The invention will now be described in more detail, by way of example only, with reference to the accompanying drawings, in which:
a and 4b illustrate the bending images of the retina upwards to compensate for fan scanning distortion. The image in the square centred on the fovea of the normal subject is used in the processing of the RPE layer as described below.
OCT can be implemented, for example, with apparatus described in detail in our U.S. Pat. No. 5,975,697, the contents of which are incorporated herein by reference. The described processing can be carried out with a personal computer, for example, including a Pentium™ microprocessor.
Consider the case of angular OCT scanning of the retina. Since, the retina must be scanned through the pupil of the eye, fan scanning must be employed. The fan of rays converges at a point C, as illustrated in
The frame grabber of the OCT system places the B-scan image in the plane (h,v), where:
h=khα,
v=kvz (2a,b)
z is the axial movement of the reference mirror from the initial position. kh and kv are scanning scaling factors for the transverse and axial scanner respectively. kh is given by the number of sample pixels along the horizontal axis, 2H, divided by the maximum optical ray deflection angle, αM. kv is given by the number of vertical sample pixels in the image along the vertical axis V, divided by the maximum axial range, zM covered by the axial scanner in the reference arm of the OCT interferometer.
The axial scanner varies the reference path to select points within the retina, situated at a certain radial distance between r0 and r0+Δr. If the scanner moves by z, then the coherence gated spatial window advances from the initial position r0 to:
where n is the average index of refraction of the retina, considered a constant, 1.38 everywhere in the eye for brevity.
Placing the reference for OPD=0 in the top centre of the image o and also making the object space and the image space coincide in this point, lateral and vertical errors produced by the fan scanning can be computed as:
EI measures how much the image point I moves laterally relative to the corresponding object point O, while Ea signifies how much the image point I moves axially from the corresponding object point O. For a null α angle, the errors are zero.
To better understand the distortions in the fan scanning case, let us consider a simple rectangular object, such as a microscope slide glass in
For points on the anterior surface, Σ1, the polar coordinates in the object space are:
In Cartesian coordinates h and v, the points of the anterior surface Σ1 will be located in the B-scan image at points:
These equations show that the higher the angle α either side of the axis oz, the larger the vertical distortion of the image. A horizontal line in the object is represented as a downwardly curved line in the image space. Similarly, the second surface, Σ2, given by points
will be transferred to a curved line:
in the image plane, (h,v). The corrected image is shown in the right hand side of
We inversed equations of type 6 and 8 written for each point in the image to correct T-scan based B-scan images obtained from the retina. The correction exercise is exemplified on two images shown in
It is important to associate the pathology location to the eye curvature, which is correct in the images bent upwards. For the numerical values used, the axial error is 1.2 mm and lateral error 0.44 mm. Although it is possible to estimate the eye length, for more accurate results, OCT should be first used to evaluate the eye length value, and input to the evaluations above.
A second aspect of the disclosure is the correction of orientation of layers just below the foveal pit. These layers are important for correct diagnosis of eye diseases.
A B-scan OCT image of the fovea obtained with T-scanning is shown in
The OCT image sampled by the square in
z=ƒ(x) (19)
In the same system of coordinates, the equation of the middle of the RPE can be approximated by:
zp=cons tan t (20)
The ray coming from the vitreous is incident on the retina in Aj. The equation of the refracted line AjBj is written for a point Aj (xj,zj) on the inner limiting membrane (ILM) described by equation (19), as:
x−x
j
=m(z−zj) (21)
The slope is:
The incidence angle, θj is
θj=π±γ (23)
where γ is given by:
evaluated in each point Aj.
We can calculate the coordinate of each point Bj on the RPE where the line described by the equation (21) intersects the RPE described by equation (20), and obtain the points of coordinates (xp,zp). If the origin of the optical path length in the vitreous is at a coordinate z=z0 (a reference line is shown in
v=n
v(zj−z0)+nr√{square root over ((xj−xp)2+(zj−zp)2)}{square root over ((xj−xp)2+(zj−zp)2)} (25)
This determines the shape of the RPE layer in
In points such as Bf and Bp, where the ray comes along the normal to the retina (like points in the center, Af or outside the fovea region, AR respectively), the x coordinates are the same and the optical path length is:
v
f
=n
v(zf−z0)+nr(zp−zf) (26a)
or
v
R
=n
v(zR−z0)+nr(zp−zR) (26b)
The elevation of the RPE in the center of the image can be simply calculated by subtracting equation (26a) from (26b) which gives:
δ=(nr−nv)(zf−zR) (27)
Considering a normal average foveal pit value H=(zf−zR)=150 μm and the values for the indexes of refraction of the vitreous, nv=1.336, and of the retina up to the RPE, nr=1.35, δ=2.1 μm. Such a deviation is hard to be noticed in
Number | Date | Country | Kind |
---|---|---|---|
0405416.9 | Mar 2004 | GB | national |
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/CA05/00367 | 3/11/2005 | WO | 00 | 6/13/2007 |