This invention relates generally to medical fluid systems, and specifically to systems for providing IV fluid to patients.
In a typical hospital setting, when a patient is in need of intravenous fluid therapy, a healthcare practitioner: (i) obtains a pre-filled IV bag that has a large port with a flexible septum; (ii) wipes an antiseptic pad across the septum; and (iii) pierces the septum of the bag with a large, hollow spike. The septum forms a seal around the spike. The spike is connected to empty medical tubing, which in turn may be connected to a fluid connector.
The medical tubing is configured to convey medical fluid from the IV bag to the patient; however, before the tubing can be connected in fluid communication with the patient, it must be “primed” with liquid to avoid introducing air or vapors from the liquid into the patient's vasculature, which could be very harmful. To prime the tubing, the healthcare practitioner positions the IV bag at a higher location than the tubing, and sometimes squeezes the IV bag, permitting gravity and liquid pressure to drive the liquid in the IV bag downward in a column through the tubing, pushing the air and vapors in the tubing out the bottom end of the tubing. When the column of liquid reaches the bottom end of the tubing, a small amount of liquid is typically permitted to drip out of the end of the tubing, or out of the end of a connector attached to the end of the tubing, into a trashcan or sink, to ensure that no residual air or vapors remain in the fluid line.
The process of spiking the IV bag and priming the fluid line requires significant time and attention from the healthcare practitioner at the patient care site, and also requires that multiple components be inventoried at the patient care site. In addition, there are many inherent disadvantages with the bag-spiking process. The tip of the spike is sharp and can inadvertently puncture the healthcare provider or patient. The spike or the septum of the spike port can carry microbes, if not properly disinfected, that may be introduced into the IV fluid when attached. The connection between the spike and spike port may leak if connected improperly. Also, the priming process can permit a small amount of vapors and liquid to escape, which can be dangerous, especially if the liquid is potential harmful, such as chemotherapy or immunosuppressive drugs.
Moreover, in many healthcare treatment settings, a large number of IV bags need to be filled with customized amounts and types of drugs on a daily basis for many different patients. If performed manually, this repetitive task is susceptible to mistakes by healthcare technicians. It can also continuously expose healthcare technicians to harmful medicinal vapors, and it can be a physically difficult chore for a healthcare technician, especially when using a large-diameter filling syringe in which the syringe plunger is especially wide and hard to push forward and back.
In some embodiments, a portable electronic fluid dispensing system can provide pre-filled, pre-primed IV bag assemblies comprising therapeutic fluids in particular amounts and/or particular concentrations that are custom-made for a particular patient. The IV bag assembly in some embodiments can form a part of a substantially entirely closed fluid system. The same fluid line and connector in the IV bag assembly can be used to fill the IV bag as is used to withdraw fluid from the IV bag and infuse such fluid into the patient. The connection points along the IV assembly, such as the connection between the IV bag and a drip chamber, or between an IV bag and tubing, or between a drip chamber and tubing, or between tubing and a closeable, resealable connector, can each be resistant to disconnection by a user. In some embodiments, the healthcare practitioner at the patient care site is not required to attach any of these components to each other, and/or the healthcare practitioner at the patient care site is not required to introduce a spike into a spike port on the IV bag.
The components, systems, assemblies, and methods that are illustrated or described in this specification are examples. Any component, structure, feature, or step illustrated or described in one embodiment can be used in additional to or instead of any component, structure, feature, or step illustrated or described in another embodiment. There is no component, structure, feature, or step in this specification that is essential or indispensable to the invention.
As shown in
The electronic fluid delivery system 100 can be configured to receive one or more commands from a user through the user input system 100, or in some other way. The user command or commands can initiate a delivery of a particular type of one or more medications in a particular volume or concentration. Upon receiving a liquid delivery command, the electronic fluid delivery system 100 can cause the drive system 190 to actuate the liquid pump and/or measurer 140 to withdraw an amount of liquid from a liquid source container 120, move it through a fluid pathway or channel in the electronic fluid delivery system 100, and push it into a fluid destination container, such as an IV assembly 200, as shown. The fluid pathway or channel can comprise a number of different structures of the electronic fluid delivery system 100 through which liquid passes as it progresses from the liquid source container 120 to the fluid destination container 200.
An adapter, such as a vial adapter 165, as shown, can provide an interface between the liquid source container 120 and the fluid channel. The adapter can comprise a hollow spike (not shown) for penetrating a septum on a liquid vial to access the liquid contained in the vial. In some embodiments, a plurality of different liquid source containers 120, fluid channels, pumps, and/or liquid destination containers 130, 210 can be provided. In some embodiments, a plurality of different liquid source containers 120 with different medical fluids, such as different types of drugs, can be available to infuse in a specified combination into the same destination container 130, 210.
In some embodiments, a support 180 can be provided for supporting a liquid destination container. As illustrated, the support 180 can be oriented in a substantially horizontal position, but many other types of supports can be used, including substantially vertical supports such as hangers. The support 180 can provide a platform and/or a location to hold onto or secure an IV assembly 200 comprising an IV bag 210, a drip chamber 230, a tubing 220, and a fluid-line connector, such as a closeable male luer connector 150, as shown.
The IV assembly 200 can be removably attached to and/or in fluid communication with the fluid pathway or channel in the electronic fluid delivery system 100, which can include one or a plurality of an adapter, a pump, a connector (such as a resealable, needleless male or female luer connector 150, 160), tubing, and a removable, disposable fluid transfer cartridge 170. None of the foregoing components is essential or indispensable; rather, various other types of components can be used instead of or in addition to those listed. For example, in some embodiments, an open-ended male or female fluid connector can be used instead of a resealable, needleless connector, especially in embodiments involving benign fluids with clamps or stop-cocks in the fluid line. Various components of the fluid channel in the electronic fluid delivery system 100 can be removed and replaced, such as the liquid source container 120, the IV assembly 200, the syringe pump 140, and the cartridge 170, which can each form a portion of the fluid channel when attached. Each of these components can form a repeatably resealable connection with the fluid channel, and when detached from the fluid channel, each of these components can comprise a sealed end that prevents conveyed or residual fluid from leaking out after disconnection. The resealable connections at each point of removable attachment in the fluid channel can be made using a series of male and female needleless luer connectors 150, 160.
The electronic fluid delivery system 100 can be positioned in a fume hood and the fluid transfer can be performed in the fume hood; however, in some embodiments, as shown, any fumes and fluid are maintained inside of a closed system, so it is not required to use the system 100 in a fume hood. In some embodiments, fluid progresses under the influence of back and forth motion of the fluid pump 140, in a substantially closed system from a liquid source container 120, through the fluid channel of the electronic fluid transfer cartridge 170, and to the fluid destination container 200, which substantially entirely prevents liquids and vapors from escaping during fluid transfer. As used herein, unless otherwise indicated, the terms “substantially entirely prevents,” or “substantially entirely closed” or “substantially entirely closed system” and related or similar terms refer to processes and systems in which vapors and liquids are maintained inside a bounded region during normal usage and operating conditions to the degree that it is needed to avoid clinically significant adverse effects on healthcare practitioners or patients. Also, in some embodiments, each removable component (such as the cartridge 170, or the source container 120 and vial adapter 165, or the IV assembly 200) that temporarily forms part of, or that is temporarily attached to, the fluid channel in the electronic fluid transfer system 100 can separately form a resealable, substantially entirely closed fluid system within itself, which substantially entirely prevents liquids and vapors from escaping during fluid transfer, and which can dramatically lower or eliminate the amount and frequency of exposure of healthcare practitioners to harmful liquids and vapors during preparation and transfer of harmful fluids.
The fluid transfer cartridge 170 can comprise a disposable portion of the fluid channel of the electronic fluid delivery system 100 that selectively directs fluid out of one or more fluid source containers 120 and into one or more fluid destination containers 210. The fluid transfer cartridge 170 may include one or more internal valves (not shown) to permit liquid to move in one direction away from a fluid source container 120 and toward a fluid destination container 210, while resisting unintended flow in an opposing direction. As illustrated, the fluid transfer cartridge 170 can comprise one or more closeable, resealable, needleless connectors, such as closeable male and female connectors 150, 160, that are configured to connect to other portions of the fluid channel within the electronic fluid delivery system, such as an adapter on the fluid source container 120 and a connector on the fluid destination container 210. The fluid transfer cartridge can be supported or held in place by a holder 195. As used in this specification, unless otherwise specified, the terms “connect,” “attach,” “hold,” and related or similar terms are contemplated in a broad sense to refer to structures that are joined together, either directly or indirectly, to each other or to a common structure, and these terms can encompass parts that are either separable or integral with each other.
In some embodiments, as illustrated, the electronic fluid delivery system 100 can comprise multiple fluid pathways or channels for delivering different types of medicinal liquids. For example, a first pathway or channel can be configured to supply a first type of liquid (e.g., an inert or benign liquid such as water, saline solution, electrolyte liquid, or other diluent) from a first liquid source container 120, through a first fluid line 175, through a first pump or measurer 140, to a first fluid destination container 130, and a second pathway or channel can be configured to supply a second type of liquid (e.g., one or more drugs or combinations of drugs, such as a chemotherapy drug, an antibiotic, an immunosuppressive drug, and/or a pain management drug, etc.) from a second liquid source container 120, through a second fluid pathway to a second pump or measurer 140, to a second fluid destination container, such as the IV assembly 200. In some embodiments, both the first and second fluid pathways or channels can be used to fill the same fluid destination container, such as the IV assembly 200, either serially or in parallel, and/or through the same fluid port on the fluid destination container, or through multiple, different fluid ports on the same fluid destination container.
