The present disclosure relates to an on-line biomedical spectroscopy software platform for real-time cancer diagnostics at endoscopy and methods for instrument-independent measurements for quantitative analysis in fiber-optic Raman spectroscopy.
Raman spectroscopy is a technique which uses inelastic or Raman scattering of monochromatic light. Conventionally, the monochromatic light source is a laser in the visible or near infrared (“NIR”) range. The energy of the scattered photons is shifted up or down in response to interaction with vibrational modes or excitations in the illuminated material, varying the wavelength of the scattered photons. Accordingly, the spectra from the scattered light can provide information about the scattering material.
NIR Raman spectroscopy is known as a potential technique for characterisation and diagnosis of precancerous and cancerous cells and tissue in vivo in a number of organs. The technique is desirable as it can be non-invasive or minimally invasive, not requiring biopsies or the other removal of tissue. It is known to use NIR Raman spectroscopy in two wavelength ranges. The first is the so-called fingerprint (“FP”) range, with wave numbers from 800 to 1800 cm−1, owing to the wealth of highly specific bimolecular information, for example from protein, DNA and lipid contents, contained in this spectral region for tissue characterisation and diagnosis. The disadvantage of this wavelength range is, that when used with a commonly used 785 nm laser source, the strong tissue autofluorescence background signal can be generated. Further, where the probe uses optical fiber, a Raman signal is scattered from the fused silica in the optical fibers. In particular, where a charge-coupled device (“CCD”) is used to measure the scattered spectra, the autofluorescence signal can saturate the CCD and interfere with the detection of the inherently very weak Raman signals in this wavelength range.
Another problem with fiber-optic Raman spectroscopy as a technique is that of standardization of instruments. The fiber-optic Raman spectroscopy technique has mainly been limited to single systems and no attempts have been made to transfer into multi-centre clinical trials or routine medical diagnostics. This is mainly because Raman spectrometer instruments are generally dissimilar (i.e., optics, response function, alignment, throughput etc.) and in general produce very different Raman spectra. Further, fiber optic Raman probes have limited lifetimes and must be replaced or interchanged periodically. Unfortunately, Raman data acquired using different fiber optic probes cannot be compared, because each fiber optic probe has its own unique background as well as being associated with different transmissive spectral properties. The different transmissive characteristics significantly distort the spectral intensities making the tissue Raman spectra obtained with different fiber optic probes incomparable. As a consequence, multivariate diagnostic algorithms developed on a primary clinical platform cannot be applied to secondary clinical platforms. In particular, the quantitative measurement of tissue Raman intensity is one of the most challenging issues in fiber optic biomedical Raman applications. The instrument/fiber probe-independent intensity calibration and standardization is essential to the realization of global use of fiber optic Raman spectroscopy in biomedicine. For this reason, a multivariate statistical diagnostic model constructed using a ‘master’ probe cannot be applied to spectra measured with a ‘slave’ probe. In order for Raman technique to become a widespread tool for cancer screening on a global scale, there is a need to standardize both Raman spectrometers and fiber optic probes especially for biomedical applications. Most of the reported studies have focused on inter-Raman spectrometer standardization for measurements of simple chemical mixtures without fiber optic probes. In general Raman spectroscopy of simple chemical mixtures cannot be compared with the fiber optic Raman spectroscopy of heterogeneous biological tissue samples.
A further problem with standardizing results across instruments is that of spectral variation associated with the laser excitation power. Conventionally, Raman spectra are normalized which preserves the general spectrum shape, but this removes the absolute quantitative spectral characteristics. It has been known to attempt to monitor delivered laser power in fibre-optic Raman probes by, for example, embedding a diamond in the fibre tip or locating a polymer cap in the laser light path as a reference. However, these solutions are not satisfactory and may cause interferences in the required spectral regions.
A further problem in using optical spectroscopic techniques (including reflectance fluorescence and Raman) for in vivo diagnosis of cancer and precancerous conditions is that data analysis mostly been limited to post-processing and off-line algorithm development. This is true for endoscopic analysis because a large number of spectra collected during endoscopy are outliers. It would be useful to have a system that allows for real-time diagnosis for endoscopy.
According to a first aspect there is provided a method of calibrating a fiber-optic Raman spectroscope system, the system comprising a laser source, a spectroscope and a fiber optic probe to transmit light from the laser source to a target and return scattered light to the spectroscope, the method comprising transmitting light from the laser source to a standard target having a known spectrum, recording a calibration spectrum of the scattered light from the standard target, comparing the known spectrum and the calibration system and generating a transfer function, and storing the transfer function.
The method may further comprise the steps of subsequently illuminating a test subject, recording a spectrum and correcting the spectrum in accordance with the stored transfer function.
The method may comprise recording calibration spectra for each of a plurality of fiber optic probes, calculating a transfer function for the system including each of said probes, and associating the transfer function with the corresponding probe.
The spectrometer has an associated spectrometer transfer function and the probe may have an associated probe transfer function, and the transfer function may be a function of the spectrometer transfer function and the probe transfer function.
The method may comprise, on a primary spectrometer system, calculating a first transfer function with a primary fiber optic probe, and a second transfer function with a secondary fiber optic probe, and calculating a (inter-probe) calibration function based on the first transfer function and second transfer function.
The method may comprise associating the calibration function with the secondary fiber optic probe.
The method may comprise, on a secondary spectrometer system, using the primary fiber optic probe and generating a secondary system transfer function and storing the secondary system transfer function.
The method may comprise using the secondary fiber optic probe with the secondary spectrometer system and modifying the stored secondary system transfer function in accordance with the calibration function.
The method may comprise the initial step of performing a wavelength-axis calibration of the secondary spectrometer system in accordance with the primary spectrometer system.
According to a second aspect there is provided a method of operating a Raman spectroscope system, the system comprising a laser source, a spectroscope and a fiber optic probe to transmit light from the laser source to a target and return scattered light to the spectroscope, the method comprising transmitting light from the laser source to a target having a known spectrum, recording a spectrum of the scattered light from the target, and modifying the recorded spectrum in accordance with a stored transfer function.
The stored transfer function may be associated with the spectrometer and the fiber optic probe.
The stored transfer function may be associated with the spectrometer and a primary fiber optic probe and the method may further comprise modifying the stored transfer function in accordance with a stored calibration function associated with the fiber optic probe.
