N/A
Heterogeneous immunoassays typically require the separation of sought-for components bound to component-selective particles from unbound or free components of the assay. To increase the efficiency of this separation, many assays wash the solid phase (the bound component) of the assay after the initial separation (the removal or aspiration of the liquid phase). Some chemiluminescent immunoassays use magnetic separation to remove the unbound assay components from the reaction vessel prior to addition of a reagent used in producing chemiluminescence or the detectable signal indicative of the amount of bound component present. This is accomplished by using magnetizable particles including, but not restricted to, paramagnetic particles, superparamagnetic particles, ferromagnetic particles and ferrimagnetic particles. Tested-for assay components are bound to component-specific sites on magnetizable particles during the course of the assay. The associated magnetizable particles are attracted to magnets for retention in the reaction vessel while the liquid phase, containing unbound components, is aspirated from the reaction vessel.
Washing of the solid phase after the initial separation is accomplished by dispensing and then aspirating a wash solution, such as de-ionized water or a wash buffer, while the magnetizable particles are attracted to the magnet.
Greater efficiency in washing may be accomplished by moving the reaction vessels along a magnet array having a gap in the array structure proximate a wash position, allowing the magnetizable particles to be resuspended during the dispense of the wash solution. This is known as resuspension wash. Subsequent positions in the array include additional magnets, allowing the magnetizable particles to recollect on the side of the respective vessel.
Once the contents of the reaction vessel have again accumulated in a pellet on the side of the reaction vessel and the wash liquid has been aspirated, it is desirable to resuspend the particles in an acid reagent used to condition the bound component reagent. In the prior art, a single stream of acidic reagent is injected at the pellet. Because the size of the pellet and limitations on the volume and flow rate of reagent, insufficient resuspension may result. To address this inadequacy, prior art systems have resorted to the use of an additional resuspension magnet disposed on an opposite side of the process path from the previous separation magnets. The resuspension magnet is configured to assist in drawing paramagnetic particles into suspension, though the magnetic field is insufficient to cause an aggregation of particles on the opposite side of the vessel from where the pellet had been formed. In addition, since the prior art approach utilizes a resuspension magnet, there is less motivation to accurately aim the acid resuspension liquid. Any inhomogeneity in the suspended particles is addressed by the resuspension magnet.
It would be preferable to provide a system in which the use of a resuspension magnet is obviated.
An improved acid injection module includes dual, parallel injection probes. A high-precision aiming strategy is employed to ensure that complete, homogenous resuspension of accumulated solid-phase particles is achieved, obviating the need for subsequent resuspension magnet positions.
The dual, parallel injector probe nozzles are spaced by a degree necessary to provide substantially adjacent impact zones on the reaction vessel wall, also referred to as “hit zones” or “hit points.” Through careful control over lateral spacing of the two nozzles, and thus the two hit zones, and by performing an exacting analysis of the various physical tolerances which can effect hit zone location relative to the solid-phase pellet, thorough resuspension can be achieved without use of a resuspension magnet.
Other features, aspects and advantages of the above-described method and system will be apparent from the detailed description of the invention that follows.
The invention will be more fully understood by reference to the following detailed description of the invention in conjunction with the drawings of which:
The presently disclosed concept finds particular applicability to automated laboratory analytical analyzers in which paramagnetic particles are drawn into a pellet on the side of a reaction vessel as part of a separation and wash process. In particular, in an analyzer in which chemiluminescence is utilized for determining analyte concentration, the accumulated particles must be thoroughly resuspended to obtain an accurate reading. One approach in such systems is to resuspend the accumulated, washed particles in acid prior to introducing a base, and thus triggering the chemiluminescent response, at an optical measuring device such as a luminometer. However, it is noted that the presently disclosed concept is also applicable to any environment in which thorough resuspension of accumulated particles is required.
One practical aspect not accounted for in the configuration described in
With respect to
The speed v0 of the liquid at the probe tip can be derived from the pump flow rate and the needle inner diameter:
With reference to
s=√{square root over (l2−h2)}
s=3.37 mm=3.37 ·10−3m
With reference to
The first part of the term is equal to h and the second part gives the difference between the ideal shown in
Overall, there are four tolerance chains which can affect the hit point:
In the following, every tolerance chain is treated individually. Eventually, the total tolerance is estimated by adding the results of the individual tolerance chains.
