Aspects of the present invention relate to electrosurgical procedures, techniques, and devices that utilize radiofrequency energy to seal blood vessels during surgical procedures. In particular, and by no means of limitation, aspects of the present invention relate to devices and techniques for sealing precisely placed or otherwise sensitive or difficult to access blood vessels often found in micro-surgical procedures, difficult to access anatomy or general surgery in pediatric patients.
The use of electrosurgical energy in vessel sealing surgical procedures is common and has been used in conjunction with a variety of surgical techniques for many years. In general, electrosurgical devices used in vessel sealing rely on a combination of pressure and high frequency electric energy applied to biological tissue as a way to cut, coagulate, desiccate or seal tissue, for example, blood vessels. In the case of vessel sealing, the electrosurgical energy acts to create collagen melting and tissue fusion. Most RF surgical devices on the market today and described in the prior art have power delivery systems and end effectors that are sized to accommodate a wide range of situations, tissue thicknesses, tissue volume, and end effector bite sizes. In particular, vessel sealing technology has always relied upon the use of relatively high current settings and large end effector sizes to create a seal in varying surgical situations. No one has been able to produce a reliable vessel sealing instrument that operates at currents below 2 Amps.
For coagulation or blood vessel sealing with known devices, the average power density delivered by the end effector is typically reduced below the threshold of cutting. In some cases (e.g. a monopolar coagulation instrument) a modulated sine wave is used with the overall effect being a slower heating process which causes tissue to coagulate rather than burn and/or char to the point of cutting. In some simple dual function coagulation/cutting devices, a lower duty cycle is used for a coagulation mode and a higher duty cycle is used for a cutting mode with the same equipment. Coagulation and in particular vessel sealing techniques present unique challenges in electrosurgery
While some modern electrosurgical generators provide modulated waveforms with power adjusted in real time based on changes of the tissue characteristics (e.g. impedance), none have been able to address the complexities and sensitivities that arise when dealing with blood vessels located in delicate surgical sites or other difficult to reach blood vessels. None of the prior art or products currently offered combine aspects of applied pressure, low-power energy delivery, waveform modulation, and instrument size/configuration to safely and effectively seal the blood vessels that are often presented in micro-surgical procedures, difficult to access anatomy or pediatric general surgery.
Recent advances in vessel sealing technology have specifically abandoned an attempt to address this specialized market and instead focus on larger devices and techniques for sealing larger vessels more commonly found in general surgery or on “one-size-fits all” devices. U.S. Pat. No. 5,827,271 describes how earlier attempts to seal vessels with electrosurgery were unsuccessful in part because they attempted to apply relatively low currents to the vessels. As a solution, the advances described in U.S. Pat. No. 5,827,271 relate to increasing the current applied to the blood vessel above a certain threshold. The prior art in this field is consistent in its recognition that a high current and high generator power output is required to seal vessels. None of the prior art addresses the unique circumstances presented with small caliber blood vessels that present during micro-surgical procedures, difficult to access anatomy or general surgery in pediatric patients. None of the art accounts for a technique or device that effectively and safely seals blood vessels with an instrument that is smaller in size (both in the shaft and entry point features) and configured to be more effective which working in smaller spaces and with smaller vessel calibers. None of the prior art recognizes the role of the surgical instrument size and end effector surface area in the development of effective vessel sealing techniques and energy delivery sequencing.
In one example, a surgical system for fusing tissue is provided. The system has an electrosurgical generator capable of delivering electrosurgical power, a surgical instrument electrically connected to the electrosurgical generator and adapted to transfer the electrosurgical power from the electrosurgical generator to a distal end of the surgical instrument, and a power control circuit for controlling the delivery of radio frequency energy to the tissue in contact with the distal end of the surgical instrument. The surgical system is configured to: deliver the radio frequency energy at a non-pulsing power to the tissue for a period of time of 3 seconds or less. The non-pulsing power is held constant, the output current is held under 2 Amperes RMS, and the output voltage is held under 100 Volts RMS. The non-pulsing power further causes the tissue to begin to desiccate and to fuse within the period of time.
In another example, a power control system for delivering radio frequency energy to a surgical instrument is provided. The power control system has a power supply for delivering an output voltage and an output current to a distal end of the surgical instrument, a sensing circuit for detecting parameters indicative of an impedance of a tissue portion being fused, and a power sequencing module for automatically sequencing an electrosurgical power delivered to the surgical instrument. The power sequencing module is adapted to: apply non-pulsing power to the tissue portion for a period of time of 3 seconds or less. The period of time is measured from beginning application of the non-pulsing power through the beginning of a desiccation of the tissue portion and through a drying out and the fusing of the tissue portion. The non-pulsing power is held constant, the output current is held under 2 Amperes RMS, and the output voltage is held under 100 Volts RMS.
In another example, a surgical system for fusing tissue is provided. The surgical system has an electrosurgical generator capable of delivering electrosurgical power; a surgical instrument electrically connected to the electrosurgical generator and adapted to transfer electrosurgical power from the electrosurgical generator to a distal end of the surgical instrument; and a power control circuit for controlling the delivery of radio frequency energy to the tissue through the distal end of the surgical instrument. The surgical system is configured to deliver a radio frequency energy to the tissue, the radio frequency energy having a non-pulsed power having an output current and an output voltage; and apply the non-pulsed power to the tissue for a period of time while the non-pulsed power is held constant, the output current is held under 2 Amperes RMS, and the output voltage is held under 100 Volts RMS. The period of time is measured from beginning the application of the non-pulsed power through fusing of the tissue, and the non-pulsed power causes the tissue to begin to desiccate within the period of time, wherein the period of time is 3 seconds or less.
In another example, a method of fusing tissue is provided. The method includes: providing a surgical system having an electrosurgical generator and a surgical instrument electrically connected to the electrosurgical generator and adapted to transfer electrosurgical power from the electrosurgical generator to a distal end of the surgical instrument. The method also includes delivering a radio frequency energy to the tissue, the radio frequency energy having a non-pulsed power having an output current and an output voltage; and applying the non-pulsed power to the tissue for a period of time while the non-pulsed power is held constant, the output current is held under 2 Amperes RMS, and the output voltage is held under 100 Volts RMS. The period of time is measured from the beginning of the application of the non-pulsed power and continues through fusing of the tissue. The non-pulsed power causes the tissue to begin to desiccate within the period of time, and the period of time is 3 seconds or less.
Other aspects will become apparent to one of skill in the art upon a review of the following drawings and detailed description.
Various aspects, objects and advantages, and a more complete understanding of the present invention are apparent and more readily appreciated by reference to the following Detailed Description and to the appended claims when taken in conjunction with the accompanying Drawings, wherein:
Other aspects of devices and methods in accordance with the present invention will become known to those of skill in the art when read in conjunction with the following disclosure.
