This invention relates to surgical methods and apparatus in general, and more particularly to methods and apparatus for treating a hip joint.
The hip joint is a ball-and-socket joint which movably connects the leg to the torso. The hip joint is capable of a wide range of different motions, e.g., flexion and extension, abduction and adduction, medial and lateral rotation, etc. See
With the possible exception of the shoulder joint, the hip joint is perhaps the most mobile joint in the body. Significantly, and unlike the shoulder joint, the hip joint carries substantial weight loads during most of the day, in both static (e.g., standing and sitting) and dynamic (e.g., walking and running) conditions.
The hip joint is susceptible to a number of different pathologies. These pathologies can have both congenital and injury-related origins. In some cases, the pathology can be substantial at the outset. In other cases, the pathology may be minor at the outset but, if left untreated, may worsen over time. More particularly, in many cases, an existing pathology may be exacerbated by the dynamic nature of the hip joint and the substantial weight loads imposed on the hip joint.
The pathology may, either initially or thereafter, significantly interfere with patient comfort and lifestyle. In some cases, the pathology can be so severe as to require partial or total hip replacement. A number of procedures have been developed for treating hip pathologies short of partial or total hip replacement, but these procedures are generally limited in scope due to the significant difficulties associated with treating the hip joint.
A better understanding of various hip joint pathologies, and also the current limitations associated with their treatment, can be gained from a more thorough understanding of the anatomy of the hip joint.
The hip joint is formed at the junction of the leg and the hip. More particularly, and looking now at
More particularly, and looking now at
Looking next at
Both the head of the femur and the acetabular cup are covered with a layer of articular cartilage which protects the underlying bone and facilitates motion. See
Various ligaments and soft tissue serve to hold the ball of the femur in place within the acetabular cup. More particularly, and looking now at
As noted above, the hip joint is susceptible to a number of different pathologies. These pathologies can have both congenital and injury-related origins.
By way of example but not limitation, one important type of congenital pathology of the hip joint involves impingement between the neck of the femur and the rim of the acetabular cup. In some cases, and looking now at
By way of further example but not limitation, another important type of congenital pathology of the hip joint involves defects in the articular surface of the ball and/or the articular surface of the acetabular cup. Defects of this type sometimes start out fairly small but often increase in size over time, generally due to the dynamic nature of the hip joint and also due to the weight-bearing nature of the hip joint. Articular defects can result in substantial pain, induce and/or exacerbate arthritic conditions and, in some cases, cause significant deterioration of the hip joint.
By way of further example but not limitation, one important type of injury-related pathology of the hip joint involves trauma to the labrum. More particularly, in many cases, an accident or sports-related injury can result in the labrum being torn away from the rim of the acetabular cup, typically with a tear running through the body of the labrum. See
The current trend in orthopedic surgery is to treat joint pathologies using minimally-invasive techniques. Such minimally-invasive, “keyhole” surgeries generally offer numerous advantages over traditional, “open” surgeries, including reduced trauma to tissue, less pain for the patient, faster recuperation times, etc.
By way of example but not limitation, it is common to re-attach ligaments in the shoulder joint using minimally-invasive, “keyhole” techniques which do not require laying open the capsule of the shoulder joint. By way of further example but not limitation, it is common to repair torn meniscal cartilage in the knee joint, and/or to replace ruptured ACL ligaments in the knee joint, using minimally-invasive techniques.
While such minimally-invasive approaches can require additional training on the part of the surgeon, such procedures generally offer substantial advantages for the patient and have now become the standard of care for many shoulder joint and knee joint pathologies.
In addition to the foregoing, in view of the inherent advantages and widespread availability of minimally-invasive approaches for treating pathologies of the shoulder joint and knee joint, the current trend is to provide such treatment much earlier in the lifecycle of the pathology, so as to address patient pain as soon as possible and so as to minimize any exacerbation of the pathology itself. This is in marked contrast to traditional surgical practices, which have generally dictated postponing surgical procedures as long as possible so as to spare the patient from the substantial trauma generally associated with invasive surgery.
Unfortunately, minimally-invasive treatments for pathologies of the hip joint have lagged far behind minimally-invasive treatments for pathologies of the shoulder joint and the knee joint. This is generally due to (i) the constrained geometry of the hip joint itself, and (ii) the nature and location of the pathologies which must typically be addressed in the hip joint.
More particularly, the hip joint is generally considered to be a “tight” joint, in the sense that there is relatively little room to maneuver within the confines of the joint itself. This is in marked contrast to the shoulder joint and the knee joint, which are generally considered to be relatively “spacious” joints (at least when compared to the hip joint). As a result, it is relatively difficult for surgeons to perform minimally-invasive procedures on the hip joint.
