The technology described herein generally relates to systems for detecting polynucleotides in samples, particularly from biological samples. The technology more particularly relates to microfluidic systems that carry out PCR on nucleotides of interest within microfluidic channels, and detect those nucleotides.
The medical diagnostics industry is a critical element of today's healthcare infrastructure. At present, however, diagnostic analyses no matter how routine have become a bottleneck in patient care. There are several reasons for this. First, many diagnostic analyses can only be done with highly specialist equipment that is both expensive and only operable by trained clinicians. Such equipment is found in only a few locations—often just one in any given urban area. This means that most hospitals are required to send out samples for analyses to these locations, thereby incurring shipping costs and transportation delays, and possibly even sample loss. Second, the equipment in question is typically not available ‘on-demand’ but instead runs in batches, thereby delaying the processing time for many samples because they must wait for a machine to fill up before they can be run.
Understanding that sample flow breaks down into several key steps, it would be desirable to consider ways to automate as many of these as possible. For example, a biological sample, once extracted from a patient, must be put in a form suitable for a processing regime that typically involves using PCR to amplify a vector of interest. Once amplified, the presence of a nucleotide of interest from the sample needs to be determined unambiguously. Sample preparation is a process that is susceptible to automation but is also relatively routinely carried out in almost any location. By contrast, steps such as PCR and nucleotide detection have customarily only been within the compass of specially trained individuals having access to specialist equipment.
There is a need for a method and apparatus of carrying out PCR and detection on prepared biological samples, and preferably with high throughput. In particular there is a need for an easy-to-use device that can deliver a diagnostic result on several samples in a short time.
The discussion of the background to the technology herein is included to explain the context of the technology. This is not to be taken as an admission that any of the material referred to was published, known, or part of the common general knowledge as at the priority date of any of the claims.
Throughout the description and claims of the specification the word “comprise” and variations thereof, such as “comprising” and “comprises”, is not intended to exclude other additives, components, integers or steps.
The present technology addresses systems for detecting polynucleotides in samples, particularly from biological samples. In particular, the technology relates to microfluidic systems that carry out PCR on nucleotides of interest within microfluidic channels, and detect those nucleotides.
An apparatus, comprising: a receiving bay configured to receive a microfluidic cartridge; at least one heat source thermally coupled to the cartridge and configured to carry out PCR on a microdroplet of polynucleotide-containing sample, in the cartridge; a detector configured to detect presence of one or more polynucleotides in the sample; and a processor coupled to the detector and the heat source, configured to control heating of one or more regions of the microfluidic cartridge.
A method of carrying out PCR on a plurality of polynucleotide-containing samples, the method comprising: introducing the plurality of samples in to a microfluidic cartridge, wherein the cartridge has a plurality of PCR reaction chambers configured to permit thermal cycling of the plurality of samples independently of one another; moving the plurality of samples into the respective plurality of PCR reaction chambers; and amplifying polynucleotides contained with the plurality of samples, by application of successive heating and cooling cycles to the PCR reaction chambers.
The details of one or more embodiments of the technology are set forth in the accompanying drawings and further description herein. Other features, objects, and advantages of the technology will be apparent from the description and drawings, and from the claims.
Additional figures are illustrated within the examples, and are further described therein.
Like reference symbols in the various drawings indicate like elements.
The present technology relates to a system and related methods for amplifying, and carrying out diagnostic analyses on, polynucleotides (e.g., a DNA, RNA, mRNA, or rRNA) from biological samples. For example, the system and methods can determine whether a polynucleotide indicative of the presence of a particular pathogen (such as a bacterium or a virus) can be present. The polynucleotide may be a sample of genomic DNA, or may be a sample of mitochondrial DNA. The nucleotides are typically provided to the system having been isolated or released from particles such as cells in the sample. The system includes a disposable microfluidic cartridge containing multiple sample lanes in parallel and a reusable instrument platform (a PCR analyzer apparatus) that can actuate on-cartridge operations, can detect (e.g., by fluorescence detection) and analyze the products of the PCR amplification in each of the lanes separately, in all simultaneously, or in groups simultaneously, and, optionally, can display the results on a graphical user interface.
A system, microfluidic cartridge, heater unit, detector, kit, methods, and associated computer program product, are now further described.
By cartridge is meant a unit that may be disposable, or reusable in whole or in part, and that is configured to be used in conjunction with some other apparatus that has been suitably and complementarily configured to receive and operate on (such as deliver energy to) the cartridge.
By microfluidic, as used herein, is meant that volumes of sample, and/or reagent, and/or amplified polynucleotide are from about 0.1 μl to about 999 μl, such as from 1-100 μl, or from 2-25 μl. Similarly, as applied to a cartridge, the term microfluidic means that various components and channels of the cartridge, as further described herein, are configured to accept, and/or retain, and/or facilitate passage of microfluidic volumes of sample, reagent, or amplified polynucleotide.
System Overview
A schematic overview of a system 981 for carrying out analyses described herein is shown in
A processor 980, such as a microprocessor, is configured to control functions of various components of the system as shown, and is thereby in communication with each such component. In particular, processor 980 is configured to receive data about a sample to be analyzed, e.g., from a sample reader 990, which may be a barcode reader, an optical character reader, or an RFID scanner (radio frequency tag reader). For example, the sample identifier can be a handheld bar code reader. Processor 980 can be configured to accept user instructions from an input 984, where such instructions may include instructions to start analyzing the sample, and choices of operating conditions.
Processor 980 can also be configured to communicate with an optional display 982, so that, for example, information including but not limited to the current status of the system, progress of PCR thermocycling, and any warning message in case of malfunction of either system or cartridge, as well as results of analysis, are transmitted to the display. Additionally, processor 980 may transmit one or more questions to be displayed on display 982 that prompt a user to provide input in response thereto. Thus, in certain embodiments, input 984 and display 982 are integrated with one another.
Processor 980 can be optionally further configured to transmit results of an analysis to an output device such as a printer, a visual display, or a speaker, or a combination thereof, the transmission being either directly through a directly dedicated printer cable, or wirelessly, or via a network connection.
Processor 980 is still further optionally connected via a communication interface such as a network interface to a computer network 988. The communication interface can be one or more interfaces selected from the group consisting of: a serial connection, a parallel connection, a wireless network connection and a wired network connection such as an ethernet, firewire, cable connection, or one using USB connectivity. Thereby, when the system is suitably addressed on the network, a remote user may access the processor and transmit instructions, input data, or retrieve data, such as may be stored in a memory (not shown) associated with the processor, or on some other computer-readable medium that is in communication with the processor. The computer network connection may also permit extraction of data to a remote location, such as a personal computer, personal digital assistant, or network storage device such as computer server or disk farm. The apparatus may further be configured to permit a user to e-mail results of an analysis directly to some other party, such as a healthcare provider, or a diagnostic facility, or a patient.
Although not shown in
Additionally, in various embodiments, the apparatus can further comprise a data storage medium configured to receive data from one or more of the processor, an input device, and a communication interface, the data storage medium being one or more media selected from the group consisting of: a hard disk drive, an optical disk drive, or one or more removable storage media such as a CD-R, CD-RW, USB-drive, and a flash card.
Processor 980 is further configured to control various aspects of sample diagnosis, as follows in overview, and as further described in detail herein. The system is configured to operate in conjunction with a complementary cartridge 994, such as a microfluidic cartridge. The cartridge is itself configured, as further described herein, to receive one or more samples 996 containing one or more polynucleotides in a form suitable for amplification and diagnostic analysis. The cartridge has dedicated regions within which amplification, such as by PCR, of the polynucleotides is carried out when the cartridge is situated in the apparatus.
The microfluidic cartridge is received by a receiving bay 992 configured to selectively receive the cartridge. For example, the receiving bay and the microfluidic cartridge can be complementary in shape so that the microfluidic cartridge is selectively received in, e.g., a single orientation. The microfluidic cartridge can have a registration member that fits into a complementary feature of the receiving bay. The registration member can be, for example, a cut-out on an edge of the cartridge, such as a corner that is cut-off, or one or more notches that are made on one or more of the sides. By selectively receiving the cartridge, the receiving bay can help a user to place the cartridge so that the apparatus can properly operate on the cartridge. The receiving bay can also be configured so that various components of the apparatus that can operate on the microfluidic cartridge (heat sources, detectors, force members, and the like) are positioned to properly operate on the microfluidic cartridge. In some embodiments, the apparatus can further include a sensor coupled to the processor, the sensor configured to sense whether the microfluidic cartridge is selectively received.
The receiving bay is in communication with a heater unit 998 that itself is controlled by processor 980 in such a way that specific regions of the cartridge, such as individual sample lanes, are independently and selectively heated at specific times during amplification and analysis. The processor can be configured to control application of heat to the individual sample lanes, separately, in all simultaneously, or in groups simultaneously.
The heat source can be, for example, a contact heat source such as a resistive heater or a network of resistive heaters, or a Peltier device, and the like. The contact heat source can be configured to be in direct physical contact with one or more distinct locations of a microfluidic cartridge received in the receiving bay. In various embodiments, each contact source heater can be configured to heat a distinct location having an average diameter in 2 dimensions from about 1 millimeter (mm) to about 15 mm (typically about 1 mm to about 10 mm), or a distinct location having a surface area of between about 1 mm2 about 225 mm2 (typically between about 1 mm2 and about 100 mm2, or in some embodiments between about 5 mm2 and about 50 mm2).
In various embodiments, the heat source can be situated in an assembly that is removable from the apparatus, for example, to permit cleaning or to replace the heater configuration.
In various embodiments, the apparatus can include a compliant layer at the contact heat source configured to thermally couple the contact heat source with at least a portion of a microfluidic cartridge received in the receiving bay. The compliant layer at can have a thickness of between about 0.05 and about 2 millimeters and a Shore hardness of between about 25 and about 100.
In various embodiments, the apparatus can further include one or more force members (not shown in
In various embodiments, the one or more force members are configured to apply force to a plurality of locations in the microfluidic cartridge. The force applied by the one or more force members can result in an average pressure at an interface between a portion of the receiving bay and a portion of the microfluidic cartridge of between about 5 kilopascals and about 50 kilopascals, for example, the average pressure can be at least about 7 kilopascals, and still more preferably at least about 14 kilopascals. At least one force member can be manually operated. At least one force member can be mechanically coupled to a lid at the receiving bay, whereby operation of the lid operates the force member. The application of force is important to ensure consistent thermal contact between the heater wafer and the PCR reactor and microvalves in the microfluidic cartridge.
In various embodiments, the apparatus can further include a lid at the receiving bay, the lid being operable to at least partially exclude ambient light from the receiving bay. The lid can be, for example, a sliding lid. The lid can include the optical detector. A major face of the lid at the optical detector or at the receiving bay can vary from planarity by less than about 100 micrometers, for example, less than about 25 micrometers. The lid can be configured to be removable from the apparatus. The lid can include a latching member that ensures that the lid is securely closed before amplification reactions are applied to the samples in the cartridge.
The processor is also configured to receive signals from and control a detector 999 configured to detect a polynucleotide in a sample in one or more of the individual sample lanes, separately or simultaneously. The processor thereby provides an indication of a diagnosis from the cartridge 994. Diagnosis can be predicated on the presence or absence of a specific polynucleotide in a particular sample. The diagnosis can be transmitted to the output device 986 and/or the display 982, as described hereinabove.
The detector can be, for example, an optical detector that includes a light source that selectively emits light in an absorption band of a fluorescent dye, and a light detector that selectively detects light in an emission band of the fluorescent dye, wherein the fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof. Alternatively, for example, the optical detector can include a bandpass-filtered diode that selectively emits light in the absorption band of the fluorescent dye and a bandpass filtered photodiode that selectively detects light in the emission band of the fluorescent dye; or for example, the optical detector can be configured to independently detect a plurality of fluorescent dyes having different fluorescent emission spectra, wherein each fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof; or for example, the optical detector can be configured to independently detect a plurality of fluorescent dyes at a plurality of different locations on a microfluidic cartridge, wherein each fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof in a different sample.
A suitable processor 980 can be designed and manufactured according to, respectively, design principles and semiconductor processing methods known in the art.
The system in
The system shown in outline in
The system of
The system of
In still another configuration, a system is configured to accept and to process multiple cartridges, but one or more components in
In still another configuration, a system as shown in
It is further consistent with the present technology that a cartridge can be tagged, e.g., with a molecular bar-code indicative of one or more of the samples, to facilitate sample tracking, and to minimize risk of sample mix-up. Methods for such tagging are described elsewhere, e.g., in U.S. patent application Ser. No. 10/360,854, incorporated herein by reference.
In various embodiments, the apparatus can further include an analysis port. The analysis port can be configured to allow an external sample analyzer to analyze a sample in the microfluidic cartridge; for example, the analysis port can be a hole or window in the apparatus which can accept an optical detection probe that can analyze a sample in situ in the microfluidic cartridge. In some embodiments, the analysis port can be configured to direct a sample from the microfluidic cartridge to an external sample analyzer; for example, the analysis port can include a conduit in fluid communication with the microfluidic cartridge that direct a liquid sample to a chromatography apparatus, an optical spectrometer, a mass spectrometer, or the like.
Apparatus 100 may optionally comprise one or more stabilizing feet that cause the body of the device to be elevated above a surface on which system 100 is disposed, thereby permitting ventilation underneath system 100, and also providing a user with an improved ability to lift system 100. There may be 2, 3, 4, 5, or 6, or more feet, depending upon the size of system 100. Such feet are preferably made of rubber, or plastic, or metal, and in some embodiments may elevate the body of system 100 by from about 2 to about 10 mm above a surface on which it is situated. The stabilizing function can also be provided by one or more runners that run along one or more edges—or are inwardly displaced from one or more edges—of the underside of the apparatus. Such runners can also be used in conjunction with one or more feet. In another embodiment, a turntable situated on the underside permits the apparatus to be rotated in a horizontal or near-horizontal plane when positioned on, e.g., a benchtop, to facilitate access from a number of angles by a user.
Exemplary Systems
System 2000 comprises a housing 2002, which can-be made of metal, or a hardened plastic. The form of the housing shown in
System 2000 further comprises a display 2006, which may be a liquid crystal display, such as active matrix, an OLED, or some other suitable form. It may present images and other information in color or in black and white. Display 2006 may also be a touch-sensitive display and therefore may be configured to accept input from a user in response to various displayed prompts. Display 2006 may have an anti-reflective coating on it to reduce glare and reflections from overhead lights in an laboratory setting. Display 2006 may also be illuminated from, e.g., a back-light, to facilitate easier viewing in a dark laboratory.
System 2000, as shown in
Handle 2008 performs a role of permitting a user to move lid 2010 from one position to another, and also performs a role of causing pressure to be forced down on the lid, when in a closed position, so that pressure can be applied to a cartridge in the receiving bay 2014. In
In one embodiment, the handle and lid assembly are also fitted with a mechanical sensor that does not permit the handle to be depressed when there is no cartridge in the receiving bay. In another embodiment, the handle and lid assembly are fitted with a mechanical latch that does not permit the handle to be raised when an analysis is in progress.
A further configuration of system 2000 is shown in
Heater module 2020 is preferably removable, and is further described hereinbelow.
Computer readable medium input 2022 may accept one or more of a variety of media. Shown in
Features shown on the rear of system 2000 may be arranged in any different manner, depending upon an internal configuration of various components. Additionally, features shown as being on the rear of system 2000, may be optionally presented on another face of system 2000, depending on design preference. Shown in
An exploded view of an exemplary embodiment of the apparatus is shown in
Embodiments of apparatus 2000 also include software (e.g., for interfacing with users, conducting analysis and/or analyzing test results), firmware (e.g., for controlling the hardware during tests on the cartridge 812), and one or more peripheral communication interfaces shown collectively as 2031 for peripherals (e.g., communication ports such as USB/Serial/Ethernet to connect to storage such as compact disc or hard disk, to connect input devices such as a bar code reader and/or a keyboard, to connect to other computers or storage via a network, and the like).
Control electronics 840, shown schematically in the block diagram in
Microfluidic Cartridge
The present technology comprises a microfluidic cartridge that is configured to carry out an amplification, such as by PCR, of one or more polynucleotides from one or more samples. It is to be understood that, unless specifically made clear to the contrary, where the term PCR is used herein, any variant of PCR including but not limited to real-time and quantitative, and any other form of polynucleotide amplification is intended to be encompassed. The microfluidic cartridge need not be self-contained and can be designed so that it receives thermal energy from one or more heating elements present in an external apparatus with which the cartridge is in thermal communication. An exemplary such apparatus is further described herein; additional embodiments of such a system are found in U.S. patent application Ser. No. 11/940,310, entitled “Microfluidic Cartridge and Method of Making Same”, and filed on even date herewith, the specification of which is incorporated herein by reference.
By cartridge is meant a unit that may be disposable, or reusable in whole or in part, and that is configured to be used in conjunction with some other apparatus that has been suitably and complementarily configured to receive and operate on (such as deliver energy to) the cartridge.
By microfluidic, as used herein, is meant that volumes of sample, and/or reagent, and/or amplified polynucleotide are from about 0.1 μl to about 999 such as from 1-100 μl, or from 2-25 μl. Similarly, as applied to a cartridge, the term microfluidic means that various components and channels of the cartridge, as further described herein, are configured to accept, and/or retain, and/or facilitate passage of microfluidic volumes of sample, reagent, or amplified polynucleotide. Certain embodiments herein can also function with nanolitre volumes (in the range of 10-500 nanoliters, such as 100 nanoliters).
One aspect of the present technology relates to a microfluidic cartridge having two or more sample lanes arranged so that analyses can be carried out in two or more of the lanes in parallel, for example simultaneously, and wherein each lane is independently associated with a given sample.
A sample lane is an independently controllable set of elements by which a sample can be analyzed, according to methods described herein as well as others known in the art. A sample lane comprises at least a sample inlet, and a microfluidic network having one or more microfluidic components, as further described herein.
In various embodiments, a sample lane can include a sample inlet port or valve, and a microfluidic network that comprises, in fluidic communication one or more components selected from the group consisting of: at least one thermally actuated valve, a bubble removal vent, at least one thermally actuated pump, a gate, mixing channel, positioning element, microreactor, a downstream thermally actuated valve, and a PCR reaction chamber. The sample inlet valve can be configured to accept a sample at a pressure differential compared to ambient pressure of between about 70 and 100 kilopascals.
The cartridge can therefore include a plurality of microfluidic networks, each network having various components, and each network configured to carry out PCR on a sample in which the presence or absence of one or more polynucleotides is to be determined.
A multi-lane cartridge is configured to accept a number of samples in series or in parallel, simultaneously or consecutively, in particular embodiments 12 samples, wherein the samples include at least a first sample and a second sample, wherein the first sample and the second sample each contain one or more polynucleotides in a form suitable for amplification. The polynucleotides in question may be the same as, or different from one another, in different samples and hence in different lanes of the cartridge. The cartridge typically processes each sample by increasing the concentration of a polynucleotide to be determined and/or by reducing the concentration of inhibitors relative to the concentration of polynucleotide to be determined.
The multi-lane cartridge comprises at least a first sample lane having a first microfluidic network and a second lane having a second microfluidic network, wherein each of the first microfluidic network and the second microfluidic network is as elsewhere described herein, and wherein the first microfluidic network is configured to amplify polynucleotides in the first sample, and wherein the second microfluidic network is configured to amplify polynucleotides in the second sample.
In various embodiments, the microfluidic network can be configured to couple heat from an external heat source to a sample mixture comprising PCR reagent and neutralized polynucleotide sample under thermal cycling conditions suitable for creating PCR amplicons from the neutralized polynucleotide sample.
At least the external heat source may operate under control of a computer processor, configured to execute computer readable instructions for operating one or more components of each sample lane, independently of one another, and for receiving signals from a detector that measures fluorescence from one or more of the PCR reaction chambers.
For example,
In various embodiments, the PCR reagent mixture can include a positive control plasmid and a plasmid fluorogenic hybridization probe selective for at least a portion of the plasmid, and the microfluidic cartridge can be configured to allow independent optical detection of the fluorogenic hybridization probe and the plasmid fluorogenic hybridization probe.
In various embodiments, the microfluidic cartridge can accommodate a negative control polynucleotide, wherein the microfluidic network can be configured to independently carry out PCR on each of a neutralized polynucleotide sample and a negative control polynucleotide with the PCR reagent mixture under thermal cycling conditions suitable for independently creating PCR amplicons of the neutralized polynucleotide sample and PCR amplicons of the negative control polynucleotide. Each lane of a multi-lane cartridge as described herein can perform two reactions when used in conjunction with two fluorescence detection systems per lane. A variety of combinations of reactions can be performed in the cartridge, such as two sample reactions in one lane, a positive control and a negative control in two other lanes; or a sample reaction and an internal control in one lane and a negative control in a separate lane.
In preferred embodiments, the multi-sample cartridge has a size substantially the same as that of a 96-well plate as is customarily used in the art. Advantageously, then, such a cartridge may be used with plate handlers used elsewhere in the art.
The sample inlets of adjacent lanes are reasonably spaced apart from one another to prevent any contamination of one sample inlet from another sample when a user introduces a sample into any one cartridge. In an embodiment, the sample inlets are configured so as to prevent subsequent inadvertent introduction of sample into a given lane after a sample has already been introduced into that lane. In certain embodiments, the multi-sample cartridge is designed so that a spacing between the centroids of sample inlets is 9 mm, which is an industry-recognized standard. This means that, in certain embodiments the center-to-center distance between inlet holes in the cartridge that accept samples from PCR tubes, as further described herein, is 9 mm. The inlet holes can be manufactured conical in shape with an appropriate conical angle so that industry-standard pipette tips (2 μl, 20 μl, 200 μl, volumes, etc.) fit snugly therein. The cartridge herein may be adapted to suit other, later-arising, industry standards not otherwise described herein, as would be understood by one of ordinary skill in the art.
In one embodiment, an exemplary microfluidic cartridge has 12 sample lanes. The inlet ports in this embodiment have a 6 mm spacing, so that, when used in conjunction with an automated sample loader having 4 heads, spaced equidistantly at 18 mm apart, the inlets can be loaded in three batches of four inlets: e.g., inlets 1, 4, 7, and 10 together, followed by 2, 5, 8, and 11, then finally 3, 6, 9, and 12, wherein the 12 inlets are numbered consecutively from one side of the cartridge to the other as shown.
A microfluidic cartridge as used herein may be constructed from a number of layers. Accordingly, one aspect of the present technology relates to a microfluidic cartridge that comprises a first, second, third, fourth, and fifth layers wherein one or more layers define a plurality of microfluidic networks, each network having various components configured to carry out PCR on a sample in which the presence or absence of one or more polynucleotides is to be determined. In various embodiments, one or more such layers are optional.
Microfluidic cartridge 400 can be fabricated as desired. The cartridge can include a microfluidic substrate layer 424, typically injection molded out of a plastic, such as a zeonor plastic (cyclic olefin polymer), having a PCR channel and valve channels on a first side and vent channels and various inlet holes, including wax loading holes and liquid inlet holes, on a second side (disposed toward hydrophobic vent membrane 426). It is advantageous that all the microfluidic network defining structures, such as PCR reactors, valves, inlet holes, and air vents, are defined on the same single substrate 424. This attribute facilitates manufacture and assembly of the cartridge. Additionally, the material from which this substrate is formed is rigid or nondeformable, non-venting to air and other gases, and has a low autofluorescence to facilitate detection of polynucleotides during an amplification reaction performed in the microfluidic circuitry defined therein. Rigidity is advantageous because it facilitates effective and uniform contact with a heat unit as further described herein. Use of a non-venting material is also advantageous because it reduces the likelihood that the concentration of various species in liquid form will change during analysis. Use of a material having low auto-fluorescence is also important so that background fluorescence does not detract from measurement of fluorescence from the analyte of interest.
The cartridge can further include, disposed on top of the substrate 424, an oleophobic/hydrophobic vent membrane layer 426 of a porous material, such as 0.2 to 1.0 micron pore-size membrane of modified polytetrafluorethylene, the membrane being typically between about 25 and about 100 microns thick, and configured to cover the vent channels of microfluidic substrate 424, and attached thereto using, for example, heat bonding.
Typically, the microfluidic cartridge further includes a layer 428, 430 of polypropylene or other plastic label with pressure sensitive adhesive (typically between about 50 and 150 microns thick) configured to seal the wax loading holes of the valves in substrate 424, trap air used for valve actuation, and serve as a location for operator markings. In
In various embodiments, the label is a computer-readable label. For example, the label can include a bar code, a radio frequency tag or one or more computer-readable characters. The label can be formed of a mechanically compliant material. For example, the mechanically compliant material of the label can have a thickness of between about 0.05 and about 2 millimeters and a Shore hardness of between about 25 and about 100. The label can be positioned such that it can be read by a sample identification verifier as further described herein.
The cartridge can further include a heat sealable laminate layer 422 (typically between about 100 and about 125 microns thick) attached to the bottom surface of the microfluidic substrate 424 using, for example, heat bonding. This layer serves to seal the PCR channels and vent channels in substrate 424. The cartridge can further include a thermal interface material layer 420 (typically about 125 microns thick), attached to the bottom of the heat sealable laminate layer using, for example, pressure sensitive adhesive. The layer 420 can be compressible and have a higher thermal conductivity than common plastics, thereby serving to transfer heat across the laminate more efficiently. Typically, however, layer 420 is not present.
The application of pressure to contact the cartridge to the heater of an instrument that receives the cartridge generally assists in achieving better thermal contact between the heater and the heat-receivable parts of the cartridge, and also prevents the bottom laminate structure from expanding, as would happen if the PCR channel was only partially filled with liquid and the air entrapped therein would be thermally expanded during thermocycling.
In use, cartridge 400 is typically thermally associated with an array of heat sources configured to operate the components (e.g., valves, gates, actuators, and processing region 410) of the device. Exemplary such heater arrays are further described herein. Additional embodiments of heater arrays are described in U.S. patent application Ser. No. 11/940,315, entitled “Heater Unit for Microfluidic Diagnostic System” and filed on even date herewith, the specification of which is incorporated herein by reference in its entirety. In some embodiments, the heat sources are controlled by a computer processor and actuated according to a desired protocol. Processors configured to operate microfluidic devices are described in, e.g., U.S. application Ser. No. 09/819,105, filed Mar. 28, 2001, which application is incorporated herein by reference.
In various embodiments, during transport and storage, the microfluidic cartridge can be further surrounded by a sealed pouch. The microfluidic cartridge can be sealed in the pouch with an inert gas. The microfluidic cartridge can be disposable for example after one or more of its sample lanes have been used.
Embodiments of the cartridge described herein may be constructed that have high-density microfluidic circuitry on a single cartridge that thereby permit processing of multiple samples in parallel, or in sequence, on a single cartridge. Preferred numbers of such multiple samples include 20, 24, 36, 40, 48, 50, 60, 64, 72, 80, 84, 96, and 100, but it would be understood that still other numbers are consistent with the apparatus and cartridge herein, where deemed convenient and practical.
Accordingly, different configurations of lanes, sample inlets, and associated heater networks than those explicitly depicted in the FIGS. and examples that can facilitate processing such numbers of samples on a single cartridge are within the scope of the instant disclosure. Similarly, alternative configurations of detectors and heating elements for use in conjunction with such a highly multiplexed cartridge are also within the scope of the description herein.
It is also to be understood that the microfluidic cartridges described herein are not to be limited to rectangular shapes, but can include cartridges having circular, elliptical, triangular, rhombohedral, square, and other shapes. Such shapes may also be adapted to include some irregularity, such as a cut-out, to facilitate placement in a complementary apparatus as further described herein.
In an exemplary embodiment, a highly multiplexed cartridge has 48 sample lanes, and permits independent control of each valve in each lane by suitably configured heater circuitry, with 2 banks of thermo cycling protocols per lane, as shown in
The various embodiments shown in
During the design and manufacture of highly multiplexed cartridges, photolithographic processing steps such as etching, hole drilling/photo-chemical drilling/sand-blasting/ion-milling processes should be optimized to give well defined holes and microchannel pattern. Proper distances between channels should be identified and maintained to obtain good bonding between the microchannel substrate and the heat conducting substrate layer. In particular, it is desirable that minimal distances are maintained between pairs of adjacent microchannels to promote, reliable bonding of the laminate in between the channels.
The fabrication by injection molding of these complicated microfluidic structures having multiple channels and multiple inlet holes entails proper consideration of dimensional repeatability of these structures over multiple shots from the injection molding master pattern. Proper consideration is also attached to the placement of ejector pins to push out the structure from the mold without causing warp, bend or stretching of it. For example, impression of the ejector pins on the microfluidic substrate should not sink into the substrate thereby preventing planarity of the surface of the cartridge. The accurate placement of various inlet holes (such as sample inlet holes, valve inlet holes and vent holes) relative to adjacent microfluidic channels is also important because the presence of these holes can cause knit-lines to form that might cause unintended leak from a hole to a microchannel. Highly multiplexed microfluidic substrates may be fabricated in other materials such as glass, silicon.
