The invention relates to an atomization system for pressurized spray for use in various applications, including pharmaceutical applications involving pressurized metered dose inhalers, which reduces aerosol plume velocity and increases efficiency of spray delivery.
Pressurized metered dose inhalers (“pMDIs”) are common in the industry, as are conventional MDI actuators that are based on “two-orifice-and-sump” designs. Unfortunately, their commonality is accompanied by their common disadvantages: high spray velocity and inadequate particle size distribution control, which results in poor drug delivery to the patient. It is postulated that high spray velocity is one of the lead causes of high oropharyngeal drug deposition (Newman S., (2005) “Principles of Metered-Dose Inhalers”, Respiratory Care, Vol. 50. No. 9. pp. 1177-1190). Previous attempts at solving this problem are evidenced by the use of spacers to reduce spray velocity (e.g. U.S. Pat. No. 4,972,830). However, spacers are bulky, and drug deposition within the spacers leads to a decrease of actual drug delivered to the patient. Other methods for slowing plume force include introducing complicated baffles or bluff bodies into the device nozzle or mouthpiece, or introducing a flow control/mixing chamber into the mouthpiece (e.g. U.S. Pat. Nos. 6,615,826 and 6,527,151). These methods, however, also have the propensity to increase drug deposition in the mouthpiece at the site of the baffles, bluff bodies, or other airflow obstructions.
Modifying the aerosol generation mechanism itself using a vortexing chamber produces a low plume force spray, as described in U.S. Pat. No. 6,418,925, which is incorporated herein by reference. According to the present invention, modifying the actuator design by using a flat or protruding nozzle face, as opposed to a “standard” concave-conical nozzle face (common in the industry) and a diverging mouthpiece insert further reduces drug deposition on the nozzle face, the device mouthpiece, and the throat of the patient.
Applicants have discovered that a flat or protruding nozzle face affects the dynamics of the aerosol flow at the nozzle orifice, which in turn affects the particle size distribution in the aerosol spray leaving the orifice. Upon actuation using the present invention, the drug-propellant mixture from the canister enters the vortex chamber of the nozzle of the present invention at an angle. The mixture flows along the periphery of the chamber which sets up a swirling motion, until the mixture leaves the device via an axial exit orifice at a decelerated velocity. The nozzle face geometry discovered by the applicants restricts the extent as well as spread of unvaporized drug-propellant mixture around the nozzle orifice, thus limiting drug deposition around on the nozzle face. The mouthpiece insert works to further decelerate the spray.
The atomization system of the present invention incorporates a vortexing nozzle with a flat or protruding exit orifice face and a diverging mouthpiece insert, which results in high fine particle fraction and modest throat deposition in conjunction with reduced aerosol plume velocity and impact force. Various embodiments of the present invention and related methods are also disclosed in the following description. The various embodiments can be used alone or in any combination, as is desired.
The present invention will now be described with regard to the Figures.
The swirl chamber is preferably conical, where the first swirl chamber end 22 is greater than the second swirl chamber end 24. The swirl chamber cone angle B is 60 to 120 degrees, preferably 90 degrees. The second swirl chamber end 24 is connected to nozzle exit orifice 20, which is connected to nozzle face 30. Nozzle exit orifice 20 has a length 21.
As shown in
A nozzle back seal 8 is used to close the back of the nozzle after the nozzle has been manufactured. Back seal 8 is preferably attached to the nozzle using ultrasonic welding or some other method such as by interference fit that would be known to one of skill in the art. It should also be understood that the atomization system of the present invention could also be manufactured in one piece, eliminating the need for nozzle back seal 8, using standard manufacturing methods known in the art.
More specifically,
Returning to
According to one aspect of the present invention, the shape of the insert 50 is preferably round at an end 52 proximal to the nozzle 10 and gradually expands into an elliptical shape at an end 54 distal to the nozzle 10. Preferably, the cross-sectional area at the distal end 54 is more than three (3) times the cross-sectional area at the proximal end 52. For pharmaceutical/inhaler applications, the optimal insert 50 length is about one inch. Make-up air enters the insert 50 at the proximal end 52 via one or more air inlets 56 and is expelled from the device with the aerosol at the distal end 54. The increase in cross-sectional area causes a corresponding decrease in the aerosol plume velocity. The law of conservation of mass provides that the decrease in aerosol velocity is nearly inversely proportional to the increase in cross sectional area along the length of the diverging mouthpiece insert 50. According to a preferred embodiment, the diverging mouthpiece insert 50 is elliptical or oval in shape, with a minor inner diameter at the distal end 54 of approximately 0.8″ a major inner diameter at the distal end 54 of approximately 1.0″ and an inner diameter at the inlet of the proximal end 52 of the insert 50 of approximately 0.5″.