Furthermore, the pathways for entering the interior of the hip joint (i.e., the natural pathways which exist between adjacent bones and/or delicate neurovascular structures) are generally much more constraining for the hip joint than for the shoulder joint or the knee joint. This limited access further complicates effectively performing minimally-invasive procedures on the hip joint.
In addition to the foregoing, the nature and location of the pathologies of the hip joint also complicate performing minimally-invasive procedures on the hip joint. By way of example but not limitation, consider a typical detachment of the labrum in the hip joint. In this situation, instruments must generally be introduced into the joint space using an angle of approach which is offset from the angle at which the instrument addresses the tissue. This makes drilling into bone, for example, significantly more complicated than where the angle of approach is effectively aligned with the angle at which the instrument addresses the tissue, such as is frequently the case in the shoulder joint. Furthermore, the working space within the hip joint is typically extremely limited, further complicating repairs where the angle of approach is not aligned with the angle at which the instrument addresses the tissue.
As a result of the foregoing, minimally-invasive hip joint procedures are still relatively difficult to perform and relatively uncommon in practice. Consequently, patients are typically forced to manage their hip pain for as long as possible, until a resurfacing procedure or a partial or total hip replacement procedure can no longer be avoided. These procedures are generally then performed as a highly-invasive, open procedure, with all of the disadvantages associated with highly-invasive, open procedures.
As a result, there is, in general, a pressing need for improved methods and apparatus for treating pathologies of the hip joint.
During arthroscopic hip surgery, it is common to distract the hip joint so as to provide increased workspace within the joint. More particularly, during arthroscopic hip surgery, it is common to unseat the ball of the femur from the socket of the acetabular cup so as to provide (i) improved access to the interior of the joint, (ii) additional workspace within the interior of the joint, and (iii) increased visibility for the surgeon during the procedure. This hip joint distraction is normally accomplished in the same manner that the hip joint is distracted during a total hip replacement procedure, e.g., by gripping the lower end of the patient's leg near the ankle and then pulling the leg distally with substantial force so as to unseat the ball of the femur from the acetabular cup.
However, since the distracting force is applied to the lower end of the patient's leg, this approach necessitates that the distracting force be applied across substantially the entire length of the leg. As a result, the intervening tissue (i.e., the tissue located between where the distracting force is applied and the ball of the femur) must bear the distracting load for the entire time that the hip joint is distracted.
In practice, it has been found that the longer the distracting load is maintained on the leg, the greater the trauma imposed on the intervening tissue. Specifically, it has been found that temporary or even permanent neurological damage can occur if the leg is distracted for too long using conventional distraction techniques.
As a result, the standard of care in the field is for the surgeon to limit the duration of distraction during arthroscopic hip surgery to 90 minutes or less in order to minimize damage to the intervening tissue due to joint distraction. In some situations, this can mean that desirable therapeutic procedures may be curtailed, or even eliminated entirely, in order to keep the duration of the distraction to 90 minutes or less. And even where the duration of the distraction is kept to 90 minutes or less, significant complications can nonetheless occur for many patients.
In addition to the foregoing, in current hip distraction, it is common to use a perineal post to facilitate hip distraction. More particularly, and looking now at
Unfortunately, it has been found that the use of a perineal post can contribute to the damage done to the intervening tissue when the leg is distracted too long. This is because the perineal post can press against the pudendal nerve and/or the sciatic nerve (as well as other anatomy) when distraction occurs. Thus, if the distraction is held too long, neurological damage can occur. This is another reason that the standard of care in the field is for the surgeon to limit the duration of distraction during arthroscopic hip surgery to 90 minutes or less. Additionally, the perineal post can exert pressure on the blood vessels in the leg, and it has been shown that blood flow in these vessels (e.g., the femoral vein, etc.) can be reduced, or in some cases completely occluded, while the hip is in distraction, thus placing the patient in danger of forming deep vein thrombosis or developing other complications.
Additionally, current hip distraction limits the extent to which the leg can be manipulated under distraction during hip arthroscopy, since a substantial pulling force must be maintained on the distal end of the leg throughout the duration of the distraction. Due to this, and due to the fact that there are typically only 2-4 portals available for surgical access into the interior of the hip joint, visualization and access to hip joint pathology and anatomy is frequently hindered. This can limit the extent of surgical procedures available to the surgeon, and can prevent some procedures from being attempted altogether. Procedures such as mosaicplasty and autologous cartilage injection are examples of procedures which require access to extensive areas of the articular surfaces of the femoral head, but which are typically not performed arthroscopically because of the aforementioned access limitations due to leg distraction.
