In accordance with 37 C.F.R. §1.52(e)(5), a Sequence Listing in the form of a text file (entitled “Sequence_Listing,” created on Jun. 9, 2011, and 1 kilobytes) is incorporated herein by reference in its entirety.
This invention relates to high-throughput biomolecule analysis and more particularly to a system that utilizes multifunctional encoded particles.
The ability to quantify multiple proteins, cytokines, or nucleic acid sequences in parallel using a single sample allows researchers and clinicians to obtain high-density information with minimal assay time, sample volume, and cost. Such multiplexed analysis is accompanied by several challenges, including molecular encoding and the need to retain assay sensitivity, specificity, and reproducibility with the use of complex mixtures. There are two broad classes of technologies used for multiplexing: planar arrays (1-3) and suspension (particle-based) arrays (4-21), both of which have application-specific advantages. Numbers in parentheses refer to the references appended hereto, the contents of all of which are incorporated herein by reference. Planar arrays, such as DNA and protein microarrays, are best suited for applications requiring ultra-high-density analysis. In comparison, suspension arrays benefit from solution kinetics, ease of assay modification, higher sample throughput, and better quality control by batch synthesis (22). Although particle-based arrays have been used for high-density genotyping applications (23), they are most favorable over microarrays when detecting a modest number of targets over large populations or when rapid probe-set modification is desired. Whereas planar arrays rely strictly on positional encoding, suspension arrays have used a great number of encoding schemes that can be classified as spectrometric (4-11), graphical (12-16), electronic (17-19), or physical (20, 21).
Spectrometric encoding uses specific wavelengths of light or radiation [including fluorophores (4-7), chromophores (8), photonic structures (9), and Raman tags (10, 11)] to identify a species. Fluorescence-encoded microbeads (4-7) can be rapidly processed by using conventional flow cytometry [or on fiber-optic arrays (24)], making them a popular platform for multiplexing. However, there are several disadvantages of using multiple fluorescent signals as means of barcoding, including (i) the limited barcodes achievable (typically ˜100) because of spectral overlap, (ii) the lack of portability for bulky flow cytometers, (iii) added cost with each fluorescent exciter and detector needed, and (iv) potential interference of encoding fluorescence with analyte-detection fluorescence. For these reasons, single-fluorescence methods exist that use graphical techniques to spatially embed barcodes on microcarriers.
Graphical barcodes rely on the patterning of optical elements on a microcarrier; some examples include striped rods (12, 13), ridged particles (14), and dot-patterned particles (14, 15). The chemistries used to fabricate such particles (metallic or photoresist) require additional coupling chemistries to conjugate biomolecules to the surface, and, in the case of striped rods, each metallic pattern needs to be generated one batch at a time. Typically, the patterns on these particles can only be distinguished if the fluorescence of the target signal is sufficiently high. Another graphical method for microcarrier encoding is the selective photobleaching of codes into fluorescent beads (16). In this method, both particle synthesis and decoding are time-consuming, making it an unlikely candidate for high-throughput analysis. A method that eliminates fluorescence altogether uses radio frequency memory tags (17-19). This approach is very powerful because it allows for nearly unlimited barcodes (>1012) and decouples the barcoding scheme from analyte quantification (fluorescence), but the synthesis of any appreciable number (thousands or millions) of these electronic microchip-based carriers may prove to be expensive and slow. These and several other methods developed for multiplexed analysis have been thoroughly reviewed elsewhere (25, 26).
In one aspect, the invention is a method for making multifunctional particles including flowing a first monomer stream loaded with a fluorescent entity along a microfluidic channel. A second monomer stream loaded with a probe adjacent to the first monomer stream flows along the microfluidic channel. The two monomer streams are polymerized to synthesize particles having a fluorescent, graphically encoded region and a probe-loaded region. There may be more than two monomer streams if desired. In a preferred embodiment, the polymerizing step includes exposing the first and second monomer streams to ultraviolet light transmitted through a photomask to create the encoded region. The encoded region may include a plurality of open and closed coding elements. It is preferred that the coding elements be arranged in a two-dimensional grid. An exemplary grid includes 20 coding elements. The grid may also include at least one orientation indicator. In another preferred embodiment, the polymerizing step produces more than one probe-loaded region.
In a preferred embodiment, the ultraviolet light is focused by a microscope objective. A suitable material for the first and second monomer streams is poly(ethylene glycol)diacrylate. The first and second monomer streams preferably include a photoinitiator.
In yet another aspect, the invention is a multifunctional particle including a first graphically encoded region and a second region loaded with a probe.
In yet another aspect, the invention is a high-throughput screening system including multifunctional particles having a graphically encoded region and a probe-loaded region. A flow-focusing microfluidic device is provided to align and read the particles.
We introduce a technique that overcomes many of these multiplexing limitations. By exploiting laminar flows characteristic of microfluidics, we demonstrate the ability to generate multifunctional particles with distinct regions for analyte encoding and target capture (
We used poly(ethylene glycol) (PEG) (well known as a bio-inert polymer) as the particle foundation to eliminate the need to “block” surfaces after probe conjugation and as a transparent material to allow transmission of fluorescent signal from both particle faces. These properties should enhance both specificity and sensitivity of analyte detection. Increasing analyte detection signal can be achieved by making the probes more accessible by increasing surface area in the probe region of the particle. Surface area can be increased by adding, for example, ridges or dimples. Signal can also be increased by increasing pore size. We used a simple dot-coding scheme to generate particles that can bear over a million (220) codes (
To demonstrate the versatility of particle synthesis, we selectively labeled monomer streams with a fluorophore and used a variety of channel designs to generate particles bearing a single probe region, multiple probe regions, and probe-region gradients (
A key feature of our method is the direct incorporation of probes into the encoded particles. This is accomplished by simply adding acrylate-modified biomolecules into the monomer solution. After polymerization, the probes are covalently coupled to the polymer network. This process is applicable for both oligonucleotide and protein probes (30-32). We demonstrate that the short bursts of UV used to synthesize probe-conjugated particles are not detrimental to the functionality of incorporated oligonucleotides. Previously, we showed similar results with bead-bound antibodies that were incorporated into polymer structures made from nearly identical monomer constituents (28, 33).
