Occlusive vascular disease is a common ailment in people resulting in enormous costs to the health care system and this cost is growing as the population ages. Many different endovascular devices have been developed to treat various segments and various conditions of this huge market. While many of these devices are quite effective at treating plaque, blood clots, occlusions, narrowing and the like, frequently they damage the blood vessel wall which incites an exuberant reparative effort by the body which results in a process termed restenosis. Restenosis is a secondary narrowing of the blood vessels caused by the injury to the blood vessel caused by the primary treatment.
Hence, there is a need for devices that successfully treat different conditions of the blood vessels but do not damage the vessel wall and incite the exuberant healing response that may result in restenosis, or secondary narrowing.
In some instances, it is necessary to temporarily occlude the blood vessel to deliver medicaments, to retrieve blood clot, to provide embolic protection, to anchor the catheter system, and the like. Balloon catheters are used frequently for these applications, but the outward radial force of the apparatus sufficient to produce occlusion may also be sufficient to damage the vessel wall enough to cause restenosis. A safe expansion of the balloon may be insufficient to arrest blood flow and blood may leak by the occluding member, which may propel medicaments, debris, and blood clot downstream to block smaller arteries. Too much radial force may damage the vessel wall enough to cause the restenosis cascade phenomenon, which is initiated by injury and damage to the smooth muscle cells in the media of the arterial wall, a well-known complication of balloon inflations. This cascade of events resulting in restenosis has been the “Achilles Heel” of vascular intervention as many interventions which distend the vessel, including angioplasty, stent placement, and balloon occlusion, damage the vessel wall as they attempt to correct a blockage, remove a clot, or the like. Generally, this complication should be avoided at all costs.
Balloons are notorious for causing damage to the vessel wall as they apply outward radial pressure that distends the blood vessel, injuring and damaging the blood vessel. The body's reparative effort comprises a complex series of events that results in exuberant scar tissue weeks to months later which cause the vessel to narrow significantly, a process known as “re-stenosis.”
Hence, it is an object of the present disclosure to improve upon the current devices and provide blood flow occlusion in all predictable scenarios without causing damage to the blood vessel while doing so. To this end, devices and device features meeting at least some of these objectives discussed subsequently.
Additionally, improvements to other devices which may be utilized with occluding catheters as part of a system for thrombectomy, anchoring, embolic protection, vessel occlusion, drug delivery and other applications will be described.
The present disclosure relates to medical systems, devices, and methods. More specifically, the present disclosure relates to systems, devices, and methods to treat diseased or stenosed blood vessels.
Devices, systems, and methods are disclosed for occluding a blood vessel and/or capturing a blood clot or thrombus within. An occlusion apparatus may comprise inner and outer sheaths and an expandable flexible tubular sleeve. The occlusion apparatus may be advanced to a target site in the blood vessel. A dilator having a soft, compressible tip may be advanced through a lumen of the inner sheath to facilitate the advancement of the occlusion apparatus. The sheaths may be translated relative to one another to expand the flexible tubular sleeve to a funnel shape with a distal flush portion contacting the blood vessel inner wall and a proximal tapered portion. The proximal portion may be fluid permeable so that blood can pass through to apply pressure on the vessel wall through the distal portion. A capture device and/or a traction device can be advanced out of the inner sheath lumen and retracted back therein to capture thrombus. The distal portion of the device may comprise an expandable mesh braid with a memory characteristic to limit expansion.
Aspects of the present disclosure may also be directed to a funnel catheter comprising an outer tube, an inner tube slidably located within the outer tube, and a tubular sleeve having first and second ends and movable between a radially expanded, use state and a radially contracted, deployment state. The first end of the sleeve may be secured to a distal end of the outer tube. The second end of the sleeve may be secured to a distal end of the inner tube. The sleeve may have a movable, generally U-shaped direction-reversing region so that when the first and second ends move relative to one another, the position of the direction-reversing region may move relative to the distal ends of the inner and outer tubes. The direction-reversing region may comprise the distal funnel catheter end. The sleeve may be comprised of an elastomeric coating applied or covering a part of the sleeve and a part of the sleeve which may not have an elastomeric covering.
Aspects of the present disclosure may also be directed to a method for deploying a material-directing element within a tubular structure within a patient. A funnel catheter, having a distal funnel catheter end, may be selected. The funnel catheter may comprise an outer tube, an inner tube slidably located within the outer tube, and a tubular sleeve having first and second ends and movable between a radially expanded, use state and a radially contracted, deployment state. The first end of the sleeve may be secured to a distal end of the outer tube. The second end of the sleeve may be secured to a distal end of the inner tube. The sleeve may have a movable, generally U-shaped direction-reversing region, which may comprise the distal funnel catheter end. The funnel catheter may be deployed with the sleeve in the reduced diameter, deployment state and with the sleeve being generally parallel to the outer and inner tubes. The direction-reversing region may be positioned at a chosen position within a tubular structure within a patient. The distal ends of the inner and outer tubes may be moved relative to one another to cause: the position of the direction-reversing region to move relative to the first and second ends, the sleeve to form a distally-opening material-directing funnel, and the distal funnel portion to contact the inner wall of the tubular structure. The funnel may have a distal funnel portion and a proximal funnel portion. The distal funnel portion may comprise an elastomeric covering and the proximal portion may not be comprised of an elastomeric coating.
Aspects of the present disclosure may also be directed to an improved traction device which may be placed coaxially through the funnel catheters described herein. The improved traction device may comprise an outer sleeve and inner member placed coaxially within the outer sleeve. Both the sleeve and the inner member may have proximal and distal ends. The distal end of the outer sleeve may be attached to the proximal end of a length of a deformable tubular braid. The distal end of the inner member may be attached to the distal end of the length of the deformable tubular braid. When the inner member is moved proximally relative to the outer sleeve, the deformable tubular braid may contract so that the ends of the deformable tubular braid are positioned closer together and the waist of the tubular braid expands. The traction device can include further features described herein that are directed to the ease of use and more efficient function amongst other improvements. Prior traction devices are described in U.S. Pat. Nos. 6,238,412 and 6,699,260, issued to the current inventor as a co-inventor.
Aspects of the present disclosure may also be directed to occlusion apparatuses for occluding a bodily vessel. An occlusion apparatus may comprise an outer sheath, an inner sheath, and a flexible tubular sleeve having a contracted configuration and an expanded configuration. The outer sheath and the inner sheath may be translatable relative to one another to actuate the flexible tubular sleeve between the contracted and expanded configurations. The flexible tubular sleeve in the expanded configuration may have a funnel shape with a distal flush portion adapted to contact an inner wall of the bodily vessel when the flexible tubular sleeve is expanded therein and a proximal tapered portion. At least a part of the proximal tapered portion may be fluid permeable such that bodily fluid within the bodily vessel can pass through the fluid permeable part to apply pressure on the inner wall of the bodily vessel through the distal flush portion. The bodily vessel will typically comprise a blood vessel and the bodily fluid within the bodily vessel will typically be blood. The occlusion apparatus, however, may also be used with other bodily vessels such as bile ducts, esophagus, fallopian tubes, urethra, and intestines.
The inner sheath may be disposed at least partially within the outer sheath. The flexible tubular sleeve may have a first end and a second end. The first end may be coupled to a distal end of inner sheath and/or the second end may be coupled to a distal end of the outer sheath. The distal end of the inner sleeve may be distal of the distal end of the outer sleeve when the flexible tubular sleeve is in the contracted configuration. The flexible tubular sleeve may be biased to assume the expanded configuration with the funnel shape. For example, the flexible tubular sleeve comprises a shape memory material such as Nitinol. The shape memory material may be fashioned in a way to be biased to assume the funnel shape as described below.
