Modular CPR assist device

Abstract
A system for performing chest compression for Cardiopulmonary Resuscitation. The system includes a motor, drive spool and associated couplings which allow for controlling and limiting the movement of the compressing mechanism and includes a control system for controlling the operation and interaction of the various components to provide for optimal automatic operation of the system.
Description




FIELD OF THE INVENTIONS




This invention relates to the resuscitation of cardiac arrest patients.




BACKGROUND OF THE INVENTIONS




Cardiopulmonary resuscitation (CPR) is a well known and valuable method of first aid. CPR is used to resuscitate people who have suffered from cardiac arrest after heart attack, electric shock, chest injury and many other causes. During cardiac arrest, the heart stops pumping blood, and a person suffering cardiac arrest will soon suffer brain damage from lack of blood supply to the brain. Thus, CPR requires repetitive chest compression to squeeze the heart and the thoracic cavity to pump blood through the body. Very often, the patient is not breathing, and mouth to mouth artificial respiration or a bag valve mask is used to supply air to the lungs while the chest compression pumps blood through the body.




It has been widely noted that CPR and chest compression can save cardiac arrest patients, especially when applied immediately after cardiac arrest. Chest compression requires that the person providing chest compression repetitively push down on the sternum of the patient at 80-100 compressions per minute. CPR and closed chest compression can be used anywhere, wherever the cardiac arrest patient is stricken. In the field, away from the hospital, it may be accomplished by ill-trained by-standers or highly trained paramedics and ambulance personnel.




When a first aid provider performs chest compression well, blood flow in the body is typically about 25-30% of normal blood flow. This is enough blood flow to prevent brain damage. However, when chest compression is required for long periods of time, it is difficult if not impossible to maintain adequate compression of the heart and rib cage. Even experienced paramedics cannot maintain adequate chest compression for more than a few minutes. Hightower, et al., Decay In Quality Of Chest Compressions Over Time, 26 Ann. Emerg. Med. 300 (Sep. 1995). Thus, long periods of CPR, when required, are not often successful at sustaining or reviving the patient. At the same time, it appears that, if chest compression could be adequately maintained, cardiac arrest patients could be sustained for extended periods of time. Occasional reports of extended CPR efforts (45-90 minutes) have been reported, with the patients eventually being saved by coronary bypass surgery. See Tovar, et al., Successful Myocardial Revascularization and Neurologic Recovery, 22 Texas Heart J. 271 (1995).




In efforts to provide better blood flow and increase the effectiveness of bystander resuscitation efforts, modifications of the basic CPR procedure have been proposed and used. Of primary concern in relation to the devices and methods set forth below are the various mechanical devices proposed for use in main operative activity of CPR, namely repetitive compression of the thoracic cavity.




The device shown in Barkolow, Cardiopulmonary Resuscitator Massager Pad, U.S. Pat. No. 4,570,615 (Feb. 18, 1986), the commercially available Thumper device, and other such devices, provide continuous automatic closed chest compression. Barkolow and others provide a piston which is placed over the chest cavity and supported by an arrangement of beams. The piston is placed over the sternum of a patient and set to repeatedly push downward on the chest under pneumatic power. The patient must first be installed into the device, and the height and stroke length of the piston must be adjusted for the patient before use, leading to delay in chest compression. Other analogous devices provide for hand operated piston action on the sternum. Everette, External Cardiac Compression Device, U.S. Pat. No. 5,257,619 (Nov. 2, 1993), for example, provides a simple chest pad mounted on a pivoting arm supported over a patient, which can be used to compress the chest by pushing down on the pivoting arm. These devices are not clinically more successful than manual chest compression. See Taylor, et al., External Cardiac Compression, A Randomized Comparison of Mechanical and Manual Techniques, 240 JAMA 644 (Aug. 1978).




Other devices for mechanical compression of the chest provide a compressing piston which is secured in place over the sternum via vests or straps around the chest. Woudenberg, Cardiopulmonary Resuscitator, U.S. Pat. No. 4,664,098 (May 12, 1987) shows such a device which is powered with an air cylinder. Waide, et al., External Cardiac Massage Device, U.S. Pat. No. 5,399,148 (Mar. 21, 1995) shows another such device which is manually operated. In another variation of such devices, a vest or belt designed for placement around the chest is provided with pneumatic bladders which are filled to exert compressive forces on the chest. Scarberry, Apparatus for Application of Pressure to a Human Body, U.S. Pat. No. 5,222,478 (Jun. 29, 1993) and Halperin, Cardiopulmonary Resuscitation and Assisted Circulation System, U.S. Pat. No. 4,928,674 (May 29, 1990) show examples of such devices. Lach, et al., Resuscitation Method and Apparatus, U.S. Pat. No. 4,770,164 (Sep. 13, 1988) proposed compression of the chest with wide band and chocks on either side of the back, applying a side-to-side clasping action on the chest to compress the chest.




Several operating parameters must be met in a successful resuscitation device. Chest compression must be accomplished vigorously if it is to be effective. Very little of the effort exerted in chest compression actually compresses the heart and large arteries of the thorax and most of the effort goes into deforming the chest and rib cage. The force needed to provide effective chest compression creates risk of other injuries. It is well known that placement of the hands over the sternum is required to avoid puncture of the heart during CPR. Numerous other injuries have been caused by chest compression. See Jones and Fletter, Complications After Cardiopulmonary Resuscitation, 12 AM. J. Emerg. Med. 687 (Nov. 1994), which indicates that lacerations of the heart, coronary arteries, aortic aneurysm and rupture, fractured ribs, lung herniation, stomach and liver lacerations have been caused by CPR. Thus the risk of injury attendant to chest compression is high, and clearly may reduce the chances of survival of the patient vis-à-vis a resuscitation technique that could avoid those injuries. Chest compression will be completely ineffective for very large or obese cardiac arrest patients because the chest cannot be compressed enough to cause blood flow. Chest compression via pneumatic devices is hampered in its application to females due to the lack of provision for protecting the breasts from injury and applying compressive force to deformation of the thoracic cavity rather than the breasts.




CPR and chest compression should be initiated as quickly as possible after cardiac arrest to maximize its effectiveness and avoid neurologic damage due to lack of blood flow to the brain. Hypoxia sets in about two minutes after cardiac arrest, and brain damage is likely after about four minutes without blood flow to the brain, and the severity of neurologic defect increases rapidly with time. A delay of two or three minutes significantly lowers the chance of survival and increases the probability and severity of brain damage. However, CPR and ACLS are unlikely to be provided within this time frame. Response to cardiac arrest is generally considered to occur in four phases, including action by Bystander CPR, Basic Life Support, Advanced Cardiac Life Support, and the Emergency Room. By-stander CPR occurs, if at all, within the first few minutes after cardiac arrest. Basic Life Support is provided by First Responders who arrive on scene about 4-6 minutes after being dispatched to the scene. First responders include ambulance personnel, emergency medical technicians, firemen and police. They are generally capable of providing CPR but cannot provide drugs or intravascular access, defibrillation or intubation. Advanced Life Support is provided by paramedics or nurse practitioners who generally follow the first responders and arrive about 8-15 minutes after dispatch. ALS is provided by paramedics, nurse practitioners or emergency medical doctors who are generally capable of providing CPR, drug therapy including intravenous drug delivery, defibrillation and intubation. The ALS providers may work with a patient for twenty to thirty minutes on scene before transporting the patient to a nearby hospital. Though defibrillation and drug therapy is often successful in reviving and sustaining the patient, CPR is often ineffective even when performed by well trained first responders and ACLS personnel because chest compression becomes ineffective when the providers become fatigued. Thus, the initiation of CPR before arrival of first responders is critical to successful life support. Moreover, the assistance of a mechanical chest compression device during the Basic Life Support and Advanced Life Support stages is needed to maintain the effectiveness of CPR.




SUMMARY




The devices described below provide for circumferential chest compression with a device which is compact, portable or transportable, self-powered with a small power source, and easy to use by by-standers with little or no training. Additional features may also be provided in the device to take advantage of the power source and the structural support board contemplated for a commercial embodiment of the device.




The device includes a broad belt which wraps around the chest and is buckled in the front of the cardiac arrest patient. The belt is repeatedly tightened around the chest to cause the chest compression necessary for CPR. The buckle may include an interlock which must be activated by proper attachment before the device will activate, thus preventing futile belt cycles. The operating mechanism for repeatedly tightening the belt is provided in a small box locatable at the patient's side, and comprises a rolling mechanism which takes up the intermediate length of the belt to cause constriction around the chest. The roller is powered by a small electric motor, and the motor powered by batteries and/or standard electrical power supplies such as 120V household electrical sockets or 12V DC automobile power sockets (car cigarette lighter sockets). The belt is contained in a cartridge which is easily attached and detached from the motor box. The cartridge itself may be folded for compactness. The motor is connected to the belt through a transmission that includes a cam brake and a clutch, and is provided with a controller which operates the motor, clutch and cam brake in several modes. One such mode provides for limiting belt travel according to a high compression threshold, and limiting belt travel to a low compression threshold. Another such mode includes holding the belt taught against relaxation after tightening the belt, and thereafter releasing the belt. Respiration pauses, during which no compression takes place to permit CPR respiration, can be included in the several modes. Thus, numerous inventions are incorporated into the portable resuscitation device described below.