The size of the substrate relative to the number of holes is also factor during fabrication because it is easy to make a substrate having just a simple microfluidic network with a few holes (may be fewer than 10 holes) and a few microchannels, but making a substrate having over 24, or over 48, or over 72 holes, etc., is more difficult.
Microfluidic Networks
Particular components of exemplary microfluidic networks are further described herein.
Channels of a microfluidic network in a lane of cartridge typically have at least one sub-millimeter cross-sectional dimension. For example, channels of such a network may have a width and/or a depth of about 1 mm or less (e.g., about 750 microns or less, about 500 microns, or less, about 250 microns or less).
Both valves 204 and 206 are closed prior to thermocycling to prevent or reduce any evaporation of liquid, bubble generation, or movement of fluid from the PCR reactor. End vent 214 is configured to prevent a user from introducing an excess amount of liquid into the microfluidic cartridge, as well as playing a role of containing any sample from spilling over to unintended parts of the cartridge. A user may input sample volumes as small as an amount to fill the region from the bubble removal vent (if present) to the middle of the microreactor, or up to valve 204 or beyond valve 204. The use of microvalves prevents both loss of liquid or vapor thereby enabling even a partially filled reactor to successfully complete a PCR thermocycling reaction.
The reactor 210 is a microfluidic channel that is heated through a series of cycles to carry out amplification of nucleotides in the sample, as further described herein, and according to amplification protocols known to those of ordinary skill in the art. The inside walls of the channel in the PCR reactor are typically made very smooth and polished to a shiny finish (for example, using a polish selected from SPI A1, SPI A2, SPI A3, SPI B1, or SPI B2) during manufacture. This is in order to minimize any microscopic quantities of air trapped in the surface of the PCR channel, which would causing bubbling during the thermocycling steps. The presence of bubbles especially in the detection region of the PCR channel could also cause a false or inaccurate reading while monitoring progress of the PCR. Additionally, the PCR channel can be made shallow such that the temperature gradient across the depth of the channel is minimized.
The region of the cartridge 212 above PCR reactor 210 is a thinned down section to reduce thermal mass and autofluorescence from plastic in the cartridge. It permits a detector to more reliably monitor progress of the reaction and also to detect fluorescence from a probe that binds to a quantity of amplified nucleotide. Exemplary probes are further described herein. The region 212 can be made of thinner material than the rest of the cartridge so as to permit the PCR channel to be more responsive to a heating cycle (for example, to rapidly heat and cool between temperatures appropriate for denaturing and annealing steps), and so as to reduce glare, autofluorescence, and undue absorption of fluorescence.
After PCR has been carried out on a sample, and presence or absence of a polynucleotide of interest has been determined, it is preferred that the amplified sample remains in the cartridge and that the cartridge is either used again (if one or more lanes remain unused), or disposed of. Should a user wish to run a post amplification analysis, such as gel electrophoresis, the user may pierce a hole through the laminate of the cartridge, and recover an amount—typically about 1.5 microliter—of PCR product. The user may also place the individual PCR lane on a special narrow heated plate, maintained at a temperature to melt the wax in the valve, and then aspirate the reacted sample from the inlet hole of that PCR lane.
In various embodiments, the microfluidic network can optionally include at least one reservoir configured to contain waste.
Table 1 outlines typical volumes, pumping pressures, and operation times associated with various components of a microfluidic cartridge described herein.
Valves
A valve (sometimes referred to herein as a microvalve) is a component in communication with a channel, such that the valve has a normally open state allowing material to pass along a channel from a position on one side of the valve (e.g., upstream of the valve) to a position on the other side of the valve (e.g., downstream of the valve). Upon actuation of the valve, the valve transitions to a closed state that prevents material from passing along the channel from one side of the valve to the other. For example, in one embodiment, a valve can include a mass of a thermally responsive substance (TRS) that is relatively immobile at a first temperature and more mobile at a second temperature. The first and second temperatures are insufficiently high to damage materials, such as polymer layers of a microfluidic cartridge in which the valve is situated. A mass of TRS can be an essentially solid mass or an agglomeration of smaller particles that cooperate to obstruct the passage when the valve is closed. Examples of TRS's include a eutectic alloy (e.g., a solder), wax (e.g., an olefin), polymers, plastics, and combinations thereof. The TRS can also be a blend of variety of materials, such as an emulsion of thermoelastic polymer blended with air microbubbles (to enable higher thermal expansion, as well as reversible expansion and contraction), polymer blended with expancel material (offering higher thermal expansion), polymer blended with heat conducting microspheres (offering faster heat conduction and hence, faster melting profiles), or a polymer blended with magnetic microspheres (to permit magnetic actuation of the melted thermoresponsive material).
Generally, for such a valve, the second temperature is less than about 90° C. and the first temperature is less than the second temperature (e.g., about 70° C. or less). Typically, a chamber is in gaseous communication with the mass of TRS. The valve is in communication with a source of heat that can be selectively applied to the chamber of air and to the TRS. Upon heating gas (e.g., air) in the chamber and heating the mass of TRS to the second temperature, gas pressure within the chamber due to expansion of the volume of gas, forces the mass to move into the channel, thereby obstructing material from passing therealong.
An exemplary valve is shown in
In various other embodiments, a valve for use with a microfluidic network in a microfluidic cartridge herein can be a bent valve as shown in
In various other embodiments, a valve for use with a microfluidic network can include a curved valve as shown in
Gates
In various embodiments, a microfluidic network can include a narrow gate 380 as shown in
In various embodiments, the gate can be configured to minimize the effective area or footprint of the gate within the network and thus bent gate configurations, although not shown herein are consistent with the foregoing description.
Vents
In various embodiments, the microfluidic network can include at least one hydrophobic vent in addition to an end vent. A vent is a general outlet (hole) that may or may not be covered with a hydrophobic membrane. An exit hole is an example of a vent that need not be covered by a membrane.
A hydrophobic vent (e.g., a vent in
The hydrophobic vents of the present technology are preferably constructed so that the amount of air that escapes through them is maximized while minimizing the volume of the channel below the vent surface. Accordingly, it is preferable that the vent is constructed so as to have a hydrophobic membrane 1303 of large surface area and a shallow cross section of the microchannel below the vent surface.
Hydrophobic vents are useful for bubble removal and typically have a length of at least about 2.5 mm (e.g., at least about 5 mm, at least about 7.5 mm) along a channel 1305 (see
The depth of the channel within the hydrophobic vent is typically about 75% or less (e.g., about 65% or less, about 60% or less) of the depth of the channel upstream 1301 and downstream (not shown) of the hydrophobic vent. For example, in some embodiments the channel depth within the hydrophobic vent is about 150 microns and the channel depth upstream and downstream of the hydrophobic vent is about 250 microns. Other dimensions are consistent with the description herein.
A width of the channel within the hydrophobic vent is typically at least about 25% wider (e.g., at least about 50% wider) than a width of the channel upstream from the vent and downstream from the vent. For example, in an exemplary embodiment, the width of the channel within the hydrophobic vent is about 400 microns, and the width of the channel upstream and downstream from the vent is about 250 microns. Other dimensions are consistent with the description herein.
The vent in
Use of Cutaways in Cartridge and Substrate to Improve Rate of Cooling During PCR Cycling
During a PCR amplification of a nucleotide sample, a number of thermal cycles are carried out. For improved efficiency, the cooling between each application of heat is preferably as rapid as possible. Improved rate of cooling can be achieved with various modifications to the heating substrate and/or the cartridge, as shown in
One way to achieve rapid cooling is to cutaway portions of the microfluidic cartridge substrate, as shown in
Another way to achieve rapid cooling is to cutaway portions of the heater substrate, as shown in
An example of thermal cycling performance in a PCR reaction chamber obtained with a configuration as described herein, is shown in
Manufacturing Process for Cartridge
At 1802, a laminate layer is applied to a microfluidic substrate that has previously been engineered, for example by injection molding, to have a microfluidic network constructed in it; edges are trimmed from the laminate where they spill over the bounds of the substrate.
At 1804, wax is dispensed and loaded into the microvalves of the microfluidic network in the microfluidic substrate. An exemplary process for carrying this out is further described herein.
At 1806, the substrate is inspected to ensure that wax from step 1804 is loaded properly and that the laminate from step 1802 adheres properly to it. If a substrate does not satisfy either or both of these tests, it is usually discarded. If substrates repeatedly fail either or both of these tests, then the wax dispensing, or laminate application steps, as applicable, are reviewed.
At 1808, a hydrophobic vent membrane is applied to, and heat bonded to, the top of the microfluidic substrate covering at least the one or more vent holes, and on the opposite face of the substrate from the laminate. Edges of the membrane that are in excess of the boundary of the substrate are trimmed.
At 1810, the assembly is inspected to ensure that the hydrophobic vent membrane is bonded well to the microfluidic substrate without heat-clogging the microfluidic channels. If any of the channels is blocked, or if the bond between the membrane and the substrate is imperfect, the assembly is discarded, and, in the case of repeated discard events, the foregoing process step 1808 is reviewed.
At 1812, optionally, a thermally conductive pad layer is applied to the bottom laminate of the cartridge.
At 1814, two label strips are applied to the top of the microfluidic substrate, one to cover the valves, and a second to protect the vent membranes. It would be understood that a single label strip may be devised to fulfill both of these roles.
At 1816, additional labels are printed or applied to show identifying characteristics, such as a barcode #, lot # and expiry date on the cartridge. Preferably one or more of these labels has a space and a writable surface that permits a user to make an identifying annotation on the label, by hand.
Optionally, at 1818, to facilitate transport and delivery to a customer, assembled and labeled cartridges are stacked, and cartridges packed into groups, such as groups of 25, or groups of 10, or groups of 20, or groups of 48 or 50. Preferably the packaging is via an inert and/or moisture-free medium.
Wax Loading in Valves
In general, a valve as shown in, e.g.,
To accomplish those steps, a heated dispenser head can be accurately positioned over the inlet hole of the micro channel in the microfluidic device, and can dispense molten wax drops in volumes as small as 75 nanoliters with an accuracy of 20%. A suitable dispenser is also one that can deposit amounts smaller than 100 nl with a precision of +/−20%. The dispenser should also be capable of heating and maintaining the dispensing temperature of the TRS to be dispensed. For example, it may have a reservoir to hold the solution of TRS. It is also desirable that the dispense head can have freedom of movement at least in a horizontal (x-y) plane so that it can easily move to various locations of a microfluidic substrate and dispense volumes of TRS into valve inlets at such locations without having to be re-set, repositioned manually, or recalibrated in between each dispense operation.
The inlet hole of the microfluidic cartridge, or other microchannel device, is dimensioned in such a way that the droplet of 75 nl can be accurately propelled to the bottom of the inlet hole using, for example, compressed air, or in a manner similar to an inkjet printing method. The microfluidic cartridge is maintained at a temperature above the melting point of the wax thereby permitting the wax to stay in a molten state immediately after it is dispensed. After the drop falls to the bottom of the inlet hole 1901, the molten wax is drawn into the narrow channel by capillary action, as shown in the sequence of views in
PCR Reagent Mixtures
In various embodiments, the sample for introduction into a lane of the microfluidic cartridge can include a PCR reagent mixture comprising a polymerase enzyme and a plurality of nucleotides.
In various embodiments, preparation of a PCR-ready sample for use with an apparatus and cartridge as described herein can include contacting a neutralized polynucleotide sample with a PCR reagent mixture comprising a polymerase enzyme and a plurality of nucleotides (in some embodiments, the PCR reagent mixture can further include a positive control plasmid and a fluorogenic hybridization probe selective for at least a portion of the plasmid).
The PCR-ready sample can be prepared, for example, using the following steps: (1) collect sample in sample collection buffer, (2) transfer entire sample to lysis tube, mix, heat, and incubate for seven minutes, (3) place on magnetic rack, allow beads to separate, aspirate supernatant, (4) add 100 μl of Buffer 1, mix, place on magnetic rack, allow beads to separate, aspirate supernatant, (5) add 10 μl of Buffer 2, mix, place in high temperature heat block for 3 minutes, place on magnetic rack, allow beads to separate, transfer 5 μl supernatant, and (6) Add 5 μl of Buffer 3, transfer 1 to 10 μl of supernatant for PCR amplification and detection.
The PCR reagent mixture can be in the form of one or more lyophilized pellets and the steps by which the PCR-ready sample is prepared can involve reconstituting the PCR pellet by contacting it with liquid to create a PCR reagent mixture solution. In yet another embodiment, each of the PCR lanes may have dried down or lyophilized ASR reagents preloaded such that the user only needs to input prepared polynucleotide sample into the PCR. In another embodiment, the PCR lanes may have only the application-specific probes and primers pre-measured and pre-loaded, and the user inputs a sample mixed with the PCR reagents.
In various embodiments, the PCR-ready sample can include at least one probe that can be selective for a polynucleotide sequence, wherein the steps by which the PCR-ready sample is prepared involve contacting the neutralized polynucleotide sample or a PCR amplicon thereof with the probe. The probe can be a fluorogenic hybridization probe. The fluorogenic hybridization probe can include a polynucleotide sequence coupled to a fluorescent reporter dye and a fluorescence quencher dye.
In various embodiments, the PCR-ready sample further includes a sample buffer.
In various embodiments, the PCR-ready sample includes at least one probe that is selective for a polynucleotide sequence, e.g., the polynucleotide sequence that is characteristic of a pathogen selected from the group consisting of gram positive bacteria, gram negative bacteria, yeast, fungi, protozoa, and viruses.
In various embodiments, the PCR reagent mixture can further include a polymerase enzyme, a positive control plasmid and a fluorogenic hybridization probe selective for at least a portion of the plasmid.
In various embodiments, the probe can be selective for a polynucleotide sequence that is characteristic of an organism, for example any organism that employs deoxyribonucleic acid or ribonucleic acid polynucleotides. Thus, the probe can be selective for any organism. Suitable organisms include mammals (including humans), birds, reptiles, amphibians, fish, domesticated animals, wild animals, extinct organisms, bacteria, fungi, viruses, plants, and the like. The probe can also be selective for components of organisms that employ their own polynucleotides, for example mitochondria. In some embodiments, the probe is selective for microorganisms, for example, organisms used in food production (for example, yeasts employed in fermented products, molds or bacteria employed in cheeses, and the like) or pathogens (e.g., of humans, domesticated or wild mammals, domesticated or wild birds, and the like). In some embodiments, the probe is selective for organisms selected from the group consisting of gram positive bacteria, gram negative bacteria, yeast, fungi, protozoa, and viruses.
In various embodiments, the probe can be selective for a polynucleotide sequence that is characteristic of an organism selected from the group consisting of Staphylococcus spp., e.g., S. epidermidis, S. aureus, Methicillin-resistant Staphylococcus aureus (MRSA), Vancomycin-resistant Staphylococcus; Streptococcus (e.g., α, β or γ-hemolytic, Group A, B, C, D or G) such as S. pyogenes, S. agalactiae; E. faecalis, E. durans, and E. faecium (formerly S. faecalis, S. durans, S. faecium); nonenterococcal group D streptococci, e.g., S. bovis and S. equines; Streptococci viridans, e.g., S. mutans, S. sanguis, S. salivarius, S. mitior, A. milleri, S. constellatus, S. intermedius, and S. anginosus; S. iniae; S. pneumoniae; Neisseria, e.g., N. meningitides, N. gonorrhoeae, saprophytic Neisseria sp; Erysipelothrix, e.g., E. rhusiopathiae; Listeria spp., e.g., L. monocytogenes, rarely L. ivanovii and L. seeligeri; Bacillus, e.g., B. anthracis, B. cereus, B. subtilis, B. subtilus niger, B. thuringiensis; Nocardia asteroids; Legionella, e.g., L. pneumonophilia, Pneumocystis, e.g., P. carinii; Enterobacteriaceae such as Salmonella, Shigella, Escherichia (e.g., E. coli, E. coli O157:H7); Klebsiella, Enterobacter, Serratia, Proteus, Morganella, Providencia, Yersinia, and the like, e.g., Salmonella, e.g., S. typhi S. paratyphi A, B (S. schottmuelleri), and C (S. hirschfeldii), S. dublin S. choleraesuis, S. enteritidis, S. typhimurium, S. heidelberg, S. newport, S. infantis, S. agona, S. montevideo, and S. saint-paul; Shigella e.g., subgroups: A, B, C, and D, such as S. flexneri, S. sonnei, S. boydii, S. dysenteriae; Proteus (P. mirabilis, P. vulgaris, and P. myxofaciens), Morganella (M. morganii); Providencia (P. rettgeri, P. alcalifaciens, and P. stuartii); Yersinia, e.g., Y. pestis, Y. enterocolitica, Haemophilus, e.g., H. influenzae, H. parainfluenzae H. aphrophilus, H. ducreyi; Brucella, e.g., B. abortus, B. melitensis, B. suis, B. canis; Francisella, e.g., F. tularensis; Pseudomonas, e.g., P. aeruginosa, P. paucimobilis, P. putida, P. fluorescens, P. acidovorans, Burkholderia (Pseudomonas) pseudomallei, Burkholderia mallei, Burkholderia cepacia and Stenotrophomonas maltophilia; Campylobacter, e.g., C. fetus, C. jejuni, C. pylori (Helicobacter pylori); Vibrio, e.g., V. cholerae, V. parahaemolyticus, V. mimicus, V. alginolyticus, V. hollisae, V. vulnificus, and the nonagglutinable vibrios; Clostridia, e.g., C. perfringens, C. tetani, C. difficile, C. botulinum; Actinomyces, e.g., A. israelii; Bacteroides, e.g., B. fragilis, B. thetaiotaomicron, B. distasonis, B. vulgatus, B. ovatus, B. caccae, and B. merdae; Prevotella, e.g., P. melaninogenica; genus Fusobacterium; Treponema, e.g. T. pallidum subspecies endemicum, T. pallidum subspecies pertenue, T. carateum, and T. pallidum subspecies pallidum; genus Borrelia, e.g., B. burgdorferi; genus Leptospira; Streptobacillus, e.g., S. moniliformis; Spirillum, e.g., S. minus; Mycobacterium, e.g., M. tuberculosis, M. bovis, M. africanum, M. avium M. intracellulare, M. kansasii, M. xenopi, M. marinum, M. ulcerans, the M. fortuitum complex (M. fortuitum and M. chelonei), M. leprae, M. asiaticum, M. chelonei subsp. abscessus, M. fallax, M. fortuitum, M. malmoense, M. shimoidei, M. simiae, M. szulgai, M. xenopi; Mycoplasma, e.g., M. hominis, M. orale, M. salivarium, M. fermentans, M. pneumoniae, M. bovis, M. tuberculosis, M. avium, M. leprae; Mycoplasma, e.g., M. genitalium; Ureaplasma, e.g., U. urealyticum; Trichomonas, e.g., T. vaginalis; Cryptococcus, e.g., C. neoformans; Histoplasma, e.g., H. capsulatum; Candida, e.g., C. albicans; Aspergillus sp; Coccidioides, e.g., C. immitis; Blastomyces, e.g. B. dermatitidis; Paracoccidioides, e.g., P. brasiliensis; Penicillium, e.g., P. marneffei; Sporothrix, e.g., S. schenckii; Rhizopus, Rhizomucor, Absidia, and Basidiobolus; diseases caused by Bipolaris, Cladophialophora, Cladosporium, Drechslera, Exophiala, Fonsecaea, Phialophora, Xylohypha, Ochroconis, Rhinocladiella, Scolecobasidium, and Wangiella; Trichosporon, e.g., T. beigelii; Blastoschizomyces, e.g., B. capitatus; Plasmodium, e.g., P. falciparum, P. vivax, P. ovale, and P. malariae; Babesia sp; protozoa of the genus Trypanosoma, e.g., T. cruzi; Leishmania, e.g., L. donovani, L. major L. tropica, L. mexicana, L. braziliensis, L. viannia braziliensis; Toxoplasma, e.g., T. gondii; Amoebas of the genera Naegleria or Acanthamoeba; Entamoeba histolytica; Giardia lamblia; genus Cryptosporidium, e.g., C. parvum; Isospora belli; Cyclospora cayetanensis; Ascaris lumbricoides; Trichuris trichiura; Ancylostoma duodenale or Necator americanus; Strongyloides stercoralis Toxocara, e.g., T. canis, T. cati; Baylisascaris, e.g., B. procyonis; Trichinella, e.g., T. spiralis; Dracunculus, e.g., D. medinensis; genus Filarioidea; Wuchereria bancrofti; Brugia, e.g., B. malayi, or B. timori; Onchocerca volvulus; Loa boa; Dirofilaria immitis; genus Schistosoma, e.g., S. japonicum, S. mansoni, S. mekongi, S. intercalatum, S. haematobium; Paragonimus, e.g., P. Westermani, P. Skriabini; Clonorchis sinensis; Fasciola hepatica; Opisthorchis sp; Fasciolopsis buski; Diphyllobothrium latum; Taenia, e.g., T. saginata, T. solium; Echinococcus, e.g., E. granulosus, E. multilocularis; Picornaviruses, rhinoviruses echoviruses, coxsackieviruses, influenza virus; paramyxoviruses, e.g., types 1, 2, 3, and 4; adnoviruses; Herpesviruses, e.g., HSV-1 and HSV-2; varicella-zoster virus; human T-lymphotrophic virus (type I and type II); Arboviruses and Arenaviruses; Togaviridae, Flaviviridae, Bunyaviridae, Reoviridae; Flavivirus; Hantavirus; Viral encephalitis (alphaviruses [e.g., Venezuelan equine encephalitis, eastern equine encephalitis, western equine encephalitis]); Viral hemorrhagic fevers (filoviruses [e.g., Ebola, Marburg] and arenaviruses [e.g., Lassa, Machupo]); Smallpox (variola); retroviruses e.g., human immunodeficiency viruses 1 and 2; human papillomavirus [HPV] types 6, 11, 16, 18, 31, 33, and 35.
In various embodiments, the probe can be selective for a polynucleotide sequence that is characteristic of an organisms selected from the group consisting of Pseudomonas aeruginosa, Proteus mirabilis, Klebsiella oxytoca, Klebsiella pneumoniae, Escherichia coli, Acinetobacter Baumannii, Serratia marcescens, Enterobacter aerogenes, Enterococcus faecium, vancomycin-resistant enterococcus (VRE), Staphylococcus aureus, methecillin-resistant Staphylococcus aureus(MRSA), Streptococcus viridans, Listeria monocytogenes, Enterococcus spp., Streptococcus Group B, Streptococcus Group C, Streptococcus Group G, Streptococcus Group F, Enterococcus faecalis, Streptococcus pneumoniae, Staphylococcus epidermidis, Gardenerella vaginalis, Micrococcus sps., Haemophilus influenzae, Neisseria gonorrhoeee, Moraxella catarrahlis, Salmonella sps., Chlamydia trachomatis, Peptostreptococcus productus, Peptostreptococcus anaerobius, Lactobacillus fermentum, Eubacterium lentum, Candida glabrata, Candida albicans, Chlamydia spp., Camplobacter spp., Salmonella spp., smallpox (variola major), Yersina Pestis, Herpes Simplex Virus I (HSV I), and Herpes Simplex Virus II (HSV II).
In various embodiments, the probe can be selective for a polynucleotide sequence that is characteristic of Group B Streptococcus.
In various embodiments, a method of carrying out PCR on a sample can further include one or more of the following steps: heating the biological sample in the microfluidic cartridge; pressurizing the biological sample in the microfluidic cartridge at a pressure differential compared to ambient pressure of between about 20 kilopascals and 200 kilopascals, or in some embodiments, between about 70 kilopascals and 110 kilopascals.
In some embodiments, the method for sampling a polynucleotide can include the steps of: placing a microfluidic cartridge containing a PCR-ready sample in a receiving bay of a suitably configured apparatus; carrying out PCR on the sample under thermal cycling conditions suitable for creating PCR amplicons from the neutralized polynucleotide in the sample, the PCR-ready sample comprising a polymerase enzyme, a positive control plasmid, a fluorogenic hybridization probe selective for at least a portion of the plasmid, and a plurality of nucleotides; contacting the neutralized polynucleotide sample or a PCR amplicon thereof with the at least one fluorogenic probe that is selective for a polynucleotide sequence, wherein the probe is selective for a polynucleotide sequence that is characteristic of an organism selected from the group consisting of gram positive bacteria, gram negative bacteria, yeast, fungi, protozoa, and viruses; and detecting the fluorogenic probe, the presence of the organism for which the one fluorogenic probe is selective is determined.
Carrying out PCR on a PCR-ready sample can additionally include: independently contacting each of the neutralized polynucleotide sample and a negative control polynucleotide with the PCR reagent mixture under thermal cycling conditions suitable for independently creating PCR amplicons of the neutralized polynucleotide sample and PCR amplicons of the negative control polynucleotide; and/or contacting the neutralized polynucleotide sample or a PCR amplicon thereof and the negative control polynucleotide or a PCR amplicon thereof with at least one probe that is selective for a polynucleotide sequence.
In various embodiments, a method of using the apparatus and cartridge described herein can further include one or more of the following steps: determining the presence of a polynucleotide sequence in the biological sample, the polynucleotide sequence corresponding to the probe, if the probe is detected in the neutralized polynucleotide sample or a PCR amplicon thereof; determining that the sample was contaminated if the probe is detected in the negative control polynucleotide or a PCR amplicon thereof; and/or in some embodiments, wherein the PCR reagent mixture further comprises a positive control plasmid and a plasmid probe selective for at least a portion of the plasmid, the method further including determining that a PCR amplification has occurred if the plasmid probe is detected.
Kit
In various embodiments, the microfluidic cartridge as described herein can be provided in the form of a kit, wherein the kit can include a microfluidic cartridge, and a liquid transfer member (such as a syringe or a pipette). In various embodiments, the kit can further include instructions to employ the liquid transfer member to transfer a sample containing extracted nucleic acid from a sample container via a sample inlet to the microfluidic network on the microfluidic cartridge. In some embodiments, the microfluidic cartridge and the liquid transfer member can be sealed in a pouch with an inert gas.
Typically when transferring a sample from liquid dispenser, such as a pipette tip, to an inlet on the microfluidic cartridge, a volume of air is simultaneously introduced into the microfluidic network, the volume of air being between about 0.5 mL and about 5 mL. Presence of air in the microfluidic network, however, is not essential to operation of the cartridge described herein.
In various embodiments, the kit can further include at least one computer-readable label on the cartridge. The label can include, for example, a bar code, a radio frequency tag or one or more computer-readable characters. When used in conjunction with a similar computer-readable label on a sample container, such as a vial or a pouch, matching of diagnostic results with sample is thereby facilitated.
In some embodiments, a sample identifier of the apparatus described elsewhere herein is employed to read a label on the microfluidic cartridge and/or a label on the biological sample.
Heater Unit
An exemplary heater unit 2020 is shown in
Shown in
Area 2044 is configured to accept a microfluidic cartridge in a single orientation. Therefore area 2044 can be equipped with a registration member such as a mechanical key that prevents a user from placing a cartridge into receiving bay 2014 in the wrong configuration. Shown in
Also shown in
In the embodiment of
Other non-essential features of heater unit 2020 are as follows. One or more air vents 2052 can be situated on one or more sides (such as front, rear, or flanking) or faces (such as top or bottom) of heater unit 2020, to permit excess heat to escape, when heaters underneath receiving bay 2014, are in operation. The configuration of air vents in
Heater unit 2020 may further comprise one or more guiding members 2047 that facilitate inserting the heater unit into an apparatus as further described herein, for an embodiment in which heater unit 2020 is removable by a user. Heater unit is advantageously removable because it permits system 2000 to be easily reconfigured for a different type of analysis, such as employing a different cartridge with a different registration member and/or microfluidic network, in conjunction with the same or a different sequence of processing operations. In other embodiments, heater unit 2020 is designed to be fixed and only removable, e.g., for cleaning, replacement, or maintenance, by the manufacturer or an authorized maintenance agent, and not routinely by the user. Guiding members 2047 may perform one or more roles of ensuring that the heater unit is aligned correctly in the apparatus, and ensuring that the heater unit makes a tight fit and does not significantly move during processing and analysis of a sample, or during transport of the apparatus.
