The present disclosure is generally directed to neural stimulation and, more specifically, to an improved method of providing safety checks to prevent neural damage.
In 1755, LeRoy passed the discharge of a Leyden jar through the orbit of a man who was blind from cataract and the patient saw “flames passing rapidly downwards.” Ever since, there has been a fascination with electrically elicited visual perception. The general concept of electrical stimulation of retinal cells to produce these flashes of light or phosphenes has been known for quite some time. Based on these general principles, some early attempts at devising prosthesis for aiding the visually impaired have included attaching electrodes to the head or eyelids of patients. While some of these early attempts met with some limited success, these early prosthetic devices were large, bulky and could not produce adequate simulated vision to truly aid the visually impaired.
In the early 1930's, Foerster investigated the effect of electrically stimulating the exposed occipital pole of one cerebral hemisphere. He found that, when a point at the extreme occipital pole was stimulated, the patient perceived a small spot of light directly in front and motionless (a phosphene). Subsequently, Brindley and Lewin (1968) thoroughly studied electrical stimulation of the human occipital (visual) cortex. By varying the stimulation parameters, these investigators described in detail the location of the phosphenes produced relative to the specific region of the occipital cortex stimulated. These experiments demonstrated: (1) the consistent shape and position of phosphenes; (2) that increased stimulation pulse duration made phosphenes brighter; and (3) that there was no detectable interaction between neighboring electrodes which were as close as 2.4 mm apart.
As intraocular surgical techniques have advanced, it has become possible to apply stimulation on small groups and even on individual retinal cells to generate focused phosphenes through devices implanted within the eye itself. This has sparked renewed interest in developing methods and apparatuses to aid the visually impaired. Specifically, great effort has been expended in the area of intraocular retinal prosthesis devices in an effort to restore vision in cases where blindness is caused by photoreceptor degenerative retinal diseases such as retinitis pigmentosa and age related macular degeneration which affect millions of people worldwide.
Neural tissue can be artificially stimulated and activated by prosthetic devices that pass pulses of electrical current through electrodes on such a device. The passage of current causes changes in electrical potentials across visual neuronal membranes, which can initiate visual neuron action potentials, which are the means of information transfer in the nervous system.
Based on this mechanism, it is possible to input information into the nervous system by coding the information as a sequence of electrical pulses which are relayed to the nervous system via the prosthetic device. In this way, it is possible to provide artificial sensations including vision.
One typical application of neural tissue stimulation is in the rehabilitation of the blind. Some forms of blindness involve selective loss of the light sensitive transducers of the retina. Other retinal neurons remain viable, however, and may be activated in the manner described above by placement of a prosthetic electrode device on the inner (toward the vitreous) retinal surface (epiretial). This placement must be mechanically stable, minimize the distance between the device electrodes and the visual neurons, and avoid undue compression of the visual neurons.
In 1986, Bullara (U.S. Pat. No. 4,573,481) patented an electrode assembly for surgical implantation on a nerve. The matrix was silicone with embedded iridium electrodes. The assembly fit around a nerve to stimulate it.
In 1975, Dawson and Radtke stimulated a cat's retina by direct electrical stimulation of the retinal ganglion cell layer. See Investigative Ophthalmology & Visual Science Reports, vol. 16, no. 3, p. 249 (1977). These experimenters placed nine and then fourteen electrodes upon the inner retinal layer (i.e., primarily the ganglion cell layer) of two cats. Their experiments suggested that electrical stimulation of the retina with 30 to 100 μA current resulted in visual cortical responses. These experiments were carried out with needle-shaped electrodes that penetrated the surface of the retina (see also U.S. Pat. No. 4,628,933 to Michelson).
The Michelson '933 apparatus includes an array of photosensitive devices on its surface that are connected to at least one electrode positioned on the opposite surface of the device to stimulate the retina. These electrodes are disposed to form an array similar to a “bed of nails” having conductors which impinge directly on the retina to stimulate the retinal cells. U.S. Pat. No. 4,837,049 to Byers describes spike electrodes for neural stimulation. Each spike electrode pierces neural tissue for better electrical contact. U.S. Pat. No. 5,215,088 to Norman describes an array of spike electrodes for cortical stimulation. Each spike pierces cortical tissue for better electrical contact.
