This disclosure relates generally to limited use power tools and more particularly to an enclosure for such tools during use in medical procedures; the enclosure being removed and discarded during reprocessing of the tools for subsequent re-use.
Important factors for any surgical instrument include sterility, cost of acquisition, maintenance, and reliability during use in the surgical suite. Each of these factors can have a significant impact on the cost of medical care for both the patient and the provider.
In recent years, there has been significant focus on the ever increasing cost of medical care. These cost increases have led to skyrocketing insurance premiums, reduced coverage, reduced reimbursements, increased fees for services, severe reductions in services for some patient groups by some providers, and unfortunately an apparent increase in infections and medical mishaps.
In an effort to reduce costs and improve profitability, both service providers and medical device suppliers are continuously looking for ways to streamline procedures, reduce time, cost, and risk from their products and services without reducing the quality of the products or services they provide to their customers. One area to benefit from these savings and improvements has been in the orthopedic surgical field through the use of high precision, battery powered surgical instrumentation. In the late 1960's and early 1970's battery operated drills were bulky, ill-balanced and required multiple batteries to perform some surgeries due to the limited energy storage capacity and poor efficiency of the electric motors.
Since then, manufacturers have attempted to make batteries more efficient with higher energy storage capacity, reduced size, and improved rechargeable lifespans. Likewise, motor housings such as saw and drill bodies have become more ergonomic, balanced, lightweight and energy efficient. As with many standard hand tools having multiple moving components, instrument manufacturers have reduced weight by utilizing lighter materials such as plastic housings, and gears, and put weight reducing apertures in what were previously solid housings. In some cases, standard mountings for attachments have been replaced with modular fittings, allowing for greater interchangeability and component selections. Additionally, manufacturers have attempted to improve electrical components by upgrading them with more modern components wherever possible.
All of these improvements in equipment construction have improved efficiencies, costs and quality in some areas while at the same time increasing costs for acquisition, maintenance and increasing risks in other ways that were not previously seen or predicted. Often times cost and quality can be inversely proportional to one another. One example of the increased cost and patient risk is seen in the cleaning and maintenance of instruments.
Recent published reports suggest that many of the surgical instruments used in operations were not being cleaned and/or sterilized appropriately in the very hospital facilities that were established and tasked for that purpose. In numerous reports, following cleaning and sterilization, it was noted that upon closer secondary inspection, the inside of small diameter cannulas and intricate mini-components of arthroscopic shavers that are used for many of today's minimally invasive procedures, contained human tissue and bone fragments from previous surgeries. In other cases, modular components of drills and saws such as chucks, drill bits and blades were found to have similar debris or pieces of cleaning brushes and/or bristles embedded in or on them. These investigations have demonstrated that in most cases the instruments were not cleaned according to manufacturer's specifications which has likely lead to many documented cases of serious, multiple, serial infections for subsequent patients. A pilot program conducted by the Centers for Medicare and Medicaid Services (Schaefer et al., 2010; JAMA 2010; 303(22):2273-2279) inspected 1500 outpatient surgery centers and found that 28% had been cited for infectious control deficiencies associated with equipment cleaning and sterilization. The costs to the patients and the hospitals in both expense and liability to deal with these infections can be and has been staggering.
In other cases, critical battery-operated, motorized tools such as drills or bone saws have ceased to function due to dead batteries that no longer maintain their capacity to hold a charge, or due to internal part failure, often attributable to overuse or lack of proper maintenance. The resultant downtime in the operating suite is extremely costly, as the procedure step must be put on hold while replacement or substitute tools are obtained. Wait times may often exceed 20-30 minutes, resulting in additional anesthesia exposure for the patient, additional operating room time (charged to the patient) and potential delays to other procedures where the replacement or substitute equipment had been scheduled for use in a later procedure. Recent estimates (2005) establish the average cost of operating room time to range between $62/min. (range $21.80-$133.12) depending on the procedure. These figures did not include extra resources provided by the hospital for special, non-routine situations which often occur during standard procedures, and did not include the surgeon and anesthesia provider fees, (anesthesia fees are estimated to be $4/min; range $2.20-$6.10).
Hospitals and instrument manufacturers are continuously attempting to find improved ways to reduce risk associated with infection in general, and more recently, specifically from improperly cleaned instruments. One approach has been to use more disposable, single-use instruments such as drills, saw blades and plastic cannulas. Additionally, many laparoscopic devices such as, surgical staplers and trocars, are designed as single use items that are intended to be immediately disposed of after use. Unfortunately, at today's acquisition costs, the total cost of ownership and benefits are not always clear for high-use battery-operated, motorized instruments such as saws, drills and reamers used in orthopedic procedures and the idea of disposable powered instruments has not been readily embraced.