The calculations for the individual tolerance chains are performed by executing the following steps:
Identification of related parts and their respective tolerances, providing a graphical description of the tolerance chain;
Graphical vector analysis of the tolerance chain;
Generation of a table of dimensions, tolerances, maximum dimensions, minimum dimensions;
Calculation of the ideal closure dimension;
Calculation of the arithmetic maximum and minimum closure dimensions and the arithmetic tolerance;
Identification of mean values from asymmetric tolerance zones and means values of shape and positional tolerances;
Generation of closure dimension as distribution average;
Identification of deviation σ/variance σ2 for every dimension and calculation of the total error according to the theorem of error propagation; and
Evaluation of statistical closure dimension and tolerance.
The dimensions of all parts are considered to have normal, Gaussian distributions with a deviation of ±3σ. This means that 99.73% of all parts are inside the tolerance zone. This assumption is realistic for lot sizes of 60 to 100 parts and greater. The shape and position tolerances have a folded normal distribution.
For statistical calculation of the hit point tolerance, a mathematical description of the hit point depending upon linear position and angle of the probe is necessary. The arc of the liquid stream is omitted at this point for simplicity, but is factored in subsequently.
A simplified arrangement of a probe module and cuvette is shown in
The draft angle β, not taken into account in the foregoing, is 0.5°.
(Eq. 1). Substituting the projected values from
Height tolerances are now considered with respect to
For the worst case in terms of height, it is assumed all tolerances are at their maximum, so that clearance between the washer plate and the cuvette is minimal. The hit point is thus lowered towards the bottom of the cuvette. To achieve this, parts of the left side of
The vector diagram of
In the following table, all factors with the respective maximum and minimum values and resulting tolerance zones are provided:
The nominal closure dimension MOH:
The arithmetic maximum closure dimension ymax:
The arithmetic minimum closure dimension ymin:
The arithmetic closure dimension with tolerance zone is thus:
Some statistical calculations are necessary to account for component fluctuations. The mean values from asymmetric tolerance zones M2, M4, M5 and M6 are now defined. For M2:
Similar calculations for M4, M5, and M6 yield:
μ4=1.72
μ5=15.1
μ6=1.17
As for M1, M8, and M11, shape and positional tolerances are distributed with a folded normal distribution. Mean values and deviations must therefore be calculated with the following equations. A deviation of 3σ is thereby assumed.
The closure dimension μ0H is calculated as a distribution average:
The deviation σ0H of the closure dimension:
The statistical closure dimension with tolerance zone is:
M
0H
=y=μ
0H±(TSH/2)=1.974±0.405
Axial tolerances are now considered.
The vector diagram of
The const. vector is the constant value shown in
In the following table, all of the contributors with their maximum and minimum dimensions and tolerance zones are provided.
The nominal closure dimension M0A is given by:
The arithmetic maximum closure dimension xmax is given by:
The arithmetic minimum closure dimension xmin is given by:
From these values, the arithmetic closure dimension with tolerance zone is given by:
Some statistical calculations are necessary to account for component fluctuations. The mean values for shape and position for tolerance M13 are now defined.
M13: σ13=0.033 μF13=0.027 σF13=0.02
Closure dimension μ0A is given as a distribution average:
−0.3−23.03+226−0.02−215.87+14.12=0.9
The deviation σ0A of the closure dimension is determined from:
The statistical closure dimension with tolerance zones is given by:
M
0A
=x=μ
0A
±T
SA/2=0.9±0.162
Injector inclination tolerances are now addressed. The tolerance of the bores in the washer plate is M16=±0.05°. The parallelism of the axis of the probe bore and the axis of the injector outer diameter is M17=0.05 mm. With the length of 18 mm this results in an angle tolerance of:
The nominal angle φ0 is given by:
The arithmetic maximum angle φmax is given by:
The arithmetic minimum angle φmin is given by:
The closure dimension with tolerance zone is thus given by:
Some statistical calculations are necessary to account for component fluctuation. The mean values for shape and position for tolerance M17 are now defined.
M17: σ17=0.033° μF17=0.027° σF17=0.02°
The average angle distribution μ0φ is given by:
The deviation of the angle error is given by:
The statistical angle error with tolerance zone is thus given by:
The worst case calculation for hreal can now be calculated by setting the arithmetic maximum values for xmax, ymax, and φmax into Eq. 1, above.