BRIEF DESCRIPTION OF THE DRAWINGS





FIG. 1

is an overview of the resuscitation device.





FIG. 2

illustrates the installation of the belt cartridge.





FIG. 3

illustrates the operation of the belt cartridge.





FIG. 4

illustrates the operation of the belt cartridge.





FIG. 5

illustrates an alternative configuration of the belt cartridge.





FIG. 6

illustrates an alternative configuration of the belt cartridge.





FIG. 7

illustrates an alternative configuration of the belt cartridge.





FIG. 8

illustrates an alternative configuration of the belt cartridge.





FIG. 9

illustrates an alternative configuration of the belt cartridge.





FIG. 10

illustrates an alternative embodiment of the belt.





FIG. 11

illustrates an alternative embodiment of the belt.





FIG. 12

illustrates the configuration of the motor and clutch within the motor box.





FIG. 12



a


illustrates the an alternative embodiment of the device illustrated in FIG.


12


.





FIG. 12



b


illustrates a shield used in conjunction with the device of

FIG. 12



a.







FIG. 13

is a table of the motor and clutch timing in a basic embodiment with a continuously running motor.





FIG. 13



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


13


.





FIG. 14

is a table of the motor and clutch timing in a basic embodiment.





FIG. 14



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


14


.





FIG. 15

is a table of the motor and clutch timing for squeeze and hold operation of the compression belt.





FIG. 15



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


15


.





FIG. 16

illustrates the timing of the motor, clutch and cam brake in a system that does not allow the belt compression to completely relax during each cycle.





FIG. 16



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


16


.





FIG. 17

shows a timing table for use in combination with a system that uses the motor, clutch, and secondary brake or a brake on the drive wheel or the spindle itself.





FIG. 17



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


17


.





FIG. 18

is a table of the motor and clutch timing for squeeze and hold operation of the compression belt where a brake is not energized to hold the belt during compression cycles unless the upper threshold is achieved by the system.





FIG. 18



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


18


.





FIG. 19

is a table of the motor and clutch timing for squeeze and hold operation of the compression belt with a continuously running motor.





FIG. 19



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


19


.





FIG. 20

is a table of the motor and clutch timing for squeeze and hold operation of the compression belt using two brakes.





FIG. 20



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


20


.





FIG. 21

is table of the motor and clutch timing for operation of the compression belt in an embodiment in which the system timing is reset each time an upper threshold is achieved.





FIG. 21



a


is a diagram of the pressure changes developed by the system operated according to the timing diagram of FIG.


21


.











DETAILED DESCRIPTION OF THE INVENTIONS





FIG. 1

shows an overview of the resuscitation device


1


. The major components are provided in modular form, and include the motor box


2


, the belt cartridge


3


and the belt


4


. The motor box exterior includes a sprocket


5


in drive wheel


6


which releasable mates with receiving rod


7


on the cartridge. The cartridge houses the belt which will wrap around the chest of the patient. The cartridge also includes the spool


8


which is turned by the receiving rod. The spool takes up the midpoint of the belt to drive the compression cycles. A computer control system


10


may be included as shown in an enclosure mounted on the motor box. By providing the system in modular form, with the motor box releasable attached to the belt cartridge, the belt cartridge may more easily be maneuvered while slipping it under the patient.





FIG. 2

shows a more detailed view of the cartridge, including the internal mechanisms of the belt cartridge


3


. The outer body of the cartridge provides for protection of the belt during storage, and includes a back plate


11


with a left panel


11


L and a right panel


11


R (relative to the patient during use). The right plate can be folded over the left plate for storage


5


and transport. Both panels are covered with a sheet


12


of low friction material such ad PTFE (Teflon®) to reduce friction when the belt slides over the panel during operation. Under the left panel, the cartridge has a housing


13


which houses the middle portion of the belt, the spool


8


and the spindle


15


. The lateral side


14


of the cartridge (corresponding to the anatomic position when in use on a patient) houses the drive spool


8


, with its drive rod


7


which engages the drive wheel


6


of the motor box. The cartridge also houses the guide spindle


15


(visible in

FIG. 3

) for directing the belt toward the drive spool


8


. The guide spindle is located near the center of the cartridge (corresponding to the medial line of the patient when in use), so that it is located near the spine when the device is in use. This spindle reverses the belt travel for the left side of the belt, so that when it is pulled to the left by the drive spool, the portion that wraps around the left flank of the body moves to the right. The cartridge body is also hinged near the mid-line, and in this view the cartridge is hinged near the axis of the spindle. A friction liner


16


is suspended over the belt in the area of the guide spindle, and is attached to the housing at the top and bottom panels


13




t


and


13




b


and spans the area in which the left belt portions and right belt portions diverge from the cartridge. The belt


4


is shown in the open condition. Male quick release fittings


17


R on the right belt portion fit into corresponding female quick release


17


L fitting on the left belt portion to releasably secure the belt around the patient's chest. The belt length on the left and right sides of the belt may be adjusted so that the buckles fall just over the center of the patient's chest during operation, or they may be adjusted for placement of the buckles elsewhere around the chest. The handle


18


is provided for convenient handling and carrying of the device.





FIG. 3

shows a cross section of the belt cartridge. The housing


13


is relatively flat, (but may be wedge shaped to assist in sliding it under a patient) when viewed from the superior position. The left panel


11


L sits atop the housing


13


and the right panel extends from the housing. In the unfolded position, the cartridge is flat enough to be slipped under a patient from the side. In the cross section view, the guide spindle


15


can be seen, and the manner in which the belt is threaded through the slot


9


of the drive spool


8


appears more clearly. The belt


4


comprises a single long band of tough fabric threaded through the drive spool slot


9


and extending from the drive spool to the right side quick releases


17


R and also from the drive spool, over and around the guide spindle, and back toward the drive spool to the left side quick releases


17




l


. The belt is threaded through the drive spool


8


at its mid-portion, and around the guide spindle, where the left belt portion


4


L folds around the guide spindle, under the friction liner and back to the left side of the cartridge, and the right belt portion


4


R passes the spindle to reach around the patient's right side. The friction belt liner


16


is suspended above the guide spindle and belt, being mounted on the housing, and fits between the patient and the compression belt. The cartridge is placed under the patient


20


, so that the guide spindle is located close to the spine


21


and substantially parallel to the spine, and the quick release fittings may be fastened over the chest in the general area of the sternum


22


.




In use, the cartridge is slipped under the patient


20


and the left and right quick releases are connected. As shown in

FIG. 4

, when the drive spool is rotated, it takes up the middle portion of the belt and tightens the belt around the chest. The compression force exerted by the belt is more than sufficient to induce or increase intrathoracic pressure necessary for CPR. When the belt is spooled around the drive spool


8


, the chest of the patient is compressed significantly, as illustrated.




While it will usually be preferred to slide the cartridge under the patient, this is not necessary. The device may be fitted onto the patient with the buckles at the back or side, or with the motor to the side or above the patient, whenever space restrictions require it. As show in

FIG. 5

, the cartridge may be fitted onto a patient


20


with only the right belt portion


4


R and right panel


11


R slipped under the patient, and with the right panel and left panel partially unfolded. The placement of the hinge between the right side and left side panels permits flexibility in installation of the device.

FIG. 6

shows that the cartridge may also be fitted onto a patient


20


with both the right panel


11


R and the left panel


11


L slipped under the patient, but with the motor box


2


folded upward, rotated about the axis of the drive spool


8


. These configurations are permitted by the modular nature of the motor box connection to the belt cartridge, and will prove useful in close spaces such as ambulances and helicopters. (Note that, though the belt may be tightened by spooling operation in either direction, tightening in the direction of arrow


23


, clockwise when viewed from the top of the patient and the device, will cause reactive force which urges the motor box to rotate into the device, toward the body, rather than outwardly away from the body. Locking pins may be provided to prevent any rotational movement between the motor box and the cartridge. In the construction of the motor box as shown, the limited height of the box (the height of the box is less than the distance between the left flank of the patient and the drive spool) prevents contact with the patient in case the locking pins are not engaged for any reason. The rotation of the drive belt may be reversed to a counter clockwise direction, in which reactive force will urge the motor box to rotate outwardly. In this case, locking mechanisms such as locking pins can be used to protect operators from movement of the system.)




Regardless of the orientation of the panels, the reversing spindle will properly orient the travel of the belt to ensure compression. The placement of the spindle at the point where the right belt portion and the left belt portion diverge under the patient's chest, and the placement of this spindle in close proximity to the body, permits the belt to make contact with the chest at substantially all points on the circumference of the chest. The position of the spindle reverses the travel of the belt left portion


4


L from a transverse right to left direction to a transverse left to right direction, while the fact that belt right portion


4


R bypasses the spindle means that it always moves from right to left in relation to the patient when pulled by the drive spool. Thus the portions of the belt engaging the chest always pull from opposite lateral areas of the chest to a common point near a central point. In

FIGS. 3 and 4

, the opposite lateral areas correspond to the anatomic lateral area of the patient, and the central point corresponds to the spine. In

FIG. 5

, the lateral areas correspond to the spine and anterior left side of the torso, while the central point corresponds to the left lateral area of the chest. Additionally, the use of the single spindle at the center of the body, with the drive spool placed at the side of the body, permits simple construction and the detachable or modular embodiment of the motor assembly, and allows placement of the belt about the patient before attachment of the motor box to the entire device.





