The invention relates to methods and devices for occlusion or ligation of an atrial appendage.
Embolic stroke is the nation's third leading killer for adults. Embolic stroke is also a major cause of disability. The most common cause of embolic stroke is thrombus formation in the left appendage on the atrium. In almost all atrial fibrillation (AF) patients suffering from embolic stroke, a thrombus clot forms in the appendage of the left atrium.
The primary therapy for the prevention of stroke in AF patients is the administration of oral anticoagulants. Although somewhat effective, there are numerous side effects, including bleeding and lifestyle compromises. Pharmacological therapies (such as Warfarin®) are not well tolerated by patients. The introduction of biomaterials into the left atrial appendage has resulted in the biomaterials eventually breaking down by resulting in clot formation. The left atrial appendage has been removed by others via open chest and thoroscopic surgical approaches. Such a procedure is described by Johnson in U.S. Pat. No. 5,306,234 entitled “Method for Closing an Atrial Appendage.” The '234 patent discloses a method for grasping the left atrial appendage and manipulating it into position in order to sever the tissue and remove the appendage. The wound on the heart is then sewn or clamped shut.
Appriva Medical, Inc. disclosed concepts for occluding the left atrial appendage from a percutaneous endocardial approach. In U.S. Pat. No. 6,152,144 entitled “Method and Device for Left Atrial Appendage Occlusion” assigned to Appriva Medical, a device and method for isolating the left atrial appendage from the inside of the heart is disclosed. A barrier or other device is anchored in the chamber of the left atrial appendage to prevent the passage of blood into and out of the chamber and thereby prevent clot formation. However, any foreign device left in the chamber of the heart is a potential thrombosis-generating site. In addition, biomaterials are known to eventually break down and result in clotting.
Some surgeons will remove or oversew the left atrial appendage as a concomitant procedure during other cardiac surgery. This is done under general anesthesia and may result in additional trauma to the patient.
While endoscopic or percutaneous approaches reduce the invasiveness of the surgical procedure, the above-described approaches have inherent limitations. Surgical removal of the left atrial appendage is very invasive and often results in loss of atrial hormones, such as atrial natriuretic peptide (ANP), and significant bleeding. In U.S. Pat. No. 6,666,861 issued to Grabek and entitled “Atrial Remodeling Device and Method,” a method is disclosed for applying a suture lasso placed endoscopically around the left atrial appendage to isolate it from the atrium. The '861 patent describes using either wet cauterization to remove the tissue or leaving the isolated appendage in place.
Endoscopic stapling devices, suture loops tied to the base of the appendage, and clips pinching the appendage from the outside surface to the base to close the appendage are used by physicians to isolate and remove the left atrial appendage. In U.S. Pat. No. 6,488,689 issued to Kaplan and entitled “Methods and Apparatus for Transpericardial Left Atrial Appendage Closure,” a method and apparatus to close the left atrial appendage is disclosed. The '689 patent describes using a grasper and a closing loop or clip applied to the outside of the left atrial appendage. The clip is applied extending toward the chamber of the atrial appendage and extending over the outside edge of the appendage. The clips of the '689 patent are a U-shaped metal clip, having a spring tendency to hold its shape, being deformed to open while positioned to extend over the tissue, before the clip is allowed to return to its resting shape, having the tissue pinched between the ends of the clip.
Some embodiments of the invention provide a system for occluding a left atrial appendage of a patient. Some embodiments of the system can include a ring occluder that can be positioned around the left atrial appendage and a ring applicator to position the ring occluder with respect to the left atrial appendage. The ring applicator can include a ring spreader with a spreader hinge coupled to an upper spreader jaw and a lower spreader jaw. The ring occluder can be coupled between the lower spreader jaw and the upper spreader jaw. The spreader hinge can move between an open position in which the ring occluder has a first diameter and a closed position in which the ring occluder has a second diameter, the first diameter being larger than the second diameter.
Other embodiments of the invention provide a clip occluder that can be positioned around the left atrial appendage. The clip occluder can include a clip hinge coupled to an upper clip jaw and a lower clip jaw. The clip occluder can include a clip lock. A clip applicator can position the clip occluder with respect to the left atrial appendage. The clip applicator can include a clip actuator coupled to the clip occluder by an actuator suture. The actuator suture can control a distance between the upper clip jaw and the lower clip jaw. The clip applicator can be removably coupled to the clip occluder with a retention suture.
In some embodiments of the invention, a ring applicator can include a shaft having a handle on a proximal end and a distal end, and a lumen or channel extending from the handle to the distal end of the shaft, an actuator coupled to the handle, and a ring spreader assembly on the distal end of the shaft. In one embodiment, the ring spreader assembly may include a plurality of ring expanding members. In an alternative embodiment, the ring spreader assembly may include a spreader drive wire. A ring occluder may be releasably attached to the plurality of ring expanding members or to the drive wire. The ring occluder can be stretched to an open position by the actuator which is coupled to the ring expanding members or the drive wire to allow the ring occluder to be manipulated over tissue to be occluded.
In some embodiments of the invention, the occluder member, e.g., a ring occluder or clip occluder, may comprise one or more pharmacological and/or biological agents, e.g., anti-inflammatory and/or anti-arrhythmic agents and/or drugs. In some embodiments, the occluder member may include a fabric covering.
To apply the ring 16 to a patient, a ring applicator 34, as shown in
In some embodiments, the spreader hinge 46 can include a four-bar assembly 52 for use in moving the upper spreader jaw 42 and the lower spreader jaw 44 substantially in parallel. The four-bar assembly 52 can include a first distal link 54 and a second distal link 56, as shown in
The ring spreader 40 can be moved from a relaxed closed position (as shown in
In some embodiments, as shown in
As shown in
As shown in
The feed slots 65, 74 can be used to direct the sutures attached at the handle 38 to the jaw actuator 48. The sutures 66, 68, 70, 72 can be directed along the spreader jaws 42, 44 and directed to pass through the spreader jaws 42, 44 at the apertures 67, 69, 75, 79. The sutures can 66, 68, 70, 72 then loop around the ring 16 at predefined locations. The sutures 44, 42 can also pass through the apertures 53, 73, 79, 81 and along the return slots 71, 76 to return through a lumen to attach to the jaw actuator 48. The front suture 66 can have a first end attached to the jaw actuator 48. The front suture 66 can extend through the handle 38, along the first feed slot 74, through the first distal aperture 75, and around the ring 16 in a single loop defining the front portion of the ring 16. The front suture 66 can also extend back through the second distal aperture 77, along the first return slot 76, and through the handle 38 to a second end connected to the jaw actuator 48. Likewise, the upper suture 68 can extend through the handle 38, along the first feed slot 74, through the first upper aperture 79, and around the ring 16 in a single loop defining the upper portion of the ring 16. The upper suture 68 can also extend back through the second upper aperture 81, along the first return slot 76, and through the handle 38 to a second end connected to the jaw actuator 48.