At least a part of the distal flush portion may be fluid impermeable. The fluid impermeable part of the distal flush portion may be distal of the fluid permeable part of the proximal tapered portion when the flexible tubular sleeve is in the expanded configuration. The fluid impermeable part of the distal flush portion and/the fluid permeable part of the proximal tapered portion may extend over at least a full circumference of the flexible tubular sleeve. The fluid impermeable part of the distal flush portion may comprise an elastomeric covering.
The flexible tubular sleeve may comprise a mesh braid. The flexible tubular sleeve may comprise a U-shaped direction-reversing region adapted to move relative to one or more of the inner and outer sheaths as the inner and outer sheaths are translated relative to one another. The U-shaped direction-reversing region may be disposed at a distal end of the flexible tubular sleeve. The distal flush portion of the flexible tubular sleeve in the expanded configuration may be adapted to apply a radially outward force sufficient to occlude fluid flow in the bodily vessel.
Aspects of the present disclosure may also be directed to systems for treating a disease within a bodily vessel. A system may comprise any of the occlusion apparatuses described herein. The system may further comprise a dilator advancable at least partially through a lumen of the inner sheath of the occlusion apparatus so that a distal tip of the dilator is distal of the occlusion apparatus. The dilator may have a malleable and oversized distal tip with a greater cross-sectional area than a distal tip of the inner sheath of the occlusion apparatus. Alternatively or in combination, the system may further comprise a traction apparatus translatable at least partially through the lumen of the inner sheath so that an expandable mesh braid on a distal portion of the traction apparatus can be positioned and expanded distal of the occlusion apparatus.
The dilator and/or traction apparatus may have a guidewire lumen. Retraction of one or more of the dilator and/or traction apparatus when advanced distally past the occlusion apparatus can capture and retract material in the bodily vessel into the lumen of the inner sheath. The expandable mesh braid of the traction apparatus may be configured to apply a predetermined limited amount of radially outward force when expanded in the bodily vessel and elongate to minimize distending or damaging the bodily vessel as the expandable mesh braid is expanded. The bodily vessel will typically comprise a blood vessel and the bodily fluid within the bodily vessel will typically be blood. The occlusion apparatus, however, may also be used with other bodily vessels such as bile ducts, esophagus, fallopian tubes, urethra, and intestines.
Aspects of the present disclosure may also be directed to methods of treating a bodily vessel. A method may comprise the steps of advancing an occlusion apparatus to a target site in a bodily vessel and expanding a flexible tubular sleeve of the occlusion apparatus to a funnel shaped expanded configuration occluding the bodily vessel at the target site. A distal flush portion of the flexible tubular sleeve may contact an inner wall of the bodily vessel when expanded. A proximal tapered portion of the flexible tubular sleeve may allow bodily fluid within the bodily vessel to pass at least partially therethrough to apply pressure on the inner wall of the bodily vessel through the distal flush portion.
At least a part of the distal flush portion may be fluid impermeable and at least a part of the proximal tapered portion may be fluid permeable. To expand the flexible tubular sleeve, an inner sheath and an outer sheath of the occlusion apparatus may be translated relative to one another. Such translation can move a U-shaped direction reversing region of the flexible tubular sleeve relative to one or more of the inner or outer sheaths. The distal flush portion of the flexible tubular sleeve in the expanded configuration may be adapted to apply a radially outward force sufficient to occlude fluid flow in the bodily vessel. This radially outward force may be from a combination of the inherent structure of the expanded flexible tubular sleeve and the pressure from the bodily fluid passed through the proximal tapered portion. The flexible tubular sleeve may be biased to assume the funnel shaped expanded configuration. Expanding the flexible tubular sleeve may comprise allowing the flexible tubular sleeve to assume the funnel shaped expanded configuration.
The occlusion apparatus may be advanced over a guide wire. The method may further comprise advancing a dilator through the occlusion apparatus. The dilator may be advanced through a distal end of the occlusion apparatus. The dilator may further be retracted within the bodily vessel to capture and retract material in the bodily vessel into an inner lumen of the occlusion apparatus. The material captured and retracted may comprise a clot or a thrombus. The dilator may have a malleable and oversized distal tip having a greater cross-sectional area than a distal tip of the inner sheath of the occlusion apparatus.
The method may further comprise a step of advancing a traction apparatus through the occlusion apparatus. The traction apparatus may have an expandable mesh braid on a distal portion thereof. The traction apparatus may be advanced through a distal end of the occlusion apparatus. The method may further comprise a step of expanding the expandable mesh braid to appose the inner wall of the bodily vessel. The expanded expandable mesh braid may be configured to apply a limited radially outward force when expanded in the bodily vessel and elongate to minimize distending or damaging the bodily vessel as the tubular outer sleeve is expanded. The expandable mesh braid may comprise a plurality of wires configured to exert a maximum predetermined outward radial force when the expandable mesh braid is in the expanded configuration. The method may further comprise a step of retracting the expanded expandable mesh braid within the bodily vessel to capture and retract material in the bodily vessel into an inner lumen of the occlusion apparatus. The material captured and retracted may comprise a clot or a thrombus.
Expanding the flexible tubular sleeve of the occlusion apparatus to the funnel shaped expanded configuration may apply a radially outward force from the expanded flexible tubular sleeve to hold the flexible tubular sleeve in place in the bodily vessel. The radially outward force may comprise blood pressure transmitted through the expanded flexible tubular sleeve and an inherent radially outward force from the expanded flexible tubular sleeve. The radially outward force mostly may comprise the blood pressure transmitted through the expanded flexible tubular sleeve. The radially outward force may be transmitted from blood passing through at least a portion of the flexible tubular sleeve.
Aspects of the present disclosure may also be directed to devices for dilating a bodily vessel. The device may comprise a shaft and a malleable and oversized distal tip coupled to the shaft. The shaft may have an inner lumen for the passage of a guidewire and may be advanceable through a catheter lumen. The malleable and oversized distal tip may be coupled to the shaft and may have a cross-sectional area greater than that of the shaft. The oversized distal tip may be compressible to be advancable through the catheter lumen. The malleable and oversized distal tip may comprise a first material and the shaft may comprise a second material, the first material having a softer durometer than the second material. The malleable and oversized distal tip may have a radial cut-out to provide compressibility. Typically, the bodily vessel will be a blood vessel but the dilator devices may be used for other bodily vessels as well.
Aspects of the present disclosure may also be directed to methods for dilating a bodily vessel. The method may comprise the steps of positioning a dilator within a lumen of a catheter such that an oversized distal tip of the dilator is advanced out of the catheter lumen and decompresses and advancing the oversized distal tip of the dilator and the catheter to a target site. Typically, the bodily vessel will be a blood vessel but the dilator devices may be used for other bodily vessels as well. The target object will typically comprise a blood clot or a thrombus. The catheter may comprise an occlusion catheter, and may further comprise expanding the occlusion catheter at the target site to occlude the target site. The method may further comprise a step of retracting the oversized distal tip of the dilator to capture a target object at the target site at least partially within an expanded tubular sleeve of the occlusion catheter and/or the lumen of the catheter. The oversized distal tip of the dilator may have a greater cross-sectional area than that of a shaft of the dilator the oversized distal tip is coupled to.
Aspects of the present disclosure may also be directed to traction apparatuses for treating a bodily vessel. A, traction apparatus may comprise an inner shaft having a distal end, a tubular outer sleeve coaxially disposed over the inner shaft and having a distal end, and an expandable mesh braid coaxially disposed over the inner shaft and coupled to the distal ends of the inner shaft and the tubular outer sleeve. The inner shaft and the tubular outer sleeve may be translatable relative to one another to shift the expandable mesh braid between contracted and expanded configurations. The expandable mesh braid may be configured to apply a predetermined maximum amount of radially outward force when expanded in the bodily vessel and to elongate to minimize distending or damaging the bodily vessel as the tubular outer sleeve is expanded.
