The invention generally relates to systems and methods for mass spectrometry analysis of microorganisms in samples.
Mass spectrometry is a very sensitive analytical method used for important research and for applications of analytical chemistry, particularly life science. Electrospray ionization (ESI) is generally regarded as the best-characterized and most efficient method for ionization of molecules in solution phase. The process can be conveniently divided into three stages: droplet formation, droplet evaporation and ion formation (Gaskell, S. J. Journal of Mass Spectrometry 1997, 32, 677-688). When a strong electric field is applied to a solution flowing through a mass spectrometer probe, a Taylor cone is formed at the tip of the probe, resulting in a mist of small droplets being emitted from the tip of this cone. Due to the evaporation of the free droplets and Coulombic forces, ions of sample analyte are produced. The ions enter a mass spectrometer and are subsequently analyzed.
A problem with ESI is that sample preparation is still a necessary step before ESI can be used for analysis of many types of samples. Prior to analyzing a sample by ESI mass spectrometry, the sample will undergo extraction and filtration protocols to purify the sample, for example to remove salts and detergents. Such protocols are complex, time-consuming, and expensive. Further, reagents used during the purification process can interfere with subsequent analysis of a target analyte in the purified sample. Additionally, samples that are not in solution must be dissolved as well as purified prior to ESI analysis.
More recently, the concept of ambient ionization has been developed, and now this family of ambient ionization has more than twenty members, such as desorption electrospray ionization (DESI) and direct analysis in real time (DART). Ambient ionization by mass spectrometry allows the ionization of analytes under an ambient environment from condensed-phase samples without much or even any sample preparation and/or pre-separation, offering a solution for real time and in situ analysis for complex mixtures and biological samples. These ambient ionization methods are leading are extending the mass spectrometry revolution in life science, environment monitoring, forensic applications and therapeutic analysis. However, the above described ambient ionization techniques still require pneumatic assistance, a continuous flow of solvent, and a high voltage power supply for the analysis of samples.
There is an unmet need for systems and methods that can combine sample preparation and pre-treatment and the ionization process for mass analysis of samples that do not require pneumatic assistance or a continuous flow of solvent for the analysis of the samples.
The invention generally relates to new systems and methods of generating ions from fluids and solid samples for mass spectrometric analysis. Porous materials, such as filter paper or similar materials are used to hold and transfer liquids, and ions are generated directly from the edges of the materials when a high electric voltage is applied to the materials. The porous material is kept discrete (i.e., separate or disconnected from) from a flow of solvent. Instead, a sample is either spotted onto the porous material or the porous material is wetted and used to swab a surface containing the sample. The porous material with spotted or swabbed sample is then wetted and connected to a high voltage source to produce ions of the sample which are subsequently analyzed. The sample is transported through the porous material without the need of a separate solvent flow.
Devices and methods of the invention combine sample preparation and pre-treatment with the ionization process needed for mass analysis of samples. Device and methods of the invention allow for rapid and direct analysis of chemicals in raw biological samples of complex matrices, such as biofluids and tissues, without sample preparation. In particular embodiments, devices and methods of the invention allow for the analysis of a dried spots of blood or urine.
An aspect of the invention provides a mass spectrometry probe including a porous material connected to a high voltage source, in which the porous material is discrete from a flow of solvent. Exemplary porous materials include paper, e.g., filter paper, or PVDF membrane. The porous material can be of any shape. In certain embodiments, the porous material is provided as a triangular piece.
In certain embodiments, the probe further includes a discrete amount of a solvent, e.g., a droplet or droplets, applied to the porous material. The solvent is applied as a droplet or droplets, and in an amount sufficient to wet the porous material. Once applied to the porous material, the solvent can assist transport of the sample through the porous material. The solvent can contain an internal standard. The solvent/substrate combination can allow for differential retention of sample components with different chemical properties. In certain embodiments, the solvent minimizes salt and matrix effects. In other embodiments, the solvent includes chemical reagents that allow for on-line chemical derivatization of selected analytes.
Another aspect of the invention provides a system for analyzing a sample material including, a probe including a porous material connected to a high voltage source, in which the porous material is kept separate from a flow of solvent, and a mass analyzer. The mass analyzer can be that of a benchtop mass spectrometer or a handheld mass spectrometer. Exemplary mass analyzers include a quadrupole ion trap, a rectilinear ion trap, a cylindrical ion trap, a ion cyclotron resonance trap, and an orbitrap.
Another aspect of the invention includes a method for analyzing a sample including, contacting a sample to a porous material, in which the porous material is kept separate from a flow of solvent, applying a high voltage to the porous material to generate ions of an analyte in the sample that are expelled from the porous material, and analyzing the expelled ions. The method can further include applying a discrete amount, e.g., a droplet or droplets, of a solvent to the porous material. In certain embodiments, analyzing involves providing a mass analyzer to generate a mass spectrum of analytes in the sample.
In certain embodiments, the sample is a liquid. In other embodiments, the sample is a solid. In embodiments in which the sample is a solid, the porous material can be used to swab the sample from a surface. A solvent can be applied to the porous material prior to or after the solid has been swabbed. Exemplary samples include chemical species or biological species.
Another aspect of the invention provides a method of ionizing a sample including applying a high voltage to a porous material to generate ions of an analyte in the sample, in which the porous material remains separate from a solvent flow. Exemplary porous materials include paper or PVDF membrane.