To demonstrate multiplexing capabilities, we used acrylate-modified oligonucleotide probes (which are commercially available) for DNA sequence detection (
To further demonstrate the power of our multiplexing scheme, we performed the same sequence detection assay with the use of particles with multiple adjacent functionalities (
In order to prove that this method of multiplexed analysis is practical for high-throughput applications, we developed a simple scheme to scan particles in a flow-through device (
A representative particle image is shown (
Although the two-dimensional dot-pattern scheme discussed above is extremely powerful, it will require fairly sophisticated spatial detection. A potentially more practical, “one-dimensional” encoding scheme shown in
In order to accurately and rapidly scan particles in this embodiment, we will use a detection scheme that provides rapid sampling with extremely sensitive light detection. Traditional microarray scanners and flow cytometers (and their microfluidic counterparts) utilize photomultiplier tube (PMT) modules for this purpose (37, 42, 43). These systems provide an extremely high sampling rate (˜MHz) and extraordinary sensitivity, which makes them ideal for this application. These modules can be mounted directly to the camera port of a microscope and used seamlessly with the flow-focusing devices discussed above, given the appropriate illumination source and used with standard data acquisition software.
In a simple setup of this embodiment of the invention, the PMT module will collect all of the light relayed through the microscope port. Therefore, to obtain spatial information that will be used to decode the particles, it is preferred to illuminate only a thin-slit region perpendicular to the direction of particle movement. The slit size will be set by the encoding and probe feature sizes, which will likely be on the order of 10 μm. This resolution can be achieved by simply using a mask with our projection photolithography setup (27). More sophisticated illumination sources such as line-focused laser beams can also be used (44, 45). In another embodiment of the invention, broad illumination can be used with a slit-like filter placed in front of the PMT detector.
The throughput of our system is primarily determined by the detection scheme and the particle size. The particles synthesized for this study are relatively large compared with those in other flow-through methods, measuring 90 μm in width, about 30 μm in thickness, and 180 to 270 μm in length. In some embodiments, for this study, we synthesized particles that were 90 μm wide, ˜30 μm thick and 180, 270, or 450 μm long. Large size not only limits the throughput of a system but also increases the sample volume. However, the great particle-to-particle reproducibility we have demonstrated (28) will afford a much lower redundancy than is typical in flow-through systems, improving efficiency. By using conservative estimates, we found that our system should be capable of providing rapid, high-density analysis with a manageable sample volume (28) despite the seemingly large particle size.
In addition to being very reproducible, we have also shown that our system is very sensitive. With 30-min incubations, we can detect DNA oligomers comfortably at 500 attomoles without biotin-avidin-aided signal amplification (28). This leads us to believe that our system will be at least as sensitive as current, commercially available multiplexing technologies, with the added advantages of all-in-one particle synthesis, incorporation of multiple probes, low cost (28), virtually unlimited codes, and implementation using little more than a standard fluorescence microscope.
Materials
Particles synthesized in this work were made from monomer solutions based on poly(ethylene glycol)diacrylate (PEG-DA, Aldrich, Mn=700) with 2-hydroxy-2-methylpropiophenone photoinitiator (Aldrich). For hybridization experiments, we used monomer solutions of 2:1 PEG-DA:TE Buffer (10 mM Tris pH 8.0 (Rockland), 1 mM EDTA (OmniPur)) with 1-2.5% initiator and a DNA oligomer probe at a concentration of 50 μM. Oligomer probes (IDT) came modified with a reactive Acrydite group and 18-carbon spacer (Probe #1: 5′-Acrydite-C18-ATA GCA GAT CAG CAG CCA GA-3′ (SEQ ID NO: 1), Probe #2: 5′-Acrydite-C18-CAC TAT GCG CAG GTT CTC AT-3′ (SEQ ID NO: 2)). Encoded-region fluorescence was obtained by incorporating 0.005 wt % of methacryloxyethyl thiocarbamoyl rhodamine B (Polysciences) in the respective monomer solution with 1% blue food coloring to easily visualize the co-flowing streams using bright-field microscopy.
Microfluidic Devices
Microchannels were generated using standard procedures in soft lithography. Polydimethylsiloxane (PDMS) elastomer (Sylgard 184, Dow Corning) was cured on a silicon wafer bearing channel reliefs made using SU-8 photoresist (Microchem). Channels were designed such that multiple (2-7) 100 μm-wide inlet channels converged into a common, wider channel (200-500 μm wide). The channels (and reservoirs) were cut out with a scalpel, a hole was punched at each inlet, and the channels were sealed to PDMS-coated glass slides. Pipette tips (10 μl, Molecular BioProducts), connected with rubber tubing (Tygon) to a common pressure source (regulated by a pressure valve, Controlair Inc.), were each filled with ˜5 μl of polymer and inserted into the channel inlet ports. A three way solenoid valve (Burkert, operated manually) allowed for the oscillation between pressurized (˜1 psi, high velocity) and ambient-pressure (no flow) states.
Photopolymerization Setup
Photomasks designed using AUTOCAD were printed with 20 000 dpi resolution at CAD/Art Services, Inc. (Brandon, Oreg.). The masks were inserted into the field-stop position of a Zeiss Axiovert 200 microscope, equipped with a 100 W HBO mercury lamp and wide-excitation ultraviolet (UV) filter set (11000v2:UV, Chroma). A computer-driven shutter system (UniBlitz) was in place to provide the desired pulses of UV excitation. Particles were polymerized across adjacent streams in a microfluidic channel with 30-50 ms bursts of UV excitation. Using the three way valve to control flowrate, particles were polymerized in a no-flow state and immediately flushed down the channel in a pressurized state—providing clean interfaces between adjacent streams with high resolution of particle features. Visual alignment for polymerization was achieved using a CCD camera (KPM 1A, Hitachi) with NIH Image software.
Particle Recovery
Particles were collected in a reservoir after polymerization where they were cleaned of unreacted monomer. The particles were rinsed several times with TE buffer, then with PEG-DA monomer, and again with TE buffer until all residual fluorescence was gone. After each rinse, the particles were allowed to settle to the bottom of the reservoir and excess rinsing solution was pipetted from the top. Particles were typically used immediately for hybridization experiments but were occasionally used after being stored for several days in a dark environment, showing no loss of functionality.
Oligomer Hybridization
Particles were pipetted into separate PDMS reservoirs for each hybridization experiment. Complementary target DNA oligomers modified with a Cy3 fluorophore (IDT) were suspended at concentration of 1 μM in hybridization buffer (TE buffer with 0.2M NaCl (Mallinckrodt) and 0.5% sodium dodecyl sulfate (SDS, Invitrogen)). Solutions of target oligomer were pipetted into the appropriate reservoirs and the particles were incubated for 10 min at room temperature. The particles were then rinsed several times with TE buffer and visualized using an orange longpass filter set (XF101-2, Omega), which is compatible with both rhodamine B and Cy3 fluorophores. Still images were captured using a Nikon D200 digital camera.
