The invention described herein is intended for use in preparative purification and analytical separation of large molecules using a novel design for a chromatographic purification system. The invention differs from single modal chromatographic purification by utilizing multimodal chromatography in an oscillating series.
As described by use of standard gradient elution chromatography models (described herein), a single modal chromatography column operating in a gradient elution mode is limited because the solid phase interaction with the target solute occurs within a specific mobile phase concentration range. A small region of the mobile phase gradient contributes to the separation. This active region of solvent concentration is depicted in
The novel multimodal chromatography in an oscillating series utilizes the alternating solvent exchange media to reposition the active region of separation back in phase with the target solute. The novel oscillatory system is specifically designed to reposition the gradient with respect to the solute positions to move the solutes back into the region of active separation. Each solvent exchange column bed length in the sequence is designed to achieve a subtle decrease in the solvent concentration associated with the two solutes of interest which results in an extension of the active separation.
Note that the gradient elution with the novel system can be operated in one of three modes:
Each Column in a Novel Oscillatory Alternating Positive Gradient Slope Column System
Theoretical modeling is used herein to demonstrate an improved separation using the novel multimodal oscillatory chromatographic purification system. A one-dimensional model is used to describe a challenging separation between two closely related peptides. A reverse phase system is chosen for the detailed description of the invention. This invention extends to any adsorption system.
The Solute velocity dependence on organic solvent concentration in gradient elution with reverse phase chromatography is described by the following standard chromatography solute movement and reverse phase gradient equations:
Δta=Δt0(k′a+1) Equation 2b
log k′a=Sφ+c Equation 2c
k′
a=10(sφ+c) Equation 2d
Utilizing Equation 2b in velocity expression:
for example scenario
Combining Equation 2e and 2d provides an expression for solute “a” velocity:
The organic solvent gradient concentration dependence on time and column axial distance in rev phase chromatography is described by the following equation (Equation 2g). Equation 2g is a linear expression for 2-dimensional (time and column axial distance) linear elution gradient.
(change in organic solvent conc. per cm of column axial distance)
The expression for elution time, Equation 2j, in a size exclusion (SEC) column is developed from the fundamental expression describing solute movement in SEC columns, equation 2h.
The linear Expression for 2-dimensional (time and column axial distance) linear elution gradient used for the reverse phase column applies to the SEC column. The total void fraction will have a different value because SEC media typically are designed with a large void fraction compared to adsorptive media.
Note that the gradient slope (in units of change in organic solvent concentration per cm of axial distance) for the size exclusion column will be different than for the reverse phase column because the total void fraction for the size exclusion column is different than the total void fraction for the reverse phase column. The gradient slope for the size exclusion column can be determined from the reverse phase column slope using the void fraction ratio for each column per equation 2k:
A numerical computational method is used to determine the elution time of each solute of interest. Equations 2g, 2f, and 2j are used in computations to describe separation of two closely related large molecule species. Each column is numerically integrated with the output conditions used as initial conditions for the subsequent column sequence.
One operating mode of the Novel Oscillatory Chromatographic Purification System with a Positive Gradient Slope is illustrated herein by utilizing the one dimensional model per the previous description utilizing equations 2g, 2f and 2j.
The novel multimodal chromatography in an oscillating series utilizes the alternating solvent exchange media to reposition the solvent gradient active region of separation back in phase with the target solute as shown in
Table 1 provides the Novel Oscillatory Alternating Column design parameters and theoretical results for the specific model system used in the scenario to generate
Design parameters for the Novel Oscillatory Alternating Column system include gradient slope, bed depth of each adsorptive (rev. ph.) column, bed depth of each solvent exchange column, and gradient start concentration and are listed in
The novel alternating column system can be designed for any number of column pairs. This example utilizes 12 media section pairs or 12 column pairs. The media section pairs will be referred to as column pairs with the caveat that the novel alternating media hardware may be designed as media sections in a single column, or separate columns for each media. The column lengths or media section lengths are identical in each pair, thus allowing a looped configuration where the feed solution is injected into the system and recycled through a loop configuration that could be recycled 12 times through a single column pair, or 6 times through a double column pair (2 RPC and 2 SEC columns) to achieve the same results as a once-through 12 column pair system.
Note in the Table 1 list of parameters, the starting organic solvent concentration of the gradient is 0.30 or 30%. This is the organic solvent concentration that provides the largest difference in the large molecule solute velocities of the two solutes of interest in the separation scheme. A plot of the velocity difference in solute of interest 1 and 2 verses organic solvent concentration in the RPC media is shown in
In the example presented here, the cycle of sequential columns does not extend beyond 12 cycles. If the sequence of columns is established as a repeatable configuration, the system could be design as a loop with an injection port and the system could be recycled until the desired separation is achieved.
A second operating mode of the Novel Oscillatory Chromatographic Purification System with a Negative Gradient Slope is illustrated herein by again utilizing the one dimensional model per the previous description utilizing equations 2g, 2f and 2j.
Alternatively to the positive gradient slope design, the novel oscillatory chromatography configuration can be designed to accommodate a negative slope gradient. The negative gradient slope design produces an acceleration in the differential migration rate of the two solutes of interest.
The novel multimodal chromatography in an oscillating series utilizes the alternating solvent exchange media to reposition the active region of separation back in phase with the faster moving solute of interest, component 1, while the slower moving solute of interest, component 2, is exposed to a decreasing organic solvent concentration as shown in
An SEC column length for the first size exclusion column (labeled sec 1 in
The series of alternating columns produces an ever increasing difference in solvent concentration associated with each component or solute of interest if the column lengths and gradient slope are set to the conditions described by Table 2.
Table 2 provides the Novel Oscillatory Alternating Column design parameters and theoretical results for the specific model system used in the scenario to generate
Design parameters for the Novel Oscillatory Alternating Column system include gradient slope, bed depth of each adsorptive (rev. ph.) column, bed depth of each solvent exchange column, and gradient start concentration and are listed in Table 2.
In the example presented here, the cycle of sequential columns does not extend beyond 12 cycles. If the sequence of columns is established as a repeatable configuration, the system could be design as a loop with an injection port and the system could be recycled until the desired separation is achieved.