The unexpected improvements of the atomization system of the present invention are still further achieved by the design of the flat or protruding nozzle face along with careful relative dimensioning of other nozzle components as follows. Referring to
The present invention will now be described in terms of its effect on pressurized liquid that enters the nozzle 10. As shown on
The geometry of the flat or protruding nozzle face of the present invention prevents unvaporized drug/propellant mixture from pooling on the nozzle face because the mixture is not “shielded” by the design of the nozzle face as occurs with a standard “concave-conical” nozzle face. Any unvaporized mixture tends to be stripped frequently from the surface of the nozzle face and carried with the aerosol plume, avoiding the formation of very large droplets. As a result, the particle size distribution of the aerosol plume leaving the nozzle orifice of the present invention is finer, which also aids in more accurate deliverer of drug to the patient's lungs.
The standard conical nozzle face causes significant drug deposition on the nozzle face, as opposed to a flat nozzle face, which significantly reduces the drug deposition on the face of the nozzle. Therefore, using a flat or protruding nozzle face significantly increases the amount of drug that reaches the desired therapeutic target, e.g. the patient's lungs, in the case of a systemic drug. The reduction in drug residue around exit orifice also represents a lesser risk of orifice clogging with suspension formulations, and hence modest cleaning requirements.
Another study comparing the velocities and particle size distribution of the sprays emitted from three different actuation systems: (a) a standard actuator including a concave-conical nozzle face and standard mouthpiece (Inlet ID=0.44″: outlet minor ID=0.63″, major ID=0.92″, optimal length=1.0″), (b) the actuator system of the present invention including a flat nozzle face and diverging mouthpiece insert (Inlet ID=0.50″; outlet minor ID=0.8″, major ID=1.0″, optimal length=1.0″), and (c) the commercial Proventil® actuator (mouthpiece is a tapering smooth-rectangular tube: outlet minor ID=0.58″, major ID=0.77″, length from nozzle=1.35″), which was chosen because of its low plume force. The aerosol velocity and droplet size were measured using phase-Doppler anemometry. A two-component LDA/PDA system from Dantec Dynamics A/S was used. Axial/horizontal and radial/vertical components of velocity as well as droplet size were measured along the mouthpiece exit plane for the three configurations. For configurations (a) and (b), the measurements were taken 30 mm from the exit orifice. The distance between the actuator orifice and the measurement plane was about 25% greater for the Proventil® configuration (c) due to the longer mouthpiece.
The velocities of the spray for the improved system were lower and limited to a smaller central area of the plume (max velocity 12.1 m/s over approximately 5 mm2), compared to the velocities of the Proventil plume which are greater over a broader plume area in spite of its longer mouthpiece (max velocity 13.5 m/s over approximately 17.5 mm2). The average droplet particle size of the improved system measured approximately 1.29 μm at the center of the plume, compared to approximately 2.10 μm at the center of the plume for the standard nozzle/mouthpiece configuration and approximately 1.73 μm at the center of the Proventil plume. This indicates that a greater amount of the propellant/medicament mixture is aerosolized into smaller particles exiting the device of the present invention.
As discussed above, the nozzle in accordance with the present invention is preferably connected to a diverging mouthpiece insert that also helps decelerate plume velocity. Applicants have also discovered that a larger mouthpiece on an inhalation device causes the user to open their mouth wider, increasing the tongue-to-hard-palate opening in the user's oral cavity. This promotes deep lung deposition by enabling more aerosol spray to reach the deep lung, unimpeded.
In one study the diverging mouthpiece insert (Inlet ID=0.50″; outlet minor ID=0.8″, major ID=1.0″, optimal length=1.0°) slowed the plume force measured at different distances ranging from 4 to 12 cm: from 28-20 mN range (using standard mouthpiece, Inlet ID=0.44″: outlet minor ID=0.63″, major ID=0.92″, optimal length=1.0″) to about 15-8 mN range (using the atomizing nozzle of the present invention with a flat-faced configuration). Plume velocity using the optimized nozzle of the present invention decreased from about 20 m/s at the start of the mouthpiece insert to less than 12 m/s at the exit plane.