A new method of generating ions from fluids and solids for mass spectrometry analysis is described. Porous materials, such as paper (e.g. filter paper or chromatographic paper) or other similar materials are used to hold and transfer liquids and solids, and ions are generated directly from the edges of the material when a high electric voltage is applied to the material (
In certain embodiments, the porous material is any cellulose-based material. In other embodiments, the porous material is a non-metallic porous material, such as cotton, linen wool, synthetic textiles, or plant tissue. In still other embodiments, the porous material is paper. Advantages of paper include: cost (paper is inexpensive); it is fully commercialized and its physical and chemical properties can be adjusted; it can filter particulates (cells and dusts) from liquid samples; it is easily shaped (e.g., easy to cut, tear, or fold); liquids flow in it under capillary action (e.g., without external pumping and/or a power supply); and it is disposable.
In certain embodiments, the porous material is integrated with a solid tip having a macroscopic angle that is optimized for spray. In these embodiments, the porous material is used for filtration, pre-concentration, and wicking of the solvent containing the analytes for spray at the solid type.
In particular embodiments, the porous material is filter paper. Exemplary filter papers include cellulose filter paper, ashless filter paper, nitrocellulose paper, glass microfiber filter paper, and polyethylene paper. Filter paper having any pore size may be used. Exemplary pore sizes include Grade 1 (11 μm), Grade 2 (8 μm), Grade 595 (4-7 μm), and Grade 6 (3 μm), Pore size will not only influence the transport of liquid inside the spray materials, but could also affect the formation of the Taylor cone at the tip. The optimum pore size will generate a stable Taylor cone and reduce liquid evaporation. The pore size of the filter paper is also an important parameter in filtration, i.e., the paper acts as an online pretreatment device. Commercially available ultra filtration membranes of regenerated cellulose, with pore sizes in the low nm range, are designed to retain particles as small as 1000 Da. Ultra filtration membranes can be commercially obtained with molecular weight cutoffs ranging from 1000 Da to 100,000 Da.
Probes of the invention work well for the generation of micron scale droplets simply based on using the high electric field generated at an edge of the porous material. In particular embodiments, the porous material is shaped to have a macroscopically sharp point, such as a point of a triangle, for ion generation. Probes of the invention may have different tip widths. In certain embodiments, the probe tip width is at least about 5 μm or wider, at least about 10 μm or wider, at least about 50 μm or wider, at least about 150 μm or wider, at least about 250 μm or wider, at least about 350 μm or wider, at least about 400μ or wider, at least about 450 μm or wider, etc. In particular embodiments, the tip width is at least 350 μm or wider. In other embodiments, the probe tip width is about 400 μm. In other embodiments, probes of the invention have a three dimensional shape, such as a conical shape.
As mentioned above, no pneumatic assistance is required to transport the droplets. Ambient ionization of analytes is realized on the basis of these charged droplets, offering a simple and convenient approach for mass analysis of solution-phase samples.
Sample solution is directly applied on the porous material held in front of an inlet of a mass spectrometer without any pretreatment. Then the ambient ionization is performed by applying a high potential on the wetted porous material. In certain embodiments, the porous material is paper, which is a type of porous material that contains numerical pores and microchannels for liquid transport. The pores and microchannels also allow the paper to act as a filter device, which is beneficial for analyzing physically dirty or contaminated samples.
In other embodiments, the porous material is treated to produce microchannels in the porous material or to enhance the properties of the material for use as a probe of the invention. For example, paper may undergo a patterned silanization process to produce microchannels or structures on the paper. Such processes involve, for example, exposing the surface of the paper to tridecafluoro-1,1,2,2-tetrahydrooctyl-1-trichlorosilane to result in silanization of the paper. In other embodiments, a soft lithography process is used to produce microchannels in the porous material or to enhance the properties of the material for use as a probe of the invention. In other embodiments, hydrophobic trapping regions are created in the paper to pre-concentrate less hydrophilic compounds.
Hydrophobic regions may be patterned onto paper by using photolithography, printing methods or plasma treatment to define hydrophilic channels with lateral features of 200-1000 μm. See Martinez et al. (Angew. Chem. Int. Ed. 2007, 46, 1318-1320); Martinez et al. (Proc. Natl Acad. Sci. USA 2008, 105, 19606-19611); Abe et al. (Anal. Chem. 2008, 80, 6928-6934); Bruzewicz et al. (Anal. Chem. 2008, 80, 3387-3392); Martinez et al. (Lab Chip 2008, 8, 2146-2150); and Li et al. (Anal. Chem. 2008, 80, 9131-9134), the content of each of which is incorporated by reference herein in its entirety. Liquid samples loaded onto such a paper-based device can travel along the hydrophilic channels driven by capillary action.
Another application of the modified surface is to separate or concentrate compounds according to their different affinities with the surface and with the solution. Some compounds are preferably absorbed on the surface while other chemicals in the matrix prefer to stay within the aqueous phase. Through washing, sample matrix can be removed while compounds of interest remain on the surface. The compounds of interest can be removed from the surface at a later point in time by other high-affinity solvents. Repeating the process helps desalt and also concentrate the original sample.