Guiding members shown in the embodiment of
Also shown in
In particular and not shown in
Heater Configurations to Ensure Uniform Heating of a Region
The microfluidic substrates described herein are configured to accept heat from a contact heat source, such as found in a heater unit described herein. The heater unit typically comprises a heater board or heater chip that is configured to deliver heat to specific regions of the microfluidic substrate, including but not limited to one or more microfluidic components, at specific times. For example, the heat source is configured so that particular heating elements are situated adjacent to specific components of the microfluidic network on the substrate. In certain embodiments, the apparatus uniformly controls the heating of a region of a microfluidic network. In an exemplary embodiment, multiple heaters can be configured to simultaneously and uniformly heat a region, such as the PCR reaction chamber, of the microfluidic substrate. The term heater unit, as used herein, may be used interchangeably to describe either the heater board or an item such as shown in
Referring to
Referring to
It would be understood by one of ordinary skill in the art that still other configurations of one or more heater(s) situated about a PCR reaction chamber are consistent with the methods and apparatus described herein. For example, a ‘long’ side of the reaction chamber can be configured to be heated by two or more heaters. Specific orientations and configurations of heaters are used to create uniform zones of heating even on substrates having poor thermal conductivity because the poor thermal conductivity of glass, or quartz, polyimide, FR4, ceramic, or fused silica substrates is utilized to help in the independent operation of various microfluidic components such as valves and independent operation of the various PCR lanes. It would be further understood by one of ordinary skill in the art, that the principles underlying the configuration of heaters around a PCR reaction chamber are similarly applicable to the arrangement of heaters adjacent to other components of the microfluidic cartridge, such as actuators, valves, and gates.
Generally, the heating of microfluidic components, such as a PCR reaction chamber, is controlled by passing currents through suitably configured microfabricated heaters. Under control of suitable circuitry, the lanes of a multi-lane cartridge can then be controlled independently of one another. This can lead to a greater energy efficiency of the apparatus, because not all heaters are heating at the same time, and a given heater is receiving current for only that fraction of the time when it is required to heat. Control systems and methods of controllably heating various heating elements are further described in U.S. patent application Ser. No. 11/940,315, filed Nov. 14, 2007 and entitled “Heater Unit for Microfluidic Diagnostic System”.
In certain embodiments, each heater has an associated temperature sensor. In the embodiment of
In order to reduce the number of sensor or heater elements required to control a PCR heater, the heaters may be used to sense as well as heat, and thereby obviate the need to have a separate dedicated sensor for each heater. In another embodiment, each of the four heaters may be designed to have an appropriate wattage, and connect the four heaters in series or in parallel to reduce the number of electronically-controllable elements from four to just one, thereby reducing the burden on the associated electronic circuitry.
The configuration for uniform heating, shown in
Each heater can be independently controlled by a processor and/or control circuitry used in conjunction with the apparatus described herein.
The configuration for uniform heating, shown in
Heater Multiplexing (Under Software Control)
Another aspect of the heater unit described herein, relates to a control of heat within the system and its components. The method leads to a greater energy efficiency of the apparatus described herein, because not all heaters are heating at the same time, and a given heater is receiving current for only part of the time.
Generally, the heating of microfluidic components, such as a PCR reaction chamber, is controlled by passing currents through suitably configured microfabricated heaters. The heating can be further controlled by periodically turning the current on and off with varying pulse width modulation (PWM), wherein pulse width modulation refers to the on-time/off-time ratio for the current. The current can be supplied by connecting a microfabricated heater to a high voltage source (for example, 30 V), which can be gated by the PWM signal. In some embodiments, the device includes 48 PWM signal generators. Operation of a PWM generator includes generating a signal with a chosen, programmable, period (the end count) and a particular granularity. For instance, the signal can be 4000 μs (micro-seconds) with a granularity of 1 μs, in which case the PWM generator can maintain a counter beginning at zero and advancing in increments of 1 μs until it reaches 4000 μs, when it returns to zero. Thus, the amount of heat produced can be adjusted by adjusting the end count. A high end count corresponds to a greater length of time during which the microfabricated heater receives current and therefore a greater amount of heat produced. It would be understood that the granularity and signal width can take values other than those provided here without departing from the principles described herein.
Fluorescence Detection System, Including Lenses and Filters, and Multiple Parallel Detection for a Multi-Lane Cartridge
The detection system herein is configured to monitor fluorescence coming from one or more species involved in a biochemical reaction. The system can be, for example, an optical detector having a light source that selectively emits light in an absorption band of a fluorescent dye, and a light detector that selectively detects light in an emission band of the fluorescent dye, wherein the fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof, as further described elsewhere herein. Alternatively, the optical detector can include a bandpass-filtered diode that selectively emits light in the absorption band of the fluorescent dye and a bandpass filtered photodiode that selectively detects light in the emission band of the fluorescent dye. For example, the optical detector can be configured to independently detect a plurality of fluorescent dyes having different fluorescent emission spectra, wherein each fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof. For example, the optical detector can be configured to independently detect a plurality of fluorescent dyes at a plurality of different locations of, for example, a microfluidic substrate, wherein each fluorescent dye corresponds to a fluorescent polynucleotide probe or a fragment thereof. The detector further has potential for 2, 3 or 4 color detection and is controlled by software, preferably custom software, configured to sample information from the detector.
The detection system described herein is capable of detecting a fluorescence signal from nanoliter scale PCR reactions. Advantageously, the detector is formed from inexpensive components, having no moving parts. The detector can be configured to couple to a microfluidic cartridge as further described herein, and can also be part of a pressure application system, such as a sliding lid on an apparatus in which the detector is situated, that keeps the cartridge in place.
Each of the detection systems multiplexed in the assembly of
Detection Sensitivity, Time Constant and Gain
A typical circuit that includes a detector as described herein includes, in series, a preamplifier, a buffer/inverter, a filter, and a digitizer. Sensitivity is important so that high gain is very desirable. In one embodiment of the preamplifier, a very large, for example 100 GΩ, resistor is placed in parallel with the diode. Other values of a resistor would be consistent with the technology herein: such values typically fall in the range 0.5-100 GΩ, such as 1-50 GΩ, or 2-10 GΩ. An exemplary pre-amplifier circuit configured in this way is shown in
The
A resistor-capacitor circuit in
As the resistance value for R25 is very high (˜100 GΩ), the manner of assembly of this resistor on the optics board is important for the overall efficiency of the circuit. Effective cleaning of the circuit during assembly and before use is important to achieve an optimal time-constant and gain for the optics circuit.
It is also important to test each photo-diode that is used, because many do not perform according to specification.
Sensitivity and Aperturing
The LED light passes through a filter before passing through the sample in the microfluidic channel (as described herein, typically 300 μl deep). This is a very small optical path-length for the light in the sample. The generated fluorescence then also goes through a second filter, and into a photo-detector. Ultimately, then, the detector must be capable of detecting very little fluorescence. Various aspects of the detector configuration can improve sensitivity, however.
The illumination optics can be designed so that the excitation light falling on the PCR reactor is incident along an area that is similar to the shape of the reactor. As the reactor is typically long and narrow, the illumination spot should be long and narrow, i.e., extended, as well. The length of the spot can be adjusted by altering a number of factors, including: the diameter of the bore where the LED is placed (the tube that holds the filter and lens has an aperturing effect); the distance of the LED from the PCR reactor; and the use of proper lens at the right distance in between. As the width of the beam incident on the reactor is determined by the bore of the optical element (approximately 6 mm in diameter), it is typical to use an aperture (a slit having a width approximately equal to the width of the reactor, and a length equal to the length of the detection volume) to make an optimal illumination spot. A typical spot, then, is commensurate with the dimensions of a PCR reaction chamber, for example 1.5 mm wide by 7 mm long.
The optimal spot size and intensity is importantly dependent on the ability to maintain the correct position of the LED's with respect to the center of the optical axis. Special alignment procedures and checks can be utilized to optimize this. The different illuminations can also be normalized with respect to each other by adjusting the power current through each of the LED's or adjusting the fluorescence collection time (the duration for which the photodetector is on before measuring the voltage) for each detection spot. It is also important to align the detectors with the axis of the micro-channels.
The aperturing is also important for successful fluorescence detection because as the cross-sectional area of the incident beam increases in size, so the background fluorescence increases, and the fluorescence attributable only to the molecules of interest (PCR probes) gets masked. Thus, as the beam area increases, the use of an aperture increases the proportion of collected fluorescence that originates only from the PCR reactor. Note that the aperture used in the detector herein not only helps collect fluorescence only from the reaction volume but it correspondingly adjusts the excitation light to mostly excite the reaction volume. The excitation and emission aperture is, of course, the same.
Based on a typical geometry of the optical exctiation and emission system and aperturing, show spot sizes as small as 0.5 mm by 0.5 mm and as long as 8 mm×1.5 mm have been found to be effective. By using a long detector (having an active area 6 mm by 1 mm) and proper lensing, the aperture design can extend the detection spot to as long as 15-20 mm, while maintaining a width of 1-2 mm using an aperture. Correspondingly, the background fluorescence decreases as the spot size is decreased, thereby increasing the detection sensitivity.
Use of Detection System to Measure/Detect Fluid in PCR Chamber
The fluorescence detector is sensitive enough to be able to collect fluorescence light from a PCR chamber of a microfluidic substrate. The detector can also be used to detect the presence of liquid in the chamber, a measurement that provides a determination of whether or not to carry out a PCR cycle for that chamber. For example, in a multi-sample cartridge, not all chambers will have been loaded with sample; for those that are not, it would be unnecessary to apply a heating protocol thereto. One way to determine presence or absence of a liquid is as follows. A background reading is taken prior to filling the chamber with liquid. Another reading is taken after microfluidic operations have been performed that should result in filling the PCR chamber with liquid. The presence of liquid alters the fluorescence reading from the chamber. A programmable threshold value can be used to tune an algorithm programmed into a processor that controls operation of the apparatus as further described herein (for example, the second reading has to exceed the first reading by 20%). If the two readings do not differ beyond the programmed margin, the liquid is deemed to not have entered the chamber, and a PCR cycle is not initiated for that chamber. Instead, a warning is issued to a user.
Exemplary Electronics and Software
The heater unit described herein can be controlled by various electronics circuitry, itself operating on receipt of computer-controlled instructions.
In one embodiment, the Card Engine electronics module 2116 is a commercial, off the shelf “single board computer” containing a processor, memory, and flash memory for operating system storage.
The optional LCD+Touchscreen electronics module 2110 is a user interface, for example, driven through a touchscreen, such as a 640 pixel by 480 pixel 8 inch LCD and 5-wire touchscreen.
The Compact Flash electronics module 2118 is, for example, a 256 megabyte commercial, off the shelf, compact flash module for application and data storage. Other media are alternatively usable, such as USB-drive, smart media card, memory stick, and smart data-card having the same or other storage capacities.
The Backplane electronics module 2112 is a point of connection for the removable heater assembly 2102. Bare PC boards with two connectors are sufficient to provide the necessary level of connectivity.
The Control Board electronics module 2114 supports peripherals to the Card Engine electronics module 2116. In one embodiment, the peripherals include such devices as a USB host+slave or hub, a USB CDROM interface, serial ports, and ethernet ports. The Control Board 2114 can include a power monitor with a dedicated processor. The Control Board may also include a real time clock. The Control Board may further include a speaker. The Control Board 2114 also includes a CPLD to provide SPI access to all other modules and programming access to all other modules. The Control Board includes a programmable high voltage power supply. The Control Board includes a Serial-Deserializer interface to the LCD+Touchscreen electronics module 2110 and to the Optical Detection Unit electronics module 2108. The Control Board also includes module connectors.
In the exemplary embodiment, the optical detection unit electronics module 2108 contains a dedicated processor. The optical detection unit 2108 contains a serializer-deserializer interface. The optical detection unit 2108 contains LED drivers. The optical detection unit also contains high gain-low noise photodiode amplifiers. The optical detection unit can have power monitoring capability. The optical detection unit can also be remotely reprogrammable.
The Heater Board electronics module 2104 is preferably a glass heater board. The Heater Board has PCB with bonding pads for glass heater board and high density connectors.
In one embodiment, the heater mux (‘multiplex’) board electronics module 2106 has 24 high-speed ADC, 24 precision current sources, and 96 optically isolated current drivers for heating. The heater mux board has the ability to time-multiplex heating/measurement. The heater mux board has multiplexer banks to multiplex inputs to ADC, and to multiplex current source outputs. The heater mux board has a FPGA with a soft processor core and SDRAM. The heater mux board has a Power Monitor with a dedicated processor. The Heater Mux Board can be remotely reprogrammable.
In another embodiment, control electronics can be spread over four different circuit board assemblies. These include the MAIN board: Can serve as the hub of the Analyzer control electronics and manages communication and control of the other various electronic subassemblies. The main board can also serve as the electrical and communications interface with the external world. An external power supply (12V DC/10 A; UL certified) can be used to power the system. The unit can communicate via 5 USB ports, a serial port and an Ethernet port. Finally, the main board can incorporate several diagnostic/safety features to ensure safe and robust operation of the Analyzer.
MUX Board: Upon instruction from the main board, the MUX board can perform all the functions typically used for accurate temperature control of the heaters and can coordinate the collection of fluorescence data from the detector board.
LCD Board: Can contain the typical control elements to light up the LCD panel and interpret the signals from the touch sensitive screen. The LCD/touch screen combination can serve as a mode of interaction with the user via a Graphical User Interface.
Detector Board: Can house typical control and processing circuitry that can be employed to collect, digitize, filter, and transmit the data from the fluorescence detection modules.
Certain software can be executed in each electronics module. The Control Board Electronics Module executes, for example, Control Board Power Monitor software. The Card Engine electronics module executes an operating system, graphical user interface (GUI) software, an analyzer module, and an application program interface (api). The Optical Detection Unit electronics module executes an optics software module. The Heater Mux Board electronics module executes dedicated Heater Mux software, and Heater Mux Power Monitor software. Each of the separate instances of software can be modular and under a unified control of, for example, driver software.
The exemplary electronics can use Linux, UNIX, Windows, or MacOS, including any version thereof, as the operating system. The operating system is preferably loaded with drivers for USB, Ethernet, LCD, touchscreen, and removable media devices such as compact flash. Miscellaneous programs for configuring the Ethernet interface, managing USB connections, and updating via CD-ROM can also be included.
In the embodiment of
The API provides uniform access to the analyzer module driver. The API is responsible for error trapping, and interrupt handling. The API is typically programmed to be thread safe.
The GUI software can be based on a commercial, off-the-shelf PEG graphics library. The GUI can use the API to coordinate the self-test of optical detection unit and heater assembly. The GUI starts, stops, and monitors test progress. The GUI can also implement an algorithm to arrive on diagnosis from fluorescence data. The GUI provides access control to unit and in some embodiments has an HIS/LIS interface.
The Control Board Power Monitor software monitors power supplies, current and voltage, and signals error in case of a fault.
The Optics Software performs fluorescence detection which is precisely timed to turn on/off of LED with synchronous digitization of the photodetector outputs. The Optics Software can also monitor power supply voltages. The Optics Software can also have self test ability.
The Heater Mux Module software implements a “protocol player” which executes series of defined “steps” where each “step” can turn on sets of heaters to implement a desired microfluidic action. The Heater Mux Module software also has self test ability. The Heater Mux Module software contains a fuzzy logic temperature control algorithm.
The Heater Mux Power Monitor software monitors voltage and current levels. The Heater Mux Power Monitor software can participate in self-test, synchronous, monitoring of the current levels while turning on different heaters.
The following are exemplary aspects of various parts and functions of the system described herein.
Additional embodiments of a cartridge are found in U.S. patent application Ser. No. 11/940,310, entitled “Microfluidic Cartridge and Method of Making Same”, and filed on even date herewith, the specification of which is incorporated herein by reference.
Additional embodiments of heater units and arrays are described in U.S. patent application Ser. No. 11/940,315, entitled “Heater Unit for Microfluidic Diagnostic System” and filed on even date herewith, the specification of which is incorporated herein by reference in its entirety.
Further description of suitably configured detectors are described in U.S. patent application Ser. No. 11/940,321, filed on Nov. 14, 2007 and entitled “Fluorescence Detector for Microfluidic Diagnostic System”, incorporated herein by reference.
This non-limiting example describes pictorially, various embodiments of an apparatus, showing incorporation of a heater unit and a microfluidic cartridge operated on by the heater unit.
An exemplary heater substrate,
Pictures of an exemplary Mux board and assembled heater unit are shown in
In various embodiments, the operation of a PWM generator can also include a programmable start count in addition to the aforementioned end count and granularity. In such embodiments, multiple PWM generators can produce signals that can be selectively non-overlapping (e.g., by multiplexing the on-time of the various heaters) such that the current capacity of the high voltage power is not exceeded. Multiple heaters can be controlled by different PWM signal generators with varying start and end counts. The heaters can be divided into banks, whereby a bank defines a group of heaters of the same start count. For example, 36 PWM generators can be grouped into six different banks, each corresponding to a certain portion of the PWM cycle (500 ms for this example). The end count for each PWM generator can be selectively programmed such that not more than six heaters will be on at any given time. A portion of a PWM cycle can be selected as dead time (count 3000 to 4000 for this example) during which no heating takes place and sensitive temperature sensing circuits can use this time to sense the temperature. The table below represents a PWM cycle for the foregoing example:
This non-limiting example describes pictorially, various embodiments of a detection system integrated into a force member, in an apparatus for carrying out diagnostics on microfluidic samples.
Thermal interface: the cartridge bottom can have a layer of mechanically compliant heat transfer laminate that can enable thermal contact between the microfluidic substrate and the microheater substrate of the heater module. A minimal pressure of 1 psi can be employed for reliable operation of the thermal valves, gate and pumps present in the microfluidic cartridge.
Mechanicals and assembly: the Analyzer can have a simple mechanical frame to hold the various modules in alignment. The optics module can be placed in rails for easy opening and placement of cartridges in the Analyzer and error-free alignment of the optics upon closing. The heater/sensor module can be also placed on rails or similar guiding members for easy removal and insertion of the assembly.
Slider: the slider of the Analyzer can house the optical detection system as well as the mechanical assembly that can enables the optics jig to press down on the cartridge when the handle of the slider is turned down onto the analyzer. The optics jig can be suspended from the case of the slider at 4 points. Upon closing the slider and turning the handle of the analyzer down, 4 cams can turn to push down a plate that presses on 4 springs. On compression, the springs can deliver approximately 50 lb on the optical block. See
The bottom surface of the optics block can be made flat to within 100 microns, typically within 25 microns, and this flat surface can press upon the compliant (shore hardness approximately 50-70) label (approximately 1.5 mm thick under no compression) of the cartridge making the pressure more or less uniform over the cartridge. An optional lock-in mechanism can also be incorporated to prevent the slider from being accidentally knocked-off while in use.
An exemplary optics board is shown schematically in
The power board systems include: a +12V input; and +3.3V, +3.6V, +5V, and −5V outputs, configured as follows: the +3.3V output contains a linear regulator, is used to power the LVDS interface, should maintain a +/−5% accuracy, and supply an output current of 0.35 A; the +3.6V output contains a linear regulator, is used to power the MSP430, should maintain a +/−5% accuracy, and supply an output current of 0.35 A; the +5V output contains a linear regulator, is used to power the plus rail for op-amps, should maintain a +/−5% accuracy, and supply an output current of 0.35 A; the −5V output receives its power from the +5V supply, has a mV reference, is used to power the minus rail for op-amps and for the photo-detector bias, should maintain a +/−1% voltage accuracy, and supply an output current of 6.25 mA+/−10%. Additionally, the power board has an 80 ohm source resistance, and the main board software can enable/disable the regulator outputs.
The main board interface uses a single channel of the LVDS standard to communicate between boards. This takes place using SPI signaling over the LVDS interface which is connected to the main SPI port of the control processor. The interface also contains a serial port for in-system programming.
The optical detection system of
During assembly of the various components on to the PC board, such as may occur on a production line, there are the following considerations. The extremely high impedance of the photo-detection circuit means that a rigorous cleaning procedure must be employed. Such a procedure may include, for example: After surface mount components are installed, the boards are washed on a Weskleen and blow dried upon exiting conveyor. The belt speed can be set at 20-30. The boards are soaked in an alcohol bath for approximately 3 minutes, then their entire top and bottom surfaces are scrubbed using a clean, soft bristle brush. The boards are baked in a 105° F. (40° C.) oven for 30 minutes to dry out all components.
After all the components are installed: the soldered areas of the boards can be hand wash using deionized water and a soft bristle brush. The same soldered areas can be hand washed using alcohol and a soft bristle brush. The boards are allowed to air dry. Once the board is cleaned, the optical circuitry must be conformal coated to keep contaminates out.
The foregoing description is intended to illustrate various aspects of the present technology. It is not intended that the examples presented herein limit the scope of the present technology. The technology now being fully described, it will be apparent to one of ordinary skill in the art that many changes and modifications can be made thereto without departing from the spirit or scope of the appended claims.
This application is a divisional of U.S. patent application Ser. No. 16/914,109, filed Jun. 26, 2020, which is a continuation of U.S. patent application Ser. No. 16/787,977, filed Feb. 11, 2020, which is a continuation of U.S. patent application Ser. No. 14/796,239, filed Jul. 10, 2015, which is a continuation of U.S. patent application Ser. No. 13/692,929, filed Dec. 3, 2012 and issued as U.S. Pat. No. 9,080,207 on Jul. 14, 2015, which is a continuation of U.S. patent application Ser. No. 13/035,725, filed Feb. 25, 2011, issued as U.S. Pat. No. 8,323,900 on Dec. 4, 2012, which is a continuation of U.S. patent application Ser. No. 11/985,577, filed Nov. 14, 2007, issued as U.S. Pat. No. 7,998,708 on Aug. 16, 2011, which is a continuation-in-part of U.S. patent application Ser. No. 11/728,964, filed Mar. 26, 2007, issued as U.S. Pat. No. 9,040,288 on May 26, 2015, which claims the benefit of U.S. Provisional Patent Application No. 60/786,007, filed Mar. 24, 2006, and U.S. Provisional Patent Application No. 60/859,284, filed Nov. 14, 2006. U.S. patent application Ser. No. 11/985,577 claims the benefit of U.S. Provisional Patent Application No. 60/859,284, filed Nov. 14, 2006, and U.S. Provisional Patent Application No. 60/959,437, filed Jul. 13, 2007. The disclosures of U.S. patent application Ser. No. 13/692,929, U.S. patent application Ser. No. 13/035,725, U.S. patent application Ser. No. 11/985,577, U.S. patent application Ser. No. 11/728,964, U.S. Provisional Patent Application No. 60/859,284, and U.S. Provisional Patent Application No. 60/959,437 are considered part of the disclosure of this application, and are incorporated by reference herein in their entirety.
