The present invention relates to a method and apparatus for detecting ultrasound contrast agents in arterioles. In particular, the invention relates to diagnosing coronary artery disease without the need for a stress test by detecting the presence of ultrasound contrast agents microbubble in larger vessels including the arterioles.
Ultrasound Contrast agents act as intravascular tracers and are approved for various uses throughout the world. In the US, FDA has approved the use of contrast for left ventricular opacification to aid in the delineation of endocardial borders in echo studies. In Europe, there are radiology indications as well including enhancement of the macro and microvasculature. However a great deal of clinical research is ongoing for other uses of contrast agents. Myocardial Contrast Echo (MCE), the ability to assess perfusion of the myocardium with echo, is one of the hottest areas of research in echo. The first FDA approval for assessment of MCE is expected to occur in 2006 with others to follow.
In order to assess coronary artery disease (CAD) a patient typically has to undergo some form of stress test. This is due to the heart's ability to compensate for a stenosis (partial blockage) in one of the main coronary arteries to maintain resting coronary blood flow. Compensation occurs by dilation of arterioles to account for the pressure drop across the stenosis. This helps maintain capillary pressure and this is critical for maintaining perfusion. However, after about an 85-90% stenosis the body has exhausted its coronary flow reserve (i.e., its ability to dilate arterioles). For increasing stenosis above 85-90% resting blood flow begins to decrease. In order to diagnose patients with CAD and non-flow limiting stenosis at rest some form of stress test is given—either with ECG, Echo or nuclear perfusion. These tests involve a patient running on a treadmill to obtain a higher heart rate or the use of an inotropic drug (i.e., Dobutamine) or a vasodilator. All of these tests are time consuming and carry some risk and discomfort for the patient.
During the systolic portion of a cardiac cycle the contraction of the heart squeezes blood forward in the venules and backwards in the arterioles. If the blood volume of the arterioles is increased—such as in the case of a coronary stenosis, there is more blood to be squeezed from these vessels. The velocity of this blood will be much higher than that in the capillaries. This allows for the possibility to isolate the arterioles based on velocity differences during systole. Since these vessels are too small to obtain a Doppler signal in the presence of a very strong tissue signal other methods have to be used. One such method uses microbubbles to enhance the signal from blood. Also, since microbubbles can be destroyed with ultrasound this means that destruction could be used to isolate signals from arterioles.
Using ultrasound at an energy high enough to destroy the microbubbles in an imaging plane and imaging with a frame rate such that the microbubbles in the arterioles have enough time to flow back into the scan plane can isolate the microbubbles in the arterioles. At these destructive power levels and frame rates of greater than 1-2 Hz or so, microbubbles don't have enough time to reach the capillaries. However in order to make this possible, imaging techniques have to be developed that are sensitive to small number of microbubbles while suppressing tissue signals at MI's that are destructive (i.e. MI's>=0.2). Techniques based purely on harmonics often have poor tissue suppression due to the presence of tissue harmonic signal at the powers required to destroy microbubbles. Therefore the tissue signal will mask the signal from contrast agents with these techniques. Techniques to image arterioles based on imaging in between the harmonics (i.e., ultraharmonics) or with power doppler techniques were disclosed in U.S. Pat. No. 6,730,036. These techniques are effective in reducing tissue signal at these destructive MI's but suffer from insufficient signal/noise at the power levels (typically MI>0.2 and <0.8 depending on the contrast agent and patient) and frame rates (typically frame rate >=5 and <=25 Hz) required to work effectively. At even higher power levels, the signal to noise ratio of these techniques increases but since a “thicker slice” of contrast in the myocardium is destroyed a lower frame rate is required to allow a sufficient number of microbubbles to replenish the imaging plane or subvolume even in the arterioles. Forcing these techniques to work at slower frame rates allows more time for arterioles to refill but also makes it harder to isolate arteriole signals from capillary signals since capillaries also have more time to refill. Alternatively, the higher MI could be used at a higher frame rate but would require large doses of contrast agent to improve signal to noise leading to attenuation of much of the myocardium. An additional drawback of the power doppler technique is that it suffers from motion artifact if used during portions of the cardiac cycle where the heart is moving.