FIG. 7

illustrates an embodiment of the compression belt which reduces the take up speed for a given motor speed or gearing and allows for twice the compressive force for a given motor speed. The compression belt comprises a loop


24


of belt material. The loop is threaded through the complex path around spindles


25


in the quick release fasteners


26


, around the body to the guide spindle


15


, around or past the guide spindle and into the drive spool


8


. The left belt portion outer layer


27


L and right belt portion outer layer


27


R form, together with the left belt portion inner layer


28


L and right belt portion inner layer


28


R form a continues loop running inwardly from the fastener spindle, inwardly around the chest to the opposite drive spindle, outwardly from the opposite drive spindle, downwardly over the chest, past the guide spindle to the drive spool, through the drive spool slot and back under the guide spindle, reversing around the guide spindle and upwardly over the chest back to the fastener spindle. Thus both the inner and outer layers of this two layer belt are pulled toward the drive spool to exert compressive force on the body. This can provide for a decrease in friction as the belts will act on each other rather than directly on the patient. It will also allow for a lower torque, higher speed motor to exert the necessary force.




In

FIG. 8

, the double layer belt system is modified with structure which locks the inner belt portion in place, and prevents it from moving along the body surface. This has the advantage that the major portion of the belt in contact with the body does not slide relative to the body. To lock the belt inner layer in place relative to the loop pathway, the locking bar


29


is fixed within the housing


13


in parallel with the guide spindle


15


and the drive spool


8


. The inner loop may be secured and fastened to the locking bar, or it may be slidably looped over the locking bar (and the locking bar may be rotatable, as a spindle). The left belt portion outer layer


27


L and right belt portion outer layer


27


R are threaded through the drive spool


8


. With the locking bar installed, the rotation of the drive spool takes up the outer layer of the belt, and these outer layers are forced to slide over the left belt portion inner layer


28


L and right belt portion inner layer


28


R, but the inner layers do not slide relative to the surface of the patient (except, possibly, during a brief few cycles in which the belt centers itself around the patient, which will occur spontaneously due to the forces applied to the belt.)




In

FIG. 9

, the double layer belt system is modified with structure which does not lock the inner belt portion in place or prevent it from moving along the body surface, but instead provides a second drive spool to act on the inner layer of the belt. To drive the belt inner layer relative to the loop pathway, the secondary drive spool


30


is fixed within the housing


13


in parallel with the guide spindle


15


and the drive spool


8


. This secondary drive spool is driven by the motor, either through transmission geared within the housing or through a second receiving rod protruding from the housing and a secondary drive socket driven through appropriate gearing in the motor box. The inner loop may be secured and fastened to the secondary drive spool, or it may be threaded through the secondary drive spool slot


31


. The left belt portion outer layer


27


L and right belt portion outer layer


27


R are threaded through the first drive spool


8


. With the secondary drive spool, the rotation of the first drive spool


8


takes up the outer layer of the belt, and these outer layers are forced to slide over the left belt portion inner layer


28


L and right belt portion inner layer


28


R, while the secondary drive spool takes up the inner layers.




The compression belt may be provided in several forms. It is preferably made of some tough material such as parachute cloth or tyvek. In the most basic form shown in

FIG. 10

, the belt


4


is a plain band of material with fastening ends


32


L and


32


R, corresponding left and right belt portions


4


L and


41


R, and the spool engaging center portion


33


. While we have used the spool slot in combination with the belt being threaded through the spool slot as a convenient mechanism to engage the belt in the drive spool, the belt may be fixed to the drive spool in any manner. In

FIG. 11

, the compression belt is provided in two distinct pieces comprising left and right belt portions


4


L and


41


R connected with a cable


34


which is threaded through the drive spool. This construction permits a much shorter drive spool, and may eliminate friction within the housing inherent in the full width compression band of FIG.


10


. The fastening ends


32


L and


32


R are fitted with hook and loop fastening elements


35


which may be used as an alternative to other quick release mechanisms. To provide a measurement of belt pay-out and take-up during operation, the belt or cable may be modified with the addition of a linear encoder scale, such as scale


36


on the belt near the spool engaging center portion


33


. A corresponding scanner or reader may be installed on the motor box, or in the cartridge in apposition to the encoder scale.





FIG. 12

illustrates the configuration of the motor and clutch within the motor box. The exterior of the motor box includes a housing


41


, and a computer module


10


with a convenient display screen


42


for display of any parameters measured by the system. The motor


43


is a typical small battery operated motor which can exert the required belt tensioning torque. The motor shaft


44


is lined up directly to the brake


45


which includes reducing gears and a cam brake to control free spinning of the motor when the motor is not energized (or when a reverse load is applied to the gearbox output shaft). The gearbox output rotor


46


connects to a wheel


47


and chain


48


which connect to the input wheel


49


, and thereby to the transmission rotor


50


of the clutch


51


. The clutch


51


controls whether the input wheel


49


engages the output wheel


52


, and whether rotary input to the input wheel is transmitted to the output wheel. (The secondary brake


53


, which we refer to as the spindle brake, provides for control of the system in some embodiments, as explained below in reference to

FIG. 17.

) The output wheel


52


is connected to the drive spool


8


via the chain


54


and drive wheel


6


and receiving rod


7


(the drive rod is on the cartridge). The drive wheel


6


has receiving socket


5


which is sized and shaped to mate and engage with the drive rod


7


(simple hexagonal or octagonal sprocket which matches the drive rod is sufficient). While we use a wrap spring brake (a MAC 45 sold by Warner Electric) for the cam brake in the system, any form of brake may be employed. The wrap spring brake has the advantage of allowing free rotating of the shaft when de-energized, and holds only when energized. The wrap spring brake may be operated independent of the motor. While we use chains to transmit power through the system, belts, gears or other mechanisms may be employed.





FIG. 12



a


illustrates the configuration of the motor and clutch within the motor box. The exterior of the motor box includes a housing


41


which holds the motor


43


is a typical small battery operated motor which can exert the required belt tensioning torque. The motor shaft


44


is lined up directly to the brake


45


which includes reducing gears and a cam. The gearbox output rotor


46


connects to brake to the output wheel


47


and chain


48


which in turn connects directly to the drive wheel


6


and receiving rod


7


. The drive spool


8


is contained within the housing


41


. At the end of the drive spool opposite the drive wheel, the brake


55


is directly connected to the drive spool. The belt


4


is threaded through the drive spool slot


9


. To protect the belt from rubbing on the motor box, the shield


57


with the long aperture


58


is fastened to the housing so that the aperture lies over the drive spool, allowing the belt to pass through the aperture and into the drive spool slot, and return out of the housing. Under the housing, slidably disposed within a channel in the bottom of the housing, a push plate


70


is positioned so that it can slide back and forth relative to the housing. The belt right portion


4


is fitted with a pocket


71


which catches or mates with the right tip


72


of the push plate. The right tip of the push plate is sized and dimensioned to fit within the pocket. By means of this mating mechanism, the belt can be slipped onto the push plate, and with the handle


73


on the left end of the push plate, the push plate together with the right belt portion can be pushed under a patient. The belt includes the encoder scale


36


, which can be read with an encoder scanner mounted on or within the housing. In use, the belt right portion is slipped under the patient by fastening it to the push plate and sliding the push plate under the patient. The motor box can then be positioned as desired around the patient (the belt will slip through the drive spool slot to allow adjustment). The belt right side can then be connected to the belt left portion so that the fastened belt surrounds the patient's chest. In both

FIGS. 12 and 12



a


, the motor is mounted in side-by-side relationship with the clutch and with the drive spool. With the side-by-side arrangement of the motor and the roller, the motor may be located to the side of the patient, and need not be placed under the patient, or in interfering position with the shoulders or hips. This also allows a more compact storage arrangement of the device, vis-à-vis an in-line connection between the motor and the roller. A battery is placed within the box or attached to the box as space allows.




During operation, the action of the drive spool and belt draw the device toward the chest, until the shield is in contact with the chest (with the moving belt interposed between the shield and the chest). The shield also serves to protect the patient from any rough movement of the motor box, and help keep a minimum distance between the rotating drive spool and the patients skin, to avoid pinching the patient or the patient's clothing in the belt as the two sides of the belt are drawn into the housing. As illustrated in

FIG. 12



b


, the shield


57


may also include two lengthwise apertures


74


separated by a short distance. With this embodiment of the shield, one side of the belt passes through one aperture and into the drive spool slot, and the other side of the belt exits from the drive spool slot and outwardly through the other aperture in the shield. The shield as shown has an arcuate transverse cross section (relative to the body on which it is installed). This arcuate shape permits the motor box to lay on the floor during use while a sufficient width of shield extends between the box and the belt. The shield made of plastic, polyethylene, PTFE, or other tough material which allows the belt to slide easily. The motor box, may, however, be placed anywhere around the chest of the patient.




A computer module which acts as the system controller is placed within the box or attached to the box and is operably connected to the motor, the cam brake, clutch, encoder and other operating parts, as well as biological and physical parameter sensors included in the overall system (blood pressure, blood oxygen, end tidal CO2, body weight, chest circumference, etc. are parameters that can be measured by the system and incorporated into the control system for adjusting compression rates and torque thresholds, or belt pay-out and slack limits). The computer module can also be programmed to handle various ancillary tasks such as display and remote communications, sensor monitoring and feedback monitoring, as illustrated in our prior application Ser. No. 08/922,723.




