The invention addresses methods and instrumentation for using ultrasound vibrations of intra-capillary micro-bubbles to increase streaming of lymphatic fluid from primary cancer tumors to draining lymph-nodes (DLN) to increase the generation of tumor-specific CD8+ and CD4+ effector cells for the purpose of enhanced primary ablative and abscopal-immune-related therapy of cancer.
Currently, 5000 Norwegian men are diagnosed annually with prostate cancer (PCa), and 1000 men die from the disease each year. It is one of the most prevalent cancers and is now passing 50.000 men in Norway Cancer Registry. In the current clinical practice, there is convincing evidence that early chemotherapy can extend life by several months when applied together with castration (androgen deprivation therapy, ADT) in patients with metastatic PCa. In 2015, the CHAARTED study showed that a combination of docetaxel and castration increased the overall survival with 22 months compared to men receiving only castration in men with primary metastatic prostate cancer [1,2]. However, eventually all men will progress and succumb to metastatic castrate resistant prostate cancer. Recently, the STAMPEDE trial showed a clinical benefit when adding local radiation to the primary tumor in low tumor burden patients [3]. In this trial radiation was applied to the prostate gland only, leaving out other sites of metastatic spread.
Traditionally, the prostate gland is regarded as an immunologically privileged organ not accessible for immunotherapy, partly due to the presence of strongly immune-suppressive molecules like TGFb in prostatic cells, which hinder T-cell access and cytotoxic responses at tumor sites. However, several traits make PCa especially attractive for a combination of radiotherapy with immunotherapy, as the prostate
One major challenge consists of abandoning the barrier of immunologic tolerance of the prostate and to find solutions to circumvent how cancerous cells evade T-cell attack. Recent studies linking DNA damage with radiation-mediated immunogenicity provided mechanistic evidence that the immune-modulatory properties of radiotherapy decisively contribute to its therapeutic efficacy. Histologic dissection of the tumor microenvironment may pave the way to achieve this goal. Antigen-presenting cells (APCs) are a heterogeneous group of immune cells that mediate the cellular immune response by processing and presenting antigens for recognition by certain lymphocytes such as T cells. While depletion of regulatory T cells and checkpoint blockades are suggested to broadly license tumor APCs [4], there has been no evidence of strongly stimulatory APCs within the native tumor. Classical APCs include dendritic cells, macrophages, monocytes and B cells. Monocytes can internalize antigens, migrate to lymph nodes and differentiate to inflammatory activated monocytes that act as antigen-presenting cells. The mechanism of antigen capture in lymph nodes and delivery to follicular DC has been under investigation for many decades [5]. The tumor-associated draining lymph nodes (DLN) are essential to the generation of tumor-specific CD8+ and CD4+ effector cells. Antigens are processed by classical dendritic cells in the lymph node. Inside the lymph node antigens exposure trigger T cell responses to both CD4+ and CD8+ T cells [6].
Primary immune responses to a few tumor cells can be induced and augmented by increased tumor-associated antigen (TAA) and neo-antigen presented by professional APCs/DCs to the DLNs after radiation of the prostate. Stereotactic radiotherapy enhances cross-presentation of TAA in the DLNs as the major mechanism by which radiotherapy promotes antigen-specific immune responses and synergizes with immunotherapy [7]. Interestingly, exosomes derived from breast cancer cells treated with the topoisomerase I inhibitor topotecan were reported to produce exosomes that carry DNA capable to activate DCs in a STING-dependent way [8], suggesting that action-induced DNA damage may be coupled with the activation of innate immunity via a common pathway.
Radiation depending on the fraction size increases the expression of several antigens and expands the antigenic repertoire of cancer cells [9,10]. DNA exonuclease Trex1 is induced by radiation doses above 12-18 Gy in different cancer cells, and attenuates their immunogenicity by degrading DNA that accumulates in the cytosol upon radiation. However, the optimal degradation of dsDNA and loading of DCs varies for different cancer cells [11]. Solid tumors have high interstitial pressure which will be altered during the therapy by releasing of cell fragments. TAA will attract DC and these loaded antigen-presenting cells will return to the DLNs.
The invention presents methods and instrumentation that utilizes the extra-capillary acoustic radiation force (ARF) and micro-shear waves produced by intra-capillary vibrating micro-bubbles to increase the outflow of the interstitial lymphatic fluid with TAA-loaded DCs to the lymphatic organs constituting a backflow of primed returning cytotoxic T-cells to both the primary and metastatic tumor deposits constituting a local and an abscopal immune response.
