This present disclosure relates to a method of characterizing an optical system, particularly, but not exclusively where the optical system is used as an imaging system. Aspects of the invention relate to a method of imaging, to an optical system, and to an imaging system that includes the optical system.
Imaging through optical fibres is known and is becoming increasingly common. White light imaging through certain types of optical fibres, termed imaging fibre bundles or multicore fibre (MCF), is a well-established technique used in commercial boresocopes and medical endoscopes. More recently, coherent light (i.e. light from a laser) has been used to image through other kinds of fibre, such as multimode fibre (MMF). The advantage of using coherent light is that images can be formed without lenses, meaning endoscopes need not be any thicker than the fibre itself (often <500 μm). In the case of MMF imaging, this size advantage is enhanced due to the higher ‘information density’—that is, MMF provides more pixels of resolution per unit area than MCF. This opens up opportunities for minimally invasive optical imaging in previously inaccessible areas of the body, e.g. deep in the brain. As a result, there have been a number of studies examining the use of both MMF and MCF for coherent light imaging.
White-light endoscopy is the standard-of-care for inspecting large areas of the gastrointestinal (GI) tract and lung for pre-malignant change (dysplasia) and cancer. For example, Barrett's oesophagus is an acquired metaplastic condition that predisposes patients to the development of oesophageal adenocarcinoma. The cancer risk for Barrett's patients increases significantly in the presence of dysplasia, up to more than 30% per year. Early identification of dysplasia enables curative intervention through simple endoscopic resection or radiofrequency ablation. Unfortunately, the current surveillance procedure uses white-light endoscopy combined with random biopsy, which together show only a 40-64% sensitivity for dysplasia, leading to high miss rates. The 5-year survival rate for oesophageal cancer is only 15%, yet can be as high as 80% when patients are diagnosed with early-stage disease, hence improvements in endoscopic early detection methodologies are urgently needed. While application of dyes can improve contrast, their use lengthens procedure times and can lead to toxicities. Label-free approaches could better address the clinical unmet need for improved contrast of dyplastic tissue.
Localised proliferation of cells in dysplasia scatters light, creating abnormally distorted wavefronts. Phase imaging has been shown to be highly sensitive to such distortions. Furthermore, polarimetric imaging measurements of diattenuation, retardance and circularity can be modified by scattering, as well as perturbed by the higher concentrations of optically anisotropic molecules such as collagen abundant in tumours. Light scattering spectroscopy has shown promise for detecting dysplasia using phase and polarisation information, but interrogates only a narrow field-of-view. This limitation is partially mitigated by optical coherence tomography, which uses lateral scanning mechanisms, but interpretation of the resulting cross-sectional images remains challenging. In addition, both approaches require dedicated complex instrumentation, which has limited endoscopic application.
Flexible medical fibrescopes relay optical information from within the patient to the imaging system outside, which could be used to enable direct, wide-field, phase and polarisation imaging in existing clinically approved systems with comparatively simple and low-cost elements, such as coded apertures, gratings, and polarising optics. Commercial endoscopes typically use distal sensors (‘chip-on-tip’) and although prototype devices with distal optics for other modalities have been developed (e.g. holographic imaging) the additional bulk (>2-fold width) makes integration with existing endoscopic procedures difficult.
Fibre bundles are typically <1 mm in width, independent of the imaging modality, making them attractive candidates for implementation of novel medical imaging technologies in endoscopy. However, both MCF and MMF scatter light in a deterministic but highly complex manner that is a function of bending and temperature. This scattering prevents imaging in most cases, or at the very least greatly reduces imaging quality, and so must be counteracted. This can be achieved with very high accuracy using transmission matrix (TM) approaches to pre-characterise the full optical transfer properties of the fibre before imaging. Typically, the TM is measured by sending known light fields in one end of the fibre and measuring what comes out at the other. By creating many different vectors x1=[x1 . . . xM]T and measuring the corresponding output field y1=[y1 . . . yN]T, a linear system may be built up:
Y=AX
where
These equations can then be solved to determine A, the transmission matrix (TM), for example by computing:
YX
−1
=A
An actual image vector, xs, will be transformed by the fibre to produce a raw output, ys=Axs. To recover xs from measured ys, the following equation is used: xs=A−1ys.