The computer is programmed (with software or firmware or otherwise) and operated to repeatedly turn the motor and release the clutch to roll the compression belt onto the drive spool (thereby compressing the chest of the patient) and release the drive spool to allow the belt to unroll (thereby allowing the belt and the chest of the patient to expand), and hold the drive spool in a locked or braked condition during periods of each cycle. The computer is programmed to monitor input from various sensors, such as the torque sensor or belt encoders, and adjust operation of the system in response to these sensed parameters by, for example, halting a compression stroke or slipping the clutch (or brake) in response to torque limit or belt travel limits. As indicated below, the operation of the motor box components may be coordinated to provide for a squeeze and hold compression method which prolongs periods of high intrathoracic pressure. The system may be operated in a squeeze and quick release method for more rapid compression cycles and better waveform and flow characteristics in certain situations. The operation of the motor box components may be coordinated to provide for a limited relaxation and compression, to avoid wasting time and battery power to move the belt past compression threshold limits or slack limits. The computer is preferably programmed to monitor two or more sensed parameters to determine an upper threshold for belt compression. By monitoring motor torque as measured by a torque sensor and paid out belt length as determined by a belt encoder, the system can limit the belt take-up with redundant limiting parameters. The redundancy provided by applying two limiting parameters to the system avoids over-compression in the case that a single compression parameter exceed the safe threshold while the system fails to sense and response the threshold by stopping belt take-up.




An angular optical encoder may be placed on any rotating part of the system to provide feedback to a motor controller relating to the condition of the compression belt. (The encoder system may be an optical scale coupled to an optical scanner, a magnetic or inductive scale coupled to a magnetic or inductive encoder, a rotating potentiometer, or any one of the several encoder systems available.) The encoder


56


, for example, is mounted on the secondary brake


53


(in FIG.


12


), and provides an indication of the motor shaft motion to a system controller. An encoder may also be placed on the drive socket


5


or drive wheel


6


, the motor


43


and or motor shaft


44


. The system includes a torque sensor (sensing current supply to the motor, for example), and monitors the torque or load on the motor. For either or both parameters, a threshold is established above which further compression is not desired or useful, and if this occurs during the compression of the chest, then the clutch is disengaged. The belt encoder is used by the control system to track the take-up of the belt, and to limit the length of belt which is spooled upon the drive belt.




In order to control the amount of thoracic compression (change in circumference) for the cardiac compression device using the encoder, the control system must establish a baseline or zero point for belt take-up. When the belt is tight to the point where any slack has been taken up, the motor will require more current to continue to turn under the load of compressing the chest. This expected rapid increase in motor current draw (motor threshold current draw) is measured through torque sensor (an Amp meter, a voltage divider circuit or the like). This spike in current or voltage is taken as the signal that the belt has been drawn tightly upon the patient and the paid out belt length is an appropriate starting point, and the encoder measurement at this point is zeroed within the system (that is, taken as the starting point for belt take-up). The encoder then provides information used by the system to determine the change in length of the belt from this pre-tightened position. The ability to monitor and control the change in length allows the controller to control the amount of pressure exerted on the patient and the change in volume of the patient by limiting the length of belt take-up during a compression cycle.




The expected length of belt take-up for optimum compression is 1 to 6 inches. However, six inches of travel on a thin individual may create a excessive change in thoracic circumference and present the risk of injury from the device. In order to overcome this problem, the system determines the necessary change in belt length required by measuring the amount of belt travel required to become taught as described above. Knowing the initial length of the belt and subtracting off the amount required to become taught will provide a measure of the patient's size (chest circumference). The system then relies on predetermined limits or thresholds to the allowable change in circumference for each patient on which it is installed, which can be used to limit the change in volume and pressure applied to the patient. The threshold may change with the initial circumference of the patient so that a smaller patient will receive less of a change in circumference as compared to a larger patient. The encoder provides constant feedback as to the state of travel and thus the circumference of the patient at any given time. When the belt take-up reaches the threshold (change in volume), the system controller ends the compression stroke and continues into the next period of hold or release as required by the compression/decompression regimen programmed into the controller. The encoder also enables the system to limit the release of the belt so that it does not fully release. This release point can be determined by the zero point established on the pre-tightening first take-up, or by taking a percentage of the initial circumference or a sliding scale triggered by the initial circumference of the patient.




The belt could also be buckled so that it remains tight against the patient. Requiring the operator to tighten the belt provides for a method to determine the initial circumference of the patient. Again encoders can determine the amount of belt travel and thus can be used to monitor and limit the amount of change in the circumference of the patient given the initial circumference.




Several compression and release patterns may be employed to boost the effectiveness of the CPR compression. Typical CPR compression is accomplished at 60-80 cycles per minute, with the cycles constituting mere compression followed by complete release of compressive force. This is the case for manual CPR as well as for known mechanical and pneumatic chest compression devices. With our new system, compression cycles in the range of 20-70 cpm have been effective, and the system may be operated as high as 120 cpm or more. This type of compression cycle can be accomplished with the motor box with motor and clutch operation as indicated in FIG.


13


. When the system is operating in accordance with the timing table of

FIG. 13

, the motor is always on, and the clutch cycles between engagement (on) and release (off). After several compressions at time periods T


1


, T


3


, T


5


and T


7


, the system pauses for several time periods to allow brief period (several seconds) to provide a respiration pause, during which operators may provide ventilation or artificial respiration to the patient, or otherwise cause oxygenated air to flow into the patient's lungs. (The brakes illustrated in

FIG. 12

, are not used in this embodiment, though they may be installed.) The length of the clutch engagement period is controlled in the range of 0-2000 msec, and the time between periods of clutch engagement is controlled in the range of 0-2000 msec (which of course is dictated by medical considerations and may change as more is learned about the optimal rate of compression).




The timing chart of

FIG. 13



a


illustrates the intrathoracic pressure changes caused by the compression belt when operated according to the timing diagram of FIG.


13


. The chest compression is indicated by the status line


59


. The motor is always on, as indicated by motor status line


60


. The clutch is engaged or “on” according to the square wave clutch status line


61


in the lower portion of the diagram. Each time the clutch engages, the belt is tightened around the patient's chest, resulting in a high pressure spike in belt tension and intra-thoracic pressure as indicated by the compression status line


62


. Pulses p


1


, p


2


, p


3


, p


4


and p


5


are all similar in amplitude and duration, with the exception of pulse p


3


. Pulse p


3


is limited in duration in this example to show how the torque limit feedback operates to prevent excessive belt compression. (Torque limit may be replaced by belt travel or other parameter as the limiting parameter.) As an example of system response to sensing the torque limit, Pulse p


3


is shown rapidly reaching the torque limit set on the motor. When the torque limit is reached, the clutch disengages to prevent injury to the patient and excessive drain on the battery (excessive compression is unlikely to lead to additional blood flow, but will certainly drain the batteries quickly). Note that after clutch disengagement under pulse p


3


, belt tension and intra-thoracic pressure drop quickly, and the intra-thoracic pressure is increased for only a small portion of cycle. After clutch disengagement based on an over-torque condition, the system returns to the pattern of repeated compressions. Pulse P


4


occurs at the next scheduled compression period T


7


, after which the respiration pause period spanning T


8


, T


9


, and T


10


is created by maintaining the clutch in the disengaged condition. After the respiration pause, pulse P


5


represents the start of the next set of compressions. The system repeatedly performs sets of compressions followed by respiration pauses until interrupted by the operator.





FIG. 14

illustrates the timing of the motor, clutch and cam brake in a system that allows the belt compression to be reversed by reversing the motor. It also provides for compression hold periods to enhance the hemodynamic effect of the compression periods, and relaxation holds to limit the belt pay-out in the relaxation period to the point where the belt is still taut on the chest and not excessively loose. As the diagram indicates, the motor operates first in the forward direction to tighten the compression belt, then is turned off for a brief period, then operates in the reverse direction and turns off, and continues to operate through cycles of forward, off, reverse, off, and so on. In parallel with these cycles of the motor state, the cam brake is operating to lock the motor drive shaft in place, thereby locking the drive roller in place and preventing movement of the compression belt. Brake status line


63


indicates the status of the brake


45


. Thus, when the motor tightens the compression belt up to the threshold or time limit, the motor turns off and the cam brake engages to prevent the compression belt form loosening. This effectively prevents relaxation of the patient's chest, maintaining a higher intra-thoracic pressure during hold periods T


2


, T


6


and T


10


. Before the next compression cycle begins, the motor is reversed and the cam brake is disengaged, allowing the system to drive the belt to a looser length and allowing the patient's chest to relax. Upon relaxation to the lower threshold corresponding to the pre-tightened belt length, the cam brake is energized to stop the spindle and hold the belt at the pre-tightened length. The clutch is engaged at all times (the clutch may be omitted altogether if no other compression regimen is desired in the system). (This embodiment may incorporate two motors operating in different directions, connecting to the spindle through clutches.)