Throughout this specification references are made to the use of various materials, combinations, mechanical configurations, ranges and other aspects that may be used in various combinations to form one or more devices and methods in accordance with aspects of the present invention. It should be understood, both to one of skill in the art as well as the examining divisions in the United States Patent Office and Patent Offices throughout the world, that each of the lists of materials, examples, and other embodiments are included herein in order to teach one of skill in the art that they may be combined into various alternative embodiments, without requiring specific claim permutations of these individual features, and without departing from the spirit and scope of the invention. The claims as presented herein, as well as any potential future amendments to those claims, may include one or more combinations of these materials, ranges and other alternatives without departing from the spirit and scope of the invention described herein. In particular it is contemplated that one of skill in the art would recognize and find adequate support in the written description for any combination of the features disclosed herein, whether described in a single example or embodiment, or described in multiple and disjointed sections of the written description. The description of these various example and options is specifically drafted to comply with 35 U.S.C. § 112 of the United States Patent Laws, Article 123(2) of the European Patent Laws as well as other similar national country laws relating to the adequacy of the written description.
Aspects of a device constructed in accordance with the present invention generally pertain to a low-power vessel sealing system. Application of such devices have particular applicability in pediatric surgery and other situations where small caliber blood vessels may be encountered such as micro-surgery, difficult to access anatomy and all aspects of pediatric surgery. In addition, devices constructed in accordance with aspects of the present invention allow smaller profile instruments which in turn allow smaller incisions to be made during the associated surgical procedure. A system constructed in accordance with aspects of the present invention has as its major components an RF generator and a bipolar grasper/dissector/sealer instrument hand piece that plugs into the generator. Various power sequencing and control systems are incorporated into the generator and system logic that enable effective and safe vessel sealing capabilities at lower power outputs than is currently available or otherwise known in the art.
Features of an instrument and vessel sealing system constructed in accordance with aspects of the present invention may include one or more of the following, each of which is described in greater detail below.
Orientation and Operation of End Effectors
Aspects of a device constructed in accordance with the present invention include opposing electrically isolated end effector jaw members that provide energy delivery to tissue, such as a blood vessel, undergoing a surgical procedure. The jaws preferably open in a simultaneous dual action fashion, although other jaw movement dynamics are contemplated. The jaws are located on the distal end of a shaft that serves to maximize surgical site visibility in either open, endoscopic or laparoscopic procedures. The length of the instrument shaft may vary from 10 cm to 30 cm depending on the designed use. In one embodiment the instrument shaft may vary from 15 cm to 20 cm long. Furthermore, aspects of the present invention can be used in connection with both rigid shaft instruments and flexible shaft instruments such as might be found in steerable surgical systems or robotic surgical systems.
The size of the jaws may range from 1.0-3.5 mm in width and from 8.0-12.0 mm in length. The surfaces of the jaws may include one or more non-conductive elements that function to maintain electrical isolation between the jaw surfaces and keep the electrodes from shorting out. Preferably the jaws are designed to close in a parallel fashion, i.e. both jaw surfaces move in coordination with each other upon activation by a user. The instrument or end effector itself may or may not be built with a cutting instrument that serves to divide the tissue after sealing is accomplished.
In accordance with one aspect, an instrument end effector constructed in accordance with the present invention applies pressure to the tissue to be sealed in the range of 75-110 psi so that the tissue is adequately compressed prior to the sealing action taking place.
There is preferably limited play in the motion of the instrument handle. In one embodiment the ring handle design enables a close to 1-1 opening and closing of the jaws. A ratchet point may be incorporated into the handle assembly so that full jaw closure and the required pressure is applied prior to the sealing function.
Generator Capabilities and Function
With respect to the generator and electrical aspects of the present invention, features of a device constructed in accordance with aspects of the present invention include one or more of the following, each of which is described in greater detail below.
Energy is preferably applied to the tissue undergoing a surgical procedure at a constant power. The power delivery cycle may terminate when one or more of the following occurs: a) voltage reaches a maximum level not to exceed a set level such as 80 Volts RMS or 100 Volts RMS; b) impedance reaches a final value of between 180-350 ohms; and c) a maximum seal time of between 2 and 5 seconds is reached. In addition, voltage or current limits may be put in place that further confine the operational parameters of the power delivery system.
The generator has a power capability in the range of 25-35 Watts total power, but typical operation is preferably in the range of 8-15 Watts in order to control resolution and accuracy of the power output and to minimize the possibility of tissue charring and other damaging effects. A voltage limit incorporated into the generator operation minimizes the potential for tissue damage during a sealing procedure.
Maximum current delivery is under 2 Amps, and typical operation is in the range of 0.75-1.5 Amps. Maximum voltage of the system is 100 Volts RMS. The typical maximum voltage is in the range of 70-85 Volts RMS.
The above summary should be considered as one set of overall design parameters and not as any type of exclusive requirements for how a system or particular instrument is required to be designed in accordance with aspects of the present invention. In addition, any examples given throughout this specification, and any data or test results given throughout this specification, should not be used as evidence during prosecution of the claims herein of any intention by the applicant to limit the scope of any claims to those particular embodiments or examples. For example, maximum current delivery may be limited to under 1.5 Amps in one embodiment, under 1.0 amps in another embodiment and under 0.5 amps in yet another embodiment. In addition to the above, operation of the system may be in the range of 1.0-1.25 Amps. In another embodiment operation of the system may be in the range of 0.5 to 2 Amps. In other embodiments, maximum voltage of the system is 75 Volts RMS. In yet other embodiments maximum voltage of the system is 50 Volts RMS. In another embodiment the typical maximum voltage is in the range of 50-100 Volts RMS.
One of skill in the art would know to utilize various permutations of the described examples to arrive at the claim scope submitted below.
With reference to
A handle assembly 106 generally includes a stationary handle portion 108 and a thumb lever portion 110 that operate in conjunction with an actuation mechanism 119 to control the movement, physical engagement of and electrical engagement of the end effector 104. Each of the handle portion 108 and thumb lever portion 110 define an area for a user's fingers to engage the instrument and operate the handle assembly 106 in a scissor-like movement. As such, movement of the thumb lever portion 110 with respect to the handle portion 108 facilitates movement of the end effector from an open position to a closed position. In one embodiment the handle portion 108 and the thumb level portion 110 enable a single action movement of the end effector 104. A ratchet mechanism 112 is included for selectively locking the handle assembly 106 at a selected position during movement. In another embodiment, several ratchet stops are include so that a user can select a closure position and/or pressure applied by the end effector around a tissue sample. Electrical cable assembly 116 extends from a position in an electro-mechanical sub-assembly 118 and serves generally to connect the surgical instrument 100 to a power source, such as a radio frequency generator described below. Further details of the electrical cable assembly 116 are described below.
With respect to the grasper/dissector, sealer end effector 104 depicted in
With reference to
Actuation cables 151a and 151b route through the cable collar 150. Joined to each cable is a mechanically fastened ferrule crimp 200a and 200b near the proximal end of both cables, which allows the actuation cables to be pulled (to clamp/close the jaw portions 130 and 132) and pushed (to open the jaw portions 130 and 132). In the short section of actuation cables proximal to the ferrule crimp (shown as 151a), each of the two cables are electrically connected to the two poles of the electrosurgical power wires (182 and 184). This connection can be done with an additional electrical ferrule (crimped) or soldered. This is not a load bearing connection; simply electrical. The electrosurgical power wires wrap around the cable collar 150 to provide enough path length to allow the cable collar to rotate approximately one full rotation without causing undue strain on the power wires or electrical connections. The two poles of the electrosurgical trigger switch (188 and 186) form a closed loop in the flex circuit 142 (See
In accordance with proper functioning of the instrument and system is that the two poles remain electrically isolated at all times. Power wires 182 and 184, solder/crimp connections 182a and 182b (not shown), 151a and 151b, and jaw portions 130 and 132 have to remain isolated from each other.
