This disclosure relates to water control and metering. More specifically, this disclosure relates to actuators for use in a valve meter assembly.
Water is typically supplied by a water provider which is usually a municipality. Water providers deliver water to businesses and individuals via piping systems. A piping system could be an upstream piping system, including a system to carry water from a water provider to a meter, or a downstream piping system, including a system to carry water from a meter to a user terminal. Because water providers typically sell water by unit volume, there exists a need to measure water flow to a user terminal to generate a water bill. As used herein, user terminal may include an individual residence, a place of business or any other point of termination of the water flow. Typically, a water meter will be placed in the water supply line between the water source and the user terminal to measure all water flowing to that user terminal. Meters are read and checked against prior readings to determine the total flow of water to the user terminal.
When a water user has not provided payment for water already used, it is typical in the industry for a water provider to discontinue supplying water to the user terminal associated with the water user. Typically, a water provider will install a manual water supply valve in the supply line in anticipation of the need to discontinue water supply. Although the valve may be rarely operated, there are many situations where the valve is operated on a regular basis. Further, a manual valve is standard equipment for water providers and may even be required by statutes in some localities.
Typically, water meters are read manually by water meter readers who are employees or contractors of the water providers. Additionally, water supply valves are manually operated by employees or contractors of the water providers. These manual operations associated with providing water represent a significant cost for a water provider. With the advent of wireless technology, water providers have sought methods and systems for remote reading of water meters and/or remote control of water supply valves.
Mesh networks for remote reading of water meters exist currently. Systems for remotely controlling the water supply valve exist currently. However, these systems are often cumbersome to implement, requiring excavation and replacement of water supply lines to implement a remotely controlled water supply valve. Electronic remote control of valves and reading of meters has been implemented through wired connections. While wireless systems for controlling valves or for reading meters do exist, the cast ferrous materials used to make most water meter housings can interfere with wireless signals, so the wireless equipment often cannot be placed in close proximity to typical meter housings. Moreover, a remotely controlled valve typically involves a separate system and apparatus from a remotely readable water meter. Systems that integrate a shutoff valve and water meter together are often too large to be installed without excavation of the water supply lines and are typically difficult to service if parts fail. Some systems designed to fit into the standard water meter lay-length of a water meter provide inordinate head loss through the system and provide only remote control of the valve and no ability to read the meter remotely. Moreover, wireless water supply valves typically have relatively short operative lives because their operation requires large amounts of energy.
Water meters may include a valve operable with an actuator such as a solenoid or any similar component. The actuator may operate the valve by placing a plunger over an orifice bore defined in the valve, such as in a diaphragm valve. It may therefore be desirable to precisely locate the actuator relative to the orifice bore. This may be accomplished with special tools, for example with a tool that tightens a component on an assembly to a specific torque value, but it may be desired or even required in some applications that the component be installable without such special tools. Incorporating such a feature into the component (or components) itself can have practical benefits for those who purchase, control, and/or service the equipment.
Disclosed is a valve including a valve cover, the valve cover having a solenoid attachment portion, an orifice bore defined in the solenoid attachment portion, and a valve cover mechanical stop, the solenoid attachment portion having a solenoid attachment sink; and a solenoid, the solenoid having a solenoid body, a valve cover attachment portion, and a plunger, the plunger having an open position and a closed position, the plunger including an interface portion, the solenoid body including a solenoid mechanical stop, the solenoid mechanical stop engagable with the valve cover mechanical stop to space the interface portion of the plunger away from the orifice bore when the plunger is in the open position.
Also disclosed is an assembly including a housing, the housing defining at least one inlet opening, at least one outlet opening, and a channel connecting the at least one inlet opening and the at least one outlet opening, the at least one inlet opening having an inlet end and the at least one outlet opening having an outlet end, there being a linear distance between the inlet end and the outlet end, the linear distance being no greater than a standard water meter lay-length; a water meter positioned in the channel, the water meter configured to monitor a flow of water through the assembly; and a valve in communication with the channel and configured to control the flow of water through the assembly, the valve having an actuator and a valve cover, the actuator having a actuator mechanical stop and the valve cover having a valve cover mechanical stop, the actuator mechanical stop engagable with the valve cover mechanical stop to set the position of the actuator.
Also disclosed is a method of installing an actuator including preparing a valve assembly containing a valve and device housing and register device, the valve containing a valve cover and an actuator, the valve cover including a valve cover mechanical stop and orifice bore and the actuator including an actuator mechanical stop and interface portion; installing the actuator in the valve cover, the interface portion facing the orifice bore, a dimension defining a gap between the interface portion and the orifice bore; and tightening the actuator until the actuator mechanical stop engages with the valve cover mechanical stop, fixing the position of the actuator with respect to the valve cover, the actuator further capable of receiving signals to open and close the gap.
Various implementations described in the present disclosure may include additional systems, methods, features, and advantages, which may not necessarily be expressly disclosed herein but will be apparent to one of ordinary skill in the art upon examination of the following detailed description and accompanying drawings. It is intended that all such systems, methods, features, and advantages be included within the present disclosure and protected by the accompanying claims.
The features and components of the following figures are illustrated to emphasize the general principles of the present disclosure. Corresponding features and components throughout the figures may be designated by matching reference characters for the sake of consistency and clarity.
Disclosed is a valve meter assembly and associated methods, systems, devices, and various apparatus. It would be understood by one of skill in the art that the disclosed valve is described in but a few exemplary embodiments among many. No particular terminology or description should be considered limiting on the disclosure or the scope of any claims issuing therefrom.
A valve meter assembly in which the presently disclosed valve could be incorporated is shown in U.S. Patent Publication No. 2012-0305084, published Dec. 6, 2012, which is hereby incorporated by reference in its entirety.
One embodiment of a valve meter assembly 1000 is disclosed and described in
The device housing 110 of valve meter device 100 forms the main body through which water will flow, from inlet 310 to outlet 320. The threaded connections shown are inlet threaded portion 315 and outlet threaded portion 325, although other types of connections are considered within the scope of this disclosure. A valve cover 120 is attached to the device housing 110 using valve cover screws 130a-d (130c,d shown in
The valve meter device 100 includes a valve 170 and a meter 210 (shown in
The valve cover 120 and the valve portion 265 of the device housing 110 enclose a spring 250, a diaphragm assembly 260, and a support ring 263. The solenoid tamper cover 140 encloses a solenoid 270 onto the valve cover 120. The valve cover 120 includes a valve orifice cylinder 280 (shown in
The solenoid 270 is attached to the valve cover 120. The valve orifice cylinder 280 interacts with the solenoid 270 to change water flow through the media channel 520 (shown in
In various embodiments, the spring 250 may not be required for valve operation. Other parts of the valve 170, including the solenoid tamper cover 140 or support ring 263, may not be present in various embodiments of the valve meter device 100. The valve cover 120 and the valve portion 265 of the device housing 110 are screwed together to enclose the spring 250, support ring 263, and the diaphragm assembly 260 using valve cover screws 130a,b,c,d. Spring 250 may not be present in various embodiments.
Detailed embodiments of register device 2210 and wireless communication unit 2310 are described in aforementioned U.S. Patent Publication No. 2012-0305084.
The device housing 110 has an inlet 310 and an outlet 320 (shown in
To reduce head loss, the valve 170 (including the valve portion 265) and the meter 210 (placed in the meter portion 264) are oriented such that at least a portion of each of the valve 170 and the meter 210 touch an imaginary line drawn between the inlet 310 and the outlet 320 thereby forming an “in line” configuration. The “in line” configuration is not achieved by staggering valve 170 and the meter 210, as such staggering may result in unacceptable head loss. However, the “in line” configuration does not indicate that components of the valve meter device 100, including the meter 210 and valve 170, are located along the same horizontal plane. Should components or features, including the valve 170 and the meter 210, of the valve meter device 100 be staggered such that the components are not along the same horizontal plane, such a configuration typically is arranged to accommodate other requirements, such as an uneven piping system or multiple inlet or outlet configurations, and not to address the requirement of fitting the valve meter device 100 into a standard water meter lay-length.
The device housing 110 is dimensioned so that it can fit within a standard water meter lay-length. The standard water meter lay-length of a standard water meter is designated in various industry standards documents, including the American Water Works Association (AWWA). The AWWA C700 standard requires 7.5 inches standard water meter lay-length for meters with ⅝-inch piping diameter. Other AWWA standards, such as C708 and C710, also specify the same laying lengths for meters of like sizes. Additional details may be found in aforementioned U.S. Patent Publication No. 2012-0305084.
Also shown in
The meter portion 264 of the device housing 110 is sized to define a meter cavity 450. Although the current embodiment of the meter portion 264 is cylindrical, the meter portion 264 need not be a specific shape, but need only accommodate the meter 210. The thickness of each of the walls of meter portion 264 is sized to accommodate the water pressure of the piping system and may be variable or constant depending on the method of manufacture and other factors. The meter portion 264 also includes four threaded bottom plate attachment bores (not shown) for attachment of the bottom plate 150 with the bottom plate screws 160a,b,c,d.
The valve portion 265 includes four threaded valve cover bores 510a,b,c,d for attachment of the valve cover 120 to the valve portion 265 of the device housing 110. In the current embodiment, the valve cover 120 is attached using four valve cover screws 130a,b,c,d that attach through the valve cover 120 to each valve cover bore 510a,b,c,d. As noted above, the attachment could also be achieved using welding, which would obviate any need for valve cover bores 510a,b,c,d or valve cover screws 130a,b,c,d.