FIG. 14



a


illustrates the intra-thoracic pressure changes caused by the compression belt when operated according to the timing diagram of

FIG. 14



a


. The clutch, if any, is always on as indicated by clutch status line


61


. The cam brake is engaged or “on” according to the square wave in the lower portion of the diagram. The motor is on, off, or reversed according to motor state line. Each time the motor is turned on in the forward direction, the belt is tightened around the patient's chest, resulting in a high pressure spike in belt tension and intrathoracic pressure as shown in the pressure plot line. Each time the high threshold limit is sensed by the system and the motor is de-energized, the cam brake engages to prevent further belt movement. This results in a high maintained pressure or “hold pressure” during the hold periods indicated on the diagram (time period T


2


, for example). At the end of the hold period, the motor is reversed to drive the belt to a relaxed position, then de-energized. When the motor is turned off after a period of reverse operation, the cam brake engages to prevent excess slacking of the compression belt (this would waste time and battery power). The cam brake disengages when the cycle is reinitiated and the motor is energized to start another compression. Pulses p


1


, p


2


, are similar in amplitude and duration. Pulse p


3


is limited in duration in this example to show how the torque limit feedback operates to prevent excessive belt compression. Pulse p


3


rapidly reaches the torque limit set on the motor (or the take-up limit set on the belt), and the motor stops and the cam brake engages to prevent injury to the patient and excessive drain on the battery. Note that after motor stop and cam brake engagement under pulse p


3


, belt tension and intra-thoracic pressure are maintained for the same period as all other pulses, and the intra-thoracic pressure is decreased only slightly, if at all, during the high pressure hold period. After pulse, P


3


, a respiration pause may be initiated in which the belt tension is permitted to go completely slack.





FIG. 15

illustrates the timing of the motor, clutch and cam brake in a system that allows the belt compression to completely relax during each cycle. As the table indicates, the motor operates only in the forward direction to tighten the compression belt, then is turned off for a brief period, and continues to operate through on and off cycles. In the first time period T


1


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In the next time period t


2


, the motor is turned off and the cam brake is energized (with the clutch still engaged) to lock the compression belt in the tightened position. In the next time period T


3


, the clutch is disengaged to allow the belt to relax and expand with the natural relaxation of the patient's chest. In the next period t


4


, the motor is energized to come up to speed, while the clutch is disengaged and the cam brake is off. The motor comes up to speed with no effect on the compression belt in this time period. In the next time period, the cycle repeats itself. Thus, when the motor tightens the compression belt up to the threshold or time limit, the motor turns off and the cam brake engages to prevent the compression belt from loosening. This effectively prevents relaxation of the patient's chest, maintaining a higher intra-thoracic pressure. Before the next compression cycle begins, the clutch is disengaged, allowing the chest to relax and allowing the motor to come up to speed before coming under load. This provides much more rapid belt compression, leading to a sharper increase in intra-thoracic pressure.





FIG. 15



a


illustrates the intra-thoracic pressure changes caused by the compression belt when operated according to the timing table of FIG.


15


. The clutch is turned on only after the motor has come up to speed, according to the clutch status line


61


and motor status line


60


, which shows that the motor is energized for two time periods before clutch engagement. The cam brake is engaged or “on” according to the brake status line


62


in the lower portion of the diagram. Each time the clutch is engaged, the belt is tightened around the patient's chest, resulting in a sharply increasing high pressure spike in belt tension and intra-thoracic pressure as shown in the pressure plot line. Each time the motor is de-energized, the cam brake engages and clutch remains engaged to prevent further belt movement, and the clutch prevents relaxation. This results in a high maintained pressure or “hold pressure” during the hold periods indicated on the diagram. At the end of the hold period, the clutch is de-energized to allow the belt to expand to the relaxed position. At the end of the cycle, the cam brake is disengaged (with the clutch disengaged) to allow the motor to come up to speed before initiation of the next compression cycle. The next cycle is initiated when the clutch is engaged. This action produces the sharper pressure increase at the beginning of each cycle, as indicated by the steep curve at the start of each of the pressure Pulses p


1


, p


2


, and p


3


. Again, these pressure pulses are all similar in amplitude and duration, with the exception of pulse p


2


. Pulse p


2


is limited in duration in this example to show how the torque limit feedback operates to prevent excessive belt compression. Pulse p


2


rapidly reaches the torque limit set on the motor, and the motor stops and the cam brake engages to prevent injury to the patient and excessive drain on the battery. Note that after motor stop and cam brake engagement under pulse p


2


, belt tension and intra-thoracic pressure are maintained for the same period as all other pulses, and the intra-thoracic pressure is decreased only slightly during the hold period. The operation of the system according to

FIG. 15



a


is controlled to limit belt pressure to a threshold measured by high motor torque (or, correspondingly, belt strain or belt length).





FIG. 16

illustrates the timing of the motor, clutch and cam brake in a system that does not allow the belt compression to completely relax during each cycle. Instead, the system limits belt relaxation to a low threshold of motor torque, belt strain, or belt length. As the table indicates, the motor operates only in the forward direction to tighten the compression belt, then is turned off for a brief period, and continues to operate through on and off cycles. In the first time period T


1


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In the next time period t


2


, the motor is turned off and the cam brake is energized (with the clutch still engaged) to lock the compression belt in the tightened position. In the next time period T


3


, the clutch is disengaged to allow the belt to relax and expand with the natural relaxation of the patient's chest. The drive spool will rotate to pay out the length of belt necessary to accommodate relaxation of the patient's chest. In the next period t


4


, while the motor is still off, the clutch is engaged (with the cam brake still on) to prevent the belt from becoming completely slack. To start the next cycle at T


5


, the motor starts and the cam brake is turned off and another compression cycle begins.





FIG. 16



a


illustrates the intrathoracic pressure and belt strain that corresponds to the operation of the system according to FIG.


16


. Motor status line


60


and the brake status line


62


indicate that when the motor tightens the compression belt up to the high torque threshold or time limit, the motor turns off and the cam brake engages to prevent the compression belt from loosening. Thus the high pressure attained during uptake of the belt is maintained during the hold period starting at T


2


. When the belt is loosened at T


3


by release of the clutch (which uncouples the cam brake), the intrathoracic pressure drops as indicated by the pressure line. At T


4


, after the compression belt has loosened to some degree, but not become totally slack, the clutch engages (and recouples the cam brake) to hold the belt at some minimum level of belt pressure. This effectively prevents total relaxation of the patient's chest, maintaining a slightly elevated intra-thoracic pressure even between compression cycles. A period of low level compression is created within the cycle. Note that after several cycles (four or five cycles) a respiration pause is incorporated into the compression pattern, during which the clutch is off, the cam brake is off to allow for complete relaxation of the belt and the patient's chest. (The system may be operated with the low threshold in effect, and no upper threshold in effect, creating a single low threshold system.) The motor may be energized between compression period, as shown in time periods T


11


and T


12


, to bring it up to speed before the start of the next compression cycle.





FIG. 17

shows a timing table for use in combination with a system that uses the motor, clutch, and secondary brake


53


or a brake on drive wheel or the spindle itself. The brake


45


is not used in this embodiment of the system (though it may be installed in the motor box). As the table indicates, the motor operates only in the forward direction to tighten the compression belt, and is always on. In the first time period T


1


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In the next time period t


2


, the motor is on but the clutch is disengaged and the brake


53


is energized to lock the compression belt in the tightened position. In the next time period T


3


, the clutch is disengaged and the brake is off to allow the belt to relax and expand with the natural relaxation of the patient's chest. The drive spool will rotate to pay out the length of belt necessary to accommodate relaxation of the patient's chest. In the next period t


4


, while the motor is still on, the clutch is disengaged, but energizing the spindle brake is effective to lock the belt prevent the belt from becoming completely slack (in contrast to the systems described above, the operation of the spindle brake is effective when the clutch is disengaged because the spindle brake is downstream of the clutch). To start the next cycle at T


5


, the motor starts and the spindle brake is turned off, the clutch is engaged and another compression cycle begins. During pulse P


3


, the clutch is engaged for time periods T


11


and T


12


while the torque threshold limit is not achieved the system. This provides an overshoot compression period, which can be interposed amongst the torque limited compression periods.





FIG. 17



a


illustrates the intrathoracic pressure and belt strain that corresponds to the operation of the system according to FIG.


17


. Motor status line


60


and the brake status line


62


indicate that when the motor tightens the compression belt up to the high torque threshold or time limit, the spindle brake engages (according to spindle brake status line


64


) and the clutch disengages to prevent the compression belt from loosening. Thus the high pressure attained during uptake of the belt is maintained during the hold period starting at T


2


. When the belt is loosened at T


3


by release of the spindle brake, the intrathoracic pressure drops as indicated by the pressure line. At T


4


, after the compression belt has loosened to some degree, but not become totally slack, the spindle brake engages to hold the belt at some minimum level of belt pressure. This effectively prevents total relaxation of the patient's chest, maintaining a slightly elevated intra-thoracic pressure even between compression cycles. A period of low level compression is created within the cycle. At P


3


, the upper threshold is not achieved but the maximum time allowed for compression is reached, so and the clutch is engaged for two time periods T


9


and T


10


until the system releases the clutch based on the time limit. At T


9


and T


10


, the spindle brake, though enabled, is not turned on.