According to a third aspect there is provided a Raman spectroscope system comprising a laser source, a spectroscope and a fiber optic probe to transmit light from the laser source to a target and return scattered light to the spectroscope, and a stored transfer function, the system being operable to transmit light from the laser source to a target having a known spectrum, record a spectrum of the scattered light from the target, and modify the recorded spectrum in accordance with the stored transfer function.
The stored transfer function may be associated with the spectrometer and the fiber optic probe.
The stored transfer function may be associated with the spectrometer and a primary fiber optic probe and the method may further comprise modifying the stored transfer function in accordance with a stored calibration function associated with the fiber optic probe.
According to a fourth aspect there is provided a method of estimating the laser power transmitted in a Raman spectrometer system, the system comprising a laser source, a spectroscope and a fiber optic probe to transmit light from the laser source to a target and return scattered light to the spectroscope, the method comprising transmitting light from the laser source to a plurality of targets, for each target, measuring the transmitted power of the light from the laser source and the spectrum of the scattered light at the spectroscope, performing a multivariate analysis of the captured spectra with the measured transmitted power as a dependent variable, and storing a resulting model.
The method may comprise the step of transmitting laser light to a test target, supplying a captured spectrum to the model, and calculating an estimate of the transmitted power.
According to a fifth aspect there is provided a method of subtracting a background signal from a fiber-optic Raman spectroscope system having a fiber-optic probe, the method comprising the steps of;
a) storing a background spectrum,
b) receiving a test spectrum,
c) estimating a background contribution using one or more reference peaks,
d) multiplying the background spectrum by a correction factor based on the estimated background contribution and subtracting it from the test spectrum,
e) checking the test spectrum for a remaining background contribution, and
f) if the background contribution is negligible, outputting the test spectrum, otherwise repeating steps (c) to (e).
The one or more reference peaks may comprise one or more peaks corresponding to silica or sapphire in the fiber-optic probe.
According to a sixth aspect there is provided a computer implemented method for real-time diagnosis using Raman spectroscopy during endoscopy. The method comprises receiving at least one spectrum associated with a tissue; analyzing the at least one spectrum in a model that uses the spectrum to determine a score wherein said score indicates a likelihood of the tissue being cancerous; and outputting said score.
In some embodiments the model is generated using an interpretation function selected from the group consisting of partial least squares-discriminant analysis, principal component analysis linear discriminant analysis, ant colony optimization linear discriminant analysis, classification and regression trees, support vector machine, and adaptive boosting.
In some embodiments the at least one spectrum is generated by Raman spectroscopy. Analyzing the at least one spectrum in a model may comprise analyzing the at least one spectrum in a first model and a second model. In some embodiments the model is selected based on the tissue analyzed. In some embodiments the score indicates whether the tissue is normal, intestinal metaplasia, dysplasia or neoplasia.
In some aspects analyzing the at least one spectrum comprises: performing outlier analysis; and responsive to the outlier analysis determining that the at least one spectrum is an outlier, rejecting the spectrum. Performing outlier analysis may comprise principal component analysis.
In some aspects an audio emitting device emits an audio signal responsive to the outlier analysis determining that the at least one spectrum is an outlier. Responsive to the determination that the spectra is an outlier method instructs the spectrometer to acquire an additional at least one spectrum which is received by the system for analysis.
In some embodiments, the audio emitting device to emit an audio signal identifying the tissue as normal, dysplasia or neoplasia. In some embodiments, the audio signal associated with each diagnosis is different and also different from an audio signal associated with the determination of an outlier spectrum.
In some embodiments, the diagnosis takes place during the endoscopic procedure.
Also provided are systems for carrying out the computer-implemented methods as well as non-transitory computer readable media with instructions thereon for carrying out the computer-implemented methods.
Embodiments of the disclosed system and methods are described by way of example only with reference to the accompanying drawings.
Provided herein is an on-line system and method for biomedical spectroscopy (i.e., reflectance, fluorescence and Raman spectroscopy) for realizing real-time detection of neoplastic lesions in different organs (e.g., gastrointestinal tracts (stomach, esophagus, colon), bladder, lung, oral cavity, nasopharynx, larynx, cervix, liver, skin, etc.) at endoscopy. The diagnostic method integrates excitation source synchronization, integration-time adjustment, data acquisition, preprocessing, outlier analysis and probabilistic multivariate diagnostics (i.e., partial least squares-discriminant analysis (PLS-DA), principal component analysis (PCA)-linear discriminant analysis (LDA), ant colony optimization (ACO)-LDA, classification and regression trees (CART), support vector machine (SVM), adaptive boosting (AdaBoost) etc.) including multi-class diagnostics based on comprehensive spectral databases (i.e., Raman, fluorescence, reflectance) of different organs.
In one embodiment, the disclosed system and method integrates the on-line diagnostic framework with the recently developed multimodal image-guided (WLR/NBI/AFI) Raman spectroscopic platform for early diagnosis and detection of precancer and cancer in the upper GI at endoscopy. The accumulation of tissue Raman spectra and automatic scaling of integration time with a predefined upper limit of 0.5 s allows instant acquisition of in vivo tissue spectra with improved SNR while preventing CCD signal saturation. This is especially important for endoscopic diagnostics where the autofluorescence intensity varies significantly among different anatomical regions (e.g., antrum and body in the gastric, bronchi in the lung) likely caused by distinct endogenous fluorophores in the tissue.
With specific reference now to the drawings in detail, it is stressed that the particulars shown are by way of example and for purposes of illustrative discussion of the preferred embodiments, and are presented in the cause of providing what is believed to be the most useful and readily understood description of the principles and conceptual aspects of the disclosed system and methods. In this regard, no attempt is made to show structural details of the disclosed system and methods in more detail than is necessary for a fundamental understanding of the disclosed system and methods, the description taken with the drawings making apparent to those skilled in the art how the several forms of the disclosed system and methods may be embodied in practice.
Before explaining at least one embodiment of the disclosed system and methods in detail, it is to be understood that the disclosure is not limited in its application to the details of construction and the arrangement of the components set forth in the following description or illustrated in the drawings. The disclosed system and methods are applicable to other embodiments or of being practiced or carried out in various ways. Also, it is to be understood that the phraseology and terminology employed herein is for the purpose of description and should not be regarded as limiting.
