This invention relates generally to methods and systems for simulating progressive lenses, more specifically to simulate progressive lenses with a guided lens design exploration system.
Prescriptions for eyeglasses, or spectacles, are generated today by devices and methods that were developed many decades ago, their foundations going back even farther in time. While these methods and technologies are mature, and served the patient population well, they did not benefit from the remarkable progress of optoelectronics and computational methods that revolutionized so many areas in telecommunications, consumer electronics, and interactive functionalities. Viewing optometry from the vantage point of modem opto-electronics and computer science, several areas of needs and opportunities can be identified. Some of the leading challenges and opportunities are listed and reviewed below.
(1) No trial before purchase: Patients who are asking for a progressive lens prescription are examined today only with single-power, non-progressive lenses, one for distance vision, and one for near vision. Therefore, the patients do not experience the prescribed progressive lenses in their totality before they purchase it. Since the patients do not “test-drive” the progressive lens prescriptions, they discover problems or inconveniences too late, only after the glasses have been provided to them.
(2) Only analog optometric devices are used: The single power lenses and other optical systems used by optometrists today have been developed long time ago. They are analog optical systems, and have not adapted and adopted the many advances of modem optoelectronic technology. This is to be contrasted by the very successful adaptation of optoelectronic technologies into other areas of ophthalmology, such as retinal imaging by Optical Coherence Tomography, aberrometric diagnostic devices, and optical diagnostic devices used in preparation for cataract surgery to determine the appropriate intraocular lenses.
(3) Only two distances tested: These analog lens systems test a patient's vision only at two distances, the near and the distance vision. In contrast, most patients have unique usage patterns, and often need the optimization of their spectacles for three or more distances depending on their individual habits.
(4) Eyes are only tested individually: Most of the diagnostic methods are applied for single eyes, blocking the other eye. Such approaches disregard the coordination between the eyes when they create the visual experiences, as well as the various effects of the vergence that are quite important for the full evaluation of the binocular visual acuity.
(5) Progressive lens prescriptions are under-defined: The design of progressive lenses is a complex process. Different companies have different proprietary optimization algorithms, with many parameters that use different search and optimization strategies. In contrast, the optometrists only determine 2-3 parameters about a patient's vision during an office visit. In a mathematical sense, providing only 2-3 parameters for the design of a progressive lens seriously under-defines the optimization algorithm of the lens design. When the design optimization algorithms have insufficient information, the algorithms can, and often do, stop at designs that are not truly optimal, and they are unable to identify the truly optimal design.
(6) Determining more parameters would increase treatment time per patient: Optometrists could run more tests to determine more parameters. However, doing so would extend the time spent with individual patients. This would have a negative effect on the economic model of the optometrists.
(7) Patients often need to return glasses for adjustments. Patients are unsatisfied with their progressive lenses in a statistically relevant fraction of the cases. Therefore, patients often return to the optometrist office asking for adjustments. It is not uncommon that a progressive lens has to be readjusted 3-4-5 times. The time and cost associated with these return visits seriously impacts the satisfaction of the patient, and undermines the economic model of the optometrist.
(8) Lens design verification groups are small: Lens design algorithms are typically optimized in interaction with a small test group, from less than a hundred to a couple hundred patients. Using only such small test groups for optimizing such a complex problem can lead to lens design algorithms that are not optimal. Subsequent complaints from the larger number of real patients yields feedback from a larger group, but this feedback is incomplete and uni-directional.
(9) Testing images do not reflect patient's actual visual needs: Eye testing uses standardized letters that are rarely reflective of a patient's actual needs. Testing just about never involves images that are relevant for the individual patient.
(10) Peripheral vision rarely tested: Optometrists rarely tests peripheral vision, whereas for some professions, peripheral vision might be a high value component of the overall visual acuity.
(11) Modern search algorithms are not yet utilized: Recent advances that greatly boost the efficiency of search algorithms over complex merit-landscapes, have not yet been adapted to the design of progressive lenses.
(12) Artificial Intelligence is not used: Recent advances in implementing Artificial Intelligence for system improvements also have not found their way yet into optometry.
At least the above dozen problems demonstrate that optometry could greatly benefit in a large number of ways from implementing modern optoelectronic technologies, in a patient-centric, customized manner, that also uses progress in modern computer science.
To address the above described medical needs, some embodiments of the invention include a Progressive Lens Simulator, comprising: an Eye Tracker, for tracking an eye axis direction to determine a gaze distance, an Off-Axis Progressive Lens Simulator, for generating an Off-Axis progressive lens simulation (Off-Axis PLS); and an Axial Power-Distance Simulator, for simulating a progressive lens power in the eye axis direction, thereby creating a Comprehensive progressive lens simulation from the Off-Axis PLS.
Embodiments also include a method of operating a Progressive Lens Simulator, the method comprising: tracking an eye axis direction by an Eye Tracker to determine a gaze distance; generating an off-axis progressive lens simulation (Off-Axis PLS) by an Off-Axis Progressive Lens Simulator; and creating a Comprehensive progressive lens simulation from the Off-Axis PLS by simulating a progressive lens power in the eye axis direction by an Axial Power-Distance Simulator.
Embodiments further include a Progressive Lens Simulator, comprising: an Eye Tracker, for tracking, an eye axis direction to determine a gaze distance; an Integrated Progressive Lens Simulator, for creating a Comprehensive Progressive Lens Simulation (PLS) by simulating a progressive lens power in the eye axis direction, in combination with venerating an Off-Axis progressive lens simulation (Off-Axis PLS).
Embodiments also include a Head-mounted Progressive Lens Simulator, comprising: an Eye Tracker, for tracking an eye axis direction to determine a gaze distance; an Integrated Progressive Lens Simulator, for creating a Comprehensive Progressive Lens Simulation (PLS) by simulating a progressive lens power in the eye axis direction, in combination with generating an Off-axis progressive lens simulation (Off-Axis PLS); wherein the Eye Tracker and the Integrated Progressive Lens Simulator are implemented in a head-mounted display.
Embodiments further include a Guided Lens Design Exploration System for Progressive Lens Simulator, comprising: a Progressive Lens Simulator, for generating a progressive lens simulation for a patient with a progressive lens design; a Feedback-Control Interface, for inputting at least one of control and feedback by the patient, in response to the progressive lens simulation; and a Progressive Lens Design processor, coupled to the Feedback-Control Interface, for receiving the at least one of control and feedback from the patient, and modifying the progressive lens design in response to the receiving, wherein the Progressive Lens Simulator is configured to generate a modified progressive lens simulation for the patient with the modified progressive lens design.
Embodiments also include a method of Progressive Lens Simulation, comprising: (a) activating a lens design with a Progressive Lens Design Processor; (b) generating an image by an Image Generator of a Progressive Lens Simulator; (c) generating a Comprehensive PLS, simulated from the generated image by the Progressive Lens Simulator, utilizing the lens design; (d) acquiring a visual feedback via a Feedback-Control Interface, responsive to the generating of the Comprehensive PLS with the lens design; (e) modifying the lens design by the Progressive Lens Design Processor in relation to the visual feedback; and (f) re-generating the Comprehensive PLS with the modified lens design by the Progressive Lens Simulator.
Embodiments further include a Deep Learning Method for an Artificial Intelligence Engine for a Progressive Lens Design Processor, comprising: activating a Visual Feedback-Design Factor Neural Network for a Progressive Lens Design Processor; receiving as input a visual feedback vector into the Visual Feedback-Design Factor Neural Network; outputting a design factor vector with the Visual Feedback-Design Factor Neural Network in response to the inputting; wherein coupling matrices of the Visual Feedback-Design Factor Neural Network of the Progressive Lens Design Processor were trained by performing a deep learning cycle.
Embodiments also include a Supervised Multi-station system of Progressive Lens Simulators, comprising: a set of Progressive Lens Simulators, individually including an Eye Tracker, for tracking an eye axis direction to determine a gaze distance; an Off-Axis Progressive Lens Simulator, for generating an Off-Axis progressive lens simulation (Off-Axis PLS); and an Axial Power-Distance Simulator, for simulating a progressive lens power in the eye axis direction, thereby creating a Comprehensive progressive lens simulation from the Off-Axis PLS; and a Central Supervision Station, in communication with the Progressive Lens Simulators, for providing supervision for an operation of the individual Progressive Lens Simulators.
The systems and methods described in the present patent document address the above articulated medical needs at least in the following aspects. These aspects are organized in a contrasting format to the previously described challenges of the state of the art.
(1) No trial before purchase: In embodiments, the visual experience of progressive lenses is simulated by a Progressive Lens Simulator. This system empowers the patient to actively and interactively explore and experience progressive lenses with different designs in real time. The patient can explore as many simulated progressive lens designs as she or he wishes before settling on a particular design. In short, the patient can explore, “test drive”, and “try on” the progressive lens before purchasing it.
(2) Only analog optometric devices are used: Embodiments use modem digital optoelectronic technology instead of analog optical technologies.
(3) Only two distances tested: The patient can explore the performance of the various progressive lens designs at as many distances as he or she wishes by virtue of modern optoelectronic designs.
(4) Eyes are tested individually: The patient can explore the visual experience with the simulated progressive lenses with both eyes simultaneously. This approach allows the lens design selection process to include and optimize for the patient's experience of the effects of vergence.