Referring to
In some cases, if the fluids are passed through a single fluid line and connector, different types of fluids can be used to fill the IV assembly 200 in series (e.g., first the active drugs and/or potentially harmful drugs, and second the benign liquids such as water, saline, or other diluent). In some cases, if the fluids are passed through different fluid lines and connectors, different types of fluids can be used to fill the IV assembly 200 in parallel or generally simultaneously, and such an arrangement can in some embodiments accomplish the filling in a shorter amount of time. Further, in some cases, filling different drugs through different ports in the same IV assembly 200 can help to achieve a more thorough mixing and generally uniform dilution and concentration of the liquids in the IV bag. Also, when the active or potentially harmful drugs are passed through the alternative or non-patient infusion line 160, 250, parallel infusion can help to avoid positioning harmful drugs in the leading fluid connector 150 or infusion fluid tubing 220 during transportation and initial opening of the fluid line in the IV assembly at the patient care site. In some embodiments, as shown, the IV assembly 200 can comprise a drip chamber 230 in the patient infusion fluid line.
In the illustrated embodiment, the drip chamber 230 is connected, or bonded directly, permanently, and irreversibly, to a bottom opening in the IV bag 210 at the IV bag-drip chamber bond 260, the long tubing 300 is connected, or bonded directly, permanently, and irreversibly, to a bottom opening in the drip chamber 230 at the drip chamber-tubing bond 290, and a needleless closeable connector, such as a closeable male luer connector 150 or a closeable female luer connector (not shown in
In some embodiments, the tubing 220 can be sufficiently long so as to extend comfortably, without creating a disruptive degree of fluid-line tension and without unduly constraining the patient's movement, from a typical position on or near the IV bag 210 in a hanging position on an IV pole (see
As illustrated, in some embodiments, the fluid pathway within the IV assembly 200 is configured to be open and in continuous fluid communication from a location within the IV bag 210, through the drip chamber 230 and/or tubing 220, to the closeable connector 150, without any blockage, valve, obstruction and/or resistance to fluid flow. In some embodiments, the fluid pathway in the IV assembly 200 is integral: the connections between the components forming the fluid pathway within the IV assembly 200, from a location within the IV bag 210, through the drip chamber 230 and/or tubing 220, to the closeable connector 150, are configured to be permanent and/or to resist disconnection by a user at any point within the fluid delivery pathway within the IV assembly 200 from the IV bag 210 to the patient.
A first connector 150 can be a closeable male luer connector that is configured to prevent fluid from escaping from or entering into the connector when it is not engaged with a corresponding female connector 160, but that allows fluid to flow when it is engaged with a corresponding female connector 160. In the embodiments shown, the first connector 150 is a version of the Spiros® closeable male connector manufactured by ICU Medical, Inc., of San Clemente, Calif. The first connector 150 can be configured to be attachable to a second connector 160, which is illustrated as a version of the Clave® closeable female connector, also manufactured by ICU Medical, Inc.
The first connector 150 can include a housing 398, a valve member 400, a resilient member 402, a sealing ring 404, an end cap 406, and an O-ring 407. The housing 398 can be generally tubular in shape, and can include a passageway 408 that extends axially through the housing. The housing 398 can include a male luer tip 410 that connects to the rest of the housing 398 at a base 412. The luer tip 410 can be generally tubular in shape, and the luer tip 410 can include a hole 414 at its end, providing access to the passageway 408. As illustrated, in some embodiments, connectors can be used to substantially entirely prevent fluid therein to leak, vaporize, or otherwise escape through apertures in the fluid pathway when the connectors are closed.
The luer tip 410 can be generally surrounded by a shroud 418 as illustrated. In some embodiments, the luer tip 410 extends a distance beyond the edge 420 of the shroud. The shroud 418 can include inner threads 422 on its interior surface. The inner threads 422 can be used for securing a female connector 160. The shroud can include an indented portion 424 that has a smaller outer diameter than the other portions of the housing. The indented portion 424 can be configured to engage a portion of the resilient member 402.
The housing 398 can be constructed from a variety of materials. The housing 398 can be constructed from a rigid material such as polycarbonate or other polymeric materials. In some embodiments, the housing 398 can be constructed from a hydrophobic material such as Bayer Makrolon, or any other suitable material. In some embodiments, the housing 398 can be formed from a substantially transparent material.
The connector 150 can include a fluid passageway 440 extending axially from an opening formed in a base portion 444 and into a tube 446. In some embodiments, the passageway 440 can be wider in the base portion 444 than in the tube 446. In some embodiments, the tube 446 includes a narrowed tip 448. In some embodiments, the tip 448 can have a tapered outer surface. In some embodiments, the tip 448 can be made from a flexible or compressible material, such as silicone rubber, to facilitate formation of the fluid seal between the tip 448 and the shelf 416. In some embodiments, the tube can include one or more holes 450 for providing access to the fluid passageway 440. The holes 450 can be formed, for example, in the tip 448 of the tube 446.
In some embodiments, the valve member 400 can include two struts 452a, 452b extending out from the base 444 and positioned on either side of tube 446, so that an open space is defined on either side of the tube. In some embodiments, the tube 446 can extend axially past the ends of the struts 452a, 452b. The valve member 400 can be constructed from a variety of materials, such as polycarbonate or other polymeric materials. In some embodiments, the valve member 400 can be constructed from the same material as the housing 398. In some embodiments, the valve member 400 and housing 398 can be constructed from different materials. In some embodiments, the valve member 400 can be constructed from multiple materials or from multiple pieces. For example, the tip 448 can be constructed from a material that is more flexible than the remainder of the valve member 400. In some embodiments, the valve member 400 can be formed from a substantially opaque material.
The resilient member 402 can include a first ring and a second ring connected to each other by longitudinally extending elastic members (not shown). The elastic members can be made from an elastic material that exerts a restoring force when stretched, such as silicone. Thus, if the rings are pulled apart, the elastic members function to restore the rings to their relaxed configuration. In some embodiments, the rings are also constructed from an elastic material, such as the same material used to form the elastic members. In some embodiments, the second ring can have a greater diameter than the first ring. In some embodiments, the second ring can have a tapered outer surface so that the end of the second ring that is closest to the first ring is wider than the end of the second ring that is furthest from the first ring.
The sealing ring 404 can be generally cylindrical in shape, and can have a bore extending axially therethrough. The sealing ring 404 can have a cylindrical body section and an O-ring located at one end of the body section 468. The sealing ring 404 can be constructed from a variety of materials. In some embodiments, the sealing ring 404 can be constructed from a deformable or elastic material such as a silicone rubber.
The end cap 406 can include a first end cap member 405 and a second end cap member 409. The second end cap member 409 can include an attachment portion, a plunger, and a disk portion located between the attachment portion and the plunger. The second end cap member 409 can have a fluid passageway axially positioned therein. In some embodiments, the plunger can be generally tubular in shape. In some embodiments, the outer surface of the plunger includes an indented region, which can be configured to receive the O-ring 407 therein.
The O-ring 407 can be constructed from an elastic material such as silicone rubber so that it can be stretched over the edge of the plunger and be seated in the indented region. In some embodiments, the O-ring can be constructed from the same material as the resilient member 402 and/or the sealing ring. In some embodiments, the O-ring can be sized so that when seated in the indented region, the thickest portion of the O-ring 407 extends radially outwardly a distance past the outer surface of the plunger.
The valve member 400 can be slidably inserted into the housing 398 so that the tube 446 enters the passageway 408. The narrowed tip 448 of the tube 446 can pass through the bore of the sealing ring 404 and into the male luer tip 410 until it abuts against the shelf 416. The tube 446 can have a width that substantially fills the bore 446 and presses against the O-ring 470 portion of the sealing ring 404 to form a fluid seal therebetween. The struts 452a, 452b can pass through holes in the housing 398, so that the struts 452a, 452b are positioned between the male luer tip 410 and the shroud 418.
The resilient member 402 can function to bias the valve member 400 against the housing 398. The first ring 460 can fit onto the lower portion 458 of the base 444 of the valve member 400, so that a surface of the ring 460 abuts against the protrusions 454. The second ring 462 can fit into the indented portion 424 of the housing. The elastic members can be positioned in channels.
The first connector 150 can be configured to engage a second connector 160. In some embodiments, the second connector 160 can be a closeable, resealable, needleless female connector, as illustrated. A variety of types of female connectors 332 can be used. The female connector 160 shown includes a housing 490, a spike 492, a base 494, and a resilient seal element 496. A fluid passageway 498 can pass through the base 494 and through the spike 492. The spike 492 can include one or more holes 500 providing fluid communication between the passageway 498 and the area outside the spike 492. The seal element 496 can be shaped and positioned to substantially surround the spike 492. The seal element 496 can include a closable aperture 502 or slit that can open to allow the tip of the spike 492 to pass through then end of the seal element 496 when the seal element 496 is compressed. The housing can include external threads 504 configured to engage the inner threads 422 on the housing 398 of the male connector 150. An end of the tubing 334 can be connected to the end of the female connector 160 by an adhesive, clamp, friction or pressure fitting, or other suitable manner to form a fluid tight connection.