FIG. 18

shows a timing table for use in combination with a system that uses the motor, clutch, and secondary brake


53


or a brake on drive wheel or the spindle itself. The brake


45


is not used in this embodiment of the system (though it may be installed in the motor box). As the table indicates, the motor operates only in the forward direction to tighten the compression belt, and is always on. In the time periods T


1


and T


2


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In contrast to the timing chart of

FIG. 17

, the brake is not energized to hold the belt during the compression periods (T


1


and T


2


) unless the upper threshold is achieved by the system. In the next time period T


3


, the clutch is disengaged and the brake is off to allow the belt to relax and expand with the natural relaxation of the patient's chest. The drive spool will rotate to pay out the length of belt necessary to accommodate relaxation of the patient's chest. During T


3


, the belt pays out to the zero point, so the system energizes the spindle brake. During T


4


, the motor remains on, the clutch is disengaged, and the spindle brake is effective to lock the belt to prevent the belt from becoming completely slack (in contrast to the systems using the cam brake, the operation of the spindle brake is effective when the clutch is disengaged because the spindle brake is downstream of the clutch). To start the next cycle at T


5


, the motor starts and the spindle brake is turned off, the clutch is engaged and another compression cycle begins. The system achieves the high threshold during time period T


6


, at peak P


2


, and causes the clutch to release and the spindle brake to engage, thereby holding the belt tight in the high compression state for the remainder of the compression period (T


5


and T


6


). At the end of the compression period, the brake is momentarily disengaged to allow the belt to expand to the low threshold or zero point, and the brake is engaged again to hold the belt at the low threshold point. Pulse P


3


is created with another compression period in which brake is released and the clutch is engaged in T


9


and T


10


, until the threshold is reached, whereupon the clutch disengages and the brake engages to finish the compression period with the belt held in the high compression state. In time period T


11


and T


12


, the clutch is disengaged and the brake is release to allow the chest to relax completely. This provides for a respiration pause in which the patient may be ventilated.





FIG. 18



a


illustrates the intrathoracic pressure and belt strain that corresponds to the operation of the system according to FIG.


18


. In time periods T


1


and T


2


, the motor status line


60


and the brake status line


62


indicate that the motor tightens the compression belt up the end of the compression period (the system will not initiate a hold below the upper threshold). When the belt is loosened at T


3


by release of the spindle brake, the intrathoracic pressure drops as indicated by the pressure line. At T


3


, after the compression belt has loosened to some degree, but not become totally slack, the spindle brake engages to hold the belt at some minimum level of belt pressure. This effectively prevents total relaxation of the patient's chest, maintaining a slightly elevated intra-thoracic pressure even between compression cycles. A period of low level compression is created within the cycle. Motor status line


60


and the brake status line


62


indicate that when the motor tightens the compression belt up to the high torque threshold or time limit, the spindle brake engages (according to spindle brake status line


64


) and the clutch disengages to prevent the compression belt from loosening. Thus the high pressure attained during uptake of the belt is maintained during the hold period starting at T


6


. When the belt is loosened at T


7


by release of the spindle brake, the intrathoracic pressure drops as indicated by the pressure line. At T


7


, after the compression belt has loosened to some degree, but not become totally slack, the spindle brake engages to hold the belt at the lower threshold. At P


3


, the upper threshold is again achieved, so and the clutch is disengaged and the brake is engaged at time T


10


to initiate the high compression hold.





FIG. 19

shows a timing table for use in combination with a system that uses the motor, clutch, and secondary brake


53


or a brake on drive wheel or the spindle itself. The brake


45


is not used in this embodiment of the system (though it may be installed in the motor box). As the table indicates, the motor operates only in the forward direction to tighten the compression belt, and is always on. In the first time period T


1


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In the next time period t


2


, the motor is on, the clutch is disengaged in response to the sensed threshold, and the brake


53


is enabled and energized to lock the compression belt in the tightened position only if the upper threshold is sensed during the compression period. In the next time period T


3


, the clutch is disengaged and the brake is off to allow the belt to relax and expand with the natural relaxation of the patient's chest. The drive spool will rotate to pay out the length of belt necessary to accommodate relaxation of the patient's chest. In the next period t


4


, while the motor is still on, the clutch is disengage, but energizing the spindle brake is effective to lock the belt prevent the belt from becoming completely slack (in contrast to the systems described above, the operation of the spindle brake is effective when the clutch is disengaged because the spindle brake is downstream of the clutch). To start the next cycle at T


5


, the motor starts and the spindle brake is turned off, the clutch is engaged and another compression cycle begins. During pulse P


3


, the clutch is on in time period T


9


. The clutch remains engaged and the brake is enabled but not energized in time period T


10


. The clutch and brake are controlled in response to the threshold, meaning that the system controller is awaiting until the high threshold is sensed before switching the system to the hold configuration in which the clutch is released and the brake is energized. In this example, the high threshold is not achieved during the compression period T


9


and T


10


, so the system does not initiate a hold.





FIG. 19



a


illustrates the intrathoracic pressure and belt strain that corresponds to the operation of the system according to FIG.


19


. Motor status line


60


and the brake status line


62


indicate that when the motor tightens the compression belt up to the high torque threshold or time limit, where the clutch disengages and the spindle brake engages (according to spindle brake status line


64


) to prevent the compression belt from loosening. Thus the high pressure attained during uptake of the belt is maintained during the hold period starting at T


2


. Thus the period of compression comprises a period of active compressing of the chest followed by a period of static compression. When the belt is loosened at T


3


by release of the spindle brake, the intrathoracic pressure drops as indicated by the pressure line. At T


4


, after the compression belt has loosened to some degree, but not become totally slack, the spindle brake engages to hold the belt at some minimum level of belt pressure. This effectively prevents total relaxation of the patient's chest, maintaining a slightly elevated intrathoracic pressure even between compression cycles. A period of low level compression is created within the cycle. Note that in cycles where the upper threshold is not achieved, the compression period does not include a static compression (hold) period, and the clutch is engaged for two time periods T


9


and T


10


, and the system eventually ends the active compression based on the time limit set by the system.





FIG. 20

shows a timing table for use in combination with a system that uses the motor, clutch, the cam brake within gearbox


45


and secondary brake


53


or a brake on drive wheel or the spindle itself. Both brakes are used in this embodiment of the system. As the table indicates, the motor operates only in the forward direction to tighten the compression belt. In the first time period T


1


, the motor is on and the clutch is engaged, tightening the compression belt about the patient. In the next time period t


2


, the upper threshold is achieved and the motor is turned off in response to the sensed threshold, the clutch is still engaged, and the secondary brake


53


is enabled and energized to lock the compression belt in the tightened position (these events happens only if the upper threshold is sensed during the compression period). In the next time period T


3


, with the clutch disengaged and the brakes off, the belt relaxes and expands with the natural relaxation of the patient's chest. The drive spool will rotate to pay out the length of belt necessary to accommodate relaxation of the patient's chest. In the next period t


4


(while the motor is still on), the clutch remains disengaged, but energizing the secondary brake is effective to lock the belt to prevent the belt from becoming completely slacks. To start the next cycle at T


5


, the spindle brake is turned off, the clutch is engaged and another compression cycle begins (the motor has been energized earlier, in time period T


3


or T


4


, to bring it up to speed). During pulse P


3


, the clutch is on in time period T


9


. The clutch remains engaged and the brake is enabled but not energized in time period T


10


. The clutch and brake are controlled in response to the threshold, meaning that the system controller is awaiting until the high threshold is sensed before switching the system to the hold configuration in which the clutch is released and the brake is energized. In this example, the high threshold is not achieved during the compression period T


9


and T


10


, so the system does not initiate a hold. The cam brake serves to hold the belt in the upper threshold length, and the spindle brake serves to hold the belt in the lower threshold length.





FIG. 20



a


illustrates the intrathoracic pressure and belt strain that corresponds to the operation of the system according to FIG.


20


. Motor status line


60


and the brake status line


62


indicate that when the motor tightens the compression belt up to the high torque threshold or time limit, the motor turns off and the cam brake engages (according to cam brake status line


63


) to prevent the compression belt from loosening (the clutch remains engaged). Thus the high pressure attained during uptake of the belt is maintained during the hold period starting at T


2


. Thus the period of compression comprises a period of active t


5


compressing of the chest followed by a period of static compression. When the belt is loosened at T


3


by release of the clutch, the intrathoracic pressure drops as indicated by the pressure line. At T


4


, after the compression belt has loosened to some degree, but not become totally slack, the spindle brake engages to hold the belt at some minimum level of belt pressure, as indicated by the spindle brake status line


64


. This effectively prevents total relaxation of the patient's chest, maintaining a slightly elevated intra-thoracic pressure even between compression cycles. A period of low level compression is created within the cycle. Note that in cycles where the upper threshold is not achieved, the compression period does not include a static compression (hold) period, and the clutch is engaged for two time periods T


9


and T


10


, and the system eventually ends the active compression based on the time limit set by the system.




The previous figures have illustrated control systems in a time dominant system, even where thresholds are used to limit the active compression stroke. We expect the time dominant system will be preferred to ensure a consistent number of compression periods per minute, as is currently preferred in the ACLS. Time dominance also eliminates the chance of a runaway system, where the might be awaiting indication that a torque or encoder threshold has been met, yet for some reason the system does not approach the threshold. However, it may be advantageous in some systems, perhaps with patients closely attended by medical personnel, to allow the thresholds to dominate partially or completely. An example of partial threshold dominance is indicated in the table of FIG.