Located on an interior surface of grasper portion 130 are a series of non-conductive raised elements 134a, 134b, and 134c that ensure that the interior conductive surfaces of grasper portions 130 and 132 do not physically touch and are maintained at a constant and predictable separation distance from each other when the grasper portions 130 and 132 are in a closed position. Raised elements 134a, 134b and 134c may be on either of the interior surfaces of grasper portions 130 or 132.
In other embodiments, non-conductive elements 134a, 134b and 134c may be made of a variety of non-conductive material. For example the non-conductive elements may be made of a ceramic material, nylon, polystryenes, Nylons, Syndiotacticpolystryrene (SPS), Polybutylene Terephthalate (PBT), Polycarbonate (PC), Acrylonitrile Butadiene Styrene (ABS), Polyphthalamide (PPA), Polymide, Polyethylene Terephthalate (PET), Polyamide-imide (PAl), Acrylic (PMMA), Polystyrene (PS and HIPS), Polyether Sulfone (PES), Aliphatic Polyketone, Acetal (POM) Copolymer, Polyurethane (PU and TPU), Nylon with Polyphenylene-oxide dispersion and Acrylonitrile Styrene Acrylate.
Handle portion 108 includes an exterior shell portion 144 that forms part of a casing that surrounds several of the components within electrical connection and linkage system 118. A second portion of the casing is indicated as reference number 146. Casing portion 146 connects to thumb lever 110. Enclosed in a top portion of casing portion 146 are a shaft collar 148, a cable collar 150, and a ring cable collar 154 that serve to actively engage the shaft assembly 102 with the overall handle assembly 118. Cable assembly 116 is coupled with a flex circuit 142. A length of cable 140 extends from the cable assembly 116 and is coiled about cable collar 150 in order to allow the shaft assembly 102 to rotate freely without undue resistance to the user. Rotational element 114 couples to the housing portions 144 and 146 and allow a user to rotate the shaft assembly 102 and alter the presentation of the end effector 104 during a procedure.
Various fastening devices secure the components of the handle assembly together such as screws 158, 160, 162, and 164 as well as pin 156. Alternatively, sonic welding or other connection techniques may be utilized to secure the components together. Switch 152 engages with the thumb lever 110 and is adapted to engage the power delivery through cable assembly 116 as delivered from an electrosurgical generator.
The cable assembly 116 houses four conductors, 182, 184, 186, and 188. In general the four conductors provide the RF power 182 and 184 to the end effector 104 a common/ground conductor 186 and a switch conductor 188 that couples with an identification resistor that signals the general as to what type of device is plugged in to the power source.
With reference to
Generator 300 inputs AC power from a wall outlet 302 and delivers that AC power to a power module 304 such as an IEC power entry module. Power module 304 may include such components as a DPDT switch 306 and 2 replaceable fuses 308. Power is transferred from power module 304 to a power supply unit 310. Preferably in connection with medical applications the power supply unit 310 is a medical grade power supply such as a CSS65-24 from Lambda. DC to DC converters 312 and 314 convert a 24 volt output of the power supply 310 into a pair of 12 volt outputs that are used to power control circuits. Power supply 310 then serves as the inputs 320 and 330 to a Buck-Boost Converter 340. The output of the converter 340 is passed to an H-Bridge circuit 345 which then passes that signal to a resonant LC transformer circuit 350. The transformer circuit 350 includes capacitor 352, inductor 354 and transformer 356. In one embodiment, transformer 356 is a 1:9 transformer. From the transformer 356, power is delivered to a load 400. In one embodiment, load 400 is a grasper/sealer instrument 100 as described above in connection with
Controller 360, in conjunction with the components described below, is used to sense various aspects of the energy applied to load 400, and adjust one or more of the energy characteristics in response to the vessel sealing process. Coupled to the load 400 (e.g. an electrosurgical instrument) are a current sensor circuit 390 and a voltage sensor circuit 395. Based on the sensed current and voltage, op-amps 392 and 394 pass these values to conversion circuits 380 that includes voltage RMS-DC Converter 382 and a current RMS-DC Converter 384. A multiplier 386 and a divider 388 derive power and impedance respectively from the sensed voltage and current. Each of the circuits 382, 384, 386 and 388 provide input data to the controller 360 which can then process and analyze the various input to indicate the state of the vessel sealing process. These circuits 382, 384, 386 and 388 also may be used as direct feedback signals to the Buck-Boost converter controller.
In some embodiment a user input panel 368 may be included that allows operator interaction with the controller and may include a display and some form of input device (such as a keyboard, mouse, pointer, dials or buttons). Data from the controller 360 may be output via analog or digital means such as an RS-232 connector 362, programming port 364 or memory chip 366.
Power Delivery Sequencing
Throughout the sealing process, the impedance of the vessel is calculated in real-time by measuring the voltage and current to the jaws. (Z=V/I). Since it is known that the impedance will follow the prescribed format, impedance thresholds can be set which act as trip points to cause the power profile curve to advance to the next phase of its settings. However, if the impedance limit is not met before the time has elapsed, the curve advances to the next phase. Each phase of the curve advances to the next phase according to an “OR” logic. If any one of the parameters is met according to impedance or time, the power profile is advanced.
In accordance with one aspect of the present invention it is desirable to achieve a single set of parameters that would be effective at sealing blood vessels with a diameter under 6 mm, typical in micro-surgery applications, pediatric patients, and surgical sites that are often difficult to reach and/or visualize.
The following examples are illustrative of aspects of the present invention but are not meant to be limiting under 35 U.S.C. § 112 of the United States Patent Laws, Article 123(2) of the European Patent Laws or any corresponding national country patent laws concerning the adequacy of the written description. By giving these examples, it is submitted that variations in the scope of the test results and corresponding implementations and claim scope are clearly and unambiguously disclosed to one of skill in the art.
Vessel Sealing Results
Table 1 shows the electrical characteristics associated with various seals performed according to aspects of the present invention.
With respect to the test results described in Table 1, the examples were completed by the power being held constant while current and voltage were allowed to float to the maximum values. The power was disengaged when Vmax reached 80 Volts RMS or Impedance reached 200 ohms. An initial cycle was established with a one (1) second ramp to maximum power followed by a two (2) second hold and then a one (1) second hold at a minimum power of 5 watts. If the power is left on too long, the additional power applied to the vessel will rapidly increase the impedance. In one embodiment it is envisioned that the power is deactivated within 100 msec. Complete and translucent seals were obtained at as low as 7 watts maximum power.