In some embodiments, the housing 398, sealing ring 404, resilient member 402, valve member 400, and first end cap member 405 can rotate about the longitudinal axis with respect to the second end cap member 409. Thus, as the female connector 160 of the IV bag assembly is attached to the male connector 150, the female connector 160 can be held still while the housing 398 of the male connector 150 can rotate causing the threads 504, 422 to engage. Since the female connector 322 is not required to rotate during engagement and disengagement with the male connector 150, the tubing 334 can avoid being twisted or kinked and the user is not required to twist the IV bag to accommodate rotation of the female connector 322.
When not engaged with the female connector 160, the male connector 150 can be sealed. In some embodiments, fluid can enter the male connector 150 at the male connector 352 and pass through the passageway 478 of the end cap 406, through the passageway 440 of the valve member 400, through the holes 450, and into the portion of the passageway 408 of the male luer tip 410.
When the male connector 150 is engaged with the female connector 160, the external threads 504 of the female luer connector 332 can engage the inner threads 422 on the shroud 418, securing the female connector 160 to the male connector 150. The edge of the male luer tip 410 can press against and compress the resilient seal element 496 so that the spike 492 passes through the aperture 502 until the holes 500 are exposed. The end of the housing 490 of the female luer connector 332 can enter the space between the male luer tip 410 and the shroud 418 until it contacts the struts 452a, 452b. As the female luer connector 332 further engages the male connector 150, it can push on the struts 452a, 452b causing the entire valve member 400 to retract. As the valve member 400 retracts, the elastic members 464a, 464b of the resilient member 402 stretch. When the valve member 400 retracts, the tip 448 disengages from the shelf 416, breaking the fluid seal and allowing fluid pass from the passageway 408 in the housing 398 of the male connector 150 to the passageway 498 in the female connector 160 via the holes 500. When engaged, the resilient seal element 496 exerts a restoring force toward the male connector 150 that presses the end of the seal element 496 against the end of the male luer tip 410, forming a fluid seal therebetween. Thus, the fluid can be kept isolated from the external environment while it is transferred from the male connector 150 to the female connector 160.
The female connector 160 can be disengaged from the male connector 150. The restoring force exerted by the resilient seal element 496 of the female connector 160 can cause it to return to its closed position, sealing off its passageway 498. The elastic members of the resilient member 402 can exert a restoring force on the valve member 400, causing the valve member 400 to return to its closed position with its tip 448 abutted against the shelf 416 as the female connector 160 is disengaged. Any suitable connection can be provided at the fluid line connection points illustrated and/or described in this specification, including but not limited to those illustrated and/or described in connection with
Returning to
(i) obtaining, or instructing a user to obtain, a request from a healthcare practitioner or from a computer system for a custom-filled IV bag or a plurality of IV bags, each with a particular drug or drugs, and/or a particular dose or concentration of a drug or drugs, for a particular patient or patients;
(ii) obtaining, or instructing a user to obtain, an integral IV assembly 200 with one or more permanent, pre-attached bonds along an internal fluid path from an IV bag 210 through a resealable needleless connector 150, the IV assembly initially being empty of liquid or not completely filled with liquid;
(iii) attaching, or instructing a user to attach, one or more fluid inlets in the IV assembly 200 to one or more fluid dispensing outlets in the electronic fluid dispensing system 100, such as attaching a connector (e.g., a resealable, needleless male connector 150) of the IV assembly 200 to at least a first fluid channel of an electronic fluid dispensing system 100, such as by way of a complimentary connector 160 (e.g., a resealable, needleless female connector 160) in a disposable fluid cartridge 170 of the electronic fluid dispensing system 100, or by way of a suitable fluid connection adaptor, in the event that the connectors are not capable of forming a fluid connection (such as when a male-to-male connection or a female-to-female connection needs to be made);
(iv) providing, or instructing a user to provide, one or more user commands to the electronic fluid dispensing system to enable the transfer of a specific amount and/or concentration of a liquid or liquids from one or more fluid source containers 120 to the IV bag 210, such as by first passing fluid from at least one fluid source container 120 through a combination of one or more of an adaptor 165, one or more connectors 150, 160, a fluid cartridge 170, a liquid pump and/or measurer 140, and/or one or more connectors 150, 160, to the IV assembly 200, then permitting the fluid to continue progressing into the connector 150 of the IV assembly 200, into the long tubing 220 of the IV assembly 200, through the drip chamber 230 (if included) of the IV assembly 200, and into the IV bag 210, until the desired amount and/or concentration of the drug or drugs is received in the IV bag 210 (as illustrated, for example, in
(v) detaching, or instructing a user to detach, the IV assembly 200 from the electronic fluid dispensing system 100, such as by detaching the connector 150 of the IV assembly 200 from the corresponding portion of the fluid channel of the electronic fluid dispensing system, and detaching any other fluid connections from any other fluid channels;
(vi) transporting, or instructing a user to transport, the custom-filled IV assembly 200 to a patient care site;
(vii) hanging or otherwise positioning, or instructing a user to hang or otherwise position, the IV bag 220 on an IV pole near a patient, such as by using a hanger 240 on the IV bag 220, or positioning the IV bag 220 on or in another type of holder; and/or
(iix) extending, or instructing a user to extend, the pre-primed long tubing 220 from the hanging IV bag to an infusion pump (not shown), which is in turn configured to pump fluid to a patient, or extending, or instructing a user to extend, the pre-primed long tubing 220 from the IV bag directly to an infusion site near a patient, and attaching the pre-primed, connector 150 of the IV assembly 200 to a complimentary connector 160 on a fluid line that leads into the fluid pump or directly into the patient's vasculature.
As illustrated in
All of the foregoing steps, or any combination of these steps, can be performed in some embodiments without requiring a healthcare practitioner to penetrate a spike-port on an IV bag with a hollow spike leading to a patient infusion line, and/or without requiring a healthcare practitioner to prime the fluid tubing, drip chamber (if used), and/or fluid connector (if used), of an IV assembly at or near the patient administration site and outside of a hazardous fume hood. As illustrated, in some embodiments, an IV bag can be provided without any spike-port at all or without any spike-port that has a septum configured to be pierced by a sharp spike. Rather, in some embodiments, the same fluid line in the IV assembly 200 can be used to fill the IV bag 210 with liquid in an upstream direction from the connector 150, through the long tubing 220 and drip chamber 230 (if used), to the IV bag 210, and then later to infuse the liquid from the IV bag 210 in a downstream direction through the same drip chamber 230 (if used), long tubing 220, and connector 150, to the patient injection site. The movement of liquid within the electronic fluid dispensing system 100 and/or the filling of the IV assembly 200 can be accomplished with any pump, such as a syringe pump or a peristaltic pump. In some embodiments, a syringe pump or a peristaltic pump is not used.
In some embodiments, during the IV assembly filling stage, after a desired quantity of therapeutic agent is delivered to the IV assembly 200, an additional volume of liquid, such as an inert or benign liquid, can be pushed through or made to “chase” the drug through the lower fluid pathway of the IV assembly 200, such as through the bottom connector 150, the tubing 520, the drip chamber 530 (if used), and/or into the IV bag 210, so as to flush or move any potentially harmful liquid or vapors an effective distance into the IV assembly 200 and away from persons who are transporting, storing, and/or administering the therapeutic liquid to a patient.
This application is a continuation application of U.S. patent application Ser. No. 14/746,048, filed Jun. 22, 2015, pending, which claims the benefit under 35 U.S.C. § 120 and 35 U.S.C. § 365(c) as a continuation of International Application No. PCT/US2014/065972, designating the United States, with an international filing date of Nov. 17, 2014, entitled “Methods and Systems for Filling IV Bags with Therapeutic Fluid,” which claims the benefit of U.S. Provisional Patent Application No. 61/908,674, filed on Nov. 25, 2013, and entitled “Method and Systems for Filling IV Bags with Therapeutic Fluid.” The entirety of each of the above-mentioned applications is hereby incorporated by reference herein and made a part of this disclosure.
