This invention relates generally to the field of optical coherence tomographic imaging and more specifically to optical coherence techniques for diagnosing and treating vascular stenoses.
Coronary artery disease is one of the leading causes of death worldwide. The ability to better diagnose, monitor, and treat coronary artery diseases can be of life saving importance. Intravascular optical coherence tomography (OCT) is a catheter-based imaging modality that employs safe, non-ionizing near-infrared light to peer into coronary artery walls and present images valuable for the study of the vascular wall architecture. Utilizing broad-band coherent light, interferometry, and micro-optics, OCT can provide video-rate in-vivo tomography within a diseased vessel with resolution down to the micrometer level. This level of detail enables OCT to diagnose as well as monitor the progression of coronary artery disease.
The quantitative assessment of vascular pathology and its progression involves the calculation of different quantitative measures such as the vessel cross-sectional area, mean diameter, and blood flow resistance, all of which rely on the accurate identification of the luminal border. While the luminal border in OCT images is clearly identifiable by the human eye, it is tedious, expensive, and time consuming to manually trace the luminal border. Thus there is a need for a reliable technique that can automatically identify the luminal border.
OCT produces images that are higher in resolution and contrast compared to those of intravascular ultrasound (IVUS). As opposed to IVUS which images through blood, OCT images are typically acquired with blood cleared from the view of the optical probe. This is one reason the luminal border in OCT images is sharper and more defined compared to that in IVUS images.
Cross-sectional diameter and area measurements provide interventional cardiologists with useful guidance for stent sizing and placement. However, the relationship of these geometric measurements to clinically relevant variables, such as ability of the artery to supply an adequate flow of blood to the myocardium when metabolic demands are high, is not well understood. In early studies, the percent stenosis of an individual coronary lesion measured by angiography was found to be a relatively poor predictor of the physiological significance of the lesion. In contrast, several later studies demonstrated that lumen measurements made by quantitative coronary angiography (QCA) and IVUS correlate closely with physiologic measurements of coronary obstruction, including coronary flow reserve (CFR) and fractional flow reserve (FFR). For example, several studies found a high correlation between area stenosis, measured by QCA, and CFR measured by a Doppler flow wire. It appears that the standard angiographic (and IVUS) measures of lesion severity—the minimum lumen area (MLA), percentage stenosis, and lesion length—do indeed convey physiologically relevant information. However, the value of any single geometrical measure as an independent predictor of the physiological significance of a lesion in a wide patient population is not clear.
Several factors contribute to the limitation of standard angiographic and IVUS lumen measurements for assessment of the physiological significance of coronary stenoses. First, the accuracy and reproducibility with which cross-sectional areas can be measured with angiography, which generally has a spatial resolution of 0.2-0.4 mm, are relatively low. The angle of the X-ray projection, in addition to the shadowing effect of lesions with irregular contours, can increase errors significantly beyond the theoretical minimums. Even state-of-the-art IVUS imaging systems, which have resolutions of approximately 0.15 mm in the axial dimension and 0.3 mm in the angular dimension, cannot accurately measure the cross-sectional areas of small eccentric lesions or lesions with irregular borders.
Intravascular OCT imaging, applied in combination with new clinical parameters based on advanced analysis of lesion morphology, has the potential to overcome many of the limitations of conventional measures of lesion severity based on angiography and IVUS. The high resolution of OCT enables accurate measurement of the shape and dimensions of the vessel lumen over the length of the lesion and its adjacent reference segments. Furthermore, advanced models of flow dynamics enable the physiological significance of lesions to be estimated under both normal and hyperemic conditions. It should be realized, however, that the clinical value of quantitative lesion morphology measurements—even when accurate—may be limited by physiological conditions in certain patients. Finally, high-frequency OCT imaging has the advantage that it can precisely delineate three-dimensional contours of long segments of coronary arteries in a few seconds to assist cardiologists in their real-time diagnosis and treatment during PCI procedures.
In spite of advances in intravascular imaging, cardiologists frequently do not take full advantage of the capabilities of OCT and IVUS for planning and evaluating stent deployment, because the measurements currently derived from the images provide insufficient information to predict the effectiveness of treatment. For example, many cardiologists rely on minimum lumen area (MLA) as a key variable in their stenting decisions. If MLA measurements are sufficiently low, the cardiologist may decide to implant a stent. Based on the diameters and locations of reference vessel segments, the cardiologist must then choose the proper position, length, and diameter of the stent. The wrong choice of the size or location of the stent may lead to the failure to restore blood flow and may even cause potentially serious clinical complications, such as stent migration, stent thrombosis, or dissection of the vessel wall. There is a need for new methods for optimization of stent sizing and positioning based on measurements derived from intravascular images. To achieve maximum clinical benefit, these new methods should enable cardiologists to predict the potential physiological consequences of implanting stents of different diameters and lengths in different locations.