An overview of the invention is presented, where the overview is a short form and by no means represents limitations of the invention, which in its broadest aspect is defined by the claims appended hereto.
The invention presents methods and instrumentation for increasing immune response against an existing cancer in a patient, comprising
a) first starting a primary action of the primary tumor of the patient in a way that generates professional APCs/DCs with tumor associated antigens (TAA) of the primary cancer cells in the interstitial fluid of the primary cancer region, and
b) where the start of said primary action is with a selectable delay followed by a secondary action, comprising
After the primary action, professional APCs/DCs with tumor-associated antigen (TAA) of the cancer cells can leak into the interstitial fluid of tissue close to the well-defined tumors, and the secondary action therefore addresses a region larger than the primary tumor, which we call the cancer region. The cancer region can also comprise several tumors within fairly healthy tissue, like typically found in early stage prostate cancer, or a cancer tumor can fill the whole region.
The invention further specifies
The invention further can use ultrasound to measure the vibration amplitude of the micro-bubbles, and the measured amplitude is used to determine the transmit amplitude of said ultrasound beams used to vibrate the bubbles, to obtain capillary wall vibration amplitudes within an interval.
The invention also devices a system for increasing immune response against an existing cancer in a patient, comprising
a) means for a primary action of the primary tumor of the patient in a way that generates professional APCs/DCs with tumor associated antigens (TAA) of the primary cancer cells in the interstitial fluid of the primary cancer region, and
b) means for with a selectable delay after the start of said primary action to perform a secondary action, said means comprising
The invention further describes more details of the system where
Said system further comprising,
This section gives a more detailed description of example embodiments of the invention, where the examples by no means represents limitations of the invention, which in its broadest aspect is defined by the claims appended hereto.
Radiation therapy on primary prostate cancer or in combination with CTLA-4 has in several patients shown to improve immune response towards metastatic cancer cells [13]. An hypothesis is that the radiotherapy of the primary cancer generates professional Antigen-Presenting Cells/Dendritic Cells (APCs/DCs) with tumor-associated antigen (TAA) to the primary tumor that follows lymphatic flow from the primary tumor to the draining lymph nodes (DLNs) in the patient. The radiation therapy also breaks down the tumor structure that reduces the intra-tumor pressure and improves lymphatic flow and also leaks APCs/DCs with TAAs to interstitial fluid (IF) of near surrounding tissue that also increases lymphatic flow with APCs/DCs with TAAs to the DLNs. This introduction of TAA to the DLNs stimulates the production of tumor specific CD8+ and CD4+ T-cells in the DLNs that follows the lymphatic flow to the veins and are brought back via the blood circulation to the primary tumor and also metastatic tumor residuals, and stimulate increased immune response for killing the cancer cells both in the primary and the distant metastatic deposits (abscopal effect) [12].
The invention presents methods and instrumentation for increasing immune response against an existing cancer in a patient by increasing lymphatic flow from the cancer region to the DLNs, using ultrasound vibrations of micro-bubbles in the cancer region. We define a cancer region as comprising the primary tumor with some region around of practical dimension for the action as further discussed below. The invention comprises first to provide a primary action of the primary tumor in a way that generates professional antigen-presenting cells/dendritic cells (APCs/DCs) with tumor-associated antigen (TAA) from the primary tumor in the interstitial fluid (IF) in the primary tumor and also preferentially in the IF of tissue around the primary tumor which has larger lymphatic drainage to the DLNs and helps to increase the lymphatic drainage of APCs/DCs with TAA to the DLNs. The start of the primary action is followed by a secondary action comprising i) introducing micro-bubbles into the capillary system of the cancer region, and ii) vibrating said micro bubbles with incident ultrasound beams with appropriate frequency and amplitude onto said cancer region, for the purpose of increasing lymphatic flow of APCs/DCs with TAAs from the primary tumor out of the cancer region to its DLNs. The cancer region can also comprise several tumors within fairly healthy tissue, like typically found in early stage prostate cancer, or one cancer tumor can fill the whole region, potentially also including near-by normal tissue.
