An area of increasing development in the field of gas chromatography (GC) is instrument miniaturization. Notable examples of such advances include portable field GC units and GC separations achieved on a micro-analytical chip. In conjunction with these efforts, there is also a growing interest in developing sensitive miniaturized detection methods that can be incorporated into micro-analytical devices. A number of such miniaturized or ‘micro’ detection methods have been reported based on a variety of principals including surface acoustic wave transmission, thermal conductivity, and plasma-based optical emission. Although flame-based detectors are prevalent in many conventional GC applications, relatively few have been adapted to micro-analytical formats. Since the latter tend to utilize very small (nL range) channels, this may be partly attributed to difficulties encountered in operating a stable flame within these dimensions. In this regard, however, a very interesting and useful system has been successfully demonstrated. The method employs low gas flows to support a high energy premixed flame (about 3 mm tall×1 mm wide) that can perform atomic emission/hydrocarbon ionization detection on the surface of a micro-analytical chip.
The flame photometric detector (FPD) is a widely used GC sensor for determining sulfur, phosphorus, tin, and other elements in volatile organic compounds based on their chemiluminescence within a low-temperature, hydrogen-rich flame. Very recently, we introduced a novel method of generating a similar flame environment using counter-flowing streams of gas [K. B. Thurbide, B. W. Cooke, W. A. Aue, J. Chromatogr. 1029 (2004) 193.]. This ‘counter-current’ FPD was demonstrated to provide similar sensitivity and response characteristics to that of a conventional FPD when operated in the hydrogen-rich mode. As well, it was also found to yield useful flame ionization detector (FID) signals when operated in the air-rich mode. Most notably, unlike a conventional FPD, this method produced remarkably stable flames at relatively low and high gas flows of varying stoichiometry. In fact, this aspect of the detector was employed in the primary focus of the study, which explored changes in transition metal response as a function of flame size derived from gas flows that differed by several hundred mL/min.
Subsequent to this work (but actually reported earlier) we exploited the great stability of counter-current flames in a new way by using them to create an enclosed hydrogen-rich micro-flame [K. B. Thurbide, C. D. Anderson, Analyst 128 (2003) 616]. The flame was supported on a fused silica capillary by only a few mL/min of gas flow and encompassed a very small volume of 30 nL. As well, it produced qualitatively similar response characteristics toward sulfur and phosphorus-containing analytes as that of a conventional FPD. The method was employed in a novel micro-Flame Photometric Detector (μFPD) which was operated either inside the end of a capillary gas chromatography column (on-column) or within a length of capillary quartz tubing after the separation column (post-column), with each mode displaying similar characteristics.
In general, the dimensions and qualities of the micro counter-current flame indicated that it could be a potentially useful method of producing chemiluminescent molecular emission, similar to a conventional FPD, within small channels and analytical devices of reduced proportions. However, unlike the larger counter-current flame, the primary disadvantage to the micro-flame method was the relatively large detection limits that it produced for sulfur and phosphorus due to an elevated background emission. The spectrum, intensity, and orange appearance of the emission indicated that the fused silica capillary burner was glowing from contact with the flame. Despite efforts to prevent this it was observed under all conditions investigated.
Therefore there is disclosed a μFPD device with enhanced response by removing interference from an elevated background emission. A μFPD flame detector is provided with similar performance to a conventional FPD flame, even though the two differ in size by about 3 orders of magnitude. Further, a flame detector is provided with photometric tin response and flame ionization response.
In accordance with a further aspect of this invention, there is provided an improved μFPD response that is obtained by using a metallic capillary burner to support a micro counter-current flame, as for example a stainless steel capillary burner. The μFPD has satisfactory response for many elements such as sulfur, phosphorus, and tin. Additionally, by polarizing the burner, the micro counter-current flame detector produces a satisfactory ionization response toward carbon. The μFPD as discussed herein is convenient for use in chemical weapons detection, sulfur measurements in, for example, oil and gas or pulp and paper, measuring amounts of H2S or SO2 in the environment, analyzing pesticides containing sulfur, phosphorus, and other elements, performing general gas analysis for hydrocarbons present, or detecting other elements such as transition metals and main group elements such as selenium, tin, lead, tellurium, and halogens such as chlorine or bromine.
These and other aspects of the invention are described in the detailed description and claimed in the claims.