The mesh braid may comprise a plurality of wires. The plurality of wires may have been treated to provide a combination of stiffness and flexibility so that the expandable mesh braid applies the predetermined maximum amount of radially outward force when expanded in the bodily vessel and elongates to minimize distending or damaging the bodily vessel as the tubular outer sleeve is expanded. The plurality of wires may comprise a plurality of Nitinol wires. The plurality of wires may have been heat treated. The plurality of wires may be configured to exert the predetermined maximum amount of radially outward force when the expandable mesh braid is in the expanded configuration. The plurality of wires may be configured to exert the predetermined maximum amount of radially outward force independently of the inner diameter of the bodily vessel.
The expandable mesh braid in the expanded configuration may be configured to expand to appose an inner wall of the bodily vessel while minimizing resultant expansion of the bodily vessel. The expandable mesh braid may be expandable to a maximum predetermined outer diameter to apply the predetermined maximum amount of radially outward force. The expandable mesh braid may have a shape memory characteristic to limit expansion of the expandable mesh braid to the maximum predetermined outer diameter. Further translation of the inner shaft and the tubular outer sleeve relative to one another after the expandable mesh braid has been expanded to the maximum predetermined outer diameter may axially lengthen an outer surface of the expandable mesh braid. That is, a degree of translation of the inner shaft and the tubular outer sleeve relative to one another may have a non-linear relationship(s) with an outer diameter of the expandable mesh braid and/or a radially outward pressure exerted by the expandable mesh braid. The expandable mesh braid may have a permeable portion to allow fluid within the bodily vessel to enter an interior of the expandable mesh braid and may apply radially outward pressure on the bodily vessel through an impermeable portion of the expandable mesh braid. Typically, the bodily vessel comprises a blood vessel.
Aspects of the present disclosure may also be directed to methods of treating a bodily vessel. A catheter may be introduced to a target site in the bodily vessel. A traction apparatus from within an inner lumen of the catheter may be advanced to position an expandable mesh braid of the traction apparatus. The expandable mesh braid may be expanded within the target site. The expanded mesh braid may be retracted to capture and retract material in the target site into the inner lumen of the catheter. The bodily vessel will typically comprise a blood vessel. The captured and retracted material may comprise a blood clot or a thrombus.
To expand the expandable mesh braid, an outer surface of the expandable mesh braid may be apposed against an inner wall of the bodily vessel. The expandable mesh braid may be configured to apply a predetermined maximum amount of radially outward force when expanded in the bodily vessel and to elongate to minimize distending or damaging the bodily vessel as the tubular outer sleeve is expanded. The expandable mesh braid may comprise a plurality of wires configured to exert the predetermined maximum amount of radially outward force when the expandable mesh braid is in the expanded configuration. The plurality of wires may be configured to exert the predetermined maximum amount of radially outward force independently of the inner diameter of the bodily vessel. The expandable mesh braid in the expanded configuration may be configured to expand to appose an inner wall of the bodily vessel while minimizing resultant expansion of the bodily vessel.
The expandable mesh braid may be expanded to a maximum predetermined outer diameter. The expandable mesh braid may have a shape memory characteristic to limit expansion of the expandable mesh braid to the maximum predetermined outer diameter. To expand the expandable mesh braid, an inner shaft and a tubular outer sleeve of the traction apparatus may be translated relative to one another. Further translation of the inner shaft and the tubular outer sleeve relative to one another after the expandable mesh braid has been expanded to the maximum predetermined outer diameter may axially lengthen an outer surface of the expandable mesh braid. That is, a degree of translation of the inner shaft and the tubular outer sleeve relative to one another may have a non-linear relationship(s) with an outer diameter of the expandable mesh braid and/or a radially outward pressure exerted by the expandable mesh braid.
To expand the expandable mesh braid, fluid within the bodily vessel may be allowed to enter an interior of the expandable mesh braid and apply radially outward pressure on the bodily vessel through the expandable mesh braid.
All publications, patents, and patent applications mentioned in this specification are herein incorporated by reference to the same extent as if each individual publication, patent, or patent application was specifically and individually indicated to be incorporated by reference.
The novel features of the present disclosure are set forth with particularity in the appended claims. A better understanding of the features and advantages of the present disclosure will be obtained by reference to the following detailed description that sets forth illustrative embodiments, in which the principles of the present disclosure are utilized, and the accompanying drawings of which:
There is a continuing need for improved medical devices and methods to meet some or all the following objectives.
An objective may be to reduce cost. Cost reduction may be particularly important in recent years where it is clear for safety and sanitary reasons that many devices used in the vasculature should be single use devices. Devices, even if performing a function in some improved manner, will not be widely used if they are considerably more costly than the alternatives available.
Another objective may be ease of use and simplicity of understanding. Ease of use and simplicity can encourage device adoption and use by medical personnel. They can also tend to keep cost low.
Another objective may be to provide devices that entail procedures with which the medical profession is already familiar. By doing so, the skills that have been learned from previous experience can continue to have applicability.
Another objective relates to the effectiveness and thoroughness with which the device performs, such as blockage removal and vessel occlusion. For example, it is generally important that a maximum amount of the blockage be removed, recognizing that no device is likely to provide one hundred percent removal.
Another objective concerns safety—a matter, which is often so critical as to trump the other considerations. It is generally important to avoid tissue trauma. In many circumstances, it is critically important to, for example, avoid breaking up a blockage in a fashion that leads to flushing elements of the blockage throughout the body such as by damaging the blood vessels by applying excess pressure on the vessel wall in an attempt to fully occlude the vessel so that blockages can be removed and debris can be prevented from flushing downstream. A goal of the devices disclosed herein is to do just that: provide effective occlusion of a blood vessel to remove blockages and prevent debris from flushing downstream in a manner which does not damage the vessel wall in doing so.
There are often trade-offs in design considerations to achieve the above interrelated objectives. Extreme simplicity and a very simple procedure might over compromise safety. Addressing all of these considerations often calls for some trade-off between the objectives while maintaining the effectiveness and doing so safely.
However, in cases with an exaggerated systolic blood pressure of 160 mm Hg or in a larger vessel with exaggerated flow such as the carotid or renal arteries, the outward radial force and the length of braid contacting the vessel wall may not be sufficient to arrest the blood flow on a consistent basis. By providing an outward radial force adequate to arrest blood flow, the outward radial force may cause compression of the vessel wall 210, damaging the wall 210 similar to the forces exerted by balloon inflation which is known to damage the wall and result in restenosis. Hence, there may be a need for improved occlusion devices which do arrest blood flow in cases of exaggerated systolic blood pressure or exaggerated high flow states on a consistent and reliable basis, but without exerting an undue or exaggerated amount of force on the vessel wall to do so.