The present invention addresses these needs.
In one aspect, the invention relates to an automated computer-based method of evaluating a region of a lumen. The method comprises the steps of collecting a set of data regarding a vessel segment of length L using an optical coherence tomography system, the set comprising a plurality of cross sectional areas at a plurality of positions along the length; determining a vascular resistance ratio (VRR) using a processor and at least a portion of the set of data; and determining a characteristic of at least a portion of the region disposed along the length L in relation to the vascular resistance ratio.
In one embodiment, the method is applied to the region that contains a stenotic lesion. In another embodiment, the method further comprises the step of displaying at least one numerical or graphical measure of stent length used to treat the stenotic lesion. In yet another embodiment, the step of determining the vascular resistance ratio is performed using a lumped resistor model.
In another aspect, the invention relates to a method for automatically identifying the luminal border in an in-situ OCT vascular image. In one embodiment, the method comprises the steps of generating a mask of the OCT lumen image using a computer; defining a plurality of scan lines in the mask; identifying a region as tissue on each scan line; defining contour segments in response to the plurality of scan lines and the region of tissue on each scan line; identifying valid neighboring contour segments; interpolating missing contour data between valid neighboring contour segments; assessing the likely correctness of the computed contour and indicating to the user on which image frames the computed contour may require manual adjustment.
In one embodiment, the method includes the step of detecting and removing guide wire and similar artifacts. In another embodiment, the identification of a tissue region includes the steps of finding a plurality of start/stop pairs on each scan line; calculating thickness and gap of each said start/stop pair; calculating a weight based on said thickness and said gap; and defining the tissue region based on the largest weight of tissue and gap. In another embodiment, the step of defining connected contour includes finding the scan line with the largest weight; searching for discontinuities in both directions from the scan line to define a valid segment; and identifying the root of the contour as the longest of the valid segments. In still yet another embodiment, the step of identifying valid neighboring contour includes finding the nearest clockwise and counter-clockwise neighbors of each of the contour segments that pass angular, radial, and Euclidean distance thresholds.
In another embodiment, the step of detection and removal of guide wire shadow artifact comprises the steps of clearing an image binary mask by fitting an ellipse to the foreground data of the mask and blanking the area inside ellipse; building an intensity profile using the cleared mask; identifying the guide wire shadow region in the intensity profile; detecting a guide wire offset within the shadow region; collecting the midpoint of detected guide wires on all frames; building a minimum spanning tree using the collected midpoints; and pruning the resulting minimum spanning tree to remove outliers resulting from non-guide wire shadows. In another embodiment, the step of interpolating missing data includes the steps of identifying required interpolation control points with valid contour data on both ends of the missing contour segment; and using the control points to interpolate the missing contour segment. In still yet another embodiment, the steps are performed on all missing contour segments that need to be interpolated. In yet another embodiment, the step of searching for discontinuities comprises the steps of calculating a scan line-to-scan line offset change histogram; smoothing said histogram; identifying the smallest change with zero count from the histogram; and using the smallest change as a continuity measure.
In another embodiment, the step of evaluating the correctness of the computed contour comprises the steps of computing an “Error Measure” by fitting an ellipse to the computed contour; computing the root mean square error between the computed contour and the fitted ellipse; normalizing the root mean square error to the average diameter of the ellipse; and multiplying the normalized root mean square error by the ratio of the number of scan lines where the lumen was successfully detected to the total number of scan lines in the image frame. In another embodiment, the resulting Error Measure parameter is compared to a threshold and, for image frames where the threshold is exceeded, the user is notified that manual contour correction may be required. In yet another embodiment, the notification can take the form of “alert frames” drawn on a longitudinal display of the images of the pullback region.