There will now be described preferred embodiments of the invention by reference to the figures, by way of illustration only, in which:
In this patent document, the word “comprising” does not exclude other elements being present and the use of the indefinite article “a” before an element does not exclude others of the same element being present. For the purposes of this patent document, including the claims, a flame photometric detector is considered to be a micro-flame photometric detector, or μFPD, if the flame volume is less 1 μL (1×10−6 L), which for example is satisfied when the flame dimensions are less than 0.1 mm×0.1 mm×0.1 mm.
A quartz capillary sleeve 33 (0.9 mm i.d.) extends vertically from bottom port 12 through to top port 36. In the lower port 12, the capillary sleeve 33 surrounds the hydrogen sleeve 22 and a capillary GC column 20. Above the lower port 12, in the flame region 40, the capillary sleeve 33 conducts the hydrogen and column effluent (analyte plus carrier) from capillary sleeve 22 towards the flame 42. Through a septum 34 in the top port 36, a length of stainless steel capillary tubing 38 (0.01″ i.d.×0.018″ o.d.) carrying oxygen extends downward into the quartz sleeve 33 to the center 40 of the union 10, directly in front of both the light guide port 26 and the viewing port 24. Under typical operating conditions, the micro-flame 40 is situated on the end of this oxygen capillary 38 burning ‘upside down’ within a counter flowing stream of hydrogen and column effluent from the bottom. The arrangement for delivering hydrogen and analyte may be varied considerably from what is described here. A tube in tube arrangement with hydrogen in the annulus between the tubes may be used as described here. Also, hydrogen may be supplied through a capillary column 20 along with the analyte. Other arrangements will occur to a person skilled in the art.
The separation column 20 employed is an EC-5 ((5% Phenyl)-95% Methylpolysiloxane) megabore column (30 mm×0.53 mm i.d.; 1.00 μm thickness; Alltech, Deerfield, Ill., U.S.A.) that extends vertically upward from the GC instrument and into the detector housing 10 through the connecting stainless steel tube 22 carrying the hydrogen. Typical separations employ 5 mL/min of helium as the carrier gas. Normally, about 2-3 mm separates the end of the column 20 from the oxygen burner 38. For μFID experiments, electrical leads from a Shimadzu GC are used such that the polarizer 44 of the GC is connected to the stainless steel oxygen burner 38 and the collector 46 is connected to the stainless steel hydrogen tube 22 surrounding the separation column 20.
High purity helium, hydrogen, and oxygen may be obtained from any suitable source such as Praxair. Tetrahydrothiophene (99%), trimethyl phosphite (99%), benzene (99%), decane (99%), and tetramethyl tin (95%) are obtained from any suitable source, such as Aldrich.
Stainless steel is an improvement over fused silica because it has a higher heat capacity. As such, the heat of the flame does not cause it to glow from being incandecently heated. Glowing creates a large background response in the detector, which decreases its sensitivity. The improvement offered by stainless steel include improved detection limits and the simultaneous FID method and allow the method to be useful in more situations. In addition to a stainless steel capillary, the flame could also be supported on other metals that have a sufficiently high melting point, such as nickel, or some alloys. Typical flame volume for the stainless steel example given here was about 30 nL.
The flame 40 is lit by introducing hydrogen, and igniting the flame as a diffusion flame at the top of the chimney. The oxygen containing capillary 38 is then drawn through the flame, ignites, and is pushed into the hydrogen stream, keeping it lit. The original flame either extinguishes or can be blown out like a candle. Once lit, the flame generally stays stable for hours.
Use of pure oxygen is preferred as a supply of oxygen. Experiments with air found that a flame was difficult to establish and prone to extinguish depending on the flame conditions. Therefore, the influence of gas flows on the effective operating region was explored.
Upon using oxygen in the upper burner a remarkable difference in flame dynamics was observed as the flame rarely extinguished, if at all, once inside of the capillary column 33.
Burner Characteristics
Stainless steel capillary tubing of both 0.01″ i.d. and 0.005″ i.d. was investigated for its properties as a μFPD burner 38. Respectively, these dimensions are the same as and smaller than the fused silica tubing i.d. used previously. It was found that both tubing sizes were able to support a stable flame. However, the 0.005″ i.d. (0.009″ o.d.) tubing was observed to glow considerably, yielding a similar background emission to that noted earlier for the fused silica burner. In terms of relative wall thickness, this capillary burner (0.002″) was slightly smaller compared to the fused silica tubing (0.003″) used originally.