The art of implanting an intraocular prosthetic device to electrically stimulate the retina was advanced with the introduction of retinal tacks in retinal surgery. De Juan, et al. at Duke University Eye Center inserted retinal tacks into retinas in an effort to reattach retinas that had detached from the underlying choroid, which is the source of blood supply for the outer retina and thus the photoreceptors. See, e.g., E. de Juan, et al., 99 Am. J. Ophthalmol. 272 (1985). These retinal tacks have proved to be biocompatible and remain embedded in the retina, and choroid/sclera, effectively pinning the retina against the choroid and the posterior aspects of the globe. Retinal tacks are one way to attach a retinal array to the retina. U.S. Pat. No. 5,109,844 to de Juan describes a flat electrode array placed against the retina for visual stimulation. U.S. Pat. No. 5,935,155 to Humayun describes a retinal prosthesis for use with the flat retinal array described in de Juan.
In electrically stimulating neural tissue it is usually important to prevent over stimulation and unbalanced stimulation which would cause damage to the neural tissue, the electrode, or both. It is typically critical that neural tissue in not subjected to any direct current or alternating current above a safe threshold. Further, it is usually important to identify defective electrodes as continued use may result in neural and further electrode damage. The present disclosure presents system and stimulator control mechanisms to prevent damage to neural tissue.
According to a first aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; testing the impedance of the at least one electrodes; calculating an average impedance; calculating a difference from said average impedance for each said at least one electrode; and determining bad electrodes based on said difference and a predetermined value.
According to a second aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; determining a safe level for the at least one electrode; receiving input for neural stimulation; calculating a sum of stimulation on the at least one electrode based on said input for neural stimulation; reducing the stimulation on each said at least one electrode by the amount necessary not to exceed said safe level for the at least one electrode; and stimulating said neural tissue according to a reduced stimulation level.
According to a third aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; receiving input for neural stimulation; determining a desired cathodic pulse based on said input; calculating a anodic pulse to balance said cathodic pulse; and stimulating neural tissue according to said cathodic and anodic pulses.
According to a fourth aspect, a neural stimulator is provided, comprising: a neural stimulation controller; at least one electrode suitable to stimulate neural tissue and driven by said neural stimulation controller; an impedance meter for measuring the impedance of the at least one electrode; a calculating device for calculating the average impedance of the at least one electrode; a calculating device for calculating the impedance difference from said average impedance for each said at least one electrode; and a means for determining bad electrodes based on said difference and a predetermined value.
According to a fifth aspect, a neural stimulator is provided, comprising a neural stimulation controller; at least one driver controlled by said neural stimulation controller; at least one electrode suitable to stimulate neural tissue driven by said at least one driver; at least one switch connecting the output of said at least one driver to a common line; and a comparator comparing voltage on said common line to a predetermined reference voltage.
According to a sixth aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator and at least one electrode suitable to stimulate neural tissue; testing the capacitance of the at least one electrode; recording changes in capacitance over time; and determining bad electrodes based on said change in capacitance.
According to a seventh aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; determining a maximum charge per phase for the at least one electrode; receiving input for neural stimulation; calculating a discrete integral of the wave form of the input; reducing the stimulation on each said at least one electrode by the amount necessary not to exceed said safe level for the at least one electrode; and stimulating said neural tissue according to a reduced stimulation level.
According to an eighth aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; stimulating neural tissue in accordance with external input; testing total power dissipation of said neural stimulator; comparing said total power dissipation to a predetermined maximum power dissipation; and reducing stimulation if the total power dissipation exceeds said predetermined maximum power dissipation.
According to a ninth aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one stimulating electrode and one common electrode, suitable to stimulate neural tissue; stimulating neural tissue in accordance with external input using balanced biphasic pulses in opposite phases at the same time across said at least one stimulating electrode; and disallowing stimulation waveforms on said at least one electrode that overlap non-contiguously during opposite phases with other said electrodes in the array in a bipolar or multipolar configuration.
