The embodiments described below relate to compensation methods and, more particularly, to a method of compensating the mass flow measurement of a Coriolis sensor, for temperature, using a known fluid density and the drive frequency.
Coriolis meters are designed to provide accurate measurements of both mass flow rate and density of fluids flowing through a pipeline under a wide range of process fluid and environmental conditions. There are many applications where Coriolis meters are used as a reference for the calibration of other flow measurement devices. Naturally, this requires a high degree of accuracy.
One such application is for rocket engine testing, for example. In this application, there is a need to calibrate Venturi flow meters used to measure the flow of liquid Hydrogen and Oxygen into rocket test stands. Accuracy requirements for a reference sensor, at process conditions, for such an application are high, with a need of 0.35% in some applications.
Currently, sensors in cryogenic applications, including LNG, liquid Argon, Nitrogen and Oxygen, have shown mass measurement accuracy, using traditional temperature correction, to be 1% at best.
It is understood that the relationship between mass flow rate and the flow measurement signal of a Coriolis sensor is highly dependent upon the stiffness of the vibrating tube or tubes. It is also understood temperature can affect the stiffness of the tube or tubes of a Coriolis flow sensor by three different mechanisms.
The first mechanism is the change in the Modulus of Elasticity with temperature. This effect was recognized many years ago and a linear temperature compensation was developed, as given in Equation 1:
{dot over (m)}=FCF·(Δt−zero)·(1−ϕ·ΔT) (1)
where:
{dot over (m)}=mass flow
FCF=Flow Calibration Factor (units: g/s per μs)
Δt=fundamental Coriolis time measurement
zero=Δt at no-flow conditions
ϕ=temperature coefficient for changing Modulus of Elasticity
ΔT=temperature difference (° C.).
Over the range of applications for most Coriolis sensors, the change in the Modulus of Elasticity with temperature is near linear, so this correction works well in most applications. When Coriolis sensors were first applied to Cryogenic applications, it was recognized that the modulus was non-linear below 0° C.; as shown in
{dot over (m)}=FCF·(Δt−zero)·(ϕ+ϕ1·ΔT+ϕ2·ΔT2+ϕ3·ΔT3) (2)
where each ϕ term is a polynomial coefficient that characterizes the non-linear modulus behavior, especially at low temperatures. This is illustrated in
The second mechanism influencing tube stiffness is the dilatation of the material with changes in temperature. If the tube is unconstrained, its length, cross section and the internal volume all change, effectively changing the stiffness.
It was empirically observed that the temperature correction on tube period squared, used in density measurement, was not the same as that for Flow Calibration Factor, FCF. It was determined that this was due to thermal expansion. General forms of the mass flow and density equations were developed, idealizing a U-tube Coriolis sensor as a cantilevered beam. The relationships for flow and density measurement, accounting for this mechanism are given in Equations 3 and 4; respectively.
{dot over (m)}=FCF·(Δt−zero)·(1−f(ϕ)·ΔT)·(1+α·ΔT) (3)
where:
α=thermal expansion coefficient
f(ϕ)=the polynomial expressed in Equation 2.
where:
K2=Period squared
C1 & C2=calibration constants
f(ϕ)=the polynomial expressed in equation 2.
The third mechanism affecting tube stiffness is thermally induced stress. If the tube or tubes are not free to move with changing temperature the thermal strain will be counteracted by a mechanically restoring strain. This effect is significant with straight tube or low-profile Coriolis sensors. For other sensor geometries this mechanism is essentially negligible.
A method for determining a mass flow measurement is provided according to an embodiment. A flowmeter sensor is calibrated at a first temperature. A fluid having a second temperature that is different from the first temperature is flowed through the flowmeter sensor. A density of the fluid is input into a flowmeter electronics. A compensated mass flow value of the fluid is determined with the meter electronics, wherein the Modulus of Elasticity of the flowmeter sensor is unknown.
A flowmeter comprising meter electronics configured to receive a process fluid having a second temperature, the meter electronics configured to communicate with a sensor assembly of the flowmeter is provided. At least one flow conduit is configured to receive the process fluid. At least one driver is configured to vibrate the at least one flow conduit. At least one pickoff for detecting vibrations of the at least one flow conduit is provided, wherein the flowmeter is calibrated at a first temperature. A density of the fluid is input into meter electronics, and the meter electronics is configured to determine a compensated mass flow value of the fluid wherein the Modulus of Elasticity of the at least one flow conduit is unknown.