The time to Pmax for the series was 1 second. The observations were that 5 W did not provide an adequate seal and both 10 W and 15 W began to show charring of the vessel. It was found that translucency wasn't necessarily a good indication of a good seal. Burst tests of some clear seals did not achieve the desired 500 mm Hg (˜10 psi level). It was found that some charring of the seals seemed to produce stronger seals that reach a 500 mm Hg burst strength. A series of 30 tests were then run where the time to Pmax was reduced to 0.5 seconds in order to reduce the total time of the seal and to get more energy into the vessel more quickly.
The time to Pmax was fixed at 0.5 seconds and the time to Pmin was fixed at 0.2 seconds. Pmax and Pmin were then incrementally adjusted to optimize the seal results. This testing established Pmax at 12 W and Pmin at 5 W, with Pmin being as low as 0 Watts as providing reasonable and consistent seals. Subsequent testing attempted to bracket the range for which good seals could be achieved. This ranged from 10 W to 14 W. In other embodiments the range for Pmax and Pmin is between 5 W to 10 W, alternatively between 2 W to 15 W.
In a second example, the goal was to enable seal times under 3 seconds. In this embodiment, power delivery of 12 watts for 1.1 seconds followed, by 5 watts for 1.5 seconds gave good results. These settings were initially set based on observations that seals at 7 W took about 3 seconds. If shorter sealing times were desired, it was understood that a higher power setting would be required since the seals are the result of the total energy delivered to the vessel.
Approximately 100 seals (See Table 1) were performed with varying results. The primary problem with obtaining repeatable seals at a constant power setting was that there was often arcing between the jaws and a subsequent burn-through of the vessel. In particular, tests performed using the parameters that worked on large vessels when applied to smaller vessels often had arcing at the end of the seal cycle. Experimentation realized that settings were not simply power and time dependent. Further testing incorporated the impedance threshold and voltage and current limits described below.
In another example seals were effected based on real-time measurements of the impedance. Examples of viable seals were created at 7 W peak power for 2.9 sec followed by 2 W of sustained minimum power.
During the seals shown in
Testing confirmed that controlled RF sealing of blood vessels can be achieved with as little as 7 Watts of power applied. Voltage and current applied during these seals were less than 50 volts and 0.6 A, respectively. The square area of the jaws used is 0.023 in2/14.84 mm2. The maximum current density applied by the system and as represented by I/A=0.6 Amps/14.84 mm2 or 0.040 A/mm2.
These tests and examples demonstrated that vessels up to 6 mm in size can be sealed with a low power RF energy output that utilizes a small bipolar grasper jaw end effector. Parameters of such a system include applied pressure, current density, low voltage, impedance monitor and midpoint. Steps for sealing may include one or more of the following:
As demonstrated, vessel sealing is possible using a low power system by limiting jaw size and applying high pressure. A jaw of approximately 3 mm in width and 10-12 mm in length with a cross-sectional area of approximately 15-22 mm2 is preferred but other geometries are contemplated.
In accordance with other testing aspects, non-conductive spacers, such as ceramic beads, were incorporated into the instruments to stop the jaws before they touch and prevent short circuits in the electrical systems. The ceramic beads also provided a means of keeping the jaws parallel since the moving jaw will land on the top of the beads.
In the testing environment, a silk suture material (0.006″ diameter) that is non-conductive and will not melt during sealing was utilized to maintain the separation between jaw surfaces and prevent arcing. In the examples described in Table 1, two wraps of the suture were placed around the upper jaw so that the jaws would be separated by 0.006″ when they were closed by the spring.
In accordance with another example, and with the non-conductive sutures in place, the power was set at 12 W as a nominal starting point and bracketed from 10 W to 14 W. When full jaw bites were sealed (100% filling of jaws with tissue), there were high quality seals, excellent translucency, with minimal sticking and charring. However, attempts to perform seals using these settings on smaller vessels that fill less (50-75%) of the jaw still resulted in arcing. Additional testing bracketed Pmax between 7 W and 15 W. Pmin was set at 5 W and bracketed between 4 W and 7 W. Since there was no control limit on parameters other than power and time, the lower power settings were used to produce reasonable seals on smaller vessels in the range of 50-80% jaw tissue fill. When the vessels filled 100% of the jaws, the 7 W was not sufficient to seal and a higher power near 12 W was necessary to adequately seal these larger vessels.
In accordance with other examples, further seals were provided by setting impedance thresholds in the user interface. It was then possible to set break points in the power curve that could limit the power delivered to the jaws by watching the real-time rise in impedance. Additional seals were performed with power and time as the sole modifiers of the power curve. An impedance trigger point added to the sequence and control of power delivery enabled the generator to quickly respond to changes in impedance and make the necessary adjustments.
Current and Voltage Clamping
In another aspect of testing the device, current and voltage limits were implemented. Based on observation, by limiting the voltage to 100 V arcing was eliminated in vessels ranging from 2 to 7 mm in width. These parameters made it possible to seal vessels that filled any percentage of the jaws thus the process and device were not limited to fully filling the jaws with tissue material. As an experimental validation, raising the voltage limit to 150 volts again caused arcing. This confirmed that the voltage limit is correctly stopping the arcing from occurring. The voltage maximum was set at between 75 and 100 volts for the next series of seals performed.
As another method to determine a maximum current for sealing, a system energy check was performed to see what current level is required to boil off saline that is placed between the jaws. It was found that 1.8 A was needed to cause the saline to begin to steam and boil away when RF power is applied to the jaws and this was used as the maximum current for sealing.
In another example, it was determined that a fast application of energy with a high influx of current creates good seals. Examples included following the high current energy with a one (1) second burst of low energy with Pmin set at 5 W.
In another example, a set of jaws 0.409 inches in length were installed in the system. The old jaws were removed and found to have experienced significant pitting due to the arcing incurred while there was no voltage limit in place. In addition to being longer than the original jaws, the top edges of the jaws were eased to reduce any sharp corner effects and high current concentrations that can occur with sharp edges.
In another example, seals were made with Vmax set to 75 volts and Imax set to 1.8 A. There were never any incidents of arcing observed in subsequent seals. Seals were created with the following set of parameters:
A series of thirty-two (32) seals were performed on two animals. Vessels ranged from 2 mm to 6 mm wide. Twenty (20) of the 32 seals from the series were burst tested. All seals successfully withstood a minimum of 360 mm Hg. Seals were also successfully performed on mesentery tissue using the same settings.
Those skilled in the art can readily recognize that numerous variations and substitutions may be made in the invention, its use and its configuration to achieve substantially the same results as achieved by the embodiments described herein. Accordingly, there is no intention to limit the invention to the disclosed exemplary forms. Many variations, modifications and alternative constructions fall within the scope and spirit of the disclosed invention as expressed in the claims.
This application is a continuation of U.S. patent application Ser. No. 14/990,584 filed Jan. 7, 2016 and entitled “Low-Power Tissue Sealing Device and Method” which is a continuation of U.S. patent application Ser. No. 14/824,607 filed Aug. 12, 2015 and entitled “Low Power Tissue Sealing Device and Method,” and issued as U.S. Pat. No. 9,265,561 on Feb. 23, 2016, which is a continuation of U.S. patent application Ser. No. 13/153,513, now U.S. Pat. No. 9,144,455, which was filed on Jun. 6, 2011 and entitled “LOW POWER TISSUE SEALING DEVICE AND METHOD,” which claims the benefit of Provisional U.S. Application No. 61/352,114 filed on Jun. 7, 2010, the entire disclosures of which are hereby incorporated by reference into the present application in its entirety and for all proper purposes.