(5) Progressive lens prescriptions are under-defined: The patient can explore the many possible progressive lens designs exhaustively. The search can concentrate on many particular aspects of the performance of the lens. Monitoring the search process in detail provides an extensive amount of data about the patient's vision for the advanced lens design software. The acquisition of this larger amount of data turns the lens design process from under-defined to appropriately defined in a mathematical sense. The exploration process can be continued until the lens design software concludes that it has sufficient data to zoom in on the most optimal lens design.
(6) Determining more parameters would increase treatment time per patient: The selection process of the best progressive lens design with the here-described embodiments may take longer, or even substantially longer than the duration of a typical office visit today. This may be perceived as an economic “excessively high costs” argument against the described system. However, much of the search is guided by smart software and thus does not require the active presence of the optometrist. Rather, the optometrists play a supervisory role and thus can supervise even a greater number of patients per day with these Progressive Lens Simulators than with the traditional methods.
(7) Patients often need to return glasses for adjustments: Since the patients explore all relevant and possible progressive lens designs in real time, the here-described Progressive Lens Simulator system minimizes the patient complaints and returns. This dramatically improves patient satisfaction and greatly boosts the economic model of the optometrists.
(8) Lens design verification groups are small: The search data are collected from all participating optometrist offices. Therefore, the progressive lens design software will have access to the recorded search patterns, patient behaviors, and eventual patient choices from the test group of millions of patients. Access to such a huge and rapidly growing database will be used to rapidly and efficiently improve the progressive lens design algorithms.
(9) Testing images do not reflect patient's, actual visual needs: The Progressive Lens Simulators offer to patients the testing of their vision on any images they choose. A promising implementation is presenting use-relevant images that are relevant for and simulate the patient's actual activities.
(10) Peripheral vision rarely tested: The digital image-projection system can present both central and peripheral images for a full characterization of a patient's vision.
(11) Modern search algorithms are not yet utilized: The Guidance System for Patient Exploration in some embodiments uses modern search techniques that were developed to explore complex, multiply interdependent and constrained merit-landscapes.
(12) Artificial Intelligence is not used: Artificial Intelligence Engines are used for constantly improving and upgrading the system's software block-by-block.
The here-described systems can only yield positive outcomes, because they can be first used to determine the traditional progressive lens prescription with the traditional procedure. Subsequently, the Progressive Lens Simulator can perform various advanced searches by here-described embodiments, and guide the patient to the advanced progressive lens prescription best suited for him/her. Finally, the Progressive Lens Simulator can simulate the traditional progressive lens experience, followed by the advanced progressive lens experience, and alternate back-and-forth so that the patient can compare the two progressive lens experiences to make her/his final choice. Since the patient can always return to and select the traditional progressive lens design, the overall outcome of the process cannot be worse than the traditional process, only better.
This patent document describes many embodiments, techniques and methods. It also describes more than a dozen advantages over existing traditional systems. Thus, the described advantages are not limiting for all embodiments. Indeed, embodiments which possess only one or a few of the advantages are already novel over existing systems. Also, several other, not-yet-listed advantages exist that make the system novel. Moreover, several of the described aspects can be combined in various embodiments for additional synergetic advantages.
(1) Patients can explore and “try on” many different progressive lens designs before selecting one for purchase.
(2) Embodiments use modern digital optoelectronic technology.
(3) Patients can test their vision at as many distances as they wish.
(4) The two eyes of the patients can be tested synchronously, thus factoring their vergence into the overall visual experience.
(5) The eventually selected progressive lens designs and prescriptions are well-defined because a sufficient number of parameters are determined.
(6) Since patients are exploring the progressive lens designs on their own, guided by a smart software, the demand on the time of the optometrists actually decreases relative to present systems, as the optometrists are expected only to supervise the patients' exploration.
(7) Since patients select their own design, glasses are returned for adjustments much less frequently.
(8) Lens design verification groups are large and ever expanding.
(9) Testing images reflect patient's actual visual needs.
(10) Patients' peripheral vision is tested as extensively as the patients desire.
(11) Cutting edge search algorithms are utilized to guide the patient exploration.
(12) Artificial Intelligence is used to continuously upgrade and improve both the lens design and the patient guidance systems.
These system level concepts of the GPS 10 are described in general in
The PLS 100 can further include an Axial Power-Distance Simulator (ADS) 130, that simulates a combination of the distance of the viewed image and the axial power of the progressive lens. Since the PLS 100 simulates progressive lenses, the optical power varies substantially over the lens surface. In multistage embodiments, the PLS 100 simulates this with a combination of simulating the most important axial power with the ADS 130, and the off-axis power with the separate OPS 120. In integrated embodiments, the PLS 100 simulates the spatially varying optical power with an Integrated Progressive Lens Simulator IPLS 200.
The PLS 100 can further include a Vergence-Distance Simulator (VDS) 140 that simulates the viewing distance in a different manner. The VDS 140 can present the images for the two eyes not dead ahead, but moved closer to each other, in order to create the visual experience of the target image being at a closer viewing distance. Finally, a Zoom-Distance Simulator (ZDS) 150 can simulate a change of the viewing distance (changed by the PLS 100 from a first distance to a second distance) by zooming the image in or out. Doing so can further increase the sense of reality of the visual experiences generated by the PLS 100 for the patients.
The GPS 10 can include the Guided Lens Design Exploration System for Progressive Lens Simulator (LDES) 300, to guide the exploration of the large number of possible progressive lens designs by the patient with an efficient and informed strategy. The LDES 300 can include a Feedback-Control Interface (FCI) 310. This FCI 310 can be used by the patient to enter feedback and control signals for the PLS 100, in order to express preferences and provide feedback on the simulated progressive lens designs 123. In some embodiments, the LDES 300 can include a Progressive Lens Design Processor (PLD) 320. The PLD 320 can create the specific progressive lens design based on measurements of the patient's eyes, based on the patient's input, feedback, and control signals, and on lens design algorithms. The created specific progressive lens designs can be communicated by the PLD 320 to the PLS 100 to create the corresponding progressive lens visual experience for the patient.
The LDES 300 can further include a Search Guidance Engine (SGE) 330. The patient often, or even typically may not know how to change the design of the progressive lens to improve its optical performance. The patient typically only senses that the last change of the design made the visual experience better or worse. Or, the patient can articulate what improvements she/he is looking for. But since the progressive lens design can be modified in many different ways to bring, about such changes, a guidance system to affect the desired change in an informed and strategic manned can be useful and in fact necessary. Providing such guidance is one of the functions of the SGE 330. The SGE 330 can receive a desired improvement or preference from a patient, and then suggest in return to the patient how to translate the requested improvement into a change of the lens design.
Some embodiments of the LDES 300 can further include a Synchronous Eye Exploration Controller (SEC) 340, that oversees and controls the visual experience of the two eyes synchronously, and plays an important role in integrating desired design improvements from the two eyes. Finally, the LDES 300 can also include a Peripheral Vision Explorer (PVE) 350, that evaluates the patient's vision in the peripheral zones, and feeds back this information into the simulation of the progressive lenses by the PLS 100.
Finally, the Artificial Intelligence Engine for GPS (AI-GPS) 500 can be included into some embodiments of the GPS 10, for monitoring the performance of the components of the PLS 100 and the LDES 300, and for developing suggested adjustments to improve the performance of the managed components of the GPS system 10. In some detail, the GPS 10 can include an Artificial Intelligence (AI) Engine for the Progressive Lens Design Processor (AI-PLD) 510, to monitor and improve the performance of the PLD 320. Other embodiments can include an AI engine for the Search Guidance Engine (AI-SGE) 520. Finally, some embodiments of the GPS 10 can include an AI Engine for the Progressive Lens Simulator (AI-PLS) 530. Each of the three AI engines 510/520/530 can be configured to monitor the functioning of the corresponding system blocks PLD 320, SGE 330, and PLS 100, and then perform AI-based training cycles to improve the performance of these blocks. In some embodiments, these AI engines are implemented by neural networks.
The above, system-level description of the GPS 10 is now expanded with the detailed description of a number of specific embodiments in more detail. For clarity, the presentation of these embodiments is organized into titled sections.
In some embodiments, the Off-Axis Progressive Lens Simulator OPS 120 can include an Image Generator 121, for generating an image 21; an Off-Axis Progressive Lens Simulator Processor, or OPS processor, 122, for transforming the generated image 21 into Off-Axis PLS signals 20-1 and 20-2 according to according to the progressive lens design 123; and Off-Axis Progressive Lens Simulator Displays 124-1/124-2, for displaying the Off-Axis Progressive Lens Simulation (Off-Axis PLS) 50 according to the Off-Axis PLS signal 20. Here and in the following, many items X are included in the GPS 10 pairwise, one for each eye. They will be typically labeled as items X-1 and X-2. Sometimes, for brevity, the collection of items X-1 and X-2 will be simply referred to as X.
In some PLS 100, the Off-Axis Progressive Lens Simulator Display 124 includes a pair of Off-Axis Progressive Lens Simulator Screens 124-1 and 124-2, each displaying an Off-Axis PLS 50-1 and 50-2, the two PLS together providing a stereoscopic Off-Axis PLS 50 for a first/left eye 1 and a second/right eye 2.