21


. The compression period is not timed, and ends only when the upper threshold is sensed at point a. The system operates the clutch and brake to allow relaxation to the lower threshold at point b, and then initiates the low threshold hold period. At a set time after the peak compression, a new compression stroke is initiated at point c, and maintained until the peak compression is reached at point d. The actual time spent in the active compression varies depending on how long it takes the system to achieve the threshold. Thus cycle time (a complete period of active compression, release and low threshold hold, until the start of the next compression) varies with each cycle depending on how long it takes the system to achieve the threshold, and the low threshold relaxation period floats accordingly. To avoid extended periods in which the system stalls while awaiting an upper threshold that is never achieved, outer time limit is imposed on each compression period, as illustrated at point g, where the compression is ended before reaching the maximum allowed compression. In essence, the system clock is reset each time the upper threshold is achieved. The preset time limits


75


for low compression hold periods are shifted leftward on the diagram of

FIG. 21



a


, to floating time limits


76


. This approach can be combined with each of the previous control regimens by resetting the timing whenever those systems reach the upper threshold.




The arrangement of the motor, cam brake and clutch may be applied to other systems for belt driven chest compressions. For example, Lach, Resuscitation Method And Apparatus, U.S. Pat. No. 4,770,164 (Sep. 13, 1988) proposes a hand-cranked belt that fits over the chest and two chocks under the patient's chest. The chocks hold the chest in place while the belt is cranked tight. Torque and belt tightness are limited by a mechanical stop which interferes with the rotation of the large drive roller. The mechanical stop merely limits the tightening roll of the spool, and cannot interfere with the unwinding of the spool. A motor is proposed for attachment to the drive rod, and the mate between the motor shaft and the drive roller is a manually operated mechanical interlock referred to as a clutch. This “clutch” is a primitive clutch that must be set by hand before use and cannot be operated during compression cycles. It cannot release the drive roller during a cycle, and it cannot be engaged while the motor is running, or while the device is in operation. Thus application of the brake and clutch arrangements described above to a device such as Lach will be necessary to allow that system to be automated, and to accomplish the squeeze and hold compression pattern.




Lach, Chest Compression Apparatus for Cardiac Arrest, PCT App. PCT/US96/18882 (Jun. 26, 1997) also proposes a compression belt operated by a scissor-like lever system, and proposes driving that system with a motor which reciprocatingly drives the scissor mechanism back and forth to tighten and loosen the belt. Specifically, Lach teaches that failure of full release is detrimental and suggests that one cycle of compression would not start until full release has occurred. This system can also be improved by the application of the clutch and brake systems described above. It appears that these and other belt tensioning means can be improved upon by the brake and clutch system. Lach discloses a number of reciprocating actuators for driving the belt, and requires application of force to these actuators. For example, the scissor mechanism is operated by applying downward force on the handles of the scissor mechanism, and this downward force is converted into belt tightening force by the actuator. By motorizing this operation, the advantages of our clutch and brake system can be obtained with each of the force converters disclosed in Lach. The socketed connection between the motor and drive spool can be replaced with a flexible drive shaft connected to any force converter disclosed in Lach.




Thus, while the preferred embodiments of the devices and methods have been described in reference to the environment in which they were developed, they are merely illustrative of the principles of the inventions. Other embodiments and configurations may be devised without departing from the spirit of the inventions and the scope of the appended claims.