According to a tenth aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; measuring electrode impedance using driver output voltage; stimulating neural tissue according to external input; measuring an after stimulation driver output voltage; and calculating leakage current using electrode impedance and after stimulation driver output voltage.
According to an eleventh aspect, an apparatus for controlling the stimulation of neural tissue is provided, comprising: a neural stimulator; at least one electrode suitable to stimulate neural tissue; a meter device for testing the impedance of the at least one electrode; a calculating device for calculating an average impedance; a calculating device for calculating a difference from said average impedance for each electrode.
According to a twelfth aspect, an apparatus for controlling the stimulation of neural tissue is provided, comprising: a neural stimulator; at least one electrode suitable to stimulate neural tissue; a meter device for testing the capacitance of the at least one electrode; a data storage device for recording changes in capacitance over time.
According to a thirteenth aspect, a neural stimulator is provided, comprising: a neural stimulation controller capable of providing a biphasic current pulse to each the electrodes; at least one electrode suitable to stimulate neural tissue driven by said neural stimulation controller; a meter device for taking at least one measurement of an IR drop of the at least one electrode when subjected to the biphasic current pulse; a calculating device for calculating a resistance for each of the at least one electrode; a calculating device for calculating a threshold leakage voltage value for each of the at least one electrode; and a data storage device for logging failed electrodes based on at least the IR drop measurement and the threshold leakage voltage values.
According to a fourteenth aspect, a method of controlling the stimulation of neural tissue is disclosed, comprising: providing a neural stimulator including at least one electrode suitable to stimulate neural tissue; testing the impedance of the at least one electrode; calculating an average impedance; calculating a difference from said average impedance for each said at least one electrode; determining bad electrodes based on said difference and a predetermined value; stimulating an electrode of the at least one electrode with at least one test current pulse; taking at least one measurement of an IR drop voltage; calculating a pre-short maximum voltage from at least one IR drop voltage measurement; and comparing the pre-short maximum voltage to a threshold leakage voltage value.
According to a fifteenth aspect, a device is provided for implementing any one of the methods and/or method steps disclosed in the present specification, drawings or claims, is disclosed.
Further embodiments are shown in the specification, drawings and claims of the present application.
The following description includes the best mode presently contemplated for carrying out the present disclosure. This description is not to be taken in a limiting sense, but is made merely for the purpose of describing the general principles of the invention. The scope of the invention should be determined with reference to the claims.
The electronics package 14 is electrically coupled to a secondary inductive coil 16. Preferably the secondary inductive coil 16 is made from wound wire. Alternatively, the secondary inductive coil 16 may be made from a flexible circuit polymer sandwich with wire traces deposited between layers of flexible circuit polymer. The electronics package 14 and secondary inductive coil 16 are held together by a molded body 18. The molded body 18 may also include suture tabs 20. The molded body 18 narrows to form a strap 22 which surrounds the sclera and holds the molded body 18, the secondary inductive coil 16, and the electronics package 14 in place. The molded body 18, suture tabs 20 and strap 22 are preferably an integrated unit made of silicone elastomer. Silicone elastomer can be formed in a pre-curved shape to match the curvature of a typical sclera. However, silicone remains flexible enough to accommodate implantation and to adapt to variations in the curvature of an individual sclera. The secondary inductive coil 16 and molded body 18 are preferably oval shaped. A strap 22 can better support an oval shaped coil.
A preferred prosthesis includes an external portion (not shown in
The electronics package 14 includes an integrated circuit for controlling stimulation. The integrated circuit includes an excessive direct current flow (EDCF) detection circuit as shown in
Electrode bubbling can occur when the voltage across the double layer of the electrode-tissue interface exceeds a threshold voltage window over certain time duration during the stimulation cycles. This threshold voltage window is found to be around ±1.5V for a flexible circuit electrode array 10. Therefore, in an implant environment, the electrode voltage across the double layer shall never exceed this window.
