Not Applicable.
Energy storage devices (e.g., batteries, fuel cells, ultracapacitors, etc.) have become significantly more prevalent in many government and commercial applications (e.g., automotive, military, space, electric utilities, medical, etc.). Consequently, there has also been an increased interest in smart monitoring systems that can effectively manage energy storage devices (ESDs) so as to optimize performance and extend life. An important aspect of these smart monitoring systems is the ability to estimate the response of an ESD to an anticipated load.
For example, the Lumped Parameter Model (LPM) has been used extensively by the Idaho National Laboratory (INL) to estimate the voltage response of a battery to a constant-current pulse for automotive applications. The LPM is an equivalent circuit model that recursively solves for the voltage behavior based on a given excitation current and a set of difference equations. It has been shown that the LPM is sensitive to variations in pulse amplitude and duration, and could therefore be a useful measure of state-of-health (Christophersen, 2003).
However, the excitation signals required to obtain estimates of the ESD response are not well-suited for in-situ applications since it generally requires a pulse test which may cause larger state-of-charge (SOC) swings than desired and even adversely affect the ESD (Christophersen, 2006). A need still exists to estimate the response of an ESD to an anticipated load using benign measurement techniques.
The INL has also shown that the pulse resistance for batteries is strongly correlated with the growth observed from corresponding electrochemical impedance spectroscopy (EIS) measurements (Christophersen, 2002). It has also been shown that EIS techniques are more benign than pulse tests (Christophersen, 2006) since it is a low-level, charge neutral signal that minimally perturbs the ESD. Suitable means for obtaining in-situ impedance spectra have already been developed. The Impedance Noise Identification method (U.S. Pat. No. 7,675,293 B2) uses a random signal excitation to acquire a high resolution impedance spectrum, but at the expense of computationally intensive data processing. An alternative approach is known as Compensated Synchronous Detection (U.S. Pat. No. 7,395,163 B1), and it incorporates a wideband sum-of-sines input signal to measure the impedance. It yields a faster measurement, but at the expense of lower resolution. A variant of Compensated Synchronous Detection is Fast Summation Transformation (FST). The principal attributes of FST (Morrison, 2009) are that it only requires one period of the lowest frequency to complete the measurement, and the data processing algorithm is very simple.
All patents, patent applications, provisional patent applications and publications referred to or cited herein, are incorporated by reference in their entirety to the extent they are not inconsistent with the teachings of the specification.
The subject invention involves a method by which the response of an energy storage device (ESD) to a pulse excitation can be predicted using impedance measurement techniques. This method assumes that the amplitude and duration of the anticipated or desired pulse excitation is known a priori, or can be inferred based on historical data (e.g., an average pulse profile based on typical automotive driving cycles). Assuming a periodic behavior of the desired pulse profile, the Fourier series coefficients can then be determined (note that the assumption of a periodic signal is for analytical purposes), and combined with measured impedance data to estimate the response.
The Fourier coefficients of the desired pulse profile are first used to establish the frequency range of the impedance measurement. For example, the period of the lowest frequency for the impedance measurement should be less than or equal to the period of the pulse profile. The maximum frequency of the impedance measurement should be greater than or equal to the largest desired harmonic value used in the Fourier coefficients used to recreate the desired pulse profile.
Knowing the desired frequency range, the ESD impedance spectrum can then be measured using any available methodology. For rapid, in-situ applications, techniques such as Impedance Noise Identification, Compensated Synchronous Detection, or Fast Summation Transformation can be easily implemented. The frequencies in the impedance measurement spectrum should correspond to the Fourier coefficients from the simulated pulse. In some cases (e.g., with Fast Summation Transformation), the impedance spectra will be lower resolution than desired due to the need for a very rapid measurement. However, the use of linear interpolation, cubic spline functions, or other similar types of curve fitting techniques can be used to estimate the impedance at other desired frequencies within the measured range.
Using a constant current pulse as an example, the Fourier coefficients of the desired or anticipated pulse profile are multiplied by the corresponding impedance measurements at each frequency. These data will provide the voltage response at each frequency of interest, and the results can then be summed to determine the overall voltage response of the ESD to the anticipated current pulse profile.
Thus, the ESD response of a pulse excitation can be estimated based on a simple impedance measurement combined with the Fourier coefficients of a simulated pulse. The estimated response behavior can be used by smart monitoring systems to more effectively manage ESD usage. For example, if the estimated response exceeds a desired threshold, the smart monitoring system can either shutdown operations, or iteratively determine a pulse excitation level than can be successfully applied to the ESD without violating operational limits (e.g., managing how much power assist is provided by the ESD in automotive applications). A smart system can also use this information to know when warning signals should be sent to a user prior to a demand being placed on the ESD.
a is a plot of the impedance spectrum magnitude for both the ideal and simulated conditions.
b is a plot of the impedance spectrum phase for both the ideal and simulated conditions.