In some PLS 100, the Off-Axis Progressive Lens Simulator Display 124 includes a single stereoscopic alternating Off-Axis Progressive Lens Simulator Screen 124, controlled by an image-alternator, for alternating the displaying the Off-Axis PLS 50-1 for the first eye 1, and subsequently PLS 50-2 for the second eye 2, with suitable stereoscopic adjustments. This rapidly alternating display on left eye-right eye images, with synchronized image-alternation, i.e. blocking the image for the non-targeted eye, allows the use of a single screen to generate stereoscopic images and viewing experience. This image-alternating technology has mechanical embodiments involving shuttering or rotating wheels, opto-electronic embodiments involving rapid polarization changes, and liquid crystal embodiments to block the images. Any one of these embodiments can be utilized in the stereoscopic alternating Off-Axis Progressive Lens Simulator Screen 124.
Analogously, in some embodiments of the PLS 100, the OPS Processor 122 can be configured (1) to receive the generated image 21 from the Image Generator 121; and (2) to transform the generated image 21 into the Off-Axis PLS signal 20 by introducing a locally varying curvature, or swim 127, representative of the progressive lens design 123.
The OPS Processor 122 can perform a detailed ray tracing computation of light emanating from the generated image 21 through the progressive lens design 123 to quantitatively produce this blur 126 and swim 127. In alternate embodiments, wavefront propagation methods can be used by the OPS Processor 122. Generating the correct blur 126 and swim 127 are key functions of the OPS Processor 122 to generate the life-like Off-Axis Progressive Lens Simulation (PLS) 50 at least because of the followings. When a patient evaluates the visual experience of the particular progressive lens design 123, the primary positive experience is the customized increase of the optical power from OPd in the distance region to OPn in the near region, while the primary negative experience is “the price of the positives”, the concomitant blur 126 and swim 127, induced by the power progression. The GPS 10 simulates different progressive lens designs 123 for a patient. The search for the optimal progressive lens design 123 is performed by the patient evaluating the balance of the positives against the negatives of the Comprehensive Progressive Lens Simulations 30 of the individual simulated designs 123, eventually identifying his/her most preferred design 123. The OPS Processor 122 crucially helps this search process by the most lite-like simulation of the blur 126 and swim 127 of the designs 123. In some PLS 100, at least two of the Image Generator 121, the Off-Axis Progressive Lens Simulator Processor 122, and the Off-Axis Progressive Lens Simulator Display 124 can be integrated.
The Vergence-Distance Simulator VDS 140 can be configured to simulate the vergence for the displayed Off-Axis PLS 50 at the gaze distance by (1) moving a screen of the Off-Axis Progressive Lens Simulator Display 124 dominantly laterally, or by (2) shifting the displayed Off-Axis PLS 50 on the Off-Axis Progressive Lens Simulator Display 124 dominantly laterally. In the latter embodiment, the Off-Axis PLS 50 is typically displayed only on a portion of the OPS display 124, thus leaving room to electronically moving the image of the Off-Axis PLS 50 laterally. Some PLS 100 includes the combination of (1) and (2). Other solutions can be used as well, such as rotating mirrors in the optical path of the Off-Axis PLS 50.
With modern opto-electronic techniques, the above described simulators can be integrated to various degrees. In some PLS 100, at least one of the Off-Axis Progressive Lens Simulator Processor 122, the Off-Axis Progressive Lens Simulator Display 124, and the Axial Power-Distance Simulator ADS 130 can include at least one of the Vergence-Distance Simulator 140 and the Zoom-Distance Simulator 150. In some cases, only the VDS processor 142 or the ZDS processor 152 can be included.
Next, the description turns to various embodiments of the Axial Power-Distance Simulator ADS 130. In general, the ADS 130 can be an adjustable optical system that has an adjustable optical refractive power. This adjustable optical refractive power of the ADS 130 can be adjustable to be consistent with the gaze distance, determined by the Eye Tracker 110. In other embodiments, the adjustable optical power of the ADS 130 can be adjusted to an intended gaze distance, such as when the patient, or an operator of the GPS 10 decides to explore and test vision at a different distance.
In some embodiments, the ADS 130 uses optical elements such as lenses and mirrors whose optical power is adjustable, but whose position is fixed. In other embodiments, the ADS 130 can use optical elements whose position is also adjustable to simulate a vergence corresponding to the eye axis direction, or visual axis. A simple embodiment of the ADS can include a pair of adjustable lenses or mirrors that are laterally translatable, or rotatable, to increase the life-likeness of the simulation of the vergence.
Many other embodiments of the ADS 130 exist, including a shape-changing lens, an index-of-retraction-changing lens, a variable mirror, a deformable optic, an aspheric lens with adjustable aperture, a liquid crystal optical element, an adjustable reflective optical element, an adjustable opto-electronic element, and an optical system with an optical component with adjustable relative position.
In the PLS 100 of
The Bye Tracker 110 can further include infrared (IR) telescopes 113-1 and 113-2, with infrared (IR) cameras 114-1 and 114-2, to detect the infrared eye-tracking beams and the infrared imaging light 11-1 and 11-2, reflected from the eyes 1 and 2. The IR cameras 114-1 and 114-2 then generate the eye-tracking images 14-1 and 14-2, and send them to the eye-tracking processor 115. The eye-tracking processor 115 can process and analyze these eye-tracking images to generate eye-tracking image/data 15-1 and 15-2, or jointly 15. In some detail, the IR beams of the IR LEDs 112 are reflected as Purkinje reflections, or Purkinje spots, that reflect from the various surfaces of the eye, starting with the cornea. Tracking these Purkinje spots delivers pin-point information to track the eye position and orientation. The IR light source 111, on the other hand, generates a wide-angle IR light that can be used to image the entire frontal region of the cornea. The Eye Tracker 110 can use both the pin-point information from the reflected Purkinje spots, and the wide-angle image from the reflected imaging light (together referenced as 11) to develop the comprehensive eye-tracking image/data 15.
In the embodiment of
In operation, the OPS Processor 122 can receive the generated image 21 from the image generator 121, adjust it to generate the Off-Axis Progressive Lens Simulation signals 20-1 and 20-2, and send these Off-Axis PLS signals 20 to the OPS Displays 124-1 and 124-2, so that the OPS Displays 124 generate the Off-Axis PLS 50-1 and 50-2.
As shown, in some embodiments, the VDS processor 142 and the ZDS processor 152 can be integrated in the OPS Processor 122. In these embodiments, the Off-Axis PLS signal 20 also includes vergence and zoom components. The vergence component can instruct the VDS actuators 144-1 and 144-2 to laterally move, or rotate, the OPS Displays 124-1 and 124-2, in order to simulate the needed vergence. In these embodiments, the Off-Axis PLS 50 includes Vergence and Zoom, as indicated.
It is noted that the infrared-transmissive visible mirrors 146-1 and 146-2 reflect visible light, while transmitting infrared light. Therefore, mirrors 146 are configured to reflect the Off-Axis PLS 50 towards the eyes, while transmitting the reflected infrared eye tracking beam and the infrared imaging light 11-1 and 11-2, from the eyes.
In the described embodiment of the PLS 100, the OPS Display screens 124-1 and 124-2 can be positioned peripheral to the main optical pathway of the PLS 100, while the infrared telescopes 113-1 and 113-2 of the Eye Tracker 110 can be positioned in the main optical pathway, as shown. In other embodiments, the positioned can be reversed. The mirrors 146 can be IR reflective and visible transmissive, in which case the IR telescopes 113 can be positioned peripherally, while the OPS Displays 124 can be positioned in the main optical pathway, in effect trading places.
(a) tracking 102m of an eye axis direction by an Eye Tracker 110 to determine a gaze distance of the eye;
(b) generating 103m an Off-Axis Progressive Lens Simulation (Off-Axis PLS) 50 by an Off-Axis Progressive Lens Simulator OPS 120, including blur and swim, according to a progressive lens design 123;
(c) creating 104m a Comprehensive Progressive Lens Simulation (Comprehensive PLS) 30 from the Off-Axis PLS 50 by simulating a progressive lens power in the eye axis direction by an Axial Power-Distance Simulator ADS 130;
(d) shifting 105m the Off-Axis PLS 50 Vergence-Distance Simulator VDS 140 to vergence appropriate for the gaze distance; and
(e) zooming the Off-Axis PLS 50 by a Zoom-Distance Simulator ZDS 150 to simulate transitions of gaze distance.
Various aspects of these steps have been described before in relation to the PLS 100 embodiments of
The generating 103m of an Off-Axis PLS 50 can include the following:
(a) generating an image 21 by an Image Generator 121;
(b) transforming the generated image 21 into an Off-Axis PLS signal 20 by an Off-Axis Progressive Lens Simulator Processor 122 according to the progressive lens design 123; and
(c) displaying an Off-Axis PLS 50 according to the Off-Axis PLS signal 20 by an Off-Axis Progressive Lens Simulator Display 124.
Various aspects of these steps have been described before in relation to the PLS 100 embodiments of
As described earlier, the displaying can include providing a stereoscopic Off-Axis PLS 50-1 and 50-2 for the first eye 1 and the second eye 2 by a pair of Off-Axis Progressive Lens Simulator Displays, or Screens 124-1 and 124-2.
In other embodiments, the displaying can include alternating the displaying the Off-Axis PLSs 50-1 and 50-2 for the first eye 1, and subsequently for the second eye 2, with suitable stereoscopic adjustments, by a stereoscopic alternating Off-Axis Progressive Lens Simulator Screen 124, that is controlled by an image-alternator.