Particle Reading
Flow-focusing microfluidic channels were plasma-sealed at the inlets on a PDMS-coated glass slide. Prior to particle reading, the channel was flushed with “read buffer” (TE with 25% PEG-DA and 0.5% Tween 20) for 5 min. Multi-probe particles used in hybridization experiments were resuspended in read buffer and pipetted into the channel reservoir. The pressure system described previously was used with a column of water to pull a vacuum on the channel and induce fluid flow. Particles were observed flowing through a 100 μm channel constriction with a Zeiss Axiovert 200 microscope and imaged with an exposure of 1/250 sec under fluorescence with a CCD camera (Hitachi KPM1A). Movies were taken using NIH image at 10 frames per second—particle velocities were calculated and intensity profiles were taken along the 5 “lanes” of each particle. The average and standard deviation of the intensity along the control region of a particle were taken. A positive reading for an oligomer target was defined as the average plus three standard deviations of the negative control signal.
Investigation of Particle Composition
Probe Concentration
We used serial dilution of an oligomer probe in monomer to determine the effect of concentration on signal detection. We synthesized particles with probe concentrations of 150, 75, 37.5, and 18.75 μM. Control particles (0 μM probe) were used to determine the background intensity (Ib). The particles were incubated for 30 min with a fluorescent-labelled target at 1 rinsed, and imaged to determine fluorescent intensity (Ip). The fluorescent signal is reported in arbitrary units (AUs) taken as the difference IP−Ib. The results are shown in
As can be seen, the intensity increases linearly with probe concentration. This finding is expected when considering the binding of two complementary oligomers—at equilibrium, the relationship is given as:
T+P≈TP
where T=target, P=probe, and T P=double stranded complex. At equilibrium, the concentrations of the species can be characterized by a dissociation constant, Kd, such that:
If [T]o>>[P]o as is the case in our experiment, then [T]≈[T]o. Using this, and the relationship [P]=[P]o−[PT], we obtain:
Thus, for a given initial concentration of target, we can expect the signal (which is proportional to [PT]) to be linear with respect to the probe concentration. Although we could obtain a much higher signal with increased probe concentration, we chose to incorporate 50 μM probe for proof of principle experiments—this concentration gave sufficient signal for target detection with minimal usage of oligomer.
Monomers
We investigated the use of several polymers for barcoded particles, including poly(ethylene glycol)diacrylate of three chain lengths (Mn=200, 400, 700) and also a blend of acrylamide with PEG-DA as crosslinker. The characteristics investigated when selecting a polymer blend were (1) fast polymerization kinetics, (2) low background fluorescence before hybridization, and (3) a strong fluorescence signal after hybridization. All solutions we made consisted of monomer (pure or at 2:1 with TE buffer), 2.5% photoinitiator, and DNA oligomer at 50 μM. Particles were made using 30 ms UV exposure with a 20× objective.
We found that hybridization signals were significantly higher when TE buffer was included in the monomer blend. This is consistent with our previous work (33), in which we discovered that bead-bound proteins incorporated in polymerized PEG hydrogels lost functionality in the absence of buffer. We found that PEG-DA (Mn=700) had significantly faster reaction kinetics than the other monomers and also showed significantly less background signal. The final monomer solution we chose for hybridization experiments was a 2:1 blend of PEG-DA:TE with 1-2.5% photoinitiator.
Particle Characterization
Polydispersity
It is very important that the particles being used for quantitative biomolecule analysis be consistent both morphologically and functionally. We have already shown that particles synthesized using continuous-flow lithography have a very low coefficient of variation (<2%) with respect to physical size (27). We also performed experiments to investigate the variation of fluorescent signal for particles used in a hybridization study.
The fluorescent intensity of particles used in a hybridization study gives evidence of the “functional” polydispersity of the particles. However, this measure is not only dependent on the particles, but also on the hybridization experiment itself—the numbers we present should be a conservative estimate. We found the COV of fluorescent signal to range from 6-10% when incubating with target at concentrations of 1 μM-10 nM and increase to ˜15-30% for lower concentrations (down to 10 pM).
Active Probe Concentration
In order to calculate the concentration of probe on the particle surfaces, we used the concentration of probe incorporated into our monomer solution (50 μM) with an estimate of the depth to which oligomers can diffuse into particles and react. Shown in
As can be seen, the particle shows a significantly higher signal at the edges of its sensing region—the coded region does not show this characteristic because the fluorescent dye incorporated has been homogeneously distributed throughout. We assumed that this “edge” signal was proportional to the amount of oligomer bound to the 30 μm-thick edge of the particle. The intensity in the interior of the particle was almost exactly ½ that of the edge for the several particles we investigated. Assuming that all surfaces are similar, we deduced that the active binding thickness in the interior of the particle must be
or ˜7.5 μm per face. As validation, we can see that the width of the high-intensity region at each particle edge is ˜20 pixels=8 μm. Physically, this can be explained by the pore size. The polymerization of particles occurs most rapidly at the center (where oxygen must diffuse the furthest and the concentration of free radicals is the highest) and slower at the particle surface. Therefore, we can assume that the particles are more tightly crosslinked at the center than near the surfaces, providing larger pores into which short oligomers can diffuse and react with complementary probes. Based on the work of others, the PEG polymer we used should have a pore size of ˜10 Å when completely crosslinked (34, 35).
Using the probe concentration of 50 μM (in monomer), with an active thickness of 7.5 μm per face and a 50% incorporation efficiency, we estimated the effective surface concentration to be:
Because the particles are transparent, the fluorescence of both faces should give a projected probe concentration of
This is a similar “surface” concentration to those reported by others (36). As mentioned previously, we can incorporate a substantially greater amount of probe into our particles to make the surface concentration much higher, thus increasing the sensitivity of our system.
Estimation of Throughput
Recently, microfluidic-based flow cytometers have been developed with integrated photomultiplier tubes to achieve a very high throughput. The fluid velocities in these systems can be on the order of 10 m/s (similar to conventional flow cytometers) while detection is carried out at a high sample rate of 5 MHz, allowing a particle read rate of 17,000/sec (37). We use this as a basis to estimate the throughput we can achieve with our system when incorporating more sophisticated sensing schemes. Conservatively, we estimated a flow velocity of 1 m/s and a spacing of 10 particle lengths between particles (each 200 μm in length). Thus, we can calculate a throughput of
Although this “particle/sec” throughput seems lower than that typically seen in flow cytometry, we discuss why on a “target/sec” basis, the technology should provide a sufficient throughput for high-density analysis.