Referring now to
The Raman spectroscopy system is generally shown at 20. A monochromatic laser source is shown at 21, in the present example a diode laser with an output wavelength of about 785 nm. Light from the laser diode 21 is passed through a proximal band pass filter 22, comprising a narrowband pass filter being centred at 785 nm with a full width half max of ±2.5 nm. The light is passed through a coupling 23 into an excitation optical fiber 25 provided as part of a fiber bundle. The excitation fiber 25 has a diameter of 200 μm and a numerical aperture (‘NA’) of 0.22. Light transmitted by the excitation fiber 25 enters a ball lens 26 at the end of the endoscope 11, in the present example comprising a sapphire ball lens with a diameter of about 1.0 mm and a refractive index n=1.77. As illustrated in
Collected scattered light returned by collection fibers 28 is passed through a long pass inline collection filter 29 which similarly has a cutoff at ˜800 nm. The configuration of sapphire ball lens 26, excitation and collection fibers 25, 28, band-pass filters 22, and long-pass filter 29 provides a good system for selectively collecting backscattered Raman photons from the tissue 27.
The scattered returned light is then separated at spectrograph 30 and the resulting spectrum is imaged at a light-sensing array 34, in the present example a charge-couple device (‘CCD’). A computer shown at 35 controls the operation of the system, processes and stores the spectra and control data, and provides results and data to a user.
In one embodiment, the computer 35 comprises at least one processor coupled to a chipset. Also coupled to the chipset are a memory, a storage device, a keyboard, a graphics adapter, a pointing device, an audio emitting device and a network adapter. A display is coupled to the graphics adapter. In one embodiment, the functionality of the chipset is provided by a memory controller hub and an I/O controller hub. In another embodiment, the memory is coupled directly to the processor instead of the chipset.
The storage device is any device capable of holding data, like a hard drive, compact disk read-only memory (CD-ROM), DVD, or a solid-state memory device. The memory holds instructions and data used by the processor. The pointing device may be a mouse, track ball, or other type of pointing device, and is used in combination with the keyboard to input data into the computer system. The graphics adapter displays images and other information on the display. The network adapter couples the computer system to a local or wide area network.
As is known in the art, a computer 35 can have different and/or other components than those described previously. In addition, the computer can lack certain components. Moreover, the storage device can be local and/or remote from the computer (such as embodied within a storage area network (SAN)).
As is known in the art, the computer is adapted to execute computer program modules for providing functionality described herein. As used herein, the term “module” refers to computer program logic utilized to provide the specified functionality. Thus, a module can be implemented in hardware, firmware, and/or software. In one embodiment, program modules are stored on the storage device, loaded into the memory, and executed by the processor.
Embodiments of the entities described herein can include other and/or different modules than the ones described here. In addition, the functionality attributed to the modules can be performed by other or different modules in other embodiments. Moreover, this description occasionally omits the term “module” for purposes of clarity and convenience.
The computer 35 also performs preprocessing the spectral data. As the measured tissue Raman spectra are substantially obscured by the tissue autofluorescence background, preprocessing of in vivo tissue Raman spectra is necessary to extract the weak Raman signals. The raw Raman spectra measured from in vivo tissue represent a combination of the weak Raman signal, intense autofluorescence background, and noise. The spectra are first normalized to the integration time and laser power. The spectra are then smoothed using a first-order Savitzky-Golay smoothing filter (window width of 3 pixels) to reduce the noise. A fifth-order polynomial was found to be optimal for fitting the autofluorescence background in the noise-smoothed spectrum, and this polynomial is then subtracted from the raw spectrum to yield the tissue Raman spectrum alone. The computer 35 can also including diagnostic algorithms for precancer and cancer detection.
Spectrometer and Fibre-Optic Probe Calibration
It is known that different spectrometers will have different transfer functions, i.e. will show differing intensity variations within spectra even when illuminated using the same source. As illustrated in
Using a fluorescent standard source, the transfer function, i.e. the wavelength-dependent response of the spectrometer, can be given by
(eqn. 1) where F(λ) is the correct fluorescent standard spectrum, S(λ) is the measured spectrum of the fluorescent standard source and T(λ) is the transfer function of the spectrometer. Accordingly, as T(λ) is known, a correctly calibrated Raman spectrum of a new sample R(λ) can be calculated by
(eqn. 2) where S(λ) is the measured sample spectrum.
The transfer function T(λ) is a function both of the spectrometer transfer function TS(λ) and a probe transfer function TP(λ). Equation 2 can therefore be written as
(eqn. 3). As fibre-optic probes are replaceable and may be consumables, it will be apparent that when a new probe with a new probe transfer function TP is inserted, the overall transfer function of the system will change.
Referring now to
A first method of calibration is shown in
When probe 53a is discarded and it is desired to carry out tests on a new subject, a replacement probe 53b may be substituted, in which case the method of
In an alternative process as illustrated in
An alternative approach is illustrated with reference to
(eqn. 4) and a spectrum SSP(λ) using the secondary probe, where
(eqn. 5). Equations 4 and 5 can be divided to relate the two probe transfer values through a probe calibration function TCF, where
(eqn. 6). Consequently, from equations 2 and 6, when the secondary spectrometer is used with the secondary probe, the measured spectrum S(λ) and Raman spectrum R(λ) are related by
(eqn. 7) where T(λ)=TS (λ)TPP(λ) is the stored system transfer function measured for the secondary spectrometer using the master probe.
As illustrated in
As shown in
As illustrated in
In any of the methods therefore, by matching the secondary spectrometer characteristics back to the primary spectrometer characteristics, and storing either the transfer function for the spectrometer and probe combination or a transfer function for the system incorporating a master probe and a calibration function for use with a secondary probe, spectra captured using different spectrometer and probe combinations will nevertheless be consistent and comparable.
This is apparent from
Raman spectra were measured from the gastric with two different probes (n=902 spectra). A principal component analysis (PCA) was conducted before and after calibration of the secondary probe.
Monitoring Laser Power
Advantageously, it has been found that the spectral characteristics of the fibre probe and sapphire ball lens in captured Raman spectra can be used as internal reference to derive the transmitted laser power without requiring the provision of any additional components in the optical train. As shown in the method of
In the present example, a graph of the number of latent variables included against the root mean square error is shown in
Iterative Background Subtraction
A method of subtracting the background Raman spectrum resulting from fluorescence, Raman scattering in the silica of the probe and the sapphire of the lens is shown with reference to
As shown at step 110 in
At step 114, the spectrum is checked for the presence of remaining background. If the background has been fully removed, (i.e., when the silica and sapphire signal contributes negligible to the tissue Raman spectrum), the spectrum is passed for output or further analysis as shown at step 115. If a background signal is still present, then steps 112 to 114 are repeated as shown by arrow 116.
The method need not be limited to single silica/sapphire peaks. Multivariate analysis (e.g., partial least squares and curve resolution methods etc.) can also be used for this purpose.
