A field method, device and system are provided for quantitatively and rapidly determining the status of a variety of significant chemical constituents and parameters associated with mixed streams of liquid and gas including, for example, geothermal steam and geothermal steam condensate. As will be appreciated by those skilled in the art, the method, apparatus and system disclosed herein may be adapted for analyzing mixed liquid and vapor compositions in other industrial and environmental contexts including, for example, aqueous vapors and condensates generating in connection with chemical manufacturing, petroleum refining and synthetic fuel production.
Geothermal steam and/or its associated condensate typically include a variety of constituents and exhibit a range of parameters that are of varying degrees of concern in the context of exploitation of geothermal energy. Conventional geothermal installations will include procedures for the quantitative monitoring of some or all of these constituents and parameters including, for example, the aggregate amount of gases remaining in the gas phase in relation to the amount of liquid condensate after the steam is condensed (the non-compressible gas ratio (NCG Ratio); the conductivity of the condensate, the pH of the condensate, the aggregate amount of relatively insoluble and unreactive gases (typically including nitrogen, argon, methane and hydrogen) as well as more soluble and/or reactive gases (typically carbon dioxide, hydrogen sulfide, ammonia and hydrogen chloride).
Conventional practice in the geothermal industry involves condensing a sample of the steam on site using a small condenser (typically a coil of stainless steel tubing in a bucket of water and ice) with the resulting two phase samples being collected in suitable sample bombs. These sample bombs are, in turn, sent to one or more on-site and/or off-site laboratories for the requisite chemical analysis. The associated collection, transportation, testing and reporting activities can significantly delay receipt of the test data by field operations personnel. Indeed, in some instances complete analytical results may not be routinely available for weeks or even months.
The delays in obtaining the desired analytical data from the steam analysis can severely limit the ability to field personnel to determine if a good sample has been collected or if another sample needs to be collected. These delays in receiving analytical data can be particularly problematic with regard to samples taken while wells are being drilled or tested given the unpredictable nature of the resulting mixed streams. As will be appreciated, substantially “real time” analytical results would be quite beneficial in guiding drilling operations or determining the results of operational testing and will tend to reduce the costs associated with remote testing.
For example, the real time analysis of the mixed stream of condensate and a gas phase primarily comprising a mixture of noncondensable gas (NCG) ejected from the condenser and produced by a steam-cycle geothermal power plant will be particularly beneficial in monitoring atmospheric hydrogen sulfide emissions and the performance of hydrogen sulfide abatement efforts. As will be appreciated, the real time analysis will be helpful in monitoring the performance and/or abatement efforts with respect to volatile and/or soluble constituents in streams associated with other chemical processes.
Disclosed are an integrated analytical system and a method of operating the system to obtain improved control over processes that involve or produce mixed streams of gas and liquid or streams of condensable and non-condensable gases. The systems include an intake module configured for receiving and/or generating a mixed stream of liquid and gas, at least one pretreatment module configured for mixing a reagent with the mixed stream to form a treated stream, at least one gas purge module for stripping the treated stream with a first purge gas to separate the treated stream into a liquid stream and a gas stream, a liquid level cell and an associated controller for receiving the liquid stream and maintaining a substantially constant volume within the cell, parametric test modules for receiving and analyzing the liquid stream before and/or after the liquid stream has passed through the liquid level cell, a parametric test module for receiving and analyzing the gas stream, and a data acquisition element for collecting parametric test data from the parametric test modules.
The disclosure will be more fully understood from the following detailed description, taken in connection with the accompanying drawings, in which:
It should be noted that these Figures are intended to illustrate the general characteristics of methods, apparatus and/or systems utilized in certain example embodiments and to supplement the written description provided below. These drawings are not, however, to scale and may not precisely reflect the precise structural or performance characteristics of any given element or embodiment, and, accordingly, should not be interpreted as defining or limiting the range of methods, apparatus and systems that are fairly encompassed by the present disclosure.