Number | Name | Date | Kind |
---|---|---|---|
4438766 | Bowers | Mar 1984 | A |
4569345 | Manes | Feb 1986 | A |
4574801 | Manes | Mar 1986 | A |
4617927 | Manes | Oct 1986 | A |
4658819 | Harris et al. | Apr 1987 | A |
4658820 | Klicek | Apr 1987 | A |
4850353 | Stasz et al. | Jul 1989 | A |
4903696 | Stasz et al. | Feb 1990 | A |
4905691 | Rydell | Mar 1990 | A |
4936281 | Stasz | Jun 1990 | A |
4961739 | Thompson | Oct 1990 | A |
5007908 | Rydell | Apr 1991 | A |
5013312 | Parins et al. | May 1991 | A |
5047026 | Rydell | Sep 1991 | A |
5222971 | Willard et al. | Jun 1993 | A |
5258006 | Rydell et al. | Nov 1993 | A |
5282799 | Rydell | Feb 1994 | A |
5304190 | Reckelhoff et al. | Apr 1994 | A |
5342359 | Rydell | Aug 1994 | A |
5342381 | Tidemand | Aug 1994 | A |
5352222 | Rydell | Oct 1994 | A |
5356408 | Rydell | Oct 1994 | A |
5403312 | Yates et al. | Apr 1995 | A |
5441516 | Wang et al. | Aug 1995 | A |
5445638 | Rydell et al. | Aug 1995 | A |
5447513 | Davison et al. | Sep 1995 | A |
5449372 | Schmaltz et al. | Sep 1995 | A |
5458598 | Feinberg et al. | Oct 1995 | A |
5462546 | Rydell | Oct 1995 | A |
5484409 | Atkinson et al. | Jan 1996 | A |
5484435 | Fleenor et al. | Jan 1996 | A |
5496317 | Goble et al. | Mar 1996 | A |
5514134 | Rydell et al. | May 1996 | A |
5540684 | Hassler, Jr. | Jul 1996 | A |
5540685 | Parins et al. | Jul 1996 | A |
5571100 | Goble et al. | Nov 1996 | A |
5573424 | Poppe | Nov 1996 | A |
5575789 | Bell et al. | Nov 1996 | A |
5584872 | LaFontaine et al. | Dec 1996 | A |
5599344 | Paterson | Feb 1997 | A |
5599350 | Schulze et al. | Feb 1997 | A |
5603711 | Parins et al. | Feb 1997 | A |
5611798 | Eggers | Mar 1997 | A |
5624452 | Yates | Apr 1997 | A |
5637111 | Sutcu et al. | Jun 1997 | A |
5647869 | Goble et al. | Jul 1997 | A |
5666035 | Basire et al. | Sep 1997 | A |
5667480 | Knight et al. | Sep 1997 | A |
5669907 | Platt et al. | Sep 1997 | A |
5674220 | Fox et al. | Oct 1997 | A |
5688270 | Yates et al. | Nov 1997 | A |
5693051 | Schulze et al. | Dec 1997 | A |
5709680 | Fates et al. | Jan 1998 | A |
5716366 | Yates | Feb 1998 | A |
5722934 | Knight et al. | Mar 1998 | A |
5725479 | Knight et al. | Mar 1998 | A |
5735848 | Yates et al. | Apr 1998 | A |
5735849 | Baden et al. | Apr 1998 | A |
5738648 | Lands et al. | Apr 1998 | A |
5743906 | Parins et al. | Apr 1998 | A |
5766170 | Eggers | Jun 1998 | A |
5769849 | Eggers | Jun 1998 | A |
5776128 | Eggers | Jul 1998 | A |
5776130 | Buysse et al. | Jul 1998 | A |
H001745 | Paraschac | Aug 1998 | H |
5797938 | Paraschac et al. | Aug 1998 | A |
5807392 | Eggers | Sep 1998 | A |
5807393 | Williamson, IV et al. | Sep 1998 | A |
5810805 | Sutcu et al. | Sep 1998 | A |
5810808 | Eggers | Sep 1998 | A |
5810811 | Yates et al. | Sep 1998 | A |
5817093 | Williamson, IV et al. | Oct 1998 | A |
5827271 | Buysse et al. | Oct 1998 | A |
5833690 | Yates et al. | Nov 1998 | A |
5860975 | Goble et al. | Jan 1999 | A |
5876401 | Schulze et al. | Mar 1999 | A |
5891141 | Rydell | Apr 1999 | A |
5891142 | Eggers et al. | Apr 1999 | A |
5902301 | Olig | May 1999 | A |
5902328 | LaFontaine et al. | May 1999 | A |
5908420 | Parins et al. | Jun 1999 | A |
5911719 | Eggers | Jun 1999 | A |
5921984 | Sutcu et al. | Jul 1999 | A |
5928135 | Knight et al. | Jul 1999 | A |
5928138 | Knight et al. | Jul 1999 | A |
6024741 | Williamson, IV et al. | Feb 2000 | A |
6024744 | Kese et al. | Feb 2000 | A |
6039733 | Buysse et al. | Mar 2000 | A |
6050996 | Schmaltz et al. | Apr 2000 | A |
D424694 | Tetzlaff et al. | May 2000 | S |
D425201 | Tetzlaff et al. | May 2000 | S |
6063086 | Benecke et al. | May 2000 | A |
6093186 | Goble | Jun 2000 | A |
RE36795 | Rydell | Jul 2000 | E |
6096037 | Mulier et al. | Aug 2000 | A |
6096058 | Boche | Aug 2000 | A |
6102909 | Chen et al. | Aug 2000 | A |
6110171 | Rydell | Aug 2000 | A |
6113598 | Baker | Sep 2000 | A |
6117152 | Huitema | Sep 2000 | A |
6126675 | Shchervinsky et al. | Oct 2000 | A |
6165175 | Wampler et al. | Dec 2000 | A |
6168594 | LaFontaine et al. | Jan 2001 | B1 |
6174309 | Wrublewski et al. | Jan 2001 | B1 |
6179834 | Buysse et al. | Jan 2001 | B1 |
6187003 | Buysse et al. | Feb 2001 | B1 |
6190383 | Schmaltz | Feb 2001 | B1 |
6190386 | Rydell | Feb 2001 | B1 |
6193653 | Evans et al. | Feb 2001 | B1 |
6206823 | Kolata et al. | Mar 2001 | B1 |
6206878 | Bishop et al. | Mar 2001 | B1 |
6228080 | Gines | May 2001 | B1 |
6238387 | Miller, III | May 2001 | B1 |
6277117 | Tetzlaff et al. | Aug 2001 | B1 |
6293945 | Parins et al. | Sep 2001 | B1 |
D449886 | Tetzlaff et al. | Oct 2001 | S |
6296640 | Wampler et al. | Oct 2001 | B1 |
6352536 | Buysse et al. | Mar 2002 | B1 |
6358249 | Chen et al. | Mar 2002 | B1 |
6361534 | Chen et al. | Mar 2002 | B1 |
6364879 | Chen et al. | Apr 2002 | B1 |
6398779 | Buysse et al. | Jun 2002 | B1 |
6419675 | Gallo, Sr. | Jul 2002 | B1 |
6436096 | Hareyama | Aug 2002 | B1 |
6440130 | Mulier et al. | Aug 2002 | B1 |
6443952 | Mulier et al. | Sep 2002 | B1 |