In another aspect, the invention relates to an automated method for quantifying a vascular resistance including the steps of selecting proximal and distal frames of an OCT image; calculating actual vascular resistance of the vascular segment enclosed by said proximal and the distal frames; calculating a total vascular resistance of the vascular segment; and calculating vascular resistance ratio using the actual vascular resistance and said total vascular resistance. In one embodiment, the step of calculating actual vascular resistance comprises the steps of extracting luminal contours of all frames enclosed by the proximal and the distal frames inclusive; calculating cross-sectional areas from the extracted contours; constructing a smooth area graph; and using the smooth area graph in the actual vascular resistance calculation. In another embodiment, the step of calculating the total vascular resistance comprises the steps of: fitting a shape between said proximal and said distal frames; and calculating cross-sectional areas of the shape at all frame positions enclosed by the proximal and the distal frames inclusive. In yet another embodiment, the step of constructing a smooth area graph includes the steps of constructing a graph using the cross-sectional areas; interpolating missing area values on the graph; and smoothing the resulting graph. In still yet another embodiment, vascular resistance is calculated by computational fluid dynamics from the detected three-dimensional luminal border between the proximal and distal planes.
Another aspect of the invention is a method of placing a stent including the steps of: (a) measuring the parameters in the region of interest in an OCT image of a vessel; (b) simulating the placement of the stent in the region of interest; (c) recalculating the parameters in the region of interest; and repeating Steps b and c until the desired result is obtained.
The invention is pointed out with particularity in the appended claims. The advantages of the invention described above, together with further advantages, may be better understood by referring to the following description taken in conjunction with the accompanying drawings. In the drawings, like reference characters generally refer to the same parts throughout the different views. The drawings are not necessarily to scale, emphasis instead generally being placed upon illustrating the principles of the invention.
In one embodiment a computer or processor is part of the OCT system 36 or is included as a separate subsystem 40 in electrical communication with the OCT system 36. The computer or processor 40 includes memory, storage, buses and other components suitable for processing data. for lumen detection and pull back data collection as discussed below. In one embodiment, the computer or processor includes software implementations or programs 41 of the methods described herein that are stored in memory and execute using a processor. A display 42 is part of the overall system 10 for showing cross-sectional scan data as longitudinal scans or in other suitable formats.
In brief overview, the present invention provides a method and apparatus of automatically locating a lumen boundary at a position in a vessel of interest (using an OCT image or the underlying data) and from that measuring the diameter of the vessel. From the diameter of the vessel and calculated blood flow rate a number of clinically significant physiological parameters are then determined and various images of interest generated. One use of these images and parameters is to aid the clinician in the placement of a stent. The system, in one embodiment, uses these measurements to allow the clinician to simulate the placement of a stent and determine the effect of the placement. In addition, from these patient parameters various patient treatments are then performed.
As a first step, the system determines the lumen boundary. Generally, data taken by an OCT system is used with the methods described herein to recognize and avoid residual blood, guide wire reflections, and other structures that may appear to be part of the vessel wall. Interpolation of a continuous boundary is accomplished by imposing continuity of the inner surface of the vessel across neighboring frames.
The mean diameter of each cross-section is calculated either as the diameter of a circle with an area equal to that of the cross section or as the mean of the chord lengths at all angles drawn through the centroid of the lumen cross-section. In one embodiment, the minimum lumen area (MLA), proximal and distal reference areas, percent diameter stenosis, and the length between references are displayed numerically in the same panel.
In one embodiment, the system then also generates a three-dimensional rendering of the shape of the vessel lumen as calculated from the cross-sectional measurements. An example is shown in
Referring to
In more detail and referring to
A cross-sectional image of the vessel is created for each complete rotation of the optical probe. These images are individually preprocessed and a suitable threshold is applied to create a binary foreground/background image mask, wherein the foreground is defined to contain the potentially relevant image information (e.g. the vessel wall) and the background represents the empty luminal space between the catheter and vessel wall, as well the ‘noise floor’ beyond the deepest imaging depth within the wall. The image mask is further processed by convolving the image mask with a median filter that has a suitable width W and a suitable height H. This operation fills in the gaps and removes the noise as each of the image mask values is replaced by the median value in its W×H neighborhood window. An example of a resulting mask is shown in
In still more detail, in one embodiment of the invention, each scan line of the mask is processed to find all pairs of start and stop samples as shown in
In any one scan line it is possible to have more than one region identified as tissue, as shown in
Weight=(gap*thickness2) (1)
so as to favor the thickest isolated region, as blood artifacts are thinner than the imaged vessel wall. It should be appreciated that this invention is not limited to this particular weight calculation.
At this point in the procedure, every scan line in a given cross-sectional image should have, at most, one sample that will be on the lumen contour. The calculated weight associated with the sample on any given scan line is kept for further utilization. Some scan lines such as those in a side branch of a vessel might not have detected samples corresponding to a contour.