In vitro testing of the mouthpiece insert of the present invention, described below, has shown that a larger mouthpiece results in decreased drug deposition inside the mouthpiece as well.
The following data is the result of in vitro testing using a suspension drug formulation for three different variations of the diverging mouthpiece insert, each using a standard “concave-conical” nozzle face to isolate the difference in results based on different geometrical configurations of the insert alone. The difference in the number of data points used for the first and second configurations is the result of faulty equipment used to measure the deposition; in the case of the third configuration, however, a “round” insert was used for comparison purposes only. Preferably, the present invention contemplates an elliptical diverging mouthpiece insert for ergonomic purposes.
The data in Table 1 shows that drug deposition in the horn and nozzle decreases as the inlet and outlet diameters of the mouthpiece insert increase.
Another study was performed to illustrate the effects of the flat or protruding nozzle face in combination with the diverging mouthpiece insert, as compared to the standard convex conical nozzle face and smaller mouthpiece. Applicant used an Andersen Cascade Impactor to measure the fine particle fraction captured, simulating fine particle distribution to the lung of a patient.
Configuration A used a flat inner nozzle face with an elliptical mouthpiece horn of the following dimensions: 0.44″ internal diameter of the mouthpiece horn inlet, a 0.8″ minor axis of the mouthpiece horn outlet, and a 1.0″ major axis internal diameter of the mouthpiece horn outlet.
Configuration B used a flat inner nozzle face with the preferred dimensions of the diverging mouthpiece insert: 0.5″ internal diameter of the mouthpiece horn inlet, a 0.8″ minor axis of the mouthpiece horn outlet, and a 1.0″ major axis internal diameter of the mouthpiece horn outlet.
Configuration C used a standard concave-conical nozzle face with an elliptical mouthpiece horn of the following dimensions: 0.44″ internal diameter of the mouthpiece horn inlet, a 0.63″ minor axis of the mouthpiece horn outlet, and a 0.92″ major axis internal diameter of the mouthpiece horn outlet.
Configuration D used a protruding convex-conical face (see
The Table 2 data shows the mean percentage of particles per spray of 35 U/actuation of rh-insulin that were captured. Deposition of large particles in the nozzle, horn, and patient throat (simulated using the uppermost plates, 0, 1 and 2, of the cascade impactor) is decreased in configurations A, B and D, which utilize the flat and protruding inner nozzle faces and elliptical mouthpiece inserts, compared to the standard inner nozzle face and smaller mouthpiece insert (configuration C). Further, the fine particle fraction captured at lower plates using cascade impact testing is greater using the flat and protruding inner nozzle faces and mouthpiece inserts (configurations A, B and D) than that using the standard inner nozzle face and standard mouthpiece (configuration C). This data illustrates that a greater amount of the drug administered to a patient in each puff or spray using one of the optimized configurations A, B or D, reaches the deep lung.
Yet another study compared drug deposition and fine particle fraction captured from the actuation system of the present invention, hereinafter termed the “VNA”, to another commercialized actuator, using an Andersen Cascade Impactor. The following systems were compared:
The results of tests on these two systems using 60 U/actuation insulin and 20 U/actuation insulin were observed. Three “sprays” or samples were taken from each canister, at both the “start” of canister life, after three priming sprays, and at the “end” of canister life, after 114 sprays total (three priming sprays, three sample sprays, and 108 “wasting” sprays).
Canisters 221, 222, and 223 contained a “rapid release” rh-insulin and HFA 134a formulation and were used for testing the Kos VNA. Table 3 provides the results of testing Cans 221, 222, and 223 containing 60 U/actuation tested in the Kos VNA at the beginning of canister life, after three priming sprays. Three samples were taken from each canister and averaged.
Canisters 224, 225, and 226 contained the same rh-insulin and HFA 134a formulation as tested in Canisters 221 through 223 and were used for testing the Valois device. Table 4 provides the results of testing Cans 224, 225, and 226 containing 60 U/actuation in the Valois device at the beginning of canister life, after three priming sprays. Three samples were taken from each canister and averaged.