Number | Name | Date | Kind |
---|---|---|---|
1923501 | Perry | Aug 1933 | A |
3157201 | Littmann | Nov 1964 | A |
3344785 | Hamilton | Oct 1967 | A |
D222956 | Sato | Feb 1972 | S |
D222957 | Sato | Feb 1972 | S |
D236163 | Manno | Jul 1975 | S |
4005710 | Zeddies et al. | Feb 1977 | A |
4084606 | Mittleman | Apr 1978 | A |
4187890 | Stach et al. | Feb 1980 | A |
4190048 | Sampson | Feb 1980 | A |
4262671 | Kersten | Apr 1981 | A |
4306705 | Svensson | Dec 1981 | A |
4336802 | Stone et al. | Jun 1982 | A |
4367736 | Gupton | Jan 1983 | A |
D268206 | Kosako | Mar 1983 | S |
D268284 | Manno et al. | Mar 1983 | S |
4397335 | Doblar et al. | Aug 1983 | A |
4410321 | Pearson et al. | Oct 1983 | A |
4423741 | Levy | Jan 1984 | A |
4519792 | Dawe | May 1985 | A |
4534758 | Akers et al. | Aug 1985 | A |
4559043 | Whitehouse et al. | Dec 1985 | A |
4561856 | Cochran | Dec 1985 | A |
4666429 | Stone | May 1987 | A |
4670007 | Wheeldon et al. | Jun 1987 | A |
4683916 | Raines | Aug 1987 | A |
4755172 | Baldwin | Jul 1988 | A |
4759756 | Forman et al. | Jul 1988 | A |
4768568 | Fournier et al. | Sep 1988 | A |
4778450 | Kamen | Oct 1988 | A |
4819684 | Zaugg et al. | Apr 1989 | A |
4863429 | Baldwin | Sep 1989 | A |
D305165 | Rudolph et al. | Dec 1989 | S |
4922975 | Polaschegg | May 1990 | A |
4936841 | Aoki et al. | Jun 1990 | A |
4969874 | Michel et al. | Nov 1990 | A |
4972876 | Kabata et al. | Nov 1990 | A |
4976590 | Baldwin | Dec 1990 | A |
4995268 | Ash et al. | Feb 1991 | A |
5024347 | Baldwin | Jun 1991 | A |
5037390 | Raines et al. | Aug 1991 | A |
5114580 | Ahmad et al. | May 1992 | A |
D328952 | Arioka | Aug 1992 | S |
5176658 | Ranford | Jan 1993 | A |
5224937 | van der Heiden et al. | Jul 1993 | A |
5254096 | Rondelet et al. | Oct 1993 | A |
5256155 | Yerlikaya et al. | Oct 1993 | A |
5288290 | Brody | Feb 1994 | A |
5300044 | Classey et al. | Apr 1994 | A |
D348101 | Poli et al. | Jun 1994 | S |
5334211 | Shiber | Aug 1994 | A |
5336201 | von der Decken | Aug 1994 | A |
D352778 | Irvin | Nov 1994 | S |
5378231 | Johnson et al. | Jan 1995 | A |
5405333 | Richmond | Apr 1995 | A |
5423791 | Bartlett | Jun 1995 | A |
5431201 | Torchia et al. | Jun 1995 | A |
5439451 | Collinson et al. | Aug 1995 | A |
5466220 | Brenneman | Nov 1995 | A |
5609572 | Lang | Mar 1997 | A |
5645538 | Richmond | Jul 1997 | A |
5647845 | Haber et al. | Jul 1997 | A |
5676346 | Leinsing | Oct 1997 | A |
5685866 | Lopez | Nov 1997 | A |
5776345 | Truitt et al. | Jul 1998 | A |
5782816 | Werschmidt et al. | Jul 1998 | A |
5807312 | Dzwonkiewicz | Sep 1998 | A |
5810792 | Fangrow, Jr. et al. | Sep 1998 | A |
5830185 | Block, Jr. | Nov 1998 | A |
5871110 | Grimard et al. | Feb 1999 | A |
5871500 | Jepson et al. | Feb 1999 | A |
5885270 | Ortiz et al. | Mar 1999 | A |
D408079 | Ellis | Apr 1999 | S |
5897526 | Vaillancourt | Apr 1999 | A |
5904666 | DeDecker et al. | May 1999 | A |
5910252 | Truitt et al. | Jun 1999 | A |
5935106 | Olsen | Aug 1999 | A |
5947951 | Ortiz et al. | Sep 1999 | A |
5968014 | Neftel et al. | Oct 1999 | A |
5989237 | Fowles et al. | Nov 1999 | A |
6059747 | Bruggeman et al. | May 2000 | A |
6110153 | Davis et al. | Aug 2000 | A |
RE36871 | Epstein et al. | Sep 2000 | E |
6123685 | Reynolds | Sep 2000 | A |
6132404 | Lopez | Oct 2000 | A |
6152900 | Mayer | Nov 2000 | A |
6171484 | Schnell et al. | Jan 2001 | B1 |
6179823 | Niedospial, Jr. | Jan 2001 | B1 |
6193675 | Kraus et al. | Feb 2001 | B1 |
6193689 | Woodard | Feb 2001 | B1 |
6202708 | Bynum | Mar 2001 | B1 |
6221041 | Russo | Apr 2001 | B1 |
6245048 | Fangrow, Jr. et al. | Jun 2001 | B1 |
6287289 | Niedospial, Jr. | Sep 2001 | B1 |
6302864 | Nowosielski | Oct 2001 | B1 |
6425497 | Chu et al. | Jul 2002 | B1 |
6474375 | Spero et al. | Nov 2002 | B2 |
6485472 | Richmond | Nov 2002 | B1 |
6551299 | Miyoshi et al. | Apr 2003 | B2 |
6558365 | Zinger et al. | May 2003 | B2 |
6572256 | Seaton et al. | Jun 2003 | B2 |
6585229 | Cote, Sr. et al. | Jul 2003 | B2 |
6590167 | Clare | Jul 2003 | B2 |
6599273 | Lopez | Jul 2003 | B1 |
6623455 | Small et al. | Sep 2003 | B2 |
6629956 | Polidoro et al. | Oct 2003 | B1 |
6651956 | Miller | Nov 2003 | B2 |
6663586 | Verkaart et al. | Dec 2003 | B2 |
6689108 | Lavi et al. | Feb 2004 | B2 |
6699230 | Jaafar et al. | Mar 2004 | B2 |
6711460 | Reese | Mar 2004 | B1 |
6726672 | Hanly et al. | Apr 2004 | B1 |
6793651 | Bennett et al. | Sep 2004 | B1 |
6813868 | Baldwin et al. | Nov 2004 | B2 |
6854620 | Ramet | Feb 2005 | B2 |
6877530 | Osborne et al. | Apr 2005 | B2 |
6908459 | Harding et al. | Jun 2005 | B2 |
6915823 | Osborne et al. | Jul 2005 | B2 |
6948522 | Newbrough et al. | Sep 2005 | B2 |
6953450 | Baldwin et al. | Oct 2005 | B2 |
6985870 | Martucci et al. | Jan 2006 | B2 |
6991002 | Osborne et al. | Jan 2006 | B2 |
6994315 | Ryan et al. | Feb 2006 | B2 |
6997917 | Niedospial, Jr. et al. | Feb 2006 | B2 |
7006894 | De La Huerga | Feb 2006 | B2 |
7017623 | Tribble et al. | Mar 2006 | B2 |
7086431 | D'Antonio et al. | Aug 2006 | B2 |
7108024 | Navarro | Sep 2006 | B2 |
7117901 | Martinell Gisper-Sauch et al. | Oct 2006 | B2 |
7117902 | Osborne | Oct 2006 | B2 |
7128105 | Tribble et al. | Oct 2006 | B2 |
7163031 | Graves et al. | Jan 2007 | B2 |
7163035 | Khan et al. | Jan 2007 | B2 |
7175615 | Hanly et al. | Feb 2007 | B2 |
7194336 | DiGianfilippo et al. | Mar 2007 | B2 |
7260447 | Osborne | Aug 2007 | B2 |
7317967 | DiGianfilippo et al. | Jan 2008 | B2 |
7343224 | DiGianfilippo et al. | Mar 2008 | B2 |
7343943 | Khan et al. | Mar 2008 | B2 |
7351226 | Herskowitz | Apr 2008 | B1 |
7354426 | Young | Apr 2008 | B2 |
7392638 | Baldwin et al. | Jul 2008 | B2 |
7396051 | Baldwin et al. | Jul 2008 | B2 |
7398802 | Baker | Jul 2008 | B2 |
7418981 | Baker et al. | Sep 2008 | B2 |
7442186 | Blomquist | Oct 2008 | B2 |
7488311 | Domkowski et al. | Feb 2009 | B2 |
7499581 | Tribble et al. | Mar 2009 | B2 |
7527619 | Domkowski et al. | May 2009 | B2 |
7530211 | McErlean et al. | May 2009 | B2 |
7530974 | Domkowski et al. | May 2009 | B2 |
7538858 | Mackey | May 2009 | B2 |
D594120 | Berberich et al. | Jun 2009 | S |
D596291 | Berberich et al. | Jul 2009 | S |
7566326 | Duchon et al. | Jul 2009 | B2 |
7610115 | Rob et al. | Oct 2009 | B2 |
7630788 | Reese | Dec 2009 | B1 |
7630789 | Broadfield et al. | Dec 2009 | B2 |
7632261 | Zinger et al. | Dec 2009 | B2 |
7654976 | Peterson et al. | Feb 2010 | B2 |
7681606 | Khan et al. | Mar 2010 | B2 |
7685026 | McGrady et al. | Mar 2010 | B1 |
D616092 | Domkowski et al. | May 2010 | S |
7717897 | Burg et al. | May 2010 | B2 |
D620108 | Eitenmueller et al. | Jul 2010 | S |
7753085 | Tribble et al. | Jul 2010 | B2 |
7758560 | Connell et al. | Jul 2010 | B2 |
7789850 | Roger | Sep 2010 | B2 |
7814731 | Bender et al. | Oct 2010 | B2 |
7850051 | Py et al. | Dec 2010 | B2 |
7882863 | Pestotnik et al. | Feb 2011 | B2 |