Number | Name | Date | Kind |
---|---|---|---|
D189404 | Nicolle | Dec 1960 | S |
3050239 | Williams | Aug 1962 | A |
3905772 | Hartnett et al. | Sep 1975 | A |
3985649 | Eddelman | Oct 1976 | A |
4018089 | Dzula et al. | Apr 1977 | A |
4018652 | Lanham et al. | Apr 1977 | A |
4038192 | Serur | Jul 1977 | A |
4055395 | Honkawa et al. | Oct 1977 | A |
D249706 | Adamski | Sep 1978 | S |
4139005 | Dickey | Feb 1979 | A |
D252157 | Kronish et al. | Jun 1979 | S |
D252341 | Thomas | Jul 1979 | S |
D254687 | Fadler et al. | Apr 1980 | S |
4212744 | Oota | Jul 1980 | A |
D261033 | Armbruster | Sep 1981 | S |
D261173 | Armbruster | Oct 1981 | S |
4301412 | Hill et al. | Nov 1981 | A |
4439526 | Columbus | Mar 1984 | A |
4457329 | Werley et al. | Jul 1984 | A |
4466740 | Kano et al. | Aug 1984 | A |
4472357 | Levy et al. | Sep 1984 | A |
4504582 | Swann | Mar 1985 | A |
4522786 | Ebersole | Jun 1985 | A |
D279817 | Chen et al. | Jul 1985 | S |
D282208 | Lowry | Jan 1986 | S |
4599315 | Terasaki et al. | Jul 1986 | A |
4612873 | Eberle | Sep 1986 | A |
4612959 | Costello | Sep 1986 | A |
D288478 | Carlson et al. | Feb 1987 | S |
4647432 | Wakatake | Mar 1987 | A |
4654127 | Baker et al. | Mar 1987 | A |
4673657 | Christian | Jun 1987 | A |
4678752 | Thorne et al. | Jul 1987 | A |
4683195 | Mullis et al. | Jul 1987 | A |
4683202 | Mullis | Jul 1987 | A |
4698302 | Whitehead et al. | Oct 1987 | A |
D292735 | Lovborg | Nov 1987 | S |
4720374 | Ramachandran | Jan 1988 | A |
4724207 | Hou et al. | Feb 1988 | A |
4795698 | Owen et al. | Jan 1989 | A |
4798693 | Mase et al. | Jan 1989 | A |
4800022 | Leonard | Jan 1989 | A |
4827944 | Nugent | May 1989 | A |
4841786 | Schulz | Jun 1989 | A |
D302294 | Hillman | Jul 1989 | S |
4855110 | Marker et al. | Aug 1989 | A |
4871779 | Killat et al. | Oct 1989 | A |
4889818 | Gelfand et al. | Dec 1989 | A |
4895650 | Wang | Jan 1990 | A |
4902624 | Columbus et al. | Feb 1990 | A |
4914710 | Ward et al. | Apr 1990 | A |
4919829 | Gates et al. | Apr 1990 | A |
4921809 | Schiff et al. | May 1990 | A |
4935342 | Seligson et al. | Jun 1990 | A |
4946562 | Guruswamy | Aug 1990 | A |
4948561 | Hinckley et al. | Aug 1990 | A |
4949742 | Rando et al. | Aug 1990 | A |
D310413 | Bigler et al. | Sep 1990 | S |
4963498 | Hillman | Oct 1990 | A |
4965188 | Mullis et al. | Oct 1990 | A |
4967950 | Legg et al. | Nov 1990 | A |
D312692 | Bradley | Dec 1990 | S |
4978502 | Dole et al. | Dec 1990 | A |
4978622 | Mishell et al. | Dec 1990 | A |
4989626 | Takagi et al. | Feb 1991 | A |
4994373 | Stavrianopoulos et al. | Feb 1991 | A |
4997772 | Sutton et al. | Mar 1991 | A |
5001417 | Pumphrey et al. | Mar 1991 | A |
5004583 | Guruswamy et al. | Apr 1991 | A |
5048554 | Kremer | Sep 1991 | A |
5053199 | Keiser et al. | Oct 1991 | A |
5060823 | Perlman | Oct 1991 | A |
5061336 | Soane | Oct 1991 | A |
5064618 | Baker et al. | Nov 1991 | A |
5071531 | Soane | Dec 1991 | A |
5089233 | DeVaney, Jr. et al. | Feb 1992 | A |
5091328 | Miller | Feb 1992 | A |
D324426 | Fan et al. | Mar 1992 | S |
5096669 | Lauks et al. | Mar 1992 | A |
D325638 | Sloat et al. | Apr 1992 | S |
5126002 | Iwata et al. | Jun 1992 | A |
5126022 | Soane et al. | Jun 1992 | A |
D328135 | Fan et al. | Jul 1992 | S |
D328794 | Frenkel et al. | Aug 1992 | S |
5135627 | Soane | Aug 1992 | A |
5135872 | Pouletty et al. | Aug 1992 | A |
5147606 | Charlton et al. | Sep 1992 | A |
5147777 | Sutton et al. | Sep 1992 | A |
5155166 | Danielson et al. | Oct 1992 | A |
5169512 | Wiedenmann et al. | Dec 1992 | A |
5173269 | Mon et al. | Dec 1992 | A |
D333522 | Gianino | Feb 1993 | S |
5186339 | Heissler | Feb 1993 | A |
5192507 | Taylor et al. | Mar 1993 | A |
5208163 | Charlton et al. | May 1993 | A |
5217694 | Gibler et al. | Jun 1993 | A |
5223226 | Wittmer et al. | Jun 1993 | A |
5229297 | Schnipelsky et al. | Jul 1993 | A |
5231015 | Cummins et al. | Jul 1993 | A |
D338275 | Fischer et al. | Aug 1993 | S |
5234809 | Boom et al. | Aug 1993 | A |
5250263 | Manz | Oct 1993 | A |
5252743 | Barrett et al. | Oct 1993 | A |
5256376 | Callan et al. | Oct 1993 | A |
5273716 | Northrup et al. | Dec 1993 | A |
5275787 | Yuguchi et al. | Jan 1994 | A |
5282950 | Dietze et al. | Feb 1994 | A |
5296375 | Kricka et al. | Mar 1994 | A |
5304477 | Nagoh et al. | Apr 1994 | A |
5304487 | Wilding et al. | Apr 1994 | A |
D347478 | Pinkney | May 1994 | S |
5311896 | Kaartinen et al. | May 1994 | A |
5311996 | Duffy et al. | May 1994 | A |
5316727 | Suzuki et al. | May 1994 | A |
5327038 | Culp | Jul 1994 | A |
5334499 | Burdick et al. | Aug 1994 | A |
5338671 | Scalice et al. | Aug 1994 | A |
5339486 | Persic, Jr. | Aug 1994 | A |
D351475 | Gerber | Oct 1994 | S |
D351913 | Hieb et al. | Oct 1994 | S |
5364591 | Green et al. | Nov 1994 | A |
5372946 | Cusak et al. | Dec 1994 | A |
5374395 | Robinson | Dec 1994 | A |
5384499 | Pedersen et al. | Jan 1995 | A |
5389339 | Petschek et al. | Feb 1995 | A |
D356232 | Armstrong et al. | Mar 1995 | S |
5397709 | Berndt | Mar 1995 | A |
5401465 | Smethers et al. | Mar 1995 | A |
5411708 | Moscetta et al. | May 1995 | A |
5414245 | Hackleman | May 1995 | A |
5415839 | Zaun et al. | May 1995 | A |
5416000 | Allen et al. | May 1995 | A |
5422271 | Chen et al. | Jun 1995 | A |
5422284 | Lau | Jun 1995 | A |
5427946 | Kricka et al. | Jun 1995 | A |
5443791 | Cathcart et al. | Aug 1995 | A |
5466574 | Liberti et al. | Nov 1995 | A |
5474796 | Brennan | Dec 1995 | A |
5475487 | Mariella, Jr. et al. | Dec 1995 | A |
D366116 | Biskupski | Jan 1996 | S |
5486335 | Wilding et al. | Jan 1996 | A |
5494639 | Grzegorzewski | Feb 1996 | A |
5498392 | Wilding et al. | Mar 1996 | A |
5503803 | Brown | Apr 1996 | A |
5516410 | Schneider et al. | May 1996 | A |
5519635 | Miyake et al. | May 1996 | A |
5529677 | Schneider et al. | Jun 1996 | A |
5559432 | Logue | Sep 1996 | A |
5565171 | Dovichi et al. | Oct 1996 | A |
5569364 | Hooper et al. | Oct 1996 | A |
5578270 | Reichler et al. | Nov 1996 | A |
5578818 | Kain et al. | Nov 1996 | A |
5579928 | Anukwuem | Dec 1996 | A |
5580523 | Bard | Dec 1996 | A |
5582884 | Ball et al. | Dec 1996 | A |
5582988 | Backus et al. | Dec 1996 | A |
5585069 | Zanucchi et al. | Dec 1996 | A |
5585089 | Queen et al. | Dec 1996 | A |
5585242 | Bouma et al. | Dec 1996 | A |
5587128 | Wilding et al. | Dec 1996 | A |
5589136 | Northrup et al. | Dec 1996 | A |
5593838 | Zanzucchi et al. | Jan 1997 | A |
5595708 | Berndt | Jan 1997 | A |
5599432 | Manz et al. | Feb 1997 | A |
5599503 | Manz et al. | Feb 1997 | A |
5599667 | Arnold, Jr. et al. | Feb 1997 | A |
5601727 | Bormann et al. | Feb 1997 | A |
5603351 | Cherukuri et al. | Feb 1997 | A |
5605662 | Heller et al. | Feb 1997 | A |
5609910 | Hackleman | Mar 1997 | A |
D378782 | LaBarbera et al. | Apr 1997 | S |
5628890 | Carter et al. | May 1997 | A |
5630920 | Friese et al. | May 1997 | A |
5631337 | Sassi et al. | May 1997 | A |
5632876 | Zanzucchi et al. | May 1997 | A |
5632957 | Heller et al. | May 1997 | A |
5635358 | Wilding et al. | Jun 1997 | A |
5637469 | Wilding et al. | Jun 1997 | A |
5639423 | Northrup et al. | Jun 1997 | A |
5639428 | Cottingham | Jun 1997 | A |
5643738 | Zanzucchi et al. | Jul 1997 | A |
5645801 | Bouma et al. | Jul 1997 | A |
5646039 | Northrup et al. | Jul 1997 | A |
5646049 | Tayi | Jul 1997 | A |
5647994 | Tuunanen et al. | Jul 1997 | A |
5651839 | Rauf | Jul 1997 | A |
5652141 | Henco et al. | Jul 1997 | A |
5652149 | Mileaf et al. | Jul 1997 | A |
D382346 | Buhler et al. | Aug 1997 | S |
D382647 | Staples et al. | Aug 1997 | S |
5654141 | Mariani et al. | Aug 1997 | A |
5658515 | Lee et al. | Aug 1997 | A |
5667976 | Van Ness et al. | Sep 1997 | A |
5671303 | Shieh et al. | Sep 1997 | A |
5674394 | Whitmore | Oct 1997 | A |
5674742 | Northrup et al. | Oct 1997 | A |
5681484 | Zanzucchi et al. | Oct 1997 | A |
5681529 | Taguchi et al. | Oct 1997 | A |
5683657 | Mian | Nov 1997 | A |
5683659 | Hovatter | Nov 1997 | A |
5699157 | Parce et al. | Dec 1997 | A |
5700637 | Southern | Dec 1997 | A |
5705813 | Apffel et al. | Jan 1998 | A |
5721136 | Finney et al. | Feb 1998 | A |
5725831 | Reichler et al. | Mar 1998 | A |
5726026 | Wilding et al. | Mar 1998 | A |
5726404 | Brody | Mar 1998 | A |
5726944 | Pelley et al. | Mar 1998 | A |
5731212 | Gavin et al. | Mar 1998 | A |
5744366 | Kricka et al. | Apr 1998 | A |
5746978 | Bienhaus et al. | May 1998 | A |
5747666 | Willis | May 1998 | A |
5750015 | Soane et al. | May 1998 | A |
5755942 | Zanzucchi et al. | May 1998 | A |
5762874 | Seaton et al. | Jun 1998 | A |
5763262 | Wong et al. | Jun 1998 | A |
5770029 | Nelson et al. | Jun 1998 | A |
5770388 | Vorpahl | Jun 1998 | A |
5772966 | Maracas et al. | Jun 1998 | A |
5779868 | Parce et al. | Jul 1998 | A |
5783148 | Cottingham et al. | Jul 1998 | A |
5787032 | Heller et al. | Jul 1998 | A |
5788814 | Sun et al. | Aug 1998 | A |
5800600 | Lima-Marques et al. | Sep 1998 | A |
5800690 | Chow et al. | Sep 1998 | A |
5804436 | Okun et al. | Sep 1998 | A |
D399959 | Prokop et al. | Oct 1998 | S |
5819749 | Lee et al. | Oct 1998 | A |
5827481 | Bente et al. | Oct 1998 | A |
5842106 | Thaler et al. | Nov 1998 | A |
5842787 | Kopf-Sill et al. | Dec 1998 | A |
5846396 | Zanzucchi et al. | Dec 1998 | A |
5846493 | Bankier et al. | Dec 1998 | A |
5849208 | Hayes et al. | Dec 1998 | A |
5849486 | Heller et al. | Dec 1998 | A |
5849489 | Heller | Dec 1998 | A |
5849598 | Wilson et al. | Dec 1998 | A |
5852495 | Parce | Dec 1998 | A |
5856174 | Lipshutz et al. | Jan 1999 | A |
5858187 | Ramsey et al. | Jan 1999 | A |
5858188 | Soane et al. | Jan 1999 | A |
5863502 | Southgate et al. | Jan 1999 | A |
5863708 | Zanzucchi et al. | Jan 1999 | A |
5863801 | Southgate et al. | Jan 1999 | A |
5866345 | Wilding et al. | Feb 1999 | A |
5869004 | Parce et al. | Feb 1999 | A |
5869244 | Martin et al. | Feb 1999 | A |
5872010 | Karger et al. | Feb 1999 | A |
5872623 | Stabile et al. | Feb 1999 | A |
5874046 | Megerle | Feb 1999 | A |
5876675 | Kennedy | Mar 1999 | A |
5880071 | Parce et al. | Mar 1999 | A |
5882465 | McReynolds | Mar 1999 | A |
5883211 | Sassi et al. | Mar 1999 | A |
5885432 | Hooper et al. | Mar 1999 | A |
5885470 | Parce et al. | Mar 1999 | A |
5895762 | Greenfield et al. | Apr 1999 | A |
5900130 | Benvegnu et al. | May 1999 | A |
5911737 | Lee et al. | Jun 1999 | A |
5912124 | Kumar | Jun 1999 | A |
5912134 | Shartle | Jun 1999 | A |
5914229 | Loewy | Jun 1999 | A |
5916522 | Boyd et al. | Jun 1999 | A |
5916776 | Kumar | Jun 1999 | A |
5919646 | Okun et al. | Jul 1999 | A |
5919711 | Boyd et al. | Jul 1999 | A |
5922591 | Anderson et al. | Jul 1999 | A |
5927547 | Papen et al. | Jul 1999 | A |
5928161 | Krulevitch et al. | Jul 1999 | A |
5928880 | Wilding et al. | Jul 1999 | A |
5929208 | Heller et al. | Jul 1999 | A |
D413391 | Lapeus et al. | Aug 1999 | S |
5932799 | Moles | Aug 1999 | A |
5935401 | Amigo | Aug 1999 | A |
5939291 | Loewy et al. | Aug 1999 | A |
5939312 | Baier et al. | Aug 1999 | A |
5942443 | Parce et al. | Aug 1999 | A |
5944717 | Lee et al. | Aug 1999 | A |
D413677 | Dumitrescu et al. | Sep 1999 | S |
D414271 | Mendoza | Sep 1999 | S |
5948227 | Dubrow | Sep 1999 | A |
5948363 | Gaillard | Sep 1999 | A |
5948673 | Cottingham | Sep 1999 | A |
5955028 | Chow | Sep 1999 | A |
5955029 | Wilding et al. | Sep 1999 | A |
5957579 | Kopf-Sill et al. | Sep 1999 | A |
5958203 | Parce et al. | Sep 1999 | A |
5958349 | Petersen et al. | Sep 1999 | A |
5958694 | Nikiforov | Sep 1999 | A |
5959221 | Boyd et al. | Sep 1999 | A |
5959291 | Jensen | Sep 1999 | A |
5935522 | Swerdlow et al. | Oct 1999 | A |
5964995 | Nikiforov et al. | Oct 1999 | A |
5964997 | McBride | Oct 1999 | A |
5965001 | Chow et al. | Oct 1999 | A |
5965410 | Chow et al. | Oct 1999 | A |
5965886 | Sauer et al. | Oct 1999 | A |
5968745 | Thorp et al. | Oct 1999 | A |
5972187 | Parce et al. | Oct 1999 | A |
5973138 | Collis | Oct 1999 | A |
D417009 | Boyd | Nov 1999 | S |
5976336 | Dubrow et al. | Nov 1999 | A |
5980704 | Cherukuri et al. | Nov 1999 | A |
5980719 | Cherukuri et al. | Nov 1999 | A |
5981735 | Thatcher et al. | Nov 1999 | A |
5985651 | Hunicke-Smith | Nov 1999 | A |
5989402 | Chow et al. | Nov 1999 | A |
5992820 | Fare et al. | Nov 1999 | A |
5993611 | Moroney, III et al. | Nov 1999 | A |
5993750 | Ghosh et al. | Nov 1999 | A |
5997708 | Craig | Dec 1999 | A |
6001229 | Ramsey | Dec 1999 | A |
6001231 | Kopf-Sill | Dec 1999 | A |
6001307 | Naka et al. | Dec 1999 | A |
6004450 | Northrup et al. | Dec 1999 | A |
6004515 | Parce et al. | Dec 1999 | A |
6007690 | Nelson et al. | Dec 1999 | A |
6010607 | Ramsey | Jan 2000 | A |
6010608 | Ramsey | Jan 2000 | A |
6010627 | Hood, III | Jan 2000 | A |
6012902 | Parce | Jan 2000 | A |
D420747 | Dumitrescu et al. | Feb 2000 | S |
D421130 | Cohen et al. | Feb 2000 | S |
6024920 | Cunanan | Feb 2000 | A |
D421653 | Purcell | Mar 2000 | S |
6033546 | Ramsey | Mar 2000 | A |
6033880 | Haff et al. | Mar 2000 | A |
6043080 | Lipshutz et al. | Mar 2000 | A |
6043880 | Andrews et al. | Mar 2000 | A |
6046056 | Parce et al. | Apr 2000 | A |
6048734 | Burns et al. | Apr 2000 | A |
6054034 | Soane et al. | Apr 2000 | A |
6054277 | Furcht et al. | Apr 2000 | A |
6056860 | Amigo et al. | May 2000 | A |
6057149 | Burns et al. | May 2000 | A |
6062261 | Jacobson et al. | May 2000 | A |
6063341 | Fassbind et al. | May 2000 | A |
6063589 | Kellogg et al. | May 2000 | A |
6068751 | Neukermans | May 2000 | A |
6068752 | Dubrow et al. | May 2000 | A |
6071478 | Chow | Jun 2000 | A |
6074725 | Kennedy | Jun 2000 | A |
6074827 | Nelson et al. | Jun 2000 | A |
D428497 | Lapeus et al. | Jul 2000 | S |
6086740 | Kennedy | Jul 2000 | A |
6096509 | Okun et al. | Aug 2000 | A |
6100541 | Nagle et al. | Aug 2000 | A |
6102897 | Lang | Aug 2000 | A |
6103537 | Ullman et al. | Aug 2000 | A |
6106685 | McBride et al. | Aug 2000 | A |
6110343 | Ramsey et al. | Aug 2000 | A |
6117398 | Bienhaus et al. | Sep 2000 | A |
6123205 | Dumitrescu et al. | Sep 2000 | A |
6123798 | Gandhi et al. | Sep 2000 | A |
6130098 | Handique et al. | Oct 2000 | A |
6132580 | Mathies et al. | Oct 2000 | A |
6132684 | Marino | Oct 2000 | A |
6133436 | Koster et al. | Oct 2000 | A |
D433759 | Mathis et al. | Nov 2000 | S |
6143250 | Tajima | Nov 2000 | A |
6143547 | Hsu | Nov 2000 | A |
6149787 | Chow et al. | Nov 2000 | A |
6149872 | Mack et al. | Nov 2000 | A |
6156199 | Zuk, Jr. | Dec 2000 | A |
6158269 | Dorenkott et al. | Dec 2000 | A |
6167910 | Chow | Jan 2001 | B1 |
6168948 | Anderson et al. | Jan 2001 | B1 |
6171850 | Nagle et al. | Jan 2001 | B1 |
6174675 | Chow et al. | Jan 2001 | B1 |
6180950 | Olsen | Jan 2001 | B1 |
D438311 | Yamanishi et al. | Feb 2001 | S |
6190619 | Kilcoin et al. | Feb 2001 | B1 |
6194563 | Cruickshank | Feb 2001 | B1 |
D438632 | Miller | Mar 2001 | S |
D438633 | Miller | Mar 2001 | S |
D439673 | Brophy et al. | Mar 2001 | S |
6197595 | Anderson et al. | Mar 2001 | B1 |
6211989 | Wulf et al. | Apr 2001 | B1 |
6213151 | Jacobson et al. | Apr 2001 | B1 |
6221600 | MacLeod et al. | Apr 2001 | B1 |
6228635 | Armstrong et al. | May 2001 | B1 |
6232072 | Fisher | May 2001 | B1 |
6235175 | Dubrow et al. | May 2001 | B1 |
6235313 | Mathiowitz et al. | May 2001 | B1 |
6235471 | Knapp et al. | May 2001 | B1 |
6236456 | Giebeler et al. | May 2001 | B1 |
6236581 | Foss et al. | May 2001 | B1 |
6238626 | Higuchi et al. | May 2001 | B1 |
6251343 | Dubrow et al. | Jun 2001 | B1 |
6254826 | Acosta et al. | Jul 2001 | B1 |
6259635 | Khouri et al. | Jul 2001 | B1 |
6261431 | Mathies et al. | Jul 2001 | B1 |
6267858 | Parce et al. | Jul 2001 | B1 |
D446306 | Ochi et al. | Aug 2001 | S |
6271021 | Burns et al. | Aug 2001 | B1 |
6274089 | Chow et al. | Aug 2001 | B1 |
6280967 | Ransom et al. | Aug 2001 | B1 |
6281008 | Komai et al. | Aug 2001 | B1 |
6284113 | Bjornson et al. | Sep 2001 | B1 |
6284470 | Bitner et al. | Sep 2001 | B1 |
6287254 | Dodds | Sep 2001 | B1 |
6287774 | Nikiforov | Sep 2001 | B1 |
6291248 | Haj-Ahmad | Sep 2001 | B1 |
6294063 | Becker et al. | Sep 2001 | B1 |
6300124 | Blumenfeld et al. | Oct 2001 | B1 |
6302134 | Kellogg et al. | Oct 2001 | B1 |
6302304 | Spencer | Oct 2001 | B1 |
6303343 | Kopf-sill | Oct 2001 | B1 |
6306273 | Wainright et al. | Oct 2001 | B1 |
6306590 | Mehta et al. | Oct 2001 | B1 |
6310199 | Smith et al. | Oct 2001 | B1 |
6316774 | Giebeler et al. | Nov 2001 | B1 |
6319469 | Mian et al. | Nov 2001 | B1 |
6319474 | Krulevitch et al. | Nov 2001 | B1 |
6322683 | Wolk et al. | Nov 2001 | B1 |
6326083 | Yang et al. | Dec 2001 | B1 |
6326147 | Oldham et al. | Dec 2001 | B1 |
6326211 | Anderson et al. | Dec 2001 | B1 |
6334980 | Hayes et al. | Jan 2002 | B1 |
6337435 | Chu et al. | Jan 2002 | B1 |
6353475 | Jensen et al. | Mar 2002 | B1 |
6358387 | Kopf-sill et al. | Mar 2002 | B1 |
6366924 | Parce | Apr 2002 | B1 |
6368561 | Rutishauser et al. | Apr 2002 | B1 |
6368871 | Christel et al. | Apr 2002 | B1 |
6370206 | Schenk | Apr 2002 | B1 |
6375185 | Lin | Apr 2002 | B1 |
6375901 | Robotti et al. | Apr 2002 | B1 |
6379884 | Wada et al. | Apr 2002 | B2 |
6379929 | Burns et al. | Apr 2002 | B1 |
6379974 | Parce et al. | Apr 2002 | B1 |
6382254 | Yang et al. | May 2002 | B1 |
6391541 | Petersen et al. | May 2002 | B1 |
6391623 | Besemer et al. | May 2002 | B1 |
6395161 | Schneider et al. | May 2002 | B1 |
6398956 | Coville et al. | Jun 2002 | B1 |
6399025 | Chow | Jun 2002 | B1 |
6399389 | Parce et al. | Jun 2002 | B1 |
6399952 | Maher et al. | Jun 2002 | B1 |
6401552 | Elkins | Jun 2002 | B1 |
6403338 | Knapp et al. | Jun 2002 | B1 |
6408878 | Unger et al. | Jun 2002 | B2 |
6413401 | Chow et al. | Jul 2002 | B1 |
6416642 | Alajoki et al. | Jul 2002 | B1 |
6420143 | Kopf-sill | Jul 2002 | B1 |
6425972 | McReynolds | Jul 2002 | B1 |
D461906 | Pham | Aug 2002 | S |
6428987 | Franzen | Aug 2002 | B2 |
6430512 | Gallagher | Aug 2002 | B1 |
6432366 | Ruediger et al. | Aug 2002 | B2 |
6440725 | Pourahmadi et al. | Aug 2002 | B1 |
D463031 | Slomski et al. | Sep 2002 | S |
6444461 | Knapp et al. | Sep 2002 | B1 |
6447661 | Chow et al. | Sep 2002 | B1 |
6447727 | Parce et al. | Sep 2002 | B1 |
6448047 | Dattagupta et al. | Sep 2002 | B2 |
6448064 | Vo-Dinh et al. | Sep 2002 | B1 |
6453928 | Kaplan et al. | Sep 2002 | B1 |
6458259 | Parce et al. | Oct 2002 | B1 |
6461570 | Ishihara et al. | Oct 2002 | B2 |
6465257 | Parce et al. | Oct 2002 | B1 |
6468761 | Yang et al. | Oct 2002 | B2 |
6472141 | Nikiforov | Oct 2002 | B2 |
D466219 | Wynschenk et al. | Nov 2002 | S |
6475364 | Dubrow et al. | Nov 2002 | B1 |
D467348 | McMichael et al. | Dec 2002 | S |
D467349 | Niedbala et al. | Dec 2002 | S |
6488897 | Dubrow et al. | Dec 2002 | B2 |
6495104 | Unno et al. | Dec 2002 | B1 |
6498497 | Chow et al. | Dec 2002 | B1 |
6500323 | Chow et al. | Dec 2002 | B1 |
6500390 | Boulton et al. | Dec 2002 | B1 |
D468437 | McMenamy et al. | Jan 2003 | S |
6506609 | Wada et al. | Jan 2003 | B1 |
6509186 | Zou et al. | Jan 2003 | B1 |
6509193 | Tajima | Jan 2003 | B1 |
6511853 | Kopf-sill et al. | Jan 2003 | B1 |
D470595 | Crisanti et al. | Feb 2003 | S |
6515753 | Maher | Feb 2003 | B2 |
6517783 | Horner et al. | Feb 2003 | B2 |
6520197 | Deshmukh et al. | Feb 2003 | B2 |
6521181 | Northrup et al. | Feb 2003 | B1 |
6521188 | Webster | Feb 2003 | B1 |
6524456 | Ramsey et al. | Feb 2003 | B1 |
6524532 | Northrup | Feb 2003 | B1 |
6524790 | Kopf-sill et al. | Feb 2003 | B1 |
D472324 | Rumore et al. | Mar 2003 | S |
6534295 | Tai et al. | Mar 2003 | B2 |
6537432 | Schneider et al. | Mar 2003 | B1 |
6537771 | Farinas et al. | Mar 2003 | B1 |
6540896 | Manz et al. | Apr 2003 | B1 |
6544734 | Briscoe et al. | Apr 2003 | B1 |
6547942 | Parce et al. | Apr 2003 | B1 |
6555389 | Ullman et al. | Apr 2003 | B1 |
6556923 | Gallagher et al. | Apr 2003 | B2 |
D474279 | Mayer et al. | May 2003 | S |
D474280 | Niedbala et al. | May 2003 | S |
6558916 | Veerapandian et al. | May 2003 | B2 |
6558945 | Kao | May 2003 | B1 |
6565815 | Chang et al. | May 2003 | B1 |
6569607 | McReynolds | May 2003 | B2 |
6572830 | Burdon et al. | Jun 2003 | B1 |
6575188 | Parunak | Jun 2003 | B2 |
6576459 | Miles et al. | Jun 2003 | B2 |
6579453 | Bächler et al. | Jun 2003 | B1 |
6589729 | Chan et al. | Jul 2003 | B2 |
6592821 | Wada et al. | Jul 2003 | B1 |
6597450 | Andrews et al. | Jul 2003 | B1 |
6602474 | Tajima | Aug 2003 | B1 |
6613211 | Mccormick et al. | Sep 2003 | B1 |
6613512 | Kopf-sill et al. | Sep 2003 | B1 |
6613580 | Chow et al. | Sep 2003 | B1 |
6613581 | Wada et al. | Sep 2003 | B1 |
6614030 | Maher et al. | Sep 2003 | B2 |
6620625 | Wolk et al. | Sep 2003 | B2 |
6623860 | Hu et al. | Sep 2003 | B2 |
6627406 | Singh et al. | Sep 2003 | B1 |
D480814 | Lafferty et al. | Oct 2003 | S |
6632655 | Mehta et al. | Oct 2003 | B1 |
6633785 | Kasahara et al. | Oct 2003 | B1 |
D482796 | Oyama et al. | Nov 2003 | S |
6640981 | Lafond et al. | Nov 2003 | B2 |
6649358 | Parce et al. | Nov 2003 | B1 |
6664104 | Pourahmadi et al. | Dec 2003 | B2 |
6669831 | Chow et al. | Dec 2003 | B2 |
6670133 | Knapp et al. | Dec 2003 | B2 |
6670153 | Stern | Dec 2003 | B2 |
D484989 | Gebrian | Jan 2004 | S |
6672458 | Hansen et al. | Jan 2004 | B2 |
6681616 | Spaid et al. | Jan 2004 | B2 |
6681788 | Parce et al. | Jan 2004 | B2 |
6685813 | Williams et al. | Feb 2004 | B2 |
6692700 | Handique | Feb 2004 | B2 |
6695009 | Chien et al. | Feb 2004 | B2 |
6699713 | Benett et al. | Mar 2004 | B2 |
6706519 | Kellogg et al. | Mar 2004 | B1 |
6720148 | Nikiforov | Apr 2004 | B1 |
6730206 | Ricco et al. | May 2004 | B2 |
6733645 | Chow | May 2004 | B1 |
6734401 | Bedingham et al. | May 2004 | B2 |
6737026 | Bergh et al. | May 2004 | B1 |
6740518 | Duong et al. | May 2004 | B1 |
D491272 | Alden et al. | Jun 2004 | S |
D491273 | Biegler et al. | Jun 2004 | S |
D491276 | Langille | Jun 2004 | S |
6750661 | Brooks et al. | Jun 2004 | B2 |
6752966 | Chazan | Jun 2004 | B1 |
6756019 | Dubrow et al. | Jun 2004 | B1 |
6762049 | Zou et al. | Jul 2004 | B2 |
6764859 | Kreuwel et al. | Jul 2004 | B1 |
6766817 | da Silva | Jul 2004 | B2 |
6773567 | Wolk | Aug 2004 | B1 |
6777184 | Nikiforov et al. | Aug 2004 | B2 |
6783962 | Olander et al. | Aug 2004 | B1 |
D495805 | Lea et al. | Sep 2004 | S |
6787015 | Lackritz et al. | Sep 2004 | B2 |
6787016 | Tan et al. | Sep 2004 | B2 |
6787111 | Roach et al. | Sep 2004 | B2 |
6790328 | Jacobson et al. | Sep 2004 | B2 |
6790330 | Gascoyne et al. | Sep 2004 | B2 |
6811668 | Berndt et al. | Nov 2004 | B1 |
6818113 | Williams et al. | Nov 2004 | B2 |
6819027 | Saraf | Nov 2004 | B2 |
6824663 | Boone | Nov 2004 | B1 |
D499813 | Wu | Dec 2004 | S |
D500142 | Crisanti et al. | Dec 2004 | S |
D500363 | Fanning et al. | Dec 2004 | S |
6827831 | Chow et al. | Dec 2004 | B1 |
6827906 | Björnson et al. | Dec 2004 | B1 |
6838156 | Neyer et al. | Jan 2005 | B1 |
6838680 | Maher et al. | Jan 2005 | B2 |
6852287 | Ganesan | Feb 2005 | B2 |
6858185 | Kopf-sill et al. | Feb 2005 | B1 |
6859698 | Schmeisser | Feb 2005 | B2 |
6861035 | Pham et al. | Mar 2005 | B2 |
6878540 | Pourahmadi et al. | Apr 2005 | B2 |
6878755 | Singh et al. | Apr 2005 | B2 |
6884628 | Hubbell et al. | Apr 2005 | B2 |
6887693 | McMillan et al. | May 2005 | B2 |
6893879 | Petersen et al. | May 2005 | B2 |
6900889 | Bjornson et al. | May 2005 | B2 |
6905583 | Wainright et al. | Jun 2005 | B2 |
6905612 | Dorian et al. | Jun 2005 | B2 |
6906797 | Kao et al. | Jun 2005 | B1 |
