The present apparatus and methods relate generally to energy emitting apparatus and methods for providing a medical therapy. The apparatus and methods may provide for central and peripheral nerve and other tissue modulation or stimulation therapies
The OAB and UI market in the United States is well over a $12 billion a year industry. It affects over 16% of all Americans, for a total U.S. market of approximately 34 million men and women each year. Due to social stigmas attached to OAB and UI, as well as misunderstanding of the signs and symptoms associated with OAB and UI, only 40% of those affected (13.6M) seek treatment. Of those 13.6 million individuals, nearly 30% are unsatisfied with their current therapy.
The use of pulsed electromagnetic stimulation (PES) has been well established as a beneficial therapy in a variety of medical applications. The scientific principle behind this technology is that an electric current passed through a coil will generate an electromagnetic field. These fields, in turn, have been shown to induce current within conductive materials placed within the field. When applied to the human body, pulsed electromagnetic stimulation has been found to be an effective method of stimulating nerves resting within the electromagnetic field. Recent data highlights the beneficial effects of invasive, needle-based electrostimulation (ES) of the posterior tibial nerve in individuals with OAB and UI. ES has been found to modulate bladder dysfunction through its action on the pudendal nerve and the sacral plexus which provides the major excitatory input to the bladder.
Current treatment options for OAB and UI are exercise and behavioral modifications, pharmacological therapies, surgical intervention, and neuromodulation. Although each of these treatment options targets the UI and OAB populations, each has severe limitations.
Exercise and behavioral modifications often require patients to adhere to stringent routines, including scheduled voiding, maintenance of a bladder diary, and intense exercise regiments. While this may be a viable option for a small group of highly dedicated individuals, its daily impact on one's life makes it an unattractive option for most individuals.
Pharmacological intervention is the most widely prescribed therapy for OAB and UI. Unfortunately, as with the ingestion of any chemical, patients are often subject to side effects from their drug therapy. This is especially detrimental in older and elderly patient populations where interaction with other prescribed medications can have adverse effects. Further, there is a high rate of dissatisfaction, approximately 30%, amongst individuals using pharmacological treatment.
Surgical intervention is an extremely invasive treatment and often results in the long-term, and in some cases permanent, requirement for catheterization. The high expense of these procedures, coupled with the negative impact the procedures have on the patients quality of life, make this an option only when all other treatment options have been exhausted.
Neuromodulation is another treatment alternative for OAB and UI patients. Sacral nerve stimulation (SNS) has shown itself to be an effective treatment option for those with OAB or UI. However, the procedure requires the permanent implantation of an electrical stimulation device in the patient. One estimate puts the cost at nearly $14,000 with additional routine care costs of $593 per patient per year. Additionally, SNS's risk of battery failure, implant infection, and electrode migration, lead to a high reoperation rate and make this procedure unattractive.
More recently, the introduction of a posterior tibial nerve stimulator, often referred to as SANS, has shown itself to be another neuromodulation alternative. Yet as is the case with other forms of neuromodulation, this system is invasive in its nature. It requires the insertion of a needle two inches into the patient's ankle region in order to stimulate the posterior tibial nerve. As well, it requires a minimum of 12 sessions for initial treatment, with the possibility of additional sessions needed for maintenance. Despite its high cost and invasive nature, though, an abundance of published peer-reviewed clinical trials demonstrate the safety and efficacy of the SANS therapy.
In certain variations, a method for providing transdermal electrical stimulation therapy to a patient is provided. The method may include positioning a stimulator electrode over a glabrous skin surface overlying a target nerve of a subject. Electrical stimulation may be delivered through or across the glabrous skin surface to the target nerve to stimulate the target nerve, while remaining safe and tolerable to the patient. Electrical stimulation may be delivered at frequencies that may be painful or intolerable when applied over non-glabrous surfaces of the body. The electrical stimulation may be utilized to treat various conditions, e.g., urinary incontinence and overactive bladder.
In certain variations, an applicator, e.g., an ergonomic applicator, for providing transdermal electrical stimulation therapy to a patient is provided. The applicator may be configured to position a stimulator electrode over a glabrous skin surface of the subject to deliver transdermal electrical stimulation through or across the glabrous skin surface to an underlying target nerve, resulting in stimulation of the target nerve.
In certain variations, a method for providing an energy based stimulation therapy to a subject is provided. The method may include positioning an energy emitting device in proximity to a glabrous surface overlying a target tissue. Energy may be delivered through the glabrous skin surface to the target tissue to stimulate the target tissue.
In certain variations, another method for providing an energy based stimulation therapy to a subject is proved. The method may include positioning an energy emitting device in proximity to a skin surface overlying a target nerve. Energy may be delivered at a frequency of about 1 Hz to about 30 Hz through the skin surface to the target nerve, thereby generating motor and/or sensory nerve conduction of the target nerve while remaining safe and tolerable to the subject. Optionally, energy may be delivered at less than 10 Hz to generate nerve conduction.
In certain variations, systems for electromagnetic induction therapy may include one or more conductive coils disposed within or along an applicator. The coils may be configured to generate a magnetic field focused on a target nerve, muscle or other body tissues in proximity to the coil. One or more sensors may be utilized to detect electrical conduction in the target nerve, to detect a muscular response caused by an electrical conduction in the target nerve, or to detect stimulation of a nerve, muscle or other body tissues and to provide feedback about the efficacy of the applied electromagnetic induction therapy. A controller in communication with the sensor may be adjustable to vary a current through the at least one coil so as to adjust the magnetic field focused upon the target nerve, muscle or other body tissues. Optionally, a user or patient may detect stimulation of a nerve, muscle or body tissue and the therapy may be adjusted based on feedback from the user or patient.
In certain variations, the applicator may be configured to intermittently apply or deliver pulsed magnetic fields to a target nerve, muscle or tissue without causing habituation of the target nerve, muscle or tissue.
In certain variations, methods of electromagnetic induction therapy may include one or more of the following steps. A first portion of a patient's body may be positioned relative to or in proximity to an applicator or an applicator may be positioned relative to or in proximity to a first portion of a patient's body, such that a target nerve, muscle or tissue within the first portion of the body is in proximity to one or more conductive coils disposed within or along the applicator. A current may be passed through a coil to generate a magnetic field focused on the target nerve, muscle or tissue. An electrical conduction through the target nerve, a muscular response caused by an electrical conduction through the target nerve or stimulation of a nerve, muscle, or body tissue may be detected by a sensor positioned along a second portion of the body. A signal from the sensor indicative of the electrical conduction or stimulation may be received, which provides feedback about the efficacy of the applied electromagnetic induction therapy. The current may be adjusted by a controller in communication with the conductive coils based on the feedback.
Optionally, a user may detect stimulation of a nerve, muscle or body tissue and the therapy may be adjusted based on feedback from the user. In certain variations, pulsed magnetic fields may be intermittently applied or delivered a target nerve, muscle or tissue without causing habituation of the target nerve, muscle or tissue. Such intermittent magnetic fields may be used to treat chronic conditions, e.g., chronic pain, without causing habituation.
In certain variations, applicators may be ergonomic or may be designed or configured to accommodate, approximate or be positioned relative to or in proximity to specific regions of the body or anatomy. The specific regions of the body or anatomy may be positioned relative to the applicators, or the applicators may be positioned relative to the specific regions of the body or anatomy to treat various conditions, for example, osteoarthritis, arthritis, back or neck pain, atrophy or paralysis, chronic pain, phantom or neuropathic pain, neuralgia, migraines, orthopedic conditions.
In one embodiment, the invention provides a method or device for providing transdermal electrical stimulation therapy to a subject including positioning a stimulator electrode over a glabrous skin surface overlying a palm of the subject and delivering electrical stimulation via a pulse generator transdermally through the glabrous skin surface and to a target nerve or tissue within the hand to stimulate the target nerve or tissue within the hand so that pain felt by the subject is mitigated. The pulses generated during the electrical stimulation therapy may include pulses of one or two or more different magnitudes.
In some embodiments, the electrical stimulation is delivered at a frequency of about 5 Hz to about 60 Hz, while remaining safe and tolerable to the subject. In some embodiments, the stimulator electrode is a surface electrode. The electrical stimulation may be delivered intermittently or on a chronic basis.
In some embodiments, the device includes a sensor (which may be a surface electrode, or other type of electrode) for detecting nerve stimulation which receives a signal from the sensor indicative of the detected electrical stimulation thereby providing a feedback about the efficacy of the applied electrical stimulation therapy such that the therapy is adjusted or optimized. An M-wave and/or F-wave may be detected by the sensor. A parameter of the M-wave and/or F-wave may be displayed to the subject.
In some embodiments, the feedback is queried by a controller such that the electrical stimulation therapy is adjusted to ensure that a minimum amount of energy is being applied to stimulate the target nerve within the hand while reducing the risk of burns or intolerance. Alternatively, the feedback may be queried such that the positioning of the stimulator electrode is adjusted to optimize the electrical stimulation therapy.
In some embodiments, a return electrode may be positioned on the subject to facilitate penetration of an electrical current through the glabrous surface to stimulate the target nerve within the hand. The stimulator electrode may be attached to the glabrous skin surface with an adhesive. The stimulator electrode may be positioned over the glabrous skin surface with an ergonomic applicator. The applicator may be a glove or brace configured to be positioned against the palmar surface of a hand such that the electrode delivers electrical stimulation to a target nerve within the hand.
In some embodiments, the electrical stimulation is automatically paused for a preset amount of time every 10 minutes to overcome habituation. In some embodiments, the electrical stimulation is automatically paused for a preset amount of time every 10 minutes based on feedback provided by a sensor regarding target nerve stimulation or habituation.
In some embodiments, once tetany is detected in a patient, indicating that a threshold frequency of the applied stimulation has been reached, the strength of the applied stimulation is automatically or manually decreased.
In some embodiments, electrical stimulation is provided by delivering a cycle comprising a preset number of pulses followed by a pause in stimulation, followed by a preset number of pulses, and repeating the cycle as necessary.
In some embodiments, the device is used to treat migraine pain.
Other features and advantages will appear hereinafter. The features and elements described herein can be used separately or together, or in various combinations of one or more of them.
The drawings constitute a part of this specification and include exemplary embodiments of the invention, which may be embodied in various forms. It is to be understood that in some instances various aspects of the embodiments may be shown exaggerated or enlarged to facilitate an understanding of the embodiments.
In certain variations, various apparatus and methods for providing magnetic induction therapy or electrical stimulation therapy are provided. In certain variations, various apparatus and methods may provide for central and peripheral nerve and other tissue modulation or stimulation therapies, including both excitation and blocking of nerve impulses. In certain variations, a low frequency induction therapy may be performed. In certain variations, these apparatus and methods may be useful in the treatment and prevention of urinary incontinence (UI), overactive bladder (OAB) and other conditions.
In certain variations, apparatus and methods for magnetic induction therapy, in which dosage of magnetic energy can be regulated according to conduction in a target nerve exposed to the magnetic field are provided.
In certain variations, apparatus and methods for magnetic induction therapy, in which the flow of magnetic energy can be adjusted directionally by the patient or a healthcare provider without altering the position of a housing containing conductive coils that produce the magnetic field are provided.
In certain variations, apparatus and methods for treating a variety of ailments by providing energy to a target nerve, for example magnetic energy, electrical energy or ultrasound energy, at a location and in an amount optimized by detecting conduction in the target nerve are provided.
In certain variations, an energy emitting apparatus for delivering a medical therapy that includes one or more energy generators, a logic controller electrically connected to the one or more energy generators, and one or more sensors for detecting electric conduction in a target nerve, which are connected to the logic controller is provided. The one or more energy generators produce energy focused on the target nerve upon receiving a signal from the logic controller, and the applied energy is varied by the logic controller according to an input provided by the one or more sensors based on electric conduction in the target nerve. The feedback provided by the sensors to the logic controller about the efficacy of the applied treatment causes the logic controller to modulate the current transmitted to the coils.
The applied energy may be a magnetic field, an electrical field, an ultrasound, a visible light, or an infrared or an ultraviolet energy. When a magnetic field is applied, the energy-emitting device is an apparatus that provides a magnetic induction therapy and that includes one or more conductive coils disposed in an ergonomic housing. A logic controller is electrically connected to the one or more coils, and one or more sensors detect electric conduction in the target nerve and are connected to the logic controller so to provide a feedback to the logic controller. The conductive coils receive an electric current from the logic controller and produce a magnetic field focused on a target nerve, and the electric current fed by the logic controller is varied by the logic controller according to an input provided by the sensors, thereby causing amplitude, frequency or direction of the magnetic field, or the firing sequence of the one or more coils, to be varied according to the efficiency of the treatment provided to the target nerve. In certain variations, the housing containing the conductive coils may be a flexible wrap, a cradle or a garment, and the coils may be overlapping and/or be disposed in different positions within the housing, so to generate a magnetic field on different body parts with the desired direction and amplitude.
The one or more coils may be stationary or movable within the housing, making it possible to optimize the direction of magnetic flow to the target nerve by disposing the coils in the most effective direction. In different variations, the coils may be movable manually by acting on a knob, lever, or similar type of actuator, or may be translated automatically by the logic controller in response to the input provided by the sensors. When a preferred position for the coils has been established, the coils may be locked in position and maintain that position during successive therapy sessions. In other variations, the sensors may be incorporated within the housing, or instead may be disposed on a body part of interest independently of the housing.
In still other variations, the inductive coils are disposed in a housing that is situated externally to a patient's body, and additional inductive coils are implanted into the body of the patient and are magnetically coupled to the external inductive coils. With this coil arrangement, energy may be transmitted from the external coils to the internal coils either to recharge or to activate an implantable device. In yet other variations, the electric current may varied by the logic controller both on the basis of an input provided by the one or more sensors and also an input provided by the patient according to a muscular response she has perceived, for example, the twitching of a toe after application of the magnetic field.
In yet other variations, the source of energy for nerve stimulation may be electrical energy and nerve conduction may be detected at a site sufficiently distant from the site of stimulation, so to enable detection of nerve conduction despite any interference from the direct electrical stimuli. In these variations, direct electrical stimulation of nerve and muscle may be tailored to provide optimal therapy and, in the case of electrode migration or other electrode malfunction, to report lack of stimulation of the bodily tissues. Furthermore, these variations enable a reduction in power requirement, because control of the signal is provided by the sensor to the signal generator loop.
In other variations, an energy emitting system for providing a medical therapy is provided. The system may include one or more conductive coils disposed within or along a housing and configured to generate a magnetic field focused on a target nerve in proximity to coils; one or more sensors in the form of microneedle patch configured to detect electrical conduction in the target nerve; and a controller coupled to the conductive coils and optionally in communication with the sensor.
In other variations, an energy emitting system for providing a medical therapy is provided. The system may include one or more microneedle patches having one or more microneedle arrays deposited on a surface of one or more electrodes and configured to generate or deliver an electrical or magnetic stimulus or field focused on a target nerve in proximity to the microneedle patch; one or more sensors configured to detect electrical conduction in the target nerve; and a controller coupled to the conductive coils and optionally in communication with the sensor. Optionally, the above variations may incorporate an electrode needle. Optionally, the above variations or systems may be utilized without a sensor or mechanism for detecting conduction or stimulation.
Methods of use of the above apparatus, systems and variations thereof for treating various conditions are also described herein.
Referring first to
Coil wrap 20 may be manufactured from a variety of materials suitable for wearing over ankle 22. Preferably, coil wrap is produced from a soft, body-compatible material, natural or synthetic, for example, cotton, wool, polyester, rayon, Gore-Tex®, or other fibers or materials known to a person skilled in the art as non-irritating and preferably breathable when tailored into a garment. Coil wrap 22 may even be manufactured from a molded or cast synthetic material, such as a urethane gel, to add extra comfort to the patient by providing a soft and drapable feel. Additionally, coil wrap 20 may be produced from a single layer of material or from multiple material layers and may include padding or other filling between the layers.
Coil wrap 20 contains one or more conductive coils 26 arranged to produce a pulsed magnetic field that will flow across tibial nerve 24 and generate a current that will flow along tibial nerve 24 and spread along the length of tibial nerve 24 all the way to its sacral or pudendal nerve root origins. Coils 26 may be a single coil shaped in a simple helical pattern or as a FIG. eight coil, a four leaf clover coil, a Helmholtz coil, a modified Helmholtz coil, or may be shaped as a combination of the aforementioned coils patterns. Additionally, other coil designs beyond those mentioned hereinabove might be utilized as long as a magnetic field is developed that will encompass tibial nerve 24 or any other target nerve. When a plurality of coils is utilized, such coils may be disposed on a single side of ankle 22, or may be disposed on more than one side, for example, on opposing sides, strengthening and directionalizing the flow of the magnetic field through tibial nerve 24 or other peripheral nerves of interest.
Coil wrap 20 is preferably configured as an ergonomic wrap, for example, as an essentially cylindrical band that can be pulled over ankle 22, or as an open band that can be wrapped around ankle 22 and have its ends connected with a buckle, a hoop and loop system, or any other closing system known to a person skilled in the art. By properly adjusting the position of coil wrap 20 over ankle 22, a patient or a health care provider may optimize the flow of the magnetic field through tibial nerve 24, based on system feedback or on sensory perceptions of the patient, as described in greater detail below.