By way of example,
As shown in
Combined System
The various disclosed methods can be used together. One embodiment of such a combination is illustrated in
In combining the disclosed calibration method with a diagnostic method, instrument-independent fiber optic Raman spectroscopy is possible for quantitative tissue analysis and characterization. This allows for comparison of spectra taken by different instruments and also spectra taken with the same instrument but different probes. This is important for diagnosis in that it allows use of spectra taken on different machines or with different probes to be used for comparison. This is important for increasing accuracy of diagnosis.
Real-Time Cancer Diagnostics
An on-line biomedical spectroscopy (i.e., reflectance, fluorescence and Raman spectroscopy) system and method realizes real-time cancer diagnostics at clinical endoscopy and can interface with clinicians using auditory feedback as well as graphical display of the outcome of probabilistic diagnostic algorithms with the predicted pathology. Taking Raman endoscopy in the gastric as an example (
In order to address inter-anatomical and inter-organ spectral variances the online framework implements organ specific diagnostic models and switches among the spectral databases of different organs (e.g., esophagus, gastric, colon, cervix, bladder, lung, nasopharynx, larynx, and the oral cavity (hard palate, soft palate, buccal, inner lip, ventral and the tongue)). Thus, the disclosed Raman platform is a universal diagnostic tool for cancer diagnostics at endoscopy.
Referring to
For other applications, the spectral acquisition framework could also be used for external or internal surgical interventions or to assess tissue types during surgery. The real-time capability allows on-the-spot diagnosis and could therefore be used to guide excisional margins for tumor resection. It is critical that the diagnostic information can be given online (i.e., <0.5 sec) to aid in medical decision-making. For skin measurements, there are less stringent demands to the measurement time because skin spectra are acquired under more controllable experimental settings with possibility of longer exposure times. The online software architecture can also apply to other areas that require fast spectral measurements including fluorescence, reflectance spectroscopy or in different fields such as process analytical technology, food sciences, forensics, etc., whereby uninterrupted real-time screening is needed. In step 150, the spectral acquisition module 120 determines if the signal is saturated. If so, it initiates a new data acquisition with reduced integration time to prevent saturation. In step 155, the spectra acquisition module 120 identifies and eliminates cosmic rays (i.e., using the first derivative of the spectra with a 95% confidence interval (CI) over the whole spectral range set as a maximum threshold). In one embodiment identified cosmic rays are removed by linear interpolation. Cosmic rays can be removed by other methods, including multivariate analysis, smoothing, mean filtering, median filtering, Fourier transform, wavelets, etc.
In step 160, the spectral preprocessing module 125 scales the acquired spectra with integration time and laser power. A first-order Savitzky-Golay smoothing filter is further used to remove the noise in the intensity corrected spectra. A 5th order modified polynomial constrained to the lower bound of the smoothed spectra is then subtracted to resolve the tissue Raman spectrum alone. The Raman spectrum is finally normalized to the integrated area under the curve from 800 to 1800 cm−1 to resolve the spectral line shapes and relative intensities, reducing probe handling variations at clinical endoscopy. The GUI (
In step 165, the outlier analysis module 130 detects outlier spectra using principal component (PCA) coupled with Hotelling's T2 and Q-residual statistics. The GUI (
If the spectra were verified for further analysis, they are fed to probabilistic models for in vivo cancer diagnostics. In step 170 the multivariate analysis module 135 applies probabilistic models for in vivo cancer diagnostics. The multivariate analysis module 135 switches among different pre-rendered models including partial least squares-discriminant analysis (PLS-DA), PCA-linear discriminant analysis (LDA), ant colony optimization (ACO)-LDA, classification and regression trees (CART), support vector machine (SVM), adaptive boosting (AdaBoost) etc. based on a spectral databases of large number of patients.
In step 175, the pathology module 140 implements organ specific diagnostic models that can switch among the spectral databases of different organs for probabilistic cancer diagnostics. In addition to the audio feedback, the GUI (
The database 142 stores acquired spectra as well as the stored spectra used for diagnosis.
In some embodiments multiple spectra are taken and analyzed. For example between 5-15 are taken. Each is analyzed and if more than a threshold percentage provides the same outcome—cancer vs normal—that is the determined diagnosis. For example, if 10 spectra are taken and 7 or more provide the same answer, that is the diagnosis. If only 5 or 6 provide the same answer, the process is repeated.
An integrated Raman spectroscopy and trimodal wide-field imaging system used for real-time diagnostics comprises a spectrum stabilized 785 nm diode laser (maximum output: 300 mW, B&W TEK Inc., Newark, Del., USA) electronically synchronized with a USB 6501 digital I/O (National Instruments, Austin, Tex., USA), a transmissive imaging spectrograph (Holospec f/1.8, Kaiser Optical Systems, Ann Arbor, Mich., USA) equipped with a liquid nitrogen-cooled, NIR-optimized, back-illuminated and deep depletion charge-coupled device (CCD) camera (1340 400 pixels at 20×20 per pixel; Spec-10: 400BR/LN, Princeton Instruments, Trenton, N.J., USA), and a specially designed Raman endoscopic probe for both laser light delivery and in vivo tissue Raman signal collection. The 1.8 mm Raman endoscopic probe is composed of 32 collection fibers surrounding the central light delivery fiber with two stages of optical filtering incorporated at the proximal and distal ends of the probe for maximizing the collection of tissue Raman signals, while reducing the interference of Rayleigh scattered light, fiber fluorescence and silica Raman signals. The Raman probe can easily pass down to the instrument channel of medical endoscopes and be directed to suspicious tissue sites under the guidance of wide-field endoscopic imaging (WLR/AFI/NBI) modalities. The system acquires Raman spectra in the wavenumber range of 800-1800 cm−1 from in vivo upper GI tissue within 0.5 s using the 785 nm excitation power of 1.5 W/cm2 (spot size of 200 μm) with a spectral resolution of ˜9 cm−1.