Moreover, there may be a dilemma with utilizing an outward radial force as the primary means of arresting blood flow as the optimal radial force to arrest blood flow varies as the optimal amount or size of the expanded funnel tip to arrest flow may vary in different sized vessels. The radial force and the size of the funnel tip will be different in a 2 mm coronary artery than in a 6 mm carotid artery and even still different in a 9 mm iliac artery. Hence, for different sized vessels, arresting the flow may demand the use of funnel tips with different sizes and different radial forces and different length of the wall contacting surface of the tubular mesh braid 200. This need for different sizes may create a situation in which it may be difficult to arrest flow over a continuum of sizes and flow situations, which can occur within a single vascular distribution as is the case in the lower extremity, with one design and one size of funnel tip. With the prior art occlusion device 100, while the funnel tip may oppose and contact the wall over a spectrum of sizes, the ability to occlude over this spectrum of sizes may be limited and may necessitate funnel tips sized specifically for the vessel size. The radial pressure exerted on a 2 mm diameter vessel will be greater than the radial pressure exerted on a 6 or 9 mm diameter vessel. As well, because of the different blood pressures in different patients and different flow rates in different arteries within the same patient, devices should be constructed to provide a greater force than is typically needed so to address the worst possible case which is elevated blood pressure and flow although most cases would not demand it. Hence, there may be a need for improved occlusion devices which do arrest blood flow over a range of vessel sizes with an optimized funnel tip which provides radial force sufficient to contact the wall but insufficient radial force to damage the wall. Meeting these needs can be accomplished by adding one or more features to the prior art occlusion device to improve upon it so that the occlusion device will arrest flow in cases of exaggerated systolic blood pressure, exaggerated high flow, and over a range of different sized arteries.
The sealing or occlusion efficiency of the prior art device can be summarized by the formula:
Occlusion efficiency of prior art occlusion device=
radial force of braid*length or area of contact.
Hence, the longer the length (greater area of contact), the less radial force may be needed to occlude.
In addition to the two features present in the prior art device 100 to occlude flow, the current occlusion device 100a also comprises a distinct coating or membrane free portion 215 of the tubular mesh braid 200 in addition to the coated or membrane containing section 216. Fluid or blood may flow freely though this coating free or membrane free section 215 of the tubular mesh braid, as indicated by the arrows 217, and by doing so will enhance the expansion of the wall contacting segment 212 of the tubular mesh braid 217 by adding the transmitted blood pressure to the outward or radial force of the funnel tip as demonstrated by the arrows 218 so that a seal is achieved no matter the systolic blood pressure, flow state of the vessel or the size of the vessel. The transmitted blood pressure may be additive to the inherent radial outward force utilized to expand a significant length of the funnel tip against the wall as the transmitted blood pressure essentially presses the impervious portion of the tubular mesh braid against the vessel wall. This added pressure can be summarized by the formula:
Occlusion efficiency of current occlusion device=
radial force of braid*area of contact+
nominal transmitted blood pressure*area of contact
Hence, the sealing ability of the current occlusion device 100a can be a reflection of a combination of the radial force of the braid, the length of the impervious wall contact portion, and the transmitted blood pressure. It can be important to recognize that the coating or membrane free or open portion of the tubular mesh braid 200 should be sized so that a portion of the coated or membrane comprising the impervious portion of the tubular mesh braid 200 is not compromised and that the length of contact of this portion to the wall is maximized while also providing a distinct segment of the tubular mesh braid 200 that allows ingress and egress of fluid or blood. Experiments have demonstrated that the optimal length of the open or membrane devoid portion is between 30-40% of the entire length of the tubular mesh braid 200 to accommodate a wide range of vessel sizes, although the ranges could vary between 20-60% and even 3-80% for different or specific situations. Actual overall braid length optimally for 2-6 mm vessels may be 1.6 cm with the coated or membrane containing section comprising 1.1 cm and the membrane or coating devoid portion comprising 0.5 cm. In larger vessels, the measurement may increase, but the approximate ratios will remain. This optimized length provides proper buckling of the braid 200 to form a funnel, proper ingress and egress of fluid so that the funnel tip expands and contracts easily, promptly, and without difficulty and also provides enough length of contact with the vessel wall so that the device occludes flow in any situation.
Because of the addition of these important features of providing an optimized section of the braid which does not comprise a membrane or coating to the prior art, the current devices can function consistently to arrest blood flow in blood vessels no matter the blood vessel size, the systolic blood pressure or the flow state of the vessel. As the systolic blood pressure is elevated, a corresponding pressure increase can be transmitted to the inside of the elastomeric coated braid through this section of uncoated braid to press it against the arterial wall insuring a proper seal and occlusion over varying blood pressures.
Experiments have demonstrated that the present occlusion device 100a can provide a nominal radial force against the blood vessel wall which is significantly augmented by the contribution of the patient's own blood pressure which is transmitted through the specific section of the braid 200 which is devoid of a membrane or coating. A 1.6 cm tubular mesh braid was constructed with 0.005″ diameter Nitinol wire and bonded to inner and outer tubular catheters and a silicone elastomer was applied over 1.1 cm of length maintaining a 0.5 cm length devoid of the elastomer. This arrangement caused the braid to buckle properly and form a funnel shape when the inner and outer catheter members were translated relative to one another. Compressive unloading measurements, which are reflective of the outward radial force of the device, were performed and the results are presented in the table shown by
The
Calculations reveal that a systolic blood pressure of 150 mm Hg would demand at least 285 grams of outward radial pressure from a balloon or even the prior art funnel catheter to occlude a blood vessel. At this level, a segment of the arterial wall adjacent to the upstream balloon face may experience 570 grams of force as there would be at least 285 grams of outward radial force provided by the balloon and 285 grams from the 150 mm Hg blood pressure. This may be enough to damage the vessel wall and create immediate or delayed complications. The occlusion devices of the present disclosure may subject the vessel wall to only 285 grams from the transmitted blood pressure (which it is adapted to withstand) and 1.4 grams from the inherent outward radial force of the devices. Hence, the occlusion devices of the present disclosure may add only 0.5% added pressure to the wall over the transmitted blood pressure whereas a balloon or even the prior art funnel catheter needs 100% added pressure to occlude. Hence, the occlusion devices of the present disclosure may be dramatic improvements in safety over prior art devices as vessel occlusion can be accomplished over a wide range of vessel sizes, flow rates and blood pressures by exerting a very minimal outward radial pressure on the vessel wall which is insufficient to damage to the vessel wall.
By providing a membrane devoid section of braid 200 through which the blood may flow into the space between the inner shaft 196 of the device 100a and the membrane containing braid 200, the total force of the device will generally always exceed the patient's systolic blood pressure. The systolic blood pressure will often be exceeded by the inherent outward radial force of the braid 200 (which is less than one gram in many instances) as the outward radial force of the braid 200 and the transmitted systolic blood pressure will often combine to press the membrane containing segment of the braid against the vessel wall preventing any leakage of blood around the device 100a.
To create a section of tubular braid 200 with portions that are impervious to flow and other portions in which fluid and blood may flow freely though, there may be several different methods available. Dip coating is one very practical method to apply the elastomer to the braid 200 and the braid 200 may be dipped into a solution of the elastomer just enough to make part of the section of tubular braid 200 impervious to fluids leaving the non-dipped portion free of the elastomer. Such dip coating can be somewhat inexact and tedious however. It may be appropriate to dip the entire section of tubular mesh braid 200 into the elastomer or coat the entire section in some other manner and then laser cut the elastomer from the intended porous portion of the braid 200. This latter method can be more exact and precise and can leave a portion of the tubular braided section 200 with an elastomeric membrane or coating and part of the section with no elastomer and freely porous. There are other means known in the art of creating porous and non-porous sections of tubular braid.