Shown in
Extending from the outer edge of solenoid attachment portion 820 toward the side of the valve cover including lug 875 and screw bore 810c is valve cover mechanical stop 6000. Valve cover mechanical stop 6000 includes top 6010, front 6020, and rear 6030 (shown in
As illustrated in the section view of the valve cover 120 in
The solenoid attachment portion 820 is dimensioned to define a solenoid chamber 940 between the solenoid 270 and the valve cover 120 when the solenoid 270 is attached to the valve cover 120. The valve cavity media channel 840 connects the valve cavity 905 with the solenoid chamber 940. Although the valve cavity media channel 840 is shown to connect with both the valve recess 910 and a valve bonnet 920 in the current embodiment, the valve cavity media channel 840 may connect to any portion of the valve cavity 905, including the valve recess 910. Because the valve cover media channel 830 is aligned with the center of the solenoid attachment portion 820 and solenoid attachment sink 825, the valve cover media channel 830 connects to the orifice bore 285 to the solenoid chamber 940. Valve orifice cylinder 280 is positioned at the top of valve cover media channel 830 and is part of valve cover 120 but may be manufactured as a separate component in other embodiments and added to valve cover 120 by welding, screwing, interference fit and any one of a number of other commonly available fastening methods
Also shown in
In various embodiments, the engagement of solenoid mechanical stop 3000 and/or valve cover mechanical stop 6000 may be accompanied by a tactile and/or audible snap or click or other sound to confirm for someone who is building or servicing or otherwise handling the valve 170 that engagement has occurred. This engagement of the solenoid mechanical stop 3000 and/or valve cover mechanical stop 6000 may also be accompanied by a visual indication. This visual indication may come in the form of front 3020 of solenoid mechanical stop 3000 physically interfacing with rear 6030 of valve cover mechanical stop 6000 such that no gap is visible between front 3020 and rear 6030.
In various embodiments, the engagement of the valve cover mechanical stops 1300a,b and/or solenoid mechanical stops 1320a,b may be accompanied by a tactile and/or audible snap or click or other sound to confirm for someone who is building or servicing or otherwise handling the valve 170 that engagement has occurred. This engagement of the valve cover mechanical stops 1300a,b and/or solenoid mechanical stops 1320a,b may also be accompanied by a visual indication. This visual indication may come in the form of valve cover mechanical stops 1300a,b physically moving into the solenoid mechanical stops 1320a,b. In the current embodiment, there is one orientation of the solenoid 270′ that will cause bend portions 1305a,b of valve cover mechanical stops 1300a,b to rest in an unbent, vertical position.
In various embodiments, the engagement of the solenoid mechanical stop 1500 and/or valve cover mechanical stop 1520 may be accompanied by a tactile and/or audible snap or click or other sound to confirm for someone who is building or servicing or otherwise handling the valve 170 that engagement has occurred. This engagement of the solenoid mechanical stop 1500 and/or valve cover mechanical stop 1520 may also be accompanied by a visual indication. This visual indication may come in the form of solenoid mechanical stop 1500 physically moving into the valve cover mechanical stop 1520. In the current embodiment, there is one orientation of the solenoid 270″ that will cause bend portion 1505 of solenoid mechanical stop 1500 to rest in an unbent, vertical position.
In various embodiments, the engagement of the solenoid mechanical stop 1720 and/or valve cover mechanical stop 1700 may be accompanied by a tactile and/or audible snap or click or other sound to confirm for someone who is building or servicing or otherwise handling the valve 170 that engagement has occurred. This engagement of the solenoid mechanical stop 1720 and/or valve cover mechanical stop 1700 may also be accompanied by a visual indication. This visual indication may come in the form of valve cover mechanical stop 1700 physically moving into the solenoid mechanical stop 1720. In the current embodiment, there is one orientation of the solenoid 270′″ that will cause bend portion 1705 of valve cover mechanical stop 1700 to rest in an unbent, horizontal position.
In an embodiment of the valve meter assembly 1000 (shown in
The valve orifice cylinder 280 provides the interface with the interface portion 1140 of the plunger 1130. The valve orifice cylinder 280 is chosen of an appropriate size to prevent excessive fluid flow, as excessive fluid flow will cause the diaphragm assembly 260 to lift away from the beveled edge 550 more quickly than may be desired.
In the current embodiment, the valve 170 is a pilot-operated valve. A pilot operated valve is a valve that experiences large-scale operation occurring naturally as a result of a small change in the pilot. As such, small amounts of energy can be used to control large-scale changes as the pilot changes. In the current embodiment, the pilot-operated valve is a diaphragm valve.
In use, the valve meter device 100 may assume one of two states: an “on” or “open” state and an “off” or “closed” state. A “trickle” or “reduced flow” state may be substituted for the “off” or “closed” state in various embodiments. The valve meter device 100 may be configured to assume either of the two possible states. The states correspond to the positioning of the valve 170.
The valve meter device 100 will typically be in the open state allowing a maximum, or near maximum, flow rate of water that is allowed to flow through the valve meter device 100. In one exemplary embodiment, maximum flow rate is about 25 gallons per minute, although other maximum flow rates are possible in accord with this disclosure. When the valve meter device 100 is in the open state, the valve 170 is open. When the valve 170 is open, which occurs when a flexible diaphragm of diaphragm assembly 260 is substantially lifted away from the beveled edge 550 (see
During application of valve meter assembly 1000, water travels through the valve meter device 100 originating from a water source and entering in inlet 310. Water is permitted to travel through the inlet opening 612, and to the horizontal portion 610. When water reaches the intersection of the horizontal portion 610 and vertical portion 620, water is directed vertically into the vertical portion 620 by water pressure. Water exits the vertical portion 620 by flowing over the beveled edge 550. Water fills the valve transition portion 670 and—as will be described in more detail later—the valve cavity 905 and a media channel pathway 2610, which extends from the valve cavity 905 to the valve outlet portion 340. Water exits the valve portion 265 via the valve outlet portion 340 and enters the meter portion 264. Water then enters and fills the meter cavity 450. Pressure forces water into the metering inlet (not shown), through the meter 210, and out of the metering outlet 213 to the outlet 320. Once the water exits the outlet 320, the water flows through the downstream piping system and, ultimately, to the user terminal.
In the current embodiment wherein meter 210 is a nutating disc displacement flow meter, the water passing through the meter 210 moves a nutating disc (not shown) causing a meter magnet (not shown) to rotate. The rotation of the meter magnet causes a register to log the motion, leading to a measurement of water usage and a readout of water usage from the register.
A register circuit (not shown) configured to log the readout of water usage at preset timing intervals may be included with one embodiment of the valve meter device 100. In the current embodiment, the register circuit remains in a low power mode for the majority of its operating life. Low power, as used in this disclosure, means that the register circuit is using a very small amount of power when compared to the normal operating mode. This is commonly referred to as being in a “sleep mode.” The register circuit “wakes up” at preset timing intervals to read the register and log the readout. In the current embodiment, a wireless communication unit circuit (not shown) is connected with the register circuit via wires 2360 (shown in
The valve 170 is configured in the open state when the interface portion 1140 is lifted away from the valve orifice cylinder 280 because the solenoid 270 is in the open position. The valve cavity media channel 840 provides a water pressure link between the solenoid chamber 940 and the valve cavity 905 such that the water pressure in the valve cavity 905 will be the same as the water pressure in the solenoid chamber 940. When the solenoid 270 is in the open position, the plunger 1130 is lifted so that the valve orifice cylinder 280 is open to the valve cover media channel 830. When the valve orifice cylinder 280 is uncovered, water is allowed to flow from the solenoid chamber 940 through the valve cover media channel 830 into the media channel 520 and further into the valve outlet portion 340. Therefore, the water pressure in the valve cavity 905 is substantially the same as the water pressure in the media channel 520, the solenoid chamber 940, the media channel 520, and the valve outlet portion 340. Thus, the diaphragm has no pressure behind it to close the valve 170. The valve 170 remains open. Although the current embodiment has the valve orifice cylinder 280 located on the valve cover media channel 830 such that there is a pressure link between the valve cavity 905 and the solenoid chamber 940, the valve orifice cylinder 280 may be located within the valve cavity media channel 840 in various embodiments. Other locations for the valve orifice cylinder 280 and orifice bore 285 are also contemplated by the current disclosure.
Changing the valve meter device 100 to a closed state requires the valve 170 to be changed to closed. Where a trickle state is included, the water supply valve must be changed to a trickle state, which may be the same as the closed state in various embodiments. This is accomplished by operation of the plunger 1130 moving into a closed position having the interface portion 1140 contacting the valve orifice cylinder 280, which provides a water-tight seal over the valve cover media channel 830. In the closed state, the valve meter device 100 allows no water flow through. In the trickle state, the valve meter device 100 allows minimal water flow through. In the current embodiment, the valve 170 is a diaphragm valve with a pressure-controlled pilot operation. To move the valve meter device 100 into the closed state, the solenoid 270 is engaged, or “thrown,” and closed onto the valve orifice cylinder 280. This closes or “severs” the media channel pathway 2610. Water flow is blocked from the solenoid chamber 940 to the valve cover media channel 830 as well as to the media channel 520 and media channel relief 530 thereby isolating the solenoid chamber 940, the valve cavity media channel 840, and the valve cavity 905 as one water pressure pool. Thus, the closing of the solenoid 270 is the pilot operation that triggers the dynamic state of the valve 170. The valve 170 is in the closed state when the interface portion 1140 of the plunger 1130 is in contact with the valve orifice cylinder 280 and the diaphragm assembly 260 has traveled and contacted the beveled edge 550, sealing the valve 170.
After the solenoid 270 is closed or thrown, water may no longer exit the valve cavity 905, so the valve cavity 905 no longer has media pressure behind it. Spring force provided from the diaphragm or from the optional spring 250 forces the diaphragm assembly 260 down toward the valve inlet portion 330 of the device housing 110. The spring 250 is optional because, depending on the configuration of the diaphragm, the diaphragm may already be biased toward closing the valve 170 without the spring 250. As the diaphragm assembly 260 moves toward the valve inlet portion 330, some of the water flowing through the valve portion 265 will leak through the diaphragm assembly 260 and into the valve cavity 905. The increased volume of water in the valve cavity 905 creates increased pressure in the valve cavity 905. The increased pressure in the valve cavity 905 is applied to the entire surface of the diaphragm because the valve cavity 905 extends across the entire diaphragm. This increased pressure applied over the entire diaphragm further biases the diaphragm assembly 260 in the direction of the valve inlet portion 330.