Hardware components of the Raman system (e.g., laser power control, spectrometer, CCD shutter and camera readout synchronization) were interfaced to the Matlab software through libraries for different spectrometers/cameras (e.g., PVCAM library (Princeton Instruments, Roper Scientific, Inc., Trenton, N.J., USA) and Omni Driver (Ocean Optics Inc., Dunedin, Fla., USA), etc.). The laser was electronically synchronized with the CCD shutter. The automatic adjustment of laser power, exposure time and accumulation of spectra were realized by scaling to within 85% of the total photon counts (e.g., 55,250 of 65,000 photons) based on preceding tissue Raman measurements, whereas an upper limit of 0.5 s was set to realize clinically acceptable conditions. The accumulation of multiple spectra and automatic adjustment of exposure time provides a rapid and straightforward methodology to prevent CCD saturation and to obtain high signal to noise ratio (SNR) for endoscopic applications. The Raman-shift axis (wavelength) was calibrated using a mercury/argon calibration lamp (Ocean Optics Inc., Dunedin, Fla., USA). The spectral response correction for the wavelength-dependence of the system was conducted using a standard lamp (RS-10, EG&G Gamma Scientific, San Diego, Calif., USA). The reproducibility of the platform is continuously monitored with the laser frequency and Raman spectra of cyclohexane and acetaminophen as wavenumber standards. All the system performance measures including CCD temperature, integration time, laser power, CCD alignment are accordingly logged into a central database via SQL server.
Real-time preprocessing of Raman signals was realized with the rapid detection of cosmic rays using the first derivative with a 95% confidence interval (CI) over the whole spectral range set as a maximum threshold. Data points lying outside of a threshold were interpolated to 2nd order. The spectra were further scaled with integration time and laser power. A first order, 5 point Savitzky-Golay smoothing filter was used to remove noise in the intensity corrected spectra, while a 5th order modified polynomial constrained to the lower bound of the smoothed spectra was subtracted to resolve the tissue Raman spectrum alone. The Raman spectrum was normalized to the integrated area under the curve from 800 to 1800 cm−1, enabling a better comparison of the spectral shapes and relative Raman band intensities among different tissue pathologies. The spectra were then locally mean-centered according to the specific database to remove common variations in the data. Following preprocessing, the Raman spectra were fed to a model-specific outlier analysis.
An outlier detection scheme was incorporated into biomedical spectroscopy as a high-level model-specific feedback tool in the on-line framework by using PCA coupled with Hotelling's T2 and Q-residual statistics. PCA reduces the dimension of the Raman spectra by decomposing them into linear combinations of orthogonal components (principal components (PCs)), such that the spectral variations in the dataset are maximized. The PCA model of the data matrix X is defined by:
X=TPT+E
where T and P represent scores and loadings, and E contains the residuals. The loadings correspond to the new rotated axis, whereas scores represent the data projection values. Accordingly, Hotelling's T2 statistics is a measure of variance captured by the PCA model (sample to model distance) and is defined by:
Tik2=tik(λk−1)tikT
where tik is PC scores for ith sample spectrum using component k, and λk−1λk−1 is the diagonal matrix of normalized eigenvalues of the covariance matrix for component k. Therefore, Hotelling's T2 gives an indication of extreme values within the PCA model. On the other hand, Q-residuals is a measure of variance which is not captured by the PCA model (lack of model fit statistics) and is defined by
Qik=Σ(xi−tikPkT)2
where xi is the sample spectrum, Qik is the sum of squared reconstruction error for ith sample spectrum using component k and Pk is the PC loadings. For both Hotelling's T2 and Q-residuals, the normalized 99% CI was utilized as upper thresholds to intercept anomalous Raman spectra. Accordingly, the Hotelling's T2 and Q-residuals are two independent parameters providing quantitative information about the model fit. Using these parameters as indicators of spectra quality (i.e., probe contact mode, confounding factors, white light interference etc.), auditory feedback has been integrated into the online Raman diagnostic system, facilitating real-time probe handling advice and spectroscopic screening for clinicians during clinical endoscopic procedures.
Subsequent to verification of tissue Raman spectra quality, those qualified Raman spectra were immediately fed to probabilistic models for on-line in vivo diagnostics and pathology prediction. The GUI can instantly switch among different models including partial least squares-discriminant analysis (PLS-DA), PCA-linear discriminant analysis (LDA), ant colony optimization (ACO)-LDA, classification and regression trees (CART), support vector machine (SVM), adaptive boosting (AdaBoost) etc. for prospective classification at clinical endoscopic procedures. As an example, probabilistic PLS-DA was employed for gastric cancer diagnosis. PLS-DA employs the fundamental principle of PCA but further rotates the components by maximizing the covariance between the spectral variation and group affinity to obtain the diagnostically relevant variations rather than the most prominent variations in the spectral dataset. The system supports binary classification, one-against-all and one-against-one multiclass (i.e., benign, dysplasia and cancer) probabilistic PLS-DA discriminatory analysis to predict the specific tissue pathologies.
A total of 2748 in vivo gastric tissue spectra (2465 normal and 283 cancer) were acquired from the 305 patients recruited to construct the spectral database for developing diagnostic algorithms for gastric cancer diagnostics. Tissue histopathology serves as the gold standard for evaluation of the performance of Raman technique for in vivo tissue diagnosis and characterization.
The stomach represents one of the most challenging organs presenting with many confounding factors (i.e., gastric juice, food debris, bleeding, exudates etc.) for spectroscopic diagnosis. The in vivo mean Raman spectra acquired from 305 gastric patients (normal (n 2465) and cancer (n=283)) for algorithms development are shown in
The automatic outlier detection was realized for predictive on-line analysis using PCA with Hotelling's T2 and Q-residuals statistics (99% CI). To make the online diagnostics efficient, a two-component PCA model was rendered that included the largest tissue spectral variations. These selected significant PCs (p<0.0001) accounted for maximum variance of 38.71% (PC1: 30.33%, PC2: 8.38%) of the total variability in the dataset (n=2748 Raman spectra), and the corresponding PC loadings are shown in
The prospectively acquired spectra verified by the on-line outlier analysis are further fed to probabilistic PLS-DA for instant disease prediction, achieving a diagnostic accuracy of 80.0% (60/75) for gastric cancer detection (
The receiver operating characteristic (ROC) curves were further generated to evaluate the group separations.
