The present invention concerns a method of operating a group of one or more devices on an operating site and of calibrating said devices, wherein: each device is a measurement device and calibrations of said devices are executed based on time schedules determined based on calibration time intervals determined for each of the devices.
Measurement devices are used in nearly all branches of industry. They are commonly used to measure and/or monitor measurement variables, such as e.g. physical quantities related to ongoing production processes or properties of a medium or product. Measurement results provided by measurement devices are e.g. used in process automation for monitoring, controlling and/or regulating a process performed at the operating site. Thus, measurement devices play a vital part in industry as well as in laboratories and non-compliancy of a device to a requirement specified for it, e.g. a specified measurement accuracy, may have severe consequences ranging from impaired production processes, the production of faulty products to potential hazards to people and/or the environment.
To ensure, that measurement devices fulfill the requirements specified for them, they are regularly calibrated. Calibrations are preferably performed on specially designed calibration sites capable of providing reference values of the measurement variable measured by the device with very high accuracy. A typical calibration procedure foresees the determination of a measurement error of the device based on a measured value of the measurement variable determined by the field device and the corresponding reference value provided. In case the measurement error exceeds a maximum permissible error, the device is considered not to conform. As a consequence, adjustments of the measurement indications, repair or replacement of the device is required. If the measurement error does not exceed the maximum permissible error conformity of the field device is declared and generally no further actions are taken.
Calibrations are usually performed periodically after fixed calibration time intervals. As an example, fixed calibration time intervals recommended by the manufacturer of the device can be applied. These intervals are identical for all devices of the same type, regardless of the measurement conditions individual devices may be exposed to during operation. For safety reasons, they are typically set so short, that statistically most devices, e.g. well above 90%, can be expected to be in full compliance at the end of their calibration time interval. Short calibration time intervals increase the costs involved in operating these devices. On the other hand, longer calibration time intervals increase the risk of operating a non-compliant device. In this context, the article “Calibration Intervals, A Manufacturer's Perspective” by David Deaver, Fluke Corporation, published electronically on Dec. 11, 2012 describes a method of determining fixed calibration time intervals based on a compromise between the calibration costs as well as the risk involved in operating non-compliant devices.
As an alternative to fixed calibration time intervals it is known in the art to apply individually determined calibration time intervals, wherein each next calibration time interval is determined for each device individually based on the measurement properties of the respective device determined during the present calibration. A corresponding method is e.g. described in EP 2 602 680 B1. These methods take into account, that devices found to be only just compliant during calibration can be expected to become non-compliant much sooner than devices, that were found to be fully compliant. Thus, shorter calibration time intervals are applied to compliant devices exhibiting a lower degree of compliancy to reduce the risk that these devices may become non-compliant during operation prior to their next calibration, and longer calibration time intervals are applied to compliant devices exhibiting a higher degree of compliancy to reduce the calibration cost.
Since the measurement properties of measurement devices are usually determined during each calibration anyway, individually determined calibration time intervals can be determined at very little extra cost. Scheduling and performing calibrations of measurement devices based on individually determined calibration time intervals does however involve a complex logistic and may cause extra costs and inconveniences. One reason for this is, that it is no longer possible to synchronize calibrations of several devices of the same type to be performed simultaneously. This is especially disadvantageous in applications, where a whole section of a production site has to be shut down, each time one of the devices operating on that section has to be calibrated. This can lead to a significant amount of additional costs in surplus to the costs involved in the performance of the calibrations.
It is an object of the invention to provide a method of operating a group of one or more measurement devices on an operating site and of calibrating said devices which allows to further optimize the calibration time intervals with respect to the risk and the cost involved.
