The present invention relates to a method of measurement of suprasystolic waveforms from an upper-arm blood pressure cuff. The analysis and interpretation of these waveforms can be performed directly on the measured and processed signals. It has been demonstrated that the suprasystolic waveform changes significantly and predictably with the administration of vasoactive agents, physiological challenges, normal ageing processes and morbidities.
Recent literature has noted the potential importance of central blood pressure (waveform and values such as systolic, diastolic and mean) in the management of cardiovascular risk. To that end, it would be advantageous to be able to estimate central pressures.
More particularly, the invention relates to methods of physics-based modelling of the waveform measurement system and the arterial system being studied. One of the early models of this kind, constructed by Joe El-Aklouk at the Institute of Biomedical Technologies, Auckland University of Technology, was based on acoustic pressure wave propagation in tubes.
El-Aklouk's work involved the development of models to simulate wave propagations given known material properties, geometry, end conditions and driving inputs. El-Aklouk also studied the transmission of pressure oscillations in an artery to an externally applied inflatable cuff.
More recent models have been proposed by Berend Westerhof et. al. (Am J Physiol Heart Circ Physiol 292:800-807, 2007.) who examined the pressure transfer between the subclavian root and unoccluded brachial artery assuming that the more peripheral arteries present a known end-impedance. In another recent paper (J Appl Physiol 105:1858-1863, 2008), Westerhof et al. attempt to show that changes in impedance caused by changes in the peripheral circulation have a negligible effect on the central pressure prediction. Westerhof et al. do not consider the use of a suprasystolic cuff to isolate the brachial artery from the peripheral circulation.
A principal object of the present invention is to improve the estimation of central pressure waveform by applying a suprasystolic cuff and using an inverse brachial-artery to cuff-pressure model.
A more particular object of the present invention is to extend the theoretical work done by El-Aklouk and Westerhof et al. to include a derivation of an inverse model and specifically applies it to the subclavian-brachial arterial branch. This inverse model allows the prediction of the driving input pressure (in the aortic arch) from the measurement of a suprasystolic waveform. It has been found that the inverse model problem can be solved without resorting to computationally costly and iterative numerical methods. A solution has been derived with assumptions to allow presentation in a closed-form. The model utilises only physically meaningful parameters.
This model according to the invention has been applied to clinical data collected in a study of subjects undergoing cardiac catheterisation at Auckland City Hospital, led by Dr. Wil Harrison. The results of this study, which validate the model, are set forth hereinbelow.
Model Derivation
In the arterial system under consideration, a pressure wave propagates through a volume of blood enclosed by the left subclavian and brachial arteries. At the brachial end, the artery is essentially closed by the application of a suprasystolic blood pressure cuff.
A number of assumptions, along with their justifications, have been made, as follows:
Note that many of these assumptions do not hold for other measurement methods. In particular, the use of an arterial line or tonometry method does not allow the assumption of a constant, abrupt change in impedance, nor of zero blood flow. Measurement at a radial site also introduces a significant bifurcation into the arterial system being studied, as well as further compromising the thin-walled tube assumption. Note also that the model is most correct for the left arm. The right subclavian artery is one branching generation removed from the aorta.
Using the above assumptions, it can be demonstrated that pressure propagation in the system can be described by the acoustic wave equation in one dimension:
where pt is the acoustic pressure (local deviation from ambient pressure), c is the speed of sound, x is the spacial coordinate and t is time.
where E is the elastic modulus of the artery wall material, h is wall thickness, r is tube radius and ρ is blood density.
Given a constant arterial cross section, the speed of sound is constant in time and space, and so the general solution can be described by
p
t
=p
p(x−c t)+pn(x+c t)
where pp, pn represent positive- and negative-going pressure waves respectively, the total pressure pt being a superposition of the two component waves.
It is also known that at an impedance change (causing a change in the speed of sound) a portion of a wave will be reflected, and a portion (the remainder) of the wave will be transmitted.