7900658 | Osborne et al. | Mar 2011 | B2 |
7913720 | Tribble et al. | Mar 2011 | B2 |
7963201 | Willoughby et al. | Jun 2011 | B2 |
7963954 | Kavazov | Jun 2011 | B2 |
7967202 | Durrell et al. | Jun 2011 | B2 |
7981381 | Lurvey et al. | Jul 2011 | B2 |
7997304 | Ranalletta et al. | Aug 2011 | B2 |
8034041 | Domkowski et al. | Oct 2011 | B2 |
8037659 | Osborne et al. | Oct 2011 | B2 |
8075545 | Moy et al. | Dec 2011 | B2 |
8091727 | Domkowski | Jan 2012 | B2 |
8091860 | Thompson et al. | Jan 2012 | B2 |
8104644 | Py et al. | Jan 2012 | B2 |
8117809 | McErlean et al. | Feb 2012 | B2 |
8140351 | Tribble et al. | Mar 2012 | B2 |
8141601 | Fehr et al. | Mar 2012 | B2 |
8151835 | Khan et al. | Apr 2012 | B2 |
8162903 | Reilly et al. | Apr 2012 | B2 |
8162914 | Kraushaar et al. | Apr 2012 | B2 |
8162915 | Brandenburger et al. | Apr 2012 | B2 |
D660423 | Hermle | May 2012 | S |
8172823 | Rondeau et al. | May 2012 | B2 |
8182744 | Mlodzinski et al. | May 2012 | B2 |
8197459 | Jansen et al. | Jun 2012 | B2 |
8206367 | Warrne et al. | Jun 2012 | B2 |
D664647 | Becker | Jul 2012 | S |
D664648 | Becker | Jul 2012 | S |
D664649 | Becker | Jul 2012 | S |
8209941 | Osborne et al. | Jul 2012 | B2 |
8216207 | Moy et al. | Jul 2012 | B2 |
8220503 | Tribble et al. | Jul 2012 | B2 |
8220504 | Hartman et al. | Jul 2012 | B2 |
8221382 | Moy et al. | Jul 2012 | B2 |
8225824 | Eliuk et al. | Jul 2012 | B2 |
8225826 | Horppu et al. | Jul 2012 | B2 |
8231567 | Tennican et al. | Jul 2012 | B2 |
8231749 | Dent et al. | Jul 2012 | B2 |
8241265 | Moy et al. | Aug 2012 | B2 |
D667946 | Levesque et al. | Sep 2012 | S |
8267129 | Doherty et al. | Sep 2012 | B2 |
8267912 | Ferris | Sep 2012 | B2 |
8287513 | Ellstrom et al. | Oct 2012 | B2 |
8328082 | Bochenko et al. | Dec 2012 | B1 |
8336587 | Rosenquist et al. | Dec 2012 | B2 |
8353318 | Ranalletta et al. | Jan 2013 | B2 |
8356644 | Chong et al. | Jan 2013 | B2 |
8356645 | Chong et al. | Jan 2013 | B2 |
8357137 | Yandell | Jan 2013 | B2 |
8374887 | Alexander | Feb 2013 | B1 |
8381776 | Horppu | Feb 2013 | B2 |
8382696 | Beiriger et al. | Feb 2013 | B2 |
8386070 | Eliuk et al. | Feb 2013 | B2 |
8403905 | Yow | Mar 2013 | B2 |
8409165 | Niedospial, Jr. et al. | Apr 2013 | B2 |
8414556 | Garfield et al. | Apr 2013 | B2 |
8425487 | Beiriger et al. | Apr 2013 | B2 |
8430859 | McConnell | Apr 2013 | B2 |
8449521 | Thorne, Jr. et al. | May 2013 | B2 |
D687948 | Levesque et al. | Aug 2013 | S |
8506548 | Okiyama | Aug 2013 | B2 |
8522832 | Lopez et al. | Sep 2013 | B2 |
8551037 | Suchecki et al. | Oct 2013 | B2 |
8562583 | Akerlund et al. | Oct 2013 | B2 |
8567235 | Bojan et al. | Oct 2013 | B2 |
8571708 | Rob et al. | Oct 2013 | B2 |
8562584 | Beiriger et al. | Nov 2013 | B2 |
8602067 | Kuhni et al. | Dec 2013 | B2 |
8608723 | Lev et al. | Dec 2013 | B2 |
8622985 | Ellstrom | Jan 2014 | B2 |
8636720 | Truitt et al. | Jan 2014 | B2 |
8639525 | Levine et al. | Jan 2014 | B2 |
8679075 | Lurvey et al. | Mar 2014 | B2 |
8684994 | Lev et al. | Apr 2014 | B2 |
8701696 | Guala | Apr 2014 | B2 |
8702675 | Imai | Apr 2014 | B2 |
8720496 | Huwiler et al. | May 2014 | B2 |
8721612 | Moy et al. | May 2014 | B2 |
8721614 | Takemoto et al. | May 2014 | B2 |
8721627 | Alpert | May 2014 | B2 |
D706415 | Levesque et al. | Jun 2014 | S |
8753325 | Lev et al. | Jun 2014 | B2 |
8763798 | Paul | Jul 2014 | B2 |
8795231 | Chong et al. | Aug 2014 | B2 |
8801689 | Moy et al. | Aug 2014 | B2 |
8821436 | Mosler et al. | Sep 2014 | B2 |
8834444 | Domkowski | Sep 2014 | B2 |
8852147 | Callan et al. | Oct 2014 | B2 |
8863788 | Ranalletta et al. | Oct 2014 | B2 |
8864725 | Ranalletta et al. | Oct 2014 | B2 |
8864737 | Hasegawa et al. | Oct 2014 | B2 |
8870832 | Raday et al. | Oct 2014 | B2 |
8882739 | Domkowski et al. | Nov 2014 | B2 |
8894627 | Garfield et al. | Nov 2014 | B2 |
8911421 | Domkowski et al. | Dec 2014 | B2 |
D721803 | Dubach | Jan 2015 | S |
8926554 | Okuda et al. | Jan 2015 | B2 |
8958112 | Matsui et al. | Feb 2015 | B2 |
D724198 | Oostman et al. | Mar 2015 | S |
8973622 | Lopez et al. | Mar 2015 | B2 |
8979792 | Lev et al. | Mar 2015 | B2 |
9033006 | Perazzo et al. | May 2015 | B2 |
9043019 | Eliuk et al. | May 2015 | B2 |
9056164 | Tate et al. | Jun 2015 | B2 |
9057363 | Capone | Jun 2015 | B2 |
9057370 | Mundt et al. | Jun 2015 | B2 |
9060923 | Hossainy | Jun 2015 | B2 |
9061130 | Truitt et al. | Jun 2015 | B2 |
9076115 | Utech et al. | Jul 2015 | B2 |
9079686 | Domkowski et al. | Jul 2015 | B2 |
9089474 | Cederschiöld | Jul 2015 | B2 |
9089647 | Haenggi et al. | Jul 2015 | B2 |
9101717 | Mansour et al. | Aug 2015 | B2 |
9132062 | Fangrow | Sep 2015 | B2 |
9132063 | Lev et al. | Sep 2015 | B2 |
9139316 | Husnu et al. | Sep 2015 | B2 |
9144646 | Barron, III et al. | Sep 2015 | B2 |
9149576 | Bullington et al. | Oct 2015 | B2 |
9198832 | Moy et al. | Dec 2015 | B2 |
9211231 | Mansour et al. | Dec 2015 | B2 |
9212762 | Duncan | Dec 2015 | B2 |
9220661 | Garfield et al. | Dec 2015 | B2 |
D747472 | Bradley et al. | Jan 2016 | S |
9227048 | Frattini | Jan 2016 | B2 |
9241875 | Davis et al. | Jan 2016 | B2 |
9242039 | Valk et al. | Jan 2016 | B2 |
9270890 | Okuma et al. | Feb 2016 | B2 |
9345640 | Mosler et al. | May 2016 | B2 |
9345641 | Kraus et al. | May 2016 | B2 |
9345643 | Okiyama | May 2016 | B2 |
9381135 | Reynolds et al. | Jul 2016 | B2 |
9381137 | Garfield et al. | Jul 2016 | B2 |
9382021 | Tribble et al. | Jul 2016 | B2 |
9402786 | Petrone | Aug 2016 | B2 |
9408966 | Kamen | Aug 2016 | B2 |
9466088 | Perazzo et al. | Oct 2016 | B2 |
9474690 | Ranalletta et al. | Oct 2016 | B2 |
9475019 | Kaucky et al. | Oct 2016 | B2 |
9481477 | Kjar | Nov 2016 | B2 |
D774192 | Fuchs | Dec 2016 | S |
D775325 | Larson et al. | Dec 2016 | S |
9511989 | Lopez et al. | Dec 2016 | B2 |
9561893 | Root et al. | Feb 2017 | B2 |
9579255 | Eliuk et al. | Feb 2017 | B2 |
9615997 | Fangrow | Apr 2017 | B2 |
9744102 | Kubo | Aug 2017 | B2 |
9770388 | Noike et al. | Sep 2017 | B2 |
9775778 | Qiu et al. | Oct 2017 | B2 |
9801787 | Py | Oct 2017 | B2 |
9802171 | Konrad, Jr. et al. | Oct 2017 | B2 |
9802172 | Konrad, Jr. et al. | Oct 2017 | B2 |
D803396 | Oberkircher et al. | Nov 2017 | S |
9827163 | Lopez et al. | Nov 2017 | B2 |
9827680 | Davey et al. | Nov 2017 | B2 |
D804651 | Loonan | Dec 2017 | S |
9833605 | Sanders et al. | Dec 2017 | B2 |
9849236 | Hachey et al. | Dec 2017 | B2 |
9883987 | Lopez et al. | Feb 2018 | B2 |
9930297 | Alexander et al. | Mar 2018 | B2 |
9931276 | Lopez et al. | Apr 2018 | B2 |
D819414 | Solomon | Jun 2018 | S |
10106278 | Chang et al. | Oct 2018 | B2 |
10143985 | Brown et al. | Dec 2018 | B2 |
D837983 | Fangrow | Jan 2019 | S |
10181186 | Kriheli et al. | Jan 2019 | B2 |
10188849 | Fangrow | Jan 2019 | B2 |
10189616 | Kraft | Jan 2019 | B2 |
D846146 | Amos et al. | Apr 2019 | S |
10259608 | Fianchini et al. | Apr 2019 | B2 |