In some cases, it may be advantageous to provide another device to propel a blood clot or other substance into the mouth of a catheter or even to expand to serve as a filter or an additional occlusion device. This may include a device to engage the blood clot and pull it toward the catheter mouth or opening. Prior art traction devices include balloon catheters, which are known to damage the vessel wall when fully expanded, and braided devices such as described in my prior U.S. Pat. No. 6,635,068, entitled “Occlusion, Anchoring, Tensioning and Flow Direction Apparatus and Methods for Use.” Even the latter prior art traction device may damage the vessel wall in at least some cases by a combination of the outward radial force and the irregular surface caused by the braid. A prior art traction device 300 is illustrated in
As demonstrated in
The vessel wall 310 may easily be damaged by the intended overexpansion or the inadvertent overexpansion of the braid, or other expansile element, in a vessel. This damage is illustrated in
The current inventor has conducted experiments have shown that damage to the arterial wall can be avoided by such a device if constructed of Nitinol wires treated to provide a combination of stiffness and flexibility that not only creates a limited amount of outward radial force against the wall, but also causes the braid to elongate when placed in smaller vessels. This limited radial force and elongation are illustrated by the present traction device 100a in
It is important that the improved current device provide a consistent and balanced outward radial force over a fairly wide range of vessel diameters as well. In other words, the current device, because of a combination of structural elements, should provide a limited amount of outward radial force insufficient to damage the vessel wall which is independent of the vessel wall diameter over a range of vessel diameters.
While the current disclosure discusses the specific location of the elastomer and the elastomer free portion of the braid, there may in certain instances be a need for the elastomer to cover or be applied to only a minimal section of the braid. This may especially be true when complete occlusion is not desirable. In these instances, the devices may be constructed similarly to the above examples, with the elastomer free section comprising a majority or even all of the braid. The elastomer containing section may be small or absent in at least these cases. The elastomer containing section may be placed on the braid so that a portion of the funnel is impermeable, but the funnel apparatus would typically not obstruct or occlude flow.
The traction device 300a of the present disclosure may utilize 0.004″ diameter Nitinol wires that have been heat treated to provide a combination of stiffness and flexibility and by varying the pics per inch, the number of wires, the shape of the wires, the crossing angle of the wires as well as the density of any elastomeric coating, if any, so that the expanded device will achieve a “programmed” outward radial force even if attempts are made to expand the device beyond the diameter of the vessel. As the graph in
For example, if the prior art traction device 300 were placed in a 5.5 mm diameter vessel and expanded to 5.5 mm so that it abuts the wall, it would provide a certain amount of outward radial force against the wall. If the same device were utilized in a 3.5 mm vessel and expanded to 5.5 mm diameter, the prior art traction device 300 may over-distend the wall of the vessel 2 mm or so to the 5.5 mm diameter creating a situation that may very easily cause damage to the vessel wall as previously illustrated in
An elastomer may be provided to cover all or part of the braid 304, especially if the present traction device 300a is used as a vascular occluder, and may increase or decrease the outward radial force depending on the properties, thickness and consistency of the elastomer. A braid 304 of different programmed properties may be utilized in that instance.
Yet another way of controlling the outward radial force may be to utilize an elastic or elastomeric material, such as, but not limited to, silicone, urethane, neoprene, isoprene, Pebex, chronoprene and the like for the inner member or outer sleeve, so that the inner member stretches or the outer member compresses if attempts are made to over expand the device within a vessel. Instead of the excessive outward radial force being transmitted to the wall, in this case, it can be transmitted to the at least somewhat elastic inner member by stretching said member after a selected outward radial pressure is achieved by the braid or expansile component or by compressing the outer sleeve after a selected outward radial pressure is achieved by the braid or expansile component, or by utilizing both. As well, ways to elongate the braid may also be utilized with either ways to stretch the inner member and with ways to compress the outer sleeve, or both, to prevent outward radial force from being transmitted to the vessel wall beyond a selected level.
Furthermore, since the outward radial force is usually created in a linear relationship by pulling the inner member in relation to the outer sleeve, a spring in the handle of the device attached to the inner member may prevent the full expansion of the braid and limit the outward radial pressure. Hence, a spring may be used alone or in combination with any or all of the features listed above to limit the outward radial pressure.
Yet another way of providing minimal outward radial force to the vessel wall 310 with an expandable braid device 304, as illustrated in
Referring back to
The shaft 222 of the dilator 220 may be sized to fit within the ID of the funnel catheter inner sleeve 196. The tip 221 of the dilator 220 may be the same size as the OD of the funnel catheter 100, 100a. The dilator 220 comprises a lumen 223 for receiving a guide wire. So that the tip of the dilator 220 may fit through the ID of the funnel catheter 100, 100a, it must be compressible. One way of providing compressibility may be to utilize a softer durometer substance within the tip. Another way may be to place a notch or a wedge shaped cutout 230 in the tip as demonstrated in