Thus, there is a need for a new and improved approach for distracting the hip joint which addresses the foregoing problems.
These and other objects of the present invention are addressed by the provision and use of a new method and apparatus for distracting a joint.
Among other things, the present invention provides a novel method for distracting a joint and for maintaining distraction of a joint, wherein the novel method minimizes damage to intervening tissue while maintaining distraction of the joint. In addition, the novel method allows visualization of areas in the hip joint that were not previously visible using a conventional hip distraction approach.
The present invention also provides novel apparatus for distracting a joint and for maintaining distraction of a joint, wherein the novel apparatus comprises a novel joint-spacing balloon catheter for maintaining the distraction of a joint. In addition, the novel apparatus preferably includes a novel inflatable perineal post for use in distracting the joint.
In one preferred form of the invention, there is provided a method for creating space in a joint, the method comprising:
applying force to a body part so as to distract the joint and create an intrajoint space;
inserting an expandable member into the intrajoint space while the expandable member is in a contracted condition;
expanding the expandable member within the intrajoint space; and
reducing the force applied to the body part so that the joint is supported on the expandable member.
In another preferred form of the invention, there is provided a method for creating space in a joint, the method comprising:
inserting a first expandable member into the interior of the joint while the expandable member is in a contracted condition;
expanding the first expandable member within the joint so as to create a first intrajoint space;
inserting a second expandable member into the first intrajoint space while the second expandable member is in a contracted condition; and
expanding the second expandable member within the first intrajoint space so as to create a second intrajoint space.
In another preferred form of the invention, there is provided a joint-spacing balloon catheter comprising:
a shaft having a distal end and a proximal end;
an expandable member attached to the distal end of the shaft, the expandable member being capable of supporting opposing bones of a previously-distracted joint when the distraction force is reduced; and
a handle attached to the proximal end of the shaft.
In another preferred form of the invention, there is provided a perineal post comprising a balloon.
These and other objects and features of the present invention will be more fully disclosed or rendered obvious by the following detailed description of the preferred embodiments of the invention, which is to be considered together with the accompanying drawings wherein like numbers refer to like parts, and further wherein:
In one form of the present invention, there is provided a novel joint-spacing balloon catheter for use in distracting a joint, and particularly for maintaining the distraction of a joint, as will hereinafter be discussed in detail.
More particularly, in this form of the invention, and looking next at
Elongated shaft 10 is preferably flexible, and preferably includes an internal stiffener 25 extending along at least a portion of its length so as to facilitate proper positioning of balloon 15 during use. Internal stiffener 25 could comprise a round or rectangular wire (e.g., such as shown in
Balloon 15 is preferably selectively inflatable/deflatable via an inflation/deflation lumen 45 extending through elongated shaft 10 and handle 20. An inflation/deflation control mechanism 50 is interposed between inflation/deflation lumen 45 and a supply port 55 which is connected to an appropriate fluid reservoir (not shown). By manipulating inflation/deflation control mechanism 50, the user is able to inflate/deflate balloon 15 as desired. Inflation/deflation control mechanism 50 may comprise a stopcock, a valve, a pump and/or other fluid control mechanisms. Balloon 15 preferably includes an atraumatic tip 60 at its distal end.
On account of the foregoing, joint-spacing balloon catheter 5 may have its balloon 15 set to its deflated state via inflation/deflation control mechanism 50, the deflated balloon may be advanced to a remote site using handle 20 and steering control mechanism 40, and then balloon catheter 5 may have its balloon set to its inflated state by further manipulating inflation/deflation control mechanism 50, whereby to support tissue and maintain the distraction of a joint, as will hereinafter be discussed in detail.
In another form of the present invention, there is provided a novel method for distracting a joint, preferably the hip joint, and preferably using novel joint-spacing balloon catheter 5.
More particularly, in this form of the invention, and looking now at
Next, joint-spacing balloon catheter 5, with balloon 15 set in its deflated state, is inserted into the space created between the ball of the femur and the acetabular cup. This may be done under direct visualization (i.e., using an endoscope inserted into the distracted joint), or under fluoroscopy, or both.