A recent trend in the medical community is reprocessing of single use medical instruments, by parties other than the original equipment manufacturer, instead of discarding them after use. During reprocessing, the medical instruments are disassembled, cleaned and sterilized. They are then reassembled for future use. However, because the medical instruments reprocessed for further use are specifically provided for use during a single procedure, the performance of the medical instruments tends to decline after reprocessing, because the components making up the medical instrument are not adapted for multiple uses and will degrade in performance when used beyond their intended life span. For example, reprocessing of the cutting devices on trocars is intended to extend these devices beyond their intended mission life, but often results in duller cutting edges on the blades because neither the materials used nor the reprocessing method can restore the device to the original manufacturing specifications. A greater force, therefore, is needed to make an initial incision, causing more trauma to the patient. In addition, the use of greater force increases the potential for error during the surgical procedure.
Most hospitals and surgery centers buy high-use, reusable motorized, pneumatic, wired or battery operated, orthopedic surgical equipment and are expected to clean, sterilize, and maintain them internally within the hospital. Unfortunately, the technicians hired to perform this work are typically not qualified or trained to perform this work adequately for the many varieties of powered instruments used. Further, manufacturers rarely provide the hospital/client with the training or diagnostic equipment necessary to evaluate or test the equipment. Often times the hospital employees responsible for cleaning and maintenance are not technicians at all, being paid slightly more than minimum wage, working at a fast pace to merely wash, count, and reload instruments into their appropriate system trays and flash sterilize them as quickly as possible, in an effort to keep the equipment in rotation in the hospital operating rooms, where higher throughput dictates profitability for the hospital or surgery center.
As a result of high throughput requirements, general maintenance is rarely done and preventative monitoring and maintenance is almost never done on this type of equipment. Hospital budgets for internal maintenance of equipment are generally geared toward high-end, multi-million dollar capital equipment such as x-ray and radiological equipment. It is generally assumed that it is faster, simpler, and more economical for the hospital to wait for hand-held instruments, such as drills, saws and reamers to fail, then, send them back to the manufacturer for repair or replacement.
Thus it has become apparent that there is a need for an improved system of cost-effective, battery-operated, motorized tools in conjunction with better cleaning and maintenance protocols which can provide the hospital, surgeon, and most importantly, the patient, with a higher degree of efficiency and cleanliness while reducing risk and keeping the costs of cleaning, maintenance, and repair as low as possible.
Accordingly, a reusable medical procedure power tool includes a housing and a removable, single use contamination blocking material substantially covering the power tool, wherein the power tool includes a handle portion, a power source and a tool attachment portion. The housing has a first shape and the blocking material has a second shape similar to the first shape. The handle portion includes a tactile feel portion which is movable sufficient to move an associated trigger coupled to the housing.
The embodiment of
The tool 10 includes a housing 14 comprising a handle portion 16 and in this example, a power source portion such as a receiver 18 for a portable battery pack and a tool attachment portion 20 having a chuck 21 provided for releasably receiving and holding an attachment tool such as a drill bit or a saw blade. The handle 16 includes a control portion including but not limited to an actuating trigger 22, a trigger lock 24 and a forward-reverse switch, all of which may not be visible in
The cover 12 preferably includes a two-piece hard or soft outer shell including portions 12a and 12b. The tool 10 is illustrated at 10a prior to application of the cover 12, and is illustrated at 10b after application of the cover 12. A first opening 12c is provided in cover 12 adjacent chuck 21 when the cover is applied to the tool 10. A second opening 12d, which may be closed by a sealable door 19, is provided in power source portion 18. Regardless of the material used for the cover 12, a flexible portion 12e of the cover 12 is provided on the handle 16 to provide a user with a tactile feel and operable movement of for example, the trigger 22 and the trigger lock 24.
The replaceable cover 12 is applied to tool 10 by a tool re-processor. Once the tool 10 is used in a procedure, the cover has become contaminated along with portions of the tool 10 which are adjacent the openings 12c, 12d. The tool 10, including cover 12, is returned to the tool re-processor where the cover 12 is removed and discarded. The tool 10 is then cleaned and a new cover 12 is mounted on the tool 10, rendering the tool 10 ready for re-use.