Based on the simplified electrical model of the electrode shown in
where Vc(0) is the initial voltage Vsh+ caused by the residue charge left on the capacitor at the end of shorting, τp=RpCp is the time constant of the leaky double layer, Tx is the pulse duration (see
When a DC leakage is present, the charge may increase even when the system is not stimulating. The situation is exacerbated when the DC leakage is in the same polarity as the first phase stimulation current. If the leakage current is small compared to the stimulation current, the voltage build up on the electrode occurs when the stimulating pulse happens just before the shorting starts. In this case, the maximum electrode voltage is at the start of the second current pulse, as illustrated in
V
Cmax
=I
L
R
p
+I
S
R
P
e
−Tx/τ
(1−e−Tx/τ
where IL is the leakage current, IS is the stimulation current, TP7 is P7 profile duration (see
A possible electrode voltage map is revealed in
The measured parameters Cp and Rp of the electrodes are preferably 0.25-0.3 μF and 70-80K; but 0.25 μF and 80K to handle the exacerbated condition, which may result in τp=20 ms. The shorting time Ts can be set at 1.4 ms for the current embodiment. The stimulation pulse duration Tx may also vary, but the smallest duration should be limited by the implant compliance limit in order to make the electrode less tolerant to leakage. From a 15KΩ electrode impedance assumption, this will allow a 0.275 ms duration at a current of 400 ΞA (0.35 mC/cm2 maximum charge density). The maximum allowed imbalance of 5% should also be put in the equation. Therefore, in an exacerbated condition, we have:
τp=20 ms, Tx=0.275 ms, IS=400 uA, TP7=1.4 ms, T=8.3 ms
from which we get:
V
Cmax=0.73VC(0)+21.7IL(mA)+0.43(V) (3)
This voltage should be kept below 1.5V. For a margin of safety, the maximum allowed electrode voltage across the double layer could be set to VCmax=1.0V.
The steady state value of the residue voltage on the capacitor depends on the shorting duration. It can be estimated as:
V
C(0)=VSH(0)e−T
where τS is the shorting time constant, and VSH(0) is the voltage at the end of the cycle, just before shorting is turned on. In our exacerbated situation, we have:
V
SH(0)=VCmaxe−Tx/τ
where τS=RIRCp=15K×0.25 uF=3.75 ms.
Therefore, with TSH=1.4 ms, we have from equations (4) and (5):
0.73VC(0)−22.5IL+0.06=0 (6)
From equations (3) and (6), we can calculate the maximum allowed as
I
Lmax=12 μA (7)
This maximum allowable leakage current for an individual electrode to ensure that it does not causing electrode bubbling is lower than the EDCF detection circuit (described in reference to
The electrode potentials takes some time to stabilize because the time needed to reach the balance between the charge released by shorting and the build up by charge imbalance or leakage. The above analysis on electrode voltages only addresses the steady state condition. Because a DC leakage is typically considered persistent current flow with or without the presence of stimulation, one method used is to measure the electrode voltage caused by the leakage current in quiescent condition in which all stimulation is turned off. This way the disturbance on the electrode voltage caused by the stimulation currents is avoided. However, the electrode voltage caused by the leakage current IL (if existent) is still a function of the shorting duration. When we use the same shorting strategy as the normal condition with EDCF turned off, i.e., TSH=TP7=1.7 ms, the steady state voltages on the electrode double layer are estimated as:
V
SH−
=V
SH+
e
−(T−T
)/τ
=I
L
R
p(1−e−T−T
V
SH+
=V
SH−
e
−T
/τ
≈0.635VSH− (9)
Which give us:
V
SH−=0.50V; VSH+=0.32V
Where VSH− and VSH+ are the voltages immediately before Vsh− and after Vsh+ the shorting pulse. Either VSH− or VSH+ can be used as the individual electrode leakage threshold; however, VSH− is preferred because of its higher value for better accuracy of the measurement. The voltage driver output VDO is the sum of the I-R drop caused by the real part of the electrode-tissue impedance RS 406 and the electrode voltage on the double layer capacitor VC discussed supra. The I-R drop can be calculated as ILRir, where Rir is the electrode impedance.