The increased bias causes the diaphragm assembly 260 to travel toward the valve inlet portion 330, eventually seating the bottom of an inner flat portion (not shown) of the diaphragm onto the beveled edge 550 of the top edge portion 640 of the valve inlet portion 330. When the diaphragm seats onto the beveled edge 550, the valve 170 is in the closed state.
Once the diaphragm has seated, water pressure from the valve inlet portion 330 equalizes with water pressure in the valve cavity 905 because water can pass into the valve cavity 905 through the valve cone 1210 of the diaphragm assembly 260 but cannot exit the valve cavity 905 down the media channel pathway 2610. With equalized pressure, the valve 170 remains in the closed state because the cross-section of the valve inlet portion 330 provides a smaller surface area over which to apply pressure to the diaphragm than the surface area of the diaphragm 1230 that interfaces with the valve cavity 905. With the same pressure, a smaller surface area over which the pressure is applied produces a smaller force than the same pressure applied to a larger surface area. The result is a net downward force on the diaphragm, maintaining the valve 170 in the closed state. The trickle state is accomplished by placing the diaphragm in the same position as the diaphragm 1230 is placed in the closed state. However, in the trickle state, a small amount of water is allowed to bypass the valve 170 via a leak passageway (not shown) defined in the diaphragm or a bypass channel (not shown) from the valve inlet portion 330 to the valve outlet portion 340. The bypass channel or leak passageway may be a small bore leading from the valve inlet portion 330 to the valve outlet portion 340 and may be placed in the vertical portion 620, for example. The bore would be small enough that a significant amount of water would not flow through the bore. A sealing valve may allow selective flow through the bore.
To reopen the valve 170, the solenoid 270 is actuated so that the interface portion 1140 lifts away from the valve orifice cylinder 280, opening the media channel pathway 2610. Opening the media channel pathway 2610 establishes a pressure link between all of the components of the media channel pathway 2610, including the valve cavity 905, the valve cavity media channel 840, the solenoid chamber 940, the valve cover media channel 830, the media channel relief 530, and the media channel 520. When the pressure in the valve cavity 905 is reduced, the downward force on the diaphragm and the diaphragm assembly 260 is also reduced. The pressure in the valve inlet portion 330 provides greater upward force on the bottom of the diaphragm than the downward force on the top of the diaphragm. This downward force may be provided by the spring 250 or by the inherent bias of the diaphragm. The result is a lifting of the diaphragm assembly 260, thereby opening the valve 170.
The solenoid 270 may be engaged or lifted by manual operation, by electronic actuation, or by remote control. In one embodiment, the wireless communication unit 2310 is capable of receiving electrical signals for the solenoid 270 to control its operation. Actuation of the plunger 1130 in the current embodiment is performed by a solenoid 270, which is a latching solenoid in the current embodiment. A latching solenoid is a solenoid 270 that latches in place. A latching solenoid does not utilize energy once it has achieved its desired position but does use energy to change positions. However, this actuation can be performed via a number of mechanical or electromechanical interfaces, including stepper motors, DC motors, non-latching solenoids, electromagnets and other electromagnetic devices, and spring assemblies, among others. This embodiment would allow a remotely located communicator to control operation of the valve 170, allowing the valve 170 to be changed to an open or closed state from a remote location.
One should note that conditional language, such as, among others, “can,” “could,” “might,” or “may,” unless specifically stated otherwise, or otherwise understood within the context as used, is generally intended to convey that certain embodiments include, while other embodiments do not include, certain features, elements and/or steps. Thus, such conditional language is not generally intended to imply that features, elements and/or steps are in any way required for one or more particular embodiments or that one or more particular embodiments necessarily include logic for deciding, with or without user input or prompting, whether these features, elements and/or steps are included or are to be performed in any particular embodiment.
It should be emphasized that the above-described embodiments are merely possible examples of implementations, merely set forth for a clear understanding of the principles of the present disclosure. Any process descriptions or blocks in flow diagrams should be understood as representing modules, segments, or portions of code which include one or more executable instructions for implementing specific logical functions or steps in the process, and alternate implementations are included in which functions may not be included or executed at all, may be executed out of order from that shown or discussed, including substantially concurrently or in reverse order, depending on the functionality involved, as would be understood by those reasonably skilled in the art of the present disclosure. Many variations and modifications may be made to the above-described embodiment(s) without departing substantially from the spirit and principles of the present disclosure. Further, the scope of the present disclosure is intended to cover any and all combinations and sub-combinations of all elements, features, and aspects discussed above. All such modifications and variations are intended to be included herein within the scope of the present disclosure, and all possible claims to individual aspects or combinations of elements or steps are intended to be supported by the present disclosure.
Number | Name | Date | Kind |
---|---|---|---|
691904 | Hallbergh | Jan 1902 | A |
1165429 | Mass | Dec 1915 | A |
1788618 | Cover | Jan 1931 | A |
1808209 | Earl | Jun 1931 | A |
1808212 | Earl | Jun 1931 | A |
2302529 | Cornell et al. | Nov 1942 | A |
3254660 | Ray | Jun 1966 | A |
3593957 | Dolter et al. | Jul 1971 | A |
3653261 | Feldman | Apr 1972 | A |
3672233 | Hjermstad | Jun 1972 | A |
3705385 | Batz | Dec 1972 | A |
3731534 | Painley et al. | May 1973 | A |
3795144 | Marchesi | Mar 1974 | A |
4093997 | Germer | Jun 1978 | A |
4120031 | Kincheloe et al. | Oct 1978 | A |
4291375 | Wolf | Sep 1981 | A |
4388690 | Lumsden | Jun 1983 | A |
4414633 | Churchill | Nov 1983 | A |
4442492 | Karlsson et al. | Apr 1984 | A |
4465970 | DiMassimo et al. | Aug 1984 | A |
4516213 | Gidden | May 1985 | A |
4542469 | Brandyberry et al. | Sep 1985 | A |
4591988 | Klima et al. | May 1986 | A |
4707852 | Jahr et al. | Nov 1987 | A |
4727900 | Dooling et al. | Mar 1988 | A |
4792946 | Mayo | Dec 1988 | A |
4803632 | Frew et al. | Feb 1989 | A |
4833618 | Verma et al. | May 1989 | A |
4868566 | Strobel et al. | Sep 1989 | A |
4881070 | Burrowes et al. | Nov 1989 | A |
4901751 | Story | Feb 1990 | A |
4940976 | Gastouniotis et al. | Jul 1990 | A |
4953403 | Springer | Sep 1990 | A |
4967996 | Sonoda et al. | Nov 1990 | A |
4989830 | Ratnik | Feb 1991 | A |
5056107 | Johnson et al. | Oct 1991 | A |
5075792 | Brown et al. | Dec 1991 | A |
5079715 | Venkataraman et al. | Jan 1992 | A |
5121344 | Laage et al. | Jun 1992 | A |
5239575 | White et al. | Aug 1993 | A |
5251480 | Brunson, IV et al. | Oct 1993 | A |
5267587 | Brown | Dec 1993 | A |
5298894 | Cerny et al. | Mar 1994 | A |
5381136 | Powers et al. | Jan 1995 | A |
5434911 | Gray et al. | Jul 1995 | A |
5438329 | Gastouniotis et al. | Aug 1995 | A |