This object is achieved by a method of operating a group of one or more devices on an operating site and of calibrating said devices, wherein: each device is a measurement device and calibrations of said devices are executed based on time schedules determined based on calibration time intervals determined for each of the devices, said method comprising the steps of:
This method has the advantage, that reviewing and updating of the calibration time intervals is performed based on the degree of compliancy individually determined for each of the devices. This allows for the method to profit from the cost reduction potential as well as the risk reduction potential available based on the individually determined recommended intervals. At the same it allows for the procedures predetermined for each of the ratio ranges and the rules applied to determine the updated calibration time intervals to be tailored to the needs, requirements and circumstances prevailing at the operating site. Thus, the flexibility given with respect to the calibration time intervals of devices having a higher degree of compliancy can be used in a way that best suits the operating site, as well as to synchronize calibrations. Also, devices having a lower degree of compliancy are identified and can be treated accordingly to reduce the risk.
Performing the review and the update based on the small number of predefined ratio ranges has the advantage, that the same procedure is applied to all devices falling into the same ratio range. This simplifies the logistics required and facilitates and accelerates the performance of the method.
A first refinement comprises a method, wherein the set of ratio ranges comprises:
A refinement of the first refinement comprises a method, wherein: the procedure performed with respect to all devices exhibiting ratios comprised in the lower ratio range comprises: a first procedure step of performing at least one of: adjusting, servicing and repairing the device, a second procedure step of re-calibrating the device, re-determining the recommended interval and re-determining the ratio based on the re-determined recommended interval, and a third procedure step of determining the updated calibration time interval for the respective device or a replacement device replacing the device.
A refinement of the last mentioned refinement comprises a method, wherein: the procedure performed with respect to all devices exhibiting ratios comprised in the lower ratio range further comprises at least one of the procedure steps of:
A second refinement comprises a method, wherein:
A third refinement comprises a method, further comprising the steps of: repeating the performance of the method for at least one more calibration cycle, by:
Fourths refinements comprise a method, wherein:
A fifth refinement comprises a method, further comprising the steps of:
A refinement of the fifth refinement comprises a method, wherein said indicators comprise at least one of:
A refinement of the fifth refinement further comprises the steps of:
A refinement of the fifth refinement further comprises the steps of: based on at least one of: the first asset values, the second asset values, the third asset values, the fourth asset values, the combined asset values, the scheduled asset values, the implemented asset values and/or at least one of the group-indicators each determined based on one of the first asset values, the second asset values, the third asset values, the fourth asset values, the combined asset values, the scheduled asset values or the implemented asset values performing at least one of identifying, quantifying, providing, displaying and monitoring at least one of: increased risks, increased costs, used or unused risk reduction potentials available according to the recommended intervals and used or unused cost reduction potentials available according to the recommended intervals.
A refinement of the last mentioned refinement further comprises the step of: in case one of said increased risks, increased costs, unused cost reduction potentials and/or unused risk reduction potentials exceeds a given threshold determining at least one root cause root causing the threshold to be exceeded based on the indicators and/or the group-indicators.
A refinement of the last mentioned refinement further comprises at least one of the steps of:
A refinement of the last mentioned refinement further comprises at least one of the steps of:
The invention further comprises a calibration optimization system designed to perform at least one of the method steps of the method according to the fifth refinement, wherein:
The invention and further advantages are explained in more detail using the figures of the drawing.
The invention provides a method of operating a group of one or more devices 1 on an operating site 3 and of calibrating these devices 1.
The calibrations of each of the devices 1 are executed based on time schedules determined based on calibration time intervals A determined for each of the devices 1. With respect to new devices 1 taken into operation for the first time, the calibration time interval A can e.g. be determined as or based on a manufacturer recommendation provided by the manufacturer of the device 1. With respect to devices 1, that have already been operated at the operating site 3 at least once the calibration time interval A is given by the interval currently applied to the respective device 1.