D851745 | Shauver et al. | Jun 2019 | S |
10307338 | Hellenbrand | Jun 2019 | B2 |
10314765 | Lopez et al. | Jun 2019 | B2 |
10315174 | Konrad, Jr. et al. | Jun 2019 | B2 |
10327987 | Bochenko et al. | Jun 2019 | B1 |
10327988 | Tribble et al. | Jun 2019 | B2 |
10336477 | Perazzo et al. | Jul 2019 | B2 |
10417758 | Alexander | Sep 2019 | B1 |
10420927 | Fangrow | Sep 2019 | B2 |
10494126 | Joplin | Dec 2019 | B2 |
10503873 | Prince et al. | Dec 2019 | B2 |
10512885 | Janders et al. | Dec 2019 | B2 |
D874644 | Shauver et al. | Feb 2020 | S |
10554937 | Alexander et al. | Feb 2020 | B2 |
10556062 | Simpson et al. | Feb 2020 | B2 |
10576211 | Hang et al. | Mar 2020 | B2 |
D887577 | Shor et al. | Jun 2020 | S |
10791975 | Wilkinson et al. | Oct 2020 | B2 |
D905228 | Shauver et al. | Dec 2020 | S |
11007119 | Lopez et al. | May 2021 | B2 |
11020541 | Fangrow et al. | Jun 2021 | B2 |
11033459 | Ariagno et al. | Jun 2021 | B2 |
11135416 | Fangrow | Oct 2021 | B2 |
D943732 | Shauver et al. | Feb 2022 | S |
20020017328 | Loo | Feb 2002 | A1 |
20020085952 | Ellingboe et al. | Jul 2002 | A1 |
20020095121 | Norton et al. | Jul 2002 | A1 |
20020179544 | Johnson et al. | Dec 2002 | A1 |
20020189712 | Safabash | Dec 2002 | A1 |
20030023226 | Lopez | Jan 2003 | A1 |
20030153895 | Leinsing | Aug 2003 | A1 |
20030236500 | Scheu | Dec 2003 | A1 |
20040031756 | Suzuki et al. | Feb 2004 | A1 |
20040035743 | Tighe et al. | Feb 2004 | A1 |
20040073161 | Tachibana | Apr 2004 | A1 |
20040087888 | Digianfilippo et al. | May 2004 | A1 |
20040116891 | Curutcharry | Jun 2004 | A1 |
20040118477 | Desmond | Jun 2004 | A1 |
20040225274 | Jansen et al. | Nov 2004 | A1 |
20050033260 | Kubo et al. | Feb 2005 | A1 |
20050096627 | Howard | May 2005 | A1 |
20050131357 | Denton et al. | Jun 2005 | A1 |
20050230575 | Zelenski et al. | Oct 2005 | A1 |
20050252572 | Khan et al. | Nov 2005 | A1 |
20050252574 | Khan et al. | Nov 2005 | A1 |
20060048844 | Merrill | Mar 2006 | A1 |
20060064053 | Bollish et al. | Mar 2006 | A1 |
20060169348 | Yigal | Aug 2006 | A1 |
20060259195 | Eliuk et al. | Nov 2006 | A1 |
20070007478 | Leinsing et al. | Jan 2007 | A1 |
20070017583 | Fangrow | Jan 2007 | A1 |
20070088252 | Pestotnik et al. | Apr 2007 | A1 |
20070088313 | Zinger et al. | Apr 2007 | A1 |
20070106244 | Mosler et al. | May 2007 | A1 |
20070151984 | Baker et al. | Jul 2007 | A1 |
20070169836 | Djurle et al. | Jul 2007 | A1 |
20070244447 | Capitaine et al. | Oct 2007 | A1 |
20070287953 | Ziv et al. | Dec 2007 | A1 |
20080059228 | Bossi et al. | Mar 2008 | A1 |
20080065006 | Roger et al. | Mar 2008 | A1 |
20080077116 | Dailey | Mar 2008 | A1 |
20080086094 | Peters | Apr 2008 | A1 |
20080114328 | Doherty et al. | May 2008 | A1 |
20080125897 | DiGianfilippo et al. | May 2008 | A1 |
20080169043 | Osborne et al. | Jul 2008 | A1 |
20080169044 | Osborne et al. | Jul 2008 | A1 |
20080177222 | De Marco et al. | Jul 2008 | A1 |
20080195416 | Tribble et al. | Aug 2008 | A1 |
20080199353 | Mlodzinski et al. | Aug 2008 | A1 |
20080269680 | Ibranyan et al. | Oct 2008 | A1 |
20080287920 | Fangrow | Nov 2008 | A1 |
20090012449 | Lee et al. | Jan 2009 | A1 |
20090050216 | Trocki et al. | Feb 2009 | A1 |
20090067973 | Eliuk et al. | Mar 2009 | A1 |
20090069743 | Krishnamoorthy et al. | Mar 2009 | A1 |
20090082649 | Muller et al. | Mar 2009 | A1 |
20090088687 | Yardimci et al. | Apr 2009 | A1 |
20090099547 | Radmer | Apr 2009 | A1 |
20090101576 | Rohde et al. | Apr 2009 | A1 |
20090126825 | Eliuk et al. | May 2009 | A1 |
20090145509 | Baker et al. | Jun 2009 | A1 |
20090149743 | Barron et al. | Jun 2009 | A1 |
20090154764 | Khan et al. | Jun 2009 | A1 |
20090163860 | Patrick et al. | Jun 2009 | A1 |
20090177149 | Childers et al. | Jul 2009 | A1 |
20090198215 | Chong et al. | Aug 2009 | A1 |
20090223592 | Procyshyn et al. | Sep 2009 | A1 |
20090223990 | Bailey et al. | Sep 2009 | A1 |
20090254031 | Lee | Oct 2009 | A1 |
20090270832 | Vancaillie et al. | Oct 2009 | A1 |
20090306621 | Thome, Jr. et al. | Dec 2009 | A1 |
20100024904 | Hoffman et al. | Feb 2010 | A1 |
20100049157 | Fangrow | Feb 2010 | A1 |
20100121246 | Peters et al. | May 2010 | A1 |
20100245056 | Braun et al. | Sep 2010 | A1 |
20100276034 | Gonnelli et al. | Nov 2010 | A1 |
20100280430 | Caleffi et al. | Nov 2010 | A1 |
20100286606 | Ding | Nov 2010 | A1 |
20110004143 | Beiriger et al. | Jan 2011 | A1 |
20110062703 | Lopez et al. | Mar 2011 | A1 |
20110067781 | Osborne | Mar 2011 | A1 |
20110152757 | Beck et al. | Jun 2011 | A1 |
20110175347 | Okiyama | Jul 2011 | A1 |
20110178493 | Okiyama | Jul 2011 | A1 |
20110196304 | Kramer et al. | Aug 2011 | A1 |
20110204144 | Waugh et al. | Aug 2011 | A1 |
20110229517 | Strahlendorf et al. | Sep 2011 | A1 |
20110276031 | Hoang et al. | Nov 2011 | A1 |
20110305545 | Davis et al. | Dec 2011 | A1 |
20120157914 | Stroup | Jan 2012 | A1 |
20120041391 | Fangrow et al. | Feb 2012 | A1 |
20120067429 | Mosler et al. | Mar 2012 | A1 |
20120109077 | Ryan | May 2012 | A1 |
20120123298 | Mendels et al. | May 2012 | A1 |
20120298254 | Brem et al. | Nov 2012 | A1 |
20120302986 | Brem et al. | Nov 2012 | A1 |
20130006214 | Garfield et al. | Jan 2013 | A1 |
20130018356 | Prince et al. | Jan 2013 | A1 |
20130053815 | Mucientes et al. | Feb 2013 | A1 |
20130085439 | Sansoucy | Apr 2013 | A1 |
20130102772 | Eshima et al. | Apr 2013 | A1 |
20130180618 | Py | Jul 2013 | A1 |
20130211332 | Beiriger et al. | Aug 2013 | A1 |
20130218121 | Waller et al. | Aug 2013 | A1 |
20130220484 | De Marco | Aug 2013 | A1 |
20130292002 | Lopez | Nov 2013 | A1 |
20130292004 | Ducret et al. | Nov 2013 | A1 |
20140020790 | Yuyama et al. | Jan 2014 | A1 |
20140039392 | Geipel et al. | Feb 2014 | A1 |
20140124087 | Anderson et al. | May 2014 | A1 |
20140124092 | Gonnelli et al. | May 2014 | A1 |
20140135732 | Spronken et al. | May 2014 | A1 |
20140136229 | Levine et al. | May 2014 | A1 |
20140150925 | Sjogren et al. | Jun 2014 | A1 |
20140174596 | Lopez | Jun 2014 | A1 |
20140261727 | Mansour et al. | Sep 2014 | A1 |
20140261860 | Heath | Sep 2014 | A1 |
20140261877 | Ivosevic et al. | Sep 2014 | A1 |
20140263614 | Keefe et al. | Sep 2014 | A1 |
20140276386 | Mansour et al. | Sep 2014 | A1 |
20140276649 | Ivosevic et al. | Sep 2014 | A1 |
20140299221 | Lopez | Oct 2014 | A1 |
20140323970 | Duncan | Oct 2014 | A1 |
20140350949 | Utech et al. | Nov 2014 | A1 |
20150000784 | Jamaledine | Jan 2015 | A1 |
20150008664 | Tachizaki | Jan 2015 | A1 |
20150025453 | Ledford et al. | Jan 2015 | A1 |
20150040987 | Reichert et al. | Feb 2015 | A1 |
20150040988 | Reichert et al. | Feb 2015 | A1 |
20150041531 | Vavala et al. | Feb 2015 | A1 |
20150045772 | Reichert et al. | Feb 2015 | A1 |
20150068640 | Garfield et al. | Mar 2015 | A1 |
20150082746 | Ivosevic et al. | Mar 2015 | A1 |
20150101707 | Ranalletta et al. | Apr 2015 | A1 |