However, there remains a key limitation of most work to date using TM characterisation for fibre imaging. This is that known light fields must be sent in one facet fibre and measured at the other in order to obtain the TM. This poses a problem for realistic use because one end of the fibre (the distal facet) is likely to be inaccessible, for example deep inside the body. It is undesirable to introduce a scanning mechanism at the distal facet as this would introduce additional bulk thereby reducing a key benefit of the technique: that the imaging fibre is hair thin (<500 μm). Although miniaturised scanning mechanisms have been integrated into endoscopes of diameter 1 mm, such approaches are not easily scalable to allow additional advanced processing, e.g. polarimetry or phase detection. Having all optics located at the proximal facet outside the body allows much more flexibility for advanced imaging. Such a system with no distal optics cannot simply be achieved by premeasuring the fibre TM because the fibre is extremely sensitive to minor changes in bending and temperature as would be encountered in vivo. A pre-measured TM is therefore not sufficient for image recovery in practical applications.
It is an object of embodiments of the present invention to overcome certain disadvantages associated with the prior art.
In accordance with an aspect of the present invention, there is provided a method of characterizing an optical system, wherein the optical system comprises:
the method comprising:
In certain embodiments, the reflector matrices may be determined by:
Determining the characterization transmission matrix may comprise transmitting light through the optical fibre in the characterization configuration at each of the plurality of characterization wavelengths, detecting the transmitted light at each of the plurality of characterization wavelengths, determining the characterization transmission matrix using the detected transmitted light.
Detecting reflected calibration patterns may comprise measuring the amplitude of reflected calibration patterns. Additionally or alternatively, detecting reflected calibration patterns comprises measuring the phase of reflected calibration patterns. Additionally or alternatively, detecting reflected calibration patterns comprises measuring the polarisation of reflected calibration patterns.
The method may comprise producing a square reflectance matrix from the reflected calibration patterns, wherein determining the instantaneous transmission matrix comprises using the square reflectance matrix.
In accordance with another aspect of the present invention, there is provided a method of imaging comprising:
The method may further comprise obtaining a transmission matrix of the reflector assembly, wherein determining the instantaneous transmission matrix or producing recovered image data may comprise using the transmission matrix of the reflector assembly.
Producing recovered image data may comprise producing recovered amplitude data. Additionally or alternatively, producing recovered image data may comprise producing recovered phase data. Additionally or alternatively, producing recovered image data may comprise producing recovered polarisation data. In certain non-limiting embodiments, the sample may comprise human or animal tissue. The sample may be in vivo, ex vivo or in vitro.
In accordance with another aspect of the present invention, there is provided an optical system comprising an optical fibre having a proximal end and a distal end, and a reflector assembly comprising a stack of reflectors disposed at the distal end of the optical fibre, wherein stack of reflectors is arranged to provide different reflector matrices in dependence on illumination wavelength.
The stack of reflectors may comprise a plurality of reflectors. Each of the plurality of reflectors may be separated from an adjacent reflector by an absorptive filter. The plurality of reflectors may comprise optical metasurfaces.
In certain embodiments, the optical system may further comprise:
In accordance with another aspect of the present invention, there is provided a method of determining a presence of a physiological condition in a subject, the method comprising:
The step of determining a presence of the physiological condition in the tissue sample may comprise:
using the recovered image data as contrast mechanisms to identify tissue morphology or microstructure associated with the physiological condition; and/or
using the recovered image data to produce images showing the spatial variation or spatial entropy of one or more of amplitude, phase and polarisation, and using differences thereof to delineate healthy areas of the tissue sample (e.g. areas not affected by the physiological condition) and areas of the tissue sample that are affected by the physiological condition (e.g. as described below with reference to
The physiological condition may, for example, be cancer or a pre-cancerous condition, the presence of scar tissue, or the presence of inflammation. The above described method may be useful in determining the presence of other physiological conditions.
In accordance with another aspect of the present invention, there is provided a computer implemented method comprising:
The step of determining a presence of the physiological condition in the tissue sample may comprise:
using the recovered image data as contrast mechanisms to identify tissue morphology or microstructure associated with the physiological condition; and/or
The physiological condition may, for example, be cancer or a pre-cancerous condition, the presence of scar tissue, or the presence of inflammation. The above described method may be useful in determining the presence of other physiological conditions.