The vibration of the intra-capillary micro-bubbles produces vibrations in the extra-capillary tissue that produces an outward acoustic radiation force (ARF) from the capillaries that drives fluid from the capillaries out through the tissue interstitium, increasing the interstitial pressure that further increases the flow of lymphatic fluid out from the cancer region to the DLNs. The tissue vibrations also generate micro-shear waves in the extra-capillary tissue with the potential to break elements of the extracellular matrix (ECM), and/or reduce the viscosity of the interstitial fluid (IF), for example by breaking down molecular connections within hyaluronan acid (HA), also up to HA gel. These effects further increase the flow of lymphatic fluid containing professional APCs/DCs with TAA out of the cancer region to the DLNs.
Example primary actions that produce such professional APCs/DCs with TAA from the primary tumor are stereotactic radiation-therapy, brachytherapy using implanted radioactive seeds, proton-therapy, chemo-therapy also including radiopharmaceuticals used in the same way as chemical drugs. The availability in the IF of the cancer region of such professional APCs/DCs with TAAs increases with time after the start of the primary action, and the secondary action with introduction and vibration of intra-capillary micro-bubbles can hence in many cases be done with a selectable delay after the primary action, also multiple times after the primary action and in the interval between primary actions. Vibration of intra-capillary micro-bubbles to improve drug delivery with primary chemical- or radio-pharmaceutical therapy, is currently also under investigation by Exact Therapeutics, Oslo [14-16], where we note that the secondary action according to the invention is also useful to increase the flow of lymphatic fluid containing such professional APCs/DCs with TAAs to the DLNs, also after a drug concentration in the blood is substantially reduced. The invention also devices to apply the secondary action up to multiple times between primary actions, to increase the transport of APCs/DCs with TAAs to the DLNs to further increase the immune response towards the cancer. We also note that micro-bubble vibrations in the primary tumor also increases the fluid flow from proximal capillaries into the interstitium of the tumor, further increasing the flow-loop of cytotoxic CD8+/CD4+ T-cells from the blood to the tumor cells that further enhances the immune response to the cancer. Added vibrations of micro-bubbles in the metastatic tumors can also be done to improve the transport of cytotoxic CD8+/CD4+ T-cells from the blood to the metastatic tumor cells.
To describe essential features of the invention we start with
A typical arteriolar blood pressure is ˜30 mmHg with a pressure drop of ˜15 mmHg along the capillaries to the venules. Plasma fluid containing oxygen, nutrition and other materials are transmitted as 104 across the proximal wall of capillaries to the surrounding interstitial space (IS), where the cells are bound together with a collagen matrix, the extracellular matrix (ECM). Interstitial fluid (IF) with metabolic product molecules and other are transmitted as 105 back into the distal part of the capillaries. There is hence a similar pressure drop along the IS as along the capillaries, driving the IF from the proximal region of the capillaries through the IS to the distal region of the capillaries. The Figure does not show these details. At a distance from the capillaries one have close to the average IF pressure with a typical value of PI˜22 mmHg.
In the Figure is further shown by example one lymph capillary as 106 of diameter ˜10 μm that drains fluid comprising molecules and white blood cells, called the lymphatic fluid (LF), from the region. This drainage has many functions, amongst other as a regulator of interstitial fluid volume (IFV) and interstitial pressure (IP). Lymph capillaries merges into larger lymph vessels 107, that further passes the LF through a set of draining lymph nodes 108 (DLN) and further to the blood venous system. Cancer tumors do typically have low lymphatic drainage for the center of the tumor, and leakage of the professional APCs/DCs with tumor-associated antigen (TAA) to the tumor periphery and also to the region outside the tumor, therefore increases lymphatic drainage of the APCs/DCs with TAA to the DLNs.
According to the invention, micro bubbles are introduced into the capillaries, with example methods as described below. In
The vibrations of the intra-capillary micro-bubbles set the extra-capillary tissue into vibrations, where
The ARF further interacts with the interstitial fluid (IF) in the tissue. As the IF is incompressible, the localized ARF around the capillary wall produces an IF streaming pattern throughout a large region of the IS, as shown by the simulations for the 6 μm diameter bubble in
The rapid growth of the cancer tumor leaves low density of lymph capillaries in the central region of the tumor that produces a week lymphatic drainage from the central tumor, increasing the interstitial fluid pressure in these regions. In addition the growth produces a high concentration of collagen and elastin in the extra cellular matrix (ECM) of the tumor, that also increase the intra-tumor pressure due to osmotic and capillary forces on the interstitial fluid (IF) between the collagen and elastin molecules. This intra-tumor pressure compresses both lymph and blood capillaries, also increasing both the lymphatic resistance and limiting fluid streaming from the blood capillaries. Radiation therapies and some selected chemo- or radio-pharmaceutical therapy often break down the ECM and tumor cells in the early phase of the therapy, increasing the hydraulic conductivity of the IF, and also presenting professional APCs/DCs with tumor-associated antigen (TAA) of the cancer cells into the near surrounding tissues of the tumor where the lymphatic drainage is better than in the central tumor.