While preferred embodiments of the present invention have been shown and described herein, it will be obvious to those skilled in the art that such embodiments are provided by way of example only. Numerous variations, changes, and substitutions will now occur to those skilled in the art without departing from the invention. It should be understood that various alternatives to the embodiments of the invention described herein may be employed in practicing the invention. It is intended that the following claims define the scope of the invention and that methods and structures within the scope of these claims and their equivalents be covered thereby.
This application is a continuation of U.S. application Ser. No. 14/554,348, filed on Nov. 26, 2014, which claims the benefit of U.S. Provisional Application No. 62/007,553, filed Jun. 4, 2014, which application is incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
2230226 | Auzin | Feb 1941 | A |
2259488 | Raiche | Oct 1941 | A |
3050066 | Koehn | Aug 1962 | A |
3540431 | Kazi | Nov 1970 | A |
3799172 | Szpur | Mar 1974 | A |
3831587 | Boyd | Aug 1974 | A |
3834394 | Hunter et al. | Sep 1974 | A |
3952747 | Kimmell, Jr. | Apr 1976 | A |
3978863 | Fettel et al. | Sep 1976 | A |
3996938 | Clark, III | Dec 1976 | A |
4030503 | Clark, III | Jun 1977 | A |
4372293 | Vijil-Rosales | Feb 1983 | A |
4401124 | Guess et al. | Aug 1983 | A |
4425908 | Simon | Jan 1984 | A |
4445892 | Hussein et al. | May 1984 | A |
4494531 | Gianturco | Jan 1985 | A |
4573966 | Weikl et al. | Mar 1986 | A |
4581017 | Sahota et al. | Apr 1986 | A |
4582061 | Fry | Apr 1986 | A |
4606347 | Fogarty et al. | Aug 1986 | A |
4608965 | Anspach, Jr. et al. | Sep 1986 | A |
4611594 | Grayhack et al. | Sep 1986 | A |
4621636 | Fogarty | Nov 1986 | A |
4646736 | Auth | Mar 1987 | A |
4650466 | Luther | Mar 1987 | A |
4696304 | Chin | Sep 1987 | A |
4727873 | Mobin-Uddin | Mar 1988 | A |
4762130 | Fogarty et al. | Aug 1988 | A |
4793348 | Palmaz | Dec 1988 | A |
4794925 | Mori | Jan 1989 | A |
4794928 | Kletschka | Jan 1989 | A |
4799495 | Hawkins et al. | Jan 1989 | A |
4820270 | Hardcastle et al. | Apr 1989 | A |
4832055 | Palestrant | May 1989 | A |
4858810 | Intlekofer et al. | Aug 1989 | A |
4869259 | Elkins | Sep 1989 | A |
4895560 | Papantonakos | Jan 1990 | A |
4921478 | Solano et al. | May 1990 | A |
4921484 | Hillstead | May 1990 | A |
4926858 | Gifford, III et al. | May 1990 | A |
4946440 | Hall | Aug 1990 | A |
4977897 | Hurwitz | Dec 1990 | A |
5011488 | Ginsburg | Apr 1991 | A |
5048530 | Hurwitz | Sep 1991 | A |
5059166 | Fischell et al. | Oct 1991 | A |
5078685 | Colliver | Jan 1992 | A |
5081997 | Bosley, Jr. et al. | Jan 1992 | A |
5092839 | Kipperman | Mar 1992 | A |
5100425 | Fischell et al. | Mar 1992 | A |
5102415 | Guenther et al. | Apr 1992 | A |
5108420 | Marks | Apr 1992 | A |
5112347 | Taheri | May 1992 | A |
5116352 | Schnepp-Pesch et al. | May 1992 | A |
5135484 | Wright | Aug 1992 | A |
5171305 | Schickling et al. | Dec 1992 | A |
5176659 | Mancini | Jan 1993 | A |
5183463 | Debbas | Feb 1993 | A |
5192290 | Hilal | Mar 1993 | A |
5209727 | Radisch, Jr. et al. | May 1993 | A |
5213569 | Davis | May 1993 | A |
5221269 | Miller et al. | Jun 1993 | A |
5226909 | Evans et al. | Jul 1993 | A |
5250060 | Carbo et al. | Oct 1993 | A |
5275611 | Behl | Jan 1994 | A |
5312360 | Behl | May 1994 | A |
5328471 | Slepian | Jul 1994 | A |
5329942 | Gunther et al. | Jul 1994 | A |
5330484 | Guenther et al. | Jul 1994 | A |
5334211 | Shiber | Aug 1994 | A |
5336205 | Zenzen et al. | Aug 1994 | A |
5342306 | Don Michael | Aug 1994 | A |
5370660 | Weinstein et al. | Dec 1994 | A |
5380273 | Dubrul et al. | Jan 1995 | A |
5380284 | Don Michael | Jan 1995 | A |
5382259 | Phelps et al. | Jan 1995 | A |
5383466 | Parti | Jan 1995 | A |
5383897 | Wholey | Jan 1995 | A |
5397307 | Goodin | Mar 1995 | A |
5410093 | Dorai | Apr 1995 | A |
5419774 | Willard et al. | May 1995 | A |
5423799 | Shiu | Jun 1995 | A |
5431676 | Dubrul et al. | Jul 1995 | A |
5437631 | Janzen | Aug 1995 | A |
5441485 | Peters | Aug 1995 | A |
5443454 | Tanabe et al. | Aug 1995 | A |
5454790 | Dubrul | Oct 1995 | A |
5456667 | Ham et al. | Oct 1995 | A |
5462529 | Simpson et al. | Oct 1995 | A |
5483976 | Mclaughlin et al. | Jan 1996 | A |
5490521 | Davis et al. | Feb 1996 | A |
5496275 | Sirhan et al. | Mar 1996 | A |
5497782 | Fugoso | Mar 1996 | A |
5498236 | Dubrul et al. | Mar 1996 | A |
5501408 | Kang et al. | Mar 1996 | A |
5518498 | Lindenberg et al. | May 1996 | A |
5527282 | Segal | Jun 1996 | A |
5540658 | Evans et al. | Jul 1996 | A |
5549626 | Miller et al. | Aug 1996 | A |
5556408 | Farhat | Sep 1996 | A |
5569275 | Kotula et al. | Oct 1996 | A |
5569277 | Evans et al. | Oct 1996 | A |
5571135 | Fraser et al. | Nov 1996 | A |
5591204 | Janzen et al. | Jan 1997 | A |
5606979 | Hodgson | Mar 1997 | A |
5611345 | Hibbeln | Mar 1997 | A |
5626614 | Hart | May 1997 | A |
5632754 | Farley et al. | May 1997 | A |
5643282 | Kieturakis | Jul 1997 | A |
5653684 | Laptewicz et al. | Aug 1997 | A |
5653689 | Buelna et al. | Aug 1997 | A |
5662671 | Barbut et al. | Sep 1997 | A |
5662703 | Yurek et al. | Sep 1997 | A |
5681335 | Serra et al. | Oct 1997 | A |
5683411 | Kavteladze et al. | Nov 1997 | A |
5683451 | Lenker et al. | Nov 1997 | A |
5695501 | Carol et al. | Dec 1997 | A |
5695507 | Auth et al. | Dec 1997 | A |
5709697 | Ratcliff et al. | Jan 1998 | A |
5709704 | Nott et al. | Jan 1998 | A |
5713848 | Dubrul et al. | Feb 1998 | A |
5730733 | Mortier et al. | Mar 1998 | A |
5733294 | Forber et al. | Mar 1998 | A |
5738652 | Boyd et al. | Apr 1998 | A |
5749883 | Halpern | May 1998 | A |
5766135 | Terwilliger | Jun 1998 | A |
5766191 | Trerotola | Jun 1998 | A |
5766203 | Imran et al. | Jun 1998 | A |
5769795 | Terwilliger | Jun 1998 | A |
5769871 | Mers et al. | Jun 1998 | A |
5779672 | Dormandy, Jr. | Jul 1998 | A |
5792157 | Mische et al. | Aug 1998 | A |
5795308 | Russin | Aug 1998 | A |
5795322 | Boudewijn | Aug 1998 | A |
5797960 | Stevens et al. | Aug 1998 | A |
5814064 | Daniel et al. | Sep 1998 | A |