6908594 | Schaevitz et al. | Jun 2005 | B1 |
6911183 | Handique et al. | Jun 2005 | B1 |
6914137 | Baker | Jul 2005 | B2 |
6915679 | Chien et al. | Jul 2005 | B2 |
6918404 | da Silva | Jul 2005 | B2 |
D508999 | Fanning et al. | Aug 2005 | S |
6939451 | Zhao et al. | Sep 2005 | B2 |
6940598 | Christel et al. | Sep 2005 | B2 |
6942771 | Kayyem | Sep 2005 | B1 |
6951632 | Unger et al. | Oct 2005 | B2 |
6958392 | Fomovskaia et al. | Oct 2005 | B2 |
D512155 | Matsumoto | Nov 2005 | S |
6964747 | Banerjee et al. | Nov 2005 | B2 |
6977163 | Mehta | Dec 2005 | B1 |
6979424 | Northrup et al. | Dec 2005 | B2 |
6984516 | Briscoe et al. | Jan 2006 | B2 |
D515707 | Sinohara et al. | Feb 2006 | S |
D516221 | Wohlstadter et al. | Feb 2006 | S |
7001853 | Brown et al. | Feb 2006 | B1 |
7004184 | Handique et al. | Feb 2006 | B2 |
D517554 | Yanagisawa et al. | Mar 2006 | S |
7010391 | Handique et al. | Mar 2006 | B2 |
7023007 | Gallagher | Apr 2006 | B2 |
7024281 | Unno | Apr 2006 | B1 |
7036667 | Greenstein et al. | May 2006 | B2 |
7037416 | Parce et al. | May 2006 | B2 |
7038472 | Chien | May 2006 | B1 |
7039527 | Tripathi et al. | May 2006 | B2 |
7040144 | Spaid et al. | May 2006 | B2 |
7049558 | Baer et al. | May 2006 | B2 |
D523153 | Akashi et al. | Jun 2006 | S |
7055695 | Greenstein et al. | Jun 2006 | B2 |
7060171 | Nikiforov et al. | Jun 2006 | B1 |
7066586 | da Silva | Jun 2006 | B2 |
7069952 | McReynolds et al. | Jul 2006 | B1 |
7072036 | Jones et al. | Jul 2006 | B2 |
7099778 | Chien | Aug 2006 | B2 |
D528215 | Malmsater | Sep 2006 | S |
7101467 | Spaid | Sep 2006 | B2 |
7105304 | Nikiforov et al. | Sep 2006 | B1 |
D531321 | Godfrey et al. | Oct 2006 | S |
7118892 | Ammann et al. | Oct 2006 | B2 |
7118910 | Unger et al. | Oct 2006 | B2 |
7122799 | Hsieh et al. | Oct 2006 | B2 |
7135144 | Christel et al. | Nov 2006 | B2 |
7138032 | Gandhi et al. | Nov 2006 | B2 |
D534280 | Gomm et al. | Dec 2006 | S |
7150814 | Parce et al. | Dec 2006 | B1 |
7150999 | Shuck | Dec 2006 | B1 |
D535403 | Isozaki et al. | Jan 2007 | S |
7160423 | Chien et al. | Jan 2007 | B2 |
7161356 | Chien | Jan 2007 | B1 |
7169277 | Ausserer et al. | Jan 2007 | B2 |
7169601 | Northrup et al. | Jan 2007 | B1 |
7169618 | Skold | Jan 2007 | B2 |
D537951 | Okamoto et al. | Mar 2007 | S |
D538436 | Patadia et al. | Mar 2007 | S |
7188001 | Young et al. | Mar 2007 | B2 |
7192557 | Wu et al. | Mar 2007 | B2 |
7195986 | Bousse et al. | Mar 2007 | B1 |
7205154 | Corson | Apr 2007 | B2 |
7208125 | Dong | Apr 2007 | B1 |
7235406 | Woudenberg et al. | Jun 2007 | B1 |
7247274 | Chow | Jul 2007 | B1 |
D548841 | Brownell et al. | Aug 2007 | S |
D549827 | Maeno et al. | Aug 2007 | S |
7252928 | Hafeman et al. | Aug 2007 | B1 |
7255833 | Chang et al. | Aug 2007 | B2 |
7270786 | Parunak et al. | Sep 2007 | B2 |
D554069 | Bolotin et al. | Oct 2007 | S |
D554070 | Bolotin et al. | Oct 2007 | S |
7276208 | Sevigny et al. | Oct 2007 | B2 |
7276330 | Chow et al. | Oct 2007 | B2 |
7288228 | Lefebvre | Oct 2007 | B2 |
7297313 | Northrup et al. | Nov 2007 | B1 |
D556914 | Okamoto et al. | Dec 2007 | S |
7303727 | Dubrow et al. | Dec 2007 | B1 |
D559995 | Handique et al. | Jan 2008 | S |
7315376 | Bickmore et al. | Jan 2008 | B2 |
7323140 | Handique et al. | Jan 2008 | B2 |
7332130 | Handique | Feb 2008 | B2 |
7338760 | Gong et al. | Mar 2008 | B2 |
D566291 | Parunak et al. | Apr 2008 | S |
7351377 | Chazan et al. | Apr 2008 | B2 |
D569526 | Duffy et al. | May 2008 | S |
7374949 | Kuriger | May 2008 | B2 |
7390460 | Osawa et al. | Jun 2008 | B2 |
7419784 | Dubrow et al. | Sep 2008 | B2 |
7422669 | Jacobson et al. | Sep 2008 | B2 |
7440684 | Spaid et al. | Oct 2008 | B2 |
7476313 | Siddiqi | Jan 2009 | B2 |
7480042 | Phillips et al. | Jan 2009 | B1 |
7494577 | Williams et al. | Feb 2009 | B2 |
7494770 | Wilding et al. | Feb 2009 | B2 |
7514046 | Kechagia et al. | Apr 2009 | B2 |
7518726 | Rulison et al. | Apr 2009 | B2 |
7521186 | Burd Mehta | Apr 2009 | B2 |
7527769 | Bunch et al. | May 2009 | B2 |
D595423 | Johansson et al. | Jun 2009 | S |
7553671 | Sinclair et al. | Jun 2009 | B2 |
D596312 | Giraud et al. | Jul 2009 | S |
D598566 | Allaer | Aug 2009 | S |
7578976 | Northrup et al. | Aug 2009 | B1 |
D599234 | Ito | Sep 2009 | S |
7595197 | Brasseur | Sep 2009 | B2 |
7604938 | Takahashi et al. | Oct 2009 | B2 |
7622296 | Joseph et al. | Nov 2009 | B2 |
7628902 | Knowlton et al. | Dec 2009 | B2 |
7633606 | Northrup et al. | Dec 2009 | B2 |
7635588 | King et al. | Dec 2009 | B2 |
7645581 | Knapp et al. | Jan 2010 | B2 |
7670559 | Chien et al. | Mar 2010 | B2 |
7674431 | Ganesan | Mar 2010 | B2 |
7689022 | Weiner et al. | Mar 2010 | B2 |
7704735 | Facer et al. | Apr 2010 | B2 |
7705739 | Northrup et al. | Apr 2010 | B2 |
7723123 | Murphy et al. | May 2010 | B1 |
D618820 | Wilson et al. | Jun 2010 | S |
7727371 | Kennedy et al. | Jun 2010 | B2 |
7727477 | Boronkay et al. | Jun 2010 | B2 |
7744817 | Bui | Jun 2010 | B2 |
D621060 | Handique | Aug 2010 | S |
7785868 | Yuan et al. | Aug 2010 | B2 |
D628305 | Gorrec et al. | Nov 2010 | S |
7829025 | Ganesan et al. | Nov 2010 | B2 |
7858366 | Northrup et al. | Dec 2010 | B2 |
7867776 | Kennedy et al. | Jan 2011 | B2 |
7892819 | Wilding et al. | Feb 2011 | B2 |
D637737 | Wilson et al. | May 2011 | S |
7955864 | Cox et al. | Jun 2011 | B2 |
7987022 | Handique et al. | Jul 2011 | B2 |
7998708 | Handique et al. | Aug 2011 | B2 |
8053214 | Northrup | Nov 2011 | B2 |
8071056 | Burns et al. | Dec 2011 | B2 |
8088616 | Handique | Jan 2012 | B2 |
8105783 | Handique | Jan 2012 | B2 |
8110158 | Handique | Feb 2012 | B2 |
8133671 | Williams et al. | Mar 2012 | B2 |
8182763 | Duffy et al. | May 2012 | B2 |
8246919 | Herchenbach et al. | Aug 2012 | B2 |
8273308 | Handique et al. | Sep 2012 | B2 |
D669597 | Cavada et al. | Oct 2012 | S |
8287820 | Williams et al. | Oct 2012 | B2 |
8323584 | Ganesan | Dec 2012 | B2 |
8323900 | Handique et al. | Dec 2012 | B2 |
8324372 | Brahmasandra et al. | Dec 2012 | B2 |
8415103 | Handique | Apr 2013 | B2 |
8420015 | Ganesan et al. | Apr 2013 | B2 |
8440149 | Handique | May 2013 | B2 |
8470586 | Wu et al. | Jun 2013 | B2 |
8473104 | Handique et al. | Jun 2013 | B2 |
D686749 | Trump | Jul 2013 | S |
D687567 | Jungheim et al. | Aug 2013 | S |
D692162 | Lentz et al. | Oct 2013 | S |
8592157 | Petersen et al. | Nov 2013 | B2 |
8679831 | Handique et al. | Mar 2014 | B2 |
D702854 | Nakahana et al. | Apr 2014 | S |
8685341 | Ganesan | Apr 2014 | B2 |
8703069 | Handique et al. | Apr 2014 | B2 |
8709787 | Handique | Apr 2014 | B2 |
8710211 | Brahmasandra et al. | Apr 2014 | B2 |
8734733 | Handique | May 2014 | B2 |
D710024 | Guo | Jul 2014 | S |
8765076 | Handique et al. | Jul 2014 | B2 |
8765454 | Zhou et al. | Jul 2014 | B2 |
8768517 | Handique et al. | Jul 2014 | B2 |
8852862 | Wu et al. | Oct 2014 | B2 |
8883490 | Handique et al. | Nov 2014 | B2 |
8894947 | Ganesan et al. | Nov 2014 | B2 |
8895311 | Handique et al. | Nov 2014 | B1 |
D729404 | Teich et al. | May 2015 | S |
9028773 | Ganesan | May 2015 | B2 |
9040288 | Handique et al. | May 2015 | B2 |
9051604 | Handique | Jun 2015 | B2 |
9080207 | Handique et al. | Jul 2015 | B2 |
D742027 | Lentz et al. | Oct 2015 | S |
9186677 | Williams et al. | Nov 2015 | B2 |
9217143 | Brahmasandra et al. | Dec 2015 | B2 |
9222954 | Lentz et al. | Dec 2015 | B2 |
9234236 | Thomas et al. | Jan 2016 | B2 |
9238223 | Handique | Jan 2016 | B2 |
9259734 | Williams et al. | Feb 2016 | B2 |
9259735 | Handique et al. | Feb 2016 | B2 |
9347586 | Williams et al. | May 2016 | B2 |
9480983 | Lentz et al. | Nov 2016 | B2 |
9528142 | Handique | Dec 2016 | B2 |
9618139 | Handique | Apr 2017 | B2 |
D787087 | Duffy et al. | Jun 2017 | S |
9670528 | Handique et al. | Jun 2017 | B2 |
9677121 | Ganesan et al. | Jun 2017 | B2 |
9701957 | Wilson et al. | Jul 2017 | B2 |
9745623 | Steel | Aug 2017 | B2 |
9765389 | Gubatayao et al. | Sep 2017 | B2 |
9789481 | Petersen et al. | Oct 2017 | B2 |
9802199 | Handique et al. | Oct 2017 | B2 |
9815057 | Handique | Nov 2017 | B2 |
9958466 | Dalbert et al. | May 2018 | B2 |
10065185 | Handique | Sep 2018 | B2 |
10071376 | Williams et al. | Sep 2018 | B2 |
10076754 | Lentz et al. | Sep 2018 | B2 |
10100302 | Brahmasandra et al. | Oct 2018 | B2 |
10139012 | Handique | Nov 2018 | B2 |
10179910 | Duffy et al. | Jan 2019 | B2 |
10234474 | Williams et al. | Mar 2019 | B2 |
10351901 | Ganesan et al. | Jul 2019 | B2 |
10364456 | Wu et al. | Jul 2019 | B2 |
10443088 | Wu et al. | Oct 2019 | B1 |
10494663 | Wu et al. | Dec 2019 | B1 |
10571935 | Handique et al. | Feb 2020 | B2 |
10590410 | Brahmasandra et al. | Mar 2020 | B2 |
10604788 | Wu et al. | Mar 2020 | B2 |
10619191 | Ganesan et al. | Apr 2020 | B2 |
10625261 | Williams et al. | Apr 2020 | B2 |
10625262 | Williams et al. | Apr 2020 | B2 |
10632466 | Williams et al. | Apr 2020 | B1 |
10695764 | Handique et al. | Jun 2020 | B2 |
10710069 | Handique et al. | Jul 2020 | B2 |
10717085 | Williams et al. | Jul 2020 | B2 |
10731201 | Handique et al. | Aug 2020 | B2 |
10781482 | Gubatayao et al. | Sep 2020 | B2 |
10799862 | Handique | Oct 2020 | B2 |
10821436 | Handique et al. | Nov 2020 | B2 |
10821446 | Handique et al. | Nov 2020 | B1 |
10822644 | Steel et al. | Nov 2020 | B2 |
10843188 | Handique et al. | Nov 2020 | B2 |
10844368 | Duffy et al. | Nov 2020 | B2 |
10857535 | Handique et al. | Dec 2020 | B2 |
10900066 | Handique | Jan 2021 | B2 |
20010005489 | Roach et al. | Jun 2001 | A1 |
20010012492 | Acosta et al. | Aug 2001 | A1 |
20010016358 | Osawa et al. | Aug 2001 | A1 |
20010018513 | Baker | Aug 2001 | A1 |
20010021355 | Baugh et al. | Sep 2001 | A1 |
20010023848 | Gjerde et al. | Sep 2001 | A1 |
20010038450 | McCaffrey et al. | Nov 2001 | A1 |
20010045358 | Kopf-Sill et al. | Nov 2001 | A1 |
20010046702 | Schembri | Nov 2001 | A1 |
20010048899 | Marouiss et al. | Dec 2001 | A1 |
20010051340 | Singh et al. | Dec 2001 | A1 |
20010055765 | O'Keefe et al. | Dec 2001 | A1 |
20020001848 | Bedingham et al. | Jan 2002 | A1 |
20020008053 | Hansen et al. | Jan 2002 | A1 |
20020009015 | Laugharn, Jr. et al. | Jan 2002 | A1 |
20020014443 | Hansen et al. | Feb 2002 | A1 |
20020015667 | Chow | Feb 2002 | A1 |
20020021983 | Comte et al. | Feb 2002 | A1 |
20020022261 | Anderson et al. | Feb 2002 | A1 |
20020037499 | Quake et al. | Mar 2002 | A1 |
20020039783 | McMillan et al. | Apr 2002 | A1 |
20020047003 | Bedingham et al. | Apr 2002 | A1 |
20020053399 | Soane et al. | May 2002 | A1 |
20020054835 | Robotti et al. | May 2002 | A1 |
20020055167 | Pourahmadi et al. | May 2002 | A1 |
20020058332 | Quake et al. | May 2002 | A1 |
20020060156 | Mathies et al. | May 2002 | A1 |
20020068357 | Mathies et al. | Jun 2002 | A1 |
20020068821 | Gundling | Jun 2002 | A1 |
20020086443 | Bamdad | Jul 2002 | A1 |
20020090320 | Burow et al. | Jul 2002 | A1 |
20020092767 | Bjornson et al. | Jul 2002 | A1 |
20020094303 | Yamamoto et al. | Jul 2002 | A1 |
20020131903 | Ingenhoven et al. | Sep 2002 | A1 |
20020141903 | Parunak et al. | Oct 2002 | A1 |
20020143297 | Francavilla et al. | Oct 2002 | A1 |
20020155010 | Karp et al. | Oct 2002 | A1 |
20020155477 | Ito | Oct 2002 | A1 |
20020169518 | Luoma et al. | Nov 2002 | A1 |
20020173032 | Zou et al. | Nov 2002 | A1 |
20020176804 | Strand et al. | Nov 2002 | A1 |
20020187557 | Hobbs et al. | Dec 2002 | A1 |
20020192808 | Gambini et al. | Dec 2002 | A1 |
20030008308 | Enzelberger et al. | Jan 2003 | A1 |
20030008320 | Baker | Jan 2003 | A1 |
20030019522 | Parunak | Jan 2003 | A1 |
20030022392 | Hudak | Jan 2003 | A1 |
20030036067 | Schwartz | Feb 2003 | A1 |
20030049833 | Chen et al. | Mar 2003 | A1 |
20030059823 | Matsunaga et al. | Mar 2003 | A1 |
20030064507 | Gallagher et al. | Apr 2003 | A1 |
20030072683 | Stewart et al. | Apr 2003 | A1 |
20030073106 | Johansen et al. | Apr 2003 | A1 |
20030083686 | Freeman et al. | May 2003 | A1 |
20030087300 | Knapp et al. | May 2003 | A1 |
20030088657 | Eggers | May 2003 | A1 |
20030096310 | Hansen et al. | May 2003 | A1 |
20030099954 | Miltenyi et al. | May 2003 | A1 |
20030124611 | Schwartz | Jul 2003 | A1 |
20030127327 | Kurnik | Jul 2003 | A1 |
20030134333 | Dehlinger et al. | Jul 2003 | A1 |
20030136679 | Bohn et al. | Jul 2003 | A1 |
20030156991 | Halas et al. | Aug 2003 | A1 |
20030180192 | Seippel | Sep 2003 | A1 |
20030186295 | Colin et al. | Oct 2003 | A1 |
20030190608 | Blackburn et al. | Oct 2003 | A1 |
20030199081 | Wilding et al. | Oct 2003 | A1 |
20030211517 | Carulli et al. | Nov 2003 | A1 |
20040014202 | King et al. | Jan 2004 | A1 |
20040014238 | Krug et al. | Jan 2004 | A1 |
20040018116 | Desmond et al. | Jan 2004 | A1 |
20040018119 | Massaro | Jan 2004 | A1 |
20040022689 | Wulf et al. | Feb 2004 | A1 |
20040029258 | Heaney et al. | Feb 2004 | A1 |
20040029260 | Hansen et al. | Feb 2004 | A1 |
20040037739 | McNeely et al. | Feb 2004 | A1 |
20040043479 | Briscoe et al. | Mar 2004 | A1 |
20040053290 | Terbrueggen et al. | Mar 2004 | A1 |
20040063217 | Webster et al. | Apr 2004 | A1 |
20040065655 | Brown | Apr 2004 | A1 |
20040072278 | Chou et al. | Apr 2004 | A1 |
20040072375 | Gjerde et al. | Apr 2004 | A1 |
20040076996 | Kondo et al. | Apr 2004 | A1 |
20040086427 | Childers et al. | May 2004 | A1 |
20040086956 | Bachur | May 2004 | A1 |
20040132059 | Scurati et al. | Jul 2004 | A1 |
20040141887 | Mainquist et al. | Jul 2004 | A1 |
20040151629 | Pease et al. | Aug 2004 | A1 |
20040157220 | Kurnool et al. | Aug 2004 | A1 |
20040161788 | Chen et al. | Aug 2004 | A1 |
20040171515 | Hamers et al. | Sep 2004 | A1 |
20040189311 | Glezer et al. | Sep 2004 | A1 |
20040197810 | Takenaka et al. | Oct 2004 | A1 |
20040200909 | McMillan et al. | Oct 2004 | A1 |
20040209331 | Ririe | Oct 2004 | A1 |
20040209354 | Mathies et al. | Oct 2004 | A1 |
20040224317 | Kordunsky et al. | Nov 2004 | A1 |
20040235154 | Oh et al. | Nov 2004 | A1 |
20040240097 | Evans | Dec 2004 | A1 |
20050009174 | Nikiforov et al. | Jan 2005 | A1 |
20050013737 | Chow et al. | Jan 2005 | A1 |
20050019902 | Mathies et al. | Jan 2005 | A1 |
20050037471 | Liu et al. | Feb 2005 | A1 |
20050041525 | Pugia et al. | Feb 2005 | A1 |
20050042639 | Knapp et al. | Feb 2005 | A1 |
20050048540 | Inami et al. | Mar 2005 | A1 |
20050058574 | Bysouth et al. | Mar 2005 | A1 |
20050058577 | Micklash et al. | Mar 2005 | A1 |
20050064535 | Favuzzi et al. | Mar 2005 | A1 |
20050069898 | Moon et al. | Mar 2005 | A1 |
20050084424 | Ganesan et al. | Apr 2005 | A1 |
20050106066 | Saltsman et al. | May 2005 | A1 |
20050112754 | Yoon et al. | May 2005 | A1 |
20050121324 | Park et al. | Jun 2005 | A1 |
20050129580 | Swinehart et al. | Jun 2005 | A1 |
20050130198 | Ammann et al. | Jun 2005 | A1 |
20050133370 | Park et al. | Jun 2005 | A1 |
20050135655 | Kopf-sill et al. | Jun 2005 | A1 |
20050142036 | Kim et al. | Jun 2005 | A1 |
20050158781 | Woudenberg et al. | Jul 2005 | A1 |
20050170362 | Wada et al. | Aug 2005 | A1 |
20050186585 | Juncosa et al. | Aug 2005 | A1 |
20050196321 | Huang | Sep 2005 | A1 |
20050202470 | Sundberg et al. | Sep 2005 | A1 |
20050202489 | Cho et al. | Sep 2005 | A1 |
20050202504 | Anderson et al. | Sep 2005 | A1 |
20050205788 | Itoh | Sep 2005 | A1 |
20050208676 | Kahatt | Sep 2005 | A1 |
20050214172 | Burgisser | Sep 2005 | A1 |
20050220675 | Reed et al. | Oct 2005 | A1 |
20050227269 | Lloyd et al. | Oct 2005 | A1 |
20050233370 | Ammann et al. | Oct 2005 | A1 |
20050238545 | Parce et al. | Oct 2005 | A1 |
20050239127 | Ammann et al. | Oct 2005 | A1 |
20050266489 | Ammann et al. | Dec 2005 | A1 |
20050276728 | Muller-Cohn et al. | Dec 2005 | A1 |
20060002817 | Bohm et al. | Jan 2006 | A1 |
20060003373 | Ammann et al. | Jan 2006 | A1 |
20060041058 | Yin et al. | Feb 2006 | A1 |
20060057039 | Morse et al. | Mar 2006 | A1 |
20060057629 | Kim | Mar 2006 | A1 |
20060058519 | Deggerdal et al. | Mar 2006 | A1 |
20060062696 | Chow et al. | Mar 2006 | A1 |
20060081539 | Safar et al. | Apr 2006 | A1 |
20060094004 | Nakajima et al. | May 2006 | A1 |
20060094108 | Yoder et al. | May 2006 | A1 |
20060113190 | Kurnik | Jun 2006 | A1 |
20060133965 | Tajima et al. | Jun 2006 | A1 |
20060134790 | Tanaka et al. | Jun 2006 | A1 |
20060148063 | Fauzzi et al. | Jul 2006 | A1 |
20060154341 | Chen | Jul 2006 | A1 |
20060165558 | Witty et al. | Jul 2006 | A1 |
20060165559 | Greenstein et al. | Jul 2006 | A1 |
20060177376 | Tomalia et al. | Aug 2006 | A1 |
20060177855 | Utermohlen et al. | Aug 2006 | A1 |
20060183216 | Handique | Aug 2006 | A1 |
20060201887 | Siddiqi | Sep 2006 | A1 |
20060205085 | Handique | Sep 2006 | A1 |
20060207944 | Siddiqi | Sep 2006 | A1 |
20060210435 | Alavie et al. | Sep 2006 | A1 |
20060223169 | Bedingham et al. | Oct 2006 | A1 |
20060228734 | Vann et al. | Oct 2006 | A1 |
20060246493 | Jensen et al. | Nov 2006 | A1 |
20060246533 | Fathollahi et al. | Nov 2006 | A1 |
20060269641 | Atwood et al. | Nov 2006 | A1 |
20060269961 | Fukushima et al. | Nov 2006 | A1 |
20070004028 | Lair et al. | Jan 2007 | A1 |
20070009386 | Padmanabhan et al. | Jan 2007 | A1 |
20070014695 | Yue et al. | Jan 2007 | A1 |
20070020699 | Carpenter et al. | Jan 2007 | A1 |
20070020764 | Miller | Jan 2007 | A1 |
20070026421 | Sundberg et al. | Feb 2007 | A1 |
20070042441 | Masters et al. | Feb 2007 | A1 |
20070048188 | Bigus | Mar 2007 | A1 |
20070054413 | Aviles et al. | Mar 2007 | A1 |
20070077643 | Nakamura et al. | Apr 2007 | A1 |
20070077648 | Okamoto et al. | Apr 2007 | A1 |
20070092901 | Ligler et al. | Apr 2007 | A1 |
20070098600 | Kayyem et al. | May 2007 | A1 |
20070099200 | Chow et al. | May 2007 | A1 |
20070104617 | Coulling et al. | May 2007 | A1 |
20070116613 | Elsener | May 2007 | A1 |
20070154895 | Spaid et al. | Jul 2007 | A1 |
20070177147 | Parce | Aug 2007 | A1 |
20070178603 | Takii et al. | Aug 2007 | A1 |
20070178607 | Prober et al. | Aug 2007 | A1 |
20070184463 | Molho et al. | Aug 2007 | A1 |
20070184547 | Handique et al. | Aug 2007 | A1 |
20070196237 | Neuzil et al. | Aug 2007 | A1 |
20070196238 | Kennedy et al. | Aug 2007 | A1 |
20070199821 | Chow | Aug 2007 | A1 |
20070215554 | Kreuwel et al. | Sep 2007 | A1 |
20070218459 | Miller et al. | Sep 2007 | A1 |
20070231213 | Prabhu et al. | Oct 2007 | A1 |
20070243626 | Windeyer et al. | Oct 2007 | A1 |
20070248958 | Jovanovich et al. | Oct 2007 | A1 |
20070261479 | Spaid et al. | Nov 2007 | A1 |
20070269861 | Williams et al. | Nov 2007 | A1 |
20070292941 | Handique et al. | Dec 2007 | A1 |
20080000774 | Park et al. | Jan 2008 | A1 |
20080003649 | Maltezos et al. | Jan 2008 | A1 |
20080017306 | Liu et al. | Jan 2008 | A1 |
20080056948 | Dale et al. | Mar 2008 | A1 |
20080069729 | McNeely | Mar 2008 | A1 |
20080090244 | Knapp et al. | Apr 2008 | A1 |
20080095673 | Xu | Apr 2008 | A1 |
20080118987 | Eastwood et al. | May 2008 | A1 |
20080124723 | Dale et al. | May 2008 | A1 |
20080176230 | Owen et al. | Jul 2008 | A1 |
20080192254 | Kim et al. | Aug 2008 | A1 |
20080226502 | Jonsmann et al. | Sep 2008 | A1 |
20080240898 | Manz et al. | Oct 2008 | A1 |
20080247914 | Edens et al. | Oct 2008 | A1 |
20080257882 | Turner | Oct 2008 | A1 |
20080280285 | Chen et al. | Nov 2008 | A1 |
20080308500 | Brassard | Dec 2008 | A1 |
20090047180 | Kawahara | Feb 2009 | A1 |
20090066339 | Glezer et al. | Mar 2009 | A1 |
20090136385 | Handique et al. | May 2009 | A1 |
20090148933 | Battrell et al. | Jun 2009 | A1 |
20090189089 | Bedingham et al. | Jul 2009 | A1 |
20090223925 | Morse et al. | Sep 2009 | A1 |
20090325164 | Vossenaar et al. | Dec 2009 | A1 |
20090325276 | Battrell et al. | Dec 2009 | A1 |
20100009351 | Brahmasandra et al. | Jan 2010 | A1 |
20100120129 | Amshey et al. | May 2010 | A1 |
20100233763 | Shigeura et al. | Sep 2010 | A1 |
20100284864 | Holenstein et al. | Nov 2010 | A1 |
20110008825 | Ingber et al. | Jan 2011 | A1 |
20110027151 | Handique et al. | Feb 2011 | A1 |
20110097493 | Kerr et al. | Apr 2011 | A1 |
20110127292 | Sarofim et al. | Jun 2011 | A1 |
20110158865 | Miller et al. | Jun 2011 | A1 |
20110287447 | Norderhaug | Nov 2011 | A1 |
20110300033 | Battisti | Dec 2011 | A1 |
20120122231 | Tajima | May 2012 | A1 |
20120160826 | Handique | Jun 2012 | A1 |
20120171678 | Maltezos et al. | Jul 2012 | A1 |
20120258463 | Duffy et al. | Oct 2012 | A1 |
20130183769 | Tajima | Jul 2013 | A1 |
20130315800 | Yin et al. | Nov 2013 | A1 |
20140030798 | Wu et al. | Jan 2014 | A1 |
20140227710 | Handique et al. | Aug 2014 | A1 |
20140329301 | Handique et al. | Nov 2014 | A1 |
20150045234 | Stone et al. | Feb 2015 | A1 |
20150174579 | Iten et al. | Jun 2015 | A1 |
20150315631 | Handique et al. | Nov 2015 | A1 |
20160038942 | Roberts | Feb 2016 | A1 |
20170275702 | Dahiya et al. | Sep 2017 | A1 |
20180333722 | Handique | Nov 2018 | A1 |
20190054467 | Handique | Feb 2019 | A1 |
20190054471 | Williams et al. | Feb 2019 | A1 |
20190144849 | Duffy et al. | May 2019 | A1 |
20190145546 | Handique | May 2019 | A1 |
20190151854 | Baum et al. | May 2019 | A1 |
20190154719 | LaChance et al. | May 2019 | A1 |
20190284606 | Wu et al. | Sep 2019 | A1 |
20190324050 | Williams et al. | Oct 2019 | A1 |
20200139363 | Handique et al. | May 2020 | A1 |
20200156060 | Handique et al. | May 2020 | A1 |
20200216831 | Brahmasandra et al. | Jul 2020 | A1 |
20200291388 | Brahmasandra et al. | Sep 2020 | A1 |
20200325523 | Brahmasandra et al. | Oct 2020 | A1 |
Number | Date | Country |
---|---|---|
1357102 | Mar 2002 | AU |
3557502 | Jul 2002 | AU |
4437602 | Jul 2002 | AU |
4437702 | Jul 2002 | AU |
764319 | Aug 2003 | AU |
2574107 | Sep 1998 | CA |
2294819 | Jan 1999 | CA |
1934451 | Mar 2007 | CN |
1312287 | Apr 2007 | CN |
1942590 | Apr 2007 | CN |
1968754 | May 2007 | CN |
101466848 | Jun 2009 | CN |
101522909 | Sep 2009 | CN |
103540518 | Jan 2014 | CN |
19755479 | Jun 1999 | DE |
19929734 | Dec 1999 | DE |
19833293 | Jan 2000 | DE |
0136126 | Apr 1985 | EP |
0365828 | May 1990 | EP |
0483620 | May 1992 | EP |
0402994 | Nov 1994 | EP |
0393744 | Jan 1995 | EP |
0688602 | Dec 1995 | EP |
0707077 | Apr 1996 | EP |
0698046 | Mar 1997 | EP |
0766256 | Apr 1997 | EP |
0772494 | May 1997 | EP |
0810030 | Dec 1997 | EP |
1059458 | Dec 2000 | EP |
1064090 | Jan 2001 | EP |
1077086 | Feb 2001 | EP |
1346772 | Sep 2003 | EP |
1541237 | Jun 2005 | EP |
1574586 | Sep 2005 | EP |
1621890 | Feb 2006 | EP |
1745153 | Jan 2007 | EP |
1780290 | May 2007 | EP |
1792656 | Jun 2007 | EP |
2372367 | Oct 2011 | EP |
2672301 | Aug 1992 | FR |
2795426 | Dec 2000 | FR |
2453432 | Apr 2009 | GB |
S50-100881 | Aug 1975 | JP |
58212921 | Dec 1983 | JP |
S62-119460 | May 1987 | JP |
H01-502319 | Aug 1989 | JP |
H03181853 | Aug 1991 | JP |
04-053555 | May 1992 | JP |
06-064156 | Sep 1994 | JP |
07-020010 | Jan 1995 | JP |
H07-290706 | Nov 1995 | JP |
H08-122336 | May 1996 | JP |
H08-173194 | Jul 1996 | JP |
H08-211071 | Aug 1996 | JP |
H08-285859 | Nov 1996 | JP |
H08-337116 | Dec 1996 | JP |
H09-304385 | Nov 1997 | JP |
H09-325151 | Dec 1997 | JP |
2001-502790 | Jan 1998 | JP |
H01-219669 | Sep 1998 | JP |
H10-327515 | Dec 1998 | JP |
H11-009258 | Jan 1999 | JP |
H11-501504 | Feb 1999 | JP |
H11-503315 | Mar 1999 | JP |
H11-156231 | Jun 1999 | JP |
H11-316226 | Nov 1999 | JP |
H11-515106 | Dec 1999 | JP |
2000-180455 | Jun 2000 | JP |
2000-266760 | Sep 2000 | JP |
2000-275255 | Oct 2000 | JP |
2000-514928 | Nov 2000 | JP |
2001-502319 | Feb 2001 | JP |
2001-204462 | Jul 2001 | JP |
2001-509437 | Jul 2001 | JP |
3191150 | Jul 2001 | JP |