The lower suture 72 can have a first end attached to the jaw actuator 48. The lower suture 72 can extend through the handle 38, along the second feed slot 65, through the first lower aperture 67, and around the ring 16 in a single loop defining the lower portion of the ring 16. The lower suture 72 can also extend back through the second lower aperture 73, along the second return slot 71, and through the handle 38 to a second end connected to the jaw actuator 48. Likewise, the proximal suture 70 can extend through the handle 38, along the second feed slot 65, through the first proximal aperture 69, and around the ring 16 in a single loop defining the proximal portion of the ring 16. The proximal suture 70 can also extend back through the second proximal aperture 53, along the second return slot 71, and through the handle 38 to a second end connected to the jaw actuator 48.
In some embodiments, as shown in
The sutures 66, 68, 70, 72 can act as a retainer to hold the ring 16 in the ring spreader 40. The sutures 66, 68, 70, 72 can also act as a portion of the jaw actuator 48 by pulling the ring spreader 40 toward the handle 38 to force the spreader hinge 46 to pivot at the four-bar assembly 52, causing the spreader jaws 42, 44 to spread away from one other and stretching the ring 16 open. The sutures 66, 68, 70, 72 can open the ring 16 into a substantially rectangular shape as shown in
As shown in
The clip actuator 108 can be used to move the locking clip 96 from an open, unlocked position to a closed, locked position. The clip actuator 108 can include an actuator suture 110 having a first end attached to the handle 38 and extending through a lumen 112 (as shown in
The first feed aperture 119 can be positioned through the lower clip jaw 118 to allow the actuator suture 110 to pass through the lower clip jaw 118 to the upper clip jaw 124. The upper clip jaw 124 can include the first and second actuator apertures 125, 123 that can allow the actuator suture 110 to loop around the upper clip jaw 124, while retaining a position near the clip lock 126 on the upper clip jaw 124. The clip actuator 108 can be a movable actuator, similar to the torque screw assembly on the ring applicator shown in
The lower clip jaw 118 can include an engagement connection 121 to releasably connect the locking clip 96 to the clip applicator 98 with a retention suture 111 (as shown in
As shown in
As shown in
As shown in
As shown in
Some embodiments of the invention provide a tool designed to place a ring-style left atrial appendage occlusion device. The tool can include a handle with a long neck. An upper and a lower jaw can be attached to the handle with a four-bar assembly on a distal end and a knob 50 and a torque screw 82 on a proximal end. Some embodiments of the tool can include four separate sutures (2 front and 2 rear) that can loop around a ring and then into the upper and lower jaws. The sutures can be positioned in slots on outside edges of the upper and lower jaws and down into the handle. A retaining suture can be positioned through the lower jaw of a ring into a distal end of the handle, and the loop can be completed outside a proximal end of the tool. The retaining suture can be used to pull and then hold the ring tight against the distal end of the handle. Two holes can be positioned in an end of the lower jaw of the ring. Alignment pins can be inserted into the two holes to help hold and position the ring. After routing through the handle, the retaining suture can be positioned into the hex end of the torque screw and through to the other side where the retaining suture are then tied off.
According to some embodiments of a method of the invention, a port can be placed in the patient's chest so that when the ring is placed and opened, the left atrial appendage can be pulled into the opening with a grasper, a vacuum source (e.g. a cone), an adhesive tool tip, a cryo device for temporarily sticking to tissue, etc. A neck of the tool can include articulation to aid in placement of the ring. A distal end of the tool can be guided through the port and placed near the left atrial appendage. As the torque screw is turned, the upper and lower jaws open parallel to one another. Continuing to turn the torque screw stretches the ring open. Other methods of actuation can be used to pull the sutures, such as a trigger, slider, etc. When the ring is fully opened, the left atrial appendage can be pulled between the upper and lower jaws until properly located. The torque screw can be turned in the opposite direction to release tension on the sutures and relax the ring around the left atrial appendage. The torque screw can be tightened and relaxed multiple times, if necessary to achieve proper placement. Once the ring is properly positioned, the sutures can be cut (either near the ring or on a proximal end of the tool) to release the ring, and the tool can be retracted. An inside edge of the port can be used to close the upper and lower jaws so the tool can be removed.
Other embodiments of the invention provide a tool designed to place a clip-style left atrial appendage occlusion device. The clip can be a rigid one-piece clip with a snap-in lock on one end. The clip can also include a living hinge that is spring biased open on the other end. However, other embodiments of the tool can be used with other types of rigid clips, as well as a different hinge or latching mechanism. The tool can include a handle with a long neck. Two separate suture loops can be positioned along the length of the handle inside the neck. A stop on a distal end of the tool can be actuated on a proximal end of the tool. An actuation suture can be positioned through an upper clip jaw of the left atrial appendage clip, then through a lower clip jaw, into a distal end of the handle and complete a loop outside a proximal end of the handle. A port can be created in the patient's chest such that when the clip is positioned near the left atrial appendage, the left atrial appendage can be pulled between the upper and lower clip jaws with a grasper. The neck on the tool can articulate to aid in placement of the clip. The stop can be placed in its forward position to keep the upper clip jaw from latching while pushing it through the port. The clip can spring open after it has passed through the port. The handle can be used to position the clip near the left atrial appendage and the left atrial appendage can be pulled between the upper and lower clip jaws with a grasper. Once the clip is positioned on the left atrial appendage as desired, a stop actuation knob 50 can be pulled back to retract the stop. The actuation suture can be pulled (or actuated with a trigger, slide, screw, etc.) until the upper clip jaw snap latches into the lower clip jaw. The left atrial appendage is then occluded. The sutures can be cut and pulled through the handle, which can release the clip. The sutures can be cut with a scalpel, scissors, or other surgical instruments, or the sutures can be cut with a mechanism that is built into the tool itself. The tool can then be removed from the port.