Methods and systems of the invention use a porous material, e.g., paper, to hold and transport analytes for mass spectral analysis. Analytes in samples are pre-concentrated, enriched and purified in the porous material in an integrated fashion for generation of ions with application of a high voltage to the porous material. In certain embodiments, a discrete amount of transport solution (e.g., a droplet or a few droplets) is applied to assist movement of the analytes through the porous material. In certain embodiments, the analyte is already in a solution that is applied to the porous material. In such embodiments, no additional solvent need be added to the porous material. In other embodiments, the analyte is in a powdered sample that can be easily collected by swabbing a surface. Systems and methods of the invention allow for analysis of plant or animal tissues, or tissues in living organisms.
Methods and systems of the invention can be used for analysis of a wide variety of small molecules, including epinephrine, serine, atrazine, methadone, roxithromycin, cocaine and angiotensin I. All display high quality mass and MS/MS product ion spectra (see Examples below) from a variety of porous surfaces. Methods and systems of the invention allow for use of small volumes of solution, typically a few μL, with analyte concentrations on the order of 0.1 to 10 μg/mL (total amount analyte 50 pg to 5 ng) and give signals that last from one to several minutes.
Methods and systems of the invention can be used also for analysis of a wide variety of biomolecules, including proteins and peptides. Methods of the invention can also be used to analyze oligonucleotides from gels. After electrophoretic separation of oligonucleotides in the gel, the band or bands of interest are blotted with porous material using methods known in the art. The blotting results in transfer of at least some of the oligonucleotides in the band in the gel to the porous material. The porous material is then connected to a high voltage source and the oligonucleotides are ionized and sprayed into a mass spectrometer for mass spectral analysis.
Methods and systems of the invention can be used for analysis of complex mixtures, such as whole blood or urine. The typical procedure for the analysis of pharmaceuticals or other compounds in blood is a multistep process designed to remove as many interferences as possible prior to analysis. First, the blood cells are separated from the liquid portion of blood via centrifugation at approximately 1000×g for 15 minutes (Mustard, J. F.; Kinlough-Rathbone, R. L.; Packham, M. A. Methods in Enzymology; Academic Press, 1989). Next, the internal standard is spiked into the resulting plasma and a liquid-liquid or solid-phase extraction is performed with the purpose of removing as many matrix chemicals as possible while recovering nearly all of the analyte (Buhrman, D. L.; Price, P. I.; Rudewicz, P. J. Journal of the American Society for Mass Spectrometry 1996, 7, 1099-1105). The extracted phase is typically dried by evaporating the solvent and then resuspended in the a solvent used as the high performance liquid chromatography (HPLC) mobile phase (Matuszewski, B. K.; Constanzer, M. L.; Chavez-Eng, C. M., Ithaca, N.Y., Jul. 23-25 1997; 882-889). Finally, the sample is separated in the course of an HPLC run for approximately 5-10 minutes, and the eluent is analyzed by electrospray ionization-tandem mass spectrometry (Hopfgartner, G.; Bourgogne, E. Mass Spectrometry Reviews 2003, 22, 195-214).
Methods and systems of the invention avoid the above sample work-up steps. Methods and systems of the invention analyze a dried blood spots in a similar fashion, with a slight modification to the extraction procedure. First, a specialized device is used to punch out identically sized discs from each dried blood spot. The material on these discs is then extracted in an organic solvent containing the internal standard (Chace, D. H.; Kalas, T. A.; Naylor, E. W. Clinical Chemistry 2003, 49, 1797-1817). The extracted sample is dried on the paper substrate, and the analysis proceeds as described herein.
Examples below show that methods and systems of the invention can directly detect individual components of complex mixtures, such as caffeine in urine, 50 pg of cocaine on a human finger, 100 pg of heroin on a desktop surface, and hormones and phospholipids in intact adrenal tissue, without the need for sample preparation prior to analysis (See Examples below). Methods and systems of the invention allow for simple imaging experiments to be performed by examining, in rapid succession, needle biopsy tissue sections transferred directly to paper.
Analytes from a solution are applied to the porous material for examination and the solvent component of the solution can serve as the electrospray solvent. In certain embodiments, analytes (e.g., solid or solution) are pre-spotted onto the porous material, e.g., paper, and a solvent is applied to the material to dissolve and transport the analyte into a spray for mass spectral analysis.
In certain embodiments, a solvent is applied to the porous material to assist in separation/extraction and ionization. Any solvents may be used that are compatible with mass spectrometry analysis. In particular embodiments, favorable solvents will be those that are also used for electrospray ionization. Exemplary solvents include combinations of water, methanol, acetonitrile, and THF. The organic content (proportion of methanol, acetonitrile, etc. to water), the pH, and volatile salt (e.g. ammonium acetate) may be varied depending on the sample to be analyzed. For example, basic molecules like the drug imatinib are extracted and ionized more efficiently at a lower pH. Molecules without an ionizable group but with a number of carbonyl groups, like sirolimus, ionize better with an ammonium salt in the solvent due to adduct formation.
In certain embodiments, a multi-dimensional approach is undertaken. For example, the sample is separated along one dimension, followed by ionization in another dimension. In these embodiments, separation and ionization can be individually optimized, and different solvents can be used for each phase.