5451938 | Brennan, Jr. | Sep 1995 | A |
5459459 | Lee, Jr. | Oct 1995 | A |
5481259 | Bane | Jan 1996 | A |
5493287 | Bane | Feb 1996 | A |
5519387 | Besier et al. | May 1996 | A |
5525898 | Lee et al. | Jun 1996 | A |
5553094 | Johnson et al. | Sep 1996 | A |
5590179 | Shincovich et al. | Dec 1996 | A |
5594740 | LaDue | Jan 1997 | A |
5594776 | Dent | Jan 1997 | A |
5617084 | Sears | Apr 1997 | A |
5631554 | Briese et al. | May 1997 | A |
5654692 | Baxter, Jr. et al. | Aug 1997 | A |
5666655 | Ishikawa et al. | Sep 1997 | A |
5673252 | Johnson et al. | Sep 1997 | A |
5708195 | Kurisu et al. | Jan 1998 | A |
5714931 | Petite | Feb 1998 | A |
5748104 | Argyroudis et al. | May 1998 | A |
5751797 | Saadeh | May 1998 | A |
5754101 | Tsunetomi et al. | May 1998 | A |
5767790 | Jovellana | Jun 1998 | A |
5787358 | Takahashi | Jul 1998 | A |
5801643 | Williams et al. | Sep 1998 | A |
5815086 | Ivie et al. | Sep 1998 | A |
5852658 | Knight et al. | Dec 1998 | A |
5877703 | Bloss et al. | Mar 1999 | A |
5892441 | Woolley et al. | Apr 1999 | A |
5892758 | Argyroudis | Apr 1999 | A |
5907491 | Canada et al. | May 1999 | A |
5924051 | Provost et al. | Jul 1999 | A |
5926103 | Petite | Jul 1999 | A |
5926531 | Petite | Jul 1999 | A |
5940009 | Loy et al. | Aug 1999 | A |
5963146 | Johnson et al. | Oct 1999 | A |
5963557 | Eng | Oct 1999 | A |
5971011 | Price | Oct 1999 | A |
5979863 | Lousberg | Nov 1999 | A |
5986573 | Franklin et al. | Nov 1999 | A |
5994892 | Turino et al. | Nov 1999 | A |
5996608 | Hunter et al. | Dec 1999 | A |
6006212 | Schleich et al. | Dec 1999 | A |
6028522 | Petite | Feb 2000 | A |
6028855 | Hirsch | Feb 2000 | A |
6031455 | Grube et al. | Feb 2000 | A |
6031466 | Leshets et al. | Feb 2000 | A |
6044062 | Brownrigg et al. | Mar 2000 | A |
6058374 | Guthrie et al. | May 2000 | A |
6060994 | Chen | May 2000 | A |
6069571 | Tell | May 2000 | A |
6081204 | Lavoie et al. | Jun 2000 | A |
6115677 | Perthold et al. | Sep 2000 | A |
6150955 | Tracy et al. | Nov 2000 | A |
6152173 | Makowan | Nov 2000 | A |
6163276 | Irving et al. | Dec 2000 | A |
6172616 | Johnson et al. | Jan 2001 | B1 |
6195018 | Ragle et al. | Feb 2001 | B1 |
6208266 | Lyons et al. | Mar 2001 | B1 |
6218953 | Petite | Apr 2001 | B1 |
6233327 | Petite | May 2001 | B1 |
6246677 | Nap et al. | Jun 2001 | B1 |
6249516 | Brownrigg et al. | Jun 2001 | B1 |
6288641 | Casais | Sep 2001 | B1 |
6317051 | Cohen | Nov 2001 | B1 |
6333975 | Brunn et al. | Dec 2001 | B1 |
6373399 | Johnson et al. | Apr 2002 | B1 |
6392538 | Shere | May 2002 | B1 |
6405047 | Moon | Jun 2002 | B1 |
6424270 | Ali | Jul 2002 | B1 |
6426027 | Scarborough et al. | Jul 2002 | B1 |
6430268 | Petite | Aug 2002 | B1 |
6437692 | Petite et al. | Aug 2002 | B1 |
6453247 | Hunaidi | Sep 2002 | B1 |
6456197 | Lauritsen et al. | Sep 2002 | B1 |
6470903 | Reyman | Oct 2002 | B2 |
6493377 | Schilling et al. | Dec 2002 | B2 |
6512463 | Campbell et al. | Jan 2003 | B1 |
6528957 | Luchaco | Mar 2003 | B1 |
6536469 | Dilger et al. | Mar 2003 | B2 |
6538577 | Ehrke et al. | Mar 2003 | B1 |
6560543 | Wolfe et al. | May 2003 | B2 |
6564159 | Lavoie et al. | May 2003 | B1 |
6568416 | Tucker | May 2003 | B2 |
6577961 | Hubbard et al. | Jun 2003 | B1 |
6618578 | Petite | Sep 2003 | B1 |
6618709 | Sneeringer | Sep 2003 | B1 |
6624750 | Marman et al. | Sep 2003 | B1 |
6628207 | Hemminger et al. | Sep 2003 | B1 |
6628764 | Petite | Sep 2003 | B1 |
6633781 | Lee et al. | Oct 2003 | B1 |
6653945 | Johnson et al. | Nov 2003 | B2 |
6657552 | Belski et al. | Dec 2003 | B2 |
6675071 | Griffin, Jr. et al. | Jan 2004 | B1 |
6677861 | Henry et al. | Jan 2004 | B1 |
6710721 | Holowick | Mar 2004 | B1 |
6747557 | Petite et al. | Jun 2004 | B1 |
6798352 | Holowick | Sep 2004 | B2 |
6816072 | Zoratti | Nov 2004 | B2 |
6836737 | Petite et al. | Dec 2004 | B2 |
6847300 | Yee et al. | Jan 2005 | B2 |
6880567 | Klaver | Apr 2005 | B2 |
6891838 | Petite et al. | May 2005 | B1 |
6912472 | Mizushina et al. | Jun 2005 | B2 |
6914533 | Petite | Jul 2005 | B2 |
6914893 | Petite | Jul 2005 | B2 |
6931445 | Davis | Aug 2005 | B2 |
6946972 | Mueller et al. | Sep 2005 | B2 |
6954701 | Wolfe | Oct 2005 | B2 |
6954814 | Leach | Oct 2005 | B1 |
6972677 | Coulthard | Dec 2005 | B2 |
6978210 | Suter et al. | Dec 2005 | B1 |
6980079 | Shintani et al. | Dec 2005 | B1 |
6982651 | Fischer | Jan 2006 | B2 |
7008239 | Ju | Mar 2006 | B1 |
7009530 | Zigdon et al. | Mar 2006 | B2 |
7012546 | Zigdon et al. | Mar 2006 | B1 |
7042368 | Patterson et al. | May 2006 | B2 |
7053767 | Petite et al. | May 2006 | B2 |
7054271 | Brownrigg | May 2006 | B2 |
7061924 | Durrant et al. | Jun 2006 | B1 |
7072945 | Nieminen et al. | Jul 2006 | B1 |
7079810 | Petite et al. | Jul 2006 | B2 |
7088239 | Basinger et al. | Aug 2006 | B2 |
7089125 | Sonderegger | Aug 2006 | B2 |
7099781 | Heidl et al. | Aug 2006 | B1 |
7103511 | Petite | Sep 2006 | B2 |
7111817 | Teti et al. | Sep 2006 | B2 |
7117051 | Landry et al. | Oct 2006 | B2 |
7123628 | Hwang | Oct 2006 | B1 |
7124184 | Chung et al. | Oct 2006 | B2 |
7137550 | Petite | Nov 2006 | B1 |
7142107 | Kates | Nov 2006 | B2 |
7228726 | Kates | Jun 2007 | B2 |
7248179 | Smit | Jul 2007 | B2 |
7248181 | Patterson et al. | Jul 2007 | B2 |
7250874 | Mueller et al. | Jul 2007 | B2 |
7256704 | Yoon et al. | Aug 2007 | B2 |
7263073 | Petite et al. | Aug 2007 | B2 |
7267014 | Winter | Sep 2007 | B2 |
7272635 | Longtin et al. | Sep 2007 | B1 |
7292143 | Drake et al. | Nov 2007 | B2 |
7295128 | Petite | Nov 2007 | B2 |
7301456 | Han | Nov 2007 | B2 |
7304587 | Boaz | Dec 2007 | B2 |
7315257 | Patterson et al. | Jan 2008 | B2 |
7342504 | Crane et al. | Mar 2008 | B2 |
7346030 | Cornwall | Mar 2008 | B2 |
7349766 | Rodgers | Mar 2008 | B2 |
7353280 | Chiles et al. | Apr 2008 | B2 |
7356614 | Kim et al. | Apr 2008 | B2 |
7363031 | Aisa | Apr 2008 | B1 |
7385524 | Orlosky | Jun 2008 | B1 |
7397907 | Petite | Jul 2008 | B2 |
7412882 | Lazar et al. | Aug 2008 | B2 |
7417557 | Osterloh et al. | Aug 2008 | B2 |
7423985 | Hill | Sep 2008 | B1 |
7424527 | Petite | Sep 2008 | B2 |
7443313 | Davis et al. | Oct 2008 | B2 |
7444401 | Keyghobad | Oct 2008 | B1 |
7453373 | Cumeralto et al. | Nov 2008 | B2 |
D583692 | Ball et al. | Dec 2008 | S |
7468661 | Petite et al. | Dec 2008 | B2 |
7478108 | Townsend et al. | Jan 2009 | B2 |
7480501 | Petite | Jan 2009 | B2 |
7526539 | Hsu | Apr 2009 | B1 |
7533693 | Colton et al. | May 2009 | B2 |
7549439 | Kimura et al. | Jun 2009 | B2 |
7604216 | Gebler | Oct 2009 | B2 |
7650425 | Davis | Jan 2010 | B2 |
7671480 | Pitchford et al. | Mar 2010 | B2 |
7690393 | Nagle et al. | Apr 2010 | B2 |
7694934 | Irwin | Apr 2010 | B2 |
7696940 | MacDonald | Apr 2010 | B1 |
7697492 | Petite | Apr 2010 | B2 |
7739378 | Petite | Jun 2010 | B2 |
7740024 | Brodeur | Jun 2010 | B2 |
7746246 | Salser | Jun 2010 | B2 |
7752309 | Keyghobad et al. | Jul 2010 | B2 |
7756086 | Petite | Jul 2010 | B2 |
7760703 | Kubler et al. | Jul 2010 | B2 |
7775422 | Winter et al. | Aug 2010 | B2 |
7783738 | Keyghobad et al. | Aug 2010 | B2 |
7792946 | Keyghobad et al. | Sep 2010 | B2 |
7806382 | Palumbo et al. | Oct 2010 | B1 |
7817063 | Hawkins et al. | Oct 2010 | B2 |
7825793 | Spillman et al. | Nov 2010 | B1 |
7843379 | Menzer et al. | Nov 2010 | B2 |
7870080 | Budike, Jr. | Jan 2011 | B2 |
7880641 | Parris et al. | Feb 2011 | B2 |
7962101 | Vaswani et al. | Jun 2011 | B2 |
7980317 | Preta et al. | Jul 2011 | B1 |
8014791 | Guigne et al. | Sep 2011 | B2 |
8109131 | Winter | Feb 2012 | B2 |
8140667 | Keyghobad et al. | Mar 2012 | B2 |
8249042 | Sparr et al. | Aug 2012 | B2 |
8300626 | Thubert et al. | Oct 2012 | B2 |
8351409 | Albert et al. | Jan 2013 | B2 |
8391177 | Picard | Mar 2013 | B2 |
8407333 | Keyghobad | Mar 2013 | B2 |
8549131 | Keyghobad et al. | Oct 2013 | B2 |
8660134 | Splitz | Feb 2014 | B2 |
8823509 | Hyland et al. | Sep 2014 | B2 |