We have shown that our system exhibits excellent particle-to-particle reproducibility (as discussed above). Enzyme-linked immunosorbant assays (ELISAs), which are considered to be the gold standard in sensing, show similar precision (38) to our technology and are typically done only in duplicate. This level of redundancy is significantly lower than seen in flow cytometry-based assays. We will assume for estimation purposes that our assays can be performed in triplicate to yield accurate information. With a redundancy of 3, our “target/sec” throughput becomes
Two simple modifications that will provide an even higher throughput are (1) reduction of particle size, and (2) incorporation of several probes on a single particle. We have demonstrated in our lab the ability to produce particles with features on the order of ˜1 μm, so it is well within reason to consider the production of encoded particles that are half-sized in each dimension. This would increase the number of particles/volume by a factor of 8. Furthermore, if the particles have 3 functionalities (two probes and a control as in
From this simple analysis, it seems reasonable to analyze one million targets in less than one hour when using reduced-size and/or multi-probe particles. This analysis also highlights the importance of particle size in terms of volume—reduced-size particles will likely be necessary for high-density analysis.
Limits of Detection
We hybridized particles with fluorescent target at concentrations spanning over several orders of magnitude in order to determine a suitable range of detection. Samples of particles from a common batch (with 50 μM probe incorporated) were incubated with 50 μl of target solution for 30 min at room temperature on a vortex mixer—target concentrations ranged from 10 nM-10 pM corresponding to 500 fmol-500 amol (10−18 moles) of oligomer. After hybridization, the particles were rinsed and imaged under fluorescence for detection with an EB-CCD camera (C7-190-20, Hamamatsu) mounted to our microscope. Because the system had an 8-bit limited dynamic range, it was necessary to use three different sensitivity settings to accommodate the broad range in fluorescent signal. As such, two of the particle batches were imaged at two of the three sensitivity settings so the signals could be normalized and plotted on a common scale. The results of the detection study are shown in
As can be seen, this plot is not linear as was the case with varying probe concentrations (
We found the assay to be quite sensitive, detecting oligomer comfortably at the lowest amount tested (500 amole) without biotin/avidin-based signal amplification. This finding suggests that our system has a comparable sensitivity to the current state of the art multiplexing systems including Affymetrix (40) and Luminex (41). Furthermore, we showed previously that we can increase the sensitivity of the assay by incorporating higher probe concentrations of probe into the particles.
Cost of Materials
At standard non-bulk pricing, the raw material cost to produce 1 million single-probe particles similar to those presented in the manuscript is only $ 4.28 ($ 0.14 without DNA probe) as outlined below:
As described earlier, it would be reasonable to make particles half-sized in each dimension, which would reduce particle volume and cost by a factor of 8 (to only $ 0.54 per 106 particles). It is also important to note that very little sample is wasted due to the low-volume microfluidic processing.
The microchannels used for particle synthesis and flow-through reading are of simple design and can be generated very economically. A single 4″ wafer (<$100), generated using standard SU-8 lithography, can easily bear over 10 channels and be molded from many times (we assume 10 times for subsequent calculations). In addition, each device may be used several times (we will use 5). Therefore, using very conservative estimates, the device cost would be:
The material cost for a single multiplex experiment with a million particles would be <$5.00 ($4.28 for the particles and $0.60 for one “synthesis” and one “read” channel). This estimate does not include the buffer used in particle reading, which consists of inexpensive materials and would be negligible in the overall cost.
This application is a divisional of U.S. patent application Ser. No. 11/867,217, filed Oct. 4, 2007, now U.S. Pat. No. 7,947,487 which claims priority to Provisional Application Ser. No. 60/849,651 filed on Oct. 5, 2006, the contents of each of which are incorporated herein by reference in their entirety.
Number | Name | Date | Kind |
---|---|---|---|
3687808 | Merigan et al. | Aug 1972 | A |
4152496 | Barrett et al. | May 1979 | A |
4194066 | Kaetsu et al. | Mar 1980 | A |
4683194 | Saiki et al. | Jul 1987 | A |
4743545 | Torobin | May 1988 | A |
4845205 | Huynh Dinh et al. | Jul 1989 | A |
4929400 | Rembaum et al. | May 1990 | A |
4981957 | Lebleu et al. | Jan 1991 | A |
5117357 | Inoue | May 1992 | A |
5118800 | Smith et al. | Jun 1992 | A |
5130302 | Spielvogel et al. | Jul 1992 | A |
5134066 | Rogers et al. | Jul 1992 | A |
5175273 | Bischofberger et al. | Dec 1992 | A |
5210015 | Gelfand et al. | May 1993 | A |
5319080 | Leumann | Jun 1994 | A |
5359044 | Cook et al. | Oct 1994 | A |