With respect to the calibrations of each of the devices 1 calibration methods known in the art suitable for calibrating devices of the respective type can be applied. These calibrations can e.g. be performed at the operating site 3 or on purpose built calibration sites, as shown in
During each calibration of each of the devices 1 of the group a degree of compliancy of the respective device 1 to a requirement specified for the respective device 1 is determined. As an example, the degree of compliancy to a measurement accuracy specified for the device 1 can e.g. be determined based on a measurement error of the device 1 determined during calibration. Based on the degree of compliancy determined during the calibration of the device 1 a recommended interval R is determined for the respective device 1. To this extent, any method capable of determining individually determined next calibration time intervals after which a specific device shall be calibrated based on its degree of compliancy to a specified requirement determined during the present calibration can be applied. As an example, the method described in EP 2 602 680 B1, incorporated herein by reference, can be applied. According to this method an optimized next calibration time, at which a specific device shall be calibrated next is determined as a time which is earlier or equal to a time, at which a measurement error of the device will exceed a predetermined maximum permissible error. This time is determined based on a Monte Carlo simulation performed based on the measurement errors of the respective device 1 determined during at least two previously performed calibrations and probability density functions for determining a measurement error in the respective calibration solely due to an uncertainty inherent to the respective calibration procedure. Alternatively, other methods providing individually determined recommended intervals R based on the degree of compliancy of the respective device can be applied without deviating from the scope of the invention described herein.
Regardless of the method applied, the recommended intervals R are preferably determined such, that the recommended intervals R determined for devices 1 exhibiting a higher degree of compliancy are longer than the recommended intervals R determined for devices 1 exhibiting a lower degree of compliancy and vice versa. As a result, each recommended interval R is individually determined and recommended intervals R determined during consecutive calibrations of the same device 1 can be different. Each individually determined recommended interval R represents a device-specific optimum length for the next calibration time interval of the device 1 under calibration in terms of reducing the risk that the device 1 will become non-compliant during operation prior to its next calibration, as well as in terms of reducing calibration costs by increasing the lengths of the next calibration time interval to an extent that can be safely applied in view of the risk involved for the device 1 to become non-compliant prior to its next calibration due to its present degree of compliancy.
In difference to the methods described in the prior art, the recommended intervals R determined for each device 1 during each of its calibrations are not applied directly to schedule the next calibrations of these devices 1. Instead, for each calibration a ratio r: =R/A given by the recommended interval R divided by the calibration time interval A(tn−1) currently applied to the respective device 1 is determined. In case the present calibration performed at a calibration time tn is the first calibration of the respective device 1, the currently applied calibration time interval A(tn−1) corresponds to the interval determined for the new device 1 as described above. In case the device 1 has been calibrated at least once before, the currently applied calibration time interval A(tn−1) correspond to the calibration time interval A(tn−1) applied to the device 1 following the previous calibration performed at the previous calibration time tn−1.
In addition, the currently applied calibration time intervals A(tn−1) are reviewed based on the ratios r determined for each of the devices 1 and a predetermined set of ratio ranges I, II, III. This is done by for each device 1 performing a procedure predetermined for the ratio range I, II, III comprising ratios of the size of the ratio r determined for the respective device 1 or a replacement device 1 replacing the device 1 and by updating the calibration time interval A(tn−1) currently applied to the respective device 1 based a rule defined for the respective ratio range I, II, III and/or based on a rule applicable to the respective device 1.
The set of ratio ranges preferably comprises a small number larger or equal to two of non-overlapping ratio ranges I, II, III.
Regardless of the number of ratio ranges I, II, III and the upper and/or lower limits L1, L2 applied, the procedures are preferably determined based on the size of the ratios r comprised in the respective ratio range I, II, III and as an option, preferably also based on requirements prevailing at the operating site 3.