From (7), the I-R drop caused by the allowed maximum DC leakage current is ILmaxR≈0.12-0.56V constant for electrode impedances ranging 10-40KΩ. However, when the electrode is lifted from the retina, the impedance could be lowered to be as low as 3KΩ. For a relatively accurate estimation of the leakage current using the VDO measurement, the I-R drop should be subtracted from the VDO result using the pre-measured electrode impedance values. For simplicity, it may be preferable to ignore the I-R drop effects in this leakage detection protocol and directly use the VDO data as the electrode voltage VC. Ignoring the I-R drop will yield an approximately 20-50% inaccuracy of the leakage value estimation that will result in a more conservative monitoring. This should not compromise the safety and should simplify the measurement.
In the suggested electrode impedance measurement protocol, an electrode with impedance 65KΩ or higher is labeled as an open electrode. Stimulation to an open electrode is turned off. However, an electrode with impedance higher than 65KΩ could still bubble because of DC leakage. Assuming that the maximum allowed leakage is still 12 μA and the compliance limit for the leakage current is 7.0V, calculating with the same method used above it is shown that an electrode with impedance value as high as 450KΩ could reach electrode bubbling status. However, the measured voltage would be limited to about 4V on the anodic side, which should limit the detectable threshold leakage to electrodes with impedance up to 230KΩ. For example, if the measured electrode impedance is 200KΩ, then a measured VDO of 3.4V or more can be considered a bubbling status. In either case, the electrode impedance measurement should be able to discriminate between a high impedance electrode and an open electrode.
The following two-step method is suggested to label an electrode as a high impedance (HI) electrode: (a) After the regular impedance measurement routine, all “open” electrodes (if any) are measured again using 8.1 μA/1 ms current pulses with the VDO sampling point set at 0.9 ms after the pulse start. (b) An electrode with measured impedance 500KΩ or less (VDO<±4V) shall be tagged as HI electrode, while higher impedance electrodes are tagged as open.
For HI impedance electrodes, the build up voltage on the electrode capacitor is solely from DC leakage because they are not stimulated. Therefore, the maximum allowed electrode voltage is VSH=1V, plus the I-R component when using VDO measurement. Therefore, the threshold for an HI electrode should be VDO (measured at VSH−):
VDO|
V
<=I
Lmax
R+1≈0.012R(KΩ)+1 (10)
Considering the limited ADC range in the anodic direction, it may be preferable to check all HI electrodes with impedance 200KΩ or higher against the threshold voltage of 200KΩ, which is 3.4V. For open electrodes, this checking procedure may be omitted.
Referring to
The circuit shown in
Another way to limit current density is to provide a compromise between the monopolar mode and bipolar mode of stimulation. The method includes setting up stimulation wave forms as in bipolar stimulation mode without disconnecting the common electrode (ground). A portion of the current will flow between the electrodes, and a portion of the current will flow between the electrodes and ground. The ratio of current flow will depend on electrode impedance. This hybrid bipolar mode of stimulation could possibly result in lower thresholds than can be obtained from a true bipolar mode of stimulation, and also has the advantage of greater selectivity than can be obtained from a monopolar mode of stimulation. To set up a safe hybrid bipolar stimulation wave form, the two electrodes concerned should have balanced biphasic currents going in opposite phases at exactly the same times, with the common electrode connected. If the pulse waveforms are balanced by themselves on each electrode but overlap non-contiguously in opposite phases with other electrodes in the array, a resultant unbalanced current could flow through the tissue eventually causing neural tissue damage. A safety check method could be implemented in the external electronics to prevent such unbalanced multipolar waveforms to be sent to the stimulator. The above phenomenon and safety check method is also applicable to multipolar forms of stimulation (i.e. in addition to bipolar).
It is also important to limit the maximum stimulation across all electrodes in an electrode array. While each electrode is individually stimulating at a safe level, there can still be neural damage, or even in some cases pain, if all electrodes are stimulating at or near their individual maximum level. Hence, it is important to track and limit the sum total stimulation from all electrodes, which can be performed by a calculation. For each stimulation cycle, all stimulation values are summed and compared with a predetermined maximum. If the sum exceeds the predetermined maximum, the stimulation is reduced, either proportionally across all electrodes, or by limiting electrodes set to higher stimulation levels. This calculation can be performed by software in the external electronics.