Embodiments of the invention are further described hereinafter with reference to the accompanying drawings, in which:
The recombined beam then passes through a first non-polarising beam splitter (NPBS) 24a before being focused by a second lens 14b. The focused beam then passes into an optical fibre 26. The optical fibre 26 may comprise any suitable type of fibre that is capable of facilitating optical propagation. In certain embodiments, the optical fibre 26 may comprise a non-single-mode optical fibre (i.e. an optical fibre that does not behave as a single-mode optical fibre). In particular, the non-single-mode optical fibre may be an imaging fibre bundle or multicore fibre (MCF), a multimode fibre (MMF), dual-clad fibre, photonic crystal fibre, hollow-core fibre, or other microstructured optical fibre. Where such fibres are available in single-mode and multi-mode variants, the term “non-single-mode” in the present application is not intended to encompass to the single-mode variants.
The optical fibre 26 has a proximal end 26a and a distal end 26b. In the configuration shown in
The first holographic sub-system 15a described above enables control of optical amplitude, phase and polarization of light entering the proximal end 26a of the optical fibre 26, enabling creation of arbitrary calibration patterns during characterization, and arbitrary illuminations during imaging (described further below).
The sample 30 may be (but is not necessarily) human or animal tissue which may be in vivo, ex vivo or in vitro. The human or animal tissue may form or have formed part of the gastrointestinal tract or respiratory system of a subject.
A reflector assembly 42 (described further below) is disposed on the distal end 26b of the optical fibre 26. Light exiting the distal end 26b of the optical fibre 26 is reflected by the sample, and is returned into the distal end 26b back towards the second lens 14b which then acts to collimate the light beam. The collimated light beam is reflected by the first NPBS 24a so as to separate it from the beam propagating in the opposite direction (i.e. towards the optical fibre 26).
The separated beam passes through a second NPBS 24b before being split by a third PBS 16c. The transmitted part of the split beam is reflected by a third beam reflector 18c so that it passes through a second SLM 22b prior to passing through a third half wave plate 20c and a fourth PBS 16d. The reflected part of the beam split by the third PBS 16c passes through a fourth half wave plate 20d prior to passing through the second SLM 22b and being reflected by a fourth beam reflector 18d. This part of the beam then passes through the fourth PBS 16d so as to be combined with the transmitted part of the beam. The combination of the third PBS 16c, fourth PBS 16d, third half wave plate 20c, fourth half wave plate 20d, second SLM 22b, third beam reflector 18c and fourth beam reflector 18d form a second holographic sub-system 15b.
The combined beam then passes through a 45-degree polarizer 34, is focused by a third lens 14c, and is then detected by detection means in the form of a detector 36. The detector 36 may comprise, for example, a CCD. Second holographic sub-system 15b provides a means of enabling the detector to determine the optical phase and polarisation, in addition to amplitude/intensity of the detected beam.
Prior to first use for imaging, the optical system 10 must be characterized. In particular, the optical fibre 26 must be characterized first, followed by the reflectors 42. As is described below, the characterization of the optical system 10 may utilize an alternative configuration of the optical fibre 26 (path shown by dotted line 32 in
Processing means in the form of a processor (e.g. within a PC) 38 is communicably coupled to the first SLM 22a for controlling patterns (e.g. holograms) displayed by the first SLM 22a and is similarly coupled to the second SLM 22b. The processor 38 is also communicably coupled to the detector 36 and is configured to receive data obtained by the detector 38.
At different wavelengths, the reflector assembly 42 provides a different reflector matrix, effectively switching between reflectors 48. On the assumption that the transmission matrix of the optical fibre 26 is relatively similar at each of these different wavelengths, the transmission matrix may be determined using the reflector matrices, as is described below.
In alternative embodiments, the stack of reflectors may comprise other numbers of reflectors and/or absorptive filters, and/or be any other arrangement that is capable of providing a reflector matrix that is dependent on the wavelength of incident light.
Next, maintaining the characterization configuration, the reflector assembly 42 is inserted and butt-coupled against the distal end 26b of the optical fibre 26 (as shown in
Using the determined characterization transmission matrices of the optical fibre 26 and the measured reflectance matrices, the reflector matrices, R, can be determined (step 106) since M1=A1TR1A1, and R1=(A1T)−1M1A1−1.