The total processing time for all the aforementioned on-line data acquisition to tissue pathological prediction was 0.13 s. The processing time for each step of the flow chart in
The Raman spectroscopy system comprises a spectrum stabilized 785 nm diode laser (maximum output: 300 mW, B&W TEK Inc., Newark, Del.), a transmissive imaging spectrograph (Holospec f/1.8, Kaiser Optical Systems Inc., Ann Arbor, Mich.) equipped with a liquid nitrogen-cooled, back-illuminated and deep depletion CCD camera (1340×400 pixels at 20×20 μm per pixel; Spec-10: 400BR/LN, Princeton Instruments, Trenton, N.J.). The system also consists of a specially designed fused-silica fiber-optic Raman endoscopic probe (1.8 mm in outer diameter and 1.30 meters in length) that comprises 9×200 μm collection fibers (N.A.=0.22) surrounding the central light delivery fiber (200 μm in diameter, N.A.=0.22). A 1.0 mm sapphire ball lens (refractive index 1.76) is coupled to the fiber tip of the Raman probe for enhancing epithelial tissue Raman measurements. The system acquires Raman spectra over the range of 800-1800 cm-1 with spectral resolution of 9 cm-1. Each Raman spectrum in this study was measured with an integration time of 0.5 s under the 785 nm laser excitation. The rapid Raman spectroscopy technique was wavelength calibrated using an argon/mercury spectral lamp (AR-1 and HG-1, Ocean Optics Inc., Dunedin, Fla.). All wavelength-calibrated spectra were corrected for the intensity response of the system using a tungsten-halogen calibration lamp (RS-10, EG&G Gamma Scientific, San Diego, Calif.).
Using the system and method described in
The PLS regression was employed as a multivariate method to extract characteristic internal reference background signals from the fiber-optic Raman probe. Briefly, PLS utilizes the fundamental principle of PCA but further rotates the components LVs by maximizing the covariance between the spectral variation and the dependent variable (e.g., laser excitation power), so that the LV loadings explain the relevant variations rather than the most prominent variations in the spectral dataset. Important spectral reference signals related to the laser excitation power were retained in the first few LVs. In this study, mean-centering was performed before modeling to reduce the complexity of the PLS regression model. The optimal complexity of the PLS regression model was determined through leave-one subject-out, cross-validation, and the performance of the PLS regression model was examined by calculating the coefficient of determination (R2), root mean square error of calibration (RMSEC), root mean square error of cross validation (RMSECV) and root mean square error of prediction (RMSEP). Note that an optimal PLS model has a high R2 but with a low RMSEC, RMSECV and RMSEP. The PLS regression model developed for resolving the reference signals in this study was also implemented as an on-line laser excitation power predictor in our real-time clinical Raman software and tested prospectively in an unbiased manner. Multivariate statistical analysis was conducted in the Matlab (Mathworks Inc., Natick, Mass.) programming environment.
A total of 30 normal healthy subjects (16 female and 14 males) were recruited for in vivo tissue Raman measurements in the oral cavity. Prior to in vivo tissue Raman spectroscopy measurements, all subjects underwent extensive mouthwash to reduce confounding factors (e.g. food debris, microbial coatings etc.). In vivo tissue Raman spectra (n=783) were collected of the inner lip of 25 subjects. For the 25 subjects, in vivo oral tissue Raman spectra (n=˜5) were acquired at six power levels in the range of 5-65 mW (intervals of ˜10 mW). Before each tissue Raman measurement, the laser excitation power level was measured at the distal tip of the fiber-optic probe using a power meter with a linearity of ±0.5% and accuracy of ±3% (range of 0.1 to 100 mW). Other confounding factors (e.g., probe pressure on the tissue surface, photobleaching, tissue optical properties and bending of the fiber optic probe) were not monitored purposely but incorporated into the PLS modeling for the robust extraction of reference signals in situ. After deployment of the developed PLS model in the on-line Raman acquisition framework, the prospective and independent validation of the internal reference signal for laser excitation power monitoring was performed on the 5 new subjects (n=166 spectra) in real-time.
To further validate the quantitative value of the internal reference method developed in this work, we also conducted a tissue phantom experiment. Tissue phantoms of various gelatin concentrations were prepared from bovine skin, Type-B gelatin (G9391, Sigma, USA). The gelatin was dissolved in predefined concentrations (20, 25, 30, 35, 40, 45, and 50% by weight) in distilled H2O. The dissolved gelatin was heated to 50° C. for 1 hour in a water bath with continuous stirring. Subsequently, the molten gelatin was poured into a pre-chilled mold (4° C.) and stored for 2-3 hours to produce solid gelatin phantoms. Quantitative fiber-optic Raman spectroscopic analysis of the tissue phantoms was then performed. A total of n=133 Raman spectra were measured from the various tissue phantoms using the fiber-optic Raman probe with different laser powers. The laser excitation powers were changed in the range 10-60 mW and the measured spectra were normalized to laser excitation powers as predicted by the internal reference method.
To develop the PLS regression model and resolve internal reference signals, we measured in vivo Raman spectra of 25 subjects in the oral cavity with the laser excitation power as an independent parameter. For each subject, in vivo tissue Raman spectra (n=˜5) were acquired with different power levels in the range of 5-65 mW (intervals of ˜10 mW).
A PLS regression model to extract a broad range of characteristic internal reference peaks from the oral tissue Raman spectra. The Rayleigh scattered light was excluded from PLS analysis. The measured in vivo raw tissue Raman spectra were arranged in a matrix with row-wise spectra and column-wise wavenumbers. The reference laser power levels were arranged in a column vector representing the dependent variables. After mean-centering, a PLS regression model was developed using the leave-one subject-out, cross-validation in order to establish the optimum algorithm for rendering robust reference signals for laser excitation power prediction.
The quantitative value of the internal-reference method developed for quantitative spectral analysis of tissue phantoms. Seven tissue phantoms composed of gelatin with different concentrations (i.e., 20, 25, 30, 35, 40, 45, and 50% by weight) were constructed and tested. Raman spectra (n=133 spectra) from gelatin phantoms were measured and normalized to the laser powers predicted in real-time.
This study demonstrates the feasibility of Raman spectroscopy in transnasal endoscopic applications, providing the foundation for large-scale clinical studies in the head and neck. The image-guided Raman endoscopy platform integrated with a miniaturized fiber Raman probe developed provides a rapid and minimally invasive assessment of endogenous tissue constituents of the head and neck at the molecular level during clinical endoscopic examination. This greatly facilitates clinicians to obtain detailed biomolecular fingerprints of tissue in the head and neck, reflecting the genuine compositional and morphological signatures without introducing the artifacts caused by vascular puncturing or tissue dehydration, morphological and anatomical effects, etc.
