Reference will now be made in detail to the present preferred embodiment of the invention, an example of which is illustrated in the accompanying drawings. Wherever possible, the same reference numbers are used throughout the drawings to refer to the same or like parts.
Referring to
According to a first embodiment of the invention, a method for interpreting physiological data includes the step of obtaining capnographic data, preferably in the form of a capnogram. Such data may be obtained by a device known as a capnograph. Capnographs having data acquisition capabilities suitable for use in an embodiment of the invention are commercially available and include, for example, the capnograph monitor sold under the trademark “VitalCap” by Oridion Corp of Jerusalem, Israel, or the capnograph sold under the trademark “LifeCap” by Medtronic Emergency Response Systems Inc. of Redmond, Wash. USA. Reference is now made to
Analysis of the digital waveform measurement data is performed by a processor 22 in a manner described in detail below. In some embodiments, data stored in a read-only memory (ROM) 23 may be used by the processor 22 in its analysis. The processor 22 reports the results of its analysis through an input/output (I/O) device 24. I/O device 24 may be a visual display such as a text screen, LCD screen capable of displaying graphics or pictures, LED indicator lights, etc., or an aural indicator such as a speaker which outputs recorded voiced messages or sounds such as beeps, or other sound-emitting device. A user may input commands or information to the processor 22 through the I/O device 24.
Typically, the capnogram corresponding to a single tidal breath has been described as if it had four visible linear phases. Visually, each of these four phases (i.e., on the time axis, from T0 to T1; from T1 to T3 from T3 to T5, and from T5 to T6), has conventionally been approximated as a straight line. Thus, the capnogram for a single tidal breath of a subject with normal lung function has been described as a trapezoid shape. The two phases from T1 to T3 and from T3 to T5, and the two corresponding line segments of the trapezoidal model, are commonly referred to as the “Expiratory Rise” and the “Alveolar Plateau”. Since each of the two line segments in the trapezoidal model can mathematically be described by two independent variables (e.g.: the slope and intercept, or, equivalently, the angle and length), a total of four measures suffice to completely capture the information contained in the Expiratory Rise and the Alveolar Plateau in the trapezoidal approximation. Most commonly, the angle of the line, referenced to the horizontal, and the duration of the phase have been used to characterize the capnogram in this model. However, the segmentation of exhalation into two discrete, linear phases, characteristic of the trapezoidal model, is an artificial simplification, ignoring deviations from linearity within a phase and the character of turning points between phases.
The method of analysis performed by the processor 22 according to an embodiment of the invention will now be described. In an embodiment of the invention, the processor 22 performs an analysis in which it models one or more segments of a capnogram as a nonlinear function of time, using curve-fitting techniques. In an embodiment of the invention, modeling the expiratory rise and alveolar plateau phases of the capnogram using a non-linear curve fit characterized by four calculated parameters has been found to be advantageous. This results in diagnostic utility for the capnogram that heretofore was unavailable.
In a preferred embodiment, a cumulative Weibull function is used as the non-linear function which is fit to the capnographic data by curve-fitting techniques well-known to those in the art. A cumulative Weibull function has the following form:
where transition height=a;
The parameters a, b, c and d can be used to characterize a particular cumulative Weibull curve fit to a particular data set.
Other non-linear functions can be used in embodiments of the present invention as well. For example, a cumulative Gaussian curve or logistic curve, can be used as the non-linear function to be fit to the data points of a particular capnogram.
This data was obtained by collecting capnographic data from a population of 30 adults with normal lung function as determined by history, physical examination, and spirometry. For each subject, a 3-minute data-gathering period was commenced when the subject was observed to be consistently executing normal tidal breathing. Data was collected continuously through the subsequent recording period, resulting in approximately 30 breaths being recorded from each subject.
At the conclusion of the recording, data were transmitted to a desktop computer for analysis. A three-phase pre-analysis procedure was performed on each subject recording to identify fiducial points and to assure waveform quality prior to modeling and measurement. These fiducial points are each defined by their sequence location and associated CO2 value (P(time, amp)).
First, each continuous capnogram was scanned to identify transitions between inhalation and exhalation. The start of inhalation was identified using slope and amplitude criteria: a fiducial mark, P1(time, amp), was placed at the point in the capnogram where the CO2 concentration exceeded 4 mmHg and the rising slope exceeded +xx mmHg/sec. Another fiducial point identifying the start of exhalation P3(time, amp) was placed at the location of the end-tidal CO2 value, defined as a falling slope of more than −zz mmHg/sec and an amplitude less than 4 mmHg of the maximum value since the preceding start of inhalation.
Two additional fiducial marks were placed to mark the start of the Alveolar Plateau (P2(time, amp)) and the return to baseline on inhalation (P4)time, amp)). P2 was established at the crossing point of two regression lines fit through the Expiratory Rise and alveolar plateau, generally meeting above the “knee” of the ascending capnogram. P4 was defined as the point at which the CO2 value fell below 4 mmHg in association with a falling slope of more than −yy mmHg/sec.
Finally, the annotated record was scanned to eliminate artifacts caused by coughing, sighing, and other normal physiologic and behavioral variants.
Non-linear functions suitable for characterizing the capnogram were determined through a statistical curve-fitting procedure. All normal breaths were extracted, superimposed, and aligned on their P1 fiducial marker. A median waveform was created from the population by taking the point-by-point median values at each sample point from (P1−10) to (P1+70), which included two seconds of exhalation. An automated curve-fitting system, the TableCurve2D v5.01 software program sold by Jandel Scientific of San Rafael, Calif., was used to fit candidate functions to the median capnogram, and to rank them in order of goodness-of-fit. The cumulative Weibull distribution function was selected as the best four-parameter approximation to the exhalation capnogram and is described by parameters a and b, defining the height and centering of the inflection point of the rising phase of the curve, and c and d, which define the width and trajectory of the takeoff and plateau transitions.
It will be readily understood that parameters for cumulative Weibull curves for representative or standard capnograms for various disease states and other abnormal lung function states can be determined by sampling subjects with the particular disease states or abnormal lung function desired to be studied, and fitting a cumulative Weibull function to the median waveform using techniques similar to those described above.
In the embodiment of
In the embodiment of
Calculated parameters characterizing the cumulative Weibull or other non-linear function modeling a capnogram can also be used in other manners to provide a diagnostic aid or therapeutic feedback. For example, one or more of the parameters can be examined to see if it is outside a given range of values. An output in the form of a visual or an audible indicator or alarm can be generated to alert a device user of an out-of-range capnogram parameter.
Results of the comparison step may be used to initiate a therapy or a process performed by an associated device. For example, a nebulizer may be provided with a controller that receives the result of the comparison step and activates the nebulizer therapy when a particular comparison step result is received.
The parameters of the non-linear function may be analyzed by the processor 22 in combination with other data contained in or derivable from the capnographic data, such as respiratory rate, inhalation or exhalation time, or EtCO2, or data from other physiologic sensors, such as SpO2, to provide additional diagnostic utility.
It will be apparent to those skilled in the art that various modifications and variations can be made to the above-described embodiment(s) of the invention without departing from the spirit and scope of the invention. Thus, it is intended that the present invention cover the modifications and variations of the embodiments provided they come within the scope of the appended claims and their equivalents.
Number | Date | Country | |
---|---|---|---|
60817175 | Jun 2006 | US |