8833390 | Ball et al. | Sep 2014 | B2 |
8855569 | Splitz et al. | Oct 2014 | B2 |
8931337 | Renoud | Jan 2015 | B2 |
8931505 | Hyland et al. | Jan 2015 | B2 |
9202362 | Hyland et al. | Dec 2015 | B2 |
20010010032 | Ehlers et al. | Jul 2001 | A1 |
20010013488 | Fukunaga et al. | Aug 2001 | A1 |
20010024163 | Petite | Sep 2001 | A1 |
20010048030 | Sharood et al. | Dec 2001 | A1 |
20020013679 | Petite | Jan 2002 | A1 |
20020019725 | Petite | Feb 2002 | A1 |
20020031101 | Petite | Mar 2002 | A1 |
20020051546 | Bizjak | May 2002 | A1 |
20020062392 | Nishikawa et al. | May 2002 | A1 |
20020067717 | Raschke et al. | Jun 2002 | A1 |
20020073183 | Yoon et al. | Jun 2002 | A1 |
20020089802 | Beckwith | Jul 2002 | A1 |
20020130768 | Che et al. | Sep 2002 | A1 |
20020159434 | Gosior et al. | Oct 2002 | A1 |
20020169643 | Petite et al. | Nov 2002 | A1 |
20020190956 | Klein et al. | Dec 2002 | A1 |
20030009515 | Lee et al. | Jan 2003 | A1 |
20030018733 | Yoon et al. | Jan 2003 | A1 |
20030018776 | Yoon et al. | Jan 2003 | A1 |
20030034900 | Han | Feb 2003 | A1 |
20030036810 | Petite | Feb 2003 | A1 |
20030046377 | Daum et al. | Mar 2003 | A1 |
20030074109 | Jeong et al. | Apr 2003 | A1 |
20030076241 | Middleton | Apr 2003 | A1 |
20030093484 | Petite | May 2003 | A1 |
20030107485 | Zoratti | Jun 2003 | A1 |
20030174070 | Garrod et al. | Sep 2003 | A1 |
20040010561 | Kim et al. | Jan 2004 | A1 |
20040054747 | Breh et al. | Mar 2004 | A1 |
20040129312 | Cuzzo et al. | Jul 2004 | A1 |
20040139210 | Lee et al. | Jul 2004 | A1 |
20040158333 | Ha et al. | Aug 2004 | A1 |
20040183687 | Petite et al. | Sep 2004 | A1 |
20050067022 | Istre | Mar 2005 | A1 |
20050072214 | Cooper | Apr 2005 | A1 |
20050078631 | Cornwell | Apr 2005 | A1 |
20050084418 | Hill et al. | Apr 2005 | A1 |
20050096753 | Arling | May 2005 | A1 |
20050104747 | Silic et al. | May 2005 | A1 |
20050121880 | Santangelo | Jun 2005 | A1 |
20050159823 | Hayes | Jul 2005 | A1 |
20050190784 | Stine | Sep 2005 | A1 |
20050195768 | Petite et al. | Sep 2005 | A1 |
20050195775 | Petite et al. | Sep 2005 | A1 |
20050201379 | Zhang et al. | Sep 2005 | A1 |
20050201397 | Petite | Sep 2005 | A1 |
20050203647 | Landry et al. | Sep 2005 | A1 |
20050246295 | Cameron | Nov 2005 | A1 |
20050251367 | Kahn et al. | Nov 2005 | A1 |
20060012491 | Mahowald | Jan 2006 | A1 |
20060028355 | Patterson et al. | Feb 2006 | A1 |
20060041655 | Holloway et al. | Feb 2006 | A1 |
20060046664 | Paradiso et al. | Mar 2006 | A1 |
20060098576 | Brownrigg et al. | May 2006 | A1 |
20060158347 | Roche et al. | Jul 2006 | A1 |
20060181414 | Bandy et al. | Aug 2006 | A1 |
20060201550 | Blyth et al. | Sep 2006 | A1 |
20060218266 | Matsumoto et al. | Sep 2006 | A1 |
20060273896 | Kates | Dec 2006 | A1 |
20060284784 | Smith et al. | Dec 2006 | A1 |
20070059986 | Rockwell | Mar 2007 | A1 |
20070063866 | Webb | Mar 2007 | A1 |
20070091825 | Budampati et al. | Apr 2007 | A1 |
20070284293 | Pitchford et al. | Dec 2007 | A1 |
20070293221 | Hwang et al. | Dec 2007 | A1 |
20070298779 | Wolman et al. | Dec 2007 | A1 |
20080030319 | McKenna et al. | Feb 2008 | A1 |
20080043637 | Rahman | Feb 2008 | A1 |
20080061769 | Junk et al. | Mar 2008 | A1 |
20080084260 | Swartzentruber | Apr 2008 | A1 |
20080086560 | Monier et al. | Apr 2008 | A1 |
20080095403 | Benhammou | Apr 2008 | A1 |
20080109090 | Esmaili et al. | May 2008 | A1 |
20080149180 | Parris et al. | Jun 2008 | A1 |
20080150750 | Parris | Jun 2008 | A1 |
20080169910 | Greene et al. | Jul 2008 | A1 |
20080186898 | Petite | Aug 2008 | A1 |
20080189056 | Heidl et al. | Aug 2008 | A1 |
20080195329 | Prince et al. | Aug 2008 | A1 |
20080240078 | Thubert | Oct 2008 | A1 |
20080281534 | Hurley | Nov 2008 | A1 |
20080291054 | Groft | Nov 2008 | A1 |
20090044628 | Lotscher | Feb 2009 | A1 |
20090058676 | Orlosky | Mar 2009 | A1 |
20090066524 | Yukawa et al. | Mar 2009 | A1 |
20090068947 | Petite | Mar 2009 | A1 |
20090121860 | Kimmel et al. | May 2009 | A1 |
20090133887 | Garcia et al. | May 2009 | A1 |
20090153357 | Bushman et al. | Jun 2009 | A1 |
20090188313 | Ball | Jul 2009 | A1 |
20090215424 | Petite | Aug 2009 | A1 |
20090243840 | Petite et al. | Oct 2009 | A1 |
20090255346 | Hendey et al. | Oct 2009 | A1 |
20090271045 | Savelle et al. | Oct 2009 | A1 |
20090287838 | Keyghobad et al. | Nov 2009 | A1 |
20090301571 | Ruhs | Dec 2009 | A1 |
20090309755 | Williamson et al. | Dec 2009 | A1 |
20090322453 | Kawaguchi | Dec 2009 | A1 |
20100017465 | Brownrigg et al. | Jan 2010 | A1 |
20100039984 | Brownrigg | Feb 2010 | A1 |
20100060479 | Salter | Mar 2010 | A1 |
20100156632 | Hyland et al. | Jun 2010 | A1 |
20100194582 | Petite | Aug 2010 | A1 |
20100250054 | Petite | Sep 2010 | A1 |
20100265909 | Petite et al. | Oct 2010 | A1 |
20100295672 | Hyland et al. | Nov 2010 | A1 |
20100312881 | Davis et al. | Dec 2010 | A1 |
20100329232 | Tubb et al. | Dec 2010 | A1 |
20110018762 | Walley et al. | Jan 2011 | A1 |
20110030482 | Meeusen et al. | Feb 2011 | A1 |
20110044276 | Albert et al. | Feb 2011 | A1 |
20110079402 | Darby et al. | Apr 2011 | A1 |
20110108136 | Margalit et al. | May 2011 | A1 |
20110140909 | Olson et al. | Jun 2011 | A1 |
20120007744 | Pal et al. | Jan 2012 | A1 |
20120068476 | Bradfield | Mar 2012 | A1 |
20120068477 | Bradfield | Mar 2012 | A1 |
20120106518 | Albert et al. | May 2012 | A1 |
20120271686 | Silverman | Oct 2012 | A1 |
20120305084 | Ball | Dec 2012 | A1 |
20130083722 | Bhargava et al. | Apr 2013 | A1 |
20130094537 | Hui et al. | Apr 2013 | A1 |
20130107772 | Splitz et al. | May 2013 | A1 |
20130109319 | Splitz et al. | May 2013 | A1 |
20130181848 | Picard | Jul 2013 | A1 |
20130214883 | Yano | Aug 2013 | A1 |
Number | Date | Country |
---|---|---|
2009308949 | May 2010 | AU |
2010249499 | May 2015 | AU |
2011265675 | May 2015 | AU |
2014259545 | Nov 2015 | AU |
2476119 | Feb 2005 | CA |
1185838 | Jun 1998 | CN |
19757581 | Jul 1998 | DE |
2305333 | Apr 1997 | GB |
62-295674 | Dec 1987 | JP |
05-253316 | Oct 1993 | JP |
06-223279 | Aug 1994 | JP |
6300606 | Oct 1994 | JP |
07-116285 | May 1995 | JP |
07231363 | Aug 1995 | JP |
H10-2744 | Jan 1998 | JP |
11-046254 | Feb 1999 | JP |
2000131179 | May 2000 | JP |
2000285356 | Oct 2000 | JP |
2002310840 | Oct 2002 | JP |
2002352361 | Dec 2002 | JP |
2005315663 | Nov 2005 | JP |
2005321935 | Nov 2005 | JP |
2006062414 | Mar 2006 | JP |
2006062716 | Mar 2006 | JP |
2006285645 | Oct 2006 | JP |
2007047139 | Feb 2007 | JP |
2008198044 | Aug 2008 | JP |
2010068017 | Mar 2010 | JP |
2012507090 | Mar 2012 | JP |
2012527706 | Nov 2012 | JP |
2013528732 | Jul 2013 | JP |
H5654124 | Nov 2014 | JP |
H11210028 | Nov 2014 | JP |
9810299 | Mar 1998 | WO |
9810394 | Mar 1998 | WO |
2008087911 | Jul 2008 | WO |
2009057214 | May 2009 | WO |
2010051287 | May 2010 | WO |
2010135587 | Nov 2010 | WO |
2011159403 | Dec 2011 | WO |
Entry |
---|
Vonroll Hydro—Hydrojournal, pp. 1-16, May 2008. |
English Translation: Vonroll Hydro—Hyrdojournal, Technology with a Future for Shut-off Systems—p. 4, VonRoll Hydro (shop) GmbH—New Concepts for Apprentice Training—p. 12, May 2008. |
Von Roll Hydro—Hydrojournal, pp. 1-16, Nov. 2008. |
English Translation: Von Roll Hydro—Hyrdojournal,VonRoll Hydroalert—Provides a Warning in the Event of Any Tampering with the Water Supply, p. 3, Nov. 2008. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 14/450,452, filed Aug. 4, 2014, mailed Feb. 17, 2016, 98 pgs. |
Keyghobad, Seyamak; Examiner Interview Summary Record for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Feb. 5, 2008; 2 pages. |
Keyghobad, Seyamak; Non-Final Rejection for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Oct. 26, 2007; 35 pages. |
Keyghobad, Seyamak; Requirement for Restriction/ Election for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Feb. 27, 2006; 17 pages. |