In other embodiments, transporting the analytes on the paper is accomplished by a solvent in combination with an electric field. When a high electric potential is applied, the direction of the movement of the analytes on paper is found to be related to the polarity of their charged forms in solution. Pre-concentration of the analyte before the spray can also be achieved on paper by placing an electrode at a point on the wetted paper. By placing a ground electrode near the paper tip, a strong electric field is produced through the wetted porous material when a DC voltage is applied, and charged analytes are driven forward under this electric field. Particular analytes may also be concentrated at certain parts of the paper before the spray is initiated.
In certain embodiments, chemicals are applied to the porous material to modify the chemical properties of the porous material. For example, chemicals can be applied that allow differential retention of sample components with different chemical properties. Additionally, chemicals can be applied that minimize salt and matrix effects. In other embodiments, acidic or basic compounds are added to the porous material to adjust the pH of the sample upon spotting. Adjusting the pH may be particularly useful for improved analysis of biological fluids, such as blood. Additionally, chemicals can be applied that allow for on-line chemical derivatization of selected analytes, for example to convert a non-polar compound to a salt for efficient electrospray ionization.
In certain embodiments, the chemical applied to modify the porous material is an internal standard. The internal standard can be incorporated into the material and released at known rates during solvent flow in order to provide an internal standard for quantitative analysis. In other embodiments, the porous material is modified with a chemical that allows for pre-separation and pre-concentration of analytes of interest prior to mass spectrum analysis.
The spray droplets can be visualized under strong illumination in the positive ion mode and are comparable in size to the droplets emitted from a nano-electrospray ion sources (nESI). In the negative ion mode, electrons are emitted and can be captured using vapor phase electron capture agents like benzoquinone. Without being limited by any particular theory or mechanism of action, it is believed that the high electric field at a tip of the porous material, not the fields in the individual fluid channels, is responsible for ionization.
The methodology described here has desirable features for clinical applications, including neotal screening, therapeutic drug monitoring and tissue biopsy analysis. The procedures are simple and rapid. The porous material serves a secondary role as a filter, e.g., retaining blood cells during analysis of whole blood. Significantly, samples can be stored on the porous material and then analyzed directly from the stored porous material at a later date without the need transfer from the porous material before analysis. Systems of the invention allow for laboratory experiments to be performed in an open laboratory environment.
References and citations to other documents, such as patents, patent applications, patent publications, journals, books, papers, web contents, have been made throughout this disclosure. All such documents are hereby incorporated herein by reference in their entirety for all purposes.
Various modifications of the invention and many further embodiments thereof, in addition to those shown and described herein, will become apparent to those skilled in the art from the full contents of this document, including references to the scientific and patent literature cited herein. The subject matter herein contains important information, exemplification and guidance that can be adapted to the practice of this invention in its various embodiments and equivalents thereof.
The following examples are intended to further illustrate certain embodiments of the invention, and are not to be construed to limit the scope of the invention. Examples herein show that mass spectrometry probes of the invention can ionize chemical and biological samples, allowing for subsequent mass analysis and detection. An exemplary probe was constructed as a paper triangle, which was used to generate micron scale droplets by applying a high potential on the paper. The analytes were ionized from these electrically charged droplets and transported into a conventional mass spectrometer.
Examples below show that a wide range of samples could be directly analyzed in the ambient environment by probes of the invention in both of pure state and complex mixtures. The results showed that paper-based spray has the following benefits: it operated without sheath gas, i.e., few accessories were required for in situ analysis; biological samples (dried blood, urine) could be stored on the precut filter papers for months before analysis; filter paper minimized matrix effects seen with electrospray or nano electrospray in many samples (blood cells, salt and proteins) and enhanced the MS signal of chemicals in complex samples; powdered samples were easily collected by swabbing surfaces using paper pieces and then directly analyzed; the paper could be pretreated to contain internal standards that were released at known rates during solvent flow in quantitative analysis; and the paper could be pretreated to contain matrix suppression or absorption sites or to perform ion exchange or to allow on-line chemical derivatization of selected analytes.
Detection of most analytes was achieved as low as ppb levels (when examined as solutions) or in the low ng to pg range (when solids were examined) and the detection time was less than one minute. Certain Examples below provide a protocol for analyzing a dried blood spot, which can also be used for in situ analysis of whole blood samples. The dried blood spot method is also demonstrated to be compatible with the storage and transport of blood sample for blood screening and other clinical tests.
Devices of the invention integrated the capabilities of sampling, pre-separation, pre-concentration and ionization. Methods and systems of the invention simplify the problem of sample introduction in mass analyzers.
Filter paper was cut into triangular pieces with dimensions of 10 mm long and 5 mm wide and used as a sprayer (
Samples were directly applied to the paper surface that served as a sample purification and pre-concentration device. Filter paper allowed liquid samples to move through the hydrophilic network driven by capillary action and electric effects and to transport them to the tip of the paper. Separation could take place during this transport process. Sample solution was sprayed from the tip and resulted in ionization and MS detection when a high voltage (˜4.5 kV) was applied to the paper surface.
All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, Calif.). The typical temperature of the capillary inlet was set at 150° C. while 30° C. for heroin detection. The lens voltage was set at 65 V for sample analysis and 240 V for survival yield experiment. Tandem mass spectra were collected using collision-induced dissociation (CID) to identify analytes in tested samples, especially for complex mixtures and blood samples.