As an example the procedure performed with respect to all devices 1 exhibiting ratios r comprised in the lower ratio range I preferably comprises a first procedure step of performing at least one of: adjusting, servicing and repairing the device 1. Adjusting of the device 1 is preferably performed based on the calibration data obtained during the calibration and can for example include adjustments of: offset, gain and/or span of the measurement indications of the device 1. Servicing can e.g. include at least one of: cleaning of the device 1, performing a visual inspection, replacing at least one part of the device 1 subjected to wear and tear and of performing a special test or an inspection of individual components of the device 1. Repairing of the device can e.g. include a repair or a replacement of at least one faulty component of the device 1. Following this a second procedure step is performed comprising the steps of re-calibrating the device 1, re-determining the recommended interval R and re-determining the ratio r based on the re-determined recommended interval R. As an option, the procedure may require, that the device 1 is replaced when the re-determined recommended calibration time R is shorter than a minimum interval CTImin defined for the device 1. In these cases, the ratio r does not have to be re-determined. As a further option, the procedure may require that the first and the second procedure steps are only applied to devices 1 exhibiting a ratio r exceeding a predefined threshold rmin, and that devices 1 exhibiting a ratio r smaller than this threshold rmin are replaced. As a further option, the procedure may require that the device 1 is replaced in all cases, wherein the re-determined ratio r belongs to the same ratio range I as the previously determined ratio r. Finally, as a third procedure step, the currently applied calibration time interval A(tn−1) is updated based on the rule defined for the lower ratio range I and/or based on the rule applicable to the respective device 1 by determining the updated calibration time interval A(tn) for the respective device 1 or the replacement device 1 replacing the device 1.
The rule applicable to the respective device 1 can e.g. be a rule according to which the updated calibration time interval A(tn) is set equal to the currently applied calibration time interval A(tn−1) or a predetermined fixed calibration time interval predetermined for the respective device 1 or a class of devices 1 comprising the respective device 1. The rule applicable to the respective device 1 can e.g. be applied to all devices 1 the respective rule is applicable to regardless of the ratio r determined for them. In that case, the same rule can be comprised in each of the procedures defined for the ratio ranges I, II, III. Based on rule(s) applicable to the respective device 1, there is always a possibility, that at least one of the thus determined updated calibration time intervals A(tn) may be longer than the recommended interval R determined for the respective device 1.
The rule defined for the lower ration range I preferably requires, that the updated calibration time interval A(tn) is determined based on the re-determined recommended interval R in all cases, where the device 1 has not been replaced. To this extent, the updated calibration time interval A(tn) can e.g. be determined based on a product given by multiplying the re-determined recommended interval R with a predefined constant, e.g. by determining the updated calibration time interval A(tn) to be equal to an interval shorter than the product and longer than the minimum interval CTImin defined for the respective device 1. In addition, this rule may require, that the updated calibration time interval A(tn) is determined as described above with respect to the determination of the initial calibration time interval A for new devices 1, in all cases wherein the device 1 has been replaced by a new replacement device 1.
The intermediate ratio range II comprises ratios r, wherein the recommended interval R and the currently applied calibration time interval A are of the same order of magnitude. Since the recommended intervals R are determined based on the degree of compliancy of the respective device 1, it is sufficient for the procedure defined for the intermediate ratio range II to solely comprise the step of updating the calibration time interval A. The corresponding rule defined for the intermediate ratio range II preferably requires, that the updated calibration time interval A(tn) is determined based on the recommended interval R and/or the currently applied calibration time interval A(tn−1). As an example, the updated calibration time interval A(tn) can be determined to be equal to the currently applied calibration time interval A(tn−1), to be equal to the recommended interval R or to be equal to the longer, the shorter or the average of the currently applied calibration time interval A(tn−1) and the recommended interval R. As an option, the rule defined for the intermediate ratio range II may be defined to additionally account for at least one requirement, restriction or circumstance prevailing at the operating site, and/or at least one special requirement regarding at least one of the devices 1 of the group. Even though the single step of updating the currently applied calibration time interval A based on the rule defined for the intermediate ratio range II and/or based on the rule applicable to the respective device 1 is already sufficient, the procedure defined for the intermediate ratio range II can comprise additional steps, e.g. steps directed to enhance the long-term stability of operating the respective device 1, like e.g. at least one of: servicing the device 1, cleaning the device 1 and a replacement of at least one part subjected to wear and tear.
The higher ratio range III comprises ratios r, wherein the recommended interval R is considerably longer than the currently applied calibration time interval A. Since the recommended intervals R are determined based on the degree of compliancy of the respective device 1, it is sufficient for the procedure defined for the higher ratio range III to solely comprise the step of updating the currently applied calibration time interval A based on the rule defined for the higher ratio range III and/or based on the rule applicable to the respective device 1.
