The present invention generally relates to the treatment of connective tissue. More specifically, the present invention relates to methods and apparatus to change the mechanical behavior and properties of connective tissue, such as capsular tissue, tendons, and ligaments.
Tendons, ligaments, and capsular tissues are important for providing stability and motion to the joints within the extremities. Current surgical treatment options generally rely upon open surgical approaches or arthroscopic techniques to gain access to these tissues. Many surgical procedures focus upon reducing the size, thickness, shape, or mechanical integrity of the tissue structure. For example, capsular tissue can be damaged and inflamed as joints age. At present, the only technology available that is commonly accepted to treat capsular tissue injury are conventional techniques such as incision or debridement.
Likewise, strengthening ligament or tendon in terms of enhancing its mechanical strength, (e.g., elastic modulus) is presently an elusive goal for orthopedic surgeons. At best, relatively primitive technologies involving clips or screws are used to alter the tension across these regions. In addition, surgeons currently use various sutures to change the dimensions of these tissue structures.
Beyond conventional surgical technologies, little has been developed in the medical technology to address the needs of surgeons who operate on the joints. Approximately two decades ago, thermal methods, such as radiofrequency, were advocated as a means of altering joint tissues including the capsule. However, the diffuse nature of these thermal interactions can bring about tissue injury, inflammation, and ultimately failure of the procedure. At present, there does not appear to be a company in the current market who still uses radiofrequency to address these issues related to ligament and capsular pathoanatomy.
As can be seen, there is a need in the marketplace for a technology that can lengthen or tighten joint tissue and simultaneously alter the mechanical behavior and properties of connective tissues. In instances where connective tissue must be treated, anesthesia and joint arthroscopy is often required. A minimally invasive technique that would rely upon surgical anatomy and needle-based technologies would be extremely effective in eliminating complicated procedures. Combining these solutions with an imaging technique such a CT, MRI, or ultrasound would be extremely compelling and valuable because procedures that can be performed in the office rather than the OR would be beneficial for reducing health costs and eliminating the need for invasive procedures.
In one aspect of the present invention, a method of treating connective tissue, comprising applying a DC current to the tissue in the absence of heating the tissue.
In another aspect of the present invention, a method of treating connective tissue, comprising reducing hydrogen ions to hydrogen gas in a cathodic region of the tissue; and oxidizing hydroxide ions to oxygen gas in an anodic region of the tissue.
In a further aspect of the present invention, a method of strengthening connective tissue, comprising compressing or stretching the tissue; initiating, with the use of electrodes, electrochemical reactions in the tissue; and changing a water content in the tissue to alter mechanical properties of the tissue.
These and other features, aspects and advantages of the present invention will become better understood with reference to the following drawings, description and claims.
The following detailed description is of the best currently contemplated modes of carrying out exemplary embodiments of the invention. The description is not to be taken in a limiting sense, but is made merely for the purpose of illustrating the general principles of the invention, since the scope of the invention is best defined by the appended claims.
Broadly, the present invention provides a method of treating connective tissue such as capsular tissue, tendons, and ligaments. More particularly, the present invention generally employs a method of electromechanical reshaping (EMR). EMR uses an application of DC current to induce a shape change, while undergoing very little, unintended temperature changes, thus eliminating any tissue damage that may be caused by the diffuse nature of the heat source. Accordingly, no external heating source is applied to the tissue to intentionally cause a rise in tissue temperature.
Though the complete mechanism of EMR is not completely known, it is hypothesized that EMR works through electrochemical principles. At its simplest, EMR behaves similarly to an electrochemical cell. Because cartilage tissue is approximately 75% water, the predominant oxidation-reduction reactions are the reduction of hydrogen ions to hydrogen gas at the cathodic region and the oxidation of hydroxide ions to oxygen gas at the anodic region of the tissue. These hydrolysis reactions change the water content of the tissue, which can significantly alter mechanical properties in connective tissue. Furthermore, pH effects result from changes in the hydroxide and hydrogen ion concentration due to EMR redox effects.