Spray was produced by applying a high potential on the wetted paper triangle. One paper triangle was placed in front of the inlet of LTQ with its sharp tip facing to the inlet, separated by 3 mm or more. Typically, 10 uL sample solution was applied to wet the paper triangle. The solution can wet or saturate the paper or form a thin layer of liquid film on the surface of the paper. A high potential (3-5 kV) was applied between the paper triangle and mass inlet to generate an electric field, which induced a charge accumulation on the liquid at the tip of paper triangle. The increasing coulombic force breaks the liquid to form charged droplets and then the solvent evaporated during the flight of droplets from the paper tip to the mass analyzer. Paper spray required no sheath gas, heating or any other assistance to remove the solvent.
When liquid accumulated on the paper triangle, a Taylor cone was observed at the tip when examined with a microscope. The droplets formed were clearly visible under strong illumination. The Taylor cone and visible spray disappeared after a short time of evaporation and spray. However, the mass signal lasted for a much longer period (several minutes). This revealed that the paper triangle could work in two modes for mass analysis. In a first mode, the liquid was transported inside the paper at a rate faster than the liquid could be consumed as spray at the paper tip, resulting in a large cone being formed at the paper tip and droplets being generated. In a second mode, the liquid transport inside the paper was not able to move at a rate fast enough to keep up with the spray consumption, and droplets were not visible. However, it was observed that ionization of analytes did take place. The first mode provided ESI like mass spectra and the second mode provided spectra with some of the features APCI spectra. In the latter case, the paper triangle played a role analogous to a conductive needle to generate a high electric field to ionize the molecules in the atmosphere. It was observed that the mass signal in the first mode was stronger than the mass signal in the second mode by approximately two orders of magnitude under the conditions and for the samples tested.
Probe Materials
A number of porous materials were tested to generate charged droplets for mass spectrometry. The materials were shaped into triangles having sharp tips and sample solution was then applied to the constructed probes. Data herein show that any hydrophilic and porous substrate could be used successfully, including cotton swab, textile, plant tissues as well as different papers. The porous network or microchannels of these materials offered enough space to hold liquid and the hydrophilic environment made it possible for liquid transport by capillary action. Hydrophobic and porous substrates could also be used successfully with properly selected hydrophobic solvents.
For further investigation, six kinds of commercialized papers were selected and qualitatively tested to evaluate their capabilities in analyte detection. Filter papers and chromatography paper were made from cellulose, while glass microfiber filter paper was made from glass microfiber.
It was hypothesized that the glass fiber paper was working on mode II and prohibiting efficient droplet generation, due to the relative large thickness (˜2 mm). This hypothesis was proved by using a thin layer peeled from glass fiber paper for cocaine detection. In that case, the intensity of the background increased and a cocaine peak was observed. All filter papers worked well for cocaine detection, (
Probe Shape and Tip Angle
Many different probe shapes were investigated with respect to generating droplets. A preferred shape of the porous material included at least one tip. It was observed that the tip allowed ready formation of a Taylor cone. A probe shape of a triangle was used most often. As shown in
A paper triangle was mounted on a 2D moving stage to determine how the mass signal was affected by the relative positions of the paper triangle and the mass spectrometer inlet. The paper triangle was moved 8 cm in the y-direction in a continuous manner and 3 cm in the x-direction with a 2 mm increment for each step (
Spray duration was also tested (
In another test, the paper was sealed with a PTFE membrane to prevent evaporation of solution, which prolonged the spray time by about three times. These results indicate that paper spray offers long enough time of spray for data acquisition even using 5 uL solution, and the intensity of signal is stable during the entire spray period.
Probes of the invention include a porous material, such as paper, that can function to both separate chemicals in biological fluids before in situ ionization by mass spectrometry. In this Example, the porous material for the probe was chromatography paper. As shown in
The chromatography paper thus allowed for sample collection, analyte separation and analyte ionization. This represents a significant simplification of coupling chromatography with MS analysis. Chromatography paper is a good material for probes of the invention because such material has the advantage that solvent movement is driven by capillary action and there is no need for a syringe pump. Another advantage is that clogging, a serious problem for conventional nanoelectrospray sources, is unlikely due to its multi-porous characteristics. Therefore, chromatography paper, a multi-porous material, can be used as a microporous electrospray ionization source.
As already described, probes and methods of the invention offer a simple and convenient ionization method for mass spectrometry. Paper triangles were spotted with different compounds and connected to a high voltage source to produce ions. All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, Calif.). Data herein show that a variety of chemicals could be ionized in solution phase, including amino acid, therapeutic drugs, illegal drugs and peptides.
Data herein show that probes of the invention work well over the mass/charge range from 50 to over 1000 for detection of pure compounds. Data further shows that detection was achieved down to as low as 1 ng/mL for most chemicals, including illegal drugs, such as heroin, cocaine and methadone.
Complex mixtures such as urine, blood, and cola drink were examined using methods, devices, and systems of the invention. All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, Calif.).
Body fluids, such as plasma, lymph, tears, saliva, and urine, are complex mixtures containing molecules with a wide range of molecular weights, polarities, chemical properties, and concentrations. Monitoring particular chemical components of body fluids is important in a number of different areas, including clinical diagnosis, drug development, forensic toxicology, drugs of abuse detection, and therapeutic drug monitoring. Tests of blood, including the derived fluids plasma and serum, as well as on urine are particularly important in clinical monitoring.
