Not Applicable
Not Applicable
Not Applicable
1. Field of the Invention
The present invention relates to neutron radiation detectors. More specifically, the present invention relates to a method for calibrating neutron multiplicity counters.
2. Description of Related Art including information disclosed under 37 CFR 1.97 and 1.98
Correlated neutron counting using shift register analysis is an important and long established tool for the detection and quantification of fission events in waste and safeguards. A crucial step in the manipulation of the resultant multiplicity histograms to correlated event rate is the application of correction factors for dead-time losses. Dead time correction can limit the accuracy especially when highly efficient counters are used to measure items of high neutron output where a high proportion of the primary neutrons are random (e.g. (α, n) events).
There is presently no complete treatment for this problem. Advances in instrumentation design to lessen the effects are an important active area of interest. However, these approaches also require similar correction, albeit of lesser magnitude, until a higher event rate is reached. Thus dead time correction remains a vital area of importance in neutron counting.
At present the most popular approach to making allowance for dead time losses is based on the approximation developed by Nikolai Dytlewski. In this scheme, the losses in the signal triggered rates can be compensated for by using histogram multipliers which are functions of a single free parameter—the dead-time parameter, τ. In practice however some additional empirical adjustments are often required. The issue then becomes how to best find the optimum value of the dead-time parameter.
The traditional method of applying dead-time corrections to passive neutron coincidence counting based on shift register electronics is typified by the treatment described by Menlove and Swansen for the case of the HLNCC-II. The singles rate is corrected using the exponential of a quadratic through the origin of the observed Singles rate. The Reals rate is only a function of the Singles rate and is set equal to the Singles correction factor raised to the fourth power.
A variation used by some is to replace the quadratic by the corrected Singles rate itself so that the expression is transcendental but for all practical purposes can be evaluated using a low order (e.g. 7) nested set of exponentials. The multiplier for the Reals correction is also left as a free parameter although the expectation is that it be close to four times that of the Singles dead-time parameter. This approach is set out in Croft and Yates. However, for multiplicity counting the correction for Triples is more complex and the approximate approach described by Dytlewski is the most widely used treatment.
Accordingly, a need exists for a more efficient calibration method that allows for field calibrations and verifications of the multiplicity dead time parameter. Further, a need exists for a calibration method that reduces the required number of reference sources and lengthy measurement times. Further, a need exists for a simplified calibration method that allows for readily repeatable calibrations to be performed over the life of a counter. Finally, a need exists for a method for calibration that does not require use of costly NIST certified source. The present invention satisfies these needs and others as will be explained in the following description.
The present invention provides a new method for calibration of a neutron multiplicity counter dead time parameter. This method is based on measuring the Singles (S), Doubles (D), and Triples (T), rates for a small set of 252Cf sources spanning a wide dynamic range. The plot of the uncorrected T/D rate ratio against the observed S-rate is shown empirically to be linear with a slope of 4π. This behavior is consistent for a wide variation of counter types spanning a wide range of efficiency and a wide number of pre-amplifiers in a given system. This new approach is quick and simple to implement and allows for repeatable field calibration of virtually all available neutron multiplicity counters.
Normally, at least one well characterized 252Cf source, traceable to a national laboratory such as NIST is required to determine the neutron detection efficiency of a multiplicity counter. In practice you need several, because as it turns out most of the characterized sources available from commercial sources are suspect. With the method of the present invention, you do not need traceability; any set of 252Cf sources that have sufficient neutron output can be used in a calibration.
The present invention will be more fully understood by reference to the following detailed description of the preferred embodiments of the present invention when read in conjunction with the accompanying drawings, in which like reference numbers refer to like parts throughout the views, wherein:
All figures are drawn for ease of explanation of the basic teachings of the present invention only; the extensions of the figures with respect to number, position, relationship, and dimensions of the parts to form the preferred embodiment will be explained or will be within the skill of the art after the following teachings of the present invention have been read and understood. Further, the exact dimensions and dimensional proportions to conform to specific force, weight, strength, and similar requirements will likewise be within the skill of the art after the following teachings of the present invention have been read and understood.
Examination of non-dead-time corrected rates is helpful in dead-time calibrations. A plot of the ratio of the non dead-time corrected Triples rates, Tm, to Doubles rate, Dm, yields a linear function of the non dead-time corrected Singles rate, Sm. The preferred embodiment of the present invention utilizes this linear fit in subsequent dead-time calibrations.
In Dytlewski's approach, the Singles or Trigger rate is corrected empirically while a special formalism is required for the Doubles and Triples. Hybrid solutions suggest themselves where the traditional approaches are used for the Singles and Doubles wherever needed with only the histogram based treatment being applied to the Triples term. This last approach avoids discrepancy between standard neutron coincidence counter (NCC) and multiplicity results but introduces some complications in the propagation of rates errors.
In the preferred embodiment, the dead-time corrected Doubles and Triples neutron coincidence rates are determined as follows:
where traditionally,
where Sm is the non-dead-time corrected Singles rate, D, and T are the dead-time corrected Doubles, and Triples rates respectively, φ=τ/G, τ is the characteristic dead-time parameter, G is the coincidence gate width, c and d are empirical Doubles and Triples dead-time parameters, and pi and qi are the normalized elements of the observed (R+A) and A multiplicity histograms respectively.
Calibration for the dead-time parameters, c (or τD), d (or τT) and τ, is performed by measurement of a series of 252Cf sources spanning the expected count rate range (typically 1 kHz to 1 MHz). Because there is no significant multiplication or (α, n) reaction rate with the 252Cf sources, the ratios of T/D, T/S and D/S should be constant and, once corrected, independent of source strength. The dead-time correction parameters are then determined adjustment of the parameters to obtain the minimum Chi-Square value for each of the rates ratios. A typical data set is provided in Table 1.