A first embodiment of an integrated analytical apparatus is illustrated in
As noted above, one or more of the reagents that may be introduced into the sample stream may include acidic solutions necessary for performing one or more of the desired analytic tests. A dilute acidic solution, for example a 0.01M sulfamic acid containing an effective amount of a visual pH indicator solution, may be utilized in titrating ammonia present in the condensate fraction of the sample stream. One pH indicator solution that may be used in the dilute acidic solution comprises a mixture of bromocresol green and methyl red which provides a sharp transition point at about pH 4.2 that is useful in an ammonia titration. Similarly, a more concentrated acidic solution, for example a 0.1M sodium hydrogen sulfate solution, may be used for acidifying the condensate fraction of the sample stream in connection with a sulfide titration. This more concentrated acidic solution also preferably contains a quantity (preferably about 200 mg/L) of soluble starch sufficient to provide a visible endpoint for the iodometric titration of sulfide.
As noted above, in addition to the acidic solutions, other reagents can be introduced into the sample stream as necessary to condition the stream for subsequent analysis. For example, in connection with a iodometric titration of sulfide, a second reagent solution, for example a 0.005M solution of iodine (0.010N) that also contains a quantity of potassium iodide sufficient to keep the iodine in solution, may be used in conjunction with a sodium hydrogen sulfate solution or other suitable solution for analyzing the sulfide concentration in the sample stream. As will be appreciated by those skilled in the art, the range of reagants and the particular configuration of the apparatus illustrated in
As noted above the configuration of the apparatus illustrated in
As illustrated in
One example configuration for the reaction coil comprises at least two feet (61 cm) of ⅛ inch ID (3.2 mm) plastic tube (preferably polyethylene) with a turbulence promoting element inside. The turbulence promoting element can be made by threading a length of 0.062 inch (1.57 mm) stainless steel welding rod through an 8 inch (20.3 cm) length of a ⅛ inch (3.2 mm) OD continuous stainless steel spring, and stretching and deforming the spring to at or near its maximum possible length with the welding rod inside. At least one end of the stretched spring can then secured to the welding rod by, for example, spot-welding to form a stable turbulence promoting structure that can then be inserted into the plastic tube. As will be appreciated by those skilled in the art, the necessary mixing may be achieved with configurations in addition to the reaction coil including, for example, small vessels or channels that are stirred or otherwise agitated to sufficient degree.
The mixture of the sample stream (including both condensate and gas components) and the titrating reagent(s) then exit the reaction coil and enters the gas stripping column 24 through tube or flow path 28 that may comprise, for example, a small diameter tube such as preferably ⅛ inch (3.2 mm) OD stainless steel tubing which extends a short distance into the packed part 26 of the column in order to reduce carryover of spray.
As will be appreciated, the gas stripping column 24 may be utilized in a variety of configurations depending on a number of variable including, for example, flow rates, packing materials, differential pressure and reaction times. One configuration that is considered suitable for practicing the disclosure methods includes a larger diameter upper part, preferably about 0.75 inch (1.9 cm) ID, about 8 inches (20.3 cm) long. It also preferred that the gas stripping column be transparent or be provided with longitudinally arranged transparent regions so that the performance of the column may be observed for flooding or other undesirable conditions. The packing 26 provided within the gas stripping column 24 may comprise a plurality of white ceramic balls of diameter about 0.2 inches (5.1 mm). A digital pressure-vacuum gauge 29 may be provided near the top of the gas stripping column, preferably electronic with visual display and analog output, to allow further monitoring of the performance of the column.
Gas reaching the top or head of the gas stripping column 24 may be withdrawn through gas discharge line 30, a point that is preferably located well above the point at which the mixture of sample and reagent enters the column and may be particularly configured for reducing the passage of entrained spray into the discharge line. The purge gas source(s) 32 available for introduction into the gas stripping column may include air provided through a small air pump and gas flow pulsation damper (not shown) (suitable for iodometric titration of sulfide) and/or cylinder(s) of argon and/or another inert gas(es) can also be used (preferred when potentiometric titration of sulfide is utilized). The purge gas may be metered into the gas stripping column 24 through one or more gas flow meters 34 and may be feed into the top, bottom or both of the gas stripping column as determined by the settings of valves 38A, 38B. The gas flow meter may, for example, comprise a thermal flow meter (which gives a response proportional to moles of gas weighted by the molar heat capacity of the gas) capable of a flow range of 0-1,000 ml/min.