The rule defined for the higher ratio range III preferably requires for each updated calibration time interval A(tn) to be determined to be longer than the currently applied calibration time interval A(tn−1) and shorter or equal to the recommended interval R determined for the respective device 1. As an additional option, the flexibility given by this range extending from the currently applied calibration time interval A(tn−1) to the recommended interval R, is preferably used to optimize the efficiency and cost effectiveness of the performance of the calibrations of the devices 1 of the group. In this respect, the rule defined for the higher ratio range III may be defined to additionally account for at least one requirement, restriction or circumstance prevailing at the operating site, like e.g. the flexibility and availability of calibration times at which calibrations of the respective device 1 can be performed and/or special requirements regarding at least one of the devices 1 of the group. As an example the rule defined for the higher ratio range III may require, that calibrations time intervals A(tn) for all devices 1 or for at least one or more specific devices 1 are always determined to be longer than a given minimum interval CTImin and/or shorter than a given maximum interval CTImax. As another example the rule defined for the higher ratio range III may require that updated calibrations time intervals A(tn) for all devices 1 installed on a certain section of the operating site correspond to time intervals between regular downtimes of this section of the operating site, during which calibrations of devices 1 operating in these sections can be performed without any disturbance of the process performed at the operating site 3. As a further option, the updated calibration time intervals A(tn) can e.g. be determined based on a set of intervals comprising a limited number of intervals of different lengths. In this case, each updated calibration time interval A(tn) is determined to be equal to the longest or one of the longest intervals comprised in the set of intervals that is shorter than or equal to the recommended interval R determined for the respective device 1. This reduces the variety of different interval lengths of the thus updated calibration time intervals A(tn) down to the limited number of lengths comprised in the set of intervals. This is especially advantageous when the group comprises a large number of devices 1, e.g. more than 10 or 100 or even more than 1000 devices 1, because it allows for calibrations of larger numbers of devices 1 to be synchronized.
As an alternative or an additional option, the updated calibration time intervals A(tn) can be determined to be shorter or equal to at least one of: a maximum permissible interval length defined for the respective device 1 and a maximum permissible interval length defined for the operating site 3. This reduces the risk that any one of the devices 1, this rule is applied to, becomes non-compliant during operation prior to its next calibration.
Other rules, or rules given by a combination of a rule defined for one of the ratio ranges I, II, III and a rule applicable to the respective device 1, other sets of ratio ranges and/or other procedures can be applied without deviating from the scope of the invention.
Following the reviewing and the determination of the updated calibration time intervals A(tn) the time schedule providing scheduled calibrations times ts for each device 1, at which the respective device 1 shall be calibrated again is determined based on the updated calibration time intervals A(tn). This renders a scheduled interval S for each device 1 or the replacement device 1 replacing the respective device 1 given by the time difference between the scheduled calibration time tS scheduled for the respective device 1 and the calibration time tn, at which the respective device 1—or its predecessor in case has been replaced—has last been calibrated prior to the scheduled calibration time tS. As an option the scheduled calibration times tS and thus also the scheduled intervals S are preferably determined by additionally accounting for at least one constraint prevailing with respect to at least one of: the specific device 1, the operating site 3 and the calibration site 13 governing the flexibility and the availability of calibration times at which calibrations of the respective device 1 or replacement device 1 can be performed. Finally the next calibrations of the devices 1 are performed according to the previously determined time schedule.
This method of operating and calibrating the group of measurement devices 1 has the advantages mentioned above. The method can be terminated after the calibrations of the devices 1 have been performed according to the previously determined time schedule. Alternatively, the method can be performed repeatedly for as many calibration cycles as desired or required. In that case each cycle comprises the method steps of performing the calibrations, determining the ratios r, reviewing and updating the currently applied calibration time intervals A(tn−1) and of scheduling and performing the next calibrations as described above. Thus, during each cycle following the first performance of the method, the currently applied calibration time intervals A(tn−1) are given by the corresponding updated calibration times intervals A(tn) determined during the previous cycle.