Interaction between heterogeneity in tissue mass and shear stiffness on the cellular scale, and the tissue pressure waves produced by low frequency (˜500 kHz) ultrasound vibrations of the intra-capillary micro-bubbles, produces additional micro-shear waves in the extra-capillary tissue with a potential to also break down the ECM structure.
In addition, hyaluronan acid (HA) concentration often increases in diseased tissues like cancer tumors, a phenomenon that also increases the viscosity of the IF in the tumors. The concentration of HA in some cancer tumors can even be so high that the IF forms a gel. The micro shear waves described above have a potential to break down HA molecular cross-coupling, potentially to the level of breaking down HA-gel, reducing the IF viscosity. Such effects does also increase the hydraulic conductivity of the IF, increasing lymphatic outflow from the tumor, both during the limited therapy sessions with ARF from the bubble vibrations, and also for longer periods after the therapy sessions without ARF. We note that this reduction in IF viscosity can last for larger parts after the therapy vibration sessions, potentially up to 24 hr per day, 7 days a week, and hence have a larger effect on increasing the lymphatic flow from the cancer region to the DLNs, compared to the ARF that is found only during the therapy vibration sessions that are typically less than ˜1 hr with up to days apart.
In a preferred embodiment, the processor analyses the image data to determine regions for therapy beams in front of the probe. A probe 608 with 3D scanning of the beams as illustrated in
The invention is effective when the primary action produces professional APCs/DCs with tumor associated antigens (TAA) of the primary cancer cells, that following the primary action are found in the interstitial fluid (IF) of the cancer region in the patient. In the secondary action, micro-bubbles are introduced to the capillary system of the primary tumor and brought to vibrate by incident ultrasound of adequate frequency and amplitude. The micro-bubbles can be of different types, for example, marketed ultrasound contrast agent micro-bubbles, marketed contrast agent micro-bubbles selectively filtered to be above a defined diameter, and PFC-droplet bubbles as presented in [14-16].
After imaging and detecting the primary cancer region, the probe is used to generate therapy beams that vibrate the intra-capillary micro-bubbles. The micro-bubbles are typically vibrated close to or below their resonance frequency to get adequate vibration amplitude for the therapy. For smaller micro-bubbles the therapy beams can be generated by the HF array, while for the larger micro-bubbles where the resonance frequency is in the 500 kHz range, the therapy beams are typically generated with the LF array. The intra-capillary micro-bubble vibrations set up vibrations in the extra capillary tissue that produces an outward acoustic radiation force (ARF) that increases the interstitial pressure and the lymphatic outflow to the primary draining lymph nodes (DLN) of the cancer region. The tissue vibrations also produce micro-shear waves in the extra capillary tissue that can increase the hydraulic conductivity of the IF, and hence also the lymphatic outflow. Professional APCs/DCs with TAA of the cancer cells, produced by the first action with radiation- and/or chemical- or radio-pharmaceutical therapy on the primary tumor, is then found in the interstitial fluid and transported from the primary cancer region to the DLNs by the increased lymphatic flow produced by the secondary action. TAA incorporated in dendritic cells then present antigens at the lymph nodes (DLNs) for priming of tumor reactive CD8+/CD4+ T-cells leading to the expansion of these cell populations. This is essential, and as such has been successfully targeted therapeutically using GM-CSF (Dranoff, 2002) to increase presentation in the DLNs. As such, much of the focus has remained on the DLN despite our clear understanding that antigen-loading occurs within the vicinity of the tumor itself and likely influences the functions of tumor CTLs.
