The present invention relates to a method and device for the improvement of hearing acuity.
It has been shown that optical means can replace electrical means for neural excitation. It has also been demonstrated that this effect can provide much better auditory performance than the electrical counterpart. However the internal structure of the ear is not directly adapted to light guidance and distribution, inter alia, because of the very small angles of curvature of the inner ear. In addition, directing light directly on the nerves requires some kind of alignment that is not easy to perform.
The invention comprises several features that aim at solving the main points that have been described in the introduction: the small radius of curvature of the internal ear, and the alignment of the light source and the nerve. This invention enables a large number of excitation channels leading to higher performance and sensitivity to different sound frequencies.
The most straightforward way to guide light into the ear is to use optical fibers. However, optical fibers guide light only if they are bent less than a critical curvature. If the curvature is too high (which can be the case with a small radius of curvature as in the ear), light is no more guided. Typically fibers cannot be bent with a radius of curvature lower than 5 to 10 mm. The cochlea spiral radius of curvature is as low as 2 mm. Therefore light might leave the fiber before arriving at its target. In a first aspect of this invention, light is guided by a metallic waveguide or alternatively a photonic crystal waveguide, as will be described below.
An additional challenge is the alignment of the light waveguide output with the auditory nerve. In the absence of a proper way to align, light is delivered in a wide angular range of directions transverse to the light guide. Only part of the light would impinge on the auditory nerve, and the device energy consumption would be much higher than what is effectively needed. To solve this problem, in a second aspect of the invention, the waveguide is maintained in place using a thermally deformable device, such as a thermally activated shape memory polymer (SMP) device that deforms under heating and maintains the light sources in place facing the auditory nerve, leading to more efficient excitation. Thermally activated SMP can be deformed upon heating from its permanent shape to a temporary shape and maintain this temporary shape upon cooling. Reheating of the polymer results in its returning to its permanent shape.
Finally, it is challenging to introduce the device in a well-controlled manner that is not dependent on the operator manual abilities; it is especially challenging to introduce a straight device in a spiraling tunnel. To solve this problem, in a third aspect of the invention, the device is rolled before use and then unrolled in a controlled manner during the installation.
The present invention will be understood and appreciated more fully from the following detailed description, taken in conjunction with the drawings in which:
The objective of the device is to bring light to different regions of the auditory nerve through the cochlea. The wavelength of the light used in these embodiments of the invention may be, without limitation, in the near infrared range of approximately 1000 to 1600 nm, or may also include ultraviolet, visible, infrared, far infrared or deep infrared light.
Light is sent to different regions of the auditory nerve using an array of waveguides. One way to do it is, for example, to introduce an array of optical fibers. However, the radius of curvature of the cochlea changes from 8 mm to 2 mm. Such a bending leads light to leave the optical fiber since total internal reflection does not occur. An alternative is to guide light in a total reflecting waveguide structure. In order to do so, it is possible to surround the transparent waveguide by a metallic or by a 1 dimensional photonic crystal (Bragg mirror with high reflective index contrast). Light is therefore guided even when the radius of curvature is very small. In order to deliver light at the right locations, such waveguides of different lengths can be formed to deliver light according to a given location. In
These waveguides are made of a light transparent material (such as optical adhesive from NORLAND or solgel), that possess some elasticity in order to be able to follow the shape of the cochlea. This point will be discussed below.
The transparent waveguide is surrounded either by a metallic surface such as silver, gold or aluminum, by a very low refractive index material (such as fluoride based materials) or by a properly designed alternate of high and low refractive index layers forming a totally reflective one dimensional photonic crystal.
The waveguides are then embedded in a polymer so that the inputs and outputs of the waveguides can be optically accessed. At the output of each waveguide a micro-lens can optionally be positioned so that light that exits the waveguide is focused on the nerve. This micro-lens can be external to the waveguide or alternatively the end facet of the waveguide can be formed so that it has a lens shape (105).
One of the challenges in such a system is to ensure that light is indeed delivered to the auditory nerve. However it is difficult to precisely align the device within the cochlea. In order to assist in this fine alignment, the substrate in which the waveguides are embedded can be an SMP. This SMP is such that below a critical temperature is has the shape 201 and when heated, it takes the shape 202 and stays under this shape when returning below the critical temperature. For example one can take the polymer marketed by SMPTech (Japan) with a critical temperature of 55° C. The device therefore, when heated, deploys ribs like an umbrella and pushes the waveguides against the auditory nerve.
The SMP is heated by embedding heating wires 203 within the polymer. These wires are connected to a low voltage current source that is controlled externally during the installation of the device. When the current is activated, the device deploys its ribs and positions the light source in front of the auditory nerve. The wire temperature is continuously monitored using the thermistor/thermocouple measurement.
The combination of micro-lenses and the stent-like deployment of the device allow using much less energy for nerve activation, which is a critical issue for mobile devices.
Finally, a last issue is the introduction of the device within the cochlea. The device can be identified as a ribbon with optical waveguides embedded within it. The spiral shape of the cochlea is such that it is difficult to precisely guide the ribbon-shaped device within the organ. There are chances that the device will by itself deform and part of the light will arrive opposite to the auditory nerve. Finally, this spiral shape requires the device itself not to be straight but rather take a more complex shape so that it fits the cochlea topology once introduced. This three-dimensional shape can be either a standard or adapted to each patient. The individual cochlea shape can be obtained using standard imaging procedure such as CT or MRI.
Using the same concept as described above for opening the device, heating wires are positioned along the ribbon as indicated in
The heating of the ribbon is not done in one step. Rather, the distal part of the ribbon is first heated and together with the introduction of the device, the ribbon is progressively heated until reaching the proximal region of the device. This can be obtained by introducing heating pads along the ribbon, while transferring the heating current in a low-resistivity wire 303.
In addition, the critical temperature of the ribbon SMP can be chosen to be different from the critical temperature of the ribs (for example, lower), so that unrolling the ribbon does not affect the ribs' opening.
It should be noted that the unrolling of the device and the deployment of the “umbrella” are controlled independently and the heating of one part of the device does not influence the heating of the second part.
Therefore, the procedure is as follows:
First the operator introduces the device by progressively unrolling it by heating the central part of the ribbon (
Once the device is introduced, the operator opens the device by heating the ribs of the device.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/IB2022/050638 | 1/25/2022 | WO |
Number | Date | Country | |
---|---|---|---|
63141563 | Jan 2021 | US |