According to an aspect a method for determining a mass flow measurement comprises a flowmeter sensor is calibrated at a first temperature. A fluid having a second temperature that is different from the first temperature is flowed through the flowmeter sensor. A density of the fluid is input into a flowmeter electronics. A compensated mass flow value of the fluid is determined with the meter electronics, wherein the Modulus of Elasticity of the flowmeter sensor is unknown.
Preferably, the density is a known reference value.
Preferably, the density is calculated from an equation of state.
Preferably, the equation of state comprises a pressure term and a temperature term.
Preferably, the compensated mass flow rate is calculated as:
Preferably, the accuracy of the compensated mass flow value is ±0.5%.
Preferably, the first temperature is a non-cryogenic temperature, and the second temperature is a cryogenic temperature.
According to an aspect, a flowmeter comprising meter electronics configured to receive a process fluid having a second temperature, wherein the meter electronics configured to communicate with a sensor assembly of the flowmeter. At least one flow conduit is configured to receive the process fluid. At least one driver is configured to vibrate the at least one flow conduit. At least one pickoff for detecting vibrations of the at least one flow conduit is provided, wherein the flowmeter is calibrated at a first temperature. A density of the fluid is input into meter electronics, and the meter electronics is configured to determine a compensated mass flow value of the fluid wherein the Modulus of Elasticity of the at least one flow conduit is unknown.
Preferably, the density is a known reference value.
Preferably, the density is calculated from an equation of state.
Preferably, the equation of state comprises a pressure term and a temperature term.
Preferably, the compensated mass flow rate is calculated as:
Preferably, the accuracy of the compensated mass flow value is ±0.5%.
Preferably, the first temperature is a non-cryogenic temperature, and the second temperature is a cryogenic temperature.
The same reference number represents the same element on all drawings. It should be understood that the drawings are not necessarily to scale.
The sensor assembly 10 includes a pair of flanges 101 and 101′, manifolds 102 and 102′, a driver 104, pick-off sensors 105 and 105′, and flow conduits 103A and 103B. The driver 104 and the pick-off sensors 105 and 105′ are connected to the flow conduits 103A and 103B.
The flanges 101 and 101′ are affixed to the manifolds 102 and 102′. The manifolds 102 and 102′ can be affixed to opposite ends of a spacer 106 in some embodiments. The spacer 106 maintains the spacing between the manifolds 102 and 102′. When the sensor assembly 10 is inserted into a pipeline (not shown) which carries the process fluid being measured, the process fluid enters the sensor assembly 10 through the flange 101, passes through the inlet manifold 102 where the total amount of process fluid is directed to enter the flow conduits 103A and 103B, flows through the flow conduits 103A and 103B and back into the outlet manifold 102′, where it exits the sensor assembly 10 through the flange 101′.
The process fluid can comprise a liquid. The process fluid can comprise a gas. The process fluid can comprise a multi-phase fluid, such as a liquid including entrained gases and/or entrained solids, for example without limitation. The flow conduits 103A and 103B are selected and appropriately mounted to the inlet manifold 102 and to the outlet manifold 102′ so as to have substantially the same mass distribution, moments of inertia, and elastic moduli about the bending axes W-W and W′-W′, respectively. The flow conduits 103A and 103B extend outwardly from the manifolds 102 and 102′ in an essentially parallel fashion.
The flow conduits 103A and 103B are driven by the driver 104 in opposite directions about the respective bending axes W and W′ and at what is termed the first out of phase bending mode of the flowmeter 5. The driver 104 may comprise one of many well-known arrangements, such as a magnet mounted to the flow conduit 103A and an opposing coil mounted to the flow conduit 103B. An alternating current is passed through the opposing coil to cause both conduits to oscillate. A suitable drive signal is applied by the meter electronics 20 to the driver 104 via lead 110. Other driver devices are contemplated and are within the scope of the description and claims.
The meter electronics 20 receives sensor signals on leads 111 and 111′, respectively. The meter electronics 20 produces a drive signal on lead 110 which causes the driver 104 to oscillate the flow conduits 103A and 103B. Other sensor devices are contemplated and are within the scope of the description and claims.