2001-515216 | Sep 2001 | JP |
2001-523812 | Nov 2001 | JP |
2001-527220 | Dec 2001 | JP |
2002-503331 | Jan 2002 | JP |
2002-085961 | Mar 2002 | JP |
2002-517735 | Jun 2002 | JP |
2002-215241 | Jul 2002 | JP |
2002-540382 | Nov 2002 | JP |
2002-544476 | Dec 2002 | JP |
2003-500674 | Jan 2003 | JP |
2003-047839 | Feb 2003 | JP |
2003-047840 | Feb 2003 | JP |
2003-516125 | May 2003 | JP |
2003-164279 | Jun 2003 | JP |
2003-185584 | Jul 2003 | JP |
2003-299485 | Oct 2003 | JP |
2003-329693 | Nov 2003 | JP |
2003-329696 | Nov 2003 | JP |
2003-532382 | Nov 2003 | JP |
2004-003989 | Jan 2004 | JP |
2004-506179 | Feb 2004 | JP |
2004-150797 | May 2004 | JP |
2004-283728 | Oct 2004 | JP |
2004-531360 | Oct 2004 | JP |
2004-533838 | Nov 2004 | JP |
2004-534157 | Nov 2004 | JP |
2004-361421 | Dec 2004 | JP |
2004-536291 | Dec 2004 | JP |
2004-536689 | Dec 2004 | JP |
2005-009870 | Jan 2005 | JP |
2005-010179 | Jan 2005 | JP |
2005-511264 | Apr 2005 | JP |
2005-514718 | May 2005 | JP |
2005-518825 | Jun 2005 | JP |
2005-176613 | Jul 2005 | JP |
2005-192439 | Jul 2005 | JP |
2005-192554 | Jul 2005 | JP |
2005-519751 | Jul 2005 | JP |
2005-204661 | Aug 2005 | JP |
2005-525816 | Sep 2005 | JP |
2005-291954 | Oct 2005 | JP |
2005-532043 | Oct 2005 | JP |
2005-323519 | Nov 2005 | JP |
2005-533652 | Nov 2005 | JP |
2005-535904 | Nov 2005 | JP |
2006-021156 | Jan 2006 | JP |
2006-055837 | Mar 2006 | JP |
2006-094866 | Apr 2006 | JP |
2006-145458 | Jun 2006 | JP |
2006-167569 | Jun 2006 | JP |
2006-284409 | Oct 2006 | JP |
2007-024742 | Feb 2007 | JP |
2007-074960 | Mar 2007 | JP |
2007-097477 | Apr 2007 | JP |
2007-101364 | Apr 2007 | JP |
2007-510518 | Apr 2007 | JP |
2007-514405 | Jun 2007 | JP |
2007-178328 | Jul 2007 | JP |
2007-535933 | Dec 2007 | JP |
2009-515140 | Apr 2009 | JP |
2009-542207 | Dec 2009 | JP |
3193848 | Oct 2014 | JP |
1020060044489 | May 2006 | KR |
2418633 | May 2011 | RU |
WO 1988006633 | Sep 1988 | WO |
WO 1990012350 | Oct 1990 | WO |
WO 1992005443 | Apr 1992 | WO |
WO 1994005414 | Mar 1994 | WO |
WO 1994011103 | May 1994 | WO |
WO 1995033846 | Dec 1995 | WO |
WO 1996000228 | Jan 1996 | WO |
WO 1996004547 | Feb 1996 | WO |
WO 1996018731 | Jun 1996 | WO |
WO 1996039547 | Dec 1996 | WO |
WO 1997005492 | Feb 1997 | WO |
WO 1997016835 | May 1997 | WO |
WO 1997021090 | Jun 1997 | WO |
WO 1997022825 | Jun 1997 | WO |
WO 1997027324 | Jul 1997 | WO |
WO 1998000231 | Jan 1998 | WO |
WO 1998007019 | Feb 1998 | WO |
WO 1998022625 | May 1998 | WO |
WO 1998035013 | Aug 1998 | WO |
WO 1998038487 | Sep 1998 | WO |
WO 1998049548 | Nov 1998 | WO |
WO 1998050147 | Nov 1998 | WO |
WO 1998053311 | Nov 1998 | WO |
WO 1999001688 | Jan 1999 | WO |
WO 1999009042 | Feb 1999 | WO |
WO 1999012016 | Mar 1999 | WO |
WO 1999017093 | Apr 1999 | WO |
WO 1999029703 | Jun 1999 | WO |
WO 1999033559 | Jul 1999 | WO |
WO 1999060397 | Nov 1999 | WO |
WO 2000022436 | Apr 2000 | WO |
WO 2000075623 | Dec 2000 | WO |
WO 2000078455 | Dec 2000 | WO |
WO 2001005510 | Jan 2001 | WO |
WO 2001014931 | Mar 2001 | WO |
WO 2001027614 | Apr 2001 | WO |
WO 2001028684 | Apr 2001 | WO |
WO 2001030995 | May 2001 | WO |
WO 2001041931 | Jun 2001 | WO |
WO 2001046474 | Jun 2001 | WO |
WO 2001054813 | Aug 2001 | WO |
WO 2001089681 | Nov 2001 | WO |
WO 2001089705 | Nov 2001 | WO |
WO 2001092569 | Dec 2001 | WO |
WO 2002048164 | Jun 2002 | WO |
WO 2002052002 | Jul 2002 | WO |
WO 2002072264 | Sep 2002 | WO |
WO 2002078845 | Oct 2002 | WO |
WO 2002086454 | Oct 2002 | WO |
WO 2002094185 | Nov 2002 | WO |
WO 2003007677 | Jan 2003 | WO |
WO 2003012325 | Feb 2003 | WO |
WO 2003012406 | Feb 2003 | WO |
WO 2003048295 | Jun 2003 | WO |
WO 2003055605 | Jul 2003 | WO |
WO 2003076661 | Sep 2003 | WO |
WO 2003078065 | Sep 2003 | WO |
WO 2003080868 | Oct 2003 | WO |
WO 2003087410 | Oct 2003 | WO |
WO 2004007081 | Jan 2004 | WO |
WO 2004010760 | Feb 2004 | WO |
WO 2004048545 | Jun 2004 | WO |
WO 2004055522 | Jul 2004 | WO |
WO 2004056485 | Jul 2004 | WO |
WO 2004074848 | Sep 2004 | WO |
WO-2004074848 | Sep 2004 | WO |
WO 2004094986 | Nov 2004 | WO |
WO 2005008255 | Jan 2005 | WO |
WO 2005011867 | Feb 2005 | WO |
WO 2005030984 | Apr 2005 | WO |
WO 2005072353 | Aug 2005 | WO |
WO 2005094981 | Oct 2005 | WO |
WO 2005107947 | Nov 2005 | WO |
WO 2005108571 | Nov 2005 | WO |
WO 2005108620 | Nov 2005 | WO |
WO 2005116202 | Dec 2005 | WO |
WO 2005118867 | Dec 2005 | WO |
WO 2005120710 | Dec 2005 | WO |
WO 2006010584 | Feb 2006 | WO |
WO 2006032044 | Mar 2006 | WO |
WO 2006035800 | Apr 2006 | WO |
WO 2006043642 | Apr 2006 | WO |
WO 2006066001 | Jun 2006 | WO |
WO 2006079082 | Jul 2006 | WO |
WO 2006081995 | Aug 2006 | WO |
WO 2006113198 | Oct 2006 | WO |
WO 2006118420 | Nov 2006 | WO |
WO 2006119280 | Nov 2006 | WO |
WO 2007044917 | Apr 2007 | WO |
WO 2007050327 | May 2007 | WO |
WO 2007064117 | Jun 2007 | WO |
WO 2007075919 | Jul 2007 | WO |
WO 2007091530 | Aug 2007 | WO |
WO 2007112114 | Oct 2007 | WO |
WO 2007120240 | Oct 2007 | WO |
WO 2007120241 | Oct 2007 | WO |
WO 2008005321 | Jan 2008 | WO |
WO 2008030914 | Mar 2008 | WO |
WO 2008060604 | May 2008 | WO |
WO 2008149282 | Dec 2008 | WO |
WO 2009012185 | Jan 2009 | WO |
WO 2009054870 | Apr 2009 | WO |
WO 2010118541 | Oct 2010 | WO |
WO 2010130310 | Nov 2010 | WO |
WO 2010140680 | Dec 2010 | WO |
WO 2011009073 | Jan 2011 | WO |
WO 2011101467 | Aug 2011 | WO |
Entry |
---|
Allemand et al., “pH-Dependent Specific Binding and Combing of DNA”, Biophys J. (1997) 73(4): 2064-2070. |
Altet et al., [Eds.] “Thermal Transfer and Thermal Coupling in IC's”, Thermal Testing of Integrated Circuits; Chapter 2 (2002) Springer Science pp. 23-51. |
Anderson et al., “Microfluidic biochemical analysis system” Proc. 1997 IEEE Int. Conf. Solid-State Sens. Actuat. (1997) pp. 477-480. |
Anderson et al., “Advances in Integrated Genetic Analysis” Micro Total Analysis Systems '98 Conference Proceedings, D. Kluwer Academic Publishers (1998) in 6 pages. |
Anderson et al., “A Miniature Integrated Device for Automated Multistep Genetic Assays” Nucleic Acids Research (2000) 28(12), i-vi. |
Ateya et al., “The good, the bad, and the tiny: a review of microflow cytometry”, Anal Bioanal Chem. (2008) 391(5):1485-1498. |
Auroux et al., “Miniaturised nucleic acid analysis”, Lab Chip. (2004) 4(6):534-546. |
Baechi et al., “High-density microvalve arrays for sample processing in PCR chips”, Biomed Microdevices. (2001) 3(3):183-190. |
Baker M., “Clever PCR: more genotyping, smaller volumes.” Nature Methods (May 2010) 70(5):351-356. |
Becker H. “Fabrication of Polymer Microfluidic Devices”, in Biochip Technology (2001), Chapter 4, pp. 63-96. |
Becker H., “Microfluidic Devices Fabricated by Polymer Hot Embossing,” in Integrated Microfabricated Biodevices: Advanced Technologies for Genomics, Drug Discovery, Bioanalysis, and Clinical Diagnostics (2002), Chapter 13, 32 pages. |
Becker H., “Microfluidics: A Technology Coming of Age”, Med Device Technol. (2008) 19(3):21-24. |
Becker et al., “Portable CE system with contactless conductivity detection in an injection molded polymer chip for on-site food analysis”, SPIE Proceedings MOEMS-MEMS 2008 Micro and Nanofabrication (2008) vol. 6886 in 8 pages. |
Becker H., “Hype, hope and hubris: the quest for the killer application in microfluidics”, Lab on a Chip, The Royal Society of Chemistry (2009) 9:2119-2122. |
Becker H., “Collective Wisdom”, Lab on a Chip, The Royal Society of Chemistry (2010) 10:1351-1354. |
Belgrader et al., “Rapid PCR for Identity Testing Using a Battery-Powered Miniature Thermal Cycler”, J Forensic Sci. (1998) 43(2):315-319. |
Belgrader et al., “A minisonicator to rapidly disrupt bacterial spores for DNA analysis.”, Anal Chem. (1999) 71(19):4232-4236. |
Belgrader et al., “Real-time PCR Analysis on Nucleic Acids Purified from Plasma Using a Silicon Chip”, Micro Total Analysis Systems 2000 (pp. 525-528). Springer, Dordrecht. |
Belgrader et al., “A microfluidic cartridge to prepare spores for PCR analysis”, Biosens Bioelectron. (2000) 14(10-11):849-852. |
Belgrader et al., “A Battery-Powered Notebook Thermal Cycler for Rapid Multiplex Real-Time PCR Analysis”, Anal Chem. (2001) 73(2):286-289. |
Belgrader et al., “Rapid and Automated Cartridge-based Extraction of Leukocytes from Whole Blood for Microsatellite DNA Analysis by Capillary Electrophoresis”, Clin Chem. (2001) 47(10):1917-1933. |
Belgrader et al., “A Rapid, Flow-through, DNA Extraction Module for Integration into Microfluidic Systems”, Micro Total Analysis Systems (2002) pp. 697-699). Springer, Dordrecht. |
Belgrader et al., “Development of a Battery-Powered Portable Instrumentation for Rapid PCR Analysis”, in Integrated Microfabricated Devices, (2002) Ch. 8, pp. 183-206, CRC Press. |
Bell M., “Integrated Microsystems in Clinical Chemistry”, in Integrated Microfabricated Devices, (2002) Ch. 16, pp. 415-435, CRC Press. |
Berthier et al., “Managing evaporation for more robust microscale assays Part 1. Volume loss in high throughput assays”, Lab Chip (2008) 8(6):852-859. |
Berthier et al., “Managing evaporation for more robust microscale assays Part 2. Characterization of convection and diffusion for cell biology”, Lab Chip (2008) 8(6):860-864. |
Berthier et al., “Microdrops,” in Microfluidics for Biotechnology (2006), Chapter 2, pp. 51-88. |
BioMerieux Press Release: “bioMérieux—2018 Financial Results,” dated Feb. 27, 2019, accessed at www.biomerieux.com, pp. 13. |
Blanchard et al., “Micro structure mechanical failure characterization using rotating Couette flow in a small gap”, J Micromech Microengin. (2005) 15(4):792-801. |
Blanchard et al., “Single-disk and double-disk viscous micropumps”, Sensors and Actuators A (2005) 122:149-158. |
Blanchard et al., “Performance and Development of a Miniature Rotary Shaft Pump”, J Fluids Eng. (2005) 127(4):752-760. |
Blanchard et al., “Single-disk and double-disk viscous micropump”, ASME 2004 Inter'l Mechanical Engineering Congress & Exposition, Nov. 13-20, 2004, Anaheim, CA, IMECE2004-61705:411-417. |
Blanchard et al., “Miniature Single-Disk Viscous Pump (Single-DVP), Performance Characterization”, J Fluids Eng. (2006) 128(3):602-610. |
Bollet, C. et al., “A simple method for the isolation of chromosomal DNA from Gram positive or acid-fast bacteria”, Nucleic Acids Research, vol. 19, No. 8 (1991), p. 1955. |
Brahmasandra et al., On-chip DNA detection in microfabricated separation systems, SPIE Conference on Microfluidic Devices and Systems, 1998, vol. 3515, pp. 242-251, Santa Clara, CA. |
Brahmasandra et al., “Microfabricated Devices for Integrated DNA Analysis”, in Biochip Technology by Cheng et al., [Eds.] (2001) pp. 229-250. |
Breadmore, M.C. et al., “Microchip-Based Purification of DNA from Biological Samples”, Anal. Chem., vol. 75 (2003), pp. 1880-1886. |
Brody, et al., Diffusion-Based Extraction in a Microfabricated Device, Sensors and Actuators Elsevier, 1997, vol. A58, No. 1, pp. 13-18. |
Broyles et al., “Sample Filtration, Concentration, and Separation Integrated on Microfluidic Devices” Analytical Chemistry (American Chemical Society), (2003) 75(11): 2761-2767. |
Bu et al., “Design and theoretical evaluation of a novel microfluidic device to be used for PCR”, J Micromech Microengin. (2003) 13(4):S125-S130. |
Burns et al., “Microfabricated Structures for Integrated DNA Analysis” Proc. Natl. Acad. Sci. USA (May 1996) 93: 5556-5561. |
Burns et al., “An Integrated Nanoliter DNA Analysis Device”, Science 282:484-487 (1998). |
Cady et al., “Real-time PCR detection of Listeria monocytogenes using an integrated microfluidics platform”, Sensors Actuat B. (2005) 107:332-341. |
Carlen et al., “Paraffin Actuated Surface Micromachined Valve,” in IEEE MEMS 2000 Conference, Miyazaki, Japan, (Jan. 2000) pp. 381-385. |
Carles et al., “Polymerase Chain Reaction on Microchips” in Methods in Molecular Biology—Microfluidic Techniques, Reviews & Protocols by Minteer S.D. [Ed.] Humana Press (2006), vol. 321; Chapter 11, pp. 131-140. |
Chang-Yen et al., “A novel integrated optical dissolved oxygen sensor for cell culture and micro total analysis systems”, IEEE Technical Digest MEMS International Conference Jan. 24, 2002, 4 pages. |
Chang-Yen et al., “A PDMS microfluidic spotter for fabrication of lipid microarrays”, IEEE 3rd EMBS Special Topic Conference May 12-15, 2005; 2 pages. |
Chang-Yen et al., “Design and fabrication of a multianalyte-capable optical biosensor using a multiphysics approach”, IEEE 3rd EMBS Special Topic Conference May 12-15, 2005; 2 pages. |
Chang-Yen et al., “A Novel PDMS Microfluidic Spotter for Fabrication of Protein Chips and Microarrays”, IEEE J of Microelectromech Sys. (2006) 15(5): 1145-1151. |
Chang-Yen et al., “Design, fabrication, and packaging of a practical multianalyte-capable optical biosensor,” J Microlith Microfab Microsyst. (2006) 5(2):021105 in 8 pages. |
Chang-Yen et al., “Spin-assembled nanofilms for gaseous oxygen sensing.” Sens Actuators B: Chemical (2007), 120(2):426-433. |
Chaudhari et al., “Transient Liquid Crystal Thermometry of Microfabricated PCR Vessel Arrays”, J Microelectro Sys., (1998) 7(4):345-355. |
Chen P-C., “Accelerating micro-scale PCR (polymerase chain reactor) for modular lab-on-a-chip system”, LSU Master's Theses—Digital Commons, (2006) 111 pages. |
Chen et al., “Total nucleic acid analysis integrated on microfluidic devices,” Lab on a Chip. (2007) 7:1413-1423. |
Cheng et al., “Biochip-Based Portable Laboratory”, Biochip Tech. (2001):269-289. |
Cho et al., “A facility for characterizing the steady-state and dynamic thermal performance of microelectromechanical system thermal switches”, Rev Sci Instrum. (2008) 79(3):034901-1 to -8. |
Chong et al., “Disposable Polydimethylsiloxane Package for ‘Bio-Microfluidic System’”, IEEE Proceedings Electronic Components and Technology (2005); 5 pages. |
Chou et al., “A miniaturized cyclic PCR device—modeling and experiments”, Microelec Eng. (2002) 61-62:921-925. |
Christel et al., “Nucleic Acid Concentration and PCR for Diagnostic Applications”, in Micro Total Analysis Systems. (1998) D.J. Harrison et al. [Eds.] pp. 277-280. |
Christel et al., “Rapid, Automated Nucleic Acid Probe Assays Using Silicon Microstructures for Nucleic Acid Concentration”, J Biomech Eng. (1999) 121(1):22-27. |
Christensen et al., “Characterization of interconnects used in PDMS microfluidic systems”, J Micromech Microeng. (2005) 15:928 in 8 pages. |
Chung, Y. et al., “Microfluidic chip for high efficiency DNA extraction”, Miniaturisation for Chemistry, Biology & Bioengineering, vol. 4, No. 2 (Apr. 2004), pp. 141-147. |
Cooley et al., “Applications of Ink-Jet Printing Technology to BioMEMS and Microfluidic Systems”, Proceedings, SPIE Conference on Microfluids and BioMEMS, (Oct. 2001), 12 pages. |
Crews et al, “Rapid Prototyping of a Continuous-Flow PCR Microchip”, Proceedings of the AiChE Annual Meeting(Nov. 15, 2006) (335a) 3 pages. |
Crews et al., Thermal gradient PCR in a continuous-flow microchip. In Microfluidics, BioMEMS, and Medical Microsystems V; Jan. 2007; vol. 6465, p. 646504; 12 pages. |
Crews et al., “Continuous-flow thermal gradient PCR”, Biomed Microdevices. (2008) 10(2):187-195. |
Cui et al., “Electrothermal modeling of silicon PCR chips”, In MEMS Design, Fabrication, Characterization, and Packaging, (Apr. 2001) (vol. 4407, pp. 275-280. |
Cui et al., “Design and Experiment of Silicon PCR Chips,” Proc. SPIE 4755, Design, Test, Integration, and Packaging of MEMS/MOEMS 2002, (Apr. 19, 2002) pp. 71-76. |
Danaher Press Release: “Danaher to Acquire Cepheid for $53.00 per share, or approximately $4 Billion,” dated Sep. 6, 2016, accessed at www.danaher.com, pp. 3. |
Demchenko A.P., “The problem of self-calibration of fluorescence signal in microscale sensor systems”, Lab Chip. (2005) 5(11):1210-1223. |
Dineva et al., “Sample preparation: a challenge in the development of point-of-care nucleic acid-based assays for resource-limited settings”, Analyst. (2007) 132(12):1193-1199. |
Dishinger et al., “Multiplexed Detection and Applications for Separations on Parallel Microchips”, Electrophoresis. (2008) 29(16):3296-3305. |
Dittrich et al., “Single-molecule fluorescence detection in microfluidic channels—the Holy Grail in muTAS?”, Anal Bioanal Chem. (2005) 382(8):1771-1782. |
Dittrich et al., “Lab-on-a-chip: microfluidics in drug discovery”, Nat Rev Drug Discov. (2006) 5(3):210-218. |
Dunnington et al., “Approaches to Miniaturized High-Throughput Screening of Chemical Libraries”, in Integrated Microfabricated Devices, (2002) Ch. 15, pp. 371-414, CRC Press. |
Eddings et al., “A PDMS-based gas permeation pump for on-chip fluid handling in microfluidic devices”, J Micromech Microengin. (2006) 16(11):2396-2402. |
Edwards, “Silicon (Si),” in “Handbook of Optical Constants of Solids” (Ghosh & Palik eds., 1997) in 24 pages. |
Edwards et al., “Micro Scale Purification Systems for Biological Sample Preparation”, Biomed Microdevices (2001) 3(3):211-218. |
Edwards et al., “A microfabricated thermal field-flow fractionation system”, Anal Chem. (2002) 74(6):1211-1216. |
Ehrlich et al., “Microfluidic devices for DNA analysis”, Trends Biotechnol. (1999) 17(8):315-319. |
El-Ali et al., “Simulation and experimental validation of a SU-8 based PCR thermocycler chip with integrated heaters and temperature sensor”, Sens Actuators A: Physical (2004) 110(1-3):3-10. |
EP Communication dated Aug. 9, 2006 for European Patent Application 02723636.3, filed Mar. 27, 2002. |
Erickson et al., “Joule heating and heat transfer in poly(dimethylsiloxane) microfluidic systems”, Lab Chip (2003) 3(3):141-149. |
Erickson et al., “Integrated Microfluidic Devices”, Analytica Chim Acta. (2004) 507:11-26. |
Erill et al., “Development of a CMOS-compatible PCR chip: comparison of design and system strategies”, J Micromech Microengin. (2004) 14(11):1-11. |
European Supplemental Search Report dated Aug. 5, 2010 for Application No. EP 08826342.1, filed Jul. 11, 2008. |
European Supplemental Search Report dated Jul. 13, 2012 for Application No. EP 08843060.8, filed Jul. 14, 2008. |
European Extended Search Report dated Feb. 16, 2017 for Application No. EP 16191793.5, filed Sep. 30, 2016. |
European Extended Search Report dated May 11, 2017 for Application No. EP 16191787.7, filed Sep. 30, 2016. |
Fair R.B., Digital microfluidics: is a true lab-on-a-chip possible? Microfluidics Nanofluid. (2007) 3:245-281. |
Fan et al., “Integrated Plastic Microfluidic Devices for Bacterial Detection”, in Integrated Biochips for DNA Analysis by Liu et al. [Eds], (2007) Chapter 6, pp. 78-89. |
Fiorini et al., “Disposable microfluidic devices: fabrication, function, and application”, Biotechniques (2005) 38(3):429-446. |
Frazier et al., “Integrated micromachined components for biological analysis systems”, J Micromech. (2000) 1(1):67-83. |
Gale et al., “Micromachined electrical field-flow fractionation (mu-EFFF) system”, IEEE Trans Biomed Eng. (1998) 45(12):1459-1469. |
Gale et al., “Geometric scaling effects in electrical field flow fractionation. 1. Theoretical analysis”, Anal Chem. (2001) 73(10):2345-2352. |
Gale et al., “BioMEMS Education at Louisiana Tech University”, Biomed Microdevices, (2002) 4:223-230. |
Gale et al., “Geometric scaling effects in electrical field flow fractionation. 2. Experimental results”, Anal Chem. (2002) 74(5):1024-1030. |
Gale et al., “Cyclical electrical field flow fractionation”, Electrophoresis. (2005) 26(9):1623-1632. |
Gale et al., “Low-Cost MEMS Technologies”, Elsevier B.V. (2008), Chapter 1.12; pp. 342-372. |
Garst et al., “Fabrication of Multilayered Microfluidic 3D Polymer Packages”, IEEE Proceedings Electronic Components & Tech, Conference May/Jun. 2005, pp. 603-610. |
Gärtner et al., “Methods and instruments for continuous-flow PCR on a chip”, Proc. SPIE 6465, Microfluidics, BioMEMS, and Medical Microsystems V, (2007) 646502; 8 pages. |
Giordano et al., “Toward an Integrated Electrophoretic Microdevice for Clinical Diagnostics”, in Integrated Microfabricated Biodevices: Advanced Technologies for Genomics, Drug Discovery, Bioanalysis, and Clinical Diagnostics (2002) Chapter 1; pp. 1-34. |
Goldmeyer et al., “Identification of Staphylococcus aureus and Determination of Methicillin Resistance Directly from Positive Blood Cultures by Isothermal Amplification and a Disposable Detection Device”, J Clin Microbiol. (Apr. 2008) 46(4): 1534-1536. |
Graff et al., “Nanoparticle Separations Using Miniaturized Field-flow Fractionation Systems”, Proc. Nanotechnology Conference and Trade Show (NSTI) (2005); pp. 8-12. |
Greer et al., “Comparison of glass etching to xurography prototyping of microfluidic channels for DNA melting analysis”, J Micromech Microengin. (2007) 17(12):2407-2413. |
Grunenwald H., “Optimization of Polymerase Chain Reactions,” in Methods in Molecular Biology, PCR Protocols., Second Edition by Bartlett et al. [Eds.] Humana Press (2003) vol. 226, pp. 89-99. |
Guijt et al., “Chemical and physical processes for integrated temperature control in microfluidic devices”, Lab Chip. (2003) 3(1):1-4. |
Gulliksen A., “Microchips for Isothermal Amplification of RNA”, Doctoral Thesis (2007); Department of Mol. Biosciences—University of Oslo; 94 pages. |
Guttenberg et al., “Planar chip device for PCR and hybridization with surface acoustic wave pump”, Lab Chip. (2005) 5(3):308-317. |
Haeberle et al., “Microfluidic platforms for lab-on-a-chip applications”, Lab Chip. (2007) 7(9):1094-1110. |
Hale et al., “Optical constants of Water in the 200-nm to 200-μm Wavelength Region”, Applied Optics, 12(3): 555-563 (1973). |
Handal et al., “DNA mutation detection and analysis using miniaturized microfluidic systems”, Expert Rev Mol Diagn. (2006) 6(1):29-38. |
Handique et al, “Microfluidic flow control using selective hydrophobic patterning”, SPIE, (1997) 3224: 185-194. |
Handique et al., “On-Chip Thermopneumatic Pressure for Discrete Drop Pumping”, Anal. Chem., (2001) 73(8):1831-1838. |
Handique et al., “Nanoliter-volume discrete drop injection and pumping in microfabricated chemical analysis systems”, Solid-State Sensor and Actuator Workshop (Hilton Head, South Carolina, Jun. 8-11, 1998) pp. 346-349. |
Handique et al., “Mathematical Modeling of Drop Mixing in a Slit-Type Microchannel”, J. Micromech. Microeng., 11:548-554 (2001). |
Handique et al., “Nanoliter Liquid Metering in Microchannels Using Hydrophobic Patterns”, Anal. Chem., 72(17):4100-4109 (2000). |
Hansen et al., “Microfluidics in structural biology: smaller, faster . . . better”, Curr Opin Struct Biol. (2003) 13(5):538-544. |
Harding et al., “DNA isolation using Methidium-Spermine-Sepharose”, Meth Enzymol. (1992) 216: 29-39. |
Harding et al., “Rapid isolation of DNA from complex biological samples using a novel capture reagent—methidium-spermine-sepharose”, Nucl Acids Res. (1989) 17(17): 6947-6958. |
Harrison et al., “Capillary Electrophoresis and Sample Injection Systems Integrated on a Planar Glass Chip”, Anal. Chem., (1992) 64: 1926-1932. |
He et al., Microfabricated Filters for Microfluidic Analytical Systems, Analytical Chemistry, American Chemical Society, 1999, vol. 71, No. 7, pp. 1464-1468. |
Heid et al., “Genome Methods—Real Time Quantitative PCR”, Genome Res. (1996) 6(10):986-994. |
Henry C.S. [Ed], “Microchip Capillary electrophoresis”, Methods in Molecular Biology, Humana Press 339 (2006) Parts I-IV in 250 pages. |