Then balloon 15 is inflated. See
Next, the distal force which was previously applied to the distal end of the leg is partially or fully released. Release of the full distraction force has the beneficial effect of completely eliminating the tension load imposed on the intervening tissue, whereas a partial release of the distraction force only partially eliminates the tension load imposed on the intervening tissue—however, even such partial release of the distraction force can still meaningfully reduce the tension load imposed on the intervening tissue, and it provides a safeguard in the event that balloon 15 should prematurely deflate, e.g., mid-procedure. The aforementioned partial or full release of the external distraction force allows the ball of the femur to seat itself on the inflated balloon, with the balloon acting as a spacer so as to maintain a desired spacing between the ball of the femur and the acetabular cup. Thus, joint distraction is maintained even though a substantial distraction force is no longer being applied to the distal end of the leg. Since joint distraction can be reliably maintained without the risk of damage to the intervening tissue from a substantial externally-applied distraction force, the traditional concern to complete procedures in 90 minutes or less is substantially diminished, and complications from joint distraction are greatly reduced. This is a very significant improvement over the prior art.
With the joint so distracted, the arthroscopic surgery can then proceed in the normal fashion.
Significantly, and in accordance with another novel aspect of the invention (see
Additionally, some procedures which would normally require the creation of an additional portal to access pathology can be accomplished without the creation of the additional portal, thereby reducing the visible scar and potential morbidity of the additional portal. This is also a significant improvement over the prior art.
At the conclusion of the arthoscopic surgery, a distal force is re-applied to the distal end of the leg so as to take the load off the inflated balloon, the balloon is deflated, and then the joint-spacing balloon catheter is removed from the interior of the joint.
Finally, the distal force applied to the distal end of the leg is released, so as to allow the ball of the femur to re-seat itself in its normal position within the acetabular cup.
With respect to the foregoing method of the present invention, it should be appreciated that joint-spacing balloon catheter 5 can be specifically located in the joint space so as to preferentially bias the position of the femoral head relative to the acetabulum when the pulling force on the distal end of the leg is relaxed and the ball of the femur transfers its load to (i.e., is seated on) the inflated balloon. For example, positioning joint-spacing balloon catheter 5 so that balloon 15 is more posterior in the joint causes the femoral head to settle in a more anterior position, which can improve visualization and access to the posterior acetabular rim.
With respect to the foregoing method of the present invention, it should also be appreciated that joint-spacing balloon catheter 5 can be placed in the joint space so as to provide better visualization and access to the peripheral compartment of the hip.
Thus it will be seen that the present invention provides a safe and simple way to significantly reduce trauma to intervening tissue in the leg when practicing leg distraction, since a substantial distally-directed force only needs to be applied to the distal end of the patient's leg long enough for the deflated balloon to be positioned in the distracted joint and for the balloon to thereafter be inflated—the distally-directed distraction force does not need to be maintained on the distal end of the patient's leg during the surgery itself. As a result, trauma to the intervening tissue is greatly reduced, and the surgeon no longer needs to limit the duration of distraction to 90 minutes or less in order to avoid damage to the intervening tissue. This is a very significant improvement over the prior art.
In addition, the use of the present invention enables more of the joint anatomy and pathology to be visualized and accessed, since supporting the ball of the femur on an inflated balloon allows the initial external distraction to be relaxed, and allows the leg to be manipulated on the inflated balloon while the joint is in a distracted state. By contrast, the leg cannot be manipulated in this manner while the leg is being distracted in a conventional manner, i.e., by a pulling force applied to the distal end of the leg. Therefore, arthroscopic procedures can be performed using the present invention which cannot be performed using conventional distraction techniques. This is a very significant improvement over the prior art.
Additionally, some procedures which would normally require the creation of an additional portal to access pathology can be accomplished without the creation of the additional portal, thereby reducing the visible scar and potential morbidity of the additional portal. This is also a significant improvement over the prior art.
It will be appreciated that balloon 15 preferably serves as a both a spacer and as a pivot support to allow the manipulation of the femur while the joint is distracted. Balloon 15 is constructed so as to be atraumatic in order to avoid damaging the anatomy, including the cartilage surfaces of the joint. At the same time, and as will hereinafter be discussed in further detail, balloon 15 may be appropriately textured and/or sculpted in order to maintain its position within the joint, preferentially to either one of the acetabulum or femur, while still allowing the opposing bone to move smoothly over the balloon surface.
In one preferred form of the invention, elongated shaft 10 has an outer diameter of about 0.040″ (or less) to about 0.250″ (or more). An outer diameter of approximately 0.120″ to 0.200″ is preferred for many hip applications.
If desired, a retractable sheath (not shown) may be provided over shaft 10 in order to cover balloon 15 prior to inflation.
And if desired, the distal end of shaft 10 can be pre-shaped with a bend so as to give joint-spacing balloon catheter 5 a directional bias at its distal end.