The purge gas can then be fed into the lower portion or gas purge chamber 36 of the gas stripping column 24 through a gas dispersion frit 36A, preferably sintered glass or sintered stainless steel, or other nozzle or eductor. The gas purge chamber 36 may comprise a segment of transparent glass or plastic tubing ⅝ inch (15.9 mm) ID and about 2 inches (5.1 cm) long, thereby allowing the color of the mixture of the condensate portion of the sample stream and reagent solution(s) 40 to be visually monitored as it is stripped of dissolved gases. The temperature of the solution may be monitored using a temperature probe 37, for example, a resistance temperature detection (RTD) element. As illustrated in
As illustrated in
As illustrated in
Pump controller 54 is configured to respond to the conductance measured by the conductivity meter 50 and adjust the flow rate through pump 12A in order to maintain a substantially constant volume within liquid level cell 46. As detailed above, the output of pump 12A may be modulated using one or more pulsation dampers 16A and/or mass flow controllers 18A in order to provide improved control of the rate at which the liquid from liquid level cell 46 is introduced into the downstream measurement elements. The pump 12A used for extracting the treated condensate may, for example, be configured as a variable speed peristaltic pump while the liquid flow meter 18A may, for example, be configured as a turbine (Pelton wheel) type meter that incorporates an optical sensor and provides for a flow rate in the range of 0-100 mL/min.
As also illustrated in
As illustrated in
The liquid flow from pump 12A will eventually reach a three way valve 56 configured with exit ports C and D for selectively routing the treated condensate into an ion exchange column 59 from which the condensate will enter the conductivity probe cell 58B where it will be exposed to a conductivity probe 58A that may, for example, include platinum surfaced electrodes and exhibit a cell constant of 0.1. As will be appreciated, the conductivity probe cell 58B may be configured in a manner similar or even substantially identical to the pH probe cell 42B for ensuring sufficient contact between the condensate flow and the conductivity probe.
The ion exchange column 59 may, for example, be packed with one or more strongly acidic cation exchange resin(s) in H+ form with the output of the conductivity probe cell 58B being discharged into a waste jug 60 that may, for example, comprise a heavy walled plastic jug providing a volume of about 2 gallons (7.6 liters) volume and having a closed or closeable top. Although not required, maintenance of the apparatus may be simplified if most or all of the waste streams exiting the apparatus—both liquid and gas—are discharged to a single waste jug having sufficient capacity to support the intended operation of the apparatus as illustrated in
A vacuum pump may be provided in connection with waste jug 60 that will permit the apparatus as illustrated in
As illustrated in
The construction of the phase separation column 64 may be similar to that of gas stripping column 24. A discharge tube 64A is provided at that head of phase separation column 64 above the packed section 65 which functions as a mist catcher. The packed section may be packed with a variety of materials including, for example, 5 mm Raschig rings comprising 5 mm long cut pieces of thin walled borosilicate glass tubing. The discharge tube may be connected through a branch tube or connection 66A to a gas pulsation damper 66 for stabilizing the gas flow entering the gas flow meter 70 with the headspace of the gas pulsation damper being, in turn, connected to a waste jug 60 through pressure equalizing tube 66B.
As illustrated in
As further illustrated in
Although, as will be appreciated, the antisiphon element may be manufactured in a range of functional configurations. One such embodiment is illustrated in
Illustrated in
As illustrated in
An embodiment of a pump controller 54 is illustrated in
As illustrated in
As illustrated in
As illustrated in
As illustrated in
As further illustrated in
An embodiment of a pump speed control circuit suitable for use with reagent pumps 12B, 12C and 12D is illustrated in
An embodiment of a general scheme or system for acquiring analytical data from embodiments of the disclosed apparatus is illustrated in
Similarly, switch position data 190 with respect to the position of switches 98 and 100 within pump controller 54 may be transmitted to data acquisition unit 182 as “on-off” voltage values and a pH meter 192 may transmit the sensed value of the condensate pH as an analog voltage signal to data acquisition unit 182. The data acquisition unit 182, in turn, transmits the accumulated parametric and status data to a laptop or minicomputer 194 where the data is used in calculating derivative chemical and physical parameters and values that may be stored, displayed and/or transmitted as calculated values, raw data and switch settings and/or one or more summary analytical report(s).