The Raman spectroscopy system consists of a spectrum stabilized 785 nm diode laser (maximum output: 300 mW, B&W TEK Inc., Newark, Del.), a transmissive imaging spectrograph (Holospec f/1.8, Kaiser Optical Systems) equipped with a cryogenic cooled (−120° C.), NIR-optimized, back-illuminated and deep depletion charge-coupled device (CCD) camera (1340×400 pixels at 20×20 μm per pixel; Spec-10: 400BR/LN, Princeton Instruments). The novel spectrometer fiber input coupling consists of parabolic aligned array of 58 fibers (100 μm) to correct the spectrometer image aberration for improving both the spectral resolution and signal-to-noise ratio of Raman signals. A 1.8 mm fiber-optic Raman probe for transnasal endoscopic applications maximizing both the tissue excitation and in vivo tissue Raman collections was utilized. The Raman fiber probe fits into the instrument channel of flexible transnasal endoscopes and can be safely directed to different locations in the nasopharynx and larynx under the wide field imaging (i.e., white-light reflectance (WLR) and narrowband imaging (NBI)) guidance. The clinical Raman endoscopy platform has been integrated with our recently developed on-line data processing software to facilitate probe handling-advise and sound feedback to clinicians in real-time (processing time<0.1 s). Briefly, the on-line Raman endoscopy framework synchronizes spectral acquisition (i.e., laser exposure, integration time, CCD shutter and readout etc.) and automatically extracts the Raman signals from the raw tissue spectra (comprising strong autofluorescence background and weak Raman signals) using the established preprocessing methods including smoothing, fifth-order polynomial baseline subtraction etc. The in vivo Raman spectra and the outcome of multivariate algorithms (e.g., principal component analysis) can be displayed in real-time in a comprehensible graphical user interface (GUI) during clinical transnasal Raman endoscopy.
A total of 23 normal healthy male subjects of different races (twenty-two Asian and one Caucasian) were recruited for in vivo tissue Raman measurements at transnasal endoscopy. In these subjects recruited, no suspicious lesions were identified under the WLR and NB imaging examination. A total of three primary measurement sites of assumed normal (or benign) tissues were predefined for in vivo Raman acquisitions, including the true laryngeal vocal cords (LVC), the posterior nasopharynx (PN), and also the pharyngeal recess (i.e., fossa of Rosenmüller (FOR)) where NPC typically initiates. The fiber-optic Raman probe can be placed in gentle contact with internal tissues interrogating with the endogenous biomolecular compositions of tissue in real-time. The accurate positioning against the biopsied tissue sites was verified on the WLR/NBI monitor by the endoscopists in-charge. The probe allowed Raman spectra to be collected from an area (200 μm in diameter) with probing volume of approximately 1 mm3 and penetration depth of ˜800 μm. Each spectrum was acquired within 0.5 s using the 785 nm laser light with the power of ˜50 mW on the tissue surface.
The Raman spectra were displayed on-line and were stored for post-procedural inspection. This rapid Raman endoscopic technology is non-destructive, and can now routinely be used under endoscopic transnasal examinations for clinical evaluation. To assess the intra-tissue site variance, several Raman spectra (˜18) were also acquired from each tissue site. As a result, a total of 874 in vivo Raman spectra from 47 sites were measured at transnasal endoscopy and used for spectral analysis [PN (n=521), FOR (n=157) and LVC (n=196)] from the 23 subjects.
Prior to data-analysis, the raw Raman spectra were firstly smoothed using a linear Savitzky Golay filter, and tissue autofluorescence background was then subtracted from the smoothed spectra using a 5th order polynomial fit. The background-subtracted Raman spectra were normalized to the integrated areas under the curves to minimize the effect of Raman probe handling variations on clinical Raman measurements with respect to different subjects and tissue sites. All processed Raman spectra were assembled into a matrix, and the mean centering of the entire Raman dataset was then performed. To reduce the dimension of the spectral data, principal component analysis (PCA) was employed to extract a set of orthogonal principal components (PCs) that account for the maximum variance in the Raman spectral dataset for tissue characterization. Accordingly, loadings on the PCs represent orthogonal basis spectra of the most prominent spectral variation in the dataset accounting for progressively decreasing variance, whereas the scores on the PCs represent the projection value of the tissue Raman spectra on the corresponding loading. Thus, PCA can efficiently be used to resolve spectral variations while reducing the dimension of the dataset to a minimum. The number of retained PCs was chosen based on the analysis of variance (ANOVA) and Student's t-test at 0.05 level. We employed post-hoc Fisher's least squares differences (LSD) test to assess differences in means. Multivariate statistical analysis was performed using the PLS toolbox (Eigenvector Research, Wenatchee, Wash.) in the Matlab (Mathworks Inc., Natick, Mass.) programming environment.
High quality in vivo Raman spectra can routinely be acquired in the nasopharynx and larynx in real-time during transnasal image-guided (i.e., WLR and NBI) endoscopic inspections.
In vitro Raman spectra of blood, saliva and nasal mucus obtained from healthy volunteers were measured as shown in
Fiber-optic confocal Raman diagnostics can be achieved in real-time (<0.5 second) and uncovers the progressive biomolecular and functional changes of epithelial cells and tissues in Barrett's carcinogenesis in situ. Histopathology characterized 152 of the prospectively measured tissue sites as columnar lined epithelium (n=597 spectra), 48 as intestinal metaplasia (n=123 spectra), 9 high-grade dysplasia (n=77 spectra). Using receiver operating characteristics (ROC) analysis, identification of high-grade dysplasia could be successfully achieved yielding a sensitivity of 87.0%, and a specificity of 84.7% on spectrum basis. The area under the ROC curve was found to be 0.90. This new biomolecular specific endoscopic modality with real-time capability offers the gastroenterologist a reliable tool to objectively target high-risk tissue areas in Barrett's patients during ongoing endoscopy.
The confocal Raman spectroscopic system comprises of a near-infrared (NIR) diode laser (λex=785 nm), a high-throughput transmissive imaging spectrograph equipped with a liquid nitrogen-cooled, NIR-optimized charge-coupled device (CCD) camera and a specially designed 1.8-mm fiber-optic confocal Raman probe. The system acquires Raman spectra in the range 800-1800 cm−1 with a spectral resolution of ˜9 cm−1. The developed fiber-optic confocal Raman endoscopic probe is used for both laser light delivery and in vivo tissue Raman signal collection.