Furthermore, if desired, and looking now at
More particularly,
In one preferred form of the invention, balloon 15 is preferably approximately 28 mm in diameter, although it can also range from about 10 mm (or less) in diameter to about 50 mm (or more) in diameter if desired. Furthermore, the length of balloon 15 is preferably approximately 50 mm, although it can also range from about 10 mm (or less) in length to about 75 mm (or more) in length if desired. In this respect, it will be appreciated that balloons of various sizes may be used to address patients of different sizes, variations in anatomy, and/or different pathologies.
Balloon 15 may be inflated with a pressure of up to about 1000 psi, and is preferably inflated with a pressure of up to about 200 psi, and is most preferably inflated with a pressure of up to about 100 psi. In this respect it will be appreciated that it is generally accepted that a force of about 50-80 lbs. is sufficient to distract the hip joint. In order for joint-spacing balloon catheter 5 to support this force, it must provide sufficient pressure over a sufficient surface area (force=pressure X area). Although there are a number of different balloon sizes and operating pressures which can be envisioned, there are limitations on the balloon size and pressure to consider. On the one hand, the balloon must be large enough to cover a sufficient amount of cartilage such that the pressure on the cartilage is lower than that which would damage the cartilage. On the other hand, the balloon must be small enough so as to permit access to and visualization of the operative areas. Hence, there is an optimal range of balloon size and operating pressure, and this optimal range is dependent on tissue dynamics.
In one preferred form of the invention, balloon 15 is fabricated so as to be semi-compliant, although it can also be fabricated so as to be compliant or non-compliant if desired. Examples of semi-compliant balloon materials are polyurethane, nylon and polyether block amide (PEBA). An example of a compliant balloon material is silicone rubber. An example of a non-compliant balloon material is polyethylene terapthalate (PET). A compliant or semi-compliant balloon is generally preferred since it will deform under load to the shape of the surface which the balloon is contacting in order to help distribute load onto that surface. A semi-compliant balloon is generally most preferred since it will retain some aspects of its pre-load shape even when under load, which can be helpful in directing or maintaining bone positioning, particularly when the leg is being manipulated while in a distracted state. The thickness of the balloon material is preferably in the range of about 0.001″ to about 0.020″, and is most preferably between about 0.002″ and about 0.012″. The durometer of the balloon material is preferably in the range of about 30 Shore A to about 85 Shore D, and is most preferably between about 40 Shore D and about 85 Shore D.
If desired, the surfaces of balloon 15 can be textured (e.g., with dimples, ridges, etc.) or covered with another material (e.g., a coating or covering) so as to prevent slippage of the balloon along cartilage when the balloon is being used to support a joint. At the same time, this surface texture or non-slip covering is configured so as to engage the cartilage without causing cartilage damage. In one preferred form of the invention, only a portion of the outer surface of the balloon is textured or covered with a non-slip material. For example, the portion of the balloon which faces the acetabulum could be textured or covered with a non-slip material, but the portion of the balloon which faces the femoral head could be non-textured or non-covered, so as to keep the surface facing the acetabulum from slipping while allowing the surface facing the femoral head to slide relative to the femoral head. In another preferred form of the invention, a majority of the balloon surface is textured or covered with a non-slip material. In yet another preferred form of the invention, two or more different textures or non-slip coverings are provided on the outer surface of the balloon, e.g., depending on the particular cartilage surface which they may engage.
In yet another embodiment of the invention, the balloon is covered with a low friction material which enables slippage of the joint surface on the balloon. The low friction material may cover some or all of the balloon surface.
The balloon may comprise both low slippage and low friction coverings if desired.
Furthermore, if desired, fluoroscopic markings can be incorporated into or disposed on elongated shaft 10, or incorporated into or disposed on balloon 15, or incorporated into or disposed on another part of joint-spacing balloon catheter 5, so as to render the apparatus visible under X-ray. Such fluoroscopic markings may comprise radiopaque ink applied to the apparatus, radiopaque bands applied to the apparatus, radiopaque material incorporated in the construction of the apparatus, and/or a radiopaque fluid used to inflate the balloon (such as a contrast agent). By way of example but not limitation, a radiopaque band material could comprise platinum. By way of further example but not limitation, a radiopaque fluid could comprise a contrast agent such as Dodecafluoropentane.
In one preferred form of the invention, balloon 15 is preferably inflated with a liquid medium, e.g., saline; however, it could also be inflated with a gaseous medium, e.g., air. Among other things, the balloon can be inflated with a high viscosity fluid. This latter construction may be beneficial in the event of a balloon puncture as it would slow the pace of balloon deflation. If desired, a fluid could be used which changes viscosity when subject to changes in temperature, electrical charge, magnetic field, or other means. Alternatively, the balloon can be filled with a compound which increases in viscosity when exposed to saline. This latter construction can be advantageous in certain circumstances, e.g., during a balloon puncture, the escaping fluid would react to the saline present in the joint and could at least partially seal the puncture hole in the balloon.