In contrast to this, when trials were run using the 0.01″ i.d. (0.018″ o.d.) tubing as a burner 38, the orange glow was observed to disappear and the background emission was much less intense compared to that obtained with fused silica. This tubing has a wall thickness of 0.004″. It therefore seems advantageous to have a wall thickness greater than 0.002″ (0.05 mm) under typical conditions in order to avoid any glowing of the stainless steel burner. In routine comparisons with fused silica burners, it was found that the thicker walled stainless steel capillary tubing readily reduced the background emission observed by over an order of magnitude. Also in general, the size, stability, chemiluminescent properties, and gas flow operating regions of the flame itself did not differ between stainless steel and fused silica burners of 0.01″ i.d. under the same conditions. Therefore, this stainless steel capillary tubing provides a more effective burner for the μFPD and was used in experiments described herein.
Photometric Response of Sulfur and Phosphorus
Similar to earlier efforts using a fused silica burner, the best μFPD signal to noise ratios in this study are also generally found at lower flows of oxygen and hydrogen, the former having a much more significant impact on the background emission. Using stainless steel the optimum μFPD response for sulfur was obtained with 7 mL/min of oxygen and 45 mL/min of hydrogen, while that for phosphorus was obtained when using 9 mL/min of oxygen and 58 mL/min of hydrogen. While these oxygen flows agree within 3 to 5 mL/min of those used in the ‘post-column’ detection mode of the previous μFPD experiments, the hydrogen flows used are 30 to 40 mL/min smaller [27]. However, as demonstrated in that study, the latter is directly proportional to the inner diameter of the quartz capillary sleeve used. Since the sleeve used currently is narrower by comparison, smaller optimum hydrogen flows are to be expected.
As a result of the diminished background emission obtained using stainless steel, the signal to noise ratios realized for sulfur and phosphorus are about 100 times larger than those reported earlier for the μFPD.
Narrow band interference filters are often used to selectively monitor sulfur or phosphorus response in the conventional FPD, although this practice is known to decrease sensitivity. Since these methods were equally effective in the μFPD with a fused silica capillary burner, no differences in behavior of the narrow band interference filters were anticipated or observed from using stainless steel instead. For example, when the S2* emission of sulfur is isolated and monitored near 400 nm, the μFPD sensitivity for this element typically decreases by a factor of 2 to 10 times depending on the filter used. Comparable results are also obtained when observing the HPO* emission of phosphorus near 526 nm. Selective monitoring of sulfur and phosphorus using suitable interference filters with the μFPD is demonstrated later in this study.
Another concern that arises for monitoring sulfur using an FPD is the quenching of analyte signal that occurs in the presence co-eluting hydrocarbons [C. G. Flinn, W. A. Aue, Can. J. Spectrosc. 25 (1980) 141 and Dressler cited above]. While this phenomenon is widely observed in conventional FPD detection, it is unknown to what extent that it may occur in the counter current flame of the micro-FPD. Since very similar chemiluminescent systems, such as the reactive flow detector do not demonstrate this phenomenon, it is therefore useful to examine if response quenching by co-eluting hydrocarbons is observed in the micro-FPD. In order to investigate this, a sulfur peak was measured with and without a co-eluting solvent peak present. Table 2 displays the results and clearly indicates that as the amount of co-eluting acetone approaches 1 μL, the sulfur response reduces to approximately 30% of that which occurred without any acetone present. This amount of acetone corresponds to about 60 μg s−1 of carbon flow in the detector, which agrees with the mass flow of carbon observed to induce sulfur response quenching in a conventional FPD. Thus, similar to a conventional FPD, sulfur response quenching due to co-eluting hydrocarbons does occur in the micro-FPD and this effect appears to only be significant for carbon flows in the microgram range.
The setup used also helps avoid false positives and avoids carbon influencing the results. The simultaneous FID mode helps to identify large amounts of material as opposed to strongly responding sulfur or phosphorus compounds.