Repeated performance of this method has the advantage, that each calibration cycle contributes to the reduction of the overall calibration cost, due to all those devices 1 for which the updated calibration time interval A(tn) is longer than the currently applied calibration time interval A(tn−1). In addition, each calibration cycle contributes to the reduction of the risk of operating a non-compliant device 1, due to all compliant devices 1 having a lower degree of compliancy that were identified and for which correspondingly short updated calibration time intervals A(tn) have been determined.
In real life, it is not always possible to perform calibrations at the exact points in time given by the scheduled calibration times tS. This can have various reason, like e.g. changes of the workload of technicians performing the calibrations, unexpected delays, e.g. due to downtimes of the calibration site 13, unexpected accelerations, e.g. due to free calibration capacities available at the calibration site 13, or special circumstances occurring at the operating site 3 affecting the performance of the calibration. In consequence, the lengths of the operating time intervals OTI and the corresponding scheduled intervals S are not necessarily identical.
In consequence, the method renders a set of intervals comprising the recommended interval R, the currently applied and the updated calibration time interval A(tn−1), A(tn), the scheduled interval S and the operating time interval OTI following the respective calibration for each calibration of each device 1.
Assuming an optimum performance of each method step and procedures and rules best suited for the application, the differences between the individual intervals comprised in each set are solely due and justified by the circumstances and conditions prevailing at the operating site 3. To give an example, additional costs caused by operating time intervals OTI, that are shorter than the corresponding recommended intervals R can e.g. be justified by the costs saved by the reduction of downtimes of the operating site 3 required to perform the calibrations and/or saved by the synchronization of calibrations of several devices 1 achieved by this. In practice, the determination of the procedures and rules best suited for the application is a complex task. In addition, the logistics involved in performing calibrations, in particular with respect to groups comprising large numbers of device 1, are rather complex. Thus, there is a possibility that the execution of one or more method steps, as well as the predetermined procedures and rules can be further improved to profit from or to realize the full potential of the method, in particular the full risk reduction potential and/or the full cost reduction potential available according to the recommended intervals R. This could be done by amending the performance of at least one method step, at least one rule and/or at least one of the procedures. Identifying presently unused potential of the method and/or determining the amendment(s) required to make better use of the potential is not an easy task. One reasons for this is, that the potential available during each update of the calibration time interval of each device 1 depends on the length of the recommended intervals Rat the time. This lengths not only depends on the long term measurement properties of the respective device 1 but also on the environment conditions the device 1 has been exposed to during operation and unexpected events affecting the measurement properties of the device 1. In consequence recommended intervals R determined for the same device 1 during consecutive calibrations can be different, which makes it very difficult to decide, whether amendments of the method are worthwhile.
As an option the method is therefore further improved by recording a data set k comprising a device identification d of the respective device 1 and at least two, preferably all of: the recommended interval R, the currently applied calibration time interval A(tn−1), the updated calibration time interval A(tn), the scheduled interval S and the operating time interval OTI following the respective calibration, for at least one, preferably for each calibration of at least one, preferably of all of the devices 1 of the group. As an option, each data set k preferably also comprises the corresponding calibration time tn. Further, at least one indicator indicative of the performance of the method is determined based on at least one of the data sets k.