The electric current that produces the magnetic field by flowing through coils 26 is supplied by a programmable logic controller 28, which is connected to coils 26, for example, with a power cord 32. A sensor 30 that feeds information to logic controller 28 is also provided, in order to tailor the strength of the magnetic field and control activation of coils 26 based on nerve conduction. The purpose of sensor 30 is to detect and record the firing of the target nerve and to provide related information to logic controller 28, so to render the intended therapy most effective. For example, sensor input may cause logic controller 28 to alter the strength or pulse amplitude of the magnetic field based on sensor input, or fire the coils in a certain sequence.
In this variation, as well as in the other variations described hereinafter, sensor 30 may include one or more sensor patches and may be placed at different distances from the region of direct exposure to the magnetic field. For example, sensor 30 may be configured as a voltage or current detector in the form of an EKG patch and may be placed anywhere in the vicinity of the target nerve to detect its activation. For ease of description, the term “coils” will be used hereinafter to indicate “one or more coils” and “sensor” to indicate “one or more sensors,” unless specified otherwise.
By virtue of the above described arrangement, coil wrap 20 provides a reproducibly correct level of stimulation during an initial therapy session and during successive therapy sessions, because the presence or absence of nerve conduction is detected and, in some variations, measured when coil wrap 20 is first fitted and fine-tuned on the patient. In addition to properly modulating the applied magnetic field, the positioning of coils 26 over ankle 22 may also be tailored according to the input provided by sensor 30, so to fine-tune the direction of the magnetic field. Such an adjustment of the direction, amplitude, and level of the stimulation provided to the target nerve through the above described automated feedback loop, to ensure that peripheral nerve conduction is being achieved, is one of the key features.
If the magnetic pulse does not substantially interfere with sensor 30, sensor 30 may be placed directly within the field of stimulation, so that power supplied to the system may be conserved. This is particularly important for battery-powered systems. Alternatively, sensor 30 may also be placed at a distance from the magnetic field and still properly detect neural stimulation.
In a method of use of coil wrap 20, the amplitude and/or firing sequence of coils 26 may be ramped up progressively, so that the magnetic field is increased in strength and/or breadth until nerve conduction is detected, after which the applied stimulus is adjusted or maintained at its current level for the remainder of the therapy. The level of stimulation may be also controlled through a combination of feedback from sensor 30 and feedback based on perceptions of the patient. For example, the patient may activate a switch once she perceives an excessive stimulation, in particular, an excessive level of muscular stimulation. In one instance, the patient may be asked to push a button or turn a knob when she feels her toe twitching or when she experiences paresthesia over the sole of her foot. The patient will then continue pressing the button or keep the knob in the rotated position until she can no longer feel her toe twitching or paresthesia in her foot, indicating that that level of applied stimulation corresponds to an optimal therapy level. From that point on, the patient may be instructed to simply retain her foot, knee, or other limb within coil wrap 20 until therapy has been terminated while the system is kept at the optimal level. Adding patient input enables control of coil wrap 20 during outpatient treatments, because the patient is now able to adjust the intensity of the magnetic field herself beyond the signals provided to logic controller 28 by sensor 30.
Detecting and, if the case, measuring conduction in one or more nerves along the conduction pathways of the stimulated nerve confirms that the target nerve has been stimulated, providing an accurate assessment of the efficiency of the applied therapy on the patient. A concomitant detection of muscle contraction may also confirm that the target nerve is being stimulated and provide an indication to the patient or to a healthcare provider as to whether stimulation has been applied at an excessive level in view of the anatomical and physiological characteristics of the patient.
Based on the foregoing, coil wrap 20 allows for a consistent, user-friendly targeting and modulation of the peripheral nerves via the posterior tibial nerve on an outpatient basis, in particular, the targeting and modulation of the pudendal nerve and of the sacral plexus. When multiple coils 26 are present, coils 26 may be activated simultaneously or differentially to generate the desired magnetic field. The direction and location of each of coils 26 may be reversibly or irreversibly adjusted by the healthcare provider or by the patient, customizing the location of the applied stimulation to the anatomy and therapy needs of each patient. After a healthcare provider has optimized position and firing sequence for each of coils 26, the patient may be sent home with coil wrap 20 adjusted to consistently target the desired nerve. In one variant of the present variation, an automatic feedback system adjusts one or more of firing sequence, firing strength or position of coils 26 within coil wrap 20 during the initial setup and also during successive therapy sessions.
In summary, certain variations include the creation of a loop consisting of feeding information on nerve conduction to logic controller 28 and on logic controller 28 tailoring the electrical current sent to coil wrap 20 according to the information received from sensor 26 based on whether or not the nerve is receiving the desired stimulation and, in some variations, the desired amount of stimulation. This arrangement offers an unparalleled level of therapy control and flexibility within a home care setting, because a consistent, repeatable stimulation of the target nerve can be attained. Aside from adjusting the position of coils 26 in accordance with the patient's anatomy and physiological variations, controlling pulse amplitude is also of great importance even during different therapy sessions with the same patient. For example, a patient with leg edema will encounter difficulties in properly adjusting coil wrap 20 based on whether her legs and ankles are swollen or not swollen, and the power required to penetrate to posterior tibial nerve 24 (in the case of a VI therapy) will vary greatly due to the variable depth of the nerve. Thus, having feedback provided by sensor 26 becomes a necessity for achieving an accurate dosage of the treatment rather than an option. Benchtop testing has demonstrated that a system constructed as described herein is capable of non-invasively generating electrical currents similar to those found in therapeutic electrostimulation and to do so in different settings.
Referring now to
Having coil windings arranged along a common longitudinal axis, as required in a Helmholtz coil configuration, generates a more focused magnetic field and a more accurate targeting of tibial nerve 38 or of any other nerve. Like in the previous variation, the operation of coils 40 is controlled by a logic controller 42, which is in turn connected to sensor 44 that monitors conduction in tibial nerve 44 and that generates a feedback to logic controller 42 about the efficiency of the therapy in progress. Therefore, like in the previous variation, the coupling of sensor 44 with logic controller 42 optimizes operation of coil wrap 34 according to results measured at the level of tibial nerve 38. Also like in the previous variation, manual adjustments to the parameters of electric current provided by logic controller 42 to Helmholtz coil 40 may also be made manually by the patient or by a healthcare provider, and coil wrap 34 may be structured so that the position of Helmholtz coil 40 within coil wrap 34 is adjusted as desired either manually by the patient or by a healthcare provider, or automatically by logic controller 42.
Referring now to
Referring now to
Cradle 48 may be made from a variety of materials and may be monolithic, or have a hollow or semi-hollow structure to enable the movement of coils 52 within cradle 48, as described in greater detail below. Preferably, cradle 48 has an ergonomically design allowing the ankle and heel of the patient to be retained within cradle 48, in a position that matches the positions of stimulating coils 52 to the area of stimulation. The design of cradle 48 provides for a particularly comfortable delivery of therapy to patients that prefer to remain seated during their therapy, and enables the patient to perform the required therapy within a health care facility, or to take cradle 48 home, typically after an initial session and appropriate training in a health care facility. In that event, the patient will be trained to apply sensor 54 autonomously and to adjust stimulation to a comfortable level.
In one variant of this variation, sensor 54 may include a conductive electrode patch that provides a feedback to logic controller 56 for adjusting, if necessary, the stimulation parameters of coils 52. Alternatively, sensor 54 may be a sensor patch that is either applied to the skin of the patient or is incorporated within the structure of cradle 48.
Referring now to
A method of use of the foot cradle depicted in
During a second step illustrated in
During a third step illustrated in
During a fourth, optional step illustrated in
While the present method has been described with regard to foot cradle 70, the same method steps may be envisioned for coil wraps or cradles of different configurations, for example, for the coil wraps and cradles described with reference to the previous FIGS.
In an alternative variation, the logic controller (not shown) may automatically adjust coil positioning to optimize therapy during the initial and successive sessions. While this set-up may be more difficult to implement, it also provides for an accurate targeting of the target nerve during each therapy session, regardless of alterations in patient positioning or changes to the anatomy of the patient (for example, when a foot is swollen). In this variation, the device simply varies the orientation of coils 84 until stimulation has been sensed.
Further, coils 84 may be translated along a single direction (for example, horizontally) or along a plurality of directions, to provide for the most accurate positioning of coils 84 with respect to the target nerve.
A second method of use of the foot cradle depicted in
The position of coils 84 may be translated in different directions (in the illustrated variation, may be translated horizontally) and may be locked in an initial position once conduction in nerve 89 is detected by a sensor (for example, sensing patch 86).
More particularly,
With reference to
Referring now to
One or more of the illustrated sensors 94 may be employed over body regions being stimulated (for example, back, leg, arm, neck, head, torso, etc.) and may be either incorporated within an actual cradle or wrap or, otherwise, be applied separately from the cradle or the wrap.
Sensors 94 may be structured as disposable, single-use, EKG-type patches that are attached to the body outside of cradle 98 along the nerve conduction pathway and are then connected to the logic controller (not shown) before beginning therapy. This arrangement provides for an intimate body contact of sensors 94 without the risk of infection or other detrimental side effects that may be present with transcutaneous devices. Sensors 94 may be employed both for beginning and for monitoring the stimulation therapy; more specifically, sensors 94 may be employed during the beginning of the therapy to optimize the strength of the magnetic field and/or to adjust the positioning of coils 96 within the cradle 98. Once therapy has begun, sensors 94 continue to monitor nerve conduction to ensure that the correct level of stimulation is being provided. In the event that for some reason nerve conduction decays during therapy, the logic controller can automatically adjust the magnetic field, ensuring that the appropriate therapy is delivered for the appropriate amount of time.
One or more of sensors 94 in this variation, or any of the variations described herein, may take the form of an inductive coil designed to receive impulses from the underlying nerves, so that inductive technologies may be used to both stimulate the nerve or tissues as well as to record the effect of the stimulation on nerves or tissues. Any of sensors 94 may be connected to the logic controller through one or more connection modes, including, but not limited to, wireless signals, wired signals, radio frequencies, Bluetooth® (Bluetooth Sig, Inc., Kirkland, Wash.), infrared, ultrasound, direct switching of the current circuit, etc., so long as communication between the sensor and the device is effective.
During implementation of the present method, a healthcare provider may simply elect to use sensors 94 to adjust the device, for example, to lock coils 96 into position, during the first therapy session and not require the use of sensors 94 during each successive therapy session.
Referring now to
More particularly,
Other variations that are not illustrated include, bur are not limited to: a head worn garment, such as a cap; a neck worn garment, such as a neck brace; and a lower-back garment. Each garment and applicator may also utilize the locking, targeting coil feature described previously, without requiring the use of the any sensing components after a proper positioning of the coils in relation to the target nerve or nerves has been established.
Still other variations are depicted in
As shown, a device constructed according to the principles described herein provides a targeted and precise stimulation of the posterior tibial nerve, or of other peripheral nerves, in a non-invasive manner by employing an ergonomic wrap or cradle that specifically targets the posterior tibial nerve in a consistent and repeatable manner. For example, in patients with OAB or VI, the novel, reversibly lockable movement of the coils and the use of a logic controller-sensor loop enables the application of a magnetic field that can be varied in location, amplitude and strength according to the amount of stimulation actually induced in one or more target nerves and of the response of the patient to the therapy. An apparatus according to the variations described herein may deliver any frequency of stimulation, including low frequencies, high frequencies, mid frequencies and ultrahigh frequencies, and overlapping and non-overlapping coils may be used to generate the desired field, although overlapping or Helmholtz coils are preferred due to their ability to target a broader region and achieve more thorough stimulation.
Ailments that may be treated through the use of apparatus and methods as described herein include not only OAB and VI, but also obesity, depression, urinary incontinence, fecal incontinence, hypertension, pain, back pain, restless leg syndrome, Guillain Barre syndrome, quadriplegia, paraplegia, diabetic polyneuropthy, dyskinesias, paresthesias, dental procedure pain, knee osteoarthritis, anesthesia (pain relief during surgery), Alzheimer's disease, angina (chest pain from heart disease), ankylosing spondylitis, back pain, bum pain, cancer pain, chronic pain, dysmenorrhea (painful menstruation), headache, hemiplegia, hemiparesis (paralysis on one side of the body), labor pain, local anesthesia during gallstone lithotripsy, facial pain, trigeminal neuralgia, bruxism (tooth grinding) pain, myofascial pain, pregnancy-related nausea or vomiting, neck and shoulder pain, pain from broken bones, rib fracture or acute trauma, diabetic peripheral neuropathy, phantom limb pain, post-herpetic neuralgia (pain after shingles), postoperative ileus (bowel obstruction), irritable bowel syndrome, postoperative nausea or vomiting, postoperative pain, post-stroke rehabilitation, rheumatoid arthritis, skin ulcers, spinal cord injury, temporomandibular joint pain, detrusor instability, spinal muscular atrophy (in children), pain during hysteroscopy, gastroparesis, chronic obstructive pulmonary disease rehabilitation, carpal tunnel syndrome, soft tissue injury, multiple sclerosis, intermittent claudication, attention-deficit hyperactivity disorder (ADHD), cognitive impairment, knee replacement pain, achalasia, atopic eczema, bursitis, carpal tunnel syndrome, dementia, depression, dry mouth, dystonia, enhanced blood flow in the brain, enhanced blood perfusion of the uterus and placenta, esophageal spasm, fibromyalgia, fracture pain, Guillain-Barre syndrome, hemophilia, herpes, hip pain, interstitial cystitis, irritable bowel syndrome, pruritis, joint pain, labor induction, local anesthesia, menstrual cramps, muscle cramps, muscle spasticity, muscle strain or pain, musculoskeletal trauma, myofascial pain dysfunction syndrome, nerve damage, osteoarthritis, pain medication adjunct, pancreatitis, Raynaud's phenomenon, repetitive strain injuries, sacral pain, schizophrenia, shingles, shoulder subluxation, sickle cell anemia pain, Skin flap ischemia (during plastic surgery), sphincter of Oddi disorders, sports injuries, thrombophlebitis, tinnitus (ringing in the ear), restless legs, tremor, whiplash and neuralgias. In contrast to implantable nerve stimulators, this therapy is completely non-invasive and does not require a major surgery to implant a permanent nerve stimulation device. Moreover, this therapy can be controlled to optimize the level of therapy delivered according to power consumption and nerve stimulation requirements and need not be delivered by a professional healthcare provider.
In other variations, neural stimulation may be applied as electrical transcutaneous stimulation, for example, by inserting an invasive electrical needle into a target body part and by modulating stimulation is modulated on the basis of information sent back to the logic controller from the one or more sensors that are used to detect and/or maintain the correct level of stimulation. The transcutaneous electrical stimulation sensor may be placed in the body independently or be incorporated within the wrap and may provide, among other things, feedback as to the quality of the electrical connection to the skin, which is directly related to the burn risk inherently associated with this type of therapy. In fact, these methods of stimulation may not be optimal due to the resulting skin irritation and risk of potential burns, a very serious issue in the large percentage of patients that have neuropathies. Even when patches are applied to monitor transcutaneous stimulation very closely, the patches may still become displaced and allow a burn to occur. Moreover, potentially interfering electrical impulses may develop at the treatment site, creating a noisy environment for the detection of nerve conduction.
In still other variations, an external coil or coils may be inductively connected to an implanted coil or coils may be utilized. In these variations, an ergonomic applicator may be adjusted by the user or by a healthcare provider such to optimize inductive power transmission between the external and implanted coils. One or more sensors may be utilized to provide a feedback that the relative coil positions have been optimized, and the external coil may then be reversibly locked into position within the ergonomic applicator. Two applications of this variation relate to the transfer of power to recharge an implantable device, and to the transfer of power to activate an implantable device.
In the first application, when an implantable rechargeable device is utilized, the external coils may be used for recharging the implanted device by means of inductive fields generated by the external coils. The external coils may include circuitry that determines the amount of resistance encountered by the magnetic field or other electrical properties related to the quality and degree of the magnetic coupling that is being established. Based on this feedback, the position of the external coils may be adjusted manually or automatically to optimize the coupling achieved with during each recharging session. Alternatively, a sensor may be incorporated into the implantable device and may communicate the degree and quality of the magnetic coupling to the external coils and/or the connected circuitry via wireless communication, providing a feedback for the automatic or manual adjustment of the external recharging coils.
The coils within the ergonomic applicator may be reversibly locked into place for the duration of the recharge session, and the implantable device may also communicate to the external recharging unit that the implantable device has been fully recharged, terminating the recharging session has been completed. By providing for an intermittent recharging of an implanted device, an apparatus according as described herein enables the implantable device to devote more power to performing its intended function optimally and with a lesser concern about protecting or extending battery life.