5827324 | Cassell et al. | Oct 1998 | A |
5833650 | Imran | Nov 1998 | A |
5843022 | Willard et al. | Dec 1998 | A |
5846251 | Hart | Dec 1998 | A |
5851210 | Torossian | Dec 1998 | A |
5853422 | Huebsch et al. | Dec 1998 | A |
5868708 | Hart et al. | Feb 1999 | A |
5897567 | Ressemann et al. | Apr 1999 | A |
5904698 | Thomas et al. | May 1999 | A |
5908435 | Samuels | Jun 1999 | A |
5916235 | Guglielmi | Jun 1999 | A |
5928186 | Homsma et al. | Jul 1999 | A |
5928260 | Chin et al. | Jul 1999 | A |
5935139 | Bates | Aug 1999 | A |
5947985 | Imran | Sep 1999 | A |
5947995 | Samuels | Sep 1999 | A |
5954737 | Lee | Sep 1999 | A |
5972019 | Engelson et al. | Oct 1999 | A |
5997503 | Willis et al. | Dec 1999 | A |
6001118 | Daniel et al. | Dec 1999 | A |
6027520 | Tsugita et al. | Feb 2000 | A |
6053876 | Fisher | Apr 2000 | A |
6053900 | Brown et al. | Apr 2000 | A |
6086605 | Barbut et al. | Jul 2000 | A |
6096053 | Bates | Aug 2000 | A |
6156005 | Theron | Dec 2000 | A |
6161034 | Burbank et al. | Dec 2000 | A |
6179860 | Fulton, III et al. | Jan 2001 | B1 |
6206868 | Parodi | Mar 2001 | B1 |
6217600 | Dimatteo | Apr 2001 | B1 |
6221006 | Dubrul et al. | Apr 2001 | B1 |
6221086 | Forber | Apr 2001 | B1 |
6231544 | Tsugita et al. | May 2001 | B1 |
6238412 | Dubrul et al. | May 2001 | B1 |
6258115 | Dubrul | Jul 2001 | B1 |
6277083 | Eggers et al. | Aug 2001 | B1 |
6280414 | Shah et al. | Aug 2001 | B1 |
6287271 | Dubrul et al. | Sep 2001 | B1 |
6356782 | Sirimanne et al. | Mar 2002 | B1 |
6361545 | Macoviak et al. | Mar 2002 | B1 |
6375634 | Carroll | Apr 2002 | B1 |
6413235 | Parodi | Jul 2002 | B1 |
6443971 | Boylan et al. | Sep 2002 | B1 |
6450989 | Dubrul et al. | Sep 2002 | B2 |
6485501 | Green | Nov 2002 | B1 |
6524301 | Wilson et al. | Feb 2003 | B1 |
6530923 | Dubrul et al. | Mar 2003 | B1 |
6540768 | Diaz et al. | Apr 2003 | B1 |
6544278 | Vrba et al. | Apr 2003 | B1 |
6602204 | Dubrul et al. | Aug 2003 | B2 |
6605102 | Mazzocchi et al. | Aug 2003 | B1 |
6626886 | Barbut | Sep 2003 | B1 |
6635068 | Dubrul et al. | Oct 2003 | B1 |
6656202 | Papp et al. | Dec 2003 | B2 |
6660014 | Demarais et al. | Dec 2003 | B2 |
6695858 | Dubrul et al. | Feb 2004 | B1 |
6699260 | Dubrul et al. | Mar 2004 | B2 |
6740094 | Maitland et al. | May 2004 | B2 |
6852097 | Fulton, III | Feb 2005 | B1 |
6929652 | Andrews et al. | Aug 2005 | B1 |
6945977 | Demarais et al. | Sep 2005 | B2 |
6994677 | Buehlmann et al. | Feb 2006 | B1 |
7201770 | Johnson et al. | Apr 2007 | B2 |
7220269 | Ansel et al. | May 2007 | B1 |
7232432 | Fulton et al. | Jun 2007 | B2 |
7374561 | Barbut | May 2008 | B2 |
7422579 | Wahr et al. | Sep 2008 | B2 |
7534251 | Wasdyke | May 2009 | B2 |
7569071 | Haverkost et al. | Aug 2009 | B2 |
7645296 | Theron et al. | Jan 2010 | B2 |
7670368 | Hill et al. | Mar 2010 | B2 |
7686825 | Hauser et al. | Mar 2010 | B2 |
7780722 | Thielen et al. | Aug 2010 | B2 |
7803171 | Uflacker | Sep 2010 | B1 |
7867274 | Hill et al. | Jan 2011 | B2 |
7951189 | Haverkost et al. | May 2011 | B2 |
7959603 | Wahr et al. | Jun 2011 | B2 |
8366737 | Hancock et al. | Feb 2013 | B2 |
8657849 | Parker | Feb 2014 | B2 |
8663273 | Khairkhahan et al. | Mar 2014 | B2 |
8771289 | Mohiuddin et al. | Jul 2014 | B2 |
8777976 | Brady et al. | Jul 2014 | B2 |
8784441 | Rosenbluth et al. | Jul 2014 | B2 |
9186487 | Dubrul et al. | Nov 2015 | B2 |
9211396 | Aboytes | Dec 2015 | B2 |
9498604 | Dubrul et al. | Nov 2016 | B2 |
9561094 | Fulton | Feb 2017 | B2 |
20020007130 | Burbank et al. | Jan 2002 | A1 |
20020016555 | Ritchart et al. | Feb 2002 | A1 |
20020019640 | Mcguckin et al. | Feb 2002 | A1 |
20020022859 | Hogendijk | Feb 2002 | A1 |
20020026201 | Foerster et al. | Feb 2002 | A1 |
20020045916 | Gray et al. | Apr 2002 | A1 |
20020095169 | Maitland et al. | Jul 2002 | A1 |
20020165574 | Ressemann et al. | Nov 2002 | A1 |
20020165598 | Wahr et al. | Nov 2002 | A1 |
20030023204 | Vo et al. | Jan 2003 | A1 |
20030083608 | Evans et al. | May 2003 | A1 |
20030109896 | Dubrul et al. | Jun 2003 | A1 |
20030114879 | Euteneuer et al. | Jun 2003 | A1 |
20030163158 | White | Aug 2003 | A1 |
20030176884 | Berrada et al. | Sep 2003 | A1 |
20040015150 | Zadno-Azizi | Jan 2004 | A1 |
20040044391 | Porter | Mar 2004 | A1 |
20040153117 | Clubb et al. | Aug 2004 | A1 |
20040181237 | Forde et al. | Sep 2004 | A1 |
20040199202 | Dubrul et al. | Oct 2004 | A1 |
20040236369 | Dubrul | Nov 2004 | A1 |
20040260332 | Dubrul et al. | Dec 2004 | A1 |
20040260333 | Dubrul et al. | Dec 2004 | A1 |
20050038447 | Huffmaster | Feb 2005 | A1 |
20050059993 | Ramzipoor et al. | Mar 2005 | A1 |
20050187570 | Nguyen et al. | Aug 2005 | A1 |
20050277976 | Galdonik et al. | Dec 2005 | A1 |
20060047286 | West | Mar 2006 | A1 |
20060058836 | Bose et al. | Mar 2006 | A1 |
20060085065 | Krause et al. | Apr 2006 | A1 |
20060200074 | Zadno-Azizi | Sep 2006 | A1 |
20060253145 | Lucas | Nov 2006 | A1 |
20070126161 | Gray et al. | Jun 2007 | A1 |
20070142858 | Bates | Jun 2007 | A1 |
20070233175 | Zaver et al. | Oct 2007 | A1 |
20080058794 | MacAdam et al. | Mar 2008 | A1 |
20080058800 | Collins et al. | Mar 2008 | A1 |
20080119888 | Huffmaster | May 2008 | A1 |
20100030256 | Dubrul et al. | Feb 2010 | A1 |
20100114113 | Dubrul et al. | May 2010 | A1 |
20100228281 | Gilson et al. | Sep 2010 | A1 |
20110009943 | Paul et al. | Jan 2011 | A1 |
20110270178 | Fiorella et al. | Nov 2011 | A1 |
20110270298 | Abrams | Nov 2011 | A1 |
20110288529 | Fulton et al. | Nov 2011 | A1 |
20110288572 | Martin | Nov 2011 | A1 |
20120083824 | Berrada et al. | Apr 2012 | A1 |
20120316597 | Fitz et al. | Dec 2012 | A1 |
20130110152 | Dubrul et al. | May 2013 | A1 |
20130144326 | Brady et al. | Jun 2013 | A1 |
20130310803 | Morsi | Nov 2013 | A1 |
20130317534 | Zhou et al. | Nov 2013 | A1 |
20130345739 | Brady et al. | Dec 2013 | A1 |
20140005712 | Martin | Jan 2014 | A1 |
20140039598 | Sampognaro et al. | Feb 2014 | A1 |
20140188156 | Tekulve et al. | Jul 2014 | A1 |
20140236219 | Dubrul et al. | Aug 2014 | A1 |
20140343602 | Cox et al. | Nov 2014 | A1 |
20150066075 | Russell et al. | Mar 2015 | A1 |
20150112376 | Molaei et al. | Apr 2015 | A1 |
20150190141 | Cragg et al. | Jul 2015 | A1 |