As noted above, the embodiment of the analytical apparatus as illustrated in
Further, it should be understood that while the procedures detailed below may be described in a manner that suggests manual operation of the apparatus and manual recordation of the various measurements, the data should preferably be collected using data acquisition unit 182 and laptop computer 194 as illustrated in
“Total NCG” or “NCG Ratio” refers to the weight or number of moles of the vapor phase remaining in contact with the condensate at ambient temperature and pressure after the steam is condensed with no attempt made to correct for the amount of gas dissolved in the condensate or the amount of water vapor in vapor phase. Typically, the value is expressed in terms of weight percent or mole percent in relation to the condensate.
To determine NCG ratio in a steam sample using steam sampling module 10A illustrated in
Set switch 100 to “adjust level.” When condensate appears in level cell 46, adjust speed control potentiometer 102 as needed to maintain a stable water level inside of level cell 46. Then set switch 100 to “control level.”
When all readings have stabilized, record gas flow through gas flow meter 70 and condensate flow through liquid flow meter 18A. The ratio of these values will be proportional to the NCG ratio. (Note: the response of a thermal gas flow meter is proportional to the number of moles of gas per minutes weighted by the molar heat capacity of each gas at constant pressure. Therefore, an average heat capacity for the mixture of gases needs to be estimated in order to convert the NCG ratio to mole percent, and an average molecular weight needs to be estimated in order to convert the NCG ratio to weight fraction.)
Record condensate pH. Briefly set switch 98 to “read conductivity,” record the conductivity value and condensate temperature, then reset switch 98 to “read level.”
Set 3-way switch 56 to exit port D, forcing condensate to flow through ion exchange column 59 before it enters conductivity probe cell 58B. Read and record conductivity and condensate temperature. Repeat conductivity reading as necessary until the conductivity reading is stable. Finally, set 3-way switch 56 back to exit port C to conserve the resin in the ion exchange column.
The strongly acidic cation exchange resin in column 59 will replace all cations in the condensate with protons, thereby converting bicarbonate and hydrogen sulfide ions in the condensate to carbon dioxide and hydrogen sulfide. The chloride ion in the condensate will remain as hydrogen chloride, HCl. In the absence of brine carryover or another source of salts of cations other than ammonium, HCl will be the only electrolyte present in the condensate, whereby the measured conductivity of the condensate will be approximately proportional to the amount of chloride in the condensate.
The measurement of HCl is subject to interference by nonvolatile salts (for example sodium chloride and sodium sulfate) that may be present due to brine carryover or other sources of contamination. These salts will be converted to the corresponding acids and will contribute to the measured conductivity together with the HCl initially present in the steam. In this case, the conductivity reading recorded will be approximately proportional to the total number of equivalents of HCl, sulfuric acid and any other strong acids that may be present in the sample analyzed.
The major electrolytes present in the condensate from steam without brine carryover will be the ammonium salts of carbon dioxide, hydrogen sulfide and boric acid, all three of which are weak acids. The mole sum of these three anions can be determined as the total alkalinity of the brine to the methyl red endpoint, approximately pH 4.2. If chloride is present, it will primarily be in the form of ammonium chloride, and the ammonia associated with chloride will not register as part of the titre. The total amount of ammonia in the sample analyzed can be determined according to Equation I as:
Ammonia(moles)=Alkalinity(equivalents)+Chloride(moles) I
Determine alkalinity by titrating the condensate using 0.010N sulfamic acid or another dilute solution of a strong acid from reservoir 14B with accurately known concentration. Adjust the flow rate of pump 12B until the pH of the condensate as measured by pH probe 42A is stable within the range 4.0 to 4.5, and preferably about 4.2. Record the flow rates at flow meters 18A and 18B. The concentration of ammonia in the condensate is then determined by Equation II:
C(ammonia,M)=C(chloride,N)+C(Acid,N)×Flow(18B)/(Flow(18A)−Flow(18B)) II
The end point of the titration can conveniently be noted visually by adding a suitable amount of a pH indicator dye composition to the acidic solution from reservoir 14B; for example, a combination of bromocresol green and methyl red which has a sharp transition near pH 4.2. The operator will be able to detect the approximate location of the endpoint as a function of flow rate measured by flow meter 18B by noting the characteristic mixed color at the transition value of pH as observed inside the transparent gas purge chamber 36. If optional colorimeter 44 is provided, the endpoint will be indicated by the appropriate ratio of absorbance at two wavelengths which are representative of absorbance by the indicator dye composition above and below the transition pH value of the indicator dye composition employed.
