The invention relates to a non-invasive blood pressure (NIBP) monitor and a method of operating the same, and in particular relates to improving the comfort of the blood pressure measurements obtained by such a monitor.
Arterial blood pressure (BP) is one of the most important vital signs and is widely used in clinical practice. Non-invasive arterial blood pressure (NIBP) is usually measured by slowly varying the pressure in a cuff that is wrapped around the upper arm of a subject. The BP is determined either by measuring sound distal from the cuff (the auscultatory method, based on Korotkoff sounds) or by measuring pressure pulsations in the cuff caused by volume pulsations of the arm and brachial artery and extracting features from the envelope of these pressure pulses (the oscillometric method). The oscillometric method is easily automated and is widely used.
The principle behind a typical oscillometric method is illustrated by
The typical monitor 20 used for acquiring oscillometric NIBP measurements is illustrated in
The operation of the typical monitor 20 is often uncomfortable for the subject (and in some cases is painful), since the arm is compressed with an external pressure. In a clinical or hospital (or even home) setting where blood pressure measurements need to be obtained through the day and night, the taking of a blood pressure measurement by the monitor 20 will often disturb the sleep of the subject. NIBP monitors originally developed for high acuity subjects (e.g. those in an intensive care unit (ICU)) were optimized for accuracy and precision, but not the comfort of the subject.
In a home setting, it has been found that NIBP measurements have a relatively low acceptance by subjects (e.g. the subjects do not comply with the required measurement schedule or do not perform the measurements properly), which in some cases is due to the pain caused by the inflation of the cuff (which can relate to the duration that the cuff is inflated for and/or the peak pressure in the cuff), irritation of skin under the cuff (particularly on NIBP monitors that are continuously worn by a subject), haematomas, and disturbance of the sleep of the subject.
The comfort of the NIBP measurement can be improved in any or all of three areas: the total measurement time (where a reduction is desired), the maximum cuff pressure reached (where a lower maximum pressure is desired) and the integral of cuff pressure over time (where a smaller integral is desired). Of course, this increase in comfort should not come at the expense of the accuracy of the NIBP measurement beyond acceptable limits.
In addition to the types of monitor described above in which the BP is measured using envelope detection during deflation of the cuff (which can typically take around 45 seconds), monitors have been developed that can measure the BP while the cuff is being inflated. This can reduce the total measurement time (in some cases to around 20 seconds), since the deflation stage can be very quick once the BP measurement has been obtained, and therefore can result in a measurement that is more comfortable for the subject. However, there is further scope for improving the comfort of these inflation-based BP measurements.
Therefore there is a need for an NIBP monitor and method of operating the same that measures the blood pressure during inflation of the cuff and that is more comfortable for the subject than conventional monitors.
According to a first aspect, there is provided a method of obtaining a measurement of the blood pressure of a subject using a non-invasive blood pressure, NIBP, monitor that comprises a pulse rate sensor and a cuff that is to be placed around a limb of the subject, the method comprising using the pulse rate sensor to obtain information on the pulse rate of the subject; adapting a pressure signal filter according to the obtained information on the pulse rate of the subject; starting inflation of the cuff; obtaining a pressure signal representing the pressure in the cuff as the cuff is inflated; filtering the pressure signal using the adapted pressure signal filter during inflation of the cuff; and processing the filtered pressure signal to obtain a blood pressure measurement for the subject during inflation of the cuff.
In some embodiments the step of using the pulse rate sensor to obtain information on the pulse rate of the subject is performed prior to starting the inflation of the cuff.
In some embodiments the step of adapting the pressure signal filter is performed prior to starting the inflation of the cuff.
In some embodiments the steps of using the pulse rate sensor to obtain information on the pulse rate of the subject and adapting the pressure signal filter are also performed during inflation of the cuff.
In some embodiments the step of adapting the pressure signal filter comprises adapting the frequency characteristics of the filter according to the obtained information on the pulse rate of the subject.
In some embodiments the method further comprises the step of stopping the inflation of the cuff when a blood pressure measurement has been obtained.
In some embodiments the method further comprises the step of deflating the cuff once the blood pressure measurement has been obtained.
In some embodiments the method further comprises the step of performing a measurement of the blood pressure during deflation of the cuff if the step of processing the filtered pressure signal to obtain a blood pressure measurement for the subject does not provide a measurement of the blood pressure or an acceptable measurement.
In some embodiments the pulse rate sensor is an accelerometer, a photoplethysmograph, PPG, sensor or an electrocardiograph, ECG, sensor.
In preferred embodiments the pulse rate sensor is a different sensor to the sensor that measures the pressure in the cuff.
In preferred embodiments the information on the pulse rate of the subject is not obtained from the pressure signal.
In some embodiments the NIBP monitor further comprises a pump for inflating the cuff.