From these 60 U/actuation tests at the beginning of canister life, 67% of the total mass recovery using the Kos VNA was fine particles <4.7 μm. Only 44% of the total mass recovery from the Valois device was fine particles <4.7 μm. Less than 7% of the Total Recovery (which includes measurement from the device) was deposited in the device itself using the Kos VNA. Over 14% was deposited in the Valois device.
Table 5 is a summary of the data collected from all six canisters, 221 through 223 using the Kos VNA, and 224 through 226 using the Valois device at the end of the canister life, after three priming sprays, three samples and 108 wasting sprays.
From these 60 U/actuation tests at the end of canister life, 54% of the total mass recovery using the Kos VNA was fine particles <4.71 μm. Only 42% of the total mass recovery from the Valois device was fine particles <4.7 μm. Less than 6% of the particulate mass was deposited in the device using the Kos VNA, as opposed to 14% deposited in the Valois device.
The second iteration of this test was completed using canisters containing 20 U/spray of the rh-insulin/HFA 134a formulation. Canisters 221, 222, and 223 were used for testing the Kos VNA. Canisters 224, 225 and 226 were used for testing the Valois device. Table 6 below provides the results of testing Canisters 221 through 226 at the beginning of canister life. Three samples were taken from each device and averaged after three priming sprays.
From these 20 U/actuation tests at the beginning of canister life, 67% of the total mass recovery using the Kos VNA was fine particles 4.7 μm. Only 54% of the total mass recovery from the Valois device was fine particles <4.7 μm. Eight percent of the total recovery was measured in the Kos VNA device, compared to 14% measured in the Valois device.
Table 7 below provides the results of testing Canisters 221 through 226 containing 20 U/actuation at the end of canister life. Three samples were taken from each device after the three priming sprays and three sample sprays and 108 wasting sprays.
From these 20 U/actuation tests at the beginning of canister life, 62% of the total mass recovery using the Kos VNA was fine particles <4.7 μm. Only 52% of the total mass recovery from the Valois device was fine particles <4.7 μm. Nine percent of the total recovery was measured in the Kos VNA device, compared to 22% measured in the Valois device.
These results illustrate that at both the beginning and end of canister life, the atomization system of the present invention outperforms the Valois system. There is a significantly greater amount of drug deposited in the Valois device, from 14-22% versus less than 9% in the Kos VNA at either the beginning or end of canister life. The fine particle fraction captured from the VNA is significantly greater for both <3.3 μm and <4.7 μm. This drug deposition measurement is an indirect measurement of plume velocity, and also illustrates that the velocity of the VNA plume is less than that of the Valois plume, allowing a greater number of fine particles to reach the lower cascade impaction plate filters as opposed to being captured in the upper plates of the impactor, or the device or patient's throat.
The actuator comparison between the Kos VNA and Valois device is illustrated in
The present invention can be carried out in other specific ways than those herein set forth without departing from the scope and essential characteristics of the invention. Further, the various aspects of the disclosed device and method can be used alone or in any combination, as is desired. The disclosed embodiments are. therefore, to be considered in all respects as illustrative and not restrictive, and all changes coming within the meaning and equivalency range of the appended claims are intended to be embraced therein.
This application claims priority to U.S. Application Ser. No. 60/903,970, filed Feb. 28, 2007, and is incorporated herein by reference in its entirety.
Number | Name | Date | Kind |
---|---|---|---|
3001524 | Maison et al. | Sep 1961 | A |
3053462 | Schloz | Sep 1962 | A |
3346195 | Groth | Oct 1967 | A |
3437270 | Venus, Jr. | Apr 1969 | A |
4074861 | Magers et al. | Feb 1978 | A |
4972830 | Wong et al. | Nov 1990 | A |
5662271 | Weston | Sep 1997 | A |
5676311 | Hartman | Oct 1997 | A |
5682875 | Blower et al. | Nov 1997 | A |
6418925 | Genova et al. | Jul 2002 | B1 |
6527151 | Pavkov et al. | Mar 2003 | B1 |
6615826 | Gabrio et al. | Sep 2003 | B1 |
20080296318 | Chevalier | Dec 2008 | A1 |
Number | Date | Country |
---|---|---|
933235 | Sep 1955 | DE |
102004001222 | Mar 2006 | DE |
0071192 | Nov 2000 | WO |
Number | Date | Country | |
---|---|---|---|
20080203193 A1 | Aug 2008 | US |
Number | Date | Country | |
---|---|---|---|
60903970 | Feb 2007 | US |