Some embodiments of the invention provide a tool designed to place a loop-clip style left atrial appendage occlusion device. The tool can include a handle with a long neck. The tool can include two separate suture loops that can be positioned through the length of the handle and a shaft. The tool can include a knob 50 and a torque screw on a proximal end of the handle and the shaft. The loop clip 174 can include a rigid base with a flexible loop that can include one-way teeth molded into it. The loop can wrap around one end of the rigid member and through a slot with a locking snap (like a cable tie). This allows the loop to be pulled in, but not release. The end of the loop can include a hole that a suture can be routed through. The suture can be positioned into the distal end of the placement device, can be positioned through the handle, into the hex end of the torque screw, and tied off on an opposite end. A retaining suture can be positioned through a hole in an end of the loop clip, then into a distal end of the placement device, positioned through the handle, and tied off on a proximal end of the tool. The retaining suture can be used to pull and hold the loop clip tight against a distal end of the handle. An alignment boss on an end of the handle can be inserted into a matching slot on the loop clip to ensure proper alignment. A port can be created in the patient's chest so that when the loop clip is positioned, the left atrial appendage can be pulled between the loop and the base with a grasper. The neck on the placement tool can articulate to aid in placement of the loop clip. The flexible loop can be pushed down to place the loop clip through the port, then once it is through, the loop can return to its original shape. The handle can be used to position the loop clip near the left atrial appendage and the left atrial appendage can be pulled inside the loop with a grasper. Once the loop clip is positioned on the left atrial appendage as desired, turning the torque screw can gradually tighten the loop (a trigger, slide, etc. could also be used to tighten the loop). The torque screw can be turned until the loop is tight enough to occlude flow and remain securely placed. After the loop is tight enough, the actuation suture and the retaining suture can be cut and pulled through the handle. The tool can then be removed from the port.
Some embodiments of the invention include a device and method for occlusion or ligation of an atrial appendage or other tissue. The method and applicators disclosed herein describe a minimally-invasive approach to ligation of an atrial appendage, specifically, of the left atrial appendage of patients with atrial fibrillation. Some embodiments of the invention include a method and apparatus to access the left appendage through a small incision and the use of a delivery tool to apply an occluder to the appendage. The tool can be used to grasp the appendage to help stabilize the appendage to allow for application of the ligation device. The ligation device may be applied and left behind as a permanent implant.
Some embodiments of the invention include a device and procedure that can occlude the left atrial appendage from the body of the left atrium—thereby substantially preventing the formation of a clot within the appendage and a subsequent embolism. Some embodiments of the invention include an implantable device and applicator for substantially permanently occluding the left atrial appendage. Some embodiments of the invention include a device and procedure that is minimally invasive to apply a device as a simple and quick method to deliver therapy to prevent embolic strokes. Some embodiments of the invention include a device and procedure that does not require the use of blood-contacting biomaterials. Some embodiments of the invention include a device and procedure that results in tissue necrosis at the left atrium/left atrial appendage junction that is necessary to help prevent reentry. Some embodiments of the invention include a device and procedure that places a device to occlude while preserving the tissue of the left atrial appendage for the production of atrial hormones. Some embodiments of the invention include a device and procedure with a substantially permanently-implanted clamp used for occluding the left atrial appendage. Some embodiments of the invention include a device and procedure that is applied from the exterior of the heart, which may be accessed by a sternotomy, thoracotomy, minimally invasive, endoscopic or other means. Some embodiments of the invention include a device and procedure that may be practiced by a number of different embodiments of the clamping mechanism as disclosed herein.
The delivery tool 202 can include a ring spreader 240 having ring-expanding members 242 used to hold and expand the ring 214. The ring spreader 240 can be coupled to a distal end of shaft 220. In one embodiment, the delivery tool 202 can include multiple ring-expanding members 242, for example, four, as shown in
As shown in
As shown in
As shown in
In one embodiment, surgical access to the left atrial appendage can be through a left-sided thoracotomy or laproscopic port incision. The delivery tool and attached collapsed ring can be inserted through the left thoracotomy access. The ring can then be expanded. The left atrial appendage can be grasped and drawn or pulled through the expanded ring. The ring can be positioned toward the base of the left atrial appendage and released from the delivery tool. The delivery tool can then be removed from the patient, and the incision can be closed. Various imaging methods can be employed before, during, and after the tissue occlusion procedure. For example, contrast fluoroscopy, trans-thoracic ultrasound, and/or trans-esophageal echo (TEE) can be employed. Other surgical approaches are possible including sub-xyphoid.
As shown in
Some embodiments of the invention address a number of problems, such as surgical access to the endocardial surfaces of the atrial chambers of a beating heart and permanent closure of the atrial appendage volume. These problems have been addressed by others using various methods and devices, such as percutaneous catheters. The right atrium may be accessed via transvenous catheters placed through a femoral vein in the groin, as well as through superior veins such as the subclavian, brachiocephalic, or jugular veins. The left atrium is more difficult to reach transvenously, requiring first, right atrial access followed by a transseptal puncture through the fosa ovalis of the inter-atrial septum into the left atrium. With these transvenous methods, only relatively small diameter catheters can be passed through the vasculature. In addition, these devices must be navigated using fluoroscopic guidance or some form of electronic navigation.
Positioning and placement of the therapeutic elements of such catheters can be a challenge, because the movement is controlled remotely from the point of venous access. From point of entry into the body to the therapeutic end of the catheter, the distance may be 70-110 cm through a difficult path. In cases where tissue contact force is critical, this can be a significant problem. Stability of the catheter tip is an issue in a beating heart compared to firm control possible with much shorter and more rigid surgical implements. A number of commercially-available catheters may be able to be positioned in most areas of the atria, but only one catheter-based device has been developed to permanently close off the left atrial appendage following a procedure. This device, known as PLAATO (percutaneous left atrial appendage total occlusion) must be carefully sized to allow positioning within the left atrial appendage such that it is retained in position distal to the ostium with the left atrial chamber. This presents a significant risk that the device may be released into the atrium and pass into the left ventricle and become entangled in the chordae tendinae supporting the mitral valve or become lodged in the left ventricular outflow tract or aorta. The risks of such a procedure were documented in an abstract by Fischer at the 2005 meeting of the American College of Cardiology—Evelyn Fischer, et al., “Left Atrial Appendage Occlusion to Prevent Stroke in Suboptimal Warfarin Candidates: Current Results of the European Multicenter Registry Trial,” American College of Cardiology, Abstract presented at 2005 National Meeting, which is herein incorporated by reference in its entirety.
The atria may be accessed through sternotomy, thoracotomy, intercostals ports, or under the sub-xiphoid process. Access to the atria is important for treatment of atrial fibrillation (AF), atrial-septal-defects (ASD's), patent foramen ovalis (PFO), and mitral or tricuspid valve disease. Also of importance is the elimination of the left atrial appendage volume at the end of the procedure in order to reduce stroke risk. Surgical removal and closure of the left atrial appendate has been accomplished by using a surgical stapler/ligation device or by suturing the appendage closed followed by surgical excision of the distal appendage. This is not without risks as noted by Krum et al.—David Krum, David L. Olson, Daniel Bloomgarden, Jasbir Sra, “Visualization of Remnants of the Left Atrial Appendage following Epicardial Surgical Removal,” Heart Rhythm (2004) 1, 249. Such surgical removal may result in an incomplete reduction of the left atrial appendage and allow a volume to remain unclosed, which is herein incorporated by reference in its entirety.