In this case, the acidic solution from reservoir 14B should be a 0.1N solution of a fairly strong acid, preferably a 0.1M solution of sodium hydrogen sulfate also containing an effective amount of soluble starch, preferably about 200 mg/L. Sulfide titrant solution from reservoir 14C should preferably be a 0.005M solution of iodine also containing sufficient potassium iodide to keep the iodine in solution, and sulfide titrant solution from reservoir 14C should be added to the condensate by route 21A; that is, before reaction coil 22. Adjust pump 12B to give a stable condensate pH of 3 or lower. Adjust pump 12C until dark blue color indicative of excess iodine is visible in column packing 26 and gas purge chamber 36. Due to small variations in mixing of the condensate with sulfide titrant solution from reservoir 14C, the color of the liquid will fluctuate rapidly between blue and clear at the equivalence point. Record condensate flow at flow meter 18A and sulfide titrant flow at flow meter 18C. The concentration of hydrogen sulfide in the condensate may then be determined using Equation III:
C(Sulfide,M)=C(Iodine,M)×Flow(18C)/(Flow(18A)−Flow(18B)−Flow(18C)) III
The endpoint can also be determined electronically, using optional colorimeter 44 to note the color change.
The sum of carbon dioxide and the unreactive gases (hydrogen, argon, etc.) is determined by acidifying the condensate and purging it to strip out dissolved gases while eliminating hydrogen sulfide by reaction with iodine. Mass transfer of dissolved gases to the gas phase is accomplished in three steps. First, gases partition between liquid and gas inside reaction coil 22, and are separated when the mixture of liquid and gas enters gas stripping column 24. The liquid then flows down through column packing 26, where it meets a counter flowing stream of the purge gas. Finally, the liquid is purged with gas bubbles from frit 36A in gas purge chamber 36.
Most conveniently, this analysis should be performed immediately after determining sulfide by iodometric titration, starting with the apparatus in its condition existing at the end of the iodometric titration procedure.
Maintain the flow rate of acidic solution from reservoir 14B as needed to maintain condensate pH at 3 or lower. Adjust pump 12C to increase flow of sulfide titrant from reservoir 14C to about 10% beyond the stoichiometric end point of the sulfide titration to establish and maintain a dark blue color inside gas purge cell 36.
If the volume of gas in the sample (as indicated by gas flow meter 70 before purge gas is turned on) is less than about 4 times the condensate volume, adjust control valve 38B to provide purge gas flow sufficient to raise the combined gas flow through gas flow meter 62 to a value which is 4 to 6 times greater by volume than condensate flow. (Condensate flow rate in mL/min, and flow rates of gases in mL/min referred to ambient temperature and pressure, which corresponds to the usual output of a thermal gas flow meter.) Adjust control valve 38A to provide purge gas flow (as indicated by gas flow meter 34) through frit 36A which is 4 to 6 times greater by volume than condensate flow. Record all liquid and gas flows and then apply Equation IV:
Condensate Flow=Flow(18A)−Flow(18B)−Flow(18C) IV
Correct gas flows determined by gas flow meters 34 and 70 by subtracting water vapor content which can be reliably estimated based on the temperature and pressure of the gas and then apply Equation V:
CO2+Unreactive Gases=Corr.Flow(70)−Corr.Flow(34) V
Determine flow of CO2+ nonreactive gases in the manner detailed above.