An imaging technique that would allow good tissue suppression and signal to noise at power levels required to destroy microbubbles varies the amplitude and/or phase between pulses to suppress linear tissue signals. One possible technique was described in U.S. Pat. No. 5,577,505 but was not used in this manner. This patent describes a multi-pulse technique that involves changing amplitudes between transmit pulses and combining these pulses during receive to suppress the linear signals. With this technique as well as other multi-pulse techniques that have amplitude changes between transmit pulses and optionally phase changes as well, microbubbles exhibit a strong nonlinear response at the fundamental whereas tissue is suppressed since it behaves relatively linearly at the fundamental frequency. This tissue suppression at the fundamental frequency is opposed to purely harmonic based techniques (either pulse inversion or harmonic imaging) that have tissue harmonic signals present at low MI's (>0.1 or so)—often below the threshold required to destroy the microbubbles. There is also an improvement in signal to noise in operating at the fundamental frequency since attenuation is much lower than at the harmonic.
There are also multi-pulse techniques based on changing the amplitude and possibly phase between pulses and combining the pulses in such a manner that linear and/or non-linear tissue harmonic signals cancel (U.S. Pat. No. 6,361,498).
With these nonlinear detection techniques it is possible to image at MI's of 0.2 or higher—which are destructive for most contrast agents and to have minimal tissue signal—even when imaging at the harmonic. Frame rates of 25 Hz or lower allow enough time for some arteriole refill to occur.
As a means to further increase sensitivity and signal to noise of the arteriole contrast signal, the use of “coded” waveforms could be employed. Coded waveforms have been described in literature (e.g., U.S. Pat. No. 6,050,947) and involve transmitting a longer waveform to increase signal to noise. With proper “decoding” on receive the returning pulse can be compressed to gain back the loss of resolution. For example a “chirp” is a special case of a “coded” waveform and is a signal in which the frequency increases (‘up-chirp’) or decreases (‘down-chirp’) with time. These waveforms could be used in combination with the previous described multi-pulse detection techniques by modifying the amplitude and/or phase of the coded signal, decoding them on receive and combining them in a manner to suppress linear and/or non-linear signals.
Increased sensitivity can also be obtained by using an imaging sequence that uses an MI that is high enough to destroy contrast agent throughout the throughout the cardiac cycle (e.g., 0.2-0.8 depending on the microbubble characteristics) but then uses an even higher MI (e.g., 1.0) during systole—the portion of the cardiac cycle that has the blood in the arterioles “squeezed” into the imaging plane. This improves signal-to-noise by increasing the detection beamwidth to image more microbubbles as well as increasing the backscatter from each microbubble due to the higher power level. Other techniques could be used to get the same effect—such as increasing the beamwidth through focusing or apodization. A Matrix transducer allows for control of the elevation in this manner.
In order to make the results meaningful it is critical to calibrate the concentration of contrast. This can be accomplished by measuring the intensity in the myocardium throughout the cardiac cycle and normalizing to large intra-myocardial vessels that are typically seen during diastole (U.S. Pat. No. 6,730,036). In the scenario of the triggered imaging mentioned above this would require a 2nd triggered frame during diastole to compare to. Another possible method would be to normalize to a large blood pool that represents 100% blood volume. This can be in the Left Ventricular cavity (U.S. Pat. No. 6,258,033) or a large vessel in the myocardium. For example, if the intensity in the myocardium during systole was 20 dB lower than the LV cavity and the concentration of microbubbles was still in the linear dosing range then myocardial arteriole blood volume would be 1%.
The invention described here is a method and apparatus for ultrasound imaging of microbubbles of a contrast agent in arterioles while virtually all microbubbles of the contrast agent have been eliminated in the capillaries of a patient and tissue signal response to ultrasound imaging is suppressed. This method and apparatus permits ultrasound imaging for detecting coronary artery disease without the need for a stress test.
The invention would be used to diagnose coronary artery disease without having a stress test. It could also serve as a quick screening tool for CAD.