Splitz, David. E.; U.S Patent Application Entitled: Systems and Methods for Time-Based Hailing of Radio Frequency Devices assigned U.S. Appl. No. 13/283,526, filed Oct. 27, 2011, 51 pages. |
Splitz, David Edwin; Non-Final Office Action for U.S. Appl. No. 13/283,526, filed Oct. 27, 2011, mailed Jun. 18, 2013, 67 pgs. |
Splitz, David Edwin; Notice of Allowance for U.S. Appl. No. 13/283,526, filed Oct. 27, 2011, mailed Oct. 9, 2013, 16 pgs. |
Splitz, David Edwin; Issue Notification for U.S. Appl. No. 13/283,526, filed Oct. 27, 2011, mailed Feb. 5, 2014, 1 pg. |
Splitz, David E.; U.S. Patent Application Entitled: Systems and Methods for Dynamic Squelching in Radio Frequency Devices assigned U.S. Appl. No. 13/339,655, filed Dec. 29, 2011; 50 pgs. |
Splitz, David Edwin; Non-Final Office Action for U.S. Appl. No. 13/339,655, filed Dec. 29, 2011, mailed Sep. 16, 2013, 57 pgs. |
Splitz, David Edwin; Non-Final Office Action for U.S. Appl. No. 13/339,655, filed Dec. 29, 2011, mailed Mar. 5, 2014, 18 pgs. |
Splitz, David Edwin; Notice of Allowance for U.S. Appl. No. 13/339,655, filed Dec. 29, 2011, mailed May 23, 2014, 41 pgs. |
Splitz, David Edwin; Issue Notification for U.S. Appl. No. 13/339,655, filed Dec. 29, 2011, mailed Sep. 17, 2014, 1 pg. |
Splitz, David Edwin; U.S. Patent Application entitled: Systems and Methods for Recovering an Out-of-Service Node in a Hierarchical Network, U.S. Appl. No. 14/490,081, filed Sep. 18, 2014, 51 pgs. |
Dukes, Brent; U.S. Application entitled: Dynamic Routing in a Mesh Network, having U.S. Appl. No. 14/475,050, filed Sep. 2, 2014, 42 pgs. |
Splitz, David; International Search Report and Written Opinion for serial No. PCT/US11/58260, filed Oct. 28, 2011, mailed Feb. 7, 2012, 8 pgs. |
Splitz, David; PCT Application entitled: Systems and Methods for Time-Based Hailing of Radio Frequency having serial No. PCT/US11/058260, filed Oct. 28, 2011, 51 pgs. |
Splitz, David; International Preliminary Report on Patentability for serial No. PCT/US11/58260, filed Oct. 28, 2011, mailed May 8, 2014, 7 pgs. |
Splitz, David; International Search Report and Written Opinion for serial No. PCT/US12/22060, filed Jan. 20, 2012, mailed Mar. 29, 2012, 8 pgs. |
Splitz, David; PCT Application entitled: Systems and Methods for Dynamic Squelching in Radio Frequency Devices having serial No. PCT/US12/022060, filed Jan. 20, 2012, 50 pgs. |
Splitz, David; International Preliminary Report on Patentability for serial No. PCT/US12/22060, filed Jan. 20, 2012, mailed May 8, 2014, 6 pgs. |
Splitz, David Edwin; Extended European Search Report for serial No. 12844451.0, filed Jan. 20, 2012, mailed Apr. 21, 2015, 8 pgs. |
Splitz, David Edwin; U.S. Provisional Patent Application entitled: Automatic Discovery of Nodes in a Mesh Network, U.S. Appl. No. 61/779,892, filed Mar. 13, 2013; 110 pgs. |
Dukes, Brent; PCT Application entitled: Dynamic Routing in a Mesh Network having serial No. PCT/US15/44140, filed Aug. 7, 2015, 41 pgs. |
Dukes, Brent; International Search Report and Written Opinion for application No. PCT/US15/44140, filed Aug. 7, 2015, mailed Dec. 30, 2015, 15 pgs. |
Ball, Marty Scott; Issue Notification for U.S. Appl. No. 13/149,720, filed May 31, 2011, mailed Aug. 27, 2014, 1 pg. |
Ball, Marty Scott; Non-Final Office Action for U.S. Appl. No. 13/149,720, filed May 31, 2011, mailed Mar. 11, 2014, 75 pgs. |
Ball, Marty Scott; Notice of Allowance for U.S. Appl. No. 13/149,720, filed May 31, 2011, mailed Jun. 24, 2014, 29 pgs. |
Ball, Marty Scott; Supplemental Notice of Allowability for U.S Appl. No. 13/149,720, filed May 31, 2011, mailed Aug. 12, 2014, 4 pgs. |
Ball, Marty Scott; U.S. Patent Application Entitled: Valve Meter Assembly and Method under U.S. Appl. No. 13/149,720, filed May 31, 2011; 56 pgs. |
Ball, Marty Scott; U.S. Patent Application Entitled: Valve Meter Assembly and Method, U.S. Appl. No. 14/451,896, filed Aug. 5, 2014; 56 pgs. |
Hyland, Gregory E.; Applicant Initiated Interview Summary for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Feb. 18, 2014, 4 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Dec. 17, 2013, 54 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Apr. 10, 2013, 80 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Sep. 22, 2014, 49 pgs. |
Hyland, Gregory E.; Issue Notification for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Nov. 11, 2015, 1 pg. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Oct. 18, 2012; 44 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Apr. 18, 2014, 43 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Apr. 16, 2015, 47 pgs. |
Hyland, Gregory E.; Notice of Allowance for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Jul. 27, 2015, 19 pgs. |
Hyland, Gregory E.; Supplemental Notice of Allowability for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Oct. 13, 2015, 4 pgs. |
Hyland; U.S. Patent Application entitled: Infrastructure Monitoring System and Method, having U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, 30 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 12/606,957, filed Oct. 27, 2009, mailed Sep. 6, 2013; 53 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Feb. 11, 2014; 44 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed May 29, 2013, 71 pgs. |
Hyland, Gregory E.; Issue Notification for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Aug. 13, 2014. 1 pg. |
Hyland, Gregory E.; U.S. Continuation Application entitled: Infrastructure Monitoring System and Method having U.S. Appl. No. 14/928,725, filed Oct. 30, 2015, 28 pgs. |
Hyland, Gregory E; Non-Final Office Action for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Sep. 10, 2012, 35 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Sep. 24, 2013; 37 pgs. |
Hyland, Gregory E.; Notice of Allowance for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Apr. 23, 2014, 20 pgs. |
Hyland, Gregory E.; Supplemental Notice of Allowability for U.S. Appl. No. 12/784,300, filed May 20, 2010, mailed Aug. 1, 2014, 4 pgs. |
Hyland; U.S. Application entitled: Infrastructure Monitoring Devices, Systems, and Methods, having U.S. Appl. No. 12/784,300, filed May 20, 2010, 32 pgs. |
Hyland, Gregory E.; U.S. Continuation Application entitled: Infrastructure Monitoring Devices, Systems, and Methods, having U.S. Appl. No. 14/450,452, filed Aug. 4, 2014, 32 pgs. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002, mailed Oct. 8, 2008; 1 pg. |
Keyghobad, Seyamak; Requirement for Restriction/ Election for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Feb. 9, 2006; 11 pages. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 12/243,452, filed Oct. 1, 2008 mailed Jun. 16, 2010; 1 pg. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 12/490,867, filed Jun. 24, 2009, mailed Feb. 29, 2012; 1 pg. |
Keyghobad, Seyamak; Non Final Rejection for U.S. Appl. No. 12/490,867, filed Jun. 24, 2009, mailed Mar. 21, 2011; 9 pgs. |
Keyghobad, Seyamak; Non Final Rejection for U.S. Appl. No. 12/490,867, filed Jun. 24, 2009, mailed Oct. 4, 2010; 13 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,867, filed Jun. 24, 2006, mailed Sep. 7, 2011; 6 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,867, filed Jun. 24, 2009, mailed Nov. 2, 2011; 17 pgs. |
Keyghobad, Seyamak; U.S. Patent Application Entitled: Method and Apparatus for Inexpensively Monitoring and controlling Remotely Distributed Appliances under U.S. Appl. No. 12/490,867, filed Jun. 24, 2009; 33 pgs. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 12/490,925, filed Jun. 24, 2009; mailed Aug. 18, 2010; 1 pg. |
Keyghobad, Seyamak; Non-final office action for U.S. Appl. No. 12/490,925, filed Jun. 24, 2009; mailed Dec. 23, 2009; 17 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,925, filed Jun. 24, 2009, mailed Aug. 2, 2010, 8 pgs. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 12/490,957, filed Jun. 24, 2009; mailed Aug. 4, 2010; 1 pg. |