5367066 | Urdea et al. | Nov 1994 | A |
5393878 | Leumann | Feb 1995 | A |
5432272 | Benner | Jul 1995 | A |
5446137 | Maag et al. | Aug 1995 | A |
5457187 | Gmeiner et al. | Oct 1995 | A |
5459255 | Cook et al. | Oct 1995 | A |
5466786 | Buhr et al. | Nov 1995 | A |
5484908 | Froehler et al. | Jan 1996 | A |
5502177 | Matteucci et al. | Mar 1996 | A |
5514785 | Van Ness et al. | May 1996 | A |
5519134 | Acevedo et al. | May 1996 | A |
5525711 | Hawkins et al. | Jun 1996 | A |
5552540 | Haralambidis | Sep 1996 | A |
5567811 | Misiura et al. | Oct 1996 | A |
5576427 | Cook et al. | Nov 1996 | A |
5587469 | Cook et al. | Dec 1996 | A |
5591722 | Montgomery et al. | Jan 1997 | A |
5594121 | Froehler et al. | Jan 1997 | A |
5596091 | Switzer | Jan 1997 | A |
5597909 | Urdea et al. | Jan 1997 | A |
5610300 | Altmann et al. | Mar 1997 | A |
5614617 | Cook et al. | Mar 1997 | A |
5627053 | Usman et al. | May 1997 | A |
5639873 | Barascut et al. | Jun 1997 | A |
5645985 | Froehler et al. | Jul 1997 | A |
5646265 | McGee | Jul 1997 | A |
5658873 | Bertsch-Frank et al. | Aug 1997 | A |
5668268 | Tang et al. | Sep 1997 | A |
5670633 | Cook et al. | Sep 1997 | A |
5677196 | Herron et al. | Oct 1997 | A |
5681941 | Cook et al. | Oct 1997 | A |
5700920 | Altmann et al. | Dec 1997 | A |
5763588 | Matteucci et al. | Jun 1998 | A |
5792747 | Schally et al. | Aug 1998 | A |
5830653 | Froehler et al. | Nov 1998 | A |
5876924 | Zhang et al. | Mar 1999 | A |
5879900 | Kim et al. | Mar 1999 | A |
5916539 | Pilgrimm | Jun 1999 | A |
5945526 | Lee et al. | Aug 1999 | A |
5981180 | Chandler et al. | Nov 1999 | A |
6005096 | Matteucci et al. | Dec 1999 | A |
6150102 | Mills, Jr. et al. | Nov 2000 | A |
6159739 | Weigl et al. | Dec 2000 | A |
6251303 | Bawendi et al. | Jun 2001 | B1 |
6319426 | Bawendi et al. | Nov 2001 | B1 |
6322901 | Bawendi et al. | Nov 2001 | B1 |
6344316 | Lockhart et al. | Feb 2002 | B1 |
6376742 | Zdrahala et al. | Apr 2002 | B1 |
6423551 | Weiss et al. | Jul 2002 | B1 |
6426513 | Bawendi et al. | Jul 2002 | B1 |
6444143 | Bawendi et al. | Sep 2002 | B2 |
6488872 | Beebe et al. | Dec 2002 | B1 |
6532061 | Ortyn et al. | Mar 2003 | B2 |
6576291 | Bawendi et al. | Jun 2003 | B2 |
6592821 | Wada et al. | Jul 2003 | B1 |
6649138 | Adams et al. | Nov 2003 | B2 |
6815064 | Treadway et al. | Nov 2004 | B2 |
6934408 | Frost et al. | Aug 2005 | B2 |
7255994 | Lao | Aug 2007 | B2 |
7319003 | Cantor et al. | Jan 2008 | B2 |
7438792 | Mathies et al. | Oct 2008 | B2 |
7709544 | Doyle et al. | May 2010 | B2 |
7947487 | Doyle et al. | May 2011 | B2 |
8034629 | Chapin et al. | Oct 2011 | B2 |
8232049 | Nilsen et al. | Jul 2012 | B2 |
8535644 | Haghgooie et al. | Sep 2013 | B2 |
8609337 | Pregibon et al. | Dec 2013 | B2 |
8647742 | Dendukuri et al. | Feb 2014 | B2 |
20010023078 | Bawendi et al. | Sep 2001 | A1 |
20020001813 | Taylor et al. | Jan 2002 | A1 |
20020004573 | Domschke et al. | Jan 2002 | A1 |
20020056945 | Gelbart | May 2002 | A1 |
20020155490 | Skinner et al. | Oct 2002 | A1 |
20020165198 | Singh et al. | Nov 2002 | A1 |
20020197614 | Weir et al. | Dec 2002 | A1 |
20030032203 | Sabatini et al. | Feb 2003 | A1 |
20030045597 | Randolph et al. | Mar 2003 | A1 |
20030049629 | Edman et al. | Mar 2003 | A1 |
20030059764 | Ravkin et al. | Mar 2003 | A1 |
20030108900 | Oliphant et al. | Jun 2003 | A1 |
20030143604 | Storhoff et al. | Jul 2003 | A1 |
20040005352 | Lopez et al. | Jan 2004 | A1 |
20040038408 | Abbott et al. | Feb 2004 | A1 |
20040043506 | Haussecker et al. | Mar 2004 | A1 |
20040069857 | Leblans et al. | Apr 2004 | A1 |
20040110141 | Pusey et al. | Jun 2004 | A1 |
20040126820 | Chan et al. | Jul 2004 | A1 |
20040209376 | Natan et al. | Oct 2004 | A1 |
20040210289 | Wang et al. | Oct 2004 | A1 |
20040248163 | Kramer et al. | Dec 2004 | A1 |
20050043428 | Caneba et al. | Feb 2005 | A1 |
20050147973 | Knott | Jul 2005 | A1 |
20050172476 | Stone et al. | Aug 2005 | A1 |
20050196702 | Bryant et al. | Sep 2005 | A1 |
20050214737 | Dejneka et al. | Sep 2005 | A1 |
20050214825 | Stuelpnagel | Sep 2005 | A1 |
20050233318 | Chee et al. | Oct 2005 | A1 |
20060019258 | Yeakley | Jan 2006 | A1 |
20060094025 | Getts et al. | May 2006 | A1 |
20060121122 | Nakajima et al. | Jun 2006 | A1 |
20060147924 | Ramsing et al. | Jul 2006 | A1 |
20060201390 | Lahann et al. | Sep 2006 | A1 |
20060228386 | Stephens et al. | Oct 2006 | A1 |
20060228735 | Bobrow et al. | Oct 2006 | A1 |
20060228742 | Hashmi et al. | Oct 2006 | A1 |
20070003940 | Wang | Jan 2007 | A1 |
20070037195 | Ho | Feb 2007 | A1 |
20070054119 | Garstecki et al. | Mar 2007 | A1 |
20070065844 | Golub et al. | Mar 2007 | A1 |
20070105972 | Doyle et al. | May 2007 | A1 |
20080026394 | Labgold et al. | Jan 2008 | A1 |
20080176216 | Doyle et al. | Jul 2008 | A1 |
20080182239 | Mullinax et al. | Jul 2008 | A1 |
20080213912 | Randall et al. | Sep 2008 | A1 |
20080234144 | Ho et al. | Sep 2008 | A1 |
20080305481 | Whitman et al. | Dec 2008 | A1 |
20090036316 | Drmanac | Feb 2009 | A1 |
20090061424 | Chen | Mar 2009 | A1 |
20090063095 | Bagwell | Mar 2009 | A1 |
20090201504 | Ho et al. | Aug 2009 | A1 |
20100129855 | Kataoka et al. | May 2010 | A1 |
20110129941 | Kumacheva et al. | Jun 2011 | A1 |
Number | Date | Country |
---|---|---|
2006331607 | Jul 2007 | AU |
2388652 | Mar 2005 | GB |