Where balloon 15 is inflated with a gaseous medium, and that gaseous medium is air, inflation/deflation control mechanism 50 may comprise a pump, and supply port 55 may be open to the atmosphere.
In one aspect of the invention, and looking now at
In another embodiment (
Additionally, and looking now at
To improve resistance to kinking, or to provide the shaft with the desired stiffness and torsional characteristics, a braid or coil 71 (
In
Balloon 15 may also be in the form of an arc or other curvature (i.e., a geometry where one side has a greater curvature than the other side), or some other shape (e.g., U-shaped), so as to fit around the ligamentum teres. See
It is also possible to provide joint-spacing balloon catheter 5 with more than one balloon 15. Where more than one balloon is provided, the balloons can be disposed in series (i.e., end-to-end, such as is shown in
These multiple balloons 15 can also be disposed in a mutually-supporting configuration, as shown in
It is also possible to provide each of the balloons 15 with a plurality of separate internal chambers 75 (
If desired, balloons 15 can be formed so as to be puncture resistant in order to minimize the possibility of inadvertently deflating the balloon, e.g., with an errant surgical instrument. To this end, and looking now at
Furthermore, if desired, and looking now at
Alternatively, and looking now at
Alternatively, and looking now at
Alternatively, a different type of secondary structure can be deployed in balloon 15 in order to prevent balloon 15 from completely collapsing in the event that it is punctured. In one embodiment, and looking now at
In yet another embodiment (
If desired, joint-spacing balloon catheter 5 can include pressure regulation, e.g., a release valve (not shown) to ensure that a balloon is not inflated beyond a maximum level, or an alarm or other alert (not shown) to advise the user that a balloon has been inflated beyond a pre-determined level. This can be important to avoid damage to the patient's tissue or to reduce the risk of inadvertent balloon rupture.
Furthermore, a check valve (not shown) may be installed on the inflation port(s) 55 to enable joint-spacing balloon catheter 15 to be disconnected from the fluid reservoir while maintaining pressure in balloon 15.
It is also possible to place markings (e.g., longitudinal lines) along the body of balloon 15, or to color the balloon material, so as to improve endoscopic visualization of the balloon, including to show the degree of balloon inflation. Alternatively, the fluid used to inflate the balloon could be colored, or the balloon surface could have texture, in order to aid visualization of the balloon.
Alternatively, a transparent, thick-walled balloon 15 can be used to increase visualization of the balloon by increasing the refraction of light, which will make the balloon foggy in appearance. Alternatively, a coating could be applied to the balloon material which improves the endoscopic visualization of the balloon. Alternatively, a second balloon or an expandable extrusion could be placed over the primary balloon so as to improve endoscopic visualization. The second balloon and/or expandable extrusion may be colored for improving endoscopic visualization. This configuration can also add to the puncture resistance of the primary balloon and assist in the delivery and retrieval of the primary balloon.
The joint-spacing balloon catheter 5 may also comprise a sensor (not shown). The sensor can measure the temperature of the surrounding tissue or fluid in the joint (e.g., the sensor may be a temperature sensor). The sensor may also detect characteristics of the adjacent cartilage, such as thickness, density, and/or quality (e.g., the sensor may be an ultrasound device, etc.). The sensor could be located on shaft 10 or on balloon 15, or on another portion of joint-spacing balloon catheter 5.
In the foregoing description, the external distraction of the limb is generally discussed in the context of applying a distally-directed distraction force to the distal end of the leg. However, it should be appreciated that the distally-directed distraction force may be applied to another portion of the leg, e.g., to an intermediate portion of the leg, such as at or about the knee. Thus, as used herein, the term “distal end of the leg” is meant to include substantially any portion of the leg which is distal to the ball of the femur, such that by applying the external distraction force to the leg, a tension load is imposed on the intervening tissue. Furthermore, as used herein, the term “intervening tissue” is intended to mean the tissue which is interposed between the location where the external distraction force is applied to the leg and the ball of the femur.
The present invention also preferably comprises the provision and use of a novel inflatable perineal post for facilitating joint distraction.
More particularly, and looking now at
The inflatable balloon 130 of the inflatable perineal post 120 is preferably constructed out of a semi-compliant material, but it could also be compliant or non-compliant. The inflatable balloon 130 of the inflatable perineal post 120 may involve a covering (not shown) for contact with the patient; this covering may be a non-slip material. The inflatable balloon 130 is preferably inflated with a manual or electric pump. The inflatable perineal post 120 could include a read-out panel displaying the balloon pressure.