Using this information, the following relationships between the total pressure in the aorta and the total pressure at the occlusion may be written, for the system depicted in
where subscripts 0, 1 and 2 respectively represent locations in the aorta outside the subclavian artery (source), the root of the subclavian artery and just before the occlusion. Factor a is the reflection coefficient at the cuff and is physically constrained to be in the range 0 to 1. Constant dt is the time taken for a wave to travel from the subclavian root to the cuff occlusion.
It can be seen that under the conditions and assumptions described above, which are specific to suprasystolic measurement, theoretically, the aortic pressure waveform can be easily reconstructed from the occlusion pressure, using only two parameters, a and dt.
Measurement Model
The formulation thus far has considered only the internal physics of pressure wave propagation within the artery. The sensing system according to the invention is, however, non-invasive and relies on the transfer of internal pressure oscillations to an externally applied inflatable cuff.
It has been reported in the literature that the amplitude of the transferred pressure oscillation is determined by a large number of factors. However, the critical parameter appears to be the transmural pressure—that is, the difference in pressure between the externally applied cuff pressure and the internal arterial pressure. This relationship is shown diagrammatically in
This information permits compensation of the oscillometric waveform to estimate the arterial pressure. The procedure described herein is presented as a functional example, and may not be the only or even the best method of compensation.
The suprasystolic waveform, after filtering, posc, represents cuff pressure oscillations less than a few mmHg in amplitude. This waveform is rescaled to the measured systolic, SBP, and diastolic, DBP, pressures to generate pss.
The difference between pss and the cuff pressure, pCuff, is calculated to be the transmural pressure, ptp. In the case of suprasystolic measurement the cuff pressure can be assumed to be a value 25 mmHg above SBP.
p
TP
=p
Cuff
−p
ss
A correction amount, pΔ is calculated as a function of the transmural pressure. In order to match the relationship shown in
p
Δ=(d pTP)4
where d is a scaling factor.
Other forms of correction may also be applied. It has been found from our analysis that the form of correction should be such that if the corrected and uncorrected waveforms are normalized, then the correction function should be monotonically increasing between zero and one, but proportionally greater for pressures between zero and one, as shown in
One function that satisfies this requirement is of the form
where pSSnorm is the uncorrected normalized waveform value, pCorr is the corrected normalized waveform value and d is a control parameter that controls the amount of correction.
Once the correction amount has been calculated, the estimated arterial pressure, pArt, can be rescaled as follows
An example of an estimated pressure wave (normalized to SBP=1, DBP=0) is shown in
Blood Pressure Scaling
In the above discussion, pArt was scaled to the measured systolic and diastolic pressures. In fact, the theory suggests that this is not a correct treatment. The arterial blood pressure is actually generated by wave reflections from the peripheral circulation. However, the arterial pressures estimated by suprasystolic measurement are a function of reflections from the cuff. The arterial pressures with a suprasystolic cuff will therefore likely be higher than the pressures without a suprasystolic cuff. The degree to which this is the case may depend on a number of factors. In this model, the scaling is treated as a constant factor, c.
p
t2
=c (PArt−Mean(pArt))
Note once again that pt2 is the pressure perturbation about a mean pressure, not the gauge blood pressure. The assumptions infer that the mean pressure does not change along the length of the subclavian-brachial artery. Therefore the aortic blood pressure can be estimated by the equation:
p
AO
=p
t0+Mean(pArt)
Model Summary
In summary, the present invention is a model-based method of predicting central aortic pressures (specifically at the opening to the left subclavian artery). The input to the model is the suprasystolic, oscillometric waveform. The model has the following parameters:
Clinical data collected in a study of subjects undergoing cardiac catheterisation at Auckland City Hospital, led by Dr. Wil Harrison, were used to experimentally verify the theory and model according to the present invention.
Clinical Validation
Data for clinical validation was collected under the leadership of Dr. Wil Harrison from the cardiac investigation laboratories at Auckland City Hospital. Twenty-seven subjects were recruited from consecutive cases, with the exclusion of those with known, severe aneurysms, moderate to severe arrhythmias, or abnormal subclavian/brachial anatomy. Suprasystolic, oscillometric waveforms were collected non-invasively using a blood pressure cuff and analyzed using the model according to the present invention. Concurrently, ten seconds of invasive pressure waveforms were collected with the catheter tip near the aortic root. The non-invasive blood pressure (NIBP) was measured with a monitor using its internal, Welch Allyn oscillometric NIBP module. NIBP was determined approximately thirty seconds prior to the collection of waveform data.