54074886 | Jun 1979 | JP |
11 118819 | Apr 1999 | JP |
WO-9839352 | Sep 1998 | WO |
WO-9914226 | Mar 1999 | WO |
WO-0049385 | Aug 2000 | WO |
WO-0074927 | Dec 2000 | WO |
WO-0116152 | Mar 2001 | WO |
WO-2004026457 | Apr 2004 | WO |
WO-2005103106 | Nov 2005 | WO |
WO-2004076056 | Dec 2006 | WO |
WO-2007050704 | May 2007 | WO |
WO-2007071062 | Jun 2007 | WO |
WO-2007075894 | Jul 2007 | WO |
WO-2007050704 | Aug 2007 | WO |
WO-2007075894 | Jan 2008 | WO |
WO-2008063758 | May 2008 | WO |
WO-2008124423 | Oct 2008 | WO |
WO-2009002225 | Dec 2008 | WO |
WO-2009021923 | Feb 2009 | WO |
WO-2009029742 | Mar 2009 | WO |
WO-2009046149 | Apr 2009 | WO |
WO-2011156432 | Dec 2011 | WO |
Entry |
---|
Armstrong, B. et al. “Cytometry”. 40, No. 2, 102-108 (2000). |
Battersby, B.J. et al. “J. Am. Chem. Soc.” 122, 2138 (2000). |
Beebe et al. “Functional Hydrogel Structures for Autonomous Flow Control Inside Microfluidic Channels”. Nature, 200, 404: 588-590. |
Braeckmans, Kevin et al. “Nature Materials”, 2, No. 3, 169 (2003). |
Braeckmans, Kevin et al. “Nat. Rev. Drug Discov.”, 1, 227 (2002). |
Braeckmans, K. et al. “Nat. Matter” 2, 169 (2003). |
Cruise, G.M. et al. “Biomaterials” 19, 1287 (1998). |
Cunin, F. et al. “Nat. Mater.” 1, 39 (2002). |
De Jager W. et al. “Methods” 38, 294 (2006). |
Dendukuri, D. et al. “Nat. Mater.” 5, 365 (2006). |
Dunbar, S.A. et al. “J. Microbiol. Methods” 53, 245 (2003). |
Evans, M. et al. “Assay Drug Dev. Technol.” 1, 199 (2003). |
Eyal, and Quake. “Electrophoresis” 23, 2653 (2002). |
Fan, J.B. et al. “Nat. Rev. Genet.” 7, 632, (2006). |
Ferguson, J.A. et al. “Anal Chem.” 72, 5618 (2000). |
Fenniri, S. et al. “J. Am. Chem. Soc.” 125, 10546 (2003). |
Finkel, N.H. et al. “Anal Chem.” 76, 353A (2004). |
Fodor, S. P. et al. “Nature” 364, 555 (1993). |
Fulton, R.J. et al. “Clin. Chem”, 43, 1749 (1997). |
Gershon, D. “Nature”, 416, 885 (2002). |
Han, M. et al. “Nat Biotechnol.” 19, 631 (2001). |
Hergt, R. et al. “IEEE Trans. Magn.” 34, 3745 (1998). |
Hunt, H.C. et al. “Microfluidics and Nanofluidics” 4, No. 1-2, 53-79 (2008). |
International Search Report for PCT/US2007/080426, mailed Sep. 30, 2008. |
Irizarry, R.A. et al. “Bioinformatics” 22, 789 (2006). |
Kohara, Y. et al. “Nucleic Acids Res.” 30, e87 (2002). |
McClain et al. “Anal. Chem.” 75, 5646 (2003). |
McHugh, T. M. et al. “J. Clin. Microbial” 26, 1957 (1988). |
Mellott, M.B. “Biomaterials” 22, 929 (2001). |
Moran, E. J. et al. “J. Am. Chem. Soc.” 117, 10787 (1995). |
Nicewarner-Pena et al. “Science” 294, 137 (2001). |
Nicolaou, K. C. et al. “Agnew Chem. Int. Ed.” 34, 2289 (1995). |
Nie et al “Janus and Ternanry Particles Generated by Microfluidic Synthesis: Design, Synthesis and Self-Assembly” J. Am. Chem. Soc. Jul. 2006, 128: 9408-9412. |
Nolan, J.P. et al. “Trends Biotechnol.” 20, 9 (2002). |
Pearce, M.E. et al. “Pharmaceutical Research” 24 No. 12, 2335 (2007). |
Pregibon, D.C. et al. “Langmuir” 22, 5122 (2006). |
Pregibon, Daniel C. et al. “Science” 315, 1393 (2007). |
Roh et al “Biphasic Janus Particles with Nanoscale Anisotropy” Natuer Materials, Oct. 2005, 4: 759-763. |
Service, R.F. “Science” 270, 577 (1995). |
Sha, M. Y. et al. “Anal Bioanal. Chem.” 384, 658 (2006). |
Simonnet, C. et al. “Anal. Chem.” 78, 5653 (2006). |
Sinclair et al. “App. Optics” 43, 2079 (2004). |
Stevens, P.W. et al. “Nucleic Acids Res.” 27, 1719 (1999). |
Su, X. et al. “Nano Lett.” 5, 49 (2005). |
Rehman, F.N. et al. “Nucleic Acids Res.” 27, 649 (1999). |
Rubina, A.Y. et al. “Biotechniques” 34, 1008 (2003). |
Vaino, A. R. et al. “Natl. Acad. Sc”. U.S.A. 97, 7692 (2000). |
Vasiliskov, A.V. et al. “Biotechniques” 27, 592 (1999). |
Wang, et al. “Lab Chip” 4, 625 (2004). |
Written Opinion of the International Searching Authority for PCT/US2007/080426, mailed Sep. 30, 2008. |
Xu, H. et al. “Nucleic Acids Res.”, 31, e43 (2003). |
Zhao, X.W. et al. “Chem. Matter” 18, 2443 (2006). |
Zhi, Z. L. et al. “Anal Chem” 75, 4125 (2003). |
Australian Patent Examination Report No. 1 from AU2007324117, dated Aug. 15, 2012, 4 pages. |
Fotin et al., Parallel thermodynamic analysis of duplexes on oligodeoxyribonucleotide microchips, Nucleic Acids Research, 1998, 26(6): 1515-1521. |
Sorokin et al., Kinetics of hybridization on surface oligonucleotide microchips: Theory, experiment, and comparison with hybridization on gel-based microchips, Journal of Biomolecular Structure & Dynamics, 2006, 24(1): 57-66. |
Ciba Formulators Guide for Coatings—Photoinitiators for UV Curing. Ciba Specialty Chemicals. Available on the web Sep. 2003 at http://www.mufong.com.tw/Ciba/ciba—guid/photo—uv—2.pdf. |
Dendukuri et al., Controlled Synthesis of Nonspherical Microparticles Using Microfluids, Langmuir, 2005, 21: 2113-2116. |
Fialkowski et al., Self-assembly of Polymerc Microspheres of Complex Internal Structures, Nature Materials, 2005, 4, 93-97. |
Hillborg et al., Crosslinked Polydimethylsilozane Exposed to Ocygen Plasma Studied by Neutron Reflectrometry and Other Surface Specific Techniques, Polymer, 2000, 41(18): 6851-6863. |
Jo et al., Three-Dimensional Micro-Channel Fabrication in Polydimethylsiloxane (PDMS) Elastomer, Journal of Microelctromechanical Systems, 2000, 9(1): 76-81. |
Kenis et al., Microfabrication Inside Capillaries Using Multiphase Laminar Flow Patterning, Science, 1999, 285: 83-85. |