In some embodiments, as shown in
In some embodiments, as shown in
In some embodiments, at least two of the Image Generator 121, the Off-Axis Progressive Lens Simulator Processor 122, and the Off-Axis Progressive Lens Simulator Display 124 can be integrated.
In some embodiments, as shown in
In some embodiments, the simulating a vergence can include moving a screen of the Off-Axis Progressive Lens Simulator Display 124 dominantly laterally; and shifting the displayed Off-Axis PLS 50 on the Off-Axis Progressive Lens Simulator Display 124 dominantly laterally.
In some embodiments, as shown in
In some embodiments, at least one of the Off-Axis Progressive Lens Simulator Processor 122, the Off-Axis Progressive Lens Simulator Display 124, and the Axial Power-Distance Simulator ADS 130 can include at least one of the Vergence-Distance Simulator VDS 140 and the Zoom-Distance Simulator ZDS 150.
In some embodiments, the simulating a progressive lens power (within the creating 104m) can include adjusting an optical refractive power of an adjustable optical power system 131 of the Axial Power-Distance Simulator ADS 130.
In some embodiments, the adjusting can include adjusting the optical refractive power of the Axial Power-Distance Simulator ADS 130 to be consistent with the determined gaze distance. In some embodiments, the adjusting can include adjusting the Axial Power-Distance Simulator 130 to simulate a vergence corresponding to the eye axis direction.
In some embodiments, shown in
In some embodiments, as shown in
In this section, another embodiment of the Progressive Lens Simulator 100 will be described. Numerous elements of this embodiment have been already described in relation to
In some embodiments, the PLS 100 can include an Image Generator 121, for generating an image 21; and a Progressive Lens Simulator Processor 122, for transforming the generated image 21 into a Comprehensive PLS signal 20-1 and 20-2 according to the progressive lens design 123, and for coupling the generated PLS signal 20-1 and 20-2 into the Integrated Progressive Lens Simulator 200 for creating the Comprehensive PLS 30-1 and 30-2. In some embodiments, the Image Generator 121 and the Progressive Lens Simulator Processor 122 can be integrated into the IPLS 200.
Just as the multistage PLS 100 could include a single OPS display 124, or a pair of OPS displays 124-1 and 124-2, the PLS 100 of
In the single IPLS 200 embodiments, the IPLS 200 can be a stereoscopic alternating Integrated Progressive Lens Simulator 200, controlled by an image-alternator, for alternating the generating the Comprehensive PLS 30-1 for the first eye 1, and subsequently 30-2 for the second eye 2, with suitable stereoscopic adjustments.
In some embodiments, similarly to
In some embodiments, similarly to
In some embodiments, similarly to
In some embodiments, similarly to
The Integrated Progressive Lens Simulator 200 can be configured to simulate a primary function of the ADS 130; the optical power of the progressive lens design 123 in the eye axis direction by creating the Comprehensive PLS 30 with light rays having a vergence related to the gaze distance. As described before, the simulated optical power can be selected by combining the simulation of the distance of the viewed object with the simulation of the axial power of the progressive lens design 123.
Various embodiments of the IPLS 200 will be described next, in relation to
An aspect of the embodiments in
The IPLS 200 can further include a second/X-scanner hinge system 206h, optionally embedded in the first/Y-scanner hinge system 203, for reflecting and scanning the projected light by the XY-scanning mirror 206 in two spatial dimensions as the XY-scanned light 209. This X-scanner hinge system 206h can be driven by various embodiments of electro-mechanical actuators. The scanning speed of the MEMS Laser Scanner 201 can be high enough so that it projects the Comprehensive PLS 30 with a high enough refresh rate that the patient perceives it as realistic. Helpfully, the images used in evaluating vision, are typically static, or only slowly moving, thus the demands on the refresh rates and therefore scanning, rates are lower than, e.g. in fast-action video games or live TV.
In the embodiments of
(a) tracking 102i an eye axis direction by Eye Tracker ET 110 to determine a gaze distance;
(b) creating 103i a Comprehensive Progressive Lens Simulation (PLS) by the Integrated Progressive Lens Simulator IPLS 200, by simulating an effect of an Off-Axis Progressive Lens Simulator OPS 120, and an effect of an Axial Power-Distance Simulator ADS 130;
(c) shifting 104i the Comprehensive PLS by a Vergence-Distance Simulator VDS 140 to vergence appropriate for the gaze distance; and
(d) zooming 105i the comprehensive PLS by a Zoom-Distance Simulator ZDS 150 to simulate transitions of the gaze distance.
In some detail, embodiments of the PLS 100 can include an Eye Tracker 110, for tracking an eye axis direction to determine a gaze distance; and the Integrated Progressive Lens Simulator 200, for creating a Comprehensive Progressive Lens Simulation (PLS) 30 by simulating a progressive lens power in the eye axis direction, in combination with generating an Off-Axis progressive lens simulation (Off-Axis PLS) 50. In these embodiments, the Eye Tracker 110 and the Integrated Progressive Lens Simulator 200 can be implemented in a head-mounted display, virtual reality viewer, or goggles. Next, two specific embodiments of the head-mounted PLS 260 are described in more detail.
The field of virtual reality goggles is rapidly expanding. These goggles are more and more capable of generating lifelike visual experiences, and therefore their technology can be promisingly implemented and adapted for use in embodiments of the head-mounted PLS 260 to create the Comprehensive PLS 30.
In particular, the light field effect generated by the head-mounted PLS 260 can be three dimensional, or four dimensional. The latter (4D) technology also represents the depth of focus perception, making objects that are in front or behind the object plane blurrier. Doing so further enhances the life-likeness of the visual perception.
Some embodiments of the PLS 100 can use not only XYZ position sensors but XYZ motion sensors 263. Picking up not only the position and direction of the head-mounted PLS 260/270, but also sensing a motion of a wearer of the head-mounted PLS 260/270 cap be integrated into the control software that runs on the driver electronics 276, or in a separate computer. The XYZ motion sensor 263, or motion sensors can include at least one of an accelerometer, a gyroscope, and a magnetometer. These can all contribute to sensing, the motion, position and direction of the user's gaze.
The mics lens array head-mounted PLS 280 can further include binocular viewing lenses 285; and an eye tracker 287. As before, the eye tracker 287 can be an embodiment of the Eye Tracker 110. In some eases, the eye tracker 287 has to be able to work at high angles, or with the help of an IR reflective, visible transmissive mirror, similar to the mirror 146 in
This embodiment can further include at least one XYZ position, direction, or motion sensor 263, for sensing a position, direction, or motion of a wearer of the head-mounted PLS 280. This sensor can sense the position, the direction, or the motion of the head-mounted PLS 280, thereby aiding the generation of the Comprehensive PLS 30.
All embodiments of
Finally, in some embodiments, head-mounted PLS 260/270/280 can involve augmented reality glasses, wherein the Comprehensive Progressive Lens Simulation 30 is generated from an image viewed via the augmented reality glasses.
As mentioned in the introductory part,
The LDES 300 can further include a Progressive Lens Design processor 320, coupled to the Progressive Lens Simulator 100, for modifying the progressive lens design 123 in response to the Visual Feedback 430, and transmitting the modified progressive lens design 123 to the Progressive Lens Simulator 100 to generate a modified Comprehensive Progressive Lens Simulation 30 fur the patient with the modified progressive lens design 123. The progressive Lens Design Processor 320 can be part of, or even integrated into, the OPS 120.
The LDES 300 can further include a Feedback-Control Interface 310, coupled to the Progressive Lens Design processor 320, for receiving the Visual Feedback 430 from an operator, selected from the group consisting of a joystick, touchpad, a mouse, an audio interface, an external tablet GUI, and an internal visual-interface GUI, and forwarding the received Visual Feedback in the form of a feedback-control signal 311 to progress the Progressive Lens Design processor 320. Other embodiments may include the Eye tracker 110, coupled to the Progressive Lens Design processor 320, for receiving a Visual Feedback 430 in a form of an objective patient vision measurement. In other embodiments, other systems can provide objective feedbacks, including Purkinje-spot based imagers, Scheimpflug systems, and OCT systems.
Also, the Progressive Lens Design Processor 320 can base some of its calculations of eye modeling. This may involve imaging some of the ophthalmic layers in the eye and then building an eye model like the widely used Holladay model.
Several modes of operation of the embodiments of the LDES 300 have been contemplated regarding the feedback. (1) Some LDES 300 are operated by the patient himself/herself. The PLS 100 generates the Comprehensive PLS 30 for the patient, who evaluates the visual experience and directly enters a subjective feedback in FCI 310.
(2) In other embodiments of LDES 300, the Visual Feedback may only be indirect. The patient may only express a verbal feedback, such as the last modification made the visual experience of the Comprehensive PLS 30 better or worse, and a trained operator, technician, or the optometrist herself/himself may enter a control signal 311 via the FCI 310.
(3) Other LDES 300 can be based on objective patient feedback and do not require an active, or subjective, patient feedback. For example, the Eye Tracker 110 can monitor the patient's eye movements and draw conclusions from the monitoring. E.g. if the jitter of the patient's eye increases, or the patient struggles to focus in response to a modification of the Comprehensive PLS 30, then the Eye Tracker 110 may report this to the Progressive Lens Design Processor 320 of the LDES 300. In response, a software of the Progressive Lens Design Processor 320 may conclude that the modification was undesirable, and undo the modification, or try a different one.