In the second application, the powering coils may contain circuitry to determine the amount of resistance encountered by the applied magnetic field, or other electrical properties that may reflect the quality and degree of the magnetic coupling that is being achieved. Based on this feedback, the powering coils in the applicator may be adjusted manually or automatically to activate and optimize the coil coupling at the beginning of each therapy session. Alternatively, a sensor may be incorporated into the implantable device and communicate the degree and quality of the magnetic coupling externally via wireless communication, which may in turn provide feedback for the automatic or manual adjustment of the powering coil. In one variant of the present variation, the inductive coils may be magnetically coupled to a needle targeting the posterior tibial nerve.
An exemplary method of use of an apparatus as described herein on a patient suffering from VI and/or OAB includes the following steps:
The patient places a conductive wrap contained within a flexible material over a region of the ankle (or alternatively over the knee) to provide the required pulsed magnetic field. Alternatively, the patient may use an ergonomic foot/leg rest or cradle having embedded coils.
A sensor (for example, a sensor patch) is placed on the patient's body along the path of the nerve, ideally proximal to the stimulation site to ensure afferent nerve stimulation, and is connected to a logic controller.
A physician or healthcare provider adjusts the coils in the wrap or cradle until nerve conduction is achieved based on patient and sensor feedback. An optimal position is sought, and the coils may be reversibly locked into position within the conductive wrap or ergonomic cradle and remain in this position during subsequent use.
During the therapy session, the logic controller provides an electric current to the coils, generating an inductive magnetic field. In one variation, this field begins at low amplitude and slowly ramps up until nerve conduction exceeds a threshold level, as signaled by the sensor and possibly by the patient, who may feel motor conduction. Alternatively, one or more coils may also be activated to increase the covered area of stimulation in the event that stimulation does not occur with the initial coil configuration or is inadequate
The optimal stimulation may be determined in a variety of manners, for example, by measuring exposure to electromagnetic fields capable of generating a square wave electric signal at a frequency of 10-30 Hz at the targeted tissue depth. The square wave configuration of the signal may be generated via Fourier transformation or may be a ramped current generated in any manner.
The inductive magnetic pulses continue for an appropriate duration of use, for example, for 15-30 minutes. The sensor may remain in place during the entire therapy session to ensure that stimulation occurs consistently and to provide for appropriate corrections if nerve conduction deteriorated. The logic controller may be powered either by a portable power source such as a battery, or by or a fixed power source such as a traditional wall outlet.
The conductive wrap and/or ergonomic cradle is removed from the body when therapeutic stimulation is not being delivered, typically at the end of the therapy session.
The conductive wrap and/or ergonomic cradle is reapplied along with the sensor patch (ideally disposable) from time to time as indicated, for example, on a daily basis, and steps 4-8 are repeated.
The devices and methods described herein may be applied to any body tissues, including nerve, muscle, skin, vasculature, or any other organ or tissue within the human body. Further, the devices and methods described herein may be used to treat any conditions suited for neuromodulation regardless of whether the stimulation source is an electromagnetic field, a direct electric current, a RF field, infrared energy, visible light, ultraviolet light, ultrasound, or other energy dispensing device.
In other variations, as shown in
Referring to
Additionally, microneedles may be attached, deposited, or arrayed on an electrode surface or patch in a variety of configurations and arrangements, depending on where the particular microneedle patch will be utilized and the treatment to be delivered. The number of microneedles included in an array can vary. For example, the number of microneedles may range from about 5 to 500 or 100 to 400 or about 200 to 300 or about 256. In certain variations where microneedles are composed of strong, highly conductive material, the number of microneedles necessary may be less and may range from about 5 to 100 or 10 to 50 or 5 to 50. However, where microneedles are composed of higher resistance metal, a greater number of needles may be needed, e.g., about 100 to 500 or about 200 to 300 or greater than 500.
Referring to
In other variations, microneedles may have lengths that range from about 480 to 1450 microns, widths from about 160 to 470 microns, thicknesses from about 30 to 100 microns and tip angles from about 20 to 90 degrees, and arrays can contain from 5 to 50 microneedles. For example, microneedles having these dimensions have been shown to be less painful than hypodermic needles. Length and number of microneedles can affect the level of pain experienced. Decreasing microneedle length and/or the number of microneedles may be beneficial and act to further reduce pain and provide comfort.
In certain variations, the one or more microneedles may include an electrically conductive material such that the microneedles may transmit an electrical signal to an overlying electrode or other surface. Microneedles may be constructed of an electrically conductive material and/or coated with an electrically conductive material. Optionally, microneedles may be coated with an electrically conductive material and constructed of a non-conductive material. Microneedles may be fabricated using a variety of materials, e.g., metals, stainless steel, solid or coat of gold over NI, Pd or Pd—Co, Pt, silicon, silicon dioxide, polymers, glass, biocompatible polymers, titanium, silver, or suture materials. Biodegradable polymers may also be used such that if a tip of a microneedle were to snap or break off during insertion, it would easily biodegrade.
A microneedle array 234 may be constructed or fabricated using any variety of manufacturing methods known to persons of ordinary skill in the art. Microneedles may be arrayed, attached, etched or deposited onto a surface of an electrode. In another variation, microneedles may be etched from or deposited onto a silicon electrode, such that the microneedle patch, including electrode and microneedles, are made from one material creating a durable and stable microneedle patch.
As shown in
In certain variations, microneedle arrays are fabricated by patterning SU-8 onto glass substrates and defining needle shapes by lithography. The tips of the needles can be sharpened using reactive ion etching. Optionally, holes may be drilled, e.g., by laser, through the microneedles and base substrate. Holes may be drilled off-center, but parallel to the microneedle axis, terminating in side-opening holes along the needle shaft below the needle tip. If desired, the holes can serve as micro fluidic needle bores for injection or infusion of drugs, medicines, insulin, proteins, nanoparticles that would encapsulate a drug or demonstrate the ability to deliver a virus for vaccinations, etc. to be used separately or in combination with electrical or magnetic therapy. The microneedles may also be coated with nickel by electroplating, which can increase their mechanical strength.
In certain variations, microneedle patches or microneedle electrode arrays are made by fabricating master structures from which replicates are molded and then made electrically active. For example, SU-8 may be spun on a glass substrate bearing an array mask pattern, baked, and then exposed from the backside to from a tapered needle structure. Microneedles may be sharpened by RIE etching. A PDMS (polydimethylsiloxane) or similar material mold can then be copied from the master. A PMMA (polymethylmethacrylate) microneedle array is formed by solvent-casting and then released from the mold.
To provide the arrays with electrical functionality, a Ti/Cu seed layer may be deposited on the PMMA array and patterned by excimer laser to electrically isolate adjacent rows. A Ni layer (e.g., about 15 to 30 microns or 20 to 25 microns thick) may be electroplated on the patterned seed layer to enhance structural rigidity. A backside electrical connection to the array may be formed by backside etching of a hole and forming an electrical connection through the hole.
In another variation, the microneedle array is arranged in a 16.times.16 array (i.e., 256 needles). Each needle has a height of about 400 microns and the base diameter is about 100 microns. The pitch between microneedles can be about 250 microns. The microneedle arrays are then coated with metal and laser-etched to provide electrical functionality. Optionally, rows of microneedles can be electrically isolated from each other so that alternating rows can provide alternating electrical polarity. The arrays are also interfaced with a power source. Microneedles may be made of polymer, coated with a metal, and etched to act as alternating electrodes. In certain variations, the firing sequence of the microneedles by rows or groups may be varied or configured to alternate.
In certain variations, a microneedle array may include one or more microneedles having multiple channels. For example, a multichannel silicon microneedle may be constructed to deliver bioactive compounds into neural or other tissue while simultaneously monitoring and stimulating neurons and nerves.
In certain variations, microneedles are formed such that they are in direct contact with their corresponding or overlying electrodes. For example, a microneedle patch may include an adhesive electrode pad and may utilize a conductive gel to help hold the microneedles in place to prevent shear forces from breaking or bending the microneedles.
In certain variation, as shown in
Referring again to
Referring again to
Once the patient's foot is in position and the microneedle patch 228 (e.g., conductive microneedle patch) is in place, a current is passed from controller 218 through coils 212, and as a result, the coils 212 generate a magnetic field which is focused on the posterior tibial nerve 220. The magnetic field stimulates tibial nerve 220, generating a current that will flow along the tibial nerve 220 and spread along its length, to its sacral or pudendal nerve roots. Microneedle patch 228 detects corresponding muscle stimulation or twitching or electrical conduction through the stimulated posterior tibial nerve. Upon detection, the microneedle array may conduct and transmit an electrical signal to the overlying electrode of microneedle patch 228. The signal may be transmitted to controller 218, which can be integral or a separate controller or device, or a separate controller coupled to controller 218. The controller can then be varied or adjusted (to adjust the current or magnetic field) based on the signal received from microneedle patch 228 to ensure that adequate conduction of the posterior tibial nerve 220 occurs and an adequate and accurate dosage of treatment is being received. Although shown utilizing a sensor, it is also contemplated that the system could be used without a sensor.
Referring to
A sensor utilized in the energy emitting system 210 may be a microneedle patch 228 as described above or optionally the sensor may a sensor type known in the art (e.g., EKG sensor) or as described in any of the variations herein. It is also contemplated that energy emitting system 250 can be utilized without a sensor. Optionally, the sensor may be positioned within or along the housing, e.g., the foot cradle, along with the one or more conductive coils, or positioned at a site distant from the housing or conductive coils.
In certain variations, energy emitting system 210 my optionally include one or more conductive microneedle patches which can be positioned in proximity to the target nerve or muscle and provide an additional or supplemental electrical or magnetic stimulus to the target nerve or muscle.
Referring to
In energy emitting system 260, current density and subsequent electric field intensity generated between conductive coils 212 and percutaneous electrode needle 262 is greater than that generated by traditional percutaneous stimulators. A greater electric field intensity makes site location for conductive coils 212 and percutaneous electrode needle 262 easier. Furthermore, the load impedance through the surface of the skin is much higher than the internal impedance, and as such, the relatively high load impedance lessens the likelihood of damage to tissue and nerves due to high current pulses.
Referring again to
In another variation, as shown in
In certain variations of energy emitting system 260 as described above and shown in
In certain variations, as shown in
Referring to
Once conductive microneedle patch 252 and sensor 221 are in position, a current is passed from controller 218 through conductive microneedle patch 252, resulting in an electrical stimulus of the posterior tibial nerve 220. Alternatively, the microneedle array may be insulated or constructed of non conductive material such that the microneedle patch 252 generates a magnetic field that stimulates tibial nerve 220 in a manner similar to the one or more coils described in the variations above, without an electrical stimulus. Whether the stimulus is electrical or magnetic, either stimulus will generate a current that will flow along the tibial nerve 220 and spread along its length, to its sacral or pudendal nerve roots. Sensor 221 detects electrical conduction through the stimulated posterior tibial nerve 220, and then transmits the signal to controller 218. In certain variations, the sensor may be in the form of a microneedle patch sensor. The signal may be transmitted to controller 218, a separate controller or device, or a separate controller coupled to controller 218. The controller can then be varied or adjusted based on the signal from sensor 221 to ensure that adequate conduction of the posterior tibial nerve 220 occurs and an adequate and accurate dosage of treatment is being received.
The sensor utilized in the energy emitting system 250 may be a sensor of the type described above, with respect to other variations. Optionally, for example, the sensor may be a microneedle patch. It is also contemplated that energy emitting system 250 can be utilized without a sensor. The sensor may be placed over a portion of the subject's body suitable for detecting conduction of the target nerve (e.g., on the leg as shown) or over a muscle to detect muscle stimulation resulting from stimulating the target nerve.
In certain variations, energy emitting system 250 my optionally include one or more conductive coils disposed within or along a housing which can be positioned in proximity to the target nerve or muscle and provide an additional or supplemental stimulation of the target nerve or muscle.
Referring to
Optionally, the electrode needle may be non-percutaneous, such that it is insertable in an orifice or opening in the subject, such as a natural orifice. The percutaneous electrode needle 262 is inductively coupled to conductive microneedle patch 252. In use, a microneedle patch 252 is placed on a first portion of a patient's body, for example a foot, ankle, or leg, in proximity to posterior tibial nerve 220 within the first portion of the patient's body and down-stream or distal to a selected stimulation site 261. The percutaneous electrode needle 262 is inserted through the skin at a location and to a depth that brings the tip in close proximity to the target nerve to be stimulated.
The controller 218 is activated and a current passes through microneedle patch 252 and traverses the internal stimulation site 261 by passing from microneedle patch 252 to the internal percutaneous electrode needle 262, as indicated by arrow i. The current passing through microneedle patch 252 may also generate a magnetic field which can generate a current that traverses the internal stimulation site 261 by passing from microneedle patch 252 to the internal percutaneous electrode needle 262. Also, the percutaneous electrode needle may be positioned within the generated magnetic field, whereby the magnetic field generates a current in the percutaneous electrode which stimulates a target nerve and traverses an internal stimulation site. Optionally, a current may be passed from the controller 218 through microneedle patch 252 and/or from the controller 218 through percutaneous electrode needle 262, traversing the internal stimulation site as the current passes between the patch and needle.
Referring to
In any of the above systems, variations are contemplated where the sensors are also coupled or connected to or otherwise in communication with energy emitting devices, e.g., the conductive coils or conductive microneedle patches.
In certain variations, the one or more microneedles of the microneedle patch may include an electrically conductive material such that the microneedles may transmit an electrical signal to an overlying electrode or other surface. Microneedles may be constructed of an electrically conductive material and/or coated with an electrically conductive material. Optionally, microneedles may be coated with an electrically conductive material and constructed of a non-conductive material. Microneedles may be fabricated using a variety of materials, e.g., metals, stainless steel, solid or coat of gold over NI, Pd or Pd—Co, Pt, silicon, silicon dioxide, polymers, glass, biocompatible polymers, titanium, silver, or suture materials. Biodegradable polymers may also be used such that if a tip of a microneedle were to snap or break off during insertion, it would easily biodegrade. Optionally, the microneedle patch may be non-conductive.
In certain variations, an electrode patch for improved conductance or conduction is provided. The patch can include at least one electrode having a first surface and/or a second surface. The electrode may optionally be attached to various other materials or adhesive materials. An array of microneedles may be deposited on a surface of the electrode, or attached to a patch or other material and indirectly or directly connected to the electrode. The array of microneedles may include a conductive material. Such patches may be used as a sensor to detect muscle stimulation or electrical conduction, or to provide or deliver an electrical stimulus or magnetic field, e.g., to a target nerve, and may optionally be used in any of the variations described herein or in any application where improved conductance or conduction is desired. Microneedles yield improved reduction in impedance compared to simple abrasion and other techniques, and are less painful and more comfortable for the patient.
In certain variations, typical voltage sensed at the skin and detectable or conductable by a microneedle patch or microneedle array may range from about 1 to 400 microvolts or about 10 to 300 microvolts.
In certain variations, methods of treating a subject with urinary incontinence or various pelvic floor disorders utilizing the energy emitting systems described herein are contemplated. Symptoms associated with urinary incontinence may be observed, detected, or diagnosed. An energy emitting device having one or more energy generators, e.g., one or more conductive coils or one or more microneedle patches, may be positioned in proximity to a target nerve, e.g., the tibial or posterior tibial nerve or popliteal or sacral nerve or branches thereof of a subject or patient along a first portion of a subject's or patient's body. The subject may or may not be exhibiting symptoms associated with urinary incontinence. In the case of the conductive coils, the coils may be positioned within or along a housing, such as a foot or knee cradle, and a foot or leg may be positioned therein. In the case of a microneedle patch, the patch may be attached to a subject's skin. Optionally, the method involves positioning a first portion of a subject's body, the subject exhibiting symptoms associated with urinary incontinence, relative to an energy emitting device such that a target nerve within the first portion of the body is in proximity to at least one energy generator disposed within or along the energy emitting device.
A current is then passed through the energy generator to produce, generate or deliver energy, e.g., a magnetic or electromagnetic field or electrical or magnetic energy or stimulus, focused on the tibial or posterior tibial nerve or branches thereof. This in turn may cause the stimulation of a pudendal nerve, sacral plexus, or other nerves in the pelvic floor. Various nerves innervating the various muscles, sphincters, nerves, organs and conduits of the urinary tract and bladder may be stimulated directly or indirectly. In certain variations, a current is passed through one more coils, which generate a magnetic or electromagnetic field which stimulates the posterior tibial nerve. In certain variations, the positioning of the coils relative to the first portion of the subject's body may be adjusted to re-focus the magnetic field on the posterior tibial nerve as needed. In certain variations, a current is passed through a microneedle patch generating or delivering an electrical or magnetic stimulus or field. The positioning of the microneedle patch relative to the first portion of the subject's body may be adjusted to re-focus the electrical or magnetic stimulus or field on the posterior tibial nerve as needed.
Optionally, electrical conduction through the target nerve, e.g., the posterior tibial nerve, or muscle stimulation can be detected via at least one sensor. A conductive sensor may be positioned in proximity to the posterior tibial nerve along a second portion of the subject's body. Optionally, a sensor may be positioned over a corresponding muscle to detect muscle stimulation or twitching resulting from nerve stimulation. Optionally, the electrical conduction is detected along a second portion of the subject's body which is different from the first portion of the body. Optionally, the sensor in the form of a microneedle patch. In certain variations, the sensor may be positioned behind a subject's knee to detect the electrical conduction along the afferent posterior tibial nerve or on another portion of a patient's leg or foot. In other variations, the sensor may be positioned within or along a housing along with the one or more conductive coils.