This increased introduction of TAA to the primary DLNs stimulates the production of CD8+ and CD4+ T-cells in the DLNs that are further transported by lymphatic flow to the venous system and brought back via the blood circulation to the primary tumor and also metastatic tumor residuals, to stimulate increased immune response for killing the cancer cells both in the primary and the distant metastatic deposits (abscopal effect). For the tumor where vibrations of the micro-bubbles are in process, the vibrating micro-bubbles also increases the fluid flow from proximal capillaries into the interstitium as in
Recent studies linking DNA damage with radiation mediated immunogenicity provided mechanistic evidence that the immune modulatory properties of radiotherapy also contribute to its therapeutic efficacy [2,3]. Within the irradiated tumor micro-environment (TME), radiotherapy initiates a cascade of molecular and cellular events leading to immunogenic cell death, up-regulation of MHC class I, enhanced type I IFN signaling and dendritic cell activation [3,6]. Ultimately, effective anti-tumor adaptive immunity requires the priming and subsequent chemokine-driven trafficking of tumor-specific CD8+ CTLs into the TME to mediate tumoricidal effector functions [7-9].
To achieve desired acoustic radiation force and micro-shear wave amplitudes, one wants the tissue vibration amplitude to be within a desired range. For a given type and dimension of the micro-bubbles there are mathematical relations between the micro-bubble vibration amplitude and the vibration amplitudes of the capillary wall and hence also the extra-capillary tissue vibration amplitudes. By example, the PFC-droplet bubble fills the capillary, so that the vibration amplitude of the capillary wall is equal to the vibration amplitude of the PFC-droplet bubble. For smaller micro-bubbles that do not fill the capillary, there are mathematical relations between micro-bubble and the capillary wall vibration amplitudes, determined by the diameters of the micro-bubble and the capillary, the micro-bubble location in the capillary, and also by the shear stiffness of the extra-capillary tissue.
A first stage in the process of defining the required ultrasound transmit amplitude to obtain adequate vibration amplitude of the extra capillary tissue, is hence to determine the required micro-bubble vibration amplitude to obtain adequate vibration amplitude of the tissue. For the PFC-droplet bubbles the capillary wall vibration and bubble vibration amplitudes are the same. For micro-bubbles with diameter less than the capillary diameter, one must first determine the shear stiffness of the tissue, which can be obtained from the well known shear wave imaging or as described in US Provisional Patent Application 63,248,180, but also from determination of the nonlinear bulk elasticity parameter, for example as described in U.S. Pat. No. 9,291,493, US Patent applications 2019/0,235,076; 2020/0,191,690; that is correlated to density of collagen, elastin, and other large molecules that also determine the shear stiffness. From a mathematical model one can then calculate the required micro-bubble vibration amplitude, and measure the ultrasound transmit amplitude that produces adequate micro-bubble vibration amplitudes. Imaging the micro-bubbles with the dual frequency ultrasound as described in the cited US patents and patent applications, one can use the high frequency (HF) (>2 MHz) image to determine the micro-bubble vibration amplitude produced by the low frequency (LF) (˜0.1-2 MHz) manipulation pulses. One should then transmit several combined LF/HF pulse complexes with different phase between the HF and LF pulse, and use the amplitude of the received HF signal for the pulse complexes with the phase lag that gives maximal signal, for example as described in US Provisional Patent Application 63,248,180. This allows adjustments of the LF transmit amplitude to obtain micro-bubble vibration amplitudes within a desired interval.
Thus, while there have shown and described and pointed out fundamental novel features of the invention as applied to preferred embodiments thereof, it will be understood that various omissions and substitutions and changes in the form and details of the devices illustrated, and in their operation, may be made by those skilled in the art without departing from the spirit of the invention.
It is also expressly intended that all combinations of those elements and/or method steps which perform substantially the same function in substantially the same way to achieve the same results, are within the scope of the invention. Moreover, it should be recognized that structures and/or elements and/or method steps shown and/or described in connection with any disclosed form or embodiment of the invention may be incorporated in any other disclosed or described or suggested form or embodiment as a general matter of design choice. It is the intention, therefore, to be limited only as indicated by the scope of the claims appended hereto.
Journal of Pharmaceuticcs, 495, 2015, 1019-1027
This application claims priority from U.S. Provisional Patent Application Ser. No. 63/088,851 which was filed on Oct. 7, 2020.
Number | Date | Country | |
---|---|---|---|
63088851 | Oct 2020 | US |