Next, the second beam stop 40b is removed and the first beam stop 40a is replaced (to adopt the configuration shown in
Prior to acquisition of sample image data, some further characterization is required (this time in the imaging configuration as shown in
A physical model used for aiding understanding of certain aspects of the present invention is shown in
y=A
λ
T
R
λ
A
λ
x=
(1)
where
C
λ
=A
λ
T
R
λ
A
λ∈2M×2M (2)
is a reflection matrix (RM). Physically speaking, Cλ represents light taking a complete round-trip (or double-pass): forwards down the fibre 26, off a given reflector 48 of the reflector assembly 42 and back up the fibre 26 (as shown in
The zeroth-order model assumes that the TMs under the different reflectors 48 are the same, i.e.: A=Aλ
Physically, this corresponds to wavelength modulations significantly less than the spectral bandwidth of the fibre 26. This model has the significant advantage that if at least 3 wavelengths are used (Q≥3) producing at least 3 RMs, A can be recovered in a relatively straightforward way relying largely on analytical steps (described further below). This analytical approach further requires that the eigenvalues of each PM, Rλ, λ=λ1 . . . λQ, must be distinct for unambiguous TM recovery, in agreement with previous findings. Light must therefore be coupled between different fibre modes and polarisations, so a conventional partial mirror reflector would not work because it preserves polarisation states leading to repeated matrix eigenvalues. This insight informs reflector assembly 42 design. In reality, the relatively small spectral bandwidths of typical imaging fibres (˜5 nm) would require the use of optical components (e.g. filters) with impractically sharp spectral responses to achieve different reflectance behaviour over such a small wavelength range.
By way of further explanation in relation to the zeroth-order model, using the same sampling matrix, T, “distilled” square versions of all two (or three) refection mode matrices can be produced (taken at different wavelengths), which are denoted C1, C2, and C3. Since each of these uses a different reflector 48 but has approximately the same transmission matrix, the following can be written:
C
1
=A
T
R
1
A (3)
C
2
=A
T
R
2
A (4)
C
3
=A
T
R
3
A (5)
The following non-limiting steps may then be taken to recover the instantaneous transmission matrix, A (however any suitable alternative method may be employed).
Starting from (3), (4) and (5):
C
2
−1
=A
−1
R
2
−1(AT)−1
C
2
−1
C
1
=A
−1
R
2
−1(AT)−1ATR1A
C
2
−1
C
1
=A
−1
R
2
−1
R
1
A
C
α
=A
−1
R
α
A (6)
AC
α
=R
α
A
R
α
A−AC
α=0 (7)
where Cα=C2−1C1 and Rα=R2−1R1. It is important to note from (6) that Cα and Rα are similar matrices and so have the same eigenvalues.
In a similar fashion, the following can be derived:
C
3
−1
=A
−1
R
3
−1(AT)−1
C
3
−1
C
2
=A
−1
R
3
−1(AT)−1ATR2A
C
3
−1
C
2
=A
−1
R
3
−1
R
2
A
C
β
=A
−1
R
β
A (8)
AC
β
=R
β
A
R
β
A−AC
β=0 (9)
where Cβ=C3−1C2 and Rβ=R3−1R2.
Both (7) and (9) are examples of Sylvester equations, so known methods of solving them can be employed.
Because (7) and (9) have RHS=0, the trivial solution A=0 satisfies both equations. However, there must exist at least one other non-trivial solution for A. Therefore, an attempt may first be made to determine the space of the non-trivial solutions so that the true answer may be searched for within that space. This may be achieved by implementing a modified version of the Bartels-Stewart algorithm for solving Sylvester equations which is described in R. Bartels and G. W. Stewart (R. Bartels and G. W. Stewart, “Algorithm 432: Solution of the Matrix Equation (AX+XB=C),” Comm. ACM, vol. 15, no. 9, pp. 820-826, 1972.) which is hereby incorporated by reference in its entirety. The advantage of using this approach over other methods (e.g. using Kronecker products to solve for a vectorized matrix, vec(A)) is that its computational complexity is significantly reduced (O(n3) vs. O(n6)) as its memory usage.
Using the Schur matrix decomposition (described in R. A. Horn and C. R. Johnson, Matrix analysis. 2013, which is hereby incorporated by reference in its entirety), any square matrix, A, can be decomposed as:
A=QUQ
−1 (10)
where Q is unitary and U is an upper triangular matrix. Importantly, the diagonal of U comprises the eigenvalues of A. These are in an arbitrary order, but there are algorithms that enable these to be sorted, e.g. in descending order of magnitude.
