The present description relates generally to a current reversal mitigation and more particularly to methods and apparatus for mitigation of current reversal in capillary zone electrophoresis-electrospray devices.
As depicted in
As shown in
When the electrospray voltage is not accurately controlled as in the conventional devices, the sensitivity, reproducibility, and detection limit of the instrument suffer. To avoid the current-sinking challenge, separations are performed in low current conditions (i.e. <10 μA). When a large inner diameter capillary (e.g. 50 μm) is used in order to improve the loading capacity, a low separation voltage (typically 400 V/cm or lower) is necessary which limits the analysis throughput. When an improvement the analysis is tried throughput by increasing the separation voltage (e.g. 1,000 V/cm), it is often required to use a very small inner diameter separation capillary (e.g. 10 μm), which significantly limits the loading capacity, leading to low signal intensity and identification capacity. If the current-sinking challenge is mitigated, shorter capillaries with larger inner diameters may be employed while using higher conductivity separation buffers, which enable faster separations and larger loading amounts. Previous solutions to this problem float the capillary electrophoresis circuitry on the electrospray voltage which results in the effective capillary voltage no longer depending on the electrospray voltage. However, this first approach would increase the cost of the setup due to the need for additional safety precautions, such as isolated electronics on the capillary electrophoresis instrument.
The following description of example methods and apparatus is not intended to limit the scope of the description to the precise form or forms detailed herein. Instead the following description is intended to be illustrative so that others may follow its teachings.
Combined capillary-zone electrophoresis-electrospray ionization-mass spectrometry (CZE-ESI-MS) is attracting renewed interest in use for molecular analysis of many biomolecules and other samples. This interest is stimulated by the development of high-sensitivity interfaces for coupling CZE to MS.
Referring now to
The convential capillary electrophoresis instrumentation 102 consists of a separation capillary 112 for transporting a sample with an injection end 114 and a distal end 116, the injection end 114 being inserted into a reservoir 118 containing a background electrolyte. The separation capillary 112 introduces the sample by capillary action, pressure, siphoning, or electrokinetically. A voltage differential is applied across the injection end 114 and the distal end 116, causing the sample to migrate across the capillary 102. A number of forms of capillary electrophoresis exist including capillary zone electrophoresis (CZE), but other electrophoretic techniques including capillary gel electrophoresis (CGE), capillary isoelectric focusing (CIEF), capillary isotachophoresis and micellar electrokinetic chromatography (MEKC). In each, separation occurs when electrokinetic motion outpaces osmotic motion, sorting the ions within the same by a combination of charge and weight.
In a device shown in
Referring now to
Compared to the prior art device shown in
The separation capillary 212 is a submillimeter diameter tube, constructed of fused silica in the example shown, other materials for the capillary walls are contemplated such as polyamide walls. In some versions, like sheathless apparatus 200B, the capillary 212 is coated in an electrical conductor, called a sheathless design. In other versions, the capillary 212 is sheathed using a coaxial capillary 230 with a sheath liquid provided by a transfer capillary 224. Both the buffer liquid in the reservoir 218 and the sheath liquid in the transfer liquid reservoir 228 are made of a combination of a base and an additive. The base is water, methanol, or any other suitable material and the additive is acetic or formic acid or another suitable chemical. In some versions, a plastic or metal protective layer is used to protect these capillary layers.
In the combined devices of
In both sheathed apparatus 200A and sheathless apparatus 200B, the example device includes a single power supply 222 adapted with a protective circuit 250. In
Referring now to
The example electrospray power supply 202 with the protective circuit 250 sinks current when coupled with a diode-based protection circuit to handle situations where there is significant current reversal. In the following figures, an example system constructed in accordance with the teachings of the present invention, was subject to a number of test and the example system's capabilities were demonstrated and compared to a conventional configuration as illustrated in
For the results shown, Acetic acid and hydrofluoric acid (HF) were used in the base solution. Samples are of Bovine serum albumin (BSA), angiotensin II (human, Asp-Arg-Val-Tyr-Ile-His-Pro-Phe), and formic acid (FA) were used. In the tested example of the teachings of this disclosure the fused silica separation capillary 212 and the electrospray emitters 502 were constructed of borosilicate glass capillary. An LTQ XL™ Linear Ion Trap Mass Spectrometer, available from Thermo Fisher Scientific Inc., Waltham Mass. was used for all experiments. Only MS1 spectra were acquired. The scan range of the ion trap mass analyzer was m/z 350-1800. The separation power supply 122 was a SPELLMAN CZE-1000R, as noted above.