6443970 | Schulze et al. | Sep 2002 | B1 |
6458128 | Schulze | Oct 2002 | B1 |
6458130 | Frazier et al. | Oct 2002 | B1 |
6464702 | Schulze et al. | Oct 2002 | B2 |
6464704 | Schmaltz et al. | Oct 2002 | B2 |
6468275 | Wampler et al. | Oct 2002 | B1 |
6485490 | Wampler et al. | Nov 2002 | B2 |
6511476 | Hareyama | Jan 2003 | B2 |
6511480 | Tetzlaff et al. | Jan 2003 | B1 |
6527771 | Weadock et al. | Mar 2003 | B1 |
6547786 | Goble | Apr 2003 | B1 |
6554829 | Schulze et al. | Apr 2003 | B2 |
6572615 | Schulze et al. | Jun 2003 | B2 |
6582427 | Goble et al. | Jun 2003 | B1 |
6585735 | Frazier et al. | Jul 2003 | B1 |
6592582 | Hess et al. | Jul 2003 | B2 |
6592604 | Hess et al. | Jul 2003 | B2 |
6607529 | Jones et al. | Aug 2003 | B1 |
6610060 | Mulier et al. | Aug 2003 | B2 |
6613048 | Mulier et al. | Sep 2003 | B2 |
6616656 | Brommersma | Sep 2003 | B2 |
6616662 | Scholer et al. | Sep 2003 | B2 |
6620161 | Schulze et al. | Sep 2003 | B2 |
6623482 | Pendekanti et al. | Sep 2003 | B2 |
6652521 | Schulze | Nov 2003 | B2 |
6656176 | Hess et al. | Dec 2003 | B2 |
6667685 | Wasaki et al. | Dec 2003 | B2 |
6679892 | Guido et al. | Jan 2004 | B2 |
6682528 | Frazier et al. | Jan 2004 | B2 |
6695840 | Schulze | Feb 2004 | B2 |
6726686 | Buysse et al. | Apr 2004 | B2 |
6736813 | Yamauchi et al. | May 2004 | B2 |
6740102 | Hess et al. | May 2004 | B2 |
6743229 | Buysse et al. | Jun 2004 | B2 |
6755827 | Mulier et al. | Jun 2004 | B2 |
6770072 | Truckai et al. | Aug 2004 | B1 |
6773409 | Truckai et al. | Aug 2004 | B2 |
6773435 | Schulze et al. | Aug 2004 | B2 |
6776780 | Mulier et al. | Aug 2004 | B2 |
6790217 | Schulze et al. | Sep 2004 | B2 |
6796981 | Wham et al. | Sep 2004 | B2 |
6832998 | Goble | Dec 2004 | B2 |
6835195 | Schulze et al. | Dec 2004 | B2 |
6843789 | Goble | Jan 2005 | B2 |
6855142 | Harano et al. | Feb 2005 | B2 |
6887240 | Lands et al. | May 2005 | B1 |
6905497 | Truckai et al. | Jun 2005 | B2 |
6923805 | Lafontaine et al. | Aug 2005 | B1 |
6926716 | Baker et al. | Aug 2005 | B2 |
6929641 | Goble et al. | Aug 2005 | B2 |
6929644 | Truckai et al. | Aug 2005 | B2 |
6942662 | Goble et al. | Sep 2005 | B2 |
6960209 | Clague et al. | Nov 2005 | B2 |
6966907 | Goble | Nov 2005 | B2 |
6966909 | Marshall et al. | Nov 2005 | B2 |
6984231 | Goble et al. | Jan 2006 | B2 |
7001382 | Gallo, Sr. | Feb 2006 | B2 |
7011657 | Truckai et al. | Mar 2006 | B2 |
7033356 | Latterell et al. | Apr 2006 | B2 |
D521641 | Reschke et al. | May 2006 | S |
7041102 | Truckai et al. | May 2006 | B2 |
7063699 | Hess et al. | Jun 2006 | B2 |
7070597 | Truckai et al. | Jul 2006 | B2 |
7083619 | Truckai et al. | Aug 2006 | B2 |
7087054 | Truckai et al. | Aug 2006 | B2 |
7112201 | Truckai et al. | Sep 2006 | B2 |
7118570 | Tetzlaff et al. | Oct 2006 | B2 |
7125409 | Truckai et al. | Oct 2006 | B2 |
7131970 | Moses et al. | Nov 2006 | B2 |
7137980 | Buysse et al. | Nov 2006 | B2 |
7147637 | Goble | Dec 2006 | B2 |
7150748 | Ebbutt et al. | Dec 2006 | B2 |
7153300 | Goble | Dec 2006 | B2 |
D535396 | Reschke et al. | Jan 2007 | S |
7156842 | Sartor et al. | Jan 2007 | B2 |
7156844 | Reschke et al. | Jan 2007 | B2 |
7169146 | Truckai et al. | Jan 2007 | B2 |
7179258 | Buysse et al. | Feb 2007 | B2 |
7186253 | Truckai et al. | Mar 2007 | B2 |
7189231 | Clague et al. | Mar 2007 | B2 |
7189233 | Truckai et al. | Mar 2007 | B2 |
7204835 | Latterell et al. | Apr 2007 | B2 |
7211084 | Goble et al. | May 2007 | B2 |
7214224 | Goble | May 2007 | B2 |
7220260 | Fleming et al. | May 2007 | B2 |
7235072 | Sartor et al. | Jun 2007 | B2 |
7241296 | Buysse et al. | Jul 2007 | B2 |
7244257 | Podhajsky et al. | Jul 2007 | B2 |
7255696 | Goble et al. | Aug 2007 | B2 |
7261724 | Molitor et al. | Aug 2007 | B2 |
7267677 | Johnson et al. | Sep 2007 | B2 |
7288098 | Huitema et al. | Oct 2007 | B2 |
7297149 | Vitali et al. | Nov 2007 | B2 |
7303557 | Wham et al. | Dec 2007 | B2 |
7311709 | Truckai et al. | Dec 2007 | B2 |
7322975 | Goble et al. | Jan 2008 | B2 |
7329256 | Johnson et al. | Feb 2008 | B2 |
7354440 | Truckai et al. | Apr 2008 | B2 |
7364577 | Wham et al. | Apr 2008 | B2 |
7377920 | Buysse et al. | May 2008 | B2 |
7381209 | Truckai et al. | Jun 2008 | B2 |
7491199 | Goble | Feb 2009 | B2 |
7510556 | Nguyen et al. | Mar 2009 | B2 |
7513898 | Johnson et al. | Apr 2009 | B2 |
7553312 | Tetzlaff et al. | Jun 2009 | B2 |
7582087 | Tetzlaff et al. | Sep 2009 | B2 |
7632269 | Truckai et al. | Dec 2009 | B2 |
7648499 | Orszulak et al. | Jan 2010 | B2 |
7717910 | Goble | May 2010 | B2 |
7722602 | Mihori | May 2010 | B2 |
9532828 | Condie et al. | Jan 2017 | B2 |
9615826 | Shelton, IV et al. | Apr 2017 | B2 |
10166064 | Kennedy et al. | Jan 2019 | B2 |
10441354 | Govari et al. | Oct 2019 | B2 |
10582964 | Weinberg et al. | Mar 2020 | B2 |
20010037110 | Schmaltz et al. | Nov 2001 | A1 |
20020082596 | Buysse et al. | Jun 2002 | A1 |
20020173787 | Buysse et al. | Nov 2002 | A1 |
20030014052 | Buysse et al. | Jan 2003 | A1 |
20030014053 | Nguyen et al. | Jan 2003 | A1 |