A wide variety of chemicals from blood are routinely monitored in a clinical setting. Common examples include a basic metabolic panel measuring electrolytes like sodium and potassium along with urea, glucose, and creatine and a lipid panel for identifying individuals at risk for cardiovascular disease that includes measurements of total cholesterol, high density lipoprotein (HDL), low density lipoprotein (LDL), and triglycerides. Most laboratory tests for chemicals in blood are actually carried out on serum, which is the liquid component of blood separated from blood cells using centrifugation. This step is necessary because many medical diagnostic tests rely on colorimetric assays and therefore require optically clear fluids. After centrifugation, detection of the molecule of interest is carried in a number of ways, most commonly by an immunoassay, such as an enzyme-linked immunosorbent assay (ELISA) or radioimmunoas say (RIA), or an enzyme assay in which the oxidation of the molecule of interest by a selective enzyme is coupled to a reaction with a color change, such as the tests for cholesterol (oxidation by cholesterol oxidase) or glucose (oxidation by glucose oxidase).
There is considerable interest in the pharmaceutical sciences in the storage and transportation of samples of whole blood as dried blood spots on paper (N. Spooner et al. Anal Chem., 2009, 81, 1557). Most tests for chemicals found in blood are carried out on a liquid sample, typically serum or plasma isolated from the liquid whole blood. The required storage, transportation, and handling of liquid blood or blood components present some challenges. While blood in liquid form is essential for some tests, others can be performed on blood or other body fluids that have been spotted onto a surface (typically paper) and allowed to dry.
Probes and methods of the invention can analyze whole blood without the need for any sample preparation. The sample was prepared as follows. 0.4 uL blood was directly applied on the center of paper triangle and left to dry for about 1 min. to form a dried blood spot (
Different drugs were spiked into whole blood and the blood was applied to probes of the invention as described above. Detection of different drugs is described below.
Imatinib (GLEEVEC), a 2-phenylaminopyrimidine derivative, approved by the FDA for treatment of chronic myelogenous leukemia, is efficacious over a rather narrow range of concentrations. Whole human blood, spiked with imatinib at concentrations including the therapeutic range, was deposited on a small paper triangle for analysis (
Atenolol, a β-blocker drug used in cardiovascular diseases, was tested using the dried blood spot method to evaluate paper spray for whole blood analysis. Atenolol was directly spiked into whole blood at desired concentrations and the blood sample was used as described above for paper spray. The protonated atenolol of 400 pg (1 ug/mL atenolol in 0.4 uL whole blood) in dried blood spot was shown in mass spectra, and the MS/MS spectra indicated that even 20 pg of atenolol (50 ug/mL atenolol in 0.4 uL whole blood) could be identified in the dried blood spot (
It was also observed that the paper medium served a secondary role as a filter, retaining blood cells. Significantly, samples were analyzed directly on the storage medium rather than requiring transfer from the paper before analysis. All experiments were done in the open lab environment. Two additional features indicated that the methodology had the potential to contribute to increasing the use of mass spectrometry in primary care facilities: blood samples for analysis were drawn by means of a pinprick rather than a canula; and the experiment was readily performed using a handheld mass spectrometer (
Systems and methods of the invention were compatible with a handheld mass spectrometer. Paper spray was performed using a handheld mass spectrometer (Mini 10, custom made at Purdue University). Analysis of whole blood spiked with 10 μg/mL atenolol. Methanol/water (1:1; 10 μL) was applied to the paper after the blood (0.4 uL) had dried (˜1 min) to generate spray for mass detection (
Sample collection by paper wiping followed by analysis using probes of the invention was used for fast analysis of agrochemicals on fruit. Chromatography paper (3×3 cm) wetted with methanol was used to wipe a 10 cm2 area on the peel of a lemon purchased from a grocery store. After the methanol had dried, a triangle was cut from the center of the paper and used for paper spray by applying 10 μL methanol/water solution. The spectra recorded (
Systems and methods of the invention were used to analyze human prostate tumor tissue and normal tissue. Tumor and adjacent normal tissue sections were 15 μm thick and fixed onto a glass slide for an imaging study using desorption electrospray ionization (DESI). A metal needle was used to remove a 1 mm2×15 μm volume of tissue from the glass slide from the tumor region and then from the normal region and place them onto the surface of the paper triangle for paper spray analysis.
A droplet of methanol/water (1:1 v:v; 10 μl) was added to the paper as solvent and then 4.5 kV positive DC voltage applied to produce the spray. Phospholipids such as phosphatidylcholine (PC) and sphingomyelin (SM) were identified in the spectrum (
The administration of a drug depends on managing the appropriate dosing guidelines for achievement of a safe and effective outcome. This guideline is established during clinical trials where the pharmacokinetics (PK) and pharmacodynamics (PD) of the drug are studied. Clinical trials use PK-PD studies to establish a standard dose, which may be fixed or adjusted according formulas using variables like body mass, body surface area, etc. However, the drug exposure, i.e. the amount of drug circulating over time, is influenced by a number of factors that vary from patient to patient. For example, an individuals' metabolic rate, the type and level of plasma proteins, and pre-existing conditions such as renal and/or hepatic impairment all play a role in affecting the exposure of the drug in vivo. Further, administration of a drug in combination with other medications may also affect exposure. As a result, it is often difficult to predict and prescribe an optimum regimen of drug administration.