The 1.8 mm (in outer diameter) confocal Raman endoscopic probe comprises 9×200 μm filter-coated collection fibers (NA=0.22) surrounding the central light delivery fiber (200 μm in diameter, NA=0.22). A miniature 1.0 mm sapphire ball lens (NA=1.78) is coupled to the fiber tip of the confocal probe to tightly focus the excitation light onto tissue, enabling the effective Raman spectrum collection from the epithelial lining (<200 μm). The fiber-optic confocal Raman probe can be inserted into the instrument channel of conventional endoscopes and placed in gentle contact with the epithelium for in vivo tissue characterization and diagnosis. The depth-selectivity of this confocal Raman probe offers compelling experimental advantages, including (i) fiber-optic confocal Raman spectroscopy selectively targets the epithelial lining associated with early onset of Barrett's carcinogenesis, which is superior to conventional volume-type fiber-optic Raman probes that interrogate a larger tissue volume; (ii) the shallower tissue interrogation ability of confocal Raman technique provides a higher tissue Raman to autofluorescence background ratio due to a much reduced tissue autofluorescence contribution from deeper tissue layers (e.g., stroma), and (iii) combining this novel fiber-optic confocal Raman spectroscopy platform with well-documented multivariate analysis enables epithelial molecular information to be extracted and analyzed in real-time in vivo. The entire confocal Raman endoscopic system is controlled in an intuitive software framework that permits rapid survey in endoscopic screening settings with auditory probabilistic feedback to the endoscopist, pushing the frontier of Raman spectroscopy into routine clinical diagnostics.
A total of 450 patients have been enrolled in the Raman endoscopic examinations for surveillance or screening of various indications, including dyspepsia and upper GI neoplasia. During a typical examination of suspicious lesions, each tissue Raman measurement can be acquired within 0.5 second, which permits rapid survey of large tissue areas. The in vivo Raman spectral data acquired from 373 patients with different histological subtypes in the upper GI have been used to construct a comprehensive Raman library (>12,000 Raman spectra). For the patients recruited for screening and surveillance of BE, Raman spectra are categorized into following three histopathologically risk classes: (i) “Normal”—columnar lined epithelium (CLE), (ii) “Low-risk” BE-defined as the presence of goblet cells, (iii) “High-risk”-low-grade dysplasia (LGD) and high-grade dysplasia (HGD). For example,
Histopathology characterized 152 of the prospectively (i.e., independently) measured tissue sites as CLE (n=597 spectra), 48 as IM (n=123 spectra) and 9 as HGD (n=77 spectra).
In the above description, an embodiment is an, example or implementation of the disclosed system and methods. The various appearances of “one embodiment”, “an embodiment” or “some embodiments” do not necessarily all refer to the same embodiments.
Although various features of the disclosed system and methods may be described in the context of a single embodiment, the features may also be provided separately or in any suitable combination. Conversely, although the disclosed system and methods may be described herein in the context of separate embodiments for clarity, the disclosed system and methods may also be implemented in a single embodiment.
Furthermore, it is to be understood that the disclosed system and methods can be carried out or practiced in various ways and can be implemented in embodiments other than the ones outlined in the description above.
Meanings of technical and scientific terms used herein are to be commonly understood as by one of ordinary skill in the art belong, unless otherwise defined.
Certain aspects of the present disclosure include process steps and instructions described herein in the form of a method. It should be noted that the process steps and instructions of the present disclosure could be embodied in software, firmware or hardware, and when embodied in software, could be downloaded to reside on and be operated from different platforms used by real time network operating systems.
The present disclosure also relates to an apparatus for performing the operations herein. This apparatus may be specially constructed for the required purposes, or it may comprise a general-purpose computer selectively activated or reconfigured by a computer program stored on a computer readable medium that can be accessed by the computer. Such a computer program may be stored in a tangible non-transitory computer readable storage medium, such as, but is not limited to, any type of disk including floppy disks, optical disks, CD-ROMs, magnetic-optical disks, read-only memories (ROMs), random access memories (RAMs), EPROMs, EEPROMs, magnetic or optical cards, application specific integrated circuits (ASICs), or any type of media suitable for storing electronic instructions, and each coupled to a computer system bus. Furthermore, the computers referred to in the specification may include a single processor or may be architectures employing multiple processor designs for increased computing capability.
The methods and operations presented herein are not inherently related to any particular computer or other apparatus. Various general-purpose systems may also be used with programs in accordance with the teachings herein, or it may prove convenient to construct more specialized apparatus to perform the required method steps. The required structure for a variety of these systems will be apparent to those of skill in the art, along with equivalent variations. In addition, the present disclosure is not described with reference to any particular programming language. It is appreciated that a variety of programming languages may be used to implement the teachings of the present disclosure as described herein, and any references to specific languages are provided for disclosure of enablement and best mode of the present disclosure.
The present disclosure is well suited to a wide variety of computer network systems over numerous topologies. Within this field, the configuration and management of large networks comprise storage devices and computers that are communicatively coupled to dissimilar computers and storage devices over a network, such as the Internet, public networks, private networks, or other networks enabling communication between computing systems. Finally, it should be noted that the language used in the specification has been principally selected for readability and instructional purposes, and may not have been selected to delineate or circumscribe the inventive subject matter. Accordingly, the disclosure of the present disclosure is intended to be illustrative, but not limiting, of the scope of the disclosure, which is set forth in the following claims.
Number | Date | Country | Kind |
---|---|---|---|
1307338.2 | Apr 2013 | GB | national |
This application claims priority to the following patent applications: (1) Patent Cooperation Treaty Application PCT/SG2013/000273 filed Jul. 2, 2013; (2) U.S. Patent Application 61/667,384, filed Jul. 2, 2012; and a (3) Great Britain Patent Application 1307338.2, filed Apr. 23, 2013, each of the above cited applications is hereby incorporated by reference herein as if fully set forth in its entirety.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/SG2013/000273 | 7/2/2013 | WO | 00 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2014/007759 | 1/9/2014 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
5850623 | Carman, Jr. et al. | Dec 1998 | A |
5976885 | Cohenford et al. | Nov 1999 | A |
6351306 | Tedesco | Feb 2002 | B1 |
6621574 | Forney et al. | Sep 2003 | B1 |
20030191398 | Motz et al. | Oct 2003 | A1 |
20130162989 | Chen | Jun 2013 | A1 |
Number | Date | Country |
---|---|---|
WO2011072380 | Jun 2011 | WO |
Entry |
---|
International Publication and International Search Report for PCT/SG2013/000273. |
Written Opinion, Singapore Patent Application No. 11201408793X, dated Sep. 6, 2016. |
Second Office Action, SIPO Patent Application No. 201380035488.5, dated Mar. 20, 2017 (original Chinese and English translation). |
Number | Date | Country | |
---|---|---|---|
20150335248 A1 | Nov 2015 | US |
Number | Date | Country | |
---|---|---|---|
61667384 | Jul 2012 | US |