Broken electrode detection can be achieved with a method of monitoring impedance of the electrodes (over time (ΔI), as well as comparing impedances of electrodes to its neighboring electrodes in the media, and incorporation of physiological data as observed from physicians). When a broken electrode is found in the system, the stimulator device is commanded to halt stimulation on that broken electrode—with the history of electrode damage logged and persistent in the stimulator controller. Visually, the electrode health is represented through a color/topographical map of the electrode array on a computer system. Optimally, a movie-like data playback of the impedances can be provided. Electrode damage can also consist of shorting between electrodes (as opposed to ‘broken’) which will also be detected through impedance monitoring.
Charge imbalance can be reduced by implementing a 1.4 ms (adjustable, as mentioned previously) shorting pulse prior to any stimulation on every stimulation frame of the stimulator. Additionally, the stimulator controller can adjust and check the anodic and cathodic current level, using amplitude parameter tables determined from the manufacturing tests of the stimulator, to achieve the best balance. The stimulator controller also checks that there exist no overlapping anodic/cathodic profiles that could cause a charge imbalance at the tissue.
Another method of reducing charge imbalance on an electrode is by having the stimulator controller auto-balancing the pulse. Two ways of achieving this are by appending a square pulse after the pulse to compensate for the imbalance or by appending an inverted pulse to compensate. A check against DC flow in tissue from the electrodes is achieved through a combination of the ASIC test, the shorting function, and an initial individual electrode check implemented on startup (or periodically) by the stimulator controller. The issue of implant overheating is handled in multiple ways. Power is controlled on the retinal prosthesis through a feedback loop fed by implant back telemetry. In one embodiment, this is achieved by inferring the implant heat through the current in the shunt regulators of the implant device. However, if a thermistor is placed in the implant device, then the heat can be measured and returned to the controller through the back telemetry link.
Additionally, the instantaneous current output on the stimulator device can be limited to a constant value by the controller. This ensures that no amount of instantaneous current is allowed that might expect the stimulator to reset due to lack of power. This instantaneous current limit could also be variable (instead of constant) with appropriate back telemetry and controller design. For system operational modes that do not allow heat control via the shunt current values, the controller performs a check that ensures the implant shunt current level can be set to a safe value immediately prior to the operational mode which doesn't provide shunt current information in the back telemetry. In addition, the stimulator controller verifies the voltage setting with an ADC circuit upon any change of voltage to the RF power circuitry.
Referring to
Once all electrodes have been tested, the system can calculate and store a median and a standard deviation 718. If this test has been previously preformed 720, the impedance data is shifted to the next memory location 722. If not, calculate the standard deviation of good electrodes 724, and shift the standard deviation to the previous memory location 726. If the electrode values are close, a very low standard deviation might trigger too many electrodes marked as bad. Therefore, the standard deviation can be compared with a preset minimum 728 and replaced with the minimum if the minimum is higher 722. If the electrode values are far apart, a very high standard deviation might not mark bad electrodes. Therefore, the standard deviation is compared with a preset maximum 730 and replaced with the maximum if the maximum is not higher 732, 734.
It has been observed that an electrode which fails (for example, reading as high as 65 k Ω initially) may give subsequent impedance values which begin to fall, sometimes returning to a nominal, good value. This is apparently due to fluid leaking in the broken area and creating an alternative (and undesirable) conduction path. Therefore, any one set of readings might not show all electrodes known obviously to be broken, so a history should be maintained. If an electrode ever exceeds the limits, it could be considered permanently failed.
The history of the electrodes should not begin until the implantation of the array has stabilized. Impedance readings are very useful through the implantation process, but values could shift drastically, and false failures can impede the process.
It might appear that many failed electrodes never reach a high value, or reach it and return so quickly it may not be recorded. Therefore, additional checks are suggested.
Referring to
In one embodiment, impedance is used to determine the electrode integrity at the interface of the neural stimulator and tissue, as well as the integrity of the electrode stimulation path through the implantable device. However, data obtained both in vitro and in vivo seem to indicate that these impedance readings are time dependent, and any given temporal snapshot might show failed electrodes as having perfectly nominal values. Furthermore, potentially critical failures such as electrode movement or loss of tissue contact might not be quickly diagnosed using the current methods as the changes in impedance might not be large.