Of the twenty-seven subjects recruited, technical difficulties prevented measurements from being taken on two of the first subjects. Poor quality catheter tracings were recorded on a further two subjects. Catheter tracings for one additional subject were not available at the time of this analysis. Waveform measurements were obtained from the remaining twenty-two subjects with a mean (sd) signal to noise ratio, SNR, of 13.4 (3.06) dB. This represents very good signal quality.
Non-Invasive vs. Invasive Measurements
As a point of reference, the statistics derived directly from the non-invasive and invasive measurements were compared.
In the case of diastolic pressure, as shown in
Direct comparison between suprasystolic augmentation index as calculated by the R6.5 device (AISS) and central Augmentation Index AI is not justifiable due to the different method of calculation. However, if performed, it gives very poor limits of agreement of 57±94% and r of 0.56. The X-Y plot is shown in
Model-Predicted vs. Measured Invasive Measurements
The model described above was applied to the clinical data in an investigation into the feasibility of predicting central pressures. Model parameters were set as follows: a=0.7, c=1.25, dt=0.045 seconds and d=0.045.
As can be seen in
Similar results were obtained for diastolic blood pressure as shown in
The invasive and predicted augmentation indices were also compared. The relevant charts are shown in
Results of Clinical Study
It may be seen that the prediction of the central pressures using the model and method according to the present invention closely matches the actual values that were measured invasively. The method according to the invention does not require any calibration to central waveform parameters (such as diastolic and mean pressures). In view of the documented inaccuracies of the NIBP estimation, it appears that the central pressure variations cannot be significantly improved. Indeed, the blood pressure predictions easily pass the international standards for the accuracy of blood pressure devices (although this standard does not strictly apply to central pressure estimation).
There are a few methodological shortcomings to the current study, which are described below.
Ideally, parameters to the model would be determined by measurement of each individual subject. In the case of dt, this could be determined relatively easily using additional, non-invasive sensors, or possibly estimated from demographic information such as age, height, weight and sex. An analysis of the correlation between the derivative of the predicted waveform and the derivative of the measured waveform shows that this correlation is far more sensitive to parameter dt than to a. This is shown in
Note that the correlation shown in
Conclusions
The method according to the invention advantageously includes the steps of:
The method thus estimates central artery pressures and pressure waveforms from measurement of pressure pulse wave signals at a peripheral location using a blood pressure cuff.
The basic applications of this method are:
There has thus been shown and described a novel method for estimating a central pressure waveform obtained with a blood pressure cuff, which method fulfils all the objects and advantages sought therefor. Many changes, modifications, variations and other uses and applications of the subject invention will, however, become apparent to those skilled in the art after considering this specification and the accompanying drawings which disclose the preferred embodiments thereof. All such changes, modifications, variations and other uses and applications which do not depart from the spirit and scope of the invention are deemed to be covered by the invention, which is to be limited only by the claims which follow.
For example, additional aspects of the invention may include:
This present application claims benefit of priority from U.S. patent application Ser. No. 11/358,283, filed Feb. 21, 2006 (now U.S. Patent Publication No. 2006/0224070-A1, published Oct. 5, 2006); U.S. patent application Ser. No. 12/157,854, filed Jun. 13, 2008 (now U.S. Patent Publication No. 2009/0012411-A1, published Jan. 8, 2009); U.S. Provisional Application Ser. No. 61/127,436, filed May 15, 2008 and U.S. Provisional Application No. 61/194,193, filed Sep. 24, 2008. The invention disclosed and claimed herein is related in subject matter to that disclosed in U.S. Pat. No. 5,913,826, issued Jun. 22, 1999; U.S. Pat. No. 6,994,675, issued Feb. 7, 2006; and the aforementioned U.S. Patent Publication No. 2006/0224070-A1 and U.S. Patent Publication No. 2009/0012411-A1, all of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
61127436 | May 2008 | US | |
61127436 | May 2008 | US |