Claims
  • 1. A device for compressing a chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take-up means adapted to take up the belt upon rotation of the belt take up means, said belt take-up means operably connected to the belt; a means for sensing slack take-up said means for sensing slack take-up being capable of indicating when the belt has fit snugly about the chest of the patient; a motor operable connected to the belt take up means through a coupling means, said motor capable of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; wherein said controller is programmed to operate the motor and belt take-up means to take up slack in the belt after initial placement of the belt on the chest of the patient by rotating the motor in a tightening rotation until the means for sensing slack take-up indicates that the belt has fit snugly around the chest of the patient; and wherein said controller is further programmed to operate the motor, the coupling means, and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; and an encoder operably connected to one of the belt or belt take-up means, so that the encoder is capable of measuring the position of the belt, said encoder being capable of transmitting a belt position signal indicative of the position of the belt to the controller.
  • 2. The chest compression device of claim 1 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; and the controller is further programmed, based on the belt position signal, to operate the motor to limit loosening of the belt to the slack-limit position during operating cycles of the chest compression device.
  • 3. The chest compression device of claim 1 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; the controller is further programmed, based on the slack-limit position, to define a set amount of belt displacement to be used during compressions; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the take-up of the belt to the set amount of belt displacement.
  • 4. The chest compression device of claim 1 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the release of the belt so that the belt does not fully return to the slack limit position.
  • 5. The chest compression device of claim 1 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; the controller is further programmed, based on the belt position signal, to operate the motor to limit loosening of the belt to the slack-limit position during operating cycles of the chest compression device; the controller is further programmed, based on the slack-limit position, to define a set amount of belt displacement to be used during compressions; the controller is further programmed, based on the belt position signal, to operate the motor to limit the take-up of the belt to the set amount of belt displacement; the controller is further programmed to operate the motor to momentarily hold the belt at the set amount of belt displacement and then release the belt; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the release of the belt so that the belt does not fully return to the slack limit position.
  • 6. The chest compression device of claim 1 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 7. The chest compression device of claim 1 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 8. The chest compression device of claim 2 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 9. The chest compression device of claim 2 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 10. The chest compression device of claim 3 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 11. The chest compression device of claim 3 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 12. The chest compression device of claim 4 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 13. The chest compression device of claim 4 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 14. The chest compression device of claim 5 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 15. The chest compression device of claim 5 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 16. A device for compressing a chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take-up means adapted to take up the belt upon rotation of the belt take up means, said belt take-up means operably connected to the belt; a means for sensing slack take-up, said means for sensing slack take-up being capable of indicating when the belt has fit snugly about the chest of the patient; a motor operably connected to the belt take up means through a coupling means, said motor capable of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a means for braking the motor; a controller for controlling operation of the motor and the brake; wherein said controller is programmed to operate the motor and belt take-up means to take up slack in the belt after initial placement of the belt on the chest of the patient by rotating the motor in a tightening rotation until the means for sensing slack take-up indicates that the belt has fit snugly around the chest of the patient; wherein the controller is further programmed to operate the motor, the coupling means, and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; and wherein the controller is further programmed to set a threshold of belt tightness and control the brake to momentarily hold the belt at the threshold of belt tightness, and then to release the belt; and an encoder operably connected to one of the belt or belt take-up means, so that the encoder is capable of measuring the position of the belt, said encoder being capable of transmitting a belt position signal indicative of the position of the belt to the controller.
  • 17. The chest compression device of claim 16 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; and the controller is further programmed, based on the belt position signal, to operate the motor to limit loosening of the belt to the slack-limit position during operating cycles of the chest compression device.
  • 18. The chest compression device of claim 16 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; the controller is further programmed, based on the slack-limit position, to define a set amount of belt displacement to be used during compressions; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the take-up of the belt to the set amount of belt displacement.
  • 19. The chest compression device of claim 16 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the release of the belt so that the belt does not fully return to the slack limit position.
  • 20. The chest compression device of claim 16 wherein:the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; the controller is further programmed, based on the belt position signal, to operate the motor to limit loosening of the belt to the slack-limit position during operating cycles of the chest compression device; the controller is further programmed, based on the slack-limit position, to define a set amount of belt displacement to be used during compressions; the controller is further programmed, based on the belt position signal, to operate the motor to limit the take-up of the belt to the set amount of belt displacement; the controller is further programmed to operate the motor to momentarily hold the belt at the set amount of belt displacement and then release the belt; and the controller is further programmed, based on the belt position signal, to operate the motor to limit the release of the belt so that the belt does not fully return to the slack limit position.
  • 21. The chest compression device of claim 16 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 22. The chest compression device of claim 16 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 23. The chest compression device of claim 17 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 24. The chest compression device of claim 17 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 25. The chest compression device of claim 18 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 26. The chest compression device of claim 18 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 27. The chest compression device of claim 19 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 28. The chest compression device of claim 19 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 29. The chest compression device of claim 20 wherein the means for sensing slack take-up comprises a means for sensing the torque on the motor.
  • 30. The chest compression device of claim 20 wherein the means for sensing slack take-up comprises a means for sensing the amount of force applied to the belt.
  • 31. A device for compressing the chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a drive spool operably connected to the belt and adapted to take up the belt upon rotation of the drive spool; a motor operably connected to the drive spool through a coupling, said motor capable of operating the drive spool repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; a current sensor operably connected to the motor said current sensor adapted to sense current drawn by the motor and transmit a corresponding current signal to the controller; said controller programmed to operate the motor and drive spool to take up slack in the belt after initial placement of the belt on the patient's chest by rotating the motor in a tightening rotation until the current signal provided by the current sensor rapidly increases; said controller further programmed to operate the motor and coupling and drive spool to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; a linear encoder scale on the belt and an encoder scanner located in apposition to the scale, said encoder transmitting an encoder position to the controller; said controller further programmed to associate the encoder position to the point of rapid increase in the current signal during initial placement thereby defining a slack-limit position of the belt, and to operate the motor to limit loosening of the belt during cycles to the slack-limit position.
  • 32. A device for compressing a chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take up means adapted to take up the belt upon rotation of the belt take up means, said belt take up means operably connected to the belt; a motor operably connected to the belt take up means through a coupling means said motor ca able of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; a means for sensing the current drawn by the motor and transmitting a corresponding current signal to the controller; wherein said controller is programmed to operate the motor and belt take up means to take up slack in the belt after initial placement of the belt on the chest of the patient by rotating the motor in a tightening rotation until the Current signal provided by the current sensor rapidly increases; wherein said controller is further programmed to operate the motor and the coupling means and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; an encoder means for reading and transmitting a corresponding encoder position to the controller; wherein said controller is further programmed to associate the encoder means position to the point of rapid increase in the current signal during initial placement thereby defining a slack-limit position of the belt, and to operate the motor to limit loosening of the belt during cycles to the slack-limit position.
  • 33. A device for compressing the chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a drive spool operably connected to the belt and adapted to take up the belt upon rotation of the drive spool; a motor operably connected to the drive spool through a coupling, said motor capable of operating the drive spool repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; said controller programmed to operate the motor and drive spool to take up slack in the belt after initial placement of the belt on the patient's chest by rotating the motor in a tightening rotation until the belt is snug about the patient's chest, thereby defining a slack take up position of the belt; said controller further programmed to operate the motor and coupling and drive spool to cause repeated cycles of tightening of the belt about the chest of the patient, tighter than the slack take up position, and loosening of the belt about the chest of the patient to the slack take up position; a linear encoder scale on the belt and an encoder scanner located in apposition to the scale, said encoder transmitting an encoder position signal to the controller; said controller further programmed to associate an encoder position to the slack take up position, and to operate the motor to limit loosening of the belt during cycles to the slack-limit position.
  • 34. A device for compressing the chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a drive spool operably connected to the belt and adapted to take up the belt upon rotation of the drive spool; a motor operably connected to the drive spool through a coupling, said motor capable of operating the drive spool repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; said controller programmed to operate the motor and drive spool to take up slack in the belt after initial placement of the belt on the patient's chest by rotating the motor in a tightening rotation until the belt is snug about the patient's chest, thereby defining a slack take up position of the belt; said controller further programmed to operate the motor and coupling and drive spool to cause repeated cycles of tightening of the belt about the chest of the patient, tighter than the slack take up position, and loosening of the belt about the chest of the patient to the slack take up position; an encoder scale and an encoder scanner located in apposition to the scale, said encoder and encoder scale operably connected to a component of the chest compression device, said encoder transmitting an encoder position to the controller; said controller further programmed to associate an encoder position to the slack take up position, and to operate the motor to limit loosening of the belt during cycles to the slack-limit position.
  • 35. A device for compressing a chest of a patient comprising:a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take up means adapted to take up the belt upon rotation of the belt take up means, said belt take up means operably connected to the belt; a motor operably connected to the belt take up means through a coupling means, said motor capable of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor a means for sensing the current drawn by the motor and transmitting a corresponding current signal to the controller; wherein said controller programmed to operate the motor and drive spool to take up slack in the belt after initial placement of the belt on the patient's chest by rotating the motor in a tightening rotation until the belt is snug about the patient's chest, thereby defining a slack take up position of the belt wherein said controller is further programmed to operate the motor and the coupling means and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; an encoder means for reading and transmitting a corresponding encoder position signal, indicative of the position of the belt, to the controller; said controller further programmed to associate an encoder position to the slack take up position, and to operate the motor to limit loosening of the belt during cycles to the slack-limit position.
  • 36. A method of using a chest compression device to determine a slack-limit position of a belt of the chest compression device, said slack-limit position being set to the size of a particular patient, said method comprising the steps of:providing the chest compression device, said chest compression device comprising: a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take-up means adapted to take up the belt upon rotation of the belt take up means, said belt take-up means operably connected to the belt; a means for sensing slack take-up, said means for sensing slack take-up being capable of indicating when the belt has fit snugly about the chest of the patient; a motor operably connected to the belt take up means through a coupling means, said motor capable of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; wherein said controller is programmed to operate the motor and belt take-up means to take up slack in the belt after initial placement of the belt on the chest of the patient by rotating the motor in a tightening rotation until the means for sensing slack take-up indicates that the belt has fit snugly around the chest of the patient; wherein said controller is further programmed to operate the motor, the coupling means, and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; an encoder operably connected to one of the belt or belt take-up means, so that the encoder is capable of measuring the position of the belt, said encoder being capable of transmitting a belt position signal indicative of the position of the belt to the controller; and wherein the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; placing the belt around the chest of the patient; taking up the slack of the belt with the means for taking up slack until the means for sensing slack take up indicates that the belt has fit snugly on the chest of the patient, whereby the slack limit position of the belt is defined to be the position of the belt when the belt fits snugly on the chest of the patient.
  • 37. A method of performing chest compression, said method comprising the steps of:providing the chest compression device, said chest compression device comprising: a belt adapted to extend around the chest of the patient and fastened on the patient; a belt take-up means adapted to take up the belt upon rotation of the belt take up means, said belt take-up means operably connected to the belt; a means for sensing slack take-up, said means for sensing slack take-up being capable of indicating when the belt has fit snugly about the chest of the patient; a motor operably connected to the belt take up means through a coupling means said motor capable of operating the belt take up means repeatedly to cause the belt to tighten about the chest of the patient and loosen about the chest of the patient; a controller for controlling operation of the motor; wherein said controller is programmed to operate the motor and belt take-up means to take up slack in the belt after initial placement of the belt on the chest of the patient by rotating the motor in a tightening rotation until the means for sensing slack take-up indicates that the belt has fit snugly around the chest of the patient; an encoder operably connected to one of the belt or belt take-up means, so that the encoder is capable of measuring the position of the belt, said encoder being capable of transmitting a belt position signal indicative of the position of the belt to the controller; wherein the controller is further programmed, based on the belt position signal, to set a slack-limit position of the belt at the point when the means for sensing slack take-up indicates that the belt has fit snugly about the chest of the patient; and wherein the controller is further programmed, based on the belt position signal, to operate the motor to limit loosening of the belt to the slack-limit position during operating cycles of the chest compression device; wherein said controller is further programmed to a operate the motor, the coupling means, and the belt take up means to cause repeated cycles of tightening of the belt about the chest of the patient and loosening of the belt about the chest of the patient; placing the belt around the chest of the patient; taking up the slack of the belt with the means for belt take-up until the means for sensing slack take up indicates that the belt has fit snugly on the chest of the patient; and repeatedly tightening and loosening the belt with belt take-up means.
  • 38. The method of claim 37 comprising the further step of limiting the loosening of the belt to the slack-limit position during operating cycles of the chest compression device.
  • 39. The method of claim 37 wherein the controller of the chest compression device is further programmed to operate the motor to momentarily hold the belt at the set amount of belt displacement and then release the belt.
  • 40. The method of claim 39 comprising the further steps of:limiting the loosening of the belt to the slack-limit position during operating cycles of the chest compression device; and momentarily holding the belt at the set amount of belt displacement and then releasing the belt.
Parent Case Info

This application is a continuation of U.S. patent application Ser. No. 09/345,635 filed Jun. 30, 1999, now U.S. Pat. No. 6,398,745, which is a continuation of U.S. patent application Ser. No. 09/087,299 filed May 29, 1998, now U.S. Pat. No. 6,066,106.

US Referenced Citations (43)
Number Name Date Kind
2071215 Peterson Feb 1937 A
2486667 Meister Nov 1949 A
2699163 Engstrom Jan 1955 A
2754817 Nemeth Jul 1956 A
2780222 Polzin et al. Feb 1957 A
2853998 Emerson Sep 1958 A
2899955 Huxley III et al. Aug 1959 A
3042024 Mendelson Jul 1962 A
3120228 Huxley III Feb 1964 A
3368550 Glascock Feb 1968 A
3461860 Barkalow et al. Aug 1969 A
3481327 Drennen Dec 1969 A
3777774 Fryfogle et al. Dec 1973 A
3782371 Dorouineau Jan 1974 A
3835847 Smith Sep 1974 A
4004579 Dedo Jan 1977 A
4338924 Bloom Jul 1982 A
4349015 Alferness Sep 1982 A
4397306 Weisfeldt et al. Aug 1983 A
4570615 Barkolow Feb 1986 A
4655312 Frantom et al. Apr 1987 A
4664098 Woudenberg et al. May 1987 A
4770164 Lach et al. Sep 1988 A
4915095 Chun Apr 1990 A
4928674 Halperin et al. May 1990 A
5056505 Warwick et al. Oct 1991 A
5098369 Heilman et al. Mar 1992 A
5184606 Csorba Feb 1993 A
5217010 Tsitlik et al. Jun 1993 A
5222478 Scarberry et al. Jun 1993 A
5257619 Everete Nov 1993 A
5277194 Hosterman et al. Jan 1994 A
5295481 Geeham Mar 1994 A
5327887 Nowakowski Jul 1994 A
5359999 Kinsman Nov 1994 A
5370603 Newman Dec 1994 A
5399148 Waide et al. Mar 1995 A
5405362 Kramer et al. Apr 1995 A
5474533 Ward et al. Dec 1995 A
5490820 Schock et al. Feb 1996 A
5630789 Schock et al. May 1997 A
5664563 Schroeder et al. Sep 1997 A
5738637 Kelly et al. Apr 1998 A
Foreign Referenced Citations (1)
Number Date Country
PCTUS9618882 Jun 1997 WO
Continuations (2)
Number Date Country
Parent 09/345635 Jun 1999 US
Child 10/162462 US
Parent 09/087299 May 1998 US
Child 09/345635 US