The inflatable perineal post 120 may also comprise physiologic sensors (not shown) for monitoring parameters such as patient skin temperature and blood flow. Such parameters may be reflective of patient conditions of interest to the surgeon, e.g., a falling patient skin temperature is frequently indicative of reduced blood flow. These physiologic sensors could be incorporated into the surface of the balloon, or they could be separate sensors which are included as part of a kit provided with the inflatable perineal post. The physiologic sensors are adapted to be connected to a monitor so as to provide read-outs on the monitor.
In use, the deflated perineal post balloon is positioned between the patient's legs, the joint is distracted by pulling on the distal end of the leg so that the ball of the femur is spaced from the acetabular cup, the perineal post balloon is inflated, a joint-spacing balloon catheter 5 is inserted into the distracted joint, the balloon 15 is inflated, the force applied to the distal end of the leg is relaxed so that the ball of the femur settles back down onto the one or more inflated balloons 15, and then the perineal post balloon 130 is at least partially deflated. At this point the arthroscopic surgery can be conducted without trauma to the patient's tissue, due to either the distal distraction of the leg or due to engagement of the perineal post with the tissue of the patient. At the conclusion of the surgery, the distal end of the leg is pulled distally again, the perineal post balloon 130 is inflated, the joint-spacing balloon 15 is deflated, the joint-spacing balloon catheter 5 is removed from the joint, and the joint is reduced. Alternatively, the perineal post balloon could be inflated prior to pulling on the distal end of the leg. Or, alternatively, the perineal post balloon 130 could be deflated prior to withdrawal of the force being applied to the distal end of the leg. In some cases, only one of either (i) pulling on the leg, or (ii) inflating of the perineal post is performed in order to remove or re-position the joint-spacing balloon.
If desired the inflatable perineal post 120 may be used to replace a standard perineal post, and is used in conjunction with a standard traction table; in other words, in this form of the invention, the inflatable perineal post 120 is not used in conjunction with a joint-spacing balloon catheter 5.
In one preferred form of the present invention, the aforementioned novel method for distracting the joint is implemented using the aforementioned novel joint-spacing balloon catheter 5 and the aforementioned inflatable perineal post 120.
More particularly, in this form of the invention, the hip joint is first distracted by pulling on the distal end of the leg just above the ankle, and then inflating the inflatable perineal post, where the perineal post is positioned between the patient's legs. The leg may be adducted so as to lever the femur laterally. Alternatively, the inflatable perineal post could be inflated prior to the distal end of the leg being pulled distally.
Next, the surgeon identifies a portal location for joint-spacing balloon catheter delivery. Then a needle is placed into the joint, the stylet is removed, a guidewire is delivered through the needle, and then the needle is removed. The guidewire can be placed in the desired delivery path of the joint-spacing balloon catheter 5.
An arthroscopic cannula or outer guiding member may then be emplaced if desired; in this instance, the guidewire may be removed if desired.
Next, a joint-spacing balloon catheter 5 of the appropriate size is selected from a kit providing a range of differently-sized joint-spacing balloon catheters. Then the joint-spacing balloon catheter 5 is delivered over the guidewire (either percutaneously or through a cannula) to the target site between the femoral head and the acetabulum. The joint-spacing balloon catheter 5 may be rotated as appropriate if there is asymmetry in the balloon's shape. Alternatively, the joint-spacing balloon catheter 5 may be delivered through a cannula without the use of a guidewire.
Next, a syringe (or other inflation device) is secured to the joint-spacing balloon catheter 5, and the balloon 15 is inflated to the desired pressure and/or size. If there is more than one balloon 15, the additional balloon(s) can be inflated. If the additional balloon(s) are used to affect the direction of joint spacing, the pressure and/or size of each balloon is adjusted so as to achieve the desired joint spacing direction.
Once the balloon has been inflated to the desired pressure and/or size, the distraction force applied to the leg is at least partially removed, allowing the head of the femur to rest on the inflated balloon (which is itself supported by the acetabulum).
Additionally, the inflatable perineal post 120 is deflated as appropriate; this could occur before the leg distraction force is released.
The balloon 15 can be re-positioned by re-applying distraction force to the leg and/or re-inflating the inflatable perineal post 120, deflating balloon 15 and re-positioning the joint-spacing balloon catheter 5, re-inflating the balloon of the joint-spacing balloon catheter, then releasing the leg distraction and/or deflating the inflatable perineal post. The balloon 15 may be placed in a location which directs the distraction in a preferred direction. Alternatively, where the joint-spacing balloon catheter comprises a plurality of balloons, the balloons may be inflated to different sizes and/or pressures in order to direct the joint distraction in a preferred direction.