According to a second aspect there is provided a computer program product comprising a computer readable medium having computer readable code embodied therein, the computer readable code being configured such that, on execution by a suitable computer, processor or control unit, the computer, processor or control unit is caused to perform any of the methods described above.
According to a third aspect, there is provided a non-invasive blood pressure, NIBP, monitor for measuring the blood pressure of a subject, the NIBP monitor comprising a control unit that is configured to obtain information on the pulse rate of the subject from a pulse rate sensor; adapt a pressure signal filter according to the obtained information on the pulse rate of the subject; control a pump to start inflating a cuff; obtain a pressure signal representing the pressure in the cuff as the cuff is inflated; filter the pressure signal using the adapted pressure signal filter during inflation of the cuff; and process the filtered pressure signal to obtain a blood pressure measurement for the subject during inflation of the cuff.
In some embodiments the NIBP monitor further comprises a pulse rate sensor for measuring the pulse rate of the subject and for providing information on the pulse rate to the control unit.
In some embodiments the pulse rate sensor is an accelerometer, a photoplethysmograph, PPG, sensor or an electrocardiograph, ECG, sensor.
In some embodiments the control unit is configured to obtain the information on the pulse rate of the subject prior to controlling the pump to start inflating the cuff.
In some embodiments the control unit is configured to adapt the pressure signal filter prior to controlling the pump to start inflating the cuff.
In some embodiments the control unit is further configured to obtain information on the pulse rate of the subject from the pulse rate sensor and to adapt the pressure signal filter during inflation of the cuff.
In some embodiments the control unit is configured to adapt the pressure signal filter by adapting the frequency characteristics of the filter according to the obtained information on the pulse rate of the subject.
In some embodiments the control unit is further configured to control the pump to stop the inflation of the cuff when a blood pressure measurement has been obtained.
In some embodiments the control unit is further configured to deflate the cuff once the blood pressure measurement has been obtained.
In some embodiments the control unit is further configured to obtain a measurement of the blood pressure during deflation of the cuff if the processing of the filtered pressure signal does not provide a measurement of the blood pressure or an acceptable measurement of the blood pressure.
In preferred embodiments the pulse rate sensor is a different sensor to the sensor that measures the pressure in the cuff.
In preferred embodiments the control unit does not obtain the information on the pulse rate of the subject from the pressure signal.
In some embodiments the NIBP monitor further comprises a cuff; a pump for inflating the cuff; and a pressure sensor for measuring the pressure in the cuff and for outputting a pressure signal representing the pressure in the cuff to the control unit.
For a better understanding of the invention, and to show more clearly how it may be carried into effect, reference will now be made, by way of example only, to the accompanying drawings, in which:
As described above, measuring the blood pressure (BP) of a subject during inflation of a cuff, rather than during deflation of the cuff from a peak pressure that is sufficient to prevent blood flow in the limb, allows the BP measurement to be completed more quickly, which helps to improve the comfort of the BP measurement for the subject.
The flow chart in
In step 101 inflation of the cuff 28 is started, and the pressure signal representing the pressure in the system is buffered (step 103). In order to obtain a reliable BP measurement, the subject's blood pressure range (diastolic to systolic) needs to be sampled over a certain number of heart cycles (beats). This certain number of heart cycles puts an upper limit on the rate at which the cuff can be inflated from the diastolic pressure to the systolic pressure (since if the cuff is inflated too quickly there will not be enough heart cycles in the signal representing the pressures between diastolic and systolic pressure to obtain a reliable BP measurement). Thus, in step 101 the cuff 28 is inflated at a rate that will allow a sufficient number of heart cycles in the diastolic to systolic range to be measured (typically this rate is set based on an assumption of the lowest possible heart rate in order to ensure that a sufficient number of heart cycles are captured).
In step 105, the buffered pressure signal is processed to determine the heart rate of the subject. This step typically requires the cuff to be at a certain pressure for the heart rate signal to be observed, and a substantial length or duration of pressure signal (e.g. covering a few heart beats) is required in order to be able to extract the heart rate information. If the pressure in the cuff is not yet sufficient to determine the heart rate (as determined in step 107), then no heart rate will be determined in step 105, and the method passes to step 109 in which inflation of the cuff 28 continues and the pressure signal continues to be buffered (step 103). In some embodiments, step 105 can comprise estimating the autocorrelation function of the buffered pressure signal and then finding a peak in that function that most likely corresponds to the subject's heart rate. In some cases the amplitude of the peak can be compared to a threshold to determine if the peak has a sufficient magnitude to represent the heart rate. If the amplitude of the peak exceeds the threshold, the peak can be used as the estimate of the heart rate.