The meter electronics 20 processes the left and right velocity signals from the pick-off sensors 105 and 105′ in order to compute a flow rate, among other things. The communication path 26 provides an input and an output means that allows the meter electronics 20 to interface with an operator or with other electronic systems. The description of
The meter electronics 20 in one embodiment is configured to vibrate the flow conduit 103A and 103B. The vibration is performed by the driver 104. The meter electronics 20 further receives resulting vibrational signals from the pickoff sensors 105 and 105′. The vibrational signals comprise a vibrational response of the flow conduits 103A and 103B. The meter electronics 20 processes the vibrational response and determines a response frequency and/or phase difference. The meter electronics 20 processes the vibrational response and determines one or more flow measurements, including a mass flow rate and/or density of the process fluid. Other vibrational response characteristics and/or flow measurements are contemplated and are within the scope of the description and claims.
In one embodiment, the flow conduits 103A and 103B comprise substantially omega-shaped flow conduits, as shown. Alternatively, in other embodiments, the flowmeter can comprise substantially straight flow conduits, U-shaped conduits, delta-shaped conduits, etc. Additional flowmeter shapes and/or configurations can be used and are within the scope of the description and claims.
From
A method of determining the change in modulus with temperature using a known or assumed density at a temperature near the application is provided according to an embodiment. This embodiment also makes the assumption that the coefficient of thermal expansion is constant. From
In an embodiment, the total uncertainty of a cryogenic mass flow measurement is minimized by eliminating the Modulus of Elasticity correction and the related problems of calibrating flow at cryogenic temperatures. Using equations of state for fluid density, which are well known for single component fluids, such as liquid Hydrogen or Natural Gas for example, an accurate mass flow calculation can be achieved without the issues indicated above.
Unlike prior compensation methods, this method eliminates the dependency of the mass flow equation on the change in Modulus of Elasticity as a function of temperature. In an embodiment, an equation for Mass Flow Rate measurement, independent of the change in modulus with temperature, is provided; as given in Equation 5.
The fluid density ρf could either be input as a known quantity or, in the case of a known fluid like commercially pure Hydrogen, calculated from an equation of state using pressure and temperature inputs. Temperature and pressure may simply be input into a meter or may be measured by at least one of a temperature and pressure sensor. For a dual-tube U-tube Coriolis sensor operating in this manner, it is reasonable to expect a flow accuracy of ±0.5% while measuring a pure cryogenic fluid like liquid Hydrogen.
Turning to
In an embodiment, the first temperature (i.e. the temperature at which the flowmeter is calibrated) is non-cryogenic. This would typically correspond to a range of temperatures typical of a manufacturing facility—i.e. around “room temperature.” However, the flow fluid is cryogenic, so the temperature is between about −100° C. and −273° C. One benefit of this is a reduction in cost and difficulty of cryogenic fluid handling during calibration. For the reasons noted above, the flowmeter calibrated at standard room temperatures will still be accurate with cryogenic fluids—again a departure from the prior art. This is illustrated by Table 1, which is provided as an example only, and are no way limiting. The values therein are illustrative only for a one particular flowmeter model, and do not serve to limit embodiments.
Variables influencing Mass Flow Measurement.
{dot over (m)}=f(FCF,zero,C2,Δt,K,ΔT,α,ρ,other) (6)
Uncertainty of Mass Flow Measurement due to any single variable, x.
Total uncertainty of Mass Flow Measurement due to all variables.
Uncertainty and Portion of Flow Error due to uncertainty of the change in temperature from calibration, ΔT.
Uncertainty and Portion of Flow Error due to uncertainty in coefficient of thermal expansion, α.
Uncertainty and Portion of Flow Error due to uncertainty in fluid density, ρf.
Uncertainty and Portion of Flow Error due to uncertainty in tube period, K.
The total estimated flow error uncertainty for a Coriolis flow meter, with a C2 of 1943 kg/m3, is given in Table 1.
The detailed descriptions of the above embodiments are not exhaustive descriptions of all embodiments contemplated by the inventors to be within the scope of the present description. Indeed, persons skilled in the art will recognize that certain elements of the above-described embodiments may variously be combined or eliminated to create further embodiments, and such further embodiments fall within the scope and teachings of the present description. It will also be apparent to those of ordinary skill in the art that the above-described embodiments may be combined in whole or in part to create additional embodiments within the scope and teachings of the present description.
Thus, although specific embodiments are described herein for illustrative purposes, various equivalent modifications are possible within the scope of the present description, as those skilled in the relevant art will recognize. The teachings provided herein can be applied to other fuel consumption calculations of a fuel and water mixture and not just to the embodiments described above and shown in the accompanying figures.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2019/024530 | 3/28/2019 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62651593 | Apr 2018 | US |