Regarding the relationship between the left atrial appendage and stroke risk, Blackshear et al. stated that left atrial appendage obliteration “is a routine part of modern ‘curative’ operations for nonrheumatic atrial fibrillation, such as the maze and corridor procedures.” Joseph L. Blackshear, MD, John A. Odell, FRCS(Ed), “Appendage Obliteration to Reduce Stroke in Cardiac Surgical Patients with Atrial Fibrillation,” Ann. Thorac. Surg., 1996; 61:755-759, which is herein incorporated by reference in its entirety. To assess the potential of left atrial appendage obliteration to prevent stroke in nonrheumatic atrial fibrillation patients, they reviewed previous reports that identified the etiology of atrial fibrillation and evaluated the presence and location of left atrial thrombus by transesophageal echocardiography, autopsy, or operation.
They reviewed the results of twenty-three separate studies and found that 446 of 3,504 (13%) rheumatic atrial fibrillation patients, and 222 of 1,288 (17%) nonrheumatic atrial fibrillation patients had a documented left atrial thrombus. Anticoagulation status was variable and not controlled for. Thrombi were localized to, or were present in the left atrial appendage and extended into the left atrial cavity in 254 of 446 (57%) of patients with rheumatic atrial fibrillation. In contrast, 201 of 222 (91%) of nonrheumatic atrial fibrillation-related left atrial thrombi were isolated to, or originated in the left atrial appendage (p<0.0001). Their data suggested that left atrial appendage obliteration is a strategy of potential value for stroke prophylaxis in nonrheumatic atrial fibrillation. A device was developed that allows percutaneous left atrial appendage transcatheter occlusion (PLAATO) via transseptal catheterization. Initial studies in dogs demonstrated the ability of the device to seal the left atrial appendage. Sievert et al. reported their initial experience with PLAATO in a human clinical trial involving 15 patients. Horst Sievert, MD et al, “Percutaneous Left Atrial Appendage Transcatheter Occlusion to Prevent Stroke in High-Risk Patients with Atrial Fibrillation,” Circulation, 2002, 105:1887, which is herein incorporated by reference in its entirety.
PLAATO was purported to be a less invasive, percutaneous approach to closing the left atrial appendage. Previous animal studies of the device with follow-up of up to 1 year have demonstrated occlusion of the left atrial appendage with complete healing, absence of erosions, new thrombus formation on the device, or interference with atrial function.
In the initial cohort of 15 patients, occlusion of the left atrial appendage was successful in all, as proven by left atrial angiography. There were no complications associated with the device, either acutely during the implantation procedure or during follow-up. The only complication during the study was hemopericardium in the first patient attempted, which was not device-related. It resulted from left atrial appendage access, and should be easily avoided with increased experience. The procedure was successful in a second attempt in that patient.
All patients did well in follow-up. One theoretical concern is the development of new thrombi on the implant. However, the use of ePTFE on the implant surface should result in relatively benign healing. Histological examination in clogs undergoing PLAATO reveal partial endothelialization at 1 month, which is complete by 2 to 3 months. In these 15 patients, transesophageal echo (TEE) at 1 month showed the surface to be completely smooth and free of mobile thrombi.
A larger cohort of patients was included in the European PLAATO Registry Trial by Fischer, et al. a study that was finished in January 2003 that examined the safety and feasibility of this procedure. This study described the experience of 92 patients. Inclusion criteria were atrial fibrillation (AF) with inability to take Warfarin®, prior cerebral ischemia and/or two clinical risk factors for stroke. After implantation of the PLAATO occluder, the patients were followed with X-ray, TEE and NIH stroke scale in regular intervals. Of the 92 patients, 67% were male with a mean age of 70±9 years. All candidates were successfully implanted. The mean procedure time was 76±36 minutes and the mean left atrial appendage orifice diameter was 20±3 mm. During follow up, one patient died of a bronchial carcinoma diagnosed 3 months before the one year follow up. One patient sustained a stroke six months post implant. Thus, the yearly incidence of stroke after implantation is 1.9%. With this small number of patients, the estimated risk reduction was 55%.
Of concern, in three patients, a thrombus on the occluder was found prior to hospital discharge (2) and one month after the procedure (1). All thrombi were resolved without sequelae. One device was chosen too small and embolized into the aorta after its release. It was snared with a catheter and was retrieved successfully. Another device was implanted successfully in the very same procedure.
To summarize this group's experience with the PLAATO device, the incidence of stroke in high-risk patients may decrease after implantation of the device. Considerable risks exist with this procedure, including errant transseptal puncture resulting in aortic dissection or atrial free wall perforation resulting in tamponade, embolization of the PLAATO device resulting in device entanglement in cardiac structures, along with thrombus formation on the occluder surface that could lead to emboli production and stroke. In addition, the chronic nature of this implant must be considered. Constant flexture of the nitinol wire structure may lead to long term fatigue and potential fracture and perforation of cardiac or adjoining tissues.
The group of Odell et al, hypothesized that if the atrial appendage could be safely obliterated, then the incidence of embolic stroke may be lessened. John A. Odell, et al., “Thoracoscopic Obliteration of the Left Atrial Appendage: Potential for Stroke Reduction,” Ann. Thorac. Surg., 1996, 61:565-569, which is herein incorporated by reference in its entirety. If the appendage can be obliterated using a thoracoscopic technique, a procedure of lesser magnitude than thoracotomy, then it may offer an alternative form of management for patients ineligible for Warfarin® therapy. To determine the feasibility of atrial appendage obliteration done using the thoracoscope, they performed the procedure in mongrel dogs and in human cadavers.
Transesophageal echocardiography with emphasis on visualization of the left atrial appendage was performed pre-, intra-, and postoperatively. In five dogs, the atrial appendage was obliterated with staples, and in five the appendage was obliterated with an endoloop of 0 Vicryl suture material. Three ports were made—one in approximately the seventh interspace approximately 5 cm from the midsternum (port 1), a second inserted anteriorly in the fourth interspace (port 2), and a third more posteriorly in the fourth interspace (port 3). Carbon dioxide was instilled to a pressure of 4 to 10 mm to collapse the lung. In all animals, the pericardium was opened anterior and parallel to the phrenic nerve. Gordon N. Olinger, MD, “Carbon dioxide displacement of left heart chambers,” J. Thorac. Cardiovasc. Surg., 1995, 109:187-188, which is herein incorporated by reference in its entirety.
Through the first port, the camera was inserted; through the second port, the pericardium was grasped with an instrument; and, using scissors inserted through the third port, the pericardium was opened. The technique then varied depending upon whether the appendage was obliterated with staples or with the endoloop. In those having the appendage stapled, the camera was withdrawn from port 1 and inserted in port 3. Through port 1, a 35 endo GIA stapler (Ethicon Endosurgery, Cincinnati, Ohio) with the knife blade removed was inserted, positioned across the base of the atrial appendage, and fired. In dogs having the appendage obliterated with the endoloop (Ethicon), the camera position was not changed. The endoloop was introduced through port 3 and the appendage was grasped through the loop of the suture. The loop was positioned across the base of the appendage and then tightened.