The method of the subject invention uses wideband impedance measurements to predict the response of an energy storage device (ESD) to a pulse excitation. The impedance spectrum can be acquired by various methods, but rapid, in-situ techniques such as Fast Summation Transformation (FST) are preferred. FST is based on a computationally simple approach, and it only requires one period of the lowest frequency to complete a measurement (Morrison, 2009).
In a preferred embodiment, the anticipated or desired excitation pulse consists of a constant current square-wave profile. If it assumed that this profile is periodic (for analysis purposes only), the waveform can be decomposed into the constituent harmonic components using Fourier series methods. An example of an excitation pulse is shown in
Where:
The frequency range of the impedance measurement should be well matched with the Fourier series harmonic frequencies (Equation 4). For example, the lowest frequency for the impedance measurement should be less than or equal to the period of the simulated pulse (i.e., less than or equal to 1/T). The highest frequency for the impedance measurement should correspond to the maximum harmonic component desired to recreate the pulse waveform (i.e., the maximum value for n used to recreate f(t) in Equation 2).
If the measured frequencies in the impedance spectrum match the desired Fourier harmonic frequency components from the simulated pulse waveform, then the responses from the ESD to an excitation pulse can be obtained at each frequency. For example, the impedance of the ESD at a given frequency (i.e., ωn=nωo), is shown in Equation 5. The voltage response at that frequency is the impedance (Equation 5) multiplied by the corresponding harmonic component of the current pulse (Equation 3), as shown in Equation 6. Based on the Fourier series pair of Equation 2, the estimated voltage drop due to a current pulse is the sum of the individual frequencies, as shown in Equation 7. Given a bias voltage (VB0), the ESD voltage response (VP) can then be estimated as shown in Equation 8, where the voltage drop (VZ) is subtracted from the bias. Thus if the terminal voltage of the ESD is known or measured, and if a relatively recent impedance spectrum of the ESD is available, then an estimate of the response to an excitation pulse can be obtained.
To make use of Equation 8, the values of the impedance spectrum {right arrow over (ZB)} at the Fourier series frequencies ωn must be obtained. In most cases, however the impedance measurements will have a logarithmic frequency spread, whereas the Fourier series uses linearly increasing frequency components. This can be resolved by using linearly increasing frequencies during the impedance measurement instead, but at the expense of longer measurement durations and more computationally intensive analysis techniques. Another option is to estimate the impedance at the desired frequencies within the measurement range using techniques such as linear interpolation or cubic spline fits to obtain the values of the impedance spectrum {right arrow over (ZB)} at the Fourier series frequencies ωn.
The Lumped Parameter Model (LPM) was used to verify the effectiveness of this method. The LPM equivalent circuit is shown in
The impedance spectrum of the LPM can be simulated using the Fast Summation Transformation measurement technique (Morrison, 2009) and compared to the ideal response of the equivalent circuit. The FST algorithm was applied to the LPM at the same starting frequency as the square wave pulse profile of
However, as described above, the resolution of FST is insufficient to estimate the pulse response. To obtain a higher resolution impedance spectrum, the cubic spline fit was implemented using built-in software functions (e.g., in MATLAB matrix calculation computer software) and the resulting impedance estimations were compared to the expected response. The expected impedance can be calculated based on the frequency response of the LPM using the assumed parameters shown in Table 1. Table 2 shows the expected and estimated impedance for ten sequential odd-harmonic frequencies, with a starting frequency of 0.01 Hz. As shown, the spline fit is very good compared to the expected impedance spectrum.
Equations 6 through 8 were then implemented to estimate the voltage response of the desired pulse based on the known input current and the FST impedance measurements. The resulting voltages at each frequency were summed, and the total response is provided in
It is understood that the foregoing examples are merely illustrative of the present invention. Certain modifications of the articles and/or methods may be made and still achieve the objectives of the invention. Such modifications are contemplated as within the scope of the claimed invention.
This application claims the benefits of U.S. Provisional Patent Application No. 61/186,358; filed Jun. 11, 2009. The disclosure of this application is hereby incorporated by reference in its entirety, including all figures, tables and drawings.
This invention was made with government support under Grant No. DE-AC07-05ID14517 awarded by the United States Department of Energy. The government has certain rights in the invention.
| Number | Date | Country | |
|---|---|---|---|
| 61186358 | Jun 2009 | US |