If the heart rate can be extracted from the pressure signal in step 105, the method passes to step 111 in which the heart rate is used to adapt a filter that is to be applied to the pressure signal. Once the filter has been adapted to the heart rate of the subject, the buffered pressure signal can be filtered (step 113) and an analysis of the filtered pressure signal performed to determine the BP (step 115). The filter is a high-pass filter, similar to the high-pass filter used in the traditional oscillometric method. The heart rate (HR) is used to adapt the frequency characteristics of the filter, e.g. the cut-off frequency. In some embodiments, the cut-off frequency can be set to the heart rate (e.g. 60 beats per minute, bpm=1 Hz), but in other embodiments the cut-off frequency may be (slightly) higher or lower than the heart rate (depending on details of the filter design and other processing steps). Generally, the higher the heart rate, the higher the cut-off frequency of the filter (and vice versa). In one example, a basic moving average filter is used, whereby the width of the averaging window is set to correspond to the heart period (=1/HR). Running this filter over the pressure signal returns a low-pass filtered pressure signal, which is then subtracted from the original pressure signal to obtain the high-pass filtered pressure signal.
If the pressure in the cuff 28 is not yet sufficient to determine the BP (e.g. if the cuff pressure has not yet reached the systolic pressure) then the inflation of the cuff continues (step 109). At this stage, as the heart rate has already been determined, it may not be necessary to continue processing the pressure signal to extract the current heart rate (step 105) and adapt the filter (step 111), so the method can return to step 113 after step 109 and the filter can be applied to newly buffered pressure signal data. In other implementations, the extraction of the heart rate and adaptation of the filter can be a continuous process, in which case the method can return to step 103. If at step 115 a BP measurement is determined from the pressure signal, then the method passes to step 119 in which the inflation of the cuff 28 by the pump 22 is stopped, the cuff 28 is deflated and the BP measurement can be reported to the subject, other operator of the monitor 20 or a remote computer or base unit that collates and stores the BP measurement information for the subject.
Thus, in this method, obtaining a BP measurement during inflation of the cuff 28 requires information about the subject's heart rate in order to be able to filter the pressure signal appropriately. Since estimating the heart rate from the pressure signal during inflation requires a certain pressure in the cuff for a substantial length of time (e.g. covering a few heart beats), there can be a delay in being able to perform the BP measurement processing while the heart rate information is extracted (and hence a delay in being able to determine that inflation of the cuff 28 can be stopped), by which time the cuff 28 may have been inflated beyond the maximum pressure required to complete the BP measurement. This can cause discomfort for the subject due to excess pressure and duration of the BP measurement.
Therefore, the invention provides that heart rate information (particularly the pulse rate of the subject) used to adapt the filter applied to the pressure signal is obtained from a different, separate sensor, rather than being derived from the pressure signal itself. This means that heart rate information is available continuously throughout the inflation of the cuff, and the disadvantages with the method in
A non-invasive blood pressure (NIBP) monitor according to an embodiment of the invention is shown in
As in the conventional monitor, the pump 52 is for blowing air or other gas into the cuff 58 in order to inflate the cuff 58 and prevent blood flow in the limb around which the cuff 58 is placed. The valve 56 is used to allow air or gas out of the system and thus deflate the cuff 58.
In accordance with an embodiment of the invention, the monitor 50 further comprises a pulse rate sensor 64 that is for measuring the pulse rate of the subject while the cuff 58 is being inflated. The pulse sensor 64 can be any suitable type of sensor that measures the pulse rate of the subject, and that is able to obtain a pulse rate measurement at any required time (e.g. as required by the BP measurement algorithm).
In some embodiments, the sensor 64 is a photoplethysmography (PPG) sensor, an accelerometer or an ECG sensor, although those skilled in the art will be aware of other types of heart rate sensor that can be used (such as a camera, radar, impedance cardiogram, heart sound sensor, etc.). In the case of a PPG sensor, accelerometer and/or ECG sensor, the sensor 64 can comprise the appropriate sensing apparatus, e.g. light source and detector for a PPG sensor, accelerometer, and two or more electrodes for an ECG sensor, and the processing of the signals from those sensors to determine the pulse rate can be performed by the control unit 62. In the case of an accelerometer, the acceleration signal can be processed to extract the movements caused by the beating of the heart/pulses of blood in the circulatory system.
In use, the pulse rate sensor 64 is attached to or otherwise in contact with the appropriate part of the body of the subject in order to measure the pulse rate. It will be appreciated that in some embodiments, the sensor 64 can be integrated with the cuff 58 so the subject only has to place the cuff around their arm in order to start using the monitor 50, whereas in other embodiments the sensor 64 can be separate from the cuff 58 and placed separately on the body of the subject. In some embodiments, there can be a wired connection between the sensor 64 and the control unit 62, whereas in other embodiments, the sensor 64 can communicate with the control unit 62 wirelessly.
