions use different standards. Other standards, which could also be tested for, might be imposed by a manufacturer or by the purchaser of the containment system. Knowing the volume of space 72, risers 74, 76 and hose 78, which together make up the volume of gas circulated by fan 34 and tested by detector 38, makes it possible to calculate the concentration of helium that equals the 0.005 gallon per hour leak rate. This value is calculated and stored on computer 40.
If the tank has been charged long enough for the permeation through primary vessel 68 to be substantially steady state, i.e., after about 200-400 minutes in the curve of
But if pair of measurements indicates a leak rate higher than the benchmark, that does not necessarily mean the system failed. It is possible that, due to permeation, concentrations early in the test period might exceed the 0.005 gallon per hour limit. As a result, the present method measures changes in concentrations over time, and compares these changes to changes in concentration over time that would be expected as a result of permeation.
By way of example, assume that the system test in
The measured helium concentration, which includes permeation plus helium flow resulting from any leaks or other defects, for that time period is determined in a similar manner. Specifically, concentrations are measured between 200 and 400 minutes after charging with helium to create a curve of concentrations versus time. This curve is integrated between 200 and 400 minutes and the resulting value is scaled to the
These calculations produce two numbers. First, an average concentration that would be expected for the system over the test time based on permeation alone. And second, an average concentration that results from measured helium that would include permeation and any other helium flow that might be occurring.
Comparing the actual rate of change with the expected rate of change resulting from permeation determines whether there are leaks and/or whether a particular standard is met.
Turning now to
With a leak like that shown in
Having described and illustrated the principles of the invention in a preferred embodiment thereof, it should be apparent that the invention can be modified in arrangement and detail without departing from such principles. I claim all modifications and variation coming within the spirit and scope of the following claims.
Number | Name | Date | Kind |
---|---|---|---|
4404843 | Johnson et al. | Sep 1983 | A |
4450711 | Claude | May 1984 | A |
4796676 | Hendershot et al. | Jan 1989 | A |
4939833 | Thomas | Jul 1990 | A |
5265465 | Thomas | Nov 1993 | A |
5265652 | Brunella | Nov 1993 | A |
5343191 | McAtamney | Aug 1994 | A |
5375457 | Trapp | Dec 1994 | A |
5589631 | Spring et al. | Dec 1996 | A |
5668534 | Haboian et al. | Sep 1997 | A |
6029505 | Webb | Feb 2000 | A |
6067844 | Westbrook et al. | May 2000 | A |
6116817 | Osborne | Sep 2000 | A |
6935161 | Hutchinson | Aug 2005 | B2 |
7197950 | Maxwell et al. | Apr 2007 | B2 |
7251983 | Hutchinson | Aug 2007 | B2 |
20030037596 | Sorensen | Feb 2003 | A1 |
20040045343 | Hutchinson | Mar 2004 | A1 |
20040234338 | Monroe et al. | Nov 2004 | A1 |
20050166666 | Tsukagoshi | Aug 2005 | A1 |
20060107731 | Thomas | May 2006 | A1 |
20070101803 | Tujii et al. | May 2007 | A1 |
20070119238 | Issel | May 2007 | A1 |
20080011056 | Spaolonzi et al. | Jan 2008 | A1 |
20080011057 | Spaolonzi et al. | Jan 2008 | A1 |
Number | Date | Country |
---|---|---|
753729 | Jan 1997 | EP |
Number | Date | Country | |
---|---|---|---|
20080072659 A1 | Mar 2008 | US |