Where a sensor is used, a signal is received from the sensors and the signal is indicative of the electrical conduction of the target nerve, e.g., posterior tibial nerve. The current may be adjusted or varied using a controller which is in communication with the energy generator. Adjustments may be made in response to the nerve or muscle stimulation detected by the conductive sensor, in order to optimize or ensure adequate treatment of urinary incontinence by achieving the appropriate level of conductance and appropriate level of nerve or muscle stimulation. Appropriate levels or parameters for current, frequency, magnetic field, treatment duration, etc., are those that result in an observed or detected reduction or prevention of symptoms associated with urinary incontinence. Treatment could also be administered and the appropriate levels and parameters achieved through observing or detecting reduction or prevention of symptoms where a sensor is not used. Examples of these symptoms include but are not limited to the inability to control urinary function, urinary leakage, and loss of bladder control.
In certain variations, the amplitude, frequency, direction of a generated magnetic field, electrical or magnetic stimulus, or firing sequence of the coils or microneedles making up the microneedle array may be adjusted. Optionally, the current may be varied according to a muscular response in the patient. Thus, to treat urinary incontinence, the magnetic field or electrical stimulus is applied to a subject or patient until the desired effects (e.g., reduction of symptoms) are achieved.
In certain variations, methods of treating a subject with fecal incontinence utilizing the energy emitting systems described herein are contemplated. Symptoms associated with fecal incontinence may be observed, detected, or diagnosed. An energy emitting device having one or more energy generators, e.g., one or more conductive coils or one or more microneedle patches, may be positioned in proximity to a target nerve, e.g., the tibial or posterior tibial nerve, or popliteal or sacral nerve or branches thereof, of a subject along a first portion of a subject's body. The subject may or may not be exhibiting symptoms associated with fecal incontinence. In the case of the conductive coils, the coils may be positioned within or along a housing, such as a foot or knee cradle, and a foot or leg may be positioned therein. In the case of a microneedle patch, the patch may be attached to a subject's skin. Optionally, the method involves positioning a first portion of a subject's body, the subject exhibiting symptoms associated with fecal incontinence, relative to an energy emitting device such that a target nerve within the first portion of the body is in proximity to at least one energy generator disposed within or along the energy emitting device.
A current is then passed through the energy generator to produce, generate or deliver energy, e.g., a magnetic or electromagnetic field or electrical or magnetic energy or stimulus, focused on the tibial or posterior tibial nerve or branches thereof. This in turn causes the stimulation of a pudendal nerve, sacral plexus, or nerves in the pelvic floor. Various nerves innervating the various muscles, sphincters, rectum, nerves, organs and conduits associated with bowel movements, fecal control, and the intestines may be stimulated directly or indirectly. Optionally, a current is passed through one more coils, which generate a magnetic or electromagnetic field which stimulates the posterior tibial nerve. In certain variations, the positioning of the coils relative to the first portion of the subject's body may be adjusted to re-focus the magnetic field on the posterior tibial nerve as needed. In certain variations, a current is passed through a microneedle patch generating or delivering an electrical or magnetic stimulus or field. The positioning of the microneedle patch relative to the first portion of the subject's body may be adjusted to re-focus the electrical or magnetic stimulus or field on the posterior tibial nerve as needed.
Optionally, electrical conduction through the target nerve, e.g., the posterior tibial nerve, or muscle stimulation can be detected via at least one sensor. A conductive sensor may be positioned in proximity to the posterior tibial nerve along a second portion of the subject's body. Optionally, a sensor may be positioned over a corresponding muscle to detect muscle stimulation or twitching resulting from nerve stimulation. Optionally, the electrical conduction is detected along a second portion of the subject's body which is different from the first portion of the body. Optionally, the sensor is in the form a of a microneedle patch. In certain variations, the sensor may be positioned behind a subject's knee to detect the electrical conduction along the afferent posterior tibial nerve or on another portion of a patient's leg or foot. In other variations, the sensor may be positioned within or along a housing along with the one or more conductive coils.
Where a sensor is used, a signal is received from the sensors and the signal is indicative of the electrical conduction of the target nerve, e.g., posterior tibial nerve. The current may be adjusted or varied using a controller which is in communication with the energy generator. Adjustments may be made in response to the nerve or muscle stimulation detected by the conductive sensor, in order to optimize or ensure adequate treatment of fecal incontinence by achieving the appropriate level of conductance and appropriate level of nerve or muscle stimulation. Appropriate levels or parameters for current, frequency, magnetic field, treatment duration, etc., are those that result in an observed or detected reduction or prevention of symptoms associated with fecal incontinence. Treatment could also be administered and the appropriate levels and parameters achieved through observing or detecting reduction or prevention of symptoms where a sensor is not used. Examples of these symptoms include but are not limited: the loss of voluntary control to retain stool in the rectum; loss of fecal control; inability to control bowel movements, and fecal leaking:
In certain variations, the amplitude, frequency, direction of a generated magnetic field, electrical or magnetic stimulus, or firing sequence of the coils or microneedles making up the microneedle array may be adjusted. Optionally, the current may be varied according to a muscular response in the patient. Thus, to treat fecal incontinence, the magnetic field or electrical stimulus is applied to a subject or patient until the desired effects (e.g., reduction of symptoms) are achieved.
In certain variations, methods of treating a subject with restless leg syndrome utilizing the energy emitting systems described herein are contemplated. Victims afflicted with Restless Leg Syndrome (RLS or Ekbom's syndrome), are unable to remain seated or to stand still. Activities that require maintaining motor rest and limited cognitive stimulation, such as transportation, e.g., in a car, plane, train, etc., or attending longer meetings, lectures, movies or other performances, become difficult if not impossible. These sensations become more severe at night and RLS patients find sleep to be virtually impossible, adding to the diminishing quality of their lives. The urge to move, which increases over periods of rest, can be completely dissipated by movement, such as walking. However, once movement ceases, symptoms return with increased intensity. If an RLS patient is forced to lie still, symptoms will continue to build like a loaded spring and, eventually, the legs will involuntary move, relieving symptoms immediately.
Thus, symptoms associated with restless leg syndrome may be observed, detected, or diagnosed. An energy emitting device having one or more energy generators, e.g., one or more conductive coils or one or more microneedle patches, may be positioned in proximity to a target nerve, e.g., the tibial or posterior tibial nerve, or popliteal or sacral nerve or branches thereof or other nerves associated with restless leg syndrome, of a subject along a first portion of a subject's body. The subject may or may not be exhibiting symptoms associated with restless leg syndrome. In the case of the conductive coils, the coils may be positioned within or along a housing, such as a foot or knee cradle, and a foot or leg may be positioned therein. In the case of a microneedle patch, the patch may be attached to a subject's skin. Optionally, the method involves positioning a first portion of a subject's body, the subject exhibiting symptoms associated with restless leg syndrome, relative to an energy emitting device such that a target nerve within the first portion of the body is in proximity to at least one energy generator disposed within or along the energy emitting device.
A current is then passed through the energy generator to produce, generate or deliver energy, e.g., a magnetic field or electrical or magnetic energy or stimulus, focused on the tibial or posterior tibial nerve or branches thereof or other nerves associated with restless leg syndrome. This in turn causes the stimulation of a pudendal nerve, sacral plexus or other nerves innervating the pelvic floor or various muscles, nerves, or organs associated with restless leg syndrome. The various nerves may stimulated directly or indirectly. Optionally, a current is passed through one more coils, which generates a magnetic or electromagnetic field which stimulates the posterior tibial nerve. In certain variations, the positioning of the coils relative to the first portion of the subject's body may be adjusted to re-focus the magnetic field on the posterior tibial nerve as needed. In certain variations, a current is passed through a microneedle patch generating or delivering an electrical or magnetic stimulus or field. The positioning of the microneedle patch relative to the first portion of the subject's body may be adjusted to re-focus the electrical or magnetic stimulus or field on the posterior tibial nerve as needed.
Optionally, electrical conduction through the target nerve, e.g., the posterior tibial nerve, or muscle stimulation can be detected via at least one sensor. A conductive sensor may be positioned in proximity to the posterior tibial nerve along a second portion of the subject's body. Optionally, a sensor may be positioned over a corresponding muscle to detect muscle stimulation or twitching resulting from nerve stimulation. Optionally, the electrical conduction is detected along a second portion of the subject's body which is different from the first portion of the body. Optionally, the sensor in the form a of a microneedle patch. In certain variations, the sensor may be positioned behind a subject's knee to detect the electrical conduction along the afferent posterior tibial nerve or on another portion of a patient's leg or foot. In other variations, the sensor may be positioned within or along a housing along with the one or more conductive coils.
Where a sensor is used, a signal is received from the sensors and the signal is indicative of the electrical conduction of the target nerve, e.g., posterior tibial nerve. The current may be adjusted or varied using a controller which is in communication with the energy generator. Adjustments may be made in response to the nerve or muscle stimulation detected by the conductive sensor, in order to optimize or ensure adequate treatment of restless leg syndrome by achieving the appropriate level of conductance and appropriate level of nerve or muscle stimulation. Appropriate levels or parameters for current, frequency, magnetic field, treatment duration, etc., are those that result in an observed or detected reduction or prevention of symptoms associated with restless leg syndrome. Treatment could also be administered and the appropriate levels and parameters achieved through observing or detecting reduction or prevention of symptoms where a sensor is not used. Examples of these symptoms include but are not limited to: uncomfortable sensations in the limbs, irresistible urges to move, usually the legs; motor restlessness; when at rest, symptoms return or worsen; and symptoms worsen in the evening and at night.
In certain variations, the amplitude, frequency, direction of a generated magnetic field, electrical or magnetic stimulus, or firing sequence of the coils or microneedles making up the microneedle array may be adjusted. Optionally, the current may be varied according to a muscular response in the patient. Thus, to treat restless leg syndrome, the magnetic field or electrical stimulus is applied to a subject or patient until the desired effects (e.g., reduction of symptoms) are achieved.
In certain variations, methods of treating a subject suffering from premature ejaculation or various pelvic floor disorders utilizing the energy emitting systems described herein are contemplated. Symptoms associated with premature ejaculation may be observed, detected, or diagnosed. An energy emitting device having one or more energy generators, e.g., one or more conductive coils or one or more microneedle patches, may be positioned in proximity to a target nerve, e.g., the tibial or posterior tibial nerve or popliteal or sacral nerve or branches thereof of a subject along a first portion of a subject's body. The subject may or may not be exhibiting symptoms associated with premature ejaculation. In the case of the conductive coils, the coils may be positioned within or along a housing, such as a foot or knee cradle, and a foot or leg may be positioned therein. In the case of a microneedle patch, the patch may be attached to a subject's skin. Optionally, the method involves positioning a first portion of a subject's body, the subject exhibiting symptoms associated with premature ejaculation, relative to an energy emitting device such that a target nerve within the first portion of the body is in proximity to at least one energy generator disposed within or along the energy emitting device.
A current is then passed through the energy generator to produce, generate or deliver energy, e.g., a magnetic or electromagnetic field or electrical or magnetic energy or stimulus, focused on the tibial or posterior tibial nerve or branches thereof. This in turn may cause the stimulation of a pudendal nerve, sacral plexus, or other nerves in the pelvic floor or nerves associated with the control of ejaculation. Various nerves innervating the various muscles, sphincters, nerves, organs and conduits of the urinary tract, bladder or reproductive system, or pelvic floor may be stimulated directly or indirectly. Optionally, a current is passed through one more coils, which generates a magnetic or electromagnetic field which stimulates the posterior tibial nerve. In certain variations, the positioning of the coils relative to the first portion of the subject's body may be adjusted to re-focus the magnetic field on the posterior tibial nerve as needed. In certain variations, a current is passed through a microneedle patch generating or delivering an electrical or magnetic stimulus or field. The positioning of the microneedle patch relative to the first portion of the subject's body may be adjusted to re-focus the electrical or magnetic stimulus or field on the posterior tibial nerve as needed.
Optionally, electrical conduction through the target nerve, e.g., the posterior tibial nerve, or muscle stimulation can be detected via at least one sensor. A conductive sensor may be positioned in proximity to the posterior tibial nerve along a second portion of the subject's body. Optionally, a sensor may be positioned over a corresponding muscle to detect muscle stimulation or twitching resulting from nerve stimulation. Optionally, the electrical conduction is detected along a second portion of the subject's body which is different from the first portion of the body. Optionally, the sensor in the form of a microneedle patch. In certain variations, the sensor may be positioned behind a subject's knee to detect the electrical conduction along the afferent posterior tibial nerve or on another portion of a patient's leg or foot. In other variations, the sensor may be positioned within or along a housing along with the one or more conductive coils.
Where a sensor is used, a signal is received from the sensors and the signal is indicative of the electrical conduction of the posterior tibial nerve. The current may be adjusted or varied using a controller which is in communication with the energy generator. Adjustments may be made in response to the nerve or muscle stimulation detected by the conductive sensor, in order to optimize or ensure adequate treatment of premature ejaculation by achieving the appropriate level of conductance and appropriate level of nerve or muscle stimulation. Appropriate levels for current, frequency, magnetic field, treatment duration, etc., are levels that result in an observed or detected reduction or prevention of symptoms associated with premature ejaculation. Treatment could also be administered and the appropriate levels and parameters achieved through observing or detecting reduction or prevention of symptoms where a sensor is not used. Examples of these symptoms include but are not limited to: ejaculation that frequently occurs within one minute or less of penetration, the inability to delay ejaculation on penetrations; or persistent or recurrent ejaculation with minimal stimulation before, on or shortly after penetration.
In certain variations, the amplitude, frequency, direction of a generated magnetic field, electrical or magnetic stimulus, or firing sequence of the coils or microneedles making up the microneedle array may be adjusted. Optionally, the current may be varied according to a muscular response in the patient. Thus, to treat premature ejaculation, the magnetic field or electrical stimulus is applied to a subject or patient until the desired effects (e.g., reduction of symptoms) are achieved.
Exemplary treatment parameters for treating various conditions, e.g., urinary incontinence, using the systems and methods described herein may include the following. Operation of a conductive coil at about 10 to 20 hertz generating a magnetic field of about 0.25 to 1.5 tesla, where the coil is administered to a patient for a duration of about 30 minutes/day or 30 minutes per week, depending on the severity of the symptoms, until the symptoms subside. The above treatment parameters or variations on the parameters may be used for treatment of urinary incontinence, fecal incontinence, restless leg syndrome, or premature ejaculation or other conditions. For example, the coil may be operated at various parameter ranges falling with the following ranges: about 5 to 100 hertz, about 1 to 10 tesla, for about 15 minutes to 2 hours per day or week. In treating premature ejaculation, a patient may receive treatment about 4 to 10 hours prior to intercourse. A maintenance phase of treatment, after the initial treatment, may vary for various conditions. For example, the maintenance phase may require application of the systems and methods described herein at the parameters described herein for 30 minutes/week or 30 minutes/month. Any treatment parameter may be varied or modified based on the effect on the patient or sensor or patient feedback regarding stimulation, until the desired result of treating or preventing a condition is achieved.
In certain variations, as shown in
In certain variations, as shown in
In certain variations, the energy emitting device, e.g., foot support or cradle, knee support or cradle, etc., includes a conductive coil positioned such that a target nerve is automatically targeted. The conductive coil is configured, sized and positioned within the device such that the generated magnetic field may encompass and stimulate the target nerve in any patient based on the target nerve's anatomical location, thus providing automatic targeting of the nerve in any patient once the patient positions a particular body portion in the device.
In certain variations described herein, sensors may detect voltage or current and may be connected, coupled, wirelessly connected or coupled or otherwise in communication with housing, conductive coils, microneedle patch, energy emitting apparatus, energy generators, or electrode needles and/or controller using a variety of methods or techniques known in the art. In various variations described herein, housings, conductive coils, microneedle patches, energy emitting apparatus, energy generators, or electrode needles may be connected, coupled, wirelessly connected or coupled or otherwise in communication with each other, controllers or sensors, using a variety of methods or techniques known in the art.
Coils used in any of the variations described herein and illustrated in the corresponding FIGS. may take on a variety of shapes, sizes, and configurations. For example, a coil may be shaped as a spiral (as shown) or have a simple helical pattern or be a FIG. eight coil, a four leaf clover coil, a Helmholtz coil, a modified Helmholtz coil, or may be shaped as a combination of the aforementioned coil patterns. Additionally, other coil designs beyond those mentioned hereinabove might be utilized as long as a magnetic field is developed that will encompass a target nerve.