It will be appreciated that
The flow chart in
In the first step, step 131, the pulse rate sensor 64 is used to obtain a measurement of the pulse rate of the subject. This step is first performed at the start of the BP measurement process, preferably just prior to starting the inflation of the cuff 58 by pump 52, although in some embodiments it can be performed as, or just after, inflation of the cuff 58 is started. In any case, the measurement of the pulse rate is available much earlier in the process than in the exemplary method above in which the pulse rate information has to be derived from the pressure signal.
Once the measurement of the pulse rate has been obtained, the filter that is applied to the pressure signal as part of the process for determining the BP measurement is adapted according to the pulse rate measurement (step 133). As in the method of
The inflation of the cuff 58 is started (step 135) and a pressure signal representing the pressure in the cuff 58 is obtained (step 137).
Next, this pressure signal is filtered using the adapted filter (step 139). This filtering is performed during inflation of the cuff 58, and preferably in real time as the pressure signal is obtained (or as near to real time as is possible). The filtered pressure signal is then analyzed to determine the BP measurement (step 141). Those skilled in the art will be aware of various techniques for performing this analysis, and therefore further details are not provided herein. Again, this step is performed during inflation of the cuff 58, and preferably in real time (or as near to real time as is possible).
If a BP measurement cannot be obtained from the pressure signal (e.g. if the pressure in the cuff 58 has not yet reached the systolic blood pressure in the subject), then following step 143 the inflation of the cuff 58 continues and the method returns to step 137 where the pressure signal continues to be obtained, filtered and analyzed. If a BP measurement is obtained in step 141, the method passes to step 145 in which the inflation of the cuff 58 is stopped and the deflation of the cuff 58 is started (e.g. by opening valve 56). Since the BP measurement has been obtained, the deflation is preferably performed as fast as possible. The deflation is preferably initiated as soon as or immediately after the BP measurement has been obtained. The BP measurement can be reported to the subject, other operator of the monitor 50 or a remote computer or base unit that collates and stores the BP measurement information for the subject.
The graph in
Instead, by obtaining the heart rate information (particularly the pulse rate of the subject) from a different, separate sensor 64 (and therefore not from the pressure signal), the heart rate information is available continuously throughout the inflation of the cuff 58, which means that the processing of the pressure signal to obtain the BP measurement can start as soon as the inflation of the cuff 58 starts (and thus always starts before the systolic pressure is reached in the cuff 58) and therefore the BP measurement can be obtained at the earliest opportunity.
In some embodiments, since the pulse rate of the subject can change over the course of the BP measurement process, the pulse rate sensor 64 can also be used to obtain measurements of the pulse rate of the subject during inflation of the cuff 58, and those measurements can be used to adapt the filter to the current pulse rate of the subject. This helps to ensure that the filter is up to date for the current pulse rate of the subject and therefore provide a BP measurement that is as accurate as possible.
In some embodiments, if for any reason the analysis in step 141 is unable to obtain a BP measurement (or a BP measurement that is reliable), for example if the subject moves during the measurement process or experiences an arrhythmia, the control unit 62 can control the NIBP monitor 50 to perform a conventional deflation-based BP measurement. Thus, when the systolic pressure has been exceeded (or when a preset maximum inflation pressure has been reached without being able to obtain a BP measurement) the control unit 62 can control the pump 52 to stop inflating the cuff 58, and operate the valve 56 to release pressure from the cuff 58 in a controlled manner. The pressure signal from the pressure sensor 54 obtained while the cuff 58 deflates is analyzed by the control unit 62 using a conventional deflation-based algorithm to determine the blood pressure of the subject.
There is therefore provided an NIBP monitor and a method of operating the same that measures the blood pressure of a subject during inflation of the cuff and that is more comfortable for the subject than measurements made according to conventional methods or by conventional monitors.
While the invention has been illustrated and described in detail in the drawings and foregoing description, such illustration and description are to be considered illustrative or exemplary and not restrictive; the invention is not limited to the disclosed embodiments.
Variations to the disclosed embodiments can be understood and effected by those skilled in the art in practicing the claimed invention, from a study of the drawings, the disclosure and the appended claims. In the claims, the word “comprising” does not exclude other elements or steps, and the indefinite article “a” or “an” does not exclude a plurality. A single processor or other unit may fulfil the functions of several items recited in the claims. The mere fact that certain measures are recited in mutually different dependent claims does not indicate that a combination of these measures cannot be used to advantage. A computer program may be stored/distributed on a suitable medium, such as an optical storage medium or a solid-state medium supplied together with or as part of other hardware, but may also be distributed in other forms, such as via the Internet or other wired or wireless telecommunication systems. Any reference signs in the claims should not be construed as limiting the scope.
Number | Date | Country | Kind |
---|---|---|---|
14188411.4 | Oct 2014 | EP | regional |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2015/073407 | 10/9/2015 | WO | 00 |