At 11 weeks, the dogs were again anesthetized with sodium pentobarbital (30 mg/kg intravenously) and a midline sternotomy was made. The heart was examined using epicardial echocardiography. The dogs were euthanized, the hearts were removed, and the left atrium was inspected.
The procedure also was attempted in eight human cadavers. In the cadavers, three ports were used for access. The most appropriate sites appeared to be the second interspace anteriorly in the midclavicular line (for grasping the pericardium and the atrial appendage), the sixth interspace in the midclavicular line (for the camera or stapling instrument), and the fifth interspace in the anterior axillary line (usually for the scissors to open the pericardium, but also for the camera or for the stapling instrument). The procedure as performed in the dog and human experiments is illustrated in
The group of DiSesa investigated the use of an automatic surgical stapler for ligation of the atrial appendage in sheep, and then applied this technique in patients. V. J. DiSesa, S. Tam and L. H. Cohn, “Ligation of the Left Atrial Appendage using an Automatic Surgical Stapler,” The Annals of Thoracic Surgery, Vol. 46, 652-653, which is herein incorporated by reference in its entirety. Fourteen adult sheep underwent ligation of the left atrial appendage using a surgical stapler with a rotating head and either absorbable or stainless steel staples. In four sheep, killed after two hours, no hemorrhage or intra-atrial thrombus was observed acutely. Ten sheep were allowed to recover for 90 to 100 days, twice the expected absorption time of absorbable staples. There was complete obliteration of the left atrial appendage without evidence of intra-atrial thrombus or staple migration. The absorbable staples were completely reabsorbed. They subsequently used this technique in five patients undergoing mitral valve procedures. There were no complications, and adequate obliteration of the atrial appendage was achieved. Other reports indicate that staples may require the use of reinforcement strips to prevent bleeding and tissue tearing.
Considering the simple surgical ligation methods, the group of Katz, et al. studied the incidence of incomplete ligation of the left atrial appendage during mitral valve surgery. Edward S. Katz MD, FACC, Theofanis Tsiamtsiouris MD, Robert M. Applebaum MD, FACC, Arthur Schwartzbard MD, FACC, Paul A. Tunick MD, FACC and Itzhak Kronzon MD, FACC, “Surgical Left Atrial Appendage Ligation is Frequently Incomplete: A Transesophageal Echocardiographic Study,” Journal of the American College of Cardiology, Volume 36, Issue 2, 1 Aug. 2000, Pages 468-471, which is herein incorporated by reference in its entirety. Using transesophageal Doppler echocardiography, they studied 50 patients who underwent mitral valve surgery and ligation of the left atrial appendage. Incomplete left atrial appendage ligation was detected in 18 of 50 (36%) patients. This study demonstrated that surgical left atrial appendage ligation is frequently incomplete. Residual communication between the incompletely ligated appendage and the left atrial body may produce a milieu of stagnant blood flow within the appendage and be a potential mechanism for embolic events. Ligation of the left atrial appendage is frequently performed during mitral valve surgery to eliminate a potential source of emboli. However, the success of completely excluding the left atrial appendage from the circulation had not previously been systematically addressed. Transesophageal echocardiography offers unique visualization of the appendage in the beating heart and can evaluate the integrity of the surgical ligation. Usually, when the left atrial appendage is ligated, its cavity is obliterated with clot (since no flow enters the cavity) and cannot be seen during echocardiography. This appearance was the same whether the patient was studied in the operating room or months after the surgery. When the appendage is incompletely ligated, not only can the appendage cavity be visualized but flow can be seen within the appendage, as well as through an opening in the ligation site.
The group discovered that 36% of the time the left atrial appendage was found to be incompletely ligated after attempts at excluding it from the left atrial body. Factors, such as an enlarged left atrium or significant mitral regurgitation, which may be thought to increase left atrial tension and pressure (perhaps predisposing to incomplete ligation or dehiscence of sutures), did not appear to correlate with this finding. They also did not observe a correlation between appendage size and the incidence of incomplete ligation. In addition, the surgical procedure (mitral repair or replacement) and operative approach (traditional sternotomy or minimally invasive approach) did not change the incidence of incomplete ligation. It is possible, however, that the sample size in this report may have been too small to exclude a significant effect of these variables on the development of incomplete left atrial ligation.
Incomplete left atrial appendage ligation was as commonly seen in the operating room, evaluating the patient by transesophageal echocardiography immediately after terminating cardiopulmonary bypass, as it was seen in the laboratory evaluating patients referred for transesophageal echocardiography at various times after the surgery. This suggested that incomplete left atrial appendage ligation is not a degenerative process with suture dehiscence over time, but rather is present immediately after the initial surgery. Incomplete ligation may be secondary to several surgical factors. First, the running sutures used may not start and end exactly at the most distal edges of the atrial appendage, which may not be recognized with the appendage empty and unstretched while on cardiopulmonary bypass during surgery. In addition, caution must be taken during appendage ligation to avoid deep suture bites, which may involve the left circumflex coronary artery or its branches that may course in the area. This meticulous care may lead to shallower suture bites that may dehisce when the LA is once again filled and stretched after cardiopulmonary bypass. Both of these mechanisms may play a role, as in many cases flow was detected both at the edge of the appendage orifice (apparently around the end of the suturing line) and through an area at the midpoint of the appendage orifice (through the suture line). One group reported six cases of incomplete left atrial appendage ligation when a purse string suture was used to accomplish the ligation, a technique different from that used by surgeons. The actual incidence, however, of incomplete left atrial appendage ligation using their technique was not addressed.
The clinical significance of an incompletely ligated left atrial appendage has never been studied. Theoretically, creating a small communication between the LA and left atrial appendage may produce stagnation of low velocity blood flow within the atrial appendage. The appendage would then be a model for thrombus formation and continue to serve as a potential source of embolization since a port of entry into the systemic circulation still exists. Although the numbers in this study were small, several observations support this theory. First, spontaneous echo contrast (a marker for stagnant blood flow and a precursor of thrombus formation) was seen within the appendage in half of the patients with incomplete ligation. Second, and perhaps more importantly, in two-thirds of patients with spontaneous echo contrast within the incompletely ligated appendage, the contrast was actually denser within the appendage than within the left atrial body, suggesting a more stagnant and thrombogenic milieu. In two patients, frank thrombus was seen within the incompletely ligated appendage.
The ultimate question, however, is whether patients with incompletely ligated left atrial appendages will have a higher incidence of thromboembolic events. In the Katz study, four patients with incompletely ligated appendages had such events (one patient with Starr-Edwards prosthesis, two with St. Jude prosthesis and one patient status after mitral repair). This is quite a high number considering that only eight patients with incomplete ligation had any potential for long term follow-up (the other ten patients with incomplete ligation were discovered in the operating room). However, one cannot exclude other etiologies for embolization (as mechanical prostheses or atrial fibrillation) and referral bias still clouds this issue. Certainly, conventional ligation methods must be questioned in light of the findings of this study.