More specific information regarding the tool 10 and cover 12 of the
a. Existing product or new product uses a rigid body mechanical housing 14 in conjunction with either a hard or soft/flexible shell outer shield 12 that covers and protects the majority of the tool 10 from contamination by blood/bone/tissue during a procedure. Combinations of materials such as a hard shell with flexible inserted areas for controls actuation are also contained in this area. Additional reinforcements or seals can be used in high stress areas.
Several further embodiments are described below. More specific information regarding the tool 10 and a stretch membrane cover 12 including upper member 12a and lower member 12b, of the
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a highly stretchable membrane 12 (balloon like) to cover and protect the tool 10 from contamination by blood/bone/tissue during a procedure. This cover 12 is a removable, single use cover of contamination blocking material. Single and multiple layer configurations can be considered for this version. Single or multiple membranes may be used to protect various areas of the tool 10 (main body vs. battery pack allowing access to battery pack at the start of a procedure). Variable wall thickness or reinforcements can be used in high stress areas. Members 12a and 12b are stretched over housing 14 and combined to form cover 12. Tool 10 is shown at 10a prior to application of cover 12, and is shown at 10b after the application of cover 12.
b. Methods appropriate for applying the membrane around the outer shell include but are not limited to:
More specific information regarding the tool 10 and a shrink wrap cover 12,
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a secondary shrink-wrap element 12 to cover and protect the device from contamination by blood/bone/tissue during a procedure. Cover 12 is a removable, single-use cover of contamination blocking material. Single and multiple layer configurations can be considered for this version (see considerations for transport as non-biohazard state). Single or multiple wraps may be used to protect various areas of the tool 10 (main body vs. battery pack allowing access to battery pack at the start of a procedure). Additional reinforcements or seals can be used in high stress areas. Shrink methods can include both heat application or a chilling operation depending on the type of shrink wrap utilized. Tool 10 is shown at 10a prior to application of cover 12, and is shown at 10b after the application of cover 12 and shrink activation,
b. Methods appropriate for applying the membrane around the outer shell include but are not limited to:
In
a. This embodiment uses a rigid sub-frame 110 carrying all mechanical components. The hard shell cover 12 has minimal mechanical content and is used as a disposable single-use housing of a contamination blocking material to protect the mechanical components from contamination by blood/bone/tissue during a procedure. Cover 12 comprises cover portions 12a, 12b. The sub-frame and mechanical components are intended for multiple re-use. This configuration may also be used in conjunction with a soft/flexible outer shell allowing for return of the device in a non-biohazard state. Combinations of materials such as hard shell with flexible inserted areas for controls actuation are also contained in this area. Additional reinforcements or seals can be used in areas subject to contaminant intrusion. Thus, the hard shell, single-use disposable cover 12 functions as a combination previously provided by a traditional housing 14 and cover 12.
b. Methods appropriate for fastening the outer shell to/around the inner structure include but are no limited to:
In
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a secondary spray-on protective layer 12 to cover and protect the tool 10 from contamination by blood/bone/tissue during a procedure. Single and multiple layer configurations can be considered for this version by using a release layer between subsequent spray applications. This configuration may be used in conjunction with previously described protection systems to allow access to power source portion 18 at the start of a procedure. Additional reinforcements or seals can be used in areas subject to contaminant intrusion. Layer 12 is a removable, single-use cover of contamination blocking material.
b. Methods appropriate for applying the membrane around the outer shell include but are not limited to:
In
a. This embodiment uses a rigid body mechanical housing in conjunction with a secondary dipping operation to apply a protective layer 12 intended to cover and protect the tool 10 from contamination by blood/bone/tissue during a procedure. Single and multiple layer configurations can be considered for this version by using a release layer between subsequent dip applications. This configuration may be used in conjunction with previously described protection systems to allow access to power source portion 18 at the start of a procedure. Additional reinforcements or seals can be used in areas subject to contaminant intrusion. Layer 12 is a removable, single-use cover of contamination blocking material.
b. Methods appropriate for applying the membrane around the outer shell include but are not limited to:
In
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a formed header bag outer shielding cover 12 that protects the majority of the tool 10 from contamination by blood/bone/tissue during a procedure. Additional reinforcements or seals can be used in high stress areas. Header bag cover 12 comprises a removable, single-use cover of contamination blocking material.
b. Methods appropriate for fastening the header bag to/around the inner structure include but are not limited to:
In
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a Precut Wrap outer shielding cover 12 that once applied protects the majority of the tool 10 from contamination by blood/bone/tissue during a procedure. Additional reinforcements or seals can be used in high stress areas or areas vulnerable to contaminant intrusion.
b. Methods for cutting the wrap to conform to the device include but are not limited to:
c. Methods appropriate for securing the wrap to/around the device include but are not limited to:
In
a. This embodiment uses a rigid body mechanical housing 14 in conjunction with a two layer soft/flexible shell outer cover 112 and 212 that protects the majority of the device from contamination by blood/bone/tissue during a procedure. Following the procedure and before return shipment of the device the outermost contaminated cover 212 is removed presenting the inner cover 112 that is a non-biohazard product and can economically be returned for re-processing.