From (7) and (10), it can be written:
R
α
=Q
R
U
R
Q
R
−1
C
α
=Q
C
U
C
Q
C
−1 (12)
Substituting (11) and (12) back into (7) gives:
Q
R
U
R
Q
R
−1
A−AQ
C
U
C
Q
C
−1=0
Q
R
U
R
Q
R
−1
AQ
C
−AQ
C
U
C=0
U
R
Q
R
−1
AQ
C
−Q
R
−1
AQ
C
U
C=0
If:
A′=Q
R
−1
AQ
C (13)
then:
U
R
A′−A′U
C=0 (14)
(14) is another Sylvester equation but, crucially, UR and UC are upper triangular so it can be solved element by element. If all matrices are square and of size N×N, the element (N, 1) of the RHS zero matrix may be solved first. It can be seen that:
ur
N,N
a
N,1
−a
N,1
uc
1,1=0 (15)
where urm,n is the element in row m, column n of UR, ucm,n is the element in row m, column n of UC and am,n is the element in row m, column n of A′.
It may be assumed that the eigenvalues of Rα, are distinct. This can be arranged by suitable design of the reflectors R1 and R2. In general, typical non-unitary reflectors will likely have distinct eigenvalues when characterised experimentally.
Assuming that the eigenvalues of Rα, are distinct, it is known that they are equal to the eigenvalues of Cα, because the two are similar matrices by (6). Therefore, the values on the diagonal of UR are distinct from one another, and are equal to the values on the diagonal of UC. Therefore, the only way that (15) can hold is if aN,1=0.
Next, element (N−1, 1) is considered, giving:
ur
N-1,N-1
a
N-1,1
+ur
N-1,N
a
N,1
−a
N-1,1
uc
1,1=0
Since it is known that aN,1=0, by the above reasoning, it is concluded that aN-1,1=0. Continuing on up this column, it is found that all elements must be zero until the first row is reached, where:
ur
1,1
a
1,1
−a
1,1
uc
1,1=0
It is known that urm,m=ucm,m for every m because Rα, and Cα, are similar matrices by (6). Therefore, any value of a1,1 satisfies this equation meaning this element cannot be determined. It is found that diagonal elements, am,m, are indeterminate but that all other elements in the matrix can be computed from these. For example, the element ap,q with P<Q (i.e. the upper triangular part of A′) is given by:
Rearranging, we can write:
If P is varied from 1 to Q−1 in (16), this produces Q−1 equations. If the diagonal elements of A′ are arbitrarily fixed, Q−1 unknowns are left. This system of linear equations can be solved to find all other elements of the Pth column of A′. By varying Q and solving each resultant system of equations, all non-diagonal elements of A′ can be fully determined given a set of arbitrary diagonal elements, am,m (m=1 . . . N). Since every candidate solution of (14) is a linear function of N complex numbers (i.e. of an N-dimensional vector), it can be inferred that the true matrix, A′, can be expressed as a linear combination of at most N matrices, each of which is a solution of (14).
N orthogonal 1×N complex vectors are generated, and, for each of these, the associated solution of (14), termed A′N, is found. Equation (13) is then used to get the associated matrix AN that is a solution to (7). The problem is then reduced to finding a 1×N vector of weights, w=[w1w2 . . . wN], such that:
w
1
A
1
+ . . . +w
N
A
N
=A (17)
In effect, the task is to now solve an N-dimensional problem to find an N×N matrix, which is a significant reduction in complexity. A further reduction of the problem is achieved by noting that (17) is an over-determined problem. That is, to uniquely determine the optimal set of weights, w, it is sufficient to consider only a subset comprising the same B (≥N) elements from every AN. This reduced problem of optimizing N weights across B elements is referred to as the weight-reduced solution space.