The arithmetic minimum can be calculated using the analog:
Thus, the arithmetic derivation of the hit point with tolerance zone is given by:
The hit point μh as distribution average with μ0H=1.974, μ0A=0.9, μ0φ=6.986° and employing Eq. 1:
μh=25.812
The statistical deviation σh of the hit point, depending upon the variables σ0A, σ0H, σ0φ, can now be calculated using Eq. 1. Using partial derivatives at the distribution average:
With μ0H=1.974, μ0A=0.9, μ0φ=6.986° and σ0H=0.135, σ0A=0.054, and σ0φ=0.017°, the result is:
σh=0.435
TSH=6σh=2.61
The statistical error of the hit point with tolerance zone is thus given by:
In the embodiment in which the pellet is formed by a magnet array, the tolerance of the array relative to the cuvettes must also be accounted for. The magnets, in a preferred embodiment, are fixed in a ring which is suspended under the transport ring. Most of the tolerance of the magnets is addressed in the height tolerances previously calculated. Thus, there are only the following tolerances to be accounted for:
M18—slide bearing;
M19—magnet ring (i.e. the position of the magnet assembly in the magnet ring);
M20—magnet assembly (i.e. the tolerance of the fixture into which the magnet assembly is fixed); and
M21—the slide bearing support.
All of the above contribute to movement in the same direction.
The nominal closure dimension M0M is given by:
M0M=ΣMi
4+7.4+1=12.4
The arithmetic maximum closure dimension P0M is given by:
P0M=ΣG0i
4.1+7.45+0.05+1.1=12.7
The arithmetic minimum closure dimension P0M is given by:
P0M=ΣG0i
4.05+7.35−0.05+0.9=12.25
The arithmetic closure dimension with tolerance zone is thus given by:
Mean values from asymmetric tolerance zone M18 is given by:
μ18=4.075
The closure dimension μ0M as a distribution average is found according to:
5.075+7.4+1=12.475
The deviation σ0M of the closure dimension is given by:
The statistical closure dimension with tolerance zone is thus:
The nominal distance between the magnet centerline and the cuvette top plane at the acid injection position is 31.72 mm. This value can be calculated with the nominal dimensions listed above:
3.9+12.4+6.35+5+3+1.067=31.717
(3.9 being the distance between the upper magnet and the magnet ring, 6.35 being the magnet width).
The deviation σh and the tolerance zone TSH of the hit point relative to the cuvette top plane was estimated above as 25.98±1.305 mm. The nominal measure between hit point and magnet centerline is thus:
h
total=31.717−25.98=5.737
The total deviation σ of the difference between hit point and magnet centerline is thus calculated by:
√{square root over (0.4352+0.0422)}=0.437
Ts=6σ=2.622
The statistical error of the hit point versus magnet centerline with tolerance zone can then be written as:
Once 0.25 mm is added to compensate for the arc of the liquid stream, the acid injection is calculated to hit the cuvette wall not deeper than 4.167 mm above the magnet centerline.
One embodiment of a probe housing 100 is illustrated in
The linear dimensions in
On the back of the probe housing 100 is a mounting recess 110 for interfacing to a resuspension liquid-supplying conduit (not shown). Secure attachment of the conduit to the housing 100 is preferably through interlocking threads or other means known to one skilled in the art. Preferably a buffer zone 112 exists between the forward end of the conduit once installed in the recess 110. Liquid from the conduit passes into the buffer zone and then into each of two channels 114 which lead to respective probes 116 and the probe nozzles 102 themselves. In the illustrated embodiment, the probes 116 and nozzles 102 are 0.65±0.02 mm in diameter.
Having described preferred embodiments of the presently disclosed invention, it should be apparent to those of ordinary skill in the art that other embodiments and variations incorporating these concepts may be implemented. Accordingly, the invention should not be viewed as limited to the described embodiments but rather should be limited solely by the scope and spirit of the appended claims.
This application claims priority to U.S. Prov. Appl. No. 60/574,000, filed May 24, 2004, the entirety of which is hereby incorporated by reference.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US05/18238 | 5/24/2005 | WO | 00 | 11/15/2007 |
Number | Date | Country | |
---|---|---|---|
60574000 | May 2004 | US |