For the test, BSA in 100 mM NH4HCO3 (pH 8.0) containing 8 M urea was denatured at 37° C. for 30 min, followed by standard reduction and alkylation with DTT and IAA. After dilution with 100 mM NH4HCO3 (pH 8.0) to reduce the urea concentration below 2 M, protein digestion was performed at 37° C. with trypsin at a trypsin/protein ration of 1/30 (w/w) for 12 h. After acidification with FA, the protein digest was desalted with a C18-SepPak column and then was lyophilized with a vacuum concentrator. The dried protein digest was stored at −20° C. before use.
The example protection circuit 250 disclosed herein was designed and fabricated for the HVM amplifier to prevent current from passing into the HVM amplifier when excess current was generated during electrophoresis. As previously noted, a schematic of the HVM Technologies amplifier protection circuit 250 is illustrated in
For initial power supply evaluations, a 60 cm bare fused silica capillary (50 μm i.d., 150 μm o.d.) was used to initially test the performance of the HVM amplifier and the SPELLMAN power supply. The distal end 216 of the separation capillary 212 was not etched by HF in this initial experiment. The example electrospray emitter 242 had an opening of 10 μm. The sheath buffer used was 0.1% (v/v) FA in water containing 10% (v/v) methanol. Three background electrolytes were used. The first separation buffer was 0.1% (v/v) FA, the second separation buffer was 5% (v/v) acetic acid, and the third separation buffer was 0.5% (v/v) FA.
The power supplies 122, 222 were controlled by LabVIEW software. The injection end of the separation capillary and an electrode were fixed in an injection block. The electrode provided high voltage for capillary electrophoresis separation. Nitrogen gas was used to provide pressure for capillary flushing and sample injection; no pressure was used during separation.
A constant spray voltage (1.4 kV) was used for these experiments. The separation voltage applied at the injection block was increased from 1 to 30 kV in 2 kV increments. This procedure was performed for the background electrolytes listed above. Spray voltage was measured in the vial that supplied the sheath buffer using a Fluke 80K-40 HV Probe. Prior to applying the separation voltage, the spray voltage was set and measured with the high voltage probe.
A second series of experiments used a 31 cm long, 20-μm ID, 150 μm OD separation capillary. Approximately 5 mm of the distal end of the separation capillary was etched using HF to an o.d. of ˜45 as reported earlier. The etched capillary allowed placing the distal end of the capillary a few micrometers from the emitter opening, which increased sensitivity. The electrospray emitter used in this experiment had an opening of 20 μm. The sheath electrolyte used in the electrospray emitter was 0.1% (v/v) FA in water containing 10% (v/v) methanol. The background electrolyte was 5% (v/v) acetic acid. The separation voltage was 27 kV and the spray voltage was 1.6 kV for all experiments with the 20 μm i.d. capillary.
The length of the injection plug was estimated using:
where L is the injection length in mm, P is the pressure in mbar (1 mbar≈0.015 psi), S is the injection time in seconds, C is the capillary length in cm, and D is the capillary inner diameter in μm. The injection conditions used were 2.0 seconds at 10 psi, which gave an injection length of 5.4 mm. The first sample used was a BSA digest diluted in 0.1% (v/v) FA to a concentration of 0.5 mg/mL. The second sample used was angiotensin II diluted in 0.1% (v/v) FA to concentrations of 2, 5, 10, and 20 μM. Data were collected in triplicate.
A final set of experiments used a 31-cm long, 50-μm ID, 150 μm OD separation capillary. Approximately 10 mm of the distal end of the capillary was etched using HF to an o.d. of ˜65 The electrospray emitter, sheath buffer, and separation buffer were identical to those used in the 20 μm i.d. capillary experiment.
For this set of experiments, the spray voltage was increased to 1.7 kV. Three experimental conditions were employed, Table 1. The first condition used the HVM amplifier with a separation voltage of 19 kV. The second condition used the SPELLMAN power supply with a separation voltage of 11 kV. These first two conditions demonstrate the capabilities of each of the electrospray power supplies at the maximum voltage applied by the separation power supply that would not alter the spray voltage. The third condition used the SPELLMAN power supply with a separation voltage of 19 kV and provides a comparison of the ability of the two electrospray power supplies to mitigate current reversal.
Injection conditions were calculated using Equation 1 to maintain a constant injection length of 5.4 mm. The injection conditions used were 0.8 seconds at 4 psi. The samples used for the 50 μm i.d. capillary experiment were identical to those used for the 20 μm i.d. capillary experiment. The first sample used was a BSA digest diluted in 0.1% (v/v) FA to a concentration of 0.5 mg/mL. The second sample used was angiotensin II diluted in
A 60-cm long, 50-μm ID separation capillary was used for the initial experiments. The performance of the HVM amplifier and SPELLMAN CZE-1000 power supplies were compared for three background electrolytes 0.1% (v/v) FA, 5% (v/v) acetic acid, and 0.5% (v/v) FA.xxxx The sheath electrolyte was 0.1% (v/v) FA in water containing 10% (v/v) methanol. A spray voltage of 1.4 kV was applied to the sheath electrolyte reservoir and monitored using a high voltage probe.