In
The limited use tool 10,
The present disclosure has recognized and addressed many of the foregoing limitations and drawbacks of others concerning the need to provide hospitals and surgery centers with an improved, more reliable system of cost-effective, battery-operated, motorized tools in conjunction with better cleaning and maintenance protocols. In practice, the disclosed tooling system utilizes a concept called limited-use tools (LUT) and specifically, a new cover or enclosure system to make reprocessing of the LUT more efficient. This cover or enclosure would be used only once in the operating room, then would be removed and discarded at the reprocessing facility. A new, single-use enclosure would be installed at the reprocessing facility prior to final testing, packaging and re-sterilization of the LUT. The term “limited-use” as applied to orthopedic surgical tools can mean having a limited useful life, or a restricted lifespan for intended use. Preferably in this context, limited-use is intended to mean the number of surgeries where the useful life of the tool ranges from more than one use to less than 50 surgeries, and more preferably where the useful life of the tool ranges from more than one use to less than 30 surgeries, and most preferably where the useful life of the tool ranges from more than one use to less than 20 surgeries.
In a broad respect this disclosure teaches a method of improving (i.e. reducing) potential risk factors associated with infection control, and reduction of potential disease and infection transmission due to lapses in cleaning and infection control associated with routine maintenance of reusable powered surgical instruments. In another broad respect, the disclosure teaches a method of processing battery-operated tools used in surgery, to improve the cleanliness of instruments used in multiple surgical procedures and reduce the potential for disease and infection transmission due to lapses in cleaning and infection control procedures between procedures. In yet another broad respect, the disclosure teaches a method of logistical process of powered tools to improve cleanliness, operational efficiencies and performance. Still further it is to be understood that although this disclosure discusses the invention in terms of battery operated tools, one skilled in the art would fully appreciate that this disclosure has similar application to any pneumatic, wired or electric wall socket-powered instruments as well.
Although illustrative embodiments have been shown and described, a wide range of modification, change and substitution is contemplated in the foregoing disclosure and in some instances, some features of the embodiments may be employed without a corresponding use of other features. Accordingly, it is appropriate that the appended claims be construed broadly and in a manner consistent with the scope of the embodiments disclosed herein.
This is a Divisional Application of Non-Provisional application Ser. No. 14/551,080 filed Nov. 24, 2014, which is related to and claims priority to Provisional Application No. 61/913,266 filed Dec. 7, 2013.
Number | Name | Date | Kind |
---|---|---|---|
1497561 | Gruss | Jun 1924 | A |
1691823 | Ogilvie | Nov 1928 | A |
3013656 | Murphy, Jr. | Dec 1961 | A |
3528720 | Treace | Sep 1970 | A |
3780857 | Rosano, Jr. | Dec 1973 | A |
3934582 | Gorrie | Jan 1976 | A |
3988873 | Oliverius | Nov 1976 | A |
4050528 | Foltz | Sep 1977 | A |
4091880 | Troutner | May 1978 | A |
4128173 | Lazarus | Dec 1978 | A |
4183613 | Walchie | Jan 1980 | A |