As shown in
Considering the risk and the cost as the function of the operating time t shown in
The indicators preferably comprise at least one asset value V(k) determined for each data set k. Examples of currently preferred asset values, as well as examples of formulas for their determination are shown in
As an additional or alternative option the asset values V(k) preferably comprise a second asset value V2(k) quantifying the increased risk associated with updates, wherein the currently applied calibration time interval A(tn−1) is extended such that the updated calibration time interval A(tn) longer than the recommended interval R. In the example shown in
As an additional or alternative option the asset values V(k) preferably comprise a third asset value V3(k) quantifying the cost reduction potential used by updates, wherein the currently applied calibration time interval A(tn−1) is extended such that the updated calibration time interval A(tn) shorter or equal to the recommended interval R. In the example shown in
As an additional or alternative option the asset values V(k) preferably comprise a fourth asset value V4(k) quantifying the risk reduction potential used by updates, wherein the currently applied calibration time interval A(tn−1) is reduced such that the updated calibration time interval A(tn) is longer or equal to the recommended interval R. In the example shown in
Based on the case discriminations applied in the determination of these asset values V1(k), V2(k), V3(k), V4(k), only one of the four asset values V1(k), V2(k), V3(k), V4(k) is available for each data set k. This has the advantage, that all four asset values V1(k), V2(k), V3(k), V4(k) can be summarized in a combined asset value VC(k) given by the only one of the four asset values V1(k), V2(k), V3(k), V4(k) defined for the respective data set k. These combined asset values VC(k) are preferably displayed in a diagram, shown in
Obviously other ways of determining these asset values conveying the same, an equivalent, a corresponding or a related information content can be applied, without deviating from the scope of the present invention. As an example the unused cost reduction potential can e.g. be applied instead of the used cost reduction potential and the unused risk reduction potential can e.g. be applied instead of the used risk reduction potential.
As an additional option, the indicators preferably comprise at least one scheduled asset value VS1(k), VS2(k), VS3(k), VS4(k), VCS(k) each determined in the same way as one of the first asset value V1(k), the second asset value V2(k), the third asset value V3(k), the fourth asset value V4(k) and the combined asset value VC(k) by replacing the updated calibration time intervals A(tn) applied in the respective determination by the scheduled interval S as shown in
As an alternative or additional option, the indicators preferably comprise at least one calibration value W(k) indicative of the adherence to schedule of the performance of the respective calibration to the time schedule based on which it is performed. Each calibration value W(k) is determined based on the intervals comprised in one of the data sets k based on or as a difference or an absolute value of the difference between the operating time interval OTI and the scheduled calibration time interval S comprised in the respective data set k, e.g. by W(k):=OTI-S. As an alternative option, the calibration values W(k) can be determined as or based on a quotient given by the operating time interval OTI divided by the scheduled interval S comprised in the respective data set k, e.g. by W(k):=OTI/S.
As an alternative or additional option, the indicators preferably comprise at least one other indicator value determined based on the intervals comprised in one of the data sets k. Examples are indicator values determined for at least one or each data set k based on or as a difference or an absolute value of the difference between or a quotient of: 1) the recommended interval R and the currently applied calibration time interval A(tn−1), 2) the recommended interval R and the updated calibration time interval A(tn), 3) the updated calibration time interval A(tn) and the scheduled interval S, or 3) the updated calibration time interval A(tn) and the currently applied calibration time interval A(tn−1).
At least one of the indicators, preferably more or all the indicators listed above are preferably each determined for data sets k recorded during consecutive calibrations cycles of at least one, preferably all of the devices 1 and are preferably also displayed.
As a further option, at least one device-group is preferably defined and at least one group-indicator is determined for at least one of the device-groups. The device-groups preferably comprise at least one of: at least one device-group each comprising a single specific device 1, at least one device-group each comprising devices 1 of a given class or type, at least one device-group each comprising devices 1 operating on a specific section of the operating site 3, like for example a section, that needs to be shut down every time one of the device 1 operating on this section is calibrated, a device-group comprising all devices 1 operating on the operating site 3, and/or at least one other device-group. Each of the group-indicator is preferably determined based on, as an average or an average of the absolute values of one of the previously described indicators determined for each of the data sets k recorded for the devices 1 comprised in the device-group during a given time period Δti. Each of these group-indicators is preferably determined for several consecutive given time periods Δti. The group-indicators are preferably determined, recorded, displayed and monitored and a notification is preferably issued, when one of the monitored group-indicators exceeds a threshold defined for the respective group-indicator.