A contour segment can be defined, in one embodiment, as a group of contiguous scan lines with no discontinuities. A discontinuity is a scan line-to-scan line change in the sample number (offset) that exceeds a predetermined continuity threshold. To identify all possible contour segments, the method begins by searching for the line with the largest weight among the lines not yet grouped in segments (initially, these are all scan lines in a given cross-sectional image). A segment is identified by searching for discontinuities clockwise and counter-clockwise from the line with the largest weight as illustrated in
The luminal contour is a possible grouping of one or more contour segments. The root (first segment to add to the contour) of the contour is selected as the longest valid segment. The nearest clockwise and counter-clockwise neighboring segments of each potential contour segment are identified. Valid neighbors must pass an angular distance threshold, a radial distance threshold, and a Euclidian (direct connection) distance threshold. Each potential contour is then traversed clockwise and counter-clockwise and the longest contour is selected.
To detect and remove the guide wire and other similar artifacts from the image, an ellipse is fitted to the foreground of a median mask (shown in
Missing contour data is interpolated as shown in
Alternatively, one can use any suitable function such as the cubic function or the Hermite function to interpolate missing data using four or more control points instead of two. Interpolating a point between two points y1=f(x1) and y2=f(x1+Δx), calculates the value of the point on a preselected curve between x1 and x2. The general relation is given by (1−α)y1+(α)y2, where α is the interpolation weight ranging from 0 at x1 to 1 at x1+Δx. Using the previously described cosine weighting method, the weight of a point at given distance d from x1 is calculated by α=(1−cos(π*d/Δx))/2. It should be noted that this invention is not limited to any one particular interpolation method.
For an entire longitudinal segment of interest for a vessel, an area graph vs. longitudinal position is constructed from the individually calculated cross-sectional areas as shown in
Once the cross-sectional area of the vessel has been determined the severity of any stenotic region is then characterized. One measure of severity of a stenotic lesion imaged by OCT is provided by a parameter called the vascular resistance ratio (VRR). The VRR quantifies the blood flow resistance of a stenotic vessel segment relative to the flow resistance of the entire vessel branch, assuming maximum vasodilation of the peripheral coronary vasculature. The VRR is defined as:
where Rs is the blood flow resistance of the stenotic segment and RT is the total flow resistance of the branch vessel in which the stenotic region is located. VRR ranges from 0 (no vessel narrowing) to 1 (all flow resistance due to the stenosis).
The calculation of VRR is based on a lumped parameter model (
R
T
=R
s
+R
e
+R
mv (3)
where Rs is the blood flow resistance of the stenotic segment, Re is the blood flow resistance of the remaining epicardial length of the branch, and Rmv is the microvascular resistance of the peripheral coronary vascular bed.
In general, the values of all three resistance elements depend on blood flow, but only Rs is shown explicitly as a function of Q, because Rmv and Re are only weakly flow-dependent under conditions of maximum vasodilation. During drug-induced hyperemia, Rmv is approximately constant and is given by:
where Qmax is the maximum blood flow that can be achieved in the branch when the pressure drop across the epicardial arteries is negligible (i.e., Rs+Re→0). Qmax equals the product of the mean hyperemic Doppler blood velocity, vmax, measured in a normal reference segment of the artery and the cross-sectional area, An, of the artery measured in the same location, Qmax=vmaxAn. Velocity may also be measured using speckle caused by particulates in the stream and detected in the OCT image. Based on these relationships, Eqn. 4 can be re-formulated in terms of hyperemic velocity:
The quantity in braces, which has units of mm Hg cm−1s, is the hyperemic microvascular resistance index, designated as h-MRv. An important advantage of determining hyperemic resistance using velocity instead of flow is that velocity normalizes flow for differences in arterial diameter due to branching and is preserved between proximal and distal segments. Table 1 lists published values of h-MRv measured during PCI with a Doppler flow wire. The values lie within a relatively narrow range for both treated and untreated vessels.
In the calculations shown it is assumed that h-MRv is a constant approximately equal to 1.0 mm Hg cm−1s, a value that lies at the lower end of the distribution of resistances in Table 1 for upsized stented arteries. The value of An in Eq. 5 is assumed to equal the cross-sectional area of the proximal segment of the reference vessel. For a 3-mm diameter artery, Eqn. 5 yields Rmv=17 mm Hg cm−3s with a corresponding maximum flow of about 4.7 ml/s at an arterio-venous pressure difference of 80 mmHg.