An alternative reduced solution space can be constructed as follows. If the B relevant element from each AN are taken and put into column vectors b1 . . . bN, a matrix B can be formed:
B=[b1 . . . bN] (18)
so that we can write:
Bw
T
=b
est (19)
Where best is an estimate of the B relevant elements of the true transmission matrix, A. Since B is either square or is a tall matrix, we can pre-multiply by its Moore-Penrose pseudo-inverse, B+:
B
+
Bw
T
=B+b
est
Iw
T
=B
+
b
est
w
T
=B
+
b
est
We can then multiply both sides by B to get:
Bw
T
=BB
+
b
est
and substitute in (19):
b
et
=BB
+
b
est (20)
Clearly, the solution best must then be an eigenvector of the B×B matrix, B+B, and so we call this eigenspace.
If a 4-layer stack design of the reflector assembly 42 (as shown in
b
est
=B
α
B
α
+
b
est (21)
A different amount of reflection from 3 reflectors 48 will produce Cβ and subsequently Bβ, and we can re-derive (20) as:
b
est
=B
β
B
β
+
b
est (22)
The different amount of reflection from the three reflectors 48 is determined by the characterisation wavelengths used and the spectral profile of the absorptive filters 46. In certain embodiments these wavelengths and spectral profiles may be determined through an optimisation process to produce maximally different Rα and Rβ and hence Bα and Bβ.
In certain embodiments, best may be recovered from the above equations by one of two approaches (which are described below). In accordance with other embodiments of the invention, alternative suitable approaches may be adopted. A first approach seeks to achieve optimization using a prior. In general, the elements of transmission matrices tend to follow some pattern even under large perturbations in temperature or bending. For example, in an MCF, it is typically found that there is usually high power coupling between adjacent fibre cores and weak coupling between distant cores. In MMF, modes with similar propagation constants (degenerate modes) experience high power coupling and those with very different propagation constants experience lower coupling.
This idea can be formalized as determining a statistical prior distribution of transmission matrices. To determine such a distribution, many experimental transmission matrix realisations could be measured, i.e. various fibres under randomized bending and temperature conditions, and then fit a multivariate distribution to this data. An estimated transmission matrix may be denoted by Aest and the probability density function of this fitted prior distribution may be denoted by f(Aest). As an example, f(A) might be a complex multivariate Gaussian distribution with some covariance between each element of A. In the weight-reduced solution space, the optimization procedure then seeks to find the optimum weights (w from (17)) as follows:
The ability of this approach to find the true solution space relies on the fact that typically, the matrices An are distributed quite differently from realistic transmission matrices. For example, they might be non-sparse whereas real transmission matrices are typically very sparse. This means that random linear combinations of An will typically produce mostly non-zero elements in the transmission matrix estimate Aest (=w1A1+ . . . +wNAN). If the prior distribution favours sparsity then most combinations of weights will have a low likelihood. There is therefore a high probability that the most likely Aest is in fact the correct one. Whilst this explanation has not, at present, been rigorously proven to converge on the correct answer, it has the significant advantage that it only requires two reflectors (i.e. the reflector assembly 42 shown in
A second approach seeks to achieve optimization through the addition of a third reflector 48. By recursive substitution of (21) and (22), the following expression can be written:
b
est
=B
α
B
α
+
B
β
B
β
+
. . . B
α
B
α
+
B
β
B
β
+
b
est
b
est=(BαBα+BβBβ+)Mbest (24)
where M is typically >3.
Equation (24) represents and adapted version of the power method for finding dominant eigenvalues of a matrix (described in E. Bodewig, Matrix Calculus. 1959, which is hereby incorporated by reference in its entirety) because the eigenvalue of the matrix (BαBα+BβBβ+)M corresponding to the desired solution is known to be real (=1) and is dominant as only a single solution exists (all other eigenvalues should tend to 0). By applying the power method with large M, any input, b′est, to the right hand side of (24) will produce the desired eigenvector best when multiplied by the matrix (BαBα+BβBβ+)M. Alternatively, after only a small number of iterations, the dominant eigenvector of (BαBα+BβBβ+)M will be the desired solution and can be obtained by standard eigendecomposition of (BαBα+BβBβ+)M
This optimization approach has the advantage that no prior knowledge is required to produce a unique solution. Furthermore, it is not an iterative process and can recover transmission matrices in a single step. On the other hand, three reflectors 48 are required (i.e. utilizing the reflector assembly 42 shown in
Given the limitations of the zeroth-order approach, an alternative model is considered that relates TMs at different wavelengths, λp and λq, as:
where log(Aλ
dm=dAm (26)
where dA∈2M×2M represents an infinitesimal coupling matrix, and m∈2M×1 represents the field in each of M spatial modes over 2 orthogonal polarisations. This matrix equation represents a system of 2M linear differential equations. The solution to these equations for any arbitrary input set of modes, x, after travelling a distance along the central axis of the fibre 26 is given by a matrix exponential:
y′=
(27)
Since the TM Aλ1= where λ
Matrix logarithms in general produce degenerate solutions with eigenvalues of the form γ+i2πn/λ
Substituting Equations 28 and 29 into Equation 27 produces Equation 25 as desired.