Capacitance measurement is one possible solution to this limitation, because electro-neural interfaces have distinct capacitive characteristics. Referring to
When an electrode begins to lift off, it begins making contact with the fluid, and thus has two parallel impedance paths. A simple impedance measurement should show only a slight drop at the beginning. However, the phase characteristics of the two paths are different, in both total capacitance value and total phase shift. Thus, a waveform reading should show the effect of the loss of contact.
The difference in capacitance between the various electrical paths which occurs in both healthy and failed electrodes can be manifested as a phase difference in the various waves that sums the resultant measured waveform. This difference can be computed for each stimulation phase, based on the area of the curve of the first derivative, subtracting the baseline waveform where no phase shift is present. Healthy electrodes appear to have distinctive measurable characteristics in both anodic and cathodic phases. These characteristics change significantly when an electrode is losing contact at the neural interface (array lift-off) or is degrading, and thus electrodes which show a significant difference in phase shift are likely experiencing the manifestation of a failure. This method could be used to detect array lift-off potentially before it would be detected by the associated impedance values currently in use.
Another embodiment of low level leakage detection mechanism is based on the electrode properties and the electrode model will now be disclosed. An algorithm can be used to achieve low level leakage current measurement for every electrode with a general purpose voltage measurement mechanism. The method according to this embodiment also facilitates external control of the detection; therefore, the control can be easily updated. Based on the simplified electrical model of the electrode shown in
V
C
=V
C(0)e−t/τ
where Vc(0) is the initial voltage caused by the residue charge left on the capacitor at the end of shorting, τp=RpCp is the time constant of the leaky double layer of the electrode-tissue interface.
When there is no current flow through the electrode, the residue charge previously built across the double layer capacitor CP 402 (see
V
C
=V
C(0)e−t/τ
During biphasic stimulation, as shown in
When the stimulating pulse ends just before the shorting starts, the maximum electrode voltage, VCmax, occurs at the start of the second current pulse. Factoring in an exacerbated case imbalance of 5% between the two phases of the stimulation current, the maximum allowed DC leakage current ILmax that will keep the electrode voltage at below a threshold value of VCmax can be calculated as:
where τp is the time constant of the electrode double layer τp=RpCp, τS is the time constant of the electrode shorting path τS=(RS//Rp)Cp, ISTIM is the stimulation current, Tx is the pulse duration of the stimulation current pulses which is assumed to be biphasic with a “1+1+1” profile arrangement (i.e., the pulse duration of cathodic pulse Tx is equal to the pulse duration of the anodic pulse Tz and is also equal to the delay between the cathodic and anodic pulses Ty), TP7 is the profile duration TP7 that includes a duration of TSH for shorting following a duration of common electrode disconnection for the EDCF check in the preferred embodiment, and T is the stimulation cycle.
Equation (13) shows that, for a predefined safe VCmax value, the DC leakage threshold ILmax is inversely affected by the stimulation current ISTIM, the stimulation pulse duration Tx, the time constant of the electrode double layer τp=RpCp, and the time constant of the shorting path τS=(RS//Rp)Cp. Among them, ISTIM, and Tx are application parameters, Rp and Cp are design parameters, while RS is tissue related parameter.
The electrode parameters Cp and Rp of the electrodes can typically be 0.25-0.3 μF and 80-100K; for example, they can be 0.25 μF and 100K to handle less favorable conditions where τp=25 mS. TSH is set to 1.44 ms and TP7 to 1.7 ms in the present embodiment. In use, the stimulation current ISTIM and pulse duration Tx vary constantly. However, the exacerbated condition is still limited by the maximum charge density per phase of the electrode, which is 0.35 mC/cm2 in the present embodiment, the maximum compliance of the electrode driver, which is set to 6V in the present embodiment, and the stimulation cycle T, which is set to 8.3 ms in the present embodiment. The compliance voltage generated from the stimulation also depends on the tissue and electrode impedance that is a combination of RS, RP, and CP. Substituting with the predefined time parameters, the following equation is obtained:
An example value estimation can have a measured RS of 15KΩ, RP of 100KΩ, and CP of 0.25 μF. Therefore, τp=25 ms and τS=3.3 ms in the example. The exacerbated situation stimulation can be ISTIM at 370 μA and Tx at 300 μs, where the stimulation current reaches the maximum charge density and the output voltage hit the compliance limit at the end of the pulse. Adding a safe margin in both the estimation and application, the threshold voltage VCmax can be set to 1V. Substituting these values into the IL equation provides a DC leakage threshold of 12 μA for this electrode.