Keyghobad, Seyamak; U.S. Patent Application Entitled: Method and Apparatus for Inexpensively Monitoring and Controlling Remotely Distributed Appliances under U.S. Appl. No. 12/490,957, filed Jun. 24, 2009; 33 pgs. |
Keyghobad, Seyamak; Issue Notification for U.S. Appl. No. 13/372,408, filed Feb. 13, 2012, mailed Mar. 6, 2013, 1 pg. |
Keyghobad, Seyamak; Non-final Office Action for U.S. Appl. No. 13/372,408, filed Feb. 23, 2012; mailed May 25, 2012; 17 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 13/372,408, filed Feb. 13, 2012, mailed Jul. 27, 2012; 11 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 13/372,408, filed Feb. 13, 2012; mailed Nov. 1, 2012; 18 pgs. |
Keyghobad, Seyamak; Supplemental Notice of Allowance for U.S. Appl. No. 13/372,408, filed Feb. 13, 2012; mailed Aug. 2, 2012; 7 pgs. |
Keyghobad, Seyamak; U.S. Patent Application Entitled: Method and Apparatus for Inexpensively Monitoring and controlling Remotely Distributed Appliances under U.S. Appl. No. 13/372,408, filed Feb. 13, 2012; 34 pgs. |
Keyghobad, Seyamak, Issue Notification for U.S. Appl. No. 13/590,954, filed Aug. 21, 2012, mailed Sep. 11, 2013, 1 pg. |
Keyghobad, Seyamak; Non-Final Office Action for U.S. Appl. No. 13/590,954, filed Aug. 21, 2012, mailed Dec. 13, 2012; 39 pgs. |
Keyghobad,Seyamak; U.S. Patent Application entitled: Method and Apparatus for Inexpensively Monitoring and controlling Remotely Distributed Appliances under U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; 40 pages. |
Keyghobad, Seyamak; Non-Final Rejection for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed May 18, 2006; 13 pages. |
Keyghobad, Seyamak; Non-Final Rejection or U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Jun. 6, 2007; 32 pages. |
Keyghobad, Seyamak; Certificate of Correction for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Mar. 31, 2009; 1 page. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 10/298,300, filed Nov. 18, 2002; mailed Jul. 14, 2008; 4 pages. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/243,452, filed Oct. 1, 2008; mailed Mar. 22, 2010; 8 pages. |
Keyghobad, Seyamak; Examiner Interview Summary Record for U.S. Appl. No. 12/243,452, filed Oct. 1, 2008; mailed Dec. 7, 2009; 3 pages. |
Keyghobad, Seyamak; Non-Final Rejection for U.S. Appl. No. 12/243,452, filed Oct. 1, 2008; mailed Sep. 14, 2009; 8 pages. |
Keyghobad,Seyamak; Non-Final Rejection for U.S. Appl. No. 12/243,452, filed Oct. 1, 2008; mailed May 1, 2009; 4 pages. |
Keyghobad, Seyamak; U.S. Patent Application Entitled: Method and Apparatus for Inexpensively Monitoring and controlling Remotely Distributed Appliancesl under U.S. Appl. No. 12/243,452, filed Oct. 1, 2008; 33 pages. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,925, filed Jun. 24, 2009; mailed Jul. 19, 2010; 8 pages. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,925, filed Jun. 24, 2009; mailed Jun. 28, 2010; 10 pgs. |
Keyghobad, Seyamak; U.S. Patent Application Entitled: Method and Apparatus for Inexpensively Monitoring and Controlling Remotely Distributed Appliances under serial U.S. Appl. No. 12/490,925, filed Jun. 24, 2009; 33 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 12/490,957, filed Jun. 24, 2009; mailed Jun. 24, 2010; 10 pgs. |
Keyghobad,Seyamak; Non-Final Rejection for U.S. Appl. No. 12/490,957, filed Jun. 24, 2009; mailed Dec. 23, 2009; 17 pgs. |
Young et al. “Real-Time Intranet-Controlled Virtual Instrument Multiple-Circuit Power Monitoring,” IEEE Transactions on Instrumentation and Measurement, Jun. 2000. vol. 49, No. 3, p. 570. [Accessed Dec. 29, 2011] http://ieeexplore.ieee.org/xpls/abs—all.jsp? |
De Almeida et al. “Advanced Monitoring Technologies for the Evaluation of Demand-Side Management Programs,” IEEE Transactions on Power Systems, Aug. 1994. vol. 9, No. 3. [Accessed Dec. 29, 2011] http://ieeexplore.ieee.org/xpls/abs—all.jsp?arnumber=336086. |
Dolezilek. “Microprocessor Based Relay Information Improves the Power System,” Rural Electric Power Conference, May 1999. p. B5/1-B5/9. [Accessed Dec. 29, 2011] http://ieeexplore.ieee.org/xpls/abs—all.jsp?arnumber=768685. |
Gehami et al. “Electronic Control System I Salient Feature in Substation,” Transmission & Distrubition, Mar. 1991. vol. 43, No. 3, p. 48. [Accessed Dec. 29, 2011—ProQuest]. |
Horlent. “New Metering and Reading Techniques Based on a Modular Design Concept,” 10th International Conference on Electricity Distribution, May 1989. vol. 5, p. 455-459. [Accessed Dec. 29, 2011—IEEExplore]. |
“In Brief,” Land Mobile Radio News, Jan. 16, 1998. vol. 52, No. 3, p. 1. [Accessed Dec. 29, 2011—ProQuest] http://proquest.umi.com/pqdweb?did=25435781&sid=1&Fmt=3&clientld=31810&RQT=309&VName%20=PQD. |
“Landis & Gyr Utilities: Service Partnership Helps Utilities Use Available Resources More Effectively,” www.landisgyr.com/utilities/e/fr—press1—e.htm (archived Feb. 6, 1998) http://web.archive.org/web/19980206060801/http://www.landisgyr.com/utilities. |
Tamarkin. “Automated Meter Reading”, Sep.-Oct. 1992, vol. 50, No. 5/ [Accessed Dec. 29, 2011] http://www.usclcorp.com/news/Automatic—Power reading.pdf. |
ANSI; “Protocol Specification for ANSI Type 2 Optical Port”, American National Standard, ANSI C.12.18-2006, 11 pgs. |
Federal Communications Commission; “Understanding the FCC Regulations for Low-Power, Non-Licensed Transmitters”, Office of Engineering and Technology; Oct. 1993; 34 pgs. |
Semtech; “TN1200.4, Calculating Radiated Power and Field Strength for Conducted Power Measurements”, Semtech Corporation, Camarillo, CA, 2007, 9 pgs. |
RFM; “HX 2000 Datasheet: 916.5 MHz: Hybrid Transmitter”, RF Monolithics, Inc., Dallas, TX, USA, 1998; 2 pgs. |
General Electric; “GEH-5081 kV Meter Product Manual”, Nov. 1997, 137 pgs. |
General Electric; “kV RSX—RS232/RS485 Communications Options: Instructions Manual”; Mar. 1999, 33 pgs. |
Orfield; “Badger® Orion® System Helps Lemmon, South Dakota Reduce Read Time, Billing Cycles”, Badger connect Publication, 2004, 2 pgs. |
AMCO; “Pit Water-Meter Transponder (PWT)”; AMCO Automated Systems, LLC; PDB-14611; Sep. 2002; 2 pgs. |
AMCO; “Short-Range Programmer (SRP) VRT”; AMCO Automated Systems, LLC; PDB-14555.1; Sep. 2002; 2 pgs. |
AMCO; U.S. Appl. NRemote Water-Meter Transponder (RWT); AMCO Automated Systems, LLC; PDB-14610; Sep. 2002; 2 pgs. |
Article entitled: “Remote Meter Reading”, http://www.meter.co.uk/RMR.html; accessed on Jul. 30, 2012, 2 pgs. |
Article entitled: “Datamatic, Badger Connect for AMR Solutions”, http://www.datamatic.com/badger—partnership.html; accessed on Jul. 27, 2012, 1 pg. |
Article entitled: “OET Exhibits List”, https://apps.fcc.gov/oetcf/eas/reports/ViewExhibitReport.cfm? mode=Exhibits&RequestTimeout=500&calledFromFrame=N&application—id=194044&fcc—id=; Feb. 20, 2001, 2 pgs. |
Patterson, Tim; Request for Ex Parte Reexamination under U.S. Appl. No. 90/012,468, filed Sep. 6, 2012; 52 pgs. |
Patterson, Tim; Request for Ex Parte Reexamination under U.S. Appl. No. 90/012,449, filed Aug. 23, 2012; 51 pgs. |
Radix Corporation; “Automatic Meter Reading”, 2 pgs. |
Transparent Techcnologies; “Model M1A: Utility Radio Transmitter; M1A Operating Instructions”; 7 pgs. |
Trace; “Pit Water—Meter Transponder”; User Guide; Jan. 2003 16 pgs. |
Keyghobad, Seyamak; U.S. Patent Application entitled: Method and Apparatus for Inexpensively Monitoring and Controlling Remotely Distributed Appliances for U.S. Appl. No. 13/590,954, filed Aug. 21, 2012, 35 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 13/590,954, filed Aug. 21, 2012, mailed Mar. 21, 2013, 22 pgs. |
Keyghobad, Seyamak; Notice of Allowance for U.S. Appl. No. 13/590,954, filed Aug. 21, 2012, mailed Jul. 9, 2013, 21 pgs. |
Antenna. Merriam-Webster Dictionary, 2014 [retrieved on Jun. 1, 2014]. Retrieved from the Internet: <URL: www.merriam-webster.com/dictionary/antenna>. |