In contrast to the zeroth-order model, when the TMs are related by this first-order model it is not straightforward to solve for Aλq based on Equation 2 applied at different wavelengths. Therefore, an optimisation-based approach is presented that can compute Aλ
A solution to each sub-optimization problem within the loop, λ
As will be appreciated, in certain embodiments of the invention, the projected calibration patterns are not initially separated into modes. Rather, all incident modes are reflected without any time delays being introduced. More specifically, a random superposition of all fibre modes (since each optical wavelength is distributed across a random superposition of all fibre modes) is reflected and arbitrarily recombined/mixed. Advantageously, this approach can utilize an arbitrary number of fibre modes thereby making it particularly suitable for imaging (which requires a large number of modes).
Once the instantaneous transmission matrix has been determined (e.g. by the above-described methods), the sample 30 may be imaged.
A non-limiting example of method 200 is set out below.
Imaging is performed using a new wavelength, the imaging wavelength λQ+1, which is longer than the wavelengths used for TM characterisation, i.e. λQ+1>λ1, . . . , λQ, in order that light may pass through the reflector assembly 42. The TM at this wavelength, Aλ
y′
illum
=A
λ
x
illum (30)
Next, the physical model from the perspective of imaging (
y″
illum
=A
refl
A
λ
x
illum (31)
with y″illum∈2N, Setting N≥M allows oversampled illumination fields. y″illum can be propagated by some linear operator, G∈2N×2N, through free-space (e.g. Fresnel or Fraunhofer propagation) before reaching the target 30 where the field will be:
y″
illum
=GA
refl
A
λ
x
illum (32)
G is parameterised by the distance, d, between the sample 30 and the distal surface of the reflector assembly 42 (
Representing the target by a vector, rtarget∈2N, an element-wise (Hadamard) product is performed with the illumination to give:
x′″
target
=y′″
illum
·r
target (33)
This light reflected from the target 30 propagates backwards, first through free-space GT then through the reflector assembly 42, AreflT to produce a field at the distal facet:
x′
target
=A
refl
T
G
T
x′″
target (34)
Next, the illumination light that is reflected back from the reflector assembly 42 is also considered:
x′
refl
=R
λ
A
λ
x
illum
We sum the two fields and propagate back through the fibre 26 giving:
y
total
=A
λ
T(x′target+x′refl) (35)
as the measured quantity at the camera plane, thus obtaining image data (step 204).
The goal is to recover rtarget from the raw measured data, ytotal (Equation 35). To do this, x′″target is first recovered by rearrangement of Equation 35 and substitution of Equations 34 and 33:
(Aλ
x′
target=(Aλ
x′″
target=(AreflTGT)+((Aλ
where ( . . . )+ a generalised inverse because Arefl may be non-square. Arefl and Rλ
Whilst the prior art methods have indirectly used phase and/or polarisation information to recover amplitude-only images from MMFs, embodiments of the present invention are capable of directly retrieving wide-field en-face images of these properties together with amplitude information.
Holographic endoscopes utilizing the optical system described above (and the associated methods) may record high resolution (e.g. 9.0±2.6 μm amplitude, phase; 36.0±10.4 μm polarimetric) images at working distances up to 1-2 cm and fields-of-view of ˜1×1 cm, in line with conventional endoscopy. The ability to adaptively change working distance without distal optics is a key advantage associated with embodiments of the present invention. Embodiments of the present invention offer potential ‘red-flag’ surveying of large areas for abnormalities, followed by zooming in to perform high resolution ‘optical biopsy’, which is currently impossible in a single device.
A further advantage associated with embodiments of the present invention is that is sensitive to a wider range of optical parameters relative to prior art devices and methods. In principle, this improved sensitivity is sufficient to observe small, localised proliferations of cells associated with early tumorigenesis.