With the balloon maintaining the joint distraction, the leg may be manipulated (i.e. rotated, flexed, etc.) in order to visualize and access pathology through the established portals.
Then the arthroscopic surgery is conducted. The leg may be manipulated a number of times through the procedure in order to visualize, access and treat pathology.
At the conclusion of the arthroscopic surgery, the hip joint is distracted again, e.g., by pulling on the distal end of the leg just above the ankle, so as to lift the head of the femur off the balloon. The perineal post balloon may be inflated. The balloon 15 of the joint-spacing balloon catheter is deflated and the joint-spacing balloon catheter is removed. Thereafter, the distraction force applied to the leg may be removed, allowing the head of the femur to settle back on the acetabulum.
In another form of the invention, while the distal end of the leg is held stationary, the perineal post 120 is inflated to break the suction seal of the hip joint and enable the joint-spacing balloon catheter 5 to be placed in the joint and inflated. In this case, no pulling on the leg is performed. This would have the benefit of eliminating a piece of equipment from the surgery and reducing the corresponding surgical time associated with using that equipment.
In yet another form of the invention, the joint-spacing balloon catheter 5 can perform some or all of the joint distraction. In one embodiment, a first joint-spacing balloon catheter 5 is placed adjacent to the femoral head and the balloon is inflated. The leg is then manipulated in abduction or adduction (depending on balloon location), thus levering the femoral neck against the balloon. This levering creates a gap at the acetabular rim. A second joint-spacing balloon catheter 5 is then inserted into the gap and delivered into the joint space (the space between the femoral head and the acetabulum). The balloon of the second joint-spacing balloon catheter is then inflated and distracts the joint; that is, opens up the joint space. In one embodiment, the first balloon is placed on the lateral/superior aspect of the femoral neck. Once the second balloon is inflated, the first balloon can be deflated and withdrawn. The first balloon may be of a different size and shape as the second balloon. It also may be inflated to a different pressure.
The joint-spacing balloon catheter 5 and the inflatable perineal post 120 may be offered as part of a single kit. A guidewire or obturator, outer guiding member and a balloon inflation device may additionally be provided.
It should be appreciated that the present invention may be used for distracting the hip joint in an open, more invasive procedure. The present invention can also be used in hip joint pathologies where joint distraction is not needed but space creation is needed, e.g., to visualize and/or to address pathologies in the peripheral compartment or pathologies in the peritrochanteric space. Additionally, the present invention may be used for distracting joints other than the hip joint (e.g., it may be used to distract the shoulder joint).
It should be understood that many additional changes in the details, materials, steps and arrangements of parts, which have been herein described and illustrated in order to explain the nature of the present invention, may be made by those skilled in the art while still remaining within the principles and scope of the invention.
This patent application claims benefit of: (i) pending prior U.S. Provisional Patent Application Ser. No. 61/210,315, filed Mar. 17, 2009 by Julian Nikolchev et al. for JOINT SPACING BALLOON CATHETER (Attorney's Docket No. FIAN-28 PROV); (ii) pending prior U.S. Provisional Patent Application Ser. No. 61/268,340, filed Jun. 11, 2009 by Julian Nikolchev et al. for METHOD AND APPARATUS FOR DISTRACTING A JOINT, INCLUDING THE PROVISION AND USE OF A NOVEL JOINT-SPACING BALLOON CATHETER AND A NOVEL INFLATABLE PERINEAL POST (Attorney's Docket No. FIAN-42 PROV); (iii) pending prior U.S. Provisional Patent Application Ser. No. 61/278,744, filed Oct. 9, 2009 by Julian Nikolchev et al. for METHOD AND APPARATUS FOR DISTRACTING A JOINT, INCLUDING THE PROVISION AND USE OF A NOVEL JOINT-SPACING BALLOON CATHETER AND A NOVEL INFLATABLE PERINEAL POST (Attorney's Docket No. FIAN-49 PROV); and (iv) pending prior U.S. Provisional Patent Application Ser. No. 61/336,284, filed Jan. 20, 2010 by Julian Nikolchev et al. for METHOD AND APPARATUS FOR DISTRACTING A JOINT, INCLUDING THE PROVISION AND USE OF A NOVEL JOINT-SPACING BALLOON CATHETER AND A NOVEL INFLATABLE PERINEAL POST (Attorney's Docket No. FIAN-53 PROV). The four (4) above-identified patent applications are hereby incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
61210315 | Mar 2009 | US | |
61268340 | Jun 2009 | US | |
61278744 | Oct 2009 | US | |
61336284 | Jan 2010 | US |