Over- or underexposure to a drug can lead to toxic effects or decreased efficacy, respectively. To address these concerns, therapeutic drug monitoring (TDM) can be employed. TDM is the measurement of active drug levels in the body followed by adjustment of drug dosing or schedules to increase efficacy and/or decrease toxicity. TDM is indicated when the variability in the pharmacokinetics of a drug is large relative to the therapeutic window, and there is an established relationship between drug exposure and efficacy and/or toxicity. Another requirement for TDM is that a sufficiently precise and accurate assay for the active ingredient must be available. Immunoassays and liquid chromatography mass spectrometry (LC-MS) are commonly used methods for TDM. In comparison with immunoassay, LC-MS has advantages which include wide applicability, high sensitivity, good quantitation, high specificity and high throughput. Probes of the invention may be coupled with standard mass spectrometers for providing point-of-care therapeutic drug monitoring.
The drug Imatinib (GLEEVEC in USA and GLIVEC in Europe/Australia, for the treatment of chronic myelogenous leu-kemia) in a dried blood spot was analyzed using paper spray and a lab-scale LTQ mass spectrometer. Quantitation of Imatinib in whole blood was achieved using the MS/MS spectra with a known concentration of Imatinib-d8 being used as the internal standard (
Multiple-tip devices were fabricated and applied for high throughput analysis (
To prevent the contaminant during spray, blanks were inserted between two sample tips.
Direct analysis of chemicals in animal tissue using probes of the invention was performed as shown in
Lipid profiles were obtained for human prostate tissues (1 mm2×15 μm,
For analysis of target analytes which have relatively low ionization efficiencies and relatively low concentrations in mixtures, derivatization is often necessary to provide adequate sensitivity. On-line derivatization can be implemented by adding reagents into the spray solution, such as methanol/water solutions containing reagents appropriate for targeted analytes. If the reagents to be used are stable on paper, they can also be added onto the porous material when the probes are fabricated.
As a demonstration, 5 μL methanol containing 500 ng betaine aldehyde chloride was added onto a paper triangle and allowed to dry to fabricate a sample substrate preloaded with a derivatization reagent for the analysis of cholesterol in serum. On-line charge labeling with betaine aldehyde (BA) through its reaction with hydroxyl groups has been demonstrated previously to be very effective for the identification of cholesterol in tissue (Wu et al., Anal Chem. 2009, 81:7618-7624). When the paper triangle was used for analysis, 2 μL human serum was spotted onto the paper to form a dried spot and then analyzed by using paper spray ionization. A 10 μL ACN/CHCl3 (1:1 v:v) solution, instead of methanol/water, was used for paper spray to avoid reaction between the betaine aldehyde and methanol.
The comparison between analysis using a blank and a reagent-preloaded paper triangle is shown in
Pre-concentration of chemicals on the paper surface using photoresist treatment. Chromatography paper was rendered hydrophobic by treatment with SU-8 photoresist as described previously (Martinez et al., Angew Chem Int. Ed., 2007, 46:1318-1320). Then 5 μl bradykinin 2-9 solution (100 ppm in pure H2O) was applied on the paper surface. When the solution was dry, the paper was put into water and washed for 10 s. After washing, the paper triangle was held in front of the MS inlet, 10 μl pure MeOH was applied as solvent and the voltage was set at 4.5 kV for paper spray. The same experiment was done with untreated paper substrate for comparison.
The polarity of the voltage applied to the probe need not match that used in the mass analyzer. In particular, it is possible to operate the probes of the invention with a negative potential but to record the mass spectrum of the resulting positively changed ions. In negative ion mode, a large current of electrons (or solvated electrons) is produced in paper spray. These electrons, if of suitable energy, can be captured by molecules with appropriate electron affinities to generate radical anions.
Alternatively, these electrons might be responsible for electron ionization of the analyte to generate the radical cation or alternatively ESI might involve a solvent molecule which might then undergo charge exchange with the analyte to again generate the radical cation. If this process occurs with sufficient energy, characteristic fragment ions might be produced provided the radical cation is not collisionally deactivated before fragmentation can occur.
An experiment was done on a benchtop LTP using toluene vapor, with a probe of the invention conducted at −4.5kV with methanol:water as solvent applied to the paper. The spectrum shown in
An interesting note is that the “EI” fragment ions were most easily produced when the source of toluene vapor was placed close to the MS inlet; i.e., in the cathodic region of the discharge between the paper tip and MS inlet. This suggests that direct electron ionization by energetic electrons in the “fall” region might be at least partly responsible for this behavior.
In certain aspects, probes of the invention can be used to analyze one or more microorganisms (e.g., bacteria, viruses, protozoans (also spelled protozoon), or fungi) in a sample. An exemplary method involves contacting a sample including a microorganism to a porous material, in which the porous material is kept separate from a flow of solvent. The method further involves applying high voltage to the porous material to generate ions of the microorganism that are expelled from the porous material, and analyzing the expelled ions, thereby analyzing the microorganism. Methods of the invention may also involve applying a solvent to the porous material. Any mass spectrometer known in the art may be used, and in certain embodiments, the mass spectrometer is a miniature mass spectrometer, such as that described for example in Gao et al. (Anal. Chem., 80:7198-7205, 2008) and Hou et al. (Anal. Chem., 83:1857-1861, 2011), the content of each of which is incorporated herein by reference herein in its entirety.