It can be difficult to directly monitor the small DC leakage on each electrode. However, the relationship between the DC leakage current and the electrode voltage, VE, which is the sum of electrode capacitor voltage VC and the I-R drop through the series resistance of the electrode tissue interface can be determined. This voltage is detectable at the electrode driver output in the implant condition. In order to estimate the leakage current level using the electrode voltage measurement, predefine the shorting duration and frequency the same as the condition used in the above analysis and measure the electrode voltage at the driver output in a predefined point after the electrode reaches its steady state.
It should be noted that the test described above is used to estimate the voltage caused by the leakage current in the absence of any stimulation current. The electrode voltage map induced by the DC leakage current is shown in
Therefore, the DC leakage current threshold current ILmax is converted to an electrode voltage threshold VEth. In the previous example, the threshold voltage from 12 μA threshold current will be 0.75V. In use, this threshold voltage can be pre-calculated from equation (15) for a range of normal electrode/tissue impedances, as exampled in
When an electrode appears to have high impedance, it either has severed its connection internally from the driver or has lost effective electrode surface area interfacing with the tissue. In the first case, which shows very high impedance, DC leakage will not cause bubbling of the electrode, and therefore poses minimal safety concern. In the second case, however, DC leakage will affect the electrode in the same way as a normal electrode. When a high impedance electrode is detected, the electrode should be turned off so that no stimulation current is applied and the build up voltage on the electrode capacitor is solely from DC leakage. Therefore the DC leakage threshold does not include the effect of stimulation (ISTIM, Tx) in equation (15). With similar analysis, we have the leakage threshold for high impedance electrodes where the electrode is turned off:
A VCmax value of 1V can be applied to the high impedance electrodes to estimate the leakage threshold ILmax. The voltage threshold VEth can be calculated from equation (15) with the ILmax value from equation (16). However, a high impedance electrode can have smaller double layer capacitance value CP because of the reduced effective surface area. For this reason, the nominal CP value of high impedance electrodes is much smaller than a normal electrode. Assuming a nominal CP value of 0.05 μF, the threshold voltage calculated using equation (15) for high impedance electrodes is shown in
One method of implementing an efficient sub-EDCF monitoring system based on the algorithm discussed above is illustrated in
Accordingly, what has been shown is an improved method and apparatus of stimulating neural tissue for improved response to brightness. While the apparatus and methodology has been described by means of specific embodiments and applications thereof, it is understood that numerous modifications and variations could be made thereto by those skilled in the art without departing from the spirit and scope of the disclosure. It is therefore to be understood that within the scope of the claims, the disclosure may be practiced otherwise than as specifically described herein.
This application is a divisional application of 11/927,391, filed Oct. 29, 2007, entitled Method and Apparatus to Provide Safety Checks for Neural Stimulation, which is a divisional of application Ser. No. 11/796,433, entitled “Method and Apparatus to Provide Safety Checks for Neural Stimulation”, filed Apr. 27, 2007, which is a continuation-in-part of, and claims the benefit of, U.S. non-provisional patent application Ser. No. 11/413,771 filed on Apr. 28, 2006, for “Method and Apparatus to Provide Safety Checks for Neural Stimulation” by Robert Greenberg, M.D. PhD, et al., the disclosure of all of which is incorporated herein by reference.
The present invention was made with support from the United States Government under Grant No. R24EY12893-01, awarded by the National Institutes of Health. The government has certain rights in the patent.
Number | Date | Country | |
---|---|---|---|
Parent | 11927391 | Oct 2007 | US |
Child | 13430356 | US | |
Parent | 11796433 | Apr 2007 | US |
Child | 11927391 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11413771 | Apr 2006 | US |
Child | 11796433 | US |