Quantitative phase and resolved polarimetric properties have shown promise for enhancing cancer detection in ex vivo and in vivo rigid endoscope studies. Further, they are largely robust to variations in absolute intensity under varying measurement conditions. Retrieval of these properties through a flexible endoscope could therefore enhance contrast for pre-malignant and malignant changes during diagnostic endoscopy in a subject.
Obtaining a label-free diagnosis equivalent to that provided by fluorescence staining avoids the need for extended procedure time and potential toxicities associated with in vivo dye application. Importantly, phase and polarimetric images may contain more relevant diagnostic information than amplitude-only images, demonstrating a key advance over what is possible using current commercial endoscopes. The ability to holographically produce illumination patterns at the distal end of the fibre, once the transmission matrix is known, would also enable other modalities of imaging such as fluorescence. It could even pave the way for super-resolution imaging techniques such as STED or STORM through an endoscope.
In accordance with an aspect of the present invention, there is provided a method of determining a presence of a physiological condition in a subject, the method comprising: producing recovered image data relating to a tissue sample using a method as described above or the optical system described above; and determining, in dependence on the recovered image data, a presence of the physiological condition in the tissue sample.
The step of determining a presence of the physiological condition in the tissue sample may comprise:
using the recovered image data as contrast mechanisms to identify tissue morphology or microstructure associated with the physiological condition; and/or
using the recovered image data to produce images showing the spatial variation or spatial entropy of one or more of amplitude, phase and polarisation, and using differences thereof to delineate healthy areas of the tissue sample (e.g. areas not affected by the physiological condition) and areas of the tissue sample that are affected by the physiological condition (e.g. as described above with reference to
The physiological condition may, for example, be cancer or a pre-cancerous condition, the presence of scar tissue, or the presence of inflammation. The above described method may be useful in determining the presence of other physiological conditions.
In accordance with another aspect of the present invention, there is provided a computer implemented method comprising:
receiving recovered image data relating to a tissue sample obtained using a method as described above or the optical system described above; and
determining, in dependence on the recovered image data, a presence of a physiological condition in the tissue sample.
The step of determining a presence of the physiological condition in the tissue sample may comprise:
using the recovered image data as contrast mechanisms to identify tissue morphology or microstructure associated with the physiological condition; and/or
using the recovered image data to produce images showing the spatial variation or spatial entropy of one or more of amplitude, phase and polarisation, and using differences thereof to delineate healthy areas of the tissue sample (e.g. areas not affected by the physiological condition) and areas of the tissue sample that are affected by the physiological condition (e.g. as described above with reference to
The physiological condition may, for example, be cancer or a pre-cancerous condition, the presence of scar tissue, or the presence of inflammation. The above described method may be useful in determining the presence of other physiological conditions.
Embodiments of the present invention may form part of or be utilized in an instrument that improves early detection of cancer, and thus improve survival rates (e.g. from oesophageal cancer).
Throughout the description and claims of this specification, the words “comprise” and “contain” and variations of them mean “including but not limited to”, and they are not intended to (and do not) exclude other moieties, additives, components, integers or steps. Throughout the description and claims of this specification, the singular encompasses the plural unless the context otherwise requires. In particular, where the indefinite article is used, the specification is to be understood as contemplating plurality as well as singularity, unless the context requires otherwise.
Features, integers, characteristics, compounds, chemical moieties or groups described in conjunction with a particular aspect, embodiment or example of the invention are to be understood to be applicable to any other aspect, embodiment or example described herein unless incompatible therewith. All of the features disclosed in this specification (including any accompanying claims, abstract and drawings), and/or all of the steps of any method or process so disclosed, may be combined in any combination, except combinations where at least some of such features and/or steps are mutually exclusive. The invention is not restricted to the details of any foregoing embodiments. The invention extends to any novel one, or any novel combination, of the features disclosed in this specification (including any accompanying claims, abstract and drawings), or to any novel one, or any novel combination, of the steps of any method or process so disclosed.
The reader's attention is directed to all papers and documents which are filed concurrently with or previous to this specification in connection with this application and which are open to public inspection with this specification, and the contents of all such papers and documents are incorporated herein by reference.
Number | Date | Country | Kind |
---|---|---|---|
1818290.7 | Nov 2018 | GB | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/GB2019/053195 | 11/11/2019 | WO | 00 |