(4) Finally, in some embodiments of LDES 300, the objective patient feedback, or objective patient vision measurement, such as the rapidity of the patient's eye movement, or inability to focus, may be monitored not by a computer software but by the operator of the LDES 300, such as the optometrist herself/himself without an express, or subjective cooperation of the patient. In such embodiments, the monitoring operator can enter the feedback or control into the FCI 310 that transforms it into a feedback-control signal 311.
The combination of the PLS 100 and LDES 300, described next, can be operated in any of the above four modes, or in some combination of these modes. Sometimes the patient, technician, optometrist, or some combination of more than one of these possible sources providing a feedback or a control input together will be referred to as “the operator”. Also for this reason, the inputs into the FCI 310, and the signal 311 outputted by FCI 310 can be feedback, control, and any combination of feedback and control input and feedback and control signals. These possibilities and combinations will be inclusively referred to as feedback-control input and feedback control signal 311.
It is a non-obvious, challenging task to “translate” the feedback into an actionable command on how to modify the progressive lens design 123 in response to the feedback. Several embodiments will be described next to carry out this translation, in response to the feedback, or “Visual Feedback”. The method of exploring and changing the progressive lens design in response to a Visual Feedback in general will be referred to as a method 400 of Progressive Lens Simulation. Several embodiments of this method will be described next.
(a) activating 401 a progressive lens design 123 with Design Factors 420 by a Progressive Lens Design Processor 320;
(b) generating 402 an image 21 by an Image Generator 121 of a Progressive Lens Simulator 100;
(c) generating 403 a Comprehensive Progressive Lens Simulation (PLS) 30, simulated from the generated image 21 by the Progressive Lens Simulator 100, utilizing the progressive lens design 123;
(d) acquiring 404 a Visual Feedback 430, responsive to the generating of the Comprehensive PLS 30 with the progressive lens design 123;
(e) modifying 405 the progressive lens design 123 by the Progressive Lens Design Processor 320 in relation to the Visual Feedback 430; and
(f) re-generating 406 the Comprehensive PLS 30 with the modified progressive lens design 123 by the Progressive Lens Simulator 100.
The method 400 can typically involve repeating steps (d)-(e)-(f) until the Visual Feedback 430 indicates a satisfactory outcome of the method.
The activating 401 the progressive lens design 123 with a Progressive Lens Design Processor 320 can be in response to a progressive lens design selection based on a preparatory measurement. The generating the image 21 by an Image Generator 121 can be in response to an image selection. Embodiments of the activating 401 are broadly understood. The activating 401 of the progressive lens design 123 can include recalling the progressive lens design 123 from a memory, or from a storage medium. The activating 401 can also include that the Progressive Lens Design Processor 320 computes or models the progressive lens design 123 from some model parameters. Whichever way the progressive lens design 123 is activated, the Progressive Lens Simulator 100 can be generating 403 a Comprehensive Progressive Lens Simulation PLS 30 from the generated image 21 by utilizing the activated progressive lens design 123.
The above steps of the method 400 are typically performed by the PLS 100 and the LDES 300. FIG.18B illustrates steps of a method 410 of an operator interacting with the LDES 300 to carry out the method 400. The method 410 can include the following steps.
(a) optionally selecting 411 a progressive lens design 123 with a Progressive Lens Design Processor 320, based on a preparatory measurement;
(b) optionally selecting 412 an image 21 with an Image Generator 121;
(c) evaluating 413 a visual experience of a generated Comprehensive PLS 30, simulated by a Progressive Lens Simulator PLS 100 from a selected image 21 with a progressive lens design 123;
(d) the evaluating optionally including inspecting 414 an image region of generating a Comprehensive PLS 30 in an inspection direction; and
(e) providing 415 a Visual Feedback 430 to a Progressive Lens Design Processor 320 based on the evaluating via a Feedback-Control Interface 310.
Some sign factors can be supplemented by preparatory measurements by the optometrist. One example is that the optometrist can observe the glass-positioning height, where the patient is wearing the glasses on her/his nose. The optometrist can shift the height of the optical center 422 Design Factor to account for the patient's individual glass-positioning height.
(a) A subjective patient feedback via a Feedback-Control interface;
(b) an objective patient vision measurement;
(c) an eye tracker image/data from an Eye Tracker;
(d) a direct patient feedback;
(e) an indirect patient feedback;
(f) an operator control input;
(g) an operator command;
(h) an operator response to a proposition; and
(i) an operator selection.
The term “Visual Feedback” is used broadly in this document. It can include subjective feedbacks, like the patient expressing a subjective evaluation of the visual experience. It can be an objective feedback, like a measurement of a jitteriness of the patient's visual axis. It can be a direct feedback directly entered by the patient into the FCI 310. It can be an indirect feedback, the patient verbally stating an experience or preference and an operator entering the corresponding feedback into the FCI 310. The feedback can come from a single operator, such as the patient, or from more than one operator, such as the patient entering a partial visual feedback and the operator entering another complementary feedback. Also, the Visual Feedback 430 can be simply a feedback, or a control, or command input, where the operator translated the visual experience into a control, command, selection, or response.
In each of the above classes, there can be a large number of specific Visual Feedbacks 430. To expand by examples, the subjective Visual Feedbacks can include the patient subjectively perceiving the lower nasal quadrant blurry the lower temporal quadrant blurry, the nasal progression region has too much swim, progression corridor too long, progression corridor too wide. The subjective Visual Feedback 430 can include requests, commands, or other control input, such as the patient requesting rotate cylinder orientation angle, increase prism and decrease prism progression. The objective Visual Feedback 430 can include the optometrist observing that the patient's visual inspection, direction, or, visual axis, being too jittery, or the patient having difficulty focusing on the presented object, and inputting a corresponding feedback-control input into the FCI 310. All of these possibilities are included in the scope of “Visual Feedback 430”.
An important role of this Visual Feedback-to-Lens Design Transfer Engine FLE 325 is to “translate” the received Visual Feedback 430 into a modification of the Design Factors 420. In an example, if the patient inputs the Visual Feedback 430 that “lower nasal region too blurry” into the Feedback Control Interface 310, it is a non-obvious task to translate this specific Visual Feedback 430 into which Design Factors to change to what degree to reduce the blurriness in the lower nasal region. A large amount of optical modelling, ray-tracing, patient-testing, and refinement are needed to establish that which set of Design Factors 420 need to be modified to what degree to respond to the various Visual Feedbacks 430. This complex knowledge is embodied, managed and carried out in the Visual Feedback-to-Lens Design Transfer Engine FLE 325.
ΔDF(n)=DF(n+1)−DF(n) (1)
Visibly, the ΔDF(n) increment vectors form a search path in the Design Factor space. As the combined PLS 100 and LDES 300 system performs the method 400, in interaction with an operator, patient, or optometrist according to the method 410, this search path converges to a customized optimal Design Factor vector DF 420 that best suits the patient's needs.
The elements of the VFDF matrix 326 represent how to translate the elements of the Visual Feedback vector 430 into a change of the Design Factor vector 420. Typically, more than one element of the DF vector 420 need to be changed in a correlated manner. These correlations that translate the inputted Visual Feedback vector VF 430 into a design factor vector DF 420 constitute the elements of the VFDF matrix 326.
(a) reversing a search path in a Design Factor space;
(b) reverting to a preceding bifurcation in a search path in a Design Factor space;
(c) jumping to another Design Factor vector;
(d) changing a number of the Design Factors;
(e) fixing a design factor;
(f) changing a speed of performing the method; and
(g) evaluating whether search has been successful.
(a) proposing to an operator to select a Search Management step 450;
(b) receiving a selection of a Search Management step 450 from the operator; and
(c) initiating an execution of the selected Search Management step 450. The initiating 455(c) may involve the Search Guidance Engine 330 instructing the Progressive Lens Design Processor 320 to carry out the selected Search Management step 450.
In these embodiments, the Search Guidance Engine SGE 330 may not simply carry out a Search Management step 450 on its own directive, such as fixing a Design Factor. Instead, it may interactively propose a contemplated Search Management step 450 to the patient and act only upon receipt of a response. By way of an example, the SGE 300 may prompt the patient via the FCI 310: “Should we fix the location of the optical center, and restrict the search to the remaining Design Factors?” in the step 455(a), then receive the selection from an operator, e.g. “Yes” in step 455(b), and initiate carrying the selection, such as reduce the number of Design Factors by one, thereby reducing the dimensions of the Design Factor space by one in step 455(c). In some embodiments, the Search Guidance Engine 330 may offer alternatives: “should we go back to a preceding bifurcation in a search path in a Design Factor space, or should we reverse the search path?”, and carry out the responsive wish of the patient or the operator.
In some embodiments, the Search Management method 455 can involve
(d) the patient storing a selected first lens design 123;
(e) continuing the Lens Design Exploration by any embodiment of the method 400; for example, by reversing the search path by step 450(a), or reverting to a preceding bifurcation by step 450(b);
(f) selecting a subsequent second lens design 123; and
(g) comparing the stored first lens design with the second lens design.
In some embodiments, the Search Guidance Engine SGE 330 may be integrated with the Progressive Lens Design Processor 320.
