The invention refers to a method for the quantitative measurement of the pulse laser stability of synthetic silica glass by means of a direct absorption measurement according to the species of the patent claims.
It is well-known that the determination of the long-time stability of silica glass is performed by means of long-time or marathon irradiation nowadays combined with the simultaneous transmission measurement. For this purpose, the lithography irradiation requirements with low energy densities of about 1 mJ/cm2 or ≦1 mJ/cm2 at repetition rates of 1-4 kHz are applied in accordance with a future field of application. Due to the low absorption, sample lengths of >75 mm are required to ensure a sufficient accuracy of the transmission measurement. The marathon measurements shall confirm both the achievement of a constant, acceptable value after a long phase of absorption increase and the correlation to models of long-time ageing of silica glass. Depending on the energy density pulse numbers of several 109 and—for appropriate repetition rates and permanent irradiation—exposure times of some weeks are required for doing this. The disadvantage of this method is the fact that apart from high operation and material costs required for the study, e.g. one laser tube is necessary per examination (afterwards it is unusable), the samples can show irreversible structure changes (microchannels) even before reaching the demanded exposure period due to the required sample length. These changes make the evaluation of the measured results for model formation and technology development impossible.
Therefore, the task of the invention is to specify a procedure for determining the long-time stability of synthetic silica glass which allows to considerably reduce the time and costs of examination in particular by using short sample lengths (ca. 10 mm).
According to the present invention, this task is solved by the elements of the first patent claim. The elements of the subclaims serve the further advantageous development of the invention.
Thus, the inventive procedure aims to measure a silica glass sample for its absorption at different or continuously increasing energy densities, e.g. 5-20 mJ/cm2. These measurements result in the characteristic function α1(H) of the material before long-term ageing. Due to the energy transmission, this function reflects a non-linear dependency of the absorption coefficient on the light energy density. The subsequent permanent irradiation of an energy density being constant up to reaching a constant absorption value and having a value, e.g. 20 mJ/cm2, which is considerably higher than the ones of the state of the art, than the common values of optic lithography in typical application, which is <5 mJ/cm2, allows the reduction of the pulse number very effectively. According to the state of the art, this pulse number would be required over a long phase of absorption increase till reaching the constant absorption value. Thanks to the short sample length, the development of additional irreversible changes such as microchannels can be avoided. The absorption measurement for different energy densities subsequent to the permanent irradiation is preferentially performed after continuously decreasing the energy densities and leads to a second characteristic function α2(H), from which the first characteristic function α1(H) is subtracted. The resulting difference presents the intensity-dependent absorption increases for different light energy densities. In this way it is possible to determine the increase of absorption even for lithography-relevant intensities.
Thus, the nature of the processes acting in the defect generating and curing in synthetic silica glass exposed to UV/DUV pulse laser light offers the possibility to shorten the marathon measurements mentioned above, for example by increasing the light intensity. But, by applying state of the art technologies this would lead to an acceleration of the undesired irreversible changes (formation of microchannels). The application of direct absorption measurement methods combined with the inventive procedure allows to operate with samples of short lengths (10 mm instead of 75 mm and more) and to gain comparable study results concerning the absorption increase with reference to the state of the art. In particular, the method of laser induced diversion (e.g. by an LID construction according to DE 101 39 906) offers a great advantage. The inventive procedure offers a favorable development, if the permanent irradiation and thus the pulse laser stability measurement of the sample is performed in the range of low temperatures (T<200 K). It is part of the invention that the determination of the reached saturation of the development of absorbing defect centers cannot only be carried out by absorption or fluorescent measurements but also by transmission measurements.
Subsequently, the results of a comparison of the examinations according to the state of the art and of the inventive procedure are explained by using six diagrams. They show:
For the state-of-the-art investigation of a selected silica glass sample having a hydrogen content of >1017 Mol/cm3 by means of a pulsed laser with a repetition rate of 1000 Hz, 2.75·109, laser pulses of an intensity of 1.3 mJ/cm2 have been applied during 24 h over a period of 32 days. The result is a measurement point distribution—demonstrated in the diagram in
The irradiation of an equivalent silica glass sample performed according to the inventive procedure was made by an energy density of 20 mJ/cm2, with a repetition rate of only 250 Hz for a period of about 30 h, corresponding to 2.7·107 pulses. Subsequent to the irradiation, an intensity variation followed to determine the absorption values for other intensities. The irradiation of a further sample performed according to the inventive procedure was made by using the following irradiation parameters: energy density of about 10 mJ/cm2, repetition rate of 300 Hz, period of exposure of about 4.6 107 pulses, corresponding to about 43 h. The measurements to the inventive procedures are presented as point series A and B in the diagram of
In case of irradiation, the result is the dependency of the energy density of the point series A demonstrated in
For the laser-induced fluorescence measurement (hereinafter referred to as LIF measurement) at room temperature, the development of the LIF bands at 650 nm, being characteristic for the development of defect centers (NBOH centers), considerably depends on the hydrogen content (H2) under pulse laser irradiation, i.e. on the curing of defect centers between two laser pulses. A high hydrogen content allows a very efficient curing. Practically no increase of the LIF intensity results from this for the reference sample (H2=3.5·1018 cm−3) and the sample 1 [H2=(1.5-2.0)·1018 cm−3] according to
If the same measurements are taken at lower temperatures (T<200° K.), e.g. at −185° C., the behavior of the samples changes. Due to the low temperatures the mobility of the molecular hydrogen is drastically reduced, it is “frozen”. This effect mainly prevents the irradiation-induced defects from curing. Regardless of the H2 content, the three samples behave in the same way, that is like the sample 2 having a low H2 content at room temperature.
This fact is demonstrated in
According to
All elements presented in the description, the subsequent claims and the drawing can be decisive for the invention both as single elements and in any combination.
Number | Date | Country | Kind |
---|---|---|---|
103 31 589.6 | Jul 2003 | DE | national |