Herr et al., “Investigation of a miniaturized capillary isoelectric focusing (cIEF) system using a full-field detection approach”, Solid State Sensor and Actuator Workshop, Hilton Head Island (2000), pp. 4-8. |
Herr et al., “Miniaturized Isoelectric Focusing (μIEF) As a Component of a Multi-Dimensional Microfluidic System”, Micro Total Analysis Systems (2001) pp. 51-53. |
Herr et al., Miniaturized Capillary Isoelectric Focusing (cIEF): Towards a Portable High-Speed Separation Method. In Micro Total Analysis Systems (2000) Springer, Dordrecht; pp. 367-370. |
Holland et al., “Point-of-care molecular diagnostic systems—past, present and future”, Curr Opin Microbiol. (2005) 8(5):504-509. |
Hong et al., “Integrated nanoliter systems”, Nat Biotechnol. (2003) 21(10):1179-1183. |
Hong et al., “Molecular biology on a microfluidic chip”, J Phys.: Condensed Matter (2006) 18(18):S691-S701. |
Hong et al., “Integrated Nucleic Acid Analysis in Parallel Matrix Architecture”, in Integrated Biochips for DNA Analysis by Liu et al. [Eds], (2007) Chapter 8, pp. 107-116. |
Horsman et al., “Forensic DNA Analysis on Microfluidic Devices: A Review”, J Forensic Sci. (2007) 52(4):784-799. |
Hsieh et al., “Enhancement of thermal uniformity for a microthermal cycler and its application for polymerase chain reaction”, Sens Actuators B: Chemical. (2008) 130(2):848-856. |
Hsueh et al., “A microfabricated, electrochemiluminescence cell for the detection of amplified DNA” Proc. 1995 IEEE Int. Conf. Solid-State Sens. Actuators (1995) pp. 768-771. |
Hsueh et al., “DNA quantification with an electrochemiluminescence microcell” Proc. 1997 IEEE Int. Conf. Solid-State Sens. Actuators (1997) pp. 175-178. |
Huang et al., “Temperature Uniformity and DNA Amplification Efficiency in Micromachined Glass PCR Chip”, TechConnect Briefs; Tech Proc. of the 2005 NSTI Nanotechnology Conference and Trade Show. (2005) vol. 1:452-455. |
Huebner et al., “Microdroplets: A sea of applications?”, Lab Chip. (2008) 8(8):1244-1254. |
Ibrahim, et al., Real-Time Microchip PCR for Detecting Single-Base Differences in Viral and Human DNA, Analytical Chemistry, American Chemical Society, 1998, 70(9): 2013-2017. |
International Preliminary Report on Patentability and Written Opinion dated Jan. 19, 2010 for Application No. PCT/US2008/008640, filed Jul. 14, 2008. |
International Preliminary Report on Patentability dated Jan. 19, 2010 for Application No. PCT/US2008/069897, filed Jul. 11, 2008. |
International Search Report and Written Opinion dated Apr. 4, 2008 for PCT/US2007/007513, filed Mar. 26, 2007. |
International Search Report and Written Opinion dated Jan. 5, 2009 for PCT/US2007/024022, filed Nov. 14, 2007. |
International Search Report and Written Opinion, dated Oct. 3, 2008, issued in International Application No. PCT/US2008/069897, filed Jul. 11, 2008. |
International Search Report dated Jun. 17, 2009 for Application No. PCT/US2008/008640, filed Jul. 14, 2008. |
Iordanov et al., “PCR Array on Chip—Thermal Characterization”, IEEE Sensors (2003) Conference Oct. 22-24, 2003; pp. 1045-1048. |
Irawan et al., “Cross-Talk Problem on a Fluorescence Multi-Channel Microfluidic Chip System,” Biomed Micro. (2005) 7(3):205-211. |
Ji et al., “DNA Purification Silicon Chip”, Sensors and Actuators A: Physical (2007) 139(1-2):139-144. |
Jia et al., “A low-cost, disposable card for rapid polymerase chain reaction”, Colloids Surfaces B: Biointerfaces (2007) 58:52-60. |
Jiang et al., “Directing cell migration with asymmetric micropatterns” Proc. Natl. Acad. Sci. USA (2005) 102, 975-978. |
Kaigala et al., “An inexpensive and portable microchip-based platform for integrated RT-PCR and capillary electrophoresis”, The Analyst (2008) 133(3):331-338. |
Kajiyama et al., “Genotyping on a Thermal Gradient DNA Chip”, Genome Res. (2003) 13(3):467-475. |
Kang et al., “Simulation and Optimization of a Flow-Through Micro PCR Chip”, NSTI—Nanotech (2006) vol. 2, pp. 585-588. |
Kantak et al.,“Microfluidic platelet function analyzer for shear-induced platelet activation studies”, 2nd Annual International IEEE—EMBS Special Topic Conference on Microtechnologies in Med and Biol. (May 2002) 5 pages. |
Kantak et al., “Microfabricated cyclical electrical field flow fractionation”, 7th International Conference on Miniaturized Chomical and Biochem Analysis Sys. (2003) pp. 1199-1202. |
Kantak et al., “Platelet function analyzer: Shear activation of platelets in microchannels”, Biomedical Microdevices (2003) 5(3):207-215. |
Kantak et al., “Characterization of a microscale cyclical electrical field flow fractionation system”, Lab Chip. (2006) 6(5):645-654. |
Kantak et al., “Effect of carrier ionic strength in microscale cyclical electrical field-flow fractionation”, Anal Chem. (2006) 78(8):2557-2564. |
Kantak et al., “Improved theory of cyclical electrical field flow fractions”, Electrophoresis (2006) 27(14):2833-2843. |
Karunasiri et al.,“Extraction of thermal parameters of microbolometer infrared detectors using electrical measurement”, SPIE's Inter'l Symposium on Optical Science, Engineering, and Instrumentation; Proceedings (1998) vol. 3436, Infrared Technology and Applications XXIV; (1998) 8 pages. |
Kelly et al., “Microfluidic Systems for Integrated, High-Throughput DNA Analysis,” Analytical Chemistry, (2005), 97A-102A, Mar. 1, 2005, in 7 pages. |
Khandurina et al., Microfabricated Porous Membrane Structure for Sample Concentration and Electrophoretic Analysis, Analytical Chemistry American Chemical Society, 1999, 71(9): 1815-1819. |
Khandurina et al., “Bioanalysis in microfluidic devices,” J Chromatography A, (2002) 943:159-183. |
Kim et al., “Reduction of Microfluidic End Effects in Micro-Field Flow Fractionation Channels”, Proc. MicroTAS 2003, pp. 5-9. |
Kim et al., “Multi-DNA extraction chip based on an aluminum oxide membrane integrated into a PDMS microfluidic structure”, 3rd IEEE/EMBS Special Topic Conference on Microtechnology in Med and Biol. (May 2005). |
Kim et al., “Electrohydrodynamic Generation and Delivery of Monodisperse Picoliter Droplets Using a Poly(dimethylsiloxane) Microchip”, Anal Chem. (2006) 78: 8011-8019. |
Kim et al., “Geometric optimization of a thin film ITO heater to generate a uniform temperature distribution”, (2006), Tokyo, Japan; pp. 293-295; Abstract. |
Kim et al., “Micro-Raman thermometry for measuring the temperature distribution inside the microchannel of a polymerase chain reaction chip”, J Micromech Microeng. (2006) 16(3):526-530. |
Kim et al., “Patterning of a Nanoporous Membrane for Multi-sample DNA Extraction”, J Micromech Microeng. (2006) 16:33-39. |
Kim et al., “Performance evaluation of thermal cyclers for PCR in a rapid cycling condition”, Biotechniques. (2008) 44(4):495-505. |
Kim et al., “Quantitative and qualitative analysis of a microfluidic DNA extraction system using a nanoporous AIO(x) membrane”, Lab Chip. (2008) 8(9):1516-1523. |
Kogi et al., “Microinjection-microspectroscopy of single oil droplets in water: an application to liquid/liquid extraction under solution-flow conditions”, Anal Chim Acta. (2000)—418(2):129-135. |
Kopf-Sill et al., “Creating a Lab-on-a-Chip with Microfluidic Technologies”, in Integrated Microfabricated Biodevices: Advanced Technologies for Genomics, Drug Discovery, Bioanalysis, and Clinical Diagnostics (2002) Chapter 2; pp. 35-54. |
Kopp et al., Chemical Amplification: Continuous-Flow PCR on a Chip, www.sciencemag.org, 1998, vol. 280, pp. 1046-1048. |
Kricka L.J., “Microchips, Bioelectronic Chips, and Gene Chips—Microanalyzers for the Next Century”, in Biochip Technology by Cheng et al. [Eds]; (2006) Chapter 1, pp. 1-16. |
Krishnan et al., “Polymerase chain reaction in high surface-to-volume ratio SiO2 microstructures”, Anal Chem. (2004) 76(22):6588-6593. |
Kuo et al., “Remnant cationic dendrimers block RNA migration in electrophoresis after monophasic lysis”, J Biotech. (2007) 129: 383-390. |
Kuswandi et al., “Optical sensing systems for microfluidic devices: a review”, Anal Chim Acta. (2007) 601(2):141-155. |
Kutter et al., Solid Phase Extraction on Microfluidic Devices, J. Microcolumn Separations, John Wiley & Sons, Inc., 2000, 12(2): 93-97. |
Labchem; Sodium Hydroxide, 0,5N (0.5M); Safety Data Sheet, 2015; 8 pages. |
Lagally et al., “Monolithic integrated microfluidic DNA amplification and capillary electrophoresis analysis system” Sensors and Actuators B (2000) 63:138-146. |
Lagally et al., Single-Molecule DNA Amplification and Analysis in an Integrated Microfluidic Device, Analytical Chemistry, American Chemical Society, 2001, 73(3): 565-570. |
Lagally et al., “Genetic Analysis Using Portable PCR-CE Microsystem”, Proceedings 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems (2003) pp. 1283-1286. |
Lagally et al., “Integrated portable genetic analysis microsystem for pathogen/infectious disease detection”, Anal Chem. (2004) 76(11):3152-3170. |
Lauerman L.H., “Advances in PCR technology”, Anim Health Res Rev. (2004) 5(2):247-248. |
Lawyer et al., “High-level Expression, Purification, and Enzymatic Characterization of Full-length Thermus aquaticus DNA Polymerase and a Truncated Form Deficient in 5′to 3′Exonuclease Activity.” Genome research (1993) 2(4):275-287. |
Lee et al., “Submicroliter-volume PCR chip with fast thermal response and very power consumption”, 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems, (2003) pp. 187-190. |
Lee et al., “Bulk-micromachined submicroliter-volume PCR chip with very rapid thermal response and low power consumption”, Lab Chip. (2004) 4(4):401-407. |
Lewin et al., “Use of Real-Time PCR and Molecular Beacons to Detect Virus Replication in Human Immunodeficiency Virus Type 1-infected Individuals on Prolonged Effective Antiretroviral Therapy”. J Virol. (1999) 73(7), 6099-6103. |
Li et al., “Effect of high-aspect-ratio microstructures on cell growth and attachment”, 1st Annual Inter'l IEEE—EMBS Special Topic Conference on Microtechnologies in Med and Biol. Proceedings Cat. No. 00EX451; (Oct. 2000) Poster 66, pp. 531-536. |
Li PCH., “Micromachining Methods et al.” in Microfluidic Lab-on-a-Chip for Chemical and Biological Analysis and Discovery, CRC Press (2005), Chapter 2-3 to 2-5; pp. 10-49. |
Li PCH., “Microfluidic Flow” in Microfluidic Lab-on-a-Chip for Chemical and Biological Analysis and Discovery, CRC Press (2005), Chapter 3, pp. 55-99. |
Li PCH., “Detection Methods” in Microfluidic Lab-on-a-Chip for Chemical and Biological Analysis and Discovery, CRC Press (2005), Chapter 7, pp. 187-249. |
Li PCH., “Applications to Nucleic Acids Analysis” in Microfluidic Lab-on-a-Chip for Chemical and Biological Analysis and Discovery, CRC Press (2005), Chapter 9; pp. 293-325. |
Li et al., “A Continuous-Flow Polymerase Chain Reaction Microchip With Regional Velocity Control”, J Microelectromech Syst. (2006) 15(1):223-236. |
Liao et al., “Miniature RT-PCR system for diagnosis of RNA-based viruses,” Nucl Acids Res. (2005) 33(18):e156 in 7 pages. |
Lien et al., “Integrated reverse transcription polymerase chain reaction systems for virus detection”, Biosens Bioelectron. (2007) 22(8):1739-1748. |
Lien et al., “Microfluidic Systems Integrated with a Sample Pretreatment Device for Fast Nucleic-Acid Amplification”, J Microelectro Sys. (2008) 17(2):288-301. |
Lifesciences et al., “Microfluidics in commercial applications; an industry perspective.” Lab Chip (2006) 6:1118-1121. |
Lin et al., “Thermal Uniformity of 12-in Silicon Wafer During Rapid Thermal Processing by Inverse Heat Transfer Method,” IEEE Transactions on Semiconductor Manufacturing, (2000) 13(4):448-456. |
Lin et al., “Simulation and experimental validation of micro polymerase chain reaction chips”, Sens Actuators B: Chemical. (2000) 71(1-2):127-133. |
Linder et al., “Microfluidics at the Crossroad with Point-of-care Diagnostics”, Analyst (2007) 132:1186-1192. |
Liu et al., “Integrated portable polymerase chain reaction-capillary electrophoresis microsystem for rapid forensic short tandem repeat typing”, Anal Chem. (2007) 79(5):1881-1889. |
Liu et al. [Eds], Integrated Biochips for DNA Analysis—Biotechnology Intelligence Unit; Springer/Landes Bioscience (2007) ISBN:978-0-387-76758-1; 216 pages. |
Livache et al., “Polypyrrole DNA chip on a Silicon Device: Example of Hepatitis C Virus Genotyping”, Analytical Biochemistry, (1998) 255: 188-194. |
Locascio et al., “ANYL 67 Award Address—Microfluidics as a tool to enable research and discovery in the life sciences”, Abstract; The 236th ACS National Meeting (Aug. 2008); 2 pages. |
Mahjoob et al., “Rapid microfluidic thermal cycler for polymerase chain reaction nucleic acid amplification”, Inter'l J Heat Mass Transfer. (2008) 51(9-10):2109-2122. |
Malitson, “Interspecimen Comparison of the Refractive Index of Fused Silica,” J Optical Society of America, 55:1205-1209 (1965). |
Manz et al., “Miniaturized Total Chemical Analysis Systems: a Novel Concept for Chemical Sensing,” Sensors and Actuators B1, (1990) 244-248. |
Manz et al., “Design of an open-tubular column liquid chromatograph using silicon chip technology” Sensors and Actuators B (1990) 1:249-255. |
Manz et al., “Planar chips technology for miniaturization and integration of separation techniques into monitoring systems: Capillary electrophoresis on a chip” Journal of Chromatography A (1992) 593:253-258. |
Marcus et al., “Parallel picoliter rt-PCR assays using microfluidics”, Anal Chem. (2006) 78(3):956-958. |
Mariella R.P. Jr., “Microtechnology”, Thrust Area Report FY 96 UCRL-ID-125472; Lawrence Livermore National Lab., CA (Feb. 1997) Chapter 3 in 44 pages. |
Mariella R., “Sample preparation: the weak link in microfluidics-based biodetection”, Biomed Microdevices. (2008) 10(6):777-784. |
Mastrangelo et al., Microfabricated Devices for Genetic Diagnostics. Proceedings of the IEEE (1998) 86(8):1769-1787. |
Mascini et al., “DNA electrochemical biosensors”, Fresenius J. Anal. Chem., 369: 15-22, (2001). |
McMillan et al., “Application of advanced microfluidics and rapid PCR to analysis of microbial targets”, In Proceedings of the 8th international symposium on microbial ecology (1999), in 13 pages. |
Melin et al., “Microfluidic large-scale integration: the evolution of design rules for biological automation”, Annu Rev Biophys Biomol Struct. (2007) 36:213-231. |
Merugu et al., “High Throughput Separations Using a Microfabricated Serial Electric Split System” (2003), Proceedings of μTAS 2003, 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems, Oct. 5-9, 2003, Squaw Valley, California; 1191-1194, in 3 pages. |
Meyers, R.A., Molecular Biology and Biotechnology: A Comprehensive Desk Reference; VCH Publishers, Inc. New York, NY; (1995) pp. 418-419. |
Miao et al., “Low cost micro-PCR array and micro-fluidic integration on single silicon chip”, Int'l J Comput Eng Science (2003) 4(2):231-234. |
Miao et al., “Flip-Chip packaged micro-plate for low cost thermal multiplexing”, Int'l J Comput Eng Science. (2003) 4(2):235-238. |
Micheletti et al., “Microscale Bioprocess Optimisation”, Curr Opin Biotech. (2006) 17:611-618. |
MicroTAS 2005., “Micro Total Analysis Systems”, Proceedings 9th Int. Conference on Miniaturized Systems for Chemistry and Life Sciences; Presentations/Posters/Articles for Conference; Boston, MA in Oct. 10-12, 2005 in 1667 pages. |
MicroTAS 2007., “Micro Total Analysis Systems”, Proceedings 11th Int. Conference on Miniaturized Systems for Chemistry and Life Sciences; Presentations/Posters/Articles for Conference; Paris, France in Oct. 7-11, 2007 in 1948 pages. |
MicroTAS 2007., “Micro Total Analysis Systems”, Advance Program for the Proceedings 11th Int. Conference on Miniaturized Systems for Chemistry and Life Sciences; Presentations/Posters/Articles for Conference; Paris, France in Oct. 7-11, 2007 in 42 pages. |
Minco, “Conductive Heating Technologies for Medical Diagnostic Equipment,” (2006) in 13 pages. |
Mitchell et al., “Modeling and validation of a molded polycarbonate continuous-flow polymerase chain reaction device,” Microfluidics, BioMEMS, and Medical Microsystems, Proc. SPIE (2003) 4982:83-98. |
Myers et al., “Innovations in optical microfluidic technologies for point-of-care diagnostics”, Lab Chip (2008) 8:2015-2031. |
Nakagawa et al., Fabrication of amino silane-coated microchip for DNA extraction from whole blood, J of Biotechnology, Mar. 2, 2005, 116: 105-111. |
Namasivayam et al., “Advances in on-chip photodetection for applications in miniaturized genetic analysis systems”, J Micromech Microeng. (2004) 14:81-90. |
Narayanan et al., “A microfabricated electrical Splitt system,” Lab Chip, (2006) 6:105-114. |
Neuzil et al., “Disposable real-time microPCR device: lab-on-a-chip at a low cost, ” Mol. Biosyst., (2006) 2:292-298. |
Neuzil et al., “Ultra fast miniaturized real-time PCR: 40 cycles in less than six minutes,” Nucleic Acids Research, (2006) 34(11)e77, in 9 pages. |
Nguyen et al. [Eds], “Microfluidics for Internal Flow Control: Microfluidics” in Fundamentals and Applications of Microfluidics; 2nd Edition (2006) Introduction Chapter 1, pp. 1-9. |
Nguyen et al. [Eds], “Microfluidics for Internal Flow Control: Microvalves” in Fundamentals and Applications of Microfluidics; (2006) 2nd Edition, Chapter 6, pp. 211-254. |
Nguyen et al. [Eds], “Microfluidics for Internal Flow Control: Micropumps” in Fundamentals and Applications of Microfluidics; (2006) 2nd Edition, Chapter 7, pp. 255-309. |
Nguyen et al. [Eds], “Microfluidics for Life Sciences and Chemistry: Microdispensers” in Fundamentals and Applications of Microfluidics; (2006) , Chapter 11, pp. 395-418. |
Nguyen et al. [Eds], “Microfluidics for Life Sciences and Chemistry: Microreactors” in Fundamentals and Applications of Microfluidics; (2006) 2nd Edition, Chapter 13, pp. 443-477. |
Ning et al., “Microfabrication Processes for Silicon and Glass Chips”, in Biochip Technology, CRC-Press (2006) Chapter 2, pp. 17-38. |
Northrup et al., “A MEMS-based Miniature DNA Analysis System,” Lawrence Livermore National Laboratory, (1995), submitted to Transducers '95, Stockholm, Sweden, Jun. 25 29, 1995, in 7 pages (Prepublication). |
Northrup et al., “A MEMS-based Miniature DNA Analysis System.” Transducers '95—Eurosensors in Proc. 1995 (8th) IEEE Int. Conf. Solid-State Sens. Actuators, pp. 764-767. |
Northrup et al., “Advantages Afforded by Miniaturization and Integration of DNA Analysis Instrumentation,” Microreaction Technology, (1998) 278-288. |
Northrup et al., A Miniature Analytical Instrument for Nucleic Acids Based on Micromachined Silicon Reaction Chambers, Analytical Chemistry, American Chemical Society, 1998, 70(5): 918-922. |
Northrup et al., “A New Generation of PCR Instruments and Nucleic Acid Concentration Systems,” in PCR Applications: Protocols for Functional Genomics, (1999), Chapter 8, pp. 105-125. |
Northrup, “Microfluidics, A few good tricks,” Nature materials (2004), 3:282-283. |
Northrup et al.,“Microfluidics-based integrated airborne pathogen detection systems,” Abstract, Proceedings of the SPIE, (2006), vol. 6398, Abstract in 2 pages. |
Oh et al., “World-to-chip microfluidic interface with built-in valves for multichamber chip-based PCR assays,” Lab Chip, (2005), 5:845-850. |
Oh K.W. et al., “A Review of Microvalves”, J Micromech Microeng. (2006) 16:R13-R39. |
Ohno et al., “Microfluidics: Applications for analytical purposes in chemistry and biochemistry,” Electrophoresis (2008), 29:4443-4453. |
Oleschuk et al., Trapping of Bead-Based Reagents within Microfluidic Systems: On-Chip Solid-Phase Extraction and Electrochromatography, Analytical Chemistry, American Chemical Society, 2000, 72(3): 585-590. |
Pal et al., “Phase Change Microvalve for Integrated Devices”, Anal Chem. (2004) 76: 3740-3748. |
Pal et al., “An integrated microfluidic for influenza and other genetic analyses,” Lab Chip, (2005), 5:1024-1032. |
Palina et al., “Laser Assisted Boron Doping of Silicon Wafer Solar Cells Using Nanosecond and Picosecond Laser Pulses,” 2011 37th IEEE Photovoltaic Specialists Conference, pp. 002193-002197, IEEE (2011). |
Pamme, “Continuous flow separations in microfluidic devices,” Lab Chip, (2007), 7:1644-1659. |
Pang et al., “A novel single-chip fabrication technique for three-dimensional MEMS structures,” Institute of Microelectronics, Tsinghua University, Beijing, P.R. China, (1998), IEEE, 936-938. |
Pang et al., “The Study of Single-Chip Integrated Microfluidic System,” Tsinghua University, Beijing, P.R. China, (1998), IEEE, 895-898. |
Papautsky et al., “Effects of rectangular microchannel aspect ratio on laminar friction constant”, in Microfluidic Devices and Systems II (1999) 3877:147-158. |
Paulson et al., “Optical dispersion control in surfactant-free DNA thin films by vitamin B2 doping,” Nature, Scientific Reports 8:9358 (2018) published at www.nature.com/scientificreports, Jun. 19, 2018. |
Petersen, Kurt E., “Silicon as a Mechanical Material.” Proceedings of the IEEE, (May 1982) 70(5):420-457. |
Petersen et al., “Toward Next Generation Clinical Diagnostic Instruments: Scaling and New Processing Paradigms,” Biomedical Microdevices (1998) 1(1):71-79. |
Picard et al., Laboratory Detection of Group B Streptococcus for Prevention of Perinatal Disease, Eur. J. Clin. Microbiol. Infect. Dis., Jul. 16, 2004, 23: 665-671. |
Plambeck et al., “Electrochemical Studies of Antitumor Antibiotics”, J. Electrochem Soc.: Electrochemical Science and Technology (1984), 131(11): 2556-2563. |
Poser et al., “Chip elements for fast thermocycling,” Sensors and Actuators A, (1997), 62:672-675. |
Pourahmadi et al., “Toward a Rapid, Integrated, and Fully Automated DNA Diagnostic Assay for Chlamydia trachomatis and Neisseria gonorrhea,” Clinical Chemistry, (2000), 46(9):1511-1513. |
Pourahmadi et al., “Versatile, Adaptable and Programmable Microfluidic Platforms for DNA Diagnostics and Drug Discovery Assays,” Micro Total Analysis Systems, (2000), 243-248. |
Raisi et al., “Microchip isoelectric focusing using a miniature scanning detection system,” Electrophoresis, (2001), 22:2291-2295. |
Raja et al., “Technology for Automated, Rapid, and Quantitative PCR or Reverse Transcription—PCR Clinical Testing,” Clinical Chemistry, (2005), 51(5):882-890. |
Reyes et al., “Micro Total Analysis Systems. 1. Introduction, Theory, and Technology”, Anal Chem (2002) 74:2623-2636. |
Rhee et al., “Drop Mixing in a Microchannel for Lab-on-a-Chip Applications” Langmuir (2008) 24 (2): 590-601. |
Roche et al. “Ectodermal commitment of insulin-producing cells derived from mouse embryonic stem cells” Faseb J (2005) 19: 1341-1343. |
Rodriguez et al., “Practical integration of polymerase chain reaction amplification and electrophoretic analysis in microfluidic devices for genetic analysis,” Electrophoresis, (2003), 24:172-178. |
Rohsenow et al. [Eds.], Handbook of Heat Transfer, 3rd Edition McGraw-Hill Publishers (1998) Chapters 1 & 3; pp. 108. |
Roper et al., “Advances in Polymer Chain Reaction on Microfluidic Chips,” Anal. Chem., (2005), 77:3887-3894. |
Ross et al., Analysis of DNA Fragments from Conventional and Microfabricated PCR Devices Using Delayed Extraction MALDI-TOF Mass Spectrometry, Analytical Chemistry, American Chemical Society, 1998, 70(10): 2067-2073. |
Ross et al., “Scanning Temperature Gradient Focusing for Simultaneous Concentration and Separation of Complex Samples,” Micro Total Analysis Systems 2005, vol. 2, (2005), Proceedings of μTAS 2005, Ninth International Conference on Miniaturized Systems for Chemistry and Life Sciences, Oct. 9-13, 2005, Boston, Massachusetts; 1022-1024. |