An aspect of the described embodiments that rely on subjective Visual Feedbacks 430 is that the exploration is not guided by objective merits and measures. In practice, this can, and does, reduce the repeatability of the exploration. When the same patient repeats the same lens design exploration at a later time, it has been observed that she/he sometimes arrives at different lens designs: a potentially unsatisfying outcome.
Repeatability and the soundness of the lens selection can be enhanced by PLS 100 systems acquiring and prioritizing objective Visual Feedbacks 430. One such objective Visual Feedback 430 has been already described by the Eye tracker 110 measuring a jitter of the visual axis. Another class of objective Visual Feedbacks 430 can be generated by integrating diagnostic systems into the PLS 100. Such systems may include Purkinje-spot-based imaging systems, OCT systems, Scheimpflug imaging systems, wavefront analyzers corneal topographers, slit lamps, and related other systems.
Another class of improvements can be introduced by using eye models to evaluate, organize and analyze the diagnostic measurements. Several eye models are known in ophthalmology, often used in preparation for cataract surgery, such as the Holladay and Hoffer models.
The next class of embodiments provides an improvement in this direction. These embodiments define a quantitative metric that can prove quite useful to make the lens design exploration more effective, and lead to repeatable outcomes.
However, in today's practice, this optimization is not interactive; the patient is simply given the end product. There is no visual feedback from the patient during the design process, and no chance for the patient to indicate that the computer-designed progressive lens is far from optimal for the patient's specific needs. Thus, if the patient decides that the visual, experience with the progressive lens is unsatisfactory, then he returns the lens and a time consuming, inefficient back-and-forth process starts between the lens designing company and the patient, typically shipping the glasses back-and-forth, and involving additional grinding of the lens, often taking many weeks, and wasting valuable time from all involved.
The here-described embodiments in general offer a substantive improvement over this state of the art by simulating a progressive lens design 123 in real time in steps 401-403, then acquiring a Visual Feedback 430 from the patient in real time in step 404, and modifying the progressive lens design 123 in real time in step 405, performing all these steps iteratively. The now-described embodiments offer a further improvement by introducing objective lens merit factors 460 and weave these lens merit factors 460 into the modifying the progressive lens design step 405 of the method 400.
These lens merit factors 460 can include;
(a) an improved visual acuity in one of a near and a far vision region;
(b) a reduced astigmatism in one of a near and a far vision region;
(c) a reduced swim in one of a near and a far vision region;
(d) a reduced blur in one of a near and a far vision region;
(e) a suitable progressive region;
(f) an alignment of a cylinder;
(g) a suitable prism; and
(h) a suitable progressive prism.
Many-many more Lens Merit factors can be developed, including more technical ones, such as the integral of a measure of an astigmatism over a vision region, or the coefficient of a Zernike polynomial of the lens, possibly narrowed to a region.
In some existing systems, where the progressive lens design 123 is determined by a lens design code without a Visual Feedback 430, the lens design code optimizes the lens design only by modifying the Design Factors DF1 and DF2 so that the Design Factor vector DF points to the location of the local minimum of the Lens Merit function, shown in
The methods 465, 470 and 475 transcend previously described systems by modifying the one or more Design Factors 420 based on the Visual Feedback 430 in combination with one or more Lens Merit factors 460. Therefore, the methods 465/470/475 transcend the method 440 that primarily relies on the Visual Feedbacks 430 but does not use the Lens Merit 460. These methods also transcend the present progressive lens design codes that rely only on computing Lens Merit 460, but do not use the Visual Feedback 430. Combining the two design-drivers VF 430 and LM 460 is challenging, for several reasons, First, to be able to combine these design-drivers requires the Visual Feedbacks 430 to be quantified, but the quantification of the various subjective Visual Feedbacks 430 is far from obvious. Second, a lot of thoughtful decisions are needed to select the weight factors that combine the VF and LM factors and components. Further, communicating the contemplated lens design updates to the patient in a relatable manner is also a non-trivial task, among others.
Typically, the patient's Visual Feedback VF 430 may request, or prompt, a modification of the Design Factor vector DF 420 that is different from the DF 420 pointing to the local minimum of LM. This can happen for a variety of reasons including the following: (a) The local minimum of LM in the lens design code was determined by averaging feedback from a large number of patients, and thus may not be optimal for any specific patient, (b) The search may involve optimizing several LM components simultaneously. Compromising between these coupled searches can make a DF 420 optimal for a vector other than the minimum for the coupled search. (c) The exploration uses a combination of the VF 430 and the LM vectors 460 to modify the DF vector 420. The subjective VF 430 inputs may nudge the exploration towards new design goal compromises. Also, in some embodiments, the patient may be offered to shift the mixing and weighing parameters, again moving the local minimum.
In some embodiments, the modifying step 475 may include modifying the Design Factors locally in a Design Factor space by utilizing at least one of a gradient descent method, a conjugate descent method, and a hill-climbing method. In
However,
In the shown example, the lens design exploration may be getting temporarily stuck around a Local Optimum 482 when only local search methods 475 are utilized. It is noted that the optimum can be either a minimum or a maximum. In
(a) Reversing a search path in a Design Factor space;
(b) reverting to a preceding bifurcation in a search path in a Design Factor space;
(c) jumping to another Design Factor vector;
(d) changing a number of the Design Factors;
(e) fixing a Design Factor;
(f) changing a speed of performing the method; and
(g) evaluating whether search has been successful.
As discussed in relation to the analogous method 450, such Search Management methods, or steps 490 can be very helpful when the patient's exploration of lens designs becomes disoriented, or stuck. In other embodiments, their primary value can be that they accelerate the progressive lens design exploration, making it much more efficient. In this method 490 the added factor relative to the method 450 is that a quantitative metric is involved in the form of the Lens Merit 460. The introduction of such a quantitative metric greatly narrows down the search. For example, the large number of possible ways of modifying the Design Factors 420 that need to be evaluated when practicing the methods 440 and 450 can be reduced substantially by narrowing the search to paths that the Lens Merit 460 identifies as locally optimal. In the example of
As before,
(a) proposing to an operator to select a Search Management step;
(b) receiving a selection of a Search Management step from the operator; and
(c) initiating the execution of the selected Search Management step.
As with the interactive method 455 earlier, the method 490 can also be improved by performing it interactively. Instead of the Progressive Lens Design Processor 320 and the Search Guidance Engine 330 of the LDES 300 making choices by some algorithm, the method 495 prompts the patient, or operator, to provide Visual Feedback 420, command or control signal, and to select a Search Management step, thereby accelerating the lens design exploration via executing steps 495(a)-(b)-(c).
In some further embodiments, the modifying the progressive lens design can include modifying the progressive lens design by the Progressive Lens Design Processor 320 utilizing non-linear Visual Feedback-to-Design Factor functional relations, or non-linear Visual Feedback+Merit-to-Design Factor functional relations. These embodiments can be alternatives and complementary of the Matrix methods 445 and 470, where these relations are dominantly linear.
Some embodiments of the PLS 100+LDES 300 system and the method 400 of its operation can include additional elements and steps that make it essentially certain that the overall method 400 will not deliver inferior results for the patient relative to traditional practices. These additional steps can include determining a second, “traditional” progressive lens design that is based on traditional measurements that do not involve simulating the second progressive lens design for the patient and asking for his/her Visual Feedback. This can be followed by generating a Comprehensive PLS 30 with this second, traditional progressive lens design. In such embodiments, the patient gets to experience the progressive lens design the traditional approach delivers, as well as the lens design identified by method 400 with the help of PLS 100+LDES 300. Generating the two differently designed progressive lenses, one with simulation and feedback, the other with traditional measurements, the patient can compare the Comprehensive PLS of these two designs, and select his/her preference. With this extension of the method 400, the patient cannot end up with a lens worse than the traditional progressive lens, as he/she can elect the traditional progressive lens anytime, even at the very end of the office visit, if he/she is unsatisfied with the lens design determined by the method 400.
In some cases, the generating 402 an image 21 by an image Generator 121 can be in response to an image selection. This additional step can again improve patient satisfaction, as the patients can select images that are relevant for their activities, instead of the well-known rows of letters which do not capture life-like applications. For example, a team sports person may need extra emphasis on peripheral vision, and thus can select an image 21 that has notable features located peripherally. A long-distance truck driver may select moving images, as she/he may prioritize optimizing her/his vision for moving images. A night guard may select images with low light conditions, as she/he may need to optimize his vision for low light circumstances. Offering to test a patient's vision on use-relevant images can further improve the customization and thereby the satisfactory outcome of the method 400.
Translating the Visual Feedbacks 430 into the Design Factors 420 most effectively is a substantial challenge. The various methods 400-495 offer promising embodiments. To make these embodiments the most efficient, their many parameters need to be optimized. These include the elements of the Visual Feedback-to-Design Factor matrix 326 and the elements of the Visual Feedback+Lens Merit-to-Lens Design matrix 329. For other embodiments, the possibly non-linear connections between the Visual Feedbacks 430 and the Design Factors 420 need to be tabulated and parameterized. Also, the Search Guidance Engines 330 are best operated on the basis of past experiences. The Search Management methods 450/455 and 490/495 can also be improved by learning from the lessons of the preceding explorations. Finally, performing the local modification steps 475 and non-local modification steps 480 can be done most efficiently by utilizing the lessons of past searches.