4342392 | Cox | Aug 1982 | A |
4359052 | Staub | Nov 1982 | A |
4440317 | Clark | Apr 1984 | A |
4522196 | Cunningham | Jun 1985 | A |
4730726 | Holzwarth | Mar 1988 | A |
4741326 | Sidall | May 1988 | A |
4825850 | Opie | May 1989 | A |
4991564 | Takahashi | Feb 1991 | A |
5168863 | Kurtzer | Dec 1992 | A |
5302124 | Lansing | Apr 1994 | A |
5359991 | Takahashi | Nov 1994 | A |
5366446 | Tal | Nov 1994 | A |
5379895 | Foslien | Jan 1995 | A |
5458132 | Yabe | Oct 1995 | A |
5458133 | Yabe | Oct 1995 | A |
5487661 | Peithman | Jan 1996 | A |
5591119 | Adair | Jan 1997 | A |
5688221 | Yabe | Nov 1997 | A |
5743849 | Rice | Apr 1998 | A |
5782821 | Couch | Jul 1998 | A |
5807107 | Bright | Sep 1998 | A |
5876328 | Fox | Mar 1999 | A |
6000400 | Navis | Dec 1999 | A |
6095811 | Stearns | Aug 2000 | A |
6309358 | Okubo | Oct 2001 | B1 |
6350124 | Wade | Feb 2002 | B1 |
6558060 | Raju | May 2003 | B1 |
6716215 | David | Apr 2004 | B1 |
7357774 | Cooper | Apr 2008 | B2 |
D610429 | Olson | Feb 2010 | S |
7728262 | Faries, Jr. | Jun 2010 | B1 |
7738971 | Swayze | Jun 2010 | B2 |
9510910 | Miller | Dec 2016 | B2 |
20020045802 | Hascoet | Apr 2002 | A1 |
20030205029 | Chapolini | Nov 2003 | A1 |
20060107432 | Cicero | May 2006 | A1 |
20060111723 | Chapolini | May 2006 | A1 |
20070112336 | Aizenfeld | May 2007 | A1 |
20070112337 | Salman | May 2007 | A1 |
20070270775 | Miller | Nov 2007 | A1 |
20070281272 | Rahbari | Dec 2007 | A1 |
20080045857 | Miller | Feb 2008 | A1 |
20080045860 | Miller | Feb 2008 | A1 |
20080045861 | Miller | Feb 2008 | A1 |
20080045965 | Miller | Feb 2008 | A1 |
20080195128 | Orbay | Aug 2008 | A1 |
20080213721 | Euvrard | Sep 2008 | A1 |
20080221580 | Miller | Sep 2008 | A1 |
20090093677 | Smith | Apr 2009 | A1 |
20090176185 | Chen | Jul 2009 | A1 |
20090194446 | Miller | Aug 2009 | A1 |
20110017801 | Zemlok | Jan 2011 | A1 |
20110033137 | Gaynor | Feb 2011 | A1 |
20110082387 | Miller | Apr 2011 | A1 |
20110125138 | Malinouskas | May 2011 | A1 |
20120061262 | Merboth | Mar 2012 | A1 |
20130095227 | Bengtson | Apr 2013 | A1 |
20130184704 | Beardsley | Jul 2013 | A1 |
20130184730 | Beardsley | Jul 2013 | A1 |
20130256120 | Chao | Oct 2013 | A1 |
20140052135 | Aman | Feb 2014 | A1 |
20140121569 | Schafer | May 2014 | A1 |
20150100066 | Kostrzewski | Apr 2015 | A1 |
20150196363 | Aman | Jul 2015 | A1 |
20150202009 | Nussbaumer | Jul 2015 | A1 |
20160008080 | Beardsley | Jan 2016 | A1 |
20160095585 | Zergiebel | Apr 2016 | A1 |
20160106401 | Beardsley | Apr 2016 | A1 |
20160192989 | Aman | Jul 2016 | A1 |
20160278738 | Buchalter | Sep 2016 | A1 |
20160310134 | Contini | Oct 2016 | A1 |
20170027658 | Black | Feb 2017 | A1 |
20170055965 | Flatland | Mar 2017 | A1 |
20170196645 | Ozkarsli | Jul 2017 | A1 |
20170202487 | Casey | Jul 2017 | A1 |
Number | Date | Country |
---|---|---|
2294028 | Oct 1998 | CN |
2820102 | Sep 2006 | CN |
201572196 | Sep 2010 | CN |
2851033 | Mar 2015 | EP |
H0760835 | Mar 1995 | JP |
WO9806144 | Feb 1998 | WO |
WO0167970 | Sep 2001 | WO |
Entry |
---|
Ouyang; “Drilling to the Problem: Low-Cost Sterile Drill Covers for Surgery—Interview with Lawrence Buchan, Cofounder of Arbutus Medical”; https://www.medgadget.com/2014/09/drilling-to-the-problem-low-cost-sterile-drill-covers-for-surgery-interview-with-lawrence-buchan-cofounder-of-arbutus-medical.html; Sep. 9, 2014; pp. 1-11. |
Number | Date | Country | |
---|---|---|---|
61913266 | Dec 2013 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14551080 | Nov 2014 | US |
Child | 15965897 | US |