Kim et al., Hydrodynamic Fabrication of Polymeric Barcoded Strips as Components for ParalletII Bio-Analysis and Programmable Microactuation, Lab Chip, 2005, 5: 1168-1172. |
Millman et al., Anisotropic Particle Synthesis in Dielectrophoretically Controlled Microdroplet Reactors, Nature Materials 2005, 4: 98-102. |
Mukhoadhyay, Microparticles of all Shapes and Chemistries, Analysical Chemistry, 2006, 4247. |
Nisisako et al., Synthesis of Monodisperse Bicolored Janus Particles with Electrical Anisotropy Using a Microfluidic Co-Flow System, Adv. Mater., 2006, 18: 1152-1156. |
Perro et al., Design and Sythesis of Janus Micro- and Nonoparticles, Journal of Materials Chemistry, 2005, 15: 3745-3760. |
Research Highlights, Lab Chip, 2006, 6: 707-709. |
Research Highlights, Nature, 2006, 440: 848. |
Rolland et al., Direct Fabrication and Harvesting of Monodisperse, Shape-Specific Nanobiomaterials, J. Am. Chem. Soc, 2005, 127: 10096-10100. |
Shiku et al., Oxygen Permeability of Surface-Modified Poly(dimethylsiloxane) Charactarized by Scanning Electrochemical Microscopy, Chemistry Letters, 2006, 35(2) 234-235. |
Sugiura et al., Preparation of Monodispersed Polymeric Microspheres over 50 Micron Employing Microchanncel Emulsification, Ind. Eng. Chem. Res., 2002, 4043-4047. |
International Search Report for PCT/US2006/041668, dated Jun. 18, 2007, 5 pages. |
Written Opinion for PCT/US2006/041668,dated Jun. 18, 2007, 9 pages. |
Albretsen et al., Applications of Magnetic Beads with Covalently Attached Oligonucleotides in Hybridization, Analytical Biochemistry, 189; 40-50 (1990). |
Bong et al., Magnetic Barcoded Hydrogel Microparticles for Multiplexed Detection, Langmuir Article, American Chemical Society, 26(11):8008-8014 (2010). |
Bong, K. et al., Lock Release Lithography for 3D and Composite Microparticles, Lab on a Chip 9(70):863-866 (2009). |
Bullard et al., Direct comparison of nick-joining activity of the nucleic acid ligases from bacteriophage T4, Biochem J, 398; 135-144 (2006). |
Canadian Office Action for Application No. 2.665.536, dated Dec. 19, 2013, 2 pages. |
Castoldi et al., A sensitive array for microRNA expression profiling (miChip) based on locked nucleic acids (LNA), RNA 12(5):913-20 (2006). |
Chapin et al., Ultrasensitive Multiplexed MicroRNA Quantification on Encoded Gel Microparticles Using Rolling Circle Amplification, Analytical Chemistry 83(18):7179-85 (2011). |
Chapin, S.C. et al., High-throughput flow alignment of barcoded hydrogel microparticles, Lab on a Chip, 9(21): 3100-3109 (2009). |
Chen et al., Pre-tension generates strongly reversible adhesion of a spatula pad on substrate, J. R. Soc. Interface 6(35):529-37 (2009). |
Chen et al., Real-time quantification of microRNAs by stem-loop RT-PCR, J. Nucleic Acids Res. 33(20):e179 (2005). |
Chung, S. et al., Plastic microchip flow cytometer based on 2- and 3-dimensional hydrodynamic flow focusing, Microsystem Technno. 9(8):535-533 (2003). |
Chung, T. et al., Recent advances in miniaturized microfluidic flow cytometry for clinical use, Electrophersis, 28(24):4511-20 (2007). |
Collins et al., A DNA polymorphism discovery resource for research on human genetic variation. Genome Res. 8(12):1229-31 (1998). |
Communication pursuant to Article 94(3) EPC for EP11793064.4, 5 pages (Jul. 2, 2014). |
Crooke et al., Antisense Research and Applications, 289-302 (1993). |
Crosland-Taylor, P.J., A device for counting small particles suspended in a fluid through a tube, Nature, 171(4340):37-8 (1953). |
Dendukuri et al., Stop-flow lithography in a microfluidic device, Lab on a Chip 7(7):818-28 (2007). |
Dendukuri et al., Synthesis and self-assembly of amphiphilic polymeric microparticles, Langmuir 23(8):4669-74 (2007). |
Doerr, Mutliplexing to the Max, Nature Methods, 4(5):381 (2007). |
Englisch et al., Chemically Modified Oligonucleotides as Probes and Inhibitors, Angewandte Chemie International Edition, 30(6):613-722 (1991). |
European Office Action, Application No. 07868371.1-1408, 5 pages, Jun. 27, 2014. |
European Search Report, Application No. 11793062.8, Nov. 20, 2013, 10 pages. |
Faivre M. et al., Geometrical focusing of cells in a microfluidic device: an approach to separate blood plasma, Biorheology 43(3):147-59 (2006). |
Fisher et al., Photoinitiated Polymerization of Biomaterials, Annu. Rev. Mater. Res. 31:171-81 (2001). |
Ghosh et al., Covalent attachments of oligonucleotides to solid supports, Nucleic Acids Research, 15; 5353-5372 (1987). |
Gill et al., Encapsulation of Biologicals within Silicate, Siloxane, and Hybrid Sol-Gel Polymers: an Effecient and Generic Approach, J. Am. Chem. Soc. 120: 8587-98 (1998). |
Hall et al., Integrons found in different locations have identical 5′ ends but variable 3′ ends, Journal of Bacteriology, 179: 6286-6294 (1994). |
He, B. et al., Nanowire sensors for multiplexed detection of biomolecules, Current Opinion in Chemical Biology, Current Biology Ltd., London, GB, 12(5): 522-528 (2008). |
Huh, D., et al., Microfluidics for flow cytometric analysis of cells and particles, Physiol. Meas. 26(3):R73-98 (2005). |
International Search Report for PCT/US2009/061474 dated on May 28, 2009, 3 pages. |
International Search Report for PCT/US2009/66778 dated on Jan. 13, 2010, 2 pages. |
International Search Report for PCT/US2011/039529 dated Feb. 9, 2012, 3 pages. |