Referring now to drawings of
As shown in
The present invention provides for imaging in subvolumes to include above and beyond a plane as a subvolume is more than one plane in an elevation dimension but could represent a smaller lateral dimension. Matrix transducers are capable of using subvolumes.
An imaging mode 6 is next selected for contrast destruction and tissue suppression. In the case of the first technique this can include setting a mechanical index for microbubble destruction 7 and setting a frame rate 8 to permit sufficient time for refilling larger vessels with contrast agent such as the arterioles. The mechanical index is preferably set to a value within a range of a range of 0.2 to 0.8. The frame rate is preferably set to a value within a range of 1 to 25 Hz.
In the case of the first technique by the use of a multi-pulse technique combining amplitude and phase modulation using the controls on the console of the ultrasound imaging apparatus, linear and optionally second order non-linear tissue signals are eliminated from the imaging by combining the pulses so that tissue noise is suppressed. Power and frame rates are chosen such that microbubble signals from the capillaries are eliminated.
In
The ultrasound imaging apparatus 12 obtains images of the patient's body 13 and when all images are obtained 14, the images are calibrated or normalized, as described below for either the LV cavity 15 or the myocardial intensity and appropriate normalization for LV cavity intensity 19 or either the diastolic intensity or myocardial intensity 18 is obtained. Images or a graph of results are derived based on the normalized values 17.
The first technique of the present invention is different from that disclosed in U.S. Pat. No. 6,730,036 as the present invention discloses the use of fundamental detection techniques. U.S. Pat. No. 6,730,036 discloses the use of harmonic or ultraharmonic based techniques (filtering between harmonics). This first technique would use non-linear fundamental techniques including but not limited to those described in U.S. Pat. No. 5,577,505 and U.S. Pat. No. 6,361,498. These techniques suppress tissue very well in the mechanical index (MI) range that the present invention needs to image at (typically greater than 0.2 and less than 0.8) with the first technique of the present invention.
Calibration/normalization is necessary to assess the amount of contrast. This is true since there are many things that affect the intensity of a given frame. A higher contrast dose will give a higher intensity and a higher gain or higher power will give a higher intensity so in order to determine the concentration of contrast there must be something to compare the intensity of a given region of interest in a given frame to. In one case the intensity in the myocardium of end systolic frames can be compared to the intensity in the myocardium end diastolic frames. For example, the variation in the cardiac cycle could be 6 dB with end systole being 6 dB below end diastolic intensity. Alternatively the systolic/diastolic ratio (systolic intensity divided by diastolic intensity) could be generated. In the case of 6 dB the ratio of intensities would be 0.25. The other way to normalize is to compare locally to the LV cavity. Comparing locally is important (i.e. approximately same depth so acoustic parameters including MI and beam properties are as equal as possible in the tissue and in the cavity). Since the LV cavity is 100% blood the ratio of myocardial intensity to LV cavity will give an indication of the percent of blood (e.g., bubbles in the arterioles assuming we have isolated the arterioles by destruction of bubbles in capillaries).
The frame rate will control the time and therefore velocity of vessels that are being imaged. Velocities are higher in larger vessels so faster frame rates can also help isolate bigger coronary arteries as well as arterioles. Visualization of the larger vessels such as intramyocardial coronaries are primarily seen during diastole and help determine system settings such as imaging mode, Mechanical Index, Frame rate, and gain as well as contrast infusion rate. They also provide means for normalizing the systolic intensities.
The imaging parameters are optimized for triggered images 26—settings such as delay from R-Wave, mechanical index, focusing, etc. The other settings such as gain are optimized for the best visualization of the images 27 and the steps 29-37 are similar to the steps in
With this second technique, it is also possible that the detection technique and transmit and receive parameters are different in the triggered frames vs. the non-triggered frames. In this scenario the detection techniques include those mentioned in the first technique in
While presently preferred embodiments have been described for purposes of the disclosure, numerous changes in the arrangement of method steps and apparatus parts can be made by those skilled in the art. Such changes are encompassed within the spirit of the invention as defined by the appended claims.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/IB2006/051774 | 6/2/2006 | WO | 00 | 7/2/2008 |
Number | Date | Country | |
---|---|---|---|
60687845 | Jun 2005 | US |