The second component of the total resistance in Eqn. 3, Re, the epicardial resistance outside of the stenotic segment of the vessel, is usually small compared to Rs and Rmv. Its value can be estimated by integrating the flow resistance along the length of the vessel, Le=LT−Ls, where LT is the total length of the coronary branch and Ls is the length of the stenotic segment imaged by OCT. Assuming that no significant flow-limiting stenoses are present outside of the stenotic segment and that the mean cross-sectional area of the vessel remains the same as in the mean cross-sectional area, Ān, of the reference segments adjacent to the stenosis imaged by OCT, Re can be calculated using Poiseuille's law, as
where η is the viscosity of the blood and the mean area is given by
In this equation, the cross-sectional lumen areas Ai are measured in the frames of the OCT image located outside of the stenotic region, so that the total number of available frames M depends on the lengths of the proximal and distal reference segments in the image. Although the total lengths of the epicardial coronary branches are not, in general, the same, it is assumed that LT=8 cm for the main coronary arteries (LAD, LCX, and RCA), so that Le can be found directly by subtracting the length of the OCT image region from LT. A better estimate of the epicardial length can be obtained from lengths measured by angiography, if such data is available. The mean area is estimated as the average of the diameters of the proximal and distal reference segments.
Calculation of the stenotic resistance, Rs, in Eqn. 3 is complicated by its dependence on blood flow. Rs is composed of a flow-independent component that results from viscous losses and a flow-dependent component that results from kinetic losses. A variety of methods have been developed for calculation of the flow resistance of stenotic lesions. Three different embodiments of methods (one analytical and two numerical) by which Rs can be calculated based on measurements of lumen morphology by OCT will now be discussed.
The first embodiment of a method for calculation of Rs is adapted from a model of pressure loss in stenotic lesions developed by Kirkeeide.
R
s
=R
p
+R
v
+k
e
Q (9)
Here Rp represents losses due to viscous wall friction, calculated according to Poiseuille's law as:
This resistance equals the total integrated viscous losses along the vessel minus the losses in the exit regions where flow separation occurs. Exit regions are defined as the segments of the artery within which the exit angle (θ in
The second flow-independent component of Rs in Eq. 9, which represents the additional viscous losses that occur at the entrance of regions of sudden narrowing of the vessel wall, is given by:
where dp is the diameter of the artery on the proximal side of the stenosis, Am is the minimum lumen area of the stenosis, C2=0.45.
The flow-dependent component of Rs in Eq. 9 includes losses due to flow separation and recirculation at the exit of narrowed regions of the artery. At high flow rates and in vessels with highly irregular cross sections, the effective resistance of a blood vessel can significantly exceed that predicted by Poiseuille's law, which is based on analysis of laminar flow of a Newtonian fluid through a straight cylinder. According to Kirkeeide:
where ρ is the mass density of the blood, Ad is the area of the artery distal to the stenosis, and
Here s is the length of the stenosis, defined as the region between the wall angle inflection points on either side of the stenosis (
Now that all of the terms in Eq. 9 have been defined, the vascular resistance ratio can be expressed as
The other embodiments of the method, instead of Kirkeedee's equations, use a numerical Navier-Stokes solver such as FloWorks, (SolidWorks Corporation, Concord, MA) or Fluent (Ansys, Ann Arbor, MI) or equivalent to calculate the stenotic resistance Rs in the model in
In the second embodiment of the method, the cylindrically symmetrical computational flow model, the same area-versus-position graphs are used as in the first embodiment. The Navier-Stokes equations are solved assuming the shape is a perfect circle at each location along the OCT image. In the third embodiment, the full-3D computational flow model based on the actual OCT lumen contours is used. The wall geometry is broken into triangles spanning every other frame and every 15° around the catheter.
Studies of anatomy show that the sum of the cross sectional area of branches derived from a parent is greater than the cross sectional area of the parent. This minimizes viscous shear stress through the epicardial tree. Based on viscous losses, Murray's law states that the cube of the radius of a parent vessel equals the sum of the cubes of the radii of the daughters. Table 2 shows the area increase calculated by Murray's law when the branches are symmetric.
The steps to obtain the branch sizes are:
Rmv for the entire artery is adjusted by assuming the 1.0 mm Hg cm−1s value of h-MRv was determined based on a reference diameter of 3.4 mm. For other proximal reference diameters Rmv is adjusted downwards according to the ratio of the proximal reference diameter to the reference diameter to the ¼ power. The ¼ power equates pressure drops through the vasculature. Data on the variation of velocity data through the coronaries is limited, but the ¼ power rule seems to correlate the published data as shown in Table 3. A more sophisticated approach would adjust Rmv according to the vessel type (LAD: left anterior descending artery, RCA: right coronary artery, LCx: left circumflex).