Switch 3-way valve 62 to redirect gas flow through exit port B leading to reaction coil 72, and turn on pump 12D to provide sufficient flow of alkali solution from reservoir 14D to ensure that the liquid effluent from reaction coil 72 remains highly alkaline as indicated by the color of the pH indicator dye in alkali solution from reservoir 14D.
Measure gas flow at meter 70 and correct it by subtracting the amount of water vapor. The corrected value will be sum of unreactive gases in the steam (that is, not including ammonia, hydrogen chloride, hydrogen sulfide or carbon dioxide), with mole numbers of the several unreactive gases weighted by the corresponding molar heat capacities at constant pressure.
Subtract the corrected value for unreactive gases from the corrected value.
The sum of carbon dioxide, hydrogen sulfide and unreactive gases may be determined as follows, using an inert gas free of oxygen—preferably argon—as the purge gas.
Pump 12C is not used. Adjust pump 12B to provide 0.1M acid solution as needed to maintain condensate pH at 3 or lower. Adjust purge gas flows as described above. Measure the two gas flows, and correct the values obtained for water vapor content and then apply Equation VI:
CO2+H2S+unreactive gases=Corr.Flow(70)−Corr.Flow(34) VI
To determine unreactive gases, reset 3-way valve 62 to exit port B and pump alkali solution from reservoir 14D into the gas stream to scrub CO2 and H2S from the gas stream as described above.
To determine CO2+H2S, subtract the corrected value for unreactive gases from the corrected value for [CO2+H2S+unreactive gases].
To determine CO2, subtract the value for H2S determined by titration from the corrected value for [CO2+H2S].
Potentiometric titration of sulfide may be achieved using a sulfide ion specific electrode 74 in combination with an inert, oxygen-free purge gas including, for example, helium, argon, neon, nitrogen or mixtures thereof.
Sulfide titrant solution from reservoir 14C preferably is a 0.01 M solution of zinc acetate and/or zinc sulfate with accurately known concentration. As will be appreciated, the titration can also be performed using salts of lead, cadmium or another metal forming an insoluble sulfide precipitate, but zinc salts are preferred because zinc is nontoxic and environmentally benign, and the reaction to form zinc sulfide exhibits accurate stoichiometry with no side reactions interfering.
Set pump 12B to acidify the condensate to pH 3 or below using a 0.1M solution of a strong acid. Set purge gas flows as described above to ensure thorough stripping of CO2 and H2S from the condensate. Combine sulfide titrant solution from reservoir 14C with the gas stream by way of alternate flow path 21B, that is, before the mixture enters the second reaction coil 72.
Stepwise and gradually increase the flow rate of sulfide titrant solution from reservoir 14C, allowing enough time between increments for the potential of the sulfide electrode 74 to stabilize, and record all flows and the potential of sulfide electrode 74 at each step. Plot the potential of sulfide electrode 74 against flow of sulfide titrant from reservoir 14C as read from flow meter 18C, and note the value of titrant from reservoir 14C flow that corresponds to the maximum slope of the curve. At this point, the moles of sulfide titrant delivered will be equal or substantially equal to the moles of H2S in the sample.
Sample input module 10B is used to analyze steam condensate which contains dissolved gases with little or no vapor phase present. To sample condensate from a sample tap at sufficiently high pressure to flow through the apparatus without assistance of a pump, close valve 136A, and regulate sample flow by opening and adjusting needle valve 136B. To sample condensate at atmospheric pressure, close valve 136B, open valve 136A wide, and adjust pump 134 to provide the sample flow rate desired.
To analyze condensate or another liquid containing little gas, provide purge gas flow through each of control valves 38A and 38B equal to about 4 to 6 times the condensate flow.
Use sample input module 10C to separate steam from the brine. Adjust control valve 160 to ensure that only a liquid phase exists upstream of this valve, and calculate the steam fraction as a function of the separator pressure and the temperatures of unflashed and flashed brine recorded using T-probes 158A and 158B, respectively. Note that knowledge of these three parameters allows the steam fraction to be calculated in two different ways, and that the calculation can be made more accurate by assuming that the brine has thermodynamic properties identical to a sodium chloride solution having the same conductivity at equal temperature instead of representing the brine by pure water.