20150119820 | Kanamoto | Apr 2015 | A1 |
20150123398 | Sanders et al. | May 2015 | A1 |
20150126958 | Sanders et al. | May 2015 | A1 |
20150133879 | Kanamoto et al. | May 2015 | A1 |
20150151041 | Yodfat et al. | Jun 2015 | A1 |
20150157536 | Qiu et al. | Jun 2015 | A1 |
20150161354 | Blomquist | Jun 2015 | A1 |
20150202382 | Juretich et al. | Jul 2015 | A1 |
20150202383 | Juretich et al. | Jul 2015 | A1 |
20150202384 | Juretich et al. | Jul 2015 | A1 |
20150202385 | Juretich et al. | Jul 2015 | A1 |
20150209230 | Lev et al. | Jul 2015 | A1 |
20150209233 | Fukuoka | Jul 2015 | A1 |
20150209495 | Biset et al. | Jul 2015 | A1 |
20150209510 | Burkholz et al. | Jul 2015 | A1 |
20150209572 | Garfield et al. | Jul 2015 | A1 |
20150250680 | Browka et al. | Sep 2015 | A1 |
20150250681 | Lev et al. | Sep 2015 | A1 |
20150257977 | Bochenko et al. | Sep 2015 | A1 |
20150265500 | Russo et al. | Sep 2015 | A1 |
20150297451 | Mariei et al. | Oct 2015 | A1 |
20150297453 | Kim et al. | Oct 2015 | A1 |
20150297454 | Sanders et al. | Oct 2015 | A1 |
20150297456 | Mariei et al. | Oct 2015 | A1 |
20150297459 | Sanders et al. | Oct 2015 | A1 |
20150297460 | Mansour et al. | Oct 2015 | A1 |
20150297839 | Sanders et al. | Oct 2015 | A1 |
20150297881 | Sanders et al. | Oct 2015 | A1 |
20150314066 | Shimizu | Nov 2015 | A1 |
20150320992 | Bonnet et al. | Nov 2015 | A1 |
20150346013 | Feng et al. | Dec 2015 | A1 |
20150359709 | Kriheli et al. | Dec 2015 | A1 |
20150366758 | Noguchi et al. | Dec 2015 | A1 |
20160000653 | Kramer | Jan 2016 | A1 |
20160001003 | Perazzo et al. | Jan 2016 | A1 |
20160038373 | Ohlin | Feb 2016 | A1 |
20160038374 | Merhold et al. | Feb 2016 | A1 |
20160051446 | Lev et al. | Feb 2016 | A1 |
20160058666 | Strahlendorf et al. | Mar 2016 | A1 |
20160058667 | Kriheli | Mar 2016 | A1 |
20160081878 | Marks et al. | Mar 2016 | A1 |
20160081879 | Garfield et al. | Mar 2016 | A1 |
20160101020 | Guala | Apr 2016 | A1 |
20160114922 | Bonhora et al. | Apr 2016 | A1 |
20160136051 | Lavi | May 2016 | A1 |
20160136412 | McKinnon et al. | May 2016 | A1 |
20160140315 | Diaz et al. | May 2016 | A1 |
20160158104 | Ali et al. | Jun 2016 | A1 |
20160158437 | Biasi et al. | Jun 2016 | A1 |
20160206511 | Garfield et al. | Jul 2016 | A1 |
20160213568 | Mansour et al. | Jul 2016 | A1 |
20160213861 | Whitaker et al. | Jul 2016 | A1 |
20160213862 | Whitaker et al. | Jul 2016 | A1 |
20160250102 | Garfield et al. | Sep 2016 | A1 |
20160256632 | Fangrown | Sep 2016 | A1 |
20160262981 | Carrez et al. | Sep 2016 | A1 |
20160310362 | Lane et al. | Oct 2016 | A1 |
20160331893 | Yeh et al. | Nov 2016 | A1 |
20160354281 | O'Neill et al. | Dec 2016 | A1 |
20170007501 | Schuldt-Lieb et al. | Jan 2017 | A1 |
20170020428 | Rogers | Jan 2017 | A1 |
20170079883 | Lopez | Mar 2017 | A1 |
20170081168 | Seay et al. | Mar 2017 | A1 |
20170128666 | Davis | May 2017 | A1 |
20170129763 | Fangrow, Jr. | May 2017 | A1 |
20170146381 | Eckel et al. | May 2017 | A1 |
20170165435 | Green | Jun 2017 | A1 |
20170165436 | Haddad et al. | Jun 2017 | A1 |
20170255760 | Lee et al. | Sep 2017 | A1 |
20170312716 | Konrad, Jr. et al. | Nov 2017 | A1 |
20170354571 | David et al. | Dec 2017 | A1 |
20180043323 | Janders et al. | Feb 2018 | A1 |
20180055738 | Chen et al. | Mar 2018 | A1 |
20180065097 | Konrad, Jr. et al. | Mar 2018 | A1 |
20180133667 | Lee et al. | May 2018 | A1 |
20180161244 | Lopez | Jun 2018 | A1 |
20180168930 | Tunesi | Jun 2018 | A1 |
20180168935 | Chen et al. | Jun 2018 | A1 |
20180194505 | Amano et al. | Jul 2018 | A1 |
20180207063 | Lopez | Jul 2018 | A1 |
20180232497 | Hoffman et al. | Aug 2018 | A1 |
20180263850 | Schneider et al. | Sep 2018 | A1 |
20180272117 | Fangrow | Sep 2018 | A1 |
20180344572 | Zollinger et al. | Dec 2018 | A1 |
20180353381 | Pak et al. | Dec 2018 | A1 |
20180353382 | Zollinger et al. | Dec 2018 | A1 |
20180354662 | Feith et al. | Dec 2018 | A1 |
20180357476 | Klumph | Dec 2018 | A1 |
20180360689 | Zollinger et al. | Dec 2018 | A1 |
20190019576 | DeCiccio et al. | Jan 2019 | A1 |
20190021947 | Bomgaars et al. | Jan 2019 | A1 |
20190056419 | Procyshyn et al. | Feb 2019 | A1 |
20190070405 | Fangrow | Mar 2019 | A1 |
20190091639 | Brown et al. | Mar 2019 | A1 |
20190105619 | Wilson et al. | Apr 2019 | A1 |
20190151569 | Fangrow | May 2019 | A1 |
20190152663 | Kraft | May 2019 | A1 |
20190163876 | Remme et al. | May 2019 | A1 |
20190170663 | Pirkle et al. | Jun 2019 | A1 |
20190216683 | Yaegashi | Jul 2019 | A1 |
20190244466 | Berg et al. | Aug 2019 | A1 |
20190247280 | Hellenbrand | Aug 2019 | A1 |
20190262790 | Konrad, Jr. et al. | Aug 2019 | A1 |
20190275243 | Deck et al. | Sep 2019 | A1 |
20190307643 | Tribble et al. | Oct 2019 | A1 |
20190388302 | Schobel et al. | Dec 2019 | A1 |
20200016037 | Oda et al. | Jan 2020 | A1 |
20200066389 | Prince et al. | Feb 2020 | A1 |
20200093699 | Oda et al. | Mar 2020 | A1 |
20200113784 | Lopez | Apr 2020 | A1 |
20200113785 | Lopez | Apr 2020 | A1 |
20200206492 | Fangrow | Jul 2020 | A1 |
20200289370 | Lopez | Sep 2020 | A1 |
20200297581 | Lopez | Sep 2020 | A1 |
20210002008 | Min et al. | Jan 2021 | A1 |
20210121363 | Oda et al. | Apr 2021 | A1 |
20210259921 | Lopez | Aug 2021 | A1 |
20220008711 | Fangrow | Jan 2022 | A1 |
20220054766 | Fangrow | Feb 2022 | A1 |
Number | Date | Country |
---|---|---|
1707379 | Dec 2005 | CN |
101244297 | Aug 2008 | CN |
106860003 | Jun 2017 | CN |
107198658 | Sep 2017 | CN |
108210332 | Jun 2018 | CN |
202 16 791 | Feb 2003 | DE |
20 2004 014 868 | Nov 2004 | DE |
0 521 460 | Sep 1995 | EP |
0 974 330 | Jan 2000 | EP |
1 533 597 | May 2005 | EP |
1 563 819 | Aug 2005 | EP |
1 997 471 | Dec 2008 | EP |
3 375 427 | Sep 2018 | EP |
55-156750 | Nov 1980 | JP |
55-173339 | Dec 1980 | JP |
56-95247 | Aug 1981 | JP |
62-189072 | Aug 1987 | JP |
06-343706 | Dec 1994 | JP |
10-118158 | May 1998 | JP |
2001-190689 | Jul 2001 | JP |
2002-238979 | Aug 2002 | JP |
2002-355318 | Dec 2002 | JP |
2003-144546 | May 2003 | JP |
2003-199823 | Jul 2003 | JP |
2003-225305 | Aug 2003 | JP |
2004-049497 | Feb 2004 | JP |
2007-14618 | Jan 2007 | JP |
2007-215775 | Aug 2007 | JP |
2011-0019800 | Mar 2011 | KR |
10-1574194 | Dec 2015 | KR |
WO 199714493 | Apr 1997 | WO |
WO 199823353 | Jun 1998 | WO |
WO 199919012 | Apr 1999 | WO |
WO 199963547 | Dec 1999 | WO |
WO 200041751 | Jul 2000 | WO |
WO 200103757 | Jan 2001 | WO |
WO 2001039874 | Jun 2001 | WO |
WO 2002013890 | Feb 2002 | WO |
WO 2005041846 | May 2005 | WO |
WO 2005110007 | Nov 2005 | WO |
WO 2005123162 | Dec 2005 | WO |
WO 2007033013 | Mar 2007 | WO |
WO 2007061424 | May 2007 | WO |
WO 2007062315 | May 2007 | WO |
WO 2007079305 | Jul 2007 | WO |
WO 2007148708 | Dec 2007 | WO |
WO 2008051998 | May 2008 | WO |
WO 2008128074 | Oct 2008 | WO |
WO 2008144447 | Nov 2008 | WO |
WO 2009060419 | May 2009 | WO |
WO 2009130147 | Oct 2009 | WO |