20030018332 | Schmaltz et al. | Jan 2003 | A1 |
20030073987 | Sakurai et al. | Apr 2003 | A1 |
20030109875 | Tetzlaff et al. | Jun 2003 | A1 |
20030114874 | Craig et al. | Jun 2003 | A1 |
20030139741 | Goble et al. | Jul 2003 | A1 |
20030181910 | Dycus et al. | Sep 2003 | A1 |
20030195544 | Hess et al. | Oct 2003 | A1 |
20040006340 | Latterell et al. | Jan 2004 | A1 |
20040015163 | Buysse et al. | Jan 2004 | A1 |
20040092927 | Podhajsky et al. | May 2004 | A1 |
20040097921 | Hess et al. | May 2004 | A1 |
20040122423 | Dycus et al. | Jun 2004 | A1 |
20040143263 | Schechter et al. | Jul 2004 | A1 |
20040147925 | Buysse et al. | Jul 2004 | A1 |
20040186492 | Hess et al. | Sep 2004 | A1 |
20040225288 | Buysse et al. | Nov 2004 | A1 |
20040230262 | Sartor et al. | Nov 2004 | A1 |
20040249374 | Tetzlaff et al. | Dec 2004 | A1 |
20050010212 | McClurken et al. | Jan 2005 | A1 |
20050101945 | Sakurai et al. | May 2005 | A1 |
20050101952 | Lands et al. | May 2005 | A1 |
20050113823 | Reschke et al. | May 2005 | A1 |
20050134324 | Boyer et al. | Jun 2005 | A1 |
20050137592 | Nguyen et al. | Jun 2005 | A1 |
20050203504 | Wham et al. | Sep 2005 | A1 |
20050203507 | Truckai et al. | Sep 2005 | A1 |
20050240179 | Buysee et al. | Oct 2005 | A1 |
20050261676 | Hall et al. | Nov 2005 | A1 |
20050261677 | Hall et al. | Nov 2005 | A1 |
20050267464 | Truckai et al. | Dec 2005 | A1 |
20060030848 | Craig et al. | Feb 2006 | A1 |
20060047275 | Goble | Mar 2006 | A1 |
20060224155 | Schmaltz | Oct 2006 | A1 |
20060235440 | Huitema et al. | Oct 2006 | A1 |
20060235441 | Huitema et al. | Oct 2006 | A1 |
20060235442 | Huitema | Oct 2006 | A1 |
20060235443 | Huitema et al. | Oct 2006 | A1 |
20060235444 | Huitema et al. | Oct 2006 | A1 |
20060235468 | Huitema et al. | Oct 2006 | A1 |
20060259036 | Tetzlaff et al. | Nov 2006 | A1 |
20060264929 | Goble et al. | Nov 2006 | A1 |
20060271042 | Latterell et al. | Nov 2006 | A1 |
20070016235 | Tanaka et al. | Jan 2007 | A1 |
20070038209 | Buysse et al. | Feb 2007 | A1 |
20070062017 | Dycus et al. | Mar 2007 | A1 |
20070093810 | Sartor et al. | Apr 2007 | A1 |
20070118163 | Boudreaux et al. | May 2007 | A1 |
20070142832 | Sartor et al. | Jun 2007 | A1 |
20070146113 | Truckai et al. | Jun 2007 | A1 |
20070149987 | Wellman et al. | Jun 2007 | A1 |
20070149998 | Wicks et al. | Jun 2007 | A1 |
20070149999 | Szabo et al. | Jun 2007 | A1 |
20070150002 | Szabo et al. | Jun 2007 | A1 |
20070156140 | Baily | Jul 2007 | A1 |
20070173803 | Wham et al. | Jul 2007 | A1 |
20070173804 | Wham et al. | Jul 2007 | A1 |
20070191712 | Messerly et al. | Aug 2007 | A1 |
20070203488 | Fleming et al. | Aug 2007 | A1 |
20070208339 | Arts et al. | Sep 2007 | A1 |
20070213712 | Buysse et al. | Sep 2007 | A1 |
20070219549 | Latterell et al. | Sep 2007 | A1 |
20070273340 | Miller et al. | Nov 2007 | A1 |
20070299439 | Latterell et al. | Dec 2007 | A1 |
20080004639 | Huitema et al. | Jan 2008 | A1 |
20080009849 | Goble et al. | Jan 2008 | A1 |
20080009850 | Goble et al. | Jan 2008 | A1 |
20080015615 | Molitor et al. | Jan 2008 | A1 |
20080027465 | Vitali et al. | Jan 2008 | A1 |
20080027466 | Vitali et al. | Jan 2008 | A1 |
20080132888 | Iida et al. | Jan 2008 | A1 |
20080045942 | Truckai et al. | Feb 2008 | A1 |
20080077131 | Yates et al. | Mar 2008 | A1 |
20080132893 | D'Amelio et al. | Jun 2008 | A1 |
20080147062 | Truckai et al. | Jun 2008 | A1 |
20080188851 | Truckai et al. | Aug 2008 | A1 |
20080215051 | Buysee et al. | Sep 2008 | A1 |
20080287948 | Newton et al. | Nov 2008 | A1 |
20080294156 | Newton et al. | Nov 2008 | A1 |
20090012516 | Curtis et al. | Jan 2009 | A1 |
20090043304 | Tetzlaff et al. | Feb 2009 | A1 |
20090076506 | Baker | Mar 2009 | A1 |
20090093804 | Newton | Apr 2009 | A1 |
20090171353 | Johnson et al. | Jul 2009 | A1 |
20090182323 | Eder et al. | Jul 2009 | A1 |
20090234355 | Edwards et al. | Sep 2009 | A1 |
20090248021 | McKenna | Oct 2009 | A1 |
20090259224 | Wham et al. | Oct 2009 | A1 |
20090306660 | Johnson et al. | Dec 2009 | A1 |
20090318915 | Hosier et al. | Dec 2009 | A1 |
20100042093 | Wham et al. | Feb 2010 | A9 |
20100082026 | Curtis | Apr 2010 | A1 |
20100114090 | Hosier | May 2010 | A1 |
20100130971 | Baily | May 2010 | A1 |
20100286685 | Amoah et al. | Nov 2010 | A1 |
20110184404 | Walberg et al. | Jul 2011 | A1 |
20160192978 | Kennedy et al. | Jul 2016 | A1 |
Number | Date | Country |
---|---|---|
29805534 | Aug 1998 | DE |
0640317 | Mar 1995 | EP |
0949886 | Oct 1999 | EP |
1693015 | Aug 2006 | EP |
2575660 | Apr 2013 | EP |
2016-112456 | Jun 2016 | JO |
S61-222441 | Oct 1986 | JP |
07-171163 | Jul 1995 | JP |
07-171163 | Jul 1995 | JP |
H11-5077857 | Jul 1999 | JP |
20023252 | Nov 2002 | JP |
2005-525861 | Sep 2005 | JP |
2005525861 | Sep 2005 | JP |
2006167403 | Jun 2006 | JP |
2007195980 | Sep 2007 | JP |
2008539981 | Nov 2008 | JP |
6309762 | Apr 2018 | JP |
6363635 | Jul 2018 | JP |
6483878 | Mar 2019 | JP |
2019-084380 | Jun 2019 | JP |
9710763 | Mar 1997 | WO |