The coils may have a variety of dimensions and configurations. In certain variations, a coil may have a central aperture. The diameter of the aperture may range from about 0.5 inch to 2 inches or 1 inch to 1.5 inches or the aperture may have a diameter of about 1 inch. The diameter of the coil body may vary. For example, the diameter may range from about 3.0 to about 7 inches or from about 4 to about 5 inches or the diameter may about 4.5 inches. The coil body may include any suitable number of turns. For example, the coil body may include from about 2 to about 25 turns or from about 10 to about 20 turns or 14 to 17 turns. The adjacent turns may be spaced apart from each other, providing a gap there between. An end or cross section of a turn may have various dimensions. For example, the end or cross section may have a height that is greater than its width. An end or cross section of a turn may have a height ranging from about 1 to 5 cm or from about 10 mm to 51 mm (about 0.3 inches to 2 inches) or about 25 mm to 40 mm (about 1 inch to 1.5 inches) or about 12 mm to 40 mm (about 0.5 inch to 1.5 inch) or about 0.5 inch to 2 inch. The end or cross section of the turn may have a width ranging from about 0.5 mm to about 5 mm (about 0.019 inch to 0.19 inch) or from about 1 mm to about 2 mm (about 0.03 inch to 0.07 inch) or about 0.2 mm to about 1.6 mm (about 0.01 inch to 0.06 inch). The above are all exemplary dimensions, where other dimensions are also contemplated depending on the use and configuration of a device.
In certain variations, a system or device for electromagnetic or magnetic induction therapy may include one or more conductive coils disposed within or along an applicator. The coil may be configured to generate an electromagnetic or magnetic field focused on a target nerve, muscle or other body tissue positioned in proximity to the coil. The system may also include one or more sensors. The sensor may be configured to detect electrical conduction in the target nerve or to detect stimulation of a muscle or other body tissue. The sensor may also detect a muscular response caused by an electrical conduction in a target nerve. The sensor provides feedback about the efficacy of the applied electromagnetic or magnetic induction therapy. Optionally, a user may provide such feedback based on detection by the user, with or without the use of a sensor. The system may also include a controller which is in communication with the sensor. The controller may be adjustable to vary a current through the coil in order to adjust the magnetic field focused upon the target nerve based on feedback from the sensor or user. The various systems or devices described herein may be utilized with or without a sensor.
A variety of electromagnetic or magnetic induction applicators designed or configured to stimulate various portions of a patient's body for treating various conditions are contemplated herein.
The back applicator 330 may include several coils, which may be pulsed intermittently. In certain variations, a sensor may be placed on muscle in dermatome to provide feedback to ensure stimulation of the proper dorsal root ganglion or vertebral body. The sensor may provide feedback to channel energy or current to the proper or effective coil in an applicator, e.g., in an applicator having multiple coils.
In certain variations, an applicator may be designed to ergonomically target common nerves responsible for common neuralgias in order to treat such neuralgias. In other variations, an applicator may be used for treating neuralgias virtually anywhere on a patient's body, including in deep nerves due to the ability of magnetic fields generated by the applicator to penetrate painlessly. In certain variations, an applicator may be designed to generate a magnetic field focused on a target nerve to treat central or peripheral neuralgias.
In certain variations, an applicator may include one or two (bilateral) magnetic field generating coils, which may be positioned around the knee when the patient is in a sitting, standing or prostrate position. In certain variations, a pulse generator or logic controller 397 may send energy through one or more coils to create an electromagnetic or magnetic field. The applicator or coils may generate stimulator or non-stimulator fields. Sensor or user feedback may provide feedback to logic controller to optimize therapy, e.g., with the stimulator fields. An applicator may be utilized for generating magnetic fields focused on an area of a patient's body, e.g., the knee, to treat various orthopedic indications, e.g., knee pain or osteoarthritis. An applicator may be utilized for generating magnetic fields focused on a area of a patient's body to treat various non-orthopedic indications, via, e.g., peripheral nerve stimulation.
For any of the applicators described herein, such applicators may include one or more of the following features. The applicators may be ergonomic or contoured to the specific region of the body or anatomy to which the applicator will be delivering stimulation. The applicators may be configured or designed to be positioned relative to, on, around, or in proximity to a specific region of the body or the applicators may be configured or designed such that the targeted region of the body may be positioned relative to, on, around or in proximity to the applicator. The applicators may be openable or adjustable to allow for insertion or entrance of the targeted body part or anatomy into the applicator or to allow for placement of the applicator onto or around the targeted body part or anatomy. The applicators may be flexible or ergonomic to accommodate nearly any type of body habitus. In certain variations, a solenoid-type coil may incorporated into an applicator for delivering PEMF stimulation directly to the targeted areas or regions of a body. In certain variations, any of the applicators described herein may approximate the respective targeted body area or anatomy or the applicators may be designed such that the body region or targeted anatomy may approximate the applicator.
In certain variations, any of the applicators or systems described herein may be used to provide electromagnetic or magnetic induction therapy with or without a sensor.
In certain variations, electromagnetic stimulating devices or applicators for providing stimulation to tissues of the human body, including nerves, muscles (including superficial and deep muscles), and/or other body tissues for the treatment of various conditions, including, e.g., chronic and acute pain, are provided.
The devices may utilize an inductive coil encased within an ergonomic, body-contoured applicator to target specific regions of the body. The coils may be designed to target peripheral nerves throughout the body that have been implicated or involved in pain syndromes.
The various designs and configuration of the devices described herein allow for easier application, more consistent therapy and home use while targeting anatomic regions with therapeutic pulsed electromagnetic fields. The fields may also be delivered or applied in an intermittent manner to allow for convenience and ease of use while providing a durable benefit. With intermittent external stimulation by pulsed electromagnetic or magnetic fields, a nerve or other tissues may be stimulated in manner that provides a continued and lasting effect on nerve, muscle or tissue function, without habituation.
The electromagnetic or magnetic induction stimulation devices described herein substantially improve the state of the art electromagnetic stimulation technology and may incorporate the delivery of PEMF therapy into a user friendly, body contoured applicator. In certain variations, a delivery system for PEMF therapy may include elements such as, e.g., (1) an ergonomic, body contoured applicator which provides for repetitive application and consistent therapy onto the same body area. The applicator may be coded with clear markings to facilitate repetitive and consistent therapy onto the same body area; (2) the use of a sensor to provide feedback that stimulation is occurring effectively; and/or (3) the use of intermittent stimulation to effectively treat various conditions, e.g., chronic pain, without habituation. These elements individually or the various combinations of these elements have provided for an easy to use, ergonomically designed system that has applications within a host of clinical and home ease of use health applications.
In certain variations, an electromagnetic or magnetic induction stimulation device able to provide stimulation to tissues of the human body, including nerves, muscles (including superficial and deep muscles), and/or other body tissues without significant discomfort to the patient is provided. Conductive stimulating coils may be encased in an ergonomic, body-contoured applicator that is coded with clear markings to provide for repetitive application and consistent therapy onto the same body area. The design of the applicator allows for ease of use and also for the targeting of anatomic regions to be exposed to the impulses of the PEMFs. The electromagnetic stimulating device may provide PEMF in a manner that is patient user friendly and the device may be portable. The device may be utilized in a hospital, an outpatient clinic, a therapist's office, or at a patient's home.
In certain variations, an electromagnetic or magnetic induction stimulation device may stimulate regions of the body to treat conditions requiring both maximal stimulation (i.e., sufficient to cause contraction of muscle fibers and firing of nerves) as well as submaximal stimulation (which will be sufficient to provide therapy but not to cause contraction of muscle fibers).
The electromagnetic or magnetic induction or stimulating devices described herein may be utilized for various indications. The indications may be divided into maximal and submaximal categories, in which the former requires significantly higher levels of inducting current than the latter. The maximal applications of the device include, but are not limited to: Non-invasive stimulation (intermittent or continuous) of the peripheral nervous system for treating chronic pain; stimulation of a nerve for the up- or down-regulation of hormones or cellular proliferation; treatment and/or prevention of atrophy, which would be therapeutic during recovery after an individual sustains a fracture, experiences paralysis of a limb or other body part, or undergoes surgery, such as ACL repair in the knee; treatment of neurogenic or overactive bladder and bowel; and stimulation of the central nervous system to alter neural pathways or up/down-regulate the aforementioned factors.
Additional applications of the devices include but are not limited to: treatment of neuropathic pain (e.g., phantom pain in limbs or other neurologic pain) or orthopedic pain (back and neck pain or skeletal related pain); treatment of overactive bladder and bowel; and treatment of arthritis and/or orthopedic conditions.
In certain variations, a device is provided for delivering PEMF stimulation to selective anatomic regions of the body, utilizing an ergonomic applicator designed to facilitate accurate and targeted delivery of therapy. The applicator may be coded with clear or solid markings to provide for repetitive application and consistent therapy onto the same body area of the body. This design may facilitate the placement of the device for the stimulation of key nerves, muscles, and/or body tissues.
In certain variations, a device is provided which may be utilized to electromagnetically stimulate selective nerves, muscles, and/or body tissues, where the device is user friendly and capable of being used even by an unskilled patient in a home healthcare setting.
In certain variations, a device is provided to electromagnetically stimulate selective nerves, muscles, and body tissues to provide consistent therapy, with an ergonomic applicator targeting key nerves and eliminating the requirement for a highly trained operator to manipulate the device.
In certain variations, an electromagnetic or magnetic induction system or device may be configured or designed to provide intermittently applied pulsed magnetic fields in the treatment of chronic conditions, such as pain. For example, a device as described herein may provide shorter, intermittent stimulation to treat chronic pain or other chronic conditions. The delivery of pulsed magnetic fields may have a continued and lasting effect on nerve function in treating conditions, such as, overactive bladder as well as other chronic neurological and orthopedic conditions such as neuropathic pain, restless legs and orthopedic pain (e.g., spinal pain, back pain, etc.)
In certain variations, intermittent pulsed magnetic fields may be utilized for the treatment of chronic and acute non-orthopedic conditions such as neuropathic pain, phantom pain and chronic neuralgias, as well chronic and acute orthopedic conditions, such as back pain and neck pain. The therapeutic magnetic fields may be applied frequently (e.g., several times a day) or less frequently (e.g., once a week or once a month) depending on the durability of the effect for the individual patient. Treatment involving the use of magnetic fields does not require surgery or needles to stimulate a nerve. Also, the deliver of intermittent pulsed magnetic fields prevents the nerve from becoming habituated to the stimulator signal by ensuring that there are periods during which the nerve is not subjected to the stimulator signal. Accordingly, the electromagnetic or magnetic induction systems or devices described herein may provide unparalleled ease of use, non-invasiveness, reliability of therapy based on sensor feedback and/or ergonomic targeting, and/or a lack of habituation due to intermittent stimulation provided by certain systems and devices.
In certain variations, the electromagnetic or magnetic induction systems or devices described herein may incorporate an air-cooled coil wherein the air coolant, e.g., liquid or air, is drawn through and/or in between the turns of the inductive coil, in direct contact with conductive surfaces of the coil. Drawing air or other fluid through the coil prevents the coil from heating up to the degree that could damage the coil and the electronics of a device, or expose the patient to excessive temperatures.
In certain variations, the systems and devices described therein may be utilized to stimulate nerves for a variety of conditions, including, e.g., atrophy prevention, nerve repair/regeneration, neuromodulation, chronic pain, up or down regulation of hormones, restless legs, phantom pain, etc. The systems and devices may also be used to stimulate muscles and/or other body tissues to accelerate tissue healing, regeneration and/or growth.
In certain variations, the electromagnetic or magnetic induction systems or devices described herein and other implantable or extracorporeal devices may allow for the automatic adjusting of nerve stimulation based on feedback.
In one variation, an extracorporeal or implantable device, e.g., any of the electromagnetic or magnetic induction devices described herein, a pacemaker, defibrillator, or nerve stimulator, may include a feature that allows for automatic adjustment of nerve stimulation based on feedback provided by a sensor or user. This feature may minimize pain and power usage while ensuring optimal therapy delivery. A device may include a stimulator and a sensing component. The stimulator may be automatically adjustable based on feedback from the sensor up to a maximal (safe) threshold. Each therapy may start with lower powered pulses, followed by increasing power pulses until the sensor detects stimulation. The algorithm allows for the minimal amount of power to be used and allows for automatic adjustment of power settings as conditions change. 102411 In one variation, an implantable device may include a sensor, such that the device can stimulate tissue or nerves and sense stimulation at the same site. For example, the sensor may provide feedback to the implantable device regarding nerve conduction at the site of stimulation. As fibroses develops around an implant, at the site of stimulation, the feedback will indicate whether a target nerve is no longer being effectively stimulated due to the fibroses, which will cause the power or level of stimulation to increase or decrease, as is necessary, to effectively stimulate the target site and overcome any obstruction due to fibroses. As fibroses occurs around an implant, a patient need not report back to a physician or other operator for adjustment of the stimulator power of the device. The device will automatically adjust the stimulator power or level based on sensing stimulation of the target nerve or tissue. This eliminate the guesswork involved by the user in monitoring their therapy one day at a time on their own, as they notice the effect of the therapy wear off. This also eliminates the risk of the user being exposed to unnecessarily high power levels that might otherwise by set in order to minimize frequent return visits to a physician or operator for adjustments.
The automatic adjustment feature based on sensor feedback may be utilized in any stimulator or non-stimulator implant or extracorperal device, where the device incorporates a sensor capable of detecting the desired stimulus and a feedback loop capable of automatically adjusting parameters (e.g., power, frequency, etc.) to ensure appropriate stimulation.
In certain variations, the electromagnetic or magnetic induction systems or devices described herein and other implantable or extracorporeal devices may include a feature that allows for automatic targeting of coils.
A device may include multiple inductive coils or one or more movable inductive coils. The device may also include a sensor based feedback algorithm. In one variation, the device includes a targeting or movable coil which may be positioned over or in proximity to a patient's body at a site that elicits a response that can be sensed automatically or detected by a user. Once this response is detected, the coil may either move to its stimulation position, or in the event that a small targeting coil is used, the therapeutic coil may already overlie the treatment area. Once the response is detected, the therapy may automatically begin.
In one example, relating to the treatment of depression, the motor cortex is stimulated until the thumb is seen to move. The coil may then be advanced, e.g., about 5 cm. to about 5 inches, forward to a position over the frontal cortex. This feature eliminates the guesswork that may otherwise be involved in moving or positioning a coil, and automates therapy based on user feedback or EMG sensor or other sensor feedback, e.g., over a thumb.
In another variation shown in
The feature that allows for automatic targeting of coils may be utilized in any device designed to stimulate nerve, body or other tissues with stimulator or sub-stimulator fields in which the device may be targeted based on a detectable signal or response.
Other conditions that may be treated utilizing the various electromagnetic or magnetic induction stimulation systems and methods described herein include but are not limited to: pelvic pain, interstitial cystitis, fibromyalgia, chronic fatigue and preterm labor, pain syndromes, Irritable Bowel Syndrome, Vulvodynia, Herpetiuc neuralgia, trigeminal neuralgia and Myofascial pain.
The following Examples are provided for illustration, not limitation. One with skill in the art would be able to use these examples as guidelines for making and using comparable devices.
In each example, intermittent therapy should be applied and symptoms/scores tracked for a minimum of 6 weeks in order to determine the full extent of the therapies effect.
Empirical Testing of Efficacy in the Treatment of Neuropathic Pain: The optimal stimulus intensity for neuropathic pain treatment; the optimal application parameters, i.e. frequency of stimulation, duration of treatment, location of stimulator coils in each disposable array of coils; and the optimal coil diameter/placement within the strays can be determined using the following experimental protocol: Before, during and after treatment, patients will report scores of neuropathic pain after weekly stimulation over a minimum of 6 weeks.
Empirical Testing of Efficacy in the Treatment of Neuromuscular Pain: The efficacy of neuromuscular pain treatment can be tested by monitoring patient reported pain scores. A standardized scale may be utilized and, when feasible, local biopsy and blood tests can be useful in determining the impact of the therapeutic fields on circulating factors and local mediators. The optimal pulse amplitude, duration, site of stimulation will be assessed based on reported pain scores and diagnostic tests.
Empirical Testing of Efficacy in the Treatment of Orthopedic conditions (i.e., Arthritis, Back pain and neck pain): The efficacy of arthritis treatment can be tested by monitoring patient reported functionality scores. A standardized subjective functionality scale may be utilized and, when feasible, local biopsy may be useful in determining the impact of the therapeutic fields on the cartilage and arthritic regions treated. As cartilage destruction is a well-studied side-effect of arthritis, reduction of this degeneration will be a valuable marker for efficacy of therapeutic treatments. The optimal pulse amplitude, duration, site of stimulation will be assessed based on reported functionality scores and diagnostic tests. Pain scores may also be measured to determine the device's impact on orthopedic conditions such as back pain, neck pain, etc. A standardized pain scale may be used before and after treatment to determine potential benefit.
It is also contemplated that any of the energy emitting systems or devices described herein can be used with or without a sensor for detecting conduction of a stimulated nerve or muscle of tissue stimulation resulting from the electromagnetic or magnetic field generated by the conductive coil and delivered to a patient or an electrical stimulus delivered to a patient. Also, in any of the above variations, a controller may optionally be connected, coupled, integral to or otherwise in communication with the conductive coils and/or the sensor. Optionally, the sensor may be connected, coupled, integral to or otherwise in communication with the conductive coil.