The sample may be any type of sample and may be in any form, for example, a solid, a liquid, or a gas. In certain embodiments, the sample is a human tissue or body fluid. The sample may be an in vivo sample or an extracted sample. In certain embodiments, the methods of the invention are sensitive enough to analyze and identify microorganisms without first culturing the microorganism. In some embodiments, the microorganism is cultured prior to analysis, however, methods of the invention allow for decreased culture time over that used in standard procedures.
If the sample is a solid, as in the case of a microorganism grown on agar in a petri dish for example, the porous substrate can be contacted to the solid and swapped or moved across the surface such that microorganisms in the sample are retained by the porous substrate. There are other methods for collecting microorganisms from solid samples. For example, for colonies on petri dishes, a sterile inoculation loop is used to transfer one or more colonies to the porous substrate. Once collected, voltage and a discrete amount of solvent is applied to the probe, and ions of the microorganism are generated and expelled into a mass spectrometer for analysis.
Regardless of the collection method, voltage, and optionally a discrete amount of solvent, is applied to the probe, and ions of the microorganism are generated and expelled into a mass spectrometer for analysis.
Aspects of the invention also provide methods of identifying an organism, e.g., a microorganism. The methods include obtaining a mass spectrum of an organism using porous substrate probes of the invention and correlating/comparing the mass spectra with a database that includes mass spectra of known organisms (FIGS. 43-45A-F). With use of methods of the invention, the organism can be identified and classified not just at a genus and species level, but also at a sub-species (strain), a sub-strain, and/or an isolate level. The featured methods offer fast, accurate, and detailed information for identifying organisms. The methods can be used in a clinical setting, e.g., a human or veterinary setting; or in an environmental, industrial or forensic/public safety setting (e.g., clinical or industrial microbiology, food safety testing, ground water testing, air testing, contamination testing, and the like). In essence, the invention is useful in any setting in which the detection and/or identification of a microorganism is necessary or desirable.
A database for use in the invention can include a similarity cluster. The database can be used to establish, through linear discriminant analysis and related methods, a quantitative measure of the similarity of any two spectra, a similarity index. The database can include a mass spectrum from at least one member of the Clade of the organism. The database can include a mass spectrum from at least one subspecies of the organism. The database can include a mass spectrum from a genus, a species, a strain, a sub-strain, or an isolate of the organism. The database can include a mass spectrum with motifs common to a genus, a species, a strain, a sub-strain, or an isolate of the organism.
The database(s) used with the methods described herein includes mass spectra associated with known organisms (
To generate similarity clusters, each mass spectrum is aligned against every other mass spectrum. From these alignments, a pair-wise alignment analysis is performed to determine “percent dissimilarity” between the members of the pair (
Various organisms, e.g., viruses, and various microorganisms, e.g., bacteria, protists, and fungi, can be identified with the methods featured herein. The sample containing the organism to be identified can be a human sample, e.g., a tissue sample, e.g., epithelial (e.g., skin), connective (e.g., blood and bone), muscle, and nervous tissue, or a secretion sample, e.g., saliva, urine, tears, and feces sample. The sample can also be a non-human sample, e.g., a horse, camel, llama, cow, sheep, goat, pig, dog, cat, weasel, rodent, bird, reptile, and insect sample. The sample can also be from a plant, water source, food, air, soil, plants, or other environmental or industrial sources.
The methods described herein include correlating the mass spectrum from the unknown organism with a database that includes mass spectra of known organisms. The methods involve comparing each of the mass spectra from the unknown organism from a sample against each of the entries in the database, and then combining match probabilities across different spectra to create an overall match probability (
The present application is a continuation of U.S. nonprovisional application Ser. No. 14/481,483, filed Sep. 9, 2014, which is a continuation of U.S. nonprovisional application Ser. No. 14/151,220, filed Jan. 9, 2014, which is a continuation of U.S. nonprovisional application Ser. No. 13/926,645, filed Jun. 25, 2013, which is a continuation-in-part of U.S. nonprovisional application Ser. No. 13/265,110, filed Jan. 31, 2012, which application is a national phase application and claims the benefit of and priority to PCT/US2010/032881, filed Apr. 29, 2010, which claims priority to and the benefit of U.S. provisional patent application Ser. No. 61/174,215, filed Apr. 30, 2009, U.S. provisional patent application Ser. No. 61/246,707 filed Sep. 29, 2009, and U.S. provisional patent application Ser. No. 61/308,332, filed Feb. 26, 2010, the content of each of which is incorporated by reference herein in its entirety.
Number | Date | Country | |
---|---|---|---|
61308332 | Feb 2010 | US | |
61246707 | Sep 2009 | US | |
61174215 | Apr 2009 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14481483 | Sep 2014 | US |
Child | 14594233 | US | |
Parent | 14151220 | Jan 2014 | US |
Child | 14481483 | US | |
Parent | 13926645 | Jun 2013 | US |
Child | 14151220 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13265110 | Jan 2012 | US |
Child | 13926645 | US |