WO 2011002853 | Jan 2011 | WO |
WO 2011012313 | Feb 2011 | WO |
WO 2011014525 | Feb 2011 | WO |
WO 2011058545 | May 2011 | WO |
WO 2011058548 | May 2011 | WO |
WO 2011091542 | Aug 2011 | WO |
WO 2011091543 | Aug 2011 | WO |
WO 2011104711 | Sep 2011 | WO |
WO 2011104712 | Sep 2011 | WO |
WO 2011150037 | Dec 2011 | WO |
WO 2012119225 | Sep 2012 | WO |
WO 2013096911 | Jun 2013 | WO |
WO 2014122643 | Aug 2014 | WO |
WO 2014126473 | Aug 2014 | WO |
WO 2014177347 | Nov 2014 | WO |
WO 2014181320 | Nov 2014 | WO |
WO 2015029020 | Mar 2015 | WO |
WO 2015077184 | May 2015 | WO |
WO 2015077466 | May 2015 | WO |
WO 2015122921 | Aug 2015 | WO |
WO 2016010909 | Jan 2016 | WO |
WO 2017096072 | Jun 2017 | WO |
WO 2018009996 | Jan 2018 | WO |
WO 2018022640 | Feb 2018 | WO |
WO 2019018195 | Jan 2019 | WO |
Entry |
---|
Burrows, et al., “Intravenous (IV) Fluidmaker IV. A Disposable Device for Preparation of Sterile Water for Injection In a Field Setting,” Fort Detrick, US Army Biomedical Research & Development Laboratory, Sep. 1991. https://apps.dtc.mil/dtic/tr/fulltest/u2/a247385.pdf. |
Abbott “Plum A+,” System Operating Manual (For use with List 11971-04) in 85 pages, May 2001. |
U.S. Appl. No. 29/571,547, filed Jul. 19, 2016, Shauver et al. |
U.S. Appl. No. 29/586,575, filed Dec. 5, 2016, Fangrow. |
Autoyec 50, from KRZ, Dec. 6, 2007. |
B. Braun Medical Inc. Two-Bag Irrigation Set, Two Non-vented Spikes, dated Jul. 2012, in 1 page. |
BioExpert International Inc., Company overview, credentials for Rabih Jamaleddine, Nabil Kereknawi, and Danica Robillard Corso, copyright 2010 BioExpert International Inc. in 3 pages [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://bloexpert.ca/about.html. |
Cato (Computer Aided Therapy For Oncology)—Reference Manual—Vienna, May 2005, 255 pgs. |
ISO/Tech Design, QC, Canada, “CHEMOSPHERE,” product brochure, in 2 pages [Publication Date Unknown]. |
Clearlink Needleless IV Access System, dated Aug. 2007, in 2 pages. |
CytoCare, by Health Robotics, Brochure, Date Unknown, downloaded on May 25, 2012 from http://www.health-robotics.com/smartedit/downloads/en/cytocare7.pdf, 6 pages. |
European Supplemental Search Report, re EP Application No. 14864639, dated May 12, 2017. |
Exacta-Mix 2400, from Baxa, which appears to have a date of 2007, 2 pages. |
Flickinger, Bruce, “Misperceptions Cloud the Issue of Sterile Drug Compounding,” Jun. 2007. |
Fox, Brent I., “Pharmacy Automation and Technology: Automated Intravenous Preparation: Robots for the Pharmacy,” Hospital Pharmacy, vol. 44, Mar. 2009, pp. 255-257. |
Healthmark, “Hospital Medication Preparation, Packaging and Dispensing,” Chemo Drug Preparation/Administration in 2 pages [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/2-36-88-Chemo-Drug-Preparation-Administration_en.html. |
Healthmark, “Hospital Medication Preparation, Packaging and Dispensing,” Chemosphere, Sterile Chemo Compounding (Isolator) in 1 page [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/2-36-10-ChemoSphere_en.html?ProduitID=244. |
Healthmark, “Hospital Medication Preparation, Packaging and Dispensing” in 1 page [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/2-en-Hospital-Medication-Preparation-Packaging-and-Dispensing.html. |
Healthmark, “Hospital Medication Preparation, Packaging and Dispensing,” Oncology Preparation and Administration in 1 page [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/2-36-10-COMPANY-PROFILEHospital-en.html. |
Healthmark, “Hospital Medication Preparation, Packaging and Dispensing,”Phocus RX (Camera Verification System), Remote Rx Checking of admixtures in 2 pages [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/2-36-10-PHOCUS-Rx-Camera-verification-System-_en.html?ProduitID=229. |
Healthmark, “New Product Items” in 1 page [retrieved on Jan. 6, 2015; Publication Date Unknown]; accessed on the world wide web at http://www.healthmark.ca/home.html. |
Healthmark, “Introducing the Precifill Dispensing Pump” product brochure in 2 pages [Publication Date Unknown]. |
Integra Brochure, from Eurospital, Brochure acquired in Mar. 2012. |
International Preliminary Report on Patentability for International PCT Patent Application No. PCT/US2014/065972, dated Jun. 9, 2016. |
International Search Report and Written Opinion for International PCT Patent Application No. PCT/US2014/065972, dated Feb. 24, 2015. |
Neo Care Medical Products: Product Catalog, dated Jun. 2008, in 38 pages. |
Grifols International, S.A., “PHOCUS Rx, Remote IV Compounding Validation” product brochure and “Product Description Sheet” in 13 pages [Publication Date unknown but may be May 29, 2013]. |
Pinnacle TPN Management System, from B Braun, downloaded May 5, 2009 from http://www.bbraunusa.com/index. cfm?uuid=7386ADF065B05CDOD22AF700339AA4092, 1 page. |
“Precifill,” Trademark search (TESS) in 1 page, [retrieved on Jan. 6, 2015; Application Filing Date of Sep. 30, 2011]; accessed on the world wide web at http://tmsearch.uspto.gov/bin/showfield?ff=doc&state=4807:gz67gx.3.1. |
Product detail for “NAMIC® Closed Fluid Systems” from Navilyst Medical, downloaded on May 11, 2010 from http://www.navilystmedical.com/Products/index.cfm/19, 2 pages. |
Product detail for “RapidFill™Automated Syringe Filler,” from Baxa, downloaded on Mar. 31, 2010 from http://www.baxa/com/PharmacyProducts/AutomatedFillingSystems/ProductDetail/?id=B1, 2 pages. |
Product detail for “Summit Medical DirectFlow” micro infusion extension set from Summit Medical Technologies, downloaded on May 10, 2010 from http://summitmedtech.com/p6line.php, 1 page. |
Riva, downloaded in Apr. 2009 from http://www.rivasystem.com, 6 pages. |
SmartSite Safety Disposables, with copyright notice dated 2004. |
Smith, “Lifesaving Cancer Drugs May Put Workers' Lives at Risk,” downloaded on Jul. 12, 2010 from http://www.msnbc.msn.com/id/38114586/ns/health-cancer, 7 pages. |
Spiros—Closed Male Connector, published Jan. 22, 2008. |
Technical Data sheet for Analog Amplifiers Type VA, models V8-C and V8-D, STM Sensors dated Dec. 2007, 4 pages. |
Technical Data sheet for Through Beam Sensors Type G2, 1480 nm, STM Sensors dated Dec. 2009, 2 pages. |
Technical Data sheet for Through Beam Sensors Type G2, 645 nm, STM Sensors dated Sep. 2008, 2 pages. |
User Guide for medOC 1xx Basic, Neo Care Medical Products GmbH, Version Jun. 2008, 23 pages. |
User Manual for medOC 3xx /6xx /8xx, Neo Care Medical Products GmbH, Version May 2008, 44 pages. |
Baxa Corp. v. McGaw Inc. 981 F. Supp. 1348 (1997), Memorandum Opinion and Order, 14 pages. |
Number | Date | Country | |
---|---|---|---|
20180177940 A1 | Jun 2018 | US |
Number | Date | Country | |
---|---|---|---|
61908674 | Nov 2013 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14746048 | Jun 2015 | US |
Child | 15848368 | US | |
Parent | PCT/US2014/065972 | Nov 2014 | US |
Child | 14746048 | US |