In certain variations, transdermal electrical stimulation therapy may be provided to a patient or subject. One or more stimulator electrodes may be positioned over a glabrous skin surface overlying a target nerve of the patient. Electrical stimulation or an electrical stimulus may then be delivered via the stimulator electrode through or across the glabrous skin surface to the target nerve in the patient, to stimulate the target nerve. The electrical stimulation may be delivered at a level and in a manner sufficient to generate motor and/or sensory nerve conduction. For example, the electrical stimulation may be delivered at a frequency of about 5 Hz to about 80 Hz or about 20 HZ or 30 Hz to about 60 Hz or greater than 30 Hz, while remaining safe and tolerable to the patient. The stimulation may be delivered in a non-invasive manner.
The stimulator electrode may be positioned over or on various regions of the body. In certain variations, the stimulator electrode may be positioned over a palm or plantar skin surface. For example, the stimulator electrode may be positioned over a glabrous skin surface overlying a plantar nerve of the foot, where the electrical stimulation may stimulate the plantar nerve. Stimulation of the plantar nerve may result in stimulation of the tibial nerve to treat various conditions, such as urinary incontinence or overactive bladder.
Various types of electrodes may be utilized as a stimulator electrode, e.g., a surface electrode, microneedle electrode, a TENS (transcutaneous electrical nerve stimulation) patch or other electrode that may be positioned over or on the skin surface. Optionally, a conductive substance may be injected or implanted near the target nerve to improve electrical conductivity to the target nerve from the stimulatory electrode. Electrical stimulation may be delivered intermittently or on a chronic basis and may include one or more electrical signals designed to be constructive and/or destructive in order to improve tissue penetration and/or signal tolerance.
In certain variations, nerve stimulation resulting from the electrical stimulation therapy may be detected via at least one sensor positioned on or near the subject. The sensor may provide a signal indicative of the detected electrical stimulation thereby providing feedback about the efficacy of the applied electrical stimulation therapy such that the therapy may be adjusted or optimized. The feedback loop may be queried such that the electrical stimulation therapy may be adjusted to ensure that the minimum amount of energy is being applied to stimulate the target nerve while reducing the risk of burns or intolerance. Optionally, the feedback loop may be queried such that the positioning of the stimulator electrode may be adjusted to optimize the electrical stimulation therapy. Various sensors may be utilized, including but not limited to a surface electrode, a microneedle electrode, or motion sensor. In certain variations, the sensor may detect conduction of motor and/or sensory nerves. For example, a sensor may detect afferent or efferent nerve stimulation of the target nerve or other nerve or related nerve.
In certain variations, a ground electrode may be utilized. For example, the ground electrode may be positioned on the subject to facilitate penetration of an electrical current from the stimulatory electrode through the glabrous surface to stimulate the target nerve.
In certain variations, a stimulator electrode may be positioned over or attached to a glabrous skin surface with an adhesive or other form of attachment or fastener. In other variations, a stimulator electrode may be positioned over or held in contact with a glabrous skin surface with an ergonomic applicator.
Various applicators for positioning electrodes over various regions of the body to deliver electrical stimulation, e.g., transdermal electrical stimulation, may be utilized. For example, an applicator may be in the form of an insole configured to be positioned against, over, or in contact with the plantar surface or a glabrous surface of the foot such that an electrode positioned on the insole may deliver electrical stimulation through or across the glabrous skin surface, to a target nerve or other tissue within the foot. The insole may be configured to be positioned in an orthotic or a shoe such that electrical stimulation may be delivered to the subject while the subject is walking. In other variations, an applicator may be a foot plate or foot rest or cradle on which a foot or other portion of the leg or body may be positioned to receive electrical stimulation through or across a glabrous skin surface, from the plate, rest or cradle or from electrodes of the plate, rest or cradle.
In another variation, an applicator may be a glove, brace or other hand wrap which is configured to be positioned against, in contact with, or over the palmar surface or a glabrous surface of a hand. An electrode of the glove, brace or wrap may deliver electrical stimulation through or across the glabrous skin surface to a target nerve or tissue within the hand.
Various applicators or ergonomic applicators may be utilized to provide transdermal electrical stimulation therapy to a subject. In certain variations, an applicator may include one or more stimulator electrodes and one or more electrical pulse generators or controllers. The electrical pulse generator may be coupled to the stimulator electrode. The electrical pulse generator may be incorporated into the applicator. Optionally, the electrical pulse generator may be separate from the applicator or located remotely from the applicator or stimulator electrode. An applicator may include one or more sensor electrodes configured to detect nerve stimulation and/or provide feedback about the efficacy of the applied electrical stimulation therapy. Such feedback may allow the therapy to be adjusted, modulated and/or optimized. A sensor may detect motor and/or sensory nerve stimulation. For example, a sensor may detect afferent or efferent nerve stimulation. In certain variations, the positioning of the stimulator electrode may be adjusted based on feedback from the sensor in order to optimize the electrical stimulation therapy.
The applicator may also include one or more ground or ground electrodes. The ground electrode may be a component of the applicator or may be attached to the subject separately via a strap or other attachment. The ground electrode may be positioned on the subject at a location that facilitates penetration of an electrical current through a glabrous surface to stimulate the target nerve, e.g., on the opposite surface of a body portion relative to the stimulator electrode.
As stated supra, an applicator may be designed for various portions of the body. An ergonomic applicator may be in the form of an insole which can be positioned against or over the plantar surface or a glabrous surface of the foot such that an electrode of the insole may deliver electrical stimulation through or across a glabrous surface to a target nerve within the foot. The insole may be positioned in an orthotic or a shoe. An applicator may also be in the form of a foot plate, cradle or foot rest. In other variations, the applicator may be designed to stimulate a portion of a hand or upper extremity. For example, an ergonomic applicator in the form of a glove or brace may be configured to be positioned against or over the palmar surface of a hand such that the electrode of the glove or brace may deliver electrical stimulation across a glabrous surface to a target nerve within the hand.
Any of the various applicators or ergonomic applicators described herein may be designed or configured to position or locate one or more stimulator electrodes over or on a glabrous skin surface of a patient to deliver transdermal electrical stimulation through or across the glabrous skin surface to an underlying target nerve or other tissue to stimulate the target nerve or tissue. In certain variations, an electrical stimulation or stimulus may be delivered at various frequencies, e.g., at a frequency of about 5 Hz to about 60 Hz, or greater than 30 Hz, while remaining safe and tolerable to the subject.
In other variations, one or more stimulatory electrodes may be otherwise attached to a skin surface, e.g., as a wired or wireless patch, adhesive or microneedle electrode in the absence of an ergonomic applicator, to deliver electrical stimulation across or through a glabrous skin surface to stimulate an underlying target nerve or tissue. Such electrodes may be used in combination with one or more ground or sensor electrodes. Such stimulatory electrodes may deliver electrical stimulation or an electrical stimulus at various frequencies, e.g., at a frequency of about 5 Hz to about 60 Hz, or greater than 30 Hz, while remaining safe and tolerable to the subject.
In certain variations an applicator may include two or more stimulator electrodes or two more stimulator electrodes may be utilized without an applicator.
In certain variations, methods, systems and/or applicators for providing an energy based stimulation therapy to a subject are provided. An energy emitting device may be positioned in proximity to a glabrous surface overlying a target tissue. Energy may be delivered from the energy emitting device through or across the glabrous skin surface to the target tissue to stimulate the target tissue, such as a target nerve. Various energy sources or forms of energy may be delivered by the energy emitting device, including but not limited to, an electric current, an electromagnetic or magnetic field or electromagnetic or magnetic induction stimulation, ultrasound, or RF fields. Energy may be delivered at various frequencies. For example, the energy may be delivered at a frequency of about 5 Hz to about 60 Hz, or greater than 30 Hz, while remaining safe and tolerable to the subject.
The energy based stimulation therapy may be utilized to treat various conditions, including but not limited to, overactive bladder, urinary incontinence, fecal incontinence, chronic pain, depression, migraine, epilepsy, obesity, restless leg syndrome, or foot drop. Energy may be delivered through a glabrous skin surface to provide neuromodulation to stimulate other tissue. Tissues that may be stimulated include but are not limited a central nerve, peripheral nerve, muscle, skin, or vasculature. Optionally, conductive substance may be implanted or injected near a target tissue to improve conductivity to the target tissue.
In one variation, a method for treating urinary incontinence or overactive bladder in a subject may include one or more of the following steps, a stimulator electrode, (e.g., in an applicator or as an adhesive or attachment electrode as described above) may be positioned over a glabrous skin surface overlying a plantar nerve or other nerve in the foot. An electrical stimulation may then be delivered through or across the glabrous skin surface to the plantar nerve to stimulate the plantar nerve which results in stimulation of the tibial nerve to treat urinary incontinence or overactive bladder, e.g., via stimulation of the sacral plexus or pudendal nerve.
In other variations, any of the various electrical stimulation methods and electrical stimulator applicators or electrodes may be utilized to treat any of the conditions described herein.
In certain variations, methods, systems and/or applicators for providing an energy based stimulation therapy to a subject are provided. An energy emitting device may be positioned in proximity to or over a skin surface overlying a target nerve or tissue. The energy emitting device may include any of the devices, systems or applicators described herein and/or illustrated in any of the various
Energy delivered transdermally, through, or across a patient's skin at a frequency from about 1 Hz to about 30 Hz, or at a frequency of less than 10 Hz has unexpectedly been found to stimulate or generate motor and/or sensory conduction in a target nerve. For example, energy delivered transdermally, through, or across a patients skin at about 1 Hz to about 30 Hz, or at less than 10 Hz has unexpectedly been found to stimulate or generate motor and/or sensory nerve conduction of a tibial nerve, where such level of stimulation may be sufficient to treat a patient suffering from urinary incontinence, overactive bladder, fecal incontinence or other conditions. The energy may be delivered through or across a glabrous skin surface or non-glabrous skin surface or any other skin surface (e.g., any skin surface overlying a tibial nerve).
Various energy sources or forms of energy may be delivered by the energy emitting device, including but not limited to, an electric current, an electromagnetic or magnetic field, ultrasound, or RF fields. The energy based stimulation therapy may be utilized to treat various conditions, including but not limited to, overactive bladder, urinary incontinence, fecal incontinence, chronic pain, depression, migraine, epilepsy, obesity, restless leg syndrome, or foot drop. Energy may be delivered through a glabrous skin surface or any other skin surface to provide neuromodulation to stimulate other tissue. Tissues that may be stimulated include but are not limited a central nerve, peripheral nerve, muscle, skin, or vasculature. Optionally, conductive substance may implanted or injected near a target tissue to improve conductivity to the target tissue.
In certain variations, electrical stimulation may be delivered transdermally, through or across a glabrous surface on a subject, e.g., through or across a glabrous surface of the foot or hand. The electrical stimulation may be sufficient to generate conduction of motor and/or sensory nerves. An electrical stimulation may be delivered by placing one or more electrodes anywhere over or on a glabrous skin surface overlying one or more target nerves or other target tissue. Stimulator and/or ground electrodes may be utilized. Exemplary electrodes may include but are not limited to surface electrodes, dry electrodes, gel electrodes, microneedle electrodes or any other suitable electrode for delivering an electrical stimulus. An electrode may be adhered or otherwise attached to a glabrous surface. Optionally, an electrode may be held or positioned in contact with, in proximity to, or over a glabrous surface with a wearable garment, cradle, applicator or body portion rest or support (as described in further detail herein).
In one variation, electrical stimulation may be delivered to one or more of the stimulator sites or similar sites depicted in
Delivery of electrical stimulation through or across a glabrous surface of the body via an electrode positioned over a glabrous surface, e.g., a glabrous surface on a palmar or plantar surface, unexpectedly allows for the use of a higher frequency and/or higher amplitude electrical pulsation or electrical stimulus to deliver the electrical stimulation than would otherwise be safe and/or tolerable to deliver electrical stimulation through a non-glabrous surface of the body. For example, an electrical stimulus having a frequency of about 5 Hz to about 60 Hz (a range found to be effective for generating motor and/or sensory nerve conduction of the posterior tibial nerve) may be utilized to stimulate a target nerve (to generate motor and/or sensory nerve conduction therein) or tissue through or across a glabrous skin surface (via an electrode positioned over the glabrous skin surface) in a manner that remains safe and tolerable to the patient and avoids burns or injury. Optionally, an electrical stimulus having a frequency of about 5 Hz to about 60 Hz, or greater than 30 Hz, may be utilized.
In contrast, utilizing an electrical stimulus having a frequency of about 5 Hz to about 60 Hz or greater to stimulate a target nerve or tissue through a non-glabrous skin surface (via an electrode positioned over the non-glabrous skin surface) is intolerable and painful, resulting in burns or injury, and thus making such a procedure impractical and not feasible.
For example, delivering electrical stimulation through a non-glabrous surface of the body, for example, by stimulating a site overlying a nerve near the medial malleolus to elicit a motor response of the abductor hallucis longus, generates a painful shock to the patient. While at a single pulse, such as in the use for EMG diagnostics, such electrical stimulation may be tolerable, as the frequency increases, the shocking sensation builds and quickly becomes painful and intolerable.
It is contemplated that other energy sources, for example, an electromagnetic or magnetic stimulus having a frequency of about 5 Hz to about 60 Hz, or greater than 30 Hz, may be utilized to simulate a target nerve or tissue through a glabrous surface in a manner that remains safe and tolerable to the patient.
Various electrodes and/or applicators for applying an electrical or other energy based stimulation to a patient are described herein.
An electrical pulse generator 524 or other generator or controller may be coupled to the stimulator electrode 522. The electrical pulse generator 524 may be incorporated into or attached to the ergonomic insole 520. A ground electrode 526 or other ground may also be provided. The ground electrode 526 may be attached to a strap or band extending from or attached to the ergonomic insole, an orthotic, shoe, or shoe applicator or elsewhere on the user's body. The ground electrode may be attached to one or more wires or may be wireless and/or may be coupled to the stimulator electrode 522 and/or the pulse generator or controller. The ground electrode 526 may be positioned anywhere on the foot. For example, the ground electrode 526 may be positioned over a posterior or upper surface of the foot to encourage or facilitate deeper penetration of the electrical current or stimulation, through the glabrous surface of the foot, to a plantar nerve or other target nerve within the foot.
In certain variations, the electrical pulse generator or controller may be located distant or remotely from an ergonomic insole and the electrode positioned therein. For example, the electrical pulse generator may attached elsewhere on the body, e.g., attached to a belt, inside a pocket or strapped to the calf or other region of the body. The pulse generator may communicate with or be coupled to the stimulator electrode or other electrodes via a wire or wirelessly.
In certain variations, the ergonomic insole may be custom built into an orthotic, shoe applicator or other support for the foot. For example, the ergonomic insole may be built into an orthotic or shoe applicator providing the user with the freedom to walk around while receiving electrical stimulation therapy.
Other ergonomic applicators, designed for other regions of the body, may be utilized for delivering electrical stimulation over various regions of the body. For example,
In other variations, an applicator may be in the form of a foot or hand plate, cradle or support. For example, a user may rest or position their bare foot on a foot plate to receive electrical stimulation therapy from the plate or from one or more electrodes attached to the plate. In another variation, a user may rest or position the palm of their hand on a hand plate or support to receive electrical stimulation therapy from the plate or from one or more electrodes attached to the plate to provide upper extremity stimulation.
An electrical pulse generator 534 or other generator or controller may be coupled to the stimulator electrode 532. The electrical pulse generator 534 may be incorporated into or attached to the ergonomic insole 530. A ground electrode 536 or other ground may also be provided. The ground electrode 536 may be attached to a strap or band extending from or attached to the ergonomic insole, an orthotic, shoe, or shoe applicator or elsewhere on the user's body. The ground electrode may be attached to one or more wires or may be wireless and/or may be coupled to the stimulator electrode 532 and/or the pulse generator or controller. The ground electrode 536 may be positioned anywhere on the foot. For example, the ground electrode 534 may be positioned over a posterior or upper surface of the foot to encourage or facilitate deeper penetration of the electrical current or stimulation, through the glabrous surface of the foot, to a plantar nerve or other target nerve within the foot.
In certain variations, the electrical pulse generator or controller may be located distant or remotely from an ergonomic insole and the electrode positioned therein. For example, the electrical pulse generator may attached elsewhere on the body, e.g., attached to a belt, inside a pocket or strapped to the calf or other region of the body. The pulse generator may communicate with or be coupled to the stimulator electrode or other electrodes via a wire or wirelessly.
The ergonomic insole 530 may include one or more sensor electrodes. For example, in one variation, as shown in
In certain variations, one or more separate or dedicated sensor electrodes may be attached to or positioned in the ergonomic insole to continuously and/or intermittently sense the stimulation of a nerve. The ergonomic insole 530 may hold or position the sensor electrode against, in contact with, or in proximity to a glabrous surface of the foot, to detect stimulation of the underlying target nerve, to provide feedback regarding the efficacy of the applied electrical stimulation therapy. For example, the sensor electrode may detect motor and/or sensory nerve conduction. Detection of and feedback regarding nerve stimulation via the sensor electrode may provide for automatic adjustment of the treatment parameters or may guide or allow for manual adjustment of the treatment parameters in order to optimize stimulation therapy. The sensor electrode may be coupled or connected to the electrical pulse generator. In certain variations, the sensor electrode may be attached to one or more wires, may be wireless and/or may be coupled to a pulse generator and/or may not be attached to or positioned in an ergonomic insole.