2004032776 | Apr 2004 | WO |
2005004734 | Jan 2005 | WO |
2010025815 | Mar 2010 | WO |
2011156310 | Dec 2011 | WO |
Entry |
---|
Takagi, Akira, “Japanese Office Action re Application No. 2013-514276”, dated Feb. 7, 2017, pp. 19, Published in: JP. |
Nakamura, Kazuo, “Japanese Office Action re Application No. 2016-015986”, dated Jan. 31, 2017, pp. 8, Published in : JP. |
Schmidt, Matthias, “Extended European Search Report re Application No. 11792980.2”, dated Apr. 7, 2017, pp. 12, Published in: EP. |
Riehl, Philippe, “Supplementary European Search Report re Application No. 11835192”, dated Oct. 18, 2017, pp. 2, Published in: EP. |
Naito, Shintoku, “Japanese Office Action re Application No. 2013-514276”, dated Oct. 3, 2017, pp. 3, Published in: JP. |
Kazuo Nakamura, “Office Action re: Japanese Appl. No. 2018-016961,” Aug. 23, 2018, 14, JP. |
Kenichi Managuchi, Patent Grant re: Appln. No. 2019-024359, dated Aug. 3, 2021, 2 pages, Japan Patent Office. |
English Machine Translation of: Kenichi Managuchi, Patent Grant re: Appln. No. 2019-024359, dated Aug. 3, 2021, 2 pages, Japan Patent Office. |
Kazuo Nakamura “Office Action re Application No. 2018-016961,” dated Aug. 28, 2018, 14 pages, published in Japan. |
Kenichi, Yamaguchi, “Office Action re Japanese Application No. 2019-024359 and English Translation,” dated May 12, 2020, 8 pages. |
European Patent Office, Examination Division, “Communication relating to the Patent Prosecution Highway (PPH) Pilot Program” re Application No. 11792980.2, dated May 20, 2020, 1 page. Published in: Europe. |
Schmidt, Matthias, “Invitation Pursuant to Rule 137(4) re: Application No. 11792980.2,” dated Jun. 3, 2020, 5 pages, published in Europe. |
Kazuo Nakamura, “Notice of Intention to Grant,” dated Jan. 15, 2019, 1 page, no translation available, published in Japan. |
Kim, Eun Hwa, “Office Action re Application No. 13153513”, dated Mar. 13, 2015, pp. 46, Published in: US. |
Kim, Eun Hwa, “Office Action re Application No. 13153513”, dated May 22, 2014, pp. 79, Published in: US. |
Kim, Eun Hwa, “Office Action re Application No. 13153513”, dated Oct. 6, 2014, pp. 32, Published in: US. |
Neugeboren, Craig, “Response to Office Action re Application No. 13153513”, dated Oct. 15, 2014, pp. 13, Published in: US. |
Schneider, Laura, “Response to Office Action re Application No. 13153513”, dated Apr. 27, 2015, pp. 15, Published in: US. |
Schneider, Laura, “Response to Office Action re Application No. 13153513”, dated Aug. 22, 2014, pp. 17, Published in: US. |
Neugeboren, Craig, “Response to Office Action re Application No. 13153513”, dated Apr. 23, 2014, pp. 9, Published in: US. |
Kim, Eun Hwa, “Office Action re Application No. 13153513”, dated Feb. 28, 2014, pp. 9, Published in: US. |
Moribayashi, Hirokazu, “JP Office Action re Application No. 2013514276”, dated Sep. 29, 2015, pp. 7, Published in: JP. |
Moribayashi, Hirokazu, “Japanese Office Action re Application No. 2013-514276”, dated Oct. 28, 2014, pp. 9, Published in: JP. |
J.S. Kennedy, et al., “Controlled Radiofrequency Vessel Sealing System of Surgical Applications”, “Surgical Applications of Energy”, Jan. 1998, pp. 5, Publisher: SPIE Proceedings, Published in: US. |
J.S. Kennedy, et al., “Large Vessel Ligation Using Bipolar Energy: A Chronic Animal Study and Histologic Evaluation”, 1995, pp. 3, Publisher: Seventh International Meeting of the Society for Minimally Invasive Therapy, Published in: US. |
J.S. Kennedy, et al., “High Burst Strength, Servoregulated, Bipolar Vessel Sealing”, Jun. 1997, pp. 6, Publisher: Joint Euro Asian Congress of Endoscopic Surgery, Published in: TR. |
J.S. Kennedy, et al., “High Burst Strength, Feedback Controlled Bipolar Vessel Sealing”, “Surgical Endoscopy—Ultrasound and Interventional Techniques”, 1998, pp. 3, Publisher: Springer-Verlag, Inc., Published in: US. |
J.S. Kennedy, et al., “Recent Innovations in Bipolar Electrosurgery”, Jun. 1999, pp. 5, Publisher: Isis Medical Media, Ltd., Published in: US. |
Copenheaver, Blaine R., “International Search Report and Written Opinion re Application No. PCT/US2011/039365”, dated Nov. 16, 2011, pp. 11, Published in: US. |
Baharlou, S., “International Preliminary Report on Patentability re Application No. PCT/US2011/039365”, , Published in: CH. |
Copenheaver, Blaine R., “International Search Report and Written Opinion re Application No. PCT/US11/57191”, dated Feb. 14, 2012, pp. 13, Published in: US. |
Wu, M.P. et al., “Complications and Recommended Practices for Electrosurgery in Laparoscopy, Abstract Only”, 2000, pp. 67-73, vol. 179, Publisher: The American Journal of Surgery, Published in: US. |
Number | Date | Country | |
---|---|---|---|
20190008576 A1 | Jan 2019 | US |
Number | Date | Country | |
---|---|---|---|
61352114 | Jun 2010 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14990584 | Jan 2016 | US |
Child | 16026738 | US | |
Parent | 14824607 | Aug 2015 | US |
Child | 14990584 | US | |
Parent | 13153513 | Jun 2011 | US |
Child | 14824607 | US |