20150257775 | Gilvarry et al. | Sep 2015 | A1 |
20150343182 | Vazales et al. | Dec 2015 | A1 |
20150351770 | Fulton | Dec 2015 | A1 |
20150351775 | Fulton | Dec 2015 | A1 |
20160022293 | Dubrul et al. | Jan 2016 | A1 |
20160045202 | Ferry et al. | Feb 2016 | A1 |
20160058458 | Hansen et al. | Mar 2016 | A1 |
20160074024 | Scheule | Mar 2016 | A1 |
Number | Date | Country |
---|---|---|
0983749 | Mar 2000 | EP |
1179321 | Feb 2002 | EP |
1549229 | Jul 2005 | EP |
1761298 | Mar 2007 | EP |
1799128 | Jun 2007 | EP |
1981413 | Oct 2008 | EP |
1399089 | Dec 2008 | EP |
1629784 | Jan 2010 | EP |
2057967 | Jan 2013 | EP |
2596828 | May 2013 | EP |
2683309 | Jan 2014 | EP |
2707077 | Mar 2014 | EP |
2744423 | Jun 2014 | EP |
2801325 | Nov 2014 | EP |
2854924 | Apr 2015 | EP |
2879625 | Jun 2015 | EP |
2908901 | Aug 2015 | EP |
2341845 | Jan 2016 | EP |
2979649 | Feb 2016 | EP |
2312264 | Dec 1976 | FR |
2380018 | Sep 1978 | FR |
2020557 | Nov 1979 | GB |
H08308932 | Nov 1996 | JP |
H10328306 | Dec 1998 | JP |
2006519657 | Aug 2006 | JP |
WO-8001343 | Jun 1980 | WO |
WO-8001353 | Jul 1980 | WO |
WO-9424946 | Nov 1994 | WO |
WO-9509024 | Apr 1995 | WO |
WO-9516487 | Jun 1995 | WO |
WO-9601591 | Jan 1996 | WO |
WO-9923952 | May 1999 | WO |
WO-9944506 | Sep 1999 | WO |
WO-9944510 | Sep 1999 | WO |
WO-9944542 | Sep 1999 | WO |
WO-9944542 | Nov 1999 | WO |
WO-0012009 | Mar 2000 | WO |
WO-0012010 | Mar 2000 | WO |
WO-0149208 | Jul 2001 | WO |
WO-0197697 | Dec 2001 | WO |
WO-02055146 | Jul 2002 | WO |
WO-02087677 | Nov 2002 | WO |
WO-03002028 | Jan 2003 | WO |
WO-2004019791 | Mar 2004 | WO |
WO-2004093966 | Nov 2004 | WO |
WO-2005118050 | Dec 2005 | WO |
WO-2006031410 | Mar 2006 | WO |
WO-2007089897 | Aug 2007 | WO |
WO-2008010197 | Jan 2008 | WO |
WO-2008010197 | Apr 2008 | WO |
WO-2008124567 | Oct 2008 | WO |
WO-2010010545 | Jan 2010 | WO |
WO-2012009675 | Jan 2012 | WO |
WO-2012011518 | Jan 2012 | WO |
WO-2012120490 | Sep 2012 | WO |
WO-2012155093 | Nov 2012 | WO |
WO-2013028579 | Feb 2013 | WO |
WO-2013177383 | Nov 2013 | WO |
WO-2014022409 | Feb 2014 | WO |
WO-2014062645 | Apr 2014 | WO |
WO-2014164535 | Oct 2014 | WO |
WO-2014180702 | Nov 2014 | WO |
WO-2015057796 | Apr 2015 | WO |
WO-2015187196 | Dec 2015 | WO |
WO-2016040923 | Mar 2016 | WO |
WO-2016064077 | Apr 2016 | WO |
WO-2017161204 | Sep 2017 | WO |
Entry |
---|
“EP 14894131.3 Extended European Search Report and Search Opinion dated Jan. 17, 2018”. |
Final Office Action for U.S. Appl. No. 12/477,371 Mailed on Jun. 29, 2012. |
Final Office Action for U.S. Appl. No. 12/477,371 Mailed on May 20, 2014. |
Final Office Action for U.S. Appl. No. 13/725,871 Mailed on May 21, 2014. |
International search report and written opinion dated Jul. 27, 2017 for PCT Application No. PCT/US2017/022837. |
International search report and written report dated Apr. 21, 2015 for PCT Application No. PCT/US2014/067696. |
International search report dated Oct. 15, 2004 for PCT Application No. PCT/US2004/011584. |
Notice of allowance dated Apr. 20, 2017 for U.S. Appl. No. 14/645,830. |
Notice of allowance dated Aug. 21, 2015 for U.S. Appl. No. 13/725,871. |
Notice of Allowance dated Sep. 1, 2016 for U.S. Appl. No. 14/201,371. |
Notice of allowance dated Dec. 12, 2000 for U.S. Appl. No. 09/189,574. |
Office action dated Jan. 21, 2016 for U.S. Appl. No. 14/645,830. |
Office action dated Feb. 25, 2015 for U.S. Appl. No. 13/190,416. |
Office action dated Mar. 21, 2017 for U.S. Appl. No. 14/554,348. |
Office Action dated Apr. 6, 2015 for U.S. Appl. No. 12/477,371. |
Office action dated Apr. 6, 2015 for U.S. Appl. No. 13/725,871. |
Office action dated Apr. 6, 2016 for U.S. Appl. No. 14/265,921. |
Office action dated Apr. 9, 2014 for U.S. Appl. No. 13/190,416. |
Office action dated Apr. 12, 2016 for U.S. Appl. No. 14/554,348. |
Office action dated May 4, 2015 for U.S. Appl. No. 14/201,371. |
Office action dated Jun. 6, 2000 for U.S. Appl. No. 09/189,574. |
Office action dated Jun. 28, 2017 for U.S. Appl. No. 14/878,728. |
Office action dated Jul. 17, 2017 for U.S. Appl. No. 14/645,830. |
Office action dated Aug. 5, 2015 for U.S. Appl. No. 14/554,348. |
Office action dated Aug. 5, 2015 for U.S. Appl. No. 14/645,830. |
Office action dated Sep. 19, 2013 for U.S. Appl. No. 13/190,416. |
Office action dated Oct. 8, 2014 for U.S. Appl. No. 13/190,416. |
Office action dated Oct. 14, 1999 for U.S. Appl. No. 09/189,574. |
“Office Action dated Nov. 15, 2017 for U.S. Appl. No. 14/554,348”. |
Office action dated Nov. 20, 2014 for U.S. Appl. No. 14/201,371. |
Office action dated Nov. 28, 2014 for U.S. Appl. No. 13/725,871. |
Office Action for U.S. Appl. No. 10/747,813 dated Jan. 10, 2008. (6 pages). |
Office Action for U.S. Appl. No. 10/747,813 dated Jul. 26, 2007. |
Office Action for U.S. Appl. No. 10/747,813 dated Sep. 19, 2007. |
Office Action for U.S. Appl. No. 10/765,564 dated Oct. 9, 2007. |
Office Action for U.S. Appl. No. 10/866,980 dated Oct. 5, 2007. |
Office Action for U.S. Appl. No. 12/477,371 dated Nov. 8, 2011. |
Office Action for U.S. Appl. No. 12/477,371 dated Sep. 27, 2013. |
Office Action for U.S. Appl. No. 13/725,871 dated Jul. 15, 2014. |
Office Action for U.S. Appl. No. 13/725,871 dated Sep. 5, 2013. |
Schmitz-Rode, et al. New device for percutaneous fragmentation of pulmonary emboli. Radiology. Jul. 1991;180(1):135-7. |
Sharafuddin, et al. Current status of percutaneous mechanical thrombectomy. Part I. General principles. J Vasc Interv Radiol. Nov.-Dec. 1997;8(6):911-21. |
Supplementary European Search Report dated Jul. 23, 2008 for EP Application No. 04759873.5, filed Apr. 15, 2004. (4 pages). |
U.S. Appl. No. 14/554,348, filed Nov. 26, 2014. |
U.S. Appl. No. 14/645,830, filed Mar. 12, 2015. |
“U.S. Appl. No. 14/645,830 Office Action dated May 3, 2018”. |
“U.S. Appl. No. 14/878,728 Office Action dated Apr. 25, 2018”. |
“Velocimed, Praxis, Embolic Protection System”, http://www.velocimed.com/proxis.htm (visited Feb. 5, 2004), (2003). (4 pages). |
Number | Date | Country | |
---|---|---|---|
20200345376 A1 | Nov 2020 | US |
Number | Date | Country | |
---|---|---|---|
62007553 | Jun 2014 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14554348 | Nov 2014 | US |
Child | 16226832 | US |