A sensor or applicator with a sensor may be capable of sensing stimulation of a nerve underlying a glabrous surface allowing for manual or automatic feedback to adjust the parameters of the stimulation or the position of the area being stimulated in order to optimize the stimulator therapy.
In certain variation, as sensor electrode may be positioned or placed along the path of the nerve conduction or motor or sensory impulse. Optionally, the sensor electrode may be positioned proximal to a stimulator electrode, along the nerve conduction path
In certain variations, the electrodes may be attached to one or more wires, may be wireless and/or may be coupled to an electrical pulse generator or other generator or controller and/or may not be positioned in an applicator or support. For example, one or more stimulator electrodes may be adhered to or otherwise attached to a glabrous skin surface of a patient. One or more ground electrodes and/or one or more sensor electrodes may also be adhered to or otherwise attached to the patient. The stimulator electrode, sensor electrode and/or ground may be coupled to an electrical pulse generator and/or each other using wires or wirelessly. The electrical pulse generator may positioned in various locations or located anywhere on a patient. For example, the electrical pulse generator or controller may be held by a patient, located on a belt or strap worn by the patient, or positioned in a pocket or pouch on the patient.
For example,
In certain variations, systems and methods for providing electrostimulation (ES) of the posterior tibial nerve in individuals with OAB and UI, via transdermal electrical fields, are provided. Various electrodes for delivering the electrical stimulation may be utilized, including but not limited to surface electrodes with and without gel, microneedle electrodes, and electrodes under strong pressure.
Electrical stimulation may be directed to nerves underlying glabrous skin surfaces of the body (e.g., the palms and soles). Higher levels of power (than what would be utilized and tolerated on non-glabrous skin surfaces) may be utilized when delivering electrical stimulation transdermally, through or across a glabrous skin surface, while remaining safe and tolerable. For example, stimulation of the plantar nerves or other nerves of the foot via a surface electrode positioned over the glabrous surface of the plantar surface is highly tolerable and results in sensation similar to that found with needle-based, invasive stimulation of the posterior tibial nerve, but in a non-invasive manner.
Various devices or applicators may be applied to the feet and/or hand or other body portions in order to stimulate nerves underlying glabrous surfaces on an intermittent or continuous basis. Electrical stimulation may be delivered via surface or microneedle electrodes. Various devices may be used in conjunction with implantable or injectable substances in order to improve electrical conductivity to a target nerve. As described supra, the various methods and devices for providing transdermal electrical stimulation over a glabrous skin surface may be used to treat any disorder that is impacted by neuromodulation including, but not limited to: overactive bladder, urinary incontinence, fecal incontinence, chronic pain, depression, migraine, epilepsy, obesity, restless leg, or foot drop. The methods may be applied either intermittently or, if necessary, on a chronic basis. A stimulator electrode surface may be held in contact with a glabrous surface via an adhesive or an ergonomic applicator.
Feedback may be provided to the stimulator to indicate that stimulation is occurring as intended. This may involve an eMG type measurement device, a motion sensor or other sensing device. This feedback loop may be intermittently queried so that the stimulation may be adjusted to ensure that the minimum amount of energy is being used to stimulate to reduce the risk of burns or intolerance.
In certain variations, the device may include an ergonomic wrap or cradle. The sensor feedback may allow for optimization of the positioning of electrodes based on sensor feedback further reducing the risk of burns or intolerance while increasing the efficacy of the neuromodulation.
The methods and devices described herein may be utilized to stimulate various body tissues, including nerve, muscle, skin, vasculature, or any other organ or tissue within the human body. The methods and devices may be used to treat any suitable condition or perform any suitable function via neuromodulation regardless of whether the stimulation source is electromagnetic fields, direct electric current, ultrasound, or RF fields.
In certain variations, methods, systems and/or devices for performing neuromodulation and/or low frequency induction therapy through glabrous skin surfaces, e.g., through palmar or plantar surfaces, are provided. The methods and system may be utilized for treating or preventing various conditions, such as urinary incontinence (UI) and/or overactive bladder (OAB).
For example, a patient suffering from UI or OAB may place the glabrous surface of their foot over an insole, foot rest or foot plate applicator to provide contact between the glabrous surface of their foot and a stimulator electrode of the insole, foot rest, or foot plate applicator. Alternatively, a tethered or wired electrode may be used without an applicator or separate from an applicator. The electrode may be attached to or held in contact with the glabrous surface with an adhesive or as a cutaneous patch.
The stimulator electrode may be positioned over the glabrous surface along the course of a target nerve. The stimulator electrode may be positioned proximal to a stimulation site to ensure that afferent nerve stimulation occurs. A ground electrode may be placed on the body as well. One or more sensing electrodes may be placed along the path of the nerve conduction or motor or sensory impulse. Stimulator and/or sensing electrodes may be connected or coupled to a pulse generator. The pulse generator and electrodes may be incorporated or integrated into an insole, foot plate or other applicator. Alternatively, the electrodes may be connected or coupled to a pulse generator where the electrodes are not incorporated or integrated into the insole, foot plate or applicator.
Electrical stimulation via the stimulator electrodes may begin at a low amplitude and may slowly be ramped up or increased until nerve conduction is detected by the sensor electrode and/or detected by the patient who may signal that motor conduction has occurred, e.g., by pressing a button or other indicator.
Once stimulation is detected, the electrical pulses or electrical stimulation may continue for the directed duration of use or therapy. Electrical stimulation may be delivered intermittently to provide intermittent therapy. For example, electrical stimulation may be delivered for 15-30 minute intervals. In another example, electrical stimulation may be delivered continuously to provide continuous therapy.
A sensor may remain in place for the duration of the electrical stimulation therapy to ensure that stimulation occurs the entire time or substantially the entire time, and to allow for correction or adjustment if the signal deteriorates. A controller for operating a pulse generator or sensor may be powered by a portable power source (e.g., a battery) or a fixed power source (e.g., a traditional wall outlet).
In certain variations, the electrical stimulation may have of a square wave electric signal at a frequency of about 5 Hz to about 60 Hz at the targeted tissue depth. The square wave configuration of the signal may be generated via Fourier transformation or may be a ramped current generated in any manner.
The insole, foot rest, or foot plate may be removed from the body when therapeutic stimulation is not being delivered. The insole, foot rest, or foot plate may be reapplied along with a sensor patch (which may be disposable) as indicated. For example, the electrical stimulation therapy my be administered on a daily basis, where one or more of the above steps are repeated.
Electrical stimulation may be delivered according to any of the variations described herein, in a manner such that the stimulation provides motor, sensory and/or subthreshhold stimulation. Any of the various energy based stimulation systems described herein may be utilized to provide therapy in various settings, e.g., in home use or to provide ambulatory type therapies.
In certain variations, as described supra, a device, applicator, pad, bracelet, ring or garment e.g., a gel pad or gel pad electrode, may include one or more electrodes, or have one or more electrodes coupled thereto. The device applicator, pad, bracelet, ring, garment and/or electrodes may deliver or provide energy, e.g., electrical, magnetic or electromagnetic, stimulation through or across a glabrous skin surface, e.g., a plantar or palmar skin surface. Pain receptors may be located deep under glabrous skin surfaces (e.g., located on the foot or on the palm) allowing the patient to have a higher tolerance to pain resulting from stimulation of these surfaces, than they might have at another area of their body. Applying stimulation over these surfaces allows for stimulation to be delivered at a higher frequency or power or at a greater strength or for a longer duration because of the higher tolerance to pain that a patient exhibits at these surfaces or areas.
In addition, the device, applicator, bracelet, ring, pad and/or electrode may include, carry or be coated with one or more drugs, agents or therapeutic substances. For example, the applicator or pad 520 or electrode 522, as shown in
In certain variations, stimulation e.g., energy stimulation such as electrical, magnetic or electromagnetic stimulation, may be provided through a glabrous skin surface at parameters that overcome or avoid habituation of the targeted tissue to the stimulation. For example, stimulation may be provided, e.g., continuously, for 10 minutes, where the stimulation is automatically paused or stopped for about a 30 second interval every 10 minutes. This intermittent pausing or interruption of stimulation may prevent or overcome habituation of the target tissue, e.g., nerve, muscle or other tissue, to the delivered stimulation.
In one variation, the treatment parameters for stimulating a target nerve or other tissue through a glabrous skin surface, e.g., a plantar or palmar skin surface, may include the following: Apply electrical or magnetic stimulation for a duration of 30 minutes, pausing for about a 30 second interval after 10 minutes of stimulation and then again after another 10 minutes of stimulation. In other variations, the treatment parameters may vary, e.g., stimulation may be applied for an interval ranging from 0 to 100 minutes or greater or as necessary to treat symptoms. In certain variations, for example, stimulation may be applied or provided for 20, 40, 50 or 60 minutes. In certain variations, for example, the pause in stimulation may be for an interval ranging from 0 to 100 seconds, or 5 to 60 seconds or 20 to 40 seconds.
In other variations, stimulation at varying frequency may be applied or provided for varying times with total stimulation of 20, 40, 50 or 60 minutes. In certain variations, for example, the change in stimulation frequency may be for an interval ranging from 0 to 100 seconds, or 5 to 60 seconds or 20 to 40 seconds.
In another variation, a sensor (e.g., EMG sensor) may be utilized, which may detect signs of habituation, e.g., decreased stimulation or excitation of a target nerve, muscle or other tissue (or lack of any resulting stimulation) or a reduced response to stimulation. Through a feedback loop, detection of habituation or stimulation levels by the sensor, may automatically cause the stimulator to pause or stop delivery of stimulation to the target nerve or tissue for a predetermined interval or duration or until habituation is no longer detected. In any of the variations described herein, the stimulation may be paused automatically, manually or via other controls or via an electrical pulse generator or controller in a predetermined or preset manner and/or based on feedback provided by a sensor.
In certain variations, electrical energy or stimulation may be applied to glabrous skin surfaces through electrodes, e.g., conductive microneedle patches, needles, or other electrodes, or by utilizing any of the devices or applicators described herein. Various electrodes, devices and/or applicators for providing stimulation may be applied to plantar or palmar surfaces of a patient. For example, if applied to plantar surfaces, electrodes may be applied to one foot or both feet. In cases where the stimulation is applied to one foot, the other foot may be used as ground, i.e., stimulation may be delivered into one foot and out the other foot. Various conditions may be treated using the devices, applicators and/or electrodes and/or the various methods, therapies and treatment parameters described herein, e.g., urinary incontinence, overactive bladder, epilepsy, migraines, and depression, etc.
In certain variations, treatment of a patient using electrical stimulation may be monitored and/or adjusted based on the detection of feedback signals using primarily F-wave detection or based on the detection of F-waves or M-waves, e.g., in a foot or hand. For example, detection of an F-wave or F-wave feedback (e.g., via a sensor or electrode sensor) may be used in monitoring and/or to adjust the applied stimulation therapy. Optionally, stimulation may be ramped up until an F-wave is detected and then adjusted as necessary. F-wave detection measures nerve conduction velocity. F-wave detection may be used as opposed to H-reflex detection, which measures reflectory reaction from muscles. In other variations, optionally both F-wave and H-reflex detection may be used or solely H-reflex detection may be used. Stimulating to a sub-motor stimulation may be performed. In certain variations, F-waves, M-waves and/or H-reflex may be detected in the foot, hand or other body part.
Electrical energy may be applied or delivered using any of the devices, applicators or electrodes described herein at specified frequencies and/or parameters or using various duty cycles. In one example, electrical energy may be applied at a frequency of 5 Hz to 20 Hz (e.g., to provide supermaximal stimulation) in a train of 5 pulses followed by a ¼ second break, and then another train of 5 pulses. Up to 10 pulses per treatment cycle may be provided. The train of 10 pulses can be repeated as needed. In another example, a train of 10 pulses followed by a break and then another 10 pulses may applied, e.g., at a frequency of 5 Hz to 20 Hz or at about 20 Hz. The ¼ second break time may be varied as well as the number of pulses applied and the number of cycles. This pulsed treatment may be applied through one or more or several cycles. The pulsed treatment or stimulation may produce F-waves which may be detected. In certain variations, any of the treatment parameters as described herein may be applied where only half of the pulses are stimulatory, e.g., one or more cycles of pulses may be provided at 20 Hz, where only half of the pulses are stimulatory.
In other variations, the treatment parameters may vary, e.g., stimulation frequency may be applied for an interval ranging from 0 to 100 minutes or greater or as necessary to treat symptoms followed by a varying frequency for a time interval. In certain variations, for example, stimulation at certain frequency may be applied or provided for varying times with total stimulation of 20, 40, 50 or 60 minutes. In certain variations, for example, the change in stimulation frequency may be for an interval ranging from 0 to 100 seconds, or 5 to 60 seconds or 20 to 40 seconds.
Certain frequencies, e.g., frequencies applied at greater than 20 Hz, may lead to a painful condition called tetany (i.e., painful, involuntary muscle cramping), which is an indication that the upper level for frequency has been reached in the patient. Tetany may be monitored using a sensor, e.g., an EMG sensor or Electromyography, to provide feedback regarding tetany. Tetany may be distinguished by EMG.
In one variation, a treatment device or applicator for delivering energy, e.g., electrical energy, may include a platform with one or more microneedles, microneedle electrode, or other electrodes positioned upon the platform. Microneedles may vary in length to penetrate different layers of the skin, e.g., to penetrate a callus on the skin. A patient may place their foot upon the platform to receive treatment. Optionally, a separate strap having a feedback/monitoring electrode or electrode sensor may be wrapped over the foot to place or position a feedback/monitoring electrode on the patient. The strap and feedback electrode or electrode sensor may be coupled to the platform or be separate. In another variation, the stimulatory electrode may be positioned in a shoe, sock or glove of a patient. In certain variations, a TENS electrode or stimulator may be utilized.
Detection of stimulation, excitation, or conduction using a sensor may provide information to optimize or adjust the applied therapy. If no or reduced stimulation, excitation, electrical activity, response, F-waves, and/or EMG is detected during treatment, this may be indicative of a mis-positioned foot, palm or other body part relative to the one or more electrodes or the device or applicator.
Various conditions may be treated using the devices, applicators and/or electrodes and/or the various methods, therapies and treatment parameters for providing stimulation as described above, e.g., urinary incontinence, overactive bladder, epilepsy, migraines, and depression, etc.
In certain variations, electrical, electromagnetic, magnetic energy and/or other energy may be utilized to provide stimulation according to any of the systems and/or methods described above.
The present invention may use electrical stimulation of one or more of the Ulnar nerve, the Median nerve, the tibial nerve, or other nerves, either alone or simultaneously, separately or with other types of therapy to treat various neurological conditions, including neuropathies, migraine, pain, headaches, depression, epilepsy, post-traumatic stress disorder, overactive bladder, incontinence, subcutaneous fat, effects of stroke, memory enhancement, etc. In some embodiments, treatment can be combined with certain drugs to increase treatment effectiveness.
Devices used to stimulate one or more of the Ulnar, Median or Tibial nerves can be a completely external device, a fully implantable device or a device that comprises both external and internal components. The devices typically comprise a pulse generating component that provides an electrical wave form as output. The electrical pulse comprises certain typical parameters, such as pulse width, frequency, number of pulses and amplitude among other parameters. The choice of parameters may depend on the nerve, the location of the electrodes, the desired effect sought and the amount of time the therapy treatment occurs.
One method of using the present invention is to stimulate one or both of the Ulnar and Median nerves to treat migraine. The Ulnar and Median nerves innervate portions of the arm and hands.
For example, as is show in
A different way of visualizing the cortical homunculus in
Alternatively, all or some of the functions of the generator/controller may be incorporated into a mobile computing device, such as a smart phone, tablet, mobile phone etc. The mobile computing device may include an “app” or application installed on the mobile device, which controls the functions of the electrodes, display and/or mobile device. The electrodes may be incorporated into the casing of the generator/controller, for example in a ball or handheld format. The electrodes may be incorporated into a casing which is physically attached to a mobile computing device. The casing may be powered by the mobile computing device or may be powered separately. Electrodes may be dry and built into the case for stimulation and EMG so that the user is in contact with the electrodes when the case is gripped. Electrodes may alternatively be wet and replaceable with attachments, such as snaps, which attach to, or connect to, the case. The app may use an EMG incorporated into the case, or may use accelerometers in the mobile device itself to measure intensity of stimulation. Since the gripping of the case may cause the fingers of the user to lay on the touchscreen display, the touchscreen may sense contraction of the fingers/hand and give an indication of stimulation presence/intensity. The touchscreen may sense the manner in which the fingers move (distance, direction, number of fingers etc.) or with 3D touch, the pressure on the screen may be measured.