Recently, entirely new ways of learning from past experiences have been automated in remarkably effective new ways. In various technical treatises, these ways are called Artificial Intelligence, Deep Learning, Machine Learning, or Neural Networks. Other names are also used to capture roughly the same learning methods. The embodiments in this section integrate Artificial Intelligence-based learning methods and systems into the GPS 10 so as to improve and optimize the many parameters of the matrices 326 and 329, the operations of the SGE 330 and the methods of 450/455 and 490/495, as well as the other areas where learning from past experiences is beneficial.
The AI-GPS 500 can include an AI Engine for the Progressive Lens Design Processor (AI-PLD) 510, integrated into, or at least coupled to the Progressive Lens Design Processor 320. The AI-PLD 510 can carry out most of the functions of the Visual Feedback-to-Lens Design Transfer Engine ELF 325, and thus can be viewed as an embodiment of the FLE 325. A function of the AI-PLD 510 can be to perform any one of the many known Deep Learning methods in order to make translating the Visual Feedbacks 430 into modifications of the Design Factors 420 more efficient. As indicated, these Deep Learning methods can repeatedly update and improve the matrix elements of the Visual Feedback-to-Design Factor matrix 326, and the matrix elements of the Visual Feedbacks-Lens Merit-to-Lens Design matrix 329.
In some embodiments, the GPS 10 can also include an AI Engine for the Search Guidance Engine (AI-SGE) 530, integrated into, or at least coupled to the Search Guidance Engine SCE 330. A function of the AI-SGE 530 can be to perform any one of the many known Deep Learning methods with the SGE 330 in order to make it more efficient in guiding the exploration of the progressive lens designs 123 by any of the described methods. These methods include method step 450; performing a Search Management step with the SGE 330; method step 455; performing the method step 450 interactively with a patient; performing the method step 490; performing a Search Management step 330 with the SGE 330 that involves a Lens Merit 460; and method step 495 performing the method step 490 interactively with a patient, among others.
Finally, the AI-GPS 500 can also include an AI Engine for the Progressive Lens Simulator (AI-PLS) 550, integrated into, or at least coupled to PLS 100, in some case integrated into its OPS processor 122. Embodiments of the AI-PLS 550 can perform any of the known AI Deep Learning methods with the OPS Processor 122 to simulate the progressive lens designs 123 better, learning from the Visual Feedbacks 430 of the patients.
The AI-PLD 510 can include an input layer 515, configured to receive the Visual Feedbacks VFi(n) as inputs. (In what follows, the label 430 will not always be expressly shown for the Visual Feedbacks 430, and the label 420 for the Design Factors 420, in order to avoid clutter). The patient, the operator or an objective feedback generating system, such as the Eye tracker 110, or any other eye imaging system can input the VFi elements of the Visual Feedback vector VF(n). As before, “n” references the nth iteration of the lens design exploration method 400. A Visual Feedback—Design Factor VFDF Neural Network 520 can receive the VFi(n) Visual Feedbacks as inputs. This VFDF Neural Network 520 can be configured to output a proposed update of the Design Factor vector ΔDF(n)=DF(n+1)−DF(n) at an output layer 525. In this sense, the VFDF Neural Network 520 plays an analogous function as the Visual Feedback-to-Lens Design Factor Engine FLE 325.
One of the differences is that the input VF vector 430 is not connected to the output DF vector 420 by a linear VFDF matrix 326. Instead, it is generated by a series of transformations, performed by a series of hidden layers 521, 522, 523, as shown. Each hidden layer can include a set of neurons. In the shown example, the first hidden layer 521 includes neurons N11, N12, . . . N16. In this description, the specific number of hidden layers, and the number of neurons in each layer is for illustrational purposes only. Other embodiments may contain any other number of hidden layers and neurons per layer.
The neurons N1i in the first layer can be coupled to the neurons of the second layer N2j, with coupling strength C(1i−2j).
VF(n)T*CM(1)*CM(2)* . . . *CM(K)=ΔDF(n)T, (2)
when the superscripts T in VF(n)T and ΔDF(n)T indicate vector transposition. This is needed, as in
In words, the output vector ΔDF 420 is related to the input vector VF 430 not by a single linear relation, as in the case of the VFDF matrix 326, but by a product of several bilinear relations. It is noted, that the dimensions of the individual coupling matrices can vary in this product, or chain, of coupling matrices. Selecting these dimensions wisely can enhance the efficiency of the lens design exploration. For example, there are AI methods that advocate initially shrinking the dimensions of the CM(i) matrices, then increasing them back up, forming a sort of tunnel, or spool.
The neurons in each layer can be set up to react to their input in a variety of ways. A simple, widely used algorithm is a sharp threshold input-output relation output=f(input). The inputs from the preceding layer i to a particular neuron N(i+1)f in the (i+1)th layer can be added up, and if this sum exceeds a threshold then the neuron N(i+1)f outputs a value, whereas if the sum does not exceed the threshold T, then the output remains a different, lower number, typically zero. In formula,
N
(i+1)i
=f(ΣjNijCM(i)jf−T), (3)
where f(x) is a sharp switching function of its argument x, typically a step function centered at 0, or a somewhat smoothed step function such a tan h(x). A motivation for this switching function f(x) emerged from the operations of neurons: the inputs from several dendrites that bring in the inputs in the form of stimuli are often added up by the electric charging processes of the neuron. As carried out by the ion pumps across the cell membrane. If the sum of these stimuli, often in the form of a total charge, exceeds a threshold, then the output axon of the neuron fires. If the charging does not reach the threshold, the neuron does not produce an output.
One way to set up such a VFDF Neural Network 520 is Google's TensorFlow machine learning framework. This is a convenient sophisticated framework that can be set up with limited preparation. A designer of the VFDF Neural Network 520 can simply enter the number of hidden layers, the number of neurons in each layer, and the switching functions f(x). Then, the initial coupling constants of the VFDF Neural Network 520 can be entered or defined.
After this initialization, the VFDF Neural Network 520 is driven through many teaching, or learning cycles so that the VFDF Neural Network 520 learns from incorrect, or low fidelity results, and the initial coupling constants “learn” and evolve to reduce the incorrect outputs and produce the correct output in a high percentage of the cases.
Training the AI Engine AI-GPS 500 with this Back-propagation with Gradient Descent (BGD) 510T is just one of the many possible methods to train this VFDF Neural Network 520. Also, using this particular Neural Network implementation of the AI-GPS 500 is just one of the many embodiments of the basic idea to utilize Artificial Intelligence systems and methods to improve the GPS 10. All these combinations and alternatives share the following two basic design principles, and can be viewed as embodiments of the GPS 10 system, as long as they do so. These design principles include the followings.
(1) To fine tune the large number of parameters needed to operate the GPS 10 system, such as the elements of the VFDF matrix 326 or the Visual Feedback+Lens Merit-to-Design Factor matrix 329, utilizing the remarkable power of Artificial Intelligence in any suitable AI-GPS engine 500.
(2) To continue developing and improving these GPS 10 systems as a rapidly increasing number of GPS 10s are installed worldwide. It is envisioned that a centralized system may collect the exponentially increasing amount of data on the many-many lens design explorations performed worldwide daily by a rapidly increasing number of patients. This exponentially increasing “big data” can be analyzed by central AI-GPS engines 500. The results of this analysis can be pushed out from the central system to the individual GPS 10 systems in the form of updates for the on-board search software, including the training of the elements of the matrices 326 and 329, the training of the various steps that can be involved in the Search Management 450/455 and 490/495, and the training of the simulation of the progressive lens designs 123 by the PIS 100s. Any embodiment that shares these two basic drivers is an embodiment of the GPS 10.
(a) generating 611 a Comprehensive Progressive Lens Simulation (Comprehensive PLS) 30 for a patient with a Progressive Lens Simulator 100, based on a progressive lens design 123 with a Design Factor vector DF 420, generated by a Progressive Lens Design Processor 320;
(b) receiving 612 a Visual Feedback vector VF 430 into a Visual Feedback-Design Factor Neural Network 520 of an Artificial Intelligence Engine for the Progressive Lens Design Processor AI-PLD 510, in response to the Comprehensive PLS 30; and
(c) outputting 613 a modification ΔDF of the Design Factor vector DF 420 with the Visual Feedback-Design Factor Neural Network 520, in response to the receiving 612; wherein
(d) coupling matrices CM(i) of the Visual Feedback-Design Factor Neural Network 520 were trained by performing 614 a deep learning cycle.
The method 610, further comprising:
(e) modifying the progressive lens design 123 by the Progressive Lens Design Processor 320 using the modified Design Factor vector DF 420; and
(f) generating a modified Comprehensive PLS 30 by the Progressive Lens Simulator 100, using the modified progressive lens design 123.
Repeating steps (b)-(f) iteratively can serve as the backbone of the exploration of the progressive lens designs, as described in relation to method 400.
The Visual Feedback-Design Factor Neural Network 520 can include layers of neurons Nij, including an input layer 515, one or more hidden layers 521/522/523, and an output layer 525. The neurons Nij can having switching functions f(x), and the neurons Nij can be coupled by the coupling matrices CM(i).
The performing a learning cycle step 614 may include performing the deep learning cycle by Backpropagation with Gradient Descent (BGD) 510T. Performing the deep learning cycle step with BGD 510T can fine tune the elements of the Coupling matrices CM(i)k1 to provide the most reliable translation of the Visual Feedbacks 430 into changes of the Design Factor vector 420.