Hyland, Gregory E., Non-Final Office Action for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Jul. 31, 2013; 57 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Feb. 20, 2014; 29 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Jun. 5, 2014, 29 pgs. |
Hyland; U.S Patent Application Entitled: Infrastructure Monitoring Devices, Systems and Methods under U.S. Appl. No. 13/101,235, filed May 5, 2011; 28 pgs. |
Ball, Marty Scott; Non-Final Office Action for U.S. Appl. No. 14/451,896, filed Aug. 5, 2014, mailed Mar. 18, 2016, 98 pgs. |
Ball, Marty Scott; Non-Final Office Action for U.S. Appl. No. 14/451,896, filed Aug. 5, 2014, mailed May 12, 2016; 15 pages. |
Hyland, Gregory; Non-Final Office Action for U.S. Appl. No. 14/848,676, filed Sep. 9, 2015, mailed Mar. 4, 2016, 94 pgs. |
Hyland, Gregory E.; Notice of Allowance for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Sep. 11, 2014, 11 pgs. |
Hyland, Gregory E.; Supplemental Notice of Allowability for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Nov. 25, 2014, 5 pgs. |
Hyland, Gregory E.; Issue Notification for U.S. Appl. No. 13/101,235, filed May 5, 2011, mailed Dec. 23, 2014, 1 pg. |
Hyland, Gregory E.; U.S. Patent Application entitled: Infrastructure Monitoring Devices, Systems, and Methods having U.S. Appl. No. 14/557,754, filed Dec. 2, 2014, 28 pgs. |
Hyland, Gregory E.; Non-Final Office Action for U.S. Appl. No. 14/557,754, filed Dec. 2, 2014, mailed Jan. 19, 2016, 101 pgs. |
Hyland, Gregory; U.S. Continuation Application entitled: Infrastructure Monitoring Devices, Systems, and Methods having U.S. Appl. No. 14/848,676, filed Sep. 9, 2015, 29 pgs. |
Ball, Marty Scott; Canadian Office Action for serial No. 2,777,585, filed Jan. 19, 2009, mailed Oct. 21, 2013, 2 pgs. |
Ball, Marty Scott; Mexico Office Action for serial No. MX/a/2012/006097, filed May 25, 2012, mailed Mar. 19, 2015, 3 pgs. |
Hyland; PCT Application entitled: Infrastructure Monitoring System and Method having serial No. PCT/US09/62247, filed Oct. 27, 2009, 30 pgs. |
Hyland; International Preliminary Report on Patentability for serial No. PCT/US2009/062247, filed Oct. 27, 2009, 'mailed May 3, 2011, 7 pgs. |
Ball, Marty Scott; Mexico Office Action for serial No. MX/a/2012/006097, filed May 25, 2012, mailed May 26, 2015, 5 pgs. |
International Search Report for serial No. PCT/US2009/062247, filed on Oct. 27, 2009, mailed Dec. 18, 2009, 2 pgs. |
Hyland, Gregory E.; Canadian Office Action for serial No. 2,741,843, filed Oct. 27, 2009, mailed Dec. 8, 2015, 5 pgs. |
Hyland, Gregory;Mexico Office Action for serial No. MX/a/2011/004330, filed Apr. 25, 2011, mailed Mar. 21, 2013, 7 pgs. |
Hyland, Gregory; Mexico Office Action for serial No. MX/a/2011/004330, filed Apr. 25, 2011, mailed Jul. 18, 2013, 6 pgs. |
Hyland, Gregory; Mexico Office Action for serial No. MX/a/2011/004330, filed Apr. 25, 2011, mailed Oct. 3, 2013, 6 pgs. |
Hyland; European Examination Report for serial No. EP09824079.9, filed Oct. 27, 2009, mailed Nov. 13, 2015; 6 pgs. |
Hyland; European Search Report for serial No. EP09824079.9, filed Oct. 27, 2009, mailed May 8, 2012; 38 pages. |
Hyland, Gregory; Australian Patent Examination Report for serial No. 2009308949, filed Oct. 27, 2009, mailed Nov. 12, 2013, 3 pgs. |
Hyland, Gregory E.;Japanese Office Action for serial No. 2011-533427, filed Oct. 27, 2009, mailed Apr. 30, 2013, 15 pgs. |
Hyland, Gregory E.; Japanese Office Action for serial No. 2011-533427, filed Oct. 27, 2009, mailed Feb. 4, 2014, 50 pgs. |
Hyland, Gregory E.; Decision of Rejection for Japanese serial No. 2011-533427, filed Oct. 27, 2009, mailed Sep. 16, 2014, 4 pgs. |
Hyland, Gregory E.; Australian Examination Report for serial No. 2014259545, filed Oct. 27, 2009, mailed Jun. 10, 2015; 2 pgs. |
Hyland; PCT Application entitled: Infrastructure Monitoring Devices, Systems, and Methods having serial No. PCT/US2010/035666, filed May 20, 2010; 31 pgs. |
Hyland; International Search Report and Written Opinion for serial No. PCT/US2010/035666, filed May 20, 2010, mailed Jul. 16, 2010, 7 pgs. |
Hyland; International Preliminary Report on Patentability for serial No. PCT/US2010/035666, filed May 20, 2010, mailed Nov. 22, 2011, 6 pgs. |
Hyland, Gregory E.; Mexico Office Action for serial No. MX/A/2011/012383, filed May 20, 2010, mailed Oct. 8, 2012, 3 pgs. |
Hyland, Gregory E.; Mexico Office Action for serial No. MX/A/2011/012383, filed May 20, 2010, mailed May 9, 2013, 8 pgs. |
Hyland, Gregory E.; Mexico Office Action for serial No. MX/A/2011/012383, filed May 20, 2010, mailed Sep. 3, 2013, 10 pgs. |
Hyland, Gregory E.; Mexico Final Office Action for serial No. MX/A/2011/012383, filed May 20, 2010, mailed Jan. 9, 2014, 9 pgs. |
Hyland, Gregory E.; Australian Patent Examination report for serial No. 2010249499, filed Nov. 17, 2011, mailed Jun. 16, 2014, 5 pgs. |
Hyland, Gregory E.; Australian Patent Examination report for serial No. 2010249499, filed Nov. 17, 2011, mailed Nov. 21, 2014, 5 pgs. |
European Search Report for serial No. EP2433440, filed Nov. 18, 2011, mailed Nov. 28, 2012, 6 pgs. |
Hyland, Gregory; Japanese Office Action for serial No. 2012-512048, filed May 20, 2010, mailed Oct. 22, 2013, 51 pgs. |
Hyland, Gregory; Decision of Rejection for Japanese serial No. 2012-512048, filed May 20, 2010, mailed Apr. 22, 2014, 10 pgs. |
Hyland, Gregory; Mexico Office Action for serial No. MX/a/2012/015236, filed Dec. 19, 2012, mailed Jun. 13, 2013, 4 pgs. |
Hyland, Gregory; Mexico Office Action for serial No. MX/a/2012/015236, filed Dec. 19, 2012, mailed Oct. 3, 2013, 8 pgs. |
Hyland; U.S. Provisional Patent Application entitled: Infrastructure Monitoring Devices, Systems, and Methods, having U.S. Appl. No. 61/355,468, filed Jun. 16, 2010; 31 pgs. |
Hyland, Gregory; Mexico Office Action for serial No. MX/a/2012/015236, filed Dec. 19, 2012, mailed Dec. 3, 2013, received by foreign associate on Jan. 9, 2014, 4 pgs. |
Hyland, Gregory E.; Australian Patent Examination report for serial No. 2015202223, filed May 20, 2010, mailed Nov. 4, 2015, 4 pgs. |
Hyland; International Search Report and Written Opinion for serial No. PCT/US2011/035374, filed May 5, 2011, mailed Sep. 13, 2011; 7 pgs. |
Hyland; PCT Application Entitled: Infrastructure Monitoring Devices, Systems, and Methods having serial No. PCT/US11/35374, filed May 5, 2011, 28 pgs. |
Hyland; International Preliminary Report on Patentability for serial No. PCT/US2011/035374, filed May 5, 2011, mailed Dec. 19, 2012; 5 pgs. |
Hyland, Gregory E.; Australian Patent Examination Report for serial No. 2011265675, filed Jan. 21, 2012, mailed Oct. 1, 2014, 3 pgs. |
Hyland, Gregory E.; Japanese Office Action for serial No. 2013515338, filed Jan. 30, 2012, mailed Jun. 10, 2014, 8 pgs. |
Hyland, Gregory E.; Japanese Office Action for serial No. 2014-234642, filed May 5, 2011, mailed Jul. 7, 2015, 9 pgs. |
Hyland, Gregory E.; Japanese Office Action for serial No. 2014-234642, filed May 5, 2011, mailed Nov. 4, 2015,9 pgs. |
Hyland; U.S. Provisional Patent Application entitled: Water Supply Infrastructure Monitoring System and Method, having U.S. Appl. No. 61/108,770, filed Oct. 27, 2008, 11 pgs. |
Hyland; U.S. Provisional Patent Application entitled: Water Supply Infrastructure Monitoring System and Method, having U.S. Appl. No. 61/180,600, filed May 22, 2009, 14 pgs. |
Hyland, Gregory E.; Final Office Action for U.S. Appl. No. 14/557,754, filed Dec. 2, 2014, mailed Jun. 30, 2016, 24 pgs. |
Hyland, Gregory E.; Mexico Office Action for serial No. MX/a/2012/000347, filed May 5, 2011, mailed May 30, 2016, 4 pgs. |
Dukes, Brent; Non-Final Office Action for U.S. Appl. No. 14/475,050, filed Sep. 2, 2014, mailed May 19, 2016, 119 pgs. |
Hyland, Gregory E.; Canadian Office Action for serial No. 2,741,843, filed Oct. 27, 2009, mailed Jul. 22, 2016, 5 pgs. |
Number | Date | Country | |
---|---|---|---|
20150323091 A1 | Nov 2015 | US |