The generator/controller and/or electrodes may also be incorporated into a glove.
The type of energy delivered to the nerve or nerves during stimulation by the device may depend on the pulse parameters. For example, in one therapy type, the pulse parameters have a frequency of somewhere between about 1-20 hz, a pulse width between about 0-500 microseconds and an amplitude of between about 0 and 90 mA. In another therapy, high frequency stimulation can be used, namely frequencies from about 20 hz up to about 10,000 hz. Such high frequencies can be effective using either external electrodes, or fully implantable electrodes, or both. The electrodes are placed in a location to stimulate the Ulnar nerve, the Median nerve or both nerves.
The energy pulse amplitude may also vary during the treatment. This modulation can increase both comfort and efficacy. For example, the pulse energy applied to the electrodes may be at 100% amplitude for X pulses (for example, 5 pulses) and then at a lower amplitude (for example, 50% amplitude) for Y pulses (for example, 5 pulses). The pulse amplitude and length may vary, for example, there may be two amplitude levels, or 3 or more. The different amplitude durations may be the same, or they may be different. The amplitude variations may be preset, or may be reset during the use of the device.
The parameters of the amplitude function include amplitude as a percent of maximum, amplitude time length, amplitude number of pulses, number of different amplitudes, order of different amplitudes, etc. These parameters may be communicated to a remote server for analysis to optimize treatment of individual users in the future. The data may also be aggregated (preferably anonymously) to develop treatment algorithms for different patient populations. Effectiveness input from the individual users may also be incorporated into the algorithms, for example pain level over time, comfort level, medication taken etc.
Similarly, pulse frequency parameters can also be modulated. Similarly, pulse width parameters an also be modulated.
In some embodiments of the device, an amplitude (and/or frequency) modulator adapter may be added to a commercially available TENS (Transcutaneous Electrical Nerve Stimulation) device. This adapter would modulate the amplitude and/or frequency of the signal produced by the TENS device so that the signal delivered to the user via the electrodes can be optimized for comfort and efficacy.
Some embodiments allow the user to use the device inconspicuously, for example at work, or in a meeting. The electrodes may be part of a phone case so that the user's hand is in contact with the electrodes while holding the phone. The electrodes may be part of a beverage mug, computer mouse, steering wheel, etc.
Feasibility Study.
A feasibility study was conducted stimulating the Ulnar and Median nerve of the hand, and the Tibial nerve of the foot. The following paragraphs are the steps taken during the study. One should note that it is recognized that these steps are not exclusive and more or fewer steps may be taken. Additional, while the feasibility study occurred under a controlled setting, such limitations are not required for the invention in practice. Therefore, these steps, the parameters, etc., are to show the one implementation of the invention, but do not restrict the invention in any way.
A study was performed to assess the non-invasive peripheral nerve stimulation device to treat triggered migraine attacks. Stimulation was introduced to the palmar surface of each hand (distal branches of median and ulnar nerves) and plantar surface of each foot (distal branches of posterior tibial nerve).
Methods
Volunteer participants who met ICHD-III beta criteria for episodic migraine without aura were given an intravenous infusion of glyceryl trinitrate to trigger an attack. Triggered attacks were then treated with the peripheral nerve stimulation device: 15 minutes at each extremity, in random order. The primary outcome measure was pain relief (i.e. mild or no pain) at two hours.
Results
See
The details of the use of the device to treat a migraine follows. This is an example of using the system, but is not the only way to stimulate the Ulnar, Median or Tibial nerves.
The operator (physician, physician's representative, company representative or patient) connects the stimulator, electrode patches, and EMG electrodes to the generator/controller at their appropriate connectors. The operator then places the electrode patches on the palmar surfaces of the patient at the correct locations per the instructions for use (see
Once the electrodes are connected, the generator/controller is turned on to the lowest setting in pretreatment confirmation mode. The operator increases the setting of the power until the patient experiences maximal stimulation as demonstrated by finger-curl or until the limit of patient tolerability is reached or the maximum power setting is reached. During confirmation mode, the system displays the power setting, captured M- and F-waves, and the battery power status indicator.
Once the power setting has been determined, the power setting is set by the operator via the current control knob, if necessary. The operator actuates the mode button to start the treatment mode. The embedded software of the generator/controller then automatically completes a 15-minute session of treatment at the pre-determined output cycle. In this configuration, the operator does not adjust the settings of the output of the generator/controller during the treatment. The output screen displays the captured electromyographical waves during treatment, a timer, and a battery power status indicator.
After the 15-minute treatment is complete, as determined by the internal software timer, the generator/controller returns to non-treatment mode (no output to the treatment leads and no EMG capture/display). The operator turns the generator/controller off and removes the electrodes from the patient's hand and optionally repeats the treatment protocol on the second hand.
An example of a Set up Procedure:
Check the battery level.
Place the electrode on the palm (
Determine Current Setting for Therapy. Turn on the signal generator/controller. In order to determine stimulation of afferent pathways, intermittent pulses are masked to confirm and record M-waves and F-waves. By default, the current when the signal generator/controller is powered on is 0 mA. Slowly increase the current amplitude (90 mA maximum level in the particular generator/controller used) to the patient's maximum tolerability while observing the patient's hand for a response. Patient response is generally a finger curl, flex or fan, or an extension of the entire hand. The patient may also report a tingling sensation across the palm or heal of the hand. Once the patient's maximum tolerability is reached, reduce current amplitude setting by one level. By observation, confirm M-wave appearance and confirm active neurostimulation of the Median and Ulnar nerves via F-Wave appearance on the feedback monitor. If the incremental adjustment of current amplitude fails to elicit finger curl, flex or fan, inactivate the current (0 mA) and reposition the electrode slightly. Resume therapy using the preceding instructions. If repositioning the electrode and repeating the current step-up procedure fails to elicit patient response, discard the electrode. Open another electrode, and repeat the procedure on the other hand.
Conduct therapy. After the current level is established, the operator will initiate 15-minutes of ongong electric stimulation. Begin therapy at the current amplitude determined above. The current can be adjusted to increase or decrease the current level at any time during therapy. In order to determine stimulation of afferent pathways, intermittent pulses are masked to confirm and record M-waves and F-waves. When the therapy time has elapsed, inactivate the current (0 mA).
Repeat Treatment Session. Repeat the steps above on the opposite hand then complete treatment session.
Complete Treatment Session. Turn off the signal generator/controller. Remove the cable leads from the electrodes. The treatment session is now complete. Remove electrodes from the patient. The skin may be cleaned with wipes to remove gel residue on the skin from the electrodes.
In the study conducted, the stimulation system that comprised an electrode that connected by an electrical cable to a battery-powered pulse source (electrical signal generator/controller). The signal generator/controller produced an adjustable electrical pulse that traveled to the brachial nerve plexus via the Median and Ulnar nerves when applied to the palmar surface of the hand.
Study device details:
The electrode was placed onto the hand and used to stimulate the Median and Ulnar nerves on the hands. The system consists of a 9 volt battery powered electrical signal generator/controller with electromyographical capture, display, and patient contact leads and electrodes.
The generator/controller housing is a plastic enclosure with an on/off switch, adjustable current control, mode switch button, and LCD screen facing the operator. The generator/controller has connections for a stimulation electrode (output), return electrode (input) and two electromyographic (EMG) capture electrodes (input). The generator/controller housing has a removable door for replacement of the 9 volt battery. The housing encloses the printed circuit board assembly and prevents contact by the user with the internal electronics.
Physical Principles to External Generate Output:
The generator/controller is a 9-volt battery powered, microprocessor controlled, electrical-neural stimulator with embedded software that delivers pulses of DC energy to the patient via the stimulation and return electrodes. The device simultaneously reads electromyographical signals from the patient via the EMG electrodes and displays them on a display screen. The goal is to stimulate the Median and Ulnar nerves of the treated hand.
Electrical Current Settings
Stimulation Signal Program
See
During initial testing of the system, the 100%/50% duty cycle enabled neurostimulation of the peripheral nerves without inducing tetany (muscle pain). The maximum length of maximal neurostimulation without tetany using a 20 Hz pulse train was found to be 0.5 seconds (10 pulses at 20 Hz). Trains of pulses at this frequency were also found to be tolerable as long as there was at least 0.25 seconds of non-maximal, or resting, stimulation between trains.
The stimulation of the peripheral nerves with the system is intended to be tolerable for the patient while generating neural stimulation. In this study, the duty cycle, frequency, shape, length of each pulse, and length of the treatment program were fixed, and the amplitude of the square waves was adjustable by the operator.
Principles of Electromyography
See
In a typical F-wave study, a strong electrical stimulus (above maximal muscle stimulation levels) is applied to the skin surface above (proximally to) the distal portion of a nerve so that the impulse travels both distally (towards the muscle fiber) and proximally (back to the motor neurons of the spinal cord; these directions are also known as orthodromic and antidromic, respectively). When the orthodromic stimulus reaches the muscle fiber, it elicits a strong M-response indicative of muscle contraction (M-wave). When the antidromic stimulus reaches the motor neuron cell bodies, a small portion of the motor neurons fire in the other direction and an orthodromic wave travels back down the nerve towards the muscle. This reflected stimulus evokes a small proportion of the muscle fibers causing a small, second compound muscle action potential (CMAP) called the F-wave.
The generator/controller may capture and display both the M-wave and F-wave on the display screen during confirmation and treatment modes. Being able to view the M-waves and F-waves during treatments allows the operator to confirm maximal stimulation of the nerve in question (Ulnar, Median or Tibial) and its branches in the feet; and the median and Ulnar nerves in the hand.
Example of Data Processing System
As shown in
Typically, the input/output devices 6510 are coupled to the system through input/output controllers 6509. The volatile RAM 6505 is typically implemented as dynamic RAM (DRAM) which requires power continuously in order to refresh or maintain the data in the memory. The non-volatile memory 6506 is typically a magnetic hard drive, a magnetic optical drive, an optical drive, or a DVD RAM or other type of memory system which maintains data even after power is removed from the system. Typically, the non-volatile memory will also be a random access memory, although this is not required.
While
Some portions of the preceding detailed descriptions have been presented in terms of algorithms and symbolic representations of operations on data bits within a computer memory. These algorithmic descriptions and representations are the ways used by those skilled in the data processing arts to most effectively convey the substance of their work to others skilled in the art. An algorithm is here, and generally, conceived to be a self-consistent sequence of operations leading to a desired result. The operations are those requiring physical manipulations of physical quantities.
It should be borne in mind, however, that all of these and similar terms are to be associated with the appropriate physical quantities and are merely convenient labels applied to these quantities. Unless specifically stated otherwise as apparent from the above discussion, it is appreciated that throughout the description, discussions utilizing terms such as those set forth in the claims below, refer to the action and processes of a computer system, or similar electronic computing device, that manipulates and transforms data represented as physical (electronic) quantities within the computer system's registers and memories into other data similarly represented as physical quantities within the computer system memories or registers or other such information storage, transmission or display devices.
The techniques shown in the FIGS. can be implemented using code and data stored and executed on one or more electronic devices. Such electronic devices store and communicate (internally and/or with other electronic devices over a network) code and data using computer-readable media, such as non-transitory computer-readable storage media (e.g., magnetic disks; optical disks; random access memory; read only memory; flash memory devices; phase-change memory) and transitory computer-readable transmission media (e.g., electrical, optical, acoustical or other form of propagated signals—such as carrier waves, infrared signals, digital signals).
The processes or methods depicted in the FIGS. herein may be performed by processing logic that comprises hardware (e.g. circuitry, dedicated logic, etc.), firmware, software (e.g., embodied on a non-transitory computer readable medium), or a combination of both. Although the processes or methods are described above in terms of some sequential operations, it should be appreciated that some of the operations described may be performed in a different order. Moreover, some operations may be performed in parallel rather than sequentially.
Each of the individual variations described and illustrated herein has discrete components and features which may be readily separated from or combined with the features of any of the other variations. Modifications may be made to adapt a particular situation, material, composition of matter, process, process act(s) or step(s) to the objective(s), spirit or scope of the present invention.
Methods recited herein may be carried out in any order of the recited events which is logically possible, as well as the recited order of events. Furthermore, where a range of values is provided, every intervening value between the upper and lower limit of that range and any other stated or intervening value in that stated range is encompassed within the invention. Also, any optional feature of the inventive variations described may be set forth and claimed independently, or in combination with any one or more of the features described herein.
All existing subject matter mentioned herein (e.g., publications, patents, patent applications and hardware) is incorporated by reference herein in its entirety except insofar as the subject matter may conflict with that of the present invention (in which case what is present herein shall prevail). The referenced items are provided solely for their disclosure prior to the filing date of the present application. Nothing herein is to be construed as an admission that the present invention is not entitled to antedate such material by virtue of prior invention.
Reference to a singular item, includes the possibility that there are plural of the same items present. More specifically, as used herein and in the appended claims, the singular forms “a,” “an,” “said” and “the” include plural referents unless the context clearly dictates otherwise. It is further noted that the claims may be drafted to exclude any optional element. As such, this statement is intended to serve as antecedent basis for use of such exclusive terminology as “solely,” “only” and the like in connection with the recitation of claim elements, or use of a “negative” limitation. Unless defined otherwise, all technical and scientific terms used herein have the same meaning as commonly understood by one of ordinary skill in the art to which this invention belongs.
This disclosure is not intended to be limited to the scope of the particular forms set forth, but is intended to cover alternatives, modifications, and equivalents of the variations described herein. Further, the scope of the disclosure fully encompasses other variations that may become obvious to those skilled in the art in view of this disclosure. The scope of the present invention is limited only by the appended claims.
This application is a continuation of U.S. patent application Ser. No. 15/474,875 filed Mar. 30, 2017, which is a continuation-in-part of U.S. patent application Ser. No. 15/084,356 filed Mar. 29, 2016 (now U.S. Pat. No. 9,630,004), which is a continuation of U.S. patent application Ser. No. 13/840,936 filed Mar. 15, 2013 (now U.S. Pat. No. 9,339,641), which is a continuation in part of PCT International Patent Application Number PCT/US2011/052415 filed Sep. 20, 2011, which claims benefit of priority to U.S. Provisional Patent Application No. 61/403,680 filed Sep. 20, 2010. U.S. patent application Ser. No. 15/474,875 is also a continuation-in part of U.S. patent application Ser. No. 12/508,529 filed Jul. 23, 2009 (now abandoned), which is a continuation-in-part of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application No. 60/848,720 filed Oct. 2, 2006. U.S. patent application Ser. No. 15/474,875 also claims priority to U.S. Provisional Patent Application No. 62/350,610 filed Jun. 15, 2016. Each of the above referenced applications is incorporated herein by reference in their entirety. The following applications are also incorporated herein by reference in their entirety for all purposes: PCT Application Serial No. PCT/US10/54167 filed Oct. 26, 2010; PCT Application Serial No. PCT/US10/054353 filed Oct. 27, 2010; U.S. patent application Ser. No. 12/508,529 filed Jul. 23, 2009 (now abandoned), which is a continuation in part of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application Ser. No. 60/848,720 filed Oct. 2, 2006; U.S. patent application Ser. No. 12/695,087 filed Jan. 27, 2010 (now abandoned), which is a continuation of U.S. patent application Ser. No. 11/332,797 filed Jan. 17, 2006 (now abandoned); U.S. patent application Ser. No. 12/509,362 filed Jul. 24, 2009 (now U.S. Pat. No. 9,610,459); Ser. No. 12/469,365 filed May 20, 2009 (now abandoned), which is a continuation of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application Ser. No. 60/848,720 filed Oct. 2, 2006, and Ser. No. 12/469,625 filed May 20, 2009 (now U.S. Pat. No. 8,430,805), which is a continuation of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application Ser. No. 60/848,720 filed Oct. 2, 2006; and Ser. No. 12/509,304 filed Jul. 24, 2009 (now abandoned), which is a continuation of U.S. patent application Ser. No. 12/508,529 filed Jul. 23, 2009 (now abandoned), which is a continuation-in-part of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application Ser. No. 60/848,720 filed Oct. 2, 2006; and Ser. No. 12/509,345 filed Jul. 24, 2009 (now abandoned), which is a continuation of U.S. patent application Ser. No. 12/508,529 filed Jul. 23, 2009 (now abandoned), which is a continuation-in-part of U.S. patent application Ser. No. 11/866,329 filed Oct. 2, 2007 (now abandoned), which claims priority to U.S. Provisional Patent Application Ser. No. 60/848,720 filed Oct. 2, 2006.
Number | Date | Country | |
---|---|---|---|
61403680 | Sep 2010 | US | |
60848720 | Oct 2006 | US | |
62350610 | Jun 2016 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15474875 | Mar 2017 | US |
Child | 16732706 | US | |
Parent | 13840936 | Mar 2013 | US |
Child | 15084356 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15084356 | Mar 2016 | US |
Child | 15474875 | US | |
Parent | PCT/US2011/052415 | Sep 2011 | US |
Child | 13840936 | US | |
Parent | 12508529 | Jul 2009 | US |
Child | 15474875 | US | |
Parent | 11866329 | Oct 2007 | US |
Child | 12508529 | US |