A number of devices for occlusion of ASD's have been investigated. Melhem J. A. Sharafuddin, MD; Xiaoping Gu, MD; Jack L. Titus, MD, PhD; Myra Urness, BS; J. J. Cervera-Ceballos, MD; Kurt Amplatz, MD, “Preliminary Results With a New Self-Expanding Nitinol Prosthesis in a Swine Model Transvenous Closure of Secundum Atrial Septal Defects,” Circulation, 1997, 95:2162-2168, which is herein incorporated by reference in its entirety. Most of these concepts involve percutaneous delivery from femoral vein access. Varying levels of success have been achieved. Device dislodgment can occur if the size of the defect greatly exceeds the waist diameter of the device or approaches the diameter of the retention buttons. On the other hand, placement of a disproportionately large device may result in mushrooming of the retention buttons and weakening of the cross-clamping forces against the septal rim, which increases the risk of blood flow behind the discs and may result in incomplete endothelialization. In addition, follow-up studies of a clamshell occlusion device reported a delayed rate of metal fatigue fractures of one or more arms of about 30%, The Amplatzer device is shown in
A small introduction system, simple and reliable placement technique, and favorable initial experimental success may provide promising potential of such a device for the percutaneous closure of secundum ASDs in all age groups. Heparinization is advocated in clinical use to lower the risk of catastrophic systemic embolization.
ASD device thrombosis is likely to be similar to thrombosis to be expected on left atrial appendage closure devices. This makes the study by Krumsdorf et al, on the incidence, morphology, and clinical course of thrombus formation after catheter closure of ASD closure devices of interest regarding devices such as PLAATO. Krumsdorf U, Ostermayer s, Billinger K, Trepels T, Zadan E, Horvath K, Sievert H, “Incidence and Clinical Course of Thrombus Formation on Atrial Septal Defect and Patient Foramen Ovale Closure Devices in 1,000 Consecutive Patients,” J. Am. Coll. Cardio., Jan. 21, 2004. 43(2):302-9, which is herein incorporated by reference in its entirety.
A total of 1,000 consecutive patients were investigated after patent foramen ovate (PFO) (n=593) or atrial septal defect (ASD) (n=407) closure. Transesophageal echocardiography (TEE) was scheduled after four weeks and six months. Additional TEEs were performed as clinically indicated. Thrombus formation in the left atrium (n=11), right atrium (n=6), or both (n=3) was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p=NS). The thrombus was diagnosed in 14 of 20 patients after four weeks and in 6 of 20 patients later on. The incidence was: 7.1% in the CardioSEAL device (NMT Medical, Boston, Mass.); 5.7% in the StarFLEX device (NMT Medical); 6.6% in the PFO-Star device (Applied Biometrics Inc., Burnsville, Minn.); 3.6% in the ASDOS device (Dr. Ing, Osypka Corp., Grenzach-Wyhlen, Germany); 0.8% in the Helex device (W.L. Gore and Associates, Flagstaff, Ariz.); and 0% in the Amplatzer device (AGA Medical Corp., Golden Valley, Minn.). The difference between the Amplatzer device on one hand and the CardioSEAL device, the StarFLEX device, and the PFO-Star device on the other hand was significant (p<0.05). For a device such as PLAATO, specifically designed to reduce or eliminate thromboembolic events coming from the region of the implant, occurrence of thrombus on ASD devices is a concern.
A method and apparatus for thoracoscopic intracardiac procedures was described U.S. Pat. No. 6,401,720, entitled “Method and Apparatus for Thoracoscopic Intracardiac Procedures,” Stevens, John H.; Palo Alto, Calif. 94303, Reitz, Bruce A.; Stanford, Calif. 94305, Roth, Alex T.; Redwood City, Calif. 94061, Peters, William S.; Woodside, Calif. 94062, Gifford, Hanson S.; Woodside, Calif. 94062, which is herein incorporated by reference in its entirety. They described devices, systems, and methods provided for accessing the interior of the heart and performing procedures therein while the heart is beating. In one embodiment, a tubular access device having an inner lumen is provided for positioning through a penetration in a muscular wall of the heart, the access device having a means for sealing within the penetration to inhibit leakage of blood through the penetration. The sealing means may comprise a balloon or flange on the access device, or a suture placed in the heart wall to gather the heart tissue against the access device. An obturator is removably positionable in the inner lumen of the access device, the obturator having a cutting means at its distal end for penetrating the muscular wall of the heart. The access device is preferably positioned through an intercostal space and through the muscular wall of the heart. Elongated instruments may be introduced through the tubular access device into an interior chamber of the heart to perform procedures, such as septal defect repair and electrophysiological mapping and ablation. A method of septal defect repair includes positioning a tubular access device percutaneously through an intercostal space and through a penetration in a muscular wall of the heart, passing one or more instruments through an inner lumen of the tubular access device into an interior chamber of the heart, and using the instruments to close the septal defect. Devices and methods for closing the septal defect with either sutures or with patch-type devices are disclosed. While this concept allows access to the heart chambers similar to the present invention, it does not provide for a simple means of incisional closure as do some embodiments of the invention.
Some embodiments of the invention may provide any one or more of the following advantages: a single point of access for surgical treatment of atrial fibrillation; fewer inter-costal access ports may be needed for treating atrial fibrillation, as opposed to existing minimally-invasive methods; blunt dissection of cardiac tissue is generally not required; pericardium is left substantially intact, except for a small incision; access to the heart for delivery of therapies for various disease states; a single device can provide surgical access to the heart chambers, as well as providing a means of closing the point of access at the end of the procedure; and the left atrial appendage can be ligated and/or eliminated at the close of the procedure with little or no risk of tearing. The left atrial appendage can be eliminated at the close of the procedure such that a residual remaining volume which could lead to strokes is avoided.
Some embodiments of the invention provide a device that can provide access to the interior of the heart chambers. The device can allow for single point access to treat atrial fibrillation, atrial-septal defects, patent foramen ovalis, and valvular disease, as well as other arrhythmias. Some embodiments of the device can be used to access the ventricles from the access achieved through either appendage. Ventricular septal defects can be addressed. The device can be applied to other body structures, such as the stomach where a portion of the stomach wall could be ligated by the elastic band and excluded. This may be suitable as a treatment for obesity.
Some embodiments of the invention provide methods and devices to allow entry into the atria of a beating heart to perform delivery of therapy to the structures within the heart and endocardial surfaces and valves associated with the heart chambers. Upon removal of the device from the appendage, a permanent closure and elimination of the appendage volume can be affected. More specifically, the entry points can be located in the left and right atrial appendages. Of these, the left atrial appendage may be most appropriate, because closure and elimination of this appendage following a procedure has become a standard surgical practice performed by many surgeons.