For later reference this linear function, Γm, is described as:
From this, the result is an exponential response function of the kind:
Note that the zero rate intercept of this curve, Γ0, is the Triples rate equivalent of the well known Doubles rate parameter ρ0 used in traditional NCC counting. The value of Γ0 can be calculated using Equation 6.
where νS2 and νS3 are the 2nd and 3rd factorial moments of the spontaneous fission multiplicity distribution for 252Cf respectively, fd and ft are the Doubles and Triples coincidence gate fractions, and ε is the absolute detection efficiency for 252Cf spontaneous fission neutrons.
The method of the present invention considered neutron counters with detection efficiencies of approximately 50%. The intriguing aspect of the curves from these counters was that the slope, λ, of each of these ratio curves was approximately equal to the coincidence (Reals) dead-time parameter, a, but was consistent with 4τ to within statistical errors. The prospect of a simple determination of the multiplicity dead-time parameter, τ, warranted further examination. Similar data sets were collected from very dissimilar neutron counters spanning the efficiency range from 4.5% to 61% and from a single detector bank to 54 banks; each exhibited a simple linear response. However, rather than the simple correlation between λ and τ observed for the initial data sets, a strong efficiency dependence was observed. This is evident in
The measured histograms from the calibrations of the systems in
The fit results in the following values and it is interesting to note that the parameter b is consistent with the book value of
Equation 6 suggests that the fit to the data would be improved by including a dependence on the coincidence gate fractions. To investigate this possibility the data was refit to the following functional form:
Accordingly, the best fit to the data occurs for c=0.5 rather than the expected linear dependence. The plot of the data (
The present invention also considers the relation between the characteristic dead-time parameter, τ, and the standard coincidence dead-time. While there is a general correlation between the dead-time parameters and, on average, τ≈α/4, the deviations between the r and α/4 are typically on the order of 5% but can be as large as 40%.
Using the disclosed inventive method in the preferred embodiment, a neutron multiplicity dead-time correction calibration generally consists of the measurement of a series of 252Cf sources ranging in neutron emission rate from 1000 cps to 2,000,000 neutrons per second. Each of these sources is placed in the center of the neutron assay cavity and counted for a period of time ranging from 1 hour to 8 hours. The calibration yields Reals+Accidentals and Accidentals multiplicity histograms for each source. The histogram data is used to calculate the dead-time corrected singles, doubles, and triples neutron coincidence rates using equations 1, 2 and 3 (above). The dead-time correction parameters are then determined via a Chi-square minimization process as discussed above.
Using the equations in the present embodiment, a determination of the neutron detection efficiency can be made without knowledge of the source strength. Both the characteristic dead-time parameter, τ, and the slope, λ, of the T/D-ratio can be determined without knowing the absolute neutron emission rate of the 252Cf sources. The observed efficiency dependence of the ratio of these two values suggests that these previous measurements can be used to determine the efficiency of the neutron counter without knowing the source strength.
By treating the fit results to Equations 7 and 8 as calibrations, the efficiency for a counter can be calculated from the values τ and λ in
The approximate measurement uncertainties are given in Table 3 along with each of the efficiency values. The “measured efficiency” is based on a weighted average of several 252Cf sources of known output. The limiting factor in the uncertainty is the source certificates from the suppliers. The uncertainty in the efficiencies determined by examination of the dead-time values is presently dominated by the statistical precision of the measurement. The typical multiplicity dead-time calibrations include measurements of 6 to 9 different neutron emission rates, each lasting from 1 to 4 hours. A full calibration therefore requires about 2 calendar days to complete. Longer counting periods will dramatically improve the measurement precision.
The results show some unexpected dependencies of the dead-time parameter such as a linear response of the uncorrected Triples to Doubles ratio as a function of the Singles rate. This linear behavior can be used to greatly simplify the calibration of a detection system. Count times of several hours per source and a minimum of 6 measurements spanning a count rate range of 1 kHz to 1 MHz are expected to be necessary to achieve 1% accuracy in the efficiency value for counters with greater than 30% detection efficiency.
To determine the neutron detection efficiency without the use of traceable standards we also calculate the non dead-time corrected rates. To determine the non dead-time corrected rates, the rather complex α and β arrays used by Dytlewski are replaced by the traditional weighting factors for determination of the moments of a distribution given as follows:
The ratio of the resulting non dead-time corrected triples and doubles rates are plotted as a function of the non dead-time corrected singles rates. The ratio of non dead-time corrected T:D is a linear function of the non dead-time corrected singles rate. The slope, λ, of this linear function can be used with the characteristic dead-time parameter, τ, for the counter to determine the neutron detection efficiency.
Examination of the data sets for a large number of neutron multiplicity coincidence counting systems has shown that this ratio of λ/τ is related to the neutron detection efficiency as shown in
Although the invention hereof has been described by way of a preferred embodiment, it will be evident that other adaptations and modifications can be employed without departing from the spirit and scope thereof. For example, some of the calibration steps in the inventive method could be conducted automatically as well as by manual entry for spreadsheet analysis. The terms and expressions employed herein have been used as terms of description and not of limitation; and thus, there is no intent of excluding equivalents, but on the contrary it is intended to cover any and all equivalents that may be employed without departing from the spirit and scope of the invention 58,266.
This application claims the benefit of provisional Application No. 60/830,526, filed Jul. 13, 2006.
Number | Date | Country | |
---|---|---|---|
60830526 | Jul 2006 | US |