International Search Report for PCT/US2011/39531 dated Feb. 23, 2012, 6 pages. |
International Search Report for PCT/US2013/029854 dated Jul. 4, 2013, 4 pages. |
Ju et al., Fluorescence energy transfer dye-labeled primers for DNA sequencing and analysis, Proc. Nat'l Acad. Sci., 92; 4347 (1995). |
Kellar, K. et al., Multiplexed microsphere-based flow cytometric immunoassays for human cytokines, J. Immunol. Methods 279(1-2):277-85 (2003). |
Kellar, K.L., et al., Multiplexed microsphere-based flow cytometric assays, Exp. Hematol. 30(11): 1227-37 (2002). |
Kroschwitz, The Concise Encyclopedia of Polymer Science and Engineering, 858-9 (1990). |
Lee et al., Colour-barcoded magnetic microparticles for multiplexed bioassays, Nature Materials, 9:745-749 (2010). |
Lee et al., DNA-based bioanalytical microsystems for handheld device applications, Analytica Chemica 556(1):26-37 (2006). |
Lee, H. et al., Colour-barcoded magnetic microparticles for multiplexed bioassays, Nature Materials, 9(9): 745-749 (2010). |
Lu et al., MicroRNA expression profiles classify human cancers, Nature 435(7043):834-8 (2005). |
Lund et al., Assessment of methods for covalent binding of nucleic acids to magnetic beads, DynabeadsTM and the characteristics of the bound nucletic acids in hybridization reactions, Nucleic Acids Research, 16; 10861-80 (1988). |
Lyamichev et al., Invader Assay for SNP Genotyping, Methods in Molecular Biology 212:229-40 (2002). |
Meade, S. et al., Multiplexed DNA Detection Using Spectrally Encoded Porous SiO 2 Photonic Crystal Particles, Analytical Chemistry, 81(7): 2618-2625 (2009). |
Morgan, E. et al., Cytometric bead array: a multiplexed assay platform with applications in various areas of biology, Clin. Immunol. 110(3):252-66 (2004). |
Nailis et al., Development and evaluation of different normalization strategies for gene expression studies in Candida albicans biofilms by real-time PCR. BMC Mol. Biol. 7:25 (2006). |
Nazarenko et al., Multiplex quantitative PCR using self-quenched primers labeled with a single fluorophore, Nucleic Acids Research 30(9):e37 (2002). |
Nolan T. et al., Quantification of mRNA using realtime RT-PCR, Nat. Protoc. 1(3):1559-1582 (2006). |
Noor, M.R. et al., Electrical Detection of Single-Base DNA Mutation Using Functionalized Nanoparticles, Applied Physics Letters, 95(7), 4 pages (2009). |
O'Connell et al., Testing of the BioSeeq (Smiths Detection Handheld PCR Instrument): Sensitivity, Specificity, and Effect of Interferents on Bacillus Assay Performance (2004). |
Office Action for JP 2009-531603, dated Mar. 8, 2013, 5 pages. |
Panda et al., Stop-flow lithography to generate cell-laden microgel particles, Lab on a Chip 8(7):1056-61 (2008). |
Peck et al., A Method for High-Throughput Gene Expression Signature Analysis, Genome Biology, 7(7):R61 (2006). |
Pregibon and Doyle, Optimization of encoded hydrogel particles for nucleic acid quantification, Anal. Chem. 81(12):4873-81 (2009). |
Ray UK Startup DNA Electronics Developing Handheld Device to Detect Genetic Risk for Drug AEs, Pharmacogenomics Reporter (2009). |
Stears et al., A novel, sensitive detection system for high-density microarrays using dendrimer technology, Physiol Genomics 3: 93-9 (2000). |
Stockton et al., Multiplex PCR for typing and subtyping influenza and respiratory syncytial viruses, J. Clin. Microbiol 36(10):2990-5 (1998). |
Tamura et al., MEGA5: molecular evolutionary genetics analysis using maximum likelihood, evolutionary distance, and maximum parsimony methods, Mol. Biol. Evol. 28(10):2731-9 (2011). |
Van Doorn et al., Quantitative Multiplex Detection of Plant Pathogens Using a Novel Ligation Probe-Based System coupled with Universal, High-Throughput Real-Time PCR on OpernArrays(TM), BMC Genomics 8(1): 1-14 (2007). |
Wang et al., Direct and sensitive miRNA profiling from low-input total RNA, RNA 13(1):151-9 (2007). |
Watson et al., Cystic fibrosis population carrier screening: 2004 revision of American College of Medical Genetics mutation panel, Genet. Med. 6(5):387-91 (2004). |
Wessensteiner et al., PCR Technology: Current Innovations (2007). |
Wolf et al., Rapid hybridization kinetics of DNA attached to submicron latex particles, Nucleic Acids Research, 15; 2911-26 (1987). |
Written Opinion for PCT/US2009/66778, dated on Jan. 13, 2010, 9 pages. |
Written Opinion for PCT/US2011/039529, dated Feb. 9, 2012, 5 pages. |
Written Opinion for PCT/US2011/39531, dated Feb. 23, 2012, 8 pages. |
Written Opinion for PCT/US2013/029854, dated Jul. 4, 2013, 8 pages. |
Yang, A. et al., Hydrodynamic focusing investigation in a micro-flow cytometer, Biomed. Microdevices 9(2):113-22 (2007). |
Zhu et al., High-Sensitivity Capillary Electrophoresis of Double-Stranded DNA Fragments Using Monomeric and Dimeric Fluorescent Intercalating Dyes, Anal. Chem., 66; 1941-8 (1994). |
Pregibon, D.C., Enabling Technologies for Multiplexed Biomolecule Analysis and Cell Sorting, Massachusetts Institute of Technology, Department of Chemical Engineering, 1-122 (May 23, 2008). |
Number | Date | Country | |
---|---|---|---|
20110263747 A1 | Oct 2011 | US |
Number | Date | Country | |
---|---|---|---|
60849651 | Oct 2006 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11867217 | Oct 2007 | US |
Child | 13083496 | US |