According to embodiments of the method of the invention, pre-procedure includes placement of one or two chest wall access ports for visualization and placement of the invention. The lung can be deflated and a small opening in the pericardium can be made adjacent to the left atrial appendage.
As shown in
As shown in
In one embodiment, the occlusion ring 514 can be made of a 30A durometer silicone rubber, although other hardnesses of silicone or other materials, such as polyurethane can be used. The ring 514 can be covered with a material such as Dacron® polyester to promote tissue ingrowth or prevent ring slippage after placement or to spread the load bearing surfaces of the ring. The ring 514 can be generally expandable for placement around the atrial appendage. The delivery tool 502 can be used to expand the ring 514. The delivery tool 502 can include a shaft 520 and a handle 530 coupled to shaft 520. The delivery tool 502 can be sized for reaching an atrial appendage through an opening in the patient's chest, for example, through a sternotomy or thoracotomy. The delivery tool 502 can include a tissue-grasping tool channel 570 extending from the handle 530 to the distal end of the shaft 520. A tissue-grasping tool 580 may be movably positioned within tissue-grasping tool channel 570. For example, shaft 584 of tissue-grasping tool 580 fits within tool channel 570. In one embodiment, tissue-grasping tool 580 may include one or more hooks, barbs, suction ports and/or graspers. For example, as shown in
In one embodiment, the delivery tool 502 can include a ring spreader 540 having a pair of ring-expanding members 542 used to hold and expand the ring 514. The ring spreader 540 can be coupled to a distal end of shaft 520. The ring 514 can be releasably coupled or attached to the distal ends of ring-expanding members 542 and the distal end of shaft 520, for example, via one or more sutures 550. The sutures 550 can loop around the ring 514. The ends of the sutures 550 can pass through one or more lumens within the ring expanding members 542, the shaft 520, and the handle 530. In one embodiment, the two ring-expanding members 542 open ring 514 into a triangular shape. The proximal ends of the ring-expanding members 542 are pivotally coupled to the distal end of shaft 520, thereby allowing the ring-expanding members 542 to pivot from a closed or collapsed position, as shown in
In one embodiment, shaft 520 is approximately 12 mm in diameter and tool channel 570 is approximately 5.5 mm in diameter. In one embodiment, tool channel 570 provides guidance for positioning and manipulating tissue-grasping tool 580. In addition, tool channel 570 allows deliver tool 502 and tissue-grasping tool 580 to be positioned together through a single port, for example, a 12 mm port placed between the patient's ribs and it allows the two tools to be held by one hand.
In one embodiment of the present invention, the distal end of delivery tool 502 may be passed through a port or small incision, for example, in the chest of a patient and positioned adjacent the left atrial appendage of a heart. Next, knob 532 may be rotated, thereby opening ring 514. A tissue-grasping tool 580 may then be slid distally along tissue-grasping tool channel 570 so that graspers 582 protrude through ring 514. Graspers 582 may then be manipulated by handle 586 to grasp tissue of the left atrial appendage. Tissue-grasping tool 580 and delivery tool 502 are then manipulated so as to position a desired portion of the left atrial appendage within the triangular opening of ring 514. Ring 514 is then released from delivery tool 502 and allowed to constrict tissue of the left atrium.
The ring or band occluders and the clip occluders disclosed herein can be constructed of any one or more of the following materials: silicone rubber, polyurethane, super-elastic material, shape-memory polymer or metal, latex, nitrile, butyl, styrene-butadiene, polyacrylate, acrylic, polyisoprene, chloroprene, fluoroelastomers, or other suitable biocompatible elastomeric materials. The ring or band occluders and the clip occluders disclosed herein can incorporate any one or more of the following features: texturing to aid in mechanical stability (i.e., ridges, bumps, grooves, etc.); fabric such as Polyethyleneterapthalate (i.e., Dacron®), polyester, ePTFE, etc. to promote tissue ingrowth; other types of coatings to promote tissue ingrowth; and pharmacological agents (e.g. a controlled release agent) to aid in tissue ingrowth, local therapeutic apoptosis, local necrosis, revascularization, arrhythmia control, infection control, anti-bacterial, fluid balance (i.e., atrial natritic peptide replacement).
In some embodiments of the invention, the ring or band occluders and/or clip occluders may incorporate one or more pharmacological agents including anti-inflammatory agents (e.g., steroids, dexamethasone, beclomethasone) anti-arrhythmic agents, chemotherapeutic agents, anti-infection agents, anticoagulant agents, anti-thrombotic agents (e.g., coumadin, heparin), clotting agents, platelet agents, cytotoxic agents, growth factors, angiogenesis factors, hormones (e.g., atrial natriuretic peptide), nitric oxide, radioactive agents, radiopaque agents (e.g., barium sulfate), echogenic agents (e.g., perfluorocarbon), antibodies, antigens, immunoglobulins, enzymes, neurotransmitters, cytokines, blood agents, regulatory agents, transport agents, fibrous agents, proteins, peptides, proteoglycans, toxins, antibiotic agents, antibacterial agents, antimicrobial agents, bacterial agents, hyaluronic acid, polysaccharides, carbohydrates, fatty acids, catalysts, vitamins, DNA segments, RNA segments, nucleic acids, lectin, antiviral agents, viral agents, genetic agents, ligands, drugs and dyes (e.g., which act as biological ligands). One or more drugs or gents may be found in nature (naturally occurring) and/or may be chemically synthesized.
One or more drugs or agents may be incorporated into the ring, band or clip, e.g., within a polymeric material of the ring, band or clip. One or more drugs or agents may be incorporated into one or more coatings of the ring, band or clip, e.g., within a polymeric coating covering at least a portion of the ring, band or clip. One or more drugs or agents may be incorporated into one or more fabrics of the ring, band or clip, e.g., within or on a fabric coating covering at least a portion of the ring, band or clip. In some embodiments of the invention, one or more drugs or agents may be loaded uniformly throughout one or more materials of the ring, band or clip. In some embodiments of the invention, one or more drugs or agents may be loaded non-uniformly in one or more materials of the ring, band or clip. In some embodiments of the invention, one or more drugs or agents may be loaded within an inner circumference of the ring, band or clip. In some embodiments of the invention, one or more drugs or agents may be loaded within an outer circumference of the ring, band or clip.
In some embodiments of the invention, one or more materials incorporated into the ring, band or clip may be “smart materials” which may alter their structure in response to one or more external factors, e.g., temperature. For example, the application of heat may cause a material, e.g., a polymer, to change shape or conformation, thereby resulting in the release of a drug or agent. In one embodiment, ultrasound, e.g., focused ultrasound, may be used to create heat needed to cause the material change shape or conformation.