Ross et al., “Simple Device for Multiplexed Electrophoretic Separations Using Gradient Elution Moving Boundary Electrophoresis with Channel Current Detection,” Anal. Chem., (2008), 80(24):9467-9474. |
Sadler et al., “Thermal Management of BioMEMS: Temperature Control for Ceramic-Based PCR and DNA Detection Devices,” IEEE Transactions on Components and Packaging Technologies, (2003) 26(2):309-316. |
Sammarco et al., “Thermocapillary Pumping of Discrete Drops in Microfabricated Analysis Devices” AIChE Journal (1999) 45(2): 350-366. |
Sanchez et al., “Linear-After-The-Exponential (LATE)—PCR: An advanced method of asymmetric PCR and its uses in quantitative real-time analysis”, PNAS (2004) 101(7): 1933-1938. |
Sant et al., “An Integrated Optical Detector for Microfabricated Electrical Field Flow Fractionation System,” Proceedings of μTAS 2003, 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems, Oct. 5-9, 2003, Squaw Valley, California; pp. 1259-1262. |
Sant et al., “Geometric scaling effects on instrumental plate height in field flow fractionation”, J Chromatography A (2006) 1104:282-290. |
Sant H.J., “Reduction of End Effect-Induced Zone Broadening in Field-Flow Fractionation Channels”, Anal Chem. (2006) 78:7978-7985. |
Sant et al., “Microscale Field-Flow Fractionation: Theory and Practice”, in Microfluidic Technologies for Miniaturized Analysis Systems. (2007) Chapter 12, pp. 4710521. |
Schäferling et al., “Optical technologies for the read out and quality control of DNA and protein microarrays,” Anal Bioanal Chem, (2006), 385: 500-517. |
Serpengüzel et al., “Microdroplet identification and size measurement in sprays with lasing images”, Optics express (2002) 10(20):1118-1132. |
Shackman et al., “Gradient Elution Moving Boundary Electrophoresis for High-Throughput Multiplexed Microfluidic Devices,” Anal. Chem. (2007), 79(2), 565-571. |
Shackman et al., “Temperature gradient focusing for microchannel separations,” Anal Bioanal Chem, (2007), 387:155-158. |
Shadpour et al., “Multichannel Microchip Electrophoresis Device Fabricated in Polycarbonate with an Integrated Contact Conductivity Sensor Array,” Anal Chem., (2007), 79(3), 870-878. |
Shen et al., “A microchip-based PCR device using flexible printed circuit technology,” Sensors and Actuators B (2005), 105:251-258. |
Shoffner et al., Chip PCR.I. Surface Passivation of Microfabricated Silicon-Glass Chips for PCR, Nucleic Acids Research, Oxford University Press, (1996) 24(2): 375-379. |
Sia et al., “Microfluidic devices fabricated in poly(dimethylsiloxane) for biological studies,” Electrophoresis, (2003), 24:3563-3576. |
Sigurdson M., “AC Electrokinetic Enhancement for Assay Enhancement”, ProQuest LLC (2008) Doctoral Thesis UMI Microform 3319791 in 24 pages. |
Singh et al., “PCR thermal management in an integrated Lab on Chip,” Journal of Physics: Conference Series, (2006), 34:222-227. |
Situma et al., “Merging microfluidics with microarray-based bioassays”, Biomol Engin. (2006) 23:213-231. |
Smith, K. et al., “Comparison of Commercial DNA Extraction Kits for Extraction of Bacterial Genomic DNA from Whole-Blood Samples”, Journal of Clinical Microbiology, vol. 41, No. 6 (Jun. 2003), pp. 2440-2443. |
Smith et al., “(576d) Micropatterned fluid lipid bilayers created using a continuous flow microspotter for multi-analyte assays,” (2007), Biosensors II, 2007 AIChE Annual Meeting, Nov. 8, 2007, Abstract in 2 pages. |
Sommer et al., “Introduction to Microfluidics”, in Microfluidics for Biological Applications by Tian et al. [Eds] (2008) Chapter 1, pp. 1-34. |
Spitzack et al., “Polymerase Chain Reaction in Miniaturized Systems: Big Progress in Little Devices”, in Methods in Molecular Biology—Microfluidic Techniques, Minteer S.D. [Ed.] Humana Press (2006), Chapter 10, pp. 97-129. |
Squires et al., “Microfluidics: Fluid physics at the nanoliter scale”, Rev Modern Phys. (2005) 77(3):977-1026. |
Sundberg et al., “Solution-phase DNA mutation scanning and SNP genotyping by nanoliter melting analysis,” Biomed Microdevices, (2007), 9:159-166, in 8 pages. |
Tabeling, P. [Ed.], “Physics at the micrometric scale,” in Introduction to Microfluidics (2005) Chapter 1, pp. 24-69. |
Tabeling, P. [Ed.], “Hydrodynamics of Microfluidic Systems”, in Introduction to Microfluidics; (2005) Chapter 2, pp. 70-129. |
Tabeling, P. [Ed.], Introduction to Microfluidics; (2005) Chapters 5-7, pp. 216-297. |
Tanaka et al., “Improved Method of DNA Extraction from Seeds Using Amine-Dendrimer Modified Magnetic Particles”, Proceedings of the 74th Annual Meeting of the Electrochemical Society of Japan; Abstract #2E09 on p. 149, Mar. 29, 2007; Faculty of Engineering, Science University of Tokyo; 4 pages. |
Taylor et al., “Optimization of the performance of the polymerase chain reaction in silicon-based microstructures” Nucleic Acids Res. (1997) vol. 25, pp. 3164-3168. |
Taylor et al., Fully Automated Sample Preparation for Pathogen Detection Performed in a Microfluidic Cassette, in Micro Total Analysis Systems, Springer (2001), pp. 670-672. |
Taylor et al., “Lysing Bacterial Spores by Sonication through a Flexible Interface in a Microfluidic System,” Anal. Chem., (2001), 73(3):492-496. |
Taylor et al., “Microfluidic Bioanalysis Cartridge with Interchangeable Microchannel Separation Components,” (2001), The 11th International Conference on Solid-State Sensors and Actuators, Jun. 10-14, 2001, Munich, Germany; 1214-1247. |
Taylor et al., “Disrupting Bacterial Spores and Cells using Ultrasound Applied through a Solid Interface,” (2002), 2nd Annual International IEEE—EMBS Special Topic Conference on Microtechnologies in Medicine & Biology, May 2-4, 2002, Madison, Wisconsin; 551-555. |
Terry et al., “A Gas Chromatographic Air Analyzer Fabricated on a Silicon Wafer” IEEE T Electron Dev (1979) 26:1880-1886. |
Thorsen et al., “Microfluidic Large-scale integration,” Science, (2002), 298:580-584. |
Toriello et al., “Multichannel Reverse Transcription-Polymerase Chain Reaction Microdevice for Rapid Gene Expression and Biomarker Analysis,” Anal. Chem., (2006) 78(23):7997-8003. |
Ugaz et al., “Microfabricated electrophoresis systems for DNA sequencing and genotyping applications,” Phil. Trans. R. Soc. Lond. A, (2004), 362:1105-1129. |
Ugaz et al., “PCR in Integrated Microfluidic Systems”, in Integrated Biochips for DNA Analysis by Liu et al. [Eds]; (2007) Chapter 7, pp. 90-106. |
Ullman et al., “Luminescent oxygen channeling assay (LOCI™): sensitive, broadly applicable homogeneous immunoassay method”. Clin Chem. (1996) 42(9), 1518-1526. |
U.S. Appl. No. 60/491,264, filed Jul. 31, 2003 (41 pages). |
U.S. Appl. No. 60/491,269, filed Jul. 31, 2003 (52 pages). |
U.S. Appl. No. 60/491,539, filed Aug. 1, 2003 (45 pages). |
U.S. Appl. No. 60/553,553, filed Mar. 17, 2004 (49 pages). |
U.S. Appl. No. 60/726,066, filed Oct. 11, 2005 (54 pages). |
U.S. Appl. No. 60/786,007, filed Mar. 24, 2006 (223 pages). |
U.S. Appl. No. 60/859,284, filed Nov. 14, 2006 (114 pages). |
Velten et al., “Packaging of Bio-MEMS: Strategies, Technologies, and Applications,” IEEE Transactions on Advanced Packaging, (2005) 28(4):533-546. |
Vinet et al., “Microarrays and microfluidic devices: miniaturized systems for biological analysis,” Microelectronic Engineering, (2002), 61-62:41-47. |
Wang, “Survey and Summary, from DNA Biosensors to Gene Chips”, Nucleic Acids Research, 28(16):3011-3016, (2000). |
Wang et al., “From biochips to laboratory-on-a-chip system”, in Genomic Signal Processing and Statistics by Dougherty et al. [Eds]; (2005) Chapter 5, pp. 163-200. |
Wang et al., “A disposable microfluidic cassette for DNA amplification and detection”, Lab on a Chip (2006) 6(1):46-53. |
Wang et al., “Micromachined Flow-through Polymerase Chain Reaction Chip Utilizing Multiple Membrane-activated Micropumps,” (2006), MEMS 2006, Jan. 22-26, 2006, Istanbul, Turkey; 374-377. |
Waters et al., Microchip Device for Cell Lysis, Multiplex PCR Amplification, and Electrophoretic Sizing, Analytical Chemistry, American Chemical Society, 1998, 70(1): 158-162. |
Weigl, et al., Microfluidic Diffusion-Based Separation and Detection, www.sciencemag.org, 1999, vol. 283, pp. 346-347. |
Whitesides G.M., “The origins and the future of microfluidics” Nature (2006) 442(7101):368-373. |
Woias P., “Micropumps—past, progress and future prospects” Sensors and Actuators B (2005) 105, 28-38. |
Woolley et al., “Functional integration of PCR amplification and capillary electrophoresis in a microfabricated DNA analysis device” Anal. Chem. (1996) vol. 68, pp. 4081-4086. |
Woolley A.T., “Integrating Sample Processing and Detection with Microchip Capillary Electrophoresis of DNA”, in Integrated Biochips for DNA Analysis by Liu et al. [Eds]; (2007) Chapter 5, pp. 68-77. |
Wu et al., “Fabrication of Complex Three-dimensional Microchannel Systems in PDMS” J. Am. Chem. Soc. (2003) 125, 554-559. |
Wu et al., “Polycationic dendrimers interact with RNA molecules: polyamine dendrimers inhibit the catalytic activity of Candida ribozymes”, Chem Commun. (2005) 3: 313-315. |
Xiang et al., “Real Time PCR on Disposable PDMS Chip with a Miniaturized Thermal Cycler,” Biomedical Microdevices, (2005), 7(4):273-279. |
Xuan, “Joule heating in electrokinetic flow,” Electrophoresis, (2008), 298:33-43. |
Yang et al., “High sensitivity PCR assay in plastic micro reactors,” Lab Chip, (2002), 2:179-187. |
Yang et al., “An independent, temperature controllable-microelectrode array,” Anal. Chem., (2004), 76(5):1537-1543. |
Yang et al., “Cost-effective thermal isolation techniques for use on microfabricated DNA amplification and analysis devices,” J Micromech Microeng, (2005), 15:221-230. |
Yobas et al., Microfluidic Chips for Viral RNA Extraction & Detection, (2005), 2005 IEEE, 49-52. |
Yobas et al., “Nucleic Acid Extraction, Amplification, and Detection on Si-Based Microfluidic Platforms,” IEEE Journal of Solid-State Circuits, (2007), 42(8):1803-1813. |
Yoon et al., “Precise temperature control and rapid thermal cycling in a micromachined DNA polymer chain reaction chip,” J. Micromech. Microeng., (2002), 12:813-823. |
Yoza et al., “Fully Automated DNA Extraction from Blood Using Magnetic Particles Modified with a Hyperbranched Polyamidoamine Dendrimer”, J Biosci Bioeng, 2003, 95(1): 21-26. |
Yoza et al., DNA extraction using bacterial magnetic particles modified with hyperbranched polyamidoamine dendrimer, J Biotechnol., Mar. 20, 2003, 101(3): 219-228. |
Zhang et al, “Temperature analysis of continuous-flow micro-PCR based on FEA,” Sensors and Actuators B, (2002), 82:75-81. |
Zhang et al., “PCR Microfluidic Devices for DNA Amplification,” Biotechnology Advances, 24:243-284 (2006). |
Zhang et al, “Continuous-flow PCR Microfluidics for Rapid DNA Amplification Using Thin Film Heater with Low Thermal Mass,” Analytical Letters, (2007), 40:1672-1685, in 15 pages. |
Zhang et al, “Direct Adsorption and Detection of Proteins, Including Ferritin, onto Microlens Array Patterned Bioarrays,” J Am Chem Soc., (2007), 129:9252-9253. |
Zhang et al, “Micropumps, microvalves, and micromixers within PCR microfluidic chips: Advances and trends,” Biotechnology Advances, (2007), 25:483-514. |
Zhang et al., “Miniaturized PCR chips for nucleic acid amplification and analysis: latest advances and future trends,” Nucl Acids Res., (2007) 35(13):4223-4237. |
Zhang et al., “Parallel DNA amplification by convective polymerase chain reaction with various annealing temperatures on a thermal gradient device,” Analytical Biochemistry, (2009) 387:102-112. |
Zhao et al, “Heat properties of an integrated micro PCR vessel,” Proceedings of SPIE, (2001), International Conference on Sensor Technology, 4414:31-34. |
Zhou et al., “Cooperative binding and self-assembling behavior of cationic low molecular-weight dendrons with RNA molecules”, Org Biomol Chem. (2006) 4(3): 581-585. |
Zhou et al., “PAMAM dendrimers for efficient siRNA delivery and potent gene silencing”, Chem Comm.(Camb.) (2006) 22: 2362-2364. |
Zou et al., “A Micromachined Integratable Thermal Reactor,” technical digest from International Electron Devices Meeting, IEEE, Washington, D.C., Dec. 2-5, 2001 (6 pages). |
Zou et al., “Micro-assembled multi-chamber thermal cycler for low-cost reaction chip thermal multiplexing,” Sensors and Actuators A, (2002), 102:114-121. |
Zou et al., “Miniaturized Independently Controllable Multichamber Thermal Cycler,” IEEE Sensors Journal, (2003), 3(6):774-780. |
Petition for Inter Partes Review of U.S. Pat. No. 7,998,708 (Paper 1 in IPR2019-00488) dated Dec. 20, 2018 (94 pages). |
Declaration of Bruce K. Gale, Ph.D. (Exhibit 1001 in IPR2019-00488 and IPR2019-00490) dated Dec. 20, 2018 (235 pages). |
Patent Owner Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 7,998,708 and Exhibit List (Papers 5 and 6 in IPR2019-00488) dated Apr. 18, 2019 (79 pages). |
Decision instituting Inter Partes Review of U.S. Pat. No. 7,998,708 (Paper 8 in IPR2019-00488) dated Jul. 16, 2019 (20 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,323,900 (Paper 1 in IPR2019-00490) dated Dec. 20, 2018 (85 pages). |
Declaration of Michael G. Mauk, Ph.D. in Support of Patent Owner Preliminary Responses in IPR2019-00488 and IPR2019-00490 dated Apr. 18, 2019 (43 pages). |
Patent Owner Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,323,900 and Exhibit List (Papers 5 and 6 in IPR2019-00490) dated Apr. 18, 2019 (73 pages). |
Decision instituting Inter Partes Review of U.S. Pat. No. 8,323,900 (Paper 8 in IPR2019-00490) dated Jul. 16, 2019 (23 pages). |
Patent Owner's Response in Inter Partes Review of U.S. Pat. No. 8,323,900 and Exhibit List (Paper 25 in IPR2019-00490) dated Oct. 16, 2019 (80 pages). |
Patent Owner's Response in Inter Partes Review of U.S. Pat. No. 7,998,708 and Exhibit List (Paper 25 in IPR 2019-00488) dated Oct. 16, 2019 (93 pages). |
Transcript of Deposition of Bruce K. Gale, Ph.D., in Support of Patent Owner's Responses (Exhibit 2012 in IPR2019-00488 and IPR2019-00490), taken Sep. 24, 2019 (124 pages). |
Declaration of M. Allen Northrup, Ph.D. in Support of Patent Owner's Responses (Exhibit 2036 in IPR2019-00488 and IPR2019-00490) dated Oct. 16, 2019 (365 pages). |
Petitioner's Reply to Patent Owner's Response to Petition in Inter Partes Review of U.S. Pat. No. 7,998,708 and Exhibit List (Paper 32 in IPR 2019-00488) dated Jan. 31, 2020 (34 pages). |
Petitioner's Reply to Patent Owner's Response to Petition in Inter Partes Review of U.S. Pat. No. 8,323,900 and Exhibit List (Paper 32 in IPR 2019-00490) dated Jan. 31, 2020 (35 pages). |
Second Declaration of Bruce K. Gale, Ph.D. (Exhibit 1026 in IPR2019-00488 and IPR2019-00490) dated Jan. 31, 2020 (91 pages). |
Transcript of Deposition of M. Allen Northrup, Ph.D., (Exhibit 1027 in IPR2019-00488 and IPR2019-00490), taken Dec. 19, 2019 (109 pages). |
Patent Owner's Sur-Reply in Inter Partes Review of U.S. Pat. No. 8,323,900 (Paper 42 in IPR2019-00490) dated Mar. 12, 2020 (39 pages). |
Patent Owner's Sur-Reply in Inter Partes Review of U.S. Pat. No. 7,998,708 (Paper 43 in IPR 2019-00488) dated Mar. 12, 2020 (41 pages). |
Transcript of Second Deposition of Bruce K. Gale, Ph.D., (Exhibit 2068 in IPR2019-00488 and IPR2019-00490), taken Feb. 19, 2020 (352 pages). |
Record of Oral Hearing in IPR2019-00488 and IPR2019-00490 held Apr. 21, 2020 in 80 pages; Petitioner's Demonstratives for Oral Hearing in IPR2019-00488 and IPR2019-00490 held Apr. 21, 2020 in 72 pages; Patent Owner's Demonstratives for Oral Hearing in IPR2019-00488 and IPR2019-00490 held Apr. 21, 2020 in 88 pages; Patent Owner's Objections to Petitioner's Oral Hearing Demonstratives in IPR2019-00488 and IPR2019-00490 dated Apr. 16, 2020 (4 pages). |
Judgment/Final Written Decision Determining No Challenged Claims Unpatentable in Inter Partes Review of U.S. Pat. No. 7,998,708 (Paper No. 52 in IPR2019-00488) dated Jul. 14, 2020 (43 pages). |
Judgment/Final Written Decision Determining No Challenged Claims Unpatentable in Inter Partes Review of U.S. Pat. No. 8,323,900 (Paper No. 51 in IPR2019-00490) dated Jul. 14, 2020 (43 pages). |
Petitioner's Notice of Appeal in Inter Partes Review of U.S. Pat. No. 7,998,708 (Paper No. 54 in IPR2019-00488) dated Sep. 9, 2020 (48 pages). |
Petitioner's Notice of Appeal in Inter Partes Review of U.S. Pat. No. 8,323,900 (Paper No. 53 in IPR2019-00490) dated Sep. 9, 2020 (48 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 2 in IPR2020-01083) dated Jun. 12, 2020 (104 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 2 in IPR2020-01091) dated Jun. 12, 2020 (105 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,803,069 (Paper 2 in IPR2020-01095) dated Jun. 12, 2020 (84 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,803,069 (Paper 3 in IPR2020-01100) dated Jun. 12, 2020 (83 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,709,787 (Paper 2 in IPR2020-01132) dated Jun. 18, 2020 (96 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,415,103 (Paper 2 in IPR2020-01133) dated Jun. 18, 2020 (96 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,709,787 (Paper 2 in IPR2020-01137) dated Jun. 19, 2020 (86 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 8,415,103 (Paper 2 in IPR2020-01136) dated Jun. 19, 2020 (85 pages). |
Declaration of Mark A. Burns, Ph.D. (Exhibit N1001 in IPR2020-01083, IPR2020-01091, IPR2020-01095 and IPR2020-01100) dated Jun. 12, 2020 (378 pages). |
Declaration of Mark A. Burns, Ph.D. (Exhibit N1101 in IPR2020-01132 and IPR2020-01133) dated Jun. 17, 2020 (253 pages). |
Declaration of Mark A. Burns, Ph.D. (Exhibit N1201 in IPR2020-01136 and IPR2020-01137) dated Jun. 19, 2020 (205 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,703,069 (Paper 13 in IPR2020-01095) dated Sep. 17, 2020 (77 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 13 in IPR2020-01091) dated Sep. 17, 2020 (70 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,703,069 (Paper 14 in IPR2020-01100) dated Sep. 17, 2020 (59 pages). |
Declaration of M. Allen Northrup, Ph.D. in Support of Patent Owner Preliminary Responses in IPR2020-01091, IPR2020-01095 and IPR2020-01100 (Exhibit H2003) dated Sep. 16, 2020 (137 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 13 in IPR2020-01083) dated Oct. 22, 2020 (88 pages). |
Declaration of M. Allen Northrup, Ph.D. in support of Patent Owner Preliminary Responses in IPR2020-01083, IPR2020-01091, IPR2020-01095 and IPR2020-01100 (Exhibit H2003) dated Oct. 21, 2020 (171 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 10,625,262 (Paper 2 in IPR2021-00250) dated Nov. 25, 2020 (107 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 10,625,261 (Paper 2 in IPR2021-00251) dated Nov. 25, 2020 (117 pages). |
Petition for Inter Partes Review of U.S. Pat. No. 10,632,466 (Paper 2 in IPR2021-00253) dated Nov. 25, 2020 (121 pages). |
Declaration of Mark A. Burns, Ph.D. (Exhibit N1001 in IPR2021-00250, IPR2021-00251 and IPR2021-00253) dated Nov. 24, 2020 (311 pages). |
Declaration of James L. Mullins, Ph.D. (Exhibit N1029 in IPR2021-00250, IPR2021-00251, and IPR2021-00253) dated Nov. 18, 2020 (54 pages). |
Decision Denying Institution of Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 14 in IPR2020-01091) dated Dec. 4, 2020 (21 pages). |
Decision Denying Institution of Inter Partes Review of U.S. Pat. No. 8,703,069 (Paper 14 in IPR2020-01095) dated Dec. 4, 2020 (22 pages). |
Decision Denying Institution of Inter Partes Review of U.S. Pat. No. 8,703,069 (Paper 15 in IPR2020-01100) dated Dec. 4, 2020 (19 pages). |
Complaint filed by Becton, Dickinson et al., v. NeuModx Molecular, Inc. on Jun. 18, 2019 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS, Infringement Action involving U.S. Pat. Nos. 7,998,708; 8,273,308; 8,323,900; 8,415,103; 8,703,069; and 8,709,787 (29 pages). |
Answer to Complaint filed by NeuModx Molecular, Inc. on Aug. 9, 2019 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (24 pages). |
Amended Answer to Complaint filed by NeuModx Molecular, Inc. on Oct. 4, 2019 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (31 pages). |
First Amended and Supplemental Complaint filed by Becton, Dickinson and Company et al. on Jun. 25, 2020 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS, Infringement Action involving U.S. Pat. Nos. 7,998,708; 8,273,308; 8,323,900; 8,415,103; 8,703,069; 8,709,787; 10,494,663; 10,364,456; 10,443,088; 10,604,788; 10,625,261; 10,625,262; and 10,632,466 (55 pages). |
Answer to First Amended and Supplemental Complaint filed by NeuModx Molecular, Inc. on Jul. 16, 2020 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (42 pages). |
Defendant NeuModx's Initial Invalidity Contentions filed Sep. 30, 2020 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (47 pages). |
Defendant NeuModx's Joint Claim Construction Chart [Exhibit N1023] filed Oct. 21, 2020 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (25 pages). |
Defendant NeuModx's Initial Amended Answer, Affirmative Defenses, and Counterclaims to Plaintiffs' First Amended and Supplemental Complaint filed Nov. 23, 2020 in U.S. District Court, Delaware, Case #1:19-cv-01126-LPS (97 pages). |
Rush et al., “Dispersion by Pressure-Driven Flow in Serpentine Microfluidic Channels”, Ind Eng Chem Res., (2002) 41: 4652-4662. |
Decision Denying Institution of Inter Partes Review of U.S. Pat. No. 8,273,308 (Paper 14 in IPR2020-01083) dated Jan. 7, 2021 (24 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,415,103 (Paper 20 in IPR2020-01133) dated Jan. 20, 2021 (67 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,709,787 (Paper 19 in IPR2020-01132) dated Jan. 20, 2021 (78 pages). |
Declaration of M. Allen Northrup, Ph.D. in support of Patent Owner Preliminary Responses in IPR2020-01132 and IPR2020-01133 [Exhibit H2016] dated Jan. 20, 2021 (154 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,415,103 (Paper 19 in IPR2020-01136) dated Jan. 20, 2021 (77 pages). |
Patent Owner's Preliminary Response to Petition for Inter Partes Review of U.S. Pat. No. 8,709,787 (Paper 19 in IPR2020-01137) dated Jan. 20, 2021 (69 pages). |
Declaration of M. Allen Northrup, Ph.D. in support of Patent Owner Preliminary Responses in IPR2020-01136 and IPR2020-01137 [Exhibit H2016] dated Jan. 20, 2021 (111 pages). |
Appellants Qiagen North American Holdings, Inc. and NeuMoDx's Opening Brief [Corrected] in Appeals to IPR2019-00488, IPR2019-00490, IPR2019-01493, and IPR2019-01494 filed Jan. 22, 2021 in U.S. Court of Appeals for the Federal Circuit Case Nos. 20-2249, 20-2250, 20-2273 and 20-2276 (82 pages). |
Number | Date | Country | |
---|---|---|---|
20210139951 A1 | May 2021 | US |
Number | Date | Country | |
---|---|---|---|
60786007 | Mar 2006 | US | |
60859284 | Nov 2006 | US | |
60959437 | Jul 2007 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16914109 | Jun 2020 | US |
Child | 17146403 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16787977 | Feb 2020 | US |
Child | 16914109 | US | |
Parent | 14796239 | Jul 2015 | US |
Child | 16787977 | US | |
Parent | 13692929 | Dec 2012 | US |
Child | 14796239 | US | |
Parent | 13035725 | Feb 2011 | US |
Child | 13692929 | US | |
Parent | 11985577 | Nov 2007 | US |
Child | 13035725 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11728964 | Mar 2007 | US |
Child | 11985577 | US |