SRI=Pressure Difference Between Measurement Points(dP)/Proximal Velocity
In one embodiment, SRI is calculated by assuming a velocity. Velocity is fairly constant in human arteries. In one study of 32 patients after percutaneous coronary intervention PCI, the measured velocity was 79±17.2 cm/s. Since the velocity variation is small and the SRI curve is fairly independent of velocity, the estimate of SRI made without velocity measurements can be acceptable.
Velocity is a better way to normalize SRI than flow because pressure drop is mostly proportional to velocity. If flow is used, it typically underestimates the effect of a stenosis in a large vessel and conversely overestimates the effect of a stenosis in a small vessel. The velocity that is selected is the velocity at a reference diameter, not the stenosis velocity. The physician selects the proximal reference and the velocity measurement is taken there. The resulting SRI will give the physician the resistance that will be eliminated by the stent.
The flow through the region of interest will change if a side branch is detected. The flow down the side branch will be estimated from the side branch size and the reduction in area from the proximal to distal reference. Both the algebraic equations and the Navier-Stokes Equation are modified to include the side branches.
If SRI is reported, a different SRI will be used than that of Meuwissen et al. This index, termed the LightLab SRI (LSRI) is defined as:
LSRI=Total Pressure Difference/Velocity−Integrated Poiseuille Equation
where: total pressure difference is the static pressure at a first location plus the velocity head (ρV12/2) at the first location minus the static pressure at a second location plus the velocity head (ρV22/2) at the second location. The locations typically straddle the region of interest in the lumen. Velocity, V, is the bulk average velocity. The integrated Poiseuille equation is the laminar flow pressure drop calculated between the reference locations assuming the diameter increases linearly. This is an improvement over the standard SRI measurement because the total pressure is more reflective of the true losses in an artery than the static pressure used in standard SRI and the integrated Poiseuille equation removes the effects of the distance between measurement locations, which is a limitation of standard SRI.
Another parameter that is measurable by this technique is the fractional flow reserve (FFR). As defined by the model in
Clinically an FFR value greater than or equal to 0.75 typically is considered to mean that treatment is not required. Generally the FFR is measured following the administration of drugs that cause a maximum hyperemic response by causing the capillary beds to dilate followed by the taking of an intravenous pressure measurement.
Assuming that there are no additional stenoses proximal to the stenotic segment, VRR is inversely proportional to FFR:
A VRR of less than 0.25 means that treatment is not indicated. A benefit of VRR is that, as shown below, a VRR calculation may be made without the use of drugs or the measurement of intravascular pressure.
Once the segmental resistances on which the VRR is based have been calculated, additional information can be displayed to help the clinician select the length of a stent required to cover a stenotic lesion. One concept for displaying this information is illustrated in
is satisfied. Here N(0−ΔL) and N(0+ΔL) are the frame numbers at the distal and proximal limits of the vessel segment. Alternatively, the high-resistance regions can be identified independently of the location of the MLA cross section by sorting the resistances of the incremental segments from highest to lowest and highlighting only those segments at the top of the list that sum to a user-selectable fraction of the total vascular resistance. The advantage of this method is that more than one region of high resistance in a diffusely narrowed artery can be identified readily, as shown by the example in
Once the parameters of vessel size and blood flow resistance are calculated, the present invention also provides methods for optimizing stent choice and placement automatically or semi-automatically via interactive commands. These flow calculations, when combined with a set of a priori constraints, enable a cardiologist to optimize the length, diameter, and longitudinal position of a stent before implantation.
Referring again to
where A(x) is the area of the cross section. Alternatively, the mean diameter can be found by averaging the lengths of chords drawn through the centroid of the lumen cross section.
For interactive stent optimization, the mean-diameter display shows the position of a reconfigurable stent superimposed on the vessel profile, as illustrated in
The various embodiments of the present invention provide methods for choosing the optimal stent length, diameter, and longitudinal position in accordance with the aforementioned optimization parameters (VRR, malapposition distance, branch overlap, presence of calcium, etc.). The flow chart in
In addition to reporting the recommended diameter and position of the stent to the user, this specific embodiment of the optimization procedure also reports the predicted values of VRRopt, the vascular resistance ratio, and εr, the residual malapposition distance. If the user deems these values to be unsatisfactory, the optimization can be repeated with a longer stent length as an input. In this way, errors in the sizing and positioning of stents can be avoided before implantation.