In some embodiments of the invention, the ring, band or clip may comprise one or more radiopaque materials, e.g., barium sulfate, thereby making the ring, band or clip observable during fluoroscopic procedures. In some embodiments of the invention, the ring, band or clip may comprise one or more echogenic materials, e.g., perfluorocarbon, thereby making the ring, band or clip observable during ultrasound procedures.
In some embodiments of the invention, the ring, band or clip may release one or more drugs or agents via a diffusion-controlled mechanism. For example, a drug or agent may be uniformly or non-uniformly dispersed or dissolved in a material, e.g., a polymeric material, of the ring, band or clip and/or a coating of the ring, band or clip and/or a fabric covering of the ring, band or clip. The drug or agent may diffuse from an area of high concentration (e.g., from the material(s) of the band or clip) to an area of low concentration (e.g., an area of tissue such as the LAA).
In some embodiments of the invention, the ring, band or clip may release one or more drugs or agents via a biodegradable mechanism. For example, a drug or agent may be uniformly or non-uniformly dispersed or dissolved in a material, e.g., a polymeric material, of the ring, band or clip and/or a coating of the ring, band or clip and/or a fabric covering of the ring, band or clip. The drug or agent may be released during degradation of the material. The material may be designed to either degrade completely or to degrade partially, e.g., leaving the core structure of the material intact.
In some embodiments of the invention, the ring, band or clip may comprise a cross-sectional shape that may be round, square, rectangular, oval, triangular, star-shaped, etc. In some embodiments of the invention, the ring, band or clip may be reversibly placed, and its position may be adjusted if necessary. In some embodiments of the invention, multiple ring, band or clip may be placed more and more proximal to the base of the left atrial appendage.
In some embodiments of the invention, the ring, band or clip may include one or more sensors, for example to monitor changes in one or more tissue properties. One or more properties of surrounding tissue may change over time and/or in response to drug delivery, as described above, for example. In one embodiment, the band or clip may include a sensing electrode. Sensors may be monitored and/or controlled via wireless telemetry, for example, thereby providing wireless monitoring of one or more tissue properties over time.
In one embodiment, one or more sensors may comprise a biosensor, for example, comprising an immobilized biocatalyst, enzyme, immunoglobulin, bacterial, mammalian or plant tissue, cell and/or subcellular fraction of a cell. For example, the tip of a biosensor may comprise a mitochondrial fraction of a cell, thereby providing the sensor with a specific biocatalytic activity. In one embodiment, one or more sensors may be based on potentiometric technology or fiber optic technology. For example, a sensor may comprise a potentiometric or fiber optic transducer. An optical sensor may be based on either an absorbance or fluorescence measurement and may include an UV, a visible or an ER light source. In one embodiment, one or more sensors may be used to detect naturally detectable properties representative of one or more characteristics, e.g., chemical, physical or physiological, of a patient's bodily tissues or fluids. For example, naturally detectable properties of patient's bodily tissues or fluids may include pH, fluid flow, electrical current, impedance, temperature, pressure, components of metabolic processes, chemical concentrations, for example, the absence or presence of specific peptides, proteins, enzymes, gases, ions, etc. In one embodiment, one or more sensors may include one or more imaging systems, camera systems operating in UV, visible, or IR range; electrical sensors; voltage sensors; current sensors; piezoelectric sensors; electromagnetic interference (EMI) sensors; photographic plates, polymer-metal sensors; charge-coupled devices (CCDs); photo diode arrays; chemical sensors, electrochemical sensors; pressure sensors, vibration sensors, sound wave sensors; magnetic sensors; UV light sensors; visible light sensors; IR light sensors; radiation sensors; flow sensors; temperature sensors; or any other appropriate or suitable sensor. In one embodiment, one or more sensors may be powered by a suitable power source. In addition, one or more sensors may be coupled to any appropriate output device, for example, a LCD or CRT monitor which receives and displays information regarding one or more sensors.
A temperature sensor may incorporate one or more temperature-sensing elements such as, for example, thermocouples, thermisters, temperature-sensing liquid crystals, or temperature-sensing chemicals. A temperature sensor could be used, for example, to monitor tissue temperature.
The signals from one or more sensor may be amplified by a suitable amplifier before reaching an output device. The amplifier also may be incorporated into an output device. Alternatively, the amplifier may be a separate device. The output device may incorporate one or more processors. In one embodiment, sensors may be positioned around a perimeter of the band or clip. When sensed tissue reaches a perimeter, a corresponding sensor may send a signal. In one embodiment, a sensor may send constant signals. For example, a sensor may send a constant signal based on its voltage. As a tissue perimeter changes, the voltage of the sensor may change proportionately and the signal sent by the sensor may change proportionately.
In some embodiments, the shaft of the ring, band or clip delivery tool may comprise one or more flexible, bendable and/or articulation section and/or sections. One or more flexible, bendable and/or articulation section and/or sections of the shaft of the delivery tool allows the device to accommodate a variety of patient anatomies via flexing, bending and/or articulation of the delivery tool's shaft. Preferably, any flexing, bending and/or articulation of the shaft will not inhibit the opening and closing mechanism of the delivery tool for opening and closing the ring, band or clip.
In one embodiment, as shown in
In one embodiment, as shown in
In one embodiment, as shown in
In one embodiment, as shown in
Furthermore, other substitutions, modifications, changes, and omissions may be made in the design, operating conditions, and arrangements of the exemplary embodiments without departing from the scope of the invention as expressed in the appended claims. In addition, it will be appreciated by those skilled in the art that while the invention has been described above in connection with particular embodiments and examples, the invention is not necessarily so limited, and that numerous other embodiments, examples, uses, modifications and departures from the embodiments, examples and uses are intended to be encompassed by the claims attached hereto. The entire disclosure of each patent and publication cited herein is incorporated by reference, as if each such patent or publication were individually incorporated by reference herein.
This application is a continuation of U.S. application Ser. No. 11/788,261, filed Apr. 19, 2007, which is a continuation-in-part of U.S. application Ser. No. 11/155,699, filed Jun. 17, 2005, which claims the benefit of U.S. Application Ser. Nos. 60/581,223, filed on Jun. 18, 2004, and 60/685,681, filed on May 27, 2005, all of which are incorporated herein by reference in their entirety. This application also claims the benefit of U.S. Application Ser. No. 60/795,752 filed Apr. 28, 2006, the disclosure of which is also incorporated herein by reference in its entirety.
Number | Date | Country | |
---|---|---|---|
60581223 | Jun 2004 | US | |
60685681 | May 2005 | US | |
60795752 | Apr 2006 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11788261 | Apr 2007 | US |
Child | 12768200 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11155699 | Jun 2005 | US |
Child | 11788261 | US |