The performing a learning cycle step 614 may also include using 615 a Lens Merit function LM 460. Using a Lens Merit function LM 460 can enable the AI-GPS engine 500 and specifically the AI-PLD 510 to train and improve the Lens Merit based methods 465-495.
The performing a deep learning cycle 614 can also include evaluating 616 the outputted Design Factor vector DE 420 with an output evaluator 526 in relation to a target Design Factor vector DF 420 corresponding to the inputted Visual Feedback vector VF 430; and training the coupling matrices CM(i) according to the evaluating with a coupling trainer 528.
In some embodiments, the performing a deep learning cycle step 614 can include modifying 617 a software of the Progressive Lens Design Processor 320.
(a) generating 621 a Comprehensive Progressive Lens Simulation (Comprehensive PLS) 30 for a patient with a Progressive Lens Simulator 100, based on a progressive lens design 123 generated by a Progressive Lens Design Processor 320;
(b) receiving 522 a Visual Feedback vector VF 430 into a Visual Feedback-Search Management Neural Network 520-SGE of an Artificial Intelligence Engine for a Search Guidance Engine AI-SGE 530, in response to the Comprehensive PLS 30; and
(c) outputting 623 a Search Management step 450/455, or 490/495, with the Visual Feedback-Search Management Neural Network 520-SGE to the Progressive Lens Design Processor 320, in response to the receiving 622; wherein
(d) coupling matrices CM(i) of the Visual Feedback-Search Management Neural Network 520-SGE were trained by performing a deep learning cycle 624.
Here, and in what follows, the Visual Feedback-Search Management Neural Network 520-SGE will be referenced with the same numerical labels as the Visual Feedback-Design Factor Neural Network 520, only with the sub-label “SGE” appended to it. This is to avoid needless repetition, as the embodiments are largely analogous.
In some embodiments, the method 620 can also include the followings.
(e) modifying the progressive lens design 123 by the Progressive Lens Design Processor 320 prompted by the Search Management step 450/490; and
(f) generating a modified Comprehensive PLS 30 by the Progressive Lens Simulator 100, using the modified progressive lens design 123.
Repeating steps (b)-(f) iteratively can serve as the backbone of the exploration of the progressive lens designs, as described in relation to method 400.
The Visual Feedback-Search Management Neural Network 520-SGE of the AI-SGE 530 can include layers of neurons Nij, including an input layer 515-SGE, one or more hidden layers 521/522/523-SGE, and an output layer 525-SGE; the neurons Nij, having switching functions f(x), and the neurons Nij being coupled by the coupling matrices CM(i).
In some embodiments, the performing the deep learning cycle 624 can include performing the deep learning cycle by Backpropagation with Gradient Descent BGD 510T-SGE.
In some embodiments, the performing a deep learning cycle 624 can include using 625 a Lens Merit function LM 460.
In some embodiments, the performing a deep learning cycle 624 can include evaluating 626 the outputted Search Management step 450/455 or 490/495 with an output evaluator 526 in relation to a target Search Management step corresponding to the inputted Visual Feedback vector VF 430; and training the coupling matrices CM(i) according to the evaluating with a coupling trainer 528.
By way of an example, the Search Management step can be 490(a), the reversing the search path. The decision that at which Visual Feedback VF 430 to reverse the path can be very much a complex decision. For example, referring back to
Referring to
In some embodiments, the performing a deep learning cycle 624 can include modifying 627 a software of the Search Guidance Engine 330.
In some embodiments, the AI-GPS 500 system may include a deep learning unit that uses an eye-model, such as the Holladay or Hofer eye model, wherein the AI-GPS 500 can train the parameters of the GPS 10 based of the eye model.
Finally, very analogous methods can be practiced to train and then operate the AI Engine for the Progressive Lens Simulator (AI-PLS) 550. This AI-PLS 550 can include a Visual Feedback-Lens Simulation Neural Network 520-PLS, with components and elements very analogous shown in
As mentioned earlier, the Artificial Intelligence Engine 500 can use any of the known AI methods, and is not narrowed down to neural networks alone. Other AI methods include Supervised learning methods, Non-supervised learning, Regression analysis-based methods, Clustering, Dimensionality reduction, Structured predictions, Anomaly detection and Reinforcement training. Any one of these AI methods can be implemented in the AI-GPS 500.
One of the substantial benefits of the various embodiments of the GPS system 10, for example, the PLS 100, or the PLS 100 combined with the LDES 300, or the PLS 100 and the LDES 300 combined with the Artificial Intelligence Engine for GPS. AI-GPS 500, together referenced as the 100/300/500 embodiments of GPS 10, is that many of the functions previously executed by an optometrist are now carried out by the automated systems of the GPS 10. Therefore, an optometrist does not have to be continuously involved in the exploration of progressive lens designs 123. The patient himself/herself can engage with the 100/300/500 embodiments of the GPS 10 to execute any version of the exploration method 400, and take all the time he/she needs. The patient can go back to previously identified lens designs 123 and compare them with new ones; can chose different images 21 with the Image generator 121; keep performing Search Management steps 450/455 or 490/495 to retrace search paths, explore other choices, fix a Design Factor 420 to narrow the search, slow down the search in some region of the Design Factor space, and so on. All of these can be performed without active involvement by an optometrist.
From a time-management point of view, this aspect of the GPS 10 systems frees up an optometrist to such a degree that she/he can supervise more than one individual GPS systems 10 simultaneously, thereby eliminating the need of manning each optometry station individually with an optometrist. This aspect can substantially reduce the number of personnel required to service a given number of patients, and thus can be greatly helpful for the business model of the overall optometrist office.
The individual stations can include the Progressive Lens Simulators 720-i that can be any embodiment described earlier in this application, including the multistage PLS 100, the integrated IPLS 200, a table-top embodiment of a PLS 100, and the head-mounted PLS 260, this latter embodiment being shown in
Some elements of the PLS 720-i can be implemented in the Central Supervision station 710. The Central Supervision Station 710 can be in communication with the Progressive Lens Simulators 720-i, for providing supervision for an operation of the individual Progressive Lens Simulators 720-i.
This communication can take place via the two-way communication channels 730-i between the individual Progressive Lens Simulators 720-i and the Central Supervision Station 710; the Progressive Lens Simulators 720-i can inform the Central Supervision Station 710 about simulations of progressive lens designs 123; and the Central Supervision Station 710 supervising the simulation by the Progressive Lens Simulators 720-i. The communication channels 730-i can be wired communication channels or wireless communication channels.
In some embodiments, the Progressive Lens Simulators 720-i can individually comprise Lens Design Exploration Systems LDES 300, for guiding an exploration of progressive lens designs 123.
In other embodiments, the Central Supervision Station 710 can comprise a centralized Lens Design Exploration System LDES 300, for guiding an exploration of progressive lens designs, and communicating corresponding guidance signals to the individual Progressive Lens Simulators 720-i.
In some embodiments, the individual Progressive Lens Simulators 720-i can include dedicated individual Artificial Intelligence (AI) Engines for executing a deep learning method for Progressive Lens Design Processors 320-i of the Progressive Lens Simulators 720-i. These AI Engines can be embodiments of the AI Engines for the Progressive Lens Design Processors (AI-PLD) 510.
In other embodiments, the Central Supervision Station 710 can comprise a centralized Artificial Intelligence Engine 500, for executing a deep learning method for the Progressive Lens Design Processors 320 of the Progressive Lens Simulators 720-i, and communicating corresponding training signals to the individual Progressive Lens Simulators 720-i.
In some embodiments, the Progressive Lens Simulators 720-i can individually comprise Artificial Intelligence Engines, for executing a deep learning method for Search Guidance Engines 330 of the Progressive Lens Simulators 720-i. These AI Engines can be embodiments of the AI Engines for the Search Guidance Engines (AI-SGE) 520.
In other embodiments, the Central Supervision Station 720-i can comprise a centralized Artificial Intelligence Engine 500, for executing a deep learning method for a centralized Search Guidance Engine, and communicating corresponding guiding signals to the individual Progressive Lens Simulators 720-i. The Search Guidance Engine can be also centralized, or can reside in the individual PLS 720-i as individual SGEs 330-i.
Similarly, an AI Engine can be included for executing a deep learning method for the Off-Axis PLS 120. As before, this AI Engine can be centralized, or can reside with the individual PLS 720-i.
While this document contains many specifics, these should not be construed as limitations on the scope of an invention or of what may be claimed, but rather as descriptions of features specific to particular embodiments of the invention. Certain features that are described in this document in the context of separate embodiments can also be implemented in combination in a single embodiment. For example, to structure the presentation more clearly, the description of the embodiments was organized into eight sections. However, the features of the embodiments in any one of these sections can be combined with features and limitations of any embodiments from the other seven sections. Conversely, various features that are described in the context of a single embodiment can also be implemented in multiple embodiments separately or in any suitable subcombination. Moreover, although features may be described above as acting in certain combinations and even initially claimed as such, one or more features from a claimed combination can in some cases be excised from the combination, and the claimed combination may be directed to a subcombination or a variation of a subcombination.
The present patent application is a continuation of U.S. patent application Ser. No. 15/984,397, entitled: “Progressive lens simulator with an axial power-distance simulator”, by G. T. Zimanyi, filed May 20, 2018, which application is hereby incorporated in its entirety by reference.