The inventions detailed herein represent novel solutions to the problem of providing a timely analysis of a two phase mixture of water and vapor to determine the amount of various volatile components that are distributed with widely varying proportions between the two phases and which exhibit wide variations in their chemical properties.
As detailed above, the apparatus and procedures disclosed allow the two phases present in the sample stream to be cleanly separated and moved under positive pressure through the apparatus. A control system coupled with devices for pumping water and/or aqueous solutions at a rate substantially equal to the rate at which the liquid phase enters the system is helpful in maintaining a smooth flow through the apparatus. In the preferred embodiment, this flow control function is provided by a liquid level sensing electrodes 48 inside of liquid level sensing cell 46 coupled to pump controller 54 that regulates pump 12A so as to maintain an essentially constant liquid level inside liquid level sensing cell 46. One skilled in the art will be able to substitute other means to provide the smooth flow required; for example, a flow sensor which measures and time averages flow rate of liquid 46A and a pump controller 54 that maintains the pumping rate of pump 12A equal to the average flow rate of liquid 46A.
As will be appreciated, the use of additional dampers and flow controllers also contribute to achieving a steady flow through the apparatus.
With the exception of chloride, all other chemical components are determined using measurements of gas flows or by titration, which is based on measuring liquid flows. As will be appreciated, the apparatus could easily be modified to provide a test module configured for measuring chloride by titration as well. Because the gas flow meters and liquid flow meters employed are easily calibrated and typically show little drift or error after being calibrated, reliance on gas and liquid flow measurements enables sufficiently accurate analyses to be performed in the field without requiring the laborious calibration procedures that other sorts of analytical transducers would require.
As will be appreciated by those skilled in the art, the method and apparatus as disclosed herein can be modified in many ways while remaining true to the spirit of the invention including, for example,
reducing the interfering contribution of CO2+H2S to the conductivity of the ion exchanged condensate exiting ion exchange column 59 by providing a secondary purge of the condensate using an inert gas or air before reading the conductivity;
as a first alternative to relying on conductivity as an indication of HCl content, the pH of the ion exchanged condensate can be measured instead of conductivity;
as a second alternative to relying on conductivity as an indication of HCl content, the ion exchanged brine can be titrated in a manner similar to that employed to determine the ammonia content of the condensate, using a basic solution as the titrant;
the composition of the mixture of “unreactive gases” that remains after CO2+H2S have been removed from the gas fraction by reaction with alkali from reservoir 14D can be determined using a gas chromatograph or a mass spectrometer; the fuel value of the NCG stream can be determined using a combustible gas analyzer;
the oxygen content of the NCG stream can be determined using an oxygen meter; a range of other reagent compositions and concentrations can be used instead of those described above;
3-way valve 62 may be removed to eliminate the option of bypassing reaction coil 72;
other kinds of gas and liquid flow meters can be utilized at one or more locations throughout the apparatus;
chloride in the condensate could be determined using a chloride ion specific electrode, a change which would eliminate the need for ion exchange column 59;
ammonium ion, ammonia and/or carbon dioxide concentrations could be determined using the corresponding specific electrodes; and/or.
the proportional controller circuit 106 used for controlling water level in level cell 46 could be replaced by another kind of control circuit including, for example, a proportional integral (PI) controller or a proportional integral derivative (PID) controller could be utilized for improving the controller response.
While the present disclosure has included descriptions of various embodiments of the apparatus, method and system, it should be understood that these embodiments are not intended to limit the disclosure and that one of skill in the art, guided by the present disclosure, can adopt the methods and apparatus disclosed to accommodate the testing of a wide range of parameters associated with mixed liquid/gas flows, whether such flows are produced naturally or during the course of an industrial process. Accordingly, the present disclosure is intended to encompass such alternatives, modifications, and equivalents as may be included within the spirit and scope of the appended claims.
This application claims priority from U.S. Prov. Pat. Appl. No. 61/176,257, the contents of which are hereby incorporated, in their entirety, by reference.
Number | Date | Country | |
---|---|---|---|
61176257 | May 2009 | US |