Photometric Response of Tin
Tin is another element commonly monitored by a conventional FPD, normally producing a red and/or blue chemiluminescence in the detector. Thus far, tin response has not been examined in the μFPD or in the larger counter-current flame. However, during the course of this study, quartz sleeves contaminated with traces of tin were visually observed to yield an intense blue emission on the surface of the enclosure surrounding the flame. This same luminescence is also observed in the form of tailing peaks when picogram quantities of tetramethyl tin are introduced into the detector equipped with a regular clean quartz capillary sleeve. This is consistent with the emission of SnO* on a quartz surface, which is well known to yield a very sensitive response toward tin compounds in a conventional FPD. Incidentally, the much less sensitive red emission in the gas phase (ascribed to SnH*) was not observed here. Therefore, considering its intensity, the blue tin emission was further examined in the μFPD.
Optimum signal to noise ratios for tin were obtained using 10 mL/min of oxygen and 25 mL/min of hydrogen. These μFPD conditions provide sensitive response yielding a detection limit near 6×10−15 gSn/s. However, increasing amounts of tin were only found to linearly increase the response over an order of magnitude. For instance, with tetramethyl tin this is observed between approximately 0.1 and 1 pg of the injected compound. This narrow linear range also reproduces with other calibration standards such as tetrabutyl tin, and under a variety of gas flows investigated.
Ionization Response
By using a stainless steel capillary burner in the μFPD, it has been demonstrated that the micro counter-current flame yields chemiluminescent response that is very similar to that found in much larger conventional or counter-current FPD flames. Earlier, it was also briefly noted that the larger counter-current flame was observed to provide highly sensitive FID response toward an aliphatic and an aromatic test analyte. However, more comprehensive aspects such as the relative sensitivity toward these analytes or the linearity of these responses were not investigated in the primarily photometric study. Since the micro counter-current flame provides photometric response that is similar to its larger analogue, it was somewhat anticipated that it too might also deliver useful ionization response toward carbon. However, the fuel-rich hydrogen radical flame chemistry that supports photometric signals is unique from the air-rich oxygen radical flame chemistry that promotes hydrocarbon ionization. Since the effect of reducing counter-current flame size on these processes remains unclear, it is therefore necessary to establish and investigate the extent of FID response that can be derived from the μFPD flame. This information is also potentially beneficial since such a feature could be useful in applications where both universal and selective detection of samples is desired.
Fortunately, this is facilitated by using a stainless steel capillary burner 38 in the μFPD, which makes it very convenient to apply a potential across the flame. Using the existing FID electrical leads of the GC, this mode of response was examined by applying the polarizer 44 to the capillary burner 38 and the collector 46 to the stainless steel sleeve 22 surrounding the end of the separation column. An arrangement of leads with a polarized flame burner situated below the collector of a conventional FID is known from H. H. Hill, D. G. McMinn, in Detectors for Capillary Chromatography; eds. D. G. McMinn, H. H. Hill; John Wiley, New York, 1992, 7. While other variations such as reversing the polarizer and collector connections were explored, these were not found to yield as favorable a response.
Several gas flows were examined for their impact on the ionization response of the flame. Initially, when the capillary burner was new, about 12 mL/min of oxygen was found to provide the best sensitivity. However, after a few hours of conditioning, this value decreased and stabilized at lower flows. Ultimately, the optimum gas flows for the “μFID” response mode of this flame toward carbon were obtained using 7 mL/min of oxygen and 40 mL/min of hydrogen. It is interesting to note that this flame stoichiometry is considerably rich in hydrogen compared to that of a conventional FID, which commonly yields optimal response when operated under leaner oxygen-rich conditions [Hill]. However, the flames used in the two devices are significantly different, particularly with respect to their structure.
For instance, the FID flame normally operates in a diffusion mode where hydrogen and column effluent are introduced through a central burner supporting the flame, which is concentrically surrounded by an excess of oxygen [Hill]. In contrast to this, the μFID flame operates in the unique counter-current mode, where it is supported on a capillary delivering oxygen, and burns in a counter-flowing excess of hydrogen mixed with column effluent. In this fashion the analyte, immersed in hydrogen, is directed toward the counter-current flame's oxygen-rich inner cone through its hydrogen-rich outer mantle. This is opposed to the conventional FID where analytes, also immersed in hydrogen, enter the oxygen-rich outer mantle of the flame through its hydrogen-rich inner cone region.