The HVM amplifier (red, dashed line) maintains the spray voltage at 1.4 kV while the separation voltage is ramped to 30 kV with both 0.1% FA and 5% acetic acid background electrolytes. Only when the highest conductivity buffer is used, 0.5% FA, does the HVM amplifier fail to maintain the 1.4 kV spray voltage at the highest separation voltages. In contrast, the SPELLMAN power supply (solid, blue line) is unable to maintain the set electrospray voltage at the highest separation voltage for all three background electrolytes, and the separation voltage that produces a deviation from the set voltage tracks the electrolyte conductivity.
Following the initial evaluation, the electropherograms were compared and generated using the HMV amplifier and SPELLMAN power supply with a separation voltage of 27 kV and spray voltage of 1.6 kV with the 20 μm i.d. capillary. The purpose of this experiment was to determine if the HVM amplifier could produce comparable capillary electrophoresis data to the SPELLMAN power supply under conditions with relatively low electrophoretic separation current.
The first sample that was tested was a BSA digest at a concentration of 0.5 mg/mL. The base peak electropherograms for the HVM amplifier and SPELLMAN power supply are shown in
The next sample was angiotensin II at concentrations of 20, 10, 5, and 2 μM. Electropherograms of 2 μM angiotensin II are presented in
Unweighted least squares plots were linear for both the HVM amplifier (slope=1.799×106, R=0.9920) and the SPELLMAN power supply (slope=2.088×106, R=0.9940). The y-intercepts for the HVM amplifier and the SPELLMAN power supply were equal to zero within experimental error. Thus, in general, the two power supplies both produced similar quantitative CE-ESI-MS performance.
A more stringent evaluation of the electrospray power supplies, the experiment was repeated with a 50 μm i.d. capillary. The increased inner diameter of the separation capillary provides a ˜six-fold increase in capillary cross-section, which results in a proportional increase in current flow through the separation capillary.
In the initial test of the 50 μm i.d. capillary, the SPELLMAN power supply was not able to mitigate spray voltage instability above a separation voltage of 11 kV, causing the spray voltage to deviate from its setting. The difference between the HVM amplifier and the SPELLMAN power supply is illustrated most dramatically when comparing the electropherograms from the 2 μM angiotensin II runs at the three voltage conditions in this experiment (
The electropherogram generated by the HVM amplifier and a separation voltage of 19 kV (
Again, unweighted least squares calibration curves were linear for the HVM amplifier (slope=1.4×106, R=0.9945), the SPELLMAN power supply with a separation voltage of 11 kV (slope=1.5×106, R=0.9962), and the SPELLMAN power supply with a separation voltage of 19 kV (slope=6.12×105, R=0.9967). The y-intercepts for all three plots were equal to zero within experimental error.
Finally, SPELLMAN CZE-1000R power supplies were used widely for capillary electrophoresis separations and have been used extensively as an electrospray power supply. However, as illustrated here, the SPELLMAN power supplies are not well suited for fast separations that employ high electric fields with high ionic strength separation buffers. The protection circuit 250 of the example shown developed is clearly capable of mitigating spray voltage instability and would protect the amplifier from current reversal. In the event that some current passed into the HVM amplifier, its manufactured ability to sink current could offer a secondary protection to the integrity of the spray voltage. The limit was identified of where the HVM amplifier protection circuit failed to maintain the applied spray voltage (
The description detailed herein is described in reference to the particular testing apparatus to end at a mass spectrometer. It is contemplated that the protective circuit 250 described herein could be adapted to a series of capillary electrophoresis or other reactions in which continuous voltage control is important, but stray current flows may damage a power supply.
Although certain example methods and apparatus have been described herein, the scope of coverage of this patent is not limited thereto. On the contrary, this patent covers all methods, apparatus, and articles of manufacture fairly falling within the scope of the appended claims either literally or under the doctrine of equivalents.
This application is a non-provisional claiming priority to U.S. Patent Application No. 62/400,036 entitled “CE Electrospray Distal-End Power Supply,” filed on Sep. 26, 2016, the contents of which are incorporated herein by reference in their entirety.
This invention was made with government support under Contract No. R01GM096767 awarded by National Institutes of Health (“NIH”). The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
62400036 | Sep 2016 | US |