With respect to each of the asset values V(k) comprising the first asset value V1(k), the second asset value V2(k), the third asset value V3(k), the fourth asset value V4(k), the combined asset value VC(k), the scheduled asset value VS(k) and implemented asset value VI(k), the corresponding group-asset indicator VG is preferably determined by the user of the method or by the calibration optimization system 17 based on or as an average of the respective asset values V(k) determined for the data sets k:=1, . . . , m recorded for the devices 1 comprised in the respective device-group during the given time period Δti, e.g. by:
As shown in
With respect to the calibration values W(k) each group specific calibration indicator WG is preferably determined by the user of the method or by the calibration optimization system 17 as an average of the calibration values W(k) or as an average of the absolute values of the calibration values W(k) determined for the data sets k comprised on the respective group of data sets. Analogously, this determination of group specific indicators can also be applied with respect to the other indicator values mentioned above.
Based on at least one of: at least one of the asset values V(k) and at least one of the group-specific asset indicators VG increased risks, increased costs, used or unused risk reduction potentials and/or used or unused cost reduction potentials available according to the recommended intervals R determined during the respective calibrations are preferably identified, quantified and/or monitored. The unused cost reduction potentials are given by the difference between the full potential available (shown as +100%) and the used cost reduction potentials. The unused risk reduction potentials are given by the difference between the full potential available (shown as −100%) and the used risk reduction potentials. This step is preferably performed on a device level with respect to at least one of the devices 1, on a group-level with respect to devices comprised in one of the device-groups and/or on a site level with respect to all devices 1 operating on the operating site 3. This can be done by the user of the method or by the calibration optimization system 17. In the latter case, the thus determined increased risks, increased costs and/or the used or unused potentials are preferably provided and/or displayed by the calibration optimization system 17.
In case an increased risk, an increased cost and/or an unused potential exceeding a given threshold is identified, at least one root cause root causing the threshold to be exceeded is preferably determined based on the indicators and/or the group-indicators.
As an example, e.g. at least one of the rules and/or at least one of the procedure steps of the procedures predetermined for the ratio ranges I, II, III is preferably identified as root cause when at least one of the increased risks, increased costs, unused cost reduction potentials and/or unused risk reduction potentials quantified based on the first asset values V1(k), the second asset values V2(k), the third asset values V3(k), the fourth asset values V4(k) and/or at least one of the group indicators determined based on the first asset values V1(k), the second asset values V2(k), the third asset values V3(k) and/or the fourth asset values V4(k) exceeds the corresponding threshold. As a further option the scheduling is preferably identified as a root cause when the increased risks, the increased costs, the unused risk reduction potentials and/or the unused cost reduction potentials quantified based on least one of the scheduled asset values VS1(k), VS2(k), VS3(k), VS4(k), and/or at least one of the group indicators determined based on the scheduled asset values VS1(k), VS2(k), VS3(k), VS4(k) exceeds the value of the corresponding property quantified based on the first asset values V1(k), the second asset values V2(k), the third asset values V3(k) and/or the fourth asset values V4(k) by more than a predefined threshold.
As a further option the adherence to schedule of the performance of the calibrations to the time schedule is preferably identified as a root cause when the increased risks, the increased costs, the unused risk reduction potentials and/or the unused cost reduction potentials quantified based on least one of the implemented asset values VI1(k), VI2(k), VI3(k), VI4(k) and/or at least one of the group indicators determined based on the implemented asset values VI1(k), VI2(k), VI3(k), VI4(k) exceeds the value of the corresponding property quantified based on the scheduled asset values VS1(k), VS2(k), VS3(k), VS4(k) by more than a predefined threshold.
Once a root cause has been identified corresponding countermeasures are preferable taken. As an example, the countermeasures can e.g. comprise amending at least one of the rules and/or at least one of the procedure step of the procedures predetermined for the ratio ranges I, II, III that have been identified as root cause, amending the scheduling identified as root cause or amending the calibration performance procedure when adherence to schedule has been identified as root cause.
Thus each of the different types of indicators, as well as the corresponding group specific indicators constitute a powerful tool for monitoring the performance of the method as well as for iteratively improving the implemented method steps to further profit from the full potential of the method in a way that is best suited for the application.
Number | Date | Country | Kind |
---|---|---|---|
10 2019 134 797.1 | Dec 2019 | DE | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2020/084062 | 12/1/2020 | WO |