Considering these structural differences then, they could possibly play an important role in the strong ionization response observed in the fuel-rich μFID flame. For example, if more effective mixing were to occur in the micro counter-current flame, despite its richer stoichiometry, it might still efficiently produce oxygenated carbon species such as CHO similar to the air-rich diffusion flame of a conventional FID. Note that this species is believed in the art to be responsible for the sensitive signal of the FID, even though it is only produced by about 1 in 106 carbon atoms. Unfortunately, the actual extent of mixing in these two detectors, or their relative yields of flame species such as CHO is not presently established. However, it is interesting to point out that the above scenario is consistent with earlier reports of strong FID sensitivity being obtained from a premixed, fuel-rich, hydrogen/oxygen flame, and an even more turbulent oscillating FID flame.
Previously the larger counter-current flame was found to provide optimal ionization and optimal photometric response using respective air flows that were similar and hydrogen flows that differed by about 2 to 5 times. In this way the detector appeared flexible for dual channel operation. However, since FID response was not primarily examined in that study, the effect on sensitivity of changing the gas flow settings was not investigated. What is interesting about the optimum hydrogen-rich gas flows for ionization response in the current study, is that they are now much closer to those employed for optimal photometric response than was the case for the larger counter-current flame. Thus, it should be possible to utilize a common set of conditions that would provide both optimal, or near optimal μFPD and μFID response. This would be very useful in allowing the simultaneous screening of samples by both detection modes since conventionally one derives optimal response using an entirely different flame stoichiometry than the other. Therefore, it would be useful to know how the μFPD sensitivity for sulfur and for phosphorus may be influenced by changing the gas flows between the various optimum μFPD and μFID settings.
Table 1 illustrates the relative change in the μFPD sulfur signal when using gas flows optimized for obtaining photometric sulfur, photometric phosphorus, and hydrocarbon ionization response from the flame. Also included is a similar set of data illustrating the relative change in the μFPD phosphorus signal in each of these three operating modes. As can be seen from the table, the μFPD sensitivity for sulfur and for phosphorus changes relatively little amongst the different settings. The sulfur signal is decreased by only 4% when operated in the photometric phosphorus mode, and by 10% when operated in the hydrocarbon ionization mode. By comparison, the phosphorus signal is decreased by 15% when operated in the photometric sulfur mode. Furthermore, under μFID optimized conditions where the largest change is observed, the μFPD phosphorus signal still maintains about 70% of its optimal sensitivity.
Sample Analysis
Given that significant ionization and chemiluminescent signals can both be obtained from the same micro counter-current flame, detector performance was studied when analyzing an organic sample matrix. In order to demonstrate this, a quantity of unleaded gasoline (purchased from a local vendor) was spiked with both tetrahydrothiophene and trimethyl phosphite prior to analysis. Since gasoline typically contains a moderate variety of hydrocarbon compounds, this simple sample provides a good illustration of the detector's ability to screen a multi-component mixture for its carbon, sulfur, and phosphorus content.
In all, the attributes of this method demonstrate that the hydrogen-rich micro counter-current flame is indeed capable of delivering useful, sensitive response toward organic analytes. In spite of its very small size, it yields selective chemiluminescent and universal hydrocarbon ionization response that is similar in quantity and quality to those of conventional flame based detectors. As well, since it can deliver this as a multi-dimensional response under a common set of conditions, the micro counter-current flame method allows for more information to be obtained from a sample analysis. The properties and dimensions of the micro counter-current flame may therefore be potentially useful for application to analytical devices of reduced proportions. For instance, since the method can support a stable hydrogen-rich micro-flame within a small channel, it may be beneficial for portable or miniature GC methods where the performance of a conventional FPD and/or FID in an enclosed micro format is desirable. The apparatus and method disclosed here should also act as a useful flame source to support and adapt other micro-flame based detection methods such as Alkali Flame Detection. The apparatus and method disclosed also have utility in refinery and hydrocarbon processing plants for example in online applications.
aEach mode is optimized for the element shown in brackets. Conditions are listed in the text.
aInjected as ethyl sulfide; monitored using a 400 nm wide band colored glass filter (100 nm bandpass).
bPeak separation is 10 s.
Immaterial modifications may be made to the embodiment of the invention described here without departing from the invention.
This application claims the benefit under 35 U.S.C. 119(e) of U.S. provisional application no. 60/582,549 filed Jun. 25, 2004.
Number | Date | Country | |
---|---|---|---|
60582549 | Jun 2004 | US |