In more detail, the system first creates arrays of area and diameter for each cross-section along the unstented vessel. Next, the system creates a lookup table that has the available ranges of stent diameter, length and position. Then, progressing through each entry in the lookup table, the system calculates the VRR and maximum malapposition value. The maximum malapposition value equals the distance between the maximum unstented diameter in the segment and the diameter of the stent. Table entries that result in VRR values less than VRRmax and the maximum malapposition values are retained and then the stent length for each subset is determined. The table entry in which the stent length is a minimum defines the optimal stent parameters.
To be useful as an interactive bed side tool, the recalculation of VRR for a selected stent size needs to be almost instantaneous. The most accurate method to find the chosen stent effect of VRR would be to first measure or calculate VRR on the unstented artery using the OCT measurements above or a finite element computational fluid dynamics program and then recalculate VRR using the same finite element computational fluid dynamics program on the proposed stented artery shape. However, most computational fluid dynamics programs will not run fast enough on typical computers to quickly show the affect of the proposed stent. A method is needed to have the accuracy of computational fluid dynamics but allow the rapid recalculation of VRR with the proposed placement of a stent.
A hybrid approach is disclosed here that allows for rapid recalculation. In the region of the proposed stent, algebraic equations are used to determine pressure drop. In the regions outside of the stent, the previously obtained measured or computational fluid dynamics solution is used, modified by the effect of the stent. The rapid recalculation is obtained by only using algebraic equations during the stent sizing. Once the stent sizing is complete, a full computational fluid dynamics simulation may be run to obtain an even more accurate answer.
The initial calculation of VRR on the unstented artery is done using a finite element computational fluid dynamics program. Since there is some time between the end of the imaging procedure and the start of the stent placement, the amount of time this calculation takes is not a limiting constraint. An important output of the computational fluid dynamics program is a total pressure versus distance graph as shown in the
A more sophisticated approach takes into account that the pressure drop outside of the proposed stented area will increase because the flow increases with the elimination of the stenosis.
The resistor network in
Target values of VRR can be established according to results of published clinical studies. For example, the results of one influential study showed that adverse event rates in patients with a single stented lesion were reduced significantly when the fractional flow reserve (FFR) measured in the stented artery was in the range 0.96-1.0 compared to the adverse event rates of a similar population of patients with measured FFR values in the range 0.9-0.95. Therefore, FFRmin=0.96 is a post-stent target supported by clinical evidence. According to its definition, VRR has a simple inverse relationship with fractional flow reserve (VRR=1−FFR); it follows that, based on this study, an appropriate target maximum value of is VRRmax=1−0.96=0.04.
The present invention may be embodied in many different forms, including, but in no way limited to, computer program logic for use with a processor (e.g., a microprocessor, microcontroller, digital signal processor, or general purpose computer), programmable logic for use with a programmable logic device, (e.g., a Field Programmable Gate Array (FPGA) or other PLD), discrete components, integrated circuitry (e.g., an Application Specific Integrated Circuit (ASIC)), or any other means including any combination thereof. In a typical embodiment of the present invention, some or all of the processing of the data collected using an OCT probe and the processor-based system is implemented as a set of computer program instructions that is converted into a computer executable form, stored as such in a computer readable medium, and executed by a microprocessor under the control of an operating system. Thus, query response and input data are transformed into processor understandable instructions suitable for generating OCT data, histology images, OCT images, vascular resistance, overlays masks, signal processing, weighting artifact removal, contour detection and other features and embodiments described above.
Computer program logic implementing all or part of the functionality previously described herein may be embodied in various forms, including, but in no way limited to, a source code form, a computer executable form, and various intermediate forms (e.g., forms generated by an assembler, compiler, linker, or locator). Source code may include a series of computer program instructions implemented in any of various programming languages (e.g., an object code, an assembly language, or a high-level language such as Fortran, C, C++, JAVA, or HTML) for use with various operating systems or operating environments. The source code may define and use various data structures and communication messages. The source code may be in a computer executable form (e.g., via an interpreter), or the source code may be converted (e.g., via a translator, assembler, or compiler) into a computer executable form.