Elongation of Tendon
Methods
More specifically in
Referring back to
A tensile deformation was applied at a strain rate of 0.02 mm/s until the tissue length reached a strain of 5%. As the tissue was held at a constant strain for 15 minutes, the reaction force, displacement, and time was recorded. After reaching stress equilibrium at 5 minutes (after the tissue reached the 5% strain mark), the EMR process was initiated. Two flat platinum plates were placed on opposite sides of the sample and were held together by an insulated clamp to ensure that the entire surface of the electrode was in contact with the tissue. The leads of a DC power supply (E3646A, Agilent Technologies, Inc., Palo Alto, Calif.) were connected to the electrodes. Voltage amplitude and time duration were monitored through computer software (LabVIEW, National Instruments, Austin, Tex.). EMR was accomplished for 3 minutes at 6V. These parameters were chosen because they had previously demonstrated to produce significant mechanical behavior changes in similarly sized facial cartilage tissue in reshaping studies. The total time for the experiment was 15 minutes (including 3 minutes of active EMR). After this time interval, the tissue was returned to zero strain at a rate of 0.02 mm/s and immediately rehydrated in PBS for 15 minutes.
Mechanical testing was done again on the sample to determine the moduli after tensile load and EMR. The ability for the tendon to stretch (changes in length) was determined by finding the differences in displacement of the clamps/grips at the time points of zero loads during the increase to 5% strain and the decrease to zero strain. Changes to the stress-strain curve and differences in stiffness and length were graphed. Currently, mechanical testing using Material Testing Systems (MTS), the gold standard for orthopedic industry cases, has not been accomplished in this study, but will be utilized to further validate this technique.
Results
Noticeable changes in visual appearance during and after EMR were observed for experimental samples. During the duration of EMR, subtle changes to the tissue color and gas bubble formation were observed as noted in EMR of cartilage tissue. As compared to control samples and pre-EMR samples, the tissue became slightly translucent and on the side facing the cathode, the tissue surface was noticeably rougher in texture. After rehydration, the tendon samples retained its translucency.
Mechanical testing during the elongation process of both control and EMR samples demonstrated stress relaxation of the tissue over the duration of load (
When EMR was applied at 5 minutes, the reaction force increased due to the presence of gas bubbles pushing against the load cell, as shown by six EMR plots in the middle graph (
Post-EMR samples (
In addition, EMR does not affect the length of the tendon after elongation (
Compression of Tendon
Methods
More specifically, in
The sample was placed on the mechanical testing apparatus fitted with compressive platens. The tissue was kept rehydrated by periodic drips of PBS during the duration of the applied load. A compressive deformation was applied at a rate of 0.02 mm/s until the tissue height reached a strain of 25%. As the tissue was held at a constant strain for 15 minutes, the reaction force, displacement, and time was continuously recorded.
After reaching stress equilibrium at 5 minutes after the tissue reached the 5% strain mark, EMR was applied for 3 minutes at 6 V. Platinum plates covered the entire top and bottom surface of the tissue during EMR process. The tissue remained under compression for another 7 minutes to allow further stress relaxation to complete. After the 15 minute procedure, the tissue was brought back to zero strain at a rate of 0.02 rnm/s and immediately rehydrated in PBS for 15 minutes outside of the platens.
Digital images (CanonRebel XSI DSLR) were acquired to observe changes in tissue shape after rehydration. Specifically, changes in the tissue height and diameter were assessed. Mechanical testing was done on the samples after rehydration to observe changes in the Young's moduli after the tensile load was applied. Changes to the stress-strain curve and differences in stiffness and length were graphed.
Results
Similar to elongation experiments, visible changes to the tendon were observed as a result of EMR applied under compression. Bubble formation and tissue translucency were noticed during EMR. The tissue retained its translucency after rehydration. Also, the surface exposed to the cathode electrode was noticeably rougher in texture.
There were no significant differences in compressive modulus as a result of EMR (
Post-EMR samples showed a significant shape deformation in both the diameter and the thickness of the cylindrical shaped specimen (
In sum, the present invention may provide the following advantages:
It should be understood, of course, that the foregoing relates to exemplary embodiments of the invention and that modifications may be made without departing from the spirit and scope of the invention as set forth in the following claims.
This application claims the benefit of and priority to U.S. provisional application No. 62/732,550 filed Sep. 17, 2018 which is incorporated herein by reference in its entirety.
This invention was made with government support under contract DR090349 awarded by the Department of Defense and under contract 1R21DE019026-01A2 awarded by the National Institutes of Health. The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
62732550 | Sep 2018 | US |