The computer program may be fixed in any form (e.g., source code form, computer executable form, or an intermediate form) either permanently or transitorily in a tangible storage medium, such as a semiconductor memory device (e.g., a RAM, ROM, PROM, EEPROM, or Flash-Programmable RAM), a magnetic memory device (e.g., a diskette or fixed disk), an optical memory device (e.g., a CD-ROM), a PC card (e.g., PCMCIA card), or other memory device. The computer program may be fixed in any form in a signal that is transmittable to a computer using any of various communication technologies, including, but in no way limited to, analog technologies, digital technologies, optical technologies, wireless technologies (e.g., Bluetooth), networking technologies, and internetworking technologies. The computer program may be distributed in any form as a removable storage medium with accompanying printed or electronic documentation (e.g., shrink-wrapped software), preloaded with a computer system (e.g., on system ROM or fixed disk), or distributed from a server or electronic bulletin board over the communication system (e.g., the Internet or World Wide Web).
Hardware logic (including programmable logic for use with a programmable logic device) implementing all or part of the functionality previously described herein may be designed using traditional manual methods, or may be designed, captured, simulated, or documented electronically using various tools, such as Computer Aided Design (CAD), a hardware description language (e.g., VHDL or AHDL), or a PLD programming language (e.g., PALASM, ABEL, or CUPL).
Programmable logic may be fixed either permanently or transitorily in a tangible storage medium, such as a semiconductor memory device (e.g., a RAM, ROM, PROM, EEPROM, or Flash-Programmable RAM), a magnetic memory device (e.g., a diskette or fixed disk), an optical memory device (e.g., a CD-ROM), or other memory device. The programmable logic may be fixed in a signal that is transmittable to a computer using any of various communication technologies, including, but in no way limited to, analog technologies, digital technologies, optical technologies, wireless technologies (e.g., Bluetooth), networking technologies, and internetworking technologies. The programmable logic may be distributed as a removable storage medium with accompanying printed or electronic documentation (e.g., shrink-wrapped software), preloaded with a computer system (e.g., on system ROM or fixed disk), or distributed from a server or electronic bulletin board over the communication system (e.g., the Internet or World Wide Web).
Various examples of suitable processing modules are discussed below in more detail. As used herein a module refers to software, hardware, or firmware suitable for performing a specific data processing or data transmission task. Typically, in a preferred embodiment a module refers to a software routine, program, or other memory resident application suitable for receiving, transforming, routing and processing instructions, or various types of data such as OCT scan data, interferometer signal data, clock signals, region of interest types, formulas, and other information of interest.
Computers and computer systems described herein may include operatively associated computer-readable media such as memory for storing software applications used in obtaining, processing, storing and/or communicating data. It can be appreciated that such memory can be internal, external, remote or local with respect to its operatively associated computer or computer system.
Memory may also include any means for storing software or other instructions including, for example and without limitation, a hard disk, an optical disk, floppy disk, DVD (digital versatile disc), CD (compact disc), memory stick, flash memory, ROM (read only memory), RAM (random access memory), DRAM (dynamic random access memory), PROM (programmable ROM), EEPROM (extended erasable PROM), and/or other like computer-readable media.
In general, computer-readable memory media applied in association with embodiments of the invention described herein may include any memory medium capable of storing instructions executed by a programmable apparatus. Where applicable, method steps described herein may be embodied or executed as instructions stored on a computer-readable memory medium or memory media. These instructions may be software embodied in various programming languages such as C++, C, Java, and/or a variety of other kinds of software programming languages that may be applied to create instructions in accordance with embodiments of the invention.
While the present invention has been described in terms of certain exemplary preferred embodiments, it will be readily understood and appreciated by one of ordinary skill in the art that it is not so limited, and that many additions, deletions and modifications to the preferred embodiments may be made within the scope of the invention as hereinafter claimed. Accordingly, the scope of the invention is limited only by the scope of the appended claims.
This application is a continuation of U.S. patent application Ser. No. 17/980,323, filed on Nov. 3, 2022, which is a continuation of U.S. patent application Ser. No. 16/549,814, filed on Aug. 23, 2019, which is continuation of U.S. patent application Ser. No. 13/796,710 filed on Mar. 12, 2013, which is a continuation of U.S. application Ser. No. 12/888,347, filed on Sep. 22, 2010 (now U.S. Pat. No. 9,138,147), which claims priority to and the benefit of U.S. Provisional Application No. 61/244,992, filed on Sep. 23, 2009, and U.S. Provisional Application No. 61/334,834, filed on May 14, 2010, the disclosures of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
61334834 | May 2010 | US | |
61244992 | Sep 2009 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17980323 | Nov 2022 | US |
Child | 18524930 | US | |
Parent | 16549814 | Aug 2019 | US |
Child | 17980323 | US | |
Parent | 13796710 | Mar 2013 | US |
Child | 16549814 | US